Emco phd virtausmittarit

Transcription

Emco phd virtausmittarit
Model Vortex PhD™
Inline Vortex Flowmeter
EMCO’s Vortex PhD™ inline vortex flow meters measure a wide variety of fluids and pipe sizes. The fully welded construction
with no internal gaskets eliminates leak paths. Since the sensor is not in the flow, it can be removed without process shut down.
Features
• Reliable–no moving parts,
no fluid to sensor contact
• In-process removable sensor,
below 750 psig (52 barg)
pressure
• Simple, rugged, all welded
construction, no internal
gaskets, no leak paths
• High accuracy with wide
rangeability
• Line sizes from 1 to 12"
(25 to 300 mm)
• High temperature operation
up to 750°F (400°C)
• EZ Logic™ menu-driven user
interface
• ‘Smart’ electronics provide
excellent noise immunity
Superior Flow Measurement Capability
The EMCO Vortex PhD™ vortex shedding flowmeter measures process fluid flow
by detecting the frequency at which vortices are shed from an obstruction in the
process fluid line (the bluff body). A piezo-electric sensor mounted outside the
flow line acts as the transducer, converting vortex pulses into electrical signals.
The ‘Smart’ electronics provide excellent noise immunity while the wide range
of wetted materials provides outstanding compatibility with most process fluids.
The transmitter provides a frequency output, scaled frequency output, totalizer
pulse or a 4 to 20 mA DC signal proportional to the flow rate.
emco
fl o w
s y s t e m s
Performance Specifications
Accuracy (Linear Ranges)
Liquid
±0.7% of flow rate
Gas and Steam
±1.0% of flow rate
Analog Output Version
Add ±0.1% of full scale
Repeatability
±0.15% of flow rate
Response Time
Adjustable from 1 to 100 seconds (analog version output)
K-factor - Temperature Correction
The K-factor of the flowmeter decreases by 0.29% for each 100°F (39°C) increase in temperature
above 70°F (21°C). The converse is true below 70°F (21°C). By programming the operating fluid
temperature in the electronics, the change in the K-Factor is automatically corrected for.
Operating Specifications
Fluid Types
Liquid, gas, or steam
Line Sizes
1 to 12" (25 to 300 mm)
Process Temperature Limit
-40 to 750°F (-40 to 400°C)
Process Pressure
Flange
ANSI Class 150, 300, or 600
PN16, 40, 64
Wafer
All water bodies are rated based on appropriate flange rating
to a maximum of ANSI Class 600
Enclosure Specifications
Approved for NEMA 4X watertight and dust tight requirements
Ambient Temperature Limit
Normal
-20 to 140°F (-29 to 60°C)
Display Option
32 to 140°F (0 to 60°C)
Ambient humidity
5 to 100% RH non-condensing
Power Requirements
18 to 40 VDC; max power consumed: 1 watt
Output
2
Analog
4 to 20 mA, loop powered 2-wire system, digitally adjusted span
Frequency
Voltage pulses, 3 wire system, 1 to 10 kHz square wave, open collector FET output with selectable
pull-up resistors. The pulse output can be scaled so that 1 pulse represents a specific quantity
of fluid passing through the flowmeter. For typical applications, the amplitude of the square wave
can be determined by:
Vpulse =Vs - (I • Rload)
For other options refer to the Vortex PhD manual.
Cable Conduit Connection
ANSI 3/4" female NPT. See Dimensions (Figure 4)
DIN PG 13.5. See Dimensions (Figure 4)
Signal Cable
18, 20, or 22 gauge, shielded instrumentation cable. Belden #9939 double-shielded cable
or equivalent required for European CE rating.
g e n e r a l
s p e c i f i c a t ions
Span Setting
For analog outputs the span setting can be adjusted without the use of electronic
test equipment by programming the full scale rate using the EZ Logic™ keypad.
Measurable Flow Velocities
Liquid
1.5 to 32 ft/sec (0.5 to 9 m/sec)
74
to 79 m/sec
ρ
50 to 260 ft/sec
ρ
Gas/Steam
Where
ρ = density in
lb  kg 
ft3  m3 
Process Viscosity
Reynolds number must be > 20,000. Figure 1 shows the minimum linear flow
velocity versus kinematic viscosity. Linear range Reynolds number of 20,000
to 7,000,000.
Ambient Temperature ˚F (˚C)
180 (82)
150 (66)
140 (60)
110 (43)
70 (22)
32 (0)
-100
(-73)
-20
(-29)
100
(38)
300
(149)
500
(260)
Process Temperature ˚F (˚C)
Figure 1. Minimum Linear Flow Velocity vs. Kinematic Viscosity.
3
emco
fl o w
s y s t e m s
Flowmeter Sizing
Pressure Loss
Selecting the appropriate size flowmeter can be critical. If the flowmeter is over-
If a differential pressure transducer
sized, the signal strength could be weak – limiting rangeability. If the flowmeter is
were connected between pressure
undersized, excessive pressure drop or cavitation may occur – affecting accuracy.
taps up and downstream of a vortex
Use the minimum linear flow velocity curves and Tables 3 through 8 (minimum and
flowmeter, the measured differential
maximum flow rates for various types of applications) for general sizing. For more
pressure would be the pressure
detailed information, contact your EMCO representative or go to www.emcoflow.
loss for the device. This value can
com for the EMCO computerized sizing program.
be calculated using the following
Nominal Size
Inch
Nominal Pulse Rate
mm
Hz/ft/s
1
25
1.5
38
equations:
Nominal Meter Factor
Hz/m/s
pulses/gal
pulses/L
pulses/ft3
pulses/m3
12.0
39.4
321.2
84.9
2,403
7.5
24.6
81.7
21.6
611
English
ΔP = 2.67 x 10 -4 · p · V2
84,864
or
ΔP = 2.50 x 10 -3 · p · (Q 1)2 / D4
21,578
or
ΔP = 4.47 x 10 -5 · p · (Q 2)2 / D4
Metric
ΔP = 1.24 x 10-5 · p · V2
or
ΔP = 5572 · p · (Q 1)2 / D4
or
ΔP = 20.1 · p · (Q 2)2 / D4
2
50
6.0
19.7
39.1
10.3
293
10,349
3
80
4.0
13.1
11.7
3.1
87.2
3,080
4
100
3.0
13.1
11.7
3.1
87.2
3,080
6
150
2.0
6.6
1.5
0.4
11.0
388
8
200
1.3
4.3
0.6
0.2
4.2
148
10
250
1.2
3.9
0.3
0.1
2.2
78
12
300
1.0
3.3
0.2
0.1
1.3
46
Table 1. Nominal Pulse Rate and Flowmeter Factor.
Nominal Size
Flow Range in GPM
Flow Range in L/s
Inch
mm
Hz/ft/s
Hz/m/s
min
max
1
25
3.4
72
0.2
4.5
1.5
40
8.3
176
0.5
11.1
2
50
13.8
295
0.9
18.6
3
80
30.9
659
1.9
41.6
4
100
53.8
1,147
3.4
72.3
6
150
121.9
2,600
7.7
163.9
8
200
213.5
4,555
13.5
287.2
10
250
358.7
7,653
22.6
482.5
12
300
508.3
10,843
32.0
683.7
Table 2. Water Flow Rates.
Notes:
1
Water values are determined at standard conditions of 60°F (15°C).
2
The minimum flow rate for the 1” (25 mm), and 1.5” (40 mm) flowmeters are below the
linear range - the Reynolds number < 20,000.
3
The velocities used to calculate the minimum and maximum flow rates were 1.5 ft/sec
and 32 ft/sec (0.5 m/sec and 9 m/sec) respectively.
4
Where:
ΔP
=
Pressure loss psi (bar)
p
=
Density at operating conditions
V
=
Flow velocity in ft/s (m/s)
Q1
=
Volumetric flow rate in ACFM
lb/ft3 (kg/m3)
(Actual m3/min)
Q2
=
D
=
Volumetric flow rate in U.S.
GPM (L/s)
Flowmeter internal diameter
in inches (mm)
g e n e r a l
Physical Specifications
Minimum Back Pressure
(Liquid Service Only)
The line pressure must be sufficiently
high so that no cavitation occurs. The
minimum required line pressure can
be obtained from the equation:
Materials
Wetted Parts
Wing and Shedder Bar
Stainless steel and Carbon steel Models 316L or the cast
equivalent, CF3M
Body
Stainless steel Model (all sizes) 316L or the cast equivalent,
CF3M
PG≥(1 · 3 · Po) + (2.7 · ΔP) - PATM
Where:
PG
=
Line pressure required psig
(barg)
ΔP
=
Maximum pressure loss psig
(barg)
Po
=
Saturation liquid vapor pressure
at operating temperature psia
(bara)
PATM =
s p e c i f i c a t ions
Atmospheric pressure psia (bara)
Carbon Steel Model (flanged), 6” (150 mm) and larger;
ASTM A105, A06 Grade C, and A108 Grade 1018
Electronics Enclosure
383 Aluminum
All Other Parts
Stainless Steel
Flowmeter Classification
Standard
Designed to meet NEMA 4X watertight and dust tight
specifications
Explosion Proof Version
FM and CSA approved
Explosion Proof and
Intrinsically-Safe Version
CENELEC approved
European CE Mark
Approved
Example:
Water flow rate: 0 to 200 US GPM
Remote Mount Electronics (RMT Option)
Operating temperature: 100°F
Signal cable is provided to mount the electronics enclosure up to 30 ft (optional 50
Operating pressure: 10 psig
ft) from the flowmeter body, and must be installed within proper electrical conduit.
Atmospheric pressure: 14.7 psia
Flowmeter size: 1.939” I.D.
Local Indicator and Totalizer (LOC-TOT Option)
Specific weight: 61.96 lb/ft3
The indicator displays flow rate and total in user-selectable, engineering units. It is
a two line, 8 characters per line, LCD display. The totalizer enables a pulse output
Solution:
ΔP
=
4.47 x 10 · 61.96 · (200)
-5
2
providing a 5 to 999 msec pulse each time the totalizer increments.
/(1.939) = 7.84
4
PG
=
[(1.3) · (0.951)] + [(2.7) · (7.84)]
PG
=
7.70 psig
- 14.7
Since the operating pressure of 10 psig
This option changes the lower, ambient temperature limit to 32°F (0°C).
Below 32°F, the display intensity begins to fade but does not affect the
transmitter output.
is higher than 7.70 psig, no cavitation
occurs.
Material of Construction
All models are available in stainless steel; some models can be ordered in carbon steel.
5
emco
fl o w
s y s t e m s
FM Approval (Explosion Proof)
Straight Run Piping Requirements
Class I, Div. I, Groups B, C, D
Class II, Div. I, Groups E, F, G
Class I, Div. II, Groups A, B, C, D
Class II, III, Div. 2, Groups F, G
CSA Approval (Explosion Proof)
Class I, Div. I, Groups B, C, D
Upstream
Downstream
One 90° elbow before the meter
10 D
5D
Two 90° elbows before the meter
15 D
5D
Two 90° elbows out of plane before
the meter
30 D
5D
Reduction before the meter
10 D
5D
Regulator or valve partially closed
before the meter
30 D
5D
Recommendation: typically 10 diameters upstream, 5 diameters downstream. In
certain applications where multiple bends or butterfly valves are present, more
Dust-ignition Proof for Class II, Div. 1
straight run may be required.
Groups E, F, G
Other Installation Considerations
Class III Hazardous locations
Installation
Flowmeter can be installed vertically, horizontally, or at any angle. For liquid
CENELEC Approval
service, the flow line must be full during flowmeter operation – flow up in vertical
(Explosion Proof and Intrinsically
lines is recommended.
Safe)
Site Selection
EEX d [ib] IIC T6 and/or EEX d IIC
The flow measurement location should be selected to minimize turbulence and swirl.
T6.
The extent of these flow disturbances depends upon the piping configuration. Valves,
elbows, pumps and other piping components may add disturbances to the flow.
Accessories
Flow Processor
Serviceability
A microprocessor-based flow
The sensor element can be removed, and replaced, without removing the flowmeter
processor can be used to significantly
body from the process line, and without process shut down - for process pressures
increase the accuracy and
below 750 psig (52 bar). The flowmeter K-factor is not affected by sensor or
functionality of any flowmetering
electronics servicing.
application. See the FP-93 General
Specifications for complete details.
Pressure and Temperature Taps
Pressure tap should be mounted 3.5 to 7 pipe diameters downstream from the
flowmeter. The temperature tap should be mounted an additional 3.5 to 7 pipe
diameters downstream of the pressure tap. See Figure 2.
6
g e n e r a l
s p e c i f i c a t ions
3.5 to 7
3.5 to 7
internal diameters
internal diameters
112.4
gal/min
FLOW
Pressure
Transmitter
Temperature
Transmitter
Figure 2. Pressure and Temperature Tap Dimensions.
7
emco
fl o w
1"
s y s t e m s
1.5"
2"
3"
4"
6"
8"
10"
12"
Pres
PSIG
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
10
15
20
30
40
50
60
70
80
90
100
120
140
160
180
200
220
250
300
350
400
450
500
600
31
34
37
41
45
49
53
56
59
62
64
70
74
79
83
87
91
96
105
113
120
127
134
147
283
337
390
494
597
699
801
901
1002
1102
1201
1399
1597
1794
1991
2188
2385
2680
3175
3672
4173
4679
5189
6225
77
84
90
102
112
121
129
137
145
152
158
171
183
194
204
214
223
237
257
277
295
313
329
361
697
828
958
1214
1468
1719
1968
2215
2462
2707
2952
3439
3925
4409
4893
5377
5861
6588
7803
9025
10256
11498
12751
15299
129
140
151
170
187
202
216
229
242
253
265
286
305
323
341
357
373
395
430
463
493
522
550
602
1164
1383
1600
2029
2453
2872
3289
3702
4114
4524
4933
5747
6559
7369
8177
8986
9795
11009
13040
15082
17140
19215
21310
25568
288
313
337
380
417
452
483
513
541
567
592
639
683
723
762
799
834
884
962
1035
1103
1168
1230
1348
2604
3094
3580
4539
5486
6424
7356
8281
9202
10120
11034
12855
14670
16482
18291
20099
21908
24624
29166
33734
38337
42978
47665
57016
501
546
587
661
727
786
841
893
941
987
1031
1112
1188
1260
1327
1391
1452
1540
1676
1802
1921
2034
2142
2346
4534
5387
6233
7903
9552
11186
12807
14419
16022
17619
19211
22382
25543
28697
31846
34994
38143
42873
50781
58735
66748
72657
73548
75244
1135
1237
1331
1498
1647
1783
1908
2024
2134
2237
2336
2522
2694
2855
3008
3153
3292
3490
3798
4085
4355
4611
4856
5523
10277
12213
14129
17915
21655
25357
29033
32686
36321
39942
43550
50740
57904
65054
72193
79330
86469
97191
104535
106259
107855
109388
110729
117628
1988
2167
2331
2624
2885
3122
3341
3545
3737
3919
4092
4417
4718
5001
5268
5523
5766
6113
6653
7155
7627
8076
8505
9316
18000
21390
24746
31377
37927
44412
50849
57248
63614
69956
76275
88868
101416
113938
126441
132583
133935
135625
138329
140610
142723
144751
146525
149906
3341
3642
3918
4412
4850
5249
5616
5959
6282
6587
6878
7424
7931
8407
8856
9283
9692
10275
11183
12027
12821
13575
14296
15659
30258
35955
41597
52742
63753
74654
85474
96231
106932
117593
128214
149382
166214
168295
170158
171911
173664
175855
179362
182320
185059
187689
189990
194372
4733
5160
5550
6249
6870
7435
7955
8441
8898
9331
9743
10517
11235
11908
12545
13150
13729
14555
15841
17036
18161
19229
20251
22182
42860
50930
58923
74710
90307
105748
121075
136312
151470
166572
181617
195065
197804
200281
202498
204584
206670
209278
213450
216971
220231
223360
226098
231314
For EMCO Flowmeter Sizing download sizing program “emcosize” at www.emcoflow.com
Table 3. Saturated Steam Flow Rates at Selected Process Pressures (English).1,2
(Minimum and Maximum Operating Flow Rates in lb/hr)
25mm
40mm
50mm
80mm
100mm
150mm
200mm
250mm
300mm
Pres
BARG
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
0.7
1.0
1.4
2.1
2.8
3.4
4.1
4.8
5.5
6.2
6.9
8.3
9.7
11.0
12.4
13.8
15.2
17.2
20.7
24.1
27.6
31.0
34.5
41.4
14
15
17
19
21
22
24
25
27
28
29
32
34
36
38
39
41
44
48
51
54
58
61
67
129
153
177
224
271
317
363
409
454
500
545
635
724
814
903
992
1082
1216
1440
1666
1893
2122
2354
2824
35
38
41
46
51
55
59
62
66
69
72
78
83
88
92
97
101
107
117
126
134
142
149
164
316
375
434
551
666
780
893
1005
1117
1228
1339
1560
1780
2000
2220
2439
2658
2988
3539
4094
4652
5215
5784
6940
58
64
68
77
85
92
98
104
110
115
120
130
138
147
155
162
169
179
195
210
224
237
249
273
528
627
726
920
1113
1303
1492
1679
1866
2052
2238
2607
2975
3342
3709
4076
4443
4994
5915
6841
7774
8716
9666
11597
130
142
153
172
189
205
219
233
245
257
268
290
310
328
346
362
378
401
437
469
500
530
558
611
1181
1403
1624
2059
2489
2914
3336
3756
4174
4590
5005
5831
6654
7476
8296
9117
9937
11169
13229
15302
17389
19495
21620
25862
227
248
266
300
330
357
382
405
427
448
467
505
539
571
602
631
659
698
760
817
871
923
972
1064
2056
2444
2827
3585
4333
5074
5809
6540
7267
7992
8714
10152
11586
13017
14445
15873
17302
19447
23034
26642
30276
32957
33361
34130
515
561
604
680
747
809
865
918
968
1015
1060
1144
1222
1295
1364
1430
1493
1583
1723
1853
1975
2092
2203
2505
4662
5540
6409
8126
9822
11502
13169
14826
16475
18117
19754
23015
26265
29508
32746
35984
39222
44085
47416
48198
48922
49617
50226
53355
902
983
1057
1190
1309
1416
1515
1608
1695
1778
1856
2003
2140
2268
2390
2505
2615
2773
3018
3245
3460
3663
3858
4226
8165
9702
11225
14232
17203
20145
23065
25967
28855
31732
34598
40310
46002
51681
57353
60139
60752
61518
62745
63780
64738
65658
66463
67996
1516
1652
1777
2001
2200
2381
2547
2703
2849
2988
3120
3368
3598
3813
4017
4211
4396
4661
5073
5455
5816
6158
6485
7103
13725
16309
18868
23924
28918
33862
38771
43649
48503
53339
58157
67759
75393
76337
77182
77978
78773
79767
81357
82699
83941
85134
86178
88166
2147
2340
2517
2835
3116
3372
3608
3829
4036
4232
4419
4770
5096
5401
5690
5965
6227
6602
7185
7728
8238
8722
9186
10061
19441
23102
26727
33888
40962
47966
54919
61830
68706
75556
82380
88480
89722
90846
91851
92798
93744
94927
96820
98416
99895
101314
102557
104922
For EMCO Flowmeter Sizing download sizing program “emcosize” at www.emcoflow.com
Table 4. Saturated Steam Flow Rates at Selected Process Pressures (Metric).1,2
(Minimum and Maximum Operating Flow Rates in kg/hr)
1
Maximum flow velocity is 260 ft/sec (79 m/sec) or Reynolds number limit or 7,000,000 whichever is less.
2
Minimum flow velocity is
whichever is greater.
8
lb
50 ft/sec where ρ = density in
ft3
ρ
kg
74
ρ m/sec where ρ = density in m3
or Reynolds number limit of 20,000
g e n e r a l
1"
1.5"
2"
3"
4"
s p e c i f i c a t ions
6"
8"
10"
12"
Pres
PSIG
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
0
10
20
30
40
50
60
70
80
90
100
120
140
160
180
200
220
250
300
350
400
450
500
600
711
711
711
802
888
966
1038
1105
1169
1229
1287
1395
1495
1589
1678
1762
1843
1958
2136
2301
2454
2599
2737
2993
4675
7858
11043
14229
17417
20606
23798
26990
30185
33381
36578
42978
49384
55795
62213
68635
75063
84715
100825
116963
133125
149310
165512
197961
1130
1465
1737
1971
2181
2372
2550
2715
2871
3020
3161
3426
3673
3904
4122
4330
4528
4810
5248
5652
6030
6386
6724
7353
11486
19306
27129
34957
42789
50624
58464
66308
74156
82008
89863
105586
121323
137075
152840
168619
184411
208123
247701
287347
327054
366814
389852
389852
1888
2448
2902
3294
3645
3964
4260
4537
4798
5046
5282
5725
6137
6523
6888
7235
7566
8038
8769
9445
10076
10671
11235
12287
19194
32260
45332
58412
71499
84593
97693
110800
123914
137034
150160
176433
202729
229050
255394
281760
308148
347770
413905
480154
503949
503949
503949
503949
4224
5476
6491
7369
8152
8868
9529
10149
10732
11286
11814
12806
13728
14592
15408
16184
16925
17980
19615
21127
22539
23870
25132
27485
42934
72160
101403
130661
159934
189223
218527
247845
277179
306527
335889
394656
453478
512354
571282
630260
689287
753714
753714
753714
753714
753714
753714
753714
7352
9531
11299
12826
14190
15435
16587
17665
18681
19645
20564
22291
23894
25398
26819
28169
29459
31295
34141
36772
39231
41547
43743
47840
74730
125601
176499
227425
278378
329357
380363
431394
482452
533534
584642
686931
789315
891794
994362
994383
994383
994383
994383
994383
994383
994383
994383
994383
16669
21611
25618
29080
32173
34995
37607
40050
42354
44540
46625
50539
54175
57584
60805
63867
66791
70955
77408
83373
88947
94199
99179
108466
169433
284772
400174
515637
631161
746746
862390
978094
1093855
1209674
1325549
1497292
1497292
1497292
1497292
1497292
1497292
1497292
1497292
1497292
1497292
1497292
1497292
1497292
29201
37857
44877
50942
56360
61304
65880
70160
74196
78025
81677
88534
94903
100876
106519
111882
117004
124299
135604
146053
155818
165018
173741
190010
296813
498864
701024
903292
1105667
1308148
1510734
1713422
1916212
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
1981748
49087
63722
75490
85473
94572
102870
110546
117723
124487
130782
136902
148387
158946
168847
177936
187249
195325
207975
226127
243566
259241
274584
288580
315201
498888
838268
1177647
1517027
1856406
2195786
2535166
2576801
2577147
2572706
2573347
2574344
2571893
2570006
2560955
2571876
2560024
2573422
2558874
2561760
2552194
2555165
2548126
2545390
69532
90263
106933
121073
133962
145716
156590
166756
176337
185254
193923
210192
225149
239173
252048
265241
276680
294599
320311
345014
367217
388951
408777
446486
706680
1187415
1668149
2148884
2629619
3065336
3066018
3066539
3066950
3061665
3062428
3063615
3060698
3058452
3047681
3060678
3046573
3062517
3045204
3048639
3037255
3040791
3032414
3029158
For EMCO Flowmeter Sizing download sizing program “emcosize” at www.emcoflow.com
Table 5. Air Flow Rate at Selected Process Pressures (Metric).1,2,3
(Minimum and Maximum Operating Flow Rates in SCFH)
25mm
40mm
50mm
80mm
100mm
150mm
200mm
250mm
300mm
Pres
BARG
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
0.0
0.7
1.4
2.1
2.8
3.4
4.1
4.8
5.5
6.2
6.9
8.3
9.7
11.0
12.4
13.8
15.2
17.2
20.7
24.1
27.6
31.0
34.5
41.4
19
19
19
21
24
26
28
30
31
33
34
37
40
43
45
47
49
52
57
62
66
70
73
80
125
210
296
381
466
552
637
723
808
894
980
1151
1323
1494
1666
1838
2011
2269
2701
3133
3566
3999
4433
5302
30
39
47
53
58
64
68
73
77
81
85
92
98
105
110
116
121
129
141
151
162
171
180
197
308
517
727
936
1146
1356
1566
1776
1986
2197
2407
2828
3250
3671
4094
4516
4939
5574
6634
7696
8760
9825
10442
10442
51
66
78
88
98
106
114
122
129
135
141
153
164
175
184
194
203
215
235
253
270
286
301
329
514
864
1214
1565
1915
2266
2617
2968
3319
3670
4022
4726
5430
6135
6840
7547
8253
9315
11086
12860
13498
13498
13498
13498
113
147
174
197
218
238
255
272
287
302
316
343
368
391
413
433
453
482
525
566
604
639
673
736
1150
1933
2716
3500
4284
5068
5853
6638
7424
8210
8996
10571
12146
13723
15301
16881
18462
20188
20188
20188
20188
20188
20188
20188
197
255
303
344
380
413
444
473
500
526
551
597
640
680
718
754
789
838
914
985
1051
1113
1172
1281
2002
3364
4727
6091
7456
8822
10188
11555
12922
14290
15659
18399
21141
23886
26633
26634
26634
26634
26634
26634
26634
26634
26634
26634
446
579
686
779
862
937
1007
1073
1134
1193
1249
1354
1451
1542
1629
1711
1789
1900
2073
2233
2382
2523
2656
2905
4538
7627
10718
13811
16905
20001
23098
26197
29298
32400
35504
40104
40104
40104
40104
40104
40104
40104
40104
40104
40104
40104
40104
40104
782
1014
1202
1364
1510
1642
1765
1879
1987
2090
2188
2371
2542
2702
2853
2997
3134
3329
3632
3912
4173
4420
4653
5089
7950
13362
18776
24194
29614
35038
40464
45892
51324
53079
53079
53079
53079
53079
53079
53079
53079
53079
53079
53079
53079
53079
53079
53079
1313
1705
2020
2287
2531
2753
2958
3150
3331
3500
3663
3971
4253
4518
4762
5011
5227
5566
6051
6518
6937
7348
7723
8435
13344
22426
31508
40591
49673
58755
67837
68952
68962
68844
68861
68889
68824
68774
68532
68824
68507
68866
68477
68555
68299
68378
68190
68117
1860
2415
2861
3240
3584
3899
4190
4462
4719
4957
5189
5625
6025
6400
6745
7098
7404
7884
8572
9233
9827
10409
10939
11948
18902
31767
44632
57497
70362
82023
82042
82057
82069
81928
81949
81982
81904
81845
81557
81905
81528
81954
81492
81584
81279
81374
81150
81063
For EMCO Flowmeter Sizing download sizing program “emcosize” at www.emcoflow.com
Table 6. Air Flow Rate at Selected Process Pressures (Metric).1,2,3
(Minimum and Maximum Operating Flow Rates in NCMH)
1
2
3
At standard conditions of 60°F and 14.7 psia, SCFH: standard cubic feet per hour (0°C and 1.013 bar, NCMH: Normal cubic meters per hour).
Process temperature is 60°F (16°C).
Maximum flow velocity is 260 ft/sec (79 m/s) or Reynolds number limit or 7,000,000 whichever is less.
Minimum flow velocity is
whichever is greater.
lb
50 ft/sec where ρ = density in
ft3
ρ
kg
74
ρ m/sec where ρ = density in m3
or Reynolds number limit of 20,000
9
emco
fl o w
1"
s y s t e m s
1.5"
2"
3"
4"
6"
8"
10"
12"
Pres
PSIG
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
0
10
20
30
40
50
60
70
80
90
100
120
140
160
180
200
220
250
300
350
400
450
500
600
772
772
905
1028
1138
1238
1332
1419
1502
1580
1655
1797
1929
2053
2171
2283
2391
2546
2787
3013
3226
3430
3625
3995
4665
7851
11048
14255
17472
20700
23938
27187
30446
33715
36995
43587
50222
56900
63622
70388
77198
87496
104888
122567
140539
158808
177379
215439
1444
1873
2222
2524
2795
3042
3271
3486
3689
3882
4067
4414
4738
5043
5333
5609
5874
6254
6847
7402
7926
8426
8905
9814
11460
19289
27142
35021
42925
50855
58810
66791
74797
82829
90887
107082
123382
139789
156302
172924
189654
214956
257682
301115
345267
390150
423533
423533
2413
3130
3713
4218
4670
5083
5466
5825
6165
6487
6795
7376
7917
8427
8911
9373
9816
10450
11442
12369
13245
14079
14880
16398
19149
32231
45354
58520
71728
84978
98271
111606
124985
138407
151872
178933
206171
233585
261179
288953
316909
359188
430583
503159
547487
547487
547487
547487
5397
7002
8307
9436
10446
11370
12227
13030
13789
14511
15200
16499
17710
18851
19933
20967
21957
23376
25594
27667
29626
31493
33284
36681
42834
72096
101452
130901
160445
190084
219818
249648
279575
309597
339717
400250
461176
522500
584223
646350
708885
803456
818830
818830
818830
818830
818830
818830
9395
12188
14458
16423
18182
19791
21282
22680
24001
25257
26457
28718
30826
32812
34696
36494
38219
40688
44549
48157
51567
54816
57933
63846
74557
125490
176585
227844
279268
330857
382611
434533
486622
538879
591305
696667
802714
909452
1016887
1080291
1080291
1080291
1080291
1080291
1080291
1080291
1080291
1080291
21300
27634
32781
37236
41225
44871
48253
51423
54418
57265
59986
65112
69892
74394
78665
82742
86652
92252
101005
109186
116917
124284
131350
144757
169041
284520
400368
516588
633180
750146
867489
985209
1103309
1221791
1340656
1579542
1626648
1626648
1626648
1626648
1626648
1626648
1626648
1626648
1626648
1626648
1626648
1626648
37314
48410
57426
65230
72217
78605
84530
90083
95329
100317
105084
114062
122436
130323
137805
144947
151797
161606
176940
191271
204815
217720
230098
253585
296126
498422
701364
904957
1109202
1314104
1519664
1725887
1932775
2140331
2152958
2152958
2152958
2152958
2152958
2152958
2152958
2152958
2152958
2152958
2152958
2152958
2152958
2152958
64968
84286
99983
113570
125733
136853
147166
156833
165965
174647
182943
198569
213142
226866
239886
252312
264230
281293
307963
332885
356431
378865
400376
441178
511367
860683
1211101
1562626
1915261
2269014
2623888
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
2890521
93181
120888
143401
162889
180334
196283
211075
224938
238036
250488
262387
284500
305701
325384
344058
361881
378973
403446
441698
477442
511214
543389
574242
632763
733432
123441
1737031
2241207
2746978
2746978
3254350
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
3461702
For EMCO Flowmeter Sizing download sizing program “emcosize” at www.emcoflow.com
Table 7. Natural Gas Flow Rates at Selected Process Pressures (English).1,2,3,4
(Minimum and Maximum Operating Flow Rates in SCFH)
25mm
40mm
50mm
80mm
100mm
150mm
200mm
250mm
300mm
Pres
BARG
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
min
max
0.0
0.7
1.4
2.1
2.8
3.4
4.1
4.8
5.5
6.2
6.9
8.3
9.7
11.0
12.4
13.8
15.2
17.2
20.7
24.1
27.6
31.0
34.5
41.4
21
21
24
28
30
33
36
38
40
42
44
48
52
55
58
61
64
68
75
81
86
92
97
107
125
210
296
382
468
554
641
728
815
903
991
1167
1345
1524
1704
1885
2068
2344
2809
3283
3764
4254
4751
5770
39
50
60
68
75
81
88
93
99
104
109
118
127
135
143
150
157
168
183
198
212
226
239
263
307
517
727
938
1150
1362
1575
1789
2003
2219
2434
2868
3305
3744
4186
4632
5080
5757
6902
8065
9248
10450
11344
11344
65
84
99
113
125
136
146
156
165
174
182
198
212
226
239
251
263
280
306
331
355
377
399
439
513
863
1215
1567
1921
2276
2632
2989
3348
3707
4068
4793
5522
6256
6995
7739
8488
9621
11533
13477
14664
14664
14664
14664
145
188
222
253
280
305
327
349
369
389
407
442
474
505
534
562
588
626
686
741
794
844
891
982
1147
1931
2717
3506
4297
5091
5888
6687
7488
8292
9099
10720
12352
13995
15648
17312
18987
21520
21932
21932
21932
21932
21932
21932
252
326
387
440
487
530
570
607
643
676
709
769
826
879
929
977
1024
1090
1193
1290
1381
1468
1552
1710
1997
3361
4730
6103
7480
8862
10248
11639
13034
14433
15838
18660
21500
24359
27236
28935
28935
28935
28935
28935
28935
28935
28935
28935
571
740
878
997
1104
1202
1292
1377
1458
1534
1607
1744
1872
1993
2107
2216
2321
2471
2705
2924
3132
3329
3518
3877
4528
7621
10724
13836
16959
20092
23235
26388
29551
32725
35908
42307
43568
43568
43568
43568
43568
43568
43568
43568
43568
43568
43568
43568
999
1297
1538
1747
1934
2105
2264
2413
2553
2687
2815
3055
3279
3491
3691
3882
4066
4328
4739
5123
5486
5831
6163
6792
7931
13350
18785
24238
29709
35197
40703
46226
51768
57327
57665
57665
57665
57665
57665
57665
57665
57665
57665
57665
57665
57665
57665
57665
1743
2269
2695
3063
3393
3652
3934
4199
4448
4684
4910
5336
5732
6079
6434
6772
7097
7539
8266
8925
9566
10154
10745
11839
13711
23220
32759
42328
51929
60182
69839
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
77710
2500
3254
3865
4393
4866
5238
5643
6022
6379
6719
7043
7653
8221
8719
9228
9713
10178
10813
11855
12801
13720
14571
15411
16981
19655
33303
46985
60710
74479
86316
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
93066
For EMCO Flowmeter Sizing download sizing program “emcosize” at www.emcoflow.com
Table 8. Natural Gas Flow Rates at Selected Process Pressures (Metric).1,2,3,4
(Minimum and Maximum Operating Flow Rates in NCMH)
1
2
3
4
10
At standard conditions of 60°F and 14.7 psia, SCFH: standard cubic feet per hour (0°C and 1.013 bar, NCMH: Normal cubic meters per hour).
Process temperature is 60°F (16°C).
Maximum flow velocity is 260 ft/sec (79 m/s) or Reynolds number limit or 7,000,000 whichever is less.
Specific Gravity of natural gas = 0.61 and 0.8% N2.
Minimum flow velocity is
whichever is greater.
lb
50 ft/sec where ρ = density in
ft3
ρ
kg
74
ρ m/sec where ρ = density in m3
or Reynolds number limit of 20,000
g e n e r a l
Dimensions and Weights
1.00"
(25 mm)
4.50"
(114 mm)
3.12"
(79 mm)
s p e c i f i c a t ions
3.62"
(92 mm)
4.90"
(124 mm)
A
¾" NPT
(PG 13.5)
MAX.
B
D
C
Figure 3. Dimensions and Weights: Wafer Connection Type, Integral Mounting.
Wafer Connection
Wafer connection is available in stainless steel (sizes 1 to 4" only). The schedule of the mating pipe’s internal diameter
≥ dimension “D”.
Size
1" (25 mm)
1.5" (40 mm)
2" (50 mm)
3" (80 mm)
4" (100 mm)
A
10.8" (274 mm)
10.5" (267 mm)
10.8" (274 mm)
11.5" (292 mm)
12.2" (310 mm)
B
2.2" (55.9 mm)
3.1" (78.5 mm)
3.6" (91.9 mm)
5.0" (127 mm)
6.2" (157.2 mm)
C
4.1" (104.6 mm)
4.1" (104.6 mm)
5.0" (127.0 mm)
7.0" (177.8 mm)
9.5" (241.3 mm)
D
0.957" (24.3 mm)
1.500" (38.1 mm)
1.939" (49.3 mm)
2.900" (73.7 mm)
3.826" (97.2 mm)
Approximate Weight
14 lb (6.4 kg)
14 lb (6.4 kg)
19 lb (8.6 kg)
35 lb (15.9 kg)
51 lb (23.2 kg)
11
emco
3.12"
(79 mm)
fl o w
3.62"
(92 mm)
s y s t e m s
4.50"
(114 mm)
1.00"
(25 mm)
4.90"
(124 mm)
A
3/4" NPT
MAX.
(PG 13.5)
Display
(Optional)
C
B
Figure 4. Dimensions and Weights: Flange Connection, Integral Mounting.
Size (in)
1
1.5
2
3
4
6
8
10
12
A
10.7
10.7
13.2
13.8
14.3
15.3
16.3
17.4
18.4
B
7.6
8.1
8.5
9.0
9.5
13.6
18.5
18.5
18.5
150# 300#
0.957
1.500
1.939
2.900
3.826
5.761
7.625
10.020
12.000
600#
0.957
1.500
1.939
2.900
3.826
5.761
7.625
9.750
11.750
150#, 300#
N/A
N/A
N/A
N/A
N/A
5.761
7.625
10.020
11.938
600#
N/A
N/A
N/A
N/A
N/A
5.761
7.625
9.750
11.750
150#
18
22
31
51
55
92
144
180
265
300#
20
28
36
60
72
116
182
260
365
600#
20
28
36
60
99
140
220
440
535
Size (mm)
25
40
50
80
100
150
200
250
300
A
272
272
335
350
363
389
414
442
467
B
192.0
206.2
215.9
228.6
241.3
346.2
469.6
469.9
469.9
PN 16, PN 40
24.3
38.1
49.3
73.7
97.2
146.3
193.7
254.4
304.8
PN 64
N/A
N/A
49.3
73.7
97.2
146.3
193.7
247.7
298.5
PN 16, PN 40
N/A
N/A
N/A
N/A
N/A
146.3
193.7
245.5
303.2
PN 64
N/A
N/A
N/A
N/A
N/A
146.3
193.7
247.7
288.9
PN 16
8.2
10.0
14.1
23.1
25.0
41.7
65.3
81.6
120.2
PN 40
9.1
12.7
16.3
27.2
32.7
53.6
82.6
117.9
165.6
PN 64
N/A
N/A
16.3
27.2
44.9
63.5
99.8
199.6
242.7
Stainless Steel
C
Carbon Steel
Approzimate Weight (lb)
Stainless Steel
C
Carbon Steel
Approzimate Weight (kg)
1
2
12
The schedule of the mating pipe’s internal diameter ≥ dimension ‘C’.
N/A = Not Available.
g e n e r a l
Pipe Mount Electronics
s p e c i f i c a t ions
OR
Wall Mount Electronics
Note: For best results mount
remote transmitter below
horizontal pipe.
5"
(127 mm)
Dia. 0.281"
(7.14 mm)
4.20"
(106.7 mm)
0.4"
(10.2 mm)
6.5"
(165.1 mm)
8.5"
(215.9 mm)
8.5"
(215.9 mm)
8.5"
(215.9 mm)
5"
(127 mm)
30' Cable (9.144 m) and U-bolts supplied.
Note: Cable must be run in conduit (not supplied).
Conduit connection is ¾" NPT (PG 13.5).
7.70"
(195.6 mm)
5"
(127 mm)
5"
0.4"
(10.2 mm)
Figure 5. Dimensions and Weights: Remote Mounting.
Size
1"
1.5"
2"
3"
4"
6"
8"
10"
12"
150# Flange
24 lb
28 lb
37 lb
57 lb
72 lb
98 lb
150 lb
186 lb
271 lb
300# Flange
26 lb
34 lb
42 lb
66 lb
89 lb
122 lb
188 lb
266 lb
371 lb
600# Flange
26 lb
34 lb
42 lb
66 lb
105 lb
146 lb
226 lb
556 lb
541 lb
Wafer
19 lb
20 lb
23 lb
38 lb
57 lb
N/A
N/A
N/A
N/A
Size
25 mm
40 mm
50 mm
80 mm
100 mm
150 mm
200 mm
250 mm
300 mm
PN 16 Flange
10.9 kg
2.7 kg
16.8 kg
25.9 kg
32.7 kg
44.5 kg
68.5 kg
91.6 kg
130.2 kg
PN 40 Flange
211.8 kg
15.5 kg
19.1 kg
30.0 kg
40.5 kg
55.5 kg
85.5 kg
127.9 kg
175.5 kg
PN 64 Flange
211.8 kg
15.5 kg
19.1 kg
30.0 kg
47.7 kg
66.4 kg
102.7 kg
209.6 kg
252.7 kg
Wafer
8.6 kg
9.1 kg
10.4 kg
17.2 kg
25.9 kg
N/A
N/A
N/A
N/A
13
emco
fl o w
s y s t e m s
Wiring
Pulse Output
Figures 6 through 8 represent typical field wiring applications for the Vortex PhD.
This is an open collector pulse output
using a high
Analog Output
The Vortex PhD may be operated using a 24 VDC power supply. The 4 to 20 mA
impedance electronic counter. Vpulse
output is scalable (2 wire principle). The load resistor may be installed on supply
will vary from:
or return line. Permissible load resistance values are shown in Figure 6.
Without LOC–TOT Option
With LOC–TOT Option
1250
Max. Load Resistance (Ω)
1100


RC
0 − 1V to Vpulse = VS 

 R C + R pulse 


RC
0 − 1V to Vpulse = VS 
S 
 V
Rpulse
Note :  R
≥ pulse

C +R
 0.16 
V
Note : R pulse ≥  S 
 0.16 
1000
Where:
750
LOAD = 50 (Vs – 18)
500
Vpulse =
pulse output amplitude
RC
=
counter impedance
VS
=
power supply voltage
VC
=
minimum required voltage to
trip counter
250
Permissable Load
Resistance Range
0
0
5
10 12
15
18
25
30
35
40
P
–
Power Supply Voltage (VDC)
Rpulse
+
–
Vpulse 123456789
Figure 7. Pulse Output.
+
P
–
Rload
–
Figure 6. 4 to 20 mA Output.
14
Vs
(18-40 VDC)
+
–
Vs
(18-40 VDC)
g e n e r a l
s p e c i f i c a t ions
Cenelec Approved– Power & Signal Wiring
The Vortex PhD flowmeter may be operated using a 24 VDC power supply. The
wiring schematic shown is for the areas defined by the Cenelec approval (see
approval description in options section). I. S. zener barriers must have the following
specifications:
Vmax =
29 VDC
=
Maximum voltage which may be connected to barrier terminals
Jmax =
110 mA
=
Maximum current which may be connected to barrier terminals
Cint =
1.5 nF
=
Maximum equivalent internal capacitance across barrier terminals
L int =
0H
=
Maximum equivalent internal inductance across barrier terminals
The 4 to 20 mA output is scalable (2 wire principle.) The 4 to 20 mA loop resistor
should be 250 Ω; and the pulse output is to be connected to a high impedance,
electronic counter.
Vpulse varies from: 0 -1 VDC to Vpulse = Vs
Rpulse ≥
(
Rc
Rc+Rpulse
)
Vs
0.16
Where:
Vpulse
= Pulse output amplitude
Vs
= Power supply voltage (18 to 29 VDC)
R load
= Load resistance
Rc
= Counter impedance
Rpulse = Pull–up resistance
Max. Load Resistance (Ohms)
600
Hazardous
Area
550
500
Rload
400
+
Rpulse
LOAD = 50 (Vs – 18)
300
P
200
–
Permissable Load
Resistance Range
100
0
Safe Area or Zone 2
0
18
20
24
+
Vpulse
–
123456789
+
–
VS
(18-28.5 VDC)
Zener Barrier
29
Power Supply Voltage (VDC)
Figure 8. CENELEC Approved (Intrinsically Safe).
15
emco
fl o w
s y s t e m s
110/220 VAC– Power & Signal Wiring
The Vortex PhD flowmeter may be operated using a 110/220 VAC power supply. The
power supply board converts the 110/220 VAC to 24 VDC. The 4 to 20 mA output is
scalable (2 wire principle.) The 4 to 20 mA loop resistor should be 250Ω; and the
pulse output is to be connected to a high impedance, electronic counter. (Figure 9)
Vpulse varies from 0–1 VDC to Vpulse = 24 VDC – (I · R load)
Where:
Vpulse =
Pulse output amplitude
R load =
Load resistance (250Ω)
I
Current (4 to 20 mA)
=
L
AC
PULSE
+
123456789
Vpulse
N
ACC
Analog
(–)
GND
Analog
(+)
GND
–
Rload
Figure 9. Frequency/4 to 20 mA Output using 110/220 VAC power.
16
Transit
Puls unt
ft3
Per puls
10.0000
Pulse +
Puls unt
ft3
Per puls
10.0000
Puls wth
50 mSec
Pulse –
Puls unt
ft3
Per puls
10.0000
Puls wth
50 mSec
Freq TC
05 Sec
10 kHz
Freq TC
05 Sec
1.5
Viscos
1 cP
Temp
0100˚F
Direct
Off
Total
Reset N
Defaults
No
menu C
Pipe ID
1 in.
Meter K
2400.00
000000 –
000 WO#
Tag #
00000000
Reset
menu D
Sensor
Fluid
Density
62.4000
Ref Den
0.000000
Dens Rng
To alter data
HART
Factory programmed to fit specific application. EZ Map Version #
Scan
On
ScanTime
TotReset
Yes
Dsply TC
05 Sec
05 Sec
Show
Err Code
Show
Show
Total
Show
BarGraph
FlowRate
menu P
Display
Flow Res
8 digits
HART
On
menu P
No
Yes
Error
code 1
0000
Set Pass
menu P
Password
Exit
SavChng?
Vor Ampl
0%
Vor Freq
0 Hz
4 mA Cnt
1270
20mA Cnt
7150
Version
3.00
00000 Hz
0.000%
Freq Out
Sub Freq
0000 Hz
4-20 Out
menu D
Service
To alter data press ENTER
key within selected block
Use up/down arrows
to alter value
Use right arrow
to move cursor position
To store data press ENTER
menu C
1.
2.
3.
4.
Vor freq
0 sec
Freq out
Vor freq
On
4-20 TC
05 sec
Zero TM
4-20 mA
Output
menu C
Password
0000
To Move Through Map
1. Use up/down
and right
arrows as shown
in map
2. While in display menu hold down right
arrow
key for 2 sec. to go to programming menus
3. While in column headings press ENTER
key to
go to display menu. You will be prompted by "Exit?"
and then "Sav Chg?".
Scl freq
Customer P.O. #
003 %
37.4025
Min flow
1.87013
M-factor
0150
Noise Lv
Flw unit
gal/min
Tot unit
x1 gal
Max flow
2 SEC
Freq max
Error
code 4
10%
103.42
x1 gal
30.421
gal/min
Display
Menu
Basic
menu C
FLUID:
g e n e r a l
s p e c i f i c a t ions
EZ Logic Programming Map
Customer Name
Work Order #
17
emco
fl o w
s y s t e m s
Model and Suffix Codes
Category
Model
Suffix Codes
Stainless Steel (sizes 1 to 12")
PhD-90
Carbon Steel - Stainless steel wing
PhD-92
(sizes 6 to 12")
Fluid Type
Steam or gas
S
Liquid
L
Line Size
1"
1.5"
2"
10
15
20
3"
30
4"
40
6"
60
8"
80
10"
100
12"
120
25 mm
DN25
40 mm
DN40
50 mm
DN50
80 mm
DN80
100 mm
DN100
150 mm
DN150
200 mm
DN200
250 mm
DN250
300 mm
DN300
Connection Type
Wafer6
W
Flange
F
Connection Rating
ANSI Class 150
150
ANSI Class 300
300
ANSI Class 600
600
DIN, PN 16
PN 16
DIN, PN 40
PN 40
DIN, PN 64
PN 64
Options
No indicator/totalizer option1
STD
FM
FM approved
2
CSA approved
CENELEC approved2
Local Indicator and Totalizer3
Remote Mount Electronics 4
Integral 110 VAC input5
Integral 220 VAC input5
Sensor Wires (Internal)
Teflon, –40° to 400°F (–40° to 204°C)
Fiberglass, 150° to 750°F (65° to 400°C)
Example
18
CSA
CEN
LOC-TOT
RMT
110
220
T
F
PhD-90
S-
60-
F-
300-
LOC-TOT-
T
g e n e r a l
1
s p e c i f i c a t ions
Has 4 to 20 mA and/or 50% duty cycle frequency output. A 50% duty cycle frequency can
be scaled.
2
FM, CSA and CENELEC approved with LOC-TOT option meters come with a magnet
wand. For all other meters the wand can be ordered as an option.
3
At least one local indicator is recommended at each site. If one display will be used on
many meters, it should be purchased separately.
4
Remote mount electronics are required for high process temperatures (refer to the
EUROPEAN CE MARK
CARBON STEEL
110/220
RMT
LOC-TOT
CENELEC
CSA
Wafer style meters available in stainless steel (sizes 1 to 4" only).
FM
Not available with European CE Mark or FM, CSA and CENELEC approvals.
6
STD
5
STAINLESS STEEL
general specification). The standard remote mount option comes with 30' of cable.
STD
FM
CSA
CENELEC
LOC-TOT
RMT
110/220
CARBON STEEL
STAINLESS STEEL
EUROPEAN CE MARK
Compatible
Incompatible
Figure 10. Compatibility Chart
19
emco
fl o w
s y s t e m s