Brochure

Transcription

Brochure
Real Time Monitoring
Mirador™
Mirador™ is a web based application for 24/7, real-time monitoring.
This platform is available as a custom configuration to fit multiple
applications of environmental monitoring during the various stages of
production, storage and transportation.
This fully secure system complies with regulatory requirements of the
life science, healthcare and food industries. The Mirador™ system aids
in achieving quality assurance goals as it pertains to environmental
conditions.
Features include:











24/7 monitoring of temperature, humidity, differential
pressure and/or carbon dioxide
Secure access to data via the internet or intranet
Complies with FDA code 21 CFR part 11
Excursion notifications via text, email or pager
Automatic software updates
Multi-site configurable
9 Point NIST traceable calibration
Versatile 4 alarm thresholds
Automatic recalibration reminders
No Downtime Recalibration Exchange Program
Factory installed or self-installation using Mirador Express™
Applications:
Labs and Freezers:
Simplify data retrieval
Food Processing:
Ensure HAACP protocols
Transportation:
Record data world wide
Cold Storage Facilities:
Protect the integrity of your
product
Real Time Monitoring
xTag2 Sensor
features include:
Internal/External Sensors:

Wireless/Wired temperature and humidity
sensor

Screen + LED for local equipment

Up to 5 external sensors

Alarm notifications

Internal memory

Rechargeable battery

Very high RF range

Up to [email protected]
The xTag2 can be directly installed inside the
area to be monitored (in applications above
0C). For applications below 0C, an external
sensor probe is used. The xTag2 is capable of
up to 4 probes.
xTag2 Wireless Sensor Specifications:
TEMPERATURE
TEMPERATURE & HUMIDITY
Range of Temperature
(Accuracy/Resolution)
-40°C/+85°C (-40°F/+185°F), ±0.5°C,
±0.0625°C
-40°C/+85°C (-40°F/+185°F), ±0.5°C, ±0.0625°C
Range of Humidity
(Accuracy/Resolution)
n/a
0-100%, ±0.5°C, ±0.0625°C
Display
Reflective 5 digit LCD
Reflective 5 digit LCD
Interfaces
USB, Ext. Power
USB, Ext. Power
External wired sensors
RJ12 + 6 pins connector
RJ12 + 6 pins connector
Battery
1.85A/H, Li-ion Polymer
1.85A/H, Li-ion Polymer
Internal Memory
2Mb
2Mb
Radio
Battery life
IEEE802.15.4, 2,4GHz, 16 channels, 1 to
300mW
Up to 2 years before rechargeable
IEEE802.15.4, 2,4GHz, 16 channels, 1 to
300mW
Up to 2 years before rechargeable
Dimensions
(not including antenna)
3.5" x 2.25" x 1"
88.5mm x 57.15mm x 25.4mm
3.5" x 2.25" x 1"
88.5mm x 57.15mm x 25.4mm
Real Time Monitoring
xTag Display Reader
features include:

Wireless / wired temperature and humidity
sensor

Dot matrix LCD + Led

Up to 8 external sensors

Alarm notifications

Rechargeable battery

Very high RF range

Up to 300mW @ 2.4GHz
xTag2 Wireless Reader Specifications:
TEMPERATURE
Range of Temperature (Accuracy/Resolution)
-40°C/+85°C, ±0.5°C, ±0.0625°C
Range of Humidity (Accuracy/Resolution)
0-100%, ±0.5°C, ±0.0625°C
Display
Transmissive dot matrix LCD (128x64)
Interfaces
USB, Ext. Power
External wired sensors
RJ12 + 6 pins connector
Battery
1.85A/H, Li-ion Polymer
Internal Memory
32Mb internal + USD card holder
Radio
IEEE802.15.4, 2,4GHz, 16 channels, 1 to 300mW
Battery life
Up to 8 hours
Dimensions
5” x 3” x 1.25”
127mm x 76.2mm x 31.75mm
Mirador™ Express Self Installation Kit:
For smaller applications such as freezers or
refrigerators, Mirador™ is available as an off
-the-shelf kit. This kit includes all the
features and benefits of the full system and
is easy to install.
Mirador Express™ is available in 1, 2, 3 or 4
sensor kits.
Use this QR code for a
brief overview video
of Mirador Express™
How It Works
In a standard wireless mode, each xTag2 sends up to 300mW every 3 minutes, during 10 mill-second.
This allows us to avoid any interference with all your other equipment or devices.
We can provide sensors able to operate between -328ºF to +932ºC, and manage several parameters
such as temperature, humidity, differential pressure, CO2…etc. Our sensors could be embedded or with
an external probe, running on their internal batteries or powered on an electric outlet.
For fixed storage areas, each
sensor collect data and transmit to
a reader (xTagDisplay). The reader
is connected to the dedicated
server where Mirador has been
centrally installed or remotely
hosted.
Protocol
Zigbee
Bluetooth
Wi-Fi
IEEE
802.15.4
802.1.5.1
Memory requirement
4-32ko
250 ko+
802.11a/b/g/n/ndraft
1 Mo+
Battery life
Transfer speed
Years
250kb/s
Months
1Mb/s
Hours
20Mb/s
10001500m
(0.6-1mi)
65000+
50m
(164ft)
300m
(984ft)
7
32
Our system uses a Zigbee protocol,
Range (outdoor)
which is a specification for a suite
of high level communication
protocols using small, low power Nods
digital radios.
Reporting and alarm capabilities
Mirador is web (MySQL data base) and a multi-site application; all recorded data are stored on the
same data base or local ones, where compliant graphs and reports could be easily edited. Reports
available are text report, graph and interactive graph, pdf or .csv export to Excel or Lotus for example.
Software
The Mirador™ Main Interface is a representation of your installation with sensor locations and realtime readings received from sensors. It displays the list of recent alarms triggered by the system as
well as the alarms status messages (acknowledge information and comments). It is also used to access
other screens depending on user privileges.
An audit trails interface is used to display audit trail recorded by the Mirador™ system.
Software
Sensor Readings Screen is used for generating reports or graphics based on sensors readings
(temperature, humidity, pressure etc.) stored in database .
Example report as a graph in .csv format
Example report in .txt format
Example Report in .pdf format
System Validation
Site Acceptance Testing (SAT)
A SAT can be done once the installation is completed. Site Acceptance
Testing is the stage where the manufacturer/customer conduct testing
for the components supplied under the project scope and tests the
conformance of the delivered solution. The Site Acceptance Testing
stage includes 3 parts: integration testing, performance testing, and
user acceptance testing.
System Validation
Upon customer’s request, the Mirador™ solution can be validated
(Installation, Operation and Performance qualification) with a protocol,
as required by the regulations.
May this be the case, the protocol will be written and approved by
Cryopak Verification Technologies Inc. and the customer. The execution
can be done by CVT or the customer. Once this is completed, a final
report will be written.
The system validation is a regulatory work that is done once; if any
system’s modifications occur after the validation (adding a user,
expansion, etc.), a change control document will be inserted into the
initial protocol. This change control will describe the modification and
validate this new addition to the system.
Your One-Stop Shop
on the Path to
Compliance
Ask Us How We Can Help
[email protected]
East and Ontario: 1.888.423.7251
West: 1.888.667.2532