a non-invasive immobilization system and related quality assurance
Transcription
a non-invasive immobilization system and related quality assurance
Int. J. Radiation Oncology Biol. Phys., Vol. 43, No. 2, pp. 455– 467, 1999 Copyright © 1999 Elsevier Science Inc. Printed in the USA. All rights reserved 0360-3016/99/$–see front matter PII S0360-3016(98)00398-8 PHYSICS CONTRIBUTION A NON-INVASIVE IMMOBILIZATION SYSTEM AND RELATED QUALITY ASSURANCE FOR DYNAMIC INTENSITY MODULATED RADIATION THERAPY OF INTRACRANIAL AND HEAD AND NECK DISEASE JEN-SAN TSAI, PH.D.,* Mark J. ENGLER, PH.D.,* Marilyn N. LING, M.D.,* JULIAN K. WU, M.D.,† BRADLEY KRAMER, M.D.,* THOMAS DIPETRILLO, M.D.,* AND DAVID E. WAZER, M.D.* Departments of *Radiation Oncology and †Neurosurgery, New England Medical Center Hospital and Tufts University School of Medicine, 750 Washington Street, Boston, MA Purpose: To develop and implement a non-invasive immobilization system guided by a dedicated quality assurance (QA) program for dynamic intensity-modulated radiotherapy (IMRT) of intracranial and head and neck disease, with IMRT delivered using the NOMOS Corporation’s Peacock System and MIMiC collimator. Methods and Materials: Thermoplastic face masks are combined with cradle-shaped polyurethane foaming agents and a dedicated quality assurance program to create a customized headholder system (CHS). Plastic shrinkage was studied to understand its effect on immobilization. Fiducial points for computerized tomography (CT) are obtained by placing multiple dabs of barium paste on mask surfaces at intersections of laser projections used for patient positioning. Fiducial lines are drawn on the cradle along laser projections aligned with nasal surfaces. Lateral CT topograms are annotated with a crosshair indicating the origin of the treatment planning and delivery coordinate system, and with lines delineating the projections of superior-inferior field borders of the linear accelerator’s secondary collimators, or with those of the fully open MIMiC. Port films exposed with and without the MIMiC are compared to annotated topograms to measure positional variance (PV) in superiorinferior (SI), right-left (RL), and anterior posterior (AP) directions. MIMiC vane patterns superposed on port films are applied to verify planned patterns. A 12-patient study of PV was performed by analyzing positions of 10 anatomic points on repeat CT topograms, plotting histograms of PV, and determining average PV. Results and Discussion: A 1.5 6 0.3 mm SD shrinkage per 70 cm of thermoplastic was observed over 24 h. Average PV of 1.0 6 0.8, 1.2 6 1.1, and 1.3 6 0.8 mm were measured in SI, AP, and RL directions, respectively. Lateral port films exposed with and without the MIMiC showed PV of 0.2 6 1.3 and 0.8 6 2.2 mm in AP and SI directions. Vane patterns superimposed on port films consistently verified the planned patterns. Conclusion: The CHS provided adequately reproducible immobilization for dynamic IMRT, and may be applicable to decrease PV for other cranial and head and neck external beam radiation therapy. © 1999 Elsevier Science Inc. Non-invasive immobilization, Polyurethane cradle, Alpha cradle, Thermoplastic mask, Aquaplast, Conformal radiotherapy, Intensity-modulated radiotherapy, IMRT, Topograms, Quality assurance. The anatomic accuracy of radiotherapy depends on the ability of a positioning system to reproduce patient geometry among simulation, computerized tomography (CT) scanning, and treatment procedures. Geometric accuracy is especially critical for the radiotherapeutic modalities of intensity-modulated radiotherapy (IMRT) (1–7) and radiosurgery (SRS) (8 –13), where escalated or large single doses may be delivered. The setup accuracy of the NOMOS Corporation’s dynamic Peacock IMRT System (also called Corvus Version 1.12; NOMOS, The IMRT Corp., Sewick- ley, PA) is improved by invasive or non-invasive immobilization systems, with the former using screws surgically placed in the cranium (1– 4, 8 –16), and the latter, external devices (1, 6, 7, 17–26). Thus far, over 1500 patients with cranial or head and neck disease immobilized non-invasively have been treated with the Peacock system (1, 6). Rapid growth of these applications suggests detailed study of the effectiveness of non-invasive patient immobilization and positioning. Thermoplastics (Aquaplast; WFR/Aquaplast Corp., Wyckoff, NJ) (23) and polyurethane cradles (Smithers Medical Products, Inc., Akron, OH) are widely used as non-invasive Reprint requests to: Jen-Sai Tsai, Ph.D., Department of Radiation Oncology #246, New England Medical Center, 750 Washington Street, Boston, MA 02111. E-mail: [email protected] Presented at the 38th Annual Meeting of the American Association of Physicists in Medicine, Philadelphia, PA, July 22–26, 1996. Current address of M.N.L.: Department of Radiation Oncology, Washington University, Seattle, WA 98136. Acknowledgments—The authors acknowledge extensive support from the NOMOS Corporation. We thank Richard Behrman, Ph.D., for useful discussions about CT imaging. Accepted for publication 13 September 1998. INTRODUCTION 455 456 I. J. Radiation Oncology ● Biology ● Physics positioning devices. To increase immobilization reproducibility among simulator, CT, and treatment procedures, these two devices are applied here simultaneously to create a “customized headholder” (CH) (Fig. 1a). The use of a CH has already been described with conventional external beam radiation (25, 26). In our clinic, it has been observed that the face mask shrinks after fabrication, and is flexible during applications. Hence, mask shrinkage and flexibility is studied in an attempt to minimize its effect on immobilization reproducibility. Studies of immobilization systems and positional reproducibility are affected by immobility of the patient’s anatomy of interest (AOI), ability to locate AOIs accurately, and adaptability of the immobilization technique to a specific radiotherapeutic modality. In this investigation, the CH is augmented and tested with dedicated quality assurance (QA) to create a CH system (CHS) in an attempt to optimize positional reproducibility when delivering dynamic IMRT. METHODS AND MATERIALS Coordinate and fiducial systems Coordinate and fiducial systems are generally applied to implement non-invasive immobilization with the IMRT system, and specifically, for a 12-patient study of immobilization reproducibility. The treatment planning and delivery coordinate system axes are denoted by (X, Y, Z), and ideally coincide with the RL (right-left), AP (anterior-posterior), and SI (superior-inferior) anatomic directions. Any of the three directions are denoted by “r.” Thus, coordinate pairs of points on two-dimensional (2D) simulator radiographs, CT scans, CT topograms (or “scout” images), and megavoltage port films in orthogonal transverse, coronal, and lateral planes are specified as (AP, RL), (RL, SI), and (AP, SI), respectively. The linear accelerator (linac) gantry axis (G), ideally identical to the Z axis of the Peacock system, is also ideally coincident with the SI axis, and intercepts the treatment planning and delivery coordinate system origin (O). Anatomic fiducial points applied in the CHS include anterior (A), right and left lateral (R and L), cranial vertex (V), inferior edge of the nasal septum (IENS), and posterior edges of the tragi (ET). Primary fiducial lines and planar curves include the transverse fiducial line (TFL) in a transverse plane (“tra” view), sagittal (SFL) in a sagittal plane (“lat” view), and coronal (CFL) in a coronal plane (“cor” or AP view). Secondary fiducial curves include the sagittal midline of the nasal surface (SMNS), the mid-curve of the eyebrows (EB), the abutment of the eyelids (EL), the abutment of the lips (AL), and the external curve of the ears (EA). Second and third coordinate systems on the CH baseplate are dedicated to coronal and lateral CT topograms applied in a 12-patient study of patient immobilization reproducibility. Metrics for immobilization reproducibility A metric for immobilization, reproducibility is specified as positional variance (PV) of r, PV(r), with ranges signifying one standard deviation (SD). Radiographically deter- Volume 43, Number 2, 1999 mined average absolute values of PVs are denoted by ,Drcor. and ,Drlat., for example, where Dr signifies variance in (RL, SI), or (AP, SI) directions. Absolute values of PV are plotted in histograms and their average values ,Dr. are tabulated. Values Dr averaged over both CT (coronal and lateral) topograms and port films are termed global positional variance Dgr to characterize the CHS. Detection of setup errors in topographic and port film studies To analyze PV and model patient setup, three translational degrees of freedom were recognized relative to the X, Y, and Z coordinate axes. Three rotational degrees of freedom were recognized relative to axes coincident with or parallel to the X, Y, and Z axes. Thus, a total of 6 degrees of freedom were extractable from the combination of a 2D transverse CT image with coronal and lateral topograms. Each image may be used to detect two translational and one rotational parameter. Thus, three orthogonal images, each containing three parameters, provide up to nine parameters for characterizing 6 degrees of freedom, and one of the images may be omitted. The transverse CT was omitted in the 12-patient study of PV. Although AP (coronal) port films were desired to capture possible PV in the RL dimension, it was impractical to insert cassettes directly under the patient. Anatomy surrounding the sagittal midplane would have been obscured in films exposed below the table by the steel spline supporting the table. A posterior span of up to 70° was omitted from treatment planning and delivery to avoid irradiating through the center spline. Nevertheless, rigorous QA is implemented prior to treatment to make sure of the coincidence of the SFL with the sagittal laser, thus minimizing possible PV in the RL direction. CT topograms and port films of an anthropomorphic phantom (Alderson Rando Phantom; Radiology Support Devices Inc., Long Beach, CA) were obtained to check the consistency of the 2D scale of the topogram and the magnification factor of the port film for their mutual PV verification. Dimensions on the topogram and demagnified dimensions on port films were compared with actual dimensions of phantom structure. Posterior polyurethane cradle The cradle is fabricated in the simulator room with polyurethane foaming agents and a 16.6 3 25.4 3 7.6 cm styrofoam box lacking anterior and inferior sides. The container is placed inside a plastic bag with the open end pointing in the patient’s inferior direction. The bag and container are pressed tightly into the U-shaped bay of the CH baseplate. Foaming agents are mixed and spread evenly along the inner surfaces of the bag inside and inferior to the styrofoam box. The patient’s head, with IENS approximately perpendicular to the table top, is lowered onto the bag into the container. After the foaming agents expand and harden, the cradle surface is trimmed to allow a U-frame containing softened thermoplastic to be drawn downward along the sides of the patient’s face. A 15 3 20 3 0.08 cm A non-invasive immobilization system ● J.-S. TSAI et al. Fig. 1. Alignment of the cradle. (a) Anthropomorphic phantom head resting on its cradle and before mask placement. (b) Transverse view of the SFL and laser light projection on the superior external surface of the cradle head holder aligned with the SMNS (left) and lateral view of the TFL and laser light projection on the lateral external surface of the cradle container aligned with the IENS (right). (c) Top view of the TFL and SFL fiducials and laser light projections aligned with relative to the cranium. (d) Transverse view of misalignment of the SFL with the SMNS, creating the variance DRLtra. (e) Lateral view of possible misalignment of the TFL with the SMNS, creating the variance DSIlat (left), and possibly undetected misalignment cause by cranial rotation in the sagittal plane (right). (f) Top view of correct alignment; rotation in the coronal plane is avoided by mechanical resistance. 457 458 I. J. Radiation Oncology ● Biology ● Physics piece of rubber gauze is placed on the cradle to compensate for anticipated mask shrinkage. The SFL is drawn on the cradle along the sagittal ceiling laser projection aligned with the SMNS. The table is adjusted so that the vertical side wall lasers are tangential to the IENS (Fig. 1b). The right and left sides of the TFL are drawn on the lateral cradle surfaces along the projections of the side wall lasers. If the dose distribution may be improved with the head in the flexed position, the baseplate is set so that the ETs are aligned to intercept the intersection points of the side wall laser projections. The three fiducial lines (TFL, CFL, and TFL) are applied to reposition the patient’s head at the original cradle fabrication position during CT and treatment procedures (Fig. 1c). Anterior thermoplastic face mask To reduce the beam attenuation by the baseplate, it was modified by cutting several parallel slots in the SI direction. The modified baseplate and the attached cradle are placed on the table top so that the room lasers match the TFL, CFL, and SFL on the cradle surface (Fig. 1c). The patient is positioned on the cradle such that IENS and SMNS are aligned with the TFL and SFL, respectively. For the flexed head setup, the table is adjusted so that the inferiormost edges of the ear lobes or posteriormost ET coincide with horizontal side wall lasers. The U-frame is held with two hands so that the sheet is coplanar to the table top and its geometric center is directly above the patient’s nose, and is then lowered over the patient’s face until reaching the baseplate, where it is fixed in position with four plastic clamps, to the extent of what a patient can reasonably tolerate. The thermoplastic is molded around the chin, the bridge of the nose, and the supra-orbital ridges. Despite the tightness of the mask, the patient’s cranium may move because of the deformation of soft tissue pressed against the mask. To minimize PV of the patient relative to the mask, underlying EB, EL, ML, and EA curves are traced on the outer surface of the mask with a fine-tipped felt pen, with the intention of matching the tracings to the anatomy during subsequent mask placements. After the mask hardens, a hole is cut around the nasal area. Sharp edges of the nose hole in the mask are taped to avoid skin abrasion. The nose hole is necessary for visualization of lasers along the IENS and SMNS. In addition, the hole allows for increased patient tolerance of the tightness of the mask. Mask shrinkage To study mask shrinkage, four routinely used thermoplastic mesh sheets, 23 3 30.5 cm in size, were heated in a 66°C water bath and stretched to 70 cm, an average stretched length in applications around the anterior and lateral head. Length was then measured as a function of time. In addition, three masks shaped to the Rando head were used to study shrinkage by measuring the height from the U-frame to the nasal tip of the mask as a function of time. Mask shrinkage after fabrication for patients is compen- Volume 43, Number 2, 1999 sated for by insertion of the rubber gauze during fabrication and omission of the gauze during CT and treatments. To determine the effect on PV of neglecting mask shrinkage, some topograms were obtained with the same gauze used during mask fabrication left in the cradle during CT. QA procedures common to the simulator, CT scanner, and linear accelerator Stability of lasers and the isocenter of the treatment machine are critical to the integrity of patient positioning and treatment. Thus, laser alignments in the simulator, CT scanner, and IMRT linac are inspected regularly and adjusted when necessary. Additional QA helps minimize linac isocenter motion (1, 9, 28 –30). As part of our radiosurgery program, the location of the three major rotational axes of the linac are tested with a 3D mechanical pointer and a 3-axis micrometer for their mechanical stability (28) and with multiple film exposures of a target simulator (9, 12, 29 –31). Table tops for simulator, CT, and linac procedures are inspected with a level and adjusted when necessary. To maximize positional reproducibility of the patient’s head and the outer soft tissue anatomy relative to the mask, a best coincidence of the mask tracings with underlying anatomy is sought. If no tracings match their underlying anatomy, or more than half the tracings deviate by more than 3 mm from underlying anatomy, the mask is removed and reattached until the 3 mm criterion is met. To maximize positional reproducibility of the CH among CT and treatments, , 1 mm tolerance is allowed for coincidence of the TFL, CFL, SFL, IENS, and SMNS fiducials with laser lines, and for coincidence of laser line intersections with fiducial points (A, R, L, V). Peacock IMRT gantry axis and isocenter placement The IMRT system has been described in detail elsewhere (1–7). However, certain features of the CH were developed to overcome limitations of isocenter placement during treatment planning. The isocenter is set anatomically in a software module termed “Image Registration.” For non-invasively immobilized patients, the isocenter is digitized on a CT image manually without tools to display the relationship of the isocenter coordinates to external fiducial points needed for patient setup in CT and treatment. The isocenter is placed: (a) so as to prevent collision of the MIMiC with the table or patient, as the MIMiC surface proximal to the patient is only 36 cm from the gantry axis; (b) at a point less than 10 cm, the MIMiC field half width, from the most distal target, meaning that the IMRT beam can reach and cover the most distal target; and (c) within the target to avoid suboptimal dose distributions that may arise from underdosing at matchplanes of adjacent arced fields (5). After image registration is approved, the user moves on to delineate AOIs in the “Anatomy” software module, to prescribe dose in the “Prescription” module, and to examine and approve plans in the “Display Results” module. If the isocenter placement allows collisions, or is suboptimal for treatment planning, it is necessary to revoke all of the A non-invasive immobilization system ● J.-S. TSAI et al. 459 information entered and reviewed in all of the software modules from Image Registration through Display Results, thereby wasting considerable time and effort (more recent software allows changing isocenter in “Prescription”). Thus, two sets of 3– 4 additional fiducial points are applied for CT scanning so that the above described software limitations are eliminated. These additional fiducials are specified as (R, L, V)i, where i 5 1, 2, 3, or 4. Fiducial points and planar curves are drawn on pieces of nylon tape affixed to the mask. Patient setup for CT scanning The patient’s head is adjusted in the cradle such that the SMNS is aligned with the SFL and the sagittal laser projections and simultaneously the IENS is tangential to both side wall lasers and is aligned with the TFL as it was during cradle fabrication. These adjustments are intended to insure that head position replicates the original setup in the simulator. Under such circumstances, a possible rotation of the head around the sagittal axis of the body is corrected most easily by detection in a transverse view (Fig. 1d). A possible translation of the head in the SI direction relative to its original position in the cradle is corrected most easily by detection in a lateral view (Fig. 1e). The only possible remaining misalignment is shown in Fig. 1d, and is subsequently detected by comparing a lateral topogram with a lateral port film. Because the cradle conforms to the posterior and lateral cranial surfaces, and the SFL is aligned with SMNS, PVs in the RL and SI directions are unlikely, as illustrated in Fig. 1f. Once the head is positioned in the cradle as in the original setup in the simulator, the mask is attached and the EB, EL, AL, and EA curves drawn on the mask are verified to match the corresponding anatomic structures. Prior to CT scanning, dabs of barium paste are placed on the mask fiducial points A and (R, L, V)i for subsequent placement of the origin O which automatically defines the axis G at X 5 Y 5 0 (Fig. 2a). To insure that at least one transverse CT image shows all fiducial points A and (R, L)i, about five scans, 1-mm thick, are obtained in the region of the transverse plane defined by points A and (R, L)i. Treatment planning The origin O is assigned along a line RLi at its intersection with a line dropped from point A and perpendicular to RLi. The origin satisfies the following criteria arising from system limitations described above: (a) Potential collision is avoided (Fig. 2b); (b) the target most distal from G is within the MIMiC field half width of 10 cm; and (c) G is within the target to avoid underdose at matchplanes of unopposed anterior spans of adjacent arced beam segments. Twelve-patient study of PV using CT topograms To study PV, 2–5 CT topograms were repeated for each of 12 patients over the course of 1 h, and of 4 patients over the course of more than 3 days for assessing the influence of Fig. 2. (a) Transverse, sagittal, and coronal views of the CH with fiducial points A and (R, L, V)i marked with barium sulfate paste for CT imaging. (b) Patient setup relative to Peacock IMRT hardware, highlighting the IMRT field projected through the gantry rotational axis (G), the gantry’s range of rotation, and possible collision between the patient and the MIMiC assembly. prolonged mask shrinkage. Anatomic points Qi and Qi9, where i 5 1, 2, . . . , 5, are digitized on topograms. For each patient, the internal points were selected as those with the most unique radiographic appearance within their anatomy. The origins of the topogram coordinates are points on the baseplate of the CH, and are designated as Q0 and Q09. For clarity of display in the figures, subscripts of points are displayed as normally sized next to their symbol Q or Q9. Internal anatomic fiducial points on the coronal topogram are in the sphenoparietal suture (Q1), frontozygomatic suture (Q2), supra-orbital margin (Q3), and zygomaticomaxillary sutures (Q4, Q5) as shown in Fig. 3a. Fiducial points on the lateral topogram are in the frontal sinus (Q19), anterior sella (Q29), anterior nasal spine (Q39), coronoid process of the mandible (Q49), and mastoid process (Q59) (Fig. 3b). These points were chosen for ease and uniqueness of topographic identification. In the figures, subscripts of points are displayed next to their symbol for clarity. When points were not uniquely identified on a topogram, they were not used in the statistical analysis of PV. Coordinate pairs (RL, SI) and (AP, SI) of points Q0, Qi, Q09, and Qi9 from coronal and lateral topograms were dig- 460 I. J. Radiation Oncology ● Biology ● Physics Volume 43, Number 2, 1999 Dri, j, cor 5 max ?Qi, j, cor(r) 2 Qi, j, cor(r)?, and (1a) Dri, j, lat 5 max ?Qi, j, lat9(r) 2 Qi, j, lat9(r)?, (1b) Variances were determined from topograms obtained in , 1 h and . 3 days of CT scan interval, and were plotted in histograms. Five-point variances were sought per topogram set per patient. The variance measurement was slightly increased by a reading and digitization uncertainty of 0.5 mm inherent to CT pixel size. Average variance was specified as: ,Drcor. 5 O O f(r ) Dri, j, cor / O O f(r ) Dri, j, lat / i, j, cor i and ,Drlat. 5 j i i, j, lat i j O O f(r i, j, cor ) (2a) ), (2b) j O O f(r i, j, lat i j where f(ri, j) is the frequency of variance Dri, j. Variance was also plotted in histograms. Fig. 3. (a) Coronal topogram, coordinate system Q0, and five anatomic points, Q1–Q5. (b) Lateral topogram with points Q09– Q59. itized at the CT console. Dri, j was defined as the maximum variance of either coordinate of point Qi, or Qi9 among repeat topograms of patient j: Verification of treatment setup by comparing the IMRT port film to the CT topogram Lateral port film verification prior to IMRT is obtained to monitor the consistency between CT and IMRT setup. Two methods of obtaining port films were evaluated. In both, the gantry is set horizontally at an angle of 90° or 270° and the table, at a treatment position z. With the first method, a port film is exposed to 3 MU of radiation with the MIMiC fully open to a field size of 20 3 3.4 cm. Without disturbing the patient setup, the MIMiC assembly is removed and the film is exposed to an additional 3 MU with a field size of 30 3 30 cm. In the event that the craniocaudal edges of the 20 3 3.4 cm strip fail to reveal unique bony anatomy near the target, the same port film is exposed to a third exposure with the secondary collimators set to 30 cm wide 3 independent jaw 5 or 10 cm long. Removing the MIMiC posed a risk to the patient position and was thought to be overly time consuming for routine use. Therefore, a second method is employed using three adjacent 20 3 3.4 cm MIMiC fields. The resulting film is then compared to the lateral topogram (Fig. 4). A special cassette holder was mounted on the side rails of the table so that the port film could move with the patient and capture three strips of anatomy irradiated by fully open MIMiC fields. The cassette is fixed at a known distance (SFD), usually 127 cm, from the radiation source, projecting a 4.3 cm field length to the film. Of the three exposures, the middle one is taken at a central table position identical to that of a treatment to insure imaging a substantial portion of the target. The two neighboring exposures are made at table positions 4.3 cm from the central position, for an SFD of 127 cm. The 4.3 cm shift avoids overexposure by overlap of adjacent divergent field edges at the film. The total length of the three-exposure film is then back projected 10.2 cm to the G axis, enough to capture substantial AOI including the A non-invasive immobilization system ● J.-S. TSAI et al. 461 Fig. 5. Shrinkage of thermoplastic sheets as a function of time. Fig. 4. Port film obtained with three fully opened MIMiC fields matched at the film, where 4.3 cm is the port film table index interval, and 3.4 cm is the treatment table position interval: anatomy exposed at the gantry axis (upper left), coronal view of three diverging MIMiC beams (upper right), and reconstruction without unirradiated anatomy (lower left). target. Due to the desired matching of the three strip fields at the film cassette, two segments of about 0.9 cm of anatomy at the axis G between the adjacent diverging beams are missing from the image (Fig. 4). By accurately locating the edges of the three exposures relative to the CT coordinate origin, a verification of the IMRT setup is obtained. If a discrepancy of anatomy on the topogram and port film exceeds 2 mm, another port film is exposed and the setup is adjusted until the discrepancy is under 2 mm. To image and verify the MIMiC vane pattern at horizontal gantry positions, the MIMiC controller is programmed to deliver a narrow arc and open the MIMiC to a patientspecific vane pattern over a span of , 2° including the horizontal gantry position. The MIMiC remains shut during the first 6 –12° of rotation so that stability of gantry rotational speed, as monitored by clinometers on the controller, satisfies programmed criteria. RESULTS AND DISCUSSION Shrinkage of the mask Shrinkage of thermoplastic up to 95 h after initial elongation is shown in Fig. 5. Shrinkage of 1.5 6 0.3 mm SD in 1 day was determined from all four sheets. Shrinkage of 0.9 6 0.2 mm in height was observed in the masks fabricated for the Rando phantom. After the first day, a maximum of 0.5 mm additional shrinkage was observed in the ensuing 3 days, after which no shrinkage was observed. Because shrinkage is most significant during the first day, at least 1 day is required between mask fabrication and the treatment planning CT. Patient setup reproducibility from CT topograms Figures 6a and 6b show PV histograms taken from measurements of the coronal and lateral CT topograms of five patients immobilized with the same gauze as that used for cradle fabrication, implying that the mask shrank before the CT scans were obtained. Using Eq. 2a, data from Fig. 6a yield ,DSIcor. and ,DRLcor. of 1.2 6 1.0, and 1.6 6 1.2 mm, respectively. Similarly, from Eq. 2b and Fig. 6b, ,DSIlat. and ,DAPlat. are 1.4 6 1.2, and 1.3 6 1.3 mm respectively. Figures 6c and 6d show histograms from twelve patients’ coronal and lateral topograms scanned within 1 h, without gauze, and at least 1 day after mask fabrication with gauze, for example, with mask shrinkage offset by removal of the gauze. From Eq. 2a and Fig. 6c, ,DSIcor. and ,DRLcor. are 1.1 6 0.9 and 0.8 6 0.7 mm, respectively. From Eq. 2b and Fig. 6d, ,DSIlat. and ,DAPlat. are 1.2 6 1.3 and 1.2 6 1.4 mm, respectively. The agreement of the average discrepancy of ,DSIcor. of 1.1 6 0.9 mm for Fig. 6c with the average discrepancy of ,SIlat. of 1.2 6 1.3 mm for Fig. 6d suggests that the lateral port film is adequate at this time for IMRT setup verification without consuming additional time and resources on a coronal port film. The maximum deviation of 5 mm occurred with a patient having long hair, suggesting that positioning of the hair may be important in the use of the CH. Another source of error in this study is attributed to deficient visualization of internal anatomic points Qi and Qi9 in the attempt to identify the same point on successive topograms. Results with the CHS also indicate that the variance seen on the lateral topogram is slightly larger than that seen on the coronal topogram. This may be attributed to setting up the coronal alignment first, thus making the lateral alignment secondary. In other words, the coronal alignment accuracy may be viewed somewhat as an asymptote of the system’s overall experimental error, i.e. no worse than lateral or transverse view alignments. This served as another rationalization for omitting the coronal port film with negligible deterioration of the global QA. 462 I. J. Radiation Oncology ● Biology ● Physics Volume 43, Number 2, 1999 Fig. 6. Variance histograms: (a) ,DSIcor. and ,DRLcor. from coronal topograms obtained , 1 h from mask fabrication with gauze in place; (b) ,DSIlat. and ,DAPlat. from lateral topograms obtained , 1 h from mask fabrication with gauze in place; (c) ,DSIcor. and ,DRLcor. from coronal topograms obtained more than 1 day from mask fabrication; (d) ,DSIlat. and ,DAPlat. from lateral topograms obtained more than 1 day from mask fabrication. A non-invasive immobilization system ● J.-S. TSAI et al. 463 Fig. 7. (a) Lateral topogram with three lines at predetermined treatment table coordinates. (b) Port film exposed firstly with the fully open MIMiC field of 20 3 3.4 cm, secondly with the MIMiC removed and the field size slightly increased with an independent jaw, and thirdly with a field size of 30 3 30 cm. (c) Vane patterns and relative pencil beam intensities as a function of 10° bins of gantry rotation near the horizontal gantry position. Percent black shows the fraction of time during which the vane is open over a 10° span of gantry rotation (top); three adjacent port film exposures pieced together show anatomy relative to the edges (white lines) of adjacent fields; a vane pattern at the horizontal gantry position is superimposed on the central of the three exposures. (d) The CT topogram with pre-calculated location of the Z coordinates at the edges of anatomy in the image plane; two areas labeled with asterisks are expected missing image plane anatomy. (e) Lateral view of anatomy, with isodose lines showing the target. 464 I. J. Radiation Oncology ● Biology ● Physics Volume 43, Number 2, 1999 Lateral topograms obtained over an interval of 3 days for four patients showed ,DSIlat. of 1.0 6 0.6 mm, and ,DAPlat. of 1.7 6 1.0 mm. The former agrees with value 1.2 6 1.3 mm measured within one CT session, and the latter is slightly larger than the value 1.2 6 1.4 mm measured within one CT session. This difference, still within the reading and digitization error of 0.5 mm, and the agreement of ,DSIlat. between 1.0 6 0.6 mm over 3 days interval and (1.2 6 1.3, 1.4 6 1.2 mm) in one CT session support the QA protocol for mask shrinkage and patient setup, for example, mask shrinkage has little effect on the positional variance in the SI direction and therefore accounts for small DSIlat, but does affect AP variance due to changes of the mask vertical dimension manifested in DAPlat. In the PV study, CT topograms were used in place of simulator films because: (a) the IMRT treatment planning and delivery are based on the CT images; (b) CT topograms provide high quality images and thus a high degree of spatial resolution; and (c) CT topograms in computer storage are in numerical forms (coordinates), thus facilitating the analysis. Port film verification of patient anatomy imaged in the CT topogram Anatomic dimensions obtained from topograms and demagnified port films agreed with the actual dimensions of the anthropomorphic phantom head and neck within 6 1% SD. Thus, the combined uncertainty in comparing topograms with port films is estimated to be 6 1.4%. The lateral topogram with three delineator lines at coordinates pre-calculated for the sake of port filming is shown in Fig. 7a. The three exposure port film of the anatomy of Fig. 7a together with surrounding anatomy revealed by exposure of a larger field without the MIMiC on the linac, is shown in Fig. 7b. Figures 7c and 7d show three consecutive images with the MIMiC fully opened and port images pieced together. Two narrow strips of missing anatomy are indicated with asterisks in Fig. 7d, as shown schematically in Fig. 4. The location of specific anatomic structure from these lateral topograms and port films are compared with ruler measurements to the delineator lines and field edges. Figure 7c also reveals a radiation vane pattern at the anatomic location relative to the tumor shown in Fig. 7e. This example illustrates a vane pattern corresponding to the beam at a horizontal gantry angle with expected patterns shown in the insert of Fig. 7c. For some patients, horizontal vane pattern printouts may not be available. In these cases, the vane pattern is interpolated from printouts of patterns at gantry positions 6 5° from the horizontal gantry position. From Figs. 7c, 7d, and 7e, the patient’s setup between CT and IMRT and the vane pattern’s agreement with IMRT planned vane pattern, as well as its radiation coverage relative to the tumor, are simultaneously verified. Figure 8 shows the discrepancy histograms from a sample of 28 patients’ lateral port films compared with their lateral topograms in both the SI and AP directions. The SI and AP discrepancy histograms yield average discrepancies of ?0.2? Fig. 8. Histograms of the SI (top) and AP (bottom) variance of lateral port film relative to topogram. 6 1.3 , and ?0.8? 6 2.2 mm, respectively. Discrepancies larger than 3.5 mm may arise from negligence in alignment of the sagittal and side wall lasers with the IENS and SMNS prior to mask attachment to the patient’s face. Occasional discrepancies of 2.0 mm at the target are tolerated for fractionated IMRT, and rare discrepancies . 2.0 mm have been seen, in which case the setup is adjusted until a port film fulfilling the 2 mm criterion is obtained. For example, two patients were immobilized using face masks fabricated for conventional therapy in conjunction with a standard plastic head cup. No fiducial lines relative to the nasal septum and lasers were drawn on the plastic head cup; nor were eyebrows, eyelids, or mouth lip markings on the mask with the patient’s. As a result, discrepancies . 2 mm between the CT topogram and IMRT port film were observed. For these cases, new CHs were made, and patients were re-planned. Setup variance assessed from topograms and port films are given in Table 1. Craniocaudally, the overall discrepancy is 1.0 6 0.8 mm, based on the calculation sketched in Fig. 9, where ,DgSI. 5 1.0 6 0.8 mm SD is the global mean discrepancy in the SI direction, i.e., the sum of Eq. 2 A non-invasive immobilization system ● J.-S. TSAI et al. 465 Table 1. Positional variance in millimeter of the CHS determined from topograms 1/2 Duration Gauze* ,DSIlat. ,DAPlat. ,DSIcor. ,DRLcor. ,1 h ,1 h . 3 days Port films Average variance 1 2 2 N/A 1.4 6 1.2 1.2 6 1.3 1.0 6 0.6 0.2 6 1.3 1.3 6 1.3 1.2 6 1.4 1.7 6 1.0 0.8 6 2.2 1.2 6 1.0 1.1 6 0.9 – – 1.6 6 1.2 0.8 6 0.7 – – 1.0 6 1.2 1.3 6 1.5 1.2 6 1.0 1.2 6 1.2 *1Gauze: mask shrank in the intervening time prior to implementation. 2Gauze: masks’ shrinkage is offset during implementation. over all topograms. Globally averaged ,DgAP. and ,DgRL. were 1.2 6 1.1, and 1.3 6 0.8 mm, respectively. Analogous values reported from other facilities for noninvasive immobilization are 2.0 mm from Thornton et al. (32), 2.2 6 1.4 mm from Verhey et al. (23), and 1.02 mm from Sweeney et al. (24) using a special vacuum dental cast in conjunction with a hydraulic arm. Advantages and disadvantages of non-invasive relative to invasive immobilization To optimize immobilization for IMRT, a comparison between the invasive and non-invasive immobilization setups is useful. The non-invasive CHS has the following advantages over invasive skull screw immobilization: 1. Avoids the cost and inconvenience of surgical procedure. 2. Allows easy airway access and emergency release with less risk of injury. 3. Avoids possible medical complications, such as infection around head screws. 4. May be used for treatment at a number of anatomic sites. 5. Reduces patient discomfort. Disadvantages are: 1. Slightly more PV than PV of rigid headscrew immobilization (2 mm vs. 1 mm) (1). 2. The CHS assumes a constant patient geometry during the planning CT and over the course of therapy. The mask shrinkage (Figs. 5, 6a– d) study demonstrated at least 1 day was needed before the CT. In one case, because of facial swelling in response to high dose chemotherapy, the mask would not fit on a patient, and a new mask, CT, and treatment plan had to be obtained. Implementing the present non-invasive immobilization system in conjunction with our alignment QA procedures lessens the need for additional devices such as biteblocks (20, 23, 32), vacuum dental cast (24), or earplugs for reinforcing immobilization reproducibility. QA for placement of the gantry axis and origin in the treatment planning and delivery systems The results and consequence of QA in origin O assignment relative to the target during the IMRT treatment plan- Fig. 9. Illustration of the calculation of the mean discrepancy in craniocaudal (SI) direction from Table 1. Sample ranges of the absolute values of individual variances include negative values which are a byproduct of the estimated measurement error (top). ,DgSI. is the average of the absolute values of variance in the SI direction and sSI is one standard deviation (bottom). ning is shown in the Appendix. The use of the multiple barium dots (Fig. 2a) has allowed us to achieve optimal positioning of O/G relative to target. CONCLUSION A customized styrofoam/polyurethane cradle combined with a thermoplastic mask and guided by a carefully designed QA procedure, termed the customized headholder system (CHS), has been reliably implemented for noninvasive immobilization of dynamic IMRT of intracranial, head and neck, and esophageal targets. The present noninvasive immobilization assures safe, easy, and reproducible immobilization to within a standard deviation of 2 mm. The CHS need not be limited to dynamic IMRT, and may be applicable to other external beam radiation therapy. 466 I. J. Radiation Oncology ● Biology ● Physics Volume 43, Number 2, 1999 REFERENCES 1. Tsai J-S, Wazer DE, Ling MN, et al. Dosimetric verification of the dynamic intensity-modulated radiation therapy of 92 patients. Int J Radiat Oncol Biol Phys 1998; 40:1217–1230. 2. Carol MP. Conformal radiosurgery: Stereotactic surgery and radiosurgery. In: Purdy JA, Emami B, editors. 3D Radiation Treatment Planning and Conformal Therapy. Madison: Medical Physics Publishing; 1993. p. 249–266. 3. Carol MP. Integrated 3-D conformal multivane intensity modulation delivery system for radiotherapy. In: Hounsell AR, editor. Proceedings of the Eleventh International Conference on the Use of Computers in Radiation Therapy. Stockport, Chesire, UK: Handley Printers, Ltd.; 1994. p. 172–173. 4. Carol MP. Peacock: A system for planning and rotational delivery of intensity-modulated fields. Int J Imag Syst Tech 1995; 6:56–61. 5. Carol MP, Grant WH, Bleier AR, et al. The field-matching problem as it applies to the Peacock three-dimensional conformal system for intensity modulation. Int J Radiat Oncol Biol Phys 1996; 34:183–187. 6. Verellen D, Linthout N, Van Den Gerge D, et al. Initial experience with intensity-modulated conformal radiation therapy for treatment of the head and neck region. Int J Radiat Oncol Biol Phys 1997; 39:99–114. 7. Engler MJ, Tsai J-S, Vora S, et al. Application of dynamic intensity modulated radiation therapy to the treatment of prostate cancer. (Abstr.) Int J Radiat Oncol Biol Phys 1997; 39 (Suppl.):193. 8. Colombo F, Beneditti A, Pozza F, et al. External stereotactic irradiation by linear accelerator. Neurosurgery 1985; 16:154– 160. 9. Engler MJ, Curran BH, Tsai J-S, et al. Fine tuning of linear accelerator accessories for stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 1994; 28:1001–1008. 10. Hartmann GH, Schlegel W, Sturm V, et al. Cerebral radiation surgery using moving field irradiation at a linear accelerator facility. Int J Radiat Oncol Biol Phys 1985; 11:1185–1192. 11. Houdek PV, Schwade JG, Serago CF, et al. Computer controlled stereotactic radiotherapy system. Int J Radiat Oncol Biol Phys 1992; 22:175–180. 12. Lutz WR, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int J Radiat Oncol Biol Phys 1987; 14:373–381. 13. Podgorsak EB, Olivier A, Pla M, et al. Physical aspects of dynamic stereotactic radiosurgery. Appl Neurophysiol 1987; 50:263–268. 14. Gall K, Verhey L, Wagner M. Computer-assisted positioning of radiotherapy patients using implanted radiopaque fiducials. Med Phys 1993; 20:1153–1159. 15. Jones D, Christopherson DA, Washington JT, et al. A frameless method for stereotactic radiotherapy. Br J Radiol 1993; 66:1142–1150. 16. Podgorsak EB, Souhami L, Caron J-L, et al. A technique for fractioned stereotactic radiotherapy in the treatment of intracranial tumors. Int J Radiat Oncol Biol Phys 1993; 27:1225– 1230. 17. Delannes M, Daly NJ, Bonnet J, et al. Fractioned radiotherapy of small inoperable lesions of the brain using a noninvasive stereotactic frame. Int J Radiat Oncol Biol Phys 1991; 21: 749–755. 18. Gademann G, Schlegel W, Debus J, et al. Fractioned stereotactically guided radiotherapy of head and neck tumors: A report on clinically use of a new system in 195 cases. Radiother Oncol 1993; 29:205–213. 19. Gill SS, Thomas DGT, Warrington AP, et al. Relocatable frame for stereotactic external beam radiatherapy. Int J Radiat Oncol Biol Phys 1991; 20:599–603. 20. Hamilton RJ, Kuchnir FT, Pelizzari CA, et al. Repositioning accuracy of a noninvasive head fixation system for stereotactic radiotherapy. Med Phys 1996; 23:1909–1917. 21. Hariz MI, Henricksson R, Löfroth P-O, et al. A noninvasive method for fractionated stereotactic irradiation of brain tumors with linear accelerator. Radiother Oncol 1990; 17:57–72. 22. Kooy HM, Dunbar SF, Tarbell NJ, et al. Adaption and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 1994; 30:685–691. 23. Verhey LJ, Goitein M, McNulty P, et al. Precise positioning of patients for radiation therapy. Int J Radiat Oncol Biol Phys 1982; 8:289–294. 24. Sweeney R, Bale R, Vogele M, et al. Repositioning accuracy: Comparison of a noninvasive head holder with thermoplastic mask for fractionated radiotherapy and a case report. Int J Radiat Oncol Biol Phys 1998; 41:475–483. 25. Bentel GC, Marks LB, Sherouse GW, et al. A customized head and neck support system. Int J Radiat Oncol Biol Phys 1995; 32:245–248. 26. Bentel GC, Marks LB. Comparison of two head and neck immobilization systems. Int J Radiat Oncol Biol Phys 1997; 38:867–873. 27. Sherouse GW, Bourland JD, Reynold K, et al. Virtual simulation in the clinical setting: Some practical considerations. Int J Radiat Oncol Biol Phys 1990; 19:1059–1065. 28. Tsai J-S, Curran B, Sternick E, et al. The measurement of linear accelerator isocenter motion using a three micrometer device and an adjustable pointer. Int J Radiat Oncol Biol Phys 1996; 34:189–195. 29. Tsai J-S, Buck BA, Svensson GK, et al. Quality assurance in stereotactic radiosurgery using a standard linear accelerator. Int J Radiat Oncol Biol Phys 1996; 34:189–195. 30. Tsai J-S. Analyses of multi-irradiation film for system alignments in stereotactic radiotherapy (SRT) and radiosurgery (SRS). Phys Med Biol 1996; 41:1597–1620. 31. Drzymala RE, Klein EE, Simpson JR, et al. Assurance of high quality linac-based stereotactic radiosurgery. Int J Radiat Oncol Biol Phys 1994; 30:459–472. 32. Thornton AF, Ten Haken RK, Gerhardsson A, et al. Threedimensional motion analysis of an improved head immobilization system for simulation CT, MRI, and PET imaging. Radiother Oncol 1991; 20:224–228. 33. Rosenthal SJ, Gall KP, Jackson M, et al. A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy. Int J Radiat Oncol Biol Phys 1995; 33: 1239–1245. APPENDIX Influence of the origin location relative to the target to the quality of IMRT Figures A-1 and A-2 show one example of suboptimal origin and axis assignment. The collimator pattern on the left fails criterion (c) from the Peacock IMRT gantry axis and isocenter placement subsection in Methods and Materials above: here the gantry axis G and origin O are posterior to the anterior-most target so as to be out of reach of the MIMiC, that is, the IMRT field. As a result, the isodose distribution and dose volume histogram (DVH) are suboptimal (Fig. A-2). The DVH shows that to cover the entire target volume, an isodose surface of 78% needs to be prescribed, leading to maximum dose of 513 cGy. To im- A non-invasive immobilization system ● J.-S. TSAI et al. 467 Fig. A-1. One-mm thick transverse CT view of fiducial points A and L; by placing the gantry axis and origin using the posteriormost L3, the MIMiC is unable to irradiate the anteriormost target at all gantry angles. Fig. A-3. The gantry axis and origin at L3 in Fig. A-1, which is moved anteriorly to L1, 4 cm toward the target and within reach of the MIMiC field. prove the plan, G and O were moved 4 cm anteriorly, as shown in Fig. A-3, to the fiducial points (R, L, V)1. Resulting improvement to collimator pattern, dose distribution, and DVH are shown in Fig. A-4, where the 84% isodose line completely covers the whole target with a lower maximum dose of 476 cGy. Fig. A-2. Gantry axis and origin placement from Fig. A-1 results in suboptimal MIMiC patterns. Fifteen of the patterns including vanes at the two lateral extremes of the MIMiC appear truncated. Fig. A-4. Optimal dynamic IMRT collimator pattern, isodose distribution, and DVH obtained from gantry axis and origin placement of Fig. A-3.