Technician Training Manual - Mobile Hyperbaric Centers
Transcription
Technician Training Manual - Mobile Hyperbaric Centers
df Technician Training Manual This manual will serve as your primary tool for reference and orientation as a new Hyperbaric Technician with Mobile Hyperbaric Centers. Version 1.2 Updated June 2015 1 WELCOME Company Overview Introduction MHC Corporate Staff OPERATIONAL EQUIPMENT CONT’D 3 4 5 HYPERBARIC MEDICINE What is Hyperbarics? History of Hyperbaric Medicine Chamber Classification Indications Gas Laws Terms of Pressure Dive Terminology Treatment Protocols Contraindications Side Effects/Complications Governing Organizations 6 7 9 10 11 12 13 14 18 19 21 PROFESSIONALISM Job Description Administrative Responsibilities Employee Handbook 22 23-25 26 SAFETY Safety and Operations Manual Seizure Protocol Falls Risk Prevention Prohibited Items List Chamber Venting/Bounce Safety Director Responsibilities & Log 27 28 29 30 31 32 OPERATIONAL EQUIPMENT Mechanical Equipment Chamber Layout Hydraulic Lift Compressors ECU Chiller Oxygen Supply TV Enclosure K-Bottle FSS Control Panel Generator Entry Lock Med-Lock Maintenance Preventative Maintenance Troubleshooting Reporting a Maintenance Issue 33 33 34 35 36 36 37 37 38 39 39 40 40 41 47 48 Clinical Equipment Hoods Latex/ Nitrile Hood Creation Hood Trim Chart Hood Accessories Hood Troubleshooting Trachea Attachments Masks Mask Creation Mask Fitting Ancillary Equipment BP Cuff/ Temp/Glucometer Resupply and Ordering Corporate/Hospital 49 49 51 51 51 52 53 54 54 55 56 DAILY TREATMENT ROUTINE Typical Center Schedule Daily Treatment Flow Daily Patient Flow Daily Logs (Startup/Shutdown) ECW Electronic Charting Input of Vitals Merging Templates Check out a Patient Linking Referrals Changing Schedule Cancelling a Patient Pre-Treatment Duties Assessment/Vitals Transporting Patients During Treatment Chamber Operator Responsibilities Chamber Attendant Responsibilities 3rd Technician Responsibilities Treatment Routine Post-Treatment Duties Blood Glucose Level ECW Sign off by MD Hood Cleaning Infection Control Chamber Cleaning 57 58 59 60 61 62 63 64 65 66 67 68 69 69 70 70 71 72 72 72 73 EDUCATION Introductory Course to Hyperbaric Medicine CHT Certification Process Continuing Education (Recertification) Center Incentive 74 74 75 76 2 Welcome to Mobile Hyperbaric Centers! Our company’s mission is: “Safely Heal Patients and Have Fun Doing It”. We believe that this mission provides the foundation for establishing successful hyperbaric centers and embodies the principles behind improving healthcare delivery in the United States. Our company continues to set the bar in how to best heal patients with oxygen, and is actively involved in educating the country about Hyperbaric Medicine. This education is primarily on a local level but also includes establishing a national training program for healthcare providers and doing research to better establish the science of Hyperbaric Medicine. As we continue to establish ourselves as the world’s premier Hyperbaric Medicine providers, we always continue to recognize that our ultimate success is based on fulfilling our mission with each individual patient. Each employee of our company plays an integral and critical role in this success. Throughout the time you are at Mobile Hyperbaric Centers (and we hope that it will be a long time), there will be many opportunities for you to grow and be successful. Our hope is that you will seek out such opportunities, embrace them and work towards making your life a fulfilling one. While we recognize that we cannot make you have fun, because how you perceive what you do is only up to you, we will continue to strive towards making your days with us filled with challenges, friendly people and meaning. Welcome to our team, 3 Handbook Overview As a member of the Mobile Hyperbaric Centers Team, you must follow and support the mandatory policies and procedures set forth in this manual. Please keep this booklet for future reference. The Technician Training Manual contains important information about clinical policies and procedures pertinent to your employment with Mobile Hyperbaric Centers (MHC). This manual will serve as your primary resource for training and orientation as a new Hyperbaric Technician. It is to be used in conjunction with the MHC Safety and Operations Manual, and the Employee Handbook. Philosophy “Safely Healing Patients, and Having Fun Doing It!” In order to maintain an atmosphere in which Company goals can be accomplished, Mobile Hyperbaric Centers (MHC) provides a comfortable and progressive workplace. Most importantly, MHC encourages a workplace where communication is open and problems are discussed and resolved in a mutually respectful manner. As a member of the Mobile Hyperbaric Centers Team, you are required to perform your daily activities within applicable ethical and legal standards. These standards can be achieved and sustained only through the actions of all personnel. All individuals must maintain integrity in business conduct and avoid activities that could reflect adversely on the reputation of MHC, directors, or other employees. As professional healthcare providers, we must be professional, courteous, friendly, helpful and prompt in the attention given to patients. Clear communication with the patient and their family members when speaking of the pertinent details of their care is key to continually improving the quality of care delivered. MHC should meet or exceed patients’ expectations of Hyperbaric Oxygen Therapy. Patient care, safety, and satisfaction are major factors in the success of any program within the Mobile Hyperbaric Centers network. It is important that employees remain respectful, professional, and helpful when assisting and communicating with patients. For us to be Safe, we must always be vigilant in anticipating and avoiding things which may result in accidents. Checklists are a tried and true way to accomplish this for many routine things. For us to Heal patients, we must find ways to have our patients get their hyperbaric therapy every day. Every day they miss treatment (particularly if it is two days in a row); it negatively affects their healing. Just as importantly, our ability to heal the patient relies on us teaching them to be motivated in managing their own health. We need to empower them to gain control of their healing and motivate them to want to heal. We can do most of this simply by creating a relationship with them where they look forward to showing up each day. 4 Senior Leadership Team Ronald Gordon, M.D. (Co-President/CEO) [email protected] 216-485-2714 Charles Cowap, M.D. (Co-President/CEO) [email protected] 216-485-2705 Vince Ferrini, M.D. (Chief Medical Officer) [email protected] 440-221-3410 Neema Mayhugh, Chief Operating Officer [email protected] 216-443-0430 Steve Fischer, Corporate Compliance Manager [email protected] 440-759-6827 Mary Miller, Director of Finance/HR [email protected] 216-539-1019 Ron Shook, Manager of Patient Care Services [email protected] 216-485-2715 Ron Anselmo, Director of Facility Maintenance [email protected] 440-785-6146 Alex Hribar, Human Resources Manager [email protected] 216-443-0430 Mary Hujer, New Tech Trainer [email protected] 330-721-5127 5 What is Hyperbaric Medicine? According to the Undersea and Hyperbaric Medical Society(UHMS), the definition of hyperbaric oxygen therapy is a medical treatment in which a patient breathes 100% oxygen while inside a treatment chamber at a pressure greater than sea level (>1ATA). In contrast to attempts to force oxygen into tissues by topical applications at levels only slightly higher than atmospheric pressure, hyperbaric oxygen therapy involves the systemic delivery of oxygen at values 2-3 times greater than atmospheric pressure. How it Works? HYPEROXYGENATION provides immediate support to poorly perfused tissue in areas of compromised blood flow. The elevated pressure within the hyperbaric chamber results in a significant increase in the diffusing distance of oxygen to areas otherwise unreachable. NEOVASCULARIZATION represents an indirect and delayed response to hyperbaric oxygen exposure. Therapeutic effects include enhanced fibroblast division, new formation of collagen and capillary angiogenesis in areas of sluggish vascular supply such as late radiation damaged tissue, refractory osteomyelitis and chronic ulcers. Hyperoxia-enhanced ANTIMICROBIAL ACTIVITY has been demonstrated at a number of levels. Hyperbaric Oxygen causes toxic inhibition and toxin inactivation in certain illnesses such as Gas Gangrene. It also enhances phagocytosis and white blood cell oxidative killing. DIRECT PRESSURE utilizes the concept of Boyle’s Law to reduce the volume of intravascular or other free gas. For centuries, this mechanism has formed the basis for treatment of some emergent diagnoses such as Cerebral Arterial Gas Embolism (CAGE), and decompression sickness. Hyperoxia induced VASOCONSTRICTION is another important mechanism. It is helpful in managing intermediate compartment syndrome and other acute ischemia in injured extremities. Studies have also shown a significant reduction in interstitial edema in grafted tissues. 6 History of Hyperbaric Medicine In 1662, a British clergyman named Henshaw, without scientific basis, thought it would be a good idea to raise the ambient pressure around a patient for therapeutic purposes. He later built a "domicilium," (shown below) a sealed chamber that could either raise or lower pressure depending on adjustment of the valves. He reported that acute diseases of all kinds responded favorably to increased ambient pressure. In the 19th century, following up on Henshaw's concept, pneumatic institutes began to sprawl around the European continent. These large chambers often were able to accommodate more than one person and could sustain pressures of two or more atmospheres. These pneumatic institutes started to rival the popularity of mineral-water spas. It was not until 1879 that semi-scientific efforts were made in the field. Fontaine, a French surgeon, built a mobile operating room on wheels that could be pressurized. He performed over 20 surgeries in the unit using nitric oxide as the anesthetic. Compressed-air therapy was first introduced into the United States in 1871 by Dr. J.L. Corning. In the early 1900s, Dr. Orville Cunningham, a professor of anesthesia at the University of Kansas, noted that patients with heart disease and other circulatory disorders had difficulties acclimating to high altitudes. He postulated that increased atmospheric pressure would be beneficial for patients with heart disease. To test his hypothesis (1918), he placed a young resident physician suffering from the flu into a chamber used for animal studies. The physician was successfully oxygenated during his hypoxic crisis when compressed to 2 ATM. Cunningham, realizing that his concepts were sound, built an 88-foot-long chamber, 10 feet in diameter, in Kansas City (shown right) and began treating a multitude of diseases, most of them without scientific rationale. 7 History of Hyperbaric Medicine Shortly after, Henry Timken (of Timken roller bearing company) funded the construction of a new 64 foot “hyperbaric hotel.” This was the largest hyperbaric chamber ever built. Timkin agreed to build this chamber on his property after a close friend of his was successfully treated at Cunningham’s chamber in Kansas. The Cunningham Sanitarium was an institution that focused on clean-air breathing treatment. The steel ball contained bedrooms for patients, as well as a dining hall and recreation area. The American Medical Association (AMA) and the Cleveland Medical Society, failing to receive any scientific evidence for his rationale, forced him to close his facility in 1930. But scientific evidence did later become available. Cunningham tore the chamber down before the Nobel Prize for Medicine was awarded in 1931 to Otto Warburg, who noted the importance of oxygen in metabolism and the suggestion that higher levels of oxygen in tissues might be associated with improved health. The advent of the use of hyperbaric oxygen in modern clinical medicine began in 1955 with the work of Churchhill-Davis, who helped to attenuate the effects of radiation therapy in cancer patients using high-oxygen environments. That same year, Dr. Ite Boerma, a professor of surgery at the University of Amsterdam in Holland, proposed using hyperbaric oxygen (HBO) in cardiac surgery to help prolong a patient's tolerance to circulatory arrest. He conducted surgery under pressure, including surgical corrections of transposition of the great vessels, tetralogy of Fallot and pulmonic stenosis. In 1960, Dr. Boerma published a study on "life without blood." It involved exsanguinating pigs and removing their erythrocytes before exposing them to 3 ATM of HBO. These pigs were noted to have sufficient oxygen in the plasma to sustain life when they were given HBO at 3 ATA. 8 Chamber Classification Hyperbaric treatment chambers are divided into 3 categories; Classes A, B, and C. Class A hyperbaric chambers, also known as “multiplace” chambers, are generally the largest. The term multiplace stems from the Class A definition from the NFPA-99 stating any chamber for multiple human occupancy. Depending on chamber design, this can range from 2 patients at a time to as many as the design will allow(see right). Multiplace chambers are pressurized with compressed air versus oxygen. This creates a safer treatment environment for patients. Another NFPA-99 safety code requires multi-place chambers to maintain an environment below 23.5% of oxygen. The unique design of a multiplace chamber allows pressurization to a depth greater than that of a monoplace. Oxygen is administered to the patients through masks, or clear plastic hoods versus filling the entire chamber with oxygen. Multiplace chambers also allow for direct hands-on patient care by clinical staff during the entire treatment. as opposed to monoplace treatments where there is a delay for immediate hands on care. Class B hyperbaric chambers, also known as “monoplace” chambers, are typically referred to as single occupancy chambers. They are pressurized with 100% oxygen, as opposed to air , discussed above. Due to the high oxygen concentration found within a monoplace chamber, there are stricter safety concerns when treating patients in this environment. Another potential downside to the monoplace chamber is its relatively confined interior space. Chamber sizes have increased over time with current operating models ranging in diameter from 25”-41”. Monoplace chambers offer a more inexpensive method of treating patients. They also have less operating costs than multiplace chambers, but are unable to treat at depths greater than 3ATA. Class C hyperbaric chambers are classified for “NO HUMAN OCCUPANCY”. They are used primarily in veterinary medicine to treat animals with specific illnesses. 9 Indications Acute Carbon Monoxide Poisoning - increased COHB causes hypoxic stress Air or Gas Embolism -venous gas embolism (VGE) or arterial gas embolism (AGE) Chronic Osteomyleitis -infection of bone refractory to antibiotics and debridement Clostridial Myonecrosis (Gas Gangrene) -acute clostridial infection of the muscles (c. perfringens >95% cases) Crush Injury/Compartment Syndrome -acute traumatic injury resulting in inadequate perfusion and tissue hypoxia Decompression Sickness -formation of inert gas bubbles in tissues and/or blood Arterial Insufficiencies -Central Retinal Artery Occlusion -Enhancement of Healing in Problem Wounds Exceptional Blood Loss Anemia -patient refusal of blood, cross matching difficulties Necrotizing Fasciitis -acute, potentially fatal, infection of fascia and muscle Delayed Radiation Injury -bone or soft tissue complications from therapeutic radiation Osteoradionecrosis -mandibular necrosis from therapeutic radiation Thermal Burns -acute management: wound healing support -typically not covered by insurance Idiopathic Sudden Sensorineural Hearing Loss Compromised Grafts and Flaps -flap salvage and preservation in compromised tissues 10 Gas Laws Gas Laws play a key role in understanding how hyperbaric oxygen treatment works. While there are numerous gas laws, we will focus on the primary four, namely: Boyle’s Law, Charles’ Law, Gay-Lussac’s Law, and Henry’s Law. BOYLE’S LAW -Boyles Law illustrates the relationship between pressure and volume. It states that the pressure and volume of a gas are inversely proportional (shown right). A good example of Boyles Law is if you were to place a balloon in a hyperbaric chamber and INCREASE the pressure to 2 ATA, the balloon volume will DECREASE to half its normal size. HENRY’S LAW -Henry’s Law specifies that the partial pressure of a gas dissolved in a liquid is directly proportional to the pressure exerted on that gas. Therefore, increasing the partial pressure of oxygen results in more oxygen dissolved in the blood (shown right). A more common example of Henry’s Law is opening a can of soda for the first time. When you open the can of soda you are reducing the amount of partial pressure of nitrogen dissolved into the soda. Thus, causing the gas to escape which creates the “bubbling” effect typically seen. CHARLE’S LAW - Charles’s Law illustrates the relationship between volume and temperature. It states that a change in temperature will result a change in the volume of gas. For example, if you were to submerge a balloon into an extremely cool environment it will reduce in volume. GAY-LUSSAC’S LAW -Gay-Lussac’s law specifies that the pressure of a fixed amount of gas is directly proportional to its temperature. For example, at the beginning of a hyperbaric treatment (on descent) the increase in pressure will also cause an increase in temperature. 11 Terms of Pressure Understanding these terms of pressure that are commonly used in hyperbaric medicine will help you master key aspects of your job as a technician. The main use of these units of pressure is for interpretation of the gauges on the chamber control panel. All hyperbaric chamber gauges are unique and display different units of pressure. Another use could be to determine the total bottom time of a technician when referring to the decompression tables for Repeat Designation or “Repet”. Absoulte Atmospheres of Pressure (ATA) Expressed as Millimeters of Mercury (mmHg), Feet of Sea Water (FSW), Pounds Per Square Inch (PSI), and Atmospheres (atm) Gauge Pressure (atm) 0 1 2 3 4 Absolute Pressure (ATA) 1 2 3 4 5 mmHG 760 1520 2280 3040 3800 FSW 0 33 66 99 132 PSI 0 14.7 29.4 44.1 58.8 Air pressure can be specified in several ways; the most popular term used in scuba diving is "pounds per square inch" or "psi." At sea level, the pressure exerted by the atmosphere is 14.7 psi. "Per square inch" refers to the surface area subjected to the weight of the air above it; the units could just as well be "mmHg," "fsw" or "absolute atmospheres". 12 Dive Terminology Descent/Compression- is the beginning of the treatment characterized by the gradual increase in pressure. The air temperature during descent will increase. Ascent/Decompression- the end of the treatment. When the pressure is gradually being reduced to allow the dive to arrive back at the surface. The air temperature during ascent will decrease. Treatment Depth/ Pressure- the pressure prescribed by the physician. Decompression Stop- is a specified depth where the dive must remain for a specified length of time (stop time). This brief pause during the ascent is to allow any excess nitrogen accumulation to be eliminated. MHC’s decompression stop is typically 3 minutes but can be as long as 5 minutes. Descent Time- is the total elapsed time from the time the dive leaves the surface to the time it reaches the bottom. This portion of the treatment usually lasts between 10-15 minutes. Total Bottom Time (TBT)- is the total elapsed time from the time the dive leaves the surface to the time it leaves the bottom. This includes 90 minutes at the treatment pressure+ 10/15 minutes of descent time totaling 100-105 minutes. Total Decompression Time (TDT)- is the total elapsed time from the time the dive leaves the bottom to the time it arrives at the surface. This time is also frequently called the total ascent time. This portion of the treatment usually lasts a total of 15min, and includes the Decompression Stop. Total Time of Dive (TTD)- is the total elapsed time from the time the diver leaves the surface to the time it arrives back on the surface. Surface Interval (SI)- is the time a diver spends on the surface between dives. It begins as soon as the diver surfaces and ends as soon as he starts the next descent. TBT DT 90 min min0i n90 TTD SI TDT 13 Treatment Protocols (Treatment Table 9) Mobile Hyperbaric Center’s goal is to maintain a safe environment for technicians and physicians while offering the most appropriate treatments for the patients. MHC has established its own dive protocol based on the USN Decompression Tables (rev. 6). The standard dive schedule is 45/114 or 2.4 ATA for 90 minutes. This profile is accompanied by additional safety precautions including oxygen breathing for the last 20 minutes of the treatment and a 3 minute decompression stop at 1.6 ATA (20fsw). Certain patients will require an alternate dive protocol, including high risk seizure patients. This is left to the discretion of the supervising physician at each center. Radiation injuries are to be treated at 2.4 ATA at all times. All other diagnoses including diabetic ulcers, chronic osteomyelitis, and failed grafts can be treated at 2.0 ATA if there is a need to accommodate a specific patient. Again this will be left to the discretion of the supervising physician. Our standard protocol for attendants and patients is described below. KEY: Standard Path Air Break Oxygen in Use Standard MHC Dive Protocol - 10-15 minute compression (descent) of the chamber to depth of 2.4 ATA 85 minutes on oxygen with 5 minute air break at 45 minutes 3 minute decompression stop at 1.6 ATA (20 fsw) 10-15 minute decompression (ascent) of the chamber to 1.0 ATA (includes oxygen breathing) Diagram: Standard Dive Protocol ASCENT STEP 8 End at 1.0 ATA STEP 2 Descent STEP 1 Start at 1.0 ATA STEP 3 Reach “depth” 2.4 ATA (45fsw) STEP 4 45 Min Oxygen STEP 5 5 Min air-break STEP 6 40 Min Oxygen STEP 4 Technician breathes Oxygen for the last 20 min at depth. STEP 8 Deco-Stop 1.6 ATA for 3 min 14 Treatment Protocols (Seizure) Patients who have an increased risk of seizures are given an additional oxygen break as shown below. An additional option would be to also reduce the treatment pressure to 2.0 ATA to further minimize the risk of a seizure. This is less desirable, as it is potentially less beneficial for other patients in the treatment session. This decision is always left up to the discretion of the supervising physician. - 10-15 minute compression (descent) of the chamber to depth of 2.0 or 2.4 ATA - 30 minutes on oxygen then 5 minute oxygen break followed by - 30 minutes on oxygen then 5 minute oxygen break followed by - 20 minutes back on oxygen to complete the treatment - 3 minute decompression stop at 1.6 ATA (20fsw) - 10-15 minute decompression (ascent) of the chamber to 1.0 ATA (includes oxygen breathing) Diagram: Seizure-Risk Dive Protocol ASCENT STEP 9 End at 1.0 ATA STEP 2 Descent STEP 1 Start at 1.0 ATA STEP 3 Reach “depth” at 2.4/2.0 ATA STEP 4 30 Min Oxygen STEP 5 5 Min air-break STEP 6 30 Min Oxygen STEP 5 5 Min air-break STEP 5 20 Min Oxygen STEP 8 Deco-Stop at 1.6 ATA for 3 min STEP 7 Technician breathes Oxygen for the last 20 min at depth. **Any deviation from this protocol must be approved by the Chief Medical Officer in advance.** 15 Treatment Protocols (Repeat Dives) MHC has established a dive protocol for the instance where a technician might attend one (1) 90 minute treatment and half of a second treatment, at 2.4 ATA, in one day. This repeat dive protocol is based off of the USN Decompression Tables (Revision 6). An example would be if a technician attended DIV 1 or DIV 2 and a half of DIV 4. As seen on page 14, the standard dive schedule used is 45/114. In the event of a repeat dive, a new dive schedule must be followed thus increasing the length of the decompression stop. If the technician has no less than 3 hours and 30 minutes of surface time, the new dive schedule for the repeat dive is 45/140. The technicians will be rotated just prior to the normal air break to prevent either technician from exceeding 60 min of TBT. Both technicians will then breathe supplemental oxygen throughout the ascent (approx. 15 minutes), and during the decompression stop. The decompression stop will be increased to 5 minutes at 1.6 ATA (20fsw). Diagram: 45/140 Repeat Dive Protocol **Any deviation from this protocol must be approved by the Chief Medical Officer in advance.** 16 Treatment Protocols (No “Deco”) Mobile Hyperbaric Centers does not require a decompression stop for a dive at 2.0 ATA that has less than125 minutes of TBT. ASCENT STEP 1 Start at 1.0 ATA STEP 5 End at 1.0 ATA STEP 2 Descent STEP 3 Reach “depth” 2 ATA (33fsw) STEP 4 Technician breathes Oxygen for the last 20 min at depth. **ANY deviation from these protocols must be cleared through the Chief Medical Officer in advance. MHC recommends at least 24 hours between an airplane flight and the most recent dive. For example, if a technician or physician is in a dive from 8am-10am on Tuesday, they should not be in a flight any earlier then 10:00am on Wednesday. Any questions regarding this policy can be directed to the Chief Medical Officer. 17 Contraindications Contraindications for Hyperbaric Oxygen Therapy are divided into two categories, absolute and relative. Absolute contraindications are conditions in which there are no reasonable circumstance to undergo treatment and that the risk outweighs the potential benefit. Relative contraindications are conditions in which there are higher risks of complications, but these risks may be outweighed by substantial benefits. Contraindications of Hyperbaric Oxygen Therapy Expressed in 2 Categories, Absolute and Relative. Absolute Untreated Pneumothorax Pregnancy Relative Cis-Platinum® (chemo) URI and chronic sinusitis High Fever Emphysema/COPD with CO2 Retention History of a Spontaneous Pneumothorax History of Surgery for Otosclerosis Seizures Disorders History of Optic Neuritis Congenital Spherocytosis Thorax Surgery Pacemaker Claustrophobia Asthma Adriamycin® (chemo) Sulfamylon ® (burn ointment) Above: Pneumothorax 18 Complications and Side Effects Hyperbaric Oxygen Therapy is generally a safe procedure, and complications are rare. But, as with any medical procedure, it does carry some risk. Potential side effects are shown below. Middle Ear Barotrauma- any traumatic ear injury or pain caused by an inability to equalize ears during a change in atmospheric pressure. This is the most common side effect of hyperbaric oxygen therapy. All of the following conditions predispose patients to ear barotrauma; cleft palate disfigurement, head/neck radiation, allergies, upper respiratory infection and intubated/unconscious patients. Severity is calculated by using the TEED SCALE shown right. Also shown is the anatomy of the human ear . Prevention of ear barotrauma can be accomplished by educating the patient on good ear clearing techniques and the use of nasal decongestants. Treatment for damage to the middle ear includes a break from hyperbaric treatment. This can range from as little as a day up to one week. Sinus Barotrauma “Sinus Squeeze”- any traumatic injury or pain caused by an inability of the sinuses to equalize during a change in atmospheric pressure. This is most commonly caused by allergies, edema or mucous buildup from the common cold. Prevention of a sinus squeeze can be accomplished by educating the patient on good sinus clearing techniques and pretreating with nasal decongestants and steroids. Treatment for damage to the sinuses includes a break from hyperbaric treatment until symptoms resolve. Tooth Barotrauma “Barodontalgia”- is a rare form of tooth pain that occurs when a small amount of trapped air in a tooth expands on ascent. This is more likely to occur in patients with a new/loose filling or a recent root canal. Treatment for barodontalgia includes a break from hyperbaric treatment and to referral to a dentist. Claustrophobia/Anxiety- Hyperbaric therapy may cause some degree of confinement anxiety, even in a multiplace chamber. Occasionally, mild sedation is required for such individuals to continue to receive daily hyperbaric therapy. Treatment for claustrophobia includes a written prescription for an antianxiety medication by the Primary Care Physician (PCP). 19 Complications and Side Effects CNS Toxicity “ Oxygen Seizure”- is an extremely rare side effect of hyperbaric oxygen therapy and results from breathing high partial pressures of oxygen. It is characterized by visual and hearing abnormalities, unusual fatigue, muscle twitching, anxiety, confusion, incoordination, and convulsions. Some medications, high fever, and a history of a seizure disorder could make a patient more susceptible to an oxygen toxicity seizure (OTS). Prevention of an OTS includes identifying patients at risk and ensures they are on a seizure protocol during HBOT. Premedication with an anticonvulsant such as Ativan® has been shown to lessen the chance of an OTS in patients with an increased risk. Treatment for an OTS is to immediately remove the patient from oxygen. A more detailed outline of management of an OTS can be found in the MHC Safety and Operations Manual. Visual Changes- Progressive myopia has been observed in some patients undergoing prolonged periods of daily hyperbaric therapy. Although the exact mechanism remains obscure, it is apparently lenticular in origin and usually reverses completely within several weeks after the last therapy. Myopia is the most frequently reported side effect of hyperbaric therapy. It has been suggested that hyperbaric therapy causes the formation of cataracts. Although this has been disproven, in some cases hyperbaric therapy can cause enlargement of existing cataracts in patients who have received >100 treatments. Hypoglycemia- It is typical that a diabetic patient on glucose management will experience a significant drop in blood glucose levels (BGL( during a hyperbaric treatment. The typical drop is 50 mg/dl. Careful monitoring of all diabetic patients BGL is required to minimize the risk of a hypoglycemic reaction. Prevention of a hypoglycemic reaction includes encouraging all diabetic patients to follow a proper diabetic diet and eat a meal immediately before hyperbaric treatment. The patient can also be given juice to accompany them during treatment should signs/symptoms do arise. Initial Treatment for a hypoglycemic reaction is to quickly administer either oral glucose gel or tablets. A more detailed outline of the management of a hypoglycemic reaction can be found in the MHC Safety and Operations Manual. Pneumothorax(otherwise known as a collapsed lung)Is an abnormal collection of air or gas in the pleural space that separates the lung from the chest wall. This can spontaneously occur during hyperbaric treatment or secondary to positive pressure ventilations and bronchospasms, or holding your breath during pressure changes. Prevention of a pneumothorax could include a preliminary chest x-ray to rule out any areas of concern. Treatment for a pneumothorax can include a needle thoracostomy. The patient should be transported to the nearest emergency room for chest tube placement. 20 Governing Organizations The Undersea and Hyperbaric Medical Society (UHMS) is an international, non-profit organization that is the primary source of scientific information for diving and hyperbaric medicine physiology worldwide. More information regarding the UHMS can be found on their website at www.uhms.org The National Board of Diving and Hyperbaric Medical Technology (NBDHMT) is a nonprofit certifying organization designed to meet the clinical, technical and safety needs of the discipline of undersea and hyperbaric medicine. Opportunities exist for certification as a Diver Medic (DMT), as a Hyperbaric Technologist (CHT), as a Hyperbaric Nurse (CHRN) and as a Veterinary Hyperbaric Technologist (CHVT). More information regarding the NBDHMT and certification can be found on their website at www.nbdhmt.org. 21 Job Description The Hyperbaric Technician provides Hyperbaric Oxygen Therapy to patients referred to Mobile Hyperbaric Centers (MHC). Hyperbaric Technicians administer and monitor oxygen therapy as well as work collaboratively with all other disciplines assigned at the center, including hospital locations MHC is assigned to operate in. Position: Hyperbaric Technician Facility: Outpatient Multiplace Hyperbaric Department Administer and monitor hyperbaric oxygen therapy Prepare, educate, and orient patients about hyperbaric oxygen therapy Safely maintain, operate and attend to the hyperbaric chamber Accurately obtain and document patient assessment and vital signs Safely transport patients to/from hyperbaric chamber Accurately document patient information in Electronic Medical Record (EMR) system Clean hyperbaric chamber, clinical equipment and patient care areas with approved cleaning agents Maintain daily and periodic equipment maintenance documentation including service logs for the hyperbaric equipment Communicate any abnormal patient assessment findings with center’s physician Attend staff meetings and participate in MHC quality assurance programs and safety drills Maintain accurate hyperbaric chamber dive log Maintain annual competencies, training requirements, licensures, certifications and relevant documentation required by hospital, MHC and state/federal regulations Responsible for periodic education completion per hospital and MHC requirements, including but not limited to: annual competencies, training and on-site education workshops Understand and comply with infection control, HIPAA and Joint Commission regulatory standards per hospital, MHC and state/federal policies Assist Practice Manager/Coordinator with administrative duties when necessary Positively interact with fellow staff and patients in a friendly, professional and constructive manner Maintain professional boundaries with co-workers, patients and patient’s family members QUALIFICATIONS High school degree or equivalent Valid and current BLS certification Valid and current EMT licensure Good computer skills including working knowledge of Microsoft office suite; prior experience with an EMR system preferred Good verbal and written communication, customer service and time management skills Ability to work rotating weekend shifts as needed 22 Administrative Duties Although the Hyperbaric Technician’s role is primarily clinical, there are also some administrative duties required. The graph below shows some of the general administrative duties associated with a hyperbaric patient. It begins with receiving a referral and continues with sending follow-up letters weekly. Please discuss with your Practice Manager what your responsibilities will be at your center. Referral Process- is the initial communication from another Physician requesting that we evaluate their patient for hyperbaric therapy. If a Physician’s office calls to send us a patient follow the steps below: 1. Transfer Call to Practice Manager OR 2. Locate MHC Referral Form. Ask the Caller for the Patient’s Name, DOB, Patient Phone #, Referring Provider, Their Fax #, and insurance information (if available) 3. Inform the Caller you will be faxing the Referral Form and they need to complete the remaining fields, have the MD sign it, and return it with the patient’s last clinic note. Patient Transportation- Sometimes under unique circumstances it is necessary for MHC to provide transportation to patients who are unable to travel to treatment. Check with your Practice Manager to understand your responsibilities with transportation at your center. Referral Follow- Up Letters H&P Hyperbaric Patient Patient Transportation Patient’s 1st Day 23 Administrative Duties Follow-Up Letters- are created weekly by the Hyperbaric Physician to send out to referring providers. These letters provide an informative narrative of their patients’ progress at hyperbaric therapy. You could be asked to assist in the preparation and delivery of these letters. Patient Wound Photographs- are photos taken on a weekly basis by the Hyperbaric Physician to measure the patient’s progress and wound healing. It is possible you could be asked to obtain a patient photo during normal business hours. The proper technique for capturing an accurate photo will be shown to you during your initial training. Please refer to the steps below when capturing a patient wound photo. 1. Refer back to the patient’s prior photo to maintain accuracy and consistency 2. Locate Camera and Ruler (shown below) 3. Create a label and write in patient initials, date, and wound # and then place onto ruler 4. Put on gloves 5. Place ruler square to desired wound. Make sure to get Entire Wound and White Scale Box in the photo! 6. Focus camera, capture image 7. Upload desired image into patient’s folder and subsequently map with 4-D Imaging Software Patient Photo Camera and Ruler 24 Administrative Duties History and Physical- is the patient’s initial evaluation by the Hyperbaric Physician. During this visit it will be determined if the patient is a candidate for Hyperbaric Therapy. This initial visit or History and Physical (H&P), contains key steps to be completed by all of the hyerpabric staff. These steps are outlined using a combination of forms now called the “H&P Packet.” Please refer to the employee intranet for the latest packet. All center employees are meant to participate in this process. 25 Employee Handbook The Employee Handbook contains important information about policies affecting employment with Mobile Hyperbaric Centers (MHC). It describes many employee responsibilities, both as a healthcare professional, and MHC employee by outlining programs developed by the company to benefit employees. One of Mobile Hyperbaric Centers’ major objectives is to provide a work environment that is favorable to both personal and professional growth. 26 Safety and Operations Manual Mobile Hyperbaric Centers’ top priority is to assure our facilities are safe for both patients and employees. This manual focuses on safety policies and procedures pertinent to your employment with MHC. By following the regulations and policies set forth in this manual, all areas of our facilities are limited in risk. Please take time to familiarize yourself with this manual, specifically on key areas affecting day to day operations. These areas include; patient and facility emergencies, chamber venting procedures, and the prohibited items list. A copy of the Safety and Operations Manual must be keep at that control panel in front of the Chamber Operator at all times. 27 Safety (Seizure Protocol) Some of our patients are at an increased risk for experiencing an oxygen toxicity seizure. For these patients, MHC has developed a unique treatment profile to further minimize the occurrence. The Hyperbaric Physician will identify these patients during the H&P and distinguish this patient by checking “Seizure Protocol” on the Physician Checklist Form. This information can also be verbally expressed to the clinical staff as a secondary means of communication. It should be routine for all the technicians to familiarize themselves with any unique circumstances defined by the Hyperbaric Physician on the Physician Checklist Form PRIOR to the patient beginning treatment. If a patient is identified as being on a seizure protocol they need to be identified 2 ways: 1. Red dot is placed on the patient’s plastic rim of the hood (next to name label). This informs the chamber attendant which patients are on a seizure protocol. 2. A combined list of all patients on a seizure protocol is posted in the control panel area in plain view. Be sure to use INITIALS ONLY! (shown below) This will inform the chamber operator which patients are on a seizure protocol. 28 Safety (Falls Prevention) All patients in our care at Mobile Hyperbaric Centers will have a Fall Risk Assessment completed by the Hyperbaric Physician. The assessment will be a part of the patient’s initial history and physical. Any patient noted to be at a risk for falls will be identified. MHC identifies patients by giving them a RED silicone bracelet on their first day of treatment, and by keeping a patient list in PLAIN VIEW in the control panel area (shown below). The patient is informed this bracelet is to be worn at all times during their course of treatment, no exceptions. It is the responsibility of all staff to ensure every patient at risk for a fall is identified daily. It might be necessary to consult with your hospital on the unique identifiers utilized to remain compliant with JOINT COMMISSION standards. Falls Risk Wristband 29 Safety (Prohibited Items) Mobile Hyperbaric Centers’ top priority is to assure our facilities are safe for both patients and employees. Furthermore, never permit entry of any item into the chamber that may contaminate the environment. Open or vent any sealed containers to eliminate the risk of implosion/explosion. Items that may be subject to collapse or rupture include, but are not limited to, the following: water bottles, endotracheal tube cuffs, and Foley bags. *Note: Endotracheal tube cuffs and catheter bulbs should be filled only with normal saline, do not use air. Foley bags must be emptied prior to treatment. Certain items can result in a fire or explosion of the Hyperbaric Chamber, causing serious injury or death. Being familiar with items not allowed in the chamber is extremely important! Prohibited items within the chamber include (but not limited to): Hearing Aids/Dentures Shoes/Socks Jewelry Watches, Keys Cell Phones/Pagers Perfume/Cologne Lotions/Oils Cigarettes/Lighters Batteries Electronics Coins/Money Hair Spray/Gel/Mousse Nail Polish Make-up Medications Newspapers/Magazines Metal Objects Matches Wool/Nylon/Silk/Satin 30 Safety (Chamber Venting) In the event where the hyperbaric chamber reaches an elevated percentage of oxygen, it is necessary to “vent” or “flush” the chamber. This will allow the air to circulate and return the environment to safe operating parameters (<23.5%). Venting is mandatory to maintain a safe environment for the patients and staff during a Hyperbaric Treatment. Please follow the steps below to safely ventilate the hyperbaric chamber. Chamber Operator 1. Identifies leak by visualizing the oxygen sensor on the control panel (anything above 21 %) *Note: The oxygen sensor will not alarm until levels reach 23%. It is ideal to identify and correct a leak before the alarm sounds to minimize the potential for levels to become unmanageable. 2. Verbally inform Chamber Attendant of elevated oxygen levels and instruct them to inspect for a leak. Chamber Attendant must begin physically inspecting each patient (see steps below). 3. Monitor oxygen levels for changes 4. Vent chamber if oxygen level > 23 % 5. Venting protocol: (1) Vent the chamber for a period of 10 minutes (or less if the atmosphere returns to 21%). To do so, open the press valve and primary exhaust simultaneously. Make sure to maintain adequate pressure in the chamber. If the Chamber has not reached an oxygen level less than 23.5% in 10 minutes, turn off oxygen, and continue venting. 6. If you are unable to correct the problem inform the HBOT physician 7. Ascend the chamber if oxygen level > 28% 8. Document the chamber was vented and duration in the Dive Log under the notes section Chamber Attendant 1. Inspect for a leak by inspecting each patient for trouble areas. Trouble areas include: a. Neck ring (look for micro-tears or “bunching” and bad positioning of the patient) b. Caps (check to ensure both side caps are firmly in place) c. Hose (inspect for holes, and ensure it is firmly attached to neck ring) d. Helmet (inspect helmet for holes especially around the seal to the neckring) e. Flow Rate (Ensure flow rate is properly set for the patient needs. The typical range is 20-30 lpm) If the flow is set to high this can cause excess oxygen to leak out the bottom of the patient’s neck ring. f. As a last resort, activate the Snooper (use when necessary) 2. Inform Chamber Operator if the source has been identified and problem has been corrected 3. Maintain calm environment in chamber 31 Safety Director Course As a full time technician, you are required to participate in the Mobile Hyperbaric Centers Rotating Safety Director Program. The program is designed to allow every technician to act as the Safety Director on a monthly basis. The Safety Director Course can be found on the company Intranet and should be completed on the first day of employment. Successful completion is measured by passing the Safety Director Exam with an 80%. Pathway to access the Safety Director Exam -Intranet/Education/New Full Time Requirements 32 Operational Equipment (Mechanical) Hypertec HBO Chamber: The hyperbaric oxygen chamber is fabricated per ASME-PVHO regulations and standards. It is designed to treat as many as eleven (11) patients at one time with a technician present. The maximum pressure rating is 3 ATA or 44.1 PSI. The unit is configured with two compartments. The Main Chamber as shown in diagram above is 84” in diameter x 22’ in length. The second chamber is referred to as the “Entry Lock Chamber” which is 84” in diameter x 5’ in length. The Entry Lock is utilized periodically to transfer patients or medical personnel in or out of the main chamber in order to avoid interruption of the dive cycle. The chamber is also equipped with a “Medical Lock” which is a circular passage. It allows for quick transfer of small items in/out of the chamber. Wheelchair Lift: Description: Each unit is provided with a wheelchair lift which is utilized to raise patients from ground level to the level of the chamber control room. Each location has slight variations with facility connector and therefore loading and operations should be reviewed per location. a) Braun UVL Series The UVL or Under Vehicle Lift is cassette type mount in the side of trailer which is deployed as needed for patient loading. The unit is operated hydraulically and has a maximum capacity of 600 pounds. The usable platform size is 30” in width x 43” in length. Many safety locks and barrier features are integrated into the unit and should be thoroughly understood by the operator. Please review the manufacturer’s operation manual for complete instructions. Braun UVL Lift b) Ultron MDC 4048 The Ultron MDC 4048 is a rail style platform lift which is lowered from the exterior side of the trailer. The unit is powered by a combination of hydraulic cylinders which activate a chain and sprocket system. The usable platform size is 40” wide x 48” in length and load capacity of 2,000 pounds. Many safety locks and barrier features are integrated into the unit and should be thoroughly understood by the operator. Be sure to review the manufacturer’s operation manual for complete instructions. 33 Operational Equipment Air Compressor Package: Description: The compressor system is the primary piece of equipment for the mobile hyperbaric operation. The system is designed to meet federal standards for oil free medical grade air. The compressors are located in the very rear compartment of the trailer. The compressor package is powered by 480 volts, three phase shore power supplied by host facility. If main power is lost from host facility the system is rendered inoperable other than air stored in receivers. As with typical medical grade air systems the package is designed to deliver twice the required operational air with redundant processing equipment in order to facilitate service and repairs if necessary. The compressors are controlled by the PICO™ located on the control panel, and secondarily by a PICO™ on the breaker box below the compressors. The compressors are turned on by pressing “ALT” three times on the PICO. This should be done routinely at the beginning of every work day. To shut off the compressors select “OK” on the PICO™ controller. Should an emergency arise, the compressors can be shut off by pressing the large red button labeled “emergency shut-off.” Depending on your facility you will be equipped with one of the two compressor packages shown below. a) Champion Reciprocation Compressors This system operates four 10 HP Champion two stage reciprocating compressors delivering 35.4 CFM of free air each @ 175 PSIG for a total volume of 141.6 SCFM. The compressed air is processed through several stages beginning with an air cooled heat exchanger. Next, the air travels through a tube and shell water cooled heat exchanger, and finally through appropriate particulate and carbon filters. The compressors are controlled by the PICO™ which monitors all equipment operation and air quality. The compressed air is stored in two 33 gallon stainless steel air receivers for delivery at time of chamber pressurization and treatment cycle. b) Powerex Scroll Compressors This system consists of eight 5 HP Powerex™ Scroll compressors delivering 12.1 SCFM of free air each @ 145 PSIG for a total volume of 96.8 SCFM. The compressed air is processed through refrigerated air dryers and appropriate particulate and carbon filters. The compressors are controlled by the PICO which monitors all equipment operation and air quality. The compressed air is stored in two 33 gallon stainless steel air receivers for delivery at time of chamber pressurization and treatment cycle. Daily Visual Inspection: On a monthly basis a technician should visually inspect equipment for following: 1. Rubbing stainless steel hoses, copper tubing etc. 2. Check air cooled after-cooler “radiator” surface for blockage 3. Check V-belt belts for glazing/wear 4. Check condition of compressor base vibration mounts 5. Listen and look for air leaks or fluid leaks 6. Listen for unusual knocks, squeals, or chirping noises *Note: Should any of the above conditions develop contact the Director of Facility Maintenance. Medina has (2) Kaeser Compressors. 34 Operational Equipment Environmental Control Unit - ECU: Description: The Environmental Control Unit (ECU) provides cool air to patients inside the chamber during the treatment cycle. The unit is black and is located at the front of the chamber. The cooling process begins by drawing internal chamber air in through the bottom and over a charcoal filter. Air is then circulated across a set of chilled radiator coils by means of two pneumatic blower motors. Finally, cooled air exits on the top side of the unit above the television. Operation: The ECU is turned on manually by the black valve on the upper left hand side of the unit. The air is circulated by two pneumatic blower motors which should be set at 80 PSI by the regulator on the upper left side of the unit. It can be adjusted to the desired speed merely by opening the Flow Control Ball Valve located on the left side above the regulator. As with any air conditioning process there is condensate which accumulates internally. A collection trough is located below the chilled radiator coils and requires routine draining. During warmer months this should be done daily at the end each treatment. During the winter months draining is usually unnecessary. To drain the condensate, manually turn the drain valve so it is parallel with the brass piping (shown below). (NOTE: Must be done under pressure). The ECU drain is located on the bottom left hand side of the unit. Medina has (2) Amron units. ECU ON/OFF VALVE 35 Operational Equipment Air Cooled Refrigeration Chiller: Description: The refrigeration chiller has two functions. The primary function is to provide a cooling process for the compressed air system. This cooling process takes place when air flows through what is referred to as a tube and shell heat exchanger or “chiller”. Chilled fluid circulates in the outer shell of the heat exchanger cooling the air flow in the inner tube. The second function is to provide the ECU with chilled fluid to allow climate control inside the hyperbaric chamber. Operation: The unit is in continuous operation per manufacturer’s recommendations. The chiller and its functions are monitored by the Pico controller. In the event set parameters deviate, warnings are displayed on the control panel. The CHILLER can be manually shut off by the green power switch located on the control panel. Daily visual inspection should be performed to monitor fluid levels, temperature and inspect cleanliness of the intake filter. Daily visual inspections are mandatory. Should a malfunction arise be sure to initially check for the following; fluid leakage, excess temperature or foul odor. *Note- primary service is to refill chiller fluid reservoir with glycol or anti-freeze. Oxygen Supply/Farm Description: All clinical hyperbaric chambers require a continuous oxygen supply, typically Liquid Oxygen (LOX). Many gases liquefied at low temperature can be supplied in large tanks capable of holding tens of thousands of liters (shown below). This method is more efficient when large volumes are consumed. Cylinders of oxygen are frequently used as an emergency backup source. Oxygen Farm Oxygen Line Oxygen Pedestal 36 Operational Equipment Television Enclosure: Description: All our hyperbaric chambers are equipped with an LED HDTV. It provides entertainment for the patients during their hyperbaric treatment. The actual tuning controls for the programs and movies are performed by the Chamber Operator on the exterior of the chamber depending on your system set-up. Equalization & Venting Operation: The television is encased in a sealed Plexiglas enclosure which is designed to segregate the electrical circuitry from the pressurized atmosphere of the chamber. This creates the need to circulate air through the enclosure in order to dissipate heat generated by the television and to equalize the pressure on the Plexiglas case. The pressure equalization and heat dissipation is accomplished by means of two mechanical check valves. One allows external pressure to enter the enclosure and balance internal with external and the other is to balance pressure during dive ascent and heat dissipation resulting from the operation of the television. Another requirement for heat rejection is to create an air flow. This is accomplished by inducing regulated air into the enclosure at 5 PSI above chamber operating pressure, approximately 20 PSI. 37 Operational Equipment Fire Suppression System (FSS): In the event of a fire at a Mobile Hyperbaric Centers location, all chambers are equipped with a fire suppression system (FSS). Fire suppression systems are used in conjunction with smoke detectors and fire alarm systems to ensure safety at the centers. The extinguishing system includes spray nozzles and hoses that are located inside the chamber. Mobile Hyperbaric Centers’ systems are compliant with the National Fire Protection Association (NFPA 99). The Fire Suppression system is routinely tested twice a year. Description: Per National Fire Protection Association (NFPA-99) the hyperbaric chamber is required to operate with two separate fire suppression systems. The first is referred to as a deluge system, where upon activation, the chamber receives 2 gallons of water per internal floor space square foot within one minute (about 365 gallons). The second “system” is the ability to operate at least two hand held water nozzles with a flow rate of five gallons of water for four minutes. Both systems have been engineered to these exact specifications. These details are outlined in the System Evaluation Manual prepared by Coker Engineering, LLC. Operation: Two on-board water storage tanks hold the required amount of water for activation. The deluge system tank holds 365 gallons and is located in the lower bay directly under the area of the control room. The 40 gallon hand line tank is stored in the rear section of the chamber. 1. Deluge Activation a. First, remove the white Delran Safety Clevis b. Next, push the red “Activation Button” *Note: There are multiple locations for system activation. Activation can be done at either the control panel, entry lock, or one of two inside the hyperbaric chamber. Also, system override is possible by turning the red ball valve to the “Override” position. 2. Hand Line Activation a. Activation of Hand Line system is accomplished by opening the supply by use of the red valve on the hose. b. Turn the red valve parallel to the supply piping to charge Hand Line the hand line. *Lights and communication will immediately switch to a battery a backup system* Note: Some systems are equipped with a sapphire system. Please consult your trainer for more information related to this. 38 Operational Equipment Control Panel: The control panel houses the instruments and controls necessary for the safe and effective operation of the hyperbaric chamber and its subcomponents. The control panel layout will vary depending on your location. PICO Chamber Monitors Fault Lights Entertainment Volume Chamber Lights Communications Oxygen Sensor Main Lock Pressure Gauge Entry Lock Pressure Gauge Entry Lock Press Valve Main Lock Press Valve Main Lock Exhaust Valves FSS Pressure Gauges On-Board Generator Description: Located on the front of the trailer is a 24 KW diesel generator. This unit is not capable of supporting patient treatment cycles due to power requirements of the compressor package. The primary focus of this unit is to support necessary equipment to allow for safe abortion and evacuation of our patients. This could occur should we lose main power provided by our host facility. The unit also provides adequate climate control required for equipment and fire suppression storage tanks. The generator should be run for 30 minutes, once a month as part of the safety program. 39 Operational Equipment Entry Lock: The Entry Lock is the secondary treatment compartment of the hyperbaric chamber. Generally, the entry lock is used for emergency access and removal of patients or as an extension of the hyperbaric treatment room. There have also been instances where the entry lock is used to transfer large items in/out of the hyperbaric chamber such as hoods or blankets. Controls for operating the entry lock can be found on the control panel. Please be aware that the entry lock pressurizes at a faster rate and will need to be closely monitored when patients or staff are inside. The steps for ASCENDING in the entry lock are outlined below. 1. Chamber operator changes the secondary pressure gauge to “entry lock” 2. Entry lock attendant or physician will then turn the entry lock exhaust valve 45 degrees (Step 1) when they are ready to begin ascending. The main lock door must be closed during the ascent in the entry lock. 3. At 1.7 ATA the entry lock attendant will completely open the exhaust valve to the complete parallel position (Step 2) 4. At 1.3 ATA the chamber operator will then open the entry lock exhaust valve located on the control panel. Step 1 Step 2 MedLock: The Medlock is a quick access pass thru port designed for transferring small objects in and out of the hyperbaric chamber. It must be pressurized to the side that will be accessed before it will open. To send items into the hyperbaric chamber follow the steps below: 1. Pressurize the Medlock to the exterior by turning the black arrow so it faces toward you 2. Gently unscrew each of the four clamps to open the access door 3. Place contents into Medlock 4. Close access door 5. Line up seals on access door and begin sealing clamps a little at a time until all clamps are firmly closed Turn Here 6. Finally, turn the black arrow away from you back toward the interior of the chamber 40 Maintenance It is the responsibility of all the clinical staff to understand how to complete routine maintenance of the hyperbaric chamber and all of its subcomponents. This section briefly outlines technical guidance in troubleshooting and servicing our equipment. The process for reporting a maintenance concern or equipment failure is also described below. Any immediate maintenance or equipment questions should be directed to the Director of Facility Maintenance. To take a compressor off-line: If there is a problem with a compressor, and you have contacted the Director of Facility Maintenance for approval, take the following steps to take the compressor off-line: 1. Take a regular screwdriver and the trailer keys and locate the compartment at the bottom rear of the trailer (curb side). 2. Open the compartment door; you should then see a simple door with four hour meters on the front. 3. Using the screwdriver turn the screw above the handle at the same time you turn the handle. 4. Once opened you should see a bank of six circuit breakers. The line indicates a complete circuit and is in the “on” position. The circle indicates an interrupted circuit and is in the “off” position. The four large circuit breakers are for the compressors and are sequential beginning on the right and moving to the left. 5. Move the switch from a line to a circle to take the appropriate compressor off-line. How to fill the fire suppression system: If water is lost in the hand line or the main line for the fire suppression take the following steps to refill the system. 1. Purge air from the main tank (the tank will not refill if there is too much air pressure in the tank). 2. Locate the proper fill/drain portion of the hand line tank or main line tank fire suppression system. 3. Attach hose to a water source. 4. It may be necessary to use a female to female adapter. 5. Turn valve from the closed to the fill position. 6. Turn on the water and fill either the hand line or main line tank until the lights returns to green on the control panel. 7. Turn the valve off to the hand line tank or the main line tank. 8. Turn the water off. 9. Pressurize either tank with the K bottle. 10. Remove the hose and check that the system is maintaining pressure. 11. If pressure is not maintained then double check all valves. 12. If all valves are off then check for leaks in system. 41 Maintenance Filling the Lift Fluid (Braun ONLY): Visual inspection of the fluid levels on the hydraulic lift should be done monthly. Pull out the hydraulic fluid stick located above the lift circuit board. Should the fluid level read low, replace with appropriate fluid. Regular or Synthetic transmission fluid should be kept onsite for refilling. Check Here Fill Here Transmission Fluid 42 Maintenance Changing HVAC/ECU Filters: There are 2 types of filters for the hyperbaric chamber. One is a 1” filter for the HVAC unit located in the control panel area, and the other is a 4’ charcoal filter for the ECU. The ECU filter can be replaced by removing the four black nuts located on the bottom piece of the ECU. This filter should be replaced every six months. The HVAC filters should be checked monthly and are replaced as needed. They can be replaced by opening the overhead compartment in the closet of the control panel room and sliding out of the grooves. Figure 1-1 HVAC Filter HVAC Filter ECU Filter Filling the Chiller Fluid: Visual inspection of the fluid levels on the CHILLER should be done daily. If the fluid levels are below the half-way point on the site glass, replenish the system. Antifreeze or polypropylene glycol should be added to prevent overheating. Locate the appropriate fluid on hand and add fluid to the reservoir. Reservoir 43 Maintenance Fire Suppression System Testing Specifically trained employees are required to test the fire suppression system (FSS). Those employees should receive proper training on the process of testing the FSS. The FSS System is tested twice per year. Follow the steps below when conducting a FSS test. 1. Notify all persons in the center and adjacent facilities that a fire suppression test is being conducted. 2. Check Deluge Tank for proper water level. 3. Bleed off pressure of the Deluge Tank from 190psi to 90 psi. 4. Remove both ¾” FSS nozzles in entry lock, plug with ¾” NPT brass plug, and wrap with Teflon tape for a proper seal. 5. Inside the main lock, disconnect FSS headers from the bulkhead. Loosen the black mounting blocks along header inside main lock in order to shift headers away from bulkhead by approximately 12”. A few mounting blocks may need to be completely disconnected in order to accommodate necessary clearance. Install two SwageLok Garden Hose Adapters to the bulkhead assembly. 6. Connect two Garden hoses to the adapter assembly and route to the exterior of the building for adequate drainage. Take caution with the placement of hose ends. This prevents the potential for injury as whipping may occur when water is discharged. 7. Remove safety clevis and depress the RED FIRE SUPPRESSION PALM BUTTON on CONTROL CONSOLE. 8. Verify system actuates within 1 second. 9. Verify alarm is activated both audibly and visually. 10. Once system has been initiated, verify electrical equipment has been de-energized with the exception of the alarm, communication system, and emergency backup lighting. 11. Allow to continue for approximately 3 seconds. 12. Turn FIRE SUPPRESSION OVERRIDE valve to the override position and verify the system stops (no flow). Check to make sure the air actuated valve has closed and there is still water in the tank. 13. Once “No Flow” has been verified, return FIRE SUPPRESSION OVERRIDE valve to normal operating position. 44 Maintenance 14. Pull the RED FIRE SUPPRESSION PALM BUTTON to the OUT position, replace safety clevis, and reset electrical control. 15. Repeat steps 7-14 for the following: MAIN LOCK #1 RED FIRE SUPPRESSION PALM BUTTON. MAIN LOCK #2 RED FIRE SUPPRESSION PALM BUTTON. ENTRY LOCK RED FIRE SUPPRESION PALM BUTTON. 16. Drain residual water from all lines. 17. Remove test couplings and hoses. 18. Reconnect the FSS headers inside of main lock at bulkhead. Take caution for proper alignment of connecting nut as there is a slight offset between bulkhead penetrators and header mounting blocks. Disconnecting the first inline mounting block completely will assist with reconnecting header. 19. Check orientation of main lock nozzles and secure all mounting blocks. 20. Remove brass plugs from entry lock headers and replace FSS nozzles. Remove old Teflon tape and install new tape. 21. Return all valves to operating position 22. Fill Deluge Tank to proper level. 23. Pressurize Deluge Tank to operating pressure. Check for leaks. 24. Connect test gauge to each hand-line at nozzle/hose connection and verify line pressure is at a minimum of 50 psig above maximum treatment pressure. 25. Remove test gauge from hand-line and replace nozzle. 26. Make available a container of adequate size to receive 5 gallons of water with activation of Hand-Line. 27. Activate Hand-Line; verify flow and visual and audible alarm activation. Continue flow to verify at least 5 gallon per minute for four minutes. 28. Fill Hand-Line Tank to the proper level. 29. Pressurize Hand-Line Tank to proper operating pressure. Check for leaks. 30. Verify K-Bottle has adequate volume to return center to normal operating condition. 31. Return FSS to normal operating condition. Notify all personnel the FSS test is complete. 45 Maintenance Turn On Here Manual Start-Up of the Generator: It is a possibility that during a power loss the generator will not start automatically. In this case it will be necessary to manually start the generator. The procedure for manually starting the generator varies depending on what model your facility has. Please check with your team or Director of Facility Maintenance to familiarize yourself with the procedure for manually starting your generator. Draining ECU: Depending on usage and temperature variations the amount of condensate will vary dictating frequency of draining. During warmer months this should be done daily at the end each treatment on ascent. During the winter months draining is usually unnecessary. To drain the condensate; manually turn the drain valve so it is parallel with the brass piping (shown right). The ECU drain is located on the bottom left hand side of the unit. Manually operating the Lift (Braun ONLY): Locate the manual lift lever (shown below). Insert the lever into opening above lift circuit board underneath the control panel room and manually pump up the lift to the desired height. To lower simply turn the lift handle. 46 Troubleshooting Chamber will not ascend In the event the chamber will not surface the following steps should be taken. 1. Ensure valves are open completely. * If not then open valves completely (but do not force open). 2. Ensure there is at least 90 psi in the compressor air tanks, *If not, run the compressors until at least 90psi is achieved. (shown right) 3. Ensure all BIBS vents are in the off position. *If no, close all vents completely 4. Ensure clamps are properly tightened on the exterior side of the medical lock. *If necessary the chamber may be surfaced using the 1” ball valves. The valves may be gauged to control the desired ascent rate. Chamber will not descend In the event where the chamber will not descend the following steps should be taken: 1. Ensure PICO™ is on and the compressors are indeed running. *If not, turn on the compressors by following prompts on the PICO™ on the control panel 2. Ensure there is approximately 140-165 psi in the compressor air tanks. 3. Ensure all exhaust valves are closed. *If not, close all valves completely (but do not force open). Lift will not raise/lower In the event where the lift will not raise or lower follow the steps below to troubleshoot the issue: 1. Check Hydraulic Fluid Levels. *If low, fill with transmission fluid 2. Check to ensure the lift remote is plugged in properly. * If not, plug in the remote connection completely 3. Inspect the lift for mechanical failure. ECU is not blowing cool air In the event where the ECU is not blowing, or not blowing “cool” air, the following steps should be taken: 1. Make sure the ECU handle is in the “OPEN” position. 2. Make sure the CHILLER fluid is “FULL." 3. Inspect for mechanical failure (unusual noises, poor blowing, air temperature). 47 Reporting a Maintenance Concern Should a maintenance issue arise, it should be immediately documented on the company INTRANET under the section labeled “Maintenance Log” (shown below). Please be as thorough as possible when documenting the problem. For any immediate maintenance or equipment emergency, contact the Director of Facility Maintenance. Intranet Homepage Maintenance Log 48 Clinical Equipment Oxygen is administered by one of two delivery methods, a hood or mask. We will first discuss the oxygen hood and its accessories. Oxygen Hood An oxygen hood is a breathing device used to administer 100% oxygen to patients in a multiplace hyperbaric chamber. SEALONG’s latex-free is the brand of oxygen hood that MHC uses. A complete set-up includes, an oxygen hood, a neck-ring, and oxygen tubing. 49 Clinical Equipment Hood Creation A new hood will be cut for all new patients on their first day of treatment prior to beginning the dive. This hood will be used for their entire course of treatments with MHC. Please follow the steps outlined below to create a hood set for a patient. 1. Ensure headgear is Non-Latex 3. Measure the patient’s neck with the SEALONG measuring tape. 4. Use the SEALONG measuring chart to determine where to cut the neck-ring. 5. When cutting the neck- ring, cut one line smaller than the measured number. *Note: When cutting a neck-ring try to limit JAGGED edges, it promotes tearing. 6. Next, place the cut neck-ring over the patient’s head to determine proper fit. It might be necessary to trim an additional ¼” or less to ensure a comfortable fit. Keep in mind, a “snug” fit is necessary to maintain a proper seal around the patient’s neck. 7. Obtain the remaining supplies to complete the full hood set: a) Hood b) “2” Hoses c) “4” tubing adapters d) Electrical Tape e) Hair tie f) Scissors 8. Next measure “2” 12-13 segment lengths of oxygen tubing (ie: hose) and cut with scissors. Mark the tube tip that will be used for oxygen with green electrical tape 9. After every third segment, connect the tubing using electrical tape, start three segments from each end. 10. Using an alcohol prep, place four tubing adapters on the ends of the oxygen tubing. 11. Finish with rolling up the completed hose and placing it inside of the hood. Secure the neckring onto the bottom. The hood set is now complete. 12. Label the hood with the patient’s name on the silver name label. (AND NECK RING) *Note: If the patient is on a seizure protocol, label the hood with a red dot to distinguish he/she from the remaining patients. (shown right). Label the hood with a YELLOW dot if the patient has a DNR. 50 Clinical Equipment SEALONG Hood Measuring Chart: SEALONG Hood Accessories: 51 Clinical Equipment Frequently Asked Questions/Troubleshooting Here are some frequently asked questions regarding Sea-Long Medical Systems and its products. If you have additional questions or comments, please contact Britani Bramble at 517-581-9428. How do I keep from tearing the neck-ring? First and foremost: a smooth, non-jagged edge must be obtained. Before stretching the opening over the patient's head, fold the cut edge under slightly so the stretching occurs at the fold instead of the cut line, which is now the neck-ring’s weakest point. What is the best way to cut the neck-ring to size? Determine the patient’s neck size using a tape measure. The neck-ring should fit tight enough to prevent leaks but not be uncomfortable. Use the TRIM CHART to determine the rib that most closely matches the neck measurement. If it is between sizes, choose the next smaller rib to prevent cutting too large. Use sharp scissors to make one continuous, smooth cut. This will prevent jagged edges and tearing points. Why is it so difficult to separate the hood from the neck-ring? Most likely the o-ring needs to be lubricated with an oxygen compatible lubricant. It only takes a very small amount to make it easier. Too much lubricant will make it difficult to keep the seal in place. Place a pea size amount onto your finger and rub directly on the o-ring. Equally distribute the lubricant over the entire oring. What type of lubricant do I need for the neck-ring? An Oxygen Compatible lubricant is necessary and should be available at your facility. How does the Trachea Accessory work? The Trachea Accessory was designed especially for the trachea patient. If the patient has a traditional 'Jackson' or 'Shiley' Trachea, the PN600 fits perfectly over the opening! The swivel action deters pulling on the often sensitive skin area surrounding the trachea, and the connector piece allows the inhale and exhale tubing to connect without using the hood. A diagram for proper connection is included. What flow rate is used with the Trachea Accessory? This can be tricky and will vary, so consult your patient on what is comfortable. We have surveyed and found that a flow rate of 15 LPM to 30 LPM has been successful. You must closely monitor blood oxygen levels. What causes visor fog? Fogging occurs because of high humidity percentages inside the hood. It can often be eliminated by adjusting the flow rate. My patient has a stoma, what can I use? The "Tape-on Neckdam" is available for patients with a stoma. The Neckdam Sleeve works exclusively with our Replacaeble Series to make it convenient and cost-effective. 52 Clinical Equipment SEALONG Trachea Attachment: VI) For patients with a trachea, please follow the steps below: 1. First remove trachea accessory from bag. 2. Fit tubing on one side of the trachea “T” for the intake and the other side for exhaust. 3. Turn on O2 flow. 4. Attach the Trachea Accessory over the patient’s trachea opening. 5. Continue treatment normally. 6. If the patient exhales a substantial amount through the mouth, then a hood will be required. **When setting up for a trachea patient, use what is most comfortable for the patient. 53 Clinical Equipment SEALONG Mask Creation: For any patient or staff that requires a mask for oxygen breathing they must be properly measured and fitted. Please follow the steps below for mask creation and fitting. 1. Using the SEALONG measuring tape, measure between the patient’s eyebrows to their chin (shown right) to determine correct mask size. 2. Begin assembly of complete mask set using picture below by using the picture below as a guide. Be sure to choose appropriate reservoir bag size to prevent over inflation on exhalation. 54 Ancillary Equipment It is necessary to obtain patient vital signs during the pre-treatment assessment and possibly during emergent circumstances. You will be required to become familiar with the use of three standard pieces of medical equipment; a glucometer, thermometer and sphygmometer. The individual medical devices are described in more detail below. Blood Pressure Monitor(Sphygmometer) A sphygmometer is a device used for measuring blood pressure in the arteries. It is important to familiarize yourself with the type of sphygmometer your center carries as models differ. As a technician you will be required to obtain a patient’s blood pressure daily with either a manual or mechanical sphygmometer. Thermometer A thermometer is a device used to measure a patient’s internal body temperature. It is important you familiarize yourself with the type of thermometer your center carries as models do differ. As a technician you will be required to obtain a patient’s temperature daily. Glucometer A glucometer is a medical device used for determining the approximate concentration of glucose in the blood. MHC typically uses a hospital issued glucometer to obtain a pre and post treatment blood glucose level (BGL) on all diabetic patients. It is also mandatory to obtain a pre and post BGL on a patient’s first day of treatment. In the event where a technician needs to obtain a BGL inside the hyperbaric chamber a onetouch glucometer is used, more specifically the ACCU-CHECK ADVANTAGE(shown below). *Note: Please remember to disinfect the glucometer after every use.* One-touch Glucometer Hospital glucometer 55 Resupply There are two different processes for ordering clinical supplies. Hyperbaric treatment supplies are ordered by MHC using the corporate supply order form shown below. All other clinical supplies are ordered through the center’s hospital supply warehouse. Be sure to familiarize yourself with your center’s resupply process and your responsibilities as a technician. MHC Resupply Process Please follow the steps below when placing an order for specific hyperbaric supplies such as hoods, masks, oxygen hose, alcohol disinfectant, etc. 1. Fill out the proper MHC Purchase Order Form. This can be found on the MHC Intranet under Forms > Expense Forms and Purchase Order > Purchase Orders 2. Add desired supplies by filling out the top portion and having your center director sign and date the bottom. 3. Fax Attn: Mary Miller at 216-674-8233 *Note: Corporate orders are placed every Wednesday, please allow approx: 2 buisness days for all SEALONG orders. *Note For anything not listed on the standard order form. Please complete a special purchase order form. This can be found on the employee intranet under Forms > Expense Forms and Purchase Order > Purchase Orders. 56 Typical Center Schedule The typical hyperbaric center schedule consists of three treatments or “dives” per day Monday through Friday. On Saturday two treatment times are typically available. A sample center schedule is shown below. Patients are expected to arrive approximately 30 minutes prior to their scheduled treatment time. This allows ample time for the pre-treatment assessment and any possible discussions with the hyperbaric physician. Please familiarize yourself with your centers’ schedule. 57 Daily Treatment Flow HBOT Daily Timeline Mobile Hyperbaric Centers, Inc. Group1 7:00 AM 7:15 AM 7:30 AM 7:45 AM 8:00 AM 8:15 AM 8:30 AM 8:45 AM 9:00 AM 9:15 AM 9:30 AM 9:45 AM 10:00 AM 10:15 AM 10:30 AM 10:45 AM 11:00 AM 11:15 AM 11:30 AM 11:45 AM 12:00 PM 12:15 PM 12:30 PM 12:45 PM 1:00 PM 1:15 PM 1:30 PM 1:45 PM 2:00 PM 2:15 PM 2:30 PM 2:45 PM 3:00 PM Group2 Group3 Group4 H&P TIME Assessment Transport Begin H&P 8:30 am HBOT End Transport Assessment Transport Begin H&P 11:00 am HBOT Assessment End Transport Transport Begin H&P 1:30 pm HBOT End Transport 58 Daily Patient Flow Patient arrives for HBOT no more than 1/2 hour prior to treatment time Patient registers with hospital for billing/insurance and Practice Manager Patients should arrive in scrubs Store any valuables. And continue to triage area Patient undergoes pre-treatment assessment Patient has abnormal vitals, CP, SOB, or other complaints Patient has no complaints noted on Assessment Form (form signed by physician before treatment) Patient MUST be evaluated by HBO MD on duty Patient transported into the HBO chamber with hoods, blankets and drinks. Patient is cleared for HBOT by HBO MD on duty HBOT STARTS HBOT ENDS Patient transported back to MHC waiting area. Hoods are cleaned and stored by Technicians Patient sent directly to ED or referred to PCP for evaluation No further HBOT until clearance from Pt.'s PCP has been received Patient changes into street clothes, scrubs and blankets put into lockers or laundry. 59 Daily Logs Daily Startup/Shutdown logs are required to be completed by the end of every working day. These logs verify that all locations are operating within the safety parameters defined by MHC. Both the startup and shutdown log can be found on the Intranet. To fill out the log follow the steps below: 1. Go to the MHC Intranet at www.mhcenters.com/forum 2. Select the appropriate log on the left side of the Home Screen (shown below). 3. Select “New Log Entry” and fill out all required fields. 60 ECW Electronic Charting Input of Patient Vitals The third technician or “off tech” is responsible for entering vitals for that respective treatment. It is necessary to enter patient vital signs as part of the documentation process. 1. Click on “S” Jelly Bean 2. Double click “Patient” name 3. Next, scroll down and select the “Vitals” section (Figure 2). (NOTE: do NOT enter vitals for PHOTO) 4. Input vitals in yellow highlghted field (Figure 3). 5. When finished check the “Vitals Taken” box and close. Figure 1 Figure 2 Figure 3 61 ECW Electronic Charting Merging a Template The third technician or “off tech” is responsible for entering the data for that respective treatment. It is necessary to “merge a template” as part of the documentation process. Merging a prefilled template minimizes repetitive data entry and creates a more efficient work flow. Follow the steps below to merge a template for each patient. 1. Select the “Templates” button on the button of the screen. 2. Select a generic template unique to your center or a patient specific template from a previous dive that coincides with the current dive time. Be sure to only select visits labeled “DIV” (shown below). 2 3 4 3. Next, click “Merge Template” on the bottom of the window (shown above). 4. When the merge is complete click “Close” in the bottom right hand corner of the window. 5. Review merged content for accuracy and fill in all blank fields (ie: technician, dive duration, dive depth, air break times). 6. When finished, return to the main schedule by selecting “Resource Schedule” on the top left side of the screen. 62 ECW Electronic Charting Checking out a Patient The third technician or “off tech” is responsible for entering the data for that respective treatment. It is necessary to “check out” a patient as part of the documentation process. Every patient on the schedule must be checked out by the end of the working day. When checking out a patient you will choose from the listed criteria to acknowledge the patient’s visit status for that day. Follow the steps below for checking out a patient. 1. Double click on the desired patient’s name. 2. Note any global alerts in the pop up window (ie: Falls Risk, Linking of Referral, HIPAA). 3. Verify your facility is selected. 4. Select the drop down tab and select appropriate label (usually “Check Out”). 3 4 63 ECW Electronic Charting Linking a Referral The third technician or “off tech” is responsible for entering the data for the respective treatment. It might be necessary to “link a referral” daily, as part of the documentation process for certain patients. Note: a referral does NOT need to be linked for “ABHO”, and “Non-Billable” treatments. Follow the steps below to link a referral. 1. Double click on the desired patient. 2. Acknowledge the global alert prompt that appears, and verify if referral linking is required. 3. Next select the “Referrals” tab located on the top of the appointments window. 4. In the new window 3 select the appropriate referral by single clicking the desired line and then selecting “update” and then “close.” 4 *Note: email the patient care coordinator at least 5 days before the expiration of the auth 64 ECW Electronic Charting 5. Next in the center of the referral window select “Visit Details” (shown below). 6. Single click the first available BLANK line. 7. Choose the arrow to display the drop down menu and select correlating date (shown below). 5 6 8. If two dates are available, select the appropriate “treatment” authorization. The duplicate date should be identified with a “P” indicating it should only be used for linking a “photo” exam. 9. When done select “OK.” *Note: If done correctly BLUE lettering will be displayed on appointment window (shown below) 7, 8 65 ECW Electronic Charting Changing the Time of a Scheduled Visit It is possible a patient will want to change which dive they would like to attend. If so, you could be asked to change the patient’s scheduled visit to another time. Please follow the steps below to change a visit. 1. Double click on the patient you wish to switch. 2. Change the “Start Time”, “End Time”, and “Visit Type” (shown below). 3. Click “Ok.” Cancelling a Patient Visit Ideally we would like all of our patients to attend HBOT every day. In the event that a patient must cancel their visit, you could be asked to document the cancellation on the schedule. Patient cancellations are only completed if the patient has communicated with the center as to why they are unable to attend treatment (only cancel after you have attempted contact). Depending on your center’s routine, you could also be required to place a telephone encounter to document the cancellation in greater detail. Follow the steps below to simply cancel a patient from the daily treatment schedule. 1. Double click on the patient you wish to cancel. 2. Select the “Visit Status” dropdown. 3. Select “Cancelled.” 4. Next, in the “General Notes” section, briefly describe the reason for the patient is cancelling, when they plan to return, and your initials. 3 4 5. Select “OK.” 66 Pre-Treatment Duties Pre-Treatment Assessment Patients will begin to arrive approximately a half hour before their scheduled treatment time. This allows ample time to complete the pre-treatment assessment and allows the HBO physician to complete any photo exams. During treatment, it is the responsibility of the third or “off” technician to complete the pretreatment assessment for the upcoming patients. The pre-treatment assessment is completed in the designated triage area. Please use the Daily Treatment Assessment Form (shown below) to complete a thorough and accurate assessment. Any abnormal vital signs or complaints must be noted on the assessment form and communicated to the HBO physician before the patient can begin treatment. Abnormal vital signs are outlined below. When all assessments are complete, the completed assessment forms are then given to the HBO physician for review. Once the HBO physician has signed all of the assessment forms the treatment may begin. As a reminder, all new patients will have their BGL taken before and after their first treatment regardless of referring diagnoses. *Abnormal Vital Signs* BP >150/90 or <100/60 BGL 75-150 Non-dia (>120 or <200 for diabetics) Pulse >110 or <60 Resp >20 or <12 Temp >100F *Note: All questions must be asked, and all fields must be completed* 67 Pre-Treatment Duties Transporting Patients Safely transporting our patients to or from the hyperbaric chamber is a priority for all technicians. This includes ambulatory and non-ambulatory patients. Please pay careful attention to patients who are at an increased risk for falls. These patients should be identified by a red wristband. There are two methods of accessing the hyperbaric chamber, direct entry (i.e. stairs, level) or the hydraulic lift. Depending on your facility, the type of access you may have will vary. To minimize workplace injury, always use at least two technicians when lifting or moving a patient. For further clarification on lifting or moving a patient please refer to the Safety and Operations Manual. Transporting using the Hydraulic Lift 1. Position patient evenly on lift. 2. If in a wheelchair, lock brakes on both sides of the wheelchair (shown right). 3. Inform patient of impending raise or lowering of the platform. 4. Secure gate where applicable. 5. Raise or lower the patient using the lift remote. Note: After the patient is secure, please use the yellow warning strap before it is lowered. Prepare Treatment Essentials It is also the responsibility of the third “off” technician to prepare all of the items necessary for the upcoming treatment. This includes but is not limited to; patient hoods, patient drinks, blankets, and pillows. Ideally all of these items, along with all patient assessments, should be completed before the previous dive reaches the surface. Drink Tray 68 Treatment Responsibilities During a Treatment During a treatment each of the three technicians on duty will assume one of the following roles: 1. Chamber Operator 2. Chamber Attendant 3. Third “Off” Technician Below are the responsibilities of each of the above positions. Please familiarize yourself with each role as you will be required to perform all duties independently. Training Note: Tech should be switching out operator role at least every 30 minutes. Chamber Operator: The Chamber Operator is responsible for the safe operation of the hyperbaric chamber during a treatment. The Operator must never leave the treatment controls unattended. If a break is necessary, you may ask a fellow technician or physician to switch you positions. Responsibilities include but not limited to the following: 1. Monitors chamber pressure, and adjust accordingly to maintain 2.4 ATA at all times. 2. Monitors chamber oxygen level maintaining below <23.5% at all times. 3. Accurately documents all dive times in the treatment log book and notes any abnormal events that occur during the treatment. 4. Measures the passage of time during treatment by the use of digital timers on the control panel. 5. Maintains constant verbal communication with chamber attendant and advises he/she of all changes in pressure. 6. Maintains constant attentiveness of patients and chamber attendant at all times. 7. Visually inspects control panel for any abnormalities or fault lights prior to, and throughout the duration of the treatment. 8. Minimizes distractions and stays focused on observing treatment by the viewing monitors. Distractions include (books, cell phones, laptops, personnel). 9. Aware of any unique patient circumstances affecting normal treatment regimen (i.e. seizure protocols, new patients, illness). 69 Treatment Responsibilities Chamber Attendant: The Chamber Attendant is responsible for the safety of the patients during a hyperbaric treatment. The Attendant must remain seated near the door and maintain an unobstructed forward view of all patients. The Attendant must also breath oxygen the last 20 minutes of the treatment, responsibilities include but not limited to the following: 1. Accompany and observe the patients inside the hyperbaric chamber for the duration of the treatment. 2. Verify there are NO prohibited items in the hyperbaric chamber prior to beginning treatment (see pg 30). 3. Verify all necessary items have been placed in hyperbaric chamber prior to beginning treatment (ie: hoods, drinks, blankets) 4. Prepare patient seating areas and ensure patients are comfortable 5. Set up patient breathing devices (ie: hoods, masks) 6. Coaches patients on proper ear clearing techniques 7. Places and removes hoods between oxygen breathing periods and air breaks (Note, think about “flushing” O2 from hood by switching to air) 8. Minimizes oxygen leaks by ensuring proper fit and assembly of oxygen breathing devices 9. Responsible for cleanliness of hyperbaric chamber at the conclusion of the treatment. This includes: a) Removal of all trash b) Replacement of stools in a secure location to minimize tripping. c) Restoration of all arm rests to the upright position. Third “Off” Technician: The Third Technician acts in a supporting role during a hyperbaric treatment. Responsibilities include but not limited to the following: 1. Accurately completes all electronic charting in ECW for the ongoing treatment (ie: vitals, linking referral, checking out patient, post glucose) 2. Perform pre-treatment duties for the upcoming treatment (ie: patient assessment, drink preparation, etc) 3. Perform History and Physical (H&P) orientation and tour (i.e. Technician Checklist) 4. Upload daily assessment form to patient file in eCW 70 Treatment Routine The steps below will outline the typical treatment routine followed by the Chamber Operator and Chamber Attendant for a dive to the depth of 2.4 ATA Depth Chamber Operator Chamber Attendant Depth 1.0 ATA Perform Safety Check Fill out Dive Log 1.0 ATA >1.0 ATA Opens press valve, begin decent Choose DVD, put on booties/crocs Verify all Hoods are inside chamber Verify no prohibited items "10sec time-out" Signal operator when ready, put on gloves 1.5 ATA 2.0 ATA 2.4ATA Dim lights 1, 2 Dim lights 3, 4, 5 Close press valve, verbalize "On 02" Start Timer(s) for 45min and begin movie >1.0 ATA Begin placement of oxygen neckrings back to front Begin placement of oxygen hoods back to front Turn on each supply valve to "oxygen" Adjust flow meter between 20-30 lpm 1.5 ATA 2.0 ATA 2.4ATA Oxygen 20 Minutes Verbalize "Patient Check" Command Ask patients how they are feeling? Oxygen 25 Minutes Verbalize "Air Break" command Start 5 min. Timer Turn on each supply valve to "off" Air Break, Remove all patients' hoods back to front Oxygen 20 Minutes Verbalize "Patient Check" Command Ask patients how they are feeling? Oxygen 20 Minutes 2.4ATA 1.7 ATA 1.6 ATA <1.3 ATA 1.0 ATA Technician breathes O2 (mask or hood) Verbalize warning, with one minute remaining Put on gloves, remove oxygen Begin removing patient hoods, back to front Turn all lights on, open primary or secondary exhaust valve Remove remaining headgear Open primary or secondary exhaust valve Close primary and secondary exhaust valves Verbalize "Deco Stop" Start Timer for 3 minutes At end of the "Deco" open all exhaust valves Open Exhaust Ball Valve (primary, secondary, ball valve in entry lock) Door will open at Surface, Close Valves 2.4ATA 1.7 ATA 1.6 ATA <1.3 ATA 1.0 ATA 71 Post-Treatment Duties Post Treatment Vitals It is necessary to obtain a post treatment BGL on all diabetic patients (shown below). Patients must have a post BGL of >100 mg/dl before they can be cleared to leave the facility. Notify the HBO physician if any patient is below the threshold listed above. ECW Sign-off by MD Vitals are checked before treatment After all patient vitals have been entered into ECW and the patient has been checked out, the HBO physician can then “sign” all the records for the day. Hand Off Forms Hand off forms are required for patients being returned to a skilled care facility, nursing home or rehabilitation center. The form can be found on the company Intranet. This form should be completed, signed by the physician, and subsequently copied and filed in the patient’s chart. Hood Cleaning At the end of the day all the patient hoods that were used should be cleaned for the upcoming day. Dirty hoods should be kept separate from clean hoods whenever possible. Dirty hoods can be identified by placing them on the shelf with the neckring facing outward. This easily distinguishes them as “dirty” (see right). Hood should be cleaned with ALCOHOL ONLY! This prevents premature breakdown of the neckring. Hoods are cleaned by separating the hood from neckring and cleaned individually. When hoods have dried they can be reassembled and placed upright on the shelf with the name label facing outward. Ideally, hoods should be cleaned in between treatments if timed permits Dirty Hoods 72 Infection Control Health care employees can take steps to prevent the spread of infectious diseases. These steps are part of infection control. Proper hand washing is the most effective way to prevent the spread of infections in hospitals. Other steps health care employees can take include 1. Covering coughs and sneezes. 2. Staying up-to-date with immunizations. 3. Using gloves, masks and protective clothing. 4. Making tissues and hand cleaners available. 5. Following hospital guidelines when dealing with blood or contaminated items. 6. Proper washing of patient garments including scrubs, pillow cases, blankets. 7. Minimizing cross contamination of patient specific garments. 8. Thorough weekly cleaning of all patient care areas including hyperbaric chamber and triage area. Weekly Chamber Cleaning The hyperbaric chamber should be tidy and well-kept at all times. A daily cleaning of the hyperbaric chamber should be completed by the end of every working day. In addition, a thorough cleaning of the main and entry lock chambers, control panel area, and triage area should all be completed on a weekly basis. Follow the steps below when thoroughly cleaning the hyperbaric chamber. 1. Only use MHC approved hyperbaric cleaning agent called “BETCO” 2. First, vacuum the entire hyperbaric chamber including the floor, behind seats, and in the entry lock. 3. Next, spray down walls and begin to wipe down with washcloths from ceiling to floor *Note: Remove stools when wiping down walls. 4. Replace stools and proceed to clean the floor of the hyperbaric chamber either with a mop or by hand with a washcloth. 5. Let the chamber ventilate for at least an hour before resuming treatments 73 Education It is the goal of Mobile Hyperbaric Centers to offer continued education in the field of hyperbaric medicine. Initially all new full time technicians will be eventually scheduled for an Introductory Course in Hyperbaric Medicine. The technician will attend the soonest available Introductory Course in Hyperbaric Medicine at Nix Memorial Hospital in San Antonio, Texas. The course is designed to strengthen the knowledge and foundation of the new employee in the field of Hyperbaric Medicine. The Introductory Course in Hyperbaric Medicine is just the first step toward gaining future licensure as a Certified Hyperbaric Technologist (CHT). See the flow chart below to see the steps for certification as a CHT. New Employee is cleared to work independently by Trainer Upon successful completion of 40hr Internship, employee will then be scheduled to take NBDHMT CHT Exam within 6 months. After 440 clinical hours have been completed, Chief Training Officer will be contacted to schedule 40hr Supervised Internship with Employee. employee is scheduled for available 40hr Introductory Course to Hyperbaric Medicine. Hyperbaric Technician CHT Certification Process Upon successful completion of 40hr ICHM, employee will then complete 440 hours of clinical work experience with MHC. *Approx. 3 months 74 Education (Re-certification) After CHT certification has been obtained it is necessary to complete 12 continuing education units (CEU) within a two year time frame to be eligible for renewal. In addition to the 12 CEU’s the CHT must also work a minimum of 100 clinical hours in a hyperbaric department. Currently MHC offers 12 Category “A” CEU credits which can be found on the company Intranet under Education. Please contact the Chief Training Officer for more questions regarding recertification. 75 Incentive In order to be eligible for a bonus, the Center must always maintain a safe and quality program. Therefore the Center must meet these criteria: 1. All items related to safety and particularly the items related to the: Daily Logs, Fire Suppression System Logs, Safety Director Logs, Generator Logs, Emergency Drill Logs, Fire Drill Logs, and Annual Competencies must be 100% complete and up-to-date. 2. Patient Show Rate must exceed 70% average for the month. 3. Weekly follow-up exams and letters sent to referring physicians must exceed 90% for active patients and cannot be missing for more than two weeks for any active patient. 76