Aeria Bmat
Transcription
Aeria Bmat
quick set up guide 5 Insert the power cord into the control unit, then plug into power outlet and switch on mains power. 1 Remove all bedding from your mattress. 2 Unpack overlay and place on top of mattress (material side upwards; air hoses towards bed base). Tuck end flaps under mattress ends to secure. Ensure no cell sections are tucked beneath mattress. Overlay will compress once inflated. 3 Hang the control unit over bed end using inbuilt rear hanging hooks or place on a stable surface (ie floor beneath the bed). Overlay can be rotated 180 degrees as required. 4 Attach air hoses to the control unit, pressing firmly to ensure both connectors ‘click’ into place. Straighten any twists to ensure uninterrupted air flow between control unit and mattress. Also ensure hoses are not trapped between overlay and bed. 2 3 1 6 Turn the control unit Pressure dial to maximum and switch on (the switch illuminates to indicate power is on). Allow up to 20 minutes for complete inflation. The patient may lie on the system during inflation however the overlay will fill faster if allowed to inflate first. 7 Once overlay is inflated, bedding can be replaced. Secure sheets loosely enough to ensure they do not interfere with overlay alternation. 8 When patient is lying comfortably, turn the Pressure dial to reach the desired setting for effective patient comfort and support. After lowering, always check the system is working effectively – failure to do so could cause injury. Wait at least 3-5 minutes as the system may take a full cycle to adjust to the new pressure setting. control unit layout Power switch Turns system power on and off. Switch illuminates to indicate power is on. Pressure dial Adjusts overlay pressure from low to high, in line with patient weight and comfort requirements. Replacing the air filter 1 Switch off power, disconnect air hoses and power cord. 2 Place the control unit face down on a soft, flat surface. 3 Use a screwdriver to carefully remove the air filter cover. Clean the dust from the filter or discard and replace with a new filter. 4 Refit the air filter cover to the control unit before use. www.aeda.com Actively supporting your health system overview care & cleaning Always disconnect power supply before cleaning. system overview The Aeria Bmat offers low risk patients a practical alternative to static pressure care. 130 alternating touch points gently stimulate the skin’s surface, encouraging increased blood flow and circulation across high-risk areas. For basic care and cleaning, wipe down with warm water containing detergent (do not machine wash the overlay). For infection control, swab with a solution of sodium hypochlorite or similar (up to 10,000ppm available chlorine). Allow to dry thoroughly before use. • the Aeria Bmat is indicated for the prevention and treatment of skin breakdown and pressure ulcers in patients at low risk It is recommended the Aeria Bmat system be cleaned every two weeks if in constant use. To prevent cross contamination, it is recommended the mattress overlay be disinfected between patient use. • patient conditions for which the Aeria Bmat is contraindicated include instable spinal cord injury and cervical traction • suitable for use at home or in long term care environments technical specifications Product number MT-A1-015 • supports users to a maximum weight of 220 pounds 100 kilograms Continuous operation Class II Equipment Maximum user weight 220 lb 100 kg general safety precautions • placing layers between patient and overlay should be avoided or kept to a minimum • never use sharp objects or electrically heated blankets on / under system • avoid blocking control unit air intakes • do not use in the presence of an open flame or flammable anaesthetic mixture with air, oxygen or nitrous oxide Risk category Low Cell material PVC Rated voltage 220-240V 50/60Hz 0.2A Air output 4.5 liters per minute Cycle time 3 - 5 minutes Fuse ratingmains plug 3A control unit 1A (x1) • do not open the control unit (risk of electrical shock); refer servicing to qualified service personnel • do not use high temperature autoclave steam cleaning devices or phenolic-based products for cleaning • do not immerse the system in water or store in direct sunlight Total system weight 6.6 lb 3 kg working conditions Temperature 50°-105° F 10°-40° C Humidity 30-75% Therapeutic devices should only be used in accordance with manufacturer’s instructions and under the consent, supervision and management of a suitably qualified health professional. Visit aeda.com to view a copy of the manufacturer’s Warranty Statement AQ-9000 www.aeda.com Actively supporting your health AQ-9106 AQ-D205