series surgical lights - CAN

Transcription

series surgical lights - CAN
SERIES
SURGICAL LIGHTS
Skytron's Aurora LED Surgical Lighting provides advanced bright, cool and color
correct illumination to the surgical field and a selectable color temperature of 4,000K
(soft white) or 4,500K (bright white) based upon surgeon preference for every
procedure. Focusable LEDs extend illumination range while retaining maximum
intensity at any distance, with LED pod life retaining 80% of original intensity for an
average of 10 years or 20,000 hours of use. Typical applications include General,
Orthopedic, Plastics, Vascular, Cardiovascular, OB/GYN, Trauma, Neurosurgery and
Urology.
TEC-B-0012 REV1
8/09
AUR5
2267mm
89 1/4"
945mm
37 1/4"
70mm
2 3/4"
360º
1000mm
39 1/2"
3/4" Diameter
Conduit
360º
360º
110º
Wall
Control
1765mm
69 1/2"
2032mm
80"
360º
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 144 watts (490 BTU/hr)
• Amperage: 1.6 amps
• Intensity: 4,000K=135,000 Lux (12,542fc)
4,500K=120,000 Lux (11,148fc)
• Fixture Weight: 132 lbs./59.9 Kg.
• Moment Load: 393.2 ft. lbs./532.8 N•m
CEILING HEIGHT
8'-4" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-01 REV0
8/09
AUR5TV
2267mm
89 1/4"
945mm
37 1/4"
70mm
2 3/4"
360º
1000mm
39 1/2"
3/4" Diameter
Conduit
1" Diameter
Conduit
360º
CAMERA
READY
360º
110º
Wall
Control
1765mm
69 1/2"
Camera
Control
Connector
2032mm
80"
360º
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 144 watts (490 BTU/hr)
• Amperage: 1.6 amps
• Intensity: 4,000K=135,000 Lux (12,542fc)
4,500K=120,000 Lux (11,148fc)
• Fixture Weight: 137.6 lbs./62.4 Kg.
• Moment Load: 426 ft. lbs./577.3 N•m
CEILING HEIGHT
8'-4" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-02 REV0
8/09
AUR7
2716mm
107"
1140mm
45"
1200mm
47 1/4"
70mm
2 3/4"
360º
3/4" Diameter
Conduit
360º
360º
110º
Wall
Control
1969mm
77 1/2"
2032mm
80"
360º
425mm
16 3/4"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 144 watts (490 BTU/hr)
• Amperage: 1.6 amps
• Intensity: 4,000K=160,000 Lux (14,870fc)
4,500K=139,000 Lux (12,918fc)
• Fixture Weight: 149 lbs./67.6 Kg.
• Moment Load: 616.2 ft. lbs./835 N•m
CEILING HEIGHT
8'-4" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-03 REV0
8/09
AUR7TV
2716mm
107"
1140mm
45"
1200mm
47 1/4"
70mm
2 3/4"
360º
3/4" Diameter
Conduit
1" Diameter
Conduit
360º
CAMERA
READY
360º
110º
Wall
Control
1969mm
77 1/2"
Camera
Control
Connector
2032mm
80"
360º
425mm
16 3/4"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 144 watts (490 BTU/hr)
• Amperage: 1.6 amps
• Intensity: 4,000K=160,000 Lux (14,870fc)
4,500K=139,000 Lux (12,918fc)
• Fixture Weight: 156.5 lbs./71 Kg.
• Moment Load: 655.9 ft. lbs./888.8 N•m
CEILING HEIGHT
8'-4" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-04 REV0
8/09
AUR55
2462mm
97"
1140mm
45"
70mm
2 3/4"
945mm
37 1/4"
360º
1000mm
39 1/2"
3/4" Diameter
Conduit
360º
360º
110º
360º
360º
Wall
Control
1765mm
69 1/2"
2032mm
80"
360º
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 288 watts (980 BTU/hr)
• Amperage: 3.2 amps
• Intensity: 4,000K=270,000 Lux (25,084fc)
4,500K=240,000 Lux (22,296fc)
• Fixture Weight: 215.2 lbs./97.6 Kg.
• Moment Load: 807.6 ft. lbs./1,094.3 N•m
CEILING HEIGHT
8'-9" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-05 REV0
8/09
AUR55TV
2462mm
97"
1140mm
45"
70mm
2 3/4"
945mm
37 1/4"
360º
1000mm
39 1/2"
3/4" Diameter
Conduit
360º
360º
110º
1" Diameter
Conduit
360º
CAMERA
READY
360º
Wall
Control
1765mm
69 1/2"
Camera
Control
Connector
2032mm
80"
360º
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 288 watts (980 BTU/hr)
• Amperage: 3.2 amps
• Intensity: 4,000K=270,000 Lux (25,084fc)
4,500K=240,000 Lux (22,296fc)
• Fixture Weight: 220.7 lbs./100.1 Kg.
• Moment Load: 840.5 ft. lbs./1,138.9 N•m
CEILING HEIGHT
8'-9" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-06 REV0
8/09
AUR75
2716mm
107"
1200mm
47 1/4"
1140mm
45"
70mm
2 3/4"
945mm
37 1/4"
360º
1000mm
39 1/2"
3/4" Diameter
Conduit
360º
360º
360º
110º
360º
110º
Wall
Control
1765mm
69 1/2"
1969mm
77 1/2"
2032mm
80"
360º
360º
425mm
16 3/4"
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 288 watts (980 BTU/hr)
• Amperage: 3.2 amps
• Intensity: 4,000K=295,000 Lux (27,412fc)
4,500K=259,000 Lux (24,066fc)
• Fixture Weight: 231.5 lbs./105 Kg.
• Moment Load: 983.5 ft. lbs./1,332.6 N•m
CEILING HEIGHT
8'-9" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-07 REV0
8/09
AUR75TV
2716mm
107"
1200mm
47 1/4"
1140mm
45"
945mm
37 1/4"
70mm
2 3/4"
360º
1000mm
39 1/2"
3/4" Diameter
Conduit
360º
360º
360º
110º
CAMERA
READY
360º
110º
Wall
Control
1765mm
69 1/2"
1969mm
77 1/2"
1" Diameter
Conduit
Camera
Control
Connector
2032mm
80"
360º
360º
425mm
16 3/4"
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 288 watts (980 BTU/hr)
• Amperage: 3.2 amps
• Intensity: 4,000K=295,000 Lux (27,412fc)
4,500K=259,000 Lux (24,066fc)
• Fixture Weight: 239.6 lbs./108.7 Kg.
• Moment Load: 1,016.2 ft. lbs./1,377 N•m
CEILING HEIGHT
8'-9" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-08 REV0
8/09
AUR7TV5
2716mm
107"
1200mm
47 1/4"
1140mm
45"
70mm
2 3/4"
945mm
37 1/4"
360º
1000mm
39 1/2"
3/4" Diameter
Conduit
360º
CAMERA
READY
360º
360º
110º
360º
110º
Wall
Control
1765mm
69 1/2"
1969mm
77 1/2"
1" Diameter
Conduit
Camera
Control
Connector
2032mm
80"
360º
360º
425mm
16 3/4"
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 288 watts (980 BTU/hr)
• Amperage: 3.2 amps
• Intensity: 4,000K=295,000 Lux (27,412fc)
4,500K=259,000 Lux (24,066fc)
• Fixture Weight: 239.6 lbs./108.7 Kg.
• Moment Load: 1,023.2 ft. lbs./1,386.5 N•m
CEILING HEIGHT
8'-9" min.
10'-5" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-09 REV0
8/09
AUR555
2657mm
105"
1335mm
52 1/2"
945mm
37 1/4"
1140mm
45"
70mm
2 3/4"
3/4" Diameter
Conduit
1000mm
39 1/2"
360º
360º
360º
360º
110º
360º
360º
1765mm
69 1/2"
Wall
Control
2032mm
80"
360º
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 432 watts (1,469 BTU/hr)
• Amperage: 4.8 amps
• Intensity: 4,000K=405,000 Lux (37,626fc)
4,500K=360,000 Lux (33,444fc)
• Fixture Weight: 312.4 lbs./141.7 Kg.
• Moment Load: 1,269.5 ft. lbs./1,720.2 N•m
CEILING HEIGHT
9'-2" min.
10'-4" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-10 REV0
8/09
AUR555TV
2657mm
105"
1335mm
52 1/2"
945mm
37 1/4"
1140mm
45"
1" Diameter
Conduit
70mm
2 3/4"
3/4" Diameter
Conduit
1000mm
39 1/2"
360º
360º
360º
360º
360º
110º
CAMERA
READY
360º
1765mm
69 1/2"
Wall
Control
Camera
Control
Connector
2032mm
80"
360º
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 432 watts (1,469 BTU/hr)
• Amperage: 4.8 amps
• Intensity: 4,000K=405,000 Lux (37,626fc)
4,500K=360,000 Lux (33,444fc)
• Fixture Weight: 317.5 lbs./144 Kg.
• Moment Load: 1,269.5 ft. lbs./1,720.2 N•m
CEILING HEIGHT
9'-2" min.
10'-4" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-11 REV0
8/09
AUR575
2657mm
105"
1335mm
52 1/2"
945mm
37 1/4"
1140mm
45"
70mm
2 3/4"
1000mm
39 1/2"
1200mm
47 1/4"
360º
360º
360º
110º
360º
3/4" Diameter
Conduit
110º
360º
1765mm
69 1/2"
360º
1969mm
77 1/2"
Wall
Control
360º
360º
2032mm
80"
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 432 watts (1,469 BTU/hr)
• Amperage: 4.8 amps
• Intensity: 4,000K=430,000 Lux (39,954fc)
4,500K=379,000 Lux (35,214fc)
• Fixture Weight: 329.4 lbs./149.4 Kg.
• Moment Load: 1,445.4 ft. lbs./1,958.5 N•m
CEILING HEIGHT
9'-2" min.
10'-4" max.
425mm
16 3/4"
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-12 REV0
8/09
AUR57TV5
2657mm
105"
1335mm
52 1/2"
945mm
37 1/4"
1140mm
45"
1" Diameter
Conduit
70mm
2 3/4"
1000mm
39 1/2"
1200mm
47 1/4"
360º
360º
360º
110º
360º
3/4" Diameter
Conduit
110º
360º
1765mm
69 1/2"
360º
CAMERA
READY
1969mm
77 1/2"
Wall
Control
Camera
Control
Connector
360º
360º
2032mm
80"
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 432 watts (1,469 BTU/hr)
• Amperage: 4.8 amps
• Intensity: 4,000K=430,000 Lux (39,954fc)
4,500K=379,000 Lux (35,214fc)
• Fixture Weight: 334.9 lbs./151.9 Kg.
• Moment Load: 1,485.2 ft. lbs./2,012.4 N•m
425mm
16 3/4"
CEILING HEIGHT
9'-2" min.
10'-4" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-13 REV0
8/09
AUR575TV
2657mm
105"
1335mm
52 1/2"
945mm
37 1/4"
1140mm
45"
1" Diameter
Conduit
70mm
2 3/4"
1000mm
39 1/2"
1200mm
47 1/4"
360º
360º
360º
110º
360º
3/4" Diameter
Conduit
110º
360º CAMERA
READY
1765mm
69 1/2"
360º
1969mm
77 1/2"
Wall
Control
Camera
Control
Connector
360º
360º
2032mm
80"
610mm
24"
TECHNICAL DATA
• Electrical: 100-240 VAC, 50/60 Hz, 1 phase
• Power Consumption: 432 watts (1,469 BTU/hr)
• Amperage: 4.8 amps
• Intensity: 4,000K=430,000 Lux (39,954fc)
4,500K=379,000 Lux (35,214fc)
• Fixture Weight: 334.9 lbs./151.9 Kg.
• Moment Load: 1,445.4 ft. lbs./1,958.5 N•m
425mm
16 3/4"
CEILING HEIGHT
9'-2" min.
10'-4" max.
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI
1.800.SKYTRON • 616.656.2900 • FAX 616.656.2906
TEC-B-0012-14 REV0
8/09
AURORA II WALL CONTROL
FINISHED WALL
NOTE:
FRONT EDGE OF "BACK-BOX"
MUST BE FLUSH WITH FINISHED
WALL SURFACE
120VAC
30VDC
3/4" CONDUIT
CONTROL BOX WITH RECESS
MOUNT FLANGE INSTALLED
CONDUIT
CONNECTOR
"BACK-BOX" DIMENSIONS
SINGLE / DUAL
8-1/2"W x 11-5/8"H x 4"D
TRIPLE
10-1/2"W x 14-1/2"H x 4"D
RECESS
DEPTH
A
(2) GROMMETS
B
FRONT
PANEL
ANCHOR (4)
C
RECESS MOUNT
FLANGE
V
, 250
2A
0
0
ANCHOR (4)
1
1
2
3
4
ing
t Act
Fas
5
SCREW (4)
V
, 250
2A
2
IN
MA
3
4
ing
t Act
Fas
5
E
UR
AT
ER
MP R
TO
R TECA
LO DI
CO IN
OPTIONAL BACK-BOX INSTALLATION
ED
INAT
UM
N ILL
HE
KW
00
4,5
WER
PO
UR
0 HO R
,00 TO
20 DICA
IN
RECESS
MOUNT FLANGE
SCREW (4)
RECESSED MOUNT
DIMENSIONS
A
JUNCTION
BOX
D
FRONT
PANEL
SINGLE
OR
DUAL
TRIPLE
8"
10"
B
4"
4"
C
10"
13-1/2"
D
6-7/8"
8-5/8"
E
7-5/8"
11"
F
5-7/8"
6-3/8"
RECESS MOUNT FLANGE
E
11-3/4"W x 14"H 13-3/4"W x 17-1/4"H
3"
V
, 250
2A
0
ANCHOR (4)
0
1
2
3
4
F
ing
t Act
5
Fas
3"
V
, 250
2A
1
2
IN
3
4
ing
5
t Act
Fas
1"
E
UR
AT
ER
MP R
TO
R TECA
LO INDI
CO
ED
INAT
WER
PO
UM
C
N ILL
MA
HE
KW
00
4,5
UR
0 HO R
,00 TO
20 DICA
IN
SCREW (4)
SURFACE MOUNT
TEC-B-0012-15 REV0
8/09
A
SIDE VIEW
BACK VIEW
STANDARD MOUNTING STRUCTURE GUIDELINE
TOP VIEW
STIFFENER PLATE DETAIL
6"
12"
STIFFENER PLATE
(provided by others)
Hole for
Access to
Fixture
Junction
Box
9-1/2"
6"
2" x 2" ANGLE SWAY BRACING
90° APART - 4 EACH
(see note 2)
9-1/2"
12"
Structural Ceiling
45°
45°
(4) 2" x 2" Angle Sway
Bracing
(see note 2)
12" x 12" x 3/8"
Stiffener
Plate
3/4" Metal
Flex Conduit
(see note 3)
Fixture
Junction
Box
Finished Ceiling
1-3/8"
3/4"
All Thread Rod
6" Max/ 4” Min.
(see note 1)
2-1/4"
± 1/4"
15" dia.
Mounting Plate (SKYTRON Supplied)
16-1/2" dia.
Access Hole
18" dia.
Ceiling Cover (SKYTRON Supplied)
NOTES
1. 3/4" support rods located for total support of light, all labor and materials for fabrication supplied by General
Contractor. 3/4" nuts and washers for support of SKYTRON mounting plate supplied by contractor (8 ea. required).
2. The mounting structure must be attached to structural ceiling and BRACED TO ALLOW NO TWISTING OR
LATERAL MOTION and shall be designed not to provide a degree of rotation greater than two-tenths of a degree
at the mounting plate.
3. 3/4" metal conduit and minimum 12AWG wire size (3 wires per lighthead) required between fixture and SKYTRON
supplied wall control. All metal conduit, wiring, and other electrical materials as well as installation labor for such
materials associated with the installation of the SKYTRON surgical light to be provided by Electrical Contractor. All
installations of SKYTRON surgical lights should be under the direct supervision of a SKYTRON representative. All
wiring to be in accordance with local codes and by a certified electrician.
4. Optional Camera Ready Systems require a 1” diameter conduit from the fixture junction box to the camera control
connector junction box. 65’ camera control cable with faceplate and connector supplied by SKYTRON. Conduit and
2” x 4” junction box supplied by General or Electrical Contractor.
5. CONTRACTOR HAS FINAL RESPONSIBILITY for the strength and stability of the Mounting Structure.
This is a GENERAL GUIDELINE ONLY.
TEC-B-0012-16 REV0
8/09
RECESSED MOUNTING STRUCTURE GUIDELINE
TOP VIEW
STIFFENER PLATE DETAIL
6"
12"
STIFFENER PLATE
(provided by others)
Hole for
Access to
Fixture
Junction
Box
9-1/2"
6"
2" x 2" ANGLE SWAY BRACING
90° APART - 4 EACH
(see note 2)
9-1/2"
12"
Structural Ceiling
45°
45°
(4) 2" x 2" Angle Sway
Bracing
(see note 2)
12" x 12" x 3/8"
Stiffener
Plate
3/4" Metal
Flex Conduit
(see note 3)
Fixture
Junction
Box
3/4"
All Thread Rod
6" Max/ 4” Min.
(see note 1)
Finished Ceiling
3/4"
1/4"
15" dia.
Mounting Plate (SKYTRON Supplied)
16-1/2" dia.
Access Hole
20-1/8" dia.
Ceiling Cover (SKYTRON Supplied)
NOTES
1. 3/4" support rods located for total support of light, all labor and materials for fabrication supplied by General
Contractor. 3/4" nuts and washers for support of SKYTRON mounting plate supplied by contractor (8 ea. required).
2. The mounting structure must be attached to structural ceiling and BRACED TO ALLOW NO TWISTING OR
LATERAL MOTION and shall be designed not to provide a degree of rotation greater than two-tenths of a degree at
the mounting plate.
3. 3/4" metal conduit and minimum 12AWG wire size (3 wires per lighthead) required between fixture and SKYTRON
supplied wall control. All metal conduit, wiring, and other electrical materials as well as installation labor for such
materials associated with the installation of the SKYTRON surgical light to be provided by Electrical Contractor. All
installations of SKYTRON surgical lights should be under the direct supervision of a SKYTRON representative. All
wiring to be in accordance with local codes and by a certified electrician.
4. Optional Camera Ready Systems require a 1” diameter conduit from the fixture junction box to the camera control
connector junction box. 65’ camera control cable with faceplate and connector supplied by SKYTRON. Conduit and
2” x 4” junction box supplied by General or Electrical Contractor.
5. CONTRACTOR HAS FINAL RESPONSIBILITY for the strength and stability of the Mounting Structure.
This is a GENERAL GUIDELINE ONLY.
HEADROOM CLEARANCE INCREASE - 2"
TEC-B-0012-17 REV0
8/09
SEISMIC APPLICATION MOUNTING STRUCTURE GUIDELINE
SS-L
TOP VIEW
STIFFENER PLATE DETAIL
Typical Mounting
Structure Shown
6"
Contact SKYTRON
representative for
Seismic Calculations
if Applicable
12"
STIFFENER PLATE
(provided by others)
Hole for
Access to
Fixture
Junction
Box
9-1/2"
6"
ANGLE SWAY BRACING
90° APART - 4 EACH
(see note 2)
9-1/2"
12"
Structural Ceiling
45°
45°
(4) Angle Sway
Bracing
(see note 2)
12” x 12” x 3/8”
Stiffener
Plate
3/4" Metal
Flex Conduit
(see note 3)
Support Rods
Finished Ceiling
1-3/8"
Fixture
Junction
Box
(see note 1)
As Specified by
Seismic calculations
2-1/4"
± 1/4"
18" dia.
Mounting Plate (SKYTRON Supplied)
B2-010-20
UNIVERSAL ADAPTER
PLATE
B9-210-44
CASTLE TYPE
CEILING COVER
19" dia.
Access Hole
20-1/8" dia.
Ceiling Cover (SKYTRON Supplied)
NOTES
1. Support rods as specified per seismic calculations are to be located for total support of light, all labor and materials
for fabrication supplied by General Contractor. Nuts and washers for support of SKYTRON mounting plate supplied
by contractor (8 ea. required).
2. The mounting structure must be attached to structural ceiling and BRACED TO ALLOW NO TWISTING OR
LATERAL MOTION and shall be designed not to provide a degree of rotation greater than two-tenths of a degree
at the mounting plate.
3. 3/4" metal conduit and minimum 12AWG wire size (3 wires per lighthead) required between fixture and SKYTRON
supplied wall control. All metal conduit, wiring, and other electrical materials as well as installation labor for such
materials associated with the installation of the SKYTRON surgical light to be provided by Electrical Contractor. All
installations of SKYTRON surgical lights should be under the direct supervision of a SKYTRON representative. All
wiring to be in accordance with local codes and by a certified electrician.
4. Optional Camera Ready Systems require a 1” diameter conduit from the fixture junction box to the camera control
connector junction box. 65’ camera control cable with faceplate and connector supplied by SKYTRON. Conduit and
2” x 4” junction box supplied by General or Electrical Contractor.
5. CONTRACTOR HAS FINAL RESPONSIBILITY for the strength and stability of the Mounting Structure.
This is a GENERAL GUIDELINE ONLY.
TEC-B-0012-18 REV0
8/09
5085 Corporate Exchange Blvd. S.E. Grand Rapids, MI 49512
1.800.SKYTRON • 616.656.2906 • 616.656.2906