Convergys® 100 - Convergent Technologies
Transcription
Convergys® 100 - Convergent Technologies
Convergys® 100 Biochemistry Analyser • • • • • • • • Based on advanced electronic optical technology. Hermetic sold-state optical system for stable performance. Uses hard membrane filters for longer shelf life. 20 Key soft-touch keypad. Optional Menu and step-by-step instructions. Inbuilt 58mm high speed, low-noise thermal printer. Optional external printer. 96 programmable parameters. Test Methods: end-point (absorbance, linear concentration, nonlinear concentration with bichromatic test) fixed time, kinteic, and coagulation time tests. 10mm, 25µl cuvette. www.convergent-technologies.de Convergys® 100 – Technical Features Convergys® 100 is a Biochemistry and Coagulation analyser that is based on advanced electronic optical technology. It has ergonomic design and state-of-the-art features meeting the requirements of a modern laboratory and it‘s staff. Convergys® 100 has a reliable hardware and user-friendly software. Convergys® 100 is a result of many years of research and incorporates the suggestions of many renowned experts in the field of Biochemistry. • • • • • • • • • • • • • • • • • • • • Designed for routine clinical bio-chemistry and coagulation time tests. Hermetic solid-state optical system for stable performance. Uses hard membrane filters for longer shelf life. Light splitting after cuvette, separate signal amplifier for each wavelength. Filter selection is completed electronically without mechanical movement ensuring fast measurement for precise results, especially useful for bi-chromatic measurements Dedicated 20 keys soft touch keyboard. Optional Menu and stepby-step instructions. Check on the instrument status at any time with self-test function. 240 × 128 matrix LCD backlit character display. Simultaneous display of ASC code and curve. Readable even in very dim light. Real time clock Inbuilt 58mm high speed, low-noise thermal printer. Optional external printer Test Methods: end-point (absorbance, linear concentration, nonlinear concentration with bi-chromatic test) fixed time, kinetic, and coagulation time tests Light source: 12V 20W Tungsten halogen lamp with lifetime ≥2000 hours. Optical system: 6 hard membrane filters – 340, 405, 505, 546, 578, and 630 nm, wavelength precision ±2nm Abs. range: 0-2.5000Abs, resolution : 0.0001 Abs, drift ≤0.003Abs/hr after 10 minutes warm up. Precision: Abs. = CV <0.1% at 0.5000 Abs, <1% at 2.0000 Abs. 96 programmable parameters Temperature : Peltier control 25, 30, 32, 37 ± 0.1˚C Asp. calibrator 05 T.BILI 37.0 Cuvette: Micro-flow cell 10mm, 25µl ABS Power supply: 110/220V ±10%, 50/60Hz, <80VA 03 37.0 Standard RS232 serial port + Dimension: (L x W x H): 420 x 320 x 170mm + + Weight 10kg gross, 9 kg net + + Dealer Stamp: ENTER Exit Press U/D key select Convergent Technologies GmbH & Co. KG Ginsterweg 30 • 35066 Frankenberg • Germany Tel. +49 6451 230487 • Fax +49 6451 230569 www.convergent-technologies.de