Convergys® 100 - Convergent Technologies

Transcription

Convergys® 100 - Convergent Technologies
Convergys® 100
Biochemistry Analyser
•
•
•
•
•
•
•
•
Based on advanced electronic optical technology.
Hermetic sold-state optical system for stable performance.
Uses hard membrane filters for longer shelf life.
20 Key soft-touch keypad. Optional Menu and
step-by-step instructions.
Inbuilt 58mm high speed, low-noise thermal printer.
Optional external printer.
96 programmable parameters.
Test Methods: end-point (absorbance, linear concentration,
nonlinear concentration with bichromatic test) fixed time,
kinteic, and coagulation time tests.
10mm, 25µl cuvette.
www.convergent-technologies.de
Convergys® 100 – Technical Features
Convergys® 100 is a Biochemistry and Coagulation analyser that
is based on advanced electronic optical technology.
It has ergonomic design and state-of-the-art features meeting the
requirements of a modern laboratory and it‘s staff.
Convergys® 100 has a reliable hardware and user-friendly software. Convergys® 100 is a result of many years of research and
incorporates the suggestions of many renowned experts in the
field of Biochemistry.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Designed for routine clinical bio-chemistry and coagulation time
tests.
Hermetic solid-state optical system for stable performance.
Uses hard membrane filters for longer shelf life.
Light splitting after cuvette, separate signal amplifier for each
wavelength. Filter selection is completed electronically without
mechanical movement ensuring fast measurement for precise
results, especially useful for bi-chromatic measurements
Dedicated 20 keys soft touch keyboard. Optional Menu and stepby-step instructions. Check on the instrument status at any time with self-test function.
240 × 128 matrix LCD backlit character display. Simultaneous display of ASC code and curve. Readable even in very
dim light.
Real time clock
Inbuilt 58mm high speed, low-noise thermal printer. Optional external printer
Test Methods: end-point (absorbance, linear concentration, nonlinear concentration with bi-chromatic test) fixed time,
kinetic, and coagulation time tests
Light source: 12V 20W Tungsten halogen lamp with lifetime ≥2000 hours.
Optical system: 6 hard membrane filters – 340, 405, 505, 546, 578, and 630 nm, wavelength precision ±2nm
Abs. range: 0-2.5000Abs, resolution : 0.0001 Abs, drift ≤0.003Abs/hr after 10 minutes warm up.
Precision: Abs. = CV <0.1% at 0.5000 Abs, <1% at 2.0000 Abs.
96 programmable parameters
Temperature : Peltier control 25, 30, 32, 37 ± 0.1˚C
Asp. calibrator 05 T.BILI 37.0
Cuvette: Micro-flow cell 10mm, 25µl
ABS
Power supply: 110/220V ±10%, 50/60Hz, <80VA
03
37.0
Standard RS232 serial port
+
Dimension: (L x W x H): 420 x 320 x 170mm
+
+
Weight 10kg gross, 9 kg net
+
+
Dealer Stamp:
ENTER
Exit
Press U/D key select
Convergent Technologies GmbH & Co. KG
Ginsterweg 30 • 35066 Frankenberg • Germany
Tel. +49 6451 230487 • Fax +49 6451 230569
www.convergent-technologies.de