Manual of Operation and Instruction
Transcription
Manual of Operation and Instruction
Manual of Operation and Instruction Model 3440 Surface Moisture-Density Gauge Troxler Electronic Laboratories, Inc. 3008 Cornwallis Rd. • P.O. Box 12057 Research Triangle Park, NC 27709 Phone: 1.877.TROXLER Outside the USA: +1.919.549.8661 Fax: +1.919.549.0761 www.troxlerlabs.com Troxler products are protected by U.S. and foreign patents. Copyright © 1989–2009 Troxler Electronic Laboratories, Inc. All Rights Reserved No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording, or information storage and retrieval systems, for any purpose without the express written permission of Troxler Electronic Laboratories, Inc. Adobe and Acrobat are trademarks of Adobe Systems Incorporated, and are registered in certain jurisdictions. Fantastic is a trademark of Dow Consumer Products, Inc. 409 is a trademark of the Clorox Company. Magnalube-G is a registered trademark of Carleton-Stuart Corporation. Microsoft, Excel, Windows, Windows 95, Windows 98, and Word are registered trademarks of Microsoft Corporation. PN 104279 January 2009 Edition 17.7 ii TROXLER SERVICE CENTERS Troxler Corporate Headquarters 3008 Cornwallis Road P.O. Box 12057 Research Triangle Park, NC 27709 Phone: 1.877.TROXLER (1.877.876.9537) Outside the U.S.A.: +1.919.549.8661 Fax: +1.919.549.0761 Web: www.troxlerlabs.com Technical Support Phone: 1.877.TROXLER (1.877.876.9537) E-mail: [email protected] Florida Service Center 2376 Forsyth Road Orlando, FL 32807 Fax: 407.681.3188 Texas Service Center 2016 East Randol Mill Road Suite 406 Arlington, TX 76011 Fax: 817.275.8562 Illinois Service Center 1430 Brook Drive Downers Grove, IL 60515 Fax: 630.261.9341 Northern California Service Center 11300 Sanders Drive, Suite 7 Rancho Cordova, CA 95742 Fax: 916.631.0541 Troxler European Subsidiary Troxler Electronics GMbH Gilchinger Strasse 33 D.82239 Alling nr. Munich, Germany Phone: ++49.8141.71063 Fax: ++49.8141.80731 E-mail: [email protected] NOTE To locate an independent, Troxler-authorized service center near you, call 1.877.TROXLER (1.877.876.9537). Model 3440 iii CAUTIONS AND WARNINGS Units intended for use in countries that are members of the European Community are shipped with a CE approved AC adapter, Troxler bin number 108354. Gauge cover is to be removed by trained service personnel only. No user-serviceable components inside. Note: Components behind cover can have voltage potentials in excess of 50 volts during normal operation of the gauge. Appendix C – Radiological Safety should be read carefully and understood before using the gauge. Alkaline Battery Use: See page F–2. Safety Glasses: See page 3–2. The source rod should automatically retract to the SAFE position when the gauge is lifted. See Scraper Ring/Tungsten Sliding Block Maintenance on page F–5. iv EU DECLARATION OF CONFORMITY Application of Council EMC Directive 89/336/EEC and Low Voltage Directive 73/23/EEC Standards to which Conformity is Declared: EN 61010-1 EN 55011 Group 1, Class A EN 50082-2 An EMC Technical Report/Certificate has been issued in accordance with Part IV (Reg 50) of the UK Regulations (SI 1992 No. 2372) by a UK appointed Competent Body, namely, Interference Technology International Limited 41-42 Shrivenham, Hundred Business Park Shrivenham, Swindon, Wiltshire SN6 8TZ Certificate Number C283TRO.1ABS Troxler Document Number 108205 Dated 16th January 1997 Manufacturer: Troxler Electronic Laboratories, Inc. P.O. Box 12057 3008 Cornwallis Road Research Triangle Park, North Carolina 27709 USA Apparatus: Model 3440 Surface Moisture-Density Gauge Year of Declaration 1997 (Original) 2006 (Corrected) Model 3440 v NOTES vi TABLE OF CONTENTS CHAPTER 1. INTRODUCTION TO THE SERIES 3400 Introduction..............................................................................................1–2 Gauge and Accessories ............................................................................1–4 Control Panel Keypad ..............................................................................1–6 CHAPTER 2. GETTING STARTED Turning the Gauge On..............................................................................2–2 Basic Parameter Setup..............................................................................2–4 Taking the Standard Count.....................................................................2–13 CHAPTER 3. MEASUREMENT SITE PREPARATION Soil and Base Course Preparation ............................................................3–2 Asphalt Surface Preparation.....................................................................3–4 CHAPTER 4. TAKING A DENSITY/MOISTURE MEASUREMENT Taking a Reading .....................................................................................4–2 Storing a Reading.....................................................................................4–4 CHAPTER 5. 3440 OPERATIONS Offsetting a Reading ................................................................................5–3 Count Time Function .............................................................................5–12 Store Function ........................................................................................5–12 Proctor/Marshall/Voidless Function.......................................................5–13 Status Function.......................................................................................5–16 Mode Function .......................................................................................5–17 Project Function .....................................................................................5–17 Print Function.........................................................................................5–18 Erase Function........................................................................................5–21 Counts Function .....................................................................................5–21 Recall Function ......................................................................................5–22 Depth Function.......................................................................................5–22 Calculator Function................................................................................5–24 CHAPTER 6. SPECIAL FUNCTIONS Stat Test....................................................................................................6–3 Drift Test ..................................................................................................6–7 Recover Erase ..........................................................................................6–9 Keypad Data...........................................................................................6–10 Auto-Station ...........................................................................................6–11 Special Calibration .................................................................................6–12 Nomograph.............................................................................................6–15 Precision.................................................................................................6–22 Set Units.................................................................................................6–24 Baud Rate...............................................................................................6–25 Model 3440 vii TABLE OF CONTENTS (Continued) CHAPTER 6. SPECIAL FUNCTIONS (Continued) Communication Protocol .......................................................................6–26 Battery Monitor......................................................................................6–27 Source Decay .........................................................................................6–30 Special Roadway....................................................................................6–31 CHAPTER 7. EXTENDED FUNCTIONS Time/Date ................................................................................................7–2 Customer Name .......................................................................................7–3 Serial Number ..........................................................................................7–4 Calibration Date .......................................................................................7–5 Calibration Constants...............................................................................7–6 Rod Length...............................................................................................7–8 Erase Standard Counts .............................................................................7–9 Memory Clear ........................................................................................7–10 Burn-In...................................................................................................7–10 Calibration Standard ..............................................................................7–11 Remote Control......................................................................................7–12 APPENDIX A. THEORY OF OPERATION Density ....................................................................................................A–2 Moisture ..................................................................................................A–5 APPENDIX B. RADIATION THEORY Atomic Structure ..................................................................................... B–2 Radiation Theory .................................................................................... B–3 Radiation Statistics.................................................................................. B–4 Radiation Terminology ........................................................................... B–5 APPENDIX C. RADIOLOGICAL SAFETY Radiation Types ...................................................................................... C–2 Radiation Exposure ................................................................................. C–3 Monitoring Radiation.............................................................................. C–4 3440 Radiation Profile ............................................................................ C–4 Source Encapsulation.............................................................................. C–6 Emergency Procedures............................................................................ C–6 APPENDIX D. LICENSING REQUIREMENTS Specific License ......................................................................................D–2 Radiation Safety Program .......................................................................D–2 Personnel Monitoring..............................................................................D–2 Security ...................................................................................................D–2 Training...................................................................................................D–3 viii TABLE OF CONTENTS (Continued) APPENDIX D. LICENSING REQUIREMENTS (Continued) Leak Test.................................................................................................D–3 Notice to Employees ...............................................................................D–3 Incidents ..................................................................................................D–3 Disposal...................................................................................................D–4 Record Keeping.......................................................................................D–4 APPENDIX E. TRANSPORTATION AND SHIPPING U.S. Requirements .................................................................................. E–2 Canadian Shipping Requirements ........................................................... E–4 APPENDIX F. PERIODIC MAINTENANCE AND SERVICE Batteries .................................................................................................. F–2 Mechanical Maintenance......................................................................... F–4 Electronic Maintenance........................................................................... F–9 Replacement Parts................................................................................. F–16 Returning a Gauge for Service .............................................................. F–27 APPENDIX G. 3440 SPECIFICATIONS Measurement Specifications ...................................................................G–2 Calibration Specifications .......................................................................G–3 Radiological Specifications.....................................................................G–3 Memory Specifications ...........................................................................G–3 Electrical Specifications..........................................................................G–4 Mechanical Specifications.......................................................................G–5 APPENDIX H. LOG FORMS & COMPACTION TEST DATA APPENDIX I. UNIT CONVERSION Measurement Units .................................................................................. I–2 Radiological Units.................................................................................... I–2 APPENDIX J. PRINTING (UPLOADING) PROJECT DATA Windows Hyperterminal .......................................................................... J–2 Viewing Project Data ............................................................................... J–4 INDEX WARRANTY Model 3440 ix LIST OF FIGURES Figure Title 1–1 1–2 3440 Gauge and Standard Accessories ................................. 1–5 3440 Keypad Layout ............................................................. 1–7 2–1 2–2 3440 Gauge Showing Rod Positions ................................... 2–14 Gauge Position on Reference Standard Block..................... 2–15 3–1 3–2 Drill Rod and Extraction Tool with Scraper Plate................. 3–2 Marking the Test Area .......................................................... 3–3 5–1 5–2 5–3 5–4 Gauge and Reference Standard Block Position for Trench Offset................................................................................... 5–11 Serial Port Location............................................................. 5–18 Sample Project Printouts ..................................................... 5–20 Calculator Function Keys.................................................... 5–24 6–1 6–2 Stat Test Printout Sample...................................................... 6–6 Drift Test Printout Sample .................................................... 6–8 A–1 A–2 A–3 A–4 Direct Transmission Geometry ............................................ A–2 Backscatter Geometry .......................................................... A–3 Depth of Top Layer.............................................................. A–4 Effect of Moisture on Depth of Measurement...................... A–7 B–1 B–2 Diagram of an Atom..............................................................B–2 Variation of Radioactive Emission........................................B–4 C–1 C–2 Effect of Distance on Exposure.............................................C–3 3440 Gauge and Transport Case ...........................................C–4 D–1 Sample Notice to Employees ............................................... D–6 F–1 F–2 F–3 F–4 F–5 F–6 F–7 Retaining and Scraper Rings ................................................. F–5 3440 Baseboard Assembly .................................................... F–9 3440 Gauge Assembly (Sheet 1 of 2).................................. F–16 3440 Base Mechanical Assembly ....................................... F–20 3440 Source Rod Handle Assembly.................................... F–20 3440 Preamplifier Assembly............................................... F–22 3440 Front Panel Assembly ................................................ F–24 x Page LIST OF TABLES Table Title 1–1 3440 Keypad Functions......................................................... 1–8 A–1 Neutron Thermalization and Absorption Data ..................... A–6 C–1 Radiation Dose Equivalent Rates in Millirems per Hours.....C–5 F–1 F–2 F–3 F–4 F–5 3440 Gauge Assembly Parts................................................ F–18 3440 Base Mechanical Assembly Parts............................... F–21 3440 Source Rod Handle Assembly Parts........................... F–21 3440 Preamplifier Assembly Parts ...................................... F–23 3440 Front Panel Assembly Parts........................................ F–25 Model 3440 Page xi NOTES xii ATTENTION GAUGE OWNER This gauge contains functions that require an ACCESS CODE. This code must be entered before these functions may be used. For more information on using the access code, refer to the specific function in Chapter 7. The ACCESS CODE for this gauge is: 4688 (Denmark only 2037) This page should be removed if the access code is not to be distributed to other parties or users of this gauge. Model 3440 xiii NOTES xiv CHAPTER 1 1. INTRODUCTION INTRODUCTION TO THE SERIES 3400 This chapter will familiarize the operator with the features and capabilities of the Troxler Model 3440 Surface Moisture-Density Gauge. A brief introduction to the system, as well as illustrations of the important components of the system, are included. Taking time to review this chapter before beginning operation of the system will help the operator to gain a more complete understanding of system operations. CONTENTS Introduction..............................................................................................1–2 Gauge and Accessories ............................................................................1–4 Control Panel Keypad ..............................................................................1–6 Model 3440 1–1 INTRODUCTION The Troxler Model 3440 Surface Moisture-Density Gauge measures the moisture content, density, and compaction of soils, soil-stone aggregates, concrete, asphalt treated bases, asphalt surfacing, and other materials that approximate similar ranges of density and/or moisture content. Currently recognized as the industry standard for compacting control instrumentation, the Model 3440 complies with the following American Society of Testing and Materials (ASTM) Standards: ♦ ASTM D-2922: Standard Test Methods for Density of Soil and SoilAggregate in Place by Nuclear Methods (Shallow Depth) ♦ ASTM D-2950: Standard Test Method for Density of Bituminous Concrete in Place by Nuclear Method ♦ ASTM D-3017: Standard Test Method for Water Content of Soil and Rock in Place by Nuclear Methods (Shallow Depth) The latest engineering, design, and manufacturing techniques have been incorporated, as well as over thirty years of experience in the nuclear gauge industry. The result of this effort is the ultimate in surface moisture-density gauges – the Troxler Model 3440 Surface Moisture-Density Gauge. The versatile 3440 gauge uses two modes of operation (backscatter and direct transmission) and can store over 450 readings. The gauge also provides over 30 special functions and operator-selectable precision for all phases of construction material testing. The LCD display and easy-to-use keypad ensure the operator of fast, accurate test measurement results. This manual contains sections that cover basic gauge setup and operation, as well as advanced operating techniques. The appendices provide information on radiation theory and radiation safety. Each chapter contains a table of contents for quickly locating the function required. If a question arises about the gauge, gauge operation, or an application not covered by this manual, please contact the closest Troxler Sales/Service Center or your Troxler Representative. Changes and additions may be made to the 3440 gauge and this manual without notice. Acquire updated information or a newer version of this manual by contacting the Troxler Sales Department or a Troxler Representative. The current edition of the manual can also be downloaded and/or viewed (using Adobe® Acrobat® Reader) from the Troxler website, www.troxlerlabs.com. 1–2 While no radiation hazard is imposed on operator(s) during normal use, a potential hazard does exist if improperly used. The sections of the manual covering radiological safety should be required reading for all operators and potential operators. If these sections are not completely understood, operators should seek assistance from Troxler, an appointed Troxler representative or others designated within the operator organization. Additional nuclear safety information is available by attending a Troxler Nuclear Gauge Training Course. As changes are made to local, state, and federal regulations on a continuing basis, the owner/operator must maintain a knowledge of these regulations. The responsibility for compliance ultimately falls upon the owner. The owner may also wish to purchase and subscribe to Titles 10 and 49 of the U.S. Code of Federal Regulations in addition to applicable local/state regulations pertinent to the owner’s license. Any licensing issues discussed in this manual are for the United States. To purchase a Model 3440 in Canada, owners must obtain a radioisotope license from the Canadian Nuclear Safety Commission (CNSC). The owner should obtain copies of the CNSC Regulations and the Transportation of Dangerous Goods Act and Regulations. This manual provides a guide to Canadian shipping requirements on page E–4. NOTE International communities should follow local regulations regarding products utilizing radioactive materials. Model 3440 1–3 1. INTRODUCTION Some information contained in this manual is used in training courses offered by Troxler Electronic Laboratories, Inc., to assist purchasers in obtaining a Radioactive Materials License from the U.S. Nuclear Regulatory Commission (NRC) or an Agreement State. Owners are encouraged to require study of this manual by operator(s) before allowing any use of the instrument. GAUGE AND ACCESSORIES The Troxler 3440 gauge provides a fast, accurate, and inexpensive way to find the moisture content and density of construction type materials. Take a moment to become familiar with the features and controls of the 3440, its components, operating principles, safeguards and cautions. Use Figure 1–1 and the list below to identify the gauge and standard accessories as they are unpacked. 1. The GAUGE is the portable surface moisture-density gauge containing the radioactive sources, electronics and rechargeable battery packs. The gauge serial number appears on the gauge handle and the reference standard block, as well as on the calibration, source, and warranty certificates. 2. The REFERENCE STANDARD BLOCK is primarily used to establish standard counts against which future measurements are compared. The reference standard block is also an unchanging reference for long-term stability checks. 3. The SCRAPER PLATE/DRILL ROD GUIDE is used to prepare the test site. It is used to guide the drill rod in preparing a hole for the source rod for direct transmission measurements. 4. The DRILL ROD is used to drill a hole for direct transmission measurements. Under no circumstances should the source rod be used for this purpose. 5. The DRILL ROD EXTRACTION TOOL provides leverage to remove the drill rod from clays and other materials. 6. Two CHARGER/ADAPTERS are supplied: one for dc (12 volt negative ground systems) and one for ac. 7. The TRANSPORT CASE is specially fitted for safe transport of the 3440 and associated parts. The case is water-resistant, but is not watertight. In case of inclement weather, the case should be protected with some type of covering to prevent intrusion of rain, etc. 8. OPERATOR’S MANUAL, CALIBRATION DOCUMENTS, and GAUGE CERTIFICATE 1–4 1. INTRODUCTION GAUGE TRANSPORT CASE SCRAPER PLATE/ DRILL ROD GUIDE REFERENCE BLOCK AC CHARGER DRILL ROD DC ADAPTER EXTRACTION TOOL GAUGE CERTIFICATE & WARRANTY OPERATOR'S MANUAL W/ QUICK REFERENCE GUIDE Figure 1–1. 3440 Gauge and Standard Accessories Model 3440 1–5 CONTROL PANEL KEYPAD The 3440 gauge control panel consists of two sections: the 20-key control keypad (Figure 1–2) and the 〈ON〉 and 〈OFF〉 keys. Keystrokes result in an immediate “beep” from the gauge. If the beep is not heard, the operation is not completed. Gauge and control panel operations are described throughout the manual. Table 1–1 summarizes the functions of each key, and provides page references for more information on certain operations. The 〈SHIFT〉 and shift function keys are color-coded yellow for ease of identification. The 〈SHIFT〉 must be pressed before pressing a function key. Functions are directly addressable from any other function except the calculator mode. Pressing 〈SHIFT〉 causes the display’s top line to change to –SHIFT FUNCTION–. NOTE If a function key is not pressed within four seconds, the gauge reacts as if no key was pressed. If there is no action after pressing a key, the gauge will return to READY after two minutes. The 〈YES〉, 〈NO/CE〉, and 〈START/ENTER〉 keys apply to all gauge modes. For a complete reference to any function, consult the index. Pressing 〈START/ENTER〉 from most modes will abort the mode and begin a test. The calculator function keys are 〈MS〉, 〈MR〉, 〈+〉, 〈–〉, 〈×〉, 〈÷〉, and 〈=〉. 1–6 EXIT STORE MS NO/CE C/CE OFFSET MR PROCTOR/ MARSHALL + TIME − MODE SPECIAL 7 8 9 PROJECT PRINT ERASE 4 5 6 COUNTS DEPTH CALC. 1 2 3 . START/ ENTER 1. INTRODUCTION YES STATUS RECALL SHIFT × STANDARD ÷ 0 = Figure 1–2. 3440 Keypad Layout Model 3440 1–7 Table 1–1. 3440 Keypad Functions KEYS YES EXIT DESCRIPTION PAGE Answers display prompts. Permits exit from the calculator mode. NO/CE C/CE Answers display prompts/Clear last entry. Clears calculator entry. STATUS 7 MODE 8 SPECIAL 9 STORE MS (SHIFT function) Displays status of gauge functions. Number key. (SHIFT function) Asphalt or Soils selection. Number key. (SHIFT function) Provides access to special functions. Number key. To store data in gauge memory. Memory store function for the calculator mode. OFFSET MR Select measurement offsets. Memory recall function for the calculator mode. PROJECT 4 PRINT 5 ERASE 6 PROCTOR/ MARSHALL + TIME (SHIFT function) To enter, view, or erase a project. Number key. (SHIFT function) Download data. Number key. (SHIFT function) Erase data. COUNTS 1 DEPTH 2 CALC. 3 SHIFT × 2–6 6–1 4–4 5–3 2–8 5–18 Number key. 5–21 Proctor or Marshall value selection. 5–13 Addition sign for calculator function. Select time interval for testing and measurement. Subtraction sign for calculator function. – 5–16 (SHIFT function) Displays last moisture and density counts. Number key. (SHIFT function) Selects depth mode. Number key. (SHIFT function) To access the calculator mode. Number key. Activates all SHIFT function modes 2–5 5–21 5–22 5–24 Multiplication sign for calculator mode. STANDARD Provides access to standard count mode. ÷ Division sign for calculator mode. RECALL Number key. . Decimal point key. START/ ENTER = 1–8 (SHIFT function) Recalls last moisture and density reading. 0 See manual text and index. Equals sign for calculator mode. 5–22 CHAPTER 2 GETTING STARTED This chapter explains how to get started with the Series 3440 Surface Moisture-Density Gauge. Functions explained include turning the gauge on, basic parameter setup and taking a standard count. CONTENTS Turning the Gauge On..............................................................................2–2 Daily Inspection ................................................................................2–2 Basic Parameter Setup..............................................................................2–4 Setting Measurement Units ...............................................................2–4 Count Time Selection........................................................................2–5 Mode Selection .................................................................................2–6 Selecting a Project Number...............................................................2–8 Setting the Clock/Calendar .............................................................2–11 Taking the Standard Count.....................................................................2–13 Viewing the Last Four Standard Counts .........................................2–17 Model 3440 2–1 2. GETTING STARTED NOTE It is recommended that this chapter be read and fully understood before operating the gauge for the first time. TURNING THE GAUGE ON The gauge uses rechargeable NiCad batteries (included) as a power source. To turn the gauge on, press 〈ON〉. When first turned on, the display screen fills with test characters before proceeding to the self-test phase. After two seconds, the gauge enters a 300-second self-test phase. During the self-test, the gauge displays the gauge model number, software version, serial number, company name, and progress (in seconds) of the test. After the self-test, press any key to enter the Ready mode. READY mm/dd/yyyy Depth: xxxx Time: x.xx min Batt volts: x.x The top line of the READY display alternates between the date and time. The gauge returns to the Ready mode when the gauge is “ready” to proceed to another function or there is no activity for more than 2 minutes. After 5 hours of no activity, the gauge will automatically perform a total power shutdown. DAILY INSPECTION The gauge should be inspected daily before use to ensure proper operation of all safety features as follows: 1. 2–2 Push the source rod down into the backscatter position and then raise it back to the shielded position. Turn the gauge over and verify that the tungsten sliding block is completely shut. If the gauge base opening is not completely closed by the sliding block, clean the sliding block and verify proper operation before using, transporting, or storing the gauge. Refer to the Cleaning the Tungsten Sliding Block section on page F–4 for cleaning instructions. WARNING Do not store or transport the gauge unless the tungsten sliding block is completely closed. Increased radiation levels may violate transportation regulations, and may cause excessive personnel radiation exposure. 2. Model 3440 2–3 2. GETTING STARTED If a radiation survey instrument is available, verify that the radioactive gamma source is in place by measuring the exposure rate at the surface of the gauge. The exposure rate should be approximately 10-20 mrem per hour. A reading of about 1 mrem or less indicates that either the survey instrument is not working or the cesium-137 (Cs-137) source may be missing. Refer to the Troubleshooting information in Appendix F for further instructions. BASIC PARAMETER SETUP SETTING MEASUREMENT UNITS Before taking measurements, the operator should determine the unit of measurement that is required for screen displays and/or printouts. The available selections are Metric and PCF. To execute the Set Units function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 three times for the display: YES- Next menu 9- SET UNITS 10- BAUD RATE 11-COMM PROTOCOL Press 〈9〉 for the display: UNITS in PCF Press: 1-PCF 2-METRIC ENTER- No change or the display will be: UNITS in METRIC Press: 1-PCF 2-METRIC ENTER- No change Press either 〈1〉 or 〈2〉 for the required units. The gauge will remain in the selected mode until reset. 2–4 COUNT TIME SELECTION The 3440 gauge provides three different count times to be used for taking readings. The longest count time, four minutes, will usually provide the highest accuracy. If a quick reading is required, one of the shorter periods may be selected. Press 〈TIME〉 for the display: For example, assume that a count time of 1 minute is desired. Press 〈2〉 for the display: - Count Time 1 min The display will return to READY. Model 3440 2–5 2. GETTING STARTED Time: xx min 1 - 15 sec 2 - 1 min 3 - 4 min MODE SELECTION The Mode function provides for the selection of Soil or Asphalt mode. Under Asphalt mode, the sub-mode selection of % Marshall or % Voids may be enabled. Also, while in Asphalt mode, the % Voidless density selection may be enabled. To select the Mode function, press 〈SHIFT〉 and 〈MODE〉 for the display: MODE: SOIL Select: 1- SOIL 2- ASPHALT (CE to exit) Soil Mode Selection From the above display, press 〈1〉 to select Soil mode. The display is: SOIL MODE After a short delay, the display will return to READY. Chapter 4 provides instructions for taking measurements in Soil mode. Asphalt Mode Selection From the first display on this page, press 〈2〉 to select Asphalt mode. The display will be: ASPHALT - %MA Select: 1- %MA 2- 100-%MA (CE to exit) 2–6 ASPHALT MODE, % MARSHALL. If % Marshall (or % compaction compared to a Marshall target) is required, press 〈1〉 from the last display on page 2–6. % Marshall = (WD/Marshall) × 100 ASPHALT: %MA Do you want to ENABLE %VOIDS also? To enable % Voids, press 〈YES〉. 2. GETTING STARTED % VOIDS = 100 × [1 – (WD/Voidless)] NOTE For the gauge to calculate % Voids, a voidless density must be entered into the gauge as described on page 5–15. The voidless density is the maximum theoretical density, or Rice test result. ASPHALT: %MA %VOIDS After a brief delay, the screen display returns to the READY display. ASPHALT MODE, 100 – %MA. If 100 – %MA is required, press 〈2〉 from the last display on page 2–6. NOTE The term % Voids is defined as 100 – %Voidless. ASPHALT: 100-%MA Do you want to ENABLE %VOIDS also? 100 – %MA = 100 × [1 – (WD/Marshall)] To enable % Voids, press 〈YES〉. %VOIDS = 100 × [1 – (WD/Voidless)] After a short delay, the display returns to READY. Chapter 4 provides instructions for taking measurements in Asphalt mode. Model 3440 2–7 SELECTING A PROJECT NUMBER The Project function allows unique alphanumeric project numbers to be input into gauge memory. All subsequent readings for a project are stored under its project number. NOTE Only one project number can be current in the gauge at any one time. Other stored projects may be recalled at any time. To select the Project function, press 〈SHIFT〉 and 〈PROJECT〉for the display: Current Project: PROJECT NUMBER New Project? Make an Existing Project Number Current To make an existing project current, press 〈YES〉 to display: Do you want to make an existing Project current? Press 〈YES〉 to display: PROJECT: PROJECT NUMBER 1 - Select 2 - Next Press 〈1〉 to select the displayed project, or 〈2〉 to display the next project name. 2–8 Enter a New Project Number To enter a new project number, press 〈YES〉 at the first display on the previous page. Do you want to make an existing Project current? Press 〈NO/CE〉 to display: Press SHIFT FOR LETTERS (ENTER) The project number may contain both numeric and alphabetic characters. For each numeric character, press the desired number key. For alphabetic characters, press 〈SHIFT〉 to display: Input PR: A SHIFT- See chars YES- Select Press 〈SHIFT〉 until the desired character is displayed, then press 〈YES〉 to accept the character. When the complete project number is displayed, press 〈START/ENTER〉 to activate the project number and exit. View Old Project Data To view data for an old project, from the first display on the previous page, press 〈NO/CE〉 to display: Do you want to view and/or erase a Project? Model 3440 2–9 2. GETTING STARTED Input PR: Press 〈YES〉 to display: PROJECT NUMBER 1 - View 2 - Erase 3 - Next Press 〈3〉 (if required) until the desired project number is displayed, then press 〈1〉 to display: PROJECT NUMBER mm/dd/yyyy (Press SHIFT) Press 〈SHIFT〉 to view the data for the displayed project number. Erase Old Project Number To erase an old project number, from the first display under the View Old Project Data section on the previous page, press 〈NO/CE〉 to display: Do you want to view and/or erase a Project? Press 〈YES〉 to display: PROJECT NUMBER 1 - View 2 - Erase 3 - Next Press 〈3〉 (if required) until the desired project number is displayed, then press 〈2〉 to display: Project: PROJECT NUMBER SHIFT/YES= ERASE Press 〈SHIFT〉 and 〈YES〉 to erase the project number shown on the display. 2–10 SETTING THE CLOCK/CALENDAR The Time/Date function allows the operator to change the current time-ofday or date that is displayed on the screen and used for measurement reports. To execute the Time/Date function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. Enter the code and press 〈START/ENTER〉. The display will be: YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈1〉 to select the Time/Date function. The display will be: Current Date: mm/dd/yyyy Do you want to change Date? Press 〈YES〉 to change the current date. The display is: Current Date: mm/dd/yyyy Input Date and press ENTER Model 3440 2–11 2. GETTING STARTED Press 〈19〉 to display: Input the new date and press 〈START/ENTER〉. NOTE Leading zeros must be entered for any single-digit day or month. For example, the date June 9, 2000 must be entered as 〈0〉〈6〉〈0〉〈9〉〈2〉〈0〉〈0〉〈0〉. After pressing 〈START/ENTER〉, the display will be: Current Time: hh/mm AM Do you want to change Time? If the time of day is to be changed, press 〈YES〉. The display is: Current Time: hh/mm AM Input Time and press ENTER Input the new time and press 〈START/ENTER〉. Again, note that the leading zeros must be entered for any single digit hour or minute. For example, 7:06 must be entered as 〈0〉〈7〉〈0〉〈6〉. After pressing 〈START/ENTER〉, the display will be Current Time: 7:06 AM Select: 1- AM 2- PM Select either 〈1〉 (for AM) or 〈2〉 (for PM). After a few seconds, the display will return to the Extended Functions menu. 2–12 TAKING THE STANDARD COUNT The front of the 3440 gauge is closest to you when the gauge is placed with the source rod to the left and the control panel to the right, as shown in Figure 2–1. The handle contains the indexer trigger mechanism, used to position the source rod on the notched index rod. NOTE The source rod should always be in the SAFE position when the gauge is not in use. The gauge should be turned ON before leaving for the worksite. This allows the gauge to complete its self-test routine while in transit. The standard count set can then be taken upon arrival at the work site without any delays. The gauge automatically compares the new density and moisture standard counts to the average of the values from the last four standard counts. The new standard count will “pass” if the density standard count is within 1% of the density average and the moisture standard count is within 2% of the moisture average. After taking the standard count, be sure to enter it in the log. To take a standard count, place the reference standard block on a flat surface at least 6 ft (2 m) from any vehicle or structure, and a minimum of 30 ft (10 m) from any other radioactive source. The surface can be asphalt or concrete pavement, compacted aggregate or a similar surface with density not less than 100 pcf (1600 kg/m3). Do not use truck beds, tailgates, tabletops, and so on. Model 3440 2–13 2. GETTING STARTED NOTE All Troxler nuclear gauges use low-level radioactive sources for taking measurements. The sources in the 3440 gauge have a half-life of 30 years (cesium-137) and 432 years (americium 241:beryllium). In other words, the amount of radioactivity from the cesium source will be reduced by one-half (1/2) in 30 years. Because of this continual decay, the standard count is taken to re-adjust the gauge to compensate for the decay. This decay is a known occurrence and will not compromise the accuracy of the gauge as long as the standard counts are taken. It is important to take the standard count when a gauge is initially received from the factory and before taking measurements at the worksite. SAFE POSITION (Use for Storage and Standard Counts) INDEXER TRIGGER BACKSCATTER POSITION 50 mm (2 in) SOURCE ROD INDEX ROD DIRECT TRANSMISSION POSITIONS 300 mm (12 in) Figure 2–1. 3440 Gauge Showing Rod Positions Place the gauge on the reference standard block as shown in Figure 2–2, making sure the block top and gauge base are clean and smooth, with no soil or other material to prevent good surface-to-surface contact. The gauge must be positioned between the raised edges of the block and with the right side of the gauge firmly seated against the metal butt plate on the block. 2–14 2. GETTING STARTED METAL BUTT PLATE REFERENCE STANDARD BLOCK Figure 2–2. Gauge Position on Reference Standard Block To begin taking a standard count, press 〈STANDARD〉 for the display: -Standard CountDS= xxxx MS= xxxx Take new count? Press 〈YES〉 for the display: Is gauge on Std. Block & Source rod in SAFE pos? Model 3440 2–15 Make sure the gauge is on the reference standard block as shown in Figure 2–2. Ensure that the source rod in the SAFE position and press 〈YES〉 to begin taking the four-minute standard count. Taking Standard Count xxx seconds remaining After count completion, the display is: MS=xxxx xx.xx %P DS=xxxx xx.xx %P Do you want to use the new STD? The P indicated to the right of the percentage figures indicates that the new counts are within the 1% density and 2% moisture limits. If the percentages are not within these limits, an F will be displayed. If you do get an F, or fail, display check for the following conditions: Is the source rod in the proper position? Are any other gauges close by? Is the gauge seated correctly on the reference standard block? Are the base of the gauge and the top of the reference standard block both clean? Is the reference standard block on a recommended surface? If all other conditions are normal, do not accept the standard count just taken. Press 〈NO/CE〉 and take another standard count. NOTE If the second count also fails, the old standard count sets should be erased as described on page 7–9. Successive standard count failures are usually the result of a prolonged period between the last standard count and the new one. This causes the tolerance to be exceeded because of source decay. After the old standard counts are erased, four new standard counts should be taken. Press 〈YES〉 to accept the new standard count and enter it into memory. If 〈NO/CE〉 is pressed, the new count is not accepted and the display returns to show the old values. 2–16 VIEWING THE LAST FOUR STANDARD COUNTS To view the last four standard counts, press 〈STANDARD〉 for the display: -Standard CountDS= xxxx MS= xxxx Take new count? Press 〈NO/CE〉 for the display: 2. GETTING STARTED -Standard CountWant to view last four Standard Counts? Press 〈YES〉 for the display: Density Std Cts 1:xxxx 2:xxxx 3:xxxx 4:xxxx (Yes for Moist.) To view the moisture standard counts, press 〈YES〉. The display will be: Moist. Std Cts 1:xxxx 2: xxxx 3:xxxx 4: xxxx (Press any key) The View Standard Count function may be exited by pressing any key. Model 3440 2–17 NOTES 2–18 CHAPTER 3 MEASUREMENT SITE PREPARATION This chapter covers the basic steps required to properly prepare a soil, base, or asphalt surface for measurement. CONTENTS Soil and Base Course Preparation ............................................................3–2 Asphalt Surface Preparation.....................................................................3–4 3. SITE PREPARATION Model 3440 3–1 SOIL AND BASE COURSE PREPARATION Surface conditions for the test are critical to gauge performance and test results. The scraper plate can be used to prepare surfaces that are not smooth. Any small valleys, holes, or voids in surfaces can be filled with sand or fine material from areas near the test site. Filling of large areas on test surfaces should be avoided. Placing the scraper plate on the surface to be tested, move the plate back and forth in a small area to smooth the test site. Lift the plate to fill any small depressions or voids present. Place scraper plate on the surface again and press down slightly to level the surface. WARNING Wear safety glasses while hammering drill rod through scraper plate. Put the drill rod through the extraction tool and then through one of the guides on the scraper plate (see Figure 3–1). Wearing safety glasses, step on the scraper plate to hold it firmly and hammer the drill rod at least, but not limited to, 2 in. (50mm) deeper than the desired test depth. The drill rod increment markings include the additional 2-inch depth needed for rod clearance. Before removing the drill rod from the scraper plate, mark the test area as shown in Figure 3–2 to make it easier to position the gauge over the site. DRILL ROD EXTRACTION TOOL SCRAPER PLATE/ DRILL ROD GUIDE Figure 3–1. Drill Rod and Extraction Tool with Scraper Plate 3–2 EDGE MARKS MARK FOR SCRAPER PLATE CENTER SCRAPER PLATE MARK FOR DRILL ROD CENTER SCRAPER PLATE METHOD 2 METHOD 1 Figure 3–2. Marking the Test Area Remove the drill rod by pulling straight up and twisting the extraction tool. Do not loosen the drill rod by tapping from side to side with the hammer. This will distort the hole or cause loose material to fill the 2-inch gap from the bottom of the hole. Carefully pick up the scraper plate and set it aside. Place the gauge on the surface prepared by the scraper plate so the source rod can enter the hole without disturbing any loose material around it. Using the handle and trigger mechanism, lower the source rod into the hole. Release the trigger at the desired depth and listen for the “click,” indicating that the source rod is properly locked into position on the index rod. Lightly press the top of the handle to confirm positive source rod locking. Gently pull the gauge to the right until the source rod is firmly in contact with the side of the hole. Model 3440 3–3 3. SITE PREPARATION CAUTION DO NOT use the source rod of the Model 3440 gauge to drill holes. ASPHALT SURFACE PREPARATION Test site selection for asphalt surfaces is much the same as soils except that preparation differs slightly. Voids may be filled on coarse, open-graded mixes using soft sand or cement powder, taking care to fill only the voids and leaving the asphalt bare for surface contact with the gauge base. The base should be resting on the asphalt and not on the fill material. Try to “rock” the gauge – if you can “rock” it, place more fill material or move the gauge to a flatter surface to correct the problem; the gauge should not rock. Put the source rod in backscatter position, making sure the rod is in the proper index rod notch and not resting on the asphalt. 3–4 CHAPTER 4 TAKING A DENSITY/MOISTURE MEASUREMENT This chapter covers the basic steps required for taking and storing actual density and/or moisture measurements with the Series 3440 Surface Moisture-Density Gauge. CONTENTS Taking a Reading .....................................................................................4–2 Soil Mode..........................................................................................4–2 Asphalt Mode....................................................................................4–3 Storing a Reading.....................................................................................4–4 Storage – Asphalt Mode (Auto-Station Enabled)..............................4–4 Storage – Asphalt Mode (Auto-Station Disabled) ............................4–5 Storage – Soil Mode..........................................................................4–6 4. TAKING A MEASUREMENT Model 3440 4–1 TAKING A READING NOTE When not taking readings, always keep the source rod in the SAFE position. For added operator safety, the source rod on the 3440 gauge automatically retracts to the SAFE position when the gauge is picked up by the handle. If you do not hear a “click” when the source rod is raised to the SAFE position, look at the bottom of the gauge to verify that the tungsten sliding block is completely closed. If the gauge base opening is not completely closed by the sliding block, the sliding block may require cleaning. Refer to the Cleaning the Tungsten Sliding Block section on page F–4 for cleaning instructions. CAUTION Do not store or transport the gauge unless the tungsten sliding block is completely closed. Increased radiation levels may violate transportation regulations, and may cause excessive personnel exposure. SOIL MODE Enable the Soil mode prior to taking a measurement. Refer to Chapter 2 for more details on gauge setup. Ensure that the site to be measured has been properly prepared. Refer to Chapter 3 for more detail in site preparation. Position the gauge at the proper location and release the source rod, pushing it to the proper depth. Check to see that all parameters have been set to the correct values. To start the measurement, press 〈START/ENTER〉 for the display: Depth: xx PR: xxxx Time: xx 4–2 sec. After the gauge completes its count time, the display will be: %PR= xxx % DD= xxx WD= xxx M= xx %M= xx The reading may be stored for later use by pressing the 〈STORE〉 key. Refer to page 4–4. ASPHALT MODE Enable the Asphalt mode prior to taking a measurement. Refer to Chapter 2 for more details on gauge setup. Ensure that the asphalt surface has been properly prepared. Refer to Chapter 3 for more detail in site preparation. Place the gauge at the proper location and move the source rod to the backscatter position (refer to Figure 2–1 on page 2–14). NOTE In some applications, the source rod may be extended to place the gauge in a Direct Transmission mode for use on soft asphalt. Verify that all parameters have been set to the correct values. Start a measurement by pressing 〈START/ENTER〉. The display will be: Depth: xx MA: xxxx Time: xxx sec. %MA= xxxx % WD= xx M= xxxx %M= xx %VOIDS= xxxx The above display will remain until a function key is pressed, a new measurement is taken, the 〈NO/CE〉 key is pressed, or the gauge shuts off due to inactivity. The reading may be stored for later use by pressing the 〈STORE〉 key. Refer to page 4–4. Model 3440 4–3 4. TAKING A MEASUREMENT After the gauge completes its count time, the display will be: STORING A READING The 3440 gauge allows the operator to store measurements under specific project numbers and station numbers for later recall. This function will work in soil mode as well as in asphalt mode. The 3440 will provide for additional numeric data storage under the station number and project. The operator can also setup the gauge to prompt for the information required on U.S. Federal Highway Administration (FHWA) projects (see page 6–31). The 3440 gauge is equipped with an Auto-Station function that allows automatic indexing of the station numbers. See page 6–11 for more information on using this function. To store readings, take a measurement following the procedure discussed on pages 4–2 and 4–3. When the density and/or moisture results are displayed, the reading may be stored. STORAGE – ASPHALT MODE (AUTO-STATION ENABLED) Press 〈STORE〉. If the Auto-Station function is enabled, the display will be: Station NUMBER? Distance from center line? xxx ft. From the above screen, input the distance from the centerline and press 〈START/ENTER〉 for the display: Left or right of center line? 1- LEFT 2- RIGHT Pressing 〈1〉 or 〈2〉 yields: Station: xx is stored. Do you want to store more info? 4–4 If 〈YES〉 then: Station: 1--------------Input and Press ENTER If 〈NO/CE〉 then: Station: xx Reading is stored. The READY screen is then displayed. STORAGE – ASPHALT MODE (AUTO-STATION DISABLED) If the Auto-Station function is disabled, the display will be: Station NUMBER? Press ENTER when completed. Press 〈START/ENTER〉 for: Station: xx Distance from center line? xxx ft. 4. TAKING A MEASUREMENT Enter the number of the station and press 〈START/ENTER〉. Left or right of center line? 1- LEFT 2- RIGHT Model 3440 4–5 Choose 〈1〉 or 〈2〉: Station: xx is stored. Do you want to store more info? If 〈YES〉: Station: 1--------------Input and Press ENTER Repeat loop until additional data is complete. If 〈NO/CE〉: Station: xx Reading is stored. The READY screen is then displayed. STORAGE – SOIL MODE Press 〈STORE〉. Station NUMBER? Press ENTER when completed. Press 〈START/ENTER〉 for: Station: xx is stored. Do you want to store more info? 4–6 If 〈YES〉: Station: 1--------------Input and Press ENTER Repeat loop until additional data is complete. If not, press 〈NO/CE〉. The READY screen is then displayed. 4. TAKING A MEASUREMENT Model 3440 4–7 NOTES 4–8 CHAPTER 5 This chapter contains descriptions and instructions for functions and procedures used frequently by most gauge operators on a day-to-day basis. In many cases, these operations will be performed every time a measurement is taken. Therefore, this chapter should be reviewed and completely understood before taking measurements in the field. Several of these functions have been described earlier in this manual. CONTENTS Offsetting a Reading ................................................................................5–3 Offsets – Wet Density .......................................................................5–4 Offsets – Moisture.............................................................................5–5 Offsets – Trench..............................................................................5–10 Count Time Function .............................................................................5–12 Store Function ........................................................................................5–12 Proctor/Marshall/Voidless Function.......................................................5–13 Recall a Stored Proctor/Marshall Value..........................................5–14 Enter a New Proctor/Marshall Value ..............................................5–14 Voidless Density Value...................................................................5–15 Status Function.......................................................................................5–16 Mode Function .......................................................................................5–17 Project Function .....................................................................................5–17 Print Function.........................................................................................5–18 Printing a Single Project .................................................................5–19 Erase Function........................................................................................5–21 Counts Function .....................................................................................5–21 Model 3440 5–1 5. 3440 OPERATIONS 3440 OPERATIONS CONTENTS (Continued) Recall Function ......................................................................................5–22 Depth Function.......................................................................................5–22 Manual Mode..................................................................................5–22 Automatic Mode .............................................................................5–23 Calculator Function................................................................................5–24 Addition ..........................................................................................5–24 Subtraction ......................................................................................5–25 Division...........................................................................................5–25 Multiplication..................................................................................5–25 Chain Calculations..........................................................................5–25 Memory Storage Arithmetic ...........................................................5–25 5–2 OFFSETTING A READING NOTE When an offset has been enabled, all future readings will automatically be adjusted with the offset factor regardless of the test site. It is very important that the operator DISABLE the offset function prior to taking readings on materials that do not require an offset. Offset functions will be disabled if the gauge is turned off for more than 10 seconds. The 3440 gauge provides offsets for the following measurement types: ♦ Wet density measurements ♦ Moisture measurements ♦ Trench measurements Offsets may be required for an accurate measurement if the material to be measured meets any of the following conditions: ♦ Has a density outside the range of 70 to 170 pcf (1100 to 2700 kg/m3) ♦ Has a high concentration of elements with atomic numbers greater than 20 (such as concrete, some coals, or ferrous soils) ♦ Is high in a hydrogen-rich material (such as gypsum, coal, or lime). The 3440 gauge also requires an offset if measurements are to be taken inside a trench or close to a structure. The walls of the trench or structure may scatter the gamma photons and neutrons back to the gauge, resulting in inaccurate density or moisture readings. To enable the Offset function, press 〈OFFSET〉 to display: -OFFSET- Select: 1-Dens. -OFF2-Moist. -OFF3-Trench -OFF- Model 3440 5–3 5. 3440 OPERATIONS The Troxler 3440 Gauge is factory calibrated for soils, asphalt, and concrete. The operator can adjust gauge readings based on other performance test, such as control strip results, core samples, or sand cones. This adjustment, or shift, is known as an offset. OFFSETS – WET DENSITY To enable a wet density offset, press 〈1〉 from the display on the previous page. The display will be: Density Offset DISABLED Do you want to ENABLE? Press 〈YES〉 to display: - Wet Density Offset x.x PCF Want to change? To change the offset, press 〈YES〉 (pressing 〈NO/CE〉 leaves the factors unchanged). The display is: - WD Offset Select: + or 1 = + 2 = Select either 〈1〉 or 〈2〉 for the display: - WD Offset + Press ENTER when completed Use the numbered keys to enter a new offset factor. When the entire value is input, press 〈START/ENTER〉 for the display: Density Offset ENABLED! The offset will be stored and the gauge returns to READY. 5–4 OFFSETS – MOISTURE In rare cases, the material to be measured may contain significant amounts of neutron absorbers such as boron or cadmium. These elements will result in a moisture reading that is lower than the actual value. In these cases, the offset factor will be a plus value. To enable a moisture offset, press 〈2〉 from the display on page 5–3. The display will be: Moisture Offset DISABLED Do you want to ENABLE? Press 〈YES〉 for the display: Moisture OffsetK=x.xx Do you want a new M-Offset? To change the moisture offset, press 〈YES〉 (pressing 〈NO/CE〉 leaves the factors unchanged.) The display is: Select source of Offset: 1- gauge derived 2- stored value The operator may then select a previously stored offset value or a gaugederived value. The gauge-derived value may be a new value input via the keypad or an actual value recorded from a measurement. Model 3440 5–5 5. 3440 OPERATIONS Moisture offsets (plus or minus) are sometimes needed when testing certain materials that contain high amounts of hydrogen-rich compounds, since the 3440 measures moisture by determining the amount of hydrogen present in the material. Some hydrogen-rich materials are gypsum, coal, lime, fly ash, organic material, mica clays, and phosphates. If the material contains compounds that are high in hydrogen forms other than water, an inaccurate reading may result. Offset factors for these measurements would be a minus value. Moisture Offset – Gauge Derived From the last display on the previous page, select 〈1〉 for the display: Select: 1- True M x.x % 2-Gauge M x.x % ENTER to enable For a true gauge-derived offset, two values must be input: the true moisture value, which is determined by laboratory analysis, and the gauge moisture, which is the gauge reading on the material to be tested. If the above values are correct, press 〈START/ENTER〉. To input a new true moisture value, press 〈1〉 for the display: True Moisturex.x % Press ENTER when completed Input the laboratory value and press 〈START/ENTER〉. To input a new gauge moisture value, press 〈2〉 for the display: Gauge MoistureSelect: 1- Keypad entry 2- Measurement KEYPAD ENTRY. To use the keypad to enter the gauge moisture reading from a previous test, press 〈1〉. The display is: Gauge Moisturex.x % Press ENTER when completed 5–6 Input the gauge moisture via the keypad and press 〈START/ENTER〉. 5. 3440 OPERATIONS SELECT: 1- True M x.x % 2-Gauge M x.x % ENTER to enable Press 〈START/ENTER〉. K=x.xx Do you want to save this value or later use? f To enable the K value and store it in memory for later use, press 〈YES〉: Enter desired Memory location of M-Offset: (1-4) Enter the appropriate value (1-4) and press 〈START/ENTER〉: Moisture Offset SAVED K=x.xx location x Moisture Offset ENABLED K=x.xx f To enable the K value, but not store it in memory, press 〈NO/CE〉. The value will be enabled, but will not be saved after it is disabled or the gauge is turned off. Moisture Offset ENABLED K=x.xx The offset will be enabled, but will not be saved after it is disabled or the gauge is turned off. The display will then return to READY. Model 3440 5–7 GAUGE MEASUREMENT. If the gauge is to be used to automatically record the moisture value, press 〈2〉 from the third screen on page 5–6. The display is: Place gauge on surface to test. Press START for (4) 1 min counts Place the gauge in the backscatter position ONLY at the location to be tested. Press 〈START/ENTER〉. The display will be: Moisture Count #1 Time: 60 sec. After counting down to zero, the display will be: Moist. count #1 MC =xxxx Press ENTER for next count Position the gauge at the next test site and press 〈START/ENTER〉. Repeat these steps for tests 3 and 4. After test 4 is completed, the display will be: Gauge MoistureOld:x.xx % New:x.xx % (ENTER to save) Press 〈START/ENTER〉 to save the new value. The display is: K=x.xx Do you want to save this value or later use? f To enable the K value but not save it for future use, press 〈NO/CE〉. The value will be enabled, but will not be saved after it is disabled or the gauge is turned off. The display will return to READY. 5–8 5. 3440 OPERATIONS f To enable the K value and store it in memory for later use, press 〈YES〉. The display will be: Enter desired Memory location of M-Offset: (1-4) Enter the appropriate value (1-4) and press 〈START/ENTER〉: Moisture Offset SAVED K=x.xx location x After a short delay, the display is: Moisture Offset ENABLED K=x.xx The offset will be enabled and the display will return to READY. Moisture Offset - Stored Value To recall a previously stored moisture offset, press 〈2〉 from the last display on page 5–5. The display will be: Enter desired Memory location of M-Offset: (1-4) Enter the appropriate value (1-4) and press 〈START/ENTER〉: Moisture Offset ENABLED K=x.xx The offset will be enabled and the display will return to READY. Model 3440 5–9 OFFSETS – TRENCH Vertical structures such as the walls of a building, trench, or ditch often “echo” gamma photons and neutrons back to the gauge. This “echoing” may adversely affect density and moisture measurements. Trench offsets compensate for the influence of vertical structures on measurements. Use a trench offset if taking a measurement inside a trench or within 2 ft (0.6 m) of a large vertical structure. When used, the trench offset adjusts all moisture measurements, but only the density measurements from backscatter through 4 in. (10 cm). To perform a trench offset: As described on page 2–13, take the daily standard count outside the trench and record the moisture and density count values. Place the reference standard block and gauge in the trench. Position the gauge on top of the reference standard block as in Figure 5–1. Take a one-minute count using the Trench Offset mode as follows. NOTE During the one-minute count, the gauge must be the same distance from the wall as it will be during the anticipated test readings. Press 〈3〉 from the display on page 5–3. Trench Offset DISABLED Want to use Trench Offset? Press 〈YES〉 to perform the trench offset. Trench Offset M. constant=xxx D. constant=xxx Want to change? Press 〈YES〉 to perform a new offset. Press 〈NO/CE〉 to use the existing offset constants. If a new offset is to be performed, the display will be: Rod -> SAFE pos. Press START for 1 min. STD cnt. in Trench. 5–10 5. 3440 OPERATIONS REFERENCE STANDARD BLOCK METAL BUTT PLATE Figure 5–1. Gauge and Reference Standard Block Position for Trench Offset Place the gauge on the reference standard block in the trench with the source rod in the SAFE position. Begin the one-minute count by pressing 〈START/ENTER〉. The density and moisture trench offset constants will be calculated, enabled and stored. If the gauge is not to be used for trench measurements immediately, disable the offset. To enable or disable an offset, press 〈3〉 from the display on page 5–3. Using the numeric keys, select the offset for enabling/disabling. The gauge will display the status of the offset and inquire if the operator wishes to change the status. Answer the inquiry by pressing either 〈YES〉 or 〈NO/CE〉. The gauge will enable/disable the offset and return to the READY display. Model 3440 5–11 COUNT TIME FUNCTION The 3440 gauge provides three different count times to be used for taking measurements. The Count Time function is described on page 2–5. STORE FUNCTION The 3440 gauge allows the operator to store measurements under specific project numbers and station numbers for later recall. The Store function is described on page 4–4. 5–12 PROCTOR/MARSHALL/VOIDLESS FUNCTION Soil mode (DD ÷ PR) × 100 = %PR Asphalt mode (WD ÷ MA) × 100 = %MA Asphalt mode 100 – [(WD ÷ VD) × 100] = %VOIDS To select or change a Proctor, Marshall, or Voidless density value, press the 〈PROCTOR/MARSHALL〉 key for the display: MA=xxx.x PCF PR=xxx.x VD=xxx.x Want to change? If not, press 〈NO/CE〉. Then press 〈START/ENTER〉. The screen reverts to the READY display. If a value is to be enabled, or a new value added or changed, press 〈YES〉. The display is: Select: 1 - MA 2 - PR 3 - Voidless NOTE After selecting Marshall (MA) or Proctor (PR) from the above display, the steps for the selecting the Marshall and Proctor values are identical. Therefore, this manual illustrates only the Marshall value selection screens. To change a Marshall value, press 〈1〉 to display: Select source of Marshall value: 1- Stored value 2- New value Model 3440 5–13 5. 3440 OPERATIONS The 3440 gauge provides for up to four different Proctor values, four different Marshall values, and one Voidless density value to be stored for later use. These values are used to determine percent compactions. The percent compaction calculations for Soil and Asphalt modes are as follows: RECALL A STORED PROCTOR/MARSHALL VALUE To enable a previously stored value, press 〈1〉. The display is: Select desired Marshall: 1:x.x 2:x.x 3:x.x 4:x.x Select the desired Marshall value by pressing 〈1〉, 〈2〉, 〈3〉, or 〈4〉 for the display: Marshall: x.x PCF ENABLED! The screen reverts to the READY display. ENTER A NEW PROCTOR/MARSHALL VALUE From the third display on page 5–13, press 〈2〉. The display will be: Marshall: x.x PCF Press ENTER when completed Enter the new value and press 〈START/ENTER〉, for the display: MA=x.x PCF Do you want to save this value for later use? Press 〈YES〉 to save the new Marshall value in a memory cell, for later use. Select Marshall Memory cell: 1:x.x 2:x.x 3:x.x 4:x.x 5–14 Press 〈1〉, 〈2〉, 〈3〉, or 〈4〉 to select the desired Marshall memory cell: 5. 3440 OPERATIONS MARSHALL x.x PCF ENABLED! stored in cell x VOIDLESS DENSITY VALUE From the second display on the page 5–13, press 〈3〉 for the display: Voidless Density xxx.x PCF Press ENTER when complete. Input the Voidless Density value and press 〈START/ENTER〉. The screen returns to the READY display. Model 3440 5–15 STATUS FUNCTION The Status function lets the operator check all gauge settings without executing the particular function. Press 〈SHIFT〉 and 〈STATUS〉 for the display: -CONTROL STATUSPress YES to view settings. (CE to exit) Press 〈YES〉 to display: UNITS: xxx TIME: xxxx MODE: xxxx (Press YES) Press 〈YES〉 to display: MA=x.x PCF PR=xxx.x PCF VD=xxx.x PCF (Press YES) Press 〈YES〉 to display: D OFF: x.x TRUE %M=x.x GAUGE %M=x.x (Press YES) % % Press 〈YES〉 to display: AUTO-STATION:OFF (Press YES) Press 〈YES〉 to display: Target Prec.(Press YES) Press 〈YES〉 to display: DepthCalib- xxxx xxxx (Press YES) Press 〈YES〉 to return to the first status display. Press 〈NO/CE〉 to return to the READY display. 5–16 MODE FUNCTION PROJECT FUNCTION The Project function allows unique alphanumeric project numbers to be input into gauge memory. All subsequent readings for a project are stored under its project number. The Project function is described on page 2–8. Model 3440 5–17 5. 3440 OPERATIONS The Mode function provides for the selection of Soil or Asphalt mode. This function is described on page 2–6. PRINT FUNCTION The Print function allows stored project data to be downloaded to a printer or computer. As shown in Figure 5–2, an RS-232 serial port on the gauge accepts a standard 25-pin “D” type interface cable connector. (To access the serial port, remove the black cover on the faceplate.) Appendix J provides instructions on using the HyperTerminal program in Microsoft® Windows 95® and later operating systems to print (upload) data to a computer. Project data should be downloaded to a printer or computer before erasing. This provides a permanent record of the data for use later. CAUTION To prevent damage to the 3440 gauge, turn the gauge power off before connecting an interface cable. Turn gauge and device off. Connect serial interface cable to gauge. Then, turn both gauge and the serial device on. Press 〈SHIFT〉 and 〈PRINT〉 to display: Connect Serial device & Select: 1- one Project 2- all Projects Figure 5–2. Serial Port Location 5–18 DOWNLOADING DATA -All ProjectsPress and hold any key to abort PRINTING A SINGLE PROJECT From the first display on the previous page, press 〈1〉. The display will be: PROJECT NUMBER Press YES for next Project. ENTER to select Press 〈YES〉 to scroll through the project numbers. Press 〈START/ENTER〉 to print the project desired project. DOWNLOADING DATA PROJECT NUMBER Press and hold any key to abort Model 3440 5–19 5. 3440 OPERATIONS To print all project data stored in the gauge, press 〈2〉. The gauge immediately begins to download all project data to the printer. Figure 5–3 shows sample project printouts. While printing, the screen displays: Soils Mode ********************************* PROJECT NUMBER: 1 SN: 59441 DATE: 3/16/2000 ********************************* STA # 1 2:30 PM 3/16/2000 DEPTH: 4 inches TIME: 15 seconds UNITS: PCF Std Cnts: D 3445 M 26 Dens Cnt. 3568 Moist Cnt. 32 WD = – DD = – PR = 145.0 %PR = M = + %M = +++++ Optional Data: 1234567890.1 --------------------------------STA # 2 2:30 PM 3/16/2000 DEPTH: 6 inches TIME: 15 seconds UNITS: PCF Std Cnts: D 3445 M 26 Dens Cnt. 3559 Moist Cnt. 27 WD = DD = PR = 145.0 %PR = M = + %M = +++++ Asphalt Mode ********************************* PROJECT NUMBER: 2 SN: 59441 DATE: 3/16/2000 ********************************* STA # 1 2:34 PM 3/16/2000 DEPTH: Backscatter 2 ft left of centerline TIME: 15 seconds UNITS: PCF Std Cnts: D 3445 M 26 Dens Cnt. 3564 Moist Cnt. 29 WD = DD = MA = 145.0 %MA = M = + %M = +++++ Optional Data: 020590.01234 --------------------------------STA # 2 2:35 PM 9/16/2000 DEPTH: Backscatter 5 ft right of centerline TIME: 15 seconds UNITS: PCF Std Cnts: D 3445 M 21 Dens Cnt. 3531 Moist Cnt. 28 WD = DD = MA = 145.0 % VOIDS = + Voidless Density = 145.0 %voidless = M = + %M = +++++ Figure 5–3. Sample Project Printouts 5–20 ERASE FUNCTION To select the Erase Function, press 〈SHIFT〉 and 〈ERASE〉 for the display Press SHIFT/YES to ERASE ALL STORED DATA! (CE to abort) NOTE The following step will erase all project data stored in the gauge’s memory! Press 〈SHIFT〉 and 〈YES〉 to erase all stored data. After erasing all project data from memory, the display will return to READY. Note that “erased” data actually remains in memory until it is overwritten by new data. If project data is erased accidentally, it may be recovered using the Recover Erase function described on page 6–9. NOTE The Recover Erase function may not be able to recover erased project data if new data has been stored since the erasure took place. COUNTS FUNCTION The Counts function allows the operator to review the density and moisture counts from the last measurement taken with the gauge. To execute the Counts function, press 〈SHIFT〉 and 〈COUNTS〉 for the display: Dens ct. = xxxx Moist ct. = xxxx SHIFT/RECALL to see Readings. The moisture and density counts from the last reading will be displayed. Press 〈SHIFT〉 and 〈RECALL〉 to view the individual readings. (Refer to the following pages for a more detailed description.) Model 3440 5–21 5. 3440 OPERATIONS The Erase function allows the operator to remove all project data stored in memory. RECALL FUNCTION The Recall function allows the operator to review the last density and moisture readings. To execute the Recall function, press 〈SHIFT〉 and 〈RECALL〉. If the gauge is in Soil mode, the display will be: %PR=xxxxx % DD=xxxxx PCF WD=xxxxx PCF M=xxxxx %M=xxxxx If the gauge is in Asphalt mode, the display will be: %MA=xxxxx % WD=xxxxx PCF M=xxxxx %M=xxxxx %VOIDS=xxxxx% DEPTH FUNCTION The Depth function allows the operator to select one of two depth modes: manual or automatic. In the Manual mode, the gauge prompts the operator to manually input the source rod depth using the keypad. In the Automatic mode, the gauge automatically reads the depth strip on the index rod, and then determines the source depth. To set the Depth mode, press 〈SHIFT〉 and 〈DEPTH〉. MANUAL MODE If the Depth mode is set to Automatic, the gauge display is: Depth Ind: Auto 1- Select MANUAL 2- Calibrate (CE to exit) To set the Depth mode to Manual, press 〈1〉 for: Depth Indicator: MANUAL 5–22 AUTOMATIC MODE 5. 3440 OPERATIONS If the Depth mode is set to Manual, the gauge allows the operator to change the mode. After changing the mode to Automatic, the gauge permits the operator to calibrate the depth strip. NOTE In the Automatic depth mode, the operator should calibrate the depth strip after extreme changes in temperature. The gauge automatically calibrates the depth strip when the operator takes a standard count. The gauge also allows the operator to calibrate the depth strip as follows. To calibrate the depth strip, press 〈SHIFT〉 and 〈DEPTH〉. Depth Ind: Auto 1- Select MANUAL 2- Calibrate (CE to exit) Press 〈2〉. The screen displays: -DEPTH Calib Set rod at: SAFE POSITION and press ENTER Position the source rod in the SAFE position and press 〈START/ENTER〉. The gauge calibrates the depth strip. It then indicates that the depth indicator is in the automatic mode as shown below, and returns to the READY display. Depth Indicator: AUTOMATIC Model 3440 5–23 CALCULATOR FUNCTION The 3440 gauge can also be used as a four-function calculator. The calculator function keys are shown circled on the keypad in Figure 5–4. To execute the Calculator function, press 〈SHIFT〉 and 〈CALC.〉 for the display: - CALCULATOR Use calculator keys Pressing any numbered key will clear the display (with the exception of the top line). ADDITION To add 65 and 78: press 〈6〉, 〈5〉. Press 〈+〉, 〈7〉, 〈8〉. Press 〈=〉. The display will give an answer of 143. Once a function has been completed and the answer is displayed (with the equal sign to the left), pressing any numbered key causes the display to clear (with the exception of the top line). YES EXIT STORE MS NO/CE C/CE OFFSET MR PROCTOR/ MARSHALL + TIME − STATUS MODE SPECIAL 7 8 9 PROJECT PRINT ERASE 4 5 6 COUNTS DEPTH CALC. 1 2 3 . START/ ENTER RECALL SHIFT × STANDARD ÷ 0 Figure 5–4. Calculator Function Keys 5–24 = SUBTRACTION DIVISION To divide 39 by 3: press 〈3〉, 〈9〉, 〈÷〉, 〈3〉. Press 〈=〉 for the displayed answer of 13. MULTIPLICATION To multiply 33 by 3: press 〈3〉, 〈3〉, 〈×〉, 〈3〉. Press 〈=〉 for the displayed answer of 99. CHAIN CALCULATIONS To perform a simple chain calculation, press 〈=〉 after each operation to get an intermediate result. For example, to perform the calculation 2 + 5 × 9 – 3 and divide the result by 5: press 〈2〉 〈+〉 〈5〉. Then press 〈=〉 to get the intermediate result of 7. on the last line of the display. Press 〈×〉 〈9〉 〈=〉. The figure following the equal sign will be 63. Next, press 〈–〉 〈3〉 〈=〉 for a result of 60. To divide this result by 5: press 〈÷〉 〈5〉 〈=〉 for the final answer of 12. A chain calculation such as 3+4(6+9)/2-8 must be performed using algebraic notation: Parenthetical equations must be performed first, followed by multiplication and division operations; then by addition and subtraction. Therefore, the correct way to begin this calculation is: 〈6〉 〈+〉 〈9〉 〈=〉 to get the intermediate result of 15. Next, do the multiplication 〈×〉 〈4〉 〈=〉 to get the next result of 60. The next step is division: entering 〈÷〉 〈2〉 〈=〉 brings the next result of 30. Finally, addition and subtraction: press 〈+〉 〈3〉 〈=〉 and 〈–〉 〈8〉 〈=〉 for the final answer of 25. MEMORY STORAGE ARITHMETIC Calculator mode also includes the MS (memory storage) and MR (memory recall) functions. The MS function allows data to be stored in general memory for use with other calculations or data may be stored under a specific project and station number. Model 3440 5–25 5. 3440 OPERATIONS To subtract 65 from 78: press 〈7〉, 〈8〉, 〈–〉, 〈6〉, 〈5〉, then 〈=〉. The displayed answer will be 13. To use a numerical value as a constant (for instance, pi = 3.14159) press the number keys and decimal in the proper order. When the displayed figure is 3.14159, press 〈MS〉. The display will be: Select Storage: 1- Calculator 2- Project Press 〈1〉 to store the data. An S will appear on the display to the right of the figure, indicating the decimal number sequence of 3.14159 was stored. If data is to be stored under a station number, press 〈2〉. The display will be: Station Number? Press ENTER when completed. Input the station number and press 〈START/ENTER〉. The value will be stored. To Recall stored values for use in calculations, press 〈MR〉 and the stored figures will appear on the display. An R will appear to the right of the figure to indicate that the value is recalled from memory and not a keyboard entry. This figure can now be used as a constant. To change the constant or put a different figure in memory, enter the new figure and press 〈MS〉, or press 〈MS〉 after a final result on the last line of the display. Figures in the calculator memory can be used in calculations in the same manner as figures entered separately. The figures in calculator memory may also be recalled as long as: ♦ The gauge batteries do not become completely discharged or removed. ♦ The calculator memory is not cleared. ♦ The values are not changed or replaced with another set of figures. To clear the calculator memory, press 〈0〉 and 〈MS〉. Press 〈EXIT〉 to leave the calculator mode. The display will return to READY. 5–26 CHAPTER 6 SPECIAL FUNCTIONS CONTENTS Stat Test....................................................................................................6–3 Viewing Stat Test Data .....................................................................6–5 Printing Stat Test Data ......................................................................6–6 Drift Test ..................................................................................................6–7 Execute Drift Test .............................................................................6–7 Viewing Drift Test Data....................................................................6–8 Recover Erase ..........................................................................................6–9 Keypad Data...........................................................................................6–10 Auto-Station ...........................................................................................6–11 Special Calibration .................................................................................6–12 Enable Special Calibration..............................................................6–12 Disable a Special Calibration ..........................................................6–13 Perform a Special Calibration (Recalibration) ................................6–13 Nomograph.............................................................................................6–15 Enable Nomograph .........................................................................6–16 Disable Nomograph ........................................................................6–16 Change/View Nomograph Data ......................................................6–16 Precision.................................................................................................6–22 Model 3440 6–1 6. SPECIAL FUNCTIONS This chapter contains descriptions and operation instructions for the Special Functions of the Series 3440 Surface Moisture-Density Gauge. These functions are typically not used on a day-to-day basis. In many cases, these functions are used only one time for gauge configuration and remain unchanged. The functions are arranged in order of use by a typical gauge operator with the Stat Test and Drift Test listed first and continuing through the Source Decay and Special Rdwy functions. CONTENTS (Continued) Set Units.................................................................................................6–24 Baud Rate...............................................................................................6–25 Communication Protocol .......................................................................6–26 Battery Monitor......................................................................................6–27 Battery Monitor...............................................................................6–27 Battery Status..................................................................................6–29 Source Decay .........................................................................................6–30 Special Roadway....................................................................................6–31 6–2 STAT TEST The Stat Test, or statistical stability test, may be performed to validate the normal operation of the gauge. Erratic readings or readings that seem to fluctuate may indicate a problem with the gauge. If the readings are suspect, the stat test may be executed. A stat test should result in a ratio of actual-to-predicted standard deviation. Ideally, the ratio should be 0.25, but acceptable limits are between 0.17 and 0.33. A definite instability exists if the ratio is outside these limits. Contact the nearest Troxler Service Center for additional instructions. SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈1〉 for the display: - STAT TEST Want to view last Stat data? If 〈YES〉 is pressed to view the last stat test data, the display is: DENS STAT TEST Avg cnts: xx R: x.xxx -ENTER for Moist To view the moisture data from the last stat test, press 〈START/ENTER〉. The display is: MOIST STAT TEST Avg cnts: xx R: x.xxx View Stat. data? The data for each of the readings may be viewed on the screen as described on page 6–5. Model 3440 6–3 6. SPECIAL FUNCTIONS To execute the stat test, press 〈SHIFT〉 and 〈SPECIAL〉 for the display: If 〈NO/CE〉 is pressed from the Want to view...? screen on the previous page, the display is: Put gauge on Std Block, rod in SAFE position & press START For a stat test, place the reference standard block on a high-density surface (at least 100 pcf (1602 kg/m3)) that is 5 ft (1.5 m) from a vertical surface. Following the instructions given on the display, place the gauge on the reference standard block with the source rod in the SAFE position. Press 〈START/ENTER〉 for the display: - STAT TEST Reading #1 Time= 60 secs The gauge will take twenty 60-second counts. When all counts are complete, the screen will display the average counts and ratio. DENS STAT TEST Avg cnts: xx R: x.xxx -ENTER for Moist To view the moisture data, press 〈START/ENTER〉. The display is: MOIST STAT TEST Avg cnts: xx R: xxxx View Stat. data? Press 〈YES〉 for the display: Select method of viewing data: 1- Screen 2- Printout 6–4 VIEWING STAT TEST DATA From the Select method of viewing... screen on page 6–4, press 〈1〉 for the display: - STAT TEST Press SHIFT to roll thru data Press 〈SHIFT〉 for the display: Press 〈SHIFT〉 and the gauge will display all twenty counts on the screen. After the last count, the display will be: DENS STAT TEST AVG cnts: xx R: x.xxx -ENTER for Moist To view the moisture data, press 〈START/ENTER〉. The display is: MOIST STAT TEST Avg cnts: xx R: x.xxx View Stat. data? Press 〈NO/CE〉 and the gauge will return to the READY display. Model 3440 6–5 6. SPECIAL FUNCTIONS Stat. Counts #1 Density: x Moisture: x (Press SHIFT) PRINTING STAT TEST DATA From the Select method of viewing... screen on page 6–4, press 〈2〉 for the display: Connect Printer & press START After connecting the printer to the serial connector on the gauge, press 〈START/ENTER〉 for the following display. Figure 6–1 shows a sample stat test printout. - STAT TEST Downloading data! If the printer is not connected properly, an error message will be displayed. ******************************** Statistical Stability Test ******************************** Gauge Serial #: 18326 5/04/2000 7:21 AM READING | COUNTS # | DENSITY MOISTURE 1 3721 617 2 3725 639 3 3746 624 4 3740 617 5 3737 616 6 3744 628 7 3687 621 8 3748 628 9 3735 627 10 3737 623 11 3711 621 12 3729 622 13 3757 619 14 3732 602 15 3734 630 16 3749 623 17 3729 618 18 3726 619 19 3742 618 20 3743 618 Average: 3734 622 Ratio: 0.252 0.292 (PASS) (PASS) Figure 6–1. Stat Test Printout Sample 6–6 DRIFT TEST If the stat test has already been performed (and has passed), but gauge readings seem to be fluctuating from test to test, the Drift Test can be performed to check the long-term drift of the 3440. Pass and Fail limits are averages from the stat test and the difference in the drift test average. If the average differences exceed 0.5% for density or 1% for moisture, contact your Troxler service department. Without turning off or moving the gauge from the time the stat test was taken, wait 3 to 8 hours and implement the drift test. To execute the drift test, press 〈SHIFT〉 and 〈SPECIAL〉 for the display: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈2〉 for the display: - DRIFT TEST Want to view last Drift data? EXECUTE DRIFT TEST To execute the drift test, press 〈NO/CE〉 at the above display: Put gauge on Std Block, rod in SAFE position & press START Model 3440 6–7 6. SPECIAL FUNCTIONS NOTE The gauge should not be turned off between the stat and drift tests. The stat test must be current. The gauge must not be moved between the stat and drift tests to eliminate possible failure due to positioning changes. Place the gauge on the reference standard block. Make sure the source rod is in the SAFE position and press 〈START/ENTER〉. The display will be: - DRIFT TEST Read #x Time:xxx After the gauge completes five 4-minute counts, the display will be: DENS DRIFT TEST Avg cnts: xx Drift: x.x % P -ENTER for Moist To view the moisture data from the last drift test, press 〈START/ENTER〉. The display is: MOIST DRIFT TEST Avg cnts: xx Drift: x.x % P View Drift data? VIEWING DRIFT TEST DATA The data for each of the readings may be viewed on the screen or may be printed. This function is the same as viewing the stat test data. Refer to page 6–5 for instructions on viewing or printing drift test data. Figure 6–2 shows a sample drift test printout. ******************************** DRIFT TEST ******************************** Gauge Serial #: 123456 O5/09/2000 3:01 PM READING | COUNTS # | DENSITY MOISTURE -------------------------------1 3769 672 2 3801 661 3 3777 670 4 3740 694 5 3669 660 Average: 3751 671 Drift: 0.3 % 0.2 % (PASS) (PASS) Figure 6–2. Drift Test Printout Sample 6–8 RECOVER ERASE The Recover Erasure function may be used if project data is erased accidentally. NOTE The Recover Erasure function may not be able to recover erased data if other data has been stored since the erasure took place. To execute the Recover Erasure function, press 〈SHIFT〉 and 〈SPECIAL〉 for: 6. SPECIAL FUNCTIONS SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 one time for the display: YES- Next menu 3- RECOVER ERASE 4- KEYPAD DATA 5- AUTO STATION Press 〈3〉 for the display: All data is RECOVERED! Model 3440 6–9 KEYPAD DATA The Keypad Data function allows the operator to store numeric information under a particular station number. This numeric information may only be stored under the current project number and may be viewed or printed using the Project function. To execute the Keypad Data function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 one time for the display: YES- Next menu 3- RECOVER ERASE 4- KEYPAD DATA 5- AUTO STATION Press 〈4〉 for the display: Station number? Press ENTER when completed. Input a previously stored Station Number. Press 〈START/ENTER〉 for: Station: 1--------------Input and Press ENTER Input any sequence of numeric digits (up to 12) and press 〈START/ENTER〉. Station: 1. Do you want to store more info? Press 〈YES〉 or 〈NO/CE〉. 6–10 AUTO-STATION The Auto-Station function allows the operator to initialize a station number. When this function is enabled, the gauge will automatically increment the station number by one after each reading is stored. To enable the Auto-Station function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST 6. SPECIAL FUNCTIONS Press 〈YES〉 one time for the display: YES- Next menu 3- RECOVER ERASE 4- KEYPAD DATA 5- AUTO STATION Press 〈5〉 for the display: Auto-Station DISABLED Do you want to ENABLE? Press 〈YES〉 to enable this function. The display will request the first station number. Initial Station Number: x Input and Press ENTER Input any sequence of numeric digits and press 〈START/ENTER〉. The gauge will return to the READY display. NOTE Whenever a new project number is entered into the gauge, the Auto-Station function is automatically disabled. Model 3440 6–11 SPECIAL CALIBRATION The Special Calibration function allows the operator to re-calibrate the gauge constants temporarily for use in measuring particular materials. These materials may be limestone or other materials that do not fall within the range of a normal calibration. Once enabled, the Special Calibration will remain in effect until the gauge is turned off or Normal calibration is re-enabled. To execute Special Calibration, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 two times for the display: YES- Next menu 6-SPECIAL CALIB. 7- NOMOGRAPH 8- PRECISION Press 〈6〉 for the display: SPEC CALIB.-OFF Press: 1- enable 2- disable 3- Recalib. At this time, the operator may either enable an existing special calibration or disable an existing calibration that has been in use. The operator may also re-calibrate the gauge for use in a different material. ENABLE SPECIAL CALIBRATION Press 〈1〉 to enable the calibration. The display will be: SPECIAL SOIL CALIBRATION ENABLED! The gauge will return to the READY display. 6–12 DISABLE A SPECIAL CALIBRATION To disable a previously enabled special calibration, press 〈2〉 from the SPEC CALIB. menu. The display will be: NORMAL CALIBRATION ENABLED! The gauge will return to the READY display. Before performing a special calibration, the following steps must be taken: 1. The sample area must be prepared for testing. 2. Gauge readings must be taken on each area. 3. A wet density measurement using an alternative method must be performed. To perform a new special calibration, prepare a test site and position the gauge. Position the rod at the correct depth and press 〈3〉 from the SPEC CALIB. menu. The display will be: Input rod depth: (Enter 0 for BS) xxinches (Press ENTER) Press 〈START/ENTER〉 after input of the rod depth. The display is: SPECIAL CALIB. Depth: x inches Time: x min. Press START Model 3440 6–13 6. SPECIAL FUNCTIONS PERFORM A SPECIAL CALIBRATION (RECALIBRATION) Press 〈START/ENTER〉 to begin taking the measurement. The display will be: SPECIAL CALIB. Depth: x inches Time:xxx secs. After counting down to zero, the display will be: Dens ct.=xxxx Moist ct.=xxxx To input sample dens Press ENTER Press 〈START/ENTER〉 to input the sample area wet density obtained from the alternative test method. The display will be: Input material density x.x PCF ENTER to ENABLE Input the material density and press 〈START/ENTER〉. The display will return to READY. 6–14 NOMOGRAPH In some cases, the 3440 may be used to determine the density of thin asphalt overlays. This measurement may be performed with the gauge in backscatter mode and using the “Nomograph” method of density measurement. It should be noted that this method is not as accurate as a true thin-layer gauge. However, this method may produce satisfactory results under many conditions. Obtaining an accurate overlay density with the Nomograph method requires that the density of the bottom layer and the thickness of the top layer be determined. The simplest method of determining the density of the bottom layer is to take a nuclear gauge measurement prior to applying the top layer or overlay. With the overlay applied and compacted, the depth of the layer should be determined. At this time, the gauge may be used to calculate the overlay layer density. To access the Nomograph function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 two times for the display: YES- Next menu 6-SPECIAL CALIB. 7- NOMOGRAPH 8- PRECISION Press 〈7〉 for the display: NOMOGRAPH - OFF Press: 1- enable 2- disable 3- chg/view data At this time, the operator may enable, disable, or change the Nomograph function. Model 3440 6–15 6. SPECIAL FUNCTIONS When a nuclear gauge is used in backscatter mode on overlays under 3 in. (7.5 cm), several effects must be overcome to produce an accurate reading. With most gauges, the primary problem is the result of photons from the source penetrating deeper than 3 in. (7.5 cm) and scattering back to the gauge. These backscattered photons from the underlying material will adversely influence the reading. ENABLE NOMOGRAPH Press 〈1〉 from the display at the bottom of page 6–15 to enable the Nomograph function. The display will be: NOMOGRAPH ON The gauge will return to the READY display. DISABLE NOMOGRAPH To disable the Nomograph function, press 〈2〉 from the display at the bottom of page 6–15. The display will be: NOMOGRAPH OFF The gauge will return to the READY display. CHANGE/VIEW NOMOGRAPH DATA To change or view the function data, press 〈3〉 from the display at the bottom of page 6–15. The display will be: Thicknessx.xx in. Press ENTER when completed Input the desired overlay thickness. Press 〈START/ENTER〉. The display is: Select to set1-Bottom dens. 2-Core dens. (CE to exit) The wet density of the underlying or base material may be input using one of two methods: keypad entry or by using the gauge to record actual measurements. 6–16 Bottom Density – Keypad Entry From the Select to set– display, press 〈1〉 for: Average bottom density: x.x PCF Want to change? If the display value is correct, press 〈NO/CE〉. 6. SPECIAL FUNCTIONS Press 〈YES〉 if the value is to be changed. The display will be: Select method of Entering density 1 - Keypad 2 - Measure Press 〈1〉 to enter data with the keypad. The display will be: Bottom densityx.x PCF Press ENTER when completed Enter the density of the bottom layer. The next display will be: NOMOGRAPH ON Model 3440 6–17 Bottom Density – Measurement Entry Bottom density values may be obtained by using the gauge to record an actual measurement. Follow the instructions from the Bottom Density – Keypad Entry description on the previous page until reaching the Select method of... display. Press 〈2〉 for measurement entry. The display will be: # of Readings to average (1-20)? x Press ENTER Input the number of readings to be taken and averaged. Press 〈START/ENTER〉 for the display: - Reading # 1 Time: x min. Set rod to BS & Press START Prepare the site for measurement. Position the gauge and press 〈START/ENTER〉. The display will be: - NOMOGRAPH Reading # 1 Depth: BS Time:xx sec. After counting down to zero, the display will be: Reading # 1 WD:xxxxx PCF Move gauge and press START If more than one reading is to be taken, move the gauge to the next location and press 〈START/ENTER〉. The gauge will repeat the above sequence for each reading. When all the readings are completed, the display will be: Average of x Reading(s)xxxxx PCF Press ENTER 6–18 From the previous display, press 〈START/ENTER〉. The display will be: NOMOGRAPH ON Thin-layer overlay measurements may now be taken. Press 〈START/ENTER〉 to begin the test. The display is: 6. SPECIAL FUNCTIONS Depth: Backscat MA: xxxx (Nomograph) Time: xxx sec. After counting down to zero, the display is: %MA= xxxx % WD= xx (Nomograph) This test may be stored as any other test measurement. The recall function and/or printout will show that the Nomograph function was used. NOTE Nomograph mode should be disabled before taking any measurements not requiring the Nomograph function. Core Density – Keypad Entry The Core Density method allows the operator to input the density value of a core sample for determining the overlay density. From the Select to set– display at the bottom of page 6–16, press 〈2〉. The display is: Core Density x.x PCF Press ENTER when completed If the displayed density is correct, press 〈START/ENTER〉. If a change is required, input the new value and press 〈START/ENTER〉. Model 3440 6–19 The display will be: Average measured densityx.x PCF Want to change? If the displayed value is correct, press 〈NO/CE〉. Press 〈YES〉 if the value is to be changed. The display will be: Select method of Entering density 1 - Keypad 2 - Measure Press 〈1〉 to enter data with the keypad. The display will be: Measured dens.x.x PCF Press ENTER when completed Input the average measured density and press 〈START/ENTER〉. The gauge will return to the READY display. Measurements may now be taken and stored. These measurements should represent the density at or near the core locations. Core Density – Measurement Entry From the Select method of... display above, press 〈2〉. The display is: # of Readings to average (1-20)? x Press ENTER Input the number of readings to be taken and averaged. Press 〈START/ENTER〉 for the display: - Reading # 1 Time: x min. Set rod to BS & Press START 6–20 Prepare the site for measurement. Place the gauge near the core location and press 〈START/ENTER〉. The display will be: - NOMOGRAPH Reading # 1 Depth: BS Time:xx sec. After counting down to zero, the display will be: If more than one reading is to be taken, move the gauge to the next location and press 〈START/ENTER〉. The gauge will repeat the above sequence for each reading. When all the readings are completed, the display will be: Average of x Reading(s)xxxxx PCF Press ENTER Press 〈START/ENTER〉 to return to the READY display. Thin layer overlay measurements may be taken and stored at this time. Model 3440 6–21 6. SPECIAL FUNCTIONS Reading # 1 WD:xxxxx PCF Move gauge and press START PRECISION Test results with precision limits as low as 0.5 pcf (8 kg/m3) are possible under certain conditions. When given the desired degree of precision and a test measurement, the 3440 automatically supplies this precision data to the operator. Assuming the precision requested is within range and the time required to calculate the required result does not exceed 60 minutes, the following procedure can usually produce the required results. The operator should be certain that the gauge is in the correct mode: Soil or Asphalt. To execute the Precision function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 two times for the display: YES- Next menu 6-SPECIAL CALIB. 7- NOMOGRAPH 8- PRECISION Press 〈8〉 for the display: Precision: x.xx PCF Press ENTER when completed If the gauge is in Soil mode, the display will request that the gauge be set for the correct depth and will instruct you to press 〈START/ENTER〉. Set depth for 1-min. count. Press START when ready. If the gauge is in Asphalt mode, the display will reflect that the gauge should be in the backscatter position. Depth: xx 1 min. count. Press START when ready. 6–22 If the gauge is in the correct position, press 〈START/ENTER〉. The display will be: Precision Mode Depth: xx Time: xx sec 1-min. count After counting down to zero, the display will be: min. to precision. to begin? CE to exit 6. SPECIAL FUNCTIONS xx reach Want Press Press 〈YES〉. After counting down, the display will be: (For Soil mode) %PR= xxx % DD= xxx WD= xxx M= xx %M= xx (For Asphalt mode) %MA= xxxx WD= xx % NOTE If the required precision cannot be reached in 60 minutes, the following screen will be shown displayed. -WARNING-Time to reach precision too high! Press CE to exit The precision mode will remain enabled until the gauge is turned off for more than ten seconds. Model 3440 6–23 SET UNITS The 3440 gauge allows the operator to select whether screen displays and printouts use measurement units of Metric or PCF. The Set Units function is described on page 2–4. 6–24 BAUD RATE The speed at which characters are transmitted to a printer or computer is determined by the setting of this special function. The baud rate is normally configured at bps or bits per second. A speed of 300 baud would then be approximately 300 bps. Valid baud rates for the 3440 gauge are 300, 600, 1200, 2400, and 4800. To execute the Baud Rate function, press 〈SHIFT〉 and 〈SPECIAL〉 for: 6. SPECIAL FUNCTIONS SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 three times for the display: YES- Next menu 9- SET UNITS 10- BAUD RATE 11-COMM PROTOCOL Press 〈1〉 〈0〉 for the display: Baud Rate = 600 Do you want to make a change? Press 〈YES〉 to change the baud rate. The display will be: Baud rate = 600 Input new rate. Press ENTER when finished. Input the new baud rate and press 〈START/ENTER〉. After a few seconds, the display will return to READY. NOTE If the wrong baud rate is entered, the error message – ERROR– Illegal baud rate is briefly displayed. The gauge then returns to the Do you want to make a change? display shown above. Model 3440 6–25 COMMUNICATION PROTOCOL The Communication Protocol function allows the operator to select whether a Line Feed is transmitted with a Carriage Return when the gauge is communicating with a printer or other external device. Consult the owners manual of your particular device to determine if you need to send a line feed. To execute the Communication Protocol function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 three times for the display: YES- Next menu 9- SET UNITS 10- BAUD RATE 11-COMM PROTOCOL Press 〈1〉 〈1〉 for the display: Comm. Protocol: Send LF w/ CR. Do you want to change protocol? Press 〈YES〉 for the display: Comm. Protocol Select: 1- Send LF w/CR 2- No LF w/CR Select the appropriate protocol by pressing 〈1〉 or 〈2〉. After entering a selection, the display will reflect the new protocol. Comm. Protocol: No LF w/CR. 6–26 BATTERY MONITOR The Battery Monitor function provides the operator with multiple methods for displaying the battery status. The recommended display method is actual battery voltage. However, the other choices may be useful. The following choices are available: ♦ Display the actual battery voltage ♦ Display the estimated battery life ♦ Display the elapsed time from the last charge SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 four times for the display: YES- Next menu 12- BATTERY 13- SOURCE DECAY 14- SPECIAL RDWY Press 〈1〉 〈2〉 for the display: - Battery Select: 1-Batt. Monitor 2-Batt. Status BATTERY MONITOR To change the way the battery is monitored and displayed, press 〈1〉 for: Battery Monitor1-Batt. Life 2-Batt. Voltage 3-Elapsed Time Model 3440 6–27 6. SPECIAL FUNCTIONS To execute the Battery Monitor function, press 〈SHIFT〉 and 〈SPECIAL〉 for: Battery Monitor – Battery Life Press 〈1〉 to display battery life. The display will be: Battery MonitorBatt. Life xx Do you want to change hours? If the battery hours remaining are correct, press 〈NO/CE〉. The gauge will return to the READY display. The remaining hours will be displayed. Press 〈YES〉 if the hours are to be changed. The display is: Battery MonitorBatt. Life --Input and press ENTER After inputting the new hours, the gauge will return to READY. Battery Monitor – Battery Voltage If the battery voltage is to be displayed on the READY screen, press 〈2〉 from the first Battery Monitor– display. The gauge will return to the READY display. The battery voltage will be shown on the third line of the display. Battery Monitor – Elapsed Time If the elapsed time is to be displayed on the READY screen, press 〈3〉 from the Battery Monitor– display. The display will be: Battery MonitorElapsed Time xxx Want to reset Elapsed Time? If the elapsed time shown is to be reset to zero, press 〈YES〉. The gauge will return to the READY display. 6–28 BATTERY STATUS The battery status may be checked at any time. From the – Battery – menu on page 6–27, press 〈2〉. The display will be: Battery Status Batt volts:x.x (7.0 = Batt Low) (Press YES) Last charge: mm/dd - x.x hrs Hrs used: x /p (Press YES) Press 〈YES〉 and the gauge will return to READY. Model 3440 6–29 6. SPECIAL FUNCTIONS Press 〈YES〉 to view the date when the gauge was last charged, how long the charge was and how many hours have elapsed since that charge. The display will be: SOURCE DECAY The Source Decay function calculates a theoretical set of standard counts. The radiation intensity of the source decreases by 2.3% per year. To compensate, the last calibration standard counts are “corrected” mathematically. The correction formula uses daily increments to compensate the calibration standard count. The “corrected” counts may be compared to the existing daily standard counts to determine if the gauge is operating properly. To execute the Source Decay function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈YES〉 four times for the display: YES- Next menu 12- BATTERY 13- SOURCE DECAY 14- SPECIAL RDWY Press 〈1〉 〈3〉 for the display: Std after Decay: Current: xx Predicted: xx %Diff: xx After a brief delay, the gauge will return to the Ready mode. 6–30 SPECIAL ROADWAY After taking readings and pressing 〈STORE〉, operators can let the gauge prompt them for the information required on U.S. FHWA projects. The gauge can then download the readings and other information to a computer or serial printer. When shipped from the factory, the default storage mode is the standard storage mode. To change the storage mode, select the Special Rdwy option on the Special function menu. The gauge will display the current status. Press 〈SHIFT〉 and 〈SPECIAL〉, then 〈1〉 〈4〉 for: The 〈YES〉 key toggles the function on and off. To accept the displayed status, press 〈NO/CE〉. Once the storage mode is changed, the gauge will remain in this mode until it is changed again. In both the Asphalt and Soil mode, the gauge can prompt the operator for: ♦ ♦ ♦ ♦ ♦ ♦ ♦ FHWA number Station number Distance from centerline Lane direction Elevation above original ground Elevation below finished grade Compaction method In the Asphalt mode, the gauge can prompt the operator for: ♦ ♦ ♦ Lift number Lift thickness Test type (surface mix, intermediate mix, or base mix) In the Soil mode, the gauge can prompt the operator for: ♦ ♦ Test type (soils or stone) Further test type (embankments, pipe backfill, subgrade, abutment backfill, retaining wall, or rockfill) ♦ If backfill, then pipe size ♦ If soil, then density required Model 3440 6–31 6. SPECIAL FUNCTIONS Special Roadway Mode: OFF Want to change? NOTES 6–32 CHAPTER 7 EXTENDED FUNCTIONS This chapter contains descriptions and operational instructions for the Extended Functions of the Series 3440 Surface Moisture-Density Gauge. These functions require the use of an Access Code and are to be executed by authorized personnel only. CONTENTS Time/Date.................................................................................................7–2 Customer Name........................................................................................7–3 Serial Number ..........................................................................................7–4 Calibration Date .......................................................................................7–5 Rod Length...............................................................................................7–8 Erase Standard Counts .............................................................................7–9 Memory Clear ........................................................................................7–10 Burn-In ...................................................................................................7–10 Calibration Standard...............................................................................7–11 Remote Control ......................................................................................7–12 Model 3440 7–1 7. EXTENDED FUNCTIONS Calibration Constants ...............................................................................7–6 TIME/DATE The Time/Date function allows the operator to change the current time-ofday or date that is displayed on the screen and used for measurement reports. This function is described on page 2–11. 7–2 CUSTOMER NAME The Customer Name function allows the operator to enter or change the name of the gauge owner for printouts and reports. To execute the Customer Name function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈19〉 for the display: EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. Enter the code and press 〈START/ENTER〉. The display will be: 7. EXTENDED FUNCTIONS YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈2〉 for the following display: Customer name ---------------Do you want to change name? Press 〈YES〉 to change the name. A--------------SHIFT- See chars YES- Select ENTER- Finished Press 〈SHIFT〉 to view a character of the alphabet. Press 〈YES〉 to select the character and advance the cursor to the next space. When the name entry is complete, press 〈START/ENTER〉. The display will return to the Extended Functions menu. Model 3440 7–3 SERIAL NUMBER The Serial Number function allows the operator to change the serial number of the gauge if this value is lost in memory. To execute the Serial Number function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈19〉 for the display: EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. Enter the code and press 〈START/ENTER〉. The display will be: YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈3〉 for the following display: Serial Number-----Want to change Serial Number? Press 〈YES〉 to change the current number. The display is: Serial NumberInput new Number and press ENTER Using the numeric keys, input the new serial number. When the input is complete, press 〈START/ENTER〉. The display will return to the Extended Functions menu. 7–4 CALIBRATION DATE The Calibration Date function allows the operator to change the date of the last gauge re-calibration. To execute the Calibration Date function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈19〉 for the display: EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. Enter the code and press 〈START/ENTER〉. The display will be: 7. EXTENDED FUNCTIONS YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈YES〉 for the following display: YES- Next menu 4- Calib. Date 5- Calib. Const. 6- Rod Length Press 〈4〉 for the display: DATE OF Calib: -------Press ENTER when complete. Using the numeric keys, input the new date. Press 〈START/ENTER〉 when completed. The display will return to the Extended Functions menu. Model 3440 7–5 CALIBRATION CONSTANTS The Calibration Constants function allows the operator to change the constants that the gauge uses to calculate a test result. If the gauge has been repaired, or the memory has been lost, the constants must be re-entered. NOTE Each gauge contains a unique set of constants. These constants are developed at the factory and are reflected on the factory calibration sheet, which is shipped with every gauge. If the constants are changed or are entered incorrectly, gauge measurements will be incorrect. To execute the Calibration Constants function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈19〉 for the display: EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. Enter the code and press 〈START/ENTER〉. The display will be: YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈YES〉 for the following display: YES- Next menu 4- Calib. Date 5- Calib. Const. 6- Rod Length 7–6 Press 〈5〉 to select Calibration Constants. The display will be: -CALIB. CONST.Enter Code ---and Press ENTER The Secondary Access Code must be entered. Enter the code and press 〈START/ENTER〉. The display is: -CALIB. CONST.E=x Select: 1 = + 2 = If the value for the “E” constant on the screen equals that shown on the calibration sheet, press 〈START/ENTER〉 to select the next constant. If the “E” value is not correct, press 〈1〉 for positive number or 〈2〉 for negative number and enter the value as reflected on the factory calibration sheet. Press 〈START/ENTER〉 when completed. 7. EXTENDED FUNCTIONS NOTE If your calibration sheet lists B*1000 and F*1000 values, the calibration is in metric units. Therefore, the values on the calibration sheet can be entered into the gauge without conversion. If your calibration sheet lists B and F values, then the calibration is in English units rather than metric. Because the gauge requires metric units, the B and F values must be converted to metric units and divided by 1000. The most direct conversion method is to multiply the B and F values by 62.4298, then enter the resulting product into the gauge. The gauge will request a new value for the “F” constant. After accepting or entering a corrected value for the “F” constant, the gauge requests a depth value. The display is: -CALIB. CONST.Enter depth: - inches (Press ENTER) After the depth entry (0 for backscatter, 2 for 2 inches, etc.), accept or enter the “A”, “B*1000” and “C” constants from the calibration sheet. Repeat this procedure for every depth shown on the calibration sheet. When all constants have been entered, the display will return to the Extended Functions menu. Model 3440 7–7 ROD LENGTH If the gauge depth seems incorrect after calibrating the depth, the Rod Length function allows the operator to set the rod length manually. The operator should not need to set the rod length unless a new front panel assembly is installed or the RAM is corrupt. To set the rod length manually, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈19〉 to display: EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. Enter the code and press 〈START/ENTER〉. The display will be: YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈YES〉 for: YES- Next menu 4- Calib. Date 5- Calib. Const. 6- Rod Length To access the Rod Length function, press 〈6〉. The gauge provides the operator with two rod-length options: 8 or 12 inches. The display is Set Rod Length Select: 1- 8 Inch 2- 12 Inch Press 〈1〉 or 〈2〉 to select the correct rod length. The gauge returns to the Extended Functions menu. To return to the Ready mode, press 〈NO/CE〉. If taking measurements in the automatic indicator mode, the operator should calibrate the depth strip after manually entering the rod length. 7–8 ERASE STANDARD COUNTS The Erase Standard Counts function allows the operator or service technician to remove all (four) standard counts from gauge memory. To execute the Erase Standard Counts function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈19〉 for the display: EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. 7. EXTENDED FUNCTIONS Enter the code and press 〈START/ENTER〉. The display will be: YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈YES〉 two times for: YES- Next menu 7- Erase Std Cts 8- Memory Clear 9- Burn-in Press 〈7〉 for the display: Press SHIFT/YES to erase all Standard Counts in gauge?! Press 〈SHIFT〉 and 〈YES〉 to erase all the standard counts. The display will return to the Extended Functions menu. Model 3440 7–9 MEMORY CLEAR The Memory Clear function is to be used by qualified service personnel only! CAUTION Executing the Memory Clear function will erase critical information that is necessary for gauge operation! BURN-IN The Burn-In function is to be used by qualified service personnel only! 7–10 CALIBRATION STANDARD The Calibration Standard function allows the operator or service technician to input the factory calibration standard count. To execute Calibration Standard function, press 〈SHIFT〉 and 〈SPECIAL〉 for: SPECIAL FUNCTION YES- Next menu 1- STAT TEST 2- DRIFT TEST Press 〈19〉 to display: EXTENDED FUNCTIONS Enter Code ---and Press ENTER The Access Code must be entered. This code is for authorized personnel only. If you need assistance with the code, contact your nearest Troxler Representative. Enter the code and press 〈START/ENTER〉. The display will be: 7. EXTENDED FUNCTIONS YES- Next menu 1- Time/Date 2- Customer Name 3- Serial Number Press 〈YES〉 three times for: YES- Next menu 10- Calib. Std 11- Remote Cntrl Press 〈1〉 〈0〉 to display: Calib. Standard Count: x Input and Press ENTER Input the calibration standard count and press 〈START/ENTER〉. The display will return to the Extended Functions menu. Model 3440 7–11 REMOTE CONTROL The Remote Control function is to be used by qualified service personnel only! 7–12 APPENDIX A THEORY OF OPERATION This appendix contains brief explanations of the principle of operation of the 3440 gauge. Backscatter mode and direct transmission mode are explained, as well as the theoretical differences between the density and moisture measurement technique. CONTENTS Density ....................................................................................................A–2 Moisture ..................................................................................................A–5 A. THEORY OF OPERATION Model 3440 Appendix A–1 DENSITY The Troxler Model 3440 Surface Moisture-Density Gauge uses two modes of operation: direct transmission mode (with the source rod extended) and backscatter mode (with the source rod placed in the first notch below the SAFE position). In direct transmission mode (Figure A–1), the rod containing the cesium-137 source is lowered to the desired depth, resulting in gamma radiation emission. The detectors in the gauge base measure this radiation. Gamma photons reaching the detectors must pass through the full distance of soil, resulting in a large number of photons colliding with electrons present in the soil. These collisions reduce the number of photons that reach the detectors and thus allow the density of the material to be calculated. In short, the lower the number of photons reaching the detectors, the higher the density of the material. Since all of the photons arriving at the detectors have passed through the full distance from the source depth, the measured density is a true average for the material between the surface and the source. Figure A–1. Direct Transmission Geometry Appendix A–2 In backscatter mode (Figure A–2), the gamma photons entering the material being measured must be scattered (or reflected) at least once to reach the detectors in the gauge. This result is achieved by leaving the source rod in a retracted position in the gauge. The rod is locked in the first position below the SAFE position, placing the source and detectors in the same horizontal plane. Under ideal conditions, no photons would reach the detector in a direct path from the source. However, this is not practical and the directly transmitted photons are generally held to 10% or less of the backscattered photons. Since no hole is required for the source rod, the backscatter mode is preferred for surfaces such as concrete and asphalt. However, since the photons have been scattered at least once, the average energy of the photons reaching the detectors is lower than the average energy of the photons in direct transmission conditions. This increases the possibility of error due to the effects of chemical composition. While the direct transmission geometry measures the average density from the source depth to the surface, the backscatter geometry yields an average that is heavily weighted by the surface density. Since the photon path length increases for photons passing through deeper material, this is a normal phenomena as most of the photons reaching the detectors will have passed through the surface material and have decreasing percentages through deeper materials. A. THEORY OF OPERATION Figure A–2. Backscatter Geometry Model 3440 Appendix A–3 A normalized curve showing the percentages of photons at the detectors for various depths is shown in Figure A–3. These curves can be used to compute the gauge response to layered materials of different densities. As an example, for a 20-mm thick material with a density of 2000 kg/m3 overlaying a material with a density of 2150 kg/m3, the gauge measurement will indicate a density of 2089 kg/m3 [(2000 x 0.41) + (2150 x 0.59)]. Figure A–3. Depth of Top Layer Appendix A–4 MOISTURE The 3440 gauge uses a 40mCi americium 241:beryllium source with a yield of 70,000 neutrons per second. The neutrons are used to measure the hydrogen (water) content of the material because hydrogen in the material will thermalize the neutrons. Thermalization is the process whereby fast (or epithermal) neutrons emitted by the source are slowed to velocities where additional collisions with hydrogen or other molecules will not further slow the neutrons. Neutron thermalization requires an average of nineteen collisions with hydrogen nuclei. Other elements require a greater number of collisions for thermalization (Table A–1). Due to their high sensitivity, the He-3 detectors in the 3440 gauge are very suitable for detection of thermalized neutrons. The He-3 tube is insensitive to fast neutrons and the counts obtained are directly proportional to the hydrogen/moisture content of the material. Additionally, there are elements that absorb, or capture, neutrons into their nuclei. The extent to which neutrons interact with nuclei is described in terms of quantities known as cross-sections. The larger the cross-section, the greater the possibility of neutron capture or absorption. Boron, cadmium, and chlorine are elements that have large cross-sections and may cause problems when attempting to measure the moisture of materials containing these elements. An offset may be needed for the gauge if moisture measurements are required in these materials. Depth of Measurement, or the depth below which no more than 2% of the counted neutrons pass before reaching the detector, is generally a function of the moisture content. The normalized curve set shown in Figure A–4 illustrates the effects of moisture content on the radius of measurement. This function may be expressed by: or Depth (mm) = 280 – (0.27 × M), where M = Moisture in kg/m3 Model 3440 Appendix A–5 A. THEORY OF OPERATION Depth (in.) = 11 – (0.17 × M), where M = Moisture in pcf Table A–1. Neutron Thermalization and Absorption Data Element Weight Fractions of Earth’s Crust Collisions to Thermalization Thermal Absorption Cross Section Hydrogen 0.0014 19.0 0.33 Boron * 109.2 759.00 Carbon * 120.6 0.0034 Nitrogen * 139.5 1.90 Oxygen 0.466 158.5 0.002 Sodium 0.028 224.9 0.53 Magnesium 0.021 237.4 0.063 Aluminum 0.081 262.8 0.23 Silicon 0.277 273.3 0.16 Phosphorus 0.001 300.8 0.19 Sulfur * 311.1 0.51 Chlorine * 343.3 33.00 Potassium 0.026 378.0 2.10 Calcium 0.036 387.3 0.43 Titanium 0.004 461.6 6.10 Manganese 0.001 528.5 13.30 Iron 0.050 537.2 2.53 * 1074.6 2390.00 Cadmium Lead * 1975.5 0.17 Uranium * 2268.6 4.20 * Weight fraction is less than 0.1%. Appendix A–6 Gauge Measured Moisture Content (pcf) 40 DepthE = 11 - (0.17 X M) 35 32.3 pcf 30 25 20.8 pcf 20 15 10 5.95 pcf 5 0 2 4 6 Depth (in) 8 10 12 640 DepthS = 280 - (0.27 X M) 560 521 kg/m3 480 400 336 kg/m3 320 240 160 96 kg/m3 80 0 0 50 100 150 200 Depth (mm) 250 300 Figure A–4. Effect of Moisture on Depth of Measurement Model 3440 Appendix A–7 A. THEORY OF OPERATION Gauge Measured Moisture Content (kg/m3) 0 NOTES Appendix A–8 APPENDIX B This appendix is required reading for anyone who will operate the 3440 nuclear system. This appendix covers radiation theory, along with a brief explanation of radiation statistics and radiation terminology. CONTENTS Atomic Structure .....................................................................................B–2 Radiation Theory.....................................................................................B–3 Radiation Statistics..................................................................................B–4 Radiation Terminology ...........................................................................B–5 Model 3440 Appendix B–1 B. RADIATION THEORY RADIATION THEORY ATOMIC STRUCTURE A more detailed discussion of radiological theory can be found in the Troxler Nuclear Gauge Safety Training Program manual, provided at the Troxler safety class. All materials consist of chemical elements that can not decompose by ordinary chemical methods. Some examples are: (H) Hydrogen (U) Uranium (C) Carbon (Cf) Californium (O) Oxygen (Co) Cobalt The fundamental component of an element is an atom with a unique structure (Figure B–1). The atom consists of smaller particles such as protons, neutrons, and electrons. The protons and neutrons are grouped together in the nucleus. The electrons orbit the nucleus. An atom is normally electrically neutral because the positive charges of the protons cancel out the negative charges of the electrons. ELECTRONS NEUTRON PROTON Figure B–1. Diagram of an Atom Protons carry a positive charge and are described as having a mass of one. Neutrons have a neutral charge and also have a mass of one. Electrons carry a negative charge and essentially have no mass. MASS (ATOMIC WEIGHT SCALE) CHARGE Protons 1.0073 +1 Neutrons 1.0087 0 Electrons 0.0006 –1 Since protons and neutrons are clustered together in the nucleus, the mass of an atom is concentrated in the nucleus. The atom in Figure B–1 has two protons and two neutrons; therefore, it is a helium atom. The atomic weight of an atom is the sum of the protons and neutrons. Appendix B–2 RADIATION THEORY The strength of radioactive material is measured by its activity, or rate of decay. This activity decreases with time. The length of time it takes a given amount of radioactive material to decay to half of its original strength is referred to as the half-life. The Am-241:Be source in the Model 3440 has a half-life of 432 years, while the half-life of the Cs-137 source is thirty years. Model 3440 Appendix B–3 B. RADIATION THEORY Radioactivity is the spontaneous breakdown of unstable nuclei (radioisotopes) with the resulting emission of radiation. The basic unit of radiation used in the U.S.A. is the curie (Ci). The curie is defined as 3.7 x 1010 disintegrations of nuclei per second. In the “Special Form” encapsulated sealed sources used in the 3440 gauge, the unit of measure is the millicurie (1/1,000 of a curie). The SI unit of radiation is the becquerel (Bq). The becquerel equals one disintegration per second. Therefore, one curie equals 3.7 x 1010 becquerel. RADIATION STATISTICS Radioactive emission is a random process. The number of emissions in a given time period is not constant but varies statistically about an average value. The variation in the number of emissions follows a Poisson distribution about the true mean value. In this distribution, the standard deviation (σ) about the mean (n) is defined as: σ = √n. When the mean is greater than 100, the Poisson distribution can be closely approximated by the normal distribution (Figure B–2). The normal distribution predicts the probability that any given count rate will fall within a selected region about the mean. Normal Distribution -3σ -2σ -1σ MEAN 68.3% 95.4% 99.7% 1σ 2σ 3σ Figure B–2. Variation of Radioactive Emission Using the mean of a larger number of counts to approximate the true mean, the distribution shows that 68.3% of the time the count rate obtained will be within ±1 standard deviation of the mean. Figure B–2 shows the probabilities of counts falling within three standard deviations of the mean. The operator can perform a statistical stability test (stat test) to compare the experimental standard deviation to the theoretical standard deviation (see page 6–3). Appendix B–4 RADIATION TERMINOLOGY Model 3440 Appendix B–5 B. RADIATION THEORY The radiation absorbed dose, or rad, is the unit of absorbed dose equal to 0.01 Joules/kg in any medium. To account for the effect of various types of radiation on biological tissue, the roentgen equivalent man (rem) or, more appropriate for Troxler operators, the millirem (mrem) is used when measuring radiation dose. The unit rem is derived from scaling the radiation absorbed dose (rad) by a quality factor (QF). One rad is equal to the exposure of one rem of photon radiation. For example, the average neutron energy of an americium-241:beryllium source is 4.5 MeV. The quality factor (QF) for this radiation is about 10. The absorbed dose of 1 rad of neutron radiation produces a dose equivalent of (absorbed dose x QF) 10 rem. NOTES Appendix B–6 APPENDIX C RADIOLOGICAL SAFETY This appendix is required reading for anyone who will operate the 3440 nuclear system. This appendix provides a brief discussion of general radiation safety. The exposure profile for the Model 3440 gauge is also included, along with a discussion of the source encapsulation. Radiation Types ......................................................................................C–2 Radiation Exposure .................................................................................C–3 Monitoring Radiation ..............................................................................C–4 3440 Radiation Profile ............................................................................C–4 Source Encapsulation ..............................................................................C–6 Emergency Procedures............................................................................C–6 Model 3440 Appendix C–1 C. RADIOLOGICAL SAFETY CONTENTS RADIATION TYPES The radioactive sources in the Model 3440 gauge produce four types of radiation: Alpha Particles Beta Particles Photons (Gamma Rays) Neutrons The alpha and beta particles are stopped by the source capsule. Only the photon and neutron radiation contributes to any occupational radiation exposure. Photon (gamma) radiation is electromagnetic radiation, as are x-rays, radio waves, and visible light. Visible light and photons have no mass or electrical charge, and travel at the speed of light. Photons are energetic and penetrating. Photons originate from the nucleus of the cesium-137 atom after radioactive decay. The number of photons emitted from a 3440 gauge is very low. Dense materials (i.e., lead, cadmium, etc.) provide the best shielding against photon radiation. Neutron radiation allows measurement of the hydrogen (water) content in a material because the neutrons are slowed by collisions with materials containing hydrogen atoms (for example: water, polyethylene, etc.). Neutrons have no charge and are very penetrating. Appendix C–2 RADIATION EXPOSURE Government agencies set occupational exposure limits. The current limit in the United States and in many other countries is 5,000 mrem per year. Under average conditions, a full time employee working with the 3440 gauge will receive less than 200 mrem per year. Taking advantage of all available means to limit radiation exposure is always recommended. The three methods of limiting exposure are: These methods are a part of the ALARA (As Low As Reasonably Achievable) concept. The simplest way to reduce exposure is to keep the time spent around radioactive source to a minimum. If time is cut in half, so is the exposure, with all other factors remaining constant. Distance is another effective means to reduce radiation exposure. A formula known as the “inverse square law” relates the radiation exposure rate to distance (Figure C–1). Doubling the distance from a radiation source reduces the exposure to one-fourth its original value. If the distance is tripled, the exposure is reduced by a factor of nine, etc. Figure C–1. Effect of Distance on Exposure Shielding is any material used to reduce the radiation reaching the operator from a radioactive source. While some types of radiation such as alpha particles may be stopped by a single sheet of paper, other radiation such as photons and neutrons require much more shielding. Dense materials, such as lead, shield photons. Materials containing large amounts of hydrogen, such as polyethylene, shield neutrons. The Model 3440 has built-in shielding to reduce the exposure. Model 3440 Appendix C–3 C. RADIOLOGICAL SAFETY TIME DISTANCE SHIELDING MONITORING RADIATION In the United States, anyone working with or near radioactive materials is subject to the limits on occupational exposure mentioned earlier and must complete a radiation safety training course to be designated an authorized operator. An individual designated as an authorized operator must work in a “controlled” environment to the extent that their exposure to radiation must be monitored. Several means of personnel monitoring or dosimetry exist; the most common methods are film badges and thermoluminescent dosimeter (TLD) badges. In Canada, nuclear gauge operators are not normally classified as Atomic Radiation Workers. In such cases, the general public dose limit of 0.5 rem/yr. would apply. Operators may not be required to wear a dosimeter. To establish the personnel monitoring requirements for your application, consult the conditions of your radioisotope license and the Atomic Energy Control Board regulatory document R91 “Monitoring and Dose Recording for the Individual”. 3440 RADIATION PROFILE Table C–1 shows the radiation profile for the 3440 gauge. The table lists the radiation dose equivalent rates (in mrem/hour) for each side of the gauge and transport case shown in Figure C–2. LEFT SIDE TOP BACK LEFT SIDE TOP BACK FRONT FRONT RIGHT SIDE BOTTOM TRANSPORT CASE RIGHT SIDE BOTTOM GAUGE Figure C–2. 3440 Gauge and Transport Case Appendix C–4 Table C–1. Radiation Dose Equivalent Rates in Millirems per Hours 30 CENTIMETERS Total Gamma Neutron Total Gamma Neutron Total Gamma Neutron Total 1 METER Neutron 10 CENTIMETERS Gamma SURFACE Front 19.0 1.5 20.5 5.0 0.5 5.5 1.2 0.2 1.4 0.3 0.1 0.4 Back 18.0 0.8 18.8 6.0 0.4 6.4 1.5 0.2 1.7 0.2 0.1 0.3 Left Side 13.0 0.4 13.4 3.0 0.2 3.2 0.8 0.1 0.9 0.1 0.1 0.2 Right Side 12.0 0.6 12.6 6.0 0.3 6.3 3.0 0.2 3.2 0.5 0.15 0.65 Top 19.0 0.9 19.9 3.5 0.6 4.1 0.6 0.4 1.0 0.15 0.1 0.25 Bottom 18.0 1.9 19.9 4.0 1.2 5.2 0.7 0.4 1.1 > 0.1 0.1 Front 8.0 0.3 8.3 3.0 0.25 3.25 1.1 0.15 1.25 0.3 > 0.3 Back 7.0 0.3 7.3 3.0 0.2 3.2 1.0 0.1 1.1 0.2 > 0.2 Left Side 0.4 0.1 0.5 0.3 > 0.3 0.2 > 0.2 0.2 > 0.2 Right Side 6.0 1.4 7.4 2.0 0.7 2.7 0.6 0.25 0.85 0.2 0.1 0.3 Top 13.0 0.2 13.2 3.0 0.15 3.15 0.9 > 0.9 0.3 > 0.3 Bottom 8.0 0.3 8.3 4.0 0.1 2.5 0.1 2.6 0.6 > 0.6 GAUGE 4.1 NOTES: 1. Dose rates measured by the State of North Carolina Department of Environment, Health, and Natural Resources, Division of Radiation Protection. 2. All values in mrem/hr. 3. Indicates a reading of less than 0.1 mrem/hr. 4. Dose equivalent rates for 0.3 GBq (8 mCi) Cs-137 gamma source and 1.48 GBq (40 mCi) Am-241:Be neutron source 5. Surface indicates surface of gauge or surface of transport case. 6. Gamma measurements made with Ludlum Model 14C Survey Meter, calibrated September 1987. 7. Neutron measurements made with Nuclear Research Corporation Model NP-2 Survey Meter, calibrated March 1988. Model 3440 Appendix C–5 C. RADIOLOGICAL SAFETY TRANSPORT CASE WITH GAUGE SOURCE ENCAPSULATION The source in the 3440 gauge meets regulatory requirements of U.S. and international authorities as “Special Form” sealed source material. The sealed sources used are encapsulated to prevent leakage of the radioactive material and meet radiation safety requirements. The neutron (americium-241:beryllium) and photon (cesium-137) sources are compressed and welded inside stainless steel capsules. Proper use of this instrument (following the instructions in this manual) and the shielding design of the instrument will keep the exposure levels at a minimum under normal conditions. The operator, however, may be required to wear personnel dosimetry when using the 3440. EMERGENCY PROCEDURES If the nuclear gauge is lost or stolen, immediately notify the Radiation Safety Officer (RSO). The gauge owner should complete the emergency contact information on the lines furnished below: The company RSO is ________________________________________ Call the RSO at _____________________________________________ The regulatory agency is ______________________________________ Call the agency at ___________________________________________ If a gauge is damaged, then follow the steps below: Locate the gauge and/or source. Do not touch or move the gauge. Immediately cordon off an area around the nuclear gauge and/or source. A radius of 15 ft (5 m) will be sufficient. Do not leave the area unattended. Keep all unauthorized personnel away from the nuclear gauge. Appendix C–6 If a vehicle is involved, it must be stopped until the extent of contamination, if any, can be established. The gauge operator should perform a visual inspection of the nuclear gauge to determine if the source housing and/or shielding has been damaged. Use a survey meter to measure the dose rate at a distance of 3 ft (1 m) from the gauge. Contact the company RSO (name and number given at the beginning of this section). Provide the RSO with the following: The date, time, and location of the accident ♦ The gauge model and serial number ♦ The nature of the accident ♦ The location and condition of the gauge and/or source ♦ The dose rate at 3 ft (1 m) from the gauge If you are unable to reach the RSO, then call your regulatory agency (name and number given at the beginning of this section). Follow the instructions of the RSO. The RSO may be required to report the incident to the regulatory agency. The RSO may also be required to notify the USDOT of accidents during transport. Take a picture of the gauge. Perform a leak test and send the leak test wipe to Troxler for analysis. The Troxler RSO will contact you and give you an RGA (returned goods authorization). Do not ship a damaged gauge to Troxler without an RGA. Model 3440 Appendix C–7 C. RADIOLOGICAL SAFETY ♦ NOTES Appendix C–8 APPENDIX D LICENSING REQUIREMENTS This licensing requirements outlined in this appendix apply to all Troxler gauge operators in the United States. Operators of Troxler products outside the U.S. should contact the Troxler International, Ltd. representative in their area or the factory for more information on regulatory requirements. CONTENTS Specific License ......................................................................................D–2 Radiation Safety Program .......................................................................D–2 Personnel Monitoring..............................................................................D–2 Security ...................................................................................................D–2 Leak Test.................................................................................................D–3 Notice to Employees ...............................................................................D–3 Incidents ..................................................................................................D–3 Disposal...................................................................................................D–4 Record Keeping.......................................................................................D–4 Model 3440 Appendix D–1 D. LICENSING Training...................................................................................................D–3 SPECIFIC LICENSE Possession and use of the radioactive material found in a nuclear gauge requires the operator to maintain a specific license issued by the United States Nuclear Regulatory Commission or equivalent agreement state licensing agency. Troxler Electronic Laboratories, Inc. will provide assistance and information for obtaining the necessary license. RADIATION SAFETY PROGRAM The specific license dictates that nuclear gauge operators maintain control of the gauge in a manner to minimize radiation exposure to operators and the general public. To determine the requirements applicable to your operation, review both your licensing documents and the radiation safety procedures accompanying your license application. The following requirements are generally accepted by all licensing agencies in the U.S. PERSONNEL MONITORING Most licensing agencies do require personnel monitoring devices to be worn by nuclear gauge operators. These devices record the actual radiation to which the operator is exposed while operating the equipment. When the equipment is not in use, the personnel monitoring device should be kept in a radiation free area. The reports of radiation exposure supplied by the personnel monitoring service must be maintained on file and available for review by employees and the licensing agency. Contact Troxler or their appointed representative for help in obtaining personnel monitoring devices or radiation monitoring equipment. SECURITY Regulations further require that operators provide all necessary security to prevent unauthorized use or removal of the gauge when in storage or in transit. The transport case can be locked. A lock may be obtained from Troxler (Part Number 12176). The gauge is supplied with a lock to secure the source rod in the SAFE, shielded position. Appendix D–2 TRAINING Troxler Electronic Laboratories, Inc. continues to provide a one-day training seminar for nuclear gauge operators and individuals designated as Radiation Safety Officers. This class provides radiological physics familiarization and instruction in the safety procedures required to operate a nuclear gauge. Operators outside the U.S. should contact their Troxler representative for training and assistance. LEAK TEST NOTICE TO EMPLOYEES Figure D–1 on page D–6 is a reduced example of a “NOTICE TO EMPLOYEES” (supplied by your licensing agency) that must be posted where employees that use the gauge and others who may enter the gauge storage area can see it. This notice must also be posted wherever and whenever temporary gauge storage is necessary (in a trailer on the job site, etc.). It is not required (or advisable) to post the notice on the outside of any structure where it may be viewed by the general public. INCIDENTS If the gauge is removed from the custody of the licensee (i.e., lost, stolen, etc.) or physically damaged to the extent that the source or shielding may be or is compromised (open or danger), the licensee must notify the licensing agency immediately. Call Troxler or your appointed representative for further advice or assistance. Model 3440 Appendix D–3 D. LICENSING Unless specified otherwise by your radioactive material license, the gauge must be leak tested at intervals not exceeding 12 months to ensure the integrity of the radioactive sources contained in the gauge. First, make sure the source rod is locked into the storage, or SAFE, position. Using the Troxler Model 3880 Leak Test Kit and accompanying instructions, remove the control panel from the gauge front. Looking down into the gauge interior locate the yellow radiation label in the middle of the gauge base and on the top of the circuit board. Wipe the yellow label with the wipe disk, using the tongs to hold the disk. Next, turn the gauge on one side and locate the opening through which the source rod would protrude when indexing. Wipe around and into this opening. Pack the disk, as instructed, in the envelope and mail to Troxler Electronic Laboratories, Inc. for analysis. Reassemble the gauge and secure properly. DISPOSAL Regulations require that very strict methods be followed when disposing of radioactive materials. For specific information concerning disposal costs and available disposal sites, please contact the factory or your local Troxler representative if outside of the United States. Generally, a Troxler gauge may be transferred to another authorized licensee or returned to our factory for final disposal. RECORD KEEPING In the United States, the following documentation should be maintained and available for review by the licensing agency. (Outside the United States, refer to your local licensing authority/agency or contact your authorized Troxler representative). 1. Radioactive Material License. A valid license must be on file and should be periodically reviewed to ensure that all conditions are current. 2. Troxler Training Certificate. A certificate of attendance at a Troxler Training Course is required by each operator. 3. Regulations. A copy of the regulations furnished by your licensing agency should be maintained as part of your records. 4. Leak Test Report. As a means of verification of compliance with current regulations, a copy of each Leak Test Report for each gauge in your possession must be on file. (It is also a good idea to make a note when Leak Tests are performed and sent. If you are inspected before test results are returned, you have de facto (existing) proof of compliance.) 5. Personnel Monitoring Report. A copy of each personnel monitoring report must be maintained in your files. This report must contain the wearer’s name, social security number, date of birth, and actual exposure received. 6. Inventory, Receipt, and Transfer Records. A log to show gauge location should be maintained. Whenever a gauge is removed from a permanent to a temporary location, it should be noted on the log. When removed from a temporary site, a log should show where the gauge is being used. Operators should be required to sign out when removing gauges and sign in when returning them. The licensee is required to know where each gauge is at all times. Appendix D–4 7. Gauge Certificate. New gauges are shipped with a gauge certificate denoting the radioactive material used in the gauge. The original gauge certificate should be on file and a copy carried with the gauge at all times. 8. Transport Package Certification. The regulations require that each shipper of a (Specification 7A) Type A package “must maintain on file for at least one year after the latest shipment” testing methods and results for the Type A package. Troxler has provided this document, entitled “Testing Results for Type A Package for the 3440...” with the 3440 documentation. 9. Transportation – Bill of Lading. U.S. DOT regulations require properly prepared shipping documents when transporting a gauge (see Appendix E). 10. Emergency Procedures. The telephone number of the licensing agency and procedures for reporting incidents should be kept on file for the convenience of concerned personnel. 11. Notice to Employees Poster. A copy of this notice should be on file. A facsimile of this notice is on page D–6. For display requirements, see the Notice to Employees section on page D–3. D. LICENSING Model 3440 Appendix D–5 Figure D–1. Sample Notice to Employees Appendix D–6 APPENDIX E TRANSPORTATION AND SHIPPING Devices containing radioactive materials must be transported in accordance with the rules of the U.S. Department of Transportation (DOT) and the International Atomic Energy Agency (IAEA). The IAEA recommendations have been codified in the International Air Transport Association (IATA) Dangerous Goods Regulations. International customers should consult their local government or licensing authority for applicable regulations. CONTENTS U.S. Requirements .................................................................................. E–2 Accident Notification Requirements................................................ E–3 Hazmat Training .............................................................................. E–3 Canadian Shipping Requirements ........................................................... E–4 E. TRANSPORTATION Model 3440 Appendix E–1 U.S. REQUIREMENTS The U.S. DOT hazmat regulations (49 CFR, Parts 100–185) apply any time a nuclear device is transported by motor vehicle on a public highway or by other means of transport (rail, air, ship). The major requirements for transporting a nuclear gauge in the United States are listed below. For more detailed information about these requirements, please refer to the Troxler Transportation Guide. ♦ A copy of the current IAEA Certificate of Competent Authority for each source in the gauge (Special Form Certificate) must be kept on file. Current versions can be downloaded from the Troxler website, www.troxlerlabs.com. ♦ A copy of the results of the Type A package testing must be kept on file. ♦ Hazmat employee training records must be kept on file. ♦ An Emergency Response Information document must be in the vehicle and immediately accessible to the driver. ♦ A properly completed bill of lading must be in the vehicle and immediately accessible to the driver. The shipping papers must include a 24-hr emergency response phone number. ♦ If shipping by air, a Shipper's Declaration for Dangerous Goods must accompany the air waybill. ♦ The package must be properly marked and labeled in accordance with hazmat regulations. ♦ The package must have a tamper-evident seal. ♦ The package must be inspected prior to each shipment. ♦ The package must be securely blocked and braced in the vehicle to prevent shifting during transport. Appendix E–2 ACCIDENT NOTIFICATION REQUIREMENTS In the event of a reportable incident involving radioactive material, notify the licensing agency as soon as practical. The operator is also required to notify, at the earliest practical moment, the U.S. DOT at 1-800-424-8802 of an accident that occurs during the course of transportation (including loading, unloading, and temporary storage) in which fire, breakage, spillage, or suspected contamination occurs involving shipment of radioactive materials. HAZMAT TRAINING The U.S. DOT regulations require every hazmat employer to train, test, certify, and maintain records for each hazmat employee. Hazmat training applies to anyone who transports or prepares for transport radioactive materials. Refresher training is required every three years. E. TRANSPORTATION Model 3440 Appendix E–3 CANADIAN SHIPPING REQUIREMENTS The Transportation of Dangerous Goods Act and Regulations (TDG) and Transport Packaging of Radioactive Materials Regulations (TPRM) apply any time a nuclear device used in commerce is transported by any means in Canada. For training and accident notification requirements, consult the Transportation Of Dangerous Goods Regulations. For further information on transporting a nuclear device, contact the transportation section of The Canadian Nuclear Safety Commission (CNSC). Appendix E–4 PERIODIC MAINTENANCE AND SERVICE This appendix contains important information for maintaining and servicing the 3440 Surface Moisture-Density Gauge. The following procedures should be performed to keep the 3440 gauge in working order. In the event that a serious problem with the gauge arises, contact your nearest Troxler Service Center or Troxler representative for instructions. CONTENTS Batteries .................................................................................................. F–2 Alkaline Battery Use........................................................................ F–2 Mechanical Maintenance......................................................................... F–4 Sources............................................................................................. F–4 Inspecting Source Rod ..................................................................... F–4 Cleaning the Tungsten Sliding Block............................................... F–4 Scraper Ring/Tungsten Sliding Block Maintenance ........................ F–5 Gaskets............................................................................................. F–7 3440 Gauge Top Shell...................................................................... F–8 Electronic Maintenance........................................................................... F–9 3440 Baseboard Assembly............................................................... F–9 Troubleshooting ............................................................................. F–10 Possible Malfunction Indicators..................................................... F–14 Error Messages............................................................................... F–15 Replacement Parts................................................................................. F–16 Accessories .................................................................................... F–26 Optional Accessories...................................................................... F–26 Maintenance Supplies .................................................................... F–26 Returning a Gauge for Service .............................................................. F–27 Model 3440 Appendix F–1 F. PERIODIC MAINTENANCE APPENDIX F BATTERIES The 3440 gauge constantly updates the battery condition. Depending on operator preference, either the hours remaining or the battery voltage will be displayed. A fully charged battery will last approximately eight weeks under normal working conditions (eight-hour days) before recharging is necessary. A low battery condition is indicated by the following display: READY mm/dd/yyyy Depth: xxxx Time: x.xx min <Batteries low!> When this display appears, the battery must be recharged within a few hours. For best results, perform a full recharge with the ac charger. To access the charger port, unscrew the two captive screws that attach the black access cover to the front panel. Plug the charger into the round plug under the access cover. Plug the other end of the charger into a standard 110 V ac wall outlet. For a full charge, recharge the gauge for at least sixteen hours. If necessary, for emergency use, a 30-minute recharge with the dc or ac charger provides several hours of use. After a short recharge time, the battery life display may not always indicate an accurate update of the battery-life. Only a full sixteen-hour recharge will reset the battery life indicator. NOTE For more information on updating or resetting the battery monitor, refer to page 6–27. Although batteries cannot be “overcharged”, rechargeable batteries have a memory and repeated needless charging will shorten the battery life. Do not recharge batteries unless the battery indicator signals a low battery. ALKALINE BATTERY USE If recharging is not an option, alkaline batteries may be used. An accessory battery case assembly (PN 104148) is available as an option. CAUTION Do not mix alkaline and rechargeable batteries in the gauge. Charging may cause alkaline batteries to explode! Appendix F–2 For gauge serial numbers below 21675, the High Voltage Module must be discharged before replacing the battery packs, as described in the following paragraphs. To replace the battery packs, first turn the gauge off. Remove the display panel assembly and unplug it from the gauge. Press the high voltage discharge switch located on the baseboard. The battery leads are red and black, and are connected to white plugs on the left side of the baseboard. Unplug both battery leads and the depth indicator strip’s clear leads in order to prevent possible damage. Remove the screws holding the battery packs in place and slide the packs toward the baseboard. Lift out the battery packs and replace with the alkaline packs. Reverse the procedure for connecting all the leads and reinstalling the display panel. Remember to plug the depth indicator strip leads and battery packs back in. For Serial Numbers 21675 and Above For gauge serial numbers 21675 and above, there is no longer a need to discharge any high voltage. To replace the battery packs, first turn the gauge off. Then, remove the display panel assembly by loosening the four captive screws on the display panel. Third, disconnect the ribbon cable from the baseboard. The ribbon cable connection is a locking release. To disconnect, pull the levers on either side of the ribbon down. After disconnecting the ribbon cable, remove the four screws that attach the top shell to the gauge base and suspend the top shell from the handle. After noting the polarity of the connections, disconnect the battery packs (located on either side of the post). The connectors are white plugs with red and black wires. Remove the battery pack screws closest to the baseboard (right) for each battery pack. To remove each battery pack, slide the pack forward and lift up. Replace each battery pack by sliding a new pack beneath the left screw and washer. Replace the right screw for each battery pack. With the same polarity as noted previously, connect the battery packs to the baseboard. Replace the top shell. Be careful not to over-tighten the screws. Also, be sure to tighten them evenly to prevent warping of the top shell. Reconnect the ribbon cable to the baseboard assembly and replace the front panel assembly. Model 3440 Appendix F–3 F. PERIODIC MAINTENANCE For Serial Numbers Below 21675 MECHANICAL MAINTENANCE SOURCES The 3440 gauge contains two radioactive sources. The americium241:beryllium source in the center of the gauge base is small and shielding is not required. Never attempt to remove this source. The Cs-137 source is welded inside the tip of the source rod. In the SAFE or shielded position, the Cs-137 source is surrounded by tungsten, reducing the radiation to safe levels. INSPECTING SOURCE ROD To ensure the integrity of the source rod, Troxler recommends that a qualified Troxler service person inspect the gauge and the source rod at least every five years. This inspection includes checking for wear, corrosion, or damage that could affect the safety of gauge operation. CLEANING THE TUNGSTEN SLIDING BLOCK If the sliding block is not kept clean, it may stick partially or completely open when the source rod is raised to the SAFE (shielded) position. This will result in high radiation levels near or in line with the source rod opening on the bottom of the gauge. After cleaning and re-assembling the gauge, check the operation of the sliding block by pushing the source rod into the backscatter position and then raising it back to the shielded position. You should hear a click as the sliding block snaps shut. Inspect the opening on the base of the gauge to confirm that the sliding block is completely closed. If not, check that the spring was properly reinstalled after cleaning. If the sliding block still does not close completely, contact your nearest Troxler Service Center immediately. WARNING Do not store or transport the gauge unless the sliding block is completely closed. Increased radiation levels may violate transportation regulations, and may cause excessive personnel radiation exposure. Appendix F–4 Maintenance of the scraper ring or tungsten sliding block may be required if the following problems are encountered while using the gauge: ♦ Difficulty in lowering or raising the source rod (may indicate a need to replace the scraper ring). ♦ No “click” is heard from the tungsten sliding block striking the back of the shield cavity when the source rod is raised to the “safe” position. ♦ Erratic or incorrect density standard counts. WARNING Routine maintenance to clean the cavity beneath the source requires removal of the bottom sliding block. ALWAYS wear safety glasses and a dosimeter. Procedure – Scraper Ring/Tungsten Sliding Block Maintenance 1. With the source rod in the SAFE position, place gauge on its side. 2. Clean the heads of the four screws to make removal easier and to prevent the screw heads from stripping. 3. Using a screwdriver, remove the bottom plate. Replace if excessive wear is evident on the back of the bottom plate. 4. If a large amount of dirt has accumulated on the backside of the bottom plate, the scraper ring has become excessively worn and should be replaced. Replace the scraper ring by first removing the retaining ring with a small screwdriver or scribe (Figure F–1). Carefully remove the scraper ring. Clean the entire plate of any debris. Install the new scraper ring and replace the retaining ring. Figure F–1. Retaining and Scraper Rings Model 3440 Appendix F–5 F. PERIODIC MAINTENANCE SCRAPER RING/TUNGSTEN SLIDING BLOCK MAINTENANCE 5. Remove the sliding tungsten shield. To reduce exposure when removing the shield, Troxler recommends that the operator stand to the side of the gauge. With a steel brush or rag, remove as much dirt as possible from the cavity. 6. Clean the sliding tungsten block. Replace the sliding block with the angled side up. If installed incorrectly, the source rod will not extend into the measurement position. 7. Apply a light coating of Magnalube-G® paste to the top angled surface of the block. Do not grease any other surfaces on the sliding block or in the sliding block area. Remove any excess lubricant. 8. Replace the bottom plate. Do not over-tighten screws in the aluminum base. Ensure that the source rod moves up and down freely. Procedure – Source Rod Bearing Maintenance If the source rod does not slide up and down freely, the source rod bearing may require lubrication and cleaning. Lubricate the source rod bearing with Magnalube-G paste. NOTE Troxler Electronic Laboratories, Inc. assumes no responsibility for bearing life if lubricants other than Magnalube-G are used. 1. Carefully remove the gauge top shell by first removing the four 7/64inch head screws. Tie the top shell to the source/index rod. 2. To catch any grease, place a rag or cloth under the vent valve located below the grease fitting on the source rod tower. 3. Using a standard 16-oz. grease gun with a Magnalube-G cartridge, apply five shots of lubricant or until all dirty grease is ejected from the vent valve and clean grease is visible. NOTE The vent valve was not installed on earlier gauges. Instead, an Allen-head screw was located below the grease fitting. On these gauges, remove the Allen-head screw before applying lubricant to allow the old grease to be ejected. Failure to remove the Allen-head screw may result in severe mechanical damage to the base assembly. 4. Remove any grease that was ejected. Reseat the top shell taking care to properly seat the post gasket and replace the screws. Do not overtighten the screws. Appendix F–6 Four gaskets seal the gauge from the environment. If the gauge has been out in extremely wet or humid weather, or was used in a cold environment and stored in a heated space, condensation may form on the inside of the gauge. Normally, this is not a problem and the condensation will dissipate overnight. However, after using the gauge in any of the above conditions, it is recommended that the display panel be removed to allow air to circulate inside the gauge. Placing the gauge in a humidity-controlled or airconditioned room and loosening the display panel screws will allow the gauge to “breathe”, reducing any moisture accumulation. A hair dryer (on low heat) may also be used to dry out the inside of the gauge. CAUTION High moisture levels inside the gauge may result in damage to the electronics, as well as erratic or fluctuating moisture readings. NOTE All gaskets may be replaced without removing the sources from the gauge. To replace the control panel gasket, loosen the four screws and remove the control panel. Use care in handling the panel, trying not to touch the back of the circuit board. Unplug the connector from the keypad/display and remove the panel completely. Remove the old gasket and install a new gasket. Reassemble the connectors and panel. To replace the two post gaskets and the top shell gasket, remove the four screws that hold the top shell to the base. Lift up the base, remove the gasket in question, and replace with a new gasket. A light coating of talcum powder on the inside of the post gasket will aid in re-assembly. Model 3440 Appendix F–7 F. PERIODIC MAINTENANCE GASKETS 3440 GAUGE TOP SHELL Currently, the 3440 gauge top shell is manufactured from an engineering thermoplastic designed specifically to provide high impact strength and to offer excellent compatibility with many industrial solvents and petrochemicals. The top shell may be cleaned with mild (low alkaline) soap and water. Other approved cleaning substances include: methyl, isopropyl, or isobutyl alcohols. A cloth dampened with kerosene or diesel fuel may be used to remove heavy soils. Avoid prolonged exposure and do not soak. CAUTION The use of any unapproved cleaning agents such as methyl-ethyl-ketones, amines, and methylene chloride will damage the top shell and void the warranty. NOTE In the past, the top shell was not compatible with petrochemicals. Your top shell is compatible with petrochemicals only if, upon removal of the keypad, there is a ridge around the exposed opening. If not, use only a mild soap (such as 409™ or Fantastic™) or alcohols (such as methyl or isopropyl) to clean the top shell. Appendix F–8 CAUTION To avoid electrical shock and/or damage to the 3440 gauge components, turn off the gauge before servicing any electronic component. NOTE Electronic maintenance is to be done by Troxler Qualified Service personnel only. See page F–27 for information on returning a gauge to Troxler. 3440 BASEBOARD ASSEMBLY Figure F–2 shows the 3440 baseboard assembly with any parts requiring general electronic maintenance labeled for the operator’s reference. Figure F–2. 3440 Baseboard Assembly Model 3440 Appendix F–9 F. PERIODIC MAINTENANCE ELECTRONIC MAINTENANCE TROUBLESHOOTING GAUGE FAILS STANDARD COUNTS Ensure that the source rod opening on the gauge bottom is completely closed or covered by the tungsten sliding block. If any opening is visible, the sliding block should be cleaned. If the sliding block still does not close completely, contact the nearest Troxler Service Center. NO DENSITY READING The most likely reason for no density readings is an electronic problem, such as a failure of the detector preamplifier. However, as a precaution, ensure that the tip of the source rod is intact and undamaged (that is, the source is not missing). This can be done by checking the radiation levels at the gauge base surface with a radiation survey meter (without extending the source rod). A maximum reading of 10-20 mrem/hr is normal, and indicates the source is present. However, if the maximum reading is less than 1 mrem/hr or if a survey meter is not available, perform a visual inspection of the source rod tip as follows to confirm its integrity: 1. Extend the source rod just far enough to see the source rod tip. The tip should appear flat to slightly rounded and smooth. 2. Stay at least 3 ft (1 m) away from the tip of the unshielded source rod and complete the inspection as quickly as possible to minimize exposure (the dose rate at 3 ft from the unshielded source is about 2.7 mrem/hr). If the visual inspection indicates the source rod tip is broken off (source is missing): 1. Immediately contact your Radiation Safety Officer (RSO). 2. Initiate a search for the source starting at the location where the gauge was last used. 3. Report lost or missing radioactive sources to your state or federal radiation control agency in accordance with applicable regulatory requirements. 4. Contact the Troxler Radiation Safety Department for further advice. Appendix F–10 If taking readings in the backscatter position, ensure the source rod is properly locked in the notch and not resting on the test material. Check the inside of the gauge for moisture. To dry the gauge interior, remove the keypad. If necessary, use a hair dryer (on low heat) to circulate warm air for one to three hours. Remove any foreign objects from inside the gauge. Ensure the hardware mounting screws are tight and in place. Check count time – A 4-minute count will give the best precision with a repeatability of ± 1 pcf. Erratic density readings may be caused by a dirty sliding block. Clean the sliding block. Perform a statistical stability (stat) test. If test passes, proceed with job. If test fails, repeat two more times. If test fails two out of three times, contact Troxler Service. GAUGE MEASUREMENTS ARE UNSTABLE; TUBES TEST OKAY Possible High Voltage Module or baseboard malfunction. GAUGE TURNS OFF AFTER IT IS TURNED ON, OR WILL NOT TURN ON The gauge automatically turns off after five hours if no keys are pressed. Try to turn the gauge on again. Gauge may be wet. If the gauge is wet, dry the gauge interior with hair dryer (on low heat) for three hours. Do not turn gauge on until moisture is removed from the gauge interior or serious damage of electronic components may result! Batteries are below 6.0 V dc. Recharge or replace batteries. Model 3440 Appendix F–11 F. PERIODIC MAINTENANCE GAUGE READINGS APPEAR ERRATIC NON-FUNCTIONING DEPTH INDICATOR Possible CPU Board malfunction or defective indicating strip. Use manual override or recalibrate depth indicator. GAUGE INDICATES SHORT BATTERY LIFE AFTER RECHARGING NiCad batteries may be charged up to 100 full charge/discharge cycles. Batteries may be reaching end of life cycle-replace. Charger/adapter may not be supplying the full voltage to the batteries; check the ac outlet and the dc output (12 V dc). Check that you are using the correct charger. The ac charger may be defective. To check, use the dc charger to charge the batteries. Remove any loose screws or foreign objects from the gauge interior that may cause an electrical short to ground. CPU or baseboard malfunction. BATTERY LIFE HOURS ARE INCORRECT Charge the gauge overnight (16 hours). Check battery voltage. If battery voltage is equal or greater than 7.0 volts, manually enter the correct hour value. With a new gauge after a full 16-hour charge, enter a battery life of 200. With age the battery life will decrease, so enter 200 hours and monitor the gauge until <Batteries low!> is displayed. After the next recharge, enter the new battery life (200 - battery low life). The operator may leave the gauge in voltage mode for future reference. If battery voltage is less than 7.0 volts, change the mode to the elapsed time since last charge. Monitor gauge until <Batteries low!> is displayed. If the time is too low (less than 16 hours), replace the batteries. Appendix F–12 Check for proper baud rate. Check printer. Check cable and connectors. Check CPU jumper configuration and cable pins 2 and 3. Also, possible CPU Board malfunction. COUNTS ARE SATISFACTORY, BUT RESULTS ARE IN ERROR Ensure the measurement depth corresponds to the actual source rod depth. Check the calibration constants. Check to see if an offset (density, moisture, trench, or special) is enabled. Ensure the standard counts are correct. GARBAGE OR “+++++” IS DISPLAYED Check the standard counts in memory. If the standard counts are suspect, perform new standard counts. If counts equal zero for both systems, replace high voltage board (contact the Service Department). Check gauge for water damage. If the gauge is wet, dry the gauge interior with hair dryer (on low heat) for 3 hours. Check the calibration constants. They should match the constants on your calibration data sheet if your calibration sheet is in metric units. Refer to the Calibration Constants section on page 7–6 to determine if your calibration sheet is metric and, if not, for instructions on converting B and F values from metric to English units. If necessary, perform a statistical stability test, record the results and call Customer Service. Model 3440 Appendix F–13 F. PERIODIC MAINTENANCE RS-232 PORT DOES NOT FUNCTION CORRECTLY POSSIBLE MALFUNCTION INDICATORS CPU Board Display malfunctions No keypad response RAM test fails Batteries do not recharge Battery Low indicator does not function correctly Keyboard test fails with front panel disconnected from CPU Board Gauge does not access plug-in option module or does not function properly with module plugged in (if option module is functioning correctly) Preamp Board No moisture or density counts Batteries do not recharge Gauge fails tube test Fuse blows repeatedly Gauge fails stability or drift test High Voltage Module No moisture or density counts Moisture or density counts are unstable Batteries discharge prematurely Fuse blows repeatedly Gauge fails stability or drift test Appendix F–14 Display test fails Gauge does not turn Off Fuse blows repeatedly Beeper stops (or is erratic) Gauge does not turn On when charger is connected. Auto depth indicator, battery low, and gauge automatic shutdown do not function properly. When first turned on, the 3440 gauge enters a Self-Test phase to check the internal systems for proper operation. If a fault is encountered during this internal systems check, an error message is displayed. The following error messages may be encountered in normal day-to-day operation. These errors can usually be corrected if the operator takes proper action. Error Message Corrective Action Error #6 – Data Previously stored. Error #7 – No serial device connected! Take another reading before storing. Check cables, connectors & printer. Error #8 – Illegal baud rate! Re-enter another baud rate. (No Error #) Time to reach precision too high! Re-enter lower precision value. (No Error #) Invalid Depth! Change rod depth. The following errors are not operator serviceable. Contact your Troxler Service Department for more information. Error #1 - Data NVRAM test error! Error #2 - Key Pad test error! Error #3 - GM Tube test error! Error #4 - Helium Tube test error! Error #5 - Display test error! Model 3440 Appendix F–15 F. PERIODIC MAINTENANCE ERROR MESSAGES REPLACEMENT PARTS Figure F–3. 3440 Gauge Assembly (Sheet 1 of 2) Appendix F–16 F. PERIODIC MAINTENANCE Figure F–3. 3440 Gauge Assembly (Sheet 2 of 2) Model 3440 Appendix F–17 Table F–1. 3440 Gauge Assembly Parts Match the reference number (Ref #) with the correct part in Figure F–3 on the two previous pages. Ref # Part Number Description Qty 1 000212.4201 Screw, 4-40 × 3/4 S.H.C. SS 4 3 000408.1431 Screw, #8 × 1/2 PH Phil SS Type B 3 4 104149 3440 Battery Assembly 2 5 100528.2000 Screw, Captive 2 6 108330 3440/3430 Preamp Assembly 1 7 102070 3400 Heat Shield 2 8 102071 End Protector 2 9 102074 3400 Filter 1 11 107413 Lead Disk 1 12 106368 Gasket, Top Cover 1 14 107415 Radiation Label 1 16 000025.1026 Bushing, #10 x 0.250D x 0.187L Nylon 2 17 000001.0601 Washer, #6 Internal Lock SS 4 18 000610.1430 Screw, 10 x 5/8 PH Phil Type E 2 20 100989 Gasket - Molded for Base Post 1 21 106068 Topshell, 3440/3440 1 22 108329 3440 Nameplate 1 23 104122 Radiation Label 1 24 000900.0621 Screw, #4 (.114 Dia.) x 3/16" Drive 8 26 000310.4891 Screw, 6-32 x 5/8 SHCS w/ Insert 4 27 101603.1010 Roll Pin, 3/32 Dia. x 5/8 L 1 29 108335 3440 Front Panel Assembly 1 30 104130 Plunger Block 1 31 104106.1200 Depth Indicating Strip, 12-inch 1 104106.0800 Depth Indicating Strip, 8-inch 1 Appendix F–18 Ref # Part Number Description Qty Match 3440 Base Mechanics to Part Numbers Shown 32 105341.0801 3440 Base Mechanical Assembly 8-1 (For 104125.0801) 1 105341.0802 3440 Base Mechanical Assembly 8-2 (For 104125.0802) 1 105341.1201 3440 Base Mechanical Assembly 12-1 (For 104125.1201) 1 105341.1202 3440 Base Mechanical Assembly 12-2 (For 104125.1202) 1 105341.1235 3440 Base Mechanical Assembly 12-3-5 (For 104125.1235) 1 34 108334 Assembly, Preamp Cover Box 1 35 000206.1400 Screw, 4-40 x 3/8 PHMS Phil 4 36 000001.0401 Washer, #4 Internal Lock SS 4 37 106340 Gasket, Depth Strip Hole 1 38 102662 Scaler Label (3401-3411) 1 39 000001.0801 Washer, #8 Internal Lock SS 3 40 103484 Reference Standard Assembly (3400) 1 41 106367 Gasket, Tower Flat 1 42 107414 Source Spacer 1 43 108346 Gasket Bracket, Front 1 44 108347 Label, Caution 1 Model 3440 Appendix F–19 F. PERIODIC MAINTENANCE Table F–1. 3440 Gauge Assembly Parts (Continued) Figure F–4. 3440 Base Mechanical Assembly 5 1 3 2 4 6 Figure F–5. 3440 Source Rod Handle Assembly Appendix F–20 7 8 Match the reference number (Ref #) with the correct part in Figure F–4 on the opposite page. Ref # Part Number Description Qty 1 105256 3400 Base Machining 1 2 100996 3400 Sliding Block 1 3 102351 3400 Round Bioshield 1 4 105264 3400 Bottom Plate Assembly 1 5 102069.1000 2400/3400 Wiper Cap 1 6 104594 Source Rod Bearing 2 7 107412 Source Cap 1 8 102399 3400 Shield Spring LC-055H-6 1 9 000608.1101 Screw, 10-32 x 1/2 FHMS Phil SS 4 10 012752 Seal, Wiper (V.S. S-35-14) 1 11 012759 Seal, Oil ID 5/8" #6204 2 13 013200 Fitting, 3/16 Grease w/ Serrated Shank 1 14 001006.4090 Screw, 5/16-18 x 3/8 Set Knurl 3 15 101603.0510 3/32 Dia. x 5/16 L. Roll Pin 16 012789 Magnalube-G Lubricant 14.5 oz 17 105239 Baseplate Gasket 18 018127.0001 Loctite, 242 19 105309 Spring Guide 2 .01 1 .001 1 Table F–3. 3440 Source Rod Handle Assembly Parts Match the reference number (Ref #) with the correct part in Figure F–5 on the opposite page. Ref # 1 Part Number Description Qty 104556 Gauge Handle 3440 1 2 104553 3400 Plunger 1 3 105108 Trigger 1 4 012200 Spring, Lee #LC-0242G-7 1 5 101604.1610 Roll Pin, 1/8 Dia. x 1" 1 6 012779.3000 Plug, Heyco Cap. #2643, Black 1 7 000824.4800 Screw, 1/4 - 20 x 1-1/2 Soc. HD Cap 8 018127.0001 Loctite 242 Model 3440 1 .005 Appendix F–21 F. PERIODIC MAINTENANCE Table F–2. 3440 Base Mechanical Assembly Parts Figure F–6. 3440 Preamplifier Assembly Appendix F–22 Match the reference number (Ref #) with the correct part in Figure F–6 on the opposite page. Ref # Part Number Description Qty 1 107903 Assembly, Preamp Baseboard 2 107164 Assembly, High Voltage Board 1 3 000923.2442 Spacer, 4-40 x .75 M/F 4 4 018127.0001 Loctite 242 5 108328 Preamp Mechanical Assembly 1 6 104136 Interface PC Board Bracket 1 7 000206.1400 Screw, 4-40 x 3/8 PHMS Phil 2 8 104098 G-M Tube Interface PCB Assembly 1 1 .001 9 100156 G-M Tube 2 10 000001.1001 Washer, #10 Internal Lock SS 1 11 000204.1400 Screw, 4-40 x 1/4 PHMS Phil SS 8 12 000001.0401 Washer, #4 Internal Lock SS 10 13 104094.0001 900 V dc Moisture Tube Assembly 1 14 000001.0601 Washer, #6 Internal Lock SS 2 15 000306.1401 Screw, 6-32 x 3/8 PHMS Phil SS 2 16 107854 G-M Tube Holder 1 17 103540 Modified Tube Clamp 1 18 000604.1400 Screw, 10-32 x 1/4 PHMS Phil SZ 1 19 102119 7M Tube Pad 1 Model 3440 Appendix F–23 F. PERIODIC MAINTENANCE Table F–4. 3440 Preamplifier Assembly Parts Figure F–7. 3440 Front Panel Assembly Appendix F–24 Match the reference number (Ref #) with the correct part in Figure F–7 on the opposite page. Ref # 1 Part Number Description Qty 108331 Ground Plate, CPU PC Board 1 2 104184.1000 Cover Plate Assembly 1 3 000204.1400 Screw, 4-40 × 1/4 PHMS Phil SS 3 4 107947 EMI Gasket 3440 1 5 000001.0401 Washer, #4 Internal Lock SS 7 6 104115 3440 Overlay 1 7 104134 3440 Front Panel 1 8 106369 Front Panel Gasket 1 9 107893 3440 CPU PCB Assembly 1 10 000923.0842 Spacer, 4-40 × 1/4 Lg. M/F Hex 5 11 102888 Cable, 3400-B Series 1 12 108314.0001 Ferrite Split Half 2 13 000202.1400 Screw, 4-40 × 1/8 PHMS Phil SZP 1 14 106856 Clamp, Grounding 1 15 018128.0001 Loctite 290 Wick & Seal 16 000203.1401 Screw, 4-40 x 3/16 PHMS Phil SS 5 17 000011.0401 Washer, #4 Flat SS 5/16 OD 1 18 108314.1001 End Cap for Ferrite Split Half 2 Model 3440 .001 Appendix F–25 F. PERIODIC MAINTENANCE Table F–5. 3440 Front Panel Assembly Parts NOTE The following parts are not shown in the previous figures. ACCESSORIES Part Number 110403 104156 100421 102111 103484 103680.1000 106875.0002 106875.0004 106875.0008 106875.0010 106875.0012 103623.3440 104279 104337 Description AC Adapter, 12 V CE DC Charger Drill Rod Scraper Plate (Drill Rod Guide) Standard Assy (3400) (Reference Standard Block) Extraction Tool Concrete Adapter 3400 2” w/Acc. Concrete Adapter 3400 4” w/Acc. Concrete Adapter 3400 8” w/Acc. Concrete Adapter 3400 10” w/Acc. Concrete Adapter 3400 12” w/Acc. 3440 Shipping Case/Carton Assy 3440 Instruction Manual 3440 Reference Card OPTIONAL ACCESSORIES Part Number 104148 102868 102876.0005 102873 109661 021140 104122 104661.1000 104324.1000 104340 104341 104334.1000 104719 Description Accessory Battery Case Assembly Leak Test Kit w/4 packets Leak Test Packet (4 units) 1 oz. solution detergent Survey Meter Radiation Sign Kit Radiation Label Print Pkg. 3440 Cable, (Weigh-Tronix®) Printer-to-Gauge (M/M) 3440 Printer 3440 Printer Battery Pack Cable, PC-to-Gauge (FEM/MAL) Interfacing PGM Disk/Manual MAINTENANCE SUPPLIES Part Number 012786 012789 100761 Appendix F–26 Description Lubricant Magnalube-G Paste 1 lb. Can Lubricant Magnalube-G 14.5 oz. Source Rod Pig All shipments within the U.S. to the factory must comply with 49 CFR (see Appendix E). Items returned for service must be accompanied by a Returned Good Authorization (RGA) number, and a description of the instrument and its problem. This information is used by Troxler shipping and service personnel to expedite the repair work. To obtain an RGA number, please call or fax the factory with your request. Please have the following information available when calling or faxing Troxler for an RGA number: ♦ System model and serial number. ♦ Part number/serial number (if applicable). ♦ Is system still under warranty? ♦ Problem or difficulty you are having with the instrument. ♦ Shipment method to Troxler and for return shipment. ♦ Shipping and billing address (not P.O. Box) - street address and zip. ♦ Phone number/contact (for questions from Troxler). ♦ Will estimate be required prior to performing any work on the system? ♦ Payment method: credit card, account number, or purchase order number. All government agencies (city, county, state, and federal) must send purchase order numbers. NOTE To prevent order duplication, if an order has been placed by telephone, please write “Confirming Order” on any follow-up written requests. NOTE Returning a 3440 gauge requires special handling and shipping procedures. Follow the instructions in Appendix E. Please contact a Troxler Sales Support or Service Representative with any questions. Model 3440 Appendix F–27 F. PERIODIC MAINTENANCE RETURNING A GAUGE FOR SERVICE NOTES Appendix F–28 APPENDIX G 3440 SPECIFICATIONS This appendix contains measurement and gauge specifications for the 3440 gauge. Measurement Specifications ...................................................................G–2 U.S. Customary Units ......................................................................G–2 SI Units ............................................................................................G–2 Calibration Specifications .......................................................................G–3 Radiological Specifications.....................................................................G–3 Memory Specifications ...........................................................................G–3 Electrical Specifications..........................................................................G–4 Mechanical Specifications.......................................................................G–5 Model 3440 Appendix G–1 G. 3440 SPECIFICATIONS CONTENTS MEASUREMENT SPECIFICATIONS U.S. CUSTOMARY UNITS Direct Transmission, 6 inches Precision at 125 pcf Composition error at 125 pcf Surface error (0.05 in., 100% void) 15 sec. 1 min. 4 min. ±0.42 pcf ±1.25 pcf –1.06 pcf ±0.21 pcf ±1.25 pcf –1.06 pcf ±0.11 pcf ±1.25 pcf –1.06 pcf ±1.00 pcf ±2.50 pcf –4.68 pcf ±0.50 pcf ±2.50 pcf –4.68 pcf ±0.25 pcf ±2.50 pcf –4.68 pcf ±0.64 pcf –1.12 pcf ±0.32 pcf –1.12 pcf ±0.16 pcf –1.12 pcf 1 min. 4 min. Backscatter Density, (98%) 4 inches Precision at 125 pcf Composition error at 125 Surface error (0.05 in., 100% void) Moisture Precision at 15 pcf Surface error (0.05 in., 100% void) Depth of measurement at 15 pcf – 8.5 inches SI UNITS Direct Transmission, (150 mm) Precision at 2000 kg/m3 Composition error at 2000 kg/m3 Surface error (1.25 mm, 100% void) 15 sec. ±6.8 kg/m3 ±20.0 kg/m3 –17.0 kg/m3 ±3.4 kg/m3 ±20.0 kg/m3 –17.0 kg/m3 ±1.7 kg/m3 ±20.0 kg/m3 –17.0 kg/m3 ±16.0 kg/m3 ±40.0 kg/m3 –75.0 kg/m3 ±8.0 kg/m3 ±40.0 kg/m3 –75.0 kg/m3 ±4.0 kg/m3 ±40.0 kg/m3 –75.0 kg/m3 ±10.0 kg/m3 –18.0 kg/m3 ±5.0 kg/m3 –18.0 kg/m3 ±2.5 kg/m3 –18.0 kg/m3 Backscatter Density, (98%) 100mm Precision at 2000 kg/m3 Composition error at 2000 kg/m3 Surface error (1.25 mm, 100% void) Moisture Precision at 240 kg/m3 Surface error (1.25 mm, 100% void) Depth of measurement at 240 kg/m3 – 215 mm Appendix G–2 CALIBRATION SPECIFICATIONS Uncertainty of density standards: Uncertainty of moisture standards: Calibration standards range Density: Moisture: Gauge operational range Density: Moisture: ±0.2% ±2.0% 109 – 165 pcf (1750 – 2640 kg/m3) 0 – 40 pcf (0 – 640 kg/m3) 70 – 170 pcf (1100 – 2700kg/m3) 0 – 40 pcf (0 – 640 kg/m3) RADIOLOGICAL SPECIFICATIONS Gamma source Neutron source Source housing Shielding Surface dose rates Source rod material Shipping case Certificates of Competent Authority Registry of Radioactive Sealed Sources and Devices 0.3 ± 10% GBq (8 ± 10% mCi) Cs-137 1.48 GBq 10% (40 ± 10% mCi) Am-241:Be Stainless steel, encapsulation Tungsten, lead, and cadmium See 3440 Radiation Profile on page C–4 Stainless steel DOT 7A, Type A, Yellow II label, TI = 0.6 CS-137 – USA/0614/S or USA/0356/S Am-241:Be – USA/0632/s or CZ/1009/S NC-646-D-130-S MEMORY SPECIFICATIONS RAM ROM Model 3440 64K Resident, Non-volatile Stores over 450 complete station records 192K External 256K or 512K Resident, 512K External Appendix G–3 G. 3440 SPECIFICATIONS Method: Computer reduction of count rate data based on U.S. National Bureau of Standards Electron Cross Sections, Neutron Cross Sections, and Absorption Coefficients. Data is reduced to the form D = (1/B)1n(A/(CR + C)) for density and M = (CR – E)/F for moisture where A, B, C, E, and F are constants and CR is count ratio. The algorithm corrects for hydrogen photon-scattering coefficients and provides means for offsetting non-water hydrogen. Direct calibration entry by keypad. ELECTRICAL SPECIFICATIONS Time accuracy and stability Power supply stability Stored power Battery recharge time Charge source Liquid crystal display (LCD) readout Power consumption (watts) average Power consumption after automatic battery cutoff (watts) Serial data format Parity ±0.005% ±0.0002%/°C ±0.01%°C 30 watt-hours 16 hours, automatic cutoff 12-14 V dc, 500 mA minimum 4 x 16 alphanumeric 0.16 0.00 8 Data Bits, 2 Stop Bits, No Gauge to PC-compatible Computer Cable (Null Modem Cable): Gauge Computer (Female) (Male) Gnd (pin 1)........................Gnd (pin 1) Tx (pin 2) ..........................Rx (pin 3) Rx (pin 3)..........................Tx (pin 2) DSR (pin 6).......................DTR (pin 20) DTR (pin 20) ....................DSR (pin 6) Gnd (pin 7)........................Gnd (pin 7) Gauge to WEIGH-TRONIX® Printer Cable: Gauge Printer (Male) (Male) Gnd (pin 1)........................Chassis Gnd (pin 1) Tx (pin 2) ..........................RD (pin 2) DSR (pin 6).......................CTS (pin 5) Gnd (pin 7)........................Gnd (pin 7) Gauge returns to Ready (power saving mode) after 2 minutes of operator inactivity, except in STD, Stat Test, Drift Test, and Nomograph (30 minutes). After 5 hours of complete inactivity, gauge performs a total power shutdown. Battery packs are fully protected against overcharge and over-discharge; charge life is updated every 1.9 seconds and is indicated on the READY display. Emergency Use – Capable of operation with six “D” size alkaline batteries. Not all “D” size alkaline batteries will fit. Appendix G–4 MECHANICAL SPECIFICATIONS Vibration test Drop test 0.1 in. (2.54 mm) at 12.5 Hz 300 mm on 25 mm diameter steel ball Operating temperature – Ambient: – Surface: Storage temperature 14 to 158°F (–10 to 70°C) 350°F (175°C) –70 to 185°F (–57 to 85°C) Gauge size (excluding handles) Shipping case dimensions Weight Shipping weight Model 3440 Appendix G–5 G. 3440 SPECIFICATIONS Gauge height (including handles) 14.8 x 9.1 x 7.2 in. (376 x 231 x 183 mm) 12-in. rod – 23.25 in. (591 mm) or 8-in. rod – 19.25 in. (489 mm) 29.35 x 13.88 x 16.85 in. (74.55 x 35.26 x 42.80 cm) 29 lb (13.1 kg) 90 lb (40.8 kg) in shipping case NOTES Appendix G–6 APPENDIX H LOG FORMS & COMPACTION TEST DATA CONTENTS Standard Count Log ................................................................................H–2 Nuclear Compaction Test Data ...............................................................H–4 H. LOG FORMS & TEST DATA Model 3440 Appendix H–1 STANDARD COUNT LOG Serial Number Date Moisture - MS Density - DS Appendix H–2 Date Moisture - MS Density - DS STANDARD COUNT LOG Serial Number Date Moisture - MS Density - DS Date Moisture - MS Density - DS H. LOG FORMS & TEST DATA Model 3440 Appendix H–3 NUCLEAR COMPACTION TEST DATA PROJECT __________________ JOB NUMBER _______________ DATE ______________________ TAKEN BY __________________ TEST NUMBER STATION OFFSET ELEVATION MODE AND DEPTH DENSITY COUNT WET DENSITY MOISTURE COUNT MOISTURE DRY DENSITY % MOISTURE STANDARD DENSITY OPTIMUM MOISTURE % COMPACTION MOISTURE CORRECTION 1 2 3 4 5 6 7 8 9 10 TEST NUMBER STATION OFFSET ELEVATION MODE AND DEPTH DENSITY COUNT WET DENSITY MOISTURE COUNT MOISTURE DRY DENSITY % MOISTURE STANDARD DENSITY OPTIMUM MOISTURE % COMPACTION MOISTURE CORRECTION 11 12 13 14 15 16 17 18 19 20 REMARKS DENSITY Appendix H–4 MOISTURE APPENDIX I UNIT CONVERSION The 3440 gauge can display measurement results in either SI (metric) units or English units. Also, HM-181 of 49 CFR changes the standard units of radioactivity in the United States from the English unit of curies (Ci) to the SI unit of becquerel (Bq). This requires the shipper to convert the activity on the Bill of Lading from curies to becquerels. Until everyone is accustomed to the SI units, it is permitted to follow the SI units with the English units in parentheses to clarify the description, such as 1.49 GBq (40 mCi). To help our operators convert from English units to SI units, the table in this appendix provides SI conversion factors for common English units relevant to the 3440 gauge. CONTENTS Measurement Units .................................................................................. I–2 Radiological Units.................................................................................... I–2 I. UNIT CONVERSION Model 3440 Appendix I–1 MEASUREMENT UNITS 1 in. = 25.4 mm 1 in. = 2.54 cm 1 ft = 30.48 cm 1 ft = 0.3 m 1 pcf = 16.02 kg/m3 1 pcf = 1.6 x 10–2 g/cm3 RADIOLOGICAL UNITS 1 rem = 0.01 Sv 1 Ci = 37 GBq 1 mCi = 37 MBq 1 µCi = 37 kBq The following table is provided to assist you in converting from millicuries (mCi) to gigabequerels (GBq): mCi to GBq 8.0..................................... 0.30 40...................................... 1.48 Appendix I–2 APPENDIX J J. PRINTING DATA PRINTING (UPLOADING) PROJECT DATA Current versions of Microsoft® Windows® include a HyperTerminal communications program that can be used to transfer project data from the Model 3440 Surface Moisture-Density Gauge to a computer. This appendix provides instructions on using the HyperTerminal program to print (upload) data from the gauge. CONTENTS Windows Hyperterminal .......................................................................... J–2 HyperTerminal Setup........................................................................ J–2 Equipment Connection...................................................................... J–3 Printing (Uploading) Project Data .................................................... J–3 Viewing Project Data ............................................................................... J–4 Model 3440 Appendix J–1 WINDOWS HYPERTERMINAL To print (upload) data from the gauge to a computer equipped with the Windows operating system, use the HyperTerminal program as described in the following sections. HYPERTERMINAL SETUP To configure the Windows HyperTerminal program and set up a dedicated icon to transfer data from the gauge, perform the following steps: 1. Click on the Start button, then select the Programs menu and the Accessories folder. f In Windows 95, select the HyperTerminal icon. f In Windows 98 and later versions, select the Communications subfolder, then select the HyperTerminal icon. 2. From the HyperTerminal window, double-click on the Hypertrm icon. The Connection Description window is displayed. This window is used to set up a new connection. 3. In the Name box, enter a name (for example, Troxler Data) for the new connection. Select an icon to use for the connection, then click OK. The Connect To window is displayed. 4. In the Connect To window, do not enter a telephone number. Instead, select Direct to Com1, Direct to Com2, Direct to Com3, or Direct to Com4, depending upon the computer. The COM# Properties window is displayed, where # is the number of the selected COM port. 5. Configure the Port Settings as follows: Bits per second Select same value as gauge baud rate (see page 6–25) 8 Data bits None Parity 1 Stop bits Hardware Flow control 6. Click OK. The Troxler Data – HyperTerminal window is displayed, where Troxler Data is the icon name entered in step 3. 7. From the toolbar, click on File and select Save from the dropdown menu. 8. Close the Troxler Data – HyperTerminal window by selecting Exit from the File menu. Appendix J–2 EQUIPMENT CONNECTION PRINTING (UPLOADING) PROJECT DATA To print (upload) project data from the gauge to the computer: 1. On the computer, click on the Start button, then select the Programs menu and the Accessories folder. f In Windows 95, select the HyperTerminal icon. f In Windows 98 and later versions, select the Communications subfolder, then select the HyperTerminal icon. 2. From the HyperTerminal window, double-click on the Troxler Data – HyperTerminal icon (where Troxler Data is the icon name entered in the HyperTerminal Setup section on the previous page). The Troxler Data – HyperTerminal window is displayed. 3. From the toolbar, click on Transfer, then select Capture Text from the dropdown menu. The Capture Text window is displayed, showing the default folder and filename to be used to store the data. Troxler recommends using the default settings when first using the program. 4. On the 3440 gauge, select the project data to be printed (downloaded) to the computer, and follow the instructions in Chapter 5 to print the data to the computer. The project data will be displayed in the Troxler Data – HyperTerminal window as it is downloaded. 5. When the data transmission is complete, select Transfer from the toolbar. Then select Capture Text and Stop from the dropdown submenu. Model 3440 Appendix J–3 J. PRINTING DATA Connect the gauge to the computer using the appropriate serial cable (see the list of Optional Accessories on page C–26). Connect the gauge’s serial port (see Figure 5–2 on page 5–18) to the computer COM port selected in step 4 of the HyperTerminal Setup section on the previous page. VIEWING PROJECT DATA After project data has been printed (uploaded) to the computer, it can be viewed using either a word processing program (such as Microsoft Word®) or a spreadsheet program (such as Microsoft Excel®). To view the data, open the desired program, then open the file in which the project data was stored. NOTE The user can also use the spreadsheet program to design a spreadsheet to manipulate and manage the data. Appendix J–4 INDEX % % Marshall ................................................................................................2–7 B Backscatter ......................................................................................4–3, A–2 Base course preparation .......................................................................... 3–2 Baseboard assembly ................................................................................ F–9 Basic parameter setup ............................................................................. 2–4 Battery Alkaline............................................................................................... F–2 Charger................................................................................................ 1–4 Monitor function ............................................................................... 6–27 Recharging .......................................................................................... F–2 Baud Rate function................................................................................ 6–25 Becquerel ................................................................................................B–3 Beta particles...........................................................................................C–2 Block, reference standard.............................................................. 1–4, 2–13 Bottom density ...................................................................................... 6–15 Burn-In function.................................................................................... 7–10 Model 3440 Index–1 INDEX A Absorbers, neutron .................................................................................. 5–5 Access code...................................................................................... xiii, 7–1 Accessories.................................................................................... 1–4, F–26 Accident notification............................................................................... E–3 Addition (calculator function)............................................................... 5–24 Agreement state....................................................................................... 1–3 ALARA concept......................................................................................C–3 Alkaline batteries .................................................................................... F–2 Alpha particles ........................................................................................C–2 Asphalt Measurement ....................................................................................... 4–3 Mode ........................................................................................... 2–6, 4–3 Surface preparation ............................................................................. 3–4 ASTM standards...................................................................................... 1–2 Atom Nucleus................................................................................................B–2 Structure ..............................................................................................B–2 Automatic shutdown .......................................................................2–2, G–4 Auto-Station function.................................................................... 4–4, 6–11 C Calculator Function .............................................................................................5–24 Keys .................................................................................................... 1–6 Calibration Constants function............................................................................... 7–6 Date function....................................................................................... 7–5 Depth..................................................................................................5–23 Special................................................................................................6–12 Specifications......................................................................................G–3 Standard function ...............................................................................7–11 Canadian Nuclear Safety Commission (CNSC)...................................... E–4 Canadian shipping requirements ............................................................. E–4 Case, transport......................................................................................... 1–4 Cautions and warnings ............................................................................... iv Charger, battery....................................................................................... 1–4 Clock/calendar .......................................................................................2–11 Communication Protocol function .........................................................6–26 Control panel........................................................................................... 1–6 Core density ...........................................................................................6–19 Count Standard .............................................................................................2–13 Time .................................................................................................... 2–5 Counts function......................................................................................5–21 Curie........................................................................................................B–3 Customer Name function ........................................................................ 7–3 D Daily inspection ...................................................................................... 2–2 Declaration of Conformity .......................................................................... v Density Bottom ...............................................................................................6–15 Core....................................................................................................6–19 Measurement.......................................................................................A–2 Measurements ..................................................................................... 4–1 Offset .................................................................................................. 5–4 Department of Transportation (DOT) ..................................................... E–1 Depth Calibration..........................................................................................5–23 Function .............................................................................................5–22 Mode ..................................................................................................5–22 Of Measurement..................................................................................A–5 Strip....................................................................................................5–22 Direct transmission .................................................................................A–2 Disposal ..................................................................................................D–4 Index–2 Distance, limiting radiation exposure......................................................C–3 Division (calculator function) ............................................................... 5–25 Dosimeter ................................................................................................C–4 Downloading data ................................................................................. 5–18 Drift test .................................................................................................. 6–7 Drill rod........................................................................................... 1–4, 3–2 Extraction tool............................................................................. 1–4, 3–2 Guide................................................................................................... 1–4 F Federal regulations .................................................................................. 1–3 G Gamma radiation............................................................................C–2, A–2 Gaskets .................................................................................................... F–7 Gauge Operation............................................................................................. 5–1 Setup ................................................................................................... 2–4 H Half-life...................................................................................................B–3 Hazmat training....................................................................................... E–3 He-3 detectors ........................................................................................ A–5 HyperTerminal program...........................................................................J–2 Model 3440 Index–3 INDEX E Electrical specifications ......................................................................... G–4 Electron ...................................................................................................B–2 Electronic maintenance ........................................................................... F–9 Emergency procedures ............................................................................C–6 Encapsulation, source..............................................................................C–6 Erase Function ............................................................................................ 5–21 Project ............................................................................................... 2–10 Standard Counts function .................................................................... 7–9 Erratic readings ....................................................................................... 6–3 Error messages ...................................................................................... F–15 Extended functions.................................................................................. 7–1 Extraction tool................................................................................. 1–4, 3–2 I Incidents..................................................................................................D–3 Index rod ................................................................................................2–13 Inspection Daily.................................................................................................... 2–2 Source rod ........................................................................................... F–4 International Air Transport Association (IATA)..................................... E–1 International Atomic Energy Agency (IAEA) ........................................ E–1 K Keypad .................................................................................................... 1–6 Data function......................................................................................6–10 L Leak test..................................................................................................D–3 Licensing requirements ...........................................................................D–1 M Maintenance............................................................................................ F–1 Electronic ............................................................................................ F–9 Gaskets................................................................................................ F–7 Mechanical.......................................................................................... F–4 Scraper ring......................................................................................... F–5 Source rod bearing .............................................................................. F–6 Supplies............................................................................................. F–26 Tungsten sliding block ........................................................................ F–5 Marshall .......................................................................................... 2–7, 5–13 Measurement Asphalt ................................................................................................ 4–3 Density ........................................................................................4–1, A–2 Depth of ..............................................................................................A–5 Moisture ......................................................................................4–1, A–5 Soil ...................................................................................................... 4–2 Specifications......................................................................................G–2 Storing................................................................................................. 4–4 Unit conversion .................................................................................... I–2 Units.................................................................................................... 2–4 Mechanical Maintenance........................................................................................ F–4 Specifications......................................................................................G–5 Memory Specifications......................................................................................G–3 Memory Clear function..........................................................................7–10 Metric units ............................................................................................. 2–4 Index–4 Microsoft Excel.....................................................................................................J–4 Windows ..............................................................................................J–1 Word.....................................................................................................J–4 Mode Asphalt ........................................................................................ 2–6, 4–3 Depth................................................................................................. 5–22 Function .............................................................................................. 2–6 Ready .................................................................................................. 2–2 Soil .............................................................................................. 2–6, 4–2 Moisture Measurement ...................................................................................... A–5 Measurements ..................................................................................... 4–1 Offset................................................................................................... 5–5 Multiplication (calculator function) ...................................................... 5–25 O OFF key................................................................................................... 1–6 Offset Density ................................................................................................ 5–4 Function .............................................................................................. 5–3 Moisture .............................................................................................. 5–5 Side wall............................................................................................ 5–10 Trench ............................................................................................... 5–10 Wet density ......................................................................................... 5–4 ON key .................................................................................................... 1–6 Operation Gauge .................................................................................................. 5–1 Theory ................................................................................................ A–1 Optional accessories.............................................................................. F–26 Model 3440 Index–5 INDEX N Neutron........................................................................................... C–2, B–2 Cross-sections .................................................................................... A–5 Thermalization ................................................................................... A–5 Neutron absorbers ................................................................................... 5–5 NO/CE key.............................................................................................. 1–6 Nomograph function ............................................................................. 6–15 Notice to Employees .............................................................................. D–3 Nuclear compaction test data ................................................................. H–4 P Parts Replacement...................................................................................... F–16 Personnel monitoring ..............................................................................D–2 Photons........................................................................................... C–2, A–2 Poisson distribution.................................................................................B–4 Precision function ..................................................................................6–22 Print Function .............................................................................................5–18 Print function .........................................................................................5–18 Proctor values.........................................................................................5–13 Project Erase...................................................................................................2–10 Function .............................................................................................. 2–8 View.................................................................................................... 2–9 Proton......................................................................................................B–2 Q Quality factor ..........................................................................................B–5 R Rad ..........................................................................................................B–5 Radiation Exposure .............................................................................................C–3 Gamma....................................................................................... C–2, A–2 Monitoring ..........................................................................................C–4 Profile..................................................................................................C–4 Safety ..................................................................................................C–1 Safety program....................................................................................D–2 Statistics ..............................................................................................B–4 Terminology........................................................................................B–5 Theory........................................................................................ B–3, B–1 Types...................................................................................................C–2 Radiological Specifications......................................................................................G–3 Unit conversion .................................................................................... I–2 Ready mode ............................................................................................ 2–2 Recall Function .............................................................................................5–22 Record keeping .......................................................................................D–4 Recover Erase function ........................................................................... 6–9 Reference standard block ...............................................................1–4, 2–13 Regulations Canadian shipping............................................................................... E–4 U.S. shipping....................................................................................... E–2 Index–6 Rem .........................................................................................................B–5 Remote Control function....................................................................... 7–12 Replacement Gaskets ................................................................................................ F–7 Parts................................................................................................... F–16 Returned Good Authorization (RGA) ................................................... F–27 Rod Length function ............................................................................... 7–8 RS-232 serial port ................................................................................. 5–18 Model 3440 Index–7 INDEX S SAFE position ....................................................................................... 2–13 Scraper plate/drill rod guide............................................................ 1–4, 3–2 Scraper ring maintenance ........................................................................ F–5 Security .................................................................................................. D–2 Serial Number function ........................................................................... 7–4 Serial port.............................................................................................. 5–18 Service Centers ................................................................................................... iii Returning a gauge ............................................................................. F–27 Set Units function.................................................................................... 2–4 Shielding .................................................................................................C–3 SHIFT key............................................................................................... 1–6 Shipper's Declaration for Dangerous Goods ........................................... E–2 Shipping Canadian requirements........................................................................ E–4 U.S. requirements................................................................................ E–2 Shutdown, automatic.......................................................................2–2, G–4 Side wall offset...................................................................................... 5–10 Site preparation ....................................................................................... 3–1 Sliding block Cleaning .............................................................................................. F–4 Maintenance ........................................................................................ F–5 Soil Measurement ....................................................................................... 4–2 Mode ........................................................................................... 2–6, 4–2 Site preparation ................................................................................... 3–2 Source Decay function .................................................................................. 6–30 Encapsulation ......................................................................................C–6 Rod.................................................................................................... 2–13 Rod bearing maintenance .................................................................... F–6 Rod inspection..................................................................................... F–4 Rod length ........................................................................................... 7–8 Rod positions..................................................................................... 2–14 Special Calibration function ...........................................................................6–12 Functions............................................................................................. 6–1 Roadway function ..............................................................................6–31 Specifications..........................................................................................G–1 Calibration...........................................................................................G–3 Electrical .............................................................................................G–4 Measurement.......................................................................................G–2 Mechanical..........................................................................................G–5 Memory...............................................................................................G–3 Radiological ........................................................................................G–3 Standard Count log.............................................................................................H–2 Standard count .......................................................................................2–13 START/ENTER key ............................................................................... 1–6 Stat test.................................................................................................... 6–3 Station number ........................................................................................ 4–4 Status function .......................................................................................5–16 Store function.......................................................................................... 4–4 STORE key ............................................................................................. 4–4 Subtraction (calculator function)............................................................5–25 T Test Drift..................................................................................................... 6–7 Leak ....................................................................................................D–3 Stat ...................................................................................................... 6–3 Theory, radiation............................................................................ B–3, B–1 Thermalization ........................................................................................A–5 Thermoluminescent dosimeter (TLD).....................................................C–4 Time Count................................................................................................... 2–5 Limiting radiation exposure ................................................................C–3 Time/Date function ................................................................................2–11 Training Hazmat ................................................................................................ E–3 Troxler course .....................................................................................D–3 Transport case ......................................................................................... 1–4 Transport Packaging of Radioactive Materials Regulations (TPRM)..... E–4 Transportation Canadian requirements........................................................................ E–4 U.S. requirements................................................................................ E–2 Transportation and shipping.................................................................... E–1 Transportation of Dangerous Goods Act and Regulations (TDG) .......... E–4 Trench offset ..........................................................................................5–10 Troubleshooting .................................................................................... F–10 Index–8 U U.S. NRC ................................................................................................ 1–3 U.S. shipping requirements ..................................................................... E–2 Unit conversion ........................................................................................ I–1 Units Measurements ..................................................................................... 2–4 Metric .................................................................................................. 2–4 V View project............................................................................................ 2–9 Voidless density value................................................................. 5–13, 5–15 Voids Asphalt ................................................................................................ 3–4 Soil or base.......................................................................................... 3–2 Y YES key .................................................................................................. 1–6 Model 3440 Index–9 INDEX W Warnings and cautions ............................................................................... iv Wet density offset ................................................................................... 5–4 NOTES Index–10 NOTES WARRANTY Model 3440 Warranty–1 TROXLER ELECTRONIC LABORATORIES, INC. LIMITED WARRANTY TROXLER ELECTRONIC LABORATORIES, INC., and subsidiary, TROXLER INTERNATIONAL, LTD., hereinafter referred to as “TROXLER” warrants this instrument, model 3440 serial number __________, against defects in material and workmanship for a period of 18 mos. from date of shipment. For gauges sold through authorized TROXLER representatives, the date of shipment will be as of the transfer from representative to purchaser. During the applicable warranty period, TROXLER’s obligation under this warranty shall be limited exclusively to the repair, F.O.B. TROXLER’s plant of any instrument which may prove defective under normal use and which TROXLER’s examination shall disclose to its satisfaction to be thus defective, normal use being defined for the purpose of this warranty as operation under normal load, usage, and conditions with proper care and maintenance and competent supervision. In no event shall TROXLER be held liable for damages, delays, or losses consequential, incidental, or otherwise attributable to the failure of this instrument or radioactive material contained therein. TROXLER’s liability being specifically limited to repair as stated hereinabove. This warranty is automatically initiated except where modified by contractual or other written and signed agreement. THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF, AND THIS WARRANTY IS EXPRESSLY IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, AND TROXLER NEITHER ASSUMES, NOR AUTHORIZES ANYONE TO ASSUME FOR IT ANY OTHER LIABILITY IN CONNECTION WITH THE SALE OF THE INSTRUMENT. THIS WARRANTY SHALL NOT APPLY TO THE INSTRUMENT OR ANY PART WHICH HAS BEEN SUBJECTED TO DAMAGE BY ACCIDENT, NEGLIGENCE, ALTERATION, ABUSE, MISUSE, OR SERVICE NOT AUTHORIZED IN WRITING BY TROXLER. SUCH DAMAGE TO INCLUDE BUT NOT BE LIMITED TO BURNING OF CIRCUIT BOARDS AND HARNESS FROM IMPROPER SOLDERING TECHNIQUES AND DAMAGE TO THE INSTRUMENT DUE TO PURCHASER’S FAILURE TO PERFORM MAINTENANCE AS OUTLINED IN THE AUTHORIZED OPERATOR’S MANUAL. DUE TO THE NATURE OF THEIR USE, MECHANICAL ACCESSORY PARTS AND BATTERIES ARE WARRANTED FOR 90 DAYS FROM SHIPMENT DATE. TROXLER ELECTRONIC LABORATORIES, INC. Troxler International, Ltd. Troxler Electronics (Canada), Ltd. Troxler Electronics GmbH 3008 Cornwallis Road Post Office Box 12057 Research Triangle Park, NC 27709 USA NOTICE TO CONSUMERS Any disclaimer or limitation on the remedies expressed above shall not be effective to the extent prohibited by state or federal law. NOTE: THIS WARRANTY EXCLUDES DAMAGE INCURRED IN SHIPMENT. IF THIS CASE IS RECEIVED IN DAMAGED CONDITION, THE CARRIER SHOULD BE CONTACTED IMMEDIATELY. ALL CLAIMS FOR DAMAGE IN TRANSIT SHOULD BE FILED WITH THE CARRIER. IF REQUESTED, TROXLER WILL AID IN FILING OF CLAIMS AND/OR LOCATING GAUGES LOST IN TRANSIT.