Tests in magnetic field of conventional Hamamatsu R4998
Transcription
Tests in magnetic field of conventional Hamamatsu R4998
MICE Collaboration MICE-NOTE-DET-201 9 February 2008 Tests in magnetic field of conventional Hamamatsu R4998 PMTs R.Bertoni a, M.Bonesini a, Y.Kharadzov a,1, M.Rayner a,2 and S. Terzo a a Sezione INFN Milano Bicocca, Dipartimento di Fisica G. Occhialini, Piazza Scienza 3, Milano, Italy This paper reports the measurements done at Milano Bicocca to study the behaviour of conventional Hamamatsu R4998 PMTs with local massive iron box shieldings inside magnetic fields up to 600 G. A dedicated solenoid was built for this purpose and extensive measurements done in 2007. As a conclusion, a local PMT shielding with 1 mm µ-metal + a 6 × 6 cm2 transverse area ARMCO pure iron box shielding was found sufficient to shield TOF2 PMTs and gives a valuable alternative to more complicate and expensive global shieldings. 1 2 permanent address Dept. of Physics, University of Sofia,Bulgaria permanent address Dept. of Physics, University of Oxford, UK Contents 1 Introduction 1 2 The chosen Hamamatsu R4998 PMT 1 3 PMT’s shielding: local versus global solution 2 3.1 Global shielding solution 4 3.2 Local shielding solution a la D0 4 4 Tests on Hamamatsu R4998 PMTs 5 4.1 PMTs test setup 5 4.2 Tests results for Hamamatsu R4998 PMTs. 8 5 Conclusions 12 References 17 2 1 Introduction In the MICE experiment, precision timing measurements are required to relate the time of the incoming beam muons to the phase of the accelerating field in each RF cavity and simultaneously for particle identification (PID) by a time-of-flight (TOF) method. Three time-of-flight detectors (TOF0, TOF1, TOF2) are foreseen [1], [2]. The last two (TOF1/TOF2) are at the entrance and the exit of the MICE cooling channel; the first one (TOF0) instead is placed ∼ 10 m upstream of its entrance. The first TOF0 station will work in the fringe field of the Q6 quadrupole magnet. Estimations of the B field from [3] give a value well below 50 Gauss. The other two TOF stations (TOF1/TOF2) will work instead in the stray fields of the measuring solenoids with a magnetic field up to 0.09-0.1 T. 2 The chosen Hamamatsu R4998 PMT Due to the low residual magnetic field, in the fringe field of the Q6 quadrupole where the TOF0 detector will be placed, conventional PMTs with an elongated µ-metal shielding (extending 30 mm beyond the photocathode surface) may be used. The same apply for PMTs for TOF1/TOF2 provided an additional local or a global shielding is built. To obtain a good timing resolution, PMTs with a small transit time spread (TTS) must be used. An additional requirement is a good rate capability, up to 0.5 MHz for each counter. This has led, as a natural choice, to 1” R4998 PMTs from Hamamatsu Photonics 1 . Table 1 displays their main characteristics. R4998 PMTs have been delivered by Hamamatsu in assemblies (H6533MOD) that include the PMT tube, the voltage divider chain and a 1 mm µ−metal shielding (see figure 1 for details). To increase the countrate stability of PMTs, instead of a conventional resistive divider type, active dividers or a booster on the last dynodes were requested. In this way the theoretical rate capability is increased by a factor ∼ 8 (from ∼ 260 KHz to ≥ 1 MHz). After some tests, the performances of PMTs equipped with a booster or an active divider were found equivalent and thus the active divider option was chosen for its easier use. The distribution of gains at B=0 Gauss at a nominal -2250 V H.V., as deduced from Hamamatsu data sheets, is shown in figure 2 for the sample of R4998 PMTs to be used on TOF0. In the same figure the distribution of the anode dark current (in nA) is also shown. Variation in gain up to a factor 5 are evident, even after the PMTs selection. This points to an accurate matching of PMTs for the left-right side of each TOF scintillator bar. 1 A lower cost alternative based on 1” PET PMTs (R9800 from Hamamatsu), with a TTS ∼ 250 ps, a risetime ∼ 1 ns and a nominal gain G ∼ 1 × 10 6 , was available only after this choice was made 1 R4998 Structure Linear Focussed Tube diameter 1” Active area diameter 20 mm No. stages 10 Q.E. at peak .20 Gain (B=0 T) typ. 5.7 × 106 Risetime (ns) 0.7 Transit time (ns) 10 TTS (ns) 0.16 max hIa i (mA) Table 1 Main properties of Hamamatsu R4998 PMTs. 0.16 Fig. 1. H6533MOD assemblies, with active divider, as delivered by Hamamatsu. 3 PMT’s shielding: local versus global solution TOF1 and TOF2 will work inside the high residual magnetic field not fully shielded by the 10 cm iron shield (“the Virostek plate”) either side of the measuring solenoids. Figure 3 from [4] shows the longitudinal Bk and orthogonal B⊥ components of the magnetic field at the position of TOF2, computed with a 2D Tosca [5] calculation (z = 664 cm). 3D Tosca calculations were redone in reference [6] and results were found compatible. Orthogonal components (up to 0.1 T) and longitudinal components (up to 0.025 T) of the fringe magnetic fields have to be shielded. Local or global shielding may be 2 Events ID Entries Mean RMS 7 6 100 40 3.875 1.950 5 4 3 2 1 Events 0 0 2.5 5 7.5 10 12.5 15 17.5 20 22.5 -6 25 G(B=0 T) x 10 ID Entries Mean RMS 10 101 40 7.975 11.02 8 6 4 2 0 0 10 20 30 40 50 60 70 80 90 100 Idark (nA) Fig. 2. Absolute gain G distribution at -2250 V (top) and anode dark current in nA (bottom) for the sample of R4998 PMTs, to be used in the experiment on TOF0. devised. Fig. 3. Magnetic field components, after the “Virostek” plate at the position of TOF2 [4]. For conventional PMTs the most difficult component to be shielded is the one along the PMT’s axis. Orthogonal components are more easily shielded. This report addresses mainly the first item. 3 3.1 Global shielding solution In the global shielding solution, a magnetic cage bolted to the “Virostek plate” fully embraces the TOF1 or TOF2 detector reducing the residual B field to a maneageable value of the order of a few Gauss. The solution, albeit elegant, is quite expensive and puts severe problems for access to the TOF detector PMTs inside. Figure 4 shows it for the TOFI detector. The extraction mechanics of TOFI is quite complicate, as 16 15 14 13 12 L 11 10 ITEM QTY PART NUMBER 1 1 Hindge Setup 2 1 Top Bracket & Support 3 2 Side Restraint Bracket 4 1 Bottom Support Bracket 5 1 Bottom Split Ring 6 1 Side Split Ring 7 1 Top Split Ring 8 1 Front Cage Shield 9 1 Virostek Shield 10 2 Stopper 11 2 Guide rail 12 2 Rail Spacer 13 2 Locking Pin 14 2 Stopper 1 spacer 15 2 Slide in Stopper 1 16 1 TOF CORE K Parts List DRAWING NUMBER MC-TOFS-SA01 MC-TOFS-SA02 MC-TOFS-SA03 MC-TOFS-SA04 MC-TOFS-SA05 MC-TOFS-SA06 MC-TOFS-SA07 MC-TOFS-SA08 MC-TOFS-SA09 MC-TOFS-SA10 MC-TOFS-SA11 MC-TOFS-SA12 MC-TOFS-SA13 MC-TOFS-SA14 MC-TOFS-SA15 Supplyed by MICE 9 8 7 6 5 4 3 2 1 L DESCRIPTION K J J I I H H 2 A ( 1:6 ) G G 10 14 15 8 7 3 11 F F 13 E E 6 A D D 1 C C 9 16 5 12 4 B B SCALE UNIVERSITY OF OXFORD DEPARTMENT OF PHYSICS KEBLE ROAD OXFORD, OX1 3RH TEL. +44 (0)1865 273333 FAX. +44 (0)1865 273475 DRAWN BY 1:6 J.Tacon CREATION DATE: 22/08/2007 LATEST DATE: 28/08/2007 PROJECT/ORIGINATOR X.X 0.05 0.005 SOURCE FILE : CHECKED 0.1 X.XX X.XXX TITLE TOF Shielding MICE GENERAL TOLERANCES UNLESS STATED MATERIAL COMPONENT WEIGHT FINISH NUMBER OFF 1 JOB NO. / EST. TIME USED ON : TOF Shielding RM.iam DRAWN ACCORDING TO BS308 A 3rd ANGLE PROJECTION ALL DIMENSIONS ARE IN MM UNLESS OTHERWISE STATED 16 15 14 13 12 11 10 9 8 7 6 5 4 SIZE A0 THIS DRAWING MAY NOT BE USED FOR COMMERCIAL USE 3 2 DRAWING NUMBER A MC-TOFS-MA SHEET 1 OF 1 1 Fig. 4. Global shielding for TOFI, as designed in [7]. it includes brackets to allow access to the internal part of the cage. 3.2 Local shielding solution a la D0 In media, the magnetic field is defined by H=B- 4 π M where B is the magnetic flux density and M the moment density of the media. The magnetic permeability µ is defined by B= µ H for non-ferromagnetic materials or ferromagnets at low fields. For ferromagnetic materials in large fields B = F(H) and depends on the media history (hysteresis), but at least initially one may assume B ' µ H. For the ideal case of a spherical shell of permeability µ the field inside the shield is reduced by a factor µ−1 . Magnetic shielding materials are chosen for their characteristics in respect to permeability and saturation. As permeability increase in magnetic shielding materials, their saturation level decrease. Therefore, the highest permeability alloys (such as µ−metal) have the lowest saturation values. As a saturated shield is a poor attenuator, for a given field a proper material must be chosen. Table 2 reports some commercially available materials with their permeabilities. Up to fields ∼ 50−100 G µ-metal shieldings are enough. For higher field values, such a shield saturate and becomes ineffective. It must be surrounded by a supplementary shield, usually of 4 soft iron. The usual choice is a low carbon content iron (less than 0.01%, such as ARMCO), that has relatively high permeability (compared to other steel) and excellent saturation characteristics. As shielding is a mass effect, box-shaped soft iron shieldings are more effective than cylindrical ones. This idea pioneered in the D0 experiment [8] has been tested in the case of MICE using different geometrical configuration for the iron shielding boxes and different iron materials: Fe360, ARMCO, ... The problem is usually the longitudinal component Bk of the magnetic field, while the orthogonal component B⊥ may be more easily shielded. As an example, from reference [9] a simple mild-steel shield of inner diameter 32 mm and thickness 4 mm (rin /rext = .80) in a field B⊥ = 1000 G gives an attenuation ∼ 100. An additional inner µ−metal shield with an inner diameter 29 mm and 1-mm thickness (rin /rext = .936) gives an additional attenuation of about 500, leaving a residual magnetic field of ≤ 1 Gauss. This shows that the soft iron box configurations studied are more than enough for the shielding of B⊥ fields up to 1000 G. The only remaining problem is the magnetic shielding of Bk fields. material % of C µmax r Fe360 iron 0.25% 5000 ARMCO iron ≤ 0.01% 180000 21500 800000 8000 Super-permalloy saturation (G) µ−metal (80 % Nickel) 100000 Table 2 Properties of some commercially available shielding materials. 4 6500 Tests on Hamamatsu R4998 PMTs 4.1 PMTs test setup Systematic studies have been done, using a built on purpose solenoid of 23 cm inner diameter, 40 cm length shown in figure 5 2 . The magnetic field is created by five windings in series, giving five solenoids in series on the same spool and crossed by the same current I. Being a resistive magnet a special care was put into the thermal resistance of the assembly (up to 100 C 0 ), using special insulating paints. For part of the tests a Digimess 3040 laboratory power supply (0-32 V, 0-40 A) was used. At higher fields an Eutron power supply (0-32 V, 0-100 A) on loan was used. The main limit was the heating of the windings that limited the maximum circulating current to about 55 A, due to the increase of conductor resistance giving higher voltage 2 built by TBM srl, Uboldo (VA), Italy 5 Fig. 5. The used resistive solenoid test magnet at INFN Milano Bicocca. A PMT under test is shown inside. drops (up to the maximum allowed value of ∼ 30V ). The big open bore allows tests of the H6553 assembly both with field lines orthogonal or parallel to the PMT axis. The magnetic field was measured via a gaussmeter 3 , with a better than 1% accuracy. The B field calibration curve is shown in figure 6 at the center of the solenoid and at different z positions shifted along its longitudinal axis. From simulations using a finite length solenoid approximation and performed measurements, a field uniformity better than 3% may be assumed at the center of the test solenoid in a volume of about 5 × 5 × 10 cm3 . A fast light pulse 4 was sent to the PMT’s photocathode via a 3 m long multimode 3M TECS FT-110-LMT optical fiber (with a measured dispersion of ≤ 15 ps/m, see [10]). At the end of the fiber a small Plexiglas prism, inserted in a black plastic cover in front of the PMT window, allowed illumination at the center of the photocathode. The laser spot was focused into the optical fiber (aligned by a micrometric x-y-z flexure system 5 by a 10x Newport microscope objective, after removable absorptive neutral density filters, to give light signals of different intensities. A broadband beamsplitter (BS) divided the laser beam to give 50% of light on the fiber injection system and 50% on a monitoring detector. A fast Thorlabs DET210 photodiode (risetime ∼ 1 ns) was used in most measurements, to monitor the laser intensity. Tests were done usually with a signal corresponding to about 150-300 photoelectrons: a typical value for a minimum ionizing particle (MIP) crossing a scintillator 1”-2” thick. The optical power was periodically monitored with an OPHIR NOVA laser powermeter. The number of photoelectrons (Npe ) was estimated via absolute gain measurement. This number was cross-checked with the powermeter measurements. 3 Hirst GM04 model, with axial Hall probe a home-made system based on a Nichia NDHV310APC violet laser diode and an AvtechPulse fast pulser (type AVO-9A-C laser diode driver, with ∼ 200 ps risetime and AVX-S1 output module) was used. This system gave laser pulses at ∼ 409 nm, with a FWHM between ∼ 120 ps and ∼ 3 ns (as measured with a 6 GHz 6604B Tek scope) and a max repetition rate of 1 MHz 5 Thorlabs MBT613/M with 4 mm excursion and a resolution of ∼ 0.5 µm 4 6 Fig. 6. Upper panel: (left) calibration of the longitudinal field at the geometrical center of the test solenoid, as function of the applied current; (right) variation of the calibration slope as function of the axial displacement respect to the solenoid center. Lower panel: difference of the magnitude of the B field as respect to z = 0 as a function of the circulating current. For gain measurements the PMT signal was acquired in average mode by a Tektronix TDS 754C digital scope (500 MHz bandwidth, 2 Gs/s sampling rate) triggered by the laser output sync., that had a maximum jitter of 15 ps as respect to the delivered optical pulse. In part of the measurements the signal was sent after a passive 50% T divider to a Canberra 2005 preamplifier, followed by an EG-G Ortec 570 shaper (shaping time ∼ 1 µs, gain ∼ 200) followed by a Silena 8950 MCA analyzer, using as external trigger the sync out signal of the laser. For timing measurements, the same MCA chain was used with a Silena 7422 QVT (see figure 7 for details). The STOP signal (tST OP ) was given by the PMT anode signal after a leading edge PLS 707 discriminator, while the START signal (tST ART ) was given by the sync out of the pulser after a suitable delay and an ORTEC pulse inverter. In timing measurements what is actually measured is the time difference ∆t = tST ART − tST OP , that accounts for delay in cables and electronics and jitter in the transit time in the tested PMTs. A lack of variation in this quantity or no deterioration in the FWHM of its distribution, after increasing the magnetic field intensity, demonstrates the effectiveness of the adopted shielding. The used TDC range (up to 0.1 µs) with the MCA resolution (2K) allowed a resolution of 50 ps/count. Taking into account the TTS of the used PMTs (160 ps at B=0 G), their transit time (∼ 10 ns) and risetime (tR ∼ 700 ps), this resolution is well matched to possible incoming relevant effects. 7 Laser driver xyz flexure Sync out optical fiber Laser Head BS Delay Filters Magnet Powermeter/ Photodetector B Inverter SHAPER ext Vin trigger MCA PVC CAP HV PMT In Out QVT Splitter OUT TEK 754C READOUT Fig. 7. Layout of the test setup for PMTs measurements (not in scale). In some measurements the readout section (MCA) was replaced by a TEK 754C scope. 4.2 Tests results for Hamamatsu R4998 PMTs. 4.2.1 Absolute gain measurements − 9210 Pulser Lecroy variable attenuator Test in Preamp Camberra 2005 OUT IN ORTEC 570 V in MCA Fig. 8. Layout of the test setup for absolute Gain measurement. The absolute gain G and PMT linearity as a function of HV supply was measured for some R4998 PMTs at B=0 Gauss, by using the test setup outlined in figure 8. A continuous train of pulses was delivered by a Lecroy 9210 pulser, through an attenuator, to the test input of a Canberra 2005 preamplifier. The signal was shaped to 2µs by an Ortec 570 shaping unit and then fed into a Silena MCA. In this way the MCA scale was calibrated in pC/channel. Then the preamp was connected to the PMT output, with an illumination corresponding to a SER peak condition. From the MCA peak position, it was thus possible to determine the output charge Q in pC as a function of HV and thus the absolute gain G, shown for a typical R4998 PMT in figure 9. 8 G (x 10|6) 1 inch conv PMT/active divider - gain 8 7 6 5 4 3 2 1 2000 2100 2200 2300 2400 2500 HV (V) Fig. 9. Absolute gain G for a typical R4998 PMT, with active divider. 4.2.2 Behavior in magnetic field The PMTs were inserted in the central region of the test solenoid, where the field had a uniformity better than ∼ 3%, using a support to incline them to 00 or 900 as respect to the field lines in the magnet (Bk or B⊥ ). Environment light was accurately masked, to reduce noise. Measures were done to see gain reduction and possible deterioration in timing resolution as a function of the magnetic field (B) and the relative orientation angle (θ), between the PMT axis and the magnetic field B. The tested shielded solutions were the mu-metal shielding only option (as given by the H6553MOD standard assembly) and various options with shielding with additional soft iron in box shapes. As the magnetic shielding is mainly a “mass effect” we may expect box shieldings to be more effective than cylindrical ones, having more shielding mass. The tests have been done with different configurations: – only µ-metal shielding (1 mm thick, 15 cm long: extending 3 cm beyond the photocathode area) – a 15 cm long iron box of transverse area 5 × 5 cm2 or 6 × 6 cm2 with a central hole of 3.2 cm diameter (to accomodate inside the PMT assembly with a 1 mm µ−metal shielding) made of different iron types: Fe360, Armco, ... – the same recessing the PMT assembly 1, 2 or 3 cm inside the edge of the iron shielding Results are shown for the signal reduction and timing versus the B field value in Figures 10 to 17, starting from a simple 1 mm µ-metal shielding and going to a composite shielding that includes also an iron box. Figure 17 shows a comparison for the different shieldings (signal reduction and timing versus the magnetic field intensity B) for the average and rms of a sample of ten PMTs. As a general conclusion, we see that: 9 Fig. 10. Signal ratio at field B and B=0 G, with only the mu-metal shielding of 1 mm, as function of the magnetic field B. Plots are for individual PMTs. Top panel: longitudinal field along the PMT axis, bottom panel: orthogonal field. – very low carbon content iron (Armco) is more effective than standard Fe360, even if this is quite good up to 500 G fields – as expected, bigger shielding (from 5 × 5 cm2 to 6 × 6 cm2 transverse area shieldings) are better – extending the iron shielding beyond the µ-metal shielding improves the situation As long as the signal has a sizeable pulse-height no deterioration in timing is seen 6 . The uncertainties in these studies came mainly from the following areas: (i) (ii) (iii) (iv) uniformity of the magnetic field stability of the laser pulses error in positioning of PMTs inside the magnetic field statical errors in the measure 6 only with the 1 mm µ-metal shielding or the 1 mm µ − metal shielding + 5 × 5 cm 2 Fe360 box shielding some effect was evident in timing (∆t or its FWHM) when the signal amplitude experienced a reduction of a factor ∼ 10 in one case or two in the other 10 Fig. 11. Timing difference, measured as ∆t = t ST ART − tST OP with only the mu-metal shielding of 1 mm, as function of the magnetic field B. Plots are for individual PMTs. Top panel: longitudinal field along the PMT axis, bottom panel: orthogonal field. As regards the first item, the uniformity of the magnetic field was estimated at better than 3% in the region of measurement and cross-checked with gaussmeter measurements. The stability of the laser pulses (second item) was monitored by beam-splitting the laser light and monitoring it with a PIN photodiode 7 . The overall stability of the system was within 5%, with maximum excursion in some bad runs up to 10%. The simple mechanics of the system allowed us a reproducibility of the positioning of the different PMTs’ at the level of some mm. From all the previous sources of errors, we may conservatively estimate a measurement error below 10%. Every single measurement referred to about 500 events, giving a negligible statistical error as compared to systematics. No study was done to assess eventual azimuthal angle effects on the PMTs and hysteresis effects in the shielding. As expected PMTs behave well for orientation of the B field orthogonal to the PMT axis (900 ), where the shielding effect is maximal, while along the PMT axis (00 ) the gain reduction may be more marked. The local shielding with 6×6 cm2 ARMCO iron in addition to the 1-mm µ−metal case seems more than adequate for TOF2 PMTs. 7 DET210 from Thorlabs 11 only muMetal, B field orthogonal 1.4 V PMT(B)/VPMT(0) V PMT(B)/V PMT (0) only muMetal, B field parallel 1.2 1 1.4 1.2 1 0.8 0.8 0.6 0.6 0.4 0.4 0.2 0 0 0.2 10 20 30 40 50 60 70 80 90 Bsolenoid [Gauss] 3 2.5 40 60 80 100 120 140 160 180 200 Bsolenoid [Gauss] 2 100 120 140 160 180 200 Bsolenoid [Gauss] 4 3.5 3 2.5 2 1.5 1.5 1 1 0.5 0 0 20 only muMetal, B field orthogonal 3.5 FWHM(B)/FWHM(0) FWHM(B)/FWHM(0) only muMetal, B field parallel 0 0 0.5 10 20 30 40 50 60 0 0 70 80 90 Bsolenoid [Gauss] 20 40 60 80 Fig. 12. Signal ratio at field B and B=0 G and FWHM ratio at field B and B=0 G for the timing difference, measured as ∆t = t ST ART − tST OP with only the mu-metal shielding of 1 mm. Left panel: longitudinal field, right panel: orthogonal field. The plots show the average and rms for a sample of ten R4998 PMTs. In addition the PMTs individual shields will be magnetically connected between them and to the “Virostek” plate, giving extra mass effect and so more effective shielding. In this way, our tests on single local PMTs represents a situation worse than the real one. Figure 18 shows how the local shielding is foreseen for TOF2, using a single bar for all PMTs of one side, instead of single boxes for individual PMTs. This is a preliminary design that can be clearly optimized with suitable computations, some of which have been proposed [6]. 5 Conclusions Measurements have been done up to longitudinal fields of 600 G for a sample of R4998 PMTs, with different shielding options. While a simple 1-mm µ-metal shielding is enough up to 60 (150) G for a longitudinal (orthogonal) B field, an additional ARMCO (6 × 6 cm2 ) shield is enough for longitudinal fields up to 600 G. As a final remark, a local shielding with 6 × 6 cm2 ARMCO shielding surrounding a 1 mm µ-metal works well to shield TOF2 PMTs, avoiding a much more expensive global shielding (“cage”). In this way TOF2 and KL may be put in a much nearer configuration, avoiding dead spaces. 12 fe360, 5x5-2cm extension 1.4 0.8 PMT 1.4 1.2 0.6 0.2 0.4 100 200 300 400 0.6 0.4 0.2 0.2 500 600 Bsolenoid [Gauss] 0 0 100 200 300 400 1 0.8 0.4 0.2 0 0 1.4 1.2 1 200 300 400 600 Bsolenoid [Gauss] 400 500 600 Bsolenoid [Gauss] 300 400 500 600 Bsolenoid [Gauss] 1.2 1 0.6 0.4 0.4 0 0 500 300 1.4 0.6 0.2 0.2 100 200 0.8 0.8 0.6 100 fe360, 5x5-2cm extension FWHM(B)/FWHM(0) 1.2 0 0 500 600 Bsolenoid [Gauss] fe360, 5x5-1cm extension FWHM(B)/FWHM(0) FWHM(B)/FWHM(0) fe360, 5x5 no extension 1.4 1.2 0.8 0.8 0.4 1.4 1 1 0.6 0 0 V PMT(B)/V 1 (0) fe360, 5x5-1cm extension 1.2 V PMT(B)/VPMT(0) V PMT(B)/VPMT(0) fe360, 5x5 no extension 100 200 300 400 0 0 500 600 Bsolenoid [Gauss] 100 200 Fig. 13. Signal ratio at field B and B=0 G and FWHM ratio at field B and B=0 G for the timing difference, measured as ∆t = t ST ART − tST OP with a Fe360 iron box shieldings (transverse area 5×5 cm2 ) in addition to the the mu-metal one extending 0 cm (left panel), 1 cm (middle panel) and 2cm (right panel) beyond the end of the mu-metal shielding. The B field is along the PMTs axis. The plots show the average and rms for a sample of ten R4998 PMTs. V PMT(B)/VPMT(0) fe360, 6x6-2cm extension fe360, 6x6-1cm extension 1.4 V PMT(B)/V PMT(0) 1.2 1 0.2 0 0 0.2 100 200 300 400 500 600 Bsolenoid [Gauss] fe360, 6x6 no extension 0 0 100 200 300 400 500 600 Bsolenoid [Gauss] fe360, 6x6-1cm extension 1.4 FWHM(B)/FWHM(0) FWHM(B)/FWHM(0) 0.4 0.4 0.2 1.2 1 0.8 1.4 1.2 1 0.4 0.2 0.2 100 200 300 400 500 600 Bsolenoid [Gauss] 0 0 200 300 400 500 600 Bsolenoid [Gauss] 300 400 500 600 Bsolenoid [Gauss] fe360, 6x6-2cm extension 1.4 1.2 1 0.6 0.6 0.4 100 0.8 0.8 0.6 0 0 0.6 0.6 0.4 0 0 1 0.8 0.8 0.6 1.2 1.2 1 0.8 1.4 1.4 FWHM(B)/FWHM(0) V PMT(B)/VPMT(0) fe360, 6x6 no extension 0.4 0.2 100 200 300 400 500 600 Bsolenoid [Gauss] 0 0 100 200 Fig. 14. Signal ratio at field B and B=0 G and FWHM ratio at field B and B=0 G for the timing difference, measured as ∆t = t ST ART − tST OP with a Fe360 iron box shieldings (transverse area 6×6 cm2 ) in addition to the the mu-metal one extending 0 cm (left panel), 1 cm (middle panel) and 2cm (right panel) beyond the end of the mu-metal shielding. The B field is along the PMTs axis. The plots show the average and rms for a sample of ten R4998 PMTs. Acknowledgments The essential help of Mr. F. Chignoli and R. Mazza of INFN Milano Bicocca for the preparation of this report is acknowledged. We would like to thank Mr. M. Piselli of University of Milano Bicocca for help and the generous loan of an Eutron power 13 Fig. 15. Signal ratio at field B and B=0 G and timing difference ∆t as a function of field B and, measured with an ARMCO iron box shieldings (transverse area 6×6 cm 2 ) in addition to the the mu-metal one extending 0 cm beyond the end of the mu-metal shielding. The B field is along the PMTs axis. The plots are for a set of ten R4998 PMTs. V PMT(B)/VPMT(0) feARMCO, 6x6 no extension V PMT(B)/V PMT (0) feARMCO, 5x5 no extension 1.2 1.4 1 1.2 0.8 1 0.6 0.8 0.4 0.6 0.2 0.4 0 0 0.2 100 200 300 400 500 600 Bsolenoid [Gauss] feARMCO, 5x5 no extension 1.4 1.2 1 200 300 400 500 600 Bsolenoid [Gauss] 300 400 500 600 Bsolenoid [Gauss] 1.4 1.2 1 0.8 0.8 0.6 0.6 0.4 0.4 0.2 0.2 0 0 100 feARMCO, 6x6 no extension FWHM(B)/FWHM(0) 0 0 FWHM(B)/FWHM(0) 1.4 100 200 300 400 500 600 Bsolenoid [Gauss] 0 0 100 200 Fig. 16. Signal ratio at field B and B=0 G and FWHM ratio at field B and B=0 G for the timing difference, measured as ∆t = t ST ART − tST OP with an ARMCO iron box shieldings (transverse area 5 × 5 cm2 (left panel) or 6 × 6 cm2 (right panel)) in addition to the the mu-metal one extending 0 cm (left panel) beyond the end of the mu-metal shielding. The B field is along the PMTs axis. The plots show the average and rms for a sample of ten R4998 PMTs. suppy. We are indebted also to Dr. L. Confalonieri, Hamamatsu Italia, and Ing. L. Vernocchi for help and many enlightining discussions. 14 V PMT(B)/VPMT(0) Comparison Between Shields 1.4 1.2 1 0.8 0.6 Shield fe360, 5x5 no extension 0.4 fe360, 6x6 no extension feARMCO, 5x5 no extension 0.2 feARMCO, 6x6 no extension 00 100 200 300 400 500 600 Bsolenoid [Gauss] 300 400 500 600 Bsolenoid [Gauss] FWHM(B)/FWHM(0) Comparison Between Shields 1.4 1.2 1 0.8 0.6 Shield 0.4 0.2 0 0 fe360, 5x5 no extension fe360, 6x6 no extension feARMCO, 5x5 no extension feARMCO, 6x6 no extension 100 200 Fig. 17. Signal ratio at field B and B=0 G and FWHM ratio at field B and B=0 G for the timing difference, measured as ∆t = t ST ART − tST OP with different iron box shieldings in addition to the the mu-metal one. The B field is along the PMTs axis. The plots show the average and rms for a sample of ten R4998 PMTs. 15 Fig. 18. CAD drawing of TOF2 with local shieldings for PMTs, using a single bar of ARMCO 6 cm thickness for each side [11]. 16 References [1] M.Bonesini, “The design of MICE TOF0 detector”, internal note MICE-NOTE-DET0145, 2006. [2] M.Bonesini et al., “ Study of the MICE TOF prototypes performance at the BTF test beam”, internal note MICE-NOTE-0163, 2006. [3] K. Tilley, private communication, MICE Collaboration Meeting, RAL, October 2005. [4] H. Witte,J. Cobb “The Magnetic Field in vicinity of TOF2”, MICE-NOTE-MAGNXXX in preparation. [5] http://www.vectorfields.com [6] G. Gregoire, “ Shielding update”, MICE presentation 7/11/2007. [7] G. Gregoire, “ TOFI cage”, MICE presentation at CM18, 13/6/2007; W.Lau, “Final discussion of drawings for TOFI shielding cage”, MICE presentation 31/7/2007. [8] R. Stephens et al., D0 Note 2706, 1996 (brought to our attention by L. Tortora). [9] S.O. Flyckt and C. Marmonier, “ Photomultiplier tubes: principles and applications”, Photonis, Brive, 2002. [10] M. Bonesini et al., IEEE Trans. Nucl. Sc. NS-50 (4) (2003) 1053. [11] R. Mazza, Sezione INFN Milano Bicocca, private communication. 17