pixis: 2k - Princeton Instruments

Transcription

pixis: 2k - Princeton Instruments
SPECTROSCOPY GROUP
Now Powered
by LightField™
PIXIS: 2K
2048 x 512
The PIXIS series from Princeton Instruments (PI) are fully integrated, low noise cameras designed for quantitative scientific imaging
optical spectroscopy applications. Designed utilizing PI’s exclusive XP cooling technology, PIXIS offer thermoelectric cooling down to
-75°C, while the all-metal, hermetically sealed design, with the industry’s only lifetime vacuum guarantee, ensures maintenance free
operation. Choose either back-illuminated (B/BUV) or eXcelon versions of the 2048 x 512 CCD, for optimized performance from
UV to NIR. High QE and ultra low-noise electronics make the PIXIS: 2K series of cameras ideal for demanding, low light applications
such as Raman spectroscopy, LIBS, absorbance, emission, fluorescence and reflectance spectroscopy. Princeton Instruments’ eXcelon
technology delivers the highest sensitivity in the NIR while suppressing the etaloning that occurs in standard back illuminated CCDs.
Dual speed operation at 100kHz or 2MHz allow these cameras to be used for steady state as well as fast kinetics studies.
FEATURE
BENEFITS
eXcelon® technology
Increases detector sensitivity while suppressing etalon interference fringes observed in the
NIR with conventional back-illuminated devices.
Permanent vacuum
Our all-metal brazed seals eliminate the outgassing that occurs with epoxy seals and allow
us to offer the industry’s only lifetime vacuum guarantee.
Deep thermoelectric cooling
Low dark noise allows detection of faint signals; CoolCUBE II, a compact room temperature
coolant circulator, is available for vibration sensitive environments.
Single fused silica vacuum window
Minimizes reflection losses from the UV to the NIR;
Optional AR coating and wedge windows are available.
Optional UV phosphor coating
Enhances sensitivity throughout the UV to below 200 nm.
2048 x 512 imaging array,
13.5mm x 13.5mm pixels
CCD format provides 6.9 mm height for rapid spectral acquisition; 13.5 mm pixel size
offers superior resolution to industry-standard detectors with a 1024 pixel format.
100 kHz and 2 MHz digitization rates
Dual-speed digitization allows complete freedom to select between slow operation for low
noise and highest SNR or fast operation for rapid image acquisition.
Dual amplifiers with software-selectable system gains
High sensitivity amplifier reduces noise for weak signals;
High capacity amplifier increases dynamic range.
USB2.0 data interface
Plug-and-play operation with desktops or laptops;
Optional fiber optic interface for remote operation; Easy OEM integration.
Optional: LightField® (for Windows 8/7, 64-bit)
Or WinView/Spec (for Windows 8/7/XP, 32-bit)
Flexible software packages for data acquisition, display and analysis;
LightField offers intuitive, cutting edge user interface, IntelliCal® and more.
PICAM (64-bit) / PVCAM (32-bit)
software development kits (SDKs)
Compatible with Windows 8/7/XP, and Linux;
Universal programming interfaces for easy custom programming.
Applications:
Raman Spectroscopy, LIBS, Absorbance,
Emission, Fluorescence and Reflectance
Spectroscopy
Page 1 of 6
PIXIS: 2K eXcelon Rev. N3.1
SPECTROSCOPY GROUP
SPECIFICATIONS
Features
Dark current @ -75°C (e-/p/sec)
PIXIS: 2KB_eXcelon
PIXIS: 2KB/BUV
Back-illuminated CCD with eXcelon technology.
Enhanced sensitivity in the UV and the NIR with
low etaloning.
Back-illuminated CCD.
High sensitivity in both the visible and UV regions.
Special BUV version offers the highest sensitivity
in the UV region.
0.001 (typical)
0.001 (typical)
CCD format
2048 x 512 imaging pixels; 13.5 x 13.5-mm pixels with 100% fill factor
Imaging area
Optical mount
Deepest cooling temperature
Thermostating precision
Cooling method
Spectrometric well capacity:
High Sensitivity
High Capacity
ADC speed/bits
System read noise @100 kHz
@2 MHz
Vertical shift speed
27.6 x 6.9 mm (optically centered)
Princeton Instruments Acton spectrometer adapter with optional shutter
-75°C guaranteed
±0.05°C
Thermoelectric air or liquid cooling (CoolCUBE II required)
Non-linearity
Software selectable gains
Operating systems supported
Data interface
I/O signals
<1% @ 100 kHz
1.5, 3, 6 e- (high sensitivity); 3, 6, 12 e- (high capacity); available at all speeds
Windows 8/7/XP (32-bit), Windows 8/7 (64-bit) and Linux
USB2.0
Operating environment
Certification
Dimensions / Weight
+5 to +30°C non-condensing
CE
16.59 cm (6.53") x 11.81 cm (4.65") x 11.38 cm (4.48") (L x W x H) / 2.27kg (5lbs)
250 ke- (typical), 150 ke- (min)
750 ke- (typical), 600 ke- (min)
100kHz/16-bit and 2MHz/16-bit
3.5 e- rms (typical), 6 e- rms (max)
14 e- rms (typical), 20 e- rms (max)
15.2 msec/row
Two MCX to BNC connectors for programmable frame readout, shutter, trigger in
All specifications are subject to change.
SPECTRAL RATE
@ 100 kHz Full Vertical Binning (FVB)
@ 2 MHz Full Vertical Binning (FVB)
35 spectra/sec
90 spectra/sec
Page 2 of 6
PIXIS: 2K eXcelon Rev. N3.1
SPECTROSCOPY GROUP
QE DATA
100%
90%
80%
Quantum Efficiency (QE)
70%
60%
50%
40%
30%
20%
10%
0%
250
300
350
400
450
500
2KB_eXcelon
550
600
650
2KB
700
750
800
850
900
950
1000
1050
With optional UV coating
2KBUV
Wavelength (nm)
NOTE:
Graph shows typical Quantum Efficiency (QE) data measured at + 25°C. QE decreases at normal operating temperatures. For the best results
for your application, please discuss the specific parameters of your experiment with your sales representative.
VACUUM WINDOW AR COATINGS
100
95
NOTES:
 Standard anti-reflection (AR) coating options shown on graph
 Designed by Acton Optics, our BBAR coating offers unmatched
performance for 400 nm - 1100 nm
 Custom wedge window options and other AR coatings are also
available
Contact your local sales representative for more information
Transmission (%)
90
85
80
NIR -AR
75
NO -AR
70
MgF2
65
BBAR (400 -1100nm)
60
55
50
200
300
400
500
600
700
800
900
1000
Wavelength (nm)
Page 3 of 6
PIXIS: 2K eXcelon Rev. N3.1
SPECTROSCOPY GROUP
eXcelon Performance
Etalon Oscillations (B_eXcelon vs. B)
QE Improvement (B_eXcelon vs. B)
25
0.15
0.1
15
Relative Amplitude
QE Improvement (B_eXcelon - B)
20
10
5
0.05
0
-0.05
-0.1
0
-0.15
-5
330
370
410
450
490
530
QE Data measured at +25˚C
570
610
650
690
730
770
810
850
890
930
Wavelength (nm)
B_eXcelon provides superior QE over the standard
back illuminated (“B”) version in the UV-NIR range.
970
1010 1050 1090
600
650
700
750
800
eXcelon
B/I
850
900
950
1000
1050
1100
Wavelength (nm)
B_eXcelon provides significantly lower etaloning (unwanted
fringes) compared to standard back illuminated (“B”) version.
Data taken with white light source through a monochromator, comparing etaloning performance of eXcelon vs. back-illuminated CCDs.
Page 4 of 6
PIXIS: 2K eXcelon Rev. N3.1
SPECTROSCOPY GROUP
PIXIS DRAWINGS
(AIR COOLED)
PIXIS with
Spectroscopy Mount
/ 2.27 KG
PIXIS with
Spectroscopy Mount;
with Shutter
/ 2.27 KG
Page 5 of 6
PIXIS: 2K eXcelon Rev. N3.1
SPECTROSCOPY GROUP
PIXIS DRAWINGS
(LIQUID COOLED)
COOLANT INLET/OUT QUICK
DISCONNECT FITTINGS
0.203
2 MCX
CONNECTORS
0
0.358
0.23
5.77
0.23
5.77
35°
TYP.
17.5°
TYP.
USB 2.0
CONNECTOR
3.600 B.C.
1.851
47.02
0.821
20.85
120° TYP.
0
0
0.450
11.43
0.500
12.70
4.65
2 PIN LEMO
CONNECTOR
SHUTTER
1/4-20 UNC
9 PIN DIN
CONNECTOR
POWER
0.593
15.06
0
0.250
6.35
1.780
45.21
6.41
162.9
5.09
129.2
4.24
107.6
0.310
MOUNTING SURFACE
TO FRONT SURFACE
OF WINDOW
0.600
15.2
CCD TO MOUNT
SURFACE
0.06
1.6
0
0
2.14
54.4
0
1.45
36.8
NOTES:
1.
WEIGHT 5.5 lbs / 2.49 kg
2.
POWER DISSIPATION 110 WATTS AT FULL POWER
3.
EXTERNAL WATER CIRCULATOR COOL CUBE II 7567-0002
www.princetoninstruments.com [email protected]
TOLL-FREE +1.877.474.2286 | PHONE +1 609.587.9797
Visit the Princeton Instruments website to find a representative in your area.
Page 6 of 6
PIXIS: 2K eXcelon Rev. N3.1