Product information

Transcription

Product information
Fitting in OASIS plus
Product information
ICOS 105 BTE DM VC
ICOS 105 BTE DM
To benefit from the unique features in ICOS,
OASIS plus introduces an enhanced workflow.
The Lifestyle Profile: provides a structured
framework for identifying the specific communi-
cation needs of your client. It is this information
that drives the automatic behaviour of ICOS.
ICOS: The first hearing system with
Personalized Audio NavigationTM
Client Profile
• Sound CheckTM: offers the tools that are necessary for verifying the initial fit of ICOS.
The combination of individual needs and preferences
with audiometric data drives the automatic allocation
of programs and the behaviour of the adaptive
settings in ICOS.
• Automatic Program Allocation: OASIS plus
defines what user programs have to be assigned
to cover the client‘s lifestyle priorities.
• Global Fine-Tuning: Master tools for optimizing the response for all programs simultaneously.
• Environment Specific Controls: A dedicated
set of controls offered to specifically optimize
different environments and programs.
•
ICOS Tracker: Not only can the important information about the clients use of ICOS be viewed
but OASIS plus will offer fine-tuning suggestions based on the accumulated data.
AudioMatics
The adaptive features in ICOS are combined
as needed to maximize performance in the
recognized environment.
•
•
•
•
•
•
•
•
•
•
All ICOS System can be operated with the optional Bernafon
Remote Control RC-S.
REF. 160-02-310-00
For patients that have chosen to use more than
one Remote (i.e. one in the office and one at home)
the Custom Remote Control should be ordered.
Custom Remote Control Audio Navigation ProgramTM
3 freely configurable listening programs
Dedicated DAI/FM listening program
Multi-Dimensional Directionality
Adaptive Noise Reduction in 8 bands
Audio RecognitionTM with 4 different monitors
Soft Noise Management
Adaptive Feedback Cancellation
OpenFitTM
7-channel digital signal processing
Apart from advanced fine-tuning tools, ICOS offers
an in-depth insight into the use and behavior of the
hearing system while worn by your client. It provides
a solid foundation for clear and accurate recommendations for maximal optimization.
REF. 160-02-320-00
Innovative Hearing Solutions
www.bernafon.com
Quality System
DS/EN ISO 9001
DS/EN ISO 13485
Bernafon Australia Pty. Ltd.
Unit 12, 97 Castlemaine
Street
Milton QLD 4064
Australia
Freecall 1800 809 111
Phone +61 (0)7 3250 0300
Fax +61 (0)7 3250 0372
www.bernafon.com.au
Bernafon New Zealand Ltd.
Level 1, Building F
27-29 William Pickering
Drive
Albany, Auckland
New Zealand
Phone +64 9 920 42 40
Fax +64 9 920 42 43
www.bernafon.co.nz
7.07/BAG/951-18-010-00/subject to change
Manufacturer:
Bernafon UK
Cadzow Industrial Estate
Low Waters Road
Hamilton
ML3 7QE Scotland
Phone +44 1698 28 59 68
Fax +44 1698 42 14 56
www.bernafon.co.uk
Frequency (Hz)
ICOS 105 BTE DM VC
A compact BTE with an adaptive
directional microphone system,
a size 13 battery and rotary volume
control, suitable for mild to severe
hearing losses.
Optimization
• ICOS Tracker (Data logging) with fine-tune proposals
• Sound CheckTM
• 7-Channel TriQualizer™
Bernafon AG
Morgenstrasse 131
3018 Bern
Switzerland
Phone + 41 (0)31 998 15 15
Fax + 41 (0)31 998 15 90
105 BTE DM VC/105 BTE DM
• Lifestyle Profile
• Clinical Profile
Remote Control
Standard Remote Control
dB HL
•
Innovative Hearing Solutions
ICOS 105 BTE DM
A compact BTE with an adaptive
directional microphone system,
size 13 battery, suitable for mild
to severe hearing losses.
Frequency responses
ICOS information
ICOS BTE
Gain
Output
105 BTE DM VC 105 BTE DM
IEC 60118-7
IEC 60118-0
IEC 60118-7 IEC 60118-0
Current
Distor­
tion
General
Information
Additional
Information
(Ears.)
Gain / dB
140
90
130
80
120
70
110
60
100
50
Output / dB SPL
ICOS 105 BTE DM VC
ICOS 105 BTE DM
Gain / dB
140
90
130
80
120
70
110
60
100
50
(2cc)
OSPL 90, Peak
dB SPL
123
127
123
127
OSPL 90 , 1600 Hz
dB SPL
112
121
112
121
HFA OSPL 90, ANSI dB SPL
113
––
113
––
Full-On Gain, Peak dB
60
63
60
63
Full-On Gain, 1600 Hz dB
52
58
52
58
HFA Full-On Gain, ANSI
dB
53
––
53
––
90
40
90
40
Reference Test Gain, IEC dB
36
44
36
44
80
30
80
30
20
70
dB
35
––
35
––
Quiescent Current
mA
1.2
1.2
1.2
1.2
Operating Current, IEC mA
1.2
1.2
1.2
1.2
Operating Current, ANSI
mA
1.2
––
1.2
––
13
13
Battery type
1)
(2cc)
Output / dB SPL
IEC 60118-0
Earsimulator (IEC 60711)
Reference Test Gain, ANSI (Ears.)
IEC 60118-7
2cc Coupler (IEC 60126)
500/800/1600 Hz typ., IEC %
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
500/800/1600 Hz typ., ANSI %
<1 / <1 / <1
––
<1 / <1 / <1
––
Frequency Range, ANSI Hz
100 – 5800
100 – 5800
Equiv. Input Noise, IEC/ ANSI 1)
15/13
15/13
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
dB 83
92
83
92
Telecoil HFA SPLITS, ANSI
dB SPL
96
––
96
––
yes
yes
Remote Control Reception
Earhooks
8 dB (std.), 0 dB (opt.)
8 dB (std.), 0 dB (opt.)
Childrens Earhooks
8 dB (opt.), 0 dB (opt.)
8 dB (opt.), 0 dB (opt.)
SPIRA Thin Sound Tube 0.9
optional
optional
Programmable telecoil
yes
yes
Push button
yes
yes
FM communication
yes
yes
DAI
yes
yes
Volume control
yes
––
70
250
500
1k
1k6 2k 2k5
4k
8k
Output OSPL 90
Full-On Gain
250
500
Frequency/Hz
1k
DAI/FM Adapters
FMA3 Adapter DAI3 Adapter For FM communication REF. 399-50-570-00
systems only
For connection to Hi-Fi, TV set, PC, etc.
REF. 399-50-580-00
ICOS BTE Color Range
Style & Design
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
blue white grey platinum gold anthracite black
Young & Fun
ICOS is programmed with OASIS plus, Version 6.0 or later – a NOAH compatible, MS-Windows based PC-Fitting
software. NOAH 2.0 or NOAH 3.0 with a Hi-Pro or a NOAHlinkTM is required.
Programming adapter
Programming adapter
REF. 390-01-151-00
Traditional & Classic
green yellow rose beige brown blue
Programming cables
Nr. 2, NEW STANDARD (HiPro) Blue, left Red, right REF. 384-20-033-00
REF. 384-20-032-00
Note: Always use new batteries for programming.
Nr. 2, NEW STANDARD (NOAHlink)
Blue, left Red, right 4k
Frequency/Hz
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
All measurements are based on earhook without filter.
Adapters and cables
1k6 2k 2k5
REF. 384-20-035-00
REF. 384-20-034-00
grey brown graphite
8k
20
Frequency responses
ICOS information
ICOS BTE
Gain
Output
105 BTE DM VC 105 BTE DM
IEC 60118-7
IEC 60118-0
IEC 60118-7 IEC 60118-0
Current
Distor­
tion
General
Information
Additional
Information
(Ears.)
Gain / dB
140
90
130
80
120
70
110
60
100
50
Output / dB SPL
ICOS 105 BTE DM VC
ICOS 105 BTE DM
Gain / dB
140
90
130
80
120
70
110
60
100
50
(2cc)
OSPL 90, Peak
dB SPL
123
127
123
127
OSPL 90 , 1600 Hz
dB SPL
112
121
112
121
HFA OSPL 90, ANSI dB SPL
113
––
113
––
Full-On Gain, Peak dB
60
63
60
63
Full-On Gain, 1600 Hz dB
52
58
52
58
HFA Full-On Gain, ANSI
dB
53
––
53
––
90
40
90
40
Reference Test Gain, IEC dB
36
44
36
44
80
30
80
30
20
70
dB
35
––
35
––
Quiescent Current
mA
1.2
1.2
1.2
1.2
Operating Current, IEC mA
1.2
1.2
1.2
1.2
Operating Current, ANSI
mA
1.2
––
1.2
––
13
13
Battery type
1)
(2cc)
Output / dB SPL
IEC 60118-0
Earsimulator (IEC 60711)
Reference Test Gain, ANSI (Ears.)
IEC 60118-7
2cc Coupler (IEC 60126)
500/800/1600 Hz typ., IEC %
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
500/800/1600 Hz typ., ANSI %
<1 / <1 / <1
––
<1 / <1 / <1
––
Frequency Range, ANSI Hz
100 – 5800
100 – 5800
Equiv. Input Noise, IEC/ ANSI 1)
15/13
15/13
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
dB 83
92
83
92
Telecoil HFA SPLITS, ANSI
dB SPL
96
––
96
––
yes
yes
Remote Control Reception
Earhooks
8 dB (std.), 0 dB (opt.)
8 dB (std.), 0 dB (opt.)
Childrens Earhooks
8 dB (opt.), 0 dB (opt.)
8 dB (opt.), 0 dB (opt.)
SPIRA Thin Sound Tube 0.9
optional
optional
Programmable telecoil
yes
yes
Push button
yes
yes
FM communication
yes
yes
DAI
yes
yes
Volume control
yes
––
70
250
500
1k
1k6 2k 2k5
4k
8k
Output OSPL 90
Full-On Gain
250
500
Frequency/Hz
1k
DAI/FM Adapters
FMA3 Adapter DAI3 Adapter For FM communication REF. 399-50-570-00
systems only
For connection to Hi-Fi, TV set, PC, etc.
REF. 399-50-580-00
ICOS BTE Color Range
Style & Design
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
blue white grey platinum gold anthracite black
Young & Fun
ICOS is programmed with OASIS plus, Version 6.0 or later – a NOAH compatible, MS-Windows based PC-Fitting
software. NOAH 2.0 or NOAH 3.0 with a Hi-Pro or a NOAHlinkTM is required.
Programming adapter
Programming adapter
REF. 390-01-151-00
Traditional & Classic
green yellow rose beige brown blue
Programming cables
Nr. 2, NEW STANDARD (HiPro) Blue, left Red, right REF. 384-20-033-00
REF. 384-20-032-00
Note: Always use new batteries for programming.
Nr. 2, NEW STANDARD (NOAHlink)
Blue, left Red, right 4k
Frequency/Hz
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
All measurements are based on earhook without filter.
Adapters and cables
1k6 2k 2k5
REF. 384-20-035-00
REF. 384-20-034-00
grey brown graphite
8k
20
Fitting in OASIS plus
Product information
ICOS 105 BTE DM VC
ICOS 105 BTE DM
To benefit from the unique features in ICOS,
OASIS plus introduces an enhanced workflow.
The Lifestyle Profile: provides a structured
framework for identifying the specific communi-
cation needs of your client. It is this information
that drives the automatic behaviour of ICOS.
ICOS: The first hearing system with
Personalized Audio NavigationTM
Client Profile
• Sound CheckTM: offers the tools that are necessary for verifying the initial fit of ICOS.
The combination of individual needs and preferences
with audiometric data drives the automatic allocation
of programs and the behaviour of the adaptive
settings in ICOS.
• Automatic Program Allocation: OASIS plus
defines what user programs have to be assigned
to cover the client‘s lifestyle priorities.
• Global Fine-Tuning: Master tools for optimizing the response for all programs simultaneously.
• Environment Specific Controls: A dedicated
set of controls offered to specifically optimize
different environments and programs.
•
ICOS Tracker: Not only can the important information about the clients use of ICOS be viewed
but OASIS plus will offer fine-tuning suggestions based on the accumulated data.
AudioMatics
The adaptive features in ICOS are combined
as needed to maximize performance in the
recognized environment.
•
•
•
•
•
•
•
•
•
•
All ICOS System can be operated with the optional Bernafon
Remote Control RC-S.
REF. 160-02-310-00
For patients that have chosen to use more than
one Remote (i.e. one in the office and one at home)
the Custom Remote Control should be ordered.
Custom Remote Control Audio Navigation ProgramTM
3 freely configurable listening programs
Dedicated DAI/FM listening program
Multi-Dimensional Directionality
Adaptive Noise Reduction in 8 bands
Audio RecognitionTM with 4 different monitors
Soft Noise Management
Adaptive Feedback Cancellation
OpenFitTM
7-channel digital signal processing
Apart from advanced fine-tuning tools, ICOS offers
an in-depth insight into the use and behavior of the
hearing system while worn by your client. It provides
a solid foundation for clear and accurate recommendations for maximal optimization.
REF. 160-02-320-00
Innovative Hearing Solutions
www.bernafon.com
Quality System
DS/EN ISO 9001
DS/EN ISO 13485
Bernafon Australia Pty. Ltd.
Unit 12, 97 Castlemaine
Street
Milton QLD 4064
Australia
Freecall 1800 809 111
Phone +61 (0)7 3250 0300
Fax +61 (0)7 3250 0372
www.bernafon.com.au
Bernafon New Zealand Ltd.
Level 1, Building F
27-29 William Pickering
Drive
Albany, Auckland
New Zealand
Phone +64 9 920 42 40
Fax +64 9 920 42 43
www.bernafon.co.nz
7.07/BAG/951-18-010-00/subject to change
Manufacturer:
Bernafon UK
Cadzow Industrial Estate
Low Waters Road
Hamilton
ML3 7QE Scotland
Phone +44 1698 28 59 68
Fax +44 1698 42 14 56
www.bernafon.co.uk
Frequency (Hz)
ICOS 105 BTE DM VC
A compact BTE with an adaptive
directional microphone system,
a size 13 battery and rotary volume
control, suitable for mild to severe
hearing losses.
Optimization
• ICOS Tracker (Data logging) with fine-tune proposals
• Sound CheckTM
• 7-Channel TriQualizer™
Bernafon AG
Morgenstrasse 131
3018 Bern
Switzerland
Phone + 41 (0)31 998 15 15
Fax + 41 (0)31 998 15 90
105 BTE DM VC/105 BTE DM
• Lifestyle Profile
• Clinical Profile
Remote Control
Standard Remote Control
dB HL
•
Innovative Hearing Solutions
ICOS 105 BTE DM
A compact BTE with an adaptive
directional microphone system,
size 13 battery, suitable for mild
to severe hearing losses.
Product Information
ICOS 106 BTE DM
ICOS – Lifestyle Driven Technology
dB HL
ICOS, the only premium hearing instrument family
offering Lifestyle Profile technology, now introduces
the micro BTE: ICOS 106 BTE DM.
106 BTE DM
Cl
Fit ient B
tin
g T ased
ools
-D
l P im
ro ens
c e io
ss nal
ing
Lifestyle Driven
Technology
n
atio
v ig
Na
o
di ram
Au rog
P
Styles and Colors
le
e Profi
st y l
Life
ti
ul
M i g na
S
Innovative Technology for maximized
hearing benefits
•
•
•
•
•
Micro size BTE housing
312 size Battery
Lifestyle Profile
Audio Navigation Program
Multi-Dimensional Signal Processing
– 7 channels
– Multi-Dimensional Directionality in 4 bands
– Adaptive Noise Reduction in 8 bands
– Soft Noise Management
– Adaptive Feedback Cancellation
– OpenFit
– Adaptive Signal Unification
• Client based Fitting Tools
– 3 freely configurable listening programs
(choice out of 11)
– ICOS Tracker Data logging system
• Styles and Colors
– 10 stylish BTE colors
– Compatible with SPIRA flex Sound Tube System
– Optional Remote Control
– DAI compatibility
– Mute function
Innovative Hearing Solutions
Frequency (Hz)
ICOS 106 BTE DM
A fully programmable digital micro BTE
instrument, with an adaptive directional
microphone system, a size 312 battery,
suitable for mild to moderate hearing
losses.
ICOS Information
ICOS 106 BTE DM
Additional
Information
Current
Gain
Output
SPIRA flex
Tube 0.9
IEC 60118-7
ANSI S3.22
(2cc)
IEC 60118-0
IEC 60118-0
(Ears.)
IEC 60118-7
ANSI S3.22
(2cc)
121
126
(Ears.)
OSPL 90, Peak
dB SPL
117
123
OSPL 90, 1600 Hz
dB SPL
102
112
108
118
HFA OSPL 90
dB SPL
108
––
113
––
Full-On Gain, Peak
dB
47
54
50
57
Full-On Gain, 1600 Hz
dB
36
45
41
51
HFA Full-On Gain
dB
41
––
46
––
Reference Test Gain
dB
30
37
36
42
Quiescent Current
mA
1.2
1.2
1.2
1.2
Operating Current
mA
1.3
1.2
1.3
1.2
Battery type
312
Distortion 500/800/1600 Hz typ.
%
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
Frequency Range
Hz
100–5000
––
100–5300
––
dB
16
19
12
16
Equiv. Input Noise
1)
Remote Control reception
yes
Push button
yes
DAI
yes
Volume control
1)
SPIRA flex
Tube 1.3
by Remote Control
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
All measurements are made according to IEC 60118 and ANSI S3.22-2003 with SPIRA flex tubes #2.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
Frequency Response
IEC 60118-7
2cc Coupler (IEC 60318-5)1)
Output / dB SPL
IEC 60118-0
Earsimulator (IEC 60711)
Gain / dB
Output / dB SPL
Gain / dB
140
90
130
80
120
70
60
110
60
100
50
100
50
90
40
90
40
80
30
80
30
20
70
140
90
130
80
120
70
110
70
250
500
1k
1k6 2k 2k5
4k
ICOS 106 BTE DM
Output OSPL 90
Full-On Gain
20
250
8k
500
Frequency / Hz
1k
1k6 2k 2k5
Frequency / Hz
Measurements with Tube 1.3 mm
Measurements with Tube 0.9 mm
1)
Identical to coupler described in withdrawn standard IEC 60126
SPIRA flex Sound Tube System
SPIRA flex Sound Tube System
for more cosmetic and open fitting solutions.
Please refer to the
SPIRA flex Product Brochure
REF. 951-18-710-00
or the
SPIRA flex Product Information
REF. 951-18-610-00
SPIRA flex Master Fitting Kit
containing all SPIRA flex components
REF. 890-80-060-00
4k
8k
Quick Fitting Guide
Different acoustic options result in different benefits
Open Dome
short vent length < 3 mm
Tulip Dome
Custom Tip
Canal Mold
vent length 10 mm
SPIRA flex 0.9
SPIRA flex 1.3
Maximum Fitting Range
Minimum Feedback Risk
Maximum Cosmetics
Minimum Occlusion
Suggested maximum fittings with different acoustic combinations
SPIRA flex tube 1.3
dB HL
dB HL
SPIRA flex tube 0.9
Frequency / Hz
Frequency / Hz
with Open Dome
with Tulip Dome
with Tulip Dome
with Canal Mold, large vent
with Custom Tip or Canal Mold
with Canal Mold, no vent
ICOS 106 BTE DM Color Range
Style & Design
titan
platinum
gold
anthracite
black
grey brown
beige
brown
graphite
Traditional & Classic
grey
Cables and Adapters
ICOS 106 BTE DM is programmed with OASIS plus, Version 8.0 or later – a NOAH compatible,
MS-Windows based PC-Fitting software. NOAH 2.0 or NOAH 3.0 with a Hi-Pro or a NOAHlink is required.
Programming cables
Programming Adapter
Nr. 2, NEW STANDARD (HiPro)
Programming adapter
REF. 399-50-640-00
Blue, left
REF. 384-20-033-00
Red, right
REF. 384-20-032-00
Nr. 2, NEW STANDARD (NOAHlink)
DAI 4 Adapter
Blue, left
REF. 384-20-035-00
For connection to Hi-Fi,
TV set, PC, etc.
REF. 399-50-521-00
Red, right
REF. 384-20-034-00
Note: Always use new batteries for programming.
Fitting in OASIS plus
ICOS Tracker Data logging
The enhanced ICOS Tracker Data logging system records
a wealth of operational statistics encompassing:
• Client usage: including total wearing time and average daily use
• Listening environments: representing which environments the client
had worn their instruments
• User adjustments: including program and volume changes,
and the level of changes
• Feedback monitoring: a feature unique to ICOS Tracker, which monitors
the activity of the Adaptive Feedback Canceller, and the associated
feedback risk
• Fine-tuning recommendations: which can be instantly implemented
to optimize the fitting, ensuring efficient and accurate follow up
and reducing the number of fine tuning sessions
Remote Control
All ICOS systems can be operated with the optional
Bernafon Remote Control RC-S.
Standard Remote Control
REF. 160-02-310-00
For patients that have chosen to use more than
one Remote (i.e. one in the office and one at home)
the Custom Remote Control should be ordered.
Custom Remote Control
REF. 160-02-320-00
Innovative Hearing Solutions
Bernafon AG
Morgenstrasse 131
3018 Bern
Switzerland
Phone + 41 (0)31 998 15 15
Fax + 41 (0)31 998 15 90
www.bernafon.com
Quality System
DS/EN ISO 9001
DS/EN ISO 13485
Bernafon UK
Cadzow Industrial Estate
Low Waters Road
Hamilton
ML3 7QE Scotland
Phone +44 1698 28 59 68
Fax +44 1698 42 14 56
www.bernafon.co.uk
Bernafon Australia Pty. Ltd.
512 Wickham Street
Fortitude Valley QLD 4006
Australia
Freecall 1800 809 111
Phone +61 (0)7 3250 0300
Fax +61 (0)7 3252 2048
www.bernafon.com.au
Bernafon New Zealand Ltd.
Level 1, Building F
27-29 William Pickering Drive
Albany, Auckland
New Zealand
Phone +64 (0)9 920 42 40
Fax +64 (0)9 920 42 43
www.bernafon.co.nz
09.07/BAG/951-19-210-00/subject to change
Manufacturer:
U
A
A
IEC 60118-7
2cc Coupler
(IEC 60126)
Output/dB SPL
IEC 60118-0
Earsimulator
(IEC 60711)
Gain / dB
120
70
110
60
Output / dB SPL
ICOS 205 ITE DM VC
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
50
90
40
100
50
80
30
90
40
70
20
80
30
60
10
70
20
0
60
250
500
1k
1k6 2k 2k5
4k
8k
250
500
Frequency/Hz
1k
1k6 2k 2k5
4k
The Lifestyle Profile: provides a structured
framework for identifying the specific communi­-
cation needs of your client. It is this information
that drives the automatic behaviour of ICOS.
ICOS 205 ITE DM VC
ICOS 315 ITC DM VC
ICOS 305 ITC DM
ICOS 415 MC DM
ICOS 410 MC / ICOS 400 CIC
• Sound CheckTM: offers the tools that are necessary for verifying the initial fit of ICOS.
ICOS: The first hearing system with
Personalized Audio NavigationTM
• Automatic Program Allocation: OASIS plus
defines what user programs have to be assigned
to cover the client‘s lifestyle priorities.
Client Profile
To benefit from the unique features in ICOS,
OASIS plus introduces an enhanced workflow.
100
50
8k
10
•
• Global Fine-Tuning: Master tools for optimizing the response for all programs simultaneously.
Frequency/Hz
• Environment Specific Controls: A dedicated
set of controls offered to specifically optimize
different environments and programs.
Output/dB SPL
Gain / dB
120
70
110
60
100
50
Output / dB SPL
ICOS 315 ITC DM VC
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
100
50
90
40
80
30
90
40
70
20
80
30
60
10
70
20
50
0
60
10
250
500
1k
1k6 2k 2k5
4k
8k
250
Product information
Fitting in OASIS plus
500
Frequency/Hz
1k
1k6 2k 2k5
4k
8k
•
ICOS Tracker: Not only can the important information about the clients use of ICOS be viewed
but OASIS plus will offer fine-tuning suggestions based on the accumulated data.
• Lifestyle Profile
• Clinical Profile
AudioMatics
The adaptive features in ICOS are combined
as needed to maximize performance in the
recognized environment.
•
•
•
•
•
•
•
•
•
•
All ICOS System can be operated with the optional Bernafon
Remote Control RC-S.
REF. 160-02-310-00
For patients that have chosen to use more than
one Remote (i.e. one in the office and one at home)
the Custom Remote Control should be ordered.
Custom Remote Control REF. 160-02-320-00
Frequency/Hz
Audio Navigation ProgramTM
3 freely configurable listening programs
Additional dedicated AutoTelephone program
Multi-Dimensional Directionality
Adaptive Noise Reduction in 8 bands
Audio RecognitionTM with 4 different monitors
Soft Noise Management
Adaptive Feedback Cancellation
OpenFitTM
7-channel digital signal processing
Optimization
Apart from advanced fine-tuning tools, ICOS offers
an in-depth insight into the use and behavior of the
hearing system while worn by your client. It provides
a solid foundation for clear and accurate recommendations for maximal optimization.
Gain / dB
120
70
110
60
100
50
90
40
80
Output / dB SPL
ICOS 305 ITC DM
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
Manufacturer:
110
60
Bernafon AG
Morgenstrasse 131
3018 Bern
Switzerland
Phone +41 31 998 15 15
Fax +41 31 998 15 90
www.bernafon.com
100
50
30
90
40
70
20
80
30
60
10
70
20
0
60
50
250
500
1k
1k6 2k 2k5
Frequency/Hz
4k
8k
250
500
1k
1k6 2k 2k5
Frequency/Hz
4k
8k
10
Bernafon UK
Cadzow Industrial Estate
Off Low Waters Road
Hamilton
ML3 7QE Scotland
Phone +44 1698 285 968
Fax +44 1698 421 456
www.bernafon.co.uk
Bernafon Australia Pty. Ltd.
12/97 Castlemaine Street
Milton QLD 4064
Australia
Freecall 1800 809 111
Phone +61 7 3250 0300
Fax +61 7 3250 0372
www.bernafon.com.au
Bernafon New Zealand Ltd.
Level 1, Building F
27-29 William Pickering Drive
Albany, Auckland 0632
New Zealand
Phone +64 9 920 42 40
Fax +64 9 920 42 43
www.bernafon.co.nz
05.09/BAG/951-18-110-00/subject to change
Output/dB SPL
400/410/415
305/315/205
The combination of individual needs and preferences
with audiometric data drives the automatic allocation
of programs and the behaviour of the adaptive
settings in ICOS.
Remote Control
Standard Remote Control
dB HL
Frequency responses
• ICOS Tracker (Data logging) with fine-tune proposals
• Sound CheckTM
• 7-Channel TriQualizer™
Frequency (Hz)
ICOS 205 ITE DM VC
An adaptive directional full concha instrument with a size 13 battery and a rotary
volume control for mild to moderately
severe hearing losses.
ICOS 315 ITC DM VC
An adaptive directional half shell and canal
instrument with a size 312 battery and rotary
volume control for mild to moderately severe
hearing losses.
ICOS 305 ITC DM
An adaptive directional canal instrument
with a size 312 battery for mild to moderately
severe hearing losses.
ICOS 415 MC DM
An adaptive directional mini-canal
instrument with a size 10 battery for
mild to moderate hearing losses.
ICOS 410 MC
A mini-canal instrument with a size
10 battery for mild to moderate hearing
losses.
ICOS 400 CIC
A completely in the canal instrument
with a size 10 battery for mild to moderate
hearing losses.
U
U
A
A
ICOS information
ICOS information
ICOS ITE
305 ITC DM
415 MC DM
IEC 60118-7
IEC 60118-0
(2cc)
(Ears.)
(2cc)
(Ears.)
IEC 60118-7
(2cc)
IEC 60118-0
(Ears.)
OSPL 90, Peak
dB SPL
112
122
112
122
110
120
OSPL 90 , 1600 Hz
dB SPL
105
113
105
113
102
111
HFA OSPL 90, ANSI
dB SPL
107
––
107
––
104
––
Output
Output
IEC 60118-7
2cc Coupler
(IEC 60126)
ICOS ITE
205 ITE DM VC 315 ITC DM VC
IEC 60118-7
IEC 60118-0
IEC 60118-7
IEC 60118-0
Frequency responses
410 MC
400 CIC
IEC 60118-7
(2cc)
IEC 60118-0
(Ears.)
(2cc)
(Ears.)
104
115
104
115
106
97
105
Output/dB SPL
IEC 60118-0
Earsimulator
(IEC 60711)
Gain / dB
120
70
110
60
OSPL 90, Peak
dB SPL
OSPL 90, 1600 Hz
dB SPL
97
HFA OSPL 90, ANSI
dB SPL
98
––
98
––
100
50
Output/dB SPL
ICOS 415 MC DM
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
dB
46
58
43
54
41
52
Full-On Gain, Peak dB
34
45
34
45
90
40
100
50
Full-On Gain, 1600 Hz dB
40
49
38
47
34
43
Full-On Gain, 1600 Hz dB
30
39
30
39
80
30
90
40
70
20
80
30
60
10
70
20
0
60
Gain
Full-On Gain, Peak Gain
dB
41
––
38
––
34
––
HFA Full-On Gain, ANSI
dB
30
––
31
––
Reference Test Gain, IEC dB
29
38
29
38
27
35
Reference Test Gain, IEC dB
22
31
22
31
Reference Test Gain, ANSI dB
29
––
29
––
27
––
Reference Test Gain, ANSI dB
22
––
21
––
Quiescent Current mA
1.2
1.2
1.1
1.1
1.1
1.1
Quiescent Current
mA
1.1
1.1
0.8
0.8
Operating Current, IEC
mA
1.2
1.2
1.2
1.2
1.2
1.2
Operating Current, IEC
mA
1.1
1.1
0.8
0.8
Operating Current, ANSI
mA
1.2
––
1.2
––
1.2
––
Operating Current, ANSI
mA
1.1
––
0.8
––
500/800/1600 Hz typ., IEC General
Information
1)
312
312
%
<1 / <1 / <1 <2 / <2 / <1
<1 / <1 / <1 <2 / <2 / <1
<1 / <1 / <1 <1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
500/800/1600 Hz typ., ANSI %
Frequency Range, ANSI Hz
Equiv. Input Noise, IEC/ANSI 1)
dB
100–7400
––
––
100–7500
14/14
––
––
100–7500
14/15
––
––
14/15
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
72
81
70
79
67
75
Telecoil HFA SPLITS, ANSI
dB SPL
85
––
85
––
84
––
optional
optional
optional
Programmable Telecoil
optional
optional
optional
AutoTelephone optional
optional
optional
Push button
yes
yes
yes
Volume Control yes
yes
Remote Control reception coil
Additional
Information
13
––
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
1)
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
Battery type
Distor­tion
Battery type
Distor­tion
General
Information
500/800/1600 Hz typ., IEC <1 / <1 / <1
<1 / <1 / <1
%
Hz
110–7500
Equiv. Input Noise, IEC/ANSI 1)
dB
<1 / <1 / <2
<1 / <1 / <1
––
<1 / <1 / <1
––
14/14
250
500
1k
1k6 2k 2k5
4k
8k
250
500
Frequency/Hz
1k
1k6 2k 2k5
4k
8k
10
Frequency/Hz
110–7500
<1 / <1 / <1
––
––
19/17
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
––
––
––
––
Telecoil HFA SPLITS, ANSI
dB SPL
––
––
––
––
50
10
optional
optional
Programmable Telecoil
––
––
AutoTelephone ––
––
Push button
optional
Volume Control ––
optional
––
Output/dB SPL
Gain / dB
120
70
110
60
100
50
90
40
80
Output/dB SPL
Gain / dB
130
80
120
70
110
60
100
50
30
90
40
70
20
80
30
60
10
70
20
0
60
50
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
250
500
1k
1k6 2k 2k5
Frequency/Hz
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
4k
8k
ICOS 410 MC
ICOS 400 CIC
Output OSPL 90
Full-On Gain
250
500
1k
1k6 2k 2k5
4k
8k
10
Frequency/Hz
Adapters and cables
ICOS is programmed with OASIS plus, Version 6.0 or later – a NOAH compatible, MS-Windows based PC-Fitting software.
NOAH 3.0 with a Hi-Pro or a NOAHlinkTM is required.
All colors are available for all instruments.
Programming adapter for all ITE’s
beige %
500/800/1600 Hz typ., ANSI ICOS Colors
10
Frequency Range, ANSI
Remote Control reception coil
Additional
Information
Current
Current
HFA Full-On Gain, ANSI
pink
brown dark brown
FlexConnect
REF. 390-01-180-05
Programming cables Programming cables
Nr. 2, new standard (HiPro)
Nr. 2, new standard (NOAHlink)
Blue, left
Red, right
REF. 384-20-033-00
REF. 384-20-032-00
Blue, left
Red, right
REF. 384-20-035-00
REF. 384-20-034-00
Note: Always use new batteries for programming.
U
U
A
A
ICOS information
ICOS information
ICOS ITE
305 ITC DM
415 MC DM
IEC 60118-7
IEC 60118-0
(2cc)
(Ears.)
(2cc)
(Ears.)
IEC 60118-7
(2cc)
IEC 60118-0
(Ears.)
OSPL 90, Peak
dB SPL
112
122
112
122
110
120
OSPL 90 , 1600 Hz
dB SPL
105
113
105
113
102
111
HFA OSPL 90, ANSI
dB SPL
107
––
107
––
104
––
Output
Output
IEC 60118-7
2cc Coupler
(IEC 60126)
ICOS ITE
205 ITE DM VC 315 ITC DM VC
IEC 60118-7
IEC 60118-0
IEC 60118-7
IEC 60118-0
Frequency responses
410 MC
400 CIC
IEC 60118-7
(2cc)
IEC 60118-0
(Ears.)
(2cc)
(Ears.)
104
115
104
115
106
97
105
Output/dB SPL
IEC 60118-0
Earsimulator
(IEC 60711)
Gain / dB
120
70
110
60
OSPL 90, Peak
dB SPL
OSPL 90, 1600 Hz
dB SPL
97
HFA OSPL 90, ANSI
dB SPL
98
––
98
––
100
50
Output/dB SPL
ICOS 415 MC DM
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
dB
46
58
43
54
41
52
Full-On Gain, Peak dB
34
45
34
45
90
40
100
50
Full-On Gain, 1600 Hz dB
40
49
38
47
34
43
Full-On Gain, 1600 Hz dB
30
39
30
39
80
30
90
40
70
20
80
30
60
10
70
20
0
60
Gain
Full-On Gain, Peak Gain
dB
41
––
38
––
34
––
HFA Full-On Gain, ANSI
dB
30
––
31
––
Reference Test Gain, IEC dB
29
38
29
38
27
35
Reference Test Gain, IEC dB
22
31
22
31
Reference Test Gain, ANSI dB
29
––
29
––
27
––
Reference Test Gain, ANSI dB
22
––
21
––
Quiescent Current mA
1.2
1.2
1.1
1.1
1.1
1.1
Quiescent Current
mA
1.1
1.1
0.8
0.8
Operating Current, IEC
mA
1.2
1.2
1.2
1.2
1.2
1.2
Operating Current, IEC
mA
1.1
1.1
0.8
0.8
Operating Current, ANSI
mA
1.2
––
1.2
––
1.2
––
Operating Current, ANSI
mA
1.1
––
0.8
––
500/800/1600 Hz typ., IEC General
Information
1)
312
312
%
<1 / <1 / <1 <2 / <2 / <1
<1 / <1 / <1 <2 / <2 / <1
<1 / <1 / <1 <1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
500/800/1600 Hz typ., ANSI %
Frequency Range, ANSI Hz
Equiv. Input Noise, IEC/ANSI 1)
dB
100–7400
––
––
100–7500
14/14
––
––
100–7500
14/15
––
––
14/15
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
72
81
70
79
67
75
Telecoil HFA SPLITS, ANSI
dB SPL
85
––
85
––
84
––
optional
optional
optional
Programmable Telecoil
optional
optional
optional
AutoTelephone optional
optional
optional
Push button
yes
yes
yes
Volume Control yes
yes
Remote Control reception coil
Additional
Information
13
––
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
1)
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
Battery type
Distor­tion
Battery type
Distor­tion
General
Information
500/800/1600 Hz typ., IEC <1 / <1 / <1
<1 / <1 / <1
%
Hz
110–7500
Equiv. Input Noise, IEC/ANSI 1)
dB
<1 / <1 / <2
<1 / <1 / <1
––
<1 / <1 / <1
––
14/14
250
500
1k
1k6 2k 2k5
4k
8k
250
500
Frequency/Hz
1k
1k6 2k 2k5
4k
8k
10
Frequency/Hz
110–7500
<1 / <1 / <1
––
––
19/17
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
––
––
––
––
Telecoil HFA SPLITS, ANSI
dB SPL
––
––
––
––
50
10
optional
optional
Programmable Telecoil
––
––
AutoTelephone ––
––
Push button
optional
Volume Control ––
optional
––
Output/dB SPL
Gain / dB
120
70
110
60
100
50
90
40
80
Output/dB SPL
Gain / dB
130
80
120
70
110
60
100
50
30
90
40
70
20
80
30
60
10
70
20
0
60
50
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
250
500
1k
1k6 2k 2k5
Frequency/Hz
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
4k
8k
ICOS 410 MC
ICOS 400 CIC
Output OSPL 90
Full-On Gain
250
500
1k
1k6 2k 2k5
4k
8k
10
Frequency/Hz
Adapters and cables
ICOS is programmed with OASIS plus, Version 6.0 or later – a NOAH compatible, MS-Windows based PC-Fitting software.
NOAH 3.0 with a Hi-Pro or a NOAHlinkTM is required.
All colors are available for all instruments.
Programming adapter for all ITE’s
beige %
500/800/1600 Hz typ., ANSI ICOS Colors
10
Frequency Range, ANSI
Remote Control reception coil
Additional
Information
Current
Current
HFA Full-On Gain, ANSI
pink
brown dark brown
FlexConnect
REF. 390-01-180-05
Programming cables Programming cables
Nr. 2, new standard (HiPro)
Nr. 2, new standard (NOAHlink)
Blue, left
Red, right
REF. 384-20-033-00
REF. 384-20-032-00
Blue, left
Red, right
REF. 384-20-035-00
REF. 384-20-034-00
Note: Always use new batteries for programming.
U
U
A
A
IEC 60118-7
2cc Coupler
(IEC 60126)
Output/dB SPL
IEC 60118-0
Earsimulator
(IEC 60711)
Gain / dB
120
70
110
60
Output / dB SPL
ICOS 205 ITE DM VC
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
50
90
40
100
50
80
30
90
40
70
20
80
30
60
10
70
20
0
60
250
500
1k
1k6 2k 2k5
4k
8k
250
500
Frequency/Hz
1k
1k6 2k 2k5
4k
The Lifestyle Profile: provides a structured
framework for identifying the specific communi­-
cation needs of your client. It is this information
that drives the automatic behaviour of ICOS.
ICOS 205 ITE DM VC
ICOS 315 ITC DM VC
ICOS 305 ITC DM
ICOS 415 MC DM
ICOS 410 MC / ICOS 400 CIC
• Sound CheckTM: offers the tools that are necessary for verifying the initial fit of ICOS.
ICOS: The first hearing system with
Personalized Audio NavigationTM
• Automatic Program Allocation: OASIS plus
defines what user programs have to be assigned
to cover the client‘s lifestyle priorities.
Client Profile
To benefit from the unique features in ICOS,
OASIS plus introduces an enhanced workflow.
100
50
8k
10
•
• Global Fine-Tuning: Master tools for optimizing the response for all programs simultaneously.
Frequency/Hz
• Environment Specific Controls: A dedicated
set of controls offered to specifically optimize
different environments and programs.
Output/dB SPL
Gain / dB
120
70
110
60
100
50
Output / dB SPL
ICOS 315 ITC DM VC
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
100
50
90
40
80
30
90
40
70
20
80
30
60
10
70
20
50
0
60
10
250
500
1k
1k6 2k 2k5
4k
8k
250
Product information
Fitting in OASIS plus
500
Frequency/Hz
1k
1k6 2k 2k5
4k
8k
•
ICOS Tracker: Not only can the important information about the clients use of ICOS be viewed
but OASIS plus will offer fine-tuning suggestions based on the accumulated data.
• Lifestyle Profile
• Clinical Profile
AudioMatics
The adaptive features in ICOS are combined
as needed to maximize performance in the
recognized environment.
•
•
•
•
•
•
•
•
•
•
All ICOS System can be operated with the optional Bernafon
Remote Control RC-S.
REF. 160-02-310-00
For patients that have chosen to use more than
one Remote (i.e. one in the office and one at home)
the Custom Remote Control should be ordered.
Custom Remote Control REF. 160-02-320-00
Frequency/Hz
Audio Navigation ProgramTM
3 freely configurable listening programs
Additional dedicated AutoTelephone program
Multi-Dimensional Directionality
Adaptive Noise Reduction in 8 bands
Audio RecognitionTM with 4 different monitors
Soft Noise Management
Adaptive Feedback Cancellation
OpenFitTM
7-channel digital signal processing
Optimization
Apart from advanced fine-tuning tools, ICOS offers
an in-depth insight into the use and behavior of the
hearing system while worn by your client. It provides
a solid foundation for clear and accurate recommendations for maximal optimization.
Gain / dB
120
70
110
60
100
50
90
40
80
Output / dB SPL
ICOS 305 ITC DM
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
Manufacturer:
110
60
Bernafon AG
Morgenstrasse 131
3018 Bern
Switzerland
Phone +41 31 998 15 15
Fax +41 31 998 15 90
www.bernafon.com
100
50
30
90
40
70
20
80
30
60
10
70
20
0
60
50
250
500
1k
1k6 2k 2k5
Frequency/Hz
4k
8k
250
500
1k
1k6 2k 2k5
Frequency/Hz
4k
8k
10
Bernafon UK
Cadzow Industrial Estate
Off Low Waters Road
Hamilton
ML3 7QE Scotland
Phone +44 1698 285 968
Fax +44 1698 421 456
www.bernafon.co.uk
Bernafon Australia Pty. Ltd.
12/97 Castlemaine Street
Milton QLD 4064
Australia
Freecall 1800 809 111
Phone +61 7 3250 0300
Fax +61 7 3250 0372
www.bernafon.com.au
Bernafon New Zealand Ltd.
Level 1, Building F
27-29 William Pickering Drive
Albany, Auckland 0632
New Zealand
Phone +64 9 920 42 40
Fax +64 9 920 42 43
www.bernafon.co.nz
05.09/BAG/951-18-110-00/subject to change
Output/dB SPL
400/410/415
305/315/205
The combination of individual needs and preferences
with audiometric data drives the automatic allocation
of programs and the behaviour of the adaptive
settings in ICOS.
Remote Control
Standard Remote Control
dB HL
Frequency responses
• ICOS Tracker (Data logging) with fine-tune proposals
• Sound CheckTM
• 7-Channel TriQualizer™
Frequency (Hz)
ICOS 205 ITE DM VC
An adaptive directional full concha instrument with a size 13 battery and a rotary
volume control for mild to moderately
severe hearing losses.
ICOS 315 ITC DM VC
An adaptive directional half shell and canal
instrument with a size 312 battery and rotary
volume control for mild to moderately severe
hearing losses.
ICOS 305 ITC DM
An adaptive directional canal instrument
with a size 312 battery for mild to moderately
severe hearing losses.
ICOS 415 MC DM
An adaptive directional mini-canal
instrument with a size 10 battery for
mild to moderate hearing losses.
ICOS 410 MC
A mini-canal instrument with a size
10 battery for mild to moderate hearing
losses.
ICOS 400 CIC
A completely in the canal instrument
with a size 10 battery for mild to moderate
hearing losses.
U
U
A
A
ICOS information
ICOS information
ICOS ITE
305 ITC DM
415 MC DM
IEC 60118-7
IEC 60118-0
(2cc)
(Ears.)
(2cc)
(Ears.)
IEC 60118-7
(2cc)
IEC 60118-0
(Ears.)
OSPL 90, Peak
dB SPL
112
122
112
122
110
120
OSPL 90 , 1600 Hz
dB SPL
105
113
105
113
102
111
HFA OSPL 90, ANSI
dB SPL
107
––
107
––
104
––
Output
Output
IEC 60118-7
2cc Coupler
(IEC 60126)
ICOS ITE
205 ITE DM VC 315 ITC DM VC
IEC 60118-7
IEC 60118-0
IEC 60118-7
IEC 60118-0
Frequency responses
410 MC
400 CIC
IEC 60118-7
(2cc)
IEC 60118-0
(Ears.)
(2cc)
(Ears.)
104
115
104
115
106
97
105
Output/dB SPL
IEC 60118-0
Earsimulator
(IEC 60711)
Gain / dB
120
70
110
60
OSPL 90, Peak
dB SPL
OSPL 90, 1600 Hz
dB SPL
97
HFA OSPL 90, ANSI
dB SPL
98
––
98
––
100
50
Output/dB SPL
ICOS 415 MC DM
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
dB
46
58
43
54
41
52
Full-On Gain, Peak dB
34
45
34
45
90
40
100
50
Full-On Gain, 1600 Hz dB
40
49
38
47
34
43
Full-On Gain, 1600 Hz dB
30
39
30
39
80
30
90
40
70
20
80
30
60
10
70
20
0
60
Gain
Full-On Gain, Peak Gain
dB
41
––
38
––
34
––
HFA Full-On Gain, ANSI
dB
30
––
31
––
Reference Test Gain, IEC dB
29
38
29
38
27
35
Reference Test Gain, IEC dB
22
31
22
31
Reference Test Gain, ANSI dB
29
––
29
––
27
––
Reference Test Gain, ANSI dB
22
––
21
––
Quiescent Current mA
1.2
1.2
1.1
1.1
1.1
1.1
Quiescent Current
mA
1.1
1.1
0.8
0.8
Operating Current, IEC
mA
1.2
1.2
1.2
1.2
1.2
1.2
Operating Current, IEC
mA
1.1
1.1
0.8
0.8
Operating Current, ANSI
mA
1.2
––
1.2
––
1.2
––
Operating Current, ANSI
mA
1.1
––
0.8
––
500/800/1600 Hz typ., IEC General
Information
1)
312
312
%
<1 / <1 / <1 <2 / <2 / <1
<1 / <1 / <1 <2 / <2 / <1
<1 / <1 / <1 <1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
<1 / <1 / <1
500/800/1600 Hz typ., ANSI %
Frequency Range, ANSI Hz
Equiv. Input Noise, IEC/ANSI 1)
dB
100–7400
––
––
100–7500
14/14
––
––
100–7500
14/15
––
––
14/15
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
72
81
70
79
67
75
Telecoil HFA SPLITS, ANSI
dB SPL
85
––
85
––
84
––
optional
optional
optional
Programmable Telecoil
optional
optional
optional
AutoTelephone optional
optional
optional
Push button
yes
yes
yes
Volume Control yes
yes
Remote Control reception coil
Additional
Information
13
––
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
1)
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
Battery type
Distor­tion
Battery type
Distor­tion
General
Information
500/800/1600 Hz typ., IEC <1 / <1 / <1
<1 / <1 / <1
%
Hz
110–7500
Equiv. Input Noise, IEC/ANSI 1)
dB
<1 / <1 / <2
<1 / <1 / <1
––
<1 / <1 / <1
––
14/14
250
500
1k
1k6 2k 2k5
4k
8k
250
500
Frequency/Hz
1k
1k6 2k 2k5
4k
8k
10
Frequency/Hz
110–7500
<1 / <1 / <1
––
––
19/17
Telecoil 1 mA/m 1600 Hz, IEC
dB SPL
––
––
––
––
Telecoil HFA SPLITS, ANSI
dB SPL
––
––
––
––
50
10
optional
optional
Programmable Telecoil
––
––
AutoTelephone ––
––
Push button
optional
Volume Control ––
optional
––
Output/dB SPL
Gain / dB
120
70
110
60
100
50
90
40
80
Output/dB SPL
Gain / dB
130
80
120
70
110
60
100
50
30
90
40
70
20
80
30
60
10
70
20
0
60
50
Technical data measured with expansion, corresponding to Soft Noise Management level 3.
250
500
1k
1k6 2k 2k5
Frequency/Hz
All measurements are made according to IEC 60118 if not otherwise mentioned. ANSI refers to ANSI S3.22-2003.
The Full-On Gain setting can be programmed into the instrument from OASIS plus for verification purposes.
4k
8k
ICOS 410 MC
ICOS 400 CIC
Output OSPL 90
Full-On Gain
250
500
1k
1k6 2k 2k5
4k
8k
10
Frequency/Hz
Adapters and cables
ICOS is programmed with OASIS plus, Version 6.0 or later – a NOAH compatible, MS-Windows based PC-Fitting software.
NOAH 3.0 with a Hi-Pro or a NOAHlinkTM is required.
All colors are available for all instruments.
Programming adapter for all ITE’s
beige %
500/800/1600 Hz typ., ANSI ICOS Colors
10
Frequency Range, ANSI
Remote Control reception coil
Additional
Information
Current
Current
HFA Full-On Gain, ANSI
pink
brown dark brown
FlexConnect
REF. 390-01-180-05
Programming cables Programming cables
Nr. 2, new standard (HiPro)
Nr. 2, new standard (NOAHlink)
Blue, left
Red, right
REF. 384-20-033-00
REF. 384-20-032-00
Blue, left
Red, right
REF. 384-20-035-00
REF. 384-20-034-00
Note: Always use new batteries for programming.
U
U
A
A
IEC 60118-7
2cc Coupler
(IEC 60126)
Output/dB SPL
IEC 60118-0
Earsimulator
(IEC 60711)
Gain / dB
120
70
110
60
Output / dB SPL
ICOS 205 ITE DM VC
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
50
90
40
100
50
80
30
90
40
70
20
80
30
60
10
70
20
0
60
250
500
1k
1k6 2k 2k5
4k
8k
250
500
Frequency/Hz
1k
1k6 2k 2k5
4k
The Lifestyle Profile: provides a structured
framework for identifying the specific communi­-
cation needs of your client. It is this information
that drives the automatic behaviour of ICOS.
ICOS 205 ITE DM VC
ICOS 315 ITC DM VC
ICOS 305 ITC DM
ICOS 415 MC DM
ICOS 410 MC / ICOS 400 CIC
• Sound CheckTM: offers the tools that are necessary for verifying the initial fit of ICOS.
ICOS: The first hearing system with
Personalized Audio NavigationTM
• Automatic Program Allocation: OASIS plus
defines what user programs have to be assigned
to cover the client‘s lifestyle priorities.
Client Profile
To benefit from the unique features in ICOS,
OASIS plus introduces an enhanced workflow.
100
50
8k
10
•
• Global Fine-Tuning: Master tools for optimizing the response for all programs simultaneously.
Frequency/Hz
• Environment Specific Controls: A dedicated
set of controls offered to specifically optimize
different environments and programs.
Output/dB SPL
Gain / dB
120
70
110
60
100
50
Output / dB SPL
ICOS 315 ITC DM VC
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
110
60
100
50
90
40
80
30
90
40
70
20
80
30
60
10
70
20
50
0
60
10
250
500
1k
1k6 2k 2k5
4k
8k
250
Product information
Fitting in OASIS plus
500
Frequency/Hz
1k
1k6 2k 2k5
4k
8k
•
ICOS Tracker: Not only can the important information about the clients use of ICOS be viewed
but OASIS plus will offer fine-tuning suggestions based on the accumulated data.
• Lifestyle Profile
• Clinical Profile
AudioMatics
The adaptive features in ICOS are combined
as needed to maximize performance in the
recognized environment.
•
•
•
•
•
•
•
•
•
•
All ICOS System can be operated with the optional Bernafon
Remote Control RC-S.
REF. 160-02-310-00
For patients that have chosen to use more than
one Remote (i.e. one in the office and one at home)
the Custom Remote Control should be ordered.
Custom Remote Control REF. 160-02-320-00
Frequency/Hz
Audio Navigation ProgramTM
3 freely configurable listening programs
Additional dedicated AutoTelephone program
Multi-Dimensional Directionality
Adaptive Noise Reduction in 8 bands
Audio RecognitionTM with 4 different monitors
Soft Noise Management
Adaptive Feedback Cancellation
OpenFitTM
7-channel digital signal processing
Optimization
Apart from advanced fine-tuning tools, ICOS offers
an in-depth insight into the use and behavior of the
hearing system while worn by your client. It provides
a solid foundation for clear and accurate recommendations for maximal optimization.
Gain / dB
120
70
110
60
100
50
90
40
80
Output / dB SPL
ICOS 305 ITC DM
Output OSPL 90
Full-On Gain
Gain / dB
130
80
120
70
Manufacturer:
110
60
Bernafon AG
Morgenstrasse 131
3018 Bern
Switzerland
Phone +41 31 998 15 15
Fax +41 31 998 15 90
www.bernafon.com
100
50
30
90
40
70
20
80
30
60
10
70
20
0
60
50
250
500
1k
1k6 2k 2k5
Frequency/Hz
4k
8k
250
500
1k
1k6 2k 2k5
Frequency/Hz
4k
8k
10
Bernafon UK
Cadzow Industrial Estate
Off Low Waters Road
Hamilton
ML3 7QE Scotland
Phone +44 1698 285 968
Fax +44 1698 421 456
www.bernafon.co.uk
Bernafon Australia Pty. Ltd.
12/97 Castlemaine Street
Milton QLD 4064
Australia
Freecall 1800 809 111
Phone +61 7 3250 0300
Fax +61 7 3250 0372
www.bernafon.com.au
Bernafon New Zealand Ltd.
Level 1, Building F
27-29 William Pickering Drive
Albany, Auckland 0632
New Zealand
Phone +64 9 920 42 40
Fax +64 9 920 42 43
www.bernafon.co.nz
05.09/BAG/951-18-110-00/subject to change
Output/dB SPL
400/410/415
305/315/205
The combination of individual needs and preferences
with audiometric data drives the automatic allocation
of programs and the behaviour of the adaptive
settings in ICOS.
Remote Control
Standard Remote Control
dB HL
Frequency responses
• ICOS Tracker (Data logging) with fine-tune proposals
• Sound CheckTM
• 7-Channel TriQualizer™
Frequency (Hz)
ICOS 205 ITE DM VC
An adaptive directional full concha instrument with a size 13 battery and a rotary
volume control for mild to moderately
severe hearing losses.
ICOS 315 ITC DM VC
An adaptive directional half shell and canal
instrument with a size 312 battery and rotary
volume control for mild to moderately severe
hearing losses.
ICOS 305 ITC DM
An adaptive directional canal instrument
with a size 312 battery for mild to moderately
severe hearing losses.
ICOS 415 MC DM
An adaptive directional mini-canal
instrument with a size 10 battery for
mild to moderate hearing losses.
ICOS 410 MC
A mini-canal instrument with a size
10 battery for mild to moderate hearing
losses.
ICOS 400 CIC
A completely in the canal instrument
with a size 10 battery for mild to moderate
hearing losses.
U

Similar documents