Agilent Technologies 8753ES Option H85

Transcription

Agilent Technologies 8753ES Option H85
Agilent Technologies
8753ES Option H85
Agilent Technologies
8753ES Option H85
User’s and Service Guide Supplement
This User’s and Service Guide Supplement is intended for use
with the following documents:
Agilent 8753ES Network Analyzer’s User’s Guide
Part Number 08753-90472
Agilent 8753ES Network Analyzer’s Reference Guide
Part Number 08753-90473
Agilent 8753ES Network Analyzer’s Service Guide
Part Number 08753-90484
Agilent 8753E High Power Application Note 1287-6,
Publication Number 5966-3319E
Agilent Part Number: 08753-90522
Printed in USA May 2000
Notice
The information contained in this document is subject to change
without notice. Agilent Technologies makes no warranty of any kind
with regard to this material, including but not limited to, the implied
warranties of merchantability and fitness for a particular purpose.
Agilent Technologies shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the
furnishing, performance, or use of this material.
Agilent Technologies assumes no responsibility for the use or reliability
of its software on equipment that is not furnished by Agilent
Technologies.
Restricted Rights Legend
Use, duplication, or disclosure by the U.S. Government is subject to
restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in
Technical Data and Computer Software clause at DFARS 252.227-7013
for DOD agencies, and subparagraphs (c)(1) and (c)(2) of the
Commercial Computer Software Restricted Rights clause at FAR
52.227-19 for other agencies.
© Copyright 2000 Agilent Technologies
iv
Warranty Information
IMPORTANT
Certification
Agilent Technologies certifies that this product met its published specifications
at the time of shipment from the factory. Agilent Technologies further certifies
that its calibration measurements are traceable to the United States National
Institute of Standards and Technology (NIST, formerly NBS), to the extent allowed
by the Institute’s calibration facility, and to the calibration facilities of other
International Standards Organization members.
NOTE
The actual warranty on your instrument depends on the date it was
ordered as well as whether or not any warranty options were purchased
at that time. To determine the exact warranty on your instrument,
contact the nearest Agilent Technologies sales or service office with the
model and serial number of your instrument. See the list of sales and
service offices in this document.
This Agilent Technologies instrument product is warranted against
defects in material and workmanship for the warranty period. During
the warranty period, Agilent Technologies will, at its option, either
repair or replace products which prove to be defective.
If the warranty covers repair or service to be performed at the Buyer’s
facility, then the service or repair will be performed at the Buyer’s
facility at no charge within Agilent’s service travel areas. Outside
Agilent’s service travel areas, warranty service will be performed at
Buyer’s facility only upon Agilent’s prior agreement, and Buyer shall
pay Agilent’s round-trip travel expenses. In all other areas, products
must be returned to a service facility designated by Agilent.
If the product is to be returned to Agilent Technologies for service or
repair, it must be returned to a service facility designated by Agilent
Technologies . Buyer shall prepay shipping charges to Agilent
Technologies and Agilent Technologies shall pay shipping charges to
return the product to Buyer. However, Buyer shall pay all shipping
charges, duties, and taxes for products returned to Agilent Technologies
from another country.
Agilent Technologies warrants that its software and firmware
designated by Agilent Technologies for use with an instrument will
execute its programming instructions when properly installed on that
instrument. Agilent Technologies does not warrant that the operation
of the instrument, or software, or firmware will be uninterrupted or
error-free.
v
Limitation of Warranty. The foregoing warranty shall not apply to
defects resulting from improper or inadequate maintenance by Buyer,
Buyer-supplied software or interfacing, unauthorized modification or
misuse, operation outside of the environmental specifications for the
product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND
EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT
BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER
BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
vi
Contents
8753ES Option H85 Overview
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1. Installation, Control, and Specifications
Installing the 8753ES Option H85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Check the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Controlling the Option H85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Menu Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Manual Control of the Option H85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Attenuator Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
R Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Phase Lock Menu Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
GPIB Control of the Option H85 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Attenuator Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Example GPIB Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
R IN Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Example GPIB Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Phase Lock Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
2. Application Concepts
Application Concepts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
High Power Amplifier Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
High Power Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Determining Power Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Additional Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Selecting Power Ranges and Attenuator Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Final Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
High Power 2 Port Configuration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
High Power Forward Configuration Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
3. Servicing the Option H85 Test Set
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Adjustment Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Performing Test 53 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
Replaceable Parts and Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Option H85 Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Checking the Option H85 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Option H85 Replacement Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Major Assemblies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Coupler Bracket Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
RF Rear Panel Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Installation and Removal Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Contents-1
Contents
Switch Cable Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Deck Assembly Removal or Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
SMA to SMA Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
LED PC Board Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Coupler Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Coupler Bracket Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Cable Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Cable Installation Source and R-Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
Coupler Bracket Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
Cable Installation R-Switch to R-Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-37
Cable Installation PORT A and PORT B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39
PORT 1 and PORT 2 SWITCH to F/R Transfer Switch Cables . . . . . . . . . . . . . . . . . . . 3-41
RF IN to F/R Transfer Switch Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-43
RF OUT to Source Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-45
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-47
RF IN to RF OUT Jumper Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-49
Front Panel Jumpers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-50
Option H85 Front View Without Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-51
Option H85 Bottom View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-52
Option H85 Top View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-53
Option H85 Front Bottom View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-54
Contents-2
8753ES Option H85 Overview
1
8753ES Option H85 Overview
Introduction
Introduction
This User’s and Service Guide Supplement contains information that is
only pertinent on the setup, operation, and service of the 8753ES
Option H85. This manual supplement is intended to be used in
conjunction with the following documents:
•
8753ES Network Analyzer’s User’s Guide,
part number 08753-90472
•
8753ES Network Analyzer’s Reference Guide,
part number 08753-90473
•
8753ES Network Analyzer’s Service Guide,
part number 08753-90484
• Using a Network Analyzer to Characterize High Power Components,
Application Note 1287-6, Publication Number 5966-3319E
IMPORTANT
Refer to the 8753ES standard documentation for information on
certification, warranty, safety and regulatory information.
2
Overview
8753ES Option H85 Overview
Overview
Overview
The 8753ES Option H85 provides a built-in, high-power capable test
set. High power measurements can be made up to 20 watts (+43 dBm)
maximum at each test port. The usable frequency range of the
Option H85 is 50 MHz to 3 GHz or 6 GHz with the standard Option 006.
Frequencies below 50 MHz are degraded due to the roll-off of the
couplers.
The front panel test ports are 3.5mm male. All other front panel
connections are 3.5mm female.
The 8753ES Option H85 uses a mechanical transfer switch (A24) to
route the RF signal to Port 1 or Port 2. The inactive port is grounded
through the transfer switch. Continuous switching is not allowed
between forward and reverse measurement states.
Access to the analyzer’s RF source is located on the rear panel of the
analyzer. These connections are 3.5mm female. This allows an RF
amplifier to be connected in the RF source path to increase test port
power.
Test port coupler access to the transfer switch is located on the front
panel. Coupler access allows the user to connect high power circulators
or isolators to protect the internal transfer switch when making high
power measurements.
A and B channel access to the samplers is also located on the front
panel. Sampler access allows the user to increase the dynamic range
and lower the noise floor. The maximum power to the A and B samplers
is +10 dB. Recommended power to the sampler is –10 dB. Maximum
power to the attenuators is 1 watt (+30 dBm).
The A and B channel attenuators are built into the Option H85 test set.
The attenuators may be set with different settings. Once the values are
set, they must remain the same for both the forward and reverse
measurements. Once set, they cannot be uncoupled to allow a channel
attenuator forward measurement to differ from reverse measurement
state value settings.
These variable mechanical attenuators range from 0 to 55 dB in 5 dB
steps. The variable attenuators are used to reduce the power to the A
and B channel samplers. The A and B channel attenuators can either be
controlled manually by the 8753ES display user softkeys or by GPIB
settings. These settings can be saved in instrument states.
Do not exceed the maximum power of 10 dBm to the A, B, and R
Samplers. If the maximum power is exceeded, damage will occur.
Overview
3
8753ES Option H85 Overview
Overview
The R IN port is located on the front panel and allows direct access to
the R–Channel. R IN is used to provide an external reference signal
used for ratioing. To ensure optimum power level at the receiver, add
any needed attenuation to reduce the power level to –10 dBm. R IN’s
external maximum power is +10 dBm.
The Option H85 default is set to the internal reference channel when
the instrument is preset or during power on.
Control of the internal reference switch is provided using the
TESTSET I/O FWD and TESTSET I/O REV display keys and GPIB
commands. A decimal value of seven is for internal and a decimal value
of six is for external.
Analyzers equipped with Option H85 can be configured to measure high
power devices. This ability is useful if the required input power for a
device under test is greater than the analyzer can provide, or if the
maximum output power from an amplifier under test exceeds the safe
input limits for a standard analyzer.
Analyzers equipped with Option H85 are capable of increasing their RF
output levels by means of inserting an external, high power booster
amplifier. As a result, up to 20 watts (+43 dBm) can be delivered to a
device or amplifier under test. The maximum test port input power is
1 watt (+30 dBm) CW, but jumpers on the front panel allow the
insertion of high power attenuators or isolators. This allows test device
output levels up to the power limits of the inserted components.
4
Overview
1
Installation, Control, and
Specifications
1-1
Installation, Control, and Specifications
Installing the 8753ES Option H85
Installing the 8753ES Option H85
The Option H85 is pre-installed in the test set at the factory. Refer to
the standard 8753ES documentation for complete installation
information.
Check the Shipment
1. After you have unpacked your Option H85, it is recommended that
you keep the original packaging materials so they can be used if your
instrument needs to be transported.
2. Check the items shipped with your Option H85 against the items
listed in Table 1-1 to make sure that you have received everything.
3. Inspect the Option H85 and all accessories for any signs of damage
that may have occurred during shipment. If any of the equipment
appears to be damaged or missing, call your nearest Agilent sales or
service office. Refer to the table at the end of this manual
supplement for the nearest office.
Table 1-1
Option H85 Accessories Supplied
Description
Part Number
Quantity
User’s and Service Guide Supplement
08753-90522
1
Application Note
5966-3319E
1
1-2
Chapter 1
Installation, Control, and Specifications
Controlling the Option H85
Controlling the Option H85
The following information describes the manual control and the GPIB
control of the Option H85.
Menu Keys
The following menu keys are only available on the Option H85. The
menu keys are identified as follows:
Hardkey – A key located on the front panel and identified by the label
on the key. A hardkey is shown as a bolded word with an initial capital
letter in this document. For example: Configure Menu.
Softkey – A key located to the side of the instrument display and
identified by the wording on the display. A softkey is shown as a bolded
word in all uppercase letters in this document. For example: TTL OUT.
Manual Control of the Option H85
Attenuator Control
The following key entries will control the A and B channel attenuators.
Press System, CONFIGURE MENU, OPTION Hxx, CONFIG.
Press ATTENUATOR A or ATTENUATOR B.
Press the desired decimal value [D], attenuation X 1.
R Control
Press Seq, TTL I/O, TTL OUT, TESTSET I/O FWD [D], X 1
Press I/O REV, [D], X 1
Press desired decimal value [D].
6 = External
7 = Internal
Chapter 1
1-3
Installation, Control, and Specifications
Controlling the Option H85
Phase Lock Menu Keys
1. Press System, SERVICE MENU, SERVICE MODES, PPL AUTO OFF or
2. Press System, INSTRUMENT MODE, FREQ OFFS MENU, FREQ OFFS ON .
With the frequency offset mode on (FREQS OFF ON), the phase lock of
the Option H85 is turned off.
NOTE
Banded high power solutions require that the Phase Lock be turned OFF
or an error message will be displayed on the LCD display. Phase lock
should be established once a day or before calibration in order to
maintain the analyzer’s integrity.
1-4
Chapter 1
Installation, Control, and Specifications
GPIB Control of the Option H85
GPIB Control of the Option H85
General Purpose Interface Bus (GPIB) and Hewlett-Packard Interface
Bus (HP-IB) are different names for the same command structure.
Currently, “HP-IB” is changing to “GPIB”, so some instruments may be
labeled “HP-IB” while others are labeled “GPIB”. The format and
command structure is not affected by this name change.
Attenuator Control
The following GPIB commands will control the A and B channel
attenuators:
ATTA[D]
ATTB[D]
Example GPIB Command
OUTPUT 716;”ATTA10”!SET A atten to 10 dB
OUTPUT 716;”ATTA20”!SET B atten to 20 dB
R IN Control
The following GPIB commands will control the R-Channel switch:
TSTIOFWD[D]
TSTIOREV[D]
Example GPIB Command
OUTPUT 716;”TSTIOFWD6”;”TSTIOREV6”;!External Control
OUTPUT 716;”TSTIOFWD7”;”TSTIOREV7”;!Internal Control
Phase Lock Controls
There are no GPIB commands for controlling the phase lock through
the service modes.
Phase lock can be turned off or on by:
OUTPUT 716;”FREQOFFSON”!Phase lock off
OUTPUT 716;”FREQOFFSOFF”!Phase lock on
NOTE
Banded high power solutions require that the Phase Lock be turned OFF
or an error message will be displayed on the LCD display. Phase lock
should be established once a day or before calibration in order to
maintain the analyzer’s integrity.
Chapter 1
1-5
Installation, Control, and Specifications
Specifications
Specifications
The specifications in this section apply to the 8753ES Option H85. For
standard instrument specifications, refer to the standard analyzer
documentation.
Table 1-2 shows the uncorrected instrument specifications for the
Option H85.
Table 1-2
Option H85 Instrument Specifications, Uncorrected
Uncorrected Performance Specifications
Description
50 MHz to
500 MHz (dB)
500 MHz to
2 GHz (dB)
2 GHz to
6 GHz (dB)
Directivity
30
30
21
Source Match
16
18
14
Load Match
26
24
15
Reflection Trackinga
±3
±3
±3
Transmission
Trackinga
±3
±3
±3
Crosstalk
70
100
90
a. Excludes rolloff below 840 MHz, which is characteristically –3 dB at 500 MHz, –15 dB at 100 MHz, and –20 dB
at 50 MHz.
1-6
Chapter 1
Installation, Control, and Specifications
Specifications
Table 1-3 shows the corrected instrument specification for the
Option H85. The specifications listed describe the system performance
of the Option H85 network analyzer. The system hardware includes the
following:
Table 1-3
Calibration Kit
85033D
Cables
85131F
Option H85 Instrument Specifications, Corrected
Corrected Performance Specifications
Description
50 MHz to
500 MHz (dB)
500 MHz to
2 GHz (dB)
2 GHz to
6 GHz (dB)
Directivity
42
42
38
Source Match
37
37
31
Load Match
42
42
38
Reflection Tracking
±0.005
±0.005
±.005
Transmission
Tracking
±0.028
±0.03
±0.096
Chapter 1
1-7
Installation, Control, and Specifications
Specifications
TEST PORT INPUTS
Description
Specification
OUTPUT POWER a
Linearity (25 ±5×C)
–15 to +15 dBm
±0.2 dB (relative to 0 dBm output
level)
+5 to +10 dBm (Standard)
±0.2 dB (relative to 0 dBm output
level)
FREQUENCY
Frequency Range
Standard
50 MHz to 3 GHz
Option 006
50 MHz to 6 GHz
Damage Level
+26 dBm or > 6 Vdc
Frequency Response (25 ±5×C)
50 MHz to 500 MHz
±25 dB
500 MHz to 800 MHz
±5 dB
800 MHz to 6 GHz
±3 dB
Frequency Offset Operation
Frequency Range
50 MHz to 6 GHz
R-Channel Input Requirements
0 to –30 dBm, to 3 GHz
(required for phase-locked operation)
0 to –25 dBm, 3 GHz to 6 GHz
External Source Mode (CW Time sweep only)
Frequency Range
50 MHz to 6 GHz
R Input Requirements
Power Level
1-8
0 to –25 dBm
Chapter 1
Installation, Control, and Specifications
Specifications
TEST PORT INPUTS
Description
Specification
OUTPUT POWER a
Linearity (25 ±5×C)
–15 to +15 dBm
±0.2 dB (relative to 0 dBm output
level)
+5 to +10 dBm (Standard)
±0.2 dB (relative to 0 dBm output
level)
Harmonic Measurement Accuracy (25 ±5×C)
±18 dB 50 MHz to 300 MHz
±3.5 dB 300 MHz to 3 GHz
±3.5 dB 3 GHz to 6 GHz (Option 006 on
a. Excludes rolloff below 840 MHz, which is characteristically –3 dB at 500 MHz,
–15 dB at 100 MHz, and –20 dB at 50 MHz.
Chapter 1
1-9
Installation, Control, and Specifications
Specifications
1-10
Chapter 1
2
Application Concepts
2-1
Application Concepts
Application Concepts
Application Concepts
High Power Amplifier Testing
Analyzers equipped with Option H85 are capable of increasing their RF
output levels by means of inserting an external, high power booster
amplifier. As a result, up to 20 watts (+43 dBm) can be delivered to a
device under test (DUT) or amplifier under test. The maximum test
port input power is 1 watt (+30 dBm) CW, but jumpers on the front
panel allow the insertion of high power attenuators or isolators. This
allows test device output levels up to the power limits of the inserted
components.
To protect the analyzer from high power levels, this option allows the
addition of isolators at both test ports and includes internally controlled
step attenuators located between the couplers and samplers.
NOTE
When making high power amplifier measurements, careful planning is
required and precautions must be observed. For more information on
how to make high power measurements, refer to the chapter on
“Making Measurements” in the standard documentation.
Figure 2-1 provides the test configuration for setting up the RF Input
using automatic power meter calibration.
Figure 2-1
Test Configuration for Setting RF Input using
Automatic Power Meter Calibration
2-2
Chapter 2
Application Concepts
Application Concepts
High Power Measurements
Analyzers equipped with Option H85 can be configured to measure high
power devices. This ability is useful if the required input power for a
device under test is greater than the analyzer can provide or if the
maximum output power from an amplifier under test exceeds safe input
limits for a standard analyzer. The Option H85 can be set up in the
standard mode of operation. However, in the standard mode of
operation the Option H85 will not be able to make high power
measurements.
This section provides the following information:
1. Initial Setup
2. Determining Power Levels
3. Additional Setup
4. Selecting Power Ranges and Attenuator Settings
5. Final Setup
6. High Power 2 Port Configuration Example
7. High Power Forward Configuration Example
Chapter 2
2-3
Application Concepts
Application Concepts
Initial Setup
Figure 2-2 shows the initial setup of the Option H85 for making high
power measurements.
1. If the analyzer is in the bypass mode configuration, remove the
jumper between the RF OUT and RF IN connector on the rear panel.
2. Connect the booster amplifier RF INPUT connector to the RF OUT
connector on the rear panel of the analyzer.
3. Connect a 20 dB coupler (that operates within the frequency range of
interest) to the booster amplifier RF OUTPUT connector.
Figure 2-2
High Power Test Setup
2-4
Chapter 2
Application Concepts
Application Concepts
Determining Power Levels
The following provides the information needed to determine what the
correct power levels are and what additional attenuation is needed, if
any.
4. Switch on the analyzer and reduce the power level to −20 dBm by
pressing Menu, POWER, PWR RANGE AUTO, POWER RANGES, −20 x1.
5. Switch on the booster amplifier.
6. Using a power meter, measure the output power from the coupled
arm and the open port of the coupler.
NOTE
Depending on the power meter being used, additional attenuation may
need to be added between the coupler port and the power meter.
7. Verify the gain of the booster amplifier. For example, if the analyzer
output power level was set to −20 dBm and the output power
measured from the open end of the coupler was −5 dBm, then the
gain of the booster amplifier would be +15 dB.
8. Verify that the power measured in step 6 is well within the
acceptable limits (less than −10 dBm for the coupled arm, less than
+43 dBm for the open port).
9. Estimate the maximum power level that will be needed to force the
DUT into compression.
10.At the maximum estimated power level, determine if the maximum
output power from the coupled arm of the coupler will be higher than
the acceptable limit. If so, add the appropriate amount of
attenuation that will keep the coupled output power below 0 dBm
and above −25 dBm.
Chapter 2
2-5
Application Concepts
Application Concepts
Additional Setup
Figure 2-3 (rear view) and Figure 2-4 (front view) show how to make the
RF amplifier and coupler connections to the Option H85.
11.Switch off the booster amplifier.
12.Make a connection between the open port of the 20 dB coupler and
the RF IN connector on the rear panel of the analyzer.
13.Make a connection between the coupled arm of the 20 dB coupler
(along with any added attenuation) and the R-CHANNEL IN
connector on the front panel.
14.Turn the Phase Lock OFF. Press System, SERVICE MENU, SERVICE
MODES, PLL AUTO OFF.
Figure 2-3
High Power Test Setup (Amplifier and Coupler Connections)
Figure 2-4
High Power Test Setup (RF IN Connections)
2-6
Chapter 2
Application Concepts
Application Concepts
Selecting Power Ranges and Attenuator Settings
The following information describes how to select the correct power
range, if an isolator is needed, and what amount of attenuation is
needed. Refer to Figure 2-5 to identify the internal signal paths of the
analyzer.
15.Select a power range that will not exceed the maximum estimated
power level that will force the DUT into compression. For example, if
your booster amplifier has a gain of +15 dB and the DUT will
compress if supplied with +20 dBm, then you would adjust the
analyzer output power to not exceed +5 dBm by pressing Power,
PWR RANGE MAN, POWER RANGES, RANGE 0 −15 TO +10.
16.Adjust the analyzer power level to be below the maximum limit of
the power range by pressing 0 (or less) x1.
17.Estimate the maximum amount of gain that could be provided by the
DUT and, as a result, the maximum amount of power that could be
received by TEST PORT 2 when the DUT is in compression. For
example, if a DUT with a maximum gain of +10 dB receives an input
of +20 dBm, then the maximum amount of power that could be
received by TEST PORT 2 is +30 dBm. If this power level is greater
than +30 dBm, an isolator is needed to protect the internal
mechanical transfer switch.
Figure 2-5
Internal Signal Paths of Analyzer
Chapter 2
2-7
Application Concepts
Application Concepts
18.Calculate the amount of attenuation needed between the analyzer's
couplers and samplers in order to not exceed the optimum sampler
power level of −10 dBm.
In this example, it will be necessary to take the following into
consideration:
• Sampler A will be coupled to the analyzer RF path that could
receive power reflections as high as +20 dBm.
• Sampler B will be coupled to the analyzer RF path that will
receive a maximum of +30 dBm from the DUT.
• Analyzer coupler loss is −16 dB.
• The optimum sampler power level is −10 dBm.
With the above points in mind, the amount of attenuation can be
calculated from the following equations:
• Attenuator A = +20 dBm − 16 dB − (−10 dBm).
Attenuator A = −14 dB
• Attenuator B = +30 dBm − 16 dB − (−10 dBm).
Attenuator B = −24 dB
19.Set the internal step attenuators to the values calculated in the
previous step (rounding off to the highest 5 dB step). Press System,
CONFIGURE MENU, OPTION HXX CONFIG, ATTENUATOR A, 15 x1
ATTENUATOR B 25 x1.
20.Switch on the booster amplifier.
CAUTION
From this point forward, DO NOT press Preset unless you have first
switched off the booster amplifier. Pressing Preset will return the
analyzer to its default power level and default internal attenuator
settings. This increase in power may result in damage to the DUT or
the analyzer.
21.To activate the external reference mode, press Seq, TTL IO, TTL OUT,
TESTSETIO, 6 x1, I/O REV, 6 x1.
22.Measure the output power from TEST PORT 1 and verify that it is
as expected.
23.If you are measuring a highly reflective device, high power isolators
should be inserted in place of the jumpers located between the two
sets of front panel SWITCH and COUPLER connectors.
2-8
Chapter 2
Application Concepts
Application Concepts
Final Setup
The following steps provide the information needed to set measurement
parameters and what considerations need to be taken for interpreting
the measured data.
24.Confirm that all power and attenuator settings are correct and set
the following measurement parameters:
Meas, Trans: FWD S21 (B/R)
25.Perform a response calibration:
• Connect the test port cables of the analyzer to form a thru
configuration.
• Press Cal, CALIBRATE MENU, RESPONSE, THRU.
26.Make the connections shown in Figure 2-6. Switch on the AUT
(amplifier under test) and measure the S21 gain of the amplifier
under test to confirm the proper operation of the measurement test
set up.
Figure 2-6
High Power Test Setup
27.Make any other desired high power measurements.
Ratio measurements such as gain will be correctly displayed.
However, the displayed absolute power levels on the analyzer will
not be correct. To correctly interpret power levels, the gain of the
booster amplifier and the attenuator settings must be taken into
consideration.
Chapter 2
2-9
Application Concepts
Application Concepts
NOTE
If you are increasing the power level from the level set at calibration to
the maximum level of compression, you must press Cal, INTERPOL ON to
switch on interpolation.
If no calibration has been performed or if the instrument is in an
uncalibrated state, the following must be taken into consideration when
interpreting the measured data:
• The value of attenuation added to sampler A and B.
• The R-Channel reference level supplied from the coupled arm of the
20 dB coupler.
2-10
Chapter 2
Application Concepts
Application Concepts
High Power 2 Port Configuration Example
Figure 2-7 shows how the Option H85 can be set up for performing high
power measurements at both ports.
Figure 2-7
High Power 2 Port
IN
OUT
SOURCE
+43 dBm
MAX In
High Power Amplifier
(+40 dBm)
High Power Coupler
(+40 dBm)
20 dB
(+ 20 dBm)
High Power Atten
30 dB
(- 10 dBm)
RF IN
+10 dBm
MAX RFIn
Set A Channel
Atten to 30 dB
IN
A OUT
SW CPLR
PORT 1
+43 dBm
MAX In Test
Ports 1&2
PORT 2
OUT B IN
Set B Channel
Atten to 40 dB
CPLR SW
Loss of Port 2 to B OUT ~ 15 dB
High Power Isolator
20 dB
20 dB
High Power Isolator
(+35 dBm)
DUT
(+43 dBm)
Chapter 2
2-11
Application Concepts
Application Concepts
High Power Forward Configuration Example
Figure 2-8 shows an example of a high power forward direction
configuration of the Option H85.
In this example, the test set I/O FWD is set to 7. The test set I/O REV is
set to 6. With this setup, the R-Channel external reference is set for
forward measurements only. The reverse measurement uses the
internal R-Channel reference.
Figure 2-8
High Power Configuration
IN
OUT
SOURCE
RF Jumpers
RF IN
IN
A OUT
PORT 1
SW CPLR
PORT 2
OUT B IN
CPLR SW
High Power Isolator
High Power Amplifier
High Power Coupler
2-12
Chapter 2
3
Servicing the Option H85 Test Set
3-1
Servicing the Option H85 Test Set
Introduction
Introduction
Refer to your standard analyzer documentation for operation and
service information.
The Option H85 is verified using the verification tests for the standard
8753ES analyzer only.
Adapters, cables, and calibration kits used for verification and
adjustment tests have been changed from the standard 7mm type to the
3.5mm type.
This section provides the following information:
• Adjustment Tests
• Replaceable Parts
• Block Diagram
• Major Assemblies
• Removal and Installation Procedures
3-2
Chapter 3
Servicing the Option H85 Test Set
Adjustment Tests
Adjustment Tests
Equipment Required
• 85033D Calibration Kit
• 85131F Cables
• 10 dB Attenuators, 8493C 010, Quantity 2
• 6 dB Attenuators, 8493C 006, Quantity 2
• SMA to SMA Connector, part number 1250-1251, Quantity 2
• Flex Cables, part number 5062-6664, Quantity 2
NOTE
If any part is replaced, the following adjustment tests must be
performed. Press System, SERVICE MENU, TESTS, ADJUSTMENT TESTS.
Do all of the adjustment tests from 44 through 54. For more
information on executing the adjustment tests, refer to the service
guide in the standard documentation.
Chapter 3
3-3
Servicing the Option H85 Test Set
Adjustment Tests
Performing Test 53
Figure 3-1 shows how to configure the Option H85 for performing the
adjustment tests.
To perform test 53, refer to the service guide provided with your
standard instrument documentation. When a thru cable is needed,
calibrate, find the loss of the cable, and enter the value on the test form.
Figure 3-1
Option H85 Test 53 Configuration
A IN
B IN
PORT 1
PORT 2
16 dB
16 dB
SMA Connectors
Flex Cables
NOTE
Cable loss is measured without the 16 dB attenuators attached.
3-4
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Replaceable Parts and Procedures
Table 3-1 contains the list of replaceable parts for the Option H85. If
any internal part or assembly is replaced, you must run all performance
tests to verify conformance to specifications. Refer to the standard
instrument documentation for performance verification information.
Table 3-1
Option H85 Replaceable Parts
Description
Part Number
Quantity
Cable 10C 10F-14M (Attenuator Ribbon Cable)
5062-0701
2
3.5 mm Test Port Coupler
5086-7968
2
Switch 3PT 24V (Forward/Reverse Transfer Switch)
33311J
1
Board Assembly, LED
08720-60182
1
Coupler Bracket/RF Deck
08753-00148*
1
RF Cable, Port 1 Attenuator to A Sampler
08753-20325*
1
RF Cable, F/R Transfer Switch SW to Port 1
08753-20332*
1
RF Cable, Port 1 Coupler to Coupler Out
08753-20333*
1
RF Cable, Coupler Coupled Port to Sampler Out
08753-20334*
2
RF Cable, Port 2 Coupler to Coupler Out
08753-20335*
1
RF Cable, Port 1 Attenuator to Port 1 A IN
08753-20336*
1
RF Cable, Port 2 Attenuator to B Sampler
08753-20337*
1
RF Cable, F/R Transfer Switch to RF IN (Rear)
08753-20338*
1
RF Cable, Source to RF OUT (Rear)
08753-20339*
1
RF Cable, F/R Transfer Switch to Port 2 SW
08753-20340*
1
RF Cable, Port 2 Attenuator to Port 2 B In
08753-20918*
1
RF Cable, R Channel IN to R Channel Switch
08753-20919*
1
RF Cable, R Channel Coupler to Switch Common
Port
08753-20920*
1
RF Cable, Source to R Channel Switch
08753-20921
1
Rear Panel
08753-60243
1
Front Panel Assembly
08753-60288*
1
NOTE
The part numbers with an asterisk (*) are unique to this special option.
To order a replacement part, please contact the Agilent Technologies
Component Test Product Group Unit at (707) 577-6802 with the part
number, module/model number, and option number. If ordering parts
through your local Agilent Technologies sales or service office, specify
that parts marked with an asterisk are ordered through the Component
Test Product Group Unit.
NOTE
Special options are built to order; therefore, long lead times may be
encountered when ordering replacement parts.
Chapter 3
3-5
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
NOTE
Before opening the analyzer, read all of the warnings and cautions in
the standard documentation.
CAUTION
Before opening the instrument, make sure the instrument is unplugged
from its power source.
3-6
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Option H85 Block Diagram
Figure 3-2 is a visual representation of how the Option H85 works. The
Option H85 block diagram can be used for troubleshooting and as a
reference for theory of operation.
Figure 3-2
Option H85 Block Diagram
Port 1 SWITCH
Port 1 COUPLER
A & B STEP ATTEN
A3 SOURCE
A IN
A OUT
0-55dB
10 to 300 kHz
A24
TRANSFER
SWITCH
PORT 1
A21 COUPLER
PHASE LOCK
A22 COUPLER
PORT 2
ALC
EXT AM
B OUT
B IN
1st IF
2nd LO
R
0-55dB
Port 2 COUPLER
Port 2 SWITCH
S
R IF
RF IN rear
RF OUT
panel
outputs
A
A IF
S
B IF
S
B
A4,A5,A6 SAMPLER/MIXER
R IN
1st LO
TEST SET AREA
8753ES H85
HIGH POWER
43dBm MAX
CPU Board
The following procedure describes the process for turning on the
Option H85 if the CPU board has been replaced.
CAUTION
It is extremely important that the following procedure is done correctly.
If a mistake is made during the Peek Poke addressing, the integrity of
the instrument may be corrupted. If corruption occurs, the instrument
will need to be returned to the factory.
For A/B Attenuator control, press System, SERVICE MENU, PEEK/POKE,
PEEK/POKE, ADDRESS, Enter [1619001535], POKE, [1] X1.
For mechanical switch control, press System, SERVICE MENU,
PEEK/POKE, PEEK/POKE, SERVICE MENU, POKE ADDRESS, Enter
[1619001540], POKE, [1] X1.
Press REST MEMORY.
Chapter 3
3-7
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Checking the Option H85 Installation
Perform the following procedure to see if the Option H85 has been
installed:
Press System, CONFIGURE MENU.
Press OPTION Hxx CONFIG (this should be displayed as the fourth key
from the top).
Option H85 Replacement Procedure
If parts of Option H85 need to be replaced, the following parts are
included:
• Transfer Switch
• Couplers
• Samplers
• Rear Panel RF IN and RF OUT
• Cables
The following instructions are intended as a guide in replacing the
parts that make up the Option H85. To allow more detail to be shown of
the parts in the Option H85, the figures used do not show all parts of
the assembled analyzer.
Refer to the standard documentation for detailed instructions on how to
remove and replace assemblies from the analyzer.
Tools Required
5/16-inch wrench
Major Assemblies
This section provides information on the locations of the major
assemblies used in the Option H85. For clarity of detail, not all parts of
the assembled instrument are shown. Refer to Figure 3-5 on page 3-12,
Figure 3-6 on page 3-13, and Figure 3-7 on page 3-14.
The removal and installation procedures are described in some detail to
provide complete instruction for installing or removing any of the
Option H85 parts. Most of the following procedures remove the front
panel assembly to gain access to the Option H85 assemblies. It is
possible, however, to access the Option H85 switches and cables by
removing just the top and bottom covers leaving the front panel
assembly attached to the instrument.
3-8
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
NOTE
To work on the Option H85, make sure the instrument has been
disconnected from the ac power. Also, be sure that all work is done at an
antistatic workstation.
Set up a static safe workstation. Electrostatic discharge (ESD) can
damage or destroy components. Refer to Figure 3-3.
NOTE
Refer to the 8753ES service guide for information on removing and
installing assemblies that are not part of the Option H85.
Figure 3-3
Example of an Antistatic Workstation
Chapter 3
3-9
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Figure 3-4 shows the major assemblies of the Option H85 in the
network analyzer seen from the bottom front. This view shows all of the
components and cables installed to make the Option H85 with the
following exceptions:
• PC Board installed on the deck
• Attenuator control cables
• Coupler bracket
• Rear panel connections and cable (RF IN)
Descriptions and part numbers for the assemblies in Figure 3-4 are
provided below:
Figure 3-4
Major Assemblies
2
16
15
14
3
13
1
12
4
11
5
10
6
9
8
7
3-10
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Item
Description
Part Number
1
RF Cable, A Sampler to Port 1 Attenuator
08753-20325
2
RF Cable, B Sampler to Port 2 Attenuator
08753-20337
3
RF Cable, Port 1 A IN to Port 1 Attenuator
08753-20336
4
RF Cable, Switch Port 1 to Port 1 SWITCH
08753-20332
5
Attenuator Port 1
33321-60050
6
Switch Mechanical (F/R Transfer Switch)
33311J
7
Deck Assembly
N/A
8
RF Cable, Switch Common Port to RF IN (Rear
Panel)
08753-20338
9
Attenuator Port 2
33321-60050
10
Switch (Standard, R-Channel)
5087-7067
11
RF Cable, Switch Port 2 to Port 2 SWITCH
08753-20340
12
RF Cable, Port 2 B IN to Port 2 Attenuator
08753-20918
13
RF Cable, R-Channel IN to R-Channel Switch
08753-20919
14
RF Cable, Source to R-Channel Switch
08753-20921
15
RF Cable, R-Channel Coupler to Switch Common
Port
08753-20920
16
Cable Bracket
08753-00137
Chapter 3
3-11
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Coupler Bracket Assembly
Information on the parts used to assemble the coupler bracket is
provided below.
Figure 3-5
Coupler Bracket Assembly
9
1
2
8
3
4
7
5
6
Item
Description
Part Number
1
SMA to SMA Connector
2
Coupler Bracket
3
Coupler Port 2
4
RF Cable, Coupler Coupled Port to Sampler Out
08753-20334
5
RF Cable, Port 2 Coupler to Coupler Out
08753-20335
6
RF Cable, Port 1 Coupler to Coupler Out
08753-20333
7
RF Cable, Coupler Coupled Port to Sampler Out
08753-20334
8
Coupler Port 1
5086-7968
9
LED PC Board
08720-60182
Coupler Nut (not shown)
3-12
1250-1251
08753-20148-2
5086-7968
5022-1087
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
RF Rear Panel Assembly
The following information describes the parts used to build the
RF IN and RF OUT rear panel assembly.
Figure 3-6
RF IN, RF OUT Rear Panel Assembly
1
2
3
4
5
6
Item
Description
Part Number
1
Connection to Source
2
RF Cable, Source to RF OUT (Rear)
3
Install wire tie
4
RF Cable, F/R Transfer Switch to RF IN (Rear)
08753-20338
5
Rear Connector Bracket
08720-00089
6
SMA to SMA Connector
1250-1251
Chapter 3
08753-20339
1400-0249
3-13
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Installation and Removal Procedures
Deck Assembly
The following section provides information on how to install or remove
the parts used on the deck assembly.
Figure 3-7
Deck Assembly
11
1
10
2
9
3
8
4
5
7
6
3-14
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Item
Description
Part Number
1
Cable 10C 10F-14M (Attenuator Control Ribbon Cable)
5062-0701
2
Port 1 Attenuator
3
Attenuator Mounting Screws (2)
2200-0150
4
F/R Switch Mounting Screws (2)
0515-0667
5
F/R Switch (Refer to Figure 3-9 on page 3-19)
6
PC Board (Refer to Figure 3-8 on page 3-17)
7
R-Channel Switch (Refer to Figure 3-9 on page 3-19)
5087-7067
8
R-Channel Switch Mounting Screw (2)
0515-0430
9
Port 2 Attenuator
10
Port 2 Attenuator Mounting Screws (2)
2200-0150
11
Cable 10C 10F-14M (Attenuator Control Ribbon Cable)
5062-0701
33321-60050
33311J
08753-60280
33321-60050
1. Remove the bottom cover.
2. Remove the front panel from the instrument. refer to the standard
documentation
3. Disconnect the cables going to the coupler bracket assembly using a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly. There is one
screw on the side and two screws on the deck assembly. Six screws
hold the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35.
5. Remove the deck assembly. Refer to “Deck Assembly Removal or
Installation” on page 3-21.
PORT 1 Attenuator Installation or Removal
1. Remove the ribbon cable from J12 on the PC board.
2. The Port 1 attenuator is attached to the deck with two screws. The
screws are started on the opposite side of the deck from where the
attenuator is located. Remove the two screws.
3. To install the Port 1 attenuator, replace the ribbon cable that is
supplied with the new attenuator with the longer cable, part number
5062-0701. Reverse steps one and two. Torque the two screws to
10 inches-pound (in-lb).
Chapter 3
3-15
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
PORT 2 Attenuator Installation or Removal
1. Remove the ribbon cable from J11 on the PC board.
2. The Port 2 attenuator is attached to the deck with two screws. The
screws are started on the opposite side of the deck from where the
attenuator is located. Remove the two screws.
3. To install the Port 2 attenuator, replace the ribbon cable that is
supplied with the new attenuator with the longer cable, part number
5062-0701. Reverse steps one and two. Torque the two screws to
10 in-lb.
F/R Switch Installation or Removal
1. Remove the cable harness from J3 on the PC board.
2. Remove the two screws that hold the switch to the deck. Both screws
are located on the same side of the deck as the switch.
3. To install or remove the cable harness, refer to “Switch Cable
Installation” on page 3-19.
4. Reverse steps 1 and 2 to install the F/R switch.
R-Channel Switch Installation or Removal
1. Remove the ribbon cable from J4 on the PC board.
2. Remove the two screws that hold the switch to the deck. Both screws
are located on the same side of the deck as the switch is.
3. To install or remove the cable harness, refer to “Switch Cable
Installation” on page 3-19.
4. Reverse steps 1 and 2 to install the R-Channel switch.
3-16
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
PC Board Installation
The following section provides information on how to install or remove
the PC board from the deck assembly.
Figure 3-8
Deck Assembly PC Board Installation or Removal
1
12
2
11
10
3
9
4
8
5
7
6
Item
Description
1
PC Board
2
J3 Connector
3
Connector Not Used
4
Connector Not Used
5
J12 Connector
6
J1 50 Pin Ribbon Cable Connector
7
J11 Connector
8
Connector Not Used
9
J4 Connector
10
Connector Not Used
11
Connector Not Used
12
PC Board Mounting Screws (5)
Chapter 3
Part Number
08753-60280
0515-0372
3-17
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly. Use a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly. One screw on
the side and two screws on the deck assembly. Six screws also hold
the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35.
5. Remove the Port 1 attenuator cable from J12 on the PC board.
6. Remove the F/R switch wire harness from J3 on the PC board.
7. Remove the R-Channel switch wire harness from J4 on the PC
board.
8. Remove the Port 2 attenuator cable from J11 on the PC board.
9. Remove the five screws that hold the PC board to the deck assembly.
10.To install the PC board, reverse steps 1 through 8. Torque the
screws to 10 in-lb.
3-18
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Switch Cable Installation
The following provides information on how to install or remove the
switch harness from the switch (R-Channel and F/R transfer).
Figure 3-9
Switch Cable Installation
1
2
3
4
5
Item
1
Description
F/R Transfer Switch
R-Channel Switch
2
Wire Harness
3
Blue Wire to Terminal 2
4
Red Wire to Common Terminal
5
Green Wire to Terminal 1
Chapter 3
Part Number
33311J
5087-7067
85047-60004
3-19
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Disconnect the six cables (see the list below) on the deck assembly.
Refer to Figure 3-4 on page 3-10.
For the F/R Transfer Switch:
•
F/R Transfer Switch Common Terminal to RF IN (Rear Panel)
•
F/R Transfer Switch Port 1 to Coupler Bracket Port 1 SW,
part number 08753-20332
•
F/R Transfer Switch Port 12 to Coupler Bracket Port 2 SW,
part number 08753-20340
For the R-Channel Switch:
•
R-Channel Switch Common Terminal to R-Channel,
part number 08753-20920
•
R-Channel Switch Port 2 to Source, part number 08753-20921
•
R-Channel Switch Port 1 to Coupler Bracket, part number,
08753-20919
3. Remove the bad switch with the wire harness by removing the two
screws and disconnecting the cable from J3 or J4 on the PC board.
4. De-solder and re-solder the wire harness to the switch (both
switches are wired the same) as shown in Figure 3-9 on page 3-19.
5. Install the new switch with the wire harness onto the deck
assembly. Torque the mounting screws to 10 in-lb.
6. Re-connect the cables removed in step 1.
3-20
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Deck Assembly Removal or Installation
The following information is provided on how to remove or install the
deck assembly.
Figure 3-10
Deck Assembly Installation
3
1
2
4
Item
Description
Part Number
1
Deck Assembly
2
Mounting Screw Deck to Chassis (x2)
0515-0375
3
Mounting Screw Deck to Chassis (x2)
0515-0375
4
Mounting Screw Deck to Side Rail (x2)
0515-0458
Chapter 3
08753-00148-1
3-21
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Remove the top and bottom covers.
4. Disconnect the cables going to the coupler bracket assembly using a
5/16-inch wrench.
5. Remove the coupler bracket from the deck assembly. There is one
screw on the side and two screws on the deck assembly. Six screws
hold the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35 for placement of the cables.
6. Disconnect the six cables listed below at the deck assembly. Refer to
Figure 3-4 on page 3-10.
•
F/R Transfer Switch Common Terminal to RF IN (Rear Panel)
•
F/R Transfer Switch Port 1 to Coupler Bracket Port 1 SW
•
F/R Transfer Switch Port 2 to Coupler Bracket Port 2 SW
•
R-Channel Switch Common Terminal to R-Channel
•
R-Channel Switch Port 2 to Source
•
R-Channel Switch Port 2 to Coupler Bracket R IN
•
J1 Ribbon Cable
7. Remove the deck assembly from the instrument. Refer to Figure 3-10
on page 3-21.
8. To install the deck assembly, reverse steps one to five. Torque the
screws to 10 in-lb.
3-22
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
SMA to SMA Installation
The following section provides information on how to install or remove
the SMA to SMA connectors on the coupler bracket.
Figure 3-11
Coupler Bracket SMA to SMA Connector Installation
2
1
Item
Description
1
SMA to SMA Connectors (Nine Places)
2
Coupler Bracket
Chapter 3
Part Number
1250-1251
08753-20148-2
3-23
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly using a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly. There is one
screw on the side and two screws on the deck assembly. Six screws
hold the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35.
5. Remove the required coupler bracket cables. Refer to Figure 3-14 on
page 3-29.
6. Place the new SMA to SMA connectors into the coupler bracket.
Torque to 10 in-lb.
7. Re-install the coupler bracket and re-connect all of the cables.
Torque to 10 in-lb.
3-24
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
LED PC Board Installation
The following information is provided on how to install or remove the
LED PC board assembly from the coupler bracket assembly.
Figure 3-12
LED PC Board Installation
1
2
Item
Description
1
LED PC board
2
Screws (3)
Chapter 3
Part Number
08720-60182
0515-0430
3-25
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly using a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly. There is one
screw on the side and two screws on the deck assembly. Six screws
hold the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35.
5. Replace the LED PC board. Torque the screws to 10 in-lb.
6. Re-install the coupler bracket and re-connect all of the cables.
Torque to 10 in-lb.
3-26
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Coupler Installation
The following section provides information on installing or removing
the coupler(s) from the coupler bracket.
Figure 3-13
Coupler Installation
1
2
Item
Description
Part Number
1
Coupler Port 1 and Port 2
5086-7968
2
Coupler Nut (2)
5022-1087
Chapter 3
3-27
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly with a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly There is one
screw on the side and two screws on the deck assembly. Six screws
additional screws hold the coupler bracket to the front frame.
Refer to Figure 3-17 on page 3-35.
5. Remove the required coupler bracket cables. Refer to Figure 3-14 on
page 3-29.
6. Replace the coupler on the coupler bracket. Torque the coupler nut
to 71 inch-pounds.
7. Re-install the coupler bracket and re-connect all of the cables.
Torque to 10 in-lb.
3-28
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Coupler Bracket Assembly
Information on how to install or remove the coupler bracket cables is
provided below.
Figure 3-14
Coupler Bracket Assembly
1
4
2
3
Item
Description
Part Number
1
RF Cable, Coupler Coupled Port to B OUT
08753-20334
2
RF Cable, Port 2 Coupler to Coupler Out
08753-20335
3
RF Cable, Port 1 Coupler to Coupler Out
08753-20333
4
RF Cable, Coupler Coupled Port to A OUT
08753-20334
Chapter 3
3-29
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly using a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly. There is one
screw on the side and two screws on the deck assembly. Six screws
hold the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35.
5. Replace the defective cable on the coupler bracket using a 5/16-inch
wrench. Torque the cable connection to 10 in-lb.
6. Re-install the coupler bracket and re-connect all of the cables.
Torque to 10 in-lb.
3-30
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Cable Installation
The following section provides information on how to install or remove
the B Sampler to Port 2 Attenuator cable or the A Sampler to the Port 1
Attenuator cable.
Figure 3-15
Cable A and B Sampler to PORT 1 and 2 Attenuator Installation
1
2
Bottom Deck View
Top Front View
Item
Description
Part Number
1
RF Cable, B Sampler to Port 2 Attenuator
08753-20337
2
RF Cable, A Sampler to Port 1 Attenuator
08753-20325
Chapter 3
3-31
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the top and bottom covers.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly with a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly. There is one
screw on the side and two screws on the deck assembly. Six screws
hold the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35.
5. Remove the cable for the A Sampler using a 5/16-inch wrench.
6. Remove the A Sampler cable from the Port 1 attenuator on the deck.
7. Remove the cable for the B Sampler using a 5/16-inch wrench.
8. Remove the B Sampler cable from the Port 2 attenuator on the deck.
9. Re-install the cable(s). Torque to 10 in-lb.
3-32
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Cable Installation Source and R-Channel
Information on how to install or remove the Source to R-Channel
Switch cable or the R-Channel to R-Channel Switch cable is provided
below.
Figure 3-16
Cable Source and R-Channel to R-Channel Switch and Cable
Bracket
1
2
3
4
Top Front View
Item
Bottom Deck View
Description
Part Number
1
RF Cable, Source to R-Channel Switch Port 2
08753-20921
2
RF Cable, R-Channel to R-Channel Switch Common
Terminal
08753-20920
3
Cable Bracket
08753-00137
4
Cable Bracket Mounting Screw
Chapter 3
0515-0665
3-33
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the top and bottom covers.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly using a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly. There is one
screw on the side and two screws on the deck assembly. Six screws
hold the coupler bracket to the front frame. Refer to Figure 3-17 on
page 3-35.
5. Remove the cable bracket.
6. Remove the Source to R-Channel Switch cable from the source.
7. Remove the Source to R-Channel Switch cable from the R-Channel
Switch Port 2.
8. Replace the Source to R-Channel Switch cable by reversing steps 1
through 7. Torque the connections to 10 in-lb.
9. Remove the R-Channel to R-Channel Switch cable from the
R-Channel.
10.Remove the R-Channel to R-Channel Switch cable from the
R-Channel Switch common terminal.
11.Replace the R-Channel to R-Channel Switch cable by reversing steps
1 through 7, 9 and 10. Torque the connections to 10 in-lb.
3-34
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Coupler Bracket Installation
Provided below is information on how to install or remove the coupler
bracket.
Figure 3-17
Coupler Bracket Installation
1
4
2
3
Item
Description
Part Number
1
Coupler Bracket
2
Coupler to Deck Mounting Screws (2)
0515-1946
3
Coupler to Frame Mounting Screws (6)
0515-1946
4
Coupler to Side of Frame Mounting Screws (1)
0515-1400
Chapter 3
08753-00148-2
3-35
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the top and bottom covers.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Disconnect the cables going to the coupler bracket assembly with a
5/16-inch wrench.
4. Remove the coupler bracket from the deck assembly by removing one
screw on the side and two screws on the deck assembly. There are six
more screws holding the coupler bracket to the front frame. Refer to
Figure 3-17 on page 3-35.
5. Replace any defective part on the coupler bracket assembly as
needed.
6. To re-install the coupler bracket, reverse steps 1 through 4. Torque
to 10 in-lb.
3-36
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Cable Installation R-Switch to R-Channel
The following section provides information on how to install or remove
the R-Channel to the Port 2 cable on the R-Channel Switch.
Figure 3-18
Switch to R-Channel
1
2
3
4
Item
Description
1
SMA Connector (R-Channel)
2
RF Cable, R-Channel to R-Channel Switch
Port 2
3
R-Channel Switch Port 2
4
R-Channel Switch
Chapter 3
Part Number
1250-1251
08753-20919
5087-7067
3-37
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument.
3. Disconnect the R-Channel to R-Channel Switch Port 2 cable
attached to the coupler bracket assembly with a 5/16-inch wrench.
4. Remove the R-Channel to R-Channel Switch Port 2 cable from the
R-Channel switch port 2 on the deck assembly.
5. Replace the R-Channel to R-Channel Switch Port 2 cable to the
R-Channel SMA connector on the front panel and to the R-Channel
switch port 2. Torque to 10 in-lb.
6. Reverse steps 1 and 2 to complete the process. Torque to
10 in-lb.
3-38
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Cable Installation PORT A and PORT B
The following data is provided on how to install or remove the PORT A
and PORT B cables to the PORT A and PORT B attenuators.
Figure 3-19
Cable, PORT A and PORT B, to Attenuator Installation
1
6
2
5
3
4
Item
Description
Part Number
1
Port 1 A IN
2
RF Cable, Port 1 Attenuator to Port 1 A IN
08753-20336
3
Attenuator Port 1
33321-60050
4
Attenuator Port 2
33321-60050
5
RF Cable, Port 2 Attenuator to Port 2 B IN
08753-20918
6
Port 2 A IN
Chapter 3
3-39
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Remove the PORT 1 A IN to PORT 1 attenuator or PORT 2 B IN to
PORT 2 attenuator cables using a 5/16-inch wrench.
4. Replace the PORT 1 A IN to PORT 1 attenuator or PORT 2 B IN to
PORT 2 attenuator cables with a 5/16-inch wrench. Torque to
10 in-lb.
5. Reverse steps 1 and 2 to complete the process.
3-40
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
PORT 1 and PORT 2 SWITCH to F/R Transfer Switch Cables
The following section provides information on how to install or remove
the PORT 1 and PORT 2 SWITCH cables from the F/R Transfer switch
port 1 and port 2 connectors.
Figure 3-20
Cable Switch to PORT 1 and PORT 2 SWITCH
1
6
2
5
3
4
Item
Description
Part Number
1
Port 1 SW (Switch)
2
RF Cable, F/R Transfer Switch Port 2 to Port 1 SW
08753-20332
3
F/R Transfer Switch Port 2
4
F/R Transfer Switch Port 1
5
RF Cable, F/R Transfer Switch Port 1 to Port 2 SW
6
Port 2 SW (Switch)
Chapter 3
08753-20340
3-41
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the bottom cover.
2. Remove the front panel from the instrument. Refer to the standard
documentation.
3. Remove the defective cable attached to the SMA connector on the
coupler bracket assembly, PORT 1 SWITCH or PORT 2 SWITCH
with a 5/16-inch wrench.
4. Remove the defective cable attached to the F/R Transfer switch
port 1 or port 2 with a 5/16-inch wrench.
5. Replace the defective cable. Torque to 10 in-lb.
6. Re-install the bottom cover and front panel, then torque to
10 in-lb.
3-42
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
RF IN to F/R Transfer Switch Cable
The following section provides information on how to install or remove
the RF IN to F/R Transfer switch cable.
Figure 3-21
Cable Switch to Rear Panel RF IN
1
2
3
4
Item
Description
1
Rear Panel Bracket
2
RF IN Connector
3
RF Cable, F/R Transfer Switch Common
Port to RF IN Rear Panel Bracket
4
F/R Transfer Switch Common Port
Chapter 3
Part Number
08720-00089
1250-1251
08753-20338
3-43
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the top and bottom covers.
2. Remove the rear panel from the instrument. Refer to Figure 3-23 on
page 3-47.
3. Remove the cable from the F/R Transfer switch common port using a
5/16-inch wrench.
4. Remove the cable from the RF IN connector on the rear panel
bracket. Use a 5/16-inch wrench.
5. Replace the defective cable. Torque to 10 in-lb.
6. Re-install the cable by reversing steps 1 through 4. Torque to
10 in-lb.
3-44
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
RF OUT to Source Cable
The following information is provided on how to install or remove the
RF OUT to Source cable.
Figure 3-22
Cable Source to Rear Panel RF OUT
1
2
3
4
Item
Description
1
Source Connection
2
RF Cable, Source to RF OUT
3
SMA Connector, RF OUT
4
Rear Panel Bracket
Chapter 3
Part Number
08753-20339
1250-1251
08720-00089
3-45
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Remove the top cover.
2. Remove the rear panel from the instrument. Refer to Figure 3-23 on
page 3-47.
3. Remove the cable from the Source. Use a 5/16-inch wrench.
4. Remove the cable from the RF OUT connector on the rear panel
bracket using a 5/16-inch wrench.
5. Replace the defective cable. Torque to 10 in-lb.
6. Re-install the cable by reversing steps 1 through 4. Torque to
10 in-lb.
3-46
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Rear Panel
The following section provides information on how to install or remove
the analyzer’s rear panel.
Figure 3-23
Rear Panel Installation
1
2
3
4
Item
Description
1
Rear Panel
2
RF OUT Connector with Alignment Tool Attached
3
RF IN Connector with Alignment Tool Attached
4
RF Bracket Mounting Screws (2)
Chapter 3
Part Number
3-47
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
1. Install the alignment tools.
2. Remove the two screws that hold the RF bracket to the rear panel.
3. Remove the rear panel. Refer to the standard documentation.
4. Remove the RF IN and RF OUT cables from the RF bracket with a
5/16-inch wrench.
5. To replace an SMA to SMA connector, use a 5/16-inch wrench to
remove or install the new connector. Reverse steps 1 through 5 to
complete the assembly. Torque to 10 in-lb.
6. To replace the RF bracket, remove the RF bracket and replace it by
reversing steps 1 through 5. Torque to 10 in-lb.
3-48
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
RF IN to RF OUT Jumper Cable
Information on how to install or remove the RF IN to RF OUT jumper
cable is provided below.
Figure 3-24
Cable Rear Panel RF IN to RF OUT Installation
1
2
3
Item
1
Description
RF IN to RF OUT Jumper Cable
Part Number
08720-20098
1. Remove or install the RF IN to RF OUT jumper cable using a
5/16-inch wrench. Torque to 10 in-lb.
Chapter 3
3-49
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Front Panel Jumpers
The following data is provided on how to install or remove the front
panel jumpers.
Figure 3-25
Cable Front Panel Jumpers
Item
Description
Part Number
1
Jumper Cable, A OUT to A IN
08720-20098
2
Jumper Cable, Port 1 SWITCH to Port 1 COUPLER
08720-20098
3
Jumper Cable, Port 2 COUPLER to Port 2 SWITCH
08720-20098
4
Jumper Cable, B IN to B OUT
08720-20098
5
Hole Plug
6960-0023
1. Remove or install the front panel jumper cables using a
5/16-inch wrench. Torque to 10 in-lb.
3-50
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Option H85 Front View Without Front Panel
The following graphic provides a visual layout of what the front of the
Option H85 should look like with the front panel removed.
Figure 3-26
Front View of Instrument without Front Panel Installed
Chapter 3
3-51
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Option H85 Bottom View
The following graphic provides a visual layout of the Option H85
showing what the bottom of the instrument looks like with the front
panel, bottom cover, and the deck PC board removed.
Figure 3-27
Bottom View of Instrument
3-52
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Option H85 Top View
The following provides a visual layout showing what the top of the
Option H85 looks like with the front panel, top cover, and major
standard assemblies removed.
Figure 3-28
Top View of Instrument
Chapter 3
3-53
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Option H85 Front Bottom View
The following picture provides a visual layout of the Option H85
showing what the front and bottom of the instrument look like with the
front panel, top cover, bottom cover, and deck PC board removed.
Figure 3-29
Bottom Front View
3-54
Chapter 3
Servicing the Option H85 Test Set
Replaceable Parts and Procedures
Agilent Technologies Sales and Service Offices
UNITED STATES
Instrument Support Center
Agilent Technologies
(800) 403-0801
EUROPEAN FIELD OPERATIONS
Headquarters
Hewlett-Packard S.A.
150, Route du Nant-d’Avril
1217 Meyrin 2/ Geneva
Switzerland
(41 22) 780.8111
France
Hewlett-Packard France
1 Avenue Du Canada
Zone D’Activite De
Courtaboeuf
F-91947 Les Ulis Cedex
France
(33 1) 69 82 60 60
Germany
Hewlett-Packard GmbH
Hewlett-Packard Strasse
61352 Bad Homburg v.d.H
Germany
(49 6172) 16-0
Great Britain
Hewlett-Packard Ltd.
Eskdale Road, Winnersh
Triangle Wokingham, Berkshire
RG41 5DZ England
(44 118) 9696622
INTERCON FIELD OPERATIONS
Headquarters
Hewlett-Packard Company
3495 Deer Creek Rd.
Palo Alto, CA 94304-1316
USA
(415) 857-5027
Australia
Hewlett-Packard Australia
Ltd.
31-41 Joseph Street
Blackburn, Victoria 3130
(61 3) 895-2895
Canada
Hewlett-Packard (Canada) Ltd.
17500 South Service Road
Trans-Canada Highway
Kirkland, Quebec H9J 2X8
Canada
(514) 697-4232
Japan
Hewlett-Packard Japan, Ltd.
Measurement Assistance Center
9-1, Takakura-Cho, Hachioji-Shi,
Tokyo 192-8510, Japan
TEL (81) 426-56-7832
FAX (81) 426-56-7840
Singapore
Hewlett-Packard Singapore
(Pte.) Ltd.
150 Beach Road
#29-00 Gateway West
Singapore 0718
(65) 291-9088
Taiwan
Hewlett-Packard Taiwan
8th Floor, H-P Building
337 Fu Hsing North Road
Taipei, Taiwan
(886 2) 712-0404
China
China Hewlett-Packard Co.
38 Bei San Huan X1 Road
Shuang Yu Shu
Hai Dian District
Beijing, China
(86 1) 256-6888
Chapter 3
3-55