Shortwave Diathermy
Transcription
Shortwave Diathermy
SHORTWAVE DIATHERMY 167 7 Therapeutic Fields: Shortwave Diathermy 'Shortwave diathermy' refers to heating of deeply located tissue using electric or magnetic fields which alternate at high frequency. The term 'shortwave diathermy', is something of a misnomer as the contribution of waves, as such, to the treatment is negligible. The physiological effects are a result of electric and magnetic fields generated by the shortwave diathermy apparatus. Shortwave radiation plays little or no role in the therapy. The apparatus used by physiotherapists generates alternating electric and magnetic fields with a frequency of 27.12 MHz. Since radio waves with frequencies in the range 10 MHz to 100 MHz are termed short waves the term has been, rather inappropriately, applied to this therapeutic modality. While shortwave diathermy units do radiate waves with a frequency of 27.12 MHz, this is a side-effect. The physiological effects are due to the powerful electric or magnetic fields generated by the apparatus. PRODUCTION OF THE FIELD Shortwave diathermy apparatus consists of a sinewave generator circuit which produces alternating current with a frequency of 27.12 MHz and a resonant circuit which can be tuned to exactly the same frequency. The sinewave generator supplies energy to the resonant circuit by transformer action. Figure 7.1 illustrates the arrangement. Figure 7.1 Shortwave diathermy apparatus (schematic). to table of contents SHORTWAVE DIATHERMY 168 The sinewave generator consists of a power supply (chapter 5), an oscillator with good frequency stability (chapters 2 and 5) and a power amplifier (chapter 5). The power supply converts AC from the mains (of frequency 50 Hz) to DC which is needed to power the equipment. It consists of a transformer (to convert the 240 V AC from the mains to the voltage needed by the rest of the circuitry), and a rectifier to convert the AC to DC. The DC is used to power a sinewave generator; a resonant circuit which oscillates at 27.12 MHz and an amplifier, which boosts the current produced by the resonant circuit to higher levels, as needed for patient treatment. Electrical energy produced by the sinewave generator is coupled to the patient tuning circuit by transformer action (figure 7.1). Two inductors are placed close together so that energy produced by the power amplifier is transferred to the patient circuit. This method of coupling ensures that DC in the apparatus is unable to reach the patient and the risk of electric shock is minimized. A variable capacitor, C, is included in the patient circuit so that the resonant frequency of the patient circuit can be made equal to the frequency of the oscillator. This ensures maximum efficiency of energy transfer (chapter 2) and reliable operation of the apparatus. A power meter or indicator lamp shows when resonance is achieved and maximum power is transferred. In older machines, the variable capacitor, C, was manually adjusted with the operator adjusting a knob while observing the power meter and adjusting for maximum power. Modern machines use electronic control of the variable capacitor and are described as 'auto-tuning'. The principal advantage of automatic tuning is that if the patient should move during treatment the machine will adjust to keep the patient circuit in resonance. With manual tuning machines, movement of the patient or electrodes can result in de-tuning and a drop in output of the machine. Any mains-frequency AC produced by the apparatus is also not conducted appreciably to the patient circuit as the resonant frequency (27.12 MHz) is vastly different to the mains frequency (50 Hz). The output of the apparatus is coupled to the patient via electrodes (in the capacitor field technique represented in figure 7.1) or via an induction coil. The coil or electrodes are connected directly to the output of the machine and the part of the to table of contents SHORTWAVE DIATHERMY 169 patient to be treated is positioned in the electric or magnetic field. In figure 7.1, the area highlighted in yellow is circuitry inside the machine. The part of the patient to be treated would be positioned between the external capacitor plates shown in figure 7.1. The plates are normally in the form of two metal disks, each inside a clear plastic container or envelope. The electrical characteristics of the patient's tissue affects the capacitance of the patient circuit, as does the electrode size and spacing. For this reason it is necessary that the apparatus be tuned (by adjusting C in figure 7.1) with the patient positioned in the field. Similarly, if an induction coil is used rather than capacitor plates, tuning will be necessary. This is because when the coil is wrapped around the part of the patient to be treated, the inductance of the coil will depend on the number of turns of the coil and their radius. When an induction coil is used, the presence of biological tissue in the field is irrelevant but the tissue volume to be treated will influence the number of turns of the coil and their radius. MOLECULES IN AN ELECTRIC FIELD In shortwave diathermy treatment a high frequency AC electrical signal is produced and applied to the patient via an induction coil or electrodes. The high frequency signal will produce a corresponding high frequency alternating electric or magnetic field in the patient's tissue. We now consider what effect this has on the tissue. Since an alternating magnetic field gives rise to an induced alternating electrical field (as described in chapter 6) we first examine the effects of an alternating electric field on the different molecules found in human tissue. Charged Molecules The conductivity of tissue is determined by the number of free ions in the tissue fluid. In the presence of an electric field these ions will migrate along field lines and so constitute an electric current. The process is not unlike electrical conduction in metals. Metallic conduction results from the movement of free electrons. In electrolytes the charge carriers are not electrons but ions; these are tens of to table of contents SHORTWAVE DIATHERMY 170 thousands of times more massive than electrons. Under the influence of the electric field ions will be accelerated along field lines - but they will not travel far before colliding with other molecules and losing their acquired kinetic energy. The repeated sequence of accelerations and collisions is the way in which electrical energy is converted to heat energy, which is the random-motion energy of the molecules. At the frequencies associated with shortwave diathermy the field alternations are so rapid that the ions oscillate about a mean position rather than undergoing any large scale movement, but the alternations are not so rapid that movement is prevented and heat generation is not impaired. Dipolar Molecules Dipolar molecules such as water will orient themselves in an electrical field and if the field is alternating this will result in backwards and forwards rotation of the dipoles. In a liquid the molecules are continually in motion (due to their thermal energy) and are loosely associated with each other (coupled); thus some of the rotational energy of the molecules will be converted to heat energy by what can be thought of as a frictional drag between adjacent molecules. Non Polar Molecules Though not normally polar these molecules will undergo a distortion of their electron 'clouds'; that is, they will polarize in an electric field. In an alternating field the electron clouds will Figure 7.2 Response of molecules to a high frequency alternating electric field. to table of contents SHORTWAVE DIATHERMY 171 oscillate back and forth to each end of the molecule. Since this kind of motion does not involve transport or rotation of the molecule as a whole it can only be coupled indirectly with the gross molecular movement associated with heat energy. Figure 7.2 summarizes, by illustration, the response of ions, polar molecules and non-polar molecules to a high frequency alternating electric field. In each case there is a net back and forth movement of charge: in other words, an alternating flow of current. REAL AND DISPLACEMENT CURRENT From the previous discussion it is apparent that the different kinds of molecule in a material will each respond differently to an applied electric field. The back and forth movement of ions and the consequent collisions will result in a very efficient conversion of electrical energy into heat energy. The rotational movement of polar molecules provides a less efficient mechanism of energy conversion. The electron cloud distortion of non-polar molecules represents the least efficient means of heat production. Nonetheless each kind of molecule responds to an alternating electric field in a way which results in movement of charges and hence an alternating current. The difference is in the proportion of electrical energy converted to heat energy when the alternating current is produced. With this in mind we distinguish real and displacement current. * Real current is that associated with heat production. When real current flows through a material the rate at which electrical energy is converted to heat energy is given by Joule's law: P = V.I .... (1.4) where V is the potential difference and I is the real current flowing through the material. P is the power dissipated (in watts), in other words the amount of electrical energy dissipated per second (1 watt (W) = 1 joule per second (J.s-1)). to table of contents SHORTWAVE DIATHERMY * 172 Displacement current is current flow which does not produce any heating. In this case the power dissipated, and hence the heat generated, is zero. Ionic materials are associated principally with real current and hence substantial heat production. Polar substances are associated with both real and displacement current and hence less heat production. Non-polar materials are principally associated with displacement current and hence minimal heat production. An example which serves to illustrate the distinction between real and displacement current is given in figure 7.3. Here we have a resistor and a capacitor connected in series to a source of alternating current. In this case we suppose that the capacitor is ideal - it comprises two metal plates separated by a perfect insulator which can polarize and depolarize with no loss of electrical energy to heat energy. The magnitude of the current flowing in this circuit will depend on the voltage of the AC source and the total impedance of the resistor/capacitor combination. The actual impedance of the capacitor is calculated using equation 2.5. The real current (Ir) flowing through the resistor will result in power dissipation according to equation 1.4 and hence heat production in the resistor. The displacement current (Id ) flowing through the capacitor (assumed ideal) gives no power dissipation and hence no heat production as the material between the plates is able to polarize and depolarize with no energy loss. Figure 7.3 Real and displacement current in an AC circuit. In this case, then, the current flowing from the AC source appears as real current in the resistor R and displacement current in the (ideal) capacitor C. Charges move and heat is produced in the resistor while the charge movement (displacement current) in the capacitor produces no heating. The two currents, which are different forms of the same thing, are necessarily the same size. to table of contents SHORTWAVE DIATHERMY For a capacitor to be ideal the material between the plates must be an ideal dielectric - a substance capable of polarizing in an electric field and depolarizing on its removal without any dielectric absorption. In other words, with no conversion of electrical energy to heat energy. Biological materials, particularly those with high water and ion content are far from being ideal dielectrics. When placed in an electric field the induced current will be a combination of real and displacement current. The proportions of each kind of current will depend on the proportions of ionic, polar and non-polar molecules. We now consider biological tissue exposed to an electric or magnetic field which alternates at a frequency of 27.12 MHz, the frequency licensed for use in shortwave diathermy. As we have seen, shortwave diathermy may be applied using capacitor plates (which produce an electric field) or an inductive coil (which generates a magnetic field). 173 Most gases come close to being ideal dielectrics, as do some oils. Water being a highly polar molecule, falls short of this ideal and dielectric absorption results in significant heating at any frequency below about 1010 Hz. CAPACITOR FIELD TREATMENT Consider first the situation depicted in figure 6.19(a), where an arm or leg is positioned between two capacitor plates. Figure 6.19(a) shows the electric field pattern, which is affected by refraction and termination of field lines. The total current flowing through the tissue will be determined by the total impedance of the tissue plus the air space between tissue and capacitor plates. Current will flow in the direction of the field lines and the proportions of real and displacement current will depend on the electrical properties of the particular tissue. The amount of heating in any tissue layer will be determined by two factors: the field intensity within the layer and the amount of real, rather than displacement, current. Calculation of the proportions of real and displacement current in a particular tissue is not difficult. Measured values of dielectric constant and conductivity are all that are to table of contents SHORTWAVE DIATHERMY needed. Calculation of the field pattern is much more difficult and has only been done using simplified models: even simpler than the somewhat idealized geometries shown in figure 6.19. Useful qualitative pictures are nonetheless obtained by combining diagrams such as those shown in figure 6.19, with calculated values of real and displacement current in each tissue layer. 174 The conductivity determines the amount of real current flow, the dielectric constant determines the amount of displacement current. At a frequency of 27.12 MHz the current flow in fatty tissue and bone is approximately 50% displacement. In muscle and tissues of high water content the proportions are approximately 80% real current to 20% displacement current. Figure 7.4 shows a revised view of figure 6.19(a) which takes into account the two kinds of current flow which occur. In the air spaces the current flow is entirely displacement current. In fatty tissue and bone the current is assumed to be one half real current and one half displacement current. For simplicity, muscle is shown as having entirely real current. Figure 7.4 Current type and directions in a model for an arm or leg. to table of contents SHORTWAVE DIATHERMY 175 When viewing diagrams such as these, bear in mind the simplifications made. The pictures can be misleading if interpreted too literally. You should also bear in mind that even a single tissue layer may be inhomogeneous at both the microscopic and macroscopic level. An example of the complications introduced by tissue inhomogeneity is seen with fatty tissue in the shortwave field. Fatty Tissue A practical limitation on the amount of heat which can be produced in deeply located tissue is the heat production in fatty tissue. When using capacitor plates the rate of heating of fatty tissue is always greater than that of the underlying muscle tissue. Part of the reason is that fatty tissue is inhomogeneous. The tissue is not a uniform distribution of cells but a complex structure incorporating regions of high conductivity and dielectric constant: the lymphatic and blood vessels. The high conductivity and dielectric constant of the vessels will result in field lines being focussed or channelled into them with a resulting high local field intensity and corresponding high rate of heating in and near the vessels. The phenomenon is illustrated in figure 7.5. The localized high heat production will result in greater temperature elevation of the vessels than the fatty tissue as a whole and a greater sensation of heat than would be expected if the tissue layer was homogenous. Figure 7.5 Focussing of electric field lines in blood and lymphatic vessels in fatty tissue. INDUCTIVE COIL TREATMENT We now consider application of the shortwave field with an induction coil. The objective is to induce an electric field and hence a flow of current as a result of the alternating magnetic field produced by the coil. In the example illustrated in figure 7.6 a cable carrying the shortwave frequency current is wrapped around a patient's lower to table of contents SHORTWAVE DIATHERMY 176 limb. Figure 7.6(a) shows the inductive coil wound as a solenoid around the patient's lower limb and figure 7.6(b) shows the current pathways in the different tissues. The current pathways shown are predicted assuming that the alternating magnetic field gives rise to an induced EMF in the patient's tissue. In this case the current will follow circular paths parallel to the turns of the coil in figure 7.6(a). Note that in figure 7.6(b) the current through the fatty tissue is shown as half displacement current and half real current while muscle is assumed to have real current only. As indicated previously, this is only an approximation: while the proportion of real current in fatty tissue is about 50%. in muscle it is about 80%. If the coil in figure 7.6 had a large number of closely spaced turns and the coil diameter was small compared to its length, then the magnetic field inside the coil would be uniform and the induced EMF would be the same throughout the tissue volume. Were this the case, the relative amounts of current flow in each tissue would simply depend on the tissue impedance (which is determined by the dielectric constant and conductivity). Figure 7.6 Current flow induced in a limb by inductive coil treatment. to table of contents SHORTWAVE DIATHERMY 177 A complication is that with more widely spaced turns and a relatively large diameter, the magnetic field inside the induction coil will be non-uniform. In an arrangement like that shown in figure 7.6(a), the magnetic field would be strongest close to the coil and decreasing in intensity towards the centre. The highest field intensity is thus in the superficial tissues of the limb. Another Kind of Coil This means thaqt a greater EMF will be induced in the superficial tissues and consequently there will be a greater current flow. Most manufacturers of shortwave diathermy apparatus offer accessories which include a compact coil mounted in a plastic housing. This device is called a monode. The monode is pointed at the part of the patient to be treated so that the coil is in a plane parallel to the skin surface. With this arrangement (figure 7.7), currents are induced which flow in circular paths parallel to the skin surface. The cable supplied with the shortwave machine can, of course, also be wound into a spiral and positioned to produce a similar distribution of induced current. The spiral coil placed parallel to the skin produces more superficial heating than the solenoidal coil (figure 7.6). This is because the magnetic field intensity decreases rapidly with distance from the coil, as the field lines diverge, spreading apart and looping round to the opposite side of the coil. The field spreading is similar to that which occurs at the ends of the coils in figures 6.7 and 7.6. Magnetic field lines become more separated, indicating a weaker magnetic field further from the coil and consequently less induced EMF and less induced current. Hence although the current induced in muscle is mostly real current, the amount of current at depth is much less than with a solenoid (figure 7.6). Figure 7.7 Induced currents with a spiral coil mounted parallel to the skin surface. to table of contents SHORTWAVE DIATHERMY 178 Capacitative Effects A practical complication which occurs with inductive coil treatment, whether with a solenoid or a spiral coil (monode), is that in addition to the currents induced by the magnetic field there is also a pronounced electrostatic effect. There is a certain capacitance between the loops of the coil. In fact whenever a cable or wire is folded back on itself or coiled we have produced a situation where there are two conductors separated by a space; thus we have produced a capacitor. Although in the case of a cable wound as a coil the capacitance is very small, the effect is quite significant at MHz frequencies. The inductive coil behaves as an inductor in parallel with a capacitor. At the high frequencies used for shortwave diathermy the inductance of the coil results in a high impedance to current flow in the cable (equation 2.4). The capacitance associated with the coil presents a lower impedance pathway for current to take (equation 2.5). In consequence the induced current patterns are not as simple as those shown in figure 7.6(b). The electric field between adjacent turns (Figure 7.8(a)) results in current flow along the field lines shown in blue. Because the electric field is stronger closer to the coils, greater current flows and this adds to the current induced by the magnetic field. The consequence is greater current flow in, and greater heating of, superficial tissue (figure 7.8(b)). The electric field between adjacent loops is similar to that between two small electrodes (figure 6.1(c)). The field is most intense close to the cable. A consequence is that there is a risk of burning the superficial tissues with the electric field of the coil rather than Figure 7.8 Electric field pattern (blue lines) between adjacent turns of an inductive coil. to table of contents SHORTWAVE DIATHERMY 179 heating deeper tissue with current induced by the alternating magnetic field. A similar argument applies for a spiral coil. An electric field is produced between adjacent turns within the loop. Close to the coil, the electric field is intense and greater current flows. This adds to the current induced by the magnetic field so there is greater current flow in, and greater heating of, superficial tissue. Superficial heating due to the electric field can be minimized in three ways: (a) by winding the turns of the coil close together, (b) by keeping the cable away from the patient's skin using towelling and/or rubber spacer designed for this purpose and (c) by using an electrostatic shield. Electric field heating effects can also be minimized, in the case of a solenoid, by positioning an earthed metal cylinder between the coil and the patient's limb. If a monode is used, a flat metal plate between the monode and the patient's tissue would be needed. The plate will screen-out the electric field while having little effect on the magnetic field of the coil. The electric field inside the metal cylinder or behind the metal plate would be almost nil because the metal is a good conductor and field lines will terminate on its surface. Most metals are, however, transparent as far as magnetic fields are concerned so the magnetic field is virtually unchanged. Some, but not all, inductive coil applicators are supplied with an inbuilt electric field screen. Screening is an important feature when depth efficient heating is required. When the adjacent turns are closer together, the electric field is actually greater, but it is also more localized to the space between the turns of the coil. In summary, the options with inductive coil treatment are a coil wound around the part of the patient to be treated or a flat coil (monode) positioned over the body part. The difference is the depth efficiency of treatment. A solenoidal coil (figure 7.6) has greater depth efficiency as far as tissue within the coil s concerned. A flat spiral coil (figure 7.7) has greater effect on superficial tissues. With either method of application, there is the risk of excessive superficial heating due to the electric field between adjacent turns of the coil or spiral. the risk is minimized by spacing the coil or spiral away from the patient's superficial tissues. to table of contents SHORTWAVE DIATHERMY 180 ELECTRODE PLACEMENT - CAPACITOR FIELDS With capacitor field treatment, the therapist has more control over the field intensity in different areas than with inductive coil treatment. We have discussed previously how the combination of tissue layers in the part of the patient being treated alters the shape of the electric field. The other factors influencing the field pattern involve the placement of the electrodes. Each factor listed below must be taken into account in the practical application of shortwave diathermy using capacitor field treatment. * The shape of the part of patient in the field. Compare Figure 6.19(a) with 6.19(b). In addition, if the electrodes are placed over any prominence an undesirable concentration of the field can result. * The arrangement of tissues layers in the treated structure. As discussed previously, this factor plays a significant role in determining the final shape of the field. * The size, spacing and orientation of the electrodes. Some examples of the electric field in the absence of any object were shown in figures 6.2 and 6.3. We consider below the effect when the patient is in the field. Electrode Size In general, it is preferable to use electrodes which are somewhat larger than the structure to be treated. This results in the central, more uniform, part of the field being used (figure 6.2). The dielectric constant and conductivity of tissue are much higher than those of air (table 4.2). Thus, with large electrodes, the field lines are bent towards the limb and spreading of the field is minimised. The effect is illustrated in figure 7.9 where the effect of the different tissue layers is ignored for simplicity. to table of contents SHORTWAVE DIATHERMY 181 Compare these with figures 4.19 and 4.20. The use of small electrodes results in an undesirably high field intensity in the superficial tissues. Unequal size electrodes (figure 5.10(c)) can be used to selectively heat tissue located closer to one surface of a limb. Large differences in electrode size, however, can sometimes lead to difficulty in tuning or instability in machine operation. Electrode Spacing The electrode spacing should normally be as wide as possible. In this way the problems associated with a non-uniform field pattern are minimised. The machine itself, however, sets the limit on the maximum spacing which can be used. As the electrodes are moved further apart the capacitance of the two plates decreases. In addition the field intensity (and consequently the rate of heating) will decrease. A point will be reached where the machine can no longer be tuned or insufficient power is available for adequate heating: this sets the limit on the separation of the electrodes. Figure 7.9 Effect of electrode size: (a) correct electrode size (b) electrodes too small (c) arrangement for selective heating. By use of a wide spacing the electrical properties of the tissue have a smaller effect on the overall field pattern and the electrical properties of air play a greater role. Thus the field pattern is more uniform and less subject to variation with movement of the patient in the field. Figure 7.10 illustrates the effect of electrode spacing. In 7.10(a) the electrode to surface distance varies considerably resulting in a local high field intensity to table of contents SHORTWAVE DIATHERMY 182 in the limb. In 7.10(b) the electrode to surface distance does not vary greatly and the field within the limb is more uniform. Clearly, if a relatively uniform field pattern is required the arrangement shown in 7.10(b) is to be preferred. The arrangements shown in 7.10(c) and 7.11(c) are both suitable if we wish to selectively heat one surface of a limb. They would also be suitable for heating a structure which is located close to one surface of a limb or trunk - for example, the hip joint. Electrode Orientation In the examples considered previously the electrodes were placed parallel to each other in order to obtain a relatively uniform heating pattern. However if one part of the surface of a structure is closer to an electrode, the field lines will be concentrated in that region. Figure 7.11 shows electrodes applied to the shoulder. Compare this with figure 6.16. Electrodes which are parallel to each other as in figure 7.11(a) do not give a uniform field because the air spacing varies considerably. The dielectric constant and conductivity of each field-line pathway varies considerably, resulting in variation in the field intensity. In figure 7.11(b) the distance between the plates varies but the electrical characteristics of each pathway are similar: thus the field is relatively uniform. Clearly the arrangement shown in figure 7.11(b) is preferred when uniform heating is the objective. Figure 7.10 Effect of electrode spacing: (a) narrow spacing, (b) wide spacing and (c) unequal spacing. to table of contents SHORTWAVE DIATHERMY 183 In all of the examples discussed previously the arrangement of electrodes is contraplanar: that is, electrodes are placed over opposite sides of a structure. Such an arrangement is needed if deeply located tissue is to be heated. In some circumstances it is preferable to use a coplanar electrode arrangement. For example superficial structures, such as the spine, which are too extensive for contraplanar treatment may be treated in this way. Figure 7.12 shows a coplanar arrangement of electrodes. When using a coplanar arrangement it is very important to ensure that the spacing between electrodes is greater than double the skin to electrode distance. This results in the majority of field lines passing through tissue rather than the air space between the electrodes. Figure 7.11 Effect of electrode orientation. (a) and (c): incorrect orientation (b) correct orientation. Figure 7.12 A coplanar arrangement of electrodes. to table of contents SHORTWAVE DIATHERMY 184 Even when using a contraplanar arrangement of electrodes considerable heating occurs in the superficial tissues closest to the electrodes. This effect can be minimized by using the cross-fire technique of treatment shown in figure 7.13. Half of the treatment is given with electrodes in one position (figure 7.13(a)). The electrodes are then moved so that the new electric field is at right angles to the old one (figure 7.13(b)) and the treatment is continued. In this way deeply located tissue receives treatment for twice as long as the skin. The cross-fire treatment may be used, for example, on the knee joint or pelvic organs. It is also particularly useful for treating the walls of cavities within a structure, for example the sinuses. Figure 7.14 shows the field pattern obtained with an object of high dielectric constant which has an air-filled hollow at its centre. Figure 7.13 The cross-fire technique. The field lines are concentrated in the dielectric resulting in uneven heating of the walls of the cavity. Cross-fire treatment ensures that all of the cavity wall area is treated. HEATING OF TISSUE Earlier we discussed qualitatively and in molecular terms, the heating effect of a high frequency alternating electric field. We now consider heat production and temperature rise and take a larger scale view of matter: a view at the level of tissue rather than molecules. Figure 7.14 A hollow dielectric between capacitor plates. We saw in chapter 1 that the power dissipated by a resistor, the rate at which electrical energy is converted to heat energy, is given by equation 1.4: P = V.I .... (1.4) to table of contents SHORTWAVE DIATHERMY 185 This expression relates the current, I, flowing through a resistor to the total power, P, dissipated in the resistor. For resistors the current, I, is entirely real current and thus produces heat. When we consider biological tissues we must distinguish between real current and displacement current since only the real current results in heat production. In additional, we are usually more interested in the rate of heating at a particular point in the tissue rather than in the tissue as a whole. In this case a more useful expression of equation 1.4 is equation 7.1. Pv = E.ir .... (7.1) Here Pv is the power dissipated per unit volume of tissue at a particular point. The units of Pv are thus watts per cubic metre. E is the field strength (in volts per metre) and ir is the real component of current density (in amps per square metre) at that point. The power dissipated, Pv is equal to the rate of heat production. Hence, in order to determine the rate of heating at a particular point in tissue we need to know the electric field strength and the real current density. We begin by considering fields and currents produced using capacitor field treatment. Capacitor Field Treatment Whether electrodes are positioned in a coplanar arrangement (figure 7.12) or in a contraplanar arrangement (figures 7.9 to 7.11) the current flow in muscle will be determined by the total impedance of the tissue combination plus the air space between the tissue and capacitor plates. Figure 7.15 shows electrical equivalent circuits for the two electrode/tissue arrangements. The quantities Za, Zf, and Zm refer to the electrical impedances of air, fat and muscle respectively. to table of contents SHORTWAVE DIATHERMY 186 Figure 7.15 Electrode/tissue configurations and their electrical equivalent circuits. (a) coplanar arrangement, (b) contraplanar arrangement. In figure 7.15(a) we ignore (displacement) current flow through the air directly between the electrodes. We also ignore current flowing directly through the fatty tissue and bypassing the muscle. If the electrode spacing is at least twice the electrode to tissue spacing this will be a reasonable approximation. The impedance presented by each alternate pathway will be sufficiently high to make these currents negligible. In figure 7.15(b) we ignore current flow through the bone, directly around the fatty tissue or through the air around the tissue. Again this is because these pathways have very high impedance compared to the ones shown. With these approximations the electrical equivalent circuits in 5.16(a) and (b) are the to table of contents SHORTWAVE DIATHERMY 187 same. Each of the electrode/tissue arrangements are equivalent to a series combination of impedances. Just as with resistive circuits (described in chapter 1), when impedances are connected in series the current in each impedance is the same. We thus have the following relationship: displacement current in air = displacement + real current in fatty tissue = displacement + real current in muscle As mentioned earlier, the proportion of real current in fatty tissue is approximately 50% while in muscle the proportion is about 80%. Thus the amount of real current flow in muscle is 80/50 or about one and one half times greater than in fatty tissue. Let us take the simple case where current spreading is minimal and estimate the relative rate of heating in fatty tissue and muscle. We need to know both the real current density and field strength in each tissue. The field strength is estimated below. The real current density in muscle may be increased or decreased if the electric field lines converge or diverge. This depends on the tissue/electrode geometry see figure 6.19 for example. When resistors are connected in series the current flow in each is the same but the voltage across each resistor will, in general, be different. The largest resistor will have across it the greatest potential difference. The equivalent statement for tissues of different impedance is as follows: When tissues are arranged in series the field intensity will be greatest in the tissue with highest impedance. Inspection of table 6.2 shows that muscle has a higher conductivity and dielectric constant than fatty tissue: both figures are several times higher. Now a high conductivity and dielectric constant means a low impedance. Combining the two to table of contents SHORTWAVE DIATHERMY 188 figures from table 6.2 we calculate that fatty tissue has an electrical impedance some ten times larger than muscle. The rate of heating of each tissue is given by equation 7.1: Pv = E.ir .... (7.1) The real current density in muscle is as we have seen, about one and a half times greater than in fatty tissue, however the field intensity in fatty tissue is approximately ten times higher. Hence the rate of heating of fatty tissue is predicted to be approximately 10/1.5 times higher than muscle. We thus have the general conclusion that if spreading or converging of the field is minimal the rate of heat production in fatty tissue will be about seven times higher than in muscle. If the electrode/tissue configuration permits spreading of the field in muscle the current density will be reduced and the rate of heating of muscle correspondingly reduced. Conversely if the geometry produces convergence of the field lines in muscle the current density will be increased and the relative rate of heating will be increased accordingly. Inductive Coil Treatment With capacitor field treatment tissues are effectively in a series electrical arrangement. The current flow in muscle is thus limited by the impedance of the fatty tissue layers. When inductive coil treatment is used such is not the case. Consider the coil and tissue arrangement shown in figure 7.7. Currents are induced in the plane of the fatty tissue and in the plane of the muscle. The current loops are complete electrical pathways in the one tissue. For this reason the current flow in to table of contents SHORTWAVE DIATHERMY 189 muscle is not limited by the fatty tissue but depends only on the strength of the induced electric field and the electrical characteristics of the muscle tissue. In other words the induced currents flowing in each tissue layer are independent of each other. The real component of the current density, the current density which determines heat production, is given by equation 6.10, which can be written: ir = σ.E .... (6.10) Substituting this formula into equation 7.1 we obtain an alternate expression for the power dissipated per unit volume: Pv = σ.E2 .... (7.2) Table 6.2 shows that the conductivity, σ, of muscle is some sixteen times greater than that of fatty tissue. Hence, for the same induced electric field strength, both the real current density and the power dissipated in muscle will be sixteen times greater than in fatty tissue. How large is the magnetically induced electric field? The intensity of the induced field is determined by the rate of change of the magnetic field and the permeability, µ, of the material. The permeability is close to one for biological materials (see table 6.3) so fatty tissue and muscle are alike in this regard. For the same strength of alternating magnetic field then, both fatty tissue and muscle will have the same strength of induced electric field. Thus the rate of heating of muscle in this situation will be about sixteen times greater than that of fatty tissue. The intensity of the induced electric field is determined by the rate of change of the magnetic field and the permeability, µ, of the material. It does not depend on the electrical properties, σ and ε, of the tissue. In practice such a degree of selective heating is difficult to achieve. This is for two reasons: to table of contents SHORTWAVE DIATHERMY 190 * Muscle is located beneath fatty tissue and so is further from the induction coil. Thus the magnetic field is weaker in muscle and the strength of the induced electric field is correspondingly smaller. * Fatty tissue, being closer to the induction coil may also experience an appreciable electric field due to the capacitance between adjacent turns of the coil. This effect was described earlier (see figure 7.8). These two factors combine to increase the heating of fatty tissue relative to muscle so that a sixteen to one advantage is rarely obtained. Nonetheless efficient selective heating is achieved with close spacing of the turns of the coil and a sufficiently large coil to patient distance. One would also expect good discrimination with applicators which incorporate an electric field screen in front of the inductive coil. For more information on relative heating rates see the book 'Therapeutic Heat and Cold' J F Lehmann (ed) (1982) chapters 6 and 10. HEAT AND TEMPERATURE RISE Having described the factors determining the rate of heating of tissue we now consider the relationship between rate of heating and rate of increase of temperature. The rate of heating per unit volume is given in terms of electric field intensity and real current density by equation 7.1. Hence the amount of heat produced per unit volume, ∆Qv, in a time interval ∆t is given by equation 7.3. ∆Qv = E.ir.∆t .... (7.3) ∆Qv has units of joules per cubic meter (J.m-3). In considering the therapeutic effects of diathermy it is not the heat produced, as such, which determines the physiological response but the resulting temperature rise. Temperature is a key factor in determining the rates of chemical reactions and hence to table of contents SHORTWAVE DIATHERMY 191 physiological processes. The SI unit of temperature is the kelvin (symbol K). It is related to the perhaps more familiar degree Celsius (oC) by the expression oC = K - 273.15 Notice that from this definition the size of the degree Celsius is the same as the kelvin. In other words a change in temperature of five degrees Celsius is precisely the same as a change of five Kelvin's. When we are talking about increases in temperature brought about by diathermy treatment the terms kelvin and degrees Celsius can be used interchangeably to describe the increase. To convert from degrees Celsius to Kelvin's, simply add 273.15 to the Celsius temperature. When a fixed amount of heat is supplied to different substances the increase in temperature of each will, in general, be quite different. The factor which determines the resulting temperature increase is the specific heat capacity of the substance. The specific heat capacity is defined as the amount of heat required to raise 1 kg of a substance through one kelvin. The units of specific heat capacity are thus joules per kilogram per kelvin. This can be measured experimentally by supplying a certain amount of heat (∆Q) to a known mass (m) of the substance an measuring the resulting temperature increase (∆T). The experiment must be arranged so that all of the heat supplied is used to increase the temperature of the substance. If the loss of heat is negligible then the specific heat capacity (c) can be calculated using equation 7.4: ∆Q c= .... (7.4) m.∆T When we consider the heating of tissues by diathermy, heat transfer between tissues and to the bloodstream will have a large effect on the temperature distribution during treatment. Alternatively, when the specific heat capacity of a substance is known, equation 7.4 can be used to calculate the temperature increase resulting from the heat supplied. to table of contents SHORTWAVE DIATHERMY 192 Prior to the start of treatment the body tissues are in a state of dynamic equilibrium. Cellular activity, metabolism and muscle contraction result in the steady production of heat and the circulation of blood and tissue fluids provide an efficient means of heat transfer. The net production of heat is balanced by net transfer of heat from the tissue and a stable temperature is maintained. Once treatment is started heat is produced in the tissue according to equation 7.3 and the temperature starts to increase. An expression for the initial rate of increase in temperature is obtained below. Rearranging 7.4 we have ∆Q = m.c.∆T Dividing this expression by volume we obtain: ∆Qv = ρ.C.∆T .... (7.5) where ρ is the mass per unit volume or density of the tissue. Dividing 7.5 by ∆t gives: ∆Qv ∆T = ρ.c. ∆t ∆t .... (7.6) where ∆Qv/∆t is the volume rate of heating (in Joules per cubic metre per second) and ∆T/∆t is the rate of increase in temperature (in Kelvin's per second). This equation can be used to compare the initial rate of temperature increase in fatty tissue with that of muscle. The densities of the two tissues are similar but the heat capacity of muscle is some 50% greater than that of fatty tissue. Thus if the rate of heating of each tissue is the same, the initial rate of temperature increase in muscle will be only two thirds of that of fatty tissue. To produce the same initial rate of increase in temperature in each tissue the rate at which heat energy is produced in muscle must be 50% greater. Note that this conclusion is a general one. It applies not just to shortwave diathermy but to any diathermic modality. to table of contents SHORTWAVE DIATHERMY 193 An equation specifically applicable to shortwave diathermy is obtained by solving equations 7.5 and 7.3. We then have: ρ.c.∆T = E.ir.∆t, which on rearranging gives: ∆T E.ir = ∆t ρc .... (7.7) Equation 7.7 shows that the initial rate of increase in temperature (∆T/∆t) in shortwave diathermy depends on four factors: * * * * E, the field intensity at the point ir, the magnitude of the real current density at the point ρ, the density of the tissue c, the specific heat capacity of the tissue Once the temperature of any tissue has increased appreciably two things happen: * The body's temperature regulation mechanism responds. Blood vessels dilate, circulation is increased and more heat is transferred from the tissue. * Heat is transferred by the blood and tissue fluids to adjacent cooler tissues. Both of these effects lower the rate of increase in temperature. Eventually, the stage is reached where the temperature ceases to increase. A new dynamic equilibrium is achieved where the net production of heat is once again balanced by the net transfer from the tissue. Figure 7.17 illustrates the temperature variation during treatment. There is a transient period during which the tissue temperature increases, followed by a steady state where a constant (elevated) temperature is produced. The transient period for tissue volumes of interest in physiotherapy is typically of the order of twenty to thirty minutes to table of contents SHORTWAVE DIATHERMY 194 (see Lehmann (1982), chapter 10). Thus for treatment times of up to several minutes, equation 7.7 gives a reasonable approximation to the real physical situation. Application of equations 7.1 and 7.7 to quantitative prediction of the rate of heating and rate of temperature increase in different parts of tissue is difficult. The difficulty arises in the calculation of the field intensity in a particular area. For a review of results obtained using various approximations see A. W. Guy in J F Lehmann (1982). In patient treatment, shortwave diathermy remains something of an art as well as a science. The physiotherapist must use a knowledge of anatomy together with knowledge of the electrical properties of tissues to determine the optimum placement of electrodes or coil to give the required field pattern. Once the field pattern is selected, the physiotherapist uses a knowledge of the relative heating of the tissues and the patient's report of a sensation of warmth to adjust the intensity of the applied field to an appropriate level. With this procedure it is not possible to accurately monitor dose or dose rate for the individual tissues. Since this is a problem common to all diathermic modalities we will defer further discussion of dosage until chapter eleven. Physiological Effects The therapeutic value of shortwave diathermy arises from the physiological response of tissues to an increase in temperature. A number of physiological responses are found: * Figure 7.16 A simple model for tissue temperature variation during treatment. at the cellular level an increase in temperature increases the rate of biochemical reactions. Thus cellular metabolism is increased - there is an increased demand for oxygen and nutrients and the output of waste products is increased. to table of contents SHORTWAVE DIATHERMY * blood supply is increased. A number of factors determine this response. The increased output of cellular waste products triggers dilation of the capillaries and arterioles. The temperature increase itself causes some dilation, mainly in the superficial tissues where heating is greatest. In addition, stimulation of sensory nerve endings (again mainly in the superficial tissues) can cause a reflex dilation. * a rise in temperature can induce relaxation of muscles. If there is abnormal muscle activity caused by pain, for example, repeated treatment with shortwave diathermy can be beneficial. The treatment helps to interrupt the vicious circle of pain producing muscle activity which in turn produces more pain and so on. A number of factors may contribute to relaxation: the direct effect of heat on muscle tissue, the removal of any accumulated metabolites due to increased circulation and the sedative effect of heat on sensory nerves. * the response of sensory nerves to heat is useful for the relief of pain generally. Mild heating appears to inhibit the transmission of sensory impulses via nerve fibres. In addition, when pain results from inflammation of tissue an increase in the rate of absorption of exudate with increase in temperature can result in a secondary pain-relief effect. 195 Some claims have been made that additional non-thermal effects can be produced under the conditions used for therapy. As yet there is no clinical evidence for these claims. Non-thermal effects seem to have been demonstrated using pulsed shortwave treatment when the peak power level is significantly higher than used for diathermy. The few published comparative studies indicate little or no nonthermal effect at the low continuous power levels of conventional shortwave field treatment. These points are considered further in chapter 8 following. to table of contents SHORTWAVE DIATHERMY 196 EXERCISES 1 Figure 7.1 shows a schematic diagram of shortwave diathermy apparatus. (a) Briefly explain how the apparatus produces a high-frequency alternating electric or magnetic field. (b) What is the function of inductors L1 and L2? Why is the capacitor in the patient tuning circuit a variable one? c) 2 3 4 5 (a) Why is it necessary to tune shortwave diathermy apparatus with the patient coupled to the machine? (b) What is the advantage of automatic versus manual tuning of shortwave diathermy apparatus? Figure 7.2 illustrates the response of ions, polar molecules and non-polar molecules to a high-frequency alternating electric field. (a) Briefly describe the movement of each kind of molecule in the field. (b) How is the movement related to heat production in a material? (c) Which kind of movement is associated with greatest heat production and which with least heat production? (a) What is meant by the terms 'real current' and 'displacement current'? (b) Consider the movement of ions, polar molecules and non-polar molecules in an alternating electric field. Describe the relationship between each kind of movement and real and displacement current. (a) Consider each of fatty tissue, muscle and bone in the shortwave diathermy field. Is current flow in each tissue best described as real or displacement current? to table of contents SHORTWAVE DIATHERMY (b) On the basis of your classification which tissue would be associated with maximum heat production? (c) Describe the complications (as far as prediction of heat production is concerned) caused by the presence of Iymphatic and blood vessels in fatty tissue. 6 Figure 7.4 shows current pathways in a model for an arm or leg. Describe the principal factors determining the relative rate of heating of each tissue layer. 7 A patient's lower limb is enclosed in a solenoidally wound coil as shown in figure 7.6. (a) Describe the motion of polar, non-polar and ionic molecules when a high frequency alternating current flows through the coil. (b) Indicate (with a diagram) the direction of movement of molecules in the limb. 8 If the solenoidally wound coil in question 7 was replaced by a pair of capacitor plates (one above the knee, one below the sole of the foot), what would be the new directions of molecular motion? Draw a diagram to illustrate. 9 Figure 7.8 shows the electric field associated with two adjacent turns of an induction coil 10 (a) what is the practical significance of this electric field in patient treatment? (b) how can the effects of this electric field be minimized? 197 For shortwave diathermy it is common practice to use electrodes which are somewhat larger than the structure to be treated (figure 7.9). Explain in terms of: (a) the field pattern produced to table of contents SHORTWAVE DIATHERMY (b) 198 the pattern of heating of tissue. What are the advantages and disadvantages of using unequal size electrode (figure 7.9(c) )? 11 (a) It is normal practice to space electrodes as far apart as possible (figure 7.10(b)) in shortwave diathermy treatment. Why is this the case? (b) What is the practical limitation on the electrode spacing which can be used? (c) Is there any advantage in positioning one electrode close to the patient's tissue as shown in figure 7.10(c)? 12 Consider the electrode arrangements shown in figure 7.11. Explain why the field intensity is non-uniform in diagrams (a) and (c). Under what circumstances will the field intensity be uniform, as in (b)? 13 (a) Draw a diagram showing a coplanar arrangement of electrodes over tissue and the resulting field pattern. (b) What are the advantages and disadvantages of coplanar electrode arrangements? (c) 14 What practical limit is there on the spacing of coplanar electrodes? Consider the hollow dielectric between capacitor plates which is shown in figure 7.14. (a) Explain where heat production is greatest and why. b) What technique can be used to produce more uniform heating of the dielectric? Explain. to table of contents SHORTWAVE DIATHERMY 15 16 17 18 199 When coplanar electrodes are used for patient treatment the tissues can be considered to be in series electrically (see figure 7.15(a)). (a) what approximations are implicit in this statement? (b) draw an electrical equivalent circuit similar to that in figure 7.15(a) for the situation where the electrodes are close together. (c) how would bringing the electrodes closer together affect the relative heating rate of muscle and fatty tissue? Justify your answer. (a) Explain why, in principle, it is easier to produce selective heating of muscle with an inductive coil rather than capacitor field electrodes. (b) what practical constraints limit the selective heating of muscle with an induction coil? (a) Explain the meaning of each of the terms in equation 7.6. (b) The initial rate of temperature increase in fatty tissue in an experiment is found to be double that of muscle. Assume that the densities of each tissue are the same and that muscle has a 50% greater heat capacity and calculate the relative rate of heating of the tissues. The relationship between heat production (∆Q) and current flow (I) in a conductor is given by Joule's law: ∆Q = V.I.∆t where V is the potential difference across the conductor and ∆t is the time interval for which current I flows. (a) Show how equation 7.3 can be obtained as an alternative form of Joule's law. (b) An electric field intensity of 100 V.m-1 in a conductor results in a current density of 50 A.m-2 . Use equation 7.2 to calculate the amount of heat produced in a 30 second time interval. You may assume that the current is entirely real. to table of contents SHORTWAVE DIATHERMY 19 (a) Explain the meaning of each of the terms in equation 7.7. (b) An electric field of intensity 200 V.m-1 in a material results in a real current density of 50 A.m-2. The mass density of the material is 900 kg.m-3 and its specific heat capacity is 4.0 kJ.kg-1.K-1. Calculate the initial rate of increase in temperature (∆T/∆t) of the material. 20 A block of conducting material is placed in an electric field. The field intensity in the material is 300 V.m-1 and the resulting real current density is 120 A.m-2. If the material has a density of 1000 kg.m-3 and it has a specific heat capacity of 3.8 kJ.kg -1 .K -1 , calculate the initial rate of increase in temperature of the material (using equation 7.7). 21 Equation 7.7 describes the initial rate of increase in temperature of tissue in shortwave diathermy treatment. Describe the physiological response to the initial temperature rise and the effect this has on the subsequent rate of increase of temperature (figure 7.16). 200 to table of contents