Impella® 5.0
Transcription
Impella® 5.0
Impella 5.0 ® with the Impella Console ® Circulatory Support System Instructions for Use & Clinical Reference Manual IMPORTANT NOTICE: Read this entire manual before using the Impella® 5.0 Circulatory Support System (Impella® 5.0 System). The Impella® 5.0 System is to be used only in accordance with this manual. Information contained in this document is subject to change without notice. ©2010 Abiomed, Inc. All rights reserved. The ABIOMED logo and ABIOMED are trademarks of Abiomed, Inc. and are registered in the U.S.A. and certain foreign countries. Recovering hearts. Saving lives. is a trademark of Abiomed, Inc. Impella® is a trademark of Abiomed Europe GmbH, a wholly owned subsidiary of Abiomed, Inc., and is registered in the U.S.A. and certain foreign countries. Impella® 5.0 With the Impella® Console Instructions for Use & Clinical Reference Manual Rx Only Abiomed, Inc. 22 Cherry Hill Drive Danvers, MA 01923 978-777-5410 (voice) 978-777-8411 (fax) [email protected] (email) Abiomed Europe GmbH Neuenhofer Weg 3 52074 Aachen, Germany +49 (241) 8860-0 (voice) +49 (241) 8860-111 (fax) [email protected] (email) www.abiomed.com 24-Hour Emergency Hotlines: N. America 1-800-422-8666 Europe +49 (0) 1805 2246633 June 2010 Document No. 0041-9006 Rev. A Table of Contents Introduction Introduction.............................................................................................. I 1 Warnings and Cautions Warnings.................................................................................................. 1.1 Cautions................................................................................................... 1.2 2 Indications, Contraindications, and Potential Adverse Events Indications (United States)........................................................................ 2.1 Contraindications (United States).............................................................. 2.1 Potential Adverse Events (United States)................................................... 2.1 3 The Impella® 5.0 System Overview.................................................................................................. 3.1 Impella® 5.0 Catheter............................................................................... 3.3 Impella® Console...................................................................................... 3.7 Impella® Power Supply and Power Supply Cable....................................... 3.8 Purge System............................................................................................ 3.10 Accessories............................................................................................... 3.11 4 Using the Impella® Console Overview.................................................................................................. 4.1 Impella® Console Function Keys................................................................ 4.2 Impella® Console Connections.................................................................. 4.4 Impella® Console Display.......................................................................... 4.5 Selecting Performance Levels.................................................................... 4.6 Positioning Aids and Other Displays.......................................................... 4.7 Mobile Operation...................................................................................... 4.10 5 Using the Impella® 5.0 System Startup...................................................................................................... 5.1 Preparing the Purge Fluid.......................................................................... 5.3 Preparing the Impella® 5.0 System............................................................ 5.4 Priming the Impella® 5.0 Catheter Purge Lumen....................................... 5.7 Inserting the Impella® 5.0 Catheter........................................................... 5.9 Positioning and Starting the Impella® 5.0 Catheter................................... 5.12 Replacing the Purge Tubing....................................................................... 5.14 Purge System Troubleshooting.................................................................. 5.15 Patient Weaning....................................................................................... 5.20 6 Patient Management Topics Patient Management Overview................................................................. 6.1 General Patient Care Considerations . ...................................................... 6.1 Transport Within the Hospital................................................................... 6.2 Right Heart Failure.................................................................................... 6.2 ECG Interference....................................................................................... 6.3 Latex......................................................................................................... 6.3 Positioning and Placement Devices........................................................... 6.3 Use of Echocardiography for Positioning of the Impella® 5.0 Catheter...... 6.4 Understanding and Managing Impella® 5.0 Catheter Position Alarms....... 6.11 Suction..................................................................................................... 6.16 Hemolysis................................................................................................. 6.16 Pressure Sensor Drift and Sensor Failure................................................... 6.18 Operating the Impella® 5.0 Catheter Without Heparin in the Purge Solution.......................................................................................... 6.20 How to Change to A Backup Console........................................................ 6.20 7 I mpella® Console Alarms and Warnings Impella® 5.0 Console Alarms.................................................................... 7.1 Impella® 5.0 Console Warnings................................................................. 7.3 8 General System Information Terminology, Abbreviations, and Symbols................................................. 8.1 Ambient Conditions.................................................................................. 8.3 Impella® 5.0 Catheter Parameters............................................................. 8.3 Impella® Console and Power Supply Parameters....................................... 8.4 Connector Cable Parameters..................................................................... 8.4 Impella® 5.0 Catheter Dimensions............................................................ 8.5 Anatomic Considerations.......................................................................... 8.6 Cleaning................................................................................................... 8.6 Storing the Impella® Console and the Power Supply.................................. 8.7 Returning an Impella® 5.0 Catheter to Abiomed (United States)................ 8.7 Appendices Appendix A: Impella® System Limited Service Warranty (United States)... A.1 Appendix B: Technical Safety Inspections, Maintenance, and Repair........ B.1 Index Table of Contents Figures Figure 3.1 Figure 3.2 Figure 3.3 Figure 3.4 Figure 3.5 Figure 3.6 Impella® 5.0 Catheter in the Heart............................................ 3.1 Impella® 5.0 System Components............................................. 3.2 Impella® 5.0 Catheter............................................................... 3.3 Differential Pressure Sensor...................................................... 3.5 Electrical Signal Generated by the Cardiac Cycle....................... 3.5 Correct Impella® 5.0 Catheter Positioning and Pulsatile Placement Signal...................................................................... 3.6 Figure 3.7 Incorrect Impella® 5.0 Catheter Positioning and Flat Placement Signal...................................................................... 3.6 Figure 3.8 Impella® Console...................................................................... 3.7 Figure 3.9 Impella® Power Supply.............................................................. 3.8 Figure 3.10 Impella® Power Supply Rear Panel............................................ 3.9 Figure 3.11 Impella® Power Supply Cable.................................................... 3.9 Figure 3.12 B. Braun Vista® basic Infusion Pump......................................... 3.10 Figure 3.13Impella® Combitrans-Monitoring Set (CM-Set).......................... 3.11 Figure 3.14 Pressure Transducer Cable........................................................ 3.11 Figure 3.15 Connector Cable....................................................................... 3.11 Figure 3.160.025 inch, 260 cm Placement Guidewire................................. 3.12 Figure 3.17 Purge Tubing............................................................................ 3.12 Figure 3.18 Pump Test Plug......................................................................... 3.12 Figure 3.19 20% Dextrose in Water............................................................ 3.12 Figure 4.1 Impella® Console Function Keys................................................ 4.2 Figure 4.2 Impella® Console Display.......................................................... 4.5 Figure 4.3 Pulsatile Waveform on Placement Signal Screen....................... 4.7 Figure 4.4 Placement Monitoring Screen................................................... 4.7 Figure 4.5 Dual Signal Screen.................................................................... 4.8 Figure 4.6 Purge Pressure Screen............................................................... 4.8 Figure 4.7 Motor Current Screen............................................................... 4.9 Figure 4.8 Speed Screen............................................................................ 4.9 Figure 4.9 Flow Rate Screen...................................................................... 4.10 Figure 5.1 Placing the Impella® Power Supply Next to the Impella® Console...................................................................... 5.1 Figure 5.2 Connecting the Power Supply Cable to the Power Supply......... 5.2 Figure 5.3 Connecting the Power Supply Cable to the Console.................. 5.2 Figure 5.4 Turning the Power Supply On.................................................... 5.2 Figure 5.5 Connecting the Impella® 5.0 Catheter to the Connector Cable........................................................................................ 5.4 Figure 5.6 Snapping Plastic Clip to Connector Cable.................................. 5.5 Figure 5.7 Attaching the Connector Cable to the Console......................... 5.5 Figure 5.8 Connect Pressure Transducer Cable Screen............................... 5.6 Figure 5.9 Connecting the Pressure Transducer Cable to the Console........ 5.6 Figure 5.10 Flush Purge Line and Purge Lumen Screen................................ 5.7 Figure 5.11 Connecting CM-Set to Impella® 5.0 Catheter............................ 5.7 Figure 5.12 Ready to Insert Screen.............................................................. 5.8 Figure 5.13 Discharging Fluid from Catheter Outlet Area............................. 5.8 Figure 5.14 Figure 5.15 Figure 5.16 Figure 5.17 Figure 5.18 Figure 5.19 Figure 5.20 Figure 5.21 Figure 5.22 Figure 5.23 Figure 5.24 Figure 5.25 Figure 6.1 Cut-Down Insertion of the Impella® 5.0 Catheter...................... 5.9 Guidewire Placement................................................................ 5.10 Initial Placement Signal............................................................. 5.12 Placement Signal as the Catheter Crosses the Aortic Valve....... 5.12 Confirming Placement on the Dual Signal Screen...................... 5.13 Low Purge Pressure Screen....................................................... 5.15 Normal Purge Pressure on 5 Minute Scale Screen..................... 5.16 Abnormal Purge Pressure on 5 Minute Scale Screen.................. 5.17 High Purge Pressure Screen...................................................... 5.18 Purge Management.................................................................. 5.19 Downstream Occlusion Alarm................................................... 5.20 Disconnecting the Connector Cable from the Console............... 5.21 Labeled TEE and TTE Images of the Impella® 5.0 Catheter Position.................................................................................... 6.5 Figure 6.2 Transesophageal Echocardiographic (TEE) Illustrations of . Impella® 5.0 Catheter Position.................................................. 6.8 Figure 6.3 Transthoracic Echocardiographic (TTE) Illustrations of . Impella® 5.0 Catheter Position.................................................. 6.9 Figure 6.4 Correct and Incorrect Impella® 5.0 Catheter Position (Color Doppler TTE).................................................................. 6.10 Figure 6.5 Correct Impella® 5.0 Catheter Position..................................... 6.12 Figure 6.6 Impella® 5.0 Catheter Fully in Ventricle or Fully in Aorta........... 6.13 Figure 6.7 Impella® 5.0 Catheter Position Unknown Due to Low Pulsatility.................................................................................. 6.14 Figure 6.8 Impella® 5.0 Catheter Outlet Area on or near Aortic Valve........ 6.15 Figure 8.1 Impella® 5.0 Catheter Dimensions............................................ 8.5 Figure B.1 Inspection Sticker Showing Inspection Required in May 2010... B.1 Figure B.2 Impella® Console Information Screen........................................ B.2 Tables Table 3.1 Table 3.2 Table 4.1 Table 4.2 Table 4.3 Table 4.4 Table 5.1 Table 5.2 Table 5.3 Table 6.1 Table 6.2 Impella® 5.0 Catheter Components.......................................... 3.3 Impella® 5.0 System Accessories............................................... 3.11 Impella® Console Function Keys................................................ 4.3 Impella® Console Connections.................................................. 4.4 Impella® Console Display Elements........................................... 4.5 Performance Level Flow Rates................................................... 4.6 Handling Low Purge Pressure Alarms........................................ 5.16 Amount of Dextrose Delivered in the Purge Fluid (g/h).............. 5.19 Amount of Heparin Delivered in the Purge Fluid (IU/h).............. 5.19 Guide for Managing Hemolysis in Various Circumstances.......... 6.17 Expected Flow Rate Displayed on Console................................ 6.18 Introduction introduction Purpose of Manual This Instructions for Use & Clinical Reference Manual is designed for healthcare professionals. It contains clinical and technical information to guide healthcare professionals in their use of the Impella® 5.0 System. The system performs life-sustaining functions. To use the system you must understand and follow these instructions as well as the instructions accompanying the infusion pump you are using. The Impella® 5.0 System may be used only for its intended purpose. Manual Overview This manual provides instructions for use of the Impella® 5.0 System. The following summarizes the contents of each section of this manual. • Section 1: Warnings and Cautions discusses the warnings and cautions pertaining to the use of the Impella® 5.0 System. • Section 2: Indications, Contraindications, and Potential Adverse Events discusses indications for use of the Impella® 5.0 System, contraindications, and potential adverse events that may be associated with use of the system. • Section 3: The Impella® 5.0 System provides an overview of the system and describes its major components. • Section 4: Using the Impella® Console describes the controls and connections on the Impella® Console. • Section 5: Using the Impella® 5.0 System provides the procedures for using the system. • Section 6: Patient Management Topics provides key information on various topics related to management of patients with the Impella® 5.0 Circulatory Support System. • Section 7: Impella® Console Alarms and Warnings provides an alphabetical listing of Impella® Console alarms and warnings, as well as information on what to do to resolve them. • Section 8: General System Information contains information including definitions for key terms that appear in the manual, descriptions of the abbreviations and symbols that appear on Impella® 5.0 System components and packaging, technical information pertaining to the Impella® 5.0 System, and instructions on cleaning and storing system components as well as returning components to Abiomed. • Appendices at the end of the manual provide supplemental information about topics including the Impella® System Limited Service Warranty and technical safety inspection, maintenance and repair. Impella® 5.0 with the Impella® Console I 1 Warnings and Cautions Warnings.................................................................................................... 1.1 Cautions.....................................................................................................1.2 Warnings Warnings The Impella® 5.0 System has not been qualified in the United States (US) through clinical studies for long term use (longer than 6 hours), for example, as a bridge to transplant or for pending recovery of the natural heart. Warnings alert you to situations that can cause death or serious injury. The red symbol appears before warning messages. Fluoroscopy is required to guide placement of the Impella® 5.0 Catheter. The small placement guidewire must be reliably observed at all times. Avoid manual compression of the inlet, outlet, or sensor areas of the cannula assembly. Avoid overinserting the catheter and possibly impinging the catheter tip against the walls of the vasculature, atrium, or ventricle. Do NOT advance or withdraw the Impella® 5.0 Catheter against resistance without using fluoroscopy to determine the cause of the resistance. Doing so could result in separation of the catheter or guidewire tip, damage to the catheter or vessel, or ventricular perforation. Be sure that the stopcock on the repositioning sheath is always kept in the closed position. Significant bleed back can result if the stopcock is open. The sterile components of the Impella® 5.0 System can be used only if the sterilization indicators show that the contents have been sterilized, the packaging is not damaged, and the expiration date has not elapsed. Do NOT resterilize or reuse the Impella® 5.0 Catheter. It is a disposable device and is intended for single use only. Reuse, reprocessing, or resterilization may compromise the structural integrity of the catheter and/or lead to catheter failure which, in turn, may result in patient injury, illness, or death. Retrograde flow will occur across the aortic valve if the Impella® 5.0 Catheter is set at performance level P0. The infusion pump supplied by Abiomed must NOT be used in conjunction with any products other than Impella® products. Do NOT use saline in the purge system. Do NOT use alcohol-based fluids for lubrication. Do NOT use an Impella® 5.0 System if any part of the system is damaged. To prevent the risk of explosion, do NOT operate the Impella® 5.0 System near flammable anesthetics. If at any time during the course of support with the Impella® 5.0 Catheter, the Impella® Console alarms “Low Purge Pressure,” follow the instructions presented in “Purge System Troubleshooting” in Section 5 of this manual. Impella® 5.0 with the Impella® Console 1.1 1 WARNINGS AND CAUTIONS The Impella® 5.0 System is intended for use only by personnel trained in accordance with the Abiomed® Training Program. Do NOT subject a patient who has been implanted with an Impella® 5.0 Catheter to magnetic resonance imaging (MRI). The strong magnetic energy produced by an MRI machine may cause the Impella® 5.0 System components to stop working, and result in injuries to the patient. An MRI may also damage the electronics of the Impella® 5.0 System. During defibrillation, do NOT touch the Impella® 5.0 Catheter, cables, or Impella® Console. Cautions Cautions Cautions indicate a situation in which equipment may malfunction, be damaged, or cease to operate. The yellow symbol appears before caution messages. Handle with care. The Impella® 5.0 Catheter can be damaged during removal from packaging, preparation, insertion, and removal. Do NOT bend, pull, or place excess pressure on the catheter or mechanical components at any time. Patients with aortic stenosis or other abnormal aortic valve performance may be compromised by the use of the Impella® 5.0 Catheter. Patients with aortic valve disease should be observed for aortic insufficiency. Use only original accessories and replacement parts supplied by Abiomed. Refer to the instructions for use supplied with the infusion pump being used before using the infusion pump with the Impella® 5.0 System. To troubleshoot purge system alarms not listed in this manual, refer to the appropriate infusion pump instructions for use manual. Holding down the ON key on the Impella® Console for longer than 3 seconds during operation will cause the Impella® Console to go into emergency stop mode. The performance level of the Impella® 5.0 Catheter is automatically set to P2 during zeroing if the catheter is operating above P0. Performance level remains at P2 after zeroing and does not automatically return to its prior setting. When conducting the Impella® Console function test as part of preventive maintenance, if a pressure value of 50 (± 4) mmHg does not appear or the speed is not within the specified range of 24,000 ± 1,000 revolutions per minute (rpm), do NOT use the Impella® Console. Do NOT use damaged or contaminated connector cables. To prevent device failure, do NOT start the Impella® 5.0 Catheter until the placement guidewire has been removed. Do NOT remove the Impella® 5.0 Catheter over the length of the placement guidewire. 1.2 Instructions for Use & Clinical Reference Manual 1 WARNINGS AND CAUTIONS When replacing the purge tubing, the replacement process must be completed within 2 minutes. The Impella® 5.0 Catheter may be damaged if replacement takes longer than 2 minutes. Protect the Impella® Console and the Impella® Power Supply from direct sunlight or exposure to excessive temperature (max. 40°C) to prevent malfunction. To prevent overheating and improper operation, do NOT block the cooling vents of the Impella® Power Supply while it is operating. Do NOT place the Impella® Console on top of the Impella® Power Supply. Do NOT kink or clamp any part of the Impella® 5.0 Catheter. The Li-Ion batteries must be charged for 10 hours prior to system operation. After being unplugged, the Impella® Console will operate for at least 60 minutes after the batteries have been fully charged. Do NOT use cell phones within a radius of 1 meter (3.28 feet) of the Impella® 5.0 System. Cell phones can cause improper operation of the system and endanger the patient. Minimize exposure of Impella® 5.0 System components to sources of electromagnetic interference (EMI). Exposure to sources of EMI, such as cell phones and two-way radios, may cause operational interference. To clear interference, either increase the distance between system components and the EMI source or turn off the EMI source. Operation of Impella® 5.0 System components may interfere with the operation of other devices. If interference occurs, increase the distance between the device and system components. The use of high-frequency surgical devices may cause temporary interference with the sensor signals. If continuous interference persists, the following warning message appears on the display screen: “Sensor Value not Reliable.” Please acknowledge this message. There is no reason to discontinue use of the Impella® 5.0 Catheter. Impella® 5.0 with the Impella® Console 1.3 2 Indications, Contraindications, and Potential Adverse Events Indications (united States)....................................................................2.1 Contraindications (United states)......................................................2.1 Potential Adverse Events (United States)..........................................2.1 Indications (United States) 2 Indications, contraindications, and Potential Adverse Events The Impella® 5.0 is intended for circulatory support using an extracorporeal bypass control unit, for periods up to 6 hours. It is also intended to be used to provide circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass. The Impella® 5.0 Circulatory Support System also provides pressure measurements which are useful in determining intravascular pressure. Contraindications (United States) Patients with aortic stenosis or other abnormal aortic valve performance may be compromised by the use of the Impella® 5.0 Catheter. Patients with aortic valve disease should be observed for aortic insufficiency. • Mechanical aortic valve or heart constrictive device • Aortic valve stenosis/calcification (graded as ≥ +2 equivalent to an orifice area of 1.5 cm2 or less) • Moderate to severe aortic insufficiency (echocardiographic assessment of aortic insufficiency graded as ≥ +2) • Severe peripheral arterial obstructive disease that would preclude Impella® 5.0 device placement Potential Adverse Events (United States) • Aortic insufficiency • Limb ischemia • Aortic valve injury • Myocardial infarction • Arrhythmia • Neurologic injury • Atrial fibrillation • Perforation • Bleeding • Renal failure • Cardiac tamponade • Respiratory dysfunction • Cardiogenic shock • Sepsis • Cerebral vascular accident • Thrombocytopenia • Death • Thrombotic vascular (non-CNS) complication • Device malfunction • Transient ischemic attack (TIA) • Hemolysis • Vascular injury • Hepatic failure • Ventricular fibrillation • Insertion site infection • Ventricular tachycardia Impella® 5.0 with the Impella® Console 2.1 3 The Impella® 5.0 System Overview.....................................................................................................3.1 Reusable System Components.......................................................................... 3.1 Single-use System Components........................................................................3.2 System Configuration.......................................................................................3.2 Impella® 5.0 Catheter.............................................................................. 3.3 Catheter Components.......................................................................................3.3 Differential Pressure Sensor..............................................................................3.4 Impella® Console......................................................................................3.7 Impella® Power Supply and Power Supply Cable..............................3.8 Impella® Power Supply.....................................................................................3.8 Impella® Power Supply Cable............................................................................3.9 Purge System...........................................................................................3.10 Accessories.............................................................................................. 3.11 Overview 3The Impella® 5.0 System The Impella® 5.0 Catheter is an intravascular microaxial blood pump that supports a patient’s circulatory system. The catheter is inserted through the femoral artery and into the left ventricle. (See Figure 3.1.) Figure 3.1 Impella® 5.0 Catheter in the Heart When properly positioned, the Impella® 5.0 Catheter delivers blood from the inlet area, which sits inside the left ventricle, through the cannula to the outlet opening in the ascending aorta. Physicians and device operators monitor the correct positioning and functioning of the Impella® 5.0 Catheter on the display screen of the Impella® Console. This section describes the components of the Impella® 5.0 System. Reusable System Components The Impella® 5.0 System consists of the following reusable components: • Impella® Console—provides the user interface, alarm indications, and portable battery • Impella® Power Supply and Power Supply cable—for continuous operation of the console • Infusion pump—for purging the Impella® 5.0 Catheter • Pressure transducer cable • Pump test plug Impella® 5.0 with the Impella® Console 3.1 Single-use System Components The system also includes the following single-use components: • Impella® 5.0 Catheter • Impella® Combitrans-Monitoring Set (CM-Set) • 0.025 inch, 260 cm placement guidewire • Connector cable • Purge tubing System Configuration Figure 3.2 illustrates the Impella® 5.0 System components. D20 Dextrose Solution Power Supply Power Supply Cable Infusion Pump Pressure Transducer Cable Impella® 5.0 Catheter Impella® Console Connector Cable Purge Tubing Combitrans-Monitoring Set = connection Figure 3.2 Impella® 5.0 System Components 3.2 Instructions for Use & Clinical Reference Manual Impella® 5.0 Catheter 3The Impella® 5.0 System Catheter Components The Impella® 5.0 Catheter is an intravascular microaxial blood pump that delivers up to 5.0 liters of blood per minute from the left ventricle into the aorta. Figure 3.3 illustrates the Impella® 5.0 Catheter. Table 3.1 describes each component from the pigtail at one end to the check valve on the other end. Check Valve Pressure Reservoir Red Impella® Plug Infusion Filter Clear Sidearm Catheter Shaft Repositioning Unit Pigtail Inlet Area Cannula Differential Pressure Sensor Outlet Area Motor Housing Figure 3.3 Impella® 5.0 Catheter Table 3.1 Impella® 5.0 Catheter Components Component Description Pigtail The 6 Fr pigtail is attached to the cannula at the distal end of the inlet area. It assists with stabilizing the catheter in the correct position in the left ventricle and decreases the amount of ectopy that could be caused by a foreign body in the ventricle. Inlet area The inlet area, located at the distal tip of the cannula, has five openings (windows) that allow blood to be drawn into the inlet and channeled through the cannula. Cannula The 21 Fr cannula has a spiral-shaped reinforced body that is shaped in a 35-degree angle. The cannula is made of nitinol and covered in polyurethane. • Nitinol provides the cannula with flexibility and memory, allowing it to conform across the aortic valve. • Polyurethane provides a smooth surface for atraumatic positioning and reduced risk of damage to the aortic valve. Impella® 5.0 with the Impella® Console 3.3 Table 3.1 Impella® 5.0 Catheter Components (cont’d) Component Description Differential pressure sensor A sensor that measures the pressure difference between the inside and outside of the cannula. The pressure value is used for positioning during placement and for monitoring flow and position during catheter operation. Outlet area The proximal end of the cannula is attached to the outlet area where the blood exits the cannula. Motor housing The motor housing is 21 Fr in diameter and consists of an encapsulated motor. Catheter shaft A 9 Fr catheter shaft is located between the motor housing and . the red Impella® plug. The lumen of the catheter shaft contains a purge lumen, an electrical cable, and a differential pressure measurement cable. The catheter shaft has longitudinal and transversal marks: • The longitudinal mark along the inner radius shows correct position of the 0.025 inch, 260 cm placement guidewire once backloaded on the Impella® 5.0 Catheter. • The transversal marks at 1 cm intervals with numbers every 5 cm aid in proper positioning. Repositioning unit The repositioning unit consists of a sheath and an anticontamination sleeve with an anchoring ring. • The 9 Fr sheath (11 Fr outer diameter) with hemostatic valve is located on the catheter shaft and allows repositioning of the catheter. • The anchoring ring of the anticontamination sleeve secures the sheath to the catheter. Red Impella® plug The red Impella® plug has a clear sidearm and contains memory that retains operating parameters in case the patient needs to be transferred to another console. The plug connects the catheter to the Impella® Console through a connector cable. Clear sidearm The clear sidearm is attached to the infusion pump to purge the Impella® 5.0 Catheter. It leads to the infusion filter, the pressure reservoir, and the check valve. Infusion filter The infusion filter prevents bacterial contamination and air from entering the purge lumen. Pressure reservoir The pressure reservoir includes a flexible rubber diaphragm that provides additional filling volume by means of an expansion chamber during purge solution change. Check valve The yellow check valve ensures that purge fluid does not flow in the reverse direction when the purge solution is exchanged. Differential Pressure Sensor The Impella® 5.0 Catheter has an electronic differential pressure sensor located at the proximal end of the 21 Fr cannula (see Figure 3.4). The purpose of the pressure sensor is to generate the placement signal, which is used by operators, the console’s placement monitoring screen, and alarm algorithms to monitor the position of the Impella® 5.0 cannula relative to the aortic valve. 3.4 Instructions for Use & Clinical Reference Manual 3The Impella® 5.0 System The pressure sensor is a flexible membrane integrated into the cannula (see Figure 3.4). One side of the sensor is exposed to the blood pressure on the outside of the cannula and the other side is exposed to the pressure of the blood inside of the cannula. The sensor generates an electrical signal proportional to the difference between the pressure outside the cannula and the pressure inside. This signal is displayed on the Impella® Console as the placement signal. Differential Pressure Sensor Aortic Pressure Sensor Flow from Ventricle Ventricular Pressure Flow to Aorta Membrane Cannula Aortic Valve Figure 3.4 Differential Pressure Sensor The pressure sensor membrane flexes when the pressure on one side is different from the pressure on the other side, and the electrical properties of the membrane change when it flexes. This allows the sensor to generate an electrical signal proportional to how much the membrane is flexed, and thus proportional to the difference between the pressure on the outside of the cannula and the pressure inside. When the Impella® 5.0 Catheter is placed in the correct position across the aortic valve, the top (outer surface) of the sensor is exposed to the aortic pressure and the bottom (inner surface) of the sensor is exposed to the ventricular pressure. Therefore, the placement signal is approximately equal to the difference between the aortic pressure and the ventricular pressure (see Figure 3.5). Systole Blood Pressure (mmHg) 120 Aortic pressure 80 Pressure difference Diastole Left ventricular pressure 0 Time PDifferential = PAO - PLV Pressure difference (PDifferential ) Left ventricular pressure (PLV ) Aortic pressure (PAO ) Figure 3.5 Electrical Signal Generated by the Cardiac Cycle Impella® 5.0 with the Impella® Console 3.5 When the Impella® 5.0 Catheter is correctly positioned across the aortic valve, the changes in pressure associated with the cardiac cycle result in a pulsatile placement signal (see Figure 3.6). During diastole (time zero in Figure 3.5), the large pressure difference between the aorta and the left ventricle creates a large electrical signal. Then at the peak of systole, when the aortic valve opens, the pressure difference between the aorta and the left ventricle—and thus the electrical signal—is zero. Thus, the continual pressure changes associated with the cardiac cycle produce the pulsatile (up and down) waveform seen on the Impella® Console display. P-Perf Flow [L/min] P9 5.0 98% LV4509 35 1 Placement Monitoring Ventricle P-Perf P9 LV4509 Flow [L/min] 5.0 98% 35 1 Placement Signal [mmHg] Aorta Pump Position OK 10 s 10 s Impella® 5.0 Catheter correctly positioned Pulsatile placement signal Figure 3.6 Correct Impella® 5.0 Catheter Positioning and Pulsatile Placement Signal When the Impella® 5.0 Catheter is not properly placed across the aortic valve, or when it is fully in the aorta or fully in the ventricle, the pressures outside and inside the cannula are the same throughout the cardiac cycle. As a result, the pressure on either side of the sensor membrane is the same, resulting in a flat placement signal (see Figure 3.7). P8 98% Ventricle Aorta 2.5 LV4509 Placement Monitoring P-Perf Flow [L/min] Pump Position Wrong 98% Ventricle Aorta P8 2.5 Press MENU for Reposition P-Perf P9 LV4509 Flow [L/min] 98% 5.3 8 -2 Placement Signal [mmHg] 10 s LV4509 Placement Monitoring Pump Position Wrong Ventricle Aorta Press MENU for Reposition 10 s 10 s Impella® 5.0 Catheter position incorrect Flattened placement signal Figure 3.7 Incorrect Impella® 5.0 Catheter Positioning and Flat Placement Signal 3.6 Instructions for Use & Clinical Reference Manual Impella® Console Impella® Console operation is described in detail in Section 4 of this manual. Impella® Console Battery Power The console can operate on its internal lithium-ion (Li-Ion) battery for at least 60 minutes when fully charged. Figure 3.8 Impella® Console Impella® 5.0 with the Impella® Console 3.7 3The Impella® 5.0 System The Impella® Console (see Figure 3.8) controls the microaxial blood pump and monitors the Impella® 5.0 System. The console weighs 7 lbs (3.35 kg) and can operate on its internal battery for at least 60 minutes when fully charged. Impella® Power Supply and Power Supply Cable Impella® Power Supply The Impella® Power Supply (see Figures 3.9 and 3.10) provides power to and charges the battery within the Impella® Console. Data stored in the Impella® Console can be downloaded to a PC for analysis through the CAN-Bus interface located on the Impella® Power Supply. The Impella® Power Supply is a multivoltage AC power source for use with voltages within a range of 110 to 230 V at line frequencies of 50 and 60 Hz. The integrated charging circuit can fully charge the Li-Ion batteries in the Impella® Console in about 10 hours. Figure 3.9 Impella® Power Supply The front panel of the Impella® Power Supply has three green indicator lights: • POWER light – turns on when the unit is connected to AC power and powered on with the power switch on the rear panel. • CONNECTED light – turns on when the Impella® Console is connected to the Impella® Power Supply and the power supply is turned on. • CAN light – turns on when the Impella® Console is connected and powered on, but is not as bright as the other two lights; flickers briefly at the start of data transmission and when transmission has finished it flickers until the download program has finished or the CAN-bus connection is disconnected. 3.8 Instructions for Use & Clinical Reference Manual Impella® Console Connector Fuses AC Power Connector 3The Impella® 5.0 System CAN-Bus Interface (Service Use Only) Power Switch Figure 3.10 Impella® Power Supply Rear Panel The following items are located on the rear panel (see Figure 3.10) of the Impella® Power Supply: • CAN-Bus interface for data transfer to an external device • Impella® Console connector for cable to the Impella® Console • Fuses • AC power connector for AC power cable • Power switch (lights up when power is on) Impella® Power Supply Cable The Impella® Power Supply cable (see Figure 3.11) connects the Impella® Power Supply to the Impella® Console. Figure 3.11 Impella® Power Supply Cable Impella® 5.0 with the Impella® Console 3.9 Purge System Do not use saline in the purge system. Refer to Infusion pump IFU For more information about the infusion pump being used, refer to the instructions for use included with the infusion pump. The Impella® 5.0 System requires a purge system that delivers rinsing fluid to the Impella® 5.0 Catheter. The B. Braun Vista basic Infusion Pump (see Figure 3.12) is an example of an infusion pump Abiomed uses to serve this function. It delivers purge fluid (typically 20% dextrose solution plus heparin 50 IU/mL) through the catheter to the microaxial blood pump to prevent blood from entering the motor. Display Screen Function Keys ON/OFF Key Keypad Alarm Silence Key START/STOP Key Alarm Light Clear Key Infusion Light Figure 3.12 B. Braun Vista® basic Infusion Pump • Display screen – for communicating information; green backlighting turns on when the unit is connected to AC power; in battery operation, the backlighting turns on when a key is pressed • Keypad – for entering infusion settings • Alarm light – for alerting users to an alarm; turns on when an alarm condition occurs; display flashes “AAA.A” • Infusion light – for alerting users that the pump is infusing • Function keys – for selecting on-screen options • ON/OFF key – for turning the Vista basic power on and off • Alarm silence key – for silencing the alarm for approximately 2 minutes • START/STOP key – for starting and stopping the infusion • Clear key – for resetting a value to zero 3.10 Instructions for Use & Clinical Reference Manual Accessories 3The Impella® 5.0 System Table 3.2 illustrates and describes the Impella® 5.0 System accessories. Table 3.2 Impella® 5.0 System Accessories Component Description The CM-Set is connected to the purge tubing and to the yellow check valve on the clear sidearm of the Impella® 5.0 Catheter. The CM-Set measures the purge pressure in the purge lumen of the catheter. The purge pressure measurement is displayed on the Impella® Console and must be within the range of 300 to 700 mmHg. The CM-Set is connected to the Impella® Console with the pressure transducer cable. Figure 3.13 Impella® CombitransMonitoring Set (CM-Set) The gray pressure transducer cable connects the CM-Set to the Impella® Console. The cable is reusable and nonsterile. • The yellow plug at the end of the cable is inserted into the pressure connector socket on the left side of the Impella® Console. Earlier versions of the cable have a red hub at the end that plugs into the console. • The plug at the opposite end is connected to the CM-Set. Figure 3.14 Pressure Transducer Cable The blue, disposable connector cable connects the catheter to the Impella® Console. • The socket at the gray end of the cable connects to the red Impella® plug. • The blue plug at the opposite end of the cable is inserted into the catheter connector socket on the right side of the Impella® Console. A retainer at the end of the connector cable prevents inadvertent removal of the cable from the Impella® Console. Figure 3.15 Connector Cable Impella® 5.0 with the Impella® Console 3.11 Component Description The 0.025 inch, 260 cm guidewire is used for the placement of the Impella® 5.0 Catheter only. Figure 3.16 0 .025 inch, 260 cm Placement Guidewire The purge tubing connects to the CM-Set, which connects to the yellow check valve on the clear sidearm of the Impella® 5.0 Catheter. Purge tubing for the B. Braun Vista basic Infusion Pump is shown here. Figure 3.17 Purge Tubing The pump test plug is a test device used only for preventive maintenance to confirm Impella® Console function. Figure 3.18 Pump Test Plug Hospital Provided: Dextrose solution (typically 20% dextrose in water with 50 IU/mL of heparin) is used as the purge fluid through the Impella® 5.0 Catheter. Figure 3.19 20% Dextrose in Water 3.12 Instructions for Use & Clinical Reference Manual 4 Using the Impella® Console Overview.....................................................................................................4.1 Impella® Console Function Keys.......................................................... 4.2 Impella® Console Connections............................................................ 4.4 Impella® Console Display...................................................................... 4.5 Selecting Performance Levels..............................................................4.6 Positioning Aids and Other Displays...................................................4.7 Placement Signal (mmHg)................................................................................. 4.7 Placement Monitoring...................................................................................... 4.7 Dual Signal.......................................................................................................4.8 Purge Pressure..................................................................................................4.8 Motor Current (A).............................................................................................4.9 Speed...............................................................................................................4.9 Flow Rate....................................................................................................... 4.10 Mobile Operation....................................................................................4.10 Overview 4Using the Impella® Console The Impella® Console is the primary user control interface for the Impella® 5.0 System. It controls the Impella® 5.0 Catheter performance, monitors the Impella® 5.0 System for alarms, and provides real time catheter position information regarding the location of the catheter across the aortic valve. The console can be powered by the Impella® Power Supply or can operate on internal battery power for at least 60 minutes when fully charged. This section of the manual discusses: • Console function keys • Console connections • Console display • Performance levels (P0-P9) • Positioning aids and other displays • Mobile operation of the console Impella® 5.0 with the Impella® Console 4.1 Impella® Console Function Keys Figure 4.1 illustrates the function keys on the Impella® Console. These keys are described in Table 4.1. Display Screen P-PERF Key ZERO Key Power Supply Connector FLOW Key (not active) SIGNAL Key MENU Key ▼/▲ Keys SCALE Key ON Key OK Key Pressure Connector Socket Catheter Connector Socket Figure 4.1 Impella® Console Function Keys 4.2 Instructions for Use & Clinical Reference Manual Table 4.1 Impella® Console Function Keys Description P-PERF key Change the performance level of the Impella® 5.0 Catheter (P0 to P9) ZERO key Used to zero the differential pressure sensor (described in section 6 of this manual) MENU key Access functions that differ depending on whether the catheter is connected. Functions with the catheter connected: Flow Monitoring • Active when the adjusted flow limit appears to the left of the numeric flow value • Bell symbol crossed out when warning is muted Low Flow Limit* • Set default limit to approximately 0.3 L/min below currently displayed mean flow • Change limit in 0.1 L/min steps • Turn off flow limit monitoring MENU key The MENU key on the console accesses different functions depending on whether the catheter is connected (see Table). *NOTE: A warning is generated if values fall below the low flow limit Information • Query information about catheter type, catheter ID, Impella® Console software version, and catheter operating hours Repositioning Guide • Display recommended steps for repositioning the catheter Disable Placement Monitoring • Disable the placement monitoring function; should be used only with the assistance of a trained Abiomed representative Enable Placement Monitoring • Enable disabled placement monitoring Functions without the catheter connected: Language—select English, German, French, Spanish, or Italian Switch off console Console Information ▼ and ▲ keys • Decrease or increase selected parameters • Select a menu item OK key Acknowledge an entry value or command FLOW key Not active SIGNAL key View the following screens: • Placement signal (mmHg) • Placement monitoring • Dual signal (motor current and placement signal) • Purge pressure (mmHg) • Motor current (A) • Speed (rpm x 1000) • Flow rate (L/min) Impella® 5.0 with the Impella® Console 4.3 4Using the Impella® Console Key Table 4.1 Impella® Console Function Keys (cont’d) Key Description SCALE key Select one of three time windows: • 10 seconds • 5 minutes • 5 hours The corresponding value is displayed under the bar graph on the right side . of the display. Minimum and maximum values appear above the axis. ON key Power up the Impella® Console when not connected to the Impella® Power Supply. To turn the Impella® Console ON: (1) Press and hold ON for 3 seconds* To turn the Impella® Console OFF: (1) Press the MENU key (2) Select “Switch Off Console” (3) Press OK OR Disconnect the catheter from the Impella® Console and the console will turn off automatically after 10 minutes *NOTE: Holding down the ON key for longer than 3 seconds during operation will cause the console to go into emergency stop mode Impella® Console Connections Table 4.2 describes the three connection locations on the Impella® Console (illustrated in Figure 4.1). Table 4.2 Impella® Console Connections 4.4 Connector name Description Power supply connector (Top of console) Connection for the power supply cable that connects the Impella® Console to the Impella® Power Supply and carries console data to the CAN-Bus interface located on the Impella® Power Supply Pressure connector socket (Left side of console; yellow) Socket for the gray pressure transducer cable that connects the console to the CM-Set Catheter connector socket (Right side of console; blue) Socket for the connector cable that connects the console to the red Impella® plug on the catheter Also used to connect the pump test plug to confirm console function during preventive maintenance Instructions for Use & Clinical Reference Manual Impella® Console Display 4Using the Impella® Console The elements of the Impella® Console display are shown in Figure 4.2 and described in Table 4.3. Performance Level AC Power Icon Battery Status P-Perf P9 LV4509 Catheter Operation Icon Low Flow Alarm Limit Icon Flow [L/min] 5.0 98% 35 1 Dual Signal I Flow Rate Type of Signal Selected Signal Numeric Readout Signal Bar Graph P 10 s Catheter Type and Serial Number Time Window Display Information and Waveform Display Figure 4.2 Impella® Console Display Table 4.3 Impella® Console Display Elements Display Element Description Performance level Indicates the performance level (P0 to P9) at which the catheter is operating AC power icon Appears when the Impella® Console is connected to the Impella® Power Supply Battery status information Bar graph and numeric readout (percentage); shows the approximate battery charge level • During battery operation, the bar graph flashes when the charge level is less than 50% • When the battery is charging, the bar graph flashes between the present charge level and 100% Catheter operation icon Rotates when the Impella® 5.0 Catheter is operating Low flow alarm limit icon Indicates whether the alarm is enabled or disabled • Enabled: bell symbol and numeric alarm limit are displayed • Disabled: bell symbol is crossed out (as shown in Figure 4.2) Flow rate Indicates the calculated flow rate through the Impella® 5.0 Catheter cannula (0 to 5.0 L/min) Type of signal selected Displays the title of the signal being shown Impella® 5.0 with the Impella® Console 4.5 Table 4.3 Impella® Console Display Elements (cont’d) Display Element Description Signal numeric readout Displays the maximum and minimum values or a numeric mean value for the selected signal; numeric maximum and minimum values displayed on the dual signal and placement monitoring screens correspond to the placement signal Signal bar graph Corresponds to the selected signal screen displayed and its value Time window display Indicates the type of window displayed (10 seconds, 5 minutes, or 5 hours) Information and waveform display Shows information during the startup procedure; shows waveforms during catheter operation Catheter type and serial number Indicates the catheter type; also displays the catheter serial number Selecting Performance Levels Retrograde flow A setting of P0 will result in retrograde flow when the Impella® 5.0 Catheter is placed across the aortic valve. Retrograde flow may also occur at P1. Ten different performance levels (P0 to P9) can be selected (see Table 4.4). Select the lowest performance level that will enable you to achieve the flow rate necessary for patient support. Table 4.4 Performance Level Flow Rates *Flow Rate (L/min) Revolutions Per Minute (rpm) P0 0.0 – 0.0 0 P1 0.0 – 1.4 10,000 P2 0.5 – 2.6 17,000 P3 0.5 – 3.1 20,000 P4 0.9 – 3.4 22,000 Performance Level P5 1.4 – 3.7 24,000 P6 1.8 – 4.0 26,000 P7 2.6 – 4.4 28,000 P8 3.4 – 4.7 30,000 P9 4.2 – 5.3 33,000 *Flow rate can vary due to suction or incorrect positioning. Acknowledging Performance Level Selection The performance level selection will be canceled if you do not press OK within 10 seconds of selecting the performance level. 4.6 To select a performance level: 1.Press and hold the P-PERF key until two blinking vertical bar graphs appear next to the performance level display (P0 to P9). 2.Select the performance level by pressing ▲ to increase performance level or ▼ to decrease performance level. 3.Press OK within 10 seconds to set the performance level to the level selected. Instructions for Use & Clinical Reference Manual Positioning Aids and Other Displays 4Using the Impella® Console Placement Signal (mmHg) The placement signal screen (see Figure 4.3) shows a pressure measurement that is useful for determining the location of the differential pressure sensor with respect to the aortic valve. The placement signal is used to verify the position of the microaxial blood pump by evaluating the differential pressure as pulsatile or flattened. P-Perf P9 LV4509 Flow [L/min] 5.0 98% 35 1 Placement Signal [mmHg] 10 s Figure 4.3 Pulsatile Waveform on Placement Signal Screen Placement Monitoring The placement monitoring screen (see Figure 4.4) shows the approximate catheter position. P-Perf P9 LV4509 Ventricle Flow [L/min] 5.0 98% 35 1 Placement Monitoring Aorta Placement Monitoring The placement monitoring screen will display “Disabled Due to Low Performance Level” if catheter performance level is less than P1. If there is a problem with the pressure sensor, it will display “Disabled Due to Non Reliable Sensor.” Pump Position OK Delayed Response 10 s Figure 4.4 Placement Monitoring Screen Impella® 5.0 with the Impella® Console It may take several seconds for the graphic on the screen to show changes in catheter position. 4.7 Dual Signal The dual signal screen (see Figure 4.5) shows the motor current and placement signal waveforms simultaneously. The motor current is the upper waveform. P-Perf P9 LV4509 Flow [L/min] 5.0 98% 35 2 Dual Signal Motor current (amps) Placement signal (mmHg) I Differential pressure (mmHg) P Position of bar graph corresponds to placement signal 10 s Figure 4.5 Dual Signal Screen Purge Pressure The Impella® Console monitors purge pressure, the pressure of the purge fluid delivered through the catheter to the motor. The purge pressure display ranges from 0 to 1000 mmHg. A warning appears if purge pressure falls below 300 mmHg or exceeds 700 mmHg. P-Perf P9 LV4509 Flow [L/min] 98% 5.0 Purge Pressure [mmHg] 472 < < 10 s Figure 4.6 Purge Pressure Screen 4.8 Instructions for Use & Clinical Reference Manual Motor Current (A) 4Using the Impella® Console Motor current is a measure of the energy intake of the Impella® 5.0 Catheter motor. The energy intake varies with motor speed and the pressure difference between the inlet and outlet areas of the cannula. It is important to monitor the motor current waveform (see Figure 4.7) because it provides information about the catheter position relative to the aortic valve. When the Impella® 5.0 Catheter is positioned correctly, with the inlet area in the ventricle and the outlet area in the aorta, the motor current is pulsatile because the pressure difference between the inlet and outlet areas changes with the cardiac cycle. When the inlet and outlet areas are on the same side of the aortic valve, the motor current will be dampened or flat because there is little or no pressure difference between the inlet and outlet areas. Flow [L/min] P-Perf P9 LV4509 98% 5.0 0.74 0.64 Motor Current [A] 10 s Figure 4.7 Motor Current Screen Speed The speed screen (see Figure 4.8) shows the actual rotational speed of the Impella® 5.0 Catheter motor, located on the distal portion of the catheter. Speed ranges between 0 and 33,000 rpm. P-Perf P9 LV4509 Flow [L/min] 98% Speed [rpm x 1000] 5.0 33.0 10 s Figure 4.8 Speed Screen Impella® 5.0 with the Impella® Console 4.9 Flow Rate Flow rate is representative of the Impella® 5.0 Catheter output measured in liters per minute. It is calculated using the rotational speed (as set by the performance level) and the differential pressure (placement signal). Flow rate ranges between 0 and 5.3 L/min. P-Perf P9 LV4509 Flow [L/min] 98% 5.0 5.3 4.5 Flow [L/min] 10 s Figure 4.9 Flow Rate Screen Mobile Operation The Li-Ion batteries must be charged for 10 hours prior to system operation. After being unplugged, the Impella® Console will operate for at least 60 minutes after the batteries have been fully charged. Turning Console On in Mobile Operation If it is necessary to turn the Impella® Console on under mobile operation, the ON key must be held for 3 seconds to power up the Impella® Console. 4.10 The Impella® Console can be operated on internal battery power when it is not connected to the Impella® Power Supply: 1.Disconnect the Impella® Console from the Impella® Power Supply by disconnecting the power supply cable from the top of the Impella® Console. 2.The Impella® Console beeps once to alert you that the Impella® Console is on battery power. The AC power icon to the left of the battery icon disappears. 3.When connected back to the Impella® Power Supply, the Impella® Console beeps once to alert you that the Impella® Console is on main power and the AC power icon returns. Instructions for Use & Clinical Reference Manual 5 Using the Impella® 5.0 System Startup........................................................................................................5.1 Supplies Needed............................................................................................... 5.1 Equipment Set-up............................................................................................. 5.1 Preparing the Purge Fluid......................................................................5.3 Preparing the Impella® 5.0 System.......................................................5.4 Connecting the Impella® 5.0 Catheter and Connector Cable.............................5.4 Connecting the Pressure Transducer Cable........................................................5.6 Priming the Impella® 5.0 Catheter Purge Lumen...............................5.7 Inserting the Impella® 5.0 Catheter....................................................5.9 Positioning and Starting the Impella® 5.0 Catheter..................... 5.12 Replacing the Purge Tubing................................................................. 5.14 Purge System Troubleshooting.......................................................... 5.15 Low Purge Pressure........................................................................................ 5.15 High Purge Pressure....................................................................................... 5.18 Purge Pressure Management Strategies.......................................................... 5.19 Downstream Occlusion...................................................................................5.20 Patient Weaning......................................................................................5.20 Rapid Weaning...............................................................................................5.20 Slow Weaning................................................................................................ 5.21 1 Startup 5 Using the Impella® 5.0 System Do NOT use an Impella® 5.0 System if any part of the system is damaged. The sterile components of the Impella® 5.0 System can be used only if the sterilization indicators show that the contents have been sterilized, the packaging is not damaged, and the expiration date has not elapsed. Do NOT resterilize or reuse the Impella® 5.0 Catheter. It is a disposable device and is intended for single use only. Reuse, reprocessing, or resterilization may compromise the structural integrity of the catheter and/or lead to catheter failure which, in turn, may result in patient injury, illness, or death. Protect the Impella® 5.0 Console and the Impella® Power Supply from direct sunlight or exposure to excessive temperature (max. 40°C) to prevent malfunction. To prevent overheating and improper operation, do NOT block the cooling vents of the Impella® Power Supply while it is operating. Do NOT place the Impella® Console on top of the Impella® Power Supply. The Li-Ion batteries must be charged for 10 hours prior to system operation. After being unplugged, the Impella® Console will operate for at least 60 minutes after the batteries have been fully charged. Holding down the ON key on the Impella® Console for longer than 3 seconds during operation will cause the Impella® Console to go into emergency stop mode. Supplies Needed • Impella® 5.0 System (Impella® 5.0 Catheter, accessories, and Impella® Console) • 500 cc bag of dextrose solution for purge solution (20% recommended; 5% to 40% acceptable) with 50 IU heparin/mL Equipment Set-up Before starting the Impella® Console, check that the rechargeable batteries are fully charged. Have a backup console available in the unlikely event of console failure. 1.Place the Impella® Power Supply and the Impella® Console next to each other on a horizontal surface or on the optional Impella® Console cart. Figure 5.1 Placing the Impella® Power Supply Next to the Impella® Console 2.Connect the power cord to the AC power connector on the rear panel of the Impella® Power Supply and to an AC outlet. Impella® 5.0 with the Impella® Console 5.1 3.Connect the Impella® Power Supply cable to the Impella® Console connector on the rear panel of the Impella® Power Supply (see Figure 5.2). Figure 5.2 Connecting the Power Supply Cable to the Power Supply 4.Connect the Impella® Console to the Impella® Power Supply with the power supply cable (see Figure 5.3). Avoid overtightening the screw. Figure 5.3 Connecting the Power Supply Cable to the Console If CONNECTED Light Not On If the CONNECTED light is not on, turn the Impella® Power Supply off and on. If this does not solve the problem, turn the Impella® Power Supply off and on again. Do not use the Impella® Console if the CONNECTED light is not on after two attempts. 5.Turn on the Impella® Power Supply using the power switch located on the rear panel of the power supply. The Impella® Console is powered up automatically when the Impella® Power Supply is turned on. Power switch Figure 5.4 Turning the Power Supply On 6.Make sure that the Impella® Power Supply is functioning properly by checking for all of the following conditions: • ON/OFF switch is illuminated. • Impella® Power Supply fan is running. • Green POWER and CONNECTED lights on the Impella® Power Supply are on. • AC power icon appears on the Impella® Console display. 5.2 Instructions for Use & Clinical Reference Manual Preparing the Purge Fluid To troubleshoot purge system alarms not listed in this manual, refer to the appropriate infusion pump instructions for use manual. Shaded Steps All shaded steps require sterile technique. 1.Connect the infusion pump to an AC outlet. 2.Turn the infusion pump on. 3.Add heparin to achieve 50 IU heparin/mL in 20% dextrose solution. 4.Connect the purge tubing to the 20% dextrose solution. 5.Flush the purge tubing manually and close the roller clamp. 6.Using sterile technique, open the Impella® Combitrans-Monitoring Set (CM-Set) and secure the male luerlock to the sterile field. 7.Hand the CM-Set’s green pressure transducer off the sterile field and connect it to the purge tubing. Flush the CM-Set using one of the following methods: a.Manual flush: Open the roller clamp and apply manual pressure to the purge solution bag until fluid exits the distal end of the purge tubing and the tubing is flushed of air. Close the roller clamp. Proceed to Step 8. b.Alternative flush: After attaching the CM-Set to the infusion set, perform Steps 8–12 as shown below. Press Start/Stop to start the infusion. Then press and hold the “Bolus” soft button. A second “Bolus” soft button will be displayed. Press and hold the two “Bolus” soft buttons until the CM-Set is flushed. Press Start/Stop to stop the infusion. Turning on the Infusion Pump Do NOT turn on the infusion pump with the purge tubing in place. This will cause the downstream occlusion sensor to be miscalibrated, leading to false alarms. 8.Open the infusion pump door by pressing the Door Open button at the top right corner. Push and hold open the clamp at the top of the fluid path. 9.Insert the purge tubing into the clamp and release it. Thread the tubing down the fluid path, making sure the tubing is pressed into the Air-in-Line detector at the bottom of the fluid path. Close the door and open the roller clamp. 10.The infusion pump displays “Is This a New Therapy?” Press the YES key to confirm. 11.The infusion pump displays “Enter Infusion Rate.” Enter a rate of 7 mL/h. 12.Press VTBD (Volume To Be Delivered) and enter the volume (in mL) at which the infusion pump should alarm, indicating the need to replace the purge fluid bag. . Press OK. Impella® 5.0 with the Impella® Console Note for Vista basic Users If you are using the B. Braun Vista basic Infusion Pump and it has been used for a prior infusion, it will ask if you want to continue with prior therapy after you connect the purge tubing. Press NO. 5.3 5 Using the Impella® 5.0 System Refer to the instructions for use supplied with the infusion pump being used before using the infusion pump with the Impella® 5.0 System. Preparing the Impella® 5.0 System Fluoroscopy is required to guide placement of the Impella® 5.0 Catheter. The small placement guidewire must be reliably observed at all times. Avoid manual compression of the inlet, outlet, or sensor areas of the cannula assembly. The sterile components of the Impella® 5.0 System can be used only if the sterilization indicators show that the contents have been sterilized, the packaging is not damaged, and the expiration date has not elapsed. Do NOT remove the Impella® 5.0 Catheter over the length of the placement guidewire. Handle with care. The Impella® 5.0 Catheter can be damaged during removal from packaging, preparation, insertion, and removal. Do NOT bend, pull, or place excess pressure on the catheter or mechanical components at any time. Do NOT kink or clamp any part of the Impella® 5.0 Catheter. Protect the Clear Sidearm and Catheter The clear sidearm of the Impella® 5.0 Catheter between the pressure reservoir and red Impella® plug can become kinked. Support the pressure reservoir and the length of the catheter as you remove the Impella® 5.0 Catheter from the tray. Avoid allowing the pressure reservoir to hang freely from the red Impella® plug and avoid bending the catheter near the red Impella® plug. Sensitive Medical Device The Impella® 5.0 Catheter is a sensitive medical device with extremely fine tolerances. In particular, the inlet and outlet areas of the catheter assembly may be damaged if subjected to strong external forces. 5.4 Connecting the Impella® 5.0 Catheter and Connector Cable 1. S lowly open the package and carefully remove the Impella® 5.0 Catheter using sterile technique. Inspect the catheter, including its connector, for damage. 2.Remove the connector cable from its package using sterile technique. Inspect the cable for damage, including damage to the connector pins at the console end. 3.Secure the black end of the cable to the sterile field. 4. Insert the catheter plug into the connector cable socket (black end). The tab and the slot must be aligned during connection (see Figure 5.5). Catheter Plug Connector Cable Socket (black end) Figure 5.5 Connecting the Impella® 5.0 Catheter to the Connector Cable 5.Pull back on the connection to make sure that the plug has snapped into place. Instructions for Use & Clinical Reference Manual Connector Cable Plastic Clip Figure 5.6 Snapping Plastic Clip to Connector Cable 7.Insert the blue plug of the connector cable into the catheter connector socket on the right side of the Impella® Console. Using the arrows as guides, align the tab and slot during connection as shown in Figure 5.7. A single beep followed by a series of three beeps confirms the connection. Preventing Purge Line Kinking To prevent purge line kinking, be sure to snap the plastic clip on the pressure reservoir to the correct location on the connector cable. If the clip is placed too close to the gray connector, the purge line may become kinked and fluid will not flow freely. During the procedure, check the clip regularly to make sure it remains correctly positioned, especially after changing the purge tubing. Console Color Coding The connectors on the Impella® Console are color-coded. The pressure connector socket on the left side of the console has yellow markings that match the yellow plug on the end of the pressure transducer cable; the catheter connector socket on the right side of the console has blue markings that match the blue plug on the end of the connector cable. Align the arrow on the connector cable with the arrow on the catheter connector socket Figure 5.7 Attaching the Connector Cable to the Console Impella® 5.0 with the Impella® Console 5.5 5 Using the Impella® 5.0 System 6.Snap the plastic clip (located on the pressure reservoir of the clear sidearm) to the connector cable as shown in Figure 5.6. Connecting the Pressure Transducer Cable 1.Once the Impella® 5.0 Catheter is plugged into the console, the console automatically initiates the startup procedure for the Impella® 5.0 System and displays the screen shown in Figure 5.8. P-Perf P0 LV4509 Flow [L/min] 100% ?.? Purge Pressure [mmHg] Pump Test -500 0:01 < Connect the gray pressure measurement cable to the left socket on the console and to the purge pressure transducer... < 10 s Figure 5.8 Connect Pressure Transducer Cable Screen 2.Connect the gray pressure transducer cable to the left socket on the console (as shown in Figure 5.9) and to the purge pressure (CM-Set) transducer. Figure 5.9 Connecting the Pressure Transducer Cable to the Console 5.6 Instructions for Use & Clinical Reference Manual P-Perf P0 LV4509 5 Using the Impella® 5.0 System 3.Once the gray cable is connected to the console and transducer, the console displays the screen shown in Figure 5.10. Flow [L/min] 100% ?.? Purge Pressure [mmHg] Pump Test 24 0:34 < 1. Flush purge line and connect to reservoir / filter on catheter 2. Flush purge lumen until purge pressure exceeds 350 mmHg. < 10 s Figure 5.10 Flush Purge Line and Purge Lumen Screen Priming the Impella® 5.0 Catheter Purge Lumen 1.Connect the CM-Set to the yellow check valve on the clear sidearm (filter unit) of the catheter as shown in Figure 5.11. Alternative Method for Producing the Bolus Press and hold the “Bolus” soft button. A second “Bolus” soft button will be displayed. Press and hold the two “Bolus” soft buttons until the Impella® Console beeps, indicating that the purge pressure has reached 350 mmHg. Stopping the Bolus Figure 5.11 Connecting CM-Set to Impella® 5.0 Catheter 2.Deliver an 8 cc bolus or use the alternative method for producing the bolus described in the sidebar. Impella® 5.0 with the Impella® Console If the bolus is not stopped prior to reaching 700 mmHg, the Impella® Console will alarm due to high purge pressure. The infusion pump will alarm and stop the bolus at 1000 mmHg. 5.7 3.Once the purge pressure reaches 350 mmHg, the console will switch to the “Ready to Insert” screen shown in Figure 5.12. P-Perf Flow [L/min] P0 LV4509 100% ?.? 2 1 Placement Signal [mmHg] Pump Test 1:15 Impella catheter is ready to be inserted 10 s Figure 5.12 Ready to Insert Screen 4. Confirm fluid exiting the catheter, as shown in Figure 5.13. Figure 5.13 Discharging Fluid from Catheter Outlet Area 5. T he “Ready to Insert” screen will switch to the placement signal screen within 10 seconds. 5.8 Instructions for Use & Clinical Reference Manual Inserting the Impella® 5.0 Catheter Fluoroscopy is required to guide placement of the Impella® 5.0 Catheter. The small placement guidewire must be reliably observed at all times. Avoid manual compression of the inlet, outlet, or sensor areas of the cannula assembly. Do NOT kink or clamp any part of the Impella® 5.0 Catheter. Handle with care. The Impella® 5.0 Catheter can be damaged during removal from packaging, preparation, insertion, and removal. Do NOT bend, pull, or place excess pressure on the catheter or mechanical components at any time. 1. Identify the femoral artery and perform a cut-down of 3 to 5 cm. Use Fluoroscopy for Placement Impella® 5.0 Catheter performance will be compromised if correct placement cannot be confirmed. While other imaging techniques, such as transesophageal echocardiography (TEE), can help confirm the position of the Impella® 5.0 Catheter after placement, TEE does not allow visualization of the entire catheter assembly or guidewire and is inadequate for reliably placing the Impella® 5.0 Catheter across the aortic valve. 2. E xpose the femoral artery. Wrap vessel loops, one distal and one proximal to the subsequent point of incision, one and a half times around the artery. Make the vessel loops as far apart as possible (see Figure 5.14). Guidewire U-Stitches Femoral Artery Proximal Vessel Loop Distal Vessel Loop Figure 5.14 Cut-Down Insertion of the Impella® 5.0 Catheter 3.To prepare the repositioning sheath, remove the luer plug at the end of the sidearm tube and flush the tube with 0.9% NaCl solution. Place the luer plug back in the sidearm tube and secure the plug. Impella® 5.0 with the Impella® Console 5.9 5 Using the Impella® 5.0 System NOTE – Proper surgical procedures and techniques are the responsibility of the medical professional. The described procedure is furnished for information purposes only. Each physician must evaluate the appropriateness of the procedure based on his or her medical training and experience, the type of procedure, and the type of systems used. Impella® 5.0 Catheter Use in Open Heart Surgery 4.Make the incision as close as possible to the distal loop. Insert a 6 Fr diagnostic catheter with no side holes (AL1 or Multipurpose without side holes recommended) over a diagnostic 0.035 inch or 0.038 inch guidewire into the left ventricle. If the Impella® 5.0 Catheter is used in the OR as part of open heart surgery, manipulation may be performed only through the 9 Fr steering catheter. Direct manipulation of the catheter assembly through the aorta or ventricle may result in serious damage to the Impella® 5.0 Catheter and serious injury to the patient. 5.Remove the diagnostic guidewire and exchange it for the supplied 0.025 inch placement guidewire. 6. H old tension on the proximal vessel loop to prevent bleeding. Straighten the blue pigtail and thread it over the 0.025 inch placement guidewire (see Figure 5.15). Wet the cannula with sterile water and backload the catheter onto the placement guidewire. One or two people can load the catheter on the guidewire. 1cm Inlet Catheter (with marking) Avoid Damaging Inflow Area During placement of the Impella® 5.0 Catheter, take care to avoid damage to the inflow area while holding the catheter and loading the placement guidewire. Outlet Pigtail Guidewire Sensor Figure 5.15 Guidewire Placement One-person technique a.Advance the placement guidewire into the Impella® 5.0 Catheter and stabilize the cannula between the fingers. This prevents pinching of the inlet area. The placement guidewire must exit the outlet area on the inner radius of the cannula, as shown in Figure 5.15, and align with the straight black line on the catheter. The cannula can be hyperextended as necessary to ensure the placement guidewire exits on the inner radius of the cannula. Two-person technique b.The scrub assistant can help stabilize the catheter by holding the catheter proximal to the motor. This will allow the implanting physician to visualize the inner radius. The placement guidewire must exit the outlet area on the inner radius of the cannula and align with the straight black line on the catheter. The physician can focus on advancing the placement guidewire and, if the cannula needs to be hyperextended, the scrub assistant is available to assist. 7.Make a transverse incision at the guidewire for the 21 Fr catheter. Use U-stitches (see Figure 5.14) instead of purse string sutures to avoid stenosis of the vessel after explantation. GP IIb-IIIa Inhibitors 8. Administer heparin and achieve ACT of at least 250 seconds. If the patient is receiving a GP IIb-IIIa inhibitor, the Impella® 5.0 Catheter can be inserted when ACT is 200 or above. 9. Insert the catheter into the vessel and advance along the 0.025 inch placement guidewire until resistance is met at the proximal vessel loop. 5.10 Instructions for Use & Clinical Reference Manual 11.Advance the repositioning sheath, located on the catheter shaft, through the incision and into the femoral artery until bleeding is controlled. Secure the sheath outside of the vessel using the supplied suture loop. To prevent device failure, do not start the Impella® 5.0 Catheter until the placement guidewire has been removed. Do NOT remove the Impella® 5.0 Catheter over the length of the placement guidewire. 12.Stabilize the guidewire and repositioning sheath and advance the catheter through the sheath. Follow the catheter under fluoroscopy as it is advanced across the aortic valve, positioning the inlet area of the catheter approximately 4 cm below the aortic valve annulus and in the middle of the ventricular chamber, free from the mitral valve chordae. Be careful not to coil the guidewire in the left ventricle. (Refer to Section 6 for information about console waveforms during placement.) 13.When the catheter is correctly positioned, slightly loosen the proximal vessel loop and remove the 0.025 inch guidewire. Leave at least 2 to 3 cm of the repositioning sheath inside the vessel. 14.Tighten the prepared U-stitches to seal the sheath. 15.Loosen the distal vessel loop. Then loosen the proximal vessel loop. Impella® 5.0 with the Impella® Console Do NOT Touch Inlet, Outlet, or Sensor Areas While feeding the Impella® 5.0 Catheter through the femoral artery, hold the catheter at the cannula or motor housing. Do NOT touch the inlet, outlet, or sensor areas. Maintaining ACT After insertion of the catheter (and until explant), ACT should be maintained at 160 to 180 seconds. 5.11 5 Using the Impella® 5.0 System 10.Loosen the proximal vessel loop and advance the catheter into the vessel. When the motor housing is entirely past the proximal vessel loop, temporarily tighten the loop to control bleeding. Positioning and Starting the Impella® 5.0 Catheter Retrograde flow will occur across the aortic valve if the Impella® 5.0 Catheter is set at performance level P0. 1.While the catheter is being advanced in the aorta, the initial placement signal has the characteristics shown in Figure 5.16. The inlet area of the catheter has not passed the aortic valve. Flow [L/min] P-Perf P0 LV4509 100% 0.0 8 –2 Placement Signal [mmHg] 10 s Figure 5.16 Initial Placement Signal 2.Gently advance the catheter forward until a pulsatile waveform is present on the placement signal screen (see Figure 5.17). This signal is generated when the inlet area of the catheter crosses the aortic valve. P-Perf Importance of Proper Catheter Placement When the catheter is not correctly placed, there is no effective unloading of the ventricle (hydraulic short circuit). The patient may not be profiting from the flow rate shown on the console. P0 LV4509 Flow [L/min] 100% 0.0 33 2 Placement Signal [mmHg] 10 s Figure 5.17 Placement Signal as the Catheter Crosses the Aortic Valve 5.12 Instructions for Use & Clinical Reference Manual 4.Increase performance level from P0 to P2. The catheter operation indicator to the right of the battery symbol at the top of the screen will rotate when the catheter is operating. P-Perf P9 LV4509 Flow [L/min] 100% 5.0 35 1 Dual Signal I Confirming Differential Pressure Sensor Zero If the lower value of the differential pressure signal is not within 6 mmHg of 0 when the Impella® 5.0 Catheter is in place across the aortic valve, zero the differential pressure sensor. (The zeroing process is described in section 6 of this manual.) P 10 s Figure 5.18 Confirming Placement on the Dual Signal Screen 5.Increase the performance level to P9 to confirm correct and stable placement. Evaluate the catheter position in the aortic arch and remove any excess slack. The catheter should align against the lesser curvature of the aorta rather than the greater curvature. Verify placement with fluoroscopy and with the dual signal screen (see Figure 5.18). 6.Make sure there is no bleeding at the transition from the repositioning sheath to the femoral artery. Close and dress the wound. Secure the repositioning sheath by suturing it to the skin using the yellow eyelet on the sheath. 7.Attach the anticontamination sleeve to the sheath. Lock it in place by turning clockwise. Secure the catheter by tightening the connected anchoring ring. 8.Carefully extend the anticontamination sleeve to maximum length and secure the end closest to the red Impella® plug by tightening the anchoring ring. Vascular Closure When securing the repositioning sheath, vascular closure may be difficult in obese patients with extensive adipose tissue. 9.Reposition the catheter as necessary. Optimal catheter position is indicated on the placement monitoring screen. Impella® 5.0 with the Impella® Console 5.13 5 Using the Impella® 5.0 System 3. C onfirm that the console displays a pulsatile waveform and the inlet area of the Impella® 5.0 Catheter is approximately 4 cm below the aortic valve. Replacement Time If the purge flow is more than 7 mL/h or the dextrose concentration is less than 20%, replacement time will be less than 2 minutes. Replacement should always be performed as quickly as possible. Replacing the Purge Tubing When replacing the purge tubing, the replacement process must be completed within 2 minutes. The Impella® 5.0 Catheter may be damaged if replacement takes longer than 2 minutes. 1.Prepare a 20% dextrose solution with 50 IU heparin/mL. 2.Connect new purge tubing and CM-Set to the 20% dextrose solution. 3.Flush the purge tubing and CM-Set manually and close the roller clamp. 4.Press BOL on the infusion pump and enter a bolus release of 0.5–1 mL. Press OK to start the bolus. Multiple bolus releases may be required to reach 600 mmHg. 5.When the purge pressure on the Impella® Console reaches 600 mmHg, stop the bolus. 6.Press the red STOP key on the infusion pump to stop the infusion, then close the roller clamp. 7.Disconnect the current CM-Set from the catheter at the yellow check valve and from the gray pressure cable. 8.Open the door to the infusion pump and remove the current purge tubing. 9.Connect the new CM-Set to the catheter at the yellow check valve and to the gray pressure cable. Open the roller clamp. 10.Load the new purge tubing into the infusion pump. Close the infusion pump door. 11.Once the change out has occurred and the new purge tubing is threaded through the infusion pump, press VTBD to enter the new bag volume. 12.Press C on the keypad to clear the remaining volume and then enter the volume of the new bag. 13.Press OK to accept the new value. 14.Press the green Start button to start the infusion. 15.Press BOL on the infusion pump and enter a bolus release of 0.5–1 mL. Press OK to start the bolus to increase purge pressure back to the normal operating range. 16.If the purge pressure on the Impella® Console reaches 600 mmHg, press STOP on the infusion pump. Normal operating range is 300–700 mmHg. 5.14 Instructions for Use & Clinical Reference Manual Purge System Troubleshooting Low Purge Pressure If at any time during the course of support with the Impella® 5.0 Catheter, the Impella® Console alarms “Low Purge Pressure,” follow the instructions below. An Impella 5.0 Catheter in stable operation should never require an abrupt or significant increase in purge flow to maintain purge pressure. Stable operation is defined as operation in which the following criteria are met: ® Purge Pressure Range Purge pressure range: 300-700 mmHg • No leaks in any purge tubing connections • No recent changes in dextrose concentration • Constant performance level • Purge flow maintained with only small incremental changes in purge flow After confirming stable operation as defined above, if you observe a condition in which purge pressure drops and purge flow requires an abrupt or significant increase (eg, doubling the purge flow) to maintain purge pressure over 300 mmHg, this could be an indication of catheter damage. The steps that follow describe how to respond to the “Low Purge Pressure” alarm shown in Figure 5.19 if the criteria for stable operation listed above are met; however, the nature of the situation will determine the starting point for your response as shown in Table 5.1. Warning 0:06 Low Purge Pressure Increase Purge Rate! Check Purge System! Figure 5.19 Low Purge Pressure Screen Impella® 5.0 with the Impella® Console 5.15 5 Using the Impella® 5.0 System To troubleshoot purge system alarms not listed in this manual, refer to the appropriate infusion pump instructions for use manual. Table 5.1 Handling Low Purge Pressure Alarms Observation Action Stable operation with purge flow between 4–20 mL/h Follow the instructions outlined below, starting at Step 1. Stable operation with abrupt or Immediately go to Step 7 of the instructions outlined below. significant increase in purge flow needed to maintain purge pressure Stable operation with dextrose ≥20% and purge flow ≥30 mL/h Immediately go to Step 7 of the instructions outlined below. Heparin and Dextrose Infusion Rates 1. P ress BOL on the infusion pump and enter a bolus release of 0.5 mL. Press OK to start the bolus. The purge fluid contains heparin and dextrose. When changing the purge rate, carefully assess how the new heparin and dextrose infusion rates may affect the patient. For example, significantly increasing the purge rate may cause bleeding by increasing the amount of heparin delivered to the patient. As a result, the patient’s systemic heparin dose may need to be adjusted. 2. Inspect the purge system for leaks. Refer to Tables 5.2 and 5.3 for Abiomed-recommended dextrose and heparin concentrations in purge fluid. 3. P ress STOP if the Impella® Console shows purge pressure approximately 600 mmHg. Repeat Step 1 if necessary to reach 600 mmHg. If needed, increase infusion rate on the infusion pump until the purge pressure is maintained within the range of 300 to. 700 mmHg. 4. Switch to the 5 minute scale screen and monitor the purge pressure. 5. If the pressure remains stable, no other action is required. If the purge pressure is not stable and requires repeated boluses, proceed to Step 6. Figure 5.20 is an example of normal, stable behavior as displayed on the 5 minute . scale screen. • A single bolus is delivered to raise purge pressure to 600 mmHg. • Pressure is maintained above 300 mmHg. • Normal “settling” is shown after bolus. P-Perf P9 LV4509 Flow [L/min] 100% Purge Pressure [mmHg] 5.0 449 < < 5 min Figure 5.20 Normal Purge Pressure on 5 Minute Scale Screen 5.16 Instructions for Use & Clinical Reference Manual 5 Using the Impella® 5.0 System 6. If repeated boluses within a 5 minute period are required (see Figure 5.21) and increasing the infusion rate and increasing the dextrose concentration have not been effective, proceed to Step 7. Figure 5.21 is an example of abnormal purge pressure behavior as displayed on the. 5 minute scale screen. • Repeated boluses are given to try to maintain purge pressure above 300 mmHg. • Pressure is not maintained. • Time for the pressure to drop below 300 mmHg can vary from as little as 10 seconds to as much as 60 seconds. • Look for signature “saw tooth” pattern. Flow [L/min] P-Perf P9 LV4509 100% Purge Pressure [mmHg] 5.0 60 < < 5 min Figure 5.21 Abnormal Purge Pressure on 5 Minute Scale Screen 7. If the low purge pressure alarms remain unresolved for more than 5 minutes, this may be a sign of catheter damage. Complete the following steps immediately: a.Change the catheter performance level to P1 by pressing the P-PERF button and the down arrow until P1 is displayed in the top left corner of the screen. Press OK button. b.Slowly pull back on the Impella® 5.0 Catheter until the catheter is in the descending aorta (approximately 20 cm for an average size patient; 1 cm marks are available . on the catheter). c.Turn off the Impella® 5.0 Catheter by pressing the P-PERF button and the down arrow until P0 is displayed in the top left corner of the console screen. Press OK button. dTurn off the Impella® Console and purge flow. e.Remove the Impella® 5.0 Catheter with the use of fluoroscopic imaging. If no fluoroscopy is available, leave the catheter in the descending aorta until fluoroscopy is available for visual assistance during removal of the catheter. Impella® 5.0 with the Impella® Console 5.17 High Purge Pressure High purge pressure alarms, if left unresolved, can deprive the Impella® 5.0 Catheter of the continuous flow of purge fluid needed to protect the motor. If blood enters the motor, the motor will have to work harder to maintain the set performance level. Motor current may increase gradually or sharply and the Impella® 5.0 Catheter may stop running. If the purge pressure exceeds 700 mmHg, the Impella® Console displays the alarm message shown in Figure 5.22. To resolve the high purge pressure alarm, perform the following steps. Warning 0:06 High Purge Pressure Unresolved High Purge Pressure Alarms Decrease Purge Rate! High purge pressure alarms that are not resolved by the recommendations provided could be an indication of a kink in the catheter located within the artery. In this case, the motor is no longer being purged and will eventually stop. Consider replacing the Impella® 5.0 Catheter. Check Purge System! Figure 5.22 High Purge Pressure Screen 1.Make sure that the clip on the pressure reservoir of the clear sidearm is not attached too close to the gray connector causing the purge line on the clear sidearm to kink.. If the clear sidearm is kinked, reposition the clip further down the connector cable away from the gray connector to remove the kink and allow purge fluid to flow freely. 2.Carefully inspect the purge tubing, clear sidearm, and entire visible length of the catheter for kinks, keeping in mind that kinks may be difficult to see and the purge lumen inside the clear sidearm may remain kinked after you “unkink” the clear sidearm. 3.If you do not detect any kinks, decrease the purge flow rate by 1 mL/h until the purge pressure is less than 700 mmHg. Do not decrease purge flow rate to less than 4 mL/h. 4.If purge pressure remains above 700 mmHg, decrease dextrose concentration to 5%. 5.18 Instructions for Use & Clinical Reference Manual Purge Pressure Management Strategies 5 Using the Impella® 5.0 System Figure 5.23 summarizes purge pressure management strategies. 20% Dextrose Solution with 50 Units Heparin/mL Purge flow rate 7 mL/h 300 mmHg Low Purge Pressure less than 300 mmHg Purge flow less than 20 mL/h 700 mmHg Purge Pressure between 300 and 700 mmHg Purge flow greater than 20 mL/h High Purge Pressure greater than 700 mmHg Purge flow between 4 mL/h and 20 mL/h 1. Deliver 0.5 mL bolus. 1. Deliver 1 mL bolus. 1. Purge pressure OK. 2. Increase purge fluid infusion rate by 0.5 mL/h to maintain purge pressure > 300 mmHg. 2. Increase dextrose concentration.* 2. Monitor after changes in performance (P) level. Purge flow less than 4 mL/h Purge flow greater than 4 mL/h 1. Inspect tubing for kinks or closed clamps. 1. Decrease infusion rate by 1 mL/h. 2. Decrease dextrose concentration.* 3. Adjust infusion rate as needed. 3. Adjust infusion rate as needed. * Refer to the instructions for changing purge fluid earlier in this section. Changes in purge rate or dextrose concentration can take up to 1 hour to become effective. Figure 5.23 Purge Management Tables 5.2 and 5.3 describe the amounts of dextrose and heparin delivered in the purge fluid. The green shading in the tables indicates the Abiomed-recommended concentrations of dextrose and heparin in the purge fluid. Table 5.2 Amount of Dextrose Delivered in the Purge Fluid (g/h) Purge Fluid Infusion Rate 4 Dextrose Concentration 7 8 10 12 14 16 18 20 mL/h mL/h mL/h mL/h mL/h mL/h mL/h mL/h mL/h 5% 0.2 0.35 0.4 0.5 0.6 0.7 0.8 0.9 1 10% 0.4 0.7 0.8 1 1.2 1.4 1.6 1.8 2 20% 0.8 1.4 1.6 2 2.4 2.8 3.2 3.6 4 40% 1.6 2.8 3.2 4 4.8 5.6 6.4 7.2 8 Note: Green shading indicates the Abiomed-recommended concentration of dextrose in the purge fluid Table 5.3 Amount of Heparin Delivered in the Purge Fluid (IU/h) Purge Fluid Infusion Rate 4 Heparin Concentration 7 8 10 12 14 16 18 20 mL/h mL/h mL/h mL/h mL/h mL/h mL/h mL/h mL/h 12.5 IU/mL 50 87.5 100 125 150 175 200 225 250 25 IU/mL 100 175 200 250 300 350 400 450 500 50 IU/mL 200 350 400 500 600 700 800 900 1000 Note: Green shading indicates the Abiomed-recommended concentration of heparin in the purge fluid Impella® 5.0 with the Impella® Console 5.19 Downstream Occlusion If a “downstream occlusion” alarm occurs on the infusion pump (see Figure 5.24), the purge fluid flow stops. 1.Inspect the purge tubing for any kinks or defects. 2.Make sure the roller clamp is open. 3. Press the START key to resume the infusion. 4.If the alarm persists, inspect the tubing down the fluid path. Close the door and press the START key to resume the infusion. Figure 5.24 Downstream Occlusion Alarm Patient Weaning Weaning the patient from the Impella® 5.0 Catheter is at the discretion of the physician. The following weaning protocols are provided as guidance only. Rapid Weaning 1. Initiate rapid weaning by decreasing catheter performance level in 2-level steps at intervals of several minutes (for example, P6 to P4 to P2). Do NOT decrease the performance level to below P2 until just before removing the catheter from the ventricle. 5.20 Instructions for Use & Clinical Reference Manual 3. If the patient’s hemodynamics remain stable, decrease the performance level to P1, pull the catheter into the aorta, and stop the motor by decreasing the performance level to P0. 4. Explant the catheter. 5.Follow institutional guidelines for arterial closure. 6.Disconnect the connector cable from the Impella® Console and the catheter by sliding back the locking sleeves (at the <= => markings) and removing the plugs . (see Figure 5.25). Do NOT twist the plugs because the contact pins may be damaged. Disconnecting Connector Cable When disconnecting the connector cable from the console and catheter, do NOT twist the plugs because the contact pins may be damaged. Figure 5.25 Disconnecting the Connector Cable from the Console 7.Press MENU. Select “Switch Off Console” and press OK. Remove the Impella® 5.0 Catheter With Care Slow Weaning 1.Initiate slow weaning by decreasing catheter performance level in 2-level steps over time as cardiac function allows (for example, P6 to P4 to P2). Do NOT decrease the performance level to below P2 until just before removing the catheter from the ventricle. Removal of the Impella® 5.0 Catheter must be completed with care to avoid damage to the catheter assembly. 2.When the performance level has been reduced to P2, maintain the patient on P2 support until the patient’s hemodynamics remain stable before discontinuing circulatory support. 3.If the patient’s hemodynamics remain stable, decrease the performance level to P1, pull the catheter into the aorta, and stop the motor by decreasing the performance level to P0. 4.Explant the catheter. 5.Follow institutional guidelines for arterial closure. 6.Disconnect the connector cable from the Impella® Console and the catheter by sliding back the locking sleeves (at the <= => markings) and removing the plugs (see Figure 5.25). Do NOT twist the plugs because the contact pins may be damaged. 7.Press MENU. Select “Switch Off Console” and press OK. Impella® 5.0 with the Impella® Console 5.21 5 Using the Impella® 5.0 System 2. W hen the performance level has been reduced to P2, maintain the patient on P2 support for at least 10 minutes before discontinuing circulatory support. 6 Patient Management Topics Patient Management Overview.............................................................6.1 General Patient Care Considerations ...............................................6.1 Transport Within the Hospital............................................................ 6.2 Right Heart Failure................................................................................. 6.2 ECG Interference...................................................................................... 6.3 Latex........................................................................................................... 6.3 Positioning and Placement Devices................................................... 6.3 Use of Echocardiography for Positioning of the Impella® 5.0 Catheter.............................................................................. 6.4 Background......................................................................................................6.4 Correct Impella® 5.0 Catheter Position.............................................................6.6 Impella® 5.0 Catheter Too Far into the Left Ventricle.........................................6.6 Impella® 5.0 Catheter Inlet in the Aorta............................................................ 6.7 Impella® 5.0 Catheter in Papillary Muscle......................................................... 6.7 Color Doppler Echocardiography..................................................................... 6.10 Post-insertion Positioning (PIP) Checklist........................................................ 6.11 Understanding and Managing Impella® 5.0 Catheter Position Alarms............................................. 6.11 Correct Impella® 5.0 Catheter Position........................................................... 6.12 Impella® 5.0 Catheter Fully in Ventricle or Fully in Aorta................................. 6.13 Low Native Heart Pulsatility............................................................................ 6.14 Impella® 5.0 Catheter Outlet Area on or Near Aortic Valve............................. 6.15 Suction......................................................................................................6.16 Hemolysis..................................................................................................6.16 Pressure Sensor Drift and Sensor Failure......................................6.18 Sensor Drift and Effects on Flow..................................................................... 6.18 Sensor Failure and Effects on Flow Calculations.............................................. 6.19 Sensor Failure and Effects on Position Control................................................ 6.19 Suction Detection During Sensor Drift or Sensor Failure.................................. 6.19 Operating the Impella® 5.0 Catheter Without Heparin in the Purge Solution........................................................... 6.20 How to Change to A Backup Console................................................ 6.20 Patient Management Overview 6 Patient Management Topics The information and instructions in this section of the manual are not intended to supersede established medical procedures concerning patient care. Best practice, as determined by the medical community, should always be observed. In each case, the clinician must determine whether the application of information provided is appropriate for the particular clinical setting. General Patient Care Considerations • Do not raise the head of the bed to higher than a 30-degree angle. • Use knee immobilizer as needed to maintain access site straight. • Perform dressing changes per hospital protocol, using aseptic technique. • Assess access site for bleeding and hematoma. • Be careful not to pull on the Impella® 5.0 Catheter when transferring a patient from one bed to another. • To prevent purge line kinking, do not allow the red Impella® plug to hang freely from the catheter and do not bend the catheter near the red Impella® plug. Consider attaching the red Impella® plug and catheter to a short armboard to prevent the catheter from kinking near the plug. • Use care when moving or turning a patient; the Impella® 5.0 Catheter may move out of position and cause a positioning alarm. • Monitor pedal pulses. Impella® 5.0 with the Impella® Console 6.1 Battery Power Note If the Impella® Console is allowed to discharge completely and the system shuts down due to low battery, the console will need to charge for an extended period of time before the console will turn on again. Transport Within the Hospital Patients supported with the Impella® 5.0 System may require transport within the hospital for various reasons. Transport can be safe and simple for patients supported with Impella® 5.0 Catheter. Considerations for transport within the hospital: • The Impella® 5.0 System is designed to operate on battery power for at least 1 hour. • Confirm that the battery capacity displayed on the console is 100% and the battery capacity of the infusion pump is sufficient. • If transport time might be longer than 1 hour, bring an extension cord and confirm that you will be able to connect the console to AC power once you arrive at your destination. • Use care when rolling the Impella® 5.0 System cart, and pay close attention when going over thresholds and through elevator doors. • Do not stress the Impella® 5.0 Catheter cable from the console to the red Impella® plug. Right Heart Failure The Impella® 5.0 Catheter is a left-side support device only. Patients being supported by the Impella® 5.0 Catheter should be monitored for signs of right heart failure. Caregivers should monitor the patient closely for the following potential signs of right heart failure: • Reduced output from the Impella® 5.0 Catheter • Suction alarms • Elevated filling pressures (CVP) • Signs of liver failure • Elevated pulmonary pressures If the patient is exhibiting signs of right heart failure, the clinical team should assess the need for biventricular support. 6.2 Instructions for Use & Clinical Reference Manual ECG Interference 6 Patient Management Topics Operating the Impella® Console with the Impella® Power Supply may cause interference with electrocardiogram (ECG) signals. Check the electrode pads and leads for good fixation and contact. If interference persists, activate the 50/100 Hz band-elimination filter or the 60/120 Hz band-elimination filter (also known as notch filter) on your ECG device. The filter frequency will be based on the AC power frequency for the country in which you are operating the equipment. If your ECG device does not have the appropriate filters, disconnect the Impella® Power Supply temporarily from AC power to obtain an undisturbed signal. Carefully observe the battery status while running the Impella® Console on battery power. Latex The Impella® 5.0 Catheter and all accessories approved by Abiomed are 100% latex free. Positioning and Placement Devices High quality fluoroscopic imaging is the best method for determining catheter position and is required for Impella® 5.0 Catheter placement. If fluoroscopy is not readily available, transesophageal echocardiography (TEE) or transthoracic echocardiography (TTE) is appropriate for determining catheter position. (TEE and TTE are discussed next in this section.) Alternative methods for determining catheter position include portable C-Arm fluoroscopy and chest x-ray. Notes on Imaging: All imaging technology represents the anatomy in two dimensions (2D). It is not possible to assess the interactions between the Impella® 5.0 Catheter and the intraventricular anatomy that occur in three dimensions (3D). Abiomed strongly recommends that the Impella® 5.0 Catheter be repositioned, even if the imaging view shows correct position. Alternative imaging techniques, such as transesophageal echocardiography (TEE), can be useful to confirm the position of the Impella® 5.0 Catheter after placement. However, TEE is inadequate to reliably perform initial placement across the aortic valve. TEE does not allow visualization of the entire catheter assembly or guidewire. Impella® 5.0 with the Impella® Console Use Fluoroscopy for Placement Impella® 5.0 Catheter performance will be compromised if correct placement cannot be confirmed. While other imaging techniques, such as transesophageal echocardiography (TEE), can help confirm the position of the Impella® 5.0 Catheter after placement, TEE does not allow visualization of the entire catheter assembly or guidewire and is inadequate for reliably placing the Impella® 5.0 Catheter across the aortic valve. 6.3 Use of Echocardiography for Positioning of the Impella® 5.0 Catheter Background Echocardiography is a commonly used tool for evaluating the position of the Impella® 5.0 Catheter relative to the aortic valve and other intraventricular structures post-placement. The best echocardiographic views for positioning the Impella® 5.0 Catheter in the left ventricle are a long axis transesophageal echocardiogram (TEE) or a parasternal long axis transthoracic echocardiogram (TTE). These long axis views allow you to see both the aortic valve and Impella® 5.0 Catheter inlet area. Evaluate the position of the Impella® 5.0 Catheter if the Impella® Console displays position alarms or if you observe lower than expected flows or signs of hemolysis. If the catheter does not appear to be correctly positioned, initiate steps to reposition it. The illustrations on the following page identify the structures you would expect to see in transesophageal echocardiography (top) and transthoracic echocardiography (bottom). In these illustrations, the Impella® 5.0 Catheter is positioned correctly; however, these depictions are stylized and in actual echocardiograms the pigtail and inlet and outlet areas may not be seen as distinctly. 6.4 Instructions for Use & Clinical Reference Manual Patient Management Topics Impella® 5.0 Catheter outlet area LA Mitral valve 6 Transesophageal Echocardiogram (TEE) of Impella® Catheter Location of mitral chordae Aortic valve Papillary muscle RV LV Impella® 5.0 Catheter inlet area Transthoracic Echocardiogram (TTE) of Impella® Catheter Impella® 5.0 Catheter inlet area RV Septum Impella® 5.0 Catheter outlet area Papillary muscle Aorta LV LA Location of mitral chordae Mitral valve Figure 6.1 Labeled TEE and TTE Images of the Impella® 5.0 Catheter Position Impella® 5.0 with the Impella® Console 6.5 Four Impella® 5.0 Catheter positions you are likely to encounter when examining echocardiograms from patients supported with the Impella® 5.0 Catheter include: • Correct Impella® 5.0 Catheter position • Impella® 5.0 Catheter too far into the left ventricle • Impella® 5.0 Catheter inlet in the aorta • Impella® 5.0 Catheter in papillary muscle The following pages describe each situation. Figure 6.2 illustrates a transesophageal echocardiogram (TEE) of each situation. Figure 6.3 illustrates a transthoracic echocardiogram (TTE) of each. Correct Impella® 5.0 Catheter Position For optimal positioning of the Impella® 5.0 Catheter, the inlet area of the catheter should be about 4 cm below the aortic valve annulus and well away from papillary muscle and subannular structures. The outlet area should be well above the aortic valve. If the Impella® 5.0 Catheter is correctly positioned, echocardiography will likely show the following, as depicted in . Figures 6.2a (TEE) and 6.3a (TTE): • Catheter inlet area about 4 cm below the aortic valve • Catheter outlet area well above the aortic valve (frequently not visible on TEE or TTE images) • Catheter angled toward the left ventricular apex away from the heart wall and not curled up or blocking the mitral valve Impella® 5.0 Catheter Too Far into the Left Ventricle If the Impella® 5.0 Catheter is positioned too far into the left ventricle, the patient will not receive the benefit of Impella® 5.0 Catheter support. Blood will enter the inlet area and exit the outlet area within the ventricle. Obstruction of the Impella® 5.0 Catheter inlet area can lead to increased mechanical forces on blood cell walls and subsequent hemolysis, which often presents as dark or blood-colored urine. If the Impella® 5.0 Catheter is too far into the left ventricle, echocardiography will likely show the following, as depicted in Figures 6.2b (TEE) . and 6.3b (TTE): • Catheter inlet area more than 4 cm below the aortic valve • Catheter outlet area across or near the aortic valve • Catheter too close to the heart wall or mitral valve 6.6 Instructions for Use & Clinical Reference Manual Impella® 5.0 Catheter Inlet in the Aorta 6 Patient Management Topics If the inlet area of the Impella® 5.0 Catheter is in the aorta, the patient will not receive the benefit of Impella® 5.0 Catheter support. The catheter will pull blood from the aorta rather than the left ventricle. In addition, suction is possible if the inlet area is against the wall of the aorta or valve sinus. If the inlet area of the Impella® 5.0 Catheter is in the aorta, echocardiography will likely show the following, as depicted in Figures 6.2c (TEE) and 6.3c (TTE): • Catheter inlet area in aorta or near the aortic valve • Catheter pigtail too close to the mitral valve Impella® 5.0 Catheter in Papillary Muscle If the inlet area of the Impella® 5.0 Catheter is too close to or entangled in the papillary muscle and/or subannular structures surrounding the mitral valve, it can affect mitral valve function and negatively impact catheter flow. If the inlet area of the catheter is lodged adjacent to the papillary muscle, the inflow may be obstructed, resulting in suction alarms. This positioning is also likely to place the outlet area too close to the aortic valve, which can cause outflow at the level of the aortic valve with blood streaming back into the ventricle, resulting in turbulent flow and hemolysis. If the Impella® 5.0 Catheter is too close to or entangled in the papillary muscle, echocardiography will likely show the following, as depicted in Figures 6.2d (TEE) . and 6.3d (TTE): • Catheter pigtail in papillary muscle • Catheter inlet area more than 4 cm below the aortic valve or lodged between papillary muscle and the myocardial wall • Catheter outlet area too close to the aortic valve The following figures depict transesophageal and transthoracic echocardiographic images of these four Impella® 5.0 Catheter positions. Figure 6.2 shows four transesophageal depictions of Impella® 5.0 Catheter position and Figure 6.3 shows four transthoracic depictions of Impella® 5.0 Catheter position. Impella® 5.0 with the Impella® Console 6.7 Correct Impella® 5.0 Catheter Position (TEE) 6.2a Impella® 5.0 Catheter Too Far into Left Ventricle (TEE) 6.2b • Catheter inlet area about 4 cm below the aortic valve • Catheter inlet area more than 4 cm below the aortic valve • Catheter outlet area well above the aortic valve • Catheter outlet area across or near the aortic valve • Catheter angled toward the left ventricular apex away from the heart wall and not curled up or blocking the mitral valve • Catheter too close to the heart wall or mitral valve Impella® 5.0 Catheter Inlet in Aorta (TEE) 6.2c Impella® 5.0 Catheter in Papillary Muscle (TEE) 6.2d • Catheter inlet area in aorta or near the aortic valve • Catheter pigtail in papillary muscle • Catheter pigtail too close to the mitral valve • Catheter inlet area more than 4 cm below the aortic valve or lodged between papillary muscle and the myocardial wall • Catheter outlet area too close to the aortic valve Figure 6.2 Transesophageal Echocardiographic (TEE) Illustrations of Impella® 5.0 Catheter Position 6.8 Instructions for Use & Clinical Reference Manual Impella® 5.0 Catheter Too Far into Left Ventricle (TTE) 6.3b 6 Correct Impella® 5.0 Catheter Position (TTE) 6.3a Patient Management Topics • Catheter inlet area about 4 cm below the aortic valve • Catheter inlet area more than 4 cm below the aortic valve • Catheter outlet area well above the aortic valve • Catheter outlet area across or near the aortic valve • Catheter angled toward the left ventricular apex away from the heart wall and not curled up or blocking the mitral valve • Catheter too close to the heart wall or mitral valve Impella® 5.0 Catheter Inlet in Aorta (TTE) 6.3c Impella® 5.0 Catheter in Papillary Muscle (TTE) 6.3d Papillary muscle Aorta • Catheter inlet area in aorta or near the aortic valve • Catheter pigtail in papillary muscle • Catheter pigtail too close to the mitral valve • Catheter inlet area more than 4 cm below the aortic valve or lodged between papillary muscle and the myocardial wall • Catheter outlet area too close to the aortic valve Figure 6.3 Transthoracic Echocardiographic (TTE) Illustrations of Impella® 5.0 Catheter Position Impella® 5.0 with the Impella® Console 6.9 Color Doppler Echocardiography When moving a patient supported with an Impella® 5.0 Catheter, it is important to monitor catheter migration. Adding color Doppler to an echo is another way to verify catheter position. If the Impella® 5.0 Catheter is correctly positioned, a dense mosaic pattern of turbulence will appear above the aortic valve near the outlet area of the catheter, as shown in the top image in Figure 6.4. If, however, the echocardiogram reveals a dense mosaic pattern of turbulence beneath the aortic valve (bottom image in Figure 6.4), this likely indicates that the outlet area of the catheter is in the wrong position, that is, the catheter is too far into the ventricle or entangled in papillary muscle. (Note: If using transesophageal echocardiography [TEE], look for the mosaic patterns in the same locations relative to the aortic valve and Impella® 5.0 Catheter outlet area.) Correct Impella® 5.0 Catheter Position (Color Doppler TTE) Incorrect Impella® 5.0 Catheter Position (Color Doppler TTE) Figure 6.4 Correct and Incorrect Impella® 5.0 Catheter Position (Color Doppler TTE) 6.10 Instructions for Use & Clinical Reference Manual Post-insertion Positioning (PIP) Checklist 6 Patient Management Topics Completing the steps shown in the following post-insertion positioning checklist can help to ensure proper position of the Impella® 5.0 Catheter following insertion. Pay particular attention to positioning after the patient is moved from the operating room or catheterization laboratory. 1.Remove slack in the Impella® 5.0 Catheter by increasing performance level to P9 and align the catheter against the lesser curvature of the aorta (rather than the greater curvature). 2.Use fluoroscopy to verify that the slack has been removed. 3.Verify that the Impella® 5.0 Catheter inlet area is optimally positioned about 4 cm below the aortic valve. 4.Return to previous performance level. 5.Secure the Impella® 5.0 Catheter at a firm external fixation point in the groin area. Understanding and Managing Impella® 5.0 Catheter Position Alarms The Impella® Console continuously monitors the Impella® 5.0 Catheter based on the placement signal and the motor current. • Placement Signal: Is the signal pulsatile or flattened? • Motor Current: Is the signal pulsatile or flattened? If the system alarms with one of the positioning alarms described in this section, fluoroscopic imaging is the best method for confirming position. You can also use TEE, TTE, or a standard chest X-ray. Placement Monitoring If the Impella 5.0 Catheter is either partly (just the pigtail) or completely in the ventricle, reposition the catheter under imaging guidance. If guidance is not available, use the repositioning guide to reestablish proper placement. ® If the Impella® 5.0 Catheter is completely out of the ventricle, do not attempt to reposition the catheter across the valve without a guidewire. The tables on the following pages describe possible placement conditions and the expected alarm messages, associated signal characteristics, and placement monitoring icons. Impella® 5.0 with the Impella® Console The placement monitoring screen will display “Disabled Due to Low Performance Level” if catheter performance level is less than P1. If there is a problem with the pressure sensor, it will display “Disabled Due to Non Reliable Sensor.” 6.11 Correct Impella® 5.0 Catheter Position If the Impella® 5.0 Catheter is in the correct position, the placement signal, motor current, dual signal, and placement monitoring screens will appear as shown in Figure 6.5. Placement Signal P-Perf P9 LV4509 Motor Current Flow [L/min] 98% 5.0 35 1 Placement Signal [mmHg] P-Perf Flow [L/min] P9 5.0 98% LV4509 0.80 0.69 Motor Current [A] 10 s 10 s Placement signal will be pulsatile. Motor current signal will be pulsatile. Dual Signal P-Perf P9 LV4509 Placement Monitoring Flow [L/min] 98% 5.0 35 1 Dual Signal I P Flow [L/min] P-Perf P9 LV4509 Ventricle 5.0 98% 35 1 Placement Monitoring Aorta Pump Position OK 10 s 10 s The Impella® 5.0 Catheter icon will be displayed with the valve symbols in the middle of the cannula. Figure 6.5 Correct Impella® 5.0 Catheter Position 6.12 Instructions for Use & Clinical Reference Manual Impella® 5.0 Catheter Fully in Ventricle or Fully in Aorta 6 Patient Management Topics If the Impella® 5.0 Catheter is fully in the ventricle or fully in the aorta, the following warning will appear: Pump Position Wrong Reposition Pump! Confirm with OK and Apply MENU for Reposition! In this situation, the placement signal, motor current, dual signal, and placement monitoring screens will appear as shown in Figure 6.6. Placement Signal P-Perf P9 LV4509 Motor Current Flow [L/min] 98% 5.3 8 -2 Placement Signal [mmHg] P-Perf Flow [L/min] P9 5.3 98% LV4509 0.81 0.80 Motor Current [A] 10 s 10 s Placement signal will be dampened or flat and displayed flow rate will increase even though patient is not being supported. Dual Signal P-Perf P9 LV4509 Placement Monitoring Flow [L/min] 98% Motor current signal will be dampened or flat because there is little or no difference in afterload between the Impella® 5.0 Catheter inlet and outlet. 5.3 8 -2 Dual Signal P8 98% Ventricle Aorta LV4509 Placement Monitoring P-Perf Flow [L/min] Pump Position Wrong 98% Ventricle Aorta P8 2.5 Press MENU for Reposition I 10 s LV4509 Placement Monitoring Pump Position Wrong Ventricle P 2.5 Aorta Press MENU for Reposition 10 s 10 s The Impella® 5.0 Catheter icon will scroll through a series of three icons indicating that the system is not able to differentiate between the three possible conditions. Figure 6.6 Impella® 5.0 Catheter Fully in Ventricle or Fully in Aorta Repositioning Menu Actions to take: 1.Under fluoroscopic or echocardiographic guidance, if available, reduce the performance level to P2 and carefully reposition the Impella® 5.0 Catheter by either pushing the catheter forward or pulling it back as needed. The repositioning menu can also be used as a guide for correcting Impella® 5.0 Catheter positioning. 2.Confirm that the placement signal and motor current are both pulsatile. Impella® 5.0 with the Impella® Console 6.13 Low Native Heart Pulsatility When a patient has poor native ventricular function, the placement signal may remain pulsatile; however, the amplitude will be dampened and both the minimum and maximum values will be greater than zero because the aortic valve does not open and the Impella® 5.0 Catheter raises the aortic blood pressure above the ventricular pressure during systole. In a situation of low native heart pulsatility, the Impella® Console may not be able to determine the catheter position and the placement signal, motor current, dual signal, and placement monitoring screens will appear as shown in Figure 6.7. Placement Signal Flow [L/min] P-Perf P8 LV4509 Motor Current 98% 4.2 25 17 Placement Signal [mmHg] P-Perf Flow [L/min] P8 LV4509 4.2 98% 0.79 0.77 Motor Current [A] 10 s 10 s Placement signal will show reduced pulsatility, Motor current signal will be dampened or flat and the lower value will be elevated above zero. due to small pressure gradients between the Impella® 5.0 Catheter inlet and outlet. Dual Signal P-Perf P8 LV4509 Placement Monitoring Flow [L/min] 98% 4.2 25 17 Dual Signal I P P-Perf Flow [L/min] P8 LV4509 4.2 98% 25 17 Placement Monitoring Ventricle Aorta Pump Position Unknown Due to Low Pulsatility 10 s 10 s The Impella® 5.0 Catheter icon will be displayed without the valve symbol. Figure 6.7 Impella® 5.0 Catheter Position Unknown Due to Low Pulsatility Actions to take: 1. Assess cardiac function. 6.14 Instructions for Use & Clinical Reference Manual Impella® 5.0 Catheter Outlet Area on or Near Aortic Valve 6 Patient Management Topics If the Impella® 5.0 Catheter outlet area is on or near the aortic valve, the catheter may be too deep in the ventricle. The following warning will appear: Pump Position Wrong Pump Outlet Blocked! Confirm with OK and Apply MENU for Reposition! In this situation, the placement signal, motor current, dual signal, and placement monitoring screens will appear as shown in Figure 6.8. Placement Signal P-Perf P8 LV4509 Motor Current Flow [L/min] 98% 4.4 25 -29 Placement Signal [mmHg] P-Perf P8 LV4509 Flow [L/min] 4.4 98% 0.80 0.70 Motor Current [A] 10 s 10 s Lower value of the placement signal will be reduced significantly below zero. Motor current signal will appear normal. Dual Signal P-Perf P8 LV4509 Placement Monitoring Flow [L/min] 98% Dual Signal 4.4 25 -29 I P P-Perf P8 LV4509 Flow [L/min] 4.4 98% 25 -29 Placement Monitoring Ventricle Aorta Pump Position Wrong Pump Outlet Blocked! Press MENU for Reposition! 10 s 10 s The Impella® 5.0 Catheter icon will be displayed with the valve symbol blocking the outlet window. Figure 6.8 Impella® 5.0 Catheter Outlet Area on or near Aortic Valve Actions to take: 1.Assess and adjust Impella® 5.0 Catheter position under fluoroscopic or echocardiographic guidance, if available. 2.If fluoroscopic or echocardiographic guidance is not available, reduce the performance level to P2 and gently pull the catheter back 1 cm and see if the condition resolves. Impella® 5.0 with the Impella® Console 6.15 Suction Suction may occur if the blood volume available for the Impella® 5.0 Catheter is inadequate or restricted. Suction limits the amount of support that the Impella® 5.0 Catheter can provide to the patient and results in a decrease in arterial pressure and cardiac output. It can damage blood cells, leading to hemolysis. It may also be an indicator of right heart failure. If a suction alarm occurs, follow the recommended actions in the alarm display by decreasing the performance. 1.Check the catheter for correct positioning using imaging. Reposition the catheter by rotating or moving it into or out of the ventricle slightly. Either or both of these actions could help move the inlet of the catheter away from the interior ventricular wall. 2.Assess patient’s fluid intake and output to confirm adequate volume status. 3.Confirm right ventricular function by assessing CVP or right side function on echo. If CVP is not an option, check the pulmonary artery diastolic pressure to assess the patient volume status. 4. Return performance level to pre-alarm setting. Hemolysis When blood is pumped, it is subjected to mechanical forces. Depending on the strength of the blood cells and the amount of force applied, the cells may be damaged, allowing hemoglobin to enter the plasma. Pumping forces can be generated by a variety of medical procedures including heart lung bypass, hemodialysis, or ventricular assist device (VAD) support. Patient conditions— including catheter position, pre-existing medical conditions, and small left ventricular volumes— may also play a role in patient susceptibility to hemolysis. Patients who develop high levels of hemolysis may show signs of decreased hemoglobin levels, dark or blood-colored urine, and in some cases, acute renal failure. Plasma-free hemoglobin (PfHgb) is the best indicator to confirm whether a patient is exposed to an unacceptable level of hemolysis. Management technique may differ depending on the underlying cause of hemolysis. Table 6.1 provides guidance for various circumstances. 6.16 Instructions for Use & Clinical Reference Manual Table 6.1 Guide for Managing Hemolysis in Various Circumstances 6 Console Indicators Clinical Indicators Management Impella 5.0 inlet area in close proximity to intraventricular wall • Suction alarms • Lower than expected flows Imaging (see note) • Reposition the Impella® 5.0 Catheter by rotating or moving the catheter into or out of the ventricle slightly. Either or both of these actions could help move the inlet of the catheter away from the intraventricular wall. • If reposition will be delayed, reduce the flow by 1 or 2 performance levels if tolerated by patient hemodynamics. Return performance level after repositioning. • Reassess position after performance level has returned to desired target value. Wrong catheter position • Position alarms with higher Imaging (see note) than expected flows • Suction alarms with lower than expected flows • Pump outlet blocked alarms • Reposition the Impella® 5.0 Catheter by rotating or moving the catheter into or out of the ventricle slightly. Either or both of these actions could help move the inlet of the catheter away from the intraventricular wall. • If reposition will be delayed, reduce the flow by 1 or 2 performance levels if tolerated by patient hemodynamics. Return performance level after repositioning. • Reassess position after performance level has returned to desired target value. Higher than needed flow setting • There may be no console indicators • Suction alarms • Normal hemodynamics • Native recovery • Reduce performance level until patient pressure starts to drop. • Increase performance by 1 level. Inadequate filling volume • Position alarms • Suction alarms • Lower than expected flows • Low CVP • Low PCWP • Low AOP • High PA pressures • Right heart failure • High urine output • Increased bleeding or chest tube drainage • Reduce the flow by 1 or 2 performance levels if tolerated by patient hemodynamics. • Correct I and O balance. • Consider giving volume; additional volume will expand the end systolic ventricular volume. • Reduce PA pressure. • Improve right heart function. Pre-existing patient conditions or other medical procedures N/A • Patient past medical history • Current procedures or treatments ® Note on imaging: All imaging technology represents the anatomy in two dimensions (2D). It is not possible to assess the interactions between the Impella® 5.0 Catheter and the intraventricular anatomy that occur in three dimensions (3D). Abiomed strongly recommends that the Impella® 5.0 Catheter be repositioned, even if the imaging view shows correct position. Impella® 5.0 with the Impella® Console 6.17 Patient Management Topics Condition Accuracy of Displayed Flow Rate Pressure Sensor Drift and Sensor Failure The performance level of the Impella® 5.0 Catheter is automatically set to P2 during zeroing if the catheter is operating above P0. Performance level remains at P2 after zeroing and does not automatically return to its prior setting. Under normal operating conditions, displayed flow rate can deviate from the actual flow rate by up to 0.5 L/min. Sensor Drift and Effects on Flow Zeroing and Sensor Drift Only zero the sensor when you observe evidence of sensor drift. Abiomed does NOT recommend zeroing the differential pressure sensor when there is NO evidence of sensor drift because the performance level is reduced to P2 during zeroing. The electrical signal produced by the differential pressure sensor may drift over time. This drifting is normal behavior and is characterized by an upward or downward shift in the placement signal along the y-axis of the console display; however, the amplitude (peak-to-trough, or maximum to minimum) of the waveform will not change. When sensor drifting occurs, it affects the Impella® 5.0 Catheter flow calculations and the displayed flow rate will not match the expected flow. (Refer to Table 6.2 to see what the expected flow rates for each performance level should be.) Table 6.2 Expected Flow Rate Displayed on Console Performance Level Flow rate (L/min) P1 0.0 – 1.4 P2 0.5 – 2.6 P3 0.5 – 3.1 P4 0.9 – 3.4 P5 1.4 – 3.7 P6 1.8 – 4.0 P7 2.6 – 4.4 P8 3.4 – 4.7 P9 4.2 – 5.3 If you observe that the placement waveform has shifted up or down on the y-axis, or the expected flow does not match the current performance level setting, zero the differential pressure sensor by performing the following steps: 1.Press the SIGNAL key until the placement signal is displayed. 2.Press the ZERO key. “Placement Signal Offset Adjust” is displayed. 3.Press OK within 10 seconds to start the zeroing algorithm. The Console will display the “Calculation is Running” message. Resetting Performance Level After zeroing, return the performance level to the level at which it was set prior to zeroing. 6.18 4.Press OK to accept the new setting when the console displays the “Placement Signal Offset Adjust finished!” message. 5.The Impella® 5.0 Catheter continues running at performance level P2 and the lower value of the differential pressure signal should display about 0±6 mmHg. 6.If the adjusted value deviates from zero by more than 6 mmHg, repeat the zeroing procedure (steps 2-4). 7.If the adjusted value is 0±6 mmHg, slowly increase the performance level, step by step, until you reach the setting used before the zeroing procedure. Instructions for Use & Clinical Reference Manual Sensor Failure and Effects on Flow Calculations 6 Patient Management Topics If the pressure sensor fails, the console can no longer calculate the flow rate. The console will display the flow as “?.?”. Sensor Failure and Effects on Position Control If the pressure sensor fails, placement monitoring is switched off because it is not possible to display a precise position for the catheter. In this case, the motor current signal or imaging procedures can be used for position control. As long as the motor current signal is pulsatile, . the Impella® 5.0 Catheter is correctly positioned across the valve. This signal must be monitored closely because the catheter can become dislodged (displaced) when moving the patient or changing the patient’s position. Therefore, if patient hemodynamics change—for example, if arterial pressure falls or there are signs of left ventricular failure—check the correct positioning of the catheter using imaging procedures (eg, TEE) and the motor current signal. Suction Detection During Sensor Drift or Sensor Failure If sensor drift occurs or the pressure sensor fails, the console can no longer detect suction. . (For more information about suction, refer to the “Suction” discussion earlier in this section of the manual.) The effectiveness of Impella® 5.0 Catheter support can only be assessed by monitoring patient hemodynamics, cardiac imaging, and the Impella® 5.0 Catheter motor current. Signs of suction include: • A drop in the patient’s arterial pressure • Decreased output, if a cardiac monitor is in place • Decreased motor current If imaging reveals that the suction is caused by the Impella® 5.0 Catheter inlet area in close proximity to the ventricular wall, reposition the catheter as described in Table 6.1. If hemodynamic parameters, such as low aortic pressure or high pulmonary artery pressure, indicate suction caused by inadequate filling volume, reduce flow and follow the management strategies described in Table 6.1. Impella® 5.0 with the Impella® Console 6.19 Operating the Impella® 5.0 Catheter Without Heparin in the Purge Solution The Impella® 5.0 Catheter is designed to be operated with a purge solution that contains heparin. Operation of the system without heparin in the purge solution has not been tested. In the event that a patient is intolerant to heparin, due to heparin-induced thrombocytopenia (HIT) or bleeding, physicians should use their clinical judgment in assessing the risks versus benefits of operating the Impella® 5.0 System without heparin. If it is in the best interest of the patient to operate the system without heparin, the dextrose solution is still required, and physicians should consider an alternative anticoagulant delivered systemically. Do NOT add any alternative anticoagulant (such as a direct thrombin inhibitor) to the purge solution. The Impella® 5.0 Catheter has not been tested with any alternative anticoagulants present in the purge solution. How to Change to A Backup Console A backup Impella® Console should be available at all times when a patient is on support. In the event of a console failure, follow the steps below to transition the Impella® 5.0 System to the backup console. 1.Confirm that the backup console is powered on and ready. 2.Transfer the pressure transducer cable from the original console to the pressure connector socket (yellow) on the left side of the backup console. 3.Remove the blue connector cable from the original console and plug it into the catheter connector socket (blue) located on the right side of the backup console. 4.Once the Impella® 5.0 Catheter is connected to the backup console, the “Quick Start” message will appear on the screen asking you to confirm restarting the Impella® 5.0 Catheter at the previously set performance (P) level. 5.Press OK within 10 seconds to confirm restarting the Impella® 5.0 Catheter at the previously set performance level. 6.20 Instructions for Use & Clinical Reference Manual 7 Impella® Console Alarms and Warnings Impella® 5.0 Console Alarms................................................................. 7.1 Handling Alarms and Warnings......................................................................... 7.1 Impella® 5.0 Console Warnings............................................................7.3 Impella® 5.0 Console Alarms 7 Impella® Console Alarms and Warnings Handling Alarms and Warnings You can mute alarms and warnings by pressing the OK button once. The text of the alarm or warning will remain on the display. To acknowledge receipt of the alarm or warning and reactivate monitoring, press the OK button a second time. After muting an alarm or warning, if another alarm or warning occurs it will only be heard and displayed if it is a higher priority alarm or warning than the one that was muted. Alarm Message Battery Completely Discharged Sudden Pump STOP Possible Cause Action The battery is depleted. 1. Immediately connect the Impella® Console to the Impella® Power Supply to recharge the Impella® Console batteries. 2. Connect the Impella® Power Supply to AC power and switch it on. 3. If the AC power icon does not appear, change the Impella® Console. 4. Notify Abiomed. The battery is depleted. 1. Immediately connect the Impella® Console to the Impella® Power Supply to recharge the Impella® Console batteries. 2. Connect the Impella® Power Supply to AC power and switch it on. 3. If the AC power icon does not appear, replace the Impella® Power Supply and the connection cable. 4. If the alarm is still present or the AC power icon still does not appear, change the Impella® Console. The temperature of the Impella® Console batteries is dangerously high. 1. Change the Impella® Console immediately. 2. Notify Abiomed. Change Console Immediately! Battery Discharged Change Console or Connect Power Supply! Battery Temperature too High Change Console Immediately! Keyboard not Available Change Console! Inform Service! No External Pressure Signal Check Cable and Transducer! Impella® 5.0 with the Impella® Console The Impella® Console 1. Change the Impella® Console. keypad is defective. 2. Notify Abiomed. The catheter can no longer be controlled with the Impella® Console. The purge pressure measurement is faulty. 1. Check all cables and plug connections. 2. If connections are secure, replace the pressure transducer cable. 3. If the alarm is still present, change the Impella® Console. 4. Notify Abiomed. 7.1 Alarm Message No Pump Connected Console will be Switched off Extend for 30 minutes? ...Press OK to Confirm Please Release ON Key! Do NOT Use ON to Switch off! Emergency Shut Down of Console! Using ON May Cause Data Loss! Pump Disconnected while Running Check Connection! Pump Plug Defect Use New Pump! Pump Stopped Console Defect Change Console Immediately! Pump Stopped Console Shows Last Performance Value Confirm or Change Value! Pump Stopped Restart Pump! Monitor Motor Current! 7.2 Cause Action No catheter has been plugged into the Impella® Console. The Impella® Console will operate for 10 minutes on battery power and then shut down automatically. NOTE: The last 2 lines of this message will appear only if the console is on AC power. Plug in a catheter or switch the Impella® Console off by pressing OK twice. If the console is operating on AC power you can press OK once to extend the time to shutdown to 30 minutes. During operation the ON key was pressed and held. The Impella® Console will turn off if the ON key is held for longer than 3 seconds. Data saved on the Impella® Console may be lost. Press OK to confirm the alarm message or if the Impella® Console is off, press ON to turn the Impella® Console on again. The Catheter was disconnected during operation or the connection to the catheter was interrupted. Check the catheter connection and all cable and plug connections, including power supply connections. Error reading in catheter data. 1. Replace the catheter. 2. Notify Abiomed. The catheter has stopped due to an Impella® Console error. 1. Change the Impella® Console immediately. 2. Notify Abiomed. A catheter that was operated previously at a performance level greater than P0 has been plugged in again. 1. Press OK on the Impella® Console to restart the catheter at the previous performance level. 2. To restart the catheter at a different level, use the ▼ and ▲ keys to select the desired level. Press OK to confirm. The catheter-specific 1. Try to restart the catheter at motor current limit performance level P2. has been exceeded. 2. Replace the catheter if it does not The catheter start up after the third attempt. stopped. This may be the result of electrical or mechanical causes. Instructions for Use & Clinical Reference Manual System Error Pump Stopped Pump Restarted! System Error Pump Stopped Restart Pump! Time Out due to inactivity Console will be switched off Extend for 30 minutes? ...Press OK to Confirm Time Out due to inactivity Test Plug / Test Pump Console will be switched off Cause Action 7 Impella® Console Alarms and Warnings Alarm Message The catheter stopped Notify Abiomed. but is now running again. The catheter has stopped due to an error. 1. Try to restart the catheter at performance level P2. 2. If the catheter cannot be restarted, change the Impella® Console. 3. Notify Abiomed. The pump test plug is still in the console but is not being used. The console will be switched off within 2 minutes. NOTE: This message will appear only if the console is running on AC power. Press OK to extend the time until the console is switched off to 30 minutes. Otherwise, no action is necessary and the console will be switched off in 2 minutes. The pump test plug No action necessary is still in the console but is not being used. The console will be switched off within 2 minutes. NOTE: This message will appear only if the console is running on battery power. Impella® 5.0 Console Warnings Warning Message Console Running On Battery Remaining Time “__” Minutes Could not Hold Target Performance Pump Still Running Impella® 5.0 with the Impella® Console Cause Action The remaining battery capacity is less than 50%. The Impella® Console displays the estimated remaining run-time. 1. Observe remaining run-time. 2. Be prepared to plug the console into an AC power outlet. 1. Change the Impella® Console. The selected performance level 2. If the warning is still present, replace cannot be attained. the catheter. The catheter is not running at full power. 3. Notify Abiomed. 7.3 Warning Message Data Storage not Available Please Inform Service! Flow Below Low Flow Limit Check Flow Rate! Monitor Pump Signals! High Battery Temperature Please Cool Console! High Motor Current Value Pump Still Running Monitor Motor Current! High Purge Pressure Decrease Purge Rate! Check Purge System! Low Purge Pressure Increase Purge Rate! Check Purge System! Pump has been previously used. Use New Pump Or press OK to continue Pump Plug Defect Flow Calculation not Possible 7.4 Cause Action The data logging function of the Impella® Console is not working. Notify Abiomed. The flow rate has fallen below the low flow limit value selected by the user. 1. If the performance level has been reduced, adjust the low flow limit value accordingly. 2. If attachment due to suction has occurred, reduce the performance level first. The temperature of the Impella® Console has increased to a dangerous level. Move console to a cooler location. Note: Do not place the Impella® Console on top of the Impella® Power Supply. The catheter has 1. Observe motor current characteristics. mechanical problems. 2. If the current continues to increase, The motor current replace the catheter if possible. limit for the selected 3. Notify Abiomed. performance level has been exceeded. The catheter may stop if the motor current continues to increase. The purge pressure has increased to greater than 700 mmHg. 1. Inspect the purge system for kinks. 2. Reduce the purge flow rate at the infusion pump by 1 mL/h. Do not decrease the flow rate to less than . 4 mL/h. 3. Decrease dextrose concentration. 4. If the problem persists, contact Abiomed. The purge pressure 1. Deliver 0.5 mL bolus until the purge has decreased to less pressure reaches 600 mmHg. than 300 mmHg. 2. Inspect the purge system . for leaks. 3. Increase the purge flow . rate at the infusion pump . by 0.5 mL/h. Do not exceed . 20 mL/h. 4. Refer to the Troubleshooting discussion in section 5 of this manual for more detailed information. The Impella® 5.0 Catheter has been used previously. Use new Impella® 5.0 Catheter. Error reading flow rate performance data from the red Impella® plug. Replace the catheter. Instructions for Use & Clinical Reference Manual Pump Position Wrong Reposition Pump! Confirm with OK and Cause Action A wrong catheter position was detected. 1. Reposition the catheter according to the Repositioning Menu. 2. If step 1 does not provide correct positioning, check the catheter position by imaging methods. The placement guide was canceled because of a fault. A measuring signal (placement signal) is outside the validity range. 1. Select the placement guide again using the MENU key. 2. If the placement guide cancels again, use imaging procedures as assistance for repositioning. Apply MENU for Reposition! Pump Repositioning Function has been canceled Pump Stopped Reverse Flow! Restart Pump! Real Time Clock Error Missing Time Information Please Inform Service! Sensor Value not Reliable Flow Calculation not Possible Suction Likely Decrease Performance Slowly! Monitor Placement Signal! Time Out Reposition Guide Function has been canceled Impella® 5.0 with the Impella® Console 7 Impella® Console Alarms and Warnings Warning Message Retrograde flow will Unless the catheter is being explanted, occur because the restart the catheter immediately. Do not catheter has stopped. restart the catheter during explantation. The internal clock in Notify Abiomed. the Impella® Console is defective. Data is being recorded without a time stamp. Defective pressure sensor or signal interference. 1. Check the catheter position regularly using imaging procedures and watch for signs of suction or outlet blockage. 2. If signal interference is suspected, either increase the distance between Impella® 5.0 System components and the EMI source or turn off the EMI source. 3. If the warning is still present, refer to “Pressure Sensor Drift and Sensor Failure” in section 6 of this manual for more information and consider replacing the catheter. The Impella® Console 1. Reduce the performance level. has detected catheter 2. Check the filling of the left ventricle. suction. 3. If there are no filling problems, check the catheter for correct positioning. 4. Try to increase the performance level to reach the intended level of support. The placement guide If the incorrect positioning persists, has been canceled select the placement guide again using because of timeout. the MENU key. 7.5 8 General System Information Terminology, Abbreviations, and Symbols.......................................8.1 Terminology and Abbreviations......................................................................... 8.1 Symbols............................................................................................................8.2 Ambient Conditions................................................................................ 8.3 During Operation..............................................................................................8.3 During Storage.................................................................................................8.3 Impella® 5.0 Catheter Parameters...................................................... 8.3 Impella® Console and Power Supply Parameters........................... 8.4 Connector Cable Parameters.............................................................. 8.4 Impella® 5.0 Catheter Dimensions....................................................... 8.5 Anatomic Considerations......................................................................8.6 Cleaning.....................................................................................................8.6 Storing the Impella® Console and the Power Supply.....................8.7 Returning an Impella® 5.0 Catheter to Abiomed (United States)...........................................................................................8.7 Terminology, Abbreviations, and Symbols 8 General System Information Terminology and Abbreviations Catheter serial number Identification number of the Impella® 5.0 Catheter; stated on the package label, on the red Impella® plug, and on the display screen Dextrose and Glucose The terms “dextrose” and “glucose” are used interchangeably to refer to the solution used as purge fluid for the Impella® 5.0 Catheter Hz Hertz Motor housing (or pump housing) Enclosure of the Impella® 5.0 Catheter motor Pump Central delivery unit of the Impella® 5.0 Catheter, consisting of the motor, motor housing, cannula with inlet and outlet, and pigtail at the tip Pump test plug Tool required for checking functions during preventive maintenance; thick metal plug marked “Pump Test Plug” Purge pressure Pressure present in the purge lumen within the Impella® 5.0 Catheter and in the purge tubing Purge system Infusion pump used for rinsing the Impella® 5.0 Catheter Retrograde flow Reverse flow through the cannula when the Impella® 5.0 Catheter is at a standstill (eg, regurgitation) V Volt VA Volt ampere (Watt) Impella® 5.0 with the Impella® Console 8.1 Symbols Caution; comply with instructions for use Defibrillator-proof type CF equipment Keep dry Storage temperature (eg, 10°C to 30°C) Declares conformity with directive 93/42/EEC for medical devices 2009-02 Date of manufacture (eg, February 2009) Protect from sunlight Symbol for lot designation; the manufacturer’s lot designation must be stated after the LOT symbol REF 123456 Abiomed part number (eg, part number 123456) SN 123456 Manufacturer’s serial number (eg, serial number 123456) Non Sterile! The product is not sterile 2010-06 Use-by date (eg, use before June 2010) Do not reuse Sterilized using ethylene oxide Electric scrap; must be disposed of separately. . Must not be disposed of as domestic waste. Protective Earth Protection Class II equipment 8.2 Instructions for Use & Clinical Reference Manual Ambient Conditions 8 Impella® Console Max. temperature 40°C (104°F) Min. temperature 10°C (50°F) Max. rel. atmospheric humidity 75% General System Information During Operation Impella® 5.0 Catheter Ambient conditions that prevail within the human body During Storage Impella® Console Impella® 5.0 Catheter Max. temperature 40°C (104°F) 25°C (77°F) Min. temperature 10°C (50°F) 10°C (50°F) Max. rel. atmospheric humidity 75% 75% Impella® 5.0 Catheter Parameters Speed range 0 to 33,000 rpm Power consumption Less than 1.07 W Voltage Max. 18 V DC Flow-Maximum 5.0 L/min Purging the Impella® 5.0 Catheter Recommended purge fluid. Dextrose concentration Purge pressure Infusion rate Maximum duration of use US Reliability 20% dextrose solution with heparin concentration of 50 IU per mL 5% to 40% 300 to 700 mmHg 4 to 20 mL/h 6 hours 91.4% with 80% confidence at 10 days based on chi-squared test Latex Free All versions of the Impella® System, including all accessories, are latex free. Dimensions of Impella® 5.0 Catheter Length of invasive portion (without catheter) 140 ± 3 mm Diameter Max. 7.2 mm (nom. 7.0 mm) Classification per DIN EN 60601-1 Protection class II, degree of protection: . CF (Impella® Console and Impella® 5.0 Catheter) Classification per directive 93/42/EEC Class III Latex free Yes Impella® 5.0 with the Impella® Console 8.3 Impella® Console and Power Supply Parameters Accuracy of calculated flow rates ± 10% (relative to maximum flow rate) Characteristic values Nominal voltage Max. nominal current under load Max. power consumption under load Device fuse Running time without power supply with fully charged batteries 110/230 V (at 50/60 Hz) 1.1 A 120 VA 2x2 A slow-blow At least 60 minutes (charging duration of at least 10 hours) Classification per DIN EN 60601-1 Safety Class II, degree of protection: CF Classification per directive 93/42/EEC Class IIb Electrical system Installation in accordance with pertinent regulations is required for use in medical facilities (eg, VDE 0100, VDE 0107, or ICE stipulations). Observe country-specific regulations and national deviations. Electromagnetic compatibility Tested in accordance with DIN EN 60601-1-2 Impella® Console dimensions Height Width Depth Weight 55 mm 260 mm 305 mm 3.35 kg (7.38 lbs.) Impella® Power Supply dimensions Height Width Depth Weight 75 mm 305 mm 265 mm 3.2 kg (7.0 lbs.) Maintenance and repair interval 12 months (Work must be performed by technicians authorized by Abiomed) Connector Cable Parameters 8.4 Length 2.5 m Service life Single use only Instructions for Use & Clinical Reference Manual Impella® 5.0 Catheter Dimensions 8 General System Information 6 Fr 9 Fr 35-45 cm Length of sterile sleeve 21 Fr 10 cm Repositioning sheath 97 cm Differential pressure sensor 3.5 cm 8 cm 11.5 cm Figure 8.1 Impella® 5.0 Catheter Dimensions Impella® 5.0 with the Impella® Console 14 cm 8.5 Anatomic Considerations In a small number of cases (about 2 in every 100 patients) the Impella® 5.0 Catheter cannot be successfully placed; or can be placed, and the performance is compromised due to patients having anatomic conditions outside of the range for which the Impella® 5.0 Catheter was designed. The following table describes anatomic conditions that may affect the insertion or operation of the Impella® 5.0 Catheter. Physicians should consider these characteristics when evaluating small or very tall patients for Impella® 5.0 Catheter support. Condition Effect The size and tortuosity of the femoral and iliac arteries Limits the ability of the Impella® 5.0 Catheter to be advanced from the insertion site into the left ventricle Distance from the insertion site to the apex of the left ventricle For very tall patients, the maximum interventional length may not be sufficient to allow correct placement of the Impella® 5.0 Catheter Systolic left ventricular (LV) long The Impella® 5.0 Catheter may interfere with the mitral valve axis < 7 cm Systolic left ventricular (LV) long The Impella® 5.0 Catheter pigtail will not have a surface to push axis > 11 cm against to help stabilize its position, and may have a tendency to swing or bounce Cleaning • Clean the Impella® Console, the pump test plug, and the cables by wiping them down with disinfectant. • Do not allow any fluids to enter the connector sockets. 8.6 Instructions for Use & Clinical Reference Manual • Place the Impella® Console on a horizontal surface to prevent falling. • Connect the AC power cable from the Impella® Power Supply to an AC outlet. • The battery may be destroyed if the Impella® Console is stored with a depleted battery. To keep the Impella® Console battery charged, the console should be plugged into the Impella® Power Supply. The Impella® Power Supply should be plugged into an AC outlet and switched to the ON position when it is stored. Both the green POWER and CONNECTED lights should be lit. Returning an Impella® 5.0 Catheter to Abiomed (United States) To return an Impella® 5.0 Catheter to Abiomed, contact your local Clinical Consultant for an Abiomed-approved return kit.* The kit includes instructions for returning the catheter to Abiomed. * Only available in the United States Impella® 5.0 with the Impella® Console 8.7 General System Information The Li-Ion batteries must be charged for 10 hours prior to system operation. After being unplugged, the Impella® Console will operate for at least 60 minutes after the batteries have been fully charged. Storing the Console 8 Storing the Impella® Console and the Power Supply Appendices Appendix A: Impella® System Limited Service Warranty (United States)................................................................... A.1 Appendix B: Technical Safety Inspections, Maintenance, and Repair...........................................................................B.1 Technical Safety Inspections............................................................................. B.1 Maintenance and Repair Intervals..................................................................... B.1 Impella® Console Function Test.........................................................................B.2 Appendix A Appendix A: Impella® System Limited Service Warranty (United States) Abiomed, Inc. warrants that, at the time of installation, all Impella® Systems (the “Goods”) sold will be free from defects in material and workmanship and remain free from defects under normal use . and service for a period of one (1) year from the date of shipment. Extended warranty and service may, at Abiomed's option, be offered for an additional charge, in which event separate or additional terms and conditions may apply. This warranty provides coverage for the Impella® Console, Power Supply, and Purge System. This warranty does not cover routine preventive maintenance or replacement parts that are consumed per the console’s periodic maintenance schedule outlined in the Operator’s and Service Manuals. The express warranty set forth on this page is the only warranty given by Abiomed with respect to any goods furnished hereunder. Abiomed makes no other warranty, express, implied or arising by custom or trade usage, and specifically makes no warranty of merchantability or of fitness for any particular purpose. Said express warranty shall not be enlarged or otherwise affected by Abiomed's rendering of technical or other advice or service in connection with the Goods. Abiomed shall not be liable for incidental or consequential losses, damages or expenses, directly or indirectly arising from the sale, handling or use of the Goods, or from any other cause relating thereto, and Abiomed's sole responsibility under this warranty will be, at its option, to 1) repair or replace the Goods or any components of the Goods found to be defective in workmanship or material during the foregoing warranty period, or 2) to refund the purchase price paid. All replaced components and Goods will become the property of Abiomed. This warranty shall not apply if the Goods have been: (a) repaired or altered in any way by other than Abiomed or Abiomed authorized service personnel; (b) subjected to physical or electrical abuse or misuse; or (c) operated in a manner inconsistent with Abiomed's instructions for use of the Goods. If Abiomed determines that a claim was not caused by Abiomed . or Abiomed’s authorized service personnel, then Buyer shall pay Abiomed for all related costs incurred by Abiomed. This warranty is not transferable without the express written consent of Abiomed. Under this warranty, Abiomed will provide at no charge, updates or modifications which directly affect the safe operation of the Goods. Abiomed is not obligated to provide updates or modifications which provide (a) product improvement or enhancement; (b) new product features, or (c) options to the Goods. Abiomed has no obligation to provide a loaner system during service or maintenance of the Goods. However, at Abiomed’s sole discretion, Abiomed may provide such loaner systems. This warranty applies to the Impella® System only (Console, Power Supply, and Purger) and not to any disposable or other component of the Impella® System. Specific items excluded from this warranty include, but are not limited to, pumps, external tubing and accessories. This warranty may not be amended without the express written consent of an authorized officer of Abiomed. Impella® 5.0 with the Impella® Console A.1 Appendix B Appendix B: Technical Safety Inspections, Maintenance, and Repair Technical Safety Inspections In accordance with the specification issued by Abiomed as per §6 MPBetreibV (German regulation), the Impella® Console must be subject to yearly technical safety inspections. Technical safety inspections may only be carried out by authorized technicians in accordance with the technical safety inspection requirements presented below and must be documented . in the medical product logbook in accordance with §7 MPBetreibV. A sticker on the device indicates the date for the next required inspection. Figure B.1 shows a sticker indicating that inspection is required in May 2010. However, the stipulations in the medical product logbook are binding in any case. Figure B.1 Inspection Sticker Showing Inspection Required in May 2010 The following technical safety inspections are required for the Impella® System: • Inspection of labeling and instructions for use • Visual inspection of the device and its accessories for any signs of damage • Testing for electrical safety as per DIN VDE 751 or DIN EN 60 601 • Leakage current test • Dielectric strength test • Functional testing of all switches, keys, rotary knobs, sockets, and control lights on the device • Checking battery operation If defects become apparent during the technical safety inspections that could endanger patients, employees, or third parties, then the device must not be operated until the defects have been remedied by proper technical servicing. Maintenance and Repair Intervals The Impella® Console is subject to 1 year maintenance and repair intervals. The corresponding work must be performed by technicians authorized by Abiomed, and must be documented in the medical product logbook in accordance with §7 MPBetreibV (German regulation). Impella® 5.0 with the Impella® Console B.1 Impella® Console Function Test When conducting the Impella® Console function test as part of preventive maintenance, if a pressure value of 50 (± 4) mmHg does not appear or the speed is not within the specified range of 24,000 ± 1,000 revolutions per minute (rpm), do NOT use the Impella® Console. As part of preventive maintenance, it may be necessary to perform an Impella® Console function test using the pump test plug. 1.Insert the pump test plug, with the “up” marker oriented upward, into the catheter connector socket at the right side of the Impella® Console handle. A single beep followed by a series of three beeps confirms insertion. The Impella® Console displays the Impella® Console information screen (see Figure B.2). P-Perf P0 LV0 98% Flow [L/min] ?.? 50 50 Placement Signal [mmHg] Information Pump: Testplug Pump ID: 0 MPC: 01.203-xx.xxx 10 s Figure B.2 Impella® Console Information Screen 2.Confirm that the following items are displayed (see Figure B.2): • P-Perf: P0 • Pump: Testplug • Pump ID: 0 • Flow (L/min): ?.? • Pressure value of 50 (± 4) mmHg (above the right diagram axis) Returning to P0 To return to performance level P0, press the P-PERF key until two blinking vertical bar graphs appear next to the performance level display. Set the performance level to P0 by pressing ▼ and OK. B.2 3.Press the SIGNAL key three times to open the speed screen (Speed [RPM]). 4.Press and hold the P-PERF key until two blinking vertical bar graphs appear next to the P0 display. 5.Set the performance level to P5 by pressing ▲ and acknowledging within 10 seconds by pressing OK. Confirm that the speed lies within the range of 23,000 to 25,000 rpm. 6.Return the performance level to P0 by pressing P-PERF, ▼ and OK before the pump test plug is removed. 7.Remove the pump test plug. The Impella® Console function test is now complete. Instructions for Use & Clinical Reference Manual INDEX A Abbreviations (See Terminology and abbreviations) Activated clotting time (ACT) ACT for catheter insertion with GP IIb-IIIa inhibitor, 5.10 ACT for catheter insertion with heparin, 5.10 Maintaining ACT until explant, 5.11 Adverse events (potential), 2.1 Alarms Console alarms, 7.1 Downstream occlusion (figure), 5.20 Handling alarms and warnings, 7.1 High purge pressure, 5.18 (figure), 5.18 Low purge pressure, 5.15 (figure), 5.15 Understanding and managing position alarms, 6.11 (See also Patient management: Understanding and managing Impella® 5.0 Catheter position alarms) B B. Braun Vista basic Infusion Pump (See also Purge system) Features Alarm light, 3.10 (figure), 3.10 Alarm silence key, 3.10 (figure), 3.10 Clear key, 3.10 (figure), 3.10 Display screen, 3.10 (figure), 3.10 Function keys, 3.10 (figure), 3.10 Infusion light, 3.10 (figure), 3.10 Keypad, 3.10 (figure), 3.10 ON/OFF key, 3.10 (figure), 3.10 START/STOP key, 3.10 (figure), 3.10 Replacing the purge tubing, 5.14 Troubleshooting (See Purge system: Troubleshooting) C Catheter (See also Diagnostic catheter) Catheter components, 3.3 Cannula, 3.3 (figure), 3.3 Catheter shaft, 3.4 (figure), 3.3 Check valve, 3.4 (figure), 3.3 Clear sidearm, 3.4, 5.4 (figure), 3.3 Differential pressure sensor, 3.4 (See also Catheter: Differential pressure sensor) (figure), 3.3 Infusion filter, 3.4 (figure), 3.3 Inlet area, 3.3 (figure), 3.3 Motor housing, 3.4 (figure), 3.3 Outlet area, 3.4 (figure), 3.3 Pigtail, 3.3 (figure), 3.3 Pressure reservoir, 3.4 (figure), 3.3 Red Impella® plug, 3.4 (figure), 3.3 Memory, 3.4 Repositioning unit, 3.4 (figure), 3.3 Description, 3.3 (figure), 3.3 Differential pressure sensor, 3.4 (See also Catheter: Catheter components: Differential pressure sensor) Electrical signal generated by cardiac cycle (figure), 3.5 (figure), 3.5 Flat placement signal (figure), 3.6 Pressure sensor drift, 6.18, 6.19 Pulsatile placement signal (figure), 3.6 Sensor failure, 6.18, 6.19 Dimensions, 8.5 Kinking, 5.4, 5.18 Position alarms (See Patient management: Understanding and managing Impella® 5.0 Catheter position alarms) Returning an Impella® 5.0 Catheter to Abiomed (US), 8.7 Using the catheter (See also Impella® 5.0 System: Using the system) Inserting the catheter, 5.9 (figure), 5.9, 5.10 Positioning and starting the catheter, 5.12 (figure), 5.12, 5.13 Post-insertion positioning (PIP) checklist , 6.11 Preparing the catheter, 5.4 (figure), 5.4, 5.5 Catheter serial number, 4.6, 8.1 (figure), 4.5 Cautions, 1.2 Combitrans-Monitoring Set (CM-Set) Description, 3.11 (figure), 3.11 Replacing the CM-Set, 5.14 Connector cable Connecting the connector cable, 5.4 Connecting to the Impella® 5.0 Catheter (figure), 5.4 Connecting to the Impella® Console (figure), 5.5 Description, 3.11, 8.4 (figure), 3.11 Disconnecting connector cable from console, 5.21 (figure), 5.21 Snapping plastic clip to connector cable (figure), 5.5 Console Alarms, 7.1 Battery power, 3.7, 4.10 Changing to a backup console (See Patient management: How to change to a backup console) Cleaning the console, 8.6 Connections Catheter connector socket (figure), 4.2 Power supply connector, 4.4 (figure), 4.2 Pressure connector socket, 4.4 (figure), 4.2 Console information, 4.3 Console parameters, 8.4 Description, 3.7, 4.1 (figure), 3.7, 4.2 Display elements AC power icon, 4.5 (figure), 4.5 Battery status information, 4.5 (figure), 4.5 Catheter operation icon, 4.5 (figure), 4.5 Catheter type and serial number, 4.6 (figure), 4.5 Flow rate, 4.5 (figure), 4.5 Information and waveform display, 4.6 (figure), 4.5 Low flow alarm limit icon, 4.5 (figure), 4.5 Performance level, 4.5 (figure), 4.5 Signal bar graph, 4.6 (figure), 4.5 Signal numeric readout, 4.6 (figure), 4.5 Time window display, 4.6 (figure), 4.5 Type of signal selected, 4.5 (figure), 4.5 Display screen (figure), 4.2 Dual signal screen, 4.3, 4.8 (figure), 4.8, 5.13 Flow rate screen, 4.3, 4.10 (figure), 4.10 Function keys, 4.2 ▲ and ▼ keys, 4.3, 7.2 (figure), 4.2 FLOW key, 4.3 MENU key, 4.3 OK key, 4.3 ON key, 4.4 P-PERF key, 4.3 SCALE key, 4.4 SIGNAL key, 4.3 ZERO key, 4.3 Information screen (figure), B.2 Mobile operation, 4.10 Motor current screen, 4.3, 4.9 (figure), 4.9 Placement monitoring screen, 4.3, 4.7 (figure), 4.7 Placement signal screen, 4.3, 4.7 (figure), 4.7 Initial placement signal (figure), 5.12 Placement signal as catheter crosses the aortic valve (figure), 5.12 I.1 INDEX Purge pressure screen, 4.3, 4.8 Abnormal purge pressure on 5 minute scale screen (figure), 5.17 (figure), 4.8 Normal purge pressure on 5 minute scale screen (figure), 5.16 Speed screen, 4.3, 4.9 (figure), 4.9 Storage, 8.7 Console function test (See Impella® 5.0 System: Using the system: Console function test) Contraindications, 2.1 D Dextrose, 8.1 (See also Dextrose solution) Dextrose solution Amount of dextrose delivered in purge fluid, 5.19 Description, 3.12, 5.1, 8.1 (figure), 3.12 Dextrose and heparin infusion rates, 5.16 Heparin, 3.12 (See also Patient management: Operating the Impella® 5.0 without heparin in the purge solution) Diagnostic catheter AL1, 5.10 Multipurpose (MP), 5.10 Dual signal (See Console: Dual signal screen; See Patient management: Understanding and managing Impella® 5.0 position alarms) E ECG interference (See Patient management: ECG interference) Echocardiography Color Doppler, 6.10 (figure), 6.10 Transesophageal (TEE), 6.3, 6.4 (figure), 6.5, 6.8 Transthoracic (TTE), 6.3, 6.4 (figure), 6.5, 6.9, 6.10 Explant (See Patient management: Guidelines for explant) F Flow rate (See also Console: Flow rate screen) Accuracy of displayed flow rate, 6.18 Effects of sensor drift, 6.18 Expected flow rate displayed on console, 6.18 Fluoroscopy, 5.9, 6.3 (See also Patient management: Positioning and placement devices) Function test (See Impella® 5.0 System: Using the system: Console function test) G Glucose, 8.1 (See also Dextrose solution) GP IIb-IIIa inhibitors, 5.10 Guidewires (See also Impella® 5.0 System: Using the system: Inserting the catheter) 0.025 inch, 260 cm placement guidewire, 3.12 (figure), 3.12 H Hemolysis (See Patient management: Hemolysis) Heparin (See also Patient management: Operating the Impella® 5.0 Catheter without heparin in the purge solution) Amount of heparin delivered in purge fluid, 5.19 Heparin and dextrose infusion rates, 5.16 High purge pressure (See Purge system: Troubleshooting) I.2 I Impella 5.0 Catheter (See Catheter) Impella® 5.0 Catheter dimensions, 8.5 Impella® 5.0 Power Supply (See Power Supply) Impella® 5.0 System Abbreviations, 8.1 Anatomic considerations, 8.6 Cleaning, 8.6 General system information Abbreviations, 8.2 Ambient conditions, 8.3 Anatomic considerations, 8.6 Cleaning, 8.6 Connector cable parameters, 8.4 Impella® 5.0 Catheter dimensions, 8.5 Impella® 5.0 Catheter parameters, 8.3 Impella® 5.0 Console parameters, 8.4 Impella® 5.0 Power Supply parameters, 8.4 Returning an Impella® 5.0 Catheter to Abiomed (US), 8.7 Storing the Impella® Console, 8.7 Storing the Impella® Power Supply, 8.7 Terminology, 8.1 Impella® 5.0 Catheter parameters, 8.3 Maintenance and repair intervals, B.1 Overview, 3.1 (figure), 3.1 Preventive maintenance Console function test, B.2 Returning an Impella® 5.0 Catheter to Abiomed (US), 8.7 Reusable components, 3.1 (See also Console; See also Power Supply; See also Pressure transducer cable; See also Pump test plug; See also Purge system) Single-use components, 3.2 (See also Catheter; See also Combitrans-Monitoring Set; See also Connector cable; See also Guidewires: 0.025 inch, 260 cm placement guidewire; See also Purge system: Purge tubing) Storage, 8.7 Symbols, 8.2 System configuration (figure), 3.2 System specifications (See Impella® 5.0 System: General system information) Technical safety inspections, B.1 (figure), B.1 Terminology, 8.1 Using the system Checking Impella® Power Supply function, 5.2 Connecting pressure transducer cable (figure), 5.6 Connecting the CM-Set to the catheter (figure), 5.7 Connecting the connector cable, 5.4 (figure), 5.5 Connecting the power supply cable (figure), 5.2 Connecting the pressure transducer cable, 5.6 Console function test (figure), B.2 Disconnecting connector cable from console, 5.21 (figure), 5.21 Flush purge line and purge lumen screen (figure), 5.7 ® Inserting the catheter, 5.9 (figure), 5.9, 5.10 Maintaining ACT, 5.11 Placing the Impella® Power Supply next to the Impella® Console (figure), 5.1 Positioning aids and other displays, 4.7 Positioning and starting the catheter, 5.12 (See also Echocardiography; See also Fluoroscopy) (figure), 5.12, 5.13 Post-insertion positioning (PIP) checklist, 6.11 Preparing the catheter, 5.4 (figure), 5.4, 5.5 Preparing the purge fluid, 5.3 Priming the Impella® 5.0 Catheter purge lumen, 5.7 (figure), 5.8 Ready to insert screen (figure), 5.8 Replacing the purge tubing, 5.14 Selecting performance level, 4.6 Startup, 5.1 Equipment set-up, 5.1 Supplies needed, 5.1 Turning Power Supply on (figure), 5.2 Turning the console off, 4.4 Turning the console on, 4.4 Zeroing the differential pressure sensor, 6.18 Warranty (See Warranty) Weaning, 5.20 Rapid weaning, 5.20 Slow weaning, 5.21 Impella® Console (See Console) Indications, 2.1 L Language selection, 4.3 Latex (See Patient management: Latex) Low purge pressure (See Purge system: Troubleshooting) M Maintenance and repair intervals (See Impella® 5.0 System: Maintenance and repair intervals) Motor current (See Console: Motor current screen; See Patient management: Understanding and managing Impella® 5.0 Catheter position alarms) P Patient management ECG interference, 6.3 Echocardiography for positioning catheter, 6.4 (See also Echocardiography) General patient care considerations, 6.1 Hemolysis, 6.16 Guide for managing hemolysis, 6.17 How to change to a backup console, 6.20 Latex, 6.3 Operating the Impella® 5.0 Catheter without heparin in the purge solution, 6.20 Overview, 6.1 Positioning and placement devices, 6.3 Right heart failure, 6.2 Suction, 6.16, 6.19 Transport within the hospital, 6.2 Understanding and managing Impella® 5.0 Catheter position alarms, 6.11 INDEX Correct catheter position, 6.12 (figure), 6.12 Impella® 5.0 Catheter fully in aorta, 6.13 (figure), 6.13 Impella® 5.0 Catheter outlet area on or near aortic valve, 6.15 Impella® 5.0 fully in ventricle, 6.13 (figure), 6.13 Low native heart pulsatility, 6.14 (figure), 6.14 Performance level Flow rates, 4.6 Selecting performance level, 4.6 Placement monitoring (See Console: Placement monitoring screen; See Patient management: Understanding and managing Impella® 5.0 Catheter position alarms) Placement signal (See Console: Placement signal screen) Plasma-free hemoglobin (PfHgb), 6.16 Post-insertion positioning (PIP) checklist, 6.11 Power Supply (See also Power Supply cable) Checking Impella® Power Supply function, 5.2 Description, 3.8 (figure), 3.8 Impella® Power Supply parameters, 8.4 Indicator lights CAN light, 3.8 (figure), 3.8 CONNECTED light, 3.8 (figure), 3.8 POWER light, 3.8 (figure), 3.8 Rear panel AC power connector, 3.9 (figure), 3.9 CAN-bus interface, 3.9 (figure), 3.9 Fuses, 3.9 (figure), 3.9 Impella® Console connector, 3.9 (figure), 3.9 Power switch, 3.9 (figure), 3.9 Storage, 8.7 Turning Impella® Power Supply on (figure), 5.2 Power Supply cable (See also Impella® 5.0 System: Using the system) Connecting Impella® Power Supply cable to console (figure), 5.2 Connecting Impella® Power Supply cable to power supply (figure), 5.2 Description, 3.9 (figure), 3.9 Pressure sensor drift (See Catheter: Differential pressure sensor) Pressure transducer cable Connecting the cable to the console, 5.6 (figure), 5.6 Description, 3.11 (figure), 3.11 Pump, 8.1 Pump housing, 8.1 Pump test plug Cleaning the pump test plug, 8.6 Definition, 8.1 Description, 3.12 (figure), 3.12 Use in console function test, B.2 Purge fluid (See Impella® 5.0 System: Using the system: Preparing the purge fluid) Purge pressure (See Console: Purge pressure screen; See Purge system: Troubleshooting) Definition, 8.1 Preventing kinking, 5.5 Purge pressure management strategies, 5.19 (figure), 5.19 Purge system Description, 3.10 (See also B. Braun Vista basic Infusion Pump) (figure), 3.10 Preventing kinking, 5.4 Producing the bolus, 5.7 Purge tubing, 3.12 (figure), 3.12 Replacing the purge tubing, 5.14 Stopping the bolus, 5.7 Troubleshooting, 5.15 Dextrose delivered in purge fluid, 5.19 Downstream occlusion, 5.20 (figure), 5.20 Heparin in purge fluid, 5.19 High purge pressure, 5.18 (figure), 5.18 Low purge pressure, 5.15 (figure), 5.15 Purge pressure management strategies, 5.19 W Warnings Console warnings, 7.3 High purge pressure (figure), 5.18 Low purge pressure (figure), 5.15 System warnings, 1.1 Warranty Impella® System Limited Service Warranty (US), A.1 Weaning (See also Impella® 5.0 System: Weaning) X X-ray Use in Impella® 5.0 Catheter positioning, 6.3 Z Zeroing the differential pressure sensor (See Impella® 5.0 System: Using the system: Zeroing the differential pressure sensor) R Repositioning Guide, 4.3 Retrograde flow, 4.6, 8.1 Returning an Impella® 5.0 Catheter to Abiomed (US), 8.7 Right heart failure (See Patient management: Right heart failure) S Speed (See Console: Speed screen) Suction (See Patient management: Suction) Symbols, 8.2 T Terminology and abbreviations, 8.1 Test plug (See Pump test plug) Transesophageal echocardiography (TEE), 6.3, 6.4 (See also Echocardiography) Transport within the hospital (See Patient management: Transport within the hospital) Transthoracic echocardiography (TTE), 6.3, 6.4 (See also Echocardiography) V Vista basic Infusion Pump (See B. Braun Vista basic Infusion Pump) I.3 Clinical support 24 hours per day, 7 days a week: 1-800-422-8666 (US) +49 (0) 1805 2246633 (EU) www.abiomed.com Abiomed, Inc. 22 Cherry Hill Drive Danvers, Massachusetts 01923 USA Voice: 978-777-5410 Facsimile: 978-777-8411 Email: [email protected] Abiomed Europe GmbH Neuenhofer Weg 3 52074 Aachen, Germany Voice: +49 (241) 8860-0 Facsimile: +49 (241) 8860-111 Email: [email protected]