UltraSonic Metering System - Eastech

Transcription

UltraSonic Metering System - Eastech
CHARTMETER 5000
TM
THE FIRST FLOWMETER WITH A
BUILT-IN CHART RECORDER
Eastech Badger
ULTRASONIC FLOW MEASUREMENT
EASTECH COMBINES THREE PRODUCTS INTO ONE COST-EFFECTIVE UNIT
THAT ELIMINATES CONDUIT RUNS, WIRING, AND INSTALLATION COSTS
M
O
D
E
L
CHART RECORDER
A fully integrated 1, 7, or 31 day continuous trace
circular chart recorder housed in a corrosion and
impact resistant Nema 4, 4X enclosure.
5
0
2
2
M
O
D
E
L
CHART RECORDER
A fully integrated 1, 7, or 31 day continuous trace
circular chart recorder housed in a corrosion and
impact resistant Nema 4, 4X enclosure.
5
0
4
4
OPEN CHANNEL
ULTRASONIC FLOWMETER
WITH DATALOGGER
CLOSED PIPE
TRANSIT-TIME FLOWMETER
WITH DATALOGGER
The 5022 is a highly advanced microprocessorbased ultrasonic flow transmitter for extremely
accurate measurement of flow in open channels.
Since a 10% error in level measurement can result
in a 25% error in flow measurement, the high
accuracy (0.02”) of the Model 5022, especially
when utilized for billing purposes, directly
translates into dollars.
The Model 5044 transit-time flowmeter offers plant
operating personnel a highly reliable instrument
(±0.5% of actual rate of flow) that is nonintrusive in
design. Simple to install, with a one-piece SpeedrailTM
sensor mounting system and factory pre-programmed
for ease of start-up, the 5044 Chartmeter will consistently maintain accurate readings over changing
process conditions and time.
DIMENSIONAL DATA (inches/mm)
5022
ChartMeter
5044
TM
9.250
235
8.875
225
5.25
132
7.19
180
ORDERING GUIDE & SPECIFICATIONS
Chart Recorder
ChartMeter
TM
Meter
Sensor
Extra Cable
Options
Data Retrieval
Manhole Flume
OPEN
CHANNEL
FLOW
FB4A
50 ft.
Factory
Calibrated
Modem
(phone line)
MF4
1
2
3
4
5
6
8
9
7
FB2A
50
22
Nema 4, 4X
IP66
3
X
Flow/Level (0-16 ft)
Tefzel
(30’ cable)
A
M
Y
Heater &
Thermostat
Profibus
B
E
200 ft.
Keylock
100 ft.
0
5022
Nema 4, 4X
IP66
Circular Chart
1 Day
7 Day
31 Day
(please specify)
Flow/Level (0-25 ft)
PVC
(30’ cable)
Z
C
DeviceNet
F
4”
6”
MF6
8”
MF8
Splice Kit
10”
D
MF10
Relays
(3) P
(4) G
(5) H
MF12
YEAR
WARRANTY
One 4-20 mA
RS 232/485
2 Relays
Datalogger
12”
For Operating and Suggested Specifications, see Chartmeter 5022 brochure
CLOSED
PIPE
FLOW
1
2
3
4
5
6
8
9
7
0
5044
44
Nema 4, 4X
IP66
3
YEAR
WARRANTY
One 4-20 mA
RS 232/485
1 Relay
Datalogger
External Strap-On
(1” - 42” Dia. Pipe)
AS1
External Strap-On
(48” - 120” Dia. Pipe)
Note:
Each sensor is
equipped with
50 feet of cable
as standard.
AS2
External Strap-On
100 ft.
(High Temp.: 300°F)
Y
AS3
Instream
200 ft.
(12” - 120” Dia. Pipe)
Z
AS4
Wetted “Hot Shot”
(12” - 120” Dia. Pipe)
AS5
Factory
Calibrated
A
Modem
(phone line)
(No Charge)
M
Heater &
Thermostat
Profibus
B
Not
Applicable
E
Keylock
DeviceNet
C
F
Splice Kit
D
Relays
(2) L
(3) P
Windowed
(3” - 48” Dia. Pipe)
AS6
Dual Pipe/Multi-Path
contact factory.
For Operating and Suggested Specifications, see Chartmeter 5044 brochure
Eastech Badger
ULTRASONIC FLOW MEASUREMENT
4250 S. 76th E. Avenue
Tulsa, OK 74145
800-226-3569
918-664-1212
Fax: 918-664-8494
[email protected]
www.eastechbadger.com
PARSHALL FLUMES
Parshall Flumes
Description
The Parshall flume is one of a large class of
open channel primary elements known as
critical flow venturi flumes. A distinguishing
characteristic of the Parshall flume is the downward sloping invert of the throat. This feature
gives the Parshall flume its ability to operate at
higher ratios of downstream to upstream head
than any other such device.
The Parshall flume manufactured by Eastech
Badger is a monolithic fiberglass reinforced
polyester structure to assure maximum strength
and accuracy of dimension while minimizing
installation time. A staff gauge with 1/4” minor
divisions in inches is standard. A 2” NPT stilling
well connection is available as an option.
Application
The Parshall is recommended for those applications in which moderate concentrations of
sand, grit or other heavy solids exist and fluid
velocities entering the flume are subcritical.
The flume operates with a small energy loss or
change in channel grade, about one-fourth
that of weirs having the same crest length.
The flume is ideally suited for fluid measurement in irrigation channels or sewers.
Features
Sturdy Reinforced Fiberglass Construction#
Sizes from 3”up to 36”#
Stilling Well Connection Available#
Calibration
The Parshall flume exhibits reproducible head
rise/flow rate characteristics throughout its size
range. In order to assure the accuracy of the
device, adherence to all dimensions for construction as well as free flow hydraulic conditions is required.
Ordering Guide and Dimensional Data (inches)
Model
14
W
A
B
C
D
E
F
G
K
N
Wt.
Wall thk.
PF-3
3
18.38
18.0
7
10.9
24.0
6.0
12.0
1.0
2.25
45
.25
PF-6
6
24.44
24.0
15.5
15.63
24.0
12.0
24.0
3.0
4.50
100
.25
PF-9
9
34.63
34.0
15.0
22.63
30.0
12.0
18.0
3.0
4.50
125
.25
PF-12
12
54.0
52.89
24.0
33.25
36.0
24.0
36.0
3.0
9.0
350
.375
PF-18
18
57.0
55.88
30.0
40.38
36.0
24.0
36.0
3.0
9.0
500
.5
PF-24
24
60.0
58.88
36.0
47.50
36.0
24.0
36.0
3.0
9.0
530
.5
PF-36
36
66.0
64.75
48.0
61.88
36.0
24.0
36.0
3.0
9.0
620
.5
PARSHALL FLUMES
Sample Specifications
An Eastech Badger Model _______ Parshall flume shall be installed
as shown on the plans in accordance with the manufacturer's
recommendations. The Parshall flume shall be of single-piece
reinforced polyester construction with end flanges and reinforcing
ribs providing secure and permanent anchorage.
Sizing
Selection of a Parshall flume should be made on expected flow rates;
but for single point measurement to be valid, the design hydraulic
gradient must insure that free flow conditions exist at all flow rates.
Thus, the downstream fluid level must not exceed the values in Fig. 3
or single point measurement will not produce acceptable values.
The flume should be located in a straight reach of channel without
bends or changes in alignment directly upstream and should have a
well-distributed velocity profile. To assure free flow energy conditions exist, maintain the minimum and maximum flow rates as indicated for each flume size in capacity curve diagram. (Fig. 2)
Theoretical Discussion
The Parshall flume is an empirically derived and rated measuring
device. The discharge capacities are rated for “free flow” conditions in
which the submergence (ratio of downstream depth, to upstream
depth Hb/Ha) is less than the percentage given in Fig. 3. As the
downstream depth increases, flow condition is no longer critical, thus
two depth measurement readings (at Ha and Hb) are necessary and
a correction factor must be applied to obtain the correct discharge
under these conditions. The following graph presents these corrections necessary to ensure true discharge values.
HEAD INCHES
Parshall Flume Capacities
MIN. HD. RISE
AT MAX FLOW
Fig. 2
Discharge curves for Parshall Flumes with free flow and with submerged conditions
15
SUGGESTED SPECIFICATION
ChartMeter 5022
An ultrasonic microprocessor-based flowmeter with an integral circular chart recorder shall be installed at the location on the plans in
accordance with manufacturer’s recommendation. The ChartMeter must have a single pen chart recorder electronically integrated to an
open channel ultrasonic flowmeter. The ChartMeter 5022 shall be programmed for a _______ (size & type of primary element) and scaled
at maximum to ______ (max flow and engineering units or maximum level). The ChartMeter shall have the following features:
Enclosure:
Power:
Chart:
Outputs:
Sensors:
Accuracy:
IP66/NEMA 4, 4X (optional: Explosion proof, Class I, Groups C & D, Class II, Groups E, F & G, Divisions 1 & 2)
80 to 240 VAC, 50/60 Hz or 12 to 28 VDC @ 150 mA with surge suppression and fuse. (powered by flowmeter)
Circular 24 hour, 7 day or 31 day.
MaDC: 4-20 mA isolated into 1000 ohms, monitored to detect open circuits, with RFI and gas discharge
surge protection and two fuses.
Relays: A minimum of 2 relays rated at 0.25A @ 120 VAC or 0.5A @ 24 VDC. The relays must be assignable by the
front panel keypad.
Datalogger: There shall be a datalogger integral to the electronics. The datalogger shall have non-volatile flash
memory with a storage capacity of 32768 records. Software shall be supplied for downloading the data. The
logged data shall have the capability to be displayed on the backlit display in graphing form for daily total flow
units for the past eight days.
RS-232/485: There shall be an RS-232/485 serial port of 1200-38400 baud, Modbus RTU protocol.
Warranty: The electronics and sensor shall carry a 3-year warranty.
The sensor shall be designed for use in Class I, Division 1, Groups A, B, C & D hazardous areas. The sensor shall be
made of PVC or Tefzel and supplied with ________ feet of cable. 1000 ft. maximum cable runs allowed.
Splices shall be made waterproof.
±0.02” or ±0.05% of target distance.
The sensor cables shall be run in dedicated conduit.
The unit shall be a ChartMeter 5022 as manufactured by Eastech Badger, Tulsa OK.
Ordering Guide
Please provide the following information: Size and type primary element:___________ Maximum flow:___________Maximum level:__________
Chart Recorder
ChartMeter
TM
Meter
Sensor
Extra Cable
Options
Data Retrieval
Manhole Flume
OPEN
CHANNEL
FLOW
FB4A
50 ft.
Factory
Calibrated
Modem
(phone line)
MF4
1
2
3
4
5
6
8
9
7
5022
FB2A
22
Nema 4, 4X
IP66
3
X
Flow/Level (0-16 ft)
Tefzel
(30’ cable)
A
M
Y
Heater &
Thermostat
Profibus
B
E
200 ft.
Keylock
100 ft.
0
50
Nema 4, 4X
IP66
Circular Chart
1 Day
7 Day
31 Day
(please specify)
Flow/Level (0-25 ft)
PVC
(30’ cable)
Z
C
DeviceNet
F
4”
6”
MF6
8”
MF8
Splice Kit
10”
D
MF10
Relays
(3) P
(4) G
(5) H
MF12
YEAR
WARRANTY
One 4-20 mA
RS 232/485
2 Relays
Datalogger
12”
Ordering Example: 5022 with 0-25’ sensor, factory calibration, heater and therm. 5022FB4AAB