T7D T7DS - Persee Analytics Inc
Transcription
T7D T7DS - Persee Analytics Inc
U V- V I S S P E C T R O P H O T O M E T E R U S E R T7D T7DS M A N U A L Analytical Instruments for Science Preface Thank you for purchasing our products. To operate this instrument safeley please observe the following precautions: Please read the instruction manual carefully before operating the instrument. Do not operate the instrument before understanding the operating procedures. Pay attention to words which are emphasized in the manual, especially the safety descriptions. This manual is for the use of all versions of the T7D/T7DS UV-VIS Spectrophotometers. Copyright © 2005 Persee Analytics Inc. All rights reserved. I Contents Chapter 1. 1.1 Introduction ............................................................................................................................1 Unpacking.....................................................................................................................................1 1.2 Components description ..............................................................................................................1 1.2.1 Front view ................................................................................................................................2 1.2.2 Rear view .................................................................................................................................3 1.2.3 Side view .................................................................................................................................3 1.2.4 Display .....................................................................................................................................4 1.2.5 Keypad .....................................................................................................................................5 1.2.6 Sample cell holder ...................................................................................................................6 1.3 Specifications ..................................................................................................................................8 Chapter 2. 2.1 Installation ..............................................................................................................................9 site requirement .............................................................................................................................9 2.2 Installation ....................................................................................................................................9 2.2.1 Connect the instrument to the printer.......................................................................................9 2.2.2 Connect the instrument to PC ................................................................................................10 2.3 Turning on the instrument ........................................................................................................ 11 2.4 Performance check .................................................................................................................... 12 2.4.1 Wavelength accuracy and reproducibility ..............................................................................12 2.4.2 Baseline flatness ....................................................................................................................13 Chapter 3. 3.1 Photometric measurement ....................................................................................................14 Gain access to photometric mode ............................................................................................... 14 3.2 Setting the method parameters................................................................................................... 15 3.2.1 Set photometric mode ............................................................................................................16 3.2.2 Set measuring wavelenghth ...................................................................................................16 3.2.3 Set K factor ............................................................................................................................ 18 3.2.4 Dark current correction ..........................................................................................................18 3.2.5 Quit parameters setting .......................................................................................................... 19 3.3 Sample control ............................................................................................................................. 19 3.3.1 Select sample module (changer) ............................................................................................20 3.3.2 Set sample cell number in use (Drive Cell No.) ....................................................................20 3.3.3 No.1 cell blank correction ......................................................................................................21 3.3.4 Sample cell shifting (Moving the cell changer) .....................................................................22 3.3.5 Reset to cell No.1 (Cell reset) ................................................................................................22 3.3.6 Single cell measurement ........................................................................................................22 3.3.7 Quit sample control................................................................................................................22 3.4 Auto zero (100T% correction) .................................................................................................... 22 3.5 Measurement ................................................................................................................................ 22 3.5.1 Single cell measurements.......................................................................................................22 3.5.2 Use multi-cells .......................................................................................................................23 3.6 Looking up the measurement results ......................................................................................... 24 3.7 Delete the measurement data ...................................................................................................... 24 3.8 Printout ......................................................................................................................................... 25 II 3.9 Quit measurement ....................................................................................................................... 25 3.10 Application example ................................................................................................................... 25 Chapter 4. 4.1 Spectrum scanning ............................................................................................................... 27 Gain access to spectrum mode ................................................................................................... 27 4.2 Parameters setting ...................................................................................................................... 28 4.2.1 Set photometric mode ........................................................................................................... 29 4.2.2 Setting the scanning speed .................................................................................................... 30 4.2.3 Setting the sampling interval................................................................................................. 30 4.2.4 Setting the wavelength range ................................................................................................ 31 4.2.5 Setting the ordinate range ..................................................................................................... 32 4.2.6 Light source selection ........................................................................................................... 33 4.2.7 Dark current correction ......................................................................................................... 33 4.2.8 Quit parameters setting ......................................................................................................... 34 4.3 Baseline correction ...................................................................................................................... 34 4.4 Measurement ............................................................................................................................... 34 4.5 Peak picking ................................................................................................................................ 35 4.6 Printout ...................................................................................................................................... 36 4.7 Quit measurement ....................................................................................................................... 36 4.8 Application example ................................................................................................................. 37 Chapter 5. 5.1 Quantitative measurement ................................................................................................... 39 Gain access to quantitation mode .............................................................................................. 39 5.2 Parameters setting ...................................................................................................................... 40 5.2.1 Set measurement method ...................................................................................................... 41 5.2.2 Set measuring wavelength .................................................................................................... 41 5.2.3 Set concentration unit............................................................................................................ 43 5.2.4 Dark current correction ......................................................................................................... 43 5.2.5 Quit parameters setting ......................................................................................................... 44 5.3 Construct the calibration curve ................................................................................................. 44 5.3.1 K-Factor method ................................................................................................................... 44 5.3.2 Standards method .................................................................................................................. 46 5.4 Sample control ............................................................................................................................. 50 5.5 Auto zero ...................................................................................................................................... 50 5.6 Measurement ............................................................................................................................... 50 5.6.1 Use single cell ....................................................................................................................... 50 5.6.2 Use multi-cells ...................................................................................................................... 51 5.7 Looking up the measurement results ........................................................................................ 52 5.8 Delete the measurement data ................................................................................................... 52 5.9 Printout ...................................................................................................................................... 52 5.10 Quit measurement ....................................................................................................................... 52 5.11 Application example ................................................................................................................. 52 Chapter 6. DNA/Protein Analysis .......................................................................................................... 56 6.1 The principle of Protein/DNA analysis ............................................................................................ 56 III 6.1.1 Principle of using absorbance of UV light to determine protein: .................................................56 6.1.2 Principle of using absorbance of UV light to determine DNA: ....................................................56 6.1.3 Protein/DNA determination methods............................................................................................56 Gain access to DNA/Protein analysis .........................................................................................57 6.3 Parameters setting ....................................................................................................................... 59 6.3.1 Selection analysis method ......................................................................................................60 6.3.2 Dark current correction ..........................................................................................................61 6.4 Custom defined method .............................................................................................................. 61 6.4.1 Set measuring wavelength .....................................................................................................61 6.4.2 Set DNA factors .....................................................................................................................62 6.4.3 Set protein factors ..................................................................................................................63 6.4.4 Set background correction at 320nm ..................................................................................... 64 6.4.5 Set calibration factors ............................................................................................................65 6.5 Auto zero....................................................................................................................................... 66 6.6 Measurement ................................................................................................................................ 66 6.7 Look up the measurement results .............................................................................................. 66 6.8 Delete the measurement data .................................................................................................... 67 6.9 Printout ......................................................................................................................................... 67 6.10 Quit measurement ....................................................................................................................... 67 Chapter 7. Utilities .................................................................................................................................. 68 7.1 Gain access to utilities ................................................................................................................. 68 7.2 Set lamp changeover wavelength ................................................................................................ 68 7.3 Set spectral bandwidth ................................................................................................................ 69 7.4 Set tungsten lamp On/Off ........................................................................................................... 69 7.5 Set deuterium lamp On/Off ........................................................................................................ 70 7.6 Set control mode .......................................................................................................................... 71 7.7 Set time ......................................................................................................................................... 72 7.8 Reset system ................................................................................................................................. 72 Chapter 8. Maintenance.......................................................................................................................... 74 8.1 Points for attention: ........................................................................................................................... 74 8.2 Daily care ...................................................................................................................................... 74 8.3 Trouble shooting and maintenance ............................................................................................ 75 Appendix 1 Light source replacement .......................................................................................................... 77 Replace the tungsten lamp ..................................................................................................................... 77 Deuterium lamp replacement .................................................................................................................79 Appendix 2 Cell holder replacement .......................................................................................................... 811 Appendix 3 Fuse replacement .................................................................................................................... 833 IV Instruction manual Chapter 1. Introduction 1.1 Unpacking When unpacking, check that packing list to confirm that all the items listed are included. If any items are missing or damaged then please contact your supplier immediately. Refer to the following table 1-1 for the standard instrument set. But refer to the packing list for your configuration. Table 1-1 Standard set No. Name Amount 1 Main Unit 1 2 RS232 Communication Cable 1 3 Socket Screw Wrench 1 4 Instruction Manual 1 5 Quartz Cells (10mm)( pair) 1 6 Fuses for Power(2A for 220V) 2 7 Fuses for D2 Lamp(0.5A for 220V) 2 8 Power Cord 1 9 Black Block 1 10 Dust Cover 1 11 Packing List 1 12 Quick Installation Guide 1 1.2 Components description When unpacking please ensure that the instrument is present as shown in Fig. 1.1. 1 Instruction manual Fig. 1.1 Instrument overview 1.2.1 Front view The front view is shown as Fig. 1.2. Fig. 1.2 Front view Instrument cover Front baffle of sample compartment — standard front baffle for sample compartment. It can be replaced when using accessories requiring tubing access to the sample compartment. Chassis — The main chassis is made from steel and this can greatly reduce any vibrations present in the vicinity and also contributes to the excellent stability of the instrument. Instrument label Keypad and display — 320h240 dots high resolution LCD and 27 keystrokes. Sample compartment cover — Open the cover to gain access to the sample compartment. 2 Instruction manual 1.2.2 Rear view The rear view is shown as Fig.1.3. Fig. 1.3 Rear view Light source housing coverüOpen the cover to gain access to the light source housing, here you can replace the deuterium lamp and tungsten lamp. Feet packingüShock proof rubber washer is used. Printer interfaceüConnection to HP parallel inkjet printers. RS-232 interfaceüCommunication port for a PC. Wiring terminalüConnection to a ground wire for countires requiring this facility. Outletü Connection to power supply: AC 220V, 50Hz. FuseüCan be replaced (refer to “Appendix 3 fuse replacement”). Power switch Power labelüdescribes the power requirements. Instrument nameplate and Serial Number Fan 1.2.3 Side view The instrument side view is shown in Fig.1.4. There are no other interfaces or labels except the ventilation grills. 3 Instruction manual A: Left side view B: Right side view Fig. 1.4 Side view 1.2.4 Display The LCD display on the T70 is a high resolution LCD with 320h240 dots. This displays the measured data, spectra, method parameters, mode and prompts for operating the instrument. The display and keypad are shown in Fig.1.5. Fig. 1.5 Display and keypad There are two types of menu. One is the main menu (Fig.1.6A) which appears after initialisation and is used to select the measurement mode. The other is the operation menu (Fig.1.6B) which displays the current working mode, measured data, spectra and user prompts information. 4 Instruction manual A: Main menu B: Spectrum measurement menu Fig. 1.6 Display Menu At the top of the screen as shown in Fig 1.6B the current working mode of operation is displayed. The column on the right of the screen displays the function keys (F1~F4). The function keys may vary from page to page. The center of the screen is the operating area where the measured results are displayed. At the bottom of the screen there is the information area which informs the user about the current status of the instrument it also allows you to enter parameters. There are four measurement modes: photometric measurement, spectrum measurement, quantitative determination and DNA/Protein analysis. Please refer to the following chapters for detailed operating procedures. 1.2.5 Keypad Altogether there are 27 soft-touch keys on the keyboard of the T70 which are divided into 4 groups. The functions of the keys are described as follows: Fig. 1.7 Keypad 5 Instruction manual Function keys: The function keys are used to setup the instrument parameters and are used to operate the instrument. When one of the five function keys (F1~F5) is pressed the system executes the corresponding operation such as setting the method parameters, deleting, setting the sample cell positions, printing and dark current correction etc.. śDisplay Contrast adjustment point Here you can adjust the contrast of the display by using a 3˄3mm diameter˅screwdriver. ŜNumeric keys: These are used to input the numeric values when setting up the method parameters. ŝEdit keys: , ˅üüUsed to shift the cursor up and down while browsing data, and increase and decrease values when setting the time. ƽ Cursor keys˄ , ˅üüUsed to switch the displayed data when measuring DNA , and for setting the time and also for looking up the data when peak picking. ƽ Cursor keys˄ ƽ Cancel key˄ ĀCLEā˅ üüUsed to delete entries or to cancel a certain operation. ƽ Enter key˄ ĀENTERā˅ üüUsed to confirm and execute the current function, or to exit the method parameters setting and to quit. ƽ Return key üüUsed to quit the current function and return to the previous page. ŞOperation keys: ƽ Auto zero key : Used to set the ABS to zero (and transmittance to 100%). In the will set the ABS to zero only at the photometric mode and the quantitative mode, performs a baseline scan across the set current wavelength. In spectrum mode, wavelength range. ƽ Wavelength set key ƽ Start/Stop key : Used to set the current wavelength. : Used to start or stop a measurement. şIndicator lamp: Indicates the power supply is on. 1.2.6 Sample cell holder The sample cell holder is configured according to the user’s requirements. The general configurations are fixed cell holder, 8-cell changer, and 5-cell changer. We can also supply a range of different types of sample cell holders, please refer to “chapter eight Accessories” for detailed information. The following figure 1.8 shows the three types of sample cell holders, which are used as the standard configuration. 6 Instruction manual A: Fixed cell holder B: 8-cell changer C: 5-cell changer Fig. 1.8 frequently used cell holders Fixed cell holder: As shown in 1.8A, it is the most frequently used cell holder, having only one standard cell position. It can only measure one sample at a time. 8-cell changer: As shown in 1.8B, it has eight standard cell positions and can measure up to a maximum of 8 samples automatically. 5-cell changer: As shown in 1.8C, it has five custom defined cell positions and can measure up to a maximum of 5 samples simultaneously. You can put the samples in standard 10mm path length cells or use long-pathlength cells. Please refer to the 5-cell changer instruction manual for detailed information. It also completes measurement automatically. A: 8-cell changer B : 5-cell changer Fig. 1.9 Multi-cell changer You can use different types of cell holders according to your requirements. Please refer to “Appendix ĊThe replacement of a sample cell holder”for more information on the replacement of frequently used cell holders. 7 Instruction manual 1.3 Specifications Split beam optics Model T7 T7S Wavelength range 190nm~1100nm Spectral bandwidth 2nm 5, 2, 1, 0.5nm (4 steps) Wavelength display 0.1nm Wavelength accuracy 0.3nm (with automatic wavelength correction) Wavelength reproducibility 0.2nm Photometric system Double beam ratio monitoring system Photometric accuracy 0.002Abs (0~0.5Abs) 0.004Abs (0.5~1.0Abs) 0.3%T (0~100%T) Photometric reproducibility 0.001Abs (0~0.5Abs) 0.002Abs (0.5~1.0Abs) 0.15%T (0~100%T) Stray light 0.3%T (220nm) Baseline flatness Stability 0.002Abs (190-1100nm) 0.001A/30min (2 hours warming up, 2nm bandwidth, at 500nm) Double beam optics Model T7D DOUBLE BEAM Wavelength range T7DS DOUBLE BEAM 190nm~1100nm Spectral bandwidth 2nm 5, 2, 1, 0.5nm (4 steps) Wavelength display 0.1nm Wavelength accuracy 0.3nm (with automatic wavelength correction) Wavelength reproducibility 0.2nm Photometric system Double beam optical system Photometric accuracy 0.002Abs (0~0.5Abs) 0.004Abs (0.5~1.0Abs) 0.3%T (0~100%T) Photometric reproducibility 0.001Abs (0~0.5Abs) 0.002Abs (0.5~1.0Abs) 0.15%T (0~100%T) Stray light 0.2%T (220nm) Photometric Noise 0.001Abs (at 500nm, 2nm bandwidth 0Abs) Baseline flatness 0.0015Abs (190-1100) Stability 0.0008A/h (2 hours warming up, 2nm bandwidth, at 500nm) 8 Instruction manual Chapter 2. Installation 2.1 Site requirement To ensure a better working quality and to prolong the service life of the instrument, install the instrument in a place that satisfies the following requirements: a. Ambient temperature: 15ć~35ć b. Relative humidity: less than 80% c. The instrument should be placed on a stable workbench, avoiding strong vibrations. d. Avoid a place with strong magnetic or electric fields and corrosive gases. e. Power requirements: AC110V/220V f 10%, 50Hz/60Hz f 1Hz, single phase. A regulated power supply with at least 500VA capacity is recommended. There should be a ground line in the laboratory to ensure the instrument is well grounded. 2.2 Installation 2.2.1 Connect the instrument to the printer All the T70 UV-VIS Spectrophotometers can be connected to HP parallel printers. The instruments connected with PC can use any printer as long as your PC can recognize it. The following is the installation scheme of the connection between the instrument and the printer. Put the printer cable with 25-pin plug into the parallel interface of the instrument, and put the other end with a 36-pin plug into the printer interface. Connect all the cables firmly by locking the screw (if there is any) on the plugs. Plug in the power cords of the spectrophotometer and the printer to the mains power. Note: 1. Make sure the power supply is AC 110V/220V, 50HZ/60Hz before plugging the power cords of the instrument and the printer. For the voltage, 10% fluctuation is allowed. The instrument can’t work normally if the power supply is out of range. High voltage may cause damage to the instrument. 2. Do not insert or pull out the printer cable when the instrument is on, otherwise it may destroy the instrument and the printer. 9 Instruction manual Fig. 2.1 Installation scheme 2.2.2 Connect the instrument to PC Some of the models in the T70 UV-VIS Spectrophotometer range can be connected to a PC. The connection scheme is shown in the following figure 2.2. Plug one end of the communication cable with serial interface into the instrument RS232 interface (serial port). Plug the other end into the PC RS232 interface (serial port). The two ends of the communication cable are the same. Connect all the cables firmly by locking the screw on the plugs. Plug in the power cords of spectrophotometer and the PC to the mains power. Note: 1. Make sure the power supply is AC 110V/220V, 50HZ/60Hz before plugging the power cords of the instrument and the printer. For the voltage, 10% fluctuation is allowed. The instrument can’t work normally if the power supply is out of range. High voltage may cause damage to the instrument. 2. Do not insert or pull out the communication cable when the instrument is on, otherwise it may destroy the instrument and PC. Fig. 2.2 Connection scheme 10 Instruction manual When connecting to a PC, you will need to install the UVWin software. Please refer to the software instruction manual for UVWin software operation. 2.3 Turning on the instrument After confirming that there is nothing in the sample compartment, turn on the power switch of the main unit and then the printer (if there is one). The instruments information will appear. Then the initialisation page will be displayed on the LCD. The instrument starts self-checking and initialising. This process will take about 3 minutes. After initialsation the system displays the instrument main menu, as shown in Fig.1.6A. Note: If the initialisation fails, please first check whether there is a sample in the sample compartment. Then restart the instrument. If the instrument still can’t initialise normally, please contact your local dealer. A: Instrument information B: Initialization menu Fig. 2.3 Initialisation After the inialisation has completed you can operate the instrument. Ideally you should allow the lamps to warm up for 15 to 30 minutes prior to use this will ensure that your data is accurate and stable. Then you can start measurements. Once you have finished using the instrument first turn off the instrument, then turn off the printer power, and then turn off other appliances such as the PC. When connecting to a PC, first switch on the power of the instrument, then the PC power. Run the software after the computer finishes starting. If the instrument can work both in Local Control mode and PC mode, you should select PC mode in the menu of utilities, please refer toĀ7.6 Set control modeāthen run software. 11 Instruction manual 2.4 Performance check After installation, you need to check the performance and main specifications of the instrument. And you should check the instrument periodically to ensure its performance meets the standard requirements. The performance checks, as described in the following paragraphs, should be performed after 30-minutes warm up time. 2.4.1 Wavelength accuracy and reproducibility The wavelength accuracy should be f0.3nmˈand the wavelength reproducibility should be 0.2nm (Please refer toĀ1.4 Specificationsā). You can use two characteristic emission lines radiated by the deuterium lamp to check the wavelengths. The detailed methods are shown below (please refer to the software instruction manual when connected to a PC). Make sure the spectral bandwidth isĀ2.0nmā. After initialisation, press to set the slit width toĀ2.0nm”. utilities, and press to access the Note: It is not necessary to set the slit width if the instrument is a fixed bandwidth instrument. Only instruments with a variable slit should be set. Press measurement. to return to the main menu ˈ press to select photometric to access the method parameters menu. Then press the numeric keys to set the Press sampling interval to 0.2nm (set it to 0.1nm when connecting with a PC), set the scanning speed to middle, wavelength range to 660-480nm, y axis to 0-100, the photometric mode to Es, and select D2 lamp as the light source (refer to “3.2 parameters setting). Press Press to return to the spectrum scanning page. , input 6 as the gain value, and then press to start scanning. After scanning press to input the threshold value (1̚100) then perform a peak pick (if the picked peak is 0.000nm press the RETURN key. Then press F2 to input a smaller value for the threshold and then perform the peak pick again). Press key to printout. Repeat 1 three times and perform the peak pick each time. The two standard emission lines of the deuterium lamp are 656.1 nm and 486.0 nmDŽ ǏĆ ǐ ɋďɌ The difference between the average value and the standard value of the three times’ measured peak is equivalent to the wavelength accuracy. The difference between the maximum value and the minimum value of three times’ measured peak is the wavelength reproducibility. 12 Instruction manual 2.4.2 Baseline flatness Do not place any opaque object into the sample cell holder. Use an air blank to measure the Absorbance over the full wavelength range. The method is shown as follows (refer to the software instruction manual when connected to a PC). ķ Make sure the spectral bandwidth isĀ2.0nmā(refer to the first step 2.4.1). Press to access the menu for the photometric mode. Press to access the menu for setting the method parameters. Set the photometric mode to Abs, the sampling interval to 1nm, scanning speed to middle, wavelength range to 1100̚190nm, and y axis to -0.01̚0.01 (refer to “4.2 parameters setting”). Press Press Press key to return to the menu for photometric measurement, then press to perform a baseline correction. to start scanning. to printout. Read the Absorbance. The difference between the starting point value and the maximum deviation value over the full range is the baseline flatness (a small deviation at the wavelength lamp change is allowed). 13 Instruction manual Chapter 3. Photometric measurement In Photometric mode single wavelength measurements of absorbance and transmittance can be made. K-Factor calculation and printout are possible. 3.1 Accesing the Photometric mode From the instrument main menu, please press to access the Photometric mode. Press to return to the instrument main menu, as shown in Fig. 3.1: A: Main menu B: Photometric mode Fig. 3.1 Accessing the photometric mode When you exit the photometric mode you will be prompted to “Delete all the data? (If Yes: press to return to the instruments main menu and the measured enter. If No: press CLE.)”. Press to return to the photometric mode and you will be prompted with data will be lost. Press “press START to measure” then you can continue with photometric measurements. Note: Please print out the measured data before returning to the instrument main menu, otherwise the data will be lost. The following paragraphs are an introduction to the main menu for photometric mode: 14 Instruction manual Fig. 3.2 Photometric mode Ā PHOTO ā means photometric measurement ˗Ā 990 nm ā is the current working wavelength˗ Ā0.000 Absāis the live Absorbance Value. Working area: displays the detailed measurement data in a table format. It can display up to a maximum of 9 rows of results on each screen, if this exceeds, it will display results on the next screen. Information area: the current information is prompting the user to press begin measurement. to Time display area: time in the above picture is eight thirty four on April 15. Display area for function keys: shows the function keys functions are shown as follows: ƽ üü access to the method parameters menu; ƽ üü delete the measurement data; ƽ üü set the sample cell holder; ƽ üü print the measured data. üü . The detailed In addition, pressing from this menu allows you to set the wavelength and pressing performs an auto zero at the current set wavelength. 3.2 Setting the Method Parameters From the menu for photometric measurement, press to gain access to the method to return to the photometric mode menu, as shown in parameters setting page, and press Fig.3.3: 15 Instruction manual A: Photometric mode menu B: Parameters setting page Fig. 3.3 Gain access to the parameters setting page On this page, users can set the photometric mode, wavelength and the value of factor K by pressing the corresponding numeric keys. After you have finished setting up the parameters, press the key to return to the previous page. 3.2.1 Set photometric mode Press key to set the photometric mode. There are two modes available: Abs and %T, as is pressed. shown in Fig. 3.4. They can be selected alternately each time when key A: Photometric mode: Abs B: Photometric mode: T% Fig. 3.4 Set work mode 3.2.2 Set measuring wavelength Press key to set the measuring wavelength in photometric mode. A message prompts the user to input the wavelength, as shown in Fig.3.5: 16 Instruction manual A: Parameters setting page in photometric mode B: Wavelength setting page Fig. 3.5 Setting the wavelength in photometric mode (1) At the bottom of this page, enter the desired wavelength value by pressing the numeric keys ̣ and decimal point . The key is used to clear the previous entry. Press key to confirm the wavelength entry and quit. If nothing is to be entered press the key to quit this page and return to the photometric menu. Note: The input wavelength value should be an integer value or with one digit precision. The range is from 190 to 1100. If the input value exceeds 6 digits (including decimal or exceeds the range, the system will delete what has been entered automatically point) and you should start a new entry again. You also can press from the photometric menu (as shown in Fig. 3.2) to set the working wavelength in the photometric mode, as shown in Fig. 3.6: A: Photometric mode menu B: Wavelength setting menu Fig. 3.6 Wavelength setting page in photometric mode (2) At the bottom of this page, enter the desired wavelength value by pressing the numeric keys ̣ and decimal point . The key is used to clear the previous entry. Press key to confirm the numeric entry and quit. If nothing is to be entered, press the key to quit this page and return to the photometric menu. 17 Instruction manual Note: The input wavelength value should be an integer value or one with digit precision. The range is from 190 to 1100. If the input value exceeds 6 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you will need to start a new entry again. 3.2.3 Set K factor Press key to set K factor in the photometric mode, as shown in Fig.3.7: A: Parameters setting B: K factor setting page Fig. 3.7 K factor setting page ̣ , At the bottom of this page, enter K factor value by pressing the numeric keys and . The key is used to clear the previous entry. Press key decimal point to confirm the numeric entry and return to the parameters setting page in photometric mode. If key to quit this page and return to the method parameters nothing is to be entered press the setting page in photometric mode. Note: The input K factor should be an integer value or a decimal fraction. The range is from 190 to 1100. If the input value exceeds 6 digits (including decimal point and negative sign) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 3.2.4 Dark current correction Dark current correction is performed to ensure the accuracy of the measured results. You should perform a dark current correction when environmental and instrument conditions change, such as temperature variations and relocation of the instrument. Also when you measure highly absorbing samples, it is advisable to perform a dark current correction before you start taking measurements. The procedure is shown as follows: 18 Instruction manual A: Parameters setting page in photometric mode B: Dark current correction Fig. 3.8 Dark current correction Press to perform dark current correction from the parameters setting page. It will display “system busy”. Then about 10 seconds later, it will prompt the user with the message “Please select Item”, this means that the dark current correction has finished, as shown in Fig.3.8. 3.2.5 Quit parameters setting Press key to quit the method parameters setting and return to the main menu for photometric measurement. 3.3 Sample control For sample holders you can select the preferred sample changer (fixed cell holder, 5-cell changer and 8-cell changer are available). If you select 8-cell changer or 5-cell changer, you can set cell number, move cell and cell reset etc. to gain access to the sample control menu in the case of photometric measurement, Press as shown in Fig. 3.9: A: Photometric measurement menu B: Sample control menu e.g. 3.9 Set sample control 19 Instruction manual In the case of sample control, press the corresponding numeric keys to set cell holders. Then, to return to photometric mode menu. press 3.3.1 Select sample module (changer) Press key to select sample module. There are three types available, fixed cell-holder, key. 5-cell holder and 8-cell holder. You can alternate between them each time by pressing the Fig. 3.10 Select sample module 3.3.2 Set sample cell number in use (Drive Cell No.) This parameter is only available for the 5-cell changer and the 8-cell changer. For sample key to gain access to set the cell number in use, as shown in Fig.3-11: module, press the A: Sample cell setting page B: set sample cell number 20 Instruction manual Fig. 3.11 Set cell number in use ü At the bottom of this page, enter the sample cell number by pressing the numeric keys . Press key to clear off the previous entry. Press key to confirm the numeric entry to return to the set up parameters page for sample and quit. If nothing is to be entered, press control. A 1-digit figure entry is allowed. The range is from 1 to 8. If you enter a number which exceeds the range, the system will delete the entry. You should enter a number again. Note: Only one integer is accepted. When the sample module is 8-cell changer, the range is 1-8; when the sample module is 5-cell changer, the range is 1-5. If the entered figure exceeds the range, the system will delete what has been entered. You should start a new entry. During the measurement procedure each sample cell position is measured one by one. The serial number sequence of sample cells is expressed by n-1, n-2, … 3.3.3 No.1 cell blank correction You can set position 1 on the Cell Changer to perform a blank correction to No.l cell (only key, it available for 5-cell changer and 8-cell changer). Each time the user presses the alternates between “yes” and “no” as shown in Fig.3.12. Fig. 3.12 Set No.1 cell blank correction This parameter is used to set No. 1 cell to perform a blank correction. 21 Instruction manual 3.3.4 Sample cell shifting (Moving the cell changer) This option is only available for the 5-cell changer and 8-cell changer. Press the key to move the cell holder to the sasmple position. The numeric value displayed relates to the current cell number. Each time you press the key, it will move to the next cell position. The cell changer resets to No.1 cell once you have reached the maximum cell number. The maximum cell number for the 5-cell changer is five, and the maximum cell number for the 8-cell changer is eight. 3.3.5 Reset to cell No.1 (Cell reset) This option is only available for the 5-cell changer and 8-cell changer. Press the key to reset the cell changer to sample cell No.1. 3.3.6 Single cell measurement If you set the “Cell Number” to 1, you can choose any one of the cell positions on the multi-cell changer as the measuring cell by moving the cell changer to the preferred position as described in 3.3.4. Then, the measurement will be made only on the chosen cell. 3.3.7 Quit sample control To quit the sample control press the photometric mode as shown in Fig. 3.3. key to return to the main menu for 3.4 Auto zero (100T% correction) In photometric mode, press the key to set the Absorbance to zero at the current wavelength. Remember to insert your blank before you perform the correction. 3.5 Measurement 3.5.1 Single cell Measurements Single cell measurements include the fixed cell measurement and use of one cell position on a multi-cell changer to measure (refer to 3.3.6 single cell measurement). First set up the measurement parameters and sample cell parameters then perform an key to measure auto zero. Place your sample into the sample cell, then press the key, it will re-measure the sample. The its Abs or T%. Each time you press the measurement results will be displayed as shown in Fig. 3.13 (take Abs for example): 22 Instruction manual Fig. 3.13 Single cell measurement results ĀNo.āin the table refers to serial No. of the samples set by the user.Ā2 -1āin the table refers to the second measurement result of No.1 cell (if use multi-cell changer, it refers to current sample cell in use); “Abs” refers to absorbance value;ĀK*Absārefers to the value of measurement multiplied by K-Factor. The K factor in the above table is 10. 3.5.2 Use of multi-cells First set up the measurement parameters and sample cell parameters. The measurement procedures will vary with the settings of No. 1 cell blank. When Reagent Blank Correction in Cell No. 1 is set to “NO” When Reagent Blank Correction in Cell No. 1 is set to “NO”, you need to perform an auto zero on cell No.1. After auto zero, put the samples into sample cells, then press key to start measurement. And then the measured values will be calculated, saved and displayed on the LCD one by one. When Reagent Blank Correction in Cell No. 1 (c3.5.3) is set to “YES” When Reagent Blank Correction in Cell No. 1 is set to “YES”, you do not need to perform an auto zero. The system will first measure the set N samples. Then the measured values of samples No.2~No.N will be deducted (corrected) by that of the blank in sample cell No. 1 following the formula below. The corrected results will be recorded, calculated and displayed on LCD one by one. ƽ Absorbanceφ $EVRUEDQFHRI&HOO1RQ WKHPHDVXUHPHQWYDOXHRI&HOO1RQ 㸫$EVRI&HOO1R ĂĂĂĂ˄3-1˅ ƽ Transmittance: 7KHWUDQVPLWWDQFHRIVDPSOH1RQ PHDVXUHGYDOXHRIVDPSOH1RQ PHDVXUHGYDOXHRIVDPSOH1REODQN ĂĂĂĂ˄3-2˅ The measurement results of multi-cells are shown as Fig.3.14. ĀNo.āin the table refers to sample serial number set by the user.Ā2 -3āin the table refers to the second measurement 23 Instruction manual result of No.3 cell (if you use a multi-cell changer, it refers to current sample cell in use); “Abs” refers to absorbance value;ĀK*Absārefers to the value of measurement multiplied by K-Factor. The K factor in the above table is 10. Fig. 3.14 Multi-cell measurement results 3.6 Looking up the measurement results After measurement, you can look at the measurement results by pressing the cursor keys and . 3.7 Delete the measurement data You can press the key to delete the measured data. You will be prompted with the to confirm and information “Delete all the data? (YES: ENTER, NO: CE).” Press key to cancel the operation, and you will be prompted to delete all the data. Press “Press START to measure”, and then you can continue to take more readings. Fig. 3.15 Delete measurement data 24 Instruction manual Please remember to print the measured data before deleting your results. 3.8 Printout After measurement, you can press the key to printout your results. You will be prompted with the information “Printer is ready? (YES: ENTER, NO: CE)”, as shown in Fig.3.16: Fig. 3.16 Data printout key to print the measured data. If your printer If your printer connects well, press to cancel the operation. If your printer is not ready and you have is not ready, press key, the system will prompt you with “System busy…” and you can press pressed to exit this selection. Please refer to “Chapter 2 Installation” for printer installation. If your printer does not work normally please refer to “Chapter 9 Daily Care and Maintenance”. 3.9 Quit measurement Press to quit measurement. It will prompt the user with the information to return to the main menu of “Delete all the data? (YES: Enter; NO: CE)”. Press instrument. 3.10 Application example Vitamin B1, also known as thiamin, helps fuel your body by converting blood sugar into energy. It helps to keep your mucous membranes healthy and it is essential for the nervous system, cardiovascular and muscular function. The Spectrophotometric method tests the absorbance value of Vitamin B1 at the wavelength of 400nm. If the absorbance value is not 25 Instruction manual greater than 0.020, it means the tested product has passed and qualified. The procedure for using the T70 UV-VIS Spectrophotometer to test Vitamin B1 is as follows: Dissolve Vitamin B1 1g in 10ml-deionized water; then, put Vitamin B1 and de-ionized water into cells separately. Turn on T70 UV-VIS Spectrophotometer. After 60 minutes’ preheat time, press to gain access to photometric mode menu. Refer to “3.2 Parameters setting”, set photometric mode to Abs, the measuring wavelength to 400nm, and the factor to the default 1.000. For the first measurement, put the black block (an accessory which is supplied with to perform a dark current the instrument) into the sample cell, and then press to return to correction (refer to 3.2.4 dark current correction). Then, press the main menu of photometric measurement. to gain access to sample control. Refer to “3.3 Sample control”. After Press to return to the main menu of photometric measurement. setup, press Measurement: If the “Cell No.” is set to 1, first put the reference cell (de-ionized water) into the key to sample cell holder, close the sample compartment, and then press auto zero. The absorbance value displayed on the photometric mode menu should be 0.000 Abs. After correction, put the sample cell (Vitamin B1) instead of reference cell into the sample cell holder. Close the sample compartment. Then key to start measurement. The measured results will be displayed press in the table automatically. If the “Cell No.” is not set to 1 and you do not wish to perform a No.1 cell blank correction, first put the reference cell into the No.1 cell, close the sample key to perform an auto zero. Then place the compartment, and press to measure the samples. The sample cells into the cell holder, press measurement results will be displayed in the table. If the “Cell No.” is not set to 1 and you wish to perform a No.1 cell blank correction, put the reference cell into the No.1 Cell, then place the samples into the key, the system will automatically calculate the other cells, press absorbance values of the samples. The measured results will be displayed in the table. The value of the reference No.1 cell should be 0.000Abs. To delete or print the measurement results after measurement (please refer to 3.5 measurement). 26 Instruction manual Chapter 4. Spectrum scanning This mode allows the user to measure the absorbance, transmittance and energy of the measured sample (Es), and energy of reference (Er) in a certain wavelength range. 4.1 Gain access to spectrum mode From the main menu, press the key to access the spectrum mode and press to return to the main menu, as shown in Fig. 4.1: A: Main menu B: Spectrum mode Fig. 4.1 Gain access to spectrum mode When you exit the spectrum mode, the system will prompt the user with “Delete all the data? (YES: ENTER, NO: CE)”. If you press ˈyou will return to the main menu and the measured data will be lost. you will return to the photometric mode and you will be prompted to If you press “press START to measure”, then you can continue to perform a scan. Note: Please print out the measured data before returning to the main menu, otherwise, the data will be lost. The following paragraphs give an introduction to the main menu functions within spectrum mode: 27 Instruction manual Fig. 4.2 Main menu of spectrum mode “SPECTRUM” means that the current working modeis spectrum scanning;Ā990 nmārefers to the current working wavelength˗ andĀ0.000 Absārefers to the live absorbance value. Working area: the spectrum is shown as a graph. The abscissa refers to the wavelength value and the ordinate refers to absorbance value (it also can refer to T%, Es or Er according to the photometric mode you choose from the menu). This area displays promps and information: the displayed information is prompting to begin scanning the sample. the user to press the Time and date display area: the time in the above diagram is eight thirty four on April 15 th. Display area for the function keys: shows the function keys detailed functions are shown as follows: ƽ üü access to the method parameters menu; ƽ üü peak picking; ƽ üü spectrum printout; ƽ üü wavelength correction üü . The In addition, you can press from this menu and set the current wavelength and you can perform a baseline correction. by pressing 4.2 Parameters setting From the main menu for spectrum mode, please press to access the method to return to the main menu of spectrum mode, as parameters menu page. Press shown in Fig. 4.3: 28 Instruction manual A: Main menu of spectrum mode B: Parameters setting Fig. 4.3 Method parameters menu page From this page, users can set the photometric mode, scanning speed, scanning wavelength range, ordinate full scale range and sampling interval by pressing the corresponding numeric keys. After you have finished setting up the method parameters, press key to return to the previous page. the 4.2.1 Set photometric mode Press key to set photometric mode. Four modes are available: Abs, Er, Es and %T key. to alternate between each option press the Fig. 4.4 set photometric mode 29 Instruction manual 4.2.2 Setting the scanning speed The options are Fast, Middle and Slow. You can choose Fast if you just want to get general information about peaks in a wide range. You can choose Slow if you want to get accurate and reliable peak location and peak values in a small range. Press key to set the scanning speed Fast, Middle or Slow in turn, as shown in Fig. 4.5: Fig. 4.5 Set scanning speed 4.2.3 Setting the sampling interval Press the key to set the sampling data interval. You have 6 choices: 5nm, 2nm, key is pressed in 1nm, 0.5nm, 0.2nm or 0.1nm they can be selected in turn when the succession. 30 Instruction manual Fig. 4.6 Set sampling data interval Note: A maximum of 910 data points can be stored for one spectrum. The product of sampling interval and wavelength range should be less than 910. If this is exceeded you will prompted with the information “WL range does not match interval”. You should then choose a different data interval. 4.2.4 Setting the wavelength range Press key from the method parameters menu to set the wavelength range, as shown in Fig. 4.7: Fig. 4.7 Setting the wavelength range 31 Instruction manual At the bottom of this page enter the start wavelength by pressing the numeric keys ̣ and decimal point . The key is used to clear the previous entry. key to confirm. If the value you entered meets the requirements the system will Press key to confirm. The system will accept prompt you to input the stop wavelength. Press the wavelength range you have entered and return to the method parameters menu page. If key to quit. nothing is to be entered press the Note: The input wavelength value should be an integer value or with one digit precision. The range is from 190 to 1100. If the input value exceeds 6 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. The default wavelength range is from long wavelength to short wavelength, and the instrument will scan from long wavelength to short wavelength. If the start wavelength you entered is shorter than end wavelength, the instrument will take the long wavelength as the start wavelength and the short wavelength as the end wavelength. 4.2.5 Setting the ordinate range Press key on the parameters setting page in the case of spectrum scanning to set ordinate range, as shown in Fig. 4.8: Fig. 4.8 Setting the ordinate range 32 Instruction manual At the bottom of this page, the system prompts the user to set the lower limit of the üü and to enter the desired ordinate axis. Press the numeric keys Key is used to clear the previous entry. After setting the lower limit, the system value. key to confirm the numeric entry and quit. prompts the user to set the upper limit. Press key to quit. A 5-digit figure entry is allowed. If nothing is to be entered press the The range is from –9999~9999. If the figure entered exceeds five digits the system will delete what has been entered. You should enter it again. Note: The system accepts integral and decimal, the range of which is –999-999. If the input value exceeds 7 digits (including decimal point and negative sign) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. The system remembers the ordinate ranges you set for the different photometric modes. For example if you set the ordinate range for Abs to be -0.01̚0.01 and then you change to %T, and set its ordinate range to be 0̚100, when you change the photometric mode back to Abs again the ordinate range will still be -0.01̚0.01. 4.2.6 Light source selection For the photometric modes of Es and Er you can also select the light source you want to use to peform the spectrum scan. You should make sure that the photometric mode is set to Es or Er. Then press the key from the spectrum scan parameters menu page to select the light source. Each time you key the light source will alternate between tungsten lamp and deuterium press the lamp as shown in Fig. 4.9: Fig. 4.9 Light source selection 4.2.7 Dark current correction Dark current correction is performed to ensure the accuracy of the measured results. You should perform a dark current correction when environmental factors change. This might include temperature variations in the laboratory or a change in location or when measuring highly absorbing samples. The operational procedures are shown as follows: 33 Instruction manual A: Parameters setting page B: Dark current correction Fig. 4.10 Dark current correction Press to perform a dark current correction from the spectrum scan parameters menu page. The message “system busy…” will be displayed. After about 10 seconds the user will be a prompted to “Please select Item”, meaning the dark current correction has finished as shown in Fig.4.10. 4.2.8 Quit parameters setting Press key on parameters setting page to return to the main menu of spectrum scanning, as shown in Fig. 4.3. 4.3 Baseline correction From the main menu of spectrum scanning, press the key to start a baseline correction at the set wavelength range. Before starting a baseline correction, place the blank key can be pressed to stop the baseline sample into the sample compartment. The correction operation. If you change the wavelength range a new baseline should be performed. 4.4 Measurement First set up the method parameters and perform a baseline correction, then put the to start scanning. The measured sample cell into the cell holder and press spectrum will be displayed as a graph. Note: For “Es͇and͆Er͇the system will prompt you to input the gain when pressing key. The range is 1-15. Input your preferred gain and press to to delete the entry. start measurement. Press When the system prompts you with the information “Press START to measure” it means that the measurement has finished as shown in Fig. 4.11 (take Abs for example): 34 Instruction manual Fig. 4.11 Spectrum scanning page The abscissa in the graph above relates to the wavelength range from 1000nm to 190nm, and the ordinate range is displaying the absorbance value from –0.010 to 0.010. 4.5 Peak picking Once the scanning has completed you can perform peak picking by selecting from the main menu of spectrum scanning as shown in Fig. 4.12: Fig. 4.12 Peak picking operation page At the bottom of this page you can enter the threshold value for picking the peaks or üü , and . Press valleys by pressing the numeric keys key to confirm. If nothing is to be entered, press the to clear the previous entry. Press key to quit. Note: The threshold value, shown as percentage, is also the minimum detection limit. The system can detect whether the spectrum is beyond the setup threshold value. The system can accept integral and decimal as threshold value, the range of which is 1-100. If the input value exceeds 4 digits (including decimal point and negative sign) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 35 Instruction manual After the threshold value entered, the system will start peak-picking calculation and the calculated result will be displayed on the screen, as shown in Fig. 4.13. and to check the peaks. After peak picking, you can move cursor keys to print the result, and press to return to the main menu of spectrum Press mode. If the threshold value in the information indication area is 0.000nm 0.000, it means the key and press key to input a smaller system haven’t detected the peak, press threshold value, and then perform peak picking again. Fig. 4.13 Peak picking 4.6 Printout After measurement, you can press key to printout. Please refer to “3.8 Printout”. If there are no measurement results, pressing key is inoperative. Note: Sometimes the spectrum displayed in the LCD is different from the printed one, because the resolution of the printer is higher than that of the LCD. The spectrum data points displayed is less than that of the printed. But the peak values are the same. 4.7 Quit measurement Press to quit measurement. It will prompt the user with the information to return to the main menu of “Delete all the data? (YES: Enter; NO: CE)”. Press 36 Instruction manual instrument. 4.8 Application example Vitamin B12 has characteristic absorption peak at 361nm, which the operators always concern when they perform spectrum scanning of vitamin B12. We will take vitamin B12 for example to show the user how to perform spectrum scanning. The procedures of using T70 UV-VIS Spectrophotometer to measure the absorption peak of Vitamin B12 are shown as follows: Mix standard vitamin B12 injection with de-ionized water in a proportion of 1:10 to get 10ml dilution solution. Then pour the dilution solution and de-ionized water into two quartz cells separately (de-ionized water as reference). Turn on T70 UV-VIS Spectrophotometer. After 60 minutes’ preheat time, press to gain access to spectrum mode menu (refer to 4.1 gain access to spectrum mode). key to gain access to parameters setting page. Set the photometric Press key, set the scan speed to Middle by pressing mode to Abs by pressing , set the wavelength range to key, set the sample interval to 1nm by pressing , and set the ordinate range to 0.000-2.000 by 220nm-660nm by pressing (please refer to “4.2 Parameters setting” for the detailed pressing information). Put the black block into the sample cell on the sample optical path (only this cell on the parameters setting page to perform dark can be used now). Press to return to photometric mode current correction. After correction, press menu. Take out of the black block, put the quartz cell with de-ionized water into the to sample cell holder, cover the sample compartment, then press perform baseline correction. The indication area will display “scanning baseline…”. key during correction, the correction data will be lost and If you press you should start baseline correction again. After baseline correction, the indication area will display “Press START to measure”. Take out of the de-ionized water cell, and put the sample cell into the cell key to perform holder, cover the sample compartment, then press spectrum scanning. The indication area will display “scanning baseline…” and the again to stop measurement and the spectrum will be displayed. Press results will be lost. After measurement, it will prompt the user with the information “Press START to to set threshold value, and then perform peak picking. If measure”. Press peak value is not available (indication area displays 0.000nm 0.000Abs), press 37 Instruction manual to return and then set up threshold value again (for detailed information about peak picking, please refer to “4.5 peak picking”). Press etc.). to print measurement results (spectra, peaks values and wavelength 38 Instruction manual Chapter 5. Quantitative measurement This mode allows the user to perform single-wavelength quantitative determinations and to construct standard curves. The K-Factor Method and Calibration Curve Method are available. 5.1 Gain access to quantitation mode From the main menu, press key to access the quantitation mode, press to return to the main menu, as shown in Fig. 5.1: A: Main menu B: Quantitation mode menu Fig. 5.1 Gain access to quantitation mode When you exit the quantitation mode, the system will prompt the user with the information “Delete all the data? (YES: ENTER, NO: CE)”. Pressing will return to the main menu and the measured data will be lost. Pressing will return to the quantitation mode and a prompt will display with the information “press START to measure” then you can continue with the quant measurements. Note: Please print out the measured data before returning to the main menu, otherwise, the data will be lost. The following paragraphs are an introduction to the quantitation mode menu: 39 Instruction manual Fig. 5.2 Main menu of quantitative measurement ķ “QUANT” means the current working mode is the quantitative measurement, Ā990 nmārefers to the current working wavelength˗ andĀ0.000 Absārefers to the live display of the real time absorbance value. ĸ Working area: the measured results are shown in the table. It can display a maximum of 9 rows of measuremed values each time. Ĺ Information display area: the current information is prompting the user to press to begin measurement. ĺ Time display area: time in the above picture is eight thirty four on April 15. Ļ Display area for the function keys: show the function keys detailed functions are shown as follows: ƽ üü Gain access to method parameters set up page; ƽ üü Delete measurement data; ƽ üü Set up of sample control; ƽ üü Printout üü . The In addition you can also press to set the working wavelength and you can press to perform a baseline correction. 5.2 Parameters setting For quantitative measurement press to access the method parameters set up page. to return to the main menu for quantitative measurement as shown in Fig. 5.3: Press 40 Instruction manual A: Main menufor quantitation mode B: Method parameters set up page Fig. 5.3 Gain access to parameters setting page The quant method parameters menu allows the user to set the working mode, measuring wavelength, and concentration by pressing the corresponding numeric keys. Press to return to the previous page. 5.2.1 Set measurement method K-Factor Method and Standards Method are available. Please refer to “5.3 construct calibration curve” for constructing calibration curves. key to set the working method for quantitative From quant method page press is pressed each time, measurement. The two methods can be selected alternately when as shown in Fig. 5.4: Fig. 5.4 Set measurement method 5.2.2 Set measuring wavelength Press to set the measuring wavelength as shown in Fig. 5.5: 41 Instruction manual Fig. 5.5 Set measuring wavelength () At the bottom of this page, enter the desired wavelength value by pressing the numeric ̣ and decimal point . The key is used to clear the previous keys key to confirm the numeric entry and quit. If nothing is to be entered, entry. Press key to quit this page and return to the quant method page. press Note: The input wavelength value should be an integer value or with one digit precision. The range is from 190 to 1100. If the input value exceeds 6 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. You also can press from the the main menu for quantitation mode (as shown in Fig. 5.2) to set the working wavelength as shown in Fig. 5.6: Fig. 5.6 Set measuring wavelength () At the bottom of this page, enter the desired wavelength value by pressing the numeric ̣ and decimal point . The key is used to clear the previous keys key to confirm the numeric entry and quit. If nothing is to be entered, entry. Press key to quit this page and return to the quant menu page. press Note: The input wavelength value should be an integer value or with one digit precision. The range is from 190 to 1100. If the input value exceeds 6 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 42 Instruction manual 5.2.3 Set concentration unit There are 9 choices available: “none”, “mg/L”, “%”, “mol/mL”, “ug/L”, “g/L”, “mg/mL”, “ng/mL”, “mol/L”. key to set concentration unit. Each time you press key, the Press concentration unit will change between “none”, “mg/L”, “%”, “mol/mL”, “ug/L”, “g/L”, “mg/mL”, “ng/mL”, and “mol/L”, as shown in Fig. 5.7. Fig. 5.7 Set the concentration unit 5.2.4 Dark current correction A dark current correction is performed to ensure the accuracy of your data. You should perform a dark current correction when environment changes occur such as temperature variations and changes in location for the instrument, also when you measure highly absorbing samples. The operation procedures are shown as follows: 43 Instruction manual A: Parameters setting page B: Dark current correction Fig. 5.8 Dark current correction Press to perform a dark current correction from the parameters setting page. A message will be displayed “system busy…”. After about 10 seconds you will be prompted with the message “Please select Item” this means that the dark current correction has finished as shown in Fig.5.8. 5.2.5 Quit parameters setting From the parameters setting page for quantitative measurement press quit the menu as shown in Fig. 5.3. key to 5.3 Constructing a calibration curve Before you construct a calibration curve you should set the parameters. 5.3.1 K-Factor method The K-Factor method is used to perform quantitative measurements if you know the calibration curve. You can construct a calibration curve by inputting the factor and the intercept. First make sure that you have selected the measurement method to be “K-Factor”. Press key from the quantitation parameters setting page to access the parameters setting page. key to return to the previous page as shown in Fig. 5.9. Press 44 Instruction manual Fig. 5.9 Parameters setting page for the K-Factor method The displayed formula is “Conc˙K*Abs+B”. Please set the Slope K and Intercept B by pressing and key separately. Set the slope for the calibration curve From the parameters setting page for K-Factor method, press as shown in Fig. 5.10. key to set slope K, Fig. 5.10 Set the slope of K-Factor method At the bottom of this page, enter factor K by pressing the numeric keys ̣ . key is used to clear the entry. Press key to confirm. If nothing is to be The key to return to parameters setting page. entered, press Note: The system accepts integral and decimal, the range of which is –999-999. If the input value exceeds 7 digits (including decimal point and negative sign) or exceeds the the system will delete what has been entered automatically and you should start a range, new entry again. Set the intercept of calibration curve From the parameters setting page for K-Factor method, press intercept B as shown in Fig. 5.11. 45 key to set the Instruction manual Fig. 5.11 Set the intercept of K-Factor method At the bottom of this page enter the intercept B by pressing the numeric keys ̣ . The key is used to clear the entry. Press key to confirm. If nothing is to key to return to parameters setting page. be entered press Note: The system accepts integral and decimal, the range of which is –999-999. If the input value exceeds 7 digits (including decimal point and negative sign) or exceeds the range, the system will delete what has been entered automatically and you should start a entry again. new 5.3.2 Standards method The Standards method refers to the method of constructing a calibration curve by using the standards concentration values and absorbance values. First ensure that you have selected the measurement mode to be “Standard”. Press from the quantitation parameters setting page to access the parameters setting page for the Standards method as shown in Fig. 5.9. Fig. 5.12 Parameters setting for the Standards method From the parameters setting page press to set the number of standards, press to set the concentration and absorbance values of a specific standard sample. Press key to use cell 1 to measure the standards absorbance values. After measurement, press key to construct and display the calibration curve. 46 Instruction manual Set standards number Press key to set standards number, as shown in Fig. 5.13: Fig. 5.13 Set standard number At the bottom of this page, enter standard number by pressing the numeric keys ̣ . The key is used to clear the entry. Press key to confirm. If nothing is to key to return to parameters setting page. be entered press Note: The system accepts integral㸪the range of which is 1-10. If the input value exceeds 7 digits (including decimal point and negative sign) or exceeds the range, the system will what has been entered automatically and you should start a new entry again. delete Set concentration value and absorbance value From the parameters setting page for the standards method, press key to set the specific standard sample, as shown in 5.14. At the bottom of this page, enter standard No. ̣ . The key is used to clear the entry. Press by pressing the numeric keys key to confirm. If the entered number is eligible, the system will prompt the user to key to skip input the standard concentration. If nothing is to be entered press the it. The default standard number will be used (as displayed on the screen). At the bottom of this page enter the concentration value by pressing the numeric keys ̣ . The key is used to clear the entry. Press key to confirm. If the entered value is eligible, the system will prompt the user to input the standard absorbance key to skip it. The default concentration value. If nothing is to be entered, press will be used (displayed on the screen). At the bottom of this page, enter the absorbance value by pressing the numeric keys ̣ . The key is used to clear off the entry. Press key to confirm. If the entered value is eligible, the system will prompt the user with the information “Please select key to quit the standards setting page. The item”. If nothing is to be entered, press default absorbance value will be used (displayed on the screen). 47 Instruction manual Fig. 5.14 Set the standards concentration value and absorbance value Note: The system accepts integral as the Std. No.㸪the range of which is 1 to Std. Number. If the input value exceeds 7 digits (including decimal point and negative sign) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. The system accepts integral and decimal as the concentration and absorbance, the range of which is –999-999. If the input value exceeds 7 digits (including decimal point and negative sign) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. Measure the standards absorbance The user should perform a blank correction before measuring the standards absorbance. Put the blank into the No. 1 cell, and then press key to perform auto zero. After auto zero, press key in the parameters setting page to measure the absorbance, as shown in Fig. 5.15. Put the standards into the No. 1 sample cell, and then press key to measure the current standards absorbance. After measurement, the result will be displayed and the system will prompt the user with the information “Please Select Item”. 48 Instruction manual Fig. 5.15 Measure the standards absorbance Standards absorbance measurement is used when the user knows the standards concentration but does not know the standards absorbance. First select the standard No., then to skip the absorbance value input the standards concentration value and then press setup to measure the standards absorbance. Construct calibration curve key on the parameters setting page of standards After the standards setup, press method to construct the calibration curve, as shown in Fig. 5.16. Fig. 5.16 Calibration curve display The abscissa in the above figure is the absorbance value and the ordinate is the concentration value. The calibration curve equation will be displayed at the bottom of the key to printout (Please refer to 3.8 Printout). page. Press Press key on the calibration curve display page to return to the parameters setting page. The calibration curve equation will be displayed under the “Curve Parameter”. Note: If the standards setup does not meet the requirements of constructing a calibration curve, the system will prompt the user with “ Data Range Error, Retry.” Press key then to reset the parameters. 49 Instruction manual Quit calibration curve parameters setting Press setting page. key to quit calibration curve parameter setting, and return to parameter 5.4 Sample control From the main menu for quantitative measurement as shown in Fig. 5.2, press key to access the sample control page. The sample control operation is the same as that of the photometric measurement. Please refer to “3.3 Sample control”. 5.5 Auto zero From the main menu for quantitative measurement press the key to set the Absorbance to zero at the current measuring wavelength. Before starting auto zero place the blank into the cell. 5.6 Measurement 5.6.1 Use single cell The single cell measurement includes fixed cell measurement and using one cell position on the multi-cell changer to take measurements (refer to 3.3.6 single cell measurement). First set up measurement parameters and sample cell parameters then construct a calibration curve. Perform auto zero. Place the sample into the sample cell and press the key to measure its Absorbance value. Each time you press the key it will re-measure the sample. The measurement results will be displayed as shown in Fig. 5.17. Fig. 5.17 Single cell measurement results 50 Instruction manual ĀNo.āIn the table refers to serial number of the samples set by the user. Ā2 -1āin the table refers to the second measurement result of No.1 cell (if a multi-cell changer is being used it refers to the current sample cell in use); “Abs” refers to absorbance value;ĀConc” refers to the concentration value calculated by the calibration curve. 5.6.2 Using multi-cells First set up measurement parameters and sample cell parameters, and then construct a calibration curve. The measurement procedures will vary with the settings of No. 1 cell blank. When Reagent Blank Correction in Cell No. 1 is set to “NO” When Reagent Blank Correction in Cell No. 1 is set to “NO”, you need to perform an auto zero on position cell No.1. After auto zeroing put the samples into the sample cells, then key to start measurements. The measured values will be calculated, saved press and displayed on the LCD one by one. When Reagent Blank Correction in Cell No. 1 (c3.5.3) is set to “YES” When Reagent Blank Correction in Cell No. 1 is set to “YES”, you do not need to perform auto zero. You need to put the blank sample into Cell position No.1, and then put the samples into the other cell positions. The system will first measure the set N samples. Then the measured values of samples No.2~No.N will be deducted (corrected) by that of the blank in sample cell No. 1 following the formula below. The corrected results will be recorded, calculated and displayed on the LCD one by one. ƽ Absorbanceφ $EVRUEDQFHRI&HOO1RQ WKHPHDVXUHPHQWYDOXHRI&HOO1RQ㸫$EVRI&HOO 1R The measurement results using the multi-cell holder is shown as Fig.5.18. ĀNo.āin the table refers to the serial number of the samples set by the user. Ā2 -3āin the table refers to the second measurement result of No.3 cell (if use multi-cell changer, it refers to current sample cell in use); “Abs” refers to absorbance value; “Conc” refers to the concentration value calculated by the calibration curve. The following picture shows the results of using three cells measuring twice. Fig. 5.18 Multi-cell measurement results 51 Instruction manual 5.7 Looking up the measurement results After measurement you can look up the results by pressing cursor keys . and 5.8 Delete the measurement data After measurement press the delete the measurement data”. key to delete measured results. Please refer to “3.7 5.9 Printout After measurement press the “3.8 Printout”. key to print out the measured results. Please refer to 5.10 Quit measurement Press to quit measurement. You will be prompted to “Delete all the data? (YES: to return to the main menu of instrument. Enter; NO: Clear)”. Press 5.11 Application example Determination of nitrite in food is an important item in food safety and inspection. The quantitation mode of the T70 series spectrophotometers can be used to determine the nitrite content and the procedures are shown as follows: Reagent preparation: Distilled water should be used for preparing standard solutions. Ammonium chloride buffer: Pipette 500ml distilled water, 20.0ml hydrochloric acid into a 1L volumetric flask. Shake up. Then pipette 50ml ammonium chloride into the volumetric flask. Dilute to the mark with distilled water. If necessary, adjust the buffer PH to 9.6-9.7 with dilute hydrochloric acid and dilute ammonium hydroxide. ƽ Zinc sulfate solution (0.42̉̋̈ˋ˨): dilute 120g zinc sulfate (˶̊˯˫4g˓ˤ2 52 Instruction manual ˫) to 1000ml with distilled water. ƽ Sodium hydroxide solution (20g/L): Dilute 20g sodium hydroxide to 1L with distilled water. ƽ Sulfuric acid solution: Dilute 10g sulfuric acid into the mixed solution of 700ml water and 300 ml glacial acetic acid. Put the solution into a brown bottle, and then shake up. Preserve it at room temperature. ƽ ˪ ˉ ˍ ˉ Naphthyethylenediamine (1g/L): Dilute 0.1g ˪ ˉ ˍ ˉ Naphthyethylenediamine to 100ml with 6% acetic acid solution. Put it into a brown bottle, and shake it up, then place it into a refrigerator for a week. ƽ Chromomeric agent: Mix ˪ˉˍˉNaphthyethylenediamine solution (1g/L) with sulfuric acid solution equally. ƽ Sodium nitrite standard solution: Put 250.0mg sodium nitrite, dried in a silica gel dryer for over 24 hours, into a 500ml volumetric flask. Dissolve it with water (about 200ml). Add 200ml ammonium chloride buffer to it, and then dilute it to the mark with water. Swirl to mix. Preserve the solution at 4ć, avoiding light. Each milliliter of the solution equals 500̃ sodium nitrite. ƽ Sodium nitrite standard: Before measurement, pipette 1.0ml standard solution into a 100ml volumetric flask, and then dilute it to the mark with water. Each milliliter of the solution is equal to 5.0̃ sodium nitrite. Sample processing: Place the sample 10.00g (foodstuff 5g) into the breaker to break. Mix it with 70ml water and 12ml sodium hydroxide. Adjust the sample PH to 8 with sodium hydroxide (20g/L). And then place the sample solution into a 200ml volumetric flask. Mix it with 10ml zinc sulfate. If it doesn’t produce a white deposit add sodium hydroxide 2-5ml to mix it up. Place the sample into a water bath, heat for 10mins and then cool it to room temperature. Add water to the mark. Filter the sample solution with filter paper after half an hour. Discard the primary filter solution 20ml, and collect the filter solution. Switch on the instrument Turn on the T70 UV-VIS Spectrophotometer. After 30mins warm-up, press main menu to access the quantitation mode. from Parameters setting Please refer to “5.2 Parameters setting”, set the measurement mode to “Standards method” and working wavelength to 550nm. Standard preparation Pipette the prepared sodium nitrite standard 0.10ml, 0.20ml, 0.40ml, 0.6.ml, 0.80ml, and 1.00ml (equal to sodium nitrite 2.5Ϥg, 5Ϥg, 10Ϥg, 15Ϥg, 20Ϥg and 25Ϥg) into 25ml colorimetric tubes with stoppers. Then add 4.5ml ammonium chloride buffer, 2.5ml 60% 53 Instruction manual acetic acid solution, and 5.0ml chromomeric agent into the tubes respectively. Dilute them with water to the mark. Place the prepared standard samples into the dark for 25 minutes. Place the different concentration standards into 1cm quartz cells respectively. And then number the cells 1-6. ļ Standards setup key to access the menu From the method page for quantitative measurement press key to change the standard number to 6. for setting up the standards method. Press to perform Place the quartz cell with distilled water into the sample cell and press an auto zero. key to input the standards number, sample From the standards method page, press to skip absorbance entry. Then place the numbered sodium concentration, and press key and key. The Abs value displayed is nitrite into the cell No.1, press the current sodium nitrite Abs. Construct the calibration curve From the standards method page press key to construct a calibration curve, as shown in Fig. 5.19. If the instrument displays “Data Range Error, Retry”, press the key to return to the standards method page to reset the standard samples. Fig. 5.19 Calibration curve Measurement Pipette 10ml filter solution into a 25ml colorimetric tube with a stopper. Then perform measurement (refer to 5.6 measurement). Calculate the nitrite concentration by the measured absorbance value and the standard curve. Ŀ Result analysis After measurement, the nitrite content is calculated as 35.63mg/kg by the following formula. 54 Instruction manual A u1000 ĂĂĂĂĂĂĂĂĂĂĂ˄˅ V2 m u u 1000 V1 X ̾̾nitrite content㸪 mg / kg 㸹 X m —sample weight㸪 g 㸹 A ̾̾nitrite weight of the filter solution in use㸪 Pg ; V2 ̾̾the total filter solution volume㸪 mL 㸹 V1 ̾̾the volume of filter solution in use㸪 mL . 55 Instruction manual Chapter 6. DNA/Protein Analysis This mode allows the user to perform DNA/Protein analysis. There are two pre-set methods and a custom defined method available. 6.1 The principles of Protein/DNA analysis 6.1.1 Principles of using the absorbance of UV light to determine protein: Proteins are bio-polymers there fundamental unit is the Amino Acid, there are 20 standard Amino Acids and they are connected together by peptide bonds. Amino Acids that have aromatic groups in their structures are Histidine, Phenylalanine, Tryptophan and Tyrosine. The conjugated doublebond included in the tryptophan, phenylalanine, and tyrosine absorb UV light. The absorbance values are related to the concentration of the three kinds of amino acid. Peptide bonds also absorb UV light so the absorbance value of protein solution at 238nm is as a result of the peptide bond which is then directly related to the protein concentration. 6.1.2 Principle of using absorbance of UV light to determine DNA: There are two types of nucleic acids, DNA - DeoxyRibonucleic Acid and RNA RiboNucleic Acid. The Nucleic acids are polymers; i.e., strings of similar molecules strung together. The fundamental unit is a nucleotide and the polymer is a polynucleotide. A nucleoside is a base + sugar and a nucleotide is base + sugar + phosphate. The five bases of Adenine, Guanine, Cytosine, Thymine and Uracil do absorb in the UV. The high absorbance is due to, double bonds inside aromatic rings, lone pairs on nitrogen atoms and Oxygen double bonds. They have the maximum absorbance value at 260nm. 6.1.3 Protein/DNA determination methods Absorption method at 280nm and 260nm DNA & RNA absorb UV light intensely. Its absorbance ability is 10 times as intense as protein at 280nm. The absorption of DNA & RNA at 260nm is more intense. The extinction coefficient at 260nm is twice as much as that at 280nm. But for protein, the absorption value at 280nm is larger than that at 260nm. Because the protein absorption value and DNA & RNA absorption values are different, we can judge the ratio of protein and DNA & RNA, for example: ƽ Pure protein absorbance ratio: A280/A260 | 1.8 ƽ Mixed solution of protein and DNA & RNA: 56 A280/A260 | 1.8̚0.5 Instruction manual ƽ Pure DNA & RNA absorbance ratio: A280/A260 | 0.5 Method: As for the mixed solution of protein and DNA, we can determine their concentration values by measuring their absorption values at the wavelength of 260nm and 280nm. The formulas are shown as below: Protein concentration 㸻 1552A280㸫757.3A260 DNA concentration 㸻 62.9A260㸫36.0A280 (ug/ml) ĂĂĂĂ˄6-1˅ ǏĆ1ǐ (ug/ml) ĂĂĂ˄6-2˅ ɋ*1ɌNote: 1. The coefficients need to be amended according to different situations. 2.The method is easily interfered, such as the NaCL and the solution with conjugated doublebond. ĸ Absorption method at 230nm and 260nm The peptide bond of protein has its maximum absorption value at 238nm, and we take A230 as an approximate value. DNA&RNA also has intense absorption at 230nm. So we can determine A230 and A260 with the fixed solution of protein and DNA. Method: We can determine the concentration values of protein and DNA by measuring their absorption values at the wavelength of 230nm and 260nm. The formulas are shown below: Protein concentration 㸻 183.0A230㸫75.8A260 DNA concentration 㸻 49.1A260㸫3.48A230 (ug/ml)……………...˄6-3˅ (ug/ml)………………….˄6-4˅ The coefficients above need to be amended according to different situations. Comparing with absorption method at 280nm and 260nm, the absorption method at 230nm and 260nm is more accurate. But it is more easily interfered. There is no absorbance at the wavelength of 320nm, so background correction at 320nm can be preformed. 6.2 Gain access to DNA/Protein analysis From the instrument main menu press key to select the DNA/Protein mode. The main menu for quantitative measurement will be displayed as shown in Fig. 6.1, press to return to the instrument main menu. 57 Instruction manual A: Instrument main menu B: Main menu of DNA/Protein analysis Fig. 6.1 Gain access to DNA/Protein analysis When you exit the DNA/Protein analysis mode, the system will prompt the user with the information “Delete all the data? (YES: ENTER, NO: CLE)” If you press you will return to the main menu and the measurement data will be you will return to the main menu for DNA/Protein analysis and be lost. If you press rompted to “press START to measure”, then you can continue to the DNA/Protein analysis page. Note: Please print out the measurement data before returning to the main menu, otherwise, the data will be lost. The following paragraphs introduce the main menu for DNA/Protein analysis: Fig. 6.2 Main menu for DNA/Protein analysis ķ ĀDNA/Protāmeans that you are currently on the DNA/Protein analysis page. Ā260nm” and “230nm” refers to current working wavelengths˗ 58 Instruction manual ĸ Working area: Measurement results are shown in the table and can display a maximum of 9 rows of measurement results each time. Ĺ Information display area: The current information is prompting the user to press to begin measurement. ĺ Time display area: Time in the above picture is eight thirty four on April 15. Ļ Display area for the function keys: shows the function keys detailed functions are shown as follows: ƽ üü Access to parameters setting page; ƽ üü Delete measurement data; ƽ üü Set sample control; ƽ üü Printout üü . The In addition, pressing from this menu will perform a baseline correction. Press key and key to alternate between A1, A2, DNA concentration, and protein concentration as shown in Fig. 6.3: Fig. 6.3 Results display of DNA/Protein analysis 6.3 Parameters setting From the main menu of DNA/Protein analysis press to access the parameters to return to the main menu for DNA/Protein analysis as shown setting page. Press in Fig.6.4: 59 Instruction manual Fig. 6.4 Parameters setting page From the method menu page for DNA/Protein analysis, the user can select either the preset methods or the custom defined method. The user also can perform a dark current correction from this page. 6.3.1 Selecting the analysis method Press key and key from the parameters setting page to select between method 1, method 2 and a custom defined method as shown in Fig. 6.5: 60 Instruction manual Fig. 6.5 Select DNA/Protein analysis method When the user selects method 1 or method 2, the parameters are displayed and the user is not allowed to amend them. When the user selects custom defined method, the user can input the preferred parameters. 6.3.2 Dark current correction Dark current correction is performed to ensure accuracy of the measurement results. You should perform a dark current correction when the environmental changes occur, such as temperature variations and change of location of the instrument. Also when you measure highly absorbing samples it is recommended that you perform a dark current correction. The operational procedures are shown as follows: A: parameters setting page B: Dark current correction Fig. 6.6 Dark current correction Press to perform a dark current correction from the parameters setting. A prompt will display the message “system busy…” After about 10 seconds you will be prompted to “Please select Item”, this means that the dark current correction has completed as shown in Fig.6.6. 6.4 Custom defined method From the parameters setting page select “Custom” to gain access to the custom defined method. 6.4.1 Set measuring wavelength From the custom defined method, press shown in Fig. 6.7: 61 key to set the working wavelength as Instruction manual Fig. 6.7 Set wavelengths At the bottom of this page, enter the first wavelength value by pressing the numeric keys ̣ and decimal point . The key is used to clear the previous entry. key to confirm the numeric entry. The system will prompt the user to input the Press key to confirm the entry. If second wavelength after entering the first one. Press key to skip this and the system will adopt the nothing is to be entered press the current setting. Note: The input wavelength value should be an integer value or with one digit precision. The range is from 190 to 1100. If the input value exceeds 6 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 6.4.2 Set DNA factors From the custom defined method, press 6.8: 62 to set the DNA factors, as shown in Fig. Instruction manual Fig. 6.8 DNA factors setting At the bottom of this page, enter the first DNA factor by pressing the numeric keys ̣ and decimal point . The key is used to clear the previous entry. key to confirm the numeric entry. The system will prompt the user to input the Press key to confirm the entry and quit. other protein factor after entering the first one. Press key to and the system will adopt the current If nothing is to be entered press the setting. Note: The input protein factors should be integers or decimals. The range is from -9999 to 9999. If the input value exceeds 7 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 6.4.3 Set protein factors From the custom defined method, press 6.9: 63 to set protein factors, as shown in Fig. Instruction manual Fig. 6.9 Protein factors setting At the bottom of this page, enter the first protein factor by pressing the numeric keys ̣ and decimal point . The key is used to clear the preceding entry. key to confirm the numeric entry. The system will prompt the user to input the Press key to confirm the entry and quit. other protein factor after entering the first one. Press key and the system will adopt the current setting. If nothing is to be entered, press Note: The input protein factors should be integers or decimals. The range is from -9999 to 9999. If the input value exceeds 7 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 6.4.4 Set background correction at 320nm From the custom defined method, press key to set background correction at 320nm. Two choices are available for this item: YES and NO, which can be selected by key is pressed, as shown in Fig. 6.10: alternating between the two options each time 64 Instruction manual Fig. 6.10 Background correction at 320nm 6.4.5 Set calibration factors From the custom defined method, press shown in Fig. 6.11: key to set the calibration factors, as Fig. 6.11 Set calibration factors At the bottom of this page, enter the calibration factors by pressing the numeric keys ̣ and decimal point . The key is used to clear the preceding entry. key to confirm the numeric entry. The system will display the factors and return Press to the parameters setting page if the entered value meets the requirements. If nothing is to be key to adopt the current setting. entered, press Note: The input calibration factors should be integers or decimals. The range is from -999 to 999. If the input value exceeds 7 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 65 Instruction manual 6.5 Auto zero From the main menu for DNA/Protein analysis press the key to set the Absorbance to zero at the current measuring wavelength. Before starting auto zeroing, put the blank sample into the cell. 6.6 Measurement Only one cell can be used for DNA/Protein analysis (No.1 cell is used if a multi-cell key holder is installed). From the main menu for DNA/Protein analysis press the it re-measures the sample and to start the measurement. Each time you press displays the results as shown in Fig. 6.12: Fig. 6.12 Results display ĀNo.āin the table refers to the serial number of the samples set by the user. Ā2āin the table refers to the result of second measurement; “A1/A2” refers to the absorption ratio of the two calibration wavelengths; ĀA1ā and ĀA2ārefer to the absorption values of two key and key to display A1, A2, DNA calibration wavelengths. Press concentration and protein concentration. For detailed information refer to “6.7 Look up the measurement results”. 6.7 Look up the measurement results After measurement, you can look up the measurement results by pressing cursor keys and . The default display is A1 and A2, and the user can press key and key to alternate the display between A1, A2 and DNA concentration and protein concentration, as shown in Fig. 6.12: 66 Instruction manual 6.8 Delete the measurement data After measurement press “3.7 Delete the measurement data”. key to delete the measurement results. Please refer to 6.9 Printout After measurement press the key to print out the measurement results including A1, A2, DNA concentration and protein concentration. Please refer to “3.8 Printout”. 6.10 Quit measurement Press the key to quit measurement. You will be prompted with the information to return to the instrument main “Delete all the data? (YES: Enter; NO: Clear)”. Press menu. 67 Instruction manual Chapter 7. Utilities This mode allows the user to set the operating conditions such as lamp changeover wavelength, spectral bandwidth, tungsten lamp On/Off, deuterium lamp On/Off and wavelength calibration. 7.1 Gain access to utilities From the instrument main menu press the key to access the utilities mode and key to return to the instrument main menu as shown in Fig. 7-1: press the Fig. 7.1 Gain access to utilities mode 7.2 Set lamp changeover wavelength From the main menu for utilities press wavelength as shown in Fig. 7-2. key to set the lamp changeover Fig. 7.2 Set lamp changeover wavelength At the bottom of this page, enter the preferred wavelength by pressing the numeric keys ̣ and decimal point . The key is used to clear the previous entry. key to confirm the numeric entry and quit. If nothing is to be entered press the Press 68 Instruction manual key to return to the main menu for utilities. Note: The input wavelength values should be integers or decimals. The range is from 390 to 330. If the input value exceeds 6 digits (including decimal point) or exceeds the range, the system will delete what has been entered automatically and you should start a new entry again. 7.3 Setting the spectral bandwidth Press key to set the spectral bandwidth of the instrument. Four options are key is pressed each available: 2.0, 5.0, 0.5, 1.0nm. They can be selected in turn when time as shown in Fig. 7.3: Fig. 7.3 Setting the spectral bandwidth 7.4 Setting the tungsten lamp On/Off Press the key to access the tungsten lamp on/off function. On and Off can be key is pressed each time as shown in Fig. 7.4: changed in turn when 69 Instruction manual Fig. 7.4 Setting the tungsten lamp On/Off 7.5 Setting the deuterium lamp On/Off Press key to access the deuterium lamp on/off function. On and Off can be key is pressed each time as shown in Fig. 7.5: changed in turn when the Fig. 7.5 Setting the deuterium lamp On/Off Note: When the user turns on the deuterium lamp the system will display “System busy”. When the message appears “Please select item”, the deuterium lamp has been ignited. 70 Instruction manual 7.6 Set control mode Press the you press the key to change control mode from PC mode to MCU mode. Each time key you can alternate between the two options as shown in Fig. 7.6: Fig. 7.6 Set control mode MCU means the instrument is operating in the MCU (local control) mode. When you choose PC mode the system will quit the utilities and restart. The system will prompt the user with the information “ Waiting, press START to return”. Fig. 7.7 Waiting for PC mode 71 Instruction manual 7.7 Set time Press the key from the utilities menu page to set the time as shown in Fig. 7.8: Fig. 7.8 Setting the Time At the bottom of the screen it will display the current date. The time in the above picture key and key to increase or decrease the is 19:49 on April 20 2005. Press key and key to move the cursor. After you have finished numeric values. Press key to return to the utilities main menu. Once you have setting the time press the edited the time and date you should quit the utilities mode to check the new settings. 7.8 Reset system Press from the main menu for utilities mode to reset all of the instruments parameters to its original deafault setup. A message will be displayed “System busy…”. About three seconds later you will be prompted to “Please Select Item”, now the parameters have been reset with the original default settings. The original setup parameters are shown in Tab 7-1. Note: This series of instruments can record the system parameters. The instrument will adopt the parameters the user set last time even if the power is turned off. 72 Instruction manual Table 7-1 Original setup parameters Photometric measurement Description Parameter Name Parameter value Photometric mode Abs Measuring wavelength 260.0nm Coefficient 10 Spectrum scanning Description Parameter Name Parameter value Photometric mode Es Scanning speed Fast Sampling interval 1.0 Wavelength range 1000.0 200.0 Photometric mode: Abs Ordinate range 0.00 1.00 Photometric mode: Tˁ, Er, Es Ordinate range 0.00 100.00 Light source Deuterium lamp Quantitative measurement Description Parameter Name Parameter value Measurement method Standards method Measuring wavelength 260.0nm Concentration unit mg/L DNA/Protein analysis Description Parameters can’t be changed Parameter Name Method Parameter value 1 Sample control Description Parameter Name Parameter value Sample module 8-cell changer Drive cell No. 1 Cell No.1 blank No Move cell 1 Utilities Description Parameter Name Lamp changeover wavelength Parameter value 359.0nm Spectral bandwidth 2.0nm Tungsten lamp On Deuterium lamp On Control mode MCU 73 Instruction manual Chapter 8. Maintenance The T70 UV-VIS Spectrophotometer is a precision instrument. Careful assembling and final testing have been performed in the factory. Appropriate routine attention and maintenance will not only ensure the reliability and stability of the instrument but also extend its operating lifetime. 8.1 Points for attention: ƽ Good environmental conditions should be provided according to the requirements in SectionĀ2.1 Site requirement” ƽ The instruments leave the factory fully optimised to offer excellent performance on arrival at their destination. Users should not try to adjust the optical components by themselves nor dismantle parts from the instrument. Pay special attention not to touch and damage the surfaces of the optics or to wipe the surfaces carelessly. 8.2 Daily care ƽ Always check to see if there are any sample spillages in the sample compartment after each measurement. The sample compartment should be cleaned frequently to avoid corrosion and damage to optical system. ƽ The instrument should be covered with the anti-dust cover provided with the instrument. Silica-gel bags can be placed in the sample compartment and the light source housing to keep the instrument dry. But be sure to take the bags out before switching on the instrument. ƽ During daily operation and storage, the LCD display and keyboard should be handled with care in order to prevent damage from scratches, water, dust and corrosion. ƽ Periodically check the performance of the instrument. If anything abnormal occurs, contact the seller or the manufacturer immediately. ƽ If the instrument is not to be used for a long period of time, pay attention to environmental conditions such as temperature and humidity and periodically replace the silica-gel bags. It is recommended to turn on the instrument once a month. 74 Instruction manual 8.3 Trouble shooting and maintenance The instrument begins to initialise after turning it on as shown in Fig. 9.1. The system gives status information after each operation has completed. If no problems are detected it displays “OK”. If there is a problem detected it displays “Err” (Error) and asks whether the initialisation should continue or not. The user will know the status of the instrument once the initialization finishes. Fig. 8.1 Initialisation 75 Instruction manual Table 8-1 Phenomenom The instrument does not operate at all, after the power is turned on. Spectrum can not be printed out. Causes A) Power cable is not securely connected; B) Fuse has blown; C) Circuit failure. A) Printer failure; B) CPU failure. C) The connection cable is loose Initialisation abnormality: 1) ROM check A) ROM failure; 2) Wavelength origin; 3) Multi-cell origin; 4) Slit origin; 5) Filter origin; 6) W lamp energy; 7) D2 lamp energy; 8) Wavelength check Noise abnormality A) Abnormality of WL origin ; A) Abnormality of multi-cell holder drive origin; A) Slit mechanism failure A) Filter mechanism failure; A) Opaque object in the sample compartment; B) W lamp is not lit A) Opaque object in the sample compartment; B) D2 lamp is not lit; A) Opaque object in the sample compartment; B) Wrong positioning of D2 lamp. A) Light source ageing; B) Wrong position of the sample compartment; C) Low voltage / strong magnetic field; D) Receiver ageing ; E) Preamplifier board failure; F) Abnormality of AD converter. Diagnosis A) Check the power cable; B) Check the fuse; C) Check the instrument. A) Check printer; B) Check CPU board. C) Check the connection cable A) Switch on power again to check; A) Switch on power again to check; A) Shift cell holder to check; B) Loose mechanical connection; A) Switch on power again to check; A) Switch on power again to check; A) Open the sample compartment; B) Open the light source housing; A) Open the sample compartment; B) Open the light source housing; C) Remove the cover of instrument; A) Open the sample compartment; B) Open the light source housing. A) Measure in Energy mode; B) Observe at 550nm; C) Check voltage/source of magnetic field; D) Measure in Energy mode; E) Check the preamplifier board; F) Check the drive board. 76 Remedy A) Connect it securely; B) Replace the fuse (2A); C) Contact our service personnel. A)-B): Contact our service personnel. C) Connect the connection cable securely; A) Contact our service personnel; A) Contact our service personnel; A) Remove the obstacle; B) Tighten the connecting mechanism; A) Contact our service personnel; A) Contact our service personnel; A) Remove the opaque object; B) Replace the W lamp ; A) Remove the opaque object; B) Replace the D2 lamp; C) Replace the fuse of D2 lamp; A) Remove the opaque object; B) Reposition the D2 lamp. A) Replace the light source; B) Reposition the sample compartment; C) Add voltage stabilizer /eliminate interference; D) Contact our service personnel; E) Contact our service personnel; F) Contact our service personnel. Instruction manual Appendix 1 Light source replacement Replacing the tungsten lamp The tungsten lamp is plugged into a socket and protected by a plastic sleeve holder. It is well adjusted before delivery. If you need to replace the tungsten lamp please observe the following procedures to replace it. Fig. 9.1 Tungsten lamp set Open the light source housing cover. Push upwards where the arrow points in Fig. 10, and then remove the light source cover. Fig. 9.2 Open the light source house cover 77 Instruction manual ĸ Loosen the two holding screws on the W lamp socket and remove the W lamp holder, as shown in Fig. Fig. 9.3 Loosen the two holding screws Ĺ Replace with a new W lamp holder. Clean the surface of the W lamp with a clean cloth soaked with alcohol. Insert the W lamp holder into the W lamp socket in the light source house and tighten the two holding screws. Fig. 9.4 The filament direction ĺ If the performance of the instrument changes greatly after replacing the lamp, the height of the W lamp should be adjusted. (The horizontal position of the W lamp is adjusted automatically when the self-test is performed upon switching on the power.). First remove the W lamp holder, loosen the setscrews on the W lamp holder with a watch screwdriver and adjust the height of the W lamp. The distance between the center of the filament and the bottom should be 40mm according to the 78 Instruction manual design requirements. But it can be adjusted slightly. And then tighten the set screws and insert the W lamp holder into the socket. Deuterium lamp replacement The D2 lamp is plugged into the D2 lamp socket. To replace the D2 lamp insert a new lamp directly and securely into the socket. The procedure is shown as follows: Open the light source house. ĸ Remove the deuterium lamp. Fig; 9.5 Remove the deuterium lamp Clean the surface of the D2 lamp with a clean cloth soaked with alcohol. Insert the D2 lamp into its socket. The bulge part of the deuterium lamp and the dent of the holder should be rightly overlapped (the light beam window should face the convergent lens), as shown in Fig. 10.6. 79 Instruction manual Fig. 9.6 Insert the deuterium lamp Slightly sway the deuterium lamp and install it securely. Note: 1. The light source and light source housing are considerably hot during operation. Therefore be sure to turn off the power to the instrument so as to cool down the light source sufficiently before replacing it. 2. The convergent lens should be covered with a paper bag to avoid being contaminated by fingers. 80 Instruction manual Appendix 2 Cell holder replacement This appendix describes how to replace the cell holder. Replacing the single cell holder with an 8-cell changer is also similar to the method for replacing with a 5-cell changer and solid sample cell holder etc. Open the sample compartment, and remove the front baffle, as shown in Fig. 11.1. Fig. 10.1 Remove the front baffle Loosen the two holding screws by using a plus screwdriver and remove the solid sample cell holder. Fig. 10.2 Remove the solid sample cell holder Place the plug of the 8-cell changer into the socket on the pedestal of the sample compartment. 81 Instruction manual Position the 8-cell changer, fasten the two holding screws, install the front baffle, and cover the sample compartment, as shown in Fig. 10.3. Fig. 10.3 Installation of the 8-cell holder 82 Instruction manual Appendix 3 Fuse replacement This chapter introduces the user how to replace the fuse. If you suspect the fuse has blown please replace it according to the following procedures. Switch off the instrument. Remove the fuse holder using a big plain screwdriver, as shown in Fig. 12.1. Fig. 11.1 Remove the fuse holder Remove the old fuse, and replace it with a new one. And then install the fuse holder. The fuse replacement has finished. Fig. 11.2 Replacing the fuse Note: Please select the same fuse as the old one when when replacing it. 83 Persee Analytics Inc. Address:7270 Popham Place, Solon, OH44139, United States Contact:Sean Cao Title:Sales and Marketing Director E-mail:[email protected] Tel:(440) 591 9268 Web: www.perseena.com