Braun Infrared ear thermometer IRT4520/4020

Transcription

Braun Infrared ear thermometer IRT4520/4020
English
4
15
25
This product is manufactured by
Kaz Europe SA under a license to the ‘Braun’ trademark.
‘Braun’ is a registered trademark of
Braun GmbH, Kronberg, Germany.
ThermoScan is a registered trademark of Kaz Europe SA
Kaz Europe SA
Place Chauderon 18
CH-1003 Lausanne
Switzerland
0297
©2011, Kaz Europe SA
®
www.kaz-europe.com
Designed in Germany
Made in Mexico
US Patent Nos.
6,626,835; 6,203,194; 6,149,298; 6,022,140; 6,694,174; 7,014,358; D460,006 TW Patent Nos.
NI-159809; NI-112501; 393570; 410272; NI-147458; ND-084456
KR Patent Nos.
0628283; 100507716; 0626189; 10-0883150
JP Patent Nos.
4619499; 4546650; 4052945; 4668250; 1144805
CN Patent Nos. 00124323.3; 99803219.8; 00804341.8; 02805125.4; 01337258
HK Patent Nos. 1024053; 1042550; 1062043
AU Patent Nos.
2002250974
IRT4520 / IRT4020
31IM4520198
12AUG13
2
3
Braun Thermoscan IRT 4000 series
Indications for use:
Instrument
IRT 4000 series
The Braun ThermoScan IRT 4000 series infrared ear thermometers are indicated for the
intermittent measurement and monitoring of human body temperature by consumers in a
home use environment.
ThermoScan
ThermoScan
IRT 4520
IRT 4020
Temperature measurement within
seconds
■
■
Temperature range °C
34-42.2
34-42.2
Temperature range °F
93.2-108.0
93.2-108.0
ExacTemp technology
■
■
Feedback system
■
■
Ready beep
■
■
Small, soft tip
■
Flexible tip
The physical size of the instrument shall be approx. 152mm, by 44mm, by 33mm. The
weight shall be approx. 100g, without the batteries.
Cap
The physical size of the cap shall be approx. 35mm, by 35mm, by 20mm. The weight shall
be approx. 15g. The cap shall protect the probe against pollution. It shall not cover the LC
display.
Product specifications
Product
Physical size and weight
■
Instrument cover
The physical size of the instrument cover shall be approx. 160mm by 75mm by 38mm. The
weight shall be approx. 30g.
Range definition/ Feature matrix
IRT 4020
IRT 4520
Feature
(Mass) (6 023)
(Pharmacy) (6 022)
Sensor system
Unique ExacTemp sensor technology
■
Site offsets
Ear mode
■
Patient temp. range
34°C..42.2°C
10°C..40°C
Memory storage capacity
8
1
Ambient temp. range
Auto-off after
60 sec.
60 sec.
Operating lifetime
5 years/ 10 000 readings
Easy-to-read LCD display
■
■
Battery type
2* AA (LR6)
Lens filter detector
■
■
Lens filter ejector
■
■
Battery lifetime at
normal ambient temp.
1000 readings and 5 years shelf life
Protective case with lens filter
storage
■
Probe protection and
thermometer storage
Simple cap for probe IRT instrument cover to hold
protection only
thermometer and LF 20
■
Battery charging
N/A
Charging time
N/A
Protective cap
Batteries included
■
■
21 lens filter included
■
■
Colour
Blue
Green
Model No: IRT 4020, IRT 4520
Accessories: Protective cap (LF20)
Manufactured by: Key Tronic Corporation.
Magneto # 7824, Parque Ind. Gema. Plant 1,
Juarez, Chihuahus, Mexico
Manufactured for: Kaz Europe S.A.,
Place Chauderon 18, CH-1003 Lausanne, Switzerland
4
5
English
The Braun ThermoScan thermometer has been carefully developed
for accurate, safe and fast temperature measurements in the ear.
The shape of the thermometer prevents it from being inserted too far
into the ear canal to hurt the eardrum.
However, as with any thermometer, proper technique is critical to
obtain accurate temperatures. Therefore, read the use
instructions carefully and thoroughly.
Important
• The operating ambient temperature range for this thermometer
is 10 – 40 °C (50 – 104 °F).
• Do not expose the thermometer to temperature extremes
(below – 20 °C / –4 °F or over 50 °C / 122 °F) nor excessive humidity
(> 95 % RH).
• This thermometer must only be used with genuine Braun
ThermoScan lens filters (LF 40). Never use this thermometer without
a new, clean lens filter attached.
• Keep lens filters out of reach of children.
• This thermometer is intended for household use only.
• Use of this thermometer is not intended as a substitute for
consultation with your physician.
How does Braun ThermoScan work?
Body temperature
Normal body temperature is a range. The following table shows
that this normal range also varies by site. Therefore, readings from
different sites should not be directly compared.
Normal ranges by site:
Axillary:
34.7 – 37.3 °C
94.5 – 99.1 °F
Oral:
35.5 – 37.5 °C
95.9 – 99.5 °F
Rectal:
36.6 – 38.0 °C
97.9 – 100.4 °F
ThermoScan:
35.8 – 38.0 °C
96.4 – 100.4 °F
• As well, a person’s normal temperature range tends to decrease with
age. The following table shows normal ThermoScan ranges by age.
Normal ThermoScan ranges by age:
0 – 2 years
36.4 – 38.0 °C
97.5 – 100.4 °F
3 – 10 years
36.1 – 37.8 °C
97.0 – 100.0 °F
11 – 65 years
35.9 – 37.6 °C
96.6 – 99.7 °F
> 65 years
35.8 – 37.5 °C
96.4 – 99.5 °F
Braun ThermoScan measures the infrared heat generated by the
eardrum and surrounding tissues. To avoid gross temperature
differences, the sensor itself is brought to a temperature close to that
of the human body. When the Braun ThermoScan is placed in the ear,
it continuously monitors the infrared radiation. The measurement will
be finished and the result is displayed, when an accurate reading can
be assured.
However, the range of normal also varies from person to person
and fluctuates throughout the day. It is therefore important to determine
your normal temperature range. This is easily done using Braun
ThermoScan. Practice taking temperatures on yourself and healthy
family members to determine their normal temperature ranges.
Why measure in the ear?
Note: When consulting your physician, communicate that the
ThermoScan temperature is a temperature measured in the ear and
if possible, note the individual’s normal ThermoScan temperature
range as additional reference.
The goal of thermometry is to measure core body temperature which is
the temperature of the vital organs. Ear temperatures accurately reflect
core body temperature, since the eardrum shares blood supply with
the temperature control center in the brain, the hypothalamus.
Therefore, changes in body temperature are reflected sooner and more
accurately in the ear than at other sites.
• Axillary temperatures only measure skin temperature and therefore,
are not a reliable indicator of core body temperature.
• Oral temperatures are influenced by drinking, eating and breathing.
• Rectal temperatures lag behind changes in core body temperature
and there is a risk of cross-contamination.
6
7
1. Lens filter
2. Probe
3. Lens filter detector
4. Lens filter ejector
5. Display
6. «I/O» button
(On/memory function – IRT 4520 only)
7. «ExacTemp» light
8. «start» button
9. Battery door
10. Protective cover (IRT 4520)
11. Cap (IRT 4020)
How to use your Braun ThermoScan?
1. To achieve accurate readings, make sure a new,
clean lens filter (1) is in place before each
measurement.
start
IRT 4020
IRT 4520
2. IRT 4020: Push the «start» button (8).
IRT 4520: Push the «I/O» button (6).
mp
Te
ac
During an internal self-check, the display shows all
segments. Then the last temperature taken will be
displayed together with «MEM».
E xa
cT
Wait for the ready signal beep and the ready symbol
in the display.
en
3. Fit the probe snuggly into the ear canal, then push
the «start» button (8).
n
ca
oS
erm
Th
8
If the probe has been fitted securely into the ear
canal during the complete measuring process,
a long beep will signal the end of the measuring
process. You can be assured to have taken an
accurate temperature reading. The result is shown
on the display (5).
If you take the temperature of another person, the
«ExacTemp» light (7) will be of help. It flashes during
the measuring process as long as the probe is
securely positioned, and lights up continuously
when an accurate reading has been taken.
4. If the probe has not been constantly placed in a
stable position in the ear canal, a sequence of short
beeps will sound, the «ExacTemp» light will go out
and the display will show an error message («POS»
= position error).
5. For the next measurement, eject the used lens filter
(push ejector (4)) and put on a new, clean lens filter.
IRT 4020: Clear the display by pushing the
«start» button once.
IRT 4520: Clear the display by pushing the
«I/O» button once.
Wait for the ready signal. Fit the probe snuggly into
the ear canal, then push the «start» button.
The Braun ThermoScan ear thermometer turns off
automatically after 60 seconds of inactivity.
The IRT 4520 can also be turned off by pressing
the «I/O» button for at least three seconds.
The display will shortly flash «OFF» and after
releasing the button it will go blank.
Temperature taking hints
• Always replace disposable lens filters to maintain
accuracy and hygiene.
• The right ear reading may differ from the reading
taken at the left ear. Therefore, always take the
temperature in the same ear.
• The ear must be free from obstructions or excess
earwax build-up to take an accurate reading.
• External factors may influence ear temperatures,
including when an individual has:
– been lying on one ear or the other
– had their ears covered
– been exposed to very hot or very cold
temperatures, or
– been recently swimming or bathing.
9
Exac
T
Product description (see page 3)
en
In these cases, remove the individual from the
situation and wait 20 minutes prior to taking a
temperature.
• Use the untreated ear if prescription ear drops or
other ear medications have been placed in the
ear canal.
Care and cleaning
Memory mode
If the thermometer is ever accidentally used without a
lens filter, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft
cloth moistened with alcohol. After the alcohol has
completely dried out, you can put a new lens filter on
and measure.
If the probe tip is damaged, please contact your
authorized local service centre.
The last temperature taken is stored in its memory
and will be automatically displayed when it is turned
on again. The display will show «MEM».
IRT 4520:
This model stores the last 8 temperature readings.
To display the stored readings, the thermometer
must be turned on. Then press the «I/O» button for
at least 1 second. The display shows the memory
number (e. g. MEM 1), and when releasing the «I/O»
button, the stored temperature for that memory
number is displayed, together with «MEM».
If «I/O» button is pressed too long, the thermometer
will be switched off. Each further pressing of the
«I/O» button displays the remaining memory
numbers (up to MEM 8).
MEM 1 is the latest reading, MEM 8 would be the
oldest.
The memory mode is left automatically after
displaying the oldest reading, or after pressing the
«I/O» button for at least 1 second.
Changing the temperature scale
The probe tip is the most delicate part of the
thermometer.
It has to be clean and intact to ensure accurate
readings.
Use a soft, dry cloth to clean the thermometer display
and exterior.
Do not use abrasive cleaners. Never submerge this
thermometer in water or any other liquid.
Store thermometer and lens filters in a dry location free
from dust and contamination and away from direct
sunlight.
Additional lens filters (LF 40) are available at most
stores carrying Braun ThermoScan.
For more information, please contact your authorized
local service centre.
Hong Kong service center contact number: +852 2686 3319
Taiwan service center contact number: +886 0800278789
China service center contact number: +86 021-61280577
Singapore service center contact number: +65 65522422
Your Braun ThermoScan is shipped with the Celsius
(°C) temperature scale activated. If you wish to
switch to Fahrenheit (°F) and/or back from
Fahrenheit to Celsius, proceed as follows:
(1)Make sure the thermometer is turned off.
(2)Press and hold down the «start» button (IRT
4020) or the «I/O» button (IRT 4520). After about
3 seconds the display will show this sequence:
«°C» / «SET» / «°F» / «SET» ...
(3)Release the «start» button / «I/O» button when
the desired temperature scale is shown. There
will be a short beep to confirm the new setting,
then the thermometer is turned off automatically.
10
11
Trouble-shooting
Trouble-shooting
Error message
Situation
Solution
No lens filter is attached.
Attach new, clean lens filter.
The thermometer cannot identify
a secure position of the probe. An
accurate measurement was not
possible.
IRT 4020: Clear the display by
pushing the «start» button once.
IRT 4520: Clear the display by
pushing the «I/O» button once.
POS = position error
Take care that the positioning of
the probe is correct and remains
stable.
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).
Allow the thermometer to remain
in a room for 30 minutes where
the temperature is between
10 and 40 °C or 50 and 104 °F.
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F).
HI = too high
Make sure a new, clean lens filter
is attached and thermometer is
properly inserted. Then, take a
new temperature.
LO = too low
System error – self-check display
flashes continuously and will not
be followed by the ready beep
and the ready symbol.
Wait 1 minute until the
thermometer turns off
automatically, then turn
on again.
If error persists,
... reset the thermometer by
removing the batteries and putting
them back in.
If error still persists,
Please contact your authorized
service centre.
Hong Kong service center contact
number: +852 2686 3319
Taiwan service center contact
number: +886 0800278789
China service center contact
number: +86 021-61280577
Singapore service center contact
number: +65 65522422
12
Situation
Solution
Battery is low, but thermometer will
still operate correctly.
Insert new batteries.
Battery is too low to take correct
temperature readings.
Insert new batteries.
Do you have any further questions?
Please contact your authorized
service centre
Hong Kong service center contact
number: +852 2686 3319
Taiwan service center contact number:
+886 0800278789
China service center contact number:
+86 021-61280577
Singapore service center contact
number: +65 65522422
Error message
Replacing the batteries
The thermometer is supplied with two 1.5 V type AA (LR 06) batteries.
For best performance, we recommend Duracell alkaline batteries.
Insert new batteries when the battery symbol appears on the display.
Open the battery compartment. Remove the batteries and replace with
new batteries, making sure the poles are in the right direction.
Slide battery door back until it snaps in place.
To protect the environment, dispose of empty batteries at your
retail store or at appropriate collection sites according to national
or local regulations.
Calibration
The thermometer is initially calibrated at the time of manufacture.
If this thermometer is used according to the use instructions, periodic
re-adjustment is not required. If at any time you question the accuracy
of the temperature measurement, please contact your authorized
service center.
Hong Kong service center contact number: +852 2686 3319
Taiwan service center contact number: +886 0800278789
China service center contact number: +86 021-61280577
Singapore service center contact number: +65 65522422
The manufacturing date is given by the serial numbers located on the
product. The first three digits represent the day of the year of
manufacture. The next two digits give the last two numbers of the year
of manufacture. An example: SN: 25008knnnnn-this product has been
manufactured on the 250th day of the year 2008.
13
2
1
Product specifications
(see page 3)
Model: Displayed temperature range: Operating ambient temperature range: Display resolution: Accuracy for displayed temperature range: IRT4520/4020
34 – 42.2 °C (93.2 – 108 °F)
10 – 40 °C (50 – 104 °F)
0.1 °C or °F
± 0.2 °C (35.5 – 42 °C) (95.9 – 107.6 °F)
± 0.3 °C (outside this temperature range)
± 0.14 °C (± 0.26 °F)
2 years / 1000 measurements
Clinical repeatability:
Battery life: Type BF equipment
See instruction of use
40°C
10°C
Operating temperature
Subject to change without notice.
This appliance conforms to the following standards:
Standard Reference
Edition
Title:
EN 12470-3
2000
linical thermometers - Part 3: Performance of compact
C
electrical thermometers (non-predictive and predictive)
with maximum device
EN 60601-1
2006
edical electrical equipment – Part 1: General
M
requirements for basic safety and essential
performance.
EN ISO 14971
2007
edical devices – Application of risk management to
M
medical devices.
EN ISO 10993-1
2003
iological evaluation of medical devices – Part 1:
B
Evaluation and Testing.
EN 60601-1-2
2007
edical electrical equipment – part 1-2: General
M
requirements for basic safety and essential performance
– Collateral standard: electromagnetic compatibility
– Requirements and tests
EN 980
2008
Symbols for use in labeling of medical devices
EN 1041
2008
Information supplied by the manufacturer of medical
This product conforms to the provisions of the EC directive 93/42/EEC
(Medical device directive).
Medical electrical equipment needs special precautions regarding EMC. For detailed description
of EMC requirements please contact your authorized service centre.
Hong Kong service center contact number: +852 2686 3319
Taiwan service center contact number: +886 0800278789
China service center contact number: +86 021-61280577
Singapore service center contact number: +65 65522422
Portable and mobile RF communications equipment can affect medical electrical equipment.
Please do not dispose of the product in the household waste at the end of it useful life.
To protect the environment, dispose of empty batteries at appropriate collection sites
according to national or local regulations.
14
15
耳
耳
耳
盖
耳
start
IRT 4020
IRT 4520
E xa
cT
mp
Te
ac
n
ca
oS
erm
Th
16
17
en
Exac
T
en
耳
耳
耳
18
19
耳
耳
沒有套上耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
耳
20
21
+86 021-61280577
2
耳
1
耳
SN
SN: 25008knnnnn –
2008
22
250
23
百靈耳溫槍
Braun ThermoScan infrared ear thermometers
型號: IRT 4020, IRT 4520
使用前請務必詳閱原廠之使用說明書並遵照指示使用。
衛署醫器輸字第 022485號
產品敘述及用途:
百靈耳溫槍是藉由轉換器, 藉著電子訊號放大, 調整及顯示的方法, 來測量患者體溫的器材。
轉換器是置於可分離的探針中。
產品規格
耳
產品
ThermoScan
ThermoScan
型號
IRT 4520
IRT 4020
幾秒內可測量溫度 ■
恆茂國際貿易(上海)有限公司
上海市宜山路2016號合川大廈7樓A室
郵編: 201103
顧客咨詢熱線:+86 021-61280577
■
溫度顯示範圍°C
34-42.2
34-42.2
溫度顯示範圍°F
93.2-108.0
93.2-108.0
正確溫度技術
■
■
反饋系統
■
■
短嗶聲
■
■
微小、柔軟的尖端 ■
■
有彈性的尖端
■
■
儲存容量
8
1
使用後自動斷電
60秒
60秒
液晶螢幕
■
■
耳套偵測器
■
■
耳套彈出器
■
■
包裝外盒含耳套
■
■
保護套
含電池
■
■
含耳套
顏色
■
藍色
■
綠色
型號: IRT 4020, IRT 4520
附件: 耳套(LF20)
24
25
26
27
套彈出裝置
蓋
start
IRT 4020
IRT 4520
E xa
cT
mp
Te
ac
n
ca
oS
erm
Th
28
29
en
Exac
T
en
30
31
32
33
+886 0800-278-789
+852 2686 3319
2
1
台灣維修中心
泰豐生技股份有限公司
新北市三重區重新路五段609 巷
10 號4F 電話 : +886 0800-278-789
SN后的第一、二、三碼表
示該年份的製造日,第四、五碼表示製造年份。
SN: 25008knnnnn –
2008
34
250
香港維修中心
利豐亞洲(香港)有限公司- 醫療保健組
沙田小瀝源安平街2號利豐中心14字樓
電話 : +852 2686 3319
製造廠名稱/地址: Manufactured by: Key Tronic Corporation (Magneto # 7824, Parque Ind.
Gema, Plant 1, Juarez, Chihuahua, Mexico)
for: Kaz Europe S.A. (Place Chauderon 18, CH-1003 Lausanne, Switzerland)
藥商名稱: 香港商卡舒遠東有限公司台灣分公司
藥商地址: 台北市北投區北投路二段13號4樓之1
35
36