Maintenance Manual
Transcription
Maintenance Manual
www.megoafek.com Calibrated Gradient Sequential Compression Therapy System Lympha Press Optimal Model 1201-XX Maintenance Manual Rev. 1.0 Manual Catalog No.: L93101C6 Publication Date: November 09 Maintenance Manual Table of Contents Table of Contents Contact Information ................................................................................................................1 Manufacturer Name and Address...........................................................................................1 MDD - Authorized Representative in Europe........................................................................1 Lympha Press USA - Authorized Representative in the USA ...............................................1 Preface.......................................................................................................................................2 1201 Model Family ................................................................................................................2 Symbols and Conventions used in this Manual .....................................................................2 Labels .....................................................................................................................................3 1 Introduction ........................................................................................................................4 2 Theory of Operation...........................................................................................................5 2.1 General..........................................................................................................................5 2.2 Electric/Pneumatic System ...........................................................................................5 2.3 Description of Operation ..............................................................................................6 2.4 Cycle Diagrams ............................................................................................................9 2.5 Safety Precautions ......................................................................................................10 2.5.1 2.5.2 2.5.3 2.5.4 2.5.5 3 4 5 6 7 8 In Case of Power Failure .................................................................................................. 10 Pressure Sensors ............................................................................................................... 10 Positioning Sensor ............................................................................................................ 10 Watchdog.......................................................................................................................... 10 Single Connection Option................................................................................................. 10 2.6 Quick Test Procedure .................................................................................................11 Troubleshooting................................................................................................................12 3.1 General........................................................................................................................12 3.2 Troubleshooting Guide ...............................................................................................12 3.3 Fault Isolation .............................................................................................................12 Disassembling the Device.................................................................................................15 Reassembling the Device..................................................................................................25 Replacing the Female Connector Cover.........................................................................30 SRU Parts List ..................................................................................................................37 Technical Support ............................................................................................................65 8.1 Calibration Procedure .................................................................................................65 8.1.1 8.1.2 8.2 8.3 Required Equipment ......................................................................................................... 65 Calibration ........................................................................................................................ 66 Setting Distributor Spring Force.................................................................................70 Safety Electrical Tests ................................................................................................71 8.3.1 8.3.2 Grounding Continuity Test ............................................................................................... 71 Earth Leakage Test ........................................................................................................... 71 Lympha Press Optimal Model 1201-XX Rev. 1.0 i Copyright © 2009 by Mego Afek AC Ltd. All rights reserved. All rights reserved. No part of this manual may be reproduced or copied in any form by any means including graphic, electronic or mechanical media, photocopying, typing, or other information retrieval systems, without the written permission of Mego Afek AC Ltd. Maintenance Manual Contact Information Contact Information Manufacturer Name and Address Mego Afek AC Ltd. Kibbutz Afek 30042 Israel Tel: 972-77-9084277 Fax: 972-4-8784148 E-mail: [email protected] Web: www.megoafek.com MDD - Authorized Representative in Europe Rolf Davidsen Helseagenturer A/s Postboks 238, Vestiieimvegen 35, Bygnes, 4250 Kopervik Norway Tel: +47-52-844500 Fax: +47-52-844509 Lympha Press USA - Authorized Representative in the USA Lympha Press USA 232 Park Avenue Manalapan, New Jersey 07726 Toll-Free: 888-LYMPHA-1 Tel: 732-792-9677 Fax: 732-792-9745 Web: www.lympha-press.com E-mail: [email protected] Lympha Press Optimal Model 1201-XX Rev. 1.0 1 Preface Maintenance Manual Preface Should you have any questions or require further information regarding the pressure control unit or component in the compression therapy system, please do not hesitate to contact us. You can visit our web site at www.megoafek.com for news, updates, and details on a local dealer in your area. For product support, send an e-mail to [email protected]. 1201 Model Family Cycle Name Model -Type Catalog # Voltage/Freq. Class Language 1201-EP L10002C6 230V/50Hz I English Optimal 1201-ED L10009C6 230V/50Hz II English 1201-AP L10003C6 115V/60Hz I English Pretherapy+ Lympha/ Sequential Lympha/ Sequential √ √ Pretherapy+ Wave/ Wave/Peristaltic Peristaltic √ Symbols and Conventions used in this Manual Symbol Description Note: Describes how Lympha Press Optimal functions or provides a note on how best to use it. Caution: Refers to the safe operation of Lympha Press Optimal and provides cautions where the possibility of loss of data or damage to the equipment exists. Warning: Indicates a situation in which the operator or the patient may be in danger of injury or death. 2 Lympha Press Optimal Model 1201-XX Rev. 1.0 √ Maintenance Manual Preface Labels The following labels and symbols appear on the pressure control unit, garments and/or packaging. Label Description Location Read instructions before use On electrical data label on base of pressure control unit Read instructions before use On garment label Temperature limitations for storage and/or transport of the pressure control unit On pressure control unit carton package Level of protection type BF equipment On electrical data label on base of pressure control unit Double insulation (for Class II device) On electrical data label on base of pressure control unit Manufacturer's name and address On garment label Date of Manufacture On base of pressure control unit Separate collection for waste electrical On bottom of pressure control and electronic equipment unit CE label showing compliance with EC On electrical data label on base Directive 93/42/EEC. Medical device of pressure control unit directive. CSA c/us Certification with respect to electric shock, fire and mechanical hazards only in accordance with CAN/CSA C22.2 No. 601.1-M90 (AM1+AM2) and UL 60601-1 (1st Edition) On back of pressure control unit Serial number of the pressure control unit On back of pressure control unit, near the power connector Lympha Press Optimal Model 1201-XX Rev. 1.0 3 Introduction Maintenance Manual 1 Introduction The Lympha Press Optimal Model 1201-EP user interface (as example to all models above): Pressure display Operating Treatment Time Top pressure control Treatment Time Controls Operation Controls Treatment Mode Selector Figure 1 - Front Panel Figure 2 - Power Inlet 4 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Theory of Operation 2 Theory of Operation 2.1 General An air compressor in the unit distributes air pressure through a series of hoses. The hoses are connected to a special sleeve that contains up to 24 pressure cells. The massage treatment cycle is unidirectional, originating at the lower part of the limb and progressing towards the center of the body. The inflation and emptying cycle allows application of pressure on the limbs without causing any discomfort to the patient. 2.2 Electric/Pneumatic System (See Figure 3 and also Figure 4). □ The Pump (5) supplies air pressure. □ Solenoid valve (7) is the outlet valve. It functions also as a safety valve, and it is a normally-open valve, which ensures that pressure does not remain in the system in case of electrical power failure. □ A relief valve (6) (preset to 150 mmHg), which is connected to the pressure outlet of the compressor, ensures that the pressure in the system never rises over this limit. □ Solenoid valve (8) selects the air inlet from outside (via silencer) for inflation cycles or from the garment (via the distributor) for Active Deflation (at the end of treatment). □ Solid State Relay - SSR (4) switches On and Off the Pump (5). □ A pair of Fuses (2) in the electric power inlet makes the system safe in an event of a short circuit. □ RFI filter board (20) prevents emission of RFI interferences from the device to its environment. □ Switching Power Supply unit (3) reduces the voltage in the device to 24 VDC. □ The Main Board (14) controls the operation of the entire system. □ CPU (15) on the Main Board (14), which is factory-burned, is the "brain" of the system. □ Sensors (16a & 16b) on the Main Board (14) measure the pressures in the system continuously during the treatment. □ Fan (17) cools the system inside the device. □ The front panel (18) controls enable setting the treatment parameters as follows: ▫ Buttons to set the treatment mode (18a) - in some models are locked under professional code ▫ ▫ ▫ ▫ ▫ Buttons to set the treatment time (18b) Buttons to set the top pressure (18e) Buttons to start and stop the treatment (18d) LCD screen (18c) displays all the required information to the user. USB connection on the Communication Board (19) enables setting treatment modes and pressures via PC. □ The Distributor (10) builds the treatment cycle according to the CPU (15) software. Its functioning will be described further on. □ The controlled air is transferred into a single (or a pair of) garment(s) via the Female Connector (11) using two 12 outlets (only one of them is shown in the illustration). Lympha Press Optimal Model 1201-XX Rev. 1.0 5 Theory of Operation Maintenance Manual 2.3 Description of Operation (See Figure 3 and Figure 4). 1. When turning On the Main Switch (1) at the back of the device, the LCD screen (18c) displays the treatment time at the left side, and the higher pressure at the right side (see Figure 1). At the right of the "Select" Mode buttons (18a), an orange LED turns On displaying the active treatment mode. 2. Before starting the treatment, three parameters can be set by the user: ▫ The Treatment Mode required, using the Select or buttons (18a) (can be locked for the end user). ▫ The Pressure required, using the Pressure or buttons (18e) (Top pressure can be limited for the end user). ▫ The Treatment Time required, using the Treatment Time or buttons (18b). Note: Changing the gradient profile of pressures is described thoroughly in the Lympha Press Optimal Professional Setup Guide (Cat. No. L98102C6). 3. Upon starting the treatment by pressing Start (18d), the SSR (4) turns On the Pump (5); the Outlet Valve (7) turns the air flow towards the distributor (10) while the Inlet Valve (8) remains deactivated to allow free air into the compressor (5). 4. The Step Motor (10a) turns the Rotor (10e) via a 2-wheel gear (Pinion-10b and Gear-10c). 5. The rotor (10e) routes the compressed air to the specific outlet according to the plan designed for the actual treatment. 6. All 12 air outlets are connected to the Female Connector (11), which splits each of them to 2 outlets, for the right (11a) & left (11b) sides. 7. After the Rotor (10e) finishes the cycle, it turns to the Deflation position, allowing the entire compressed air to deflate from the Garment(s) via the Exhaust Port (10h) at the Stator (10f). 8. After each cycle during the treatment, the Home Position Sensor (10g) verifies that the Rotor starts the next cycle from the correct position. 9. Each movement of the Rotor (10e) from one cell to the next is done only after the software has verified that the previous cell had reached the required pressure. This verification is done by calculation using the pressures measured by the two sensors: The Dynamic Sensor (16a) and the Static Sensor (16b). 10. During the entire operation time, the Cooling Fan (17) verifies that the temperature inside the device remains sufficiently low. 11. At the end of treatment (either manually or automatically when Treatment Time is over) Inlet Valve (8) turns towards the distributor, enabling the compressor to suck the entire air out of the garment. At the same time, the Outlet Valve (7) is released to allow the air out of the compressor. This Active Deflation continues until the software recognizes low vacuum in the system (via the sensors) - and then the device stops. 6 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Theory of Operation Figure 3 - Electric/Pneumatic Diagram Lympha Press Optimal Model 1201-XX Rev. 1.0 7 Theory of Operation Maintenance Manual Power input RFI Filter Board Air inlet 8 Switch Coil 20 1 Inlet valve 5 4 Pump SSR 3 Switching power supply board 150 mmHg Relief valve 6 14 7 Coil Outlet valve Main board 17 Fan Deflation line 10a Step motor Relief port 1 2 19 10 Rotor Pressure sensors Display board + Controllers 10h Stator Distributor 18 Home position Communication board Compressed air outlet Legend High voltage Low voltage Compressed Air Air Optimal Gradient Block diagram_011.mvd Figure 4 - Block Diagram 8 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Theory of Operation 2.4 Cycle Diagrams Figure 5 - Lympha/Sequential Cycle Figure 6 - Wave/Peristaltic Cycle Figure 7 - Pretherapy Cycle Lympha Press Optimal Model 1201-XX Rev. 1.0 9 Theory of Operation Maintenance Manual 2.5 Safety Precautions 2.5.1 In Case of Power Failure In case of a power failure, The Safety Valve (normally open valve) opens, deflating all the cells in the sleeves. If the patient wants to release the pressure from the garments quicker, the hose bundle air connectors can be easily disconnected from the device, deflating the pressure in the garment. 2.5.2 Pressure Sensors Pressure Sensors on the board verify that the pressure in the system does not rise over the maximum pressure allowed. If the pressure does rise over, the system will stop, and display an Error message on the screen. These sensors also verify that the pressure is not below the lowest pressure allowed. Too low pressure means that there is an open air line somewhere in the system. 2.5.3 Positioning Sensor Since the treatment cycle is built on the rotor that sends the pressure to the right destination port at a time, getting out of position may cause a wrong treatment. To avoid this problem, the Home Position Sensor on the Distributor verifies that the rotor starts the cycle at the right position. If, due to mechanical problem, the rotor is out of position - the system will stop, and display an Error message on the screen. 2.5.4 Watchdog In case of any malfunction in the program Software, an electric safety device called Watchdog identifies the problem and stops the treatment immediately, allowing the air pressure to deflate from the garments, and displaying an Error message on the screen. 2.5.5 Single Connection Option The 12-prong Plug is shaped asymmetrically and features a positioning mark to ensure correct connection to the device. 10 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Theory of Operation 2.6 Quick Test Procedure The purpose of the Quick Test is to verify that the device is operational, and that it meets its performance specifications. If malfunctions are detected during the Quick Test Procedure, refer to Troubleshooting. 1. Connect the device to the test kit. 2. Seal off the unused air outlet with the 12-prong Plug. 3. Turn On the device. 4. Set the Pressure to a valid pressure in the range of 60 to 90 mmHg. 5. Set the Treatment Mode Selector to Lympha/Sequential treatment. 6. Press Start . 7. Allow the device to operate for three cycles and verify correct operation, pressure level and noise level. 8. During operation, verify that you can increase and decrease the Treatment Time (by pressing Treatment Time or ). 9. During operation, verify that you can decrease and increase back the Top Pressure (by pressing Pressure and then ). 10. When the sleeve is inflated, press Pause/Stop - verify that Pause is displayed on the right side of the screen. 11. Wait up to two minutes to verify deflation of the sleeve. 12. Press Pause/Stop again. Lympha Press Optimal Model 1201-XX Rev. 1.0 11 Troubleshooting Maintenance Manual 3 Troubleshooting 3.1 General This section describes the procedures intended for use in troubleshooting a specific Shop Replaceable Unit (SRU). Use the procedures when a problem or malfunction is detected during operation or during the testing procedures. Warning: Only an authorized technician should open the device and use the troubleshooting guide. 3.2 Troubleshooting Guide The troubleshooting guide, described in Table 1, is used when a problem or malfunction is detected during operation or during the test procedures. The troubleshooting guide consists of the following columns: □ Symptom: Describes problems and malfunctions that might be detected during operation or during test procedures. □ Problem Cause: Describes the most likely reasons for malfunction symptoms. □ Corrective Actions: Details the maintenance procedures required to solve the problem. The maintenance procedures often lead to the replacement of one or more of the Lympha Press Optimal Shop Replaceable Units (SRUs). 3.3 Fault Isolation When a malfunction symptom is detected, start with the first possible cause and perform the first appropriate recommended action. If the symptom persists, perform the next recommended action, and so on, until the problem is solved. Notes: 1. Replace a defective SRU with one known to be operational for replacement. Follow the disassembling procedure (see Chapter 4 Disassembling the Device). 2. Before opening the device, remove the electric cord from the power supply. Listed below are a few reminders of basic fault isolation: 1. Check associated equipment connections. Make sure that all connections are securely connected to the correct connector. 2. Before continuing, check all accessories visually for any defects. 3. Remove the device cover and visually check the interior of the device. Inspect for evidence of overheating or short-circuiting. Check hoses and wires for good physical connections. Look for such indications as cut wires, charred components, and loose leads. Correct any discrepancies detected. 12 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Troubleshooting Table 1 - Troubleshooting Guide Before any replacement, check all electrical connections. Line # 1 Symptom LCD does not turn On Check Yes => Go To Line Main Switch Illuminates? No => Replace Go To VI Line Notes Power cord 2 Fuses 3 Main switch 4 4 Filter card 5 5 Power inlet --- Main board 7 Wire bundle from Power supply board to Main board 8 Power supply board 9 2 6 3 6 LED on main board blinks? 7 9 8 (1) (2), (3) 9 => Flat cable 10 10 => Display board --- 12 Fix connections 12 SSR control bundle 13 SSR 14 (1) Main board 15 (2) => Compressor Assy. --- (1) => Verify that vibrating parts are not touching stationary parts! 17 11 Compressor does not operate after Wire connections to Pressing "Start" SSR OK? 12 SSR LED illuminates? 15 13 14 15 16 Irregular noise during operation Are vibrating parts touching stationary parts? (2) 17 Noisy distributor? => Distributor Assy. 18 18 Noisy compressor? => Compressor Assy. --- --- => Display board 20 --- => Main board --- (2), (3) 19 LCD On but blank 20 (3) 21 Error 01 = Firmware problem --- => Main board --- (2), (3) 22 Error 02 = Problem writing to EPROM --- => Main board --- (2), (3) Lympha Press Optimal Model 1201-XX Rev. 1.0 13 Troubleshooting Line # 23 Maintenance Manual Symptom Error 04 = Home position flag was not found at initializing Check Yes => Go To Line No => Replace Go To VI Line Notes Is rotor flag broken? 24 Wire bundle from Main board to Home Position sensor 25 24 --- => Distributor Assy. (or gear wheel) 25 25 --- => Home position sensor --- 26 Error 05 = Fan failure N/A 27 Error 06 = Pressure exceeds maximum allowed value Are all hoses in the system not kinked? 28 Fix flow in hoses 28 Outlet valve operates? => Outlet valve Assy. --- 28 (2) 29 Error 08 = Home position flag is --out of allowed range => Distributor --- (3) 30 Error 09 = Home position flag was not found during operation Is rotor flag OK? 32 Distributor Assy. (or gear wheel) 31 (3) 31 --- => Wire bundle from main board to home position sensor 32 32 --- => Home position sensor --- 34 => 11 33 Error 10 = Pressure in a cell Compressor operates? cannot reach the required value within maximum time allowed (4) 34 All hoses connected properly? 35 Fix connections 35 35 All hoses unbroken? 36 Broken hose 36 36 All white O-rings of Female Connector in place? 37 Fix Female Connector 37 37 --- => Distributor Assy. --- Very Important Notes: 1. Whenever fixing a high voltage system in the device, perform the Safety Electrical Tests (see Section 8.3 - Safety Electrical Tests). 2. Whenever reconnecting the display board flat cable, verify that power is OFF, otherwise there is a good chance of damaging the display board. 3. Whenever replacing the Main Board or Distributor, perform the Calibration Procedure using the PC wizard (see Section 8.1 - Calibration Procedure). 4. The most common reason for Error 10, from the user aspect, is wrong connection of the garment. To confirm this problem, test the device using the laboratory test bench. 14 Lympha Press Optimal Model 1201-XX Rev. 1.0 (3) Maintenance Manual Disassembling the Device 4 Disassembling the Device Disconnect the device from the power outlet before removing the top cover. Each note on a picture refers to the corresponding part ID number in the specified table. For example, "T-4 #30" refers to Table 4, part ID no. 30. Each Table refers to the consecutive figures and diagrams. 1. Use a Philips screwdriver #3 to open the six screws that fix the top cover. 2. Top cover screws (the two short screws are from the handle). T-4 #28 T-4 #27 4. Disconnect the top cover cables from the main board connectors. 3. To remove the Top cover: Pull up the top cover. T-4 #26 T-2 #23 Lympha Press Optimal Model 1201-XX Rev. 1.0 T-2 #22 15 Disassembling the Device Maintenance Manual 5. Please note: Press the snap before pulling out the connector. 6. Remove the cover and put it aside. 7. Removing the Main Board: Disconnect all electrical connectors and the hoses from the dynamic pressure sensor (on the left) and from the static pressure sensor on the board. 8. Use a Philips screwdriver #3 to open the four screws that fix the panel to the base. T-2 #11-14 &18 T-3 #15 T-3 #14 9. Remove the main board panel. 10. Removing the Distributor: Disconnect the two hoses from the solenoid valves. T-3 #11 T-3 #10 T-4 #24 T-4 #23 16 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Disassembling the Device 11. Mark the hoses of the Female Connector from 1 to 12, starting from the right. 12. Pull out the hoses of the Female Connector using bent pliers. 13. Female Connector hoses are disconnected. 14. Use a Philips screwdriver #3 to open one screw on the front of the Distributor. 15. Open two more screws on the back. 16. Carefully cut the cable tie that secures the distributor hoses. 2 nd T-2 #16 screw 1st screw T-2 #15 T-3 #15 Lympha Press Optimal Model 1201-XX Rev. 1.0 17 Disassembling the Device Maintenance Manual 17. Carefully cut the cable tie that secures the distributor wiring. 18. Take out the distributor and its hoses. T-2 #15 19. Removing the pump & valves assembly: Use pliers to pull out the four connectors from the fuse houses. T-4 #21 20. Fuse connectors are removed. T-2 #5 T-2 #1 T-2 #6 T-2 #2 21. Use a flat screwdriver #1 to remove the four plastic covers on the back of the base cover. T-4 #8 & #10 18 T-2 #3 22. Use a Philips screwdriver to open the four screws that fix the pump & valves assembly. T-4 #5 & #9 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Disassembling the Device 23. Take out the pump & valves assembly from the base. 24. Disassembling the pump & valves assembly: Disconnect the power inlet connector from the filter board. T-4 #1 T-4 #18 25. Using a Philips screwdriver #3, open and release the grounding screw of the power inlet. 26. Put the power inlet aside. T-2 #2 T-2 #1 27. Cut the cable tie that secures the air hose from the compressor and the main switch to the compressor bracket. 28. Cut the cable tie that holds the wires of the filter board to the compressor bracket. T-3 #3 T-3 #4 Lympha Press Optimal Model 1201-XX Rev. 1.0 19 Disassembling the Device Maintenance Manual 29. Remove the connectors from the filter board and the power supply board. T-2 #6 30. Cut the cable tie that holds the wires from the terminal block. T-2 #6 T-2 #5 T-2 #8 31. Use a flat screwdriver #1 to disconnect the four wires connected to the bottom side of the terminal block. T-2 #8 T-2 #11 32. Remove the Main Switch Assy. T-2 #9 33. Cut two cable ties that secure the wiring to the bracket. Verify not to damage the wires insulation! 34. Cut the cable tie that secures the Outlet Valve wiring to the bracket. T-2 #18 T-3 #3 20 T-3 #4 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Disassembling the Device 35. Disconnect the wiring from the power supply board to the main board. 36. Remove the wiring. T-2 #18 37. Disconnect the white wire from the compressor to the SSR. T-2 #10 T-2 #8 38. Use a Philips screwdriver #3 to open two screws that fix the SSR to the compressor bracket. T-2 #11 T-2 #9 39. Remove the SSR. 40. Removing the Fan: Use a 3-mm Hex driver to open the four screws that fix the Fan. T-2 #14 T-3 #15 T-2 #10 T-2 #12 Lympha Press Optimal Model 1201-XX Rev. 1.0 21 Disassembling the Device 41. Remove the Fan. T-2 #12 Maintenance Manual 42. Disconnect the two air hoses from the compressor. T-3 #4 T-3 #3 43. Both hoses disconnected from the compressor. 44. Removing the Inlet Valve: Use a 3-mm Hex driver to open two screws that fix the Inlet Valve to the front of the compressor bracket. T-2 #13 45. Inlet Valve removed. 22 46. Use a 3-mm Hex driver to open two screws that fix the Outlet Valve Assy. to the side of the compressor bracket. Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual 47. Use a Philips screwdriver #3 to open the Disassembling the Device 48. Remove the Outlet Valve Assy. screw that fixes the Outlet Valve Assy. to the front of the compressor bracket. T-9 #7 & #15 T-9 #7 & #10 49. Use a Philips screwdriver #3 to open the green screw of the compressor grounding wire. 50. Use a Philips screwdriver #3 to open three screws that fix the power supply board bracket. 51. Remove the bracket. 52. Removing the two boards: Use a Philips screwdriver #3 to open four screws that fix each board. T-9 #23 T-9 #7 & #11 Lympha Press Optimal Model 1201-XX Rev. 1.0 T-9 #12 T-9 #21 23 Disassembling the Device Maintenance Manual 53. Remove the two boards. 54. Removing the Compressor: Use a Philips screwdriver #3 to open four screws that fix the compressor. T-9 #22 55. Remove the compressor. T-9 #2 T-9 #1 T-9 #6 & #7 24 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Reassembling the Device 5 Reassembling the Device After replacing the required SRU, reassemble the device in reverse order. Connect all electric and pneumatic connectors according to the diagrams in Figure 9, Figure 10, Figure 12, and Figure 13. Make sure to tie all wires and hoses as shown in the pictures above. Note: 1. Safety electrical tests should be performed whenever an electrical reassembly takes place (see Section 8.3 - Safety Electrical Tests). 2. To verify that the device is completely fixed, perform the Calibration Procedure (see Section 8.1 - Calibration Procedure). 1. Reassembling the Compressor 3. Reassembling the power supply bracket. Lympha Press Optimal Model 1201-XX Rev. 1.0 2. Reassembling the two boards 4. Reinstalling the green screw of the compressor grounding wire. 25 Reassembling the Device Maintenance Manual 5. Reinstalling the Outlet Valve Assy. on the front of the compressor bracket. 6. Reassembling the Outlet Valve Assy. on the side of the compressor bracket. 7. Reassembling the Inlet Valve Assy. on the front of the compressor bracket. 8. Reconnect the two air hoses to the compressor. 9. Reassembling the Fan. Verify that the airflow arrow on the fan is from inside out (from left to right in this figure). 10. Reassembling the SSR on the compressor bracket. Verify that the low-voltage side is upward. 26 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Reassembling the Device 11. Reconnecting the white wire incoming the SSR from the compressor. After tightening the screws, verify that the wires cannot be pulled out easily. 12. Reconnecting the wiring from the terminal block to the power supply board. 13. Reconnecting the four wires to the bottom side of the terminal block (brown & white wires at the right side, and black wire at the left side). 14. Reattach the two connectors to the boards. 15. Reconnecting the power inlet connector & grounding wire. 16. Reconnect the pump & valves assembly to the base, and then reassemble the plastic covers on the screw heads. Lympha Press Optimal Model 1201-XX Rev. 1.0 27 Reassembling the Device Maintenance Manual 17. Reconnecting the four connectors to the fuse houses: The bundle from the filter to the inner fuse connectors, the main switch bundle connectors to the outer fuse connectors, the brown wires to the fuse on the right, and the blue wires to the fuse on the left. 18. Reinstall the distributor, tightening two screws on the back, and then the screw on the front. 19. Reconnecting the hoses to the Female Connector according to the numbering from right to left (see Figure 13). 20. Reconnecting the two hoses to the Outlet Valve (see Figure 13). T-3 #11 T-3 #10 28 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Reassembling the Device 21. Reinstalling the main board panel (four screws) 22. Reconnecting all connectors (according to Figure 9), and the hoses to the dynamic pressure sensor (on the left) and to the static pressure sensor on the board. 23. Reconnect the top cover cables to the main board connectors. 24. Reassemble the top cover. 25. Use a Philips screwdriver #3 to screw & tighten the six screws that fix the top cover (the two short screws are for the handle). Lympha Press Optimal Model 1201-XX Rev. 1.0 29 Replacing the Female Connector Cover Maintenance Manual 6 Replacing the Female Connector Cover See Table 17 & Figure 37. (Top cover removal is not necessary) 1. Tilt the device upward and remove the two black plastic covers using a flat screwdriver #1. 2. Release the two screws 3-4 turns (not more!) using a Philips screwdriver #3. T-4 #11 3. Raise the bottom side of the connector cover, using a flat screwdriver #1 and release it. 4. Replacement of O-rings. T-17 #3 T-17 #4 5. Return the Connector cover, top side first. Make sure the side with the three protrusions faces downwards. 30 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Replacing the Female Connector Cover Table 2 - Wiring Parts No. 1 Catalog No. Description Qty Notes L26900C3 Power inlet Assy. - Opt. Class I 115V 1 L26900C1 Power inlet Assy. - Optimal Class II 230V 1 L26900C2 Power inlet Assy. - Opt. Class I 230V 1 L26800C2 Main switch Assy. - Optimal 115V 1 L26800C4 Main switch Assy. - Optimal 230V 1 L26C120D Ground wire Assy. Comp. 115V 120mm 2 N.A. for Class II L26D120D Ground wire Assy. Comp. 230V 120mm 2 N.A. for Class II 4 LD290002 Fuse Holder 2 4a L36651A4 Fuse S.B. 2A 5X20 2 5 L26100C1 Filter to fuses bundle - Optimal 1 6 L26100C2 Terminal to Power sup. bundle - Optimal 1 7 L3645002 Terminal block #2 1 8 L36250C1 White SSR wire 1 9 L22000C6 Compressor Assy. Optimal Grad. 115V 1 L22000C7 Pump Assy. - Opt. Grad. 230V Class II 1 L22000C8 Pump Assy. - Optimal Gradient 230V 1 10 L36500C1 SSR Halon HGK-310D/1 1 11 L2D001C1 SSR control bundle 1 12 L26800C1 Fan Assy. Optimal 1 13 L24300C3 Inlet valve Assy. 1 14 L24300C2 Outlet Valve Assy. 1 15 L39081C5 Step Motor Assy. (LIN 4118M) 1 16 L26100C5 Home Position Sensor bundle 1 17 LD0A4009 Sensor Omron SE-SX4009 P1 1 18 L26100C6 SPS to main board bundle 1 19 L2D002C2 Switching Power SupplyVNRM65-24V 1 20 L2D004C1 RFI filter card 1 21 L2D000C6 Main Board - Optimal Gradient 1 22 L2D002C4 Display Board - Optimal 1 23 L2D011C4 Flat cable 40 pin - Optimal 1 2 3 Lympha Press Optimal Model 1201-XX Rev. 1.0 31 Replacing the Female Connector Cover Maintenance Manual Switching Power Supply board 2 1 20 Main Switch 2A S.B. High Voltage 4 Main Board 19 Low Voltage 24 Display Board 22 21 2A S.B. RFI Filter Power Inlet COM Board 17 Compressor 15 9 Sensor Step Motor SSR Inlet Valve 14 Outlet Valve 13 10 Fan 12 Figure 8 - Wiring Schematics Inlet Valve Outlet Valve Fan Y-B-R DETAIL A Red 12 Bu G R Bk 14 R O Y 13 Yellow DETAIL B 3 Compressor grounding 1 2 2 18 On Off Main switch Power inlet 15 Outlet valve 17 4 19 Switching power supply card Fan RFI Filter card 6 12 Inlet valve 20 5 7 8 Black wire 9 of compressor 21 Main board 10 A B Inner panel grounding 3 9 White wire of compressor 16 15 11 13 12 14 Figure 9 - Wiring Diagram 1 32 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Replacing the Female Connector Cover Connecting the Cover to the Base C 23 Bk G R W R Bk USB 24 Remote Stop DETAIL C 22 Figure 10 - Wiring Diagram 2 Lympha Press Optimal Model 1201-XX Rev. 1.0 33 Replacing the Female Connector Cover Maintenance Manual Table 3 - Hose List No. Cat. No. Description Notes From To 1 L24C180L Tube 180mm Distributor Optimal Distributor Female connector #4 - #9 2 L24C210L Tube 210mm Distributor Optimal Distributor Female connector #10 - #12 3 L5910610 Silicon tube 6/10 430 mm Compressor inlet Inlet valve 4 L5910610 Silicon tube 6/10 230 mm Compressor outlet Relief valve 5 L22000C6 Compressor Assy. Optimal Grad. 115V L22000C7 Pump Assy. Opt. Grad. 230V Class II L22000C8 Pump Assy - Optimal Grad. 230V 6 L3466016 Relief valve Assy. 7 L245C006 Outlet valve Assy 8 L24300C3 Inlet valve Assy 9 L29000C6 Distributor Assy - Optimal Grad. 10 L5910610 Silicon tube 6/10 230 mm Outlet valve Distributor 11 L5910610 Silicon tube 6/10 290 mm Distributor Inlet valve 12 L24C130L Tube 130mm Distributor Optimal 13 L2300121 Female Connector Assy. 14 L5910204 Silicon tube 2/4 180 mm Distributor Static sensor 15 L5910204 Silicon tube 2/4 320 mm Outlet valve Dynamic sensor 16 L24C155L Tube 155mm Distributor Optimal 34 Distributor Distributor Female connector #1 Female connector #2 - #3 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Replacing the Female Connector Cover Figure 11 - Pneumatic Schematics 7 6 5 8 4 3 S1 9 S2 10 2 x3 1 x6 15 16 x2 14 13 12 11 Figure 12 - Pneumatic Diagram 1 Lympha Press Optimal Model 1201-XX Rev. 1.0 35 Replacing the Female Connector Cover Maintenance Manual 12 11 10 9 8 7 6 5 4 3 2 1 S1 S2 Hoses numbering Distributor view 8 6 10 9 7 5 4 12 11 3 2 S 1 P S - Sensor P - Pressure (in) V - Vacuum (out) V \\Megosrv\company\\מחלקת הפיתוח\פרוייקטיםMM\1201-xx\Diagrams\Piping - 0pt. Grad.dft Figure 13 - Pneumatic Diagram 2 36 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 7 SRU Parts List Table 4 - SRU Parts List Part ID Part No. Description Qty Recommended SRU 115V All 230V 1 L27B00C4 Base for Optimal - light grey 1 2 L32521A2 RUBBER BUMPER 35Shore A M4*16mm 3 3 L2300121 Female Connector Assy. 1 4 L6100100 Nut M2.5 2 5 L6110480 Pan head screw M4x8 10 6 L6060400 Spring Washer 4mm 5 7 L6040100 Washer 4x14 2 8 L3714024 Flat Bottom Snap-Cap washer 8/8-180 5 9 L6044008N Washer DIN 433 S.S 4.3X8 5 10 L3714023 Black SNAP-CAP 8/8-180 5 11 L3714025 Screw cover cap 6-10 CAP075-BLK 2 12 L6040950 Washer 2.8d*6.8D*0.4h 2 13 L6030200 Screw M2.5x30 2 14 LD290002b Fuse Holder (cup) 2 15 L36651a4 Fuse S.B. 2A 5X20 2 16 LD290002a Fuse Holder (housing) 2 17 LD290002c Fuse Holder (nut) 2 18 L27B01C6 Pump & Valves Assy. - Optimal Grad. 115V 1 L27B01C7 Pump & Valves Assy. - Opt. Grad. Class II 230V 1 L27B01C8 Pump & Valves Assy. - Optimal Grad. 230V 1 L26800C2 Main Switch Assy. - Optimal 115V 1 L26800C4 Main Switch Assy. - Optimal 230V 1 L26900C1 Power Inlet Assy. - Optimal Class II 1 L26900C2 Power Inlet Assy. - Opt. Class I 230V 1 L26900C3 Power Inlet Assy. - Opt. Class I 115V 1 21 L29000C6 Distributor Assy. - Optimal 1 22 L6040400 Washer 5/32" 1 23 L27B02C6 Main Board Assy. - Opt. Grad. 1 24 L6020200 Screw 3.0 x 10 4 19 20 Lympha Press Optimal Model 1201-XX Rev. 1.0 37 SRU Parts List Part ID Maintenance Manual Part No. Description Qty Recommended SRU 115V 25 All 230V L37240C6 Front decal - Optimal Gradient - USA 1 L37240C7 Front decal - Optimal Gradient EU 1 L37240C8 Front decal - Optimal Gradient GER 1 26 L27C00C6 Cover Assy. - Optimal Gradient 1 27 L6020600 Screw 3.5 x 20 2 28 L6020300 Screw 3.5 x 30 4 29 L37240C0 Lympha Press Label - Dome 1 30 L2D011C4 Flat cable 40 pin - Optimal 1 38 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 29 25 26 30 27 28 Figure 14 - Device Breakdown 1 Lympha Press Optimal Model 1201-XX Rev. 1.0 39 SRU Parts List Maintenance Manual 23 24 Figure 15 - Device Breakdown 2 40 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 5 6 21 22 5 x2 6 x2 Figure 16 - Device Breakdown 3 Lympha Press Optimal Model 1201-XX Rev. 1.0 41 SRU Parts List Maintenance Manual 18 1 19 20 x4 8 x4 9 x4 5 x4 10 Figure 17 - Device Breakdown 4 42 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 4 x2 3 2 x3 1 17 x2 5 x3 x2 6 x2 7 8 16 x2 9 10 15 x2 x2 11 14 x2 x2 12 x2 13 Figure 18 - Device Breakdown 5 Lympha Press Optimal Model 1201-XX Rev. 1.0 43 SRU Parts List Maintenance Manual Table 5 - Cover Assembly (Part No. L27C00C6) (Figure 19 & Figure 20) Part ID. Part No. Description Qty Recommended SRU 1 L37134C4R Cover for Optimal - light grey 1 2 L3710121 Spacer for Mini casing 4 3 L27C01C6 Front Panel Assy. - Optimal Grad. 1 4 L6023015 Screw Dia 3.0 x 15 4 5 L6023008 Screw 3.0 x 8 1 1 4 3 Figure 19 - Cover Assembly Breakdown 1 44 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 2 P:\Mini Plus\Alfa series\îëìåì îëñä\L27C00C4.dft / MM-2 1 Figure 20 - Cover Assembly Breakdown 2 Lympha Press Optimal Model 1201-XX Rev. 1.0 45 SRU Parts List Maintenance Manual Table 6 - Front Panel Assembly (Part No. L27C01C6) (Figure 21 Figure 22 & Figure 23) Part ID. Part No. Description Qty 1 L37111C6 Front Panel-Optimal Grad. 430U 1 2 L2D003C4 COM Board - Optimal 1 3 L6023006 Screw 3.0 x 6 8 4 L37220C2 Operating keypad - Optimal 1 5 L37220C3 +/- keypad - Optimal 1 6 L37220C1 Mode keypad - Optimal 1 7 L2D002C4 Display Board - Optimal 1 8 L6023008 Screw 3.0 x 8 5 9 L37112C4 Inner Panel - Optimal Grey 430U 1 5 6 4 Recommended SRU 3 2 1 \\megosrv\solidedge\LYMPHA\Booklet drawings\Optimal Grad\Cover Ass.dft Figure 21 - Front Panel Assembly Breakdown 1 46 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 3 7 \\megosrv\solidedge\LYMPHA\Booklet drawings\Optimal Grad\Cover Ass.dft Figure 22 - Front Panel Assembly Breakdown 2 Lympha Press Optimal Model 1201-XX Rev. 1.0 47 SRU Parts List Maintenance Manual 8 9 \\megosrv\solidedge\LYMPHA\Booklet drawings\Optimal Grad\Cover Ass.dft Figure 23 - Front Panel Assembly Breakdown 3 48 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List Table 7 - Base Assembly (Part No. L27B00C4) (Figure 24) Part ID. Part No. Description Qty 1 L27B00C4 Base for Optimal - light grey 1 2 L3752115 Silicone bumper 3 3 L37100C3 Locking disk for bumper 4 4 L6020700 Screw 3.5 x 12 4 5 L3752116 Silicone bumper - cut 1 Recommended SRU 4 3 2 (x3) 5 1 Figure 24 - Base Assembly Breakdown Lympha Press Optimal Model 1201-XX Rev. 1.0 49 SRU Parts List Maintenance Manual Table 8 - Main Board Assembly (Part No. L27B02C6) (Figure 25 & Figure 26) Part ID. Part No. Description Qty 1 L37110C6 Inner panel - Optimal Grad. 1 2 L2D001C6 Main Board - Optimal Gradient 1 3 L6220306 Screw M3x6 + integral tooth washer 4 4 L26C120D Compressor earth cable 1 5 L6130301 Screw 3.5x6.5 green 1 6 L6040300 Tooth washer 3mm 1 Recommended SRU 3 2 1 Figure 25 - Main Board Assembly Breakdown 1 50 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 4 6 5 Figure 26 - Main Board Assembly Breakdown 2 Lympha Press Optimal Model 1201-XX Rev. 1.0 51 SRU Parts List Maintenance Manual Table 9 - Pump & Valves Assembly (Part No. L27B01C6) (Figure 27, Figure 28, Figure 29) Part ID. Part No. Description Qty Recommended SRU 115V All 230V 1 L37021C2 Pump & Valves bracket 1 2 L22000C6 Pump Assy. - Optimal Gradient 115V 1 L22000C7 Pump Assy. - Opt. Grad. 230V Class II 1 L22000C8 Pump Assy. - Optimal Gradient 230V 1 3 L3645002 Terminal #2 1 4 L6110312 Pan head screw M3 X 12 1 5 L6040300 Tooth washer 3mm 2 6 L6110200 Pan head screw W5/32x3/8 4 7 L6060400 Spring washer 4mm 10 8 L36500C1 SSR Halon HGK-210D 1 9 L6040600 Tooth washer 4mm 7 10 L6030100 Screw M4 X 10 3 11 L6110480 Pan head screw M4x8 3 12 L37022C2 Power supply bracket 1 13 L9200700 Grounding label 1 14 L245C006 Outlet valve Assy 1 15 L6010200 Socket head screw M4x25 2 16 L24300C3 Inlet valve Assy 1 17 L6010300 Screw M4 X 30 2 18 LD0J0003 Cable tie adhesive mount 19x19mm 2 19 L37140C0 Sponge cable holder 1 20 L6130301 Screw 3.5x6.5 green 1 21 L2D004C1 Filter board - Optimal 1 22 L6220306 Screw M3x6 + Integral tooth washer 8 23 L2D002C2 Switching Power card VNRM65-24V 1 24 L26800C1 Fan Assy. - Optimal 1 25 L6010410 Socket head screw M4X10 4 52 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 22 x8 21 23 20 9 For connection of pump grounding wire 19 24 18 x2 9 x4 25 x4 Pump & Valves Assy. Breakdown 2 Figure 27 - Pump & Valves Assembly Breakdown 1 Lympha Press Optimal Model 1201-XX Rev. 1.0 53 SRU Parts List Maintenance Manual DETAIL A A 11 7 14 12 11 x2 13 7 x2 15 x2 7 x2 7 10 16 10 x2 9 x2 8 Pump & Valves Assy. Breakdown 3 17 x2 Figure 28 - Pump & Valves Assembly Breakdown 2 54 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List 2 1 3 4 5 7 6 Figure 29 - Pump & Valves Assembly Breakdown 3 Lympha Press Optimal Model 1201-XX Rev. 1.0 55 SRU Parts List Maintenance Manual Table 10 - Outlet Valve Assy (Part No. L245C006) (Figure 30) No. Catalog No. Description 1 L34660C6 Manifold - Optimal Grad. 1 2 L3440007 Nipple 1/8 NPT to 6mm 1 3 L6051816 O-Ring N70 2-014 1 4 L3460154 5/8" OD Cartridge CV 3PSI-Green 1 5 L37022C6 Check valve cover - Optimal 1 6 L6030310 Screw M3 X 10 DIN 965A 2 7 L24300C2 Pressure line valve Assy. 1 8 L6121414M M/F Elbow + M5 outlet 1/4" 1 9 L3440012 Minimal Fitting LC-0425-M5 1 10 L3240226 Brass nipple, 1/4 BSPT for 6/9 1 11 L3420002 Silencer 1/4 1 12 L3440027 Nipple 1/4" BSP x 1/4" NPT 1 13 L3412014 Sealing ring for 1/4 BSP 1 14 L37020C6 Manifold support - Opt. Grad. 1 15 L245C016 Relief valve Assy. - Optimal Grad. 1 12 13 Qty 14 1 Recommended SRU 2 15 10 11 3 4 1/4" BSP Thread 1/4" NPT Thread 9 8 5 6 (x2) 7 Figure 30 - Outlet Valve Assembly Breakdown 56 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List Table 11 - Distributor Assembly (Part No. L29000C6) (Figure 31) Part ID. Part No. Description Qty 1 L6010200 Socket head screw M4X25 2 2 L6100300 Lock nut M4 2 Recommended SRU 2 1 \\megosrv\solidedge\LYMPHA\mechanizm\assebly\distributor\L29000C6\V-05\L29000C6-05.dft / MM-1 Figure 31 - Distributor Assembly Breakdown 1 Lympha Press Optimal Model 1201-XX Rev. 1.0 57 SRU Parts List Maintenance Manual Table 12 - Distributor Assembly (Part No. L29000C6) (Figure 32) Part ID. Part No. Description Qty 1 L39020C6 Distributor bracket - Optimal 1 2 LD0A4009 Sensor Omron SE-SX4009 P1 1 3 L6040600 Tooth washer 4mm 1 4 L6110415 Screw M4x15 1 5 L29200C4 Step Motor Assy. - Optimal 1 6 L6010308 Socket screw M3x8 3 7 L6060300 Spring washer 3.0 3 8 L6040800 Washer 1/8 3 5 4 3 2 Recommended SRU 1 \\megosrv\solidedge\LYMPHA\mechanizm\assebly\distributor\L29000C6\V-05\L29000C6-05.dft / MM-2 8 7 6 Figure 32 - Distributor Assembly Breakdown 2 58 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List Table 13 - Distributor Assembly (Part No. L29000C6) (Figure 33) Part ID. Part No. Description Qty 1 Figure 34 Stator & Rotor Assy. 1 2 L6051819 O-ring N70 2-016 2 3 L6051808 O-ring N70 2-011 2 4 L3437311 Nipple hose 6/9 - O-ring 2-011 2 5 L3460153 OD Cartridge CV 0.11PSI 150 5/8" 2 6 L6050150 O-Ring N80 5.7*13.9 13 7 L39100C6 Adaptor for nipple 2/4 1 8 L3440012 Minimal Fitting LC-0425-M5 1 9 L6100300 Lock nut M4 3 10 L24C130L Tube 130mm Distributor Optimal 1 11 L24C155L Tube 155mm Distributor Optimal 2 12 L24C180L Tube 180mm Distributor Optimal 6 13 L24C210L Tube 210mm Distributor Optimal 3 14 L39100C5 Adaptor 1/4" to CV 2 15 L39021C6 Cover for distributor Optimal Gradient 1 16 L6034020 Screw, M4x20 DIN 965 3 17 L3420002 Silencer 1/4" 2 Lympha Press Optimal Model 1201-XX Rev. 1.0 Recommended SRU 59 SRU Parts List Maintenance Manual 17 View A 16 8 6 10 9 7 5 4 12 11 3 2 1 15 14 13 View A 12 5 11 4 10 3 2 6 7 8 6 1 9 \\megosrv\solidedge\LYMPHA\mechanizm\assebly\distributor\L29000C6\V-05\L29000C6-05.dft / MM-3 Figure 33 - Distributor Assembly Breakdown 3 60 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List Table 14 - Distributor Assembly (Part No. L29000C6) (Figure 34) Part ID. Part No. Description Qty Figure 15 Stator Assy. 1 2 Figure 16 Rotor Assy. 1 3 L39020C5 Spring sit Optimal 1 4 L3918251 Spring Distributor Gradient 1 5 L6046012 Washer 6x12x1.6mm 1 6 L6106000 Lock Nut M6 1 16.3 Using gauge L715040-00 1 Recommended SRU 1 2 6 3 4 5 \\megosrv\solidedge\LYMPHA\mechanizm\assebly\distributor\L29000C6\V-05\L29000C6-05.dft / MM-4 Figure 34 - Distributor Assembly Breakdown 4 Lympha Press Optimal Model 1201-XX Rev. 1.0 61 SRU Parts List Maintenance Manual Table 15 - Distributor Assembly (Part No. L29000C6) (Figure 35) Part ID. Part No. Description Qty 1 L39040C4 Distributor axis 1 2 L6090212 Roll Pin 2x12 DIN 1481 1 3 L6048016 Washer 8x16x1.6 mm 1 4 L6088001 C retaining ring 8mm 1 5 L39173C6.S Stator - Optimal Gradient 1 6 L6046012 Washer 6x12x1.6mm 1 7 L6106000 Lock Nut M6 1 Recommended SRU 4 3 1 2 5 6 7 \\megosrv\solidedge\LYMPHA\mechanizm\assebly\distributor\L29000C6\V-05\L29000C6-05.dft / MM-5 Figure 35 - Distributor Assembly Breakdown 5 62 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual SRU Parts List Table 16 - Distributor Assembly (Part No. L29000C6) (Figure 36) Part ID. Part No. Description Qty 1 L39173C6.R Rotor 1 2 L39061C1 Gear 72T 15M 25 deg. 1 3 L39040C6 Bearing FAG 608-2RS-C3 1 Recommended SRU 3 2 1 \\megosrv\solidedge\LYMPHA\mechanizm\assebly\distributor\L29000C6\V-05\L29000C6-05.dft / MM-6 Figure 36 - Distributor Assembly Breakdown 6 Lympha Press Optimal Model 1201-XX Rev. 1.0 63 SRU Parts List Maintenance Manual Table 17 - Female Connector (Part No. L2300121) (Figure 37) Part ID. Part No. Description Qty 1 Top Cover 1 2 Bottom Cover 1 Recommended SRU 3 L6050110 White O-Ring EPDM N70 2-010 24 4 L2313321 Front cover for Female Connector 2 1 2 3 4 \\megosrv\solidedge\LYMPHA\Booklet drawings\Optimal Grad\Cover Ass.dft Figure 37 - Female Connector Breakdown 64 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Technical Support 8 Technical Support 8.1 Calibration Procedure Warning: This procedure is setting new parameters to both rotor positioning and pressure sensor. Hence, do not start this procedure without the equipment needed for pressure calibration as described in the following section. 8.1.1 Required Equipment Calibration Sleeve Simulator Lympha Press Optimal Model 1201-XX Rev. 1.0 Pressure Source: Phlebo Press device equipped with a special hose bundle and pressure gauge. 65 Technical Support Maintenance Manual 8.1.2 Calibration 1. Remove the six screws of the cover and move it aside as shown. DO NOT disconnect the cover wiring to the main board! 2. Replace the static-sensor hose with the pressure-source hose. 3. Connect the device to both hose bundles of the Calibration Sleeve Simulator. 4. Connect the device to the PC using the USB cable. 5. Turn On the main switch. Verify that the USB signal blinks at the top-right corner of the LCD. 6. Open the Optimal Wizard. 7. Wait to receive the "Device: Ready" message at the low-left corner. USB signal 66 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Technical Support 8. Click Next. 9. Click Start Backlash Calibration. The Distributor will rotate a few times back and forth. 10. When this rotation stops, click Next. 11. Setting the Home Position: The two slots of the Rotor and Stator should be aligned. If they are aligned (see Step 13), proceed to Step 14. If they are not aligned, continue with Steps 12 - 13. 12. Click → to rotate the Rotor in one direction or ← for the opposite direction. You can change the length of the rotating step by changing the value in the middle box. Lympha Press Optimal Model 1201-XX Rev. 1.0 67 Technical Support Maintenance Manual 13. When the two slots are aligned (see below), proceed to Step 14. 14. Click "Set current axis position as zero position". 15. Activate the pressure source and set it to the low pressure point - 30 mmHg. 16. Verify the pressure accuracy on the pressure gauge. 17. Type the exact pressure displayed on the gauge in the text box. (If the gauge displays 21, type 21) 18. Click "Send first measurement". 19. Set the pressure to the high pressure point 80 mmHg. 68 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Technical Support 20. Type the exact pressure displayed on the gauge in the text box. 21. Click "Send second measurement and Calibrate". 22. Set the pressure to 50 mmHg. 23. Verify that the pressure displayed on the screen is the same as the pressure displayed on the gauge, ±1 mmHg. 24. Click Next. 25. The system is now ready for Dynamic Sensor Calibration, which is performed automatically, using the cells selected in the text boxes (default: 2 & 7) 26. Click "Start Dynamic Calibration". 27. Verify that the pressure on gauges 2, 3 and 7, 8 is 50 mmHg 28. When the procedure is over, click Next. 9 4 Lympha Press Optimal Model 1201-XX Rev. 1.0 3 8 2 7 1 69 Technical Support 29. Click Finish. Maintenance Manual 30. Replace the pressure-source hose with the static-sensor hose. 8.2 Setting Distributor Spring Force The force of the distributor spring is determined by the distance between its two holders. To set this distance as required, use gauge No. L715040-00. 1. Insert the gauge onto the spring between the two holders. 2. Using a 10-mm nut driver, set the distance to fit the gauge width. 70 Lympha Press Optimal Model 1201-XX Rev. 1.0 Maintenance Manual Technical Support 8.3 Safety Electrical Tests Safety electrical tests should be performed whenever electrical reassembly takes place. 8.3.1 Grounding Continuity Test Equipment: Grounding tester that can force 25 A for 10 seconds. Test Procedure: A. Turn On the main switch of the tested device. A. Connect the ground electrode clamp to the ground pin of the power plug. B. Connect the second electrode clamp to the inner bracket. C. Operate the tester for 10 seconds and verify that the tester displays Pass. D. Move the second electrode clamp from the inner bracket to a metal part of the compressor, and repeat step D. E. Turn Off the tester and the device, and disconnect the device from the tester. 8.3.2 Earth Leakage Test Equipment: Leakage tester that can measure µA up to 500 µA. Test Procedure: A. Connect one of the tester electrodes to the grounding pin of the device. B. Connect the second electrode to the grounding of the power outlet. C. Operate the tested device and verify that the leakage to the ground does not exceed 500 µA. Lympha Press Optimal Model 1201-XX Rev. 1.0 71