Document 6533850
Transcription
Document 6533850
THE EFFECT OF ALKALINE SAMPLE SOLUTIONS ON THE STABILITY OF LIQUID SCINTILLATION COCKTAILS PREPARED WITH COMMERCIALLY AVAILABLE SCINTILLANTS: ASSAYS OF 188W-188Re Jeffrey T Cessna1 Brian E Zimmerman Physics Laboratory, National Institute of Standards and Technology, 100 Bureau Dr., Stop 8462, Gaithersburg, MD 20899, USA ABSTRACT. The effect of alkalinity on the efficiency-traced activity determinations of liquid scintillation (LS) cocktails containing 188W-188Re has been investigated for several commercial scintillator formulations using the CIEMATINIST method. Four commercial scintillants were used. Packard Ultima Gold AB (UG), Packard Hionic Fluor (HF), Packard InstaGel XF (IG), and Wallac OptiPhase Hi-Safe III (HS). The aqueous fractions of the cocktails were adjusted by the addition of between 0.02 g and 1 g of 1.0 molLl NaOH to nominally 10 mL of scintillant, with the exception of the IG cocktails, which contained 7 mL of scintillant and 5 mL of either water or 1 molL NaOH. The quenching ranges in the cocktails were adjusted by the addition of a 10 percent, by volume, dilution of nitromethane in ethanol. For efficiency tracing, additional series of samples were prepared with tritiated water in such a way as to be chemically identical with the 188W-'88Re samples. The results indicate that the Insta-Gel cocktails prepared with NaOH as the primary aqueous component decomposed, resulting in a separated milky phase, but that those prepared with water as the aqueous component were stable. In addition, Ultima Gold AB was shown to be unstable in the presence of ' 88W_188Re and gave widely varying results, depending upon the aqueous (NaOH) fraction of the cocktail. The Hionic Fluor and Hi-Safe III cocktails were all stable over time and gave good results, despite the fact there was an observed difference of about 0.4% in the averages of the massic activities determined with each cocktail. The Insta-Gel cocktails prepared with water gave activity values that were in excellent agreement with those obtained with Hi-Safe III. 1 INTRODUCTION One of the most important radionuclides emerging in the field of nuclear medicine is 188Re (IznagaEscobar 1998; Zimmerman et al. 1999; Arteaga de Murphy et al. 2001), which is conveniently obtained from a generator prepared from the parent nuclide, 188W (Knapp et al. 1997). Because both 88j and 188Re are 3-emitters, the method of choice for determining the amount of radioactivity contained in an equilibrium solution of these nuclides is liquid scintillation (LS) counting. Normally observed efficiencies would be over 99% for the 188Re and slightly more than 94% for the 188W. The chemical properties of tungsten dictate that the pH of a solution intended to be stable over long periods of time be kept alkaline to prevent the tungstate from forming insoluble tungstic acid. Most commercially available scintillation fluids are developed for neutral or acidic samples with moderate ion concentrations. Unfortunately, information as to the ability of these scintillants to handle alkaline samples with moderate to high ion concentration is not widely available. This paper presents the results of recent experiments designed to determine the best cocktail composition, including scintillant, to assay solutions of 188W in secular equilibrium with its daughter, 188Re. l A variety of cocktail compositions using commercial scintillants were examined. The choice of scintillant was based on either the manufacturer's recommendation of their ability to handle basic solutions, where available (Packard Instrument Company 1997), or previous experience with the scintillant. For the non-gel forming scintillants-UG, HF, and HS-the effect of aqueous fraction was studied. The aqueous fraction was adjusted using an alkaline solution resembling the sample composition, based on our routine practice of employing an analogous system when counting low pH solutions. In the case of the Insta-Gel, we examined whether it was better to add water or an alkaline solution to produce the gel phase. An aim of the experiment was to determine stable cocktail compositions in more than one scintillant, as it is NIST practice to make massic activity determina- 1Corresponding author. Email; [email protected]. © 2002 by the Arizona Board of Regents on behalf of the University of Arizona LSC 2001, Advances in Liquid Scintillation Spectrometry Edited by Siegurd Mobius, John Noakes, Franz Schbnhofer. Pages 159-168. 159 160 JCessna, B Zimmerman tions by LS counting in more than one scintillant, utilizing more than one spectrometer, with the intention to identify possible bias from either of these sources. METHODS Approximately 2 mL of solution containing 188W-' 88Re in NaOH (in all instances, 1 mol.L-1) with an activity concentration of nominally 925 MBq.mL-'(according to the manufacturer) was received from Oak Ridge National Laboratory (ORNL) approximately 4 months subsequent to target irradiation and solution preparation. This solution was received as part of a NIST calibration, details of which are described in Zimmerman et al. (2001). To this solution was added an additional 30 mL of NaOH to bring the total volume to about 32 mL. Because the values of the mass of dissolved target and amount of inactive target were not supplied by ORNL, the exact ion concentration of the stock solution is unknown. It is assumed, however, that there was an appreciable amount of inactive tungsten in the dissolved target, therefore no further material was added as carrier. A portion of this solution was used in a series of gravimetric dilutions with NaOH to bring the activity concentration to an appropriate level for LS counting. The total dilution factor between the starting and final solutions was nominally 1071. Liquid scintillation samples were prepared in two trials, the first being prepared with scintillants UG and HF. These samples correspond to series A, B, D, E, F, and G, as described in Table 1. The second sets of LS samples were prepared with the IG and HS scintillants. These are listed in Table 1 as series H, I, J, and K. In the first trial, each series contains eight individual cocktails in three sub-series, consisting of three 1887, three tritiated water, and two blank background samples. Series A, B, and D, and similarly series E, F, and G, were adjusted in aqueous fraction by the addition of zero mL, 0.5 mL, and 1.0 mL of NaOH, respectively, by motorized pipette. The cocktail quenching was varied over all sub-series by the addition of nominally 40 µL, 120 .tL, and 240 µL, of a 10%, by volume, dilution of nitromethane in ethanol, with an aspirating pycnometer. The background sub-series samples received 40 µL and 240 µL. Cocktail components were added in the following order: 10 mL scintillant; adjustment of aqueous fraction, if any; quench agent; and finally active solution. The mass of active solution dispensed with aspirating pycnometer was determined gravimetrically. Cocktails were not mixed until all components had been added. All cocktails were prepared in conventional 20-mL glass vials, so that any visible changes or separation of the cocktail could be monitored. All cocktails were sequentially counted for nine or ten cycles each, in three spectrometers. The spectrometers, in the order used, were a Wallac Guardian model 1414, a Beckman model 78000, and a Packard Tri-Garb model A2500. Counting times were either 600 or 900 s, with sample count rates to 3,500 1. ranging from 800 s 1 s After appropriate background subtraction and decay correction, data were first considered as count rates and then were converted to activity by the CIEMAT/NIST method of tritium efficiency tracing (Coursey et al. 1986; Zimmerman and Colle 1997). NIST uses a modified version of the program EFFY4, which is an updated version of EFFY2 (Garcia-Torano and Grau Malonda 1985), to calculate the efficiency versus an independent figure of merit for the tritium standard and the radionuclide of unknown activity, over a range of quenching. By observing the experimental change in efficiency versus the change in quench indicating parameter for the tritium standard, this method is used to calculate the expected change in efficiency based on the theoretical spectrum for the tritium and that for the radionuclide of interest. The input values used for the EFFY4 program are listed in Table 2. The half live values used for decay corrections were 69.783 days ± 0.048 days for 188W-'88Re (in secular Table Description of LS cocktail compositions. Series identifiers are assigned to cocktails of similar aqueous fraction with subscripts w, T and B referring to 188W-188Re tritium and background, respectivelY. All added components are listed. Aqueous fraction was adjusted - NaOH or with either 1 molL1 o water, where «sample» refers to the amount listed due to added activity. The quenching agent used was a 10/o by volume, dilution of nitromethane in ethanol and the volumes are nominal values based on an assumed volume of 20 µL per dispensed drop. Key: Key: UG-Packard Ultima Gold AB; HF-Packard Hionic-Fluor IG-Packard Insta-Gel XF HS-Wallac 0PtiPhase Hi-Safe III. 1 Number 188W Scintillant Scintillant volume (mL) 3 UG 10 0.03 AT 3 UG 10 BW 3 UG 10 BT 3 UG 10 DW 3 UG 10 DT 3 UG 10 AB 2 UG 10 BB 2 UG 10 DB 2 UG 10 EW 3 HF 10 ET 3 HF 10 FW 3 HF 10 FT 3 HF 10 Series of ID samples AW 1 (g) solution 3H solution (g) - 0.02 0.03 0.02 0.03 - 0.02 - 0.03 0.03 0.02 0.02 molL-' NaOH (mL) sample 0.5 0.5 H20 0.5 1.0 1.0 1.0 0.5 0.5 0.002 40,120,240 - 0.002 40,120,240 0.047 40,120,240 sample 0.047 40,120,240 0.088 40,120,240 sample 0.088 40,120,240 - 0.000 40,240 0.045 40,240 0.045 40,240 0.088 40,120,240 0.088 40,120,240 0.047 40,120,240 sample 0.047 40,120,240 sample 1.0 1.0 f Quenching agent (µL) Aqueous sample 1 Description of LS cocktail compositions. Series identifiers are assigned to cocktails of similar aqueous fraction, with subscripts T and B referring to i88W-188Re tritium and background, respectively. All added components are listed. Aqueous fraction was adjusted W> o w either 1 mo1 L1- NaOH or water, where «sample» refers to the amount listed due to added activity. The quenching agent used was a 10/o with , by volume, dilution ofnitromethane in ethanol and the volumes are nominal values based on an assumed volume of 20 µL Per dispensed drop. Table Key UG-Packard Ultima Gold AB; HF-Packard Hionic-Fluor IG-Packard Insta-Gel XF; HS-Wallac 0ptiPhase Hi-Safe III. (Continued). : Number Scintillant volume Series ID of samples Scintillant GW 3 HF 10 GT 3 HF 10 EB 2 HF 10 FB 2 HF 10 GB 2 HF 10 HW 3 IG 7 HT 3 IG 7 IW 3 IG 7 IT 3 IG 7 Ha 2 1G 7 IB 2 IG 7 JW 3 HS 10 JT 3 HS 10 1 solution g 0.03 - 0.02 - 0.02 0.02 mol L -1 solution f g 0.02 - 0.02 0.02 1.0 1.0 0.5 1.0 5.0 - 0.02 sample - - 0.088 40,120,240 0.088 40,120,240 0.000 40,240 0.045 40,240 0.086 40,240 0.403 40,120,240 5.0 sample 0.403 40,120,240 sample 5.0 0.403 40,120,240 5.0 0.417 40,120,240 0.403 40,240 0.417 40,240 0.002 40,120,240 0.002 40,120,240 - 5.0 - - sample - 5.0 sample f Table 1 Description of LS cocktail compositions. Series identifiers are assigned to cocktails of similar aqueous fraction, with subscripts w T and B referring to 188 W- 88 Re tritium and background, respectively. Al1 added components are listed. Aqueous fraction was adjusted with either 1 molLi NaOH or water, where samPle refers to the amount listed due to added activity. The quenching agent used was a 10/o o by volume, dilution of nitromethane in ethanol and the volumes are nominal values based on an assumed volume of 20 µL per dispensed drop. Key: UG-Packard Ultima Gold AB; HF-Packard Hionic-Fluor IG-Packard Insta-Gel XF HS-Wallac 0PtiPhase Hi-Safe III. (Continued). 1 _ Number 'gW solution 3H solution Scintillant Scintillant volume (mL) (g) (g) 3 HS 10 0.02 KT 3 HS 10 JB 2 HS 10 KB 2 HS 10 Series ID of samples KW - - 0.02 1 molL-1 NaOH (mL) 1.0 1.0 1.0 Aqueous Quenching agent H2O f (µL) - 0.088 40,120,240 sample 0.088 40,120,240 0.000 40,240 0.091 40,240 - 164 J Cessna, B Zimmerman equilibrium) (Unterweger 2001) and 4500 days ± 8 days for 3H (Lucas and Unterweger 2000). The solution used for the 3H samples is a dilution of a NIST standard reference material (NIST 1991). Table 2 Summary of nuclear data input parameters for efficiency calculations with EFFY4. Data are from the National Nuclear Data Center (ENSDF 2000). All transitions were considered to be allowed with the exception of that marked with an asterisk, which was taken to be first forbidden-unique. 188W 188Re Value 3H Ei,max/keV P13l% ER,max/keV PR/% ER,maxlkeV P131% 2118 70.6 349 99.0 18.591 100 1962 26.0 285* 0.15 1487.2 1.68 58.3 0.83 1033.8 0.64 657.7 0.45 354.9 0.185 179.2 0.104 The samples in the second trial were prepared in the same manner as above, with the deletion of the intermediate water fraction series. Because the supplier underestimated the stock solution activity, the amount of added'88W_188Re was reduced. The IG samples were prepared in a similar fashion as those described above, except that the amount of scintillant was 7 mL and the aqueous fraction increased to 5 mL of either NaOH or water. This amount was added to ensure that the cocktails were in the gel phase. The samples were counted sequentially for ten cycles each, in the Wallac and the Beckman LS s-1 to counters. The counting times were 600 seconds, with sample count rates ranging from 1000 s-1, described as processed data were excluding samples which were not stable. The resulting 2000 above. RESULTS AND DISCUSSION The first result was the inability of UG to accurately produce stable traced 188W-188Re activity values in cocktails containing either 0.5 mL or 1.0 mL of NaOH. As seen in Figure 1, series B and D showed a pronounced decrease over time in traced activity values, compared to series A. Series A was found to be 1.4 percent lower than the average of all cocktails deemed to be stable, and described below. The average standard deviation of 10 determinations of the 188W-188Re massic activity values for series A is 0.27% for the Wallac, 0.23% for the Beckman, and 0.21% for the Packard. These values are more than double typical values. Also shown in Figure 1 are the tritium efficiencies for the same cocktail composition series. It is interesting to note that while there is a small decrease in the efficiency for series D, these samples do not show a trend large enough to account for the change in 188W-188Re efficiency. There would be only a 0.08% change in computed'88W_188Re efficiency for a 1 percent change in 3H efficiency. Alkaline Sample Solutions and the Stability ofLiquid Scintillation Cocktails 165 The poor 188W-188Re results can also not be accounted for by the quench indicating parameter. This remained relatively constant in both the 188W-188Re and tritium samples over the cycles shown in Figure 1. Because the 3H samples in series A, B, and D were relatively stable over time, it is likely that the instability in the 188W-188Re samples can be attributed to the 188W-188Re chemical species in the cocktail and not necessarily the amount of NaOH. More studies would be required to determine the conditions under which Ultima Gold AB can be used for the assay of this radionuclide. The samples showed similarly poor results in the Beckman and Packard spectrometers. Although the samples were agitated when switched from one spectrometer to the next, the samples only experienced a few percent increase in count rate. The majority of the samples returned to the low counting efficiency for'88W_188Re by the second cycle. It is a routine practice in the application of an efficiency-tracing method, such as CIEMAT/NIST, to vary the quenching range by the addition of a quenching agent such as nitromethane to a series of LS cocktails. An unexpected, but explainable, result in this study was the formation of a yellow solution in cocktails that contained both NaOH and nitromethane. Investigation revealed the likely cause to be the formation of a stable aci- tautomer of nitromethane, with a characteristic yellow color, which occurs when nitromethane is in the presence of an alkaline solution, and which is in equilibrium with the nitro- anion (could 1959). It is not clear what effect, if any, this had on the stability of the UG cocktails, but the color made it possible to determine that this solution eventually separated to a different phase and settled to the bottom of the scintillation vial. There exists the possibility for color quenching and it is known that the aci- form of nitromethane is characterized by having a broad absorption with 2max = 280 nm ± 5 nm (Asmus and Taub 1968). The Insta-Gel cocktails prepared with NaOH as the primary aqueous component, series H, decomposed resulting in a milky solution, which eventually separated to the bottom of the vial. The measured 3H efficiencies of these samples increased over time by between 300/ and 49%, for the 10 cycles of the Wallac spectrometer. When the vials were agitated and moved to the second spectrometer, Beckman, the same behavior was observed. During the same counting intervals, the quench indicating parameters indicated that the amount of quench was decreasing. These effects might be expected if 3H remains in the clear layer of the vial as the milky layer settles. The composition of what the milky substance is not evident, but it is possible that a reduction reaction between the NaOH and the ethoxy alkylphenol component of the scintillant is forming organic salts. Due to the instability in both the 3H and the 188W-188Re samples, efficiency-tracing calculations were not carried out for series H. Cocktails prepared with Insta-Gel and with water as the primary aqueous component, series I, were stable with regard to both 3H efficiency and quench indicating parameter. The 188W-188Re samples also remained stable. This is most evident in the average standard deviation of 10 determinations of the 188W-188Re activities for each sample, in each of two spectrometers, which was found to be 0.09%. The Hionic Fluor and Hi-Safe III cocktail series were all stable over time and gave good results. Once again this can be seen in the average standard deviation of 10 determinations of the 188W_ 188Re activities for each sample, in each of two spectrometers. These are 0.08% for the HF cocktails and 0.09% for the HS cocktails. Overall results from all of the traced cocktails are shown in Figure 2. Not readily apparent in the plot, due to the large spread of UG results, is that the Hionic Fluor results, while stable, were 0.4% lower than those found with Hi-Safe III and the stable Insta-Gel samples. An F-test performed on these two populations found the difference to be significant at a confidence level of 95%, giving an F-value of 13.0, compared to a critical value of 4.7 for 14 166 JCessna, B Zimmerman 140000 130000 120000 110000 100000 U 90000 80000 70000 60000 -3.0 -2.0 -1.0 0.0 3.0 2.0 1.0 Tid 0.55 r -.-AT-1 -- AT-2 AT-3 BT-1 --BT-2 --.--BT-3 --DT-1' 0.25 0.20 --------- ----3.0 - ------- -2.0 -1.0 0.0 -e-DT-2'i - 1.0 ------ - --2.0 3.0 T/d Figure 1 Results from series A, B, and D from the Wallac spectrometer. Figure 1 a shows CA tot, total efficiency-traced 188W-'88Re activity concentrations in Bqg', as a function of measurement time in days. Figure lb shows the measured tritium efficiency as a function of measurement time, T, in days. All samples contain 10 mL of Ultima Gold AB. Series A contains only NaOH from the active solution. Series B and D contain 0.5 mL and 1 mL of NaOH per sample, respectively. Each data point represents a calculational result from a single measurement. Lines are included as a guide for the eye. All data are decay corrected to a common reference time of T = 0 days. degrees of freedom. This cocktail dependence was therefore treated as a component of uncertainty. The final activity values were comprised of series E, F, G, I, J, and K. Uncertainties evaluated in the determination of the activity of the solution in these measurements are fully explained in Zimmerman et al. (2001). Alkaline Sample Solutions and the Stability ofLiquid Scintillation Cocktails 167 140000 Packard Beckman Wallac ... 130000 ... 110000 90000 I I 60000 A B D E F . .. IG HF UG 70000 .. . .. 120000 50000 . I ]E G A B D E F G A B D E F G J K J K Series Figure 2 Overall results from series A, B, D, E, F, G, I, J, and K, from all spectrometers. Values for each series and spectrometer are the average of CA, total efficiency-traced 188W-ittRe activity concentrations in Bqg ', over normally 10 cycles for each of 3 samples. Uncertainty bars are one standard deviation of the mean. Series A B, and C contain 10 mL of Ultima Gold AB (UG) and 0.02 mL, 0.5 mL, and 1 mL of NaOH, respectively. Series E, F, and G contain 10 mL of Hionic Fluor (HF) and 0.02 mL, 0.5 mL, and 1 mL of NaOH, respectively. Series J and K contain 10 mL of OptiPhase Hi-Safe III (HS) and 0.02 mL and 1 mL of NaOH, respectively. Series I contains 7 mL of InstaGel XF (IG) and 5 mL of water. Series H is not plotted, as the cocktails were deemed too unstable for meaningful efficiency tracing. CONCLUSION The present results indicate that the Insta-Gel cocktails prepared with NaOH as the primary aqueous component decomposed, resulting in a white precipitate, but that those prepared with water as the aqueous component were stable. In addition, Ultima Gold AB was shown to be unstable in the presence of '88W-lt8Re and gave widely varying results, depending upon the magnitude of aqueous (NaOH) fraction of the cocktail. The Hionic Fluor and Hi-Safe III cocktails were all stable over time and gave good results, despite the fact there was an observed difference of about 0.4% in the averages of the activities determined for each cocktail. The Insta-Gel cocktails prepared with water gave activity values that were in excellent agreement with those obtained with Hi-Safe III. Finally, although there is no direct evidence of a detrimental effect on the LS counting one should be aware of the formation of aci- tautomers of nitromethane, when used to vary quenching in the presence of alkaline solutions, recognized by a yellow color. DISCLAIMER Certain commercial equipment, instruments, or materials are identified in this report to foster understanding. Such identification does not imply recommendation by the National Institute of Standards and Technology, nor does it imply that the materials or equipment are necessarily the best available for the purpose. ACKNOWLEDGMENTS The authors would like to thank Dr FF Knapp of Oak Ridge National Laboratory for supplying the 188W-I88Re solution and Dr R Colle for many useful discussions on the topic of LS counting. 168 J Cessna, B Zimmerman REFERENCES Arteaga de Murphy C, Ferro-Flores G, Pedraza-Lopez M, Melendez-Alafort L, Croft BY, Ramirez FM, Padilla J. 2001. Labelling of Re-ABP with' ggRe for bone pain palliation. Applied Radiation and Isotopes 54(3): 435-42. Asmus KD and Taub IA. 1968. Spectrum and kinetics of the hydroxynitromethane anion radical in pulse-irradiated alkaline nitromethane solutions. The Journal of Physical Chemistry 72(10):3382-7. Coursey BM, Mann WB, Grau Malonda A, GarciaTorai o E, Los Arcos JM, Gibson JAB, Reher D.1986. Standardization of Carbon-14 by 4i3 liquid scintillation efficiency tracing with hydrogen-3. Applied Radiation and Isotopes 5:403-8. ENSDF 2000. Data extracted using the NNDC on-line data service from the ENSDF database as of June, 2000. M. R. Bhat, Evaluated Nuclear Structure Data File (ENSDF). In: Qaim SM, editor. Nuclear data for science and technology. Berlin: Springer-Verlag. p. 817. Garcia-Torafo E, Grau Malonda A. 1985. EFFY, a new program to compute the counting efficiency of Beta particles in liquid scintillators. Computer Physics Communications 36:307-12. Gould ES. 1959. Mechanism and structure in organic chemistry. Holt, Rinehart, and Winston. p. 366. Iznaga-Escobar N. 1998.'88Re-direct labeling of monoclonal antibodies for radioimmunotherapy of solid tumors: biodistribution, normal organ dosimetry, and toxicology. Nuclear Medicine and Biology 25(5):4417. Knapp FF, Beets AL, Guhlke S, Zamora P0, Bender H, Palmedo H, Biersack HJ. 1997. Availability of rhenium-188 from the alumina-based tungsten-188/rhenium-188 generator for preparation of rhenium-188labeled radiopharmaceuticals for cancer treatment. Anticancer Research 17(3B):1783-95. Lucas LL and Unterweger MP. 2000. Comprehensive review and critical evaluation of the half-life of tritium. Journal of 'Research of the National Institute of Standards Technology 105:541-9. National Institute of Standards and Technology. 1991. Standard Reference Material 4927E, Radioactivity Standard, Hydrogen-3. Gaithersburg, Maryland. Packard Instrument Company. 1997. A catalog of chemicals and supplies for life science research. Unterweger MP. 2001. Half-life measurements results at the National Institute of Standards and Technology. Forthcoming. Zimmerman BE, Cessna JT, Unterweger MP, Li AN, Whiting JS, Knapp FF. 1999. A new experimental determination of the dose calibrator setting for ' ggRe. Journal of Nuclear Medicine 40(9):1508-16. Zimmerman BE, Colle R. 1997. Standardization of 63Ni by 4it3 liquid scintillation spectrometry with 3H-standard efficiency tracing. Journal o fResearch of the Nationallnstitute ofStandards and Technology 102:45577.