Instruction Manual MCO-19AIC-PE

Transcription

Instruction Manual MCO-19AIC-PE
Operating Instructions
CO2 Incubator
MCO-19AIC
MCO-19AIC(UV)
MCO-19AIC(UVH)
MCO-19AIC
MCO-19AICUV
MCO-19AICUVH
Series
Please read these instructions carefully before using this product, and save this manual for future use.
See page 66 for all Model Nos.
CONTENTS
INTRODUCTION
PRECAUTIONS FOR SAFE OPERATION
LABELS ON INCUBATORS
ENVIRONMENTAL CONDITIONS
INCUBATOR COMPONENTS
Control panel and keypad
Remote alarm terminal
INSTALLATION SITE
INSTALLATION
Connecting a CO2 gas cylinder
PREVENTING CONTAMINATION
PRECAUTIONS FOR CULTURES
Using unlock key
INITIAL CLEANING METHOD
REMOVING INNER ATTACHMENTS
INSTALLING INNER ATTACHMENTS
FILLING HUMIDIFYING PAN
WATER LEVEL SENSOR
ROUTINE MAINTENANCE
CORRECT OPERATION
LCD PANEL
BASIC OPERATIONS ON CONTROL PANEL
BASIC PARAMETERS
Setting chamber temperature and CO2 density
Setting key lock
Removing key lock
Setting key lock password
Setting high limit alarm temperature
ALARM PARAMETERS
UV LAMP PARAMETERS
Using UV lamp
Precautions when using UV lamp
Setting UV lamp ON period
Lighting UV lamp for 24 hours
OTHER PARAMETERS
Setting date, time, and log interval
Initial settings (LCD/DAQ parameters)
DISPLAYING THE LOG
Transferring data
H2O2 DECONTAMINATION
Precautions when handling H2O2 reagent
ALARMS, SAFETY, AND SELF-DIAGNOSIS
CALIBRATION
Temperature/CO2 calibration
TROUBLESHOOTING
DISPOSAL OF UNIT
AUTOMATIC CO2 CYLINDER CHANGEOVER
AUTOMATIC CO2 DENSITY CALIBRATION
STACKING INCUBATORS
SPECIFICATIONS
PERFORMANCE
SAFETY CHECK SHEET
1
P. 2
P. 3
P. 7
P. 7
P. 8
P. 10
P. 11
P. 12
P. 13
P. 14
P. 15
P. 16
P. 16
P. 17
P. 18
P. 20
P. 21
P. 22
P. 23
P. 23
P. 24
P. 25
P. 26
P. 27
P. 27
P. 28
P. 30
P. 31
P. 32
P. 33
P. 34
P. 35
P. 36
P. 38
P. 39
P. 40
P. 41
P. 45
P. 46
P. 49
P. 51
P. 53
P. 58
P. 60
P. 63
P. 65
P. 66
P. 67
INTRODUCTION
■ Read this operating instruction carefully before using the Product and follow the instructions for safety
operation.
■ Our company disavows any responsibility for safety if the Product is used for other than the intended
use or used with any procedures other than those given in this operating instruction.
■ Keep this operating instruction in a suitable place so that it can be referred to as necessary.
■ The contents of this operating instruction are subject to change without notice for improvement of
performance or functions.
■ Contact our sales representative or agent if any page of the operating instruction is lost or the page
order is incorrect.
■ Contact our sales representative or agent if any point in this operating instruction is unclear or if there
are any inaccuracies.
■ No part of this operating instruction may be reproduced in any form without the expressed written
permission of our company.
CAUTION
Our company guarantees the product under certain warranty conditions.
way shall be responsible for any loss of content or damage of content.
Our company in no
Microsoft, Windows, Windows 7, Windows Vista, Windows XP and Windows 2000 are registered
trademarks in the United States of Microsoft Corporation and various other countries.
2
PRECAUTIONS FOR SAFE OPERATION
It is imperative that the user complies with this operating instruction
as it contains important safety advice.
Items and procedures are described so that you can use this unit correctly and safely.
If the precautions advised are followed, this will prevent possible injury to the user and
any other person.
Precautions are illustrated in the following way:
WARNING
Failure to observe WARNING signs could result in a hazard to personnel
possibly resulting in serious injury or death.
CAUTION
Failure to observe CAUTION signs could result in injury to personnel and
damage to the unit and associated property.
Symbol shows;
this symbol means caution.
this symbol means an action is prohibited.
this symbol means an instruction must be followed.
Be sure to keep this operating instruction in a place accessible to users of this unit.
< Label on the unit >
This mark is labeled on the cover in which the electrical components of high voltage
are enclosed to prevent the electric shock.
The cover should be removed by a qualified engineer or a service personnel only.
WARNING
As with any equipment that uses CO2 gas, there is a likelihood of oxygen depletion in the vicinity
of the equipment. It is important that you assess the work site to ensure there is suitable and
sufficient ventilation. If restricted ventilation is suspected, then other methods of ensuring a
safe environment must be considered. These may include atmosphere monitoring and
warning devices.
3
PRECAUTIONS FOR SAFE OPERATION
WARNING
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to
rain water.
Only qualified engineers or service personnel should install the unit.
unqualified personnel may cause electric shock or fire.
The installation by
Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from
turning over. If the floor is not strong enough or the installation site is not adequate, this may result
in injury from the unit falling or tipping over.
Never install the unit in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
Never install the unit in a flammable or volatile location. This may cause explosion or fire.
Never install the unit where acid or corrosive gases are present as current leakage or electric
shock may result due to corrosion.
Always ground (earth) the unit to prevent electric shock. If the power supply outlet is not
grounded, it will be necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.
Connect the unit to a power source as indicated on the rating label attached to the unit. Use
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.
Never store volatile or flammable substances in this unit if the container cannot be sealed. These
may cause explosion or fire.
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the unit.
This may cause electric shock or injury by accidental contact with moving parts.
Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use
may cause bad effect on your health or environment.
Turn off the power switch (if provided) and disconnect the power supply to the unit prior to any
repair or maintenance of the unit in order to prevent electric shock or injury.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet
hand. This may cause electric shock.
4
WARNING
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of
maintenance. These may be harmful to your health.
Never splash water directly onto the unit as this may cause electric shock or short circuit.
Never put containers with liquid on the unit as this may cause electric shock or short circuit when
the liquid is spilled.
Never bind, process, or step on the power supply cord, or never damage or break the power
supply plug. A broken supply cord or plug may cause fire or electric shock.
Do not use the supply cord if its plug is loose. Such supply cord may cause fire or electric shock.
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an
unauthorized person may result in fire, or electric shock or injury due to a malfunction.
Disconnect the power supply plug if there is something wrong with the unit.
abnormal operation may cause electric shock or fire.
Continued
When removing the plug from the power supply outlet, grip the power supply plug, not the cord.
Pulling the cord may result in electric shock or fire by short circuit.
Disconnect the power supply plug before moving the unit. Take care not to damage the power
cord. A damaged cord may cause electric shock or fire.
Disconnect the power plug when the unit is not used for long periods. Keeping the connection
may cause electric shock, current leakage, or fire due to the deterioration of insulation.
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that
children do not have access and that doors cannot be closed completely.
The disposal of the unit should be accomplished by appropriate personnel. Remove doors to
prevent accidents such as suffocation.
Do not put the packing plastic bag within reach of children as suffocation may result.
Use the reagent specified by our company for H2O2 decontamination. Using a different H2O2
solution may result in explosion or damage to the incubator.
When performing H2O2 decontamination, securely close the internal and external doors. Failure to
do so may be harmful to health due to leakage of H2O2 gas.
During H2O2 decontamination, plug the access hole with the silicon cap that is provided. Failure
to do so may be harmful to health due to leakage of H2O2 gas.
5
PRECAUTIONS FOR SAFE OPERATION
CAUTION
This unit must be plugged into a dedicated circuit protected by branch circuit breaker.
Use a dedicated power source as indicated on the rating label attached to the unit. A multiple-tap
may cause fire resulting from abnormal heating.
Never store corrosive substances such as acid or alkali in this unit if the container cannot be
sealed. These may cause corrosion of inner components or electric parts.
Check the setting when starting up of operation after power failure or turning off of power
switch. The stored items may be damaged due to the change of setting.
Be careful not to tip over the unit during movement to prevent damage or injury.
Prepare a safety check sheet (copy the last page) when you request any repair or maintenance for
the safety of service personnel.
Wear rubber gloves when handling the H2O2 reagent. Direct contact with the H2O2 reagent may
result in inflammation of the skin.
H2O2 decontamination can be performed only for the chamber and chamber attachments with
standard specifications, and not for any other objects.
Perform H2O2 decontamination with the chamber attachments arranged as specified by our
company. Arranging them in a different way may result in insufficient decontamination.
After H2O2 decontamination has been completed, wear rubber gloves and use a non-woven cloth
to wipe off the residual H2O2 fluid from the bottom of the chamber, any objects that were
decontaminated, and the bottoms of ducts.
6
LABELS ON INCUBATOR
Warning and caution labels are attached to the incubator. The following table describes the labels.
This label is attached to covers that access high-voltage electrical components to
prevent electric shock. Only a qualified engineer or service personnel should be
allowed to open these covers.
This symbol indicates an ultraviolet light (UV) caution.
This symbol indicates that caution is required. Refer to product documentation for
details.
This symbol indicates a hot surface.
This symbol indicates an earth.
This symbol means “ON” for a power switch.
This symbol means “OFF” for a power switch.
ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):
■ Indoor use;
■ Altitude up to 2000 m;
■ Ambient temperature 5 oC to 40 oC
■ Maximum relative humidity 80 % for temperature up to 31 oC decreasing linearly to 50 % relative
humidity at 40 oC;
■ Mains supply voltage fluctuations not to exceed ±10 % of the nominal voltage;
■ Other supply voltage fluctuations as stated by the manufacturer;
■ Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains
supply the minimum and normal category is II;
■ Pollution degree 2 in accordance with IEC 60664.
7
INCUBATOR COMPONENTS
Handles
12
5
6
7
9 (inside)
8
2
4
1
15, 16 (inside)
For MCO-19AIC(UVH)/19AIC(UV) or
when MCO-19UVS is installed.
Handles
22
10
13 (inside) 14
3
18
21
17
11
20
19
For MCO-19AIC(UVH)/19AIC(UV) or
when MCO-19UVS is installed.
Rear Right Side
Rear Left Side
8
1. Outer door: The outer door is held to the frame with the magnetic seal. The door heater is installed in
the door panel. The door opening is reversible. Contact our sales representative or agent to change the
door hinge from left to right or vice versa.
2. Inner door: The inner door is made of tempered glass. However do not subject the glass to excessive
impacts.
3. Leveling feet: The leveling feet can be turned to adjust the height. Adjust the feet so that the incubator
is level.
4. Trays: The trays can be pulled out toward you.
5. Tray supports: The tray supports can be removed by lifting the front side and pulling toward you.
6. Side supports: The right and left side supports can be removed for disinfection. Refer to page 19 and
23.
7. Fan cover: The fan cover serves as the inlet for circulating air. It is removable.
8. Duct: The duct for the path for circulating air. It is removable.
9. Fan (inside the duct): The fan is made from polypropylene resin. It can be disinfected in an autoclave.
10. Sample air outlet: The sample air outlet also functions as an internal gas outlet. Normally, cover this
outlet with the sample air outlet cap.
11. Connecting port A/B for CO2 gas pipe: When the optional MCO-21GC gas auto changer is installed,
both ports A and B are available. If the MCO-21GC is not installed, only port A is available. Refer to page
14 for gas cylinder connection. Ensure that the gas pressure is set at 0.03 MPa(G) (0.3 kgf/cm2(G), 4.3
psi(G)). Refer to page 58 for gas auto changer.
12. Door switch: Detects the door opening/closing and stops the fan and electromagnetic valve for CO2
when the door is open. The UV lamp* is also deactivated by the door opening*.
13. Humidifying pan: Fill the humidifying pan with sterile distilled water. Install the humidifying pan
properly so that it can be covered with the humidifying pan cover.
14. Humidifying pan cover: The cover prevents the UV light entering the chamber. When filling the
humidifying pan, lift the front side and take out the humidifying pan. Refer to page 21 for details.
15. UV lamp*: This UV lamp does not generate ozone. Never look directly at the UV light. For
replacement, contact our sales representative or agent.
16. Water level sensor: This sensor detects the water level in the humidifying pan. Refer to page 22 for
details.
17. Remote alarm terminals: Refer to page 11.
18. Access port: Place the silicon caps on the port both outside and inside when the port is not being
used.
19. Power switch: This is the main switch for the incubator. It also functions as an overcurrent breaker.
20. Glow starter*: The glow is started if for the UV lamp.
21. Attachment location for electric key: This is the attachment location for the electric key included in
the optional component H2O2 decon set (MCO-HL). This kit must be attached to perform H2O2
decontamination. Refer to the installation procedure for MCO-HL for details.
22. Sample air outlet cap: Always attach this cap except when using the sample air outlet.
* MCO-19AIC(UVH)/19AIC(UV) or when an optional UV system set (MCO-19UVS) is installed.
9
INCUBATOR COMPONENTS
Control panel and keypad
1
3
11
2
4
5
8
10
9
6
7
7. CE Key (CE):
Press this key to clear the entered value
when entering a setting.
1. LCD panel
2. High limit regulator:
This regulator is used to set the high
temperature limit.
8. Cursor Keys (Up, Down, Left, and Right):
Use these keys to move the cursor on the
LCD panel.
3. Overheat indicator (OVER HEAT):
This indicator lights when the chamber
temperature reaches the high limit.
9. Enter key (ENT):
Press this key to select a menu command.
When entering a setting, pressing this key to
move to the next parameter.
4. Menu key (MENU):
Press this key to access the menu.
5. Contrast knob (CONT.):
Turn this knob to adjust the contrast of the
LCD.
10. Numeric Keys
11. H2O2 Key (H2O2):
Press this key to start the H2O2
decontamination. Refer to page 41-44 for
details.
6. Buzzer key (BUZZER):
Press this key to silence the buzzer.
Note: It is not possible to silence the buzzer
for the high limit temperature alarm.
※ The following optional accessories must be installed to perform decontamination.
・UV system set (MCO-19UVS)
(This is provided as standard equipment for the MCO-19AIC(UVH)/19AIC(UV).)
・H2O2 decon set (MCO-HL)
・H2O2 generator (MCO-HP)
10
Remote alarm terminal
The remote alarm terminal is located at the rear right side of the incubator.
Remote alarm
terminals
The alarm is outputted from this terminal. Contact capacity is DC 30 V, 2 A.
Normal:
Open
Alarm:
Closed
Note:
• When the power switch is OFF or power is interrupted, the contact output will be closed.
• The remote alarm cannot be silenced by pressing the buzzer key (BUZZER) since the remote alarm is
not conjunct with the buzzer key (BUZZER).
REMOTE ALARM
MAX DC30V 2A
Remote alarm terminal
N.O.
11
INSTALLATION SITE
For correct operation of the incubator, install it in a location with the following conditions.
WARNING
When using CO2 gas for control, make sure that there is adequate ventilation. Using CO2 gas in a
small room without adequate ventilation may cause gas poisoning or oxygen deprivation. In addition,
when opening the incubator doors, do not directly inhale the air in the chamber.
Si l’appareil est utilisé dans un evdroit restreint, le niveau de la densite CO2 de l’air peut s’élever et peut
être nocif aux humains. Evitez d’aspirer l’air provenant de l’inérieur de l’appareil quand vous ouverz la
porte.
 Normal air environment
Install the incubator in an environment with normal air.
 Do not expose to direct sunlight
Do not install the incubator in a location where it will be exposed to direct sunlight. If the incubator is
operated in direct sunlight, performance will be adversely affected.
 Separate from heat sources
Do not install the incubator near significant heat sources, such as heaters, boilers, ovens, or autoclaves.
Heat will adversely affect the performance of the incubator.
 Ambient temperature at least 5 C lower than set temperature
The control temperature of the incubator is at least 5 C higher than the ambient temperature. For
example, if the chamber is controlled at 37 C, the ambient temperature must normally be no more than
32 C. Do not allow the ambient temperature to become too high.
 Strong and level floor
Select a site with a strong and level floor. If the floor is uneven or the installation is not level, the incubator
will be unstable and this may cause accident or injury. To avoid vibration and noise, always make sure
that the installation is stable. An unstable surface may result in vibration or noise.
WARNING
Install the incubator at a location that can support the weight. If the floor is not strong enough or if the
installation is insufficient, the incubator may fall over and cause injury.
Always make sure that the floor is strong, even, and level, and that the incubator will not tip over.
An insufficient installation may result in injury due to water leakage or the incubator falling over.
 Low humidity
Select a site with a relative humidity of 80 %R.H. or lower. Using the incubator in high humidity may result
in current leakage or electric shock.
WARNING
Do not use the incubator outdoors. If the incubator is exposed to rain water, it may result in current
leakage or electric shock.
Never install the incubator in a moist location, such as near a sink or water line, or where it is
likely to be exposed to water. In addition, do not install it near water or steam pipes. Moisture can
cause the insulation to deteriorate, which may result in current leakage or electric shock.
 No inflammable or corrosive gas
Never install the incubator in a location where it will be exposed to inflammable or corrosive gas. Doing so
may result in explosion or fire. In addition, insulation may deteriorate due to corrosion of protective casing,
resulting in current leakage or electric shock.
 No falling objects
Do not install the incubator in a location where there is the possibility of objects falling from above. Doing
so may result in damage or accident.
12
INSTALLATION
1. Remove the packing tape and clean up.
Remove all the tape that is securing the doors and the inner attachments. Open the doors for ventilation.
If the outer panels are dirty, dampen a cloth with a diluted neutral detergent and wipe them. (Using an
undiluted solution may damage the plastic. Follow the diluting instructions for the detergent that is used.)
Wipe off the residual detergent with a wet cloth and then wipe off any moisture.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
WARNING
Do not leave the plastic wrapping bags within reach of children. If the bag is placed over a child’s
head, it can block the mouth and nose and cause suffocation.
2. Adjust the leveling feet.
Adjust the leveling feet by turning them counterclockwise to level
the incubator. (See the illustration on the right.)
3. Ground the incubator.
Ground the incubator during installation to prevent electric shock
in case the insulation is not sufficient. If there is no ground wire at
the location, consult with qualified service personnel.
 When a ground must be installed
If a grounded 3-pole outlet is not available, then a ground must
be installed. Consult with qualified service personnel.
Leveling feet
WARNING
To prevent electric shock, always ground the incubator. If grounding is not possible, then have a ground
installed by qualified personnel. If the incubator is not grounded, it may result in electric shock.
Never connect the ground wire to a gas pipe, water pipe, lightning rod, or telephone ground wire.
Doing so may cause electric shock.
 Installing a ground fault circuit breaker
If using the incubator in the location with moisture or humidity cannot be avoided, then it is recommended
that a ground fault circuit breaker be installed in the power supply circuit (i.e., the power supply at the
incubator). Have the circuit breaker installed by qualified service personnel.
CAUTION
Do not climb on the incubator or place objects on top of it. Doing so may damage it or cause it to fall
over, resulting in injury. If it is to be stacked, refer to page 63 and stack it securely.
 When the incubator is not in use
Empty the water from the humidifying pan and remove moisture from the chamber. Make sure that the
chamber is completely dry before closing the doors. Failure to do so may result in damage.
 Before moving the incubator
Before moving the incubator, empty the water from the humidifying pan, disconnect the power supply plug
from the outlet, and make sure that the cord will not be damaged. Failure to do so may result in electric
shock or fire.
13
INSTALLATION
Connecting a CO2 gas cylinder
WARNING
When connecting a gas cylinder to the incubator, confirm the gas type. Confirm that the connections
are secure and that no gas will escape. Be sure to use the specified pressure. Using an incorrect
gas or pressure may result in explosion or fire, or in gas poisoning or oxygen deprivation due to escaping
gas.
Install the incubator in a location with adequate ventilation. If adequate ventilation cannot be
provided, then install an alarm system using CO2 and O2 densitometers.
1. Use a liquefied CO2 gas cylinder (at least 99.5 % pure). The siphon (dip tube) type cannot be used.
2. Install a gas regulator rated at 25 MPa(G) (250 kgf/cm2(G), 3600 psi(G)) for the primary side, and 0.2
MPa(G) (2 kgf/cm2(G), 30 psi(G)) for the secondary side.
3. Using the gas tube that is provided, connect the gas regulator to the connecting port A/B for CO2 gas
pipe at the rear left side of the incubator.
Note:
If the CO2 gas is supplied to multiple CO2 incubators from a single gas cylinder, a CO2 solid will be
formed in the gas regulator. The gas regulator safety valve will operate, and there may an explosive
sound.
When the MCO-21GC is not mounted
Using the gas tube that is provided, connect the gas regulator for the CO2 gas cylinder to connecting port
A for CO2 gas pipe located at the rear side of the incubator. After connecting the gas tube, make sure that
no gas is leaking.
When the MCO-21GC is mounted
Connect a pair of CO2 gas cylinders with gas regulators independently. The gas supply line will be
switched automatically. Connect the main cylinder to the connecting port A for CO2 gas pipe and the
reserve cylinder to the connecting port B for CO2 gas pipe. After connecting the cylinders, make sure that
no gas is leaking.
 For details on installing the optional automatic CO2 cylinder changeover kit (MCO-21GC), refer to the
MCO-21GC installation guide. For details on using the MCO-21GC, refer to page 58.
4. Set the CO2 gas on the secondary side to 0.03 MPa(G) (0.3 kgf/cm2(G), 4.3 psi(G)) for gas injection. As
the pressure increases, the CO2 gas density control range will increase. Excessive pressure may cause
gas supply lines inside the incubator to come loose, which may result in gas poisoning or oxygen
deprivation due to the escaping of gas. If gas lines come loose, the incubator must be repaired.
 The gas lines connected to the incubator will degrade over time. If any deterioration or abnormalities
are found during inspection, replace the lines immediately.
14
PREVENTING CONTAMINATION
To prevent contamination of the chamber, select a suitable installation site.
 Avoid locations with high temperatures or humidity.
Avoid locations with high temperatures or humidity, because of a greater presence of microorganisms in
the air.
 Avoid locations with passers-by or drafts.
Avoid locations near doors, air conditioners, fans, etc., where passers-by or drafts can facilitate the entry
of microorganisms into the chamber.
 If possible, use a cleanroom.
To achieve a better culture, it is recommended that a cleanroom be used if one is available.
 Use clean containers.
The greatest cause of contamination is dirty containers for cultures. Be careful not to get containers or
trays dirty when taking them in and out.
 Keep the chamber clean.
Wipe off any fingerprints. If water spills from the humidifying pan, or if the doors are left open for a long
time, condensation may form on the inside of the doors. If that occurs, wipe off the condensation with a
dry sterile gauze. In particular, clean and disinfect the chamber if the culture medium is spilled. For details,
refer to “ROUTINE MAINTENANCE” on page 23.
 Use sterile distilled water in the humidifying pan.
Always use sterile distilled water in the humidifying pan. Do not use ultrapure water, because it may
cause red rust-like particles in the humidifying pan.
The water level alarm is displayed in the status display area on the control panel. Quickly refill the water
in the humidifying pan when the water level alarm is displayed. Adding low-temperature water will
significantly lower the temperature in the chamber. Clean the humidifying pan once a month.
 Keep the incubator out of direct airflows from air conditioners or fans.
Cool airflow from an air conditioner may cause condensation and lead to possible contamination.
15
PRECAUTIONS FOR CULTURES
 Leave space between culture containers.
Always leave space for ventilation between culture containers (Petri dishes, flasks, etc.). Inadequate
spacing may result in uneven temperature distribution and CO2 gas density.
 Do not place harmful materials in the chamber.
Never place samples that release acidic, alkali, or corrosive gas in the chamber. Doing so may cause
damage resulting from discoloration or corrosion.
 Close the inner door.
Always close the inner door before closing the outer door. Failure to close the inner door will adversely
affect performance even if the outer door is closed.
 Open and close the doors gently.
Always open and close the doors gently. Closing the doors forcefully may cause spillage of the culture
medium, incomplete closing, or damage to the gasket. Before opening the inner door, check through the
glass to confirm that the UV lamp is OFF (if the MCO-19AIC(UVH)/19AIC(UV) or the optional
MCO-19UVS is installed).
 Be careful when closing the outer door.
Use the handle when closing the outer door. Holding the door in other places may cause injury by getting
fingers caught in the door. Do not lean on the outer door. Doing so may result in injury from the outer door
coming loose or the incubator falling over, or it may cause current leakage or electric shock.
 Be careful of the inside of the outer door.
The inside of the outer door may become hot.
 Avoid using excessive force on the inner door.
Do not put your hand on the glass, poke it with sharp objects, or apply strong force. Doing so may result
in injury from breaking the glass.
 Check the cause of any alarm buzzer.
If an alarm buzzer sounds while the incubator is in use, immediately check the cause of the alarm. For
details on what may cause an alarm buzzer to sound, refer to page 46.
Using unlock key
 Unlocking when power is interrupted
If power is interrupted to the MCO-19AIC(UVH) or MCO-19AIC(UV)/19AIC with an MCO-HL installed, the
outer door is electrically locked. To unlock the outer door while the power is interrupted, use the unlock
key that is provided. To lock the outer door again, turn the unlock key to the lock direction while the outer
door is open. After the outer door has been locked condition manually, then close the outer door.
Note:
The outer door cannot be locked by using the unlock key while the outer door is closed. Lock the outer
door while it is open. Attempting to turn the unlock key while the outer door is closed may damage the
electric lock system.
WARNING
For the MCO-19AIC(UVH) or MCO-19AIC(UV)/19AIC with an MCO-HL installed, the outer door is
electrically locked during decontamination. The chamber is filled with H2O2 gas during decontamination.
Do not unlock the outer door during decontamination.
16
INITIAL CLEANING METHOD
Before using the incubator for the first time, clean dirt (tape residue, smear, etc.) from the
chamber and the inner attachments thoroughly. To keep the chamber clean is essential to
get the proper performance out of the incubator. Use the following steps to clean the
incubator properly.
1. Remove the inner attachments, referring to “REMOVING INNER ATTACHMENTS” on page 18.
2. Clean the removed inner attachments, the chamber inside walls and the inner door gaskets with a cloth
or sponge soaked in neutral detergent, diluted by 5 % or less. (Undiluted detergent can damage the
plastic components. For the dilution, refer to the instruction of the detergent.) (Fig. 1)
CAUTION
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
Be careful to keep the diluted detergent or water out of the temperature sensor, the CO2 gas injection port,
the inner sample air access port, the fan motor shaft bearing, and the inner sample air outlet
(Fig.2
). Also, do not wash the temperature sensor, the water level sensor and the UV lamp using
detergent. Doing so may cause failure. (Fig. 2)
3. Soak a gauze or unwoven cloth in distilled water and wring it tightly, and then wipe off the residual
detergent thoroughly.
4. Wash the silicon caps (2 pcs) for the access port and the fan using the above mentioned detergent and
rinse them with distilled water, and then autoclave them for sterilization (121 oC, 20 minutes).
5. Wipe up the inside walls and the inner attachments like trays thoroughly with a cloth or unwoven cloth
soaked in alcohol for disinfection. Be careful not to leave any residue alcohol.
6. Reinstall the inner attachments correctly and securely, referring to “INSTALLING INNER
ATTACHMENTS” on page 20.
Silicon
cap
Temp. sensor
(UV lamp)
Water level sensor
Fig. 1
Fig. 2
17
REMOVING INNER ATTACHMENTS
CAUTION
Wear rubber gloves when performing maintenance on the chamber. Failure to wear gloves may
result in cuts or abrasions from sharp edges or corners.
Be careful not to damage the water level sensor or the UV lamp in the duct (if the
MCO-19AIC(UVH)/19AIC(UV) or the optional MCO-19UVS is installed).
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
1. Turn OFF the power to the incubator.
2. Open the outer and inner doors and pull out all the trays. (Fig. 1)
3. Lift the front of the tray supports and pull them out. (Fig. 2)
4. Lift the humidifying pan cover off from the pins on the rear side. (Fig. 3)
5. Pull out the humidifying pan. (Fig. 4)
6. Loosen the two screws securing the fan cover and take off the fan cover. (Fig. 5)
7. Lift the duct and remove it from the pins on the rear side. (Fig. 6)
Pins
Tray support
Humidifying pan cover
Fig. 1
Fig. 2
Fig. 3
Duct
Fan cover
Humidifying pan
Fig. 4
Fig. 5
18
Fig. 6
8. Remove the fan by pulling out the central spring and then by pulling out the fan. (Fig. 7)
9. Remove the screw securing the clamp for the side support. (Fig. 8)
10. Remove the clamp. (Fig. 9)
11. Lift the side supports off of the pins. (Fig. 10)
12. Remove the silicon caps of the access port, one each from interior (Fig. 11) and exterior (Fig. 12).
①Pull out the central spring
Clamp
Screw
②Pull out the fan
Fig. 7
Fig. 8
Fig. 9
Silicon cap
Silicon cap
Side
support
Fig. 10
Fig. 11
19
Fig. 12
INSTALLING INNER ATTACHMENTS
Use the following steps to install the inner attachments properly.
1. To reinstall all the attachments, perform the procedure in reverse order from step 12 on page 19.
2. When installing the fan, insert it to the motor shaft securely. Lightly turn the fan manually to make sure
that it does not touch the rear panel. Then install the central spring. (Fig. 1)
CAUTION
If the fan is not inserted deep enough, the intended velocity performance cannot be achieved, which may
cause culture failure or insufficient decontamination.
3. To install the duct, confirm 4 pins are securely installed in the 4 holes of the duct. (Fig. 2)
CAUTION
If the duct is fixed insecurely, the intended velocity performance cannot be achieved, which may cause
culture failure or insufficient decontamination.
① Push the fan deeply.
② Install the central spring afterwards.
③ Lightly turn the fan manually to make
sure that it does not touch the rear panel.
Fig. 1
Fig. 2
4. To install the side support, confirm the direction of fixing holes as shown in Fig. 3.
5. As shown in Fig. 4, set the tray with only the front edge bent down. If the tray is set in the wrong
direction, it may not be level and may become unstable.
Bent down
Fig. 3
20
Fig. 4
FILLING HUMIDIFYING PAN
Use the following procedure to fill the humidifying pan or to replace the water.
1. Lift the humidifying pan cover. (Fig. 1)
2. Pull out the humidifying pan toward you. (Fig. 2)
3. Dispose of the remaining water in the humidifying
pan and clean the humidifying pan with a diluted
detergent. Then rinse it thoroughly with distilled
water and wipe it with alcohol for disinfection.
4. Wipe all moisture from the bottom of the chamber.
Fig. 1
5. Return the humidifying pan to the chamber and
pour sterile distilled water (approx. 1.5 L, preheated
to 37 C). (Fig. 3)
CAUTION
Install the humidifying pan in a longitudinal direction
as its shorter side is placed in the back. Improper
installation may cause condensations.
Fig. 2
6. Set the humidifying pan with the inner side flush
against the back, and replace the humidifying pan
cover. Close the inner door and the outer door, and
confirm that RH PAN is not displayed in reverse
video in the status display field.
Note:
• Preheat the water to 37 C to be poured into the
humidifying pan. Adding low-temperature water will
lower the temperature and humidity in the chamber.
• If the water level alarm indication “RH PAN” is
displayed, use the above procedure to replenish the
water.
Fig. 3
CAUTION
When refilling the water in the humidifying pan, always wipe off any dirt from the water level sensor
with alcohol for disinfection. While doing that, be careful not to apply excessive force to the sensor
lead wire.
Caution:
In order to adjust the humidity inside the incubator chamber optimally and prevent condensation on the
chamber surface and the inner door, it is designed to create the lower temperature area underneath the
humidifying pan to recondense the evaporated moisture in the humidifying pan. Although the
condensation may be seen round the humidifying pan on the bottom of chamber (inside the humidifying
pan cover), it is not a malfunction.
21
WATER LEVEL SENSOR
This incubator is equipped with a water level sensor. The water level sensor is set automatically when the
humidifying pan is installed. Take care not to damage the water level sensor when removing or installing
the humidifying pan.
When the humidifying pan is removed (side view)
Water level sensor
Humidifying pan
When the humidifying pan is installed (side view)
Note: Make sure that the humidifying pan is pushed all the way to the back and that the water level
sensor comes down when you install the humidifying pan.
Note:
• Lift the water level sensor before installing the humidifying pan if the water level sensor is in the lower
position after maintenance.
• When installing the humidifying pan, make sure that the humidifying pan is set properly and that water
level sensor comes down into the humidifying pan. “RH PAN” will be displayed in reverse video in the
status display field on LCD panel if the water level sensor does not come down completely. If necessary,
set the humidifying pan again in the proper location.
• The water level sensor detects the water level every 30 minutes and just after the outer door is closed. It
takes several seconds to detect the water level. Therefore, “RH PAN” may be displayed in reverse video
several times in the status display field on LCD panel after the outer door is closed even when the
humidifying pan is full.
CAUTION
Foreign particles on the water surface can adhere to the water level sensor and fittings by capillary action
because the water level sensor is always in the water. The adhered foreign particles degrade water level
sensor performance and “RH PAN” may be displayed in reverse video in the status display field on LCD
panel even though there is sufficient water in the humidifying pan. Be sure to wipe OFF any dirt on the
water level sensor with alcohol for disinfection whenever you change the humidifying water. When
cleaning the water level sensor, take care not to apply excessive force to the lead wires.
22
ROUTINE MAINTENANCE
To use this unit in a clean condition, clean the chamber and all the inner attachments at least once a
month.
1. Remove all the inner attachments by the procedures shown on page 18.
2. Clean the chamber and all the inner attachments by the procedures shown on page 17.
3. Install all the inner attachments by the procedures shown on page 20.
● When there is excessive dirt, contact our sales representative or agent.
CORRECT OPERATION
Use the following procedure to start trial operation or actual operation of the incubator.
1. Install the incubator correctly, referring to “INSTALLATION” on page 13.
2. Remove the packing materials from the chamber and inner attachments. Clean and disinfect the
chamber and all the inner attachments, referring to above “ROUTINE MAINTENANCE”.
3. Add approximately 1.5 L of sterile distilled water to the humidifying pan. (Refer to page 21.)
4. Connect the power supply cord to the outlet. Turn ON the power switch on the left side of the rear panel
of the incubator.
Note:
The humidity in the incubator chamber is adjusted to the optimum setting. To prevent condensation on the
surface inside chamber and the inner door, there is a low-temperature area under the humidifying pan in
the bottom of chamber to recondense evaporated moisture. Condensation may occur around the
humidifying pan at the bottom of the chamber (on the inside of the humidifying pan cover), but this does
not indicate a problem.
23
LCD PANEL
The following display (called the “Top Display”) will appear when the power switch is turned ON. The
default temperature is 37.0 oC and the default CO2 density is 0 %. The date and time are preset at the
factory. Refer to page 36 to change the date and time.
1
1 T e mp
2
2
3
4
5
3
0 0
1 2
1
1
3 7 . 0
o
1
2
1
3
C
1 1
4 5
1
6
1
7
C O 2
S t a t u s
7 6
O K
2
0
2
1
2
2
2
3
5 . 0 %
2
4
2
5
A
2
6
2
7
2
8
2
9
3
0
3
1
3 3 3
2 3 4
3
5
3 3
6 7
3
8
3 4
9 0
B
5.0
37.0
6 5
1 1
8 9
U V
R H
P A N
4
D o o r : O p e n
7
1. Display of set value of temperature
The set value of chamber temperature is displayed.
2. Display of current temperature
The current chamber temperature is displayed.
3. Display of set value of CO2 density
The set value of the chamber CO2 density is displayed.
4. Display of current CO2 density
The current chamber CO2 density is displayed.
5. Display of current CO2 cylinder
A and B will be displayed if an optional gas auto changer (MCO-21GC) is installed. The port of the CO2
gas pipe that is currently supplying CO2 will be displayed in reverse video. Nothing will be displayed if the
MCO-21GC is not installed.
6. Status display field
Various status or alarms are displayed.
・When the UV lamp is lit: “UV” is displayed in reverse video. (If the MCO-19AIC(UVH)/19AIC(UV) or the
optional MCO-19UVS is installed).
・When the humidifying water is low: “RH PAN” is displayed in reverse video and blinks.
・When the door is open: “Door: Open” is displayed in reverse video.
7. Message display field
A message is displayed when fault occurs. The message is displayed alternately in normal characters
and reverse video. Refer to page 46-48 for alarm details. “OK” is displayed during normal operation.
24
BASIC OPERATIONS ON CONTROL PANEL
The following operations are possible through the control panel:
 Setting the temperature・・・The chamber temperature can be set (page 26).
 Setting the CO2 density・・・The chamber CO2 density can be set (page 26).
 Setting the key lock・・・Changing the temperature and CO2 density can be disabled (page 27).
 Setting alarms・・・The temperature alarm, the CO2 density alarm (page 31), and the high limit alarm
temperature (page 30) can be set.
 Setting UV parameters・・・ON/OFF parameters and the ON period of the UV lamp can be set for the
MCO-19AIC(UVH)/19AIC(UV) or when the incubator is equipped with the optional MCO-19UVS (page
32).
 Setting other parameters・・・The initial settings of the date, time, and the log cycle (page 36), and the
LCD display and baud rate (page 38) can be set.
 Displaying the operation log and transferring data・・・A graph of past operation data can be
displayed (page 39) and operation data can be transferred to a PC (page 40).
 H2O2 decontamination・・・H2O2 decontamination of the chamber and the inner attachments can be
performed when the optional H2O2 generator (MCO-HP) is installed (page 42).
 Automatic CO2 cylinder changeover・・・The cylinder can be switched when the optional gas auto
changer (MCO-21GC) is installed (page 58).
 Setting the CO2 standard gas density・・・The CO2 standard gas density can be set when the optional
STD gas auto calibration kit (MCO-SG) is installed. (page 60)
25
BASIC PARAMETERS
Setting chamber temperature and CO2 density
The setting procedure for the chamber temperature and the CO2 density are given below. (Default
settings: Chamber temperature: 37 oC, CO2 density: 0 %)
The incubator automatically starts operation using these settings after the power is turned ON.
1. From the “Top Display”, press the menu key (MENU) to access the menu. Select “Set” and press the
enter key (ENT).
1 T e mp
2
3
3 7 . 0
o
C
C O 2
5 . 0 %
M E N U
5.0
37.0
S e t
L o g
4
5
S t a t u s
6
O K
T o o l s
7
2. The “Stand-by Setting” display will be displayed. Set the parameters.
1
S t a n d - b y
S e t t i n g
2
T e mp e r a t u r e
3
C O 2
D e n s i t y
5
K e y
L o c k
6
H i g h
3 7 . 0
o
C
5 . 0 %
( 0 . 0
o
( 0 . 0 %
M
C
O K
2 0 . 0 %
C a
C - 5 0 . 0
-
o
E N U
)
)
n c e l
4
L i mi t
0
5 2 . 0
( 0 . U n l o c k
o
1 . L o c k )
C
The setting ranges of the parameters are as follows:
●Temperature: 0 oC to 50 oC. To 36.5 oC, enter 365.
●CO2 density: 0 % to 20 %. To set 5.0 %, enter 050.
3. Press the menu key (MENU) to display the menu after setting the parameters. Select “OK” and press
the enter key (ENT). The settings will be saved.
Note:
When starting the incubator for the first time or after not using it for an extended period of time, allow at
least 4 hours for the chamber temperature, the humidity, and the CO2 sensor to become stable after
setting the desired chamber temperature at a 0 % CO2 density. Then change the setting to the desired
CO2 density.
26
Setting key lock
1. To set the key lock, change the value of the key lock parameter from 0 to 1 on the “Stand-by Setting”
display and press the enter key (ENT). The buzzer will sound briefly and the keys will be locked.
1
S t a n d - b y
S e t t i n g
2
T e mp e r a t u r e
3
C O 2
D e n s i t y
5
K e y
L o c k
6
H i g h
3 7 . 0
o
C
5 . 0 %
( 0 . 0
o
( 0 . 0 %
M
C
O K
2 0 . 0 %
C a
C - 5 0 . 0
-
o
E N U
)
)
n c e l
4
1
L i mi t
5 2 . 0
( 0 . U n l o c k
o
1 . L o c k )
C
2. “Key Lock” will be displayed in reverse video at the right side of “Stand-by Setting”, and changing the
temperature and CO2 density settings will be disabled.
Removing key lock
1. To remove the key lock, change the value of the key lock parameter from 1 to 0 on the “Stand-by
Setting” display and press the enter key (ENT).
1
S t a n d - b y
S e t t i n g
2
T e mp e r a t u r e
3
C O 2
D e n s i t y
5
K e y
L o c k
6
H i g h
3 7 . 0
o
C
5 . 0 %
K e y
L o c k
( 0 . 0
o
( 0 . 0 %
C - 5 0 . 0
-
o
C )
2 0 . 0 %)
4
0
L i mi t
5 2 . 0
( 0 . U n l o c k
o
1 . L o c k )
C
2. The cursor will move to the Password field. Input the 4-digit password (default: 0000) and press the
enter key (ENT). When key lock is release, the buzzer will sound briefly, and “Key Lock” will disappear
from the “Stand-by Setting” display.
1
S t a n d - b y
S e t t i n g
2
T e mp e r a t u r e
3
C O 2
D e n s i t y
K e y
L o c k
3 7 . 0
K e y
o
C
5 . 0 %
L o c k
( 0 - 5 0 . 0
( 0 . 0 %
o
-
4
H i g h
L i mi t
0
5 2 . 0
P a s s w o r d
o
C
27
C )
2 0 . 0 %)
BASIC PARAMETERS
3. Press the menu key (MENU) to display the menu, select “OK” and press the enter key (ENT). The
settings will be saved.
Note: The buzzer will sound for a long time if the wrong password is entered. Enter the correct password.
The password for releasing the key lock must be shared and administered among all users of the
incubator. The password is set to 0000 when the incubator is shipped from the factory. The procedure for
changing the password is given next.
Setting key lock password
1. From the “Top Display”, press the menu key (MENU) to display the menu, select “Tools”, and press the
enter key (ENT).
1 T e mp
2
3
3 7 . 0
o
C
C O 2
5 . 0 %
5.0
37.0
4
5
S t a t u s
6
O K
M E N U
S e t
L o g
T o o l s
7
2. Select “Key Lock PW Setting” from the “Select Tools” display, press the menu key (MENU) to display
the menu, select “OK”, and press the enter key (ENT).
S e l e c t
T o o l s
T e mp / C O 2
A l a r m
C a l i b r a t i o n
S e t t i n g
L C D / D A Q
U V
( 1 / 2 )
O K
S e t t i n g
S v C
S e t t i n g
K e y
D a t e
L o c k
P W
M E N U
C a n c e l
S e t t i n g
T i me
28
3. Input the current user password (4 digits), select “OK”, and press the enter key (ENT). (The default
user password is 0000 when the incubator is shipped from the factory.)
K e y
L o c k
C u r r e n t
P W
S e t t i n g
U s e r
P a s s w o r d
****
4. Enter the new user password (4 digits), select “OK”, and press the enter key (ENT).
K e y
L o c k
P W
S e t t i n g
N e w
U s e r
P a s s w o r d
****
5. Enter the new user password again, select “OK”, and press the enter key (ENT) to save. After that
press the menu key (MENU) to display the menu, and select “Cancel“ from the menu.
K e y
L o c k
P W
N e w
U s e r
P a s s w o r d
****
U s e r
****
R e
E n t e r
S e t t i n g
P a s s w o r d
Note:
Be careful not to forget the key lock password. If you have forgotten the password and cannot release the
key lock, contact our sales representative or agent.
29
BASIC PARAMETERS
Setting high limit alarm temperature
The high limit temperature alarm is provided with the incubator. The alarm temperature can be changed
by using the following procedure.
1. In the “Stand-by Setting” display, turn the high limit regulator on the control panel using a small
screwdriver to set the desired high limit alarm temperature. The high limit temperature can be set
between 34 oC and 52 oC.
1
S t a n d - b y
S e t t i n g
2
T e mp e r a t u r e
3
C O 2
D e n s i t y
5
K e y
L o c k
6
H i g h
3 7 . 0
o
C
( 0 . 0
o
( 0 . 0 %
5 . 0 %
M
C
O K
2 0 . 0 %
C a
C - 5 0 . 0
-
o
E N U
)
)
n c e l
4
L i mi t
( 0 . U n l o c k
0
5 2 . 0
o
1 . L o c k )
C
Note:
・Set the high limit alarm temperature to at least 5 oC higher than the chamber set temperature.
・Set the high limit alarm temperature to at least 50 oC when H2O2 decontamination is performed.
2. After setting the parameter, press the menu key (MENU) to display the menu, select “OK”, and press
the enter key (ENT). The alarm temperature will be saved.
Refer to the tables of “ALARMS, SAFETY, AND SELF-DIAGNOSIS” on page 46 for details.
Note:
・The high limit alarm temperature will be changed at any time by turning the high limit regulator even if
the “Stand-by Setting” display is not displayed.
・Set the high limit alarm temperature after the chamber reaches the set temperature for operation.
30
ALARM PARAMETERS
1. Press the menu key (MENU) from the “Top Display” to display the menu, select “Tools”, and press the
enter key (ENT).
1 T e mp
2
3
3 7 . 0
o
C
C O 2
5 . 0 %
M E N U
5.0
37.0
S e t
L o g
4
5
S t a t u s
6
O K
T o o l s
7
2. Select “Alarm Setting” from the “Select Tools” display, press the menu key (MENU) to display the menu,
select “OK”, and press the enter key (ENT).
S e l e c t
T o o l s
T e mp / C O 2
A l a r m
M E N U
C a l i b r a t i o n
S e t t i n g
L C D / D A Q
U V
( 1 / 2 )
O K
S e t t i n g
S v c
S e t t i n g
K e y
L o c k
D a t e
C a n c e l
P W
S e t t i n g
T i me
3. The “Alarm Setting” display will appear. The temperature alarm, the CO2 alarm, the alarm delay, the
ring back, and the door delay can be set on this display. The alarm buzzer can be silenced by pressing
the buzzer key (BUZZER). The alarm buzzer will sound again after the specified ring back time if the
condition that caused the alarm continues. The ring back time can be set.
1
A l a r m
3
T e mp
C O 2
S e t t i n g
A l a r m
A l a r m
±
1 . 0
o
±
0 . 5
%
C
( ± 1 . 0 ° C
( ± 0 . 5 %
4
5
6
A l a r m
D e l a y
R i n g
B a c k
D o o r
D e l a y
1 5 mi n
( 0
3 0 m i n
( 0 . O F F
2 mi n
( 1
・Temp Alarm: ±1.0 oC to ±5.0 oC (Default: ±1.0 oC)
・CO2 Alarm: ±0.5 % to ±5.0 % (Default: ±1.0 %)
・Alarm Delay: 0 minute to 15 minutes (Default: 15 minutes)
・Ring Back: 1 minute to 99 minutes, or OFF (Default: 30 minutes)
・Door (alarm) Delay: 1 minute to 30 minutes (Default: 2 minutes)
31
-
-
-
-±
M E N U
± 5 . 0 ° C )
O K
5. 0 %)
C a n c e l
1 5 mi n )
1 -9 9 m i n )
3 0 mi n )
UV LAMP PARAMETERS
The UV lamp is used with the MCO-19AIC(UVH)/19AIC(UV) or the MCO-19AIC with an MCO-19UVS UV
system set installed. Use the following procedures to make the settings.
Using UV lamp
The UV lamp is located inside the duct to sterilize the water in the humidifying pan and the air circulating
in the chamber. Observe the following points to use the UV lamp correctly.
 When all the inner attachments are installed correctly, only the inside of the duct and the inside of the
humidifying pan cover are exposed to UV light.
 Correctly install all of the inner attachments when starting a cell culture, and never turn ON the UV lamp
when the humidifying pan cover is removed.
 Always install the humidifying pan cover even when using the incubator without turning ON the UV
lamp. Leaving the cover uninstalled will affect the chamber temperature distribution and the humidity
recovery performance.
 The UV lamp stays lit for a preset period of time after the outer door is closed. The default setting is for
five minutes.
 If the outer door is not opened for at least 12 consecutive hours, the UV lamp will light for the preset
time period every 12 hours.
 The recommended replacement time for the UV lamp (i.e., when the UV output ratio drops to 60 % to
70 % of its initial value) is when the accumulated ON time reaches 1,000 hours. When the accumulated
ON time reaches approximately 1,000 hours, the “UV” will blink in the status display field on the LCD
panel (when the UV lamp is not lit) and “Warning :UV Life” is displayed in the message display field on the
LCD panel. It is recommended that the UV lamp be quickly replaced at this point. Contact our sales
representative or agent for information on replacing the UV lamp.
 If the UV lamp burns out, “Err18:UV Lamp Abnormal” will be displayed in the message display field. If
this occurs, replace the UV lamp. When replacing the UV lamp, replace the glow starter (type FG-7P) at
the same time. Contact our sales representative or agent for information on replacing the UV lamp.
 If the UV lamp burns out (“Err18:UV Lamp Abnormal” will be displayed in the message display field), it
will not be possible to perform H2O2 decontamination. Replace the UV lamp and the glow starter (type
FG-7P).
32
Precautions when using UV lamp
WARNING
Do not look directly at UV light. UV light is harmful to the eyes.
 Always use the humidifying pan and humidifying pan cover.
The humidifying pan and humidifying pan cover prevent UV light from escaping. Always use them even
when not humidifying. To check whether the UV lamp is lit, open the outer door and then press the door
switch with the inner door still closed. Visible blue light can be confirmed from the front of the humidifying
pan cover. UV light is harmful to the eyes, so do not light the UV lamp when the inner door or the
humidifying pan cover is open.
 Be careful when handling the UV lamp.
There is a UV lamp inside of the duct. Be careful not to damage it when installing or removing the inner
attachments, the humidifying pan, or the H2O2 generator.
33
UV LAMP PARAMETERS
Setting UV lamp ON period
Use the following procedure to change the setting of the UV lamp ON period.
1. Press the menu key (MENU) from the “Top Display” to display the menu, select “Tools”, and press the
enter key (ENT).
1 T e mp
2
3
3 7 . 0
o
C
C O 2
5 . 0 %
M E N U
5.0
37.0
S e t
L o g
4
5
S t a t u s
6
O K
T o o l s
7
2. Select “UV Setting” from the “Select Tools” display, press the menu key (MENU) to display the menu,
select “OK”, and press the enter key (ENT).
S e l e c t
T o o l s
T e mp / C O 2
A l a r m
C a l i b r a t i o n
M E N U
S e t t i n g
L C D / D A Q
U V
( 1 / 2 )
O K
S e t t i n g
S v c
S e t t i n g
K e y
L o c k
D a t e
P W
C a n c e l
S e t t i n g
T i me
3. The “UV Setting” display will appear. Set the UV timer for the UV lamp ON period.
1
U V
S e t t i n g
3
U V
T i me r
5
mi n
U V
L i f e
0
%
U V
T i me r
0
%
U V
2 4 h
0
( 0 . O F F
4
E x t .
+
( 0
-
3 0 mi n )
5
Mo d e
1 . O N )
6
4. Press the menu key (MENU) to display the menu, select “OK”, and press the enter key (ENT). The
parameter will be saved.
34
 The UV timer can be set from 0 minute to 30 minutes. The default setting is for 5 minutes. When the UV
timer is set to zero, the UV lamp will not light.
 If the outer door is opened while the UV lamp is lit, the lamp will turn OFF. Then, when the outer door is
closed, the lamp will light for the preset period of time.
 If the UV lamp ON time is set for longer than 10 minutes or if only the outer door is repeated opened
and closed, condensation may occur in the chamber and may affect temperature distribution. It will also
shorten the service life of the UV lamp.
 When replacing the UV lamp, contact our sales representative or agent.
Note: To compensate for the drop in UV ray output along with increased accumulated ON time of the UV
lamp, the incubator automatically extends the UV lamp ON time according to the accumulated ON time.
The automatically extended time is displayed as a percentage under “UV Timer Ext.” in the “UV Setting”
display.
Lighting UV lamp for 24 hours
If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to
decontaminate the chamber by lighting the UV lamp for 24 hours.
1. Remove all attachments from the chamber, including the trays, the tray supports, the side supports, the
fan cover, the duct, the fan, the humidifying pan, and the humidifying pan cover. Clean all the
attachments in an autoclave or with alcohol for disinfection.
2. Clean and wipe off inside the chamber with alcohol for disinfection.
3. Set the CO2 density to 0 % and set the UV 24-hour mode to 1 on the “UV Setting” display.
1
U V
S e t t i n g
3
U V
T i me r
5
mi n
U V
L i f e
0
%
U V
T i me r
0
%
U V
2 4 h
1
( 0 . O F F
4
E x t .
+
( 0
-
3 0 mi n )
5
Mo d e
1 . O N )
6
4. Press the menu key (MENU) to display the menu, select “OK”, and press the enter key (ENT). The
parameter will be saved.
・This procedure must be preformed only with the outer door closed and the UV lamp turned OFF.
・The UV lamp will light continuously for 24 hours after this parameter is set. The setting is canceled if the
outer door is opened. Perform the above procedure to set the UV 24-hour mode again if you open the
outer door.
・Install all attachments after completion of the UV 24-hour mode.
Note: Set the high limit alarm temperature to at least 10 oC higher than the chamber set temperature
when using the UV 24-hour mode. The UV 24-hour mode may cause the automatic set temperature
alarm because the temperature of the chamber will increase.
35
OTHER PARAMETERS
Setting date, time, and log interval
1. Press the menu key (MENU) from the “Top Display” to display the menu, select “Tools” and press the
enter key (ENT).
1 T e mp
2
3
3 7 . 0
o
C
C O 2
5 . 0 %
5.0
37.0
4
5
S t a t u s
6
O K
M E N U
S e t
L o g
T o o l s
7
2. Select “Date Time” from the “Select Tools” display, press the menu key (MENU) to display the menu,
select “OK”, and press the enter key (ENT).
S e l e c t
T o o l s
T e mp / C O 2
A l a r m
C a l i b r a t i o n
S e t t i n g
L C D / D A Q
U V
( 1 / 2 )
O K
S e t t i n g
S v c
S e t t i n g
K e y
D a t e
L o c k
P W
M E N U
C a n c e l
S e t t i n g
T i me
36
3. The “Date Time” display will appear. Set the date, time and log interval.
1
D a t e
T i me
D a t e
0 9 / 0 3 / 0 1
( Y Y /
T i me
1 0 : 0 5 : 0 0
( h h :m m :s s )
M E N U
3
4
O K
MM / D D )
C a n c e l
5
L o g
I n t e r v a l
6 mi n
( 2 - 3 0 mi n )
6
・Entering the date
Example for March 1, 2009: Enter 090301 in the “Date Field”.
・Entering the time
Example for 10:05:00: Enter 100500 in the “Time Field”.
・Entering the log interval
Example for 6 minutes: Enter 6 in “Log Interval Field”.
Note:
■ The default is 6 minutes.
■ The log interval can be set to between 2 minutes and 30 minutes.
■ Relation between the log interval and the period that can be saved
1: Log interval of 2 minutes -- About 5 days
2: Log interval of 6 minutes -- About 14 days
3: Log interval of 30 minutes -- About 70 days
For the data beyond the above period, the older data is deleted to save the newer data.
37
OTHER PARAMETERS
Initial settings (LCD/DAQ parameters)
1. Select “LCD/DAQ Setting” from the “Select Tools” display, press the menu key (MENU) to display the
menu, select “OK”, and press the enter key (ENT).
S e l e c t
T o o l s
T e mp / C O 2
A l a r m
C a l i b r a t i o n
M E N U
S e t t i n g
L C D / D A Q
U V
( 1 / 2 )
O K
S e t t i n g
S v c
S e t t i n g
K e y
D a t e
L o c k
P W
C a n c e l
S e t t i n g
T i me
2. The “LCD/DAQ Setting” display will appear. Set the initial setting for each parameter as necessary.
1
L C D / D A Q
S e t t i n g
3
L C D
B a c k
D A Q
4
5
C o l o r
M E N U
2 . Wh i t e )
1
( 1 . B l u e
S p e e d
0
( 0 . 2 4
D A Q
I D
0
( 0 . O F F
D A Q
M o d e
0
( 0 . L o c a l
0
( 0 . O F F
R e mo t e
A l a r m
2 . 9 6
O
3 K
. 3 5 0 )
1 - 2 C
5 a5 n) c e l
1 . R e mo t e )
1 . A c t i v e )
6
・LCD Back Color: Setting of the background color (1: Blue, 2: White)
Note:
DAQ is an external monitoring system of the chamber status. It is necessary to set the DAQ speed, DAQ
ID, and DAQ mode to use the optional communications software. Communications software is an
optional accessory. Contact our sales representative or agent for details.
38
DISPLAYING THE LOG
A log of the past chamber temperatures and CO2 densities can be displayed on a graph.
1. Press the menu key (MENU) from the “Top Display” to display the menu, select “Log” and press the
enter key (ENT).
1 T e mp
3 7 . 0
o
C
37.0
2
3
C O 2
5 . 0 %
5.0
4
5
S t a t u s
6
O K
M E N U
S e t
L o g
T o o l s
7
2. The log will be displayed with dots. Press the “Up Cursor key” and “Down Cursor key” to switch
between the temperature and CO2 density displays. Press the “Left Cursor key” and “Right Cursor key” to
scroll the displayed data (Left Cursor key: older data, Right Cursor key: newer data).
Temperature
5 0 ° C
1
2 0 0 9 / 0 3 / 0 1
T e mp
3
4
5
6 0
1 2
2 4
Switched with Up and Down Cursor keys
CO2 density
1
2 0 %
2 0 0 9 / 0 3 / 0 1
C O 2
3
4
5
6 0
1 2
2 4
39
DISPLAYING THE LOG
Transferring data
Use the following procedure to transfer the log data to a PC.
1. To transfer the log data for one day, press the menu key (MENU) to display the menu, select “PC 1D”,
and press the enter key (ENT). To transfer all of the log data, select “PC All” and press the enter key
(ENT).
1
5 0 ° C
2 0 0 9 / 0 3 / 0 1
3
4
T e mp
M E N U
P C
1 D
P C
A l l
C l e a r
5
C a n c e l
6 0
1 2
2 4
2. A “Progress” display will appear. On a HyperTerminal on PC, specify a transfer, text capture, and the
name of the file to save. Use TXT or CSY as the file name extension. Press the menu key (MENU) to
display the menu, select “Start”, and press the enter key (ENT).
The transfer will be started. “Finished” will be displayed when the transfer has been completed. Select
“Cancel“ from the menu after the transfer has been completed and press the enter key (ENT).
1
P r o g r e s s
M E N U
3
S t a r t
S e n d
l o g
d a t a
t o
P C .
C a n c e l
4
5
L o g
D a t a
6
F i n i s h e d .
2 0 0 9 / 0 3 / 0 1
Setting in PC side for transmission of log data (For Windows 2000 and Windows XP)
1. From the Windows Start Button, select Program - Accessories - Communications HyperTerminal to start the HyperTerminal. (If the HyperTerminal is not available from the Start Menu,
execute the following file: C:¥Program Files¥Windows NT¥hypertrm.exe.)
2. In the HyperTerminal Window, set a new connection, the name (for example, Matsushita), connection
settings, method of connection, COM1, properties of COM1, and port.
Baud rate: 9,600, Data bits: 8, Parity: None, Stop bits: 1, Flow control: Xon/Xoff.
(The communications condition of the incubator will be set automatically to the above settings when the
“Progress” display appears.)
Note: An optional interface board MTR-480 (RS-232C/RS-485) is required to transfer data.
40
H2O2 DECONTAMINATION
When the chamber has been contaminated, or when cleaning the chamber prior to starting a culture,
H2O2 decontamination can be performed using the H2O2 generator.
 The following products must be purchased separately in order to perform H2O2 decontamination. Before
performing H2O2 decontamination, make sure that these are correctly installed.
 UV system set (MCO-19UVS) (Not required for the MCO-19AIC(UVH)/19AIC(UV).)
 H2O2 decon set (MCO-HL)
 H2O2 generator (MCO-HP)
 The H2O2 decontamination function decontaminates the chamber and the inner attachments. Do not
place any objects other than the specified attachments in the chamber during decontamination.
 Before performing H2O2 decontamination, clean the chamber and the inner attachments with alcohol for
disinfection.
 Before performing H2O2 decontamination, make sure that the duct, the fan and the H2O2 generator are
securely attached. (Fig. 1 - 3 on page 52)
 The chamber temperature during H2O2 decontamination reaches 45 C. Before performing H2O2
decontamination, set the high limit temperature alarm to 50 C or higher.
 For the H2O2 reagent, use one bottle of the H2O2 reagent specified by our company (MCO-H2O2).
 After decontamination, dilute the residual H2O2 reagent in the H2O2 generator with a large quantity of
water and dispose of it, and rinse with distilled water. (Do not wash either the inside or outside of the H2O2
generator with alcohol.)
WARNING
Use the reagent specified by our company for H2O2 decontamination. Using a different H2O2 solution
may cause explosion or damage to the incubator, or insufficient decontamination.
Do not use chemicals other than the H2O2 reagent, such as alcohol. Doing so may result in damage
to the H2O2 generator.
WARNING
When performing H2O2 decontamination, make sure that the outer and inner doors are securely
closed. During H2O2 decontamination, plug the access hole with the silicon caps that are provided. Failure
to do so may be harmful to health due to leakage of H2O2 gas.
CAUTION
H2O2 decontamination can be performed only for the chamber and inner attachments with standard
specifications, and not for any other objects.
CAUTION
Perform H2O2 decontamination with the inner attachments arranged as specified by our company.
Arranging them in a different way may result in insufficient decontamination.
CAUTION
Wear rubber gloves when handling the H2O2 reagent. Direct contact with the H2O2 reagent may result
in inflammation of the skin.
CAUTION
After H2O2 decontamination has been completed, residual H2O2 solution will remain on the bottom of the
chamber, the H2O2 generator, and the bottom of the duct. Wearing protective glasses and rubber
gloves, wipe it off with a non-woven cloth. Failure to do so may result in a deficient culture.
41
H2O2 DECONTAMINATION
Use the following procedure to perform H2O2 decontamination using the H2O2 generator (MCO-HP).
1. Take out all the trays, the tray supports, the side supports, the humidifying pan cover, the humidifying
pan, the fan cover, and the duct from the chamber. Dispose of the water in the humidifying pan, wipe the
inside walls of the chamber with a gauze containing water or alcohol for disinfection.
2. Attach the duct and side supports, and set the tray supports in the 2nd, 5th, 7th, 9th holes from the top
of the side supports. Then set the trays.
3. Pour one bottle of the H2O2 reagent (MCO-H2O2) into the H2O2 generator. Set the two pins on the
H2O2 generator in the 2 holes on the lower left side of the duct (Fig. 1).
4. Connect the H2O2 generator and connector on the bottom right of the far side of the chamber with the
enclosed cable. Be sure to keep the connector cap (Fig. 2).
5. Set the humidifying pan cover, the humidifying pan and the fan cover as shown in Fig. 3, and close the
inner door and the outer door.
Note: The trays included as accessory are designed to be appropriate for decontamination. If half tray
(MCO-25ST, optional) or trays for previous models are used, decontamination may not be sufficiently
effective.
Duct end
H2O2 generator
Front
Connector
Fig. 1
Fig. 2
Humidifying
pan cover
Fan cover
Humidifying pan
H2O2
Generator
Fig. 3
42
6. Press the H2O2 key (H2O2) for 5 seconds. The system check will start.
1
H 2 O 2
D e c o n t a mi n a t i o n
3
4
H 2 O 2
D e c o n
S y s t e m
C h e c k
・・・・・
5
6
7. If the system is normal, the following display will appear. (If the system is not normal, refer to Table 2
on page 47.) Select “OK” and press the enter key (ENT) to start H2O2 decontamination. Decontamination
will be performed automatically to step 10. The buzzer will sound when decontamination has been
completed (It takes about 1 hour 40 minutes after warming-up).
Note: The outer door will be locked with the electric lock for safety until the decontamination has been
completed.
1
H 2 O 2
D e c o n t a mi n a t i o n
M E N U
3
O K
C a n c e l
4
R e a d y
t o
S t a r t ?
5
6
8. During the H2O2 mist generation, “H2O2 Decon” will blink at the top left corner of the screen.
Note: H2O2 mist generates from the H2O2 generator and then quickly changes to a gas.
1 H 2 O 2
2
3
D e c o n
7min
U V
R e s o l v e
S t e p : 4
90min
4
5
T e mp : 4 5 . 0 ℃
6
D o o r : L o c k e d
7
43
H2O2 DECONTAMINATION
9. After completion of H2O2 mist generation, “UV Resolve” will blink at the top of the screen, and H2O2 gas
resolution by UV will start.
1 H 2 O 2
2
3
D e c o n
U V
R e s o l v e
S t e p : 7
89min
0min
4
5
T e mp : 4 5 . 0 ℃
6
D o o r : L o c k e d
7
10. After H2O2 gas resolution, H2O2 decontamination has been completed. The following display will
appear. Open the outer door and the inner door, then disconnect the cable of the H2O2 generator from the
connector in the chamber. Then take out the H2O2 generator.
Note: Always put on protection glasses and rubber gloves when performing decontamination.
1
H 2 O 2
D e c o n t a mi n a t i o n
S t e p : 8
2
D e co n
3
F i n i s h e d .
4
T e mp : 4 5 . 0 ℃
5
6
P l e a s e
R e mo v e
H 2 O 2
U n i t
D o o r : U n l o c k
7
11. The following display will appear. Press the menu key (MENU) to display the menu, select “OK”, and
press the enter key (ENT) to return to the “Top Display”.
1
H 2 O 2
D e c o n t a mi n a t i o n
S t e p : 8
2
D e co n
3
F i n i s h e d .
4
T e mp : 4 5 . 0 ℃
5
6
P U S H
ME N U
O K
D o o r : U n l o c k
7
44
12. Dilute the remaining H2O2 reagent in the H2O2 generator with a large
volume of water and dispose of it. Rinse and wash the H2O2 generator with
distilled water. Then keep the H2O2 generator in a clean environment outside
of the chamber.
Note: After H2O2 decontamination, cover the connector on the chamber side
with the connector cap deeply. (Fig. 1)
13. After H2O2 decontamination, surplus H2O2 liquid will remain at the bottom
of the chamber and in the bottom part of the H2O2 generator duct. This
solution contains H2O2 at a low density (about 1 % or less), so put on
protective glasses and rubber gloves and wipe it up with a non-woven cloth.
Fig. 1
14. Ventilate the chamber sufficiently and place all the attachments back into the chamber.
Note: Be sure to attach the connector cap.
CAUTION
The electric lock will remained locked if power fails during H2O2 decontamination. After the power comes
back ON, the H2O2 gas resolution process will start execution and finish automatically. Execute the
decontamination again because the decontamination will be incomplete.
WARNING
Never open the outer door by unlocking it with the unlock key during H2O2 decontamination or H2O2 gas
resolution with the UV lamp. H2O2 gas leakage is potentially harmful to health.
Precautions when handling H2O2 reagent
Observe the following precautions when handling the H2O2 reagent (MCO-H2O2).
 Handling precautions
 Wear protective equipment, such as protective glasses and rubber gloves.
 Do not use fire in the area where the reagent is being handled.
 Do not leave any reagent in the container after it has been used or while it is being used.
 Do not place inflammable or combustible materials near the area where the reagent is handled.
 Precautions for storage
 Always close the container cover securely to prevent impurities from becoming mixed in the reagent.
 Check the container to make sure that there is no damage, corrosion, or cracking.
 Store the container with the inlet facing upwards, and make sure that the container will not tip or be
knocked over.
 If possible, store the reagent in a location where water is available and where the floor can be easily
washed.
 Precautions for disposal
 Dispose according to the rule in your country.
45
ALARMS, SAFETY, AND SELF-DIAGNOSIS
The incubator supports the following alarms, safety functions, and self-diagnostic functions.
Table 1: Alarms, safety, and self-diagnosis for culture operations
Alarm or safety
function
High limit
temperature
alarm
Conditions
Display
“Err13: Main Heater Abnormal or
The chamber temperature exceeds the Err16: CO2S Box Heater Abnormal” is
high limit alarm temperature set value. displayed in the message display field.
Overheat indicator lights.
“Warning :High Temp” or
The chamber temperature is out of the
“Warning :Low Temp”
automatic set temperature alarm setting
is displayed in the message display
o
o
range (±1.0 C to ±5.0 C).
field.
Buzzer
Continuous tone
Safety operation
Heater OFF
Remote alarm
operates
Intermittent tone
after set alarm delay
time (0 min to 15
min) has elapsed
Intermittent tone
The chamber CO2 density is out of the “Warning :High CO2 Density” or
after set alarm delay
automatic set CO2 density alarm setting “Warning :Low CO2 Density” is
time (0 min to 15
range (±0.5 % to ±5.0 %).
displayed in the message display field.
min) has elapsed
There are no key operations for approx.
(Return to the “Top Display”.)
----90 s in a setting mode.
“Key Lock” is displayed on the standby
----The key lock is ON.
operation setting screen.
Intermittent tone
after set alarm delay
“Door:Open” is displayed
The outer door is open.
in the message display field.
time (1 min to 30
min) has elapsed
The setting mode is
canceled.
Setting changes are
prohibited.
The CO2 valve is
closed.
The heater turns
OFF after 1 min.
CO2 gas
cylinder empty
The CO2 density does not increase
when the CO2 valve is opened.
Remote alarm
operates.
CO2 line
changeover
“Err01: CO2 Gas Empty” is displayed
The CO2 gas supply is switched
in the message display field.
automatically to the reserve cylinder
The text blinks for the empty CO2 gas
(only when the MCO-21GC is installed).
supply line.
Automatic set
temperature
alarm
Automatic set
CO2 density
alarm
Auto-return
Key lock
Door alarm
Chamber
temperature
sensor error
“Err01: CO2 Gas Empty” is displayed
in the message display field.
Intermittent tone
〃
The chamber temperature sensor is
disconnected.
“Err05: Temp Sensor Open”
is displayed in the message display
field.
〃
The chamber temperature sensor is
short-circuited.
“Err06: Temp Sensor Short”
is displayed in the message display
field.
〃
“Err07: CO2S Box Temp Sensor Open”
is displayed in the message display
Sensor box
field.
temperature
“Err08: CO2S Box Temp Sensor Short”
sensor error
The sensor box temperature sensor is
is displayed in the message display
short-circuited.
field.
The ambient temperature sensor is
“Err09: At Sensor Open” is displayed in
Ambient
disconnected.
the message display field.
temperature
The ambient temperature sensor is
“Err10: At Sensor Short” is displayed in
sensor error
short-circuited.
the message display field.
“Err11: CO2 Sensor Vref Abnormal” or
CO2 sensor
The Vref or Vgas output voltage for the
“Err12: CO2 Sensor Gas Abnormal” is
CO2 sensor is abnormal.
error
displayed in the message display field.
Main heater burnout occurs or the main
“Err13: Main Heater Abnormal” is
Main heater
heater SSR is short-circuited when the
displayed in the message display field.
error
high limit temperature alarm operates.
“Err14: Humidity Heater Abnormal”
Bottom heater Bottom heater burnout occurs or the
is displayed in the message display
bottom heater SSR is short-circuited.
error
field.
Door heater
Door heater burnout occurs or the door “Err15: Door Heater Abnormal” is
heater SSR is short-circuited.
displayed in the message display field.
error
Sensor box heater burnout occurs or
“Err16: CO2S Box Heater Abnormal” is
Sensor box
the sensor box heater SSR is
displayed in the message display field.
heater error
short-circuited.
Heater SSR
Main, bottom, door, or sensor box
“Err17: Heater SSR Open” is displayed
heater SSR burnout occurs.
in the message display field.
burnout
Low humidifying The water in the humidifying pan drops “RH PAN” blinks in reverse video in the
to approx. 600 ml.
status display field.
water
The sensor box temperature sensor is
disconnected.
46
〃
〃
〃
〃
〃
Remote alarm
operates after delay
of 0 min to 15 min.
Remote alarm
operates after delay
of 0 min to 15 min.
The CO2 gas supply
is switched.
Remote alarm
operates.
The heater turns
OFF.
Remote alarm
operates.
The heater turns
OFF.
Remote alarm
operates.
CO2 valve is closed.
Remote alarm
operates.
CO2 valve is closed.
Remote alarm
operates.
Remote alarm
operates.
Remote alarm
operates.
CO2 valve is closed.
Remote alarm
operates.
〃
Remote alarm
operates.
〃
Remote alarm
operates.
〃
Remote alarm
operates.
〃
Remote alarm
operates.
〃
Remote alarm
operates.
-----
-----
Alarm or safety
function
Conditions
The UV lamp burns out (only for the
UV lamp failure MCO-19AIC(UVH)/19AIC(UV) or when
the optional MCO-19UVS is installed).
New UV lamp
recommended
The accumulated ON time reaches
approx. 1,000 h (only for the
MCO-19AIC(UVH)/19AIC(UV) or when
the optional MCO-19UVS is installed).
Display
Buzzer
Safety operation
“Err18: UV Lamp Abnormal” is
displayed in the message display field.
Intermittent tone
“UV” is displayed in the status display
field. (When the UV lamp is ON)
“UV” blinks in the status display field.
(When the UV lamp is OFF)
“Warning :UV Life” is displayed in the
message display field.
-----
Remote alarm
operates.
-----
 The buzzer can be stopped by pressing the buzzer key (BUZZER), but the remote alarms will not stop.
In addition, the buzzer that sounds because of the high limit temperature alarm cannot be stopped by
pressing the buzzer key (BUZZER).
 Err01 recovery is achieved automatically by correctly connecting the new CO2 gas cylinder and pressing
the buzzer key (BUZZER) to stop the buzzer. In addition, when the MCO-21GC is installed and the gas
supply is switched to the reserve gas cylinder, press the buzzer key (BUZZER) and replace the gas
cylinder.
 If an error from Err05 to Err18 occurs (except for the high limit temperature alarm in Err13 to Err16),
then contact our sales representative or agent.
Table 2: Alarms and Safety functions for H2O2 decontamination
Alarm and safety
function
Condition
Failure during H2O2
gas resolve by UV.
Safety operation
Decontamination mode is
cancelled.
-----
Decontamination mode is
cancelled.
The door is not closed.
“H2O2 Decon System Error” is displayed
in the screen center.
“Err33: Door is Open” is displayed in the
message display field.
-----
Decontamination mode is
cancelled.
Power was interrupted.
After power is restored, return to the
“Top Display”.
-----
System check error There is no H2O2 solution or
the H2O2 level sensor has
at start of H2O2
failed (or is disconnected).
decontamination
Failure during H2O2
decontamination
“H2O2 Decon System Error” is displayed
in the screen center.
“Err31:H2O2 Unit No Connect” is
displayed in the message display field.
“H2O2 Decon System Error” is displayed
in the screen center.
“Err32: H2O2 Level is Low” is displayed
in the message display field.
Buzzer
-----
The H2O2 generator is not
connected.
Failure during
warming-up
Display
Continuous
tone
(when the
The outer door is open.
outer door
is open.)
Intermittent
The volume of H2O2 generated “Err34:H2O2 Volume” is displayed in the
tone with
is abnormal.
message display field.
15min delay
After power is restored, “Err35: Power
Failed” is displayed in the message
Intermittent
Power was interrupted during
display field.
tone
H2O2 decontamination.
After completion, “Decon Failed” is
displayed in the screen center.
Continuous
“Err38:Door Lock Failed” is displayed in
tone
the message display field.
(when the
The outer door is open.
After completion, “Decon Failed” is
outer door
displayed in the screen center.
is open.)
The UV lamp failed during
“Err36: UV Lamp Failure” is displayed in Intermittent
H2O2 gas resolve by UV.
the message display field.
tone
“Err38:Door Lock Failed” is displayed in
the message display field.
After completion, “Decon Failed” is
displayed in the screen center.
Power was interrupted during
H2O2 gas resolve by UV.
After power is restored, “Err37: Power
Failed” is displayed in the message
display field.
The outer door is open.
“Err38:Door Lock Failed” is displayed in
the message display field.
After completion, “Decon Failed” is
displayed in the screen center.
47
-----
----Decontamination mode is
cancelled.
Remote alarm operates.
(when the outer door is open.)
Moves to UV resolve.
Remote alarm operates with 15
min. delay.
Electrically locked during power
interruption. After power is
restored, moves to UV resolve.
Remote alarm operates.
H2O2 mist generation is
cancelled.
Resolve time is extended.
Remote alarm operates.
(when the outer door is open.)
Resolve time is extended.
Remote alarm operates.
Electrically locked during power
interruption. After power is
restored, UV resolve is
repeated.
Continuous
tone
Resolve time is extended.
(when the
Remote alarm operates.
outer door (when the outer door is open.)
is open.)
ALARMS, SAFETY, AND SELF-DIAGNOSIS
Table 3: Alarm and Safety functions for automatic standard gas calibration
Alarm and safety
function
System check error
at start/operating of
automatic CO2
density calibration
Condition
The gas pressure of CO2 gas
line for the standard gas
calibration is less than the
specified pressure.
Display
“Std Gas Calib Error” is displayed in the
screen center.
“Err41: CO2 Std Gas Empty” is
displayed in the message display field.
Buzzer
Safety operation
The automatic CO2 density
Intermittent
calibration is cancelled.
tone
Remote alarm operates.
Note:
For models with electric locking, an unlock key is provided in order to unlock the outer door during a
power interruption or in case the electric lock fails. Always store this key in a safe place. It is
recommended that you make a note of the key symbol and number in case the key is lost.
WARNING
Do not use the unlock key to unlock the outer door during H2O2 decontamination or during H2O2
gas resolve by UV. Doing so may cause harm to health from H2O2 gas leakage.
48
CALIBRATION
Temperature/CO2 calibration
1. Press the menu key (MENU) from the “Top Display” to display the menu, select “Tools”, and press the
enter key (ENT).
1 T e mp
3 7 . 0
o
C
C O 2
5 . 0 %
3
M E N U
5.0
37.0
2
S e t
L o g
4
5
S t a t u s
6
O K
T o o l s
7
2. Select “Temp/CO2 Calibration” from the “Select Tools (1/2)” display, press the menu key (MENU) to
display the menu, select “OK”, and press the enter key (ENT).
S e l e c t
T o o l s
T e mp / C O 2
A l a r m
M E N U
C a l i b r a t i o n
S e t t i n g
L C D / D A Q
U V
( 1 / 2 )
O K
S e t t i n g
S v c
S e t t i n g
K e y
L o c k
D a t e
C a n c e l
P W
S e t t i n g
T i me
3. The “Temp/CO2 Calibration” display will appear.
Example
Enter the value shown below if the displayed chamber temperature is 37.0 oC but the actual temperature
is 36.5 oC.
●Temperature span: Enter 365.
0
1
1
0 0 0
2 3 4
0
5
0
6
0 0
7 8
0 1
9 0
1
1
T e mp / C O 2
1
2
1
3
1
4
1
5
1
6
1
7
1
8
2 2
0 1
2
2
2 2
3 4
2
5
2
6
2
7
2 2
8 9
3
0
3
1
3
2
3
3
3
4
3
5
3
6
3
7
3
8
3
9
4
0
C a l i b r a t i o n
P V
3
3 6 . 5
o
T e mp
S p a n
4
C O 2
Z e r o
0 . 1 %
5
C O 2
S p a n
0 . 0 %
C O 2
C O 2
6
1
9
C
3 7 . 0
V o l t
o
0 . 0 %
C
0 . 0 ° C
4 . 0 2
V
5 0 0
4 .
0 1
V
5 0 0
R e f
4 .
0 2
V
G a s
4 .
0 1
V
5 . 0 %
49
CALIBRATION
Example
Enter the value shown below if the displayed chamber CO2 density is 5.0 % but the actual CO2 density is
4.5 %.
●CO2 density span: Enter 045.
0
1
1
0 0 0
2 3 4
0
5
0
6
0 0
7 8
0
9
1
0
1
1
T e mp / C O 2
1
2
1
3
1
4
1
5
1
6
1
7
1
8
2
0
2
1
2
2
2 2
3 4
2
5
2
6
2
7
2 2
8 9
3
0
3
1
3
2
3
3
3
4
3
5
3
6
3
7
3 3
8 9
4
0
C a l i b r a t i o n
P V
3
o
T e mp
S p a n
0 . 0
4
C O 2
Z e r o
0 . 0 %
5
C O 2
S p a n
4 . 5 %
C O 2
C O 2
6
1
9
C
3 7. 0
V o l t
o
C
0 . 0 ° C
4 . 0 2
V
5 0 0
4 . 0 1
V
5 0 0
R e f
4 .
0 2
V
G a s
4 . 0 1
V
5 . 0 %
4. Press the menu key (MENU) to display the menu, select “OK”, and press the enter key (ENT).
5. After the “Select Tools (1/2)” display is displayed, press menu key (MENU) to display the menu, select
“Cancel”, and press the enter key (ENT) to return to the “Top Display”.
Note:
It is important to accurately measure the temperature inside the incubator when calibrating the
temperature. Particularly, the temperature gauge used must have an accuracy of JIS 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting must not change by more than ±1.0 oC during calibration. If these values are
exceeded, an error buzzer will sound, the input data will be ignored, and the incubator will return to the
display mode. If it is necessary to change the temperature by more than 1.0 oC, perform calibration in
several stages over a period of time.
50
TROUBLESHOOTING
If the incubator does not seem to be working properly, check the following items before calling for service.
Symptom
The incubator does not
operate at all.
Items to check and countermeasures
 Is the incubator plugged in?
The key is not working.
 Is the key lock function turned OFF?
An alarm has occurred.
When starting operation:
 Is there a power outage, or has a circuit breaker interrupted the
power?
 Does the chamber temperature match the set value?
 Does the CO2 gas density in the chamber match the set value?
(1) Is the secondary pressure for the gas regulator at the specified value
2
of 0.03 MPa(G) (0.3 kgf/cm (G), 4.3 psi(G))?
(2) Is the gas tube properly connected?
During operation:
 Is the high limit alarm temperature set at least 1 C higher than the
chamber set temperature?
 Has the temperature setting been changed? Has the outer door been
left open for a long time? Has a low-temperature object been placed in
the chamber? If any of these is the case, the alarm will be
automatically cleared if you wait.
 Has the gas tube come loose, or is there a gas leak?
 Has the CO2 gas density setting been changed?
 Is the gas cylinder empty? Check the primary pressure of the gas
cylinder once a week. (When the primary pressure is 3.8 MPa(G) (38
2
kgf/cm (G), 5.5 kpsi(G)) or lower, it is a sign that there is little gas
remaining. Replace the cylinder soon.)
 Is the incubator operating beside the appliance that generates the
electromagnetic wave?
The chamber temperature
does not match the set value.
 Is the ambient temperature less than 5 C different from the set value
for the chamber temperature?
 Is the outer door closed with the inner door left open?
The CO2 gas density does
not match the set value.
 Is the secondary pressure for the gas regulator at the specified value
2
of 0.03 MPa(G) (0.3 kgf/cm (G), 4.3 psi(G))?
 Is the gas tube blocked?
The chamber temperature
does not rise.
 Is there enough water in the humidifying pan? (Be sure to use sterile
distilled water.)
A large quantity of CO2 gas is
being consumed.
 Are the outer and inner doors being frequently opened and closed?
 Check whether gas is leaking from connectors due to deterioration of
the gas tube, or whether there may be any pinhole leaks. The gas tube
is a replaceable part, and it is recommended that it be replaced once a
year.
 Is the packing seal for the inner door defective?
 Is the access hole open?
Normal cultures are not
possible, and the CO2 gas
density is suspect.
 Is the ambient air environment around the incubator normal? Is there a
source of polluted gas in the vicinity?
51
TROUBLESHOOTING
Symptom
The CO2 gas density is slow to
recover.
Items to check and countermeasures
 A HEPA filter is used for the incubator CO2 gas piping. If gas
density is slow to recover when the CO2 gas pressure is normal, it
is possible that the HEPA filter may be clogged. Contact our sales
representative or agent.
 Is there little gas remaining in the CO2 gas cylinder?
 Is the secondary pressure for the gas regulator at the specified
2
value of 0.03 MPa(G) (0.3 kgf/cm (G), 4.3 psi(G))?
 Is the gas tube blocked?
CO2 gas is not being injected.
 The CO2 control method for the incubator is the ON-OFF method.
CO2 gas is intermittently injected as the gas density in the chamber
approaches the set value. Injections may be stopped for periods of
approximately 15 seconds, but that is not an error.
The outer door does not open.
MCO-19AIC(UVH), or when the optional MCO-HL is installed:
 When the power switch is OFF, the electric lock is locked and the
outer door does not open. Either turn ON the power switch or use
the accessory unlock key to override the electric lock.
 During decontamination the outer door is electrically locked and
does not open.
H2O2 decontamination cannot be
performed.
 If the MCO-19AIC(UV)/19AIC is being used, is the optional
MCO-HL installed?
 If the MCO-19AIC is being used, are the optional MCO-HL and
MCO-19UVS installed?
 Is the UV lamp burned out? If the UV lamp is burned out, H2O2
decontamination will not be possible.
 Is the H2O2 generator cable properly connected?
 Has the entire bottle of the H2O2 reagent been used?
 Has the high limit temperature alarm been set to 50 C or higher?
In H2O2 decontamination cycle,
”Err34:H2O2 Volume “ occurred.
 Is the duct securely attached? Attach the duct properly to the 4
points hooks. (Fig. 1)
 Is the fan attached properly? Remove the central spring at the
center of the fan and confirm if the fan is pushed all the way to the
motor shaft. Then, reinstall the central spring. (Fig. 2)
 Is the H2O2 generator securely installed? Set the 2 pins of the H2O2
generator in the 2 holes at the left bottom of the duct. (Fig. 3)
① Push the fan deeply.
Fig. 1
② Install the central spring
afterwards.
③ Lightly turn the fan manually to
make sure that it does not touch
the rear panel.
Fig. 2
Fig. 3
Note: If the problem still has not been solved after trying the above checks and countermeasures, or for
any problems not covered here, contact our sales representative or agent.
52
DISPOSAL OF UNIT
When disposing of the CO2 incubator, contact our sales representative or agent.
WARNING
The CO2 incubator must be dismantled and disposed of by qualified personnel only. If the CO2
incubator is left where outsiders enter, it may result unexpected accident (for example, children to become
locked inside).
Before disposing the CO2 incubator with biohazardous danger, decontaminate the CO2 incubator to the
extent possible by the user.
53
DISPOSAL OF UNIT
Note:
This symbol mark and recycle system apply only to EU countries. They
do not applied to the countries in other areas of the world.
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
(English)
This Panasonic product is designed and manufactured with high-quality materials and components that
can be recycled and reused.
This symbol means that the electrical and electronic equipment, at their end-of-life, should be disposed of
separately from household wastes.
Dispose of this product at your local community waste collection/recycling center.
In the European Union there are separate collection systems for used electrical and electronic products.
Please help us to conserve the environment we live in.
(German)
Ihr Panasonic Produkt wurde entworfen und hergestellt mit qualitativ hochwertigen Materialien und
Komponenten, die recycelt und wiederverwendet werden können.
Dieses Symbol bedeutet, daß elektrische und elektronische Geräte am Ende ihrer Nutzungsdauer von
Hausmüll getrennt entsorgt werden sollen.
Bitte entsorgen Sie dieses Gerät bei Ihrer örtlichen kommunalen Sammelstelle oder im Recycling Centre.
In der Europäischen Union gibt es unterschiedliche Sammelsysteme für Elektrik- und Elektronikgeräte.
Helfen Sie uns bitte, die Umwelt zu erhalten, in der wir leben!
54
(French)
Votre produit Panasonic est conçu et fabriqué avec des matèriels et des composants de qualité supérieure
qui peuvent être recyclés et réutilisés.
Ce symbole signifie que les équipements électriques et électroniques en fin de vie doivent être éliminés
séparément des ordures ménagères.
Nous vous prions donc de confier cet équipement à votre centre local de collecte/recyclage.
Dans l’Union Européenne, il existe des systèmes sélectifs de collecte pour les produits électriques et
électroniques usagés.
Aidez-nous à conserver l’environnement dans lequel nous vivons !
Les machines ou appareils électriques et électroniques contiennent fréquemment des matières qui, si elles
sont traitées ou éliminées de manière inappropriée, peuvent s’avérer potentiellement dangereuses pour la
santé humaine et pour l’environnement.
Cependant, ces matières sont nécessaires au bon fonctionnement de votre appareil ou de votre machine.
Pour cette raison, il vous est demandé de ne pas vous débarrasser de votre appareil ou machine usagé
avec vos ordures ménagères.
(Spanish)
Los productos Panasonic están diseñados y fabricados con materiales y componentes de alta calidad,
que pueden ser reciclados y reutilizados.
Este símbolo significa que el equipo eléctrico y electrónico, al final de su ciclo de vida, no se debe
desechar con el resto de residuos domésticos.
Por favor, deposite su viejo “televisor” en el punto de recogida de residuos o contacte con su
administración local.
En la Unión Europea existen sistemas de recogida específicos para residuos de aparatos eléctricos y
electrónicos.
Por favor, ayúdenos a conservar el medio ambiente!
55
DISPOSAL OF UNIT
(Portuguese)
O seu produto Panasonic foi concebido e produzido com materiais e componentes de alta qualidade que
podem ser reciclados e reutilizados.
Este símbolo significa que o equipamento eléctrico e electrónico no final da sua vida útil deverá ser
descartado separadamente do seu lixo doméstico.
Por favor, entregue este equipamento no seu ponto local de recolha/reciclagem.
Na União Europeia existem sistemas de recolha separados para produtos eléctricos e electrónicos
usados.
Por favor, ajude-nos a conservar o ambiente em que vivemos!
(Italian)
Il vostro prodotto Panasonic è stato costruito da materiali e componenti di alta qualità, che sono
riutilizzabili o riciclabili.
Prodotti elettrici ed elettronici portando questo simbolo alla fine dell’uso devono essere smaltiti
separatamente dai rifiuti casalinghi.
Vi preghiamo di smaltire questo apparecchio al deposito comunale.
Nell’Unione Europea esistono sistemi di raccolta differenziata per prodotti elettrici ed elettronici.
Aiutateci a conservare l’ambiente in cui viviamo!
56
(Dutch)
Panasonic producten zijn ontwikkeld en gefabriceerd uit eerste kwaliteit materialen, de onderdelen kunnen
worden gerecycled en weer worden gebruikt.
Het symbool betekent dat de elektrische en elektronische onderdelen wanneer deze vernietigd gaan
worden , dit separaat gebeurt van het normale huisafval.
Zorg ervoor dat het verwijderen van de apparatuur bij de lokaal erkende instanties gaat gebeuren.
In de Europese Unie wordt de gebruikte elektrische en elektronische apparatuur bij de daarvoor wettelijke
instanties aangeboden.
Alstublieft help allen mee om het milieu te beschermen.
(Swedish)
Din Panasonic produkt är designad och tillverkad av material och komponenter med hög kvalitet som kan
återvinnas och återanvändas.
Denna symbol betyder att elektriska och elektroniska produkter, efter slutanvändande, skall sorteras och
lämnas separat från Ditt hushållsavfall.
Vänligen, lämna denna produkt hos Din lokala mottagningstation för avfall/återvinningsstation.
Inom den Europeiska Unionen finns det separata återvinningssystem för begagnade elektriska och
elektroniska produkter.
Vänligen, hjälp oss att bevara miljön vi lever i!
57
AUTOMATIC CO2 CYLINDER CHANGEOVER
A gas auto changer (MCO-21GC) is available as an optional accessory. This kit switches the gas supply
lines when one of the CO2 gas cylinders becomes empty.
Note: The installation of the MCO-21GC must be performed only by qualified service personnel.
Install the MCO-21GC and then perform the following procedure.
1. Connect 2 gas tubes to the connecting ports A and B for CO2 gas pipe. The connecting ports A and B
for CO2 gas pipes are located on the left back of the incubator. (Refer to page 8).
Note: Only the port A can be used when the MCO-21GC is not installed. There will be 2 ports, the ports A
and B, after the MCO-21GC is installed.
2. Connect a CO2 gas cylinder equipped with a gas regulator to each gas tube. Refer to page 14 for
information on connecting the gas cylinders. (“Gas cylinder A” is the CO2 gas cylinder connected to the
connecting port A for CO2 gas pipe, and “gas cylinder B” is the CO2 gas cylinder connected to the
connecting port B for CO2 gas pipe.)
3. Open the valves of both gas cylinders.
4. The current CO2 gas supply line is displayed on the “Top Display”.
(“A” and “B” will be displayed only when the MCO-21GC is installed. Nothing will be displayed if
MCO-21GC is not installed.)
Current CO2 gas supply line
1 T e mp
2
3
3 7 . 0
o
C
37.0
C O 2
5 . 0 %
A
B
5.0
4
5
S t a t u s
6
O K
7
5. The MCO-21GC detects that there is no more CO2 gas in a cylinder when the CO2 density in the
chamber does not increase for a while even after opening the CO2 gas valve in the incubator. It switches
the gas supply line when it detects that there is no more gas. At that time, the indications of switching the
gas cylinder are as follows:
Message display field
When CO2 gas supply line switches
automatically (Only when MCO-21GC
is installed.)
Err01: CO2 Gas Empty is
displayed.
Display of gas supply
The character for gas supply line
that is empty blinks.
The gas supply line that is
currently being used is displayed
in reverse video.
●To silence the buzzer, press the buzzer key (BUZZER). The alarm condition is cleared and normal
operation continues.
58
6. When the gas cylinder A is switched to the gas cylinder B, remove gas cylinder A and replace it with
new one.
Note: Be careful when handling the empty CO2 gas cylinder. Some gas may still be left in the cylinder.
●When the gas cylinder B becomes empty, the line is switched to the gas cylinder A again.
The MCO-21GC detects that there is no more CO2 gas in a cylinder when the CO2 density in the chamber
does not increase for a while even after opening the CO2 gas valve in the incubator. The supply line may
also be switched for other reasons even though there is gas left in the cylinder: blocking or restricting of
gas tube, reduction of CO2 gas pressure, or improper opening of CO2 gas cylinder. Always check the gas
quantity in the cylinder before disconnecting it.
● Use the following procedure to manual switch the line between CO2 cylinders A and B (only when
MCO-21GC is installed). This procedure shows switching the line from the gas cylinder B to the gas
cylinder A as an example.
1. From the “Top Display”, press the “Left Cursor Key” for 5 seconds.
1 T e mp
2
3
3 7 . 0
o
C
C O 2
5 . 0 %
A
B
5.0
37.0
4
5
S t a t u s
6
O K
7
2. The CO2 gas supply line will be switched from the gas cylinder B to the gas cylinder A.
1 T e mp
2
3
3 7 . 0
o
C
37.0
C O 2
5 . 0 %
A
B
5.0
4
5
S t a t u s
6
O K
7
(When the line needs to be switched from the gas cylinder A to the gas cylinder B, press the “Right
Cursor Key” for 5 seconds.)
59
AUTOMATIC CO2 DENSITY CALIBRATION
To calibrate the CO2 density, a STD gas auto calibration kit (MCO-SG) is available. By connecting a CO2
standard gas cylinder that has the density that is the same as the set CO2 density (usually 5 % CO2), the
CO2 sensor can be calibrated to maintain the accuracy of the CO2 density display as necessary.
● Setting the connected standard CO2 gas density
1. Press the menu key (MENU) from the “Top Display” to display the menu, select “Tools”, and press the
enter key (ENT).
1 T e mp
2
3
3 7 . 0
o
C
C O 2
5 . 0 %
5.0
37.0
4
5
S t a t u s
6
O K
M E N U
S e t
L o g
T o o l s
7
2. Press the “Down Cursor Key” from the “Select Tools (1/2)” display to display the “Select Tools (2/2)”
display, and then select “Standard Gas Setting”, press the menu key (MENU) to display the menu, select
“OK”, and press the enter key (ENT).
S e l e c t
T o o l s
T e mp / C O 2
A l a r m
C a l i b r a t i o n
S e t t i n g
L C D / D A Q
U V
( 1 / 2 )
O K
S e t t i n g
S v c
S e t t i n g
K e y
D a t e
L o c k
P W
M E N U
C a n c e l
S e t t i n g
T i me
S e l e c t
T o o l s
( 2 / 2 )
S t a n d a r d
G a s
S e t t i n g
S t a n d a r d
G a s
C a l i b r a t i o n
M E N U
O K
S v c
C a n c e l
60
3. Input the CO2 density of standard gas to be used, press the menu key (MENU) to display the menu,
select “OK”, and press the enter key (ENT).
1
S t a n d a r d
G a s
S e t t i n g
M E N U
3
CO 2
ST D
G a s
D e n s i t y
5 . 0
4
%
O K
%
C a n c e l
5
6
● Calibration method for CO2 gas density
1. Select “Standard Gas Calibration” from the “Select Tools (2/2)” display, press the menu key (MENU) to
display the menu, select “OK”, and press the enter key (ENT).
S e l e c t
T o o l s
( 2 / 2 )
S t a n d a r d
G a s
S e t t i n g
S t a n d a r d
G a s
C a l i b r
at i o n
M E N U
O K
S v c
C a n c e l
2. The system check will start.
1
S t a n d a r d
G a s
C a l i b r a t i o n
S t e p :
3
S t d
G a s
C a l i b
S y s t e m
C h e c k ・・・・・
4
5
6
61
C O 2 :
0 . 0 %
S T G :
5 . 0 %
1
AUTOMATIC CO2 DENSITY CALIBRATION
3. If the system is normal, the following display will appear. Select “OK” and press the enter key (ENT) to
start calibrating the CO2 density. After calibration, return to normal cultivating operation.
1
S t a n d a r d
G a s
C a l i b r a t i o n
M E N U
3
O K
R e a d y
t o
S t a r t ?
C a n c e l
4
5
C O 2 :
0 . 0 %
S T G :
5 . 0 %
6
1
S t a n d a r d
G a s
C a l i b r a t i o n
S t e p :
3
S t d
G a s
C a l i b r a t i n g ・・・・・
4
5
6
1 T e mp
2
3
3 7 . 0
o
C
37.0
C O 2
5 . 0 %
5.0
4
5
S t a t u s
6
O K
7
62
C O 2 :
4 . 8 %
S T G :
5 . 0 %
3
STACKING INCUBATORS
Use the following procedure to stack incubators. This work is potentially dangerous, so contact our sales
representative or agent.
CAUTION
 Select a floor that is strong enough to support the stacked incubators.
 Never stack 3 or more incubators. Doing so is dangerous.
1. Take out 2 stacking plates A, 2 stacking plates B, 4 protective stickers, and 4 screws from the
accessory bag.
Note:
When stacking 2 incubators, use 2 stacking plates (A and B) for each one. Only one pair of stacking
plates is provided per an incubator, so take out the stacking plates (A and B) from both incubators.
Similarly only 2 protective stickers are provided per an incubator
2. Unplug both of the incubators that are to be stacked.
3. Remove the caps at the 2 places in the front of the top surface of the incubator that is to be on the
bottom, attach the protective stickers to the 4 corners on the top surface of the incubator that is to be on
the bottom. (Fig. A)
4. Secure stacking plate A with the screws that are provided.
5. Remove the front panel (5 screws) from the incubator that is to be on the top.
6. Place the upper incubator on top of the lower one, being careful not to pinch your fingers.
7. Make the lower incubator level by adjusting the leveling feet, and then level the upper incubator in the
same way.
8. Secure the stacking plate A fixed to the lower incubator to the upper incubator with the accessory
screws.
9. Remove from the back of the lower incubator the 2 hooks (4 screws) and remove from the upper
incubator the 2 screws on the lower rear panel.
10. Use the 6 screws that were removed in step 9 to secure stacking plate B to the rear panels of the
upper and lower incubators.
11. Replace the front panel (5 screws) to the upper incubator.
12. Use the hooks on the back of the upper incubator to prevent the incubators from falling over.
63
STACKING INCUBATORS
Front panel
Stacking plate B
Front
Stacking plate A
Protective sticker
Hook
Fig. A
64
SPECIFICATIONS
External dimensions
CO2 incubator
CO2 incubator
CO2 incubator
MCO-19AIC
MCO-19AIC(UV)
MCO-19AIC(UVH)
W620 mm x D710 mm x H900 mm (W24.4 inch x D28.0 inch x H35.4 inch)
Internal dimensions
W490 mm x D523 mm x H665 mm (W19.3 inch x D20.6 inch x H26.2 inch)
Product name
170 L (6.00 cu.ft.)
Interior volume
Exterior
Painted steel
Interior
Stainless steel containing copper
Outer door
Painted steel
Tempered glass
Inner door
Trays
Access port
4 trays made of stainless steel containing copper
W450 mm x D450 mm x H12 mm (W17.7 inch x D17.7 inch x H0.47 inch)
Maximum load: 7 kg/tray
Inner diameter: 30 mm (1.18 inch), On the back side
Rigid polyurethane foamed-in place
Insulation
DHA system (heater jacket + air jacket system)
Heating system
314 W
Heater
Natural evaporation with humidifying pan
Humidifying system
Temperature controller
PID control system
Temperature display
Digital display
PID control system
CO2 controller
CO2 density display
Digital display
Air circulation
Fan assisted
Air filter
UV lamp
Water level sensor
Alarms
Remote alarm contacts
CO2 inlet connection
CO2 inlet pressure
-----
0.3 m, Efficiency: 99.97 % or higher
4W x 1 (ozone-free emission)
Optical
Automatic set temperature alarm, Automatic set CO2 density alarm,
High limit temperature alarm, CO2 gas, various sensor/heater alarms
Allowable contact capacity: DC 30 V, 2 A
4 mm to 6 mm (0.157 inch to 0.236 inch) diameter tube can be connected
0.03 MPa(G) (0.3 kgf/cm2(G), 4.3 psi(G))
93 kg
Weight
Accessories
4 trays, 4 sets of tray supports, 1 gas tube, 1 humidifying pan,
1 pair of stacking plates A and B, 2 protective stickers, 2 tube bands
65
SPECIFICATIONS
Product name
CO2 incubator
MCO-19AIC
CO2 incubator
MCO-19AIC(UV)
H2O2 generator (MCO-HP)
H2O2 decon set (MCO-HL)
UV system set
(MCO-19UVS)
CO2 incubator
MCO-19AIC(UVH)
H2O2 generator and
H2O2 decon set are
standard equipment
UV system set is a standard equipment.
H2O2 reagent (MCO-H2O2)
STD gas auto calibration kit (MCO-SG)
Optional accessories
Gas regulator (MCO-100L)
Tray (MCO-47ST, includes 2 tray supports)
Half tray (MCO-25ST), Gas auto changer (MCO-21GC)
Roller base (MCO-18RB), UV replacement kit (MCO-20UV)
Stacking plate (MCO-18SB, when stacking on the MCO-175)
Stacking plate (MCO-21SB, when stacking on the MCO-20AIC)
Interface board (MTR-L03), Interface board (MTR-480)
Data acquisition system (MTR-5000)
interface board (MCO-420MA)
Note: Refer to the updated catalog when ordering an optional component.
Designs and specifications are subject to change without notice.
PERFORMANCE
Product name
Model No.
Temperature control range
CO2 Incubator MCO-19AIC
MCO-19AIC(UV)
MCO-19AIC(UVH)
MCO-19AIC-PT
MCO-19AIC-PA
MCO-19AIC-PK
MCO-19AICUV-PT
MCO-19AICUV-PA
MCO-19AICUV-PK
MCO-19AICUVH-PT
MCO-19AICUVH-PA
MCO-19AICUVH-PK
Ambient temperature+5 oC to 50 oC (ambient temperature: 5 oC to 35 oC)
Temperature distribution
±0.25 oC (ambient temperature: 25 oC, setting: 37 oC, CO2: 5 %, no load)
Temperature variation
±0.1 oC (ambient temperature: 25 oC, setting: 37 oC, CO2: 5 %, no load)
CO2 control range
0 % to 20 %
±0.15 % (ambient temperature: 25 oC, setting: 37 oC, CO2: 5 %, no load)
CO2 variation
95 %R.H.±5 %R.H.
Chamber humidity
o
Applicable environment
condition
Temperature: 5 C to 35 oC, Humidity: 80 %R.H. max.
(The designed performance may not be obtained
If the ambient temperature is equal or less than 15 oC)
30 dB (A scale)
Noise level
Power consumption
Max. 350 W
Max. 1,120 kJ/h
Heat emission
Rated voltage, frequency
Amperage
Max. 310 W
110 V to 120 V, 60 Hz
Max. 2.8 A
Note: The unit with CE mark complies with EC directives.
Based on our measuring method.
66
220 V, 60 Hz
Max. 1.4 A
CAUTION
Please fill in this form before servicing.
Hand over this form to the service engineer to keep for his and your safety.
Safety check sheet
1. Unit contents :
Risk of infection:
Risk of toxicity:
Risk from radioactive sources:
□Yes
□Yes
□Yes
□No
□No
□No
(List all potentially hazardous materials that have been stored in this unit.)
Notes :
2. Contamination of the unit
Unit interior
No contamination
Decontaminated
Contaminated
Others:
□Yes
□Yes
□No
□No
□Yes
□No
3. Instructions for safe repair/maintenance of the unit
a) The unit is safe to work on
□Yes
□No
b) There is some danger (see below)
□Yes
□No
Procedure to be adhered to in order to reduce safety risk indicated in b) below.
Date :
Signature :
Address, Division :
Telephone :
Product name :
Model No.
CO2 incubator
MCO-
Serial number :
Please decontaminate the unit yourself before calling the service engineer.
67
Date of Installation :
1-1-1, Sakata Oizumi-Machi Ora-Gun, Gunma 370-0596, Japan
© Panasonic Healthcare Co., Ltd. 2012
Printed in Japan
7FB6P15152900
S0412-1042