Type UB1 Load Cell Lastcell UB1

Transcription

Type UB1 Load Cell Lastcell UB1
FLINTEC
...the right weigh
Lastcell UB1
Type UB1 Load Cell
En produkt från Flintec
IP 68
Flintec load cells are designed to meet the most stringent
accuracy requirements. Certifications have been obtained
from Weights & Measures Authorities, worldwide.
UB1 universal load cells are available in the capacities 10 kN to
50 kN (1020 kg to 5099 kg) and include Accuracy
Classifications GP, C1 and C3 according to OIML R 60.
They offer total stainless steel construction and complete
hermetic sealing, making them suitable for use in the toughest
industrial environments.
The Flintec calibration technique (in mV/V/) eliminates time
consuming corner calibration in multiple load cell systems.
The UB1 is available for use in hazardous areas zone 1, 2 (gas)
and 21, 22 (dust) according to EEx ia IIC T6...T4 T130°C ATEX.
www.flintab.se
Kabelvägen 4 • 553 02 Jönköping
Tfn 036-31 42 00 • Fax 036-18 50 79
E 10
[email protected]
11/02
1/2
www.flintec.net
APPROVED
Important Features
• Capacities: 10 kN to 50 kN.
• High accuracy.
• Total stainless steel construction.
• Complete hermetic sealing.
• Protection IP 68.
• High input resistance: 1100 .
• W&M certified for 3000 intervals.
• Calibration in mV/V/.
• Easy cable replacement.
• Complete range of loading hardware
available.
• Factory mutual approved.
Option
• Explosion protection zone 1, 2, 21 and 22
ATEX.
Lastcell UB1
FLINTEC
...the right weigh
Maximum capacity (=Emax )
Metric equivalents (1 N=0.10197 kg)
Rated Output (=RO)
Calibration in mV/V/ (A...I classified)
Accuracy class according to OIML R 60
Maximum number of verification intervals (n)
Minimum load cell verification interval (vmin )
Combined error
Creep error (30 minutes) / DR
Temperature effect on minimum dead load output
Temperature effect on sensitivity
Excitation voltage
Zero balance
Input resistance
Output resistance
Insulation resistance (100 V DC)
Compensated temperature range
Operating temperature range
Safe load limit
Ultimate load
Load cell material
Sealing
Protection according DIN 40.050
kN
kg
mV/V
%RO
%RO
%RO
%RO/°C
%/°C
V
%RO
M
°C
°C
%Emax
%Emax
10 / 20 / 50
1020 / 2039 / 5099
2 ± 0.1%
± 0.05 ( ± 0.005)
(GP)
C1
C3
n.a.
1000
3000
n.a.
Emax /5000
Emax /5700
± 0.040
± 0.030
± 0.020
± 0.060
± 0.049
± 0.016
± 0.0040
± 0.0028
± 0.0024
± 0.0020
± 0.0016
± 0.0011
5...15
± 1.0
1106 ± 5
1000 ± 1
5000
–10...+40
–40...+80
200
300
stainless steel 17-4 PH (1.4548)
complete hermetic sealing; cable entry sealed by glass to metal header
IP 68
Dimensions
L
41
W
H
L
86
143
H
92
136
W
30
43
Thread D
M16
M242
All dimensions in mm. Dimensions and specifications are subject to change without notice.
*
E 10
11/02
2/2
Unified thread 5/8-18 UNF (10...20 kN) and 1-12 UNF (50 kN) is available.
Wiring
• The load cell is provided with a shielded,
4 conductor cable (AWG 24).
• Cable length:
6 m.
• Cable diameter: 5 mm.
• The shield is floating.
+ Input
(green)
+ Output
(white)
– Input
(black)
– Output
(red)
Shield
Huvudkontor:
Flintab AB
(yellow)
2004-04-01
Type
UB1-10k N/20 kN
UB1-50 kN
Vi förbehåller oss rätten att ändra tekniska och/eller kommersiella
delar av detta dokument utan föregående avisering.
D*
Distriktskontor:
www.flintec.net
KabelvägenFlintab
4 553
JÖNKÖPING
TelJönköping
036-31 42 00 MALMÖ
www.flintab.se 040
[email protected]
AB,02
Kabelvägen
4 • 553 02
- 18 10 30
GÖTEBORG 031 - 48 00 25
Tfn 036-31 42 00 • Fax 036-18 50 79
www.flintab.se • [email protected]
ESKILSTUNA
SUNDSVALL
016 - 13 76 20
060 - 58 05 20
STOCKHOLM 016 - 13 76 20
Utgåva 2013-04-22. Mått i mm där ej annat anges. Vi förbehåller oss rätten att ändra tekniska och/eller kommersiella delar av detta dokument utan föregående avisering.
UB1 Specifications