selectra 1/ 2/e

Transcription

selectra 1/ 2/e
SELECTRA 1/ 2/E
ÁCIDO ÚRICO
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME
PREDILUITION:
SLOPE BLANK:
DELAY, MIN.TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
INCUBATION TIME:
37ºC
ÁCIDO ÚRICO
AUR
ENDPOINT
546 nm
mg/dL
1
0.00 mg/dL
25 mg/dL
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF.MALE LOW:
REF.MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT.FACTOR:
CORRELAT. OFFSET:
*
NO
2.5
7
1.5
6
*
*
*
*
*
1,000
0.000 mg/dL
AUR
NO
25 mL
250 µL
250 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENT BLANK:
CAL. LOW LIMIT:
CAL. HIGH LIMIT:
FACTOR:
(- 0.100 )ABS
3.000 ABS
(- 0.100) ABS
3.000 ABS
YES #
0.000
3,000
#
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENT BLANK:
CAL. LOW LIMIT:
CAL. HIGH LIMIT:
FACTOR:
(- 0.100 )ABS
3.000 ABS
(- 0.100) ABS
3.000 ABS
YES #
0.000
3,000
#
6 µL
5 µL
5 mL
0 µL
0 µL
NO
NO
4.5 MIN
AUR
NO
25 mL
250 µL
250 µL
6 µL
2 µL
4.5 MIN.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
ALBUMINA
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
INCUBATION TIME:
37 ºC
ALBUMINA
ALB
ENDPOINT
620 nm
g/dL
2
0.00 g/dL
6.00 g/dL
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
3.5
5.5
3.5
5.5
*
*
*
*
*
1,000
0.000 g/dL
ALB
NO
25 mL
300 µL
301 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENT BLANK:
CAL. L. LIMIT:
CAL. H. LIMIT
FACTOR:
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENT BLANK:
CAL. L. LIMIT:
CAL. H. LIMIT:
FACTOR:
(- 0.100 ABS)
3.000 ABS
(- 0.100 ABS)
3.000 ABS
YES #
(-0.100 )ABS
3,000
#
3 µL
2 µL
5 mL
0 µL
0 µL
No
No
11.5 MIN.
ALB
NO
25 mL
300 µL
301 µL
3 µL
2 µL
11.5 MIN.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
AMILASE
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37 ºC
AMILASE
AMIL
KINETIC
405 nm
U/L
0
10 U/L
1500 U/L
*
*
*
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
30
130
30
130
*
*
*
*
*
1,000
0.000 U/L
AMI
NO
25 mL
250 µL
253 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENT BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
0.600 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENT BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
0.600 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
5 µL
3 µL
5 mL
0 µL
0 µL
NO
NO
50, 186 SEC.
AMILASE
NO
25 mL
250 µL
253 µL
5 µL
3 µL
50,1860 SEC.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
BILIRRUBINA DIRETA
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
37ºC
BILIRRUBINA DIRETA
BD
ENDPOINT
525 nm
mg/dL
2
0
18
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 mg/dL
BIL
YES
25 mL
400 µL
401 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.200 ABS
3.000 ABS
YES #
*
25 µL
12 µL
5 mL
40 µL
40 µL
NO
NO
6.5 MIN.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: ( 11) 3429-2555/ [email protected]
SELECTRA1/ 2/E
BILIRRUBINA TOTAL
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
37ºC
BILIRRUBINA TOTAL
BT
ENDPOINT
546 nm
mg/dL
2
0
18
*
1
1
*
*
CUT OFF:
*
PROZONE CHECK: NO
REF. MALE LOW:
*
REF. MALE HIGH:
*
REF. FEMALE LOW: *
REF. FEMALE HIGH: *
REF. PED. LOW:
*
REF. PED HIGH:
*
CONTROL 1:
*
CONTROL 2:
*
CONTROL 3:
*
CORRELAT. FACTOR: 1,000
CORRELAT. OFFSET: 0.000 mg/dL
BIL
YES
25 mL
400 µL
401 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
25 µL
12 µL
5 mL
40 µL
40 µL
NO
NO
6.5 MIN.
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.200 ABS
3.000 ABS
YES #
*
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
CÁLCIO ARSENAZO
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
CÁLCIO ARSENAZO
CAL ARS
ENDPOINT
620 nm
mg/dL
1
0.0 mg/dL
30 mg/dL
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
8.5
10.5
8.5
10.5
*
*
*
*
*
1,000
0.000 mg/dL
CAL ARS
NO
5 mL
300 µl
303 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100 )ABS
3.000 ABS
(-0.100 )ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
6 µL
3 µL
5 mL
0 µL
0 µL
NO
NO
4.5 MIN
CAL ARS
NO
5 mL
300 µl
303 µL
25 µL
12 µL
4.5 MIN
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: ( 11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
CÁLCIO
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
CÁLCIO
CAL
ENDPOINT
578 nm
mg/dL
1
0.0 mg/dL
20 mg/dL
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
8.8
11
8.8
11
*
*
*
*
*
1,000
0.000 mg/dL
CAL
NO
25 mL
200 µl
200 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.200 ABS
3.000 ABS
YES #
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.200 ABS
3.000 ABS
YES #
#
4 µL
3 µL
25 mL
200 µL
200 µL
NO
NO
11.5 MIN.
CAL
NO
25 mL
300 µL
303 µl
3 µL
2 µL
11.5 min.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
COLESTEROL HDL
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
COLESTEROL HDL
HDL
ENDPOINT
505 nm
mg/dL
0
0
150
PADRÃO
1
1
40
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
HDL
NO
25 mL
230 µL
243 µL
LINEARITY LIMIT:
15%
LOW ABSORBANCE:
(-0.100) ABS
HIGH ABSORBANCE:
3.000 ABS
R. ABS. L. LIMIT:
(-0.100) ABS
R. ABS. H. LIMIT:
3.000 ABS
R. ABS. DEVIATION
3.000 ABS
REAGENTE BLANK:
YES #
CAL. L. LIMIT CAL. L. LIMIT (-0.100 )ABS
CAL. H. LIMITCAL. H. LIMIT3.000 ABS
FACTOR:
#
23 µL
20 µL
5 mL
0 µL
0 µL
NO
NO
11.5 MIN.
HDL
NO
25 mL
230 µL
243 µL
23 µL
20 µL
11.5 MIN.
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 mg/dL
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: ( 11) 3429-2555/ [email protected]
SELECTRA 1 / 2/E
COLESTEROL HDL DIRETO
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
37ºC
HDL
HDL
TWO POINT
578 nm
mg/dL
0
0
130
*
*
*
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 mg/dL
HDL
NO
25 mL
270 µl
275 µL
POINT 1.2
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
6.236 SEC
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
3µL
2 µL
5 mL
90 µL
85 µL
NO
NO
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1 / 2/E
COLESTEROL TOTAL
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
COLESTEROL TOTAL
COL
ENDPOINT
505 nm
mg/dL
0
30
500
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
0
200
0
200
*
*
*
*
*
1,000
0.000 mg/dL
COL
NO
25 mL
300 µl
301 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100 )ABS
3.000 ABS
(-0.100 )ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.200 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
3 µL
2 µL
5 mL
0 µL
0 µL
NO
NO
11.5 MIN.
COL
NO
25 mL
300 µL
301 µL
3 µL
2 µL
11.5 MIN.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: ( 11) 3429-2555/ [email protected]
SELECTRA1/ 2/E
CK MB
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
CK MB
CK MB
KINETIC
340 nm
U/L
0
0
600
*
*
*
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
NO
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
CK MB
NO
25 mL
250 µL
252 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
10%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
8254
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
10%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
3.000 ABS
YES #
8254
10 µL
8 µL
5 mL
0 µL
0 µL
NO
NO
226.196 SEC
CK MB
NO
25 mL
250 µL
252 µL
10 µL
8 µL
226.196 SEC.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
CK TOTAL
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
CK TOTAL
CK
KINETIC
340 nm
U/L
0
0
1000
*
*
*
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
NO
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
CK Total
NO
25 mL
250 µL
265 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
10%
(-0.100) ABS
3.000 ABS
1.100 ABS
3.000 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
10%
(-0.100) ABS
3.000 ABS
1.100 ABS
3.000 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
10 µL
5 µL
5 mL
0 µL
0 µL
NO
NO
129.293 SEC.
CK Total
NO
25 mL
250 µL
265 µL
10 µL
5 µL
129. 293 SEC.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: ( 11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
CREATININA
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
CREATININA
CRE
TWO POINT
505 nm
mg/dL
1
0.0 mg/dL
10 mg/dL
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 mg/dL
CRE
NO
25 mL
230 µl
248 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
23 µL
20 µL
5 mL
0 µL
0 µL
NO
NO
24, 77 SEC.
CRE
NO
25 mL
230 µL
248 µL
23 µL
20 uL
24, 77 SEC.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555/ [email protected]
SELECTRA1/ 2/E
DESIDROGENASE LÁCTICA
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37 ºC
LDH UV
LDH
KINETIC
340 nm
U/L
0
0 U/L
1500 U/L
*
*
*
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
LDH
NO
25 mL
250 µL
253 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
1.000 ABS
3.000 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 Abs
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.000 ABS
3.000 ABS
1.000 ABS
3.000 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
5 µL
2 µL
5 mL
0 µL
0 µL
NO
NO
50,186 SEC.
LDH
NO
25 mL
250 µL
253 µL
5 µL
2 µL
50,186 SEC.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/2/E
FERRO CROMAZUROL
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37 ºC
FERRO CRX
FE
ENDPOINT
620 nm
ug/dL
0
0
500
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
NÃO
NÃO
0
0
0
0
0
0
*
*
*
1,000
0.000
FERROCRX
NO
25 mL
250 µL
255 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100) ABS
3.000 ABS
(-0.100) ABS
(-0.100) ABS
(-0.100) ABS
(-0.100) ABS
(-0.100) ABS
(-0.100) ABS
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
YES #
(-0.100) ABS
3.000 ABS
#
10 µL
8 µL
5 mL
0 µL
0 µL
NO
4.5 MIN.
FERROCRX
NO
25 mL
250 µL
255 µL
10 µL
8 µL
4.5 MIN.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA1/ 2/E
FERRO
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
DELAY, MIN. TIME:
37 ºC
FERRO
FE
ENDPOINT
576 nm
µg/dL
0
10
800
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
NO
NO
0
0
0
0
0
0
*
*
*
1,000
0.000
FERRO
NO
25 mL
200 µL
200 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
-0.1
3,000
-0.1
1,000
YES (-0.010)
*
*
50 µL
50 µL
5 mL
10 µL
10 µL
NO
6.5 MIN
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/2/E
FOSFATASE ALCALINA
TEMPERATURE: 37ºC
NAME:
FOSF. ALC.
ABBR. NAME:
FAL
MODE:
KINETIC
WAVELENGTH:
405 nm
UNITS:
U/L
DECIMALS:
0
LOW CONC:
15 U/L
HIGH CONC:
1200 U/L
CALIBRATOR NAME*
REPEAT:
*
NUMBER:
*
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
FOSF. ALC.
NO
25 mL
250 µL
253 µL
5 µL
2 µL
5 mL
0 µL
0 µL
No
No
50, 186 SEC.
FOSF. ALC.
NO
25 mL
250 µL
253 µL
5 µL
2 µL
50, 186 SEC.
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.500 ABS
0.300 ABS
YES #
6599.68
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.500 ABS
0.300 ABS
YES #
6599.68
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
FÓSFORO
TEMPERATURE: 37ºC
NAME:
FÓSFORO
ABBR. NAME:
FOS
MODE:
ENDPOINT
WAVELENGTH: 340 nm
UNITS:
mg/dL
DECIMALS:
0
LOW CONC:
0
HIGH CONC:
15
CALIBRATOR NAM
*
REPEAT:
1
NUMBER:
1
CONCENTRATION*
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
FOS
NO
25 mL
300 µL
301 µL
3 µL
2 µL
5 mL
0 µL
0 µL
No
No
11.5 MIN
FOS
NO
25 mL
300 µL
301 µL
3 µL
2 µL
11.5 MIN.
CUT OFF:
*
PROZONE CHECK: NO
REF. MALE LOW:
*
REF. MALE HIGH:
*
REF. FEMALE LOW: *
REF. FEMALE HIGH: *
REF. PED. LOW:
*
REF. PED HIGH:
*
CONTROL 1:
*
CONTROL 2:
*
CONTROL 3:
*
CORRELAT. FACTOR 1,000
CORRELAT. OFFSET 0.000 mg/dL
LOW ABSORBANCE: (-0.100) ABS
HIGH ABSORBANCE: 3.000 ABS
R. ABS. L. LIMIT:
(-0.100) ABS
R. ABS. H. LIMIT:
3.000 ABS
REAGENTE BLANK: YES #
CAL. L. LIMIT
(-0.100 )ABS
CAL. H. LIMIT
3.000 ABS
FACTOR:
16817
LOW ABSORBANCE: (-0.100) ABS
HIGH ABSORBANCE: 3.000 ABS
R. ABS. L. LIMIT:
(-0.100) ABS
R. ABS. H. LIMIT:
3.000 ABS
REAGENTE BLANK: YES #
CAL. L. LIMIT
(-0.100 )ABS
CAL. H. LIMIT
3.000 ABS
FACTOR:
16817
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
GAMA GT
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
GAMA GT
GGT
KINETIC
405 nm
U/L
0
0 U/L
700 U/L
*
*
*
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
GAMA GT
NO
25 mL
260 µL
265 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
1.600 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
16817
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.600 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
16817
13 µL
8 µL
5 mL
0 µL
0 µL
NO
NO
77, 132 SEC.
GAMA GT
NO
25 mL
260 µL
265 µL
13 µL
8 µL
77, 132 SEC.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11 ) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
GLICOSE
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
GLICOSE
GLI
ENDPOINT
505 nm
mg/dL
0
0.0 mg/dL
400 mg/dL
*
*
*
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
70
99
70
99
*
*
*
*
*
1,000
0.000 mg/dL
GLI
NO
25 mL
300 µl
301 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100 )ABS
3.000 ABS
(-0.100 )ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.200 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
3 µL
2 µL
5 mL
0 µL
0 µL
NO
NO
11.5 MIN.
GLI
NO
25 mL
300 µL
301 µL
3 µL
2 µL
11.5 MIN.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11 ) 3429-2555/ [email protected]
SELECTRA 1/2/E
LACTATO
TEMPERATURE: 37ºC
NAME:
LACTATO
ABBR. NAME:
LAC
MODE:
ENDPOINT
WAVELENGTH:
540 nm
UNITS:
mmol/L *
DECIMALS:
0
LOW CONC:
*
HIGH CONC:
*
CALIBRATOR NAME*
REPEAT:
*
NUMBER:
*
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
MAG
NO
25 mL
300 µL
301 µL
3 µL
2 µl
5 mL
0 µL
0 µL
No
No
11.5 MIN.
LAC
NO
25 mL
300 µL
301 µL
3 µL
2 µL
11.5 MIN
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/ 2/E
LIPASE
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
37ºC
LIP
LIP
TWO POINT
578 nm
U/L
0
3
300
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 mg/dL
LIP
NO
25 mL
220 µL
220 µL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
FACTOR:
0.000 ABS
3.000 ABS
0.000 ABS
3.000 ABS
YES #
#
4 µL
2 µL
5 mL
110 µL
110 µL
NO
NO
130, 130 SEC.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA 1/2/E
MAGNÉSIO
TEMPERATURE:
NAME:
ABBR. NAME:
MODE:
WAVELENGTH:
UNITS:
DECIMALS:
LOW CONC:
HIGH CONC:
CALIBRATOR NAME:
REPEAT:
NUMBER:
CONCENTRATION:
INTERVAL:
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
37ºC
MAGNÉSIO
MAG
ENDPOINT
505 nm
mg/dL
0
0
4.5
*
1
1
*
*
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
MAG
NO
25 mL
300 µL
301 µL
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
(-0.100) ABS
3.000 ABS
(-0.100) ABS
3.000 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.000 ABS
3.000 ABS
0.000 ABS
1.200 ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
3 µL
2 µL
5 mL
0 µL
0 µL
NO
NO
4.5 MIN.
MAG
NO
25 mL
300 µL
301 µL
3 µL
2 µL
4.5 MIN.
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555 / [email protected]
SELECTRA 1/ 2/E
PROTEÍNAS TOTAIS
TEMPERATURE:
37ºC
NAME:
PROTEÍNAS TOTAIS
ABBR. NAME:
PT
MODE:
ENDPOINT
546 nm
WAVELENGTH:
UNITS:
g/dL
DECIMALS:
1
LOW CONC:
0
HIGH CONC:
11
CALIBRATOR NAME *
REPEAT:
1
NUMBER:
1
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
PT
NO
25 mL
300 µL
301 µL
6 µL
5 µL
5 mL
0 µL
0 µL
NO
NO
11.5 MIN.
PT
YES
25 mL
300 µL
301 µL
6 µL
5 µL
11,5 MIN.
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
NO
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 g/dL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100 )ABS
3.000 ABS
(-0.100 )ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(- 0.100) ABS
3.000 ABS
(- 0.100) ABS
0.300 ABS
YES
(- 0.100) ABS
3,000
#
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]
SELECTRA1/ 2/E
PROTEÍNA URINÁRIA
TEMPERATURE:
37ºC
NAME:
PROTURI
ABBR. NAME:
PROTURI
ENDPOINT
MODE:
WAVELENGTH:
620 nm
UNITS:
mg/dL
DECIMALS:
1
LOW CONC:
0
HIGH CONC:
100
CALIBRATOR NAME *
REPEAT:
1
NUMBER:
1
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
PROTURI
YES
25 mL
240 µL
248 µL
12 µL
3 µL
5 mL
0 µL
0 µL
No
No
11.5 MIN.
PROTURI
YES
25 mL
240 µL
248 µL
12 µL
3 µL
11,5 MIN.
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
NO
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 g/dL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100 )ABS
3.000 ABS
(-0.100 )ABS
0.500
YES #
(-0.100 )ABS
3.000 ABS
#
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(- 0.100) ABS
3.000 ABS
(- 0.100) ABS
0.500 ABS
YES
(- 0.100) ABS
3,000
#
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555 / [email protected]
SELECTRA 1/ 2/E
TGP / ALT
TEMPERATURE:
37ºC
NAME:
TGP / ALT
ABBR. NAME:
ALT
MODE:
KINETIC
WAVELENGTH:
340 nm
UNITS:
U/L
DECIMALS:
0
LOW CONC:
0
HIGH CONC:
440
CALIBRATOR NAME *
REPEAT:
*
NUMBER:
*
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
TGP / ALT
NO
25 mL
230 µl
248 µL
23 µL
5 µL
5 mL
0 µL
0 µL
NO
NO
50, 186 SEC.
TGP / ALT
NO
25 mL
230 µL
248 µL
23 µL
5 µL
50, 186 SEC.
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.800 ABS
3.000 ABS
1.000 ABS
3.000 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
(-1750.3)
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.800 ABS
3.000 ABS
1.000 ABS
3.000 ABS
0.200 ABS
YES #
(-0.100 )ABS
3.000 ABS
(-1750.3)
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555 / [email protected]
SELECTRA 1/2/E
AST / TGO
TEMPERATURE:
37ºC
NAME:
ASTGOT
ABBR. NAME:
AST
MODE:
KINETIC
WAVELENGTH:
340 nm
UNITS:
U/L
DECIMALS:
0
LOW CONC:
0
HIGH CONC:
440
CALIBRATOR NAME *
REPEAT:
1
NUMBER:
1
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
ASTGOT
NO
25 mL
250 µL
265 µL
25 µL
10µL
5 mL
0 µL
0 µL
NO
NO
103, 133 SEC.
ASTGOT
NO
25 mL
250 µL
265 µL
25 µL
10 µL
103, 133 SEC.
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.800 ABS
3.000 ABS
1.000 ABS
3.000 ABS
0.200 ABS
YES #
(-0.100 )ABS
3.000 ABS
(-1565.3)
LINEARITY LIMIT:
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
15%
0.800 ABS
3.000 ABS
1.000 ABS
3.000 ABS
0.200 ABS
YES #
(-0.100 )ABS
3.000 ABS
(-1565.3)
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555 / [email protected]
SELECTRA 1/2/E
TRIGLICÉRIDES
TEMPERATURE:
37ºC
NAME:
TRIGLICÉRIDES
ABBR. NAME:
TRIG
MODE:
ENDPOINT
WAVELENGTH:
505 nm
UNITS:
mg/dL
DECIMALS:
0
LOW CONC:
0
1100
HIGH CONC:
CALIBRATOR NAME *
REPEAT:
1
NUMBER:
1
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
TRIG
No
25 mL
300 µL
301 µL
3 µL
2 µL
5 mL
0 µL
0 µL
NO
NO
11.5 MIN.
TRIG
NO
25 mL
300 µL
301 µL
3 µL
2 µL
11,5 MIN.
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
NO
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 U/L
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(-0.100 )ABS
3.000 ABS
(-0.100 )ABS
3.000 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
(- 0.100) ABS
3.000 ABS
(- 0.100) ABS
0.300 ABS
YES
0.000
3,000
#
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL
SAC: (11) 3429-2555 / [email protected]
SELECTRA 1/ 2/ E
URÉIA UV
TEMPERATURE:
37ºC
URÉIA UV
NAME:
ABBR. NAME:
URÉIA
MODE:
KINETIC
WAVELENGTH:
340 nm
UNITS:
mg/dL
DECIMALS:
0
LOW CONC:
0
HIGH CONC:
250
CALIBRATOR NAME *
1
REPEAT:
NUMBER:
1
CONCENTRATION: *
INTERVAL:
*
DUAL MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
R2 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
PREDILUITION:
SLOPE BLANK:
DELAY, MIN. TIME:
MONO MODE
NAME:
SAMPLE BLANK:
R1 BOTTLE:
NORMAL VOLUME:
RERUN VOLUME:
SAMPLE:
NORMAL VOLUME:
RERUN VOLUME:
DELAY, MIN. TIME:
URÉIA
NO
25 mL
300 µL
303 µL
3 µL
2 µL
5 mL
0 µL
0 µL
No
NO
77, 132 SEC.
URÉIA
NO
25 mL
300 µL
301 µL
3 µL
2 µL
77, 132 SEC.
CUT OFF:
PROZONE CHECK:
REF. MALE LOW:
REF. MALE HIGH:
REF. FEMALE LOW:
REF. FEMALE HIGH:
REF. PED. LOW:
REF. PED HIGH:
CONTROL 1:
CONTROL 2:
CONTROL 3:
CORRELAT. FACTOR:
CORRELAT. OFFSET:
*
NO
*
*
*
*
*
*
*
*
*
1,000
0.000 mg/dL
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
CAL. L. LIMIT
CAL. H. LIMIT
FACTOR:
0.800 ABS
3.000 ABS
1.000 ABS
3.000 ABS
0.300 ABS
YES #
(-0.100 )ABS
3.000 ABS
#
LOW ABSORBANCE:
HIGH ABSORBANCE:
R. ABS. L. LIMIT:
R. ABS. H. LIMIT:
R. ABS. DEVIATION
REAGENTE BLANK:
FACTOR:
0.800 ABS
3.000 ABS
1.000 ABS
3.000 ABS
0.200 ABS
YES #
#
* Indica que os dados serão inseridos pelo operador. # Calculado pelo aparelho.
Para esclarecimentos de dúvidas, favor entrar em contato com o SAC.
Assessoria Técnica KATAL:
SAC: (11) 3429-2555/ [email protected]