AU680 Sample Rack Usage

Transcription

AU680 Sample Rack Usage
AU680 Sample Rack Usage
AU680
Racks
Racks are identified by a rack barcode label. The barcode label identifies
the type of sample (serum, urine, other-1, other-2 or whole blood)
designated for that rack. Sample racks are configured in the AU
software.
Racks are available in multiple colors to assist the operator in determining
rack function. Rack function is identified by the system by the placement
of magnets on the bottom of the rack.
Reagent Blanks (Blue)
•
Use any validated container to
place DI Water in position 1
Calibration (Yellow)
•
Multiple racks may be required for
multiple chemistries
•
Use any validated container for
calibrators
Quality Control (Green)
•
Multiple racks may be required
•
Use any validated container for
QC material
Note: Calibrators and QC may be barcoded or assigned to a rack and position.
AU680 Sample Rack Usage and Validated Sample Containers, Customer
Version A, 072111
Page 1 of 3
Brea Training Center
Patient Samples
•
•
•
AU680
Racks
(Cont’d)
Patient samples can be placed in white, red or orange racks
Barcode Mode: Barcoded samples may be placed in any position
Sequential Mode: Samples without barcodes must be placed in
sequential order. Do not leave empty spaces in racks when using non
barcoded samples
Patient Samples (White)
•
Each sample type must be placed in
racks designated for that sample
type
− Routine Serum
− Routine Urine
− Routine Other 1
− Routine Other 2
− Routine Whole Blood
•
Any validated container can be used
within a rack
•
Used with the Automatic Repeat
feature
Extra (Red)
•
Extra Samples: used to designate a
different sample category (user
defined)
•
Each sample type must be placed in
racks designated for that sample
type
Manual Repeats (Orange)
•
If using the Manual Repeat feature,
you must place the sample in the
rack designated for that sample type
•
Any validated sample container may
be used in an orange rack
AU680 Sample Rack Usage and Validated Sample Containers, Customer
Version A, 072111
Page 2 of 3
Brea Training Center
AU680
Validated
Sample
Containers
Container
Placement
Description
Part Number
Photo
• Outside diameter:
Primary
Tubes
Rack/
STAT
table
11.5-16 mm
• Inside diameter:
9-15 mm
• Height :
55-102 mm
N/A
2.0 mL Sample Cup
(Hitachi)
MU853200
Sample Cups, Glass
ZM006200
2.0 mL Sample Cup
(Hitachi)
MU 853200
Sample Cups, Glass
ZM006200
2.0 mL Sample Cup
(Hitachi)
Dead volume = 65 uL
(Do not use as insert
on the STAT table)
MU 853200
Centrifichem Cups
Dead volume = 50 uL
Evergreen
Scientific
P/N 1270093-010
Rack
Sample
Cups
STAT
table
16 mm
primary
tube
Nested
Cups
EZ NEST Sample
Cups
(fits 16 x 100/
16 x 75 mm tubes)
12/13 mm
primary
tube
1270016000
EZ NEST Sample
Cups
( fits 13 x 100/
1270013000
13 x 75 mm tubes)
Dead volume = 50 uL
AU680 Sample Rack Usage and Validated Sample Containers, Customer
Version A, 072111
Page 3 of 3