Valve Diameter Dimensions
Transcription
Valve Diameter Dimensions
Valve Diameter Dimensions The Aortic Valve Guide The stent outer diameter alone is not enough to make clinically relevant decisions at the time of implant due to differences existing between the labeled size and other measurements of the bioprothesis.1 For any valve, three dimensions of diameter may be taken into account: OD, ED, and ID. Measurements that affect valve selection For a given ED, the labeled size and the ID can vary significantly from product to product. Examples with same ED (mm) ED Mitroflow Mosaic Ultra / Hancock II Ultra PERIMOUNT valve Magna Ease valve OD ID 26 23 19 26 21 18.5 26 19 18 26 21 20 (Labeled Size) 1. Dalmau, MJ (2007). One year hemodynamic performance of the Perimount Magna pericardial xenograft and the Medtronic Mosaic bioprothesis in the aortic position: a prospective randomized study. Interact CardioVasc Thorac Surg, 6, 345-349. 2. Sievers, HH (2005). Prosthetic aortic valve replacement. J Thorac Cardiovasc Surg, 129, 961-965. The ID identifies the largest potential diameter of the valve opening; Brief Summary: Aortic Bioprostheses Indications: For use in patients whose aortic valvular disease warrants replacement of their natural or previously placed prosthetic valve. Contraindications: Do not use if surgeon believes it would be contrary to the patient’s best interests. Complications and Side Effects: stenosis, regurgitation, endocarditis, hemolysis, thromboembolism, valve thrombosis, nonstructural dysfunction, structural valve deterioration, anemia, arrhythmia, hemorrhage, transient ischemic attack/stroke, congestive heart failure, myocardial infarction, angina, any of which could lead to reoperation, explantation, permanent disability, and death. Warnings: Alternative therapies should be considered in the presence of conditions affecting calcium metabolism or when calcium containing chronic drug therapies are used, including children, adolescents, young adults, and patients on a high calcium diet or maintenance hemodialysis. Should be used with caution in the presence of severe systemic hypertension or when anticipated patient longevity is longer than the known longevity of the prosthesis. the larger the ID, the better the hemodynamic performance.1 Examples of OD (mm) Mitroflow OD (Labeled Size) Mosaic Ultra / Hancock II Ultra PERIMOUNT valve Magna Ease valve Internal Diameter (ID) For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards Lifesciences Irvine, USAand Switzerland Tokyo, Japan SingaporeCorporation. Brazil I Nyon, I Singapore, I São Paulo, Edwards Edwards Magna Ease are Itrademarks of Edwards Lifesciences Lifesciences, the stylized E logo, edwards.com Carpentier-Edwards, PERIMOUNT, and Magna are trademarks of Edwards Lifesciences Corporation and are registered in the United States Patent and Trademark Office. 19 15.4 17.5 18 18 21 17.3 18.5 20 20 23 19 20.5 22 22 Mosaic, Mosaic ULTRA, Hancock II, and Hancock II ULTRA are trademarks of Medtronic, Inc. SJM Biocor and Epic are trademarks of St. Jude Medical, Inc. Mitroflow is a trademark of Sorin Bio-medica Cardio Srl. © 2010 Edwards Lifesciences Corporation. All rights reserved. AR05888 Edwards Lifesciences Irvine, USA I Nyon, Switzerland edwards.com I Tokyo, Japan I Singapore, Singapore I São Paulo, Brazil edwards.com There is More to Valve Dimension Than the Labeled Size 1,2 Bioprosthetic Aortic Valve Dimensions measured in mm* I I I Edwards Lifesciences Irvine, USA I Nyon, Switzerland edwards.com LABELED SIZE (OD) (mm) In choosing an aortic bioprosthesis, labeled size alone is not a sufficient indicator to determine patient relevancy.1 This Aortic Valve Guide outlines the valve dimensions of the Carpentier-Edwards PERIMOUNT Magna Ease pericardial aortic bioprosthesis, the Carpentier-Edwards PERIMOUNT Magna pericardial aortic bioprosthesis, the Carpentier-Edwards PERIMOUNT 19 pericardial aortic bioprosthesis, as well as competitive valves from Tokyo, Japan Singapore, Singapore AORTIC BIOPROSTHESES ID ED ID/ED RATIO PROFILE HEIGHT Magna Ease valve 18 24 0.75 Magna valve 18 24 PERIMOUNT valve 18 Mosaic Mosaic Ultra Epic Supra / SJM Biocor Supra St. Jude Medical, Medtronic, and Sorin. São Paulo, Brazil Mitroflow LABELED SIZE (OD) (mm) AORTIC BIOPROSTHESES ID ED ID/ED RATIO PROFILE HEIGHT 13 Magna Ease valve 24 30 0.80 16 0.75 14 Magna valve 24 30 0.80 17 26 0.69 14 PERIMOUNT valve 24 32 0.75 17 17.5 25 0.70 13.5 Mosaic / Hancock II 22.5 33 0.68 17.5 17.5 24 0.73 13.5 Mosaic Ultra / Hancock II Ultra 22.5 30 0.75 17.5 19 25 0.76 14 Epic / SJM Biocor 23 29 0.79 16 15.4 21 0.73 11 Epic Supra / SJM Biocor Supra 25 31 0.81 17 Mitroflow 21 29 0.72 15 25 • Stent Outer Diameter (OD): Diameter of the tissue annulus; also known as the labeled size • External Sewing Ring Diameter (ED): The full diameter of the valve • Stent Internal Diameter (ID): The largest potential diameter of the valve opening; the larger the ID, the better the hemodynamic performance1 21 • Profile Height: The total height of the valve Magna Ease valve 20 26 0.77 14 Magna Ease valve 26 32 0.81 17 Magna valve 20 26 0.77 15 Magna valve 26 32 0.81 18 PERIMOUNT valve 20 28 0.71 15 PERIMOUNT valve 26 35 0.74 18 Mosaic / Hancock II 18.5 27 0.69 15 Mosaic / Hancock II 24 36 0.67 18.5 Mosaic Ultra / Hancock II Ultra 18.5 26 0.71 15 Mosaic Ultra / Hancock II Ultra 24 32 0.75 18.5 Epic / SJM Biocor 19 25 0.76 14 Epic / SJM Biocor 25 31 0.81 17 Epic Supra / SJM Biocor Supra 21 28 0.75 15 Epic Supra / SJM Biocor Supra 27 33 0.82 19 17.3 23 0.75 13 Mitroflow 22.9 31 0.74 16 Magna Ease valve 22 28 0.79 15 Magna Ease valve 28 34 0.82 18 Magna valve 22 28 0.79 16 Magna valve 28 34 0.82 19 PERIMOUNT valve 22 31 0.71 16 PERIMOUNT valve 28 37 0.76 19 Mosaic / Hancock II 20.5 30 0.68 16 Mosaic / Hancock II 26 39 0.67 20 Mosaic Ultra / Hancock II Ultra 20.5 28 0.73 16 Mosaic Ultra / Hancock II Ultra 26 34 0.76 20 Epic / SJM Biocor 21 27 0.78 15 Epic / SJM Biocor 27 33 0.82 19 Epic Supra / SJM Biocor Supra 23 29 0.79 16 Mitroflow 24.7 33 0.75 16 Mitroflow 19 26 0.73 14 Mitroflow Profile Height 23 ID OD Labeled Size ED *All dimensions reflect the manufacturers’ published information. Edwards Lifesciences and Medtronic indicated that the dimensions are nominal dimensions. 27 29 edwards.com There is More to Valve Dimension Than the Labeled Size 1,2 Bioprosthetic Aortic Valve Dimensions measured in mm* I I I Edwards Lifesciences Irvine, USA I Nyon, Switzerland edwards.com LABELED SIZE (OD) (mm) In choosing an aortic bioprosthesis, labeled size alone is not a sufficient indicator to determine patient relevancy.1 This Aortic Valve Guide outlines the valve dimensions of the Carpentier-Edwards PERIMOUNT Magna Ease pericardial aortic bioprosthesis, the Carpentier-Edwards PERIMOUNT Magna pericardial aortic bioprosthesis, the Carpentier-Edwards PERIMOUNT 19 pericardial aortic bioprosthesis, as well as competitive valves from Tokyo, Japan Singapore, Singapore AORTIC BIOPROSTHESES ID ED ID/ED RATIO PROFILE HEIGHT Magna Ease valve 18 24 0.75 Magna valve 18 24 PERIMOUNT valve 18 Mosaic Mosaic Ultra Epic Supra / SJM Biocor Supra St. Jude Medical, Medtronic, and Sorin. São Paulo, Brazil Mitroflow LABELED SIZE (OD) (mm) AORTIC BIOPROSTHESES ID ED ID/ED RATIO PROFILE HEIGHT 13 Magna Ease valve 24 30 0.80 16 0.75 14 Magna valve 24 30 0.80 17 26 0.69 14 PERIMOUNT valve 24 32 0.75 17 17.5 25 0.70 13.5 Mosaic / Hancock II 22.5 33 0.68 17.5 17.5 24 0.73 13.5 Mosaic Ultra / Hancock II Ultra 22.5 30 0.75 17.5 19 25 0.76 14 Epic / SJM Biocor 23 29 0.79 16 15.4 21 0.73 11 Epic Supra / SJM Biocor Supra 25 31 0.81 17 Mitroflow 21 29 0.72 15 25 • Stent Outer Diameter (OD): Diameter of the tissue annulus; also known as the labeled size • External Sewing Ring Diameter (ED): The full diameter of the valve • Stent Internal Diameter (ID): The largest potential diameter of the valve opening; the larger the ID, the better the hemodynamic performance1 21 • Profile Height: The total height of the valve Magna Ease valve 20 26 0.77 14 Magna Ease valve 26 32 0.81 17 Magna valve 20 26 0.77 15 Magna valve 26 32 0.81 18 PERIMOUNT valve 20 28 0.71 15 PERIMOUNT valve 26 35 0.74 18 Mosaic / Hancock II 18.5 27 0.69 15 Mosaic / Hancock II 24 36 0.67 18.5 Mosaic Ultra / Hancock II Ultra 18.5 26 0.71 15 Mosaic Ultra / Hancock II Ultra 24 32 0.75 18.5 Epic / SJM Biocor 19 25 0.76 14 Epic / SJM Biocor 25 31 0.81 17 Epic Supra / SJM Biocor Supra 21 28 0.75 15 Epic Supra / SJM Biocor Supra 27 33 0.82 19 17.3 23 0.75 13 Mitroflow 22.9 31 0.74 16 Magna Ease valve 22 28 0.79 15 Magna Ease valve 28 34 0.82 18 Magna valve 22 28 0.79 16 Magna valve 28 34 0.82 19 PERIMOUNT valve 22 31 0.71 16 PERIMOUNT valve 28 37 0.76 19 Mosaic / Hancock II 20.5 30 0.68 16 Mosaic / Hancock II 26 39 0.67 20 Mosaic Ultra / Hancock II Ultra 20.5 28 0.73 16 Mosaic Ultra / Hancock II Ultra 26 34 0.76 20 Epic / SJM Biocor 21 27 0.78 15 Epic / SJM Biocor 27 33 0.82 19 Epic Supra / SJM Biocor Supra 23 29 0.79 16 Mitroflow 24.7 33 0.75 16 Mitroflow 19 26 0.73 14 Mitroflow Profile Height 23 ID OD Labeled Size ED *All dimensions reflect the manufacturers’ published information. Edwards Lifesciences and Medtronic indicated that the dimensions are nominal dimensions. 27 29 edwards.com There is More to Valve Dimension Than the Labeled Size 1,2 Bioprosthetic Aortic Valve Dimensions measured in mm* I I I Edwards Lifesciences Irvine, USA I Nyon, Switzerland edwards.com LABELED SIZE (OD) (mm) In choosing an aortic bioprosthesis, labeled size alone is not a sufficient indicator to determine patient relevancy.1 This Aortic Valve Guide outlines the valve dimensions of the Carpentier-Edwards PERIMOUNT Magna Ease pericardial aortic bioprosthesis, the Carpentier-Edwards PERIMOUNT Magna pericardial aortic bioprosthesis, the Carpentier-Edwards PERIMOUNT 19 pericardial aortic bioprosthesis, as well as competitive valves from Tokyo, Japan Singapore, Singapore AORTIC BIOPROSTHESES ID ED ID/ED RATIO PROFILE HEIGHT Magna Ease valve 18 24 0.75 Magna valve 18 24 PERIMOUNT valve 18 Mosaic Mosaic Ultra Epic Supra / SJM Biocor Supra St. Jude Medical, Medtronic, and Sorin. São Paulo, Brazil Mitroflow LABELED SIZE (OD) (mm) AORTIC BIOPROSTHESES ID ED ID/ED RATIO PROFILE HEIGHT 13 Magna Ease valve 24 30 0.80 16 0.75 14 Magna valve 24 30 0.80 17 26 0.69 14 PERIMOUNT valve 24 32 0.75 17 17.5 25 0.70 13.5 Mosaic / Hancock II 22.5 33 0.68 17.5 17.5 24 0.73 13.5 Mosaic Ultra / Hancock II Ultra 22.5 30 0.75 17.5 19 25 0.76 14 Epic / SJM Biocor 23 29 0.79 16 15.4 21 0.73 11 Epic Supra / SJM Biocor Supra 25 31 0.81 17 Mitroflow 21 29 0.72 15 25 • Stent Outer Diameter (OD): Diameter of the tissue annulus; also known as the labeled size • External Sewing Ring Diameter (ED): The full diameter of the valve • Stent Internal Diameter (ID): The largest potential diameter of the valve opening; the larger the ID, the better the hemodynamic performance1 21 • Profile Height: The total height of the valve Magna Ease valve 20 26 0.77 14 Magna Ease valve 26 32 0.81 17 Magna valve 20 26 0.77 15 Magna valve 26 32 0.81 18 PERIMOUNT valve 20 28 0.71 15 PERIMOUNT valve 26 35 0.74 18 Mosaic / Hancock II 18.5 27 0.69 15 Mosaic / Hancock II 24 36 0.67 18.5 Mosaic Ultra / Hancock II Ultra 18.5 26 0.71 15 Mosaic Ultra / Hancock II Ultra 24 32 0.75 18.5 Epic / SJM Biocor 19 25 0.76 14 Epic / SJM Biocor 25 31 0.81 17 Epic Supra / SJM Biocor Supra 21 28 0.75 15 Epic Supra / SJM Biocor Supra 27 33 0.82 19 17.3 23 0.75 13 Mitroflow 22.9 31 0.74 16 Magna Ease valve 22 28 0.79 15 Magna Ease valve 28 34 0.82 18 Magna valve 22 28 0.79 16 Magna valve 28 34 0.82 19 PERIMOUNT valve 22 31 0.71 16 PERIMOUNT valve 28 37 0.76 19 Mosaic / Hancock II 20.5 30 0.68 16 Mosaic / Hancock II 26 39 0.67 20 Mosaic Ultra / Hancock II Ultra 20.5 28 0.73 16 Mosaic Ultra / Hancock II Ultra 26 34 0.76 20 Epic / SJM Biocor 21 27 0.78 15 Epic / SJM Biocor 27 33 0.82 19 Epic Supra / SJM Biocor Supra 23 29 0.79 16 Mitroflow 24.7 33 0.75 16 Mitroflow 19 26 0.73 14 Mitroflow Profile Height 23 ID OD Labeled Size ED *All dimensions reflect the manufacturers’ published information. Edwards Lifesciences and Medtronic indicated that the dimensions are nominal dimensions. 27 29 Valve Diameter Dimensions The Aortic Valve Guide The stent outer diameter alone is not enough to make clinically relevant decisions at the time of implant due to differences existing between the labeled size and other measurements of the bioprothesis.1 For any valve, three dimensions of diameter may be taken into account: OD, ED, and ID. Measurements that affect valve selection For a given ED, the labeled size and the ID can vary significantly from product to product. Examples with same ED (mm) ED Mitroflow Mosaic Ultra / Hancock II Ultra PERIMOUNT valve Magna Ease valve OD ID 26 23 19 26 21 18.5 26 19 18 26 21 20 (Labeled Size) 1. Dalmau, MJ (2007). One year hemodynamic performance of the Perimount Magna pericardial xenograft and the Medtronic Mosaic bioprothesis in the aortic position: a prospective randomized study. Interact CardioVasc Thorac Surg, 6, 345-349. 2. Sievers, HH (2005). Prosthetic aortic valve replacement. J Thorac Cardiovasc Surg, 129, 961-965. The ID identifies the largest potential diameter of the valve opening; Brief Summary: Aortic Bioprostheses Indications: For use in patients whose aortic valvular disease warrants replacement of their natural or previously placed prosthetic valve. Contraindications: Do not use if surgeon believes it would be contrary to the patient’s best interests. Complications and Side Effects: stenosis, regurgitation, endocarditis, hemolysis, thromboembolism, valve thrombosis, nonstructural dysfunction, structural valve deterioration, anemia, arrhythmia, hemorrhage, transient ischemic attack/stroke, congestive heart failure, myocardial infarction, angina, any of which could lead to reoperation, explantation, permanent disability, and death. Warnings: Alternative therapies should be considered in the presence of conditions affecting calcium metabolism or when calcium containing chronic drug therapies are used, including children, adolescents, young adults, and patients on a high calcium diet or maintenance hemodialysis. Should be used with caution in the presence of severe systemic hypertension or when anticipated patient longevity is longer than the known longevity of the prosthesis. the larger the ID, the better the hemodynamic performance.1 Examples of OD (mm) Mitroflow OD (Labeled Size) Mosaic Ultra / Hancock II Ultra PERIMOUNT valve Magna Ease valve Internal Diameter (ID) For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards Lifesciences Irvine, USAand Switzerland Tokyo, Japan SingaporeCorporation. Brazil I Nyon, I Singapore, I São Paulo, Edwards Edwards Magna Ease are Itrademarks of Edwards Lifesciences Lifesciences, the stylized E logo, edwards.com Carpentier-Edwards, PERIMOUNT, and Magna are trademarks of Edwards Lifesciences Corporation and are registered in the United States Patent and Trademark Office. 19 15.4 17.5 18 18 21 17.3 18.5 20 20 23 19 20.5 22 22 Mosaic, Mosaic ULTRA, Hancock II, and Hancock II ULTRA are trademarks of Medtronic, Inc. SJM Biocor and Epic are trademarks of St. Jude Medical, Inc. Mitroflow is a trademark of Sorin Bio-medica Cardio Srl. © 2010 Edwards Lifesciences Corporation. All rights reserved. AR05888 Edwards Lifesciences Irvine, USA I Nyon, Switzerland edwards.com I Tokyo, Japan I Singapore, Singapore I São Paulo, Brazil Valve Diameter Dimensions The Aortic Valve Guide The stent outer diameter alone is not enough to make clinically relevant decisions at the time of implant due to differences existing between the labeled size and other measurements of the bioprothesis.1 For any valve, three dimensions of diameter may be taken into account: OD, ED, and ID. Measurements that affect valve selection For a given ED, the labeled size and the ID can vary significantly from product to product. Examples with same ED (mm) ED Mitroflow Mosaic Ultra / Hancock II Ultra PERIMOUNT valve Magna Ease valve OD ID 26 23 19 26 21 18.5 26 19 18 26 21 20 (Labeled Size) 1. Dalmau, MJ (2007). One year hemodynamic performance of the Perimount Magna pericardial xenograft and the Medtronic Mosaic bioprothesis in the aortic position: a prospective randomized study. Interact CardioVasc Thorac Surg, 6, 345-349. 2. Sievers, HH (2005). Prosthetic aortic valve replacement. J Thorac Cardiovasc Surg, 129, 961-965. The ID identifies the largest potential diameter of the valve opening; Brief Summary: Aortic Bioprostheses Indications: For use in patients whose aortic valvular disease warrants replacement of their natural or previously placed prosthetic valve. Contraindications: Do not use if surgeon believes it would be contrary to the patient’s best interests. Complications and Side Effects: stenosis, regurgitation, endocarditis, hemolysis, thromboembolism, valve thrombosis, nonstructural dysfunction, structural valve deterioration, anemia, arrhythmia, hemorrhage, transient ischemic attack/stroke, congestive heart failure, myocardial infarction, angina, any of which could lead to reoperation, explantation, permanent disability, and death. Warnings: Alternative therapies should be considered in the presence of conditions affecting calcium metabolism or when calcium containing chronic drug therapies are used, including children, adolescents, young adults, and patients on a high calcium diet or maintenance hemodialysis. Should be used with caution in the presence of severe systemic hypertension or when anticipated patient longevity is longer than the known longevity of the prosthesis. the larger the ID, the better the hemodynamic performance.1 Examples of OD (mm) Mitroflow OD (Labeled Size) Mosaic Ultra / Hancock II Ultra PERIMOUNT valve Magna Ease valve Internal Diameter (ID) For professional use. CAUTION: Federal (United States) law restricts this device to sale by or on the order of a physician. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events. Edwards Lifesciences Irvine, USAand Switzerland Tokyo, Japan SingaporeCorporation. Brazil I Nyon, I Singapore, I São Paulo, Edwards Edwards Magna Ease are Itrademarks of Edwards Lifesciences Lifesciences, the stylized E logo, edwards.com Carpentier-Edwards, PERIMOUNT, and Magna are trademarks of Edwards Lifesciences Corporation and are registered in the United States Patent and Trademark Office. 19 15.4 17.5 18 18 21 17.3 18.5 20 20 23 19 20.5 22 22 Mosaic, Mosaic ULTRA, Hancock II, and Hancock II ULTRA are trademarks of Medtronic, Inc. SJM Biocor and Epic are trademarks of St. Jude Medical, Inc. Mitroflow is a trademark of Sorin Bio-medica Cardio Srl. © 2010 Edwards Lifesciences Corporation. All rights reserved. AR05888 Edwards Lifesciences Irvine, USA I Nyon, Switzerland edwards.com I Tokyo, Japan I Singapore, Singapore I São Paulo, Brazil