INSTRUCTION MANUAL KC/5-S Rotating Consistency Transmitter
Transcription
INSTRUCTION MANUAL KC/5-S Rotating Consistency Transmitter
INSTRUCTION MANUAL KC/5-S Rotating Consistency Transmitter W41040300 V3.01 September 2006 This manual W41040300 V2.34 is applicable for KC/5 firmware V1.10 or higher (KC/5 transmitters delivered since August 2005 or upgraded) Manual V2.3. New Measurement Chamber and installation, Appendix 4 & 5 modified, print new manual. V2.31, Zero torque setup instructions updated, print pages 17 and 31-32. V2.32, Default calibration parameters corrected. Print page 29. V2.33, Specifications updated Software changes : V1.02ÆV1.03: Start-up procedure added to Calibration menu (In-air torque) V1.03ÆV1.04: Reset of Output Filter after the process stop Possible to edit Friction alarm limit in Factory Settings Added Power supply min. and max. temperature to Data Log Decreased Auto reverse timing, now minimum times 5seconds after each 1 minutes V1.04ÆV1.05: Corrected Year in Data Log, V1.04 displayed Year sometimes wrong V1.05ÆV1.10: Auto-zero function V1.10 -> V1.11: Spike removal software using std deviation added Manual V3.0. KC/5-S sensor, Measurement Chamber V1.1, Insertion Housing 67. ISO-torq adapter removed, fixed (flange) installation removed. KC/5 Safety Notice! PLEASE FOLLOW THESE INSTRUCTIONS CAREFULLY TO AVOID POSSIBLE INJURY! Kajaani Process Measurements’ KC/5 standard retractable version is designed for applications where line pressure is less than 10/16 bar (150/232 pounds per square inch (psi). A special, non-retractable model is available for use in applications with higher line pressures. The KC/5 is shipped in completed sub-assemblies. All safety related components are checked and verified to work correctly, before the product is shipped from our factory. The Customer is responsible for proper, safe installation by following the procedures outlined in the KC5 “Instruction Manual” included with each KC/5 shipment. Special care must be taken to ensure that the “Locking Ring” part #H41040131V1.0, part of the installation hardware, is properly tightened at all times according to the “Instruction Manual”. If the “Jack Assembly” part number A41040175V1.0, has been damaged and neither the “Seal Sleeve Assembly” nor the “Locking Ring” is secured according to the “Instruction Manual”, Section 3.5, the KC/5 Sensor may be propelled out of the line by the line pressure causing possible injury. The KC/5 Sensor must be securely locked into place with the “Locking Ring” after every instance where the sensor is removed and re-inserted into the process, or after the “Seal Sleeve Assembly” is disassembled and re-assembled. All persons who work on the KC/5 must follow all instructions in the “Instruction Manual”, Section 3.5, and in this Safety Notice, to eliminate any chance for injury when working on the KC/5. The “Locking Ring” and the “Jack Assembly” are critically important to the safe use of the KC/5 transmitter. Each must be properly attached, and securely tightened in place, while the sensor is inserted into the process line. When the Gate Valve is opened, the Sensor may be withdrawn from the line by slightly loosening the “Locking Ring”. The “Jack” then withdraws the Sensor from the process so the Gate valve may be closed. Only after the closed Gate Valve has isolated the line pressure from the withdrawn Sensor, is it safe to loosen the “Jack Assembly” and remove the “Locking Ring”. It is critically important to make certain that both the “Locking Ring” and the “Jack Assembly” are secured according to the procedures outlined in the “Instruction Manual”, Section 3.5, during initial installation and before each attempt to re - insert the Sensor into the process line. Note! Make sure all Bolts and Nuts are tightened properly before opening the gate valve. September 2006 W41040300 V3.01 Page 2 of 67 Table of contents KC/5 Safety Notice! .............................................2 PLEASE FOLLOW THESE INSTRUCTIONS CAREFULLY TO AVOID POSSIBLE INJURY! ....................................................2 1.1 Introduction ..................................................4 1.2 Contact information......................................4 1.3 Unpacking and inspection............................5 1.4 KC/5 system description ..............................6 1.5 Measuring principle......................................6 1.6 Sensing element ..........................................8 2.1 General notes ..............................................9 2.2 Selecting the optimum measuring site.........9 2.3 Dimensions and clearance requirements of the sensor ..............................................11 3.1 Process connections..................................12 3.2 Installing the Measurement Chamber........12 3.3 Installation to a BTG MEK process connection..................................................14 3.4 Installing the gate valve assembly .............15 3.5 Insertion of the sensor ...............................17 Preparing for insertion .............................17 Inserting the sensor.................................17 Safety Check after each insertion to the process line .............................................18 3.6 Flange Installation PN25............................20 3.7 Seal water connection ...............................20 3.8.1. Low pressure water flushing....................20 3.8.2. Self contained seal water system............21 3.8 Insertion depth adjustment ........................22 3.9 Installation of the display unit.....................22 4.1 Display unit connections ............................23 4.2 Sensor unit connections ............................24 5.1 Operation of the KC/5 – operator interface .....................................................24 5.2 Menu structure ...........................................26 5.3 Set-up ........................................................30 6.1 Calibration menu........................................32 6.2 Torque Offset (Zeroing of the torque measurement system) ...............................32 6.3 Single-point calibration ..............................33 6.4 Zero adjustment .........................................33 6.5 Sampling procedure...................................33 6.6 Changing calibration parameters...............34 6.6.1. Manual adjustment ..................................34 6.6.2. Transmitter calculated parameter change.....................................................34 7.1 Regular maintenance of the KC/5..............35 Seal water reservoir (optional, when mill seal water is not suitable) ...............................35 7.2 Maintenance menu ....................................35 7.3 Motor controls ............................................37 7.4 Troubleshooting .........................................38 September 2006 7.5 7.6 7.7 7.8 Removing the KC/5 sensor ....................... 40 On-line Zero check .................................... 41 On-line sensitivity check............................ 41 Checking the functionality of the KC/5 by removing the sensor from the process...... 42 7.9 Torque sensitivity calibration..................... 42 7.10 Cleaning the sensing element................... 42 7.10.1. Cleaning debris wrapped around the sensing element using “Auto Reverse.... 42 7.10.2. Cleaning the sensing element manually43 7.11 Replacing elastic shield, rod seal and needle roller bearing ................................. 43 7.11.1. Removing elastic shield ........................ 44 7.11.2. Removing rod seal and needle roller bearing.................................................... 44 7.11.3. Installing elastic shield, rod seal and needle roller bearing ............................... 45 7.12 Replacement of the mechanical seals ...... 46 7.12.1. Removing front mechanical seal........... 47 7.12.2. Removing rear mechanical seal ........... 48 7.12.3. Installation of front and rear mechanical seals .................................... 48 7.13 Electronics................................................. 50 7.14 Replacing Sensor Board ........................... 52 7.15 Replacing Optics Board ............................ 53 7.16 Replacing Power Supply Unit.................... 53 7.17 Replacing Connection Board .................... 54 7.18 Replacing LC Display Board ..................... 55 Appendix 3.1.: Standard Meas. Chamber, PN10 Gate Valve .................................... 58 Appendix 3.2.: Standard Meas. Chamber, PN16 Gate Valve .................................... 59 Appendix 3.3.: Standard Meas. Chamber, PN25 ...................................................... 61 Appendix 3.4.: BTG Adapter, PN10 Gate Valve62 Appendix 3.5.: BTG Adapter, PN16 Gate Valve64 Appendix 3.6.: Sensor Front Assembly, PN16SS .................................................. 63 Appendix 3.7.: Moment shaft and motor assembly................................................. 64 W41040300 V3.01 Page 3 of 67 1. Introduction 1.1 Introduction The robust design meets the demands of the harsh pulp and paper mill environment and provides the user with an accurate reading of pulp consistency for use in many control applications. In order to get the best results from the KC/5 transmitter, please read this manual carefully. The information provided in this manual provides clear and useful information to better understand installation requirements, set-up procedures, and the effective operation of the KC/5 rotating consistency transmitter in your application. By reading and understanding this information, your facility will be best able to optimize the performance of KC/5 transmitter to suit your specific needs and requirements. 1.2 Contact information Europe, Asia and South America: Kajaani Process Measurements Ltd. PO BOX 94 FI – 87101 Kajaani, Finland Tel: +358 8 633 1961 Fax: +358 8 612 0683 E-mail: [email protected] North America: Kajaani Process Measurements Inc. 636 U.S. Route 1, Box # 4 Scarborough, ME 04074, U.S.A Tel: 207 883 1095 1 800 consist (266 7478) Fax: +1 207 883 1104 E-mail: [email protected] Japan: Kajaani Process Measurements Ltd., Japan Branch Office K. Doi 4-444-5 Nishimiyashita Ageo-shi, Saitama-Ken 362-0043 Japan Tel: +81 487 776 7695 Fax: +81 487 776 8469 Mobile: +81 90 7633 8960 E-mail: [email protected] Please find your local supplier on www.prokajaani.com September 2006 W41040300 V3.01 Page 4 of 67 1.3 Unpacking and inspection Note: When unpacking the KC/5, check for shipping damage and verify the contents against the packing list. Immediately report any damage or missing items to ensure prompt repair and/or replacement as needed. The standard system contains following items in 4 separate boxes (See figure 1-1): Sensor A41040157V1.0 Interconnect Cable A41040095V1.0 Gate Valve Ass'y, PN 10 A41040224V1.0 Or Gate Valve Ass'y, BTG Adapter A41040159V1.0 Insertion Jack A41040175V1.0 Display Unit A410400179V1.0 Instruction Manual W41040300V1.0 Figure 1-1. KC/5 standard system components September 2006 W41040300 V3.01 Page 5 of 67 1.4 KC/5 system description The KC/5 consists of the Sensor Unit with the Mounting Assembly and the Display Unit (See figure 1-2). KPM offers five (5) installation options: 1. KPM Standard Measurement Chamber with Gate Valve and installation jack. 2. BTG MEK adapter with Gate Valve and Installation Jack. 3. Flange installation with KPM Measurement Chamber without Gate valve and Installation Jack. 4. Mounting bracket for mounting on open vessel (i.e. stuff box or saveall drop chest). 5. Flanged Adaptor for ISO-torq installation without KPM measurement Chamber or new Gate Valve, and new Installation Jack The Measurement Chamber is welded to the process pipeline (min. diameter 150mm, 6”). The DN80 (3”) gate valve is mounted to the Measurement Chamber. Sensor probe is inserted into the process by the heavy duty insertion jack which is mounted to the gate valve assembly. The display unit is connected to the sensor unit by a 10 m (33 ft) long, shielded 10-conductor cable (maximum length available is 30 m (100 ft)). The cable includes a connector that can be quickly coupled to the sensor unit. Display unit is powered by 85-264 VAC/47-63 Hz and supplies 12/48 VDC to the sensor. All external electrical connections are made in the display unit (4 – 20 mA analog output, binary inputs). The sensor motor can be interlocked with “stock pump running” information (pump) via binary input in display unit. Figure 1-2. The KC/5 sensor unit with the display unit. 1.5 Measuring principle The KC/5 sensor unit rotates the sensing element (see Figure. 1-3) in the pulp slurry. The slurry resists the movement. To overcome this resistance, a torque force is created. The higher the consistency, the greater September 2006 W41040300 V3.01 Page 6 of 67 is the force. The KC/5 measures the torque and converts this torque measurement into a consistency measurement value. KC/5 uses a “direct drive” servomotor. The “stator” is mounted in the body of the sensor unit, and the “rotor” integrated into drive shaft assembly. There is no drive belt or gearbox in the KC/5 design. The motor rotates two optical discs. One, the drive disc, is attached to the drive shaft. The other, the moment disc is coupled to the drive disc with coil springs. The moment disc is mounted to the moment shaft – the sensing element is located in the opposite end of moment shaft. When the drive shaft rotates, the springs force the moment disc to follow. The force-resisting movement varies, depending upon the consistency of the slurry within which the sensing element rotates. With higher resisting force, the springs stretch more and the moment disc lags farther behind the drive disc. At less resistance, discs follow closer to each other. The discs rotate at the same speed. As Cs increases, however, the “phase shift” between the two discs increases in direct proportion to the degree of torque applied. Figure 1-3. Measuring principle A high-resolution optical sensor detects the Phase Shift of the windows as consistency changes. Measurement electronics convert the optical signal into a calibrated torque value. The KC/5 system measures torque from zero - no torque applied to the sensor element - to a maximum 2000 mNm(170) ounces per inch. The accuracy is factory calibrated using torque standards. Each KC/5, therefore, has the same Cs sensitivity, making all sensors interchangeable without re-calibration. Torque is generated by shear force that changes when a rotating sensing element cuts through a moving fiber network (see Figure 1-4). As the curve shows, the torque generated by a rotating sensing element is not linear with consistency, but rather, increases more steeply at higher consistencies. The measurement algorithm in the KC/5 linearizes the relationship between consistency and the torque value. The KC/5 then calculates the output consistency as follows: C% = S * M + Z Where: C S M Z = = = = Consistency Slope Measurement signal (linearized torque reading) Zero offset Slope (S) and offset (Z) are grade specific. KC/5 comes with seven pre-calibrated consistency curves that fit most applications in the mill. One additional “user defined” grade is for special applications. September 2006 W41040300 V3.01 Page 7 of 67 100 Torque [mNm] Torque curvature gets steeper with increasing consistency Linearized measurement signal Reading in water at 0% Reading "in air" 0 Consistency [%] 0 Figure 1-4. Relationship between torque and consistency. 1.6 Sensing element KC/5 transmitter is provided with the paddle type sensing element as standard (See figure 1-5). Sensitivity of shear force measurement is directly proportional to the amount of “cutting surface” of a sensor element. The paddle-style sensing element has very large cutting surface area providing excellent sensitivity over full measurement range. 2 Paddle sensing element is for general purpose and provides similar signal as paddle sensing element. 4 Paddle sensing element is used to give additional sensitivity on very low consistencies. Plane sensing element is stream-lined and is less likely to pick up stringy debris as are typically encountered in unscreened recycle and “broke-line” applications. Half-Plane is for medium consistency (>8% Cs) applications. Paddle 2 Paddle 4 Paddle Plane Half-Plane Figure 1-5. Sensing elements September 2006 W41040300 V3.01 Page 8 of 67 2. Location and installation requirements 2.1 General notes The KPM transmitters are designed and manufactured to provide accurate and reliable measurements over a long period of time. Correct installation will ensure maximum performance and minimum cost of ownership. Avoid installing the transmitter around heavy vibration sources (e.g. cavitating or unbalanced pumps).The KC/5 has two seal water options, Mill seal water or integrated seal water reservoir. Mill seal water is recommended in applications with fillers and with recycled pulp. Install a reliable sampling valve close to the transmitter. Poor sampling leads to inaccurate calibration and underperformance of the measurement. KPM KS- samplers provide consistent and reliable solution for any sampling application. 2.2 Selecting the optimum measuring site Selecting the ideal location and installing the KC/5 properly are keys to successful measurement. Look for the site that: 1: Optimizes the performance of the instrument. 2: Provides clear access for maintenance. KC/5 rotating type transmitter is installed in the Measurement Chamber. The insertion depth of the sensing element is adjustable allowing to locate the sensing element to best measuring position also in difficult flow conditions. See fig. 3.10. Minimum straight pipe sections for transmitter: Low consistency 1.5 – 8%: Calming length Lbefore = 3 * D or 1 m (3 feet) whichever is longer Calming length Lafter = 1 * D Medium consistency 8 – 16%: Calming length Lbefore = 1.5 * D or 0.5 m (2 feet) whichever is longer Calming length Lafter = 1 * D The KC/5 rotating consistency is not sensitive for disturbances caused by turbulence, allowing installation immediately after the process pump when the straight pipe section is not available. September 2006 W41040300 V3.01 Page 9 of 67 x = Distance between the sensor unit and the sampling valve is about 500 - 1500 mm (20 - 59") Sampling valve 45° Sampling valve Sensor unit Lafter x Alternative position Sensor unit Lbefore Tank L = Straight line to pipe elbow, valve or pump Dilution water * Valve Valve Lab sample Installation location rules: 1. The axis of the sensor and the pump shaft should be perpendicular to each other. 2. Align pump shaft with valve stem. 3. Respect straight pipe section. Fig.2-1. Installation in a vertical pipe. Lbefore Lafter x Sensor unit Sampling valve L = Straight line to pipe elbow, valve or pump Tank Dilution water * x = Distance between the sensor unit and the sampling valve is about 500 - 1500 mm (20 - 59") Valve Lab sample Installation location rules: 1. Align pump shaft with valve stem . 2. Respect straight pipe section. Fig.2-2. Installation in a horizontal pipe. September 2006 W41040300 V3.01 Page 10 of 67 2.3 Dimensions and clearance requirements of the sensor 285 520 304 286 502 8 82 50 33 25 28 67 Figure 2-3. Dimensions of the KC/5 sensor. BTG Measuring Vessel 460 mm (18") 520 mm (20,5") 520 mm (20,5") 200 mm (8") 660 mm (26") KPM Measurement Chamber V1.1 375 mm (15") 520 mm (20,5") 805 mm (32") 723 mm (28,5") Figure 2-4. Clearance requirements of the KC/5 sensor Note: Always make sure the sensor is easily accessible for service. At the point of installation, the user should leave a minimum of 805 mm (32”) clearance perpendicular from the pipe at the point of installation (See figure 2-3), or 723 mm (28,5”) from the edge of a BTG weld-in stud. September 2006 W41040300 V3.01 Page 11 of 67 3. Sensor unit and display unit installations 3.1 Process connections There are 5 alternative process connections: 1. KPM Standard Measurement Chamber with Gate Valve and installation jack. This is the KPM recommended process connection. See 3.2 for installation instructions 2. Flange installation with KPM Measurement Chamber without Gate valve and Installation Jack 3. Existing BTG measurement chamber (Appendix 3.3 and 3.4)The KPM Adapter fits the Gate Valve and the Sensor Mounting Assembly to the existing BTG chamber. 4. Mounting bracket for mounting on open vessel or weir (i.e. stuff box or saveall drop chest). 5. Flanged Adaptor for ISO-torq installation without KPM measurement Chamber or new Gate Valve, and new Installation Jack 3.2 Installing the Measurement Chamber 1. Cut the chamber (Fig. 3-1.) so that the distance from flange to outer surface of pipe is 82 mm/ 3,23”, unless cut already by Kajaani Process Measurements. 2. Verify that the curvature of the chamber fits properly over the outside diameter of the pipe. 3. Cut a hole in the side of the pipe to match to the shape of the chamber as shown in figure 3-2. To prevent fiber debris from collecting on the edge of the cut-out, ensure the edges of the hole are ground smooth. Debris collected at this location can interfere with proper measurement (See appendix 1). 4. Place the chamber over the hole in the pipe. Verify that the hole is centered with respect to the chamber. Fillet weld around the entire perimeter of the chamber to attach the chamber to the pipe. (See welding drawing appendix 1). 4 A ø 215 (8.5") September 2006 ø 71 (2.8") 7.2 ø 168.3 (6.3") ø 71 (2.8") ø 168.3 (6.3") A 130 (5.1") A-A (PN10/16) W41040300 V3.01 33 (1.5") 130 (5.1") 33 (1.5") A-A (PN25) Page 12 of 67 Figure 3-1. Uncut Measuring Chamber Curvature of the chamber must match with the process pipe Cut hole into pipeline to match with the shape of the meas. chamber A A ø 215 ø 8.46" a4 A-A (PN10/16) a7 82± 2.5 3,23"± 0.1" Cut the chamber to match the process pipe. Center the chamber over the cut hole and fillet weld around entire perimeter A-A (PN25) 82± 2.5 ± 0.1" 3,23" Cross Sectional View of the Installation Chamber Figure 3-2. Chamber installation. (see appendix 1) September 2006 W41040300 V3.01 Page 13 of 67 3.3 Installation to a BTG MEK process connection Bolt the KC/5 adapter assembly to BTG weld-in stud or measuring vessel. Figure 3-2. Adapter assembly for BTG weld-in stud. Figure 3-3. Installation to weld-in stud or measuring vessel. September 2006 W41040300 V3.01 Page 14 of 67 3.4 Installing the gate valve assembly Note: The gate valve is installed in specific positions for either horizontal or vertical process pipes. In a horizontal process pipe, the gate valve is installed in the upright position. In the vertical pipe, the gate valve is placed in the horizontal position. Gate Valve Guide Pins (4) Measurement Chamber Bolts (4) Mounting Flange Nuts (4) Insertion Guide Gaskets Seal Assembly Lock Ring Figure 3-4. Gate valve assembly, horizontal pipe. Note: Installation jack must be installed at 12 o’clock position above the sensor. The sensor interconnect cable connector is at 3 o’clock position as in Figure 3-5. Figure 3-5. Gate valve assembly, vertical pipe. In the vertical process pipe installation, all parts - from the installation cone to the mounting flange - are rotated 90 degrees compared to the position shown in Figure 3.4. The direction of the Seal Assembly remains the same, jack must always be installed at 12 o’clock position above the sensor. September 2006 W41040300 V3.01 Page 15 of 67 See figure 3-5, 3-6 and appendix 3. 1. Screw the 4 stud bolts into the Measurement Chamber flange. 2. Install Gate Valve and gaskets to Measurement Chamber. Valve to horizontal position in vertical pipe, vertical position in horizontal pipe. 3. Place the Mounting Flange assembly (flange with Seal Assembly and Lock Ring preassembled) on stud bolts, fix the mounting flange to its place with 4 nuts. 4. Check that the gaskets are properly placed in the gate valve. Inspect opening to make certain the gaskets are not in the way of the sensor. 5. Check that jack mounting holes on Seal Assembly are horizontal (jack must always be installed at 12 o’clock position above sensor), if not remove Seal Assembly and turn to correct position. 6. Check orientation of Lock Ring opening is downwards, so that Radial Locking Bolt can be tightened (see Fig. 3.6), if not remove Lock Ring and turn to correct position. Figure. 3.6. Radial locking Bolt orientation. Safety Warning: Jack and Locking Ring must be secured and attached in place after inserting the sensor as per chapter 3.5. Note: Remember to close the gate valve before re-pressurizing the process pipe. 7. Test the gate valve installation for leakages at process pressure. Mounting the insertion jack Once the gate valve assembly has been installed and tested under process pressure, the system is ready for mounting the jack. Install the jack over the seal sleeve with 2 bolts provided. Lock the bolts with locking pins. Note: The jack is always installed in vertical position independent of gate valve position. September 2006 W41040300 V3.01 Page 16 of 67 Mounting Bolts to Seal Sleeve Mounting Bolts to Sensor Figure 3-7. KC/5 insertion assembly. 3.5 Insertion of the sensor Preparing for insertion 1. Check that the inside of seal sleeve is free of dried pulp and debris. Remove any material inside 2. Check that the O-ring inside the seal sleeve are secure and undamaged 3. Lubricate O-ring surfaces with grease to reduce friction (Vaseline or silicone grease can be used for lubrication) Inserting the sensor If the sensor has previously been in use make sure the sensing element and seal sleeve inside are clean. Every time the KC/5 is taken out of line, the seal sleeve and sensing element should be rinsed with clean water to remove any pulp and debris. Note: Before inserting the sensor the first time, please read the chapter 6.2. before insertion to follow correct setup procedure depending on the application. Note: To prevent the sensing element from pushing against the gate valve, make sure that the jack is in FULLY OPEN position. 1. With the gate valve still in the closed position, lift and slide the KC/5 unit into the seal sleeve until the bolts in the jack can be mounted to the sensor body. (See figure 3-8). The jack must be completely open. 2. Fasten sensor probe to the insertion jack with two bolts and insert the locking pins to prevent the bolts from turning by vibration. 3. Connect the seal water to the sensor. 4. Open the gate valve. 5. Jack the sensor in until the insertion depth adjustment ring stops against the locking ring. 6. Fasten the locking screw to lock the sensor safely into its measurement position. 7. Connect the interconnect cable to the Sensor September 2006 W41040300 V3.01 Page 17 of 67 Seal Sleeve Jacking Screw Locking Ring Jack Bolt Hole Depth Adjustment Ring Seal Water Seal Water Inlet Figure 3-8. Insertion of the sensor unit. Safety Check after each insertion to the process line Safety Warning: Jack and Locking Ring must be secured and attached in place after re-inserting the sensor. Proper installation of the locking ring is required to prevent: 1. Injury – secures the sensor in place to keep it from accidently backing out of the process. 2. Sensor Damage – stabilizes the sensor and minimizes vibration that may damage the jack and the sensor unit. AT ALL TIMES, IT IS CRITICAL TO ENSURE THAT THE LOCKING RING IS SECURELY TIGHTENED IN PLACE. The locking ring secures that the sensor does not come accidentally out from the process and that the sensor doesn’t vibrate inside installation parts. Excessive vibration can damage the jack and the sensor. It is important the Locking Ring is securely tightened. September 2006 W41040300 V3.01 Page 18 of 67 Figure 3-9. Locking ring of the Sensor The KC/5 sensor locking ring has to be tightened securely following way: 1. Make sure the orientation of the Locking ring is so that the Radial Locking Bolt can be tightened with allen key, if not, rotate the Locking ring on the Insertion Housing so that the Radial Locking Bolt is accessible. 2. Tighten the Radial Tightening Bolt of the Locking Ring. Figure 3-10. Radial locking of the Locking Ring. 3. Tighten the Locking Ring Fastening Bolts to lock the sensor to the Insertion Housing assy. Figure 3-11. Fastening the Locking Ring to the Insertion Housing. Note: Locking ring has to be tightened every time the sensor is installed after maintenance. September 2006 W41040300 V3.01 Page 19 of 67 3.6 PN25 Installation Note: The standard Gate Valve Assembly is designed for pressures up to 16 bar (240 psi). PN25 (360 psi) requires high pressure Jack, PN25 (360 psi) Gate Valve, high pressure insertion housing, and a special adapter for the sensor to mount the Jack to the sensor. PN25 (360 psi) Gate Valve PN25 (360 psi) Jack PN25 (360 psi) Adapter H41040375 V1.0 PN25 (360 psi) Lock Ring PN25 (360 psi) Insertion Housing PN25 (360 psi) Measurement Chamber Figure 3-12. PN25 (360 psi) Installation Parts. 3.7 Seal water connection KC/5 drive shaft is equipped with mechanical seals (double seal, tandem sealing system). The seal manufacturers recommend flushing of seals with water to extend the seal life. The water provides flushing, lubricating and heat transfer. KC/5 has two alternatives for seal flushing 1. Low pressure water flushing. 2. Self contained seal cooling and lubrication system (reservoir) 3.7.1. Low pressure water flushing It is recommended to connect flushing water if the water quality is good enough to keep the seal water system clean. Mechanical seals last longer with seal water when pulp contains fillers. The rate of flow is limited by the flow regulator. Make sure that seal water flow can be checked and monitored easily. Recommended water pressure is 1.0 -10 bar (15 -150 psi), max. particle size 200 µm. Flow is 0.5 to 2.0 l/min (0.13 to 0.5 US gal/min) within specified pressure. September 2006 W41040300 V3.01 Page 20 of 67 Connect the seal water to the ¼” male fitting of the sensor using flexible ¼” reinforced plastic tube. Flexible tube must be used to facilitate the move of the sensor. Install a shut-off valve close to the transmitter. Connect another ¼” reinforced plastic tube to the outlet, and draw it to the floor drain and fasten it just above the floor outlet so that the flow can be observed. It is a mandatory to install an anti-siphon prevention tube to prevent siphon phenomena in case of seal water loss. Reinforced flexible plastic tube 8/6 mm Elbow fitting Anti-siphon tube Flow regulator 0,5 - 2 l/min 0,13 - 0,5 gpm Valve Open flow to channel Note: As long as the seal water is running during a process shutdown there is no need to turn off the sensor motor 1 - 10 bar 15 - 150 psi Figure 3-13. Low pressure water flushing arrangement. 3.7.2. Self contained seal water system The self containing cooling liquid system is optional. This approach is recommended if water quality is poor. The cooling water reservoir is mounted in the insertion jack and connected with flexible tube to the seal water inlet. Volume is 1 liter (1/4 US gallon), use drinking water to fill the reservoir if ambient temperature stays above 0ºC (+32ºF). If freezing conditions are possible, use glycol/water mixture 50/50% Water reservoir is emptied when ever tubes to the sensor are decoupled. Refill the reservoir before power to the sensor is turned on. September 2006 W41040300 V3.01 Page 21 of 67 Figure 3-14. Self contained seal water system. 3.8 Insertion depth adjustment KC/5 sensor can be installed at adjustable depth into the process. To move the depth adjustment ring, loosen the set screws, slide the adjustment ring to desired position and relock the set screw again. Figure 310 gives most commonly used values. Unless specified in the order, KC/5 comes with standard depth adjustment for KPM Installation Chamber. With MEK adapter sensor need to be fully inserted. Depth adjustment becomes useful if flow pattern or consistency profile changes for one reason or another. By inserting the sensing element deeper inside one can avoid the layer where consistency profile varies. A Depth adjustment A: KPM Chamber 12mm BTG Adapter 0 mm Figure 3-15. Insertion depth adjustment. 3.9 Installation of the display unit The display unit can be located anywhere near the sensor unit so that it can be easily accessed by mill personnel. Connection cable standard length is 10 m (33’). Protection class of the display unit enclosure is IP66 (NEMA 4X). Mount the display unit to the wall with four (4) screws. September 2006 W41040300 V3.01 Page 22 of 67 150 6.9" Ø6 1/4" 195 7.7" 400 15.7" 350 11.8" 200 7.9" Figure 3-16. Dimensions of the display unit. 4. Electrical connections 4.1 Display unit connections The display unit has two functions: to supply 12/48 VDC power to the electronics/direct drive motor, and to serve as an easy-to-use interface. It is connected to the sensor unit via 10 m (33 ft) cable. Connections are shown in figure 4.1. Wire the display unit as follows: 1. Open the box cover. The connectors are in the connection board. 2. Install the interconnect cable coming from the sensor unit and 4 - 20 mA cable coming from DCS to terminals as shown in figure 4.1. The 4-20mA can be powered by sensor (default) or by DCS, see fig 7.14 how to change the setting. Note, if powered by DCS polarity is reversed (- pin 12, + pin 13). 3. Connect 220/110 VAC (85-264 VAC) September 2006 W41040300 V3.01 Page 23 of 67 + - Grade selection codes Grade pin 14 15 16 SW HW TMP CTMP GW Recycle Eucalypt User def 0 1 0 1 0 1 0 1 0 0 1 1 0 0 1 1 0 0 0 0 1 1 1 1 DCS Analog input 4 - 20 mA Binary Output +24 VDC Grade selection Sampler +24 VDC Binary Output +24 VDC Process stop Binary Input +24 VDC Alarm Display unit S AMPLE ES C Sensor Connections DCS Connections 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 Interconnect cable 10 m (33') shield current loop + current loop grade 0 + grade 1 + grade 2 + grade shield sample + sample shield process stop + process stop alarm on alarm GND alarm off shield current loop + current loop sensor supply +12 VDC sensor supply GND RS485 A RS485 B pwm motor supply + motor supply - ENTER ON L N 9 8 10 7 6 5 4 3 2 1 OFF 110/240 VAC 50/60 Hz Sensor unit Figure 4.1. Electrical connections. 4.2 Sensor unit connections Sensor unit is connected to display unit via supplied interconnect cable. All the internal connections are factory made. 1. Couple the seal water tubes, see fig 3-8. Water must be present on seals before power is connected to the system. 2. Plug in the quick connector at motor end of the cable. 5. Operating and setup instructions 5.1 Operation of the KC/5 – operator interface All the functions of the KC/5 are easily configured through the clear screen (See figure 5.1). Operation of KC/5 is menu driven. The arrow keys allow movement between the menus. September 2006 W41040300 V3.01 Page 24 of 67 The menu structure is divided into three main function blocks: Calibration, Set-up and Maintenance. The Factory Settings can be used to change some default parameters in special cases. The menu structure of the KC/5 user interface is shown in the next chapter. SW Cs: 3.5 % OUT 15.4 m A 1 2 4 3 5 Figure 5.1. The KC/5 user interface. 1. The display is 2 lines high x 16 characters in length. Main display shows both consistency value in percent and the analog output in mA. Display also shows current grade in the left upper corner. Menu display returns to main display automatically if not operated in 30 seconds. 2. Sample key. When activated, a 30 second countdown timer starts to show so as to identify when to take the sample. The display unit stores date, time and the measured consistency value averaged over 30 seconds time in its memory for later comparison to the lab values. This way it is easy to identify laboratory samples so that it corresponds in time to the consistency readings of the sensor unit. 3. Arrow keys - Arrow keys are used to move between menus, move cursor, or change values. Please refer to menu structure for more information about the arrow keys. 4. Esc key - press to cancel changes and/or return back to the previous menu. 5. Enter key - press to accept data and input changes. September 2006 W41040300 V3.01 Page 25 of 67 5.2 Menu structure SW Cs 0.00 % Out 4.0 mA ENTER PASSWORD 000 APPEARS IF PASSWORD OTHER THAN 000 PASSWORD IS SET IN SET-UP MENU. (PASSWORD 633 WORKS ALWAYS) CALIBRATION Change zero and slope Zero: Slope: xx.x xx.x Lab values 0000-00-00 Csxx.x 0:00 DIF 00.00 Enter lab value 0000-00-00 Csxx.x 0:00 DIF 00.00 Delete sample enter confirms SET UP MAINTENANCE Old zero: Enter new xx.x xx.x Old zero: New zero xx.x xx.x 0000-00-00 Csxx.x 0:00 LAB 00.00 Old slope: Enter new xx.x xx.x Old slope: New slope xx.x xx.x FACTORY SETTINGS (requires password 633) Calibration calculation Start up x sample ok enter continues S: xx.xx std. error: Z: xx.xx x.xx TORQUE OFFSET: T: XX.X O: +/- XX.X MEAS. TORQ. OFFSET ENTER CONTINUES Edit Torque Offset xxxx mNm Edit Torque Offset xxx mNm Target Cs ENTER TARGET Cs: xx:x % Edit Target Cs xx:x % September 2006 Edit Target Cs xx:x % W41040300 V3.01 WAIT A MOMENT RESULT: XX.X mNm ENTER CONFIRMS S= Z= Enter confirms S= Z= Enter confirms Page 26 of 67 CALIBRATION SET-UP GRADE: SW LOCAL CONTROL Grade: XXX S: XXX Z: XXX MAINTENANCE LOW LIMIT: xx.x % OLD LIMIT: ENTER NEW: XX.XX XX.XX FACTORY SETTINGS (requires password 633) HIGH LIMIT: xx.x % OLD LIMIT: ENTER NEW: XX.XX XX.XX OUTPUT FILTER: XX s EDIT FILTER: s XX SET CLOCK year-mo-dd hh:mm EDIT DATE & TIME year-mo-dd hh:mm GRADE SELECTION MODE: LOCAL SELECT MODE: REMOTE MENU LANGUAGE: ENGLISH SELECT LANGUAGE ENGLISH ELEMENT TYPE: PADDLE SELECT SELECT ELEMENT ELEMENT 8-PIN PADDLE ERROR OUTPUT CURRENT: NO EFF. SELECT ERROR CURRENT: 3.50 mA PRC.STOP OUTPUT CURRENT: SELECT PRC.STOP CURRENT: 4.00 mA PASSWORD 000 EDIT PASSWORD 000 AUTO REVERSE OFF EDIT GRADE NAME 4-PADDLE GRADE: S: X.XX SW Z: XX.X GRADE: S: X.XX SW Z: XX.X PLANE AUTOZERO SELF CLEANING September 2006 W41040300 V3.01 Page 27 of 67 SET-UP MAINTENANCE DEVICE IDENTIFICATION FACTORY SETTINGS (requires password 633) CHECK ON-LINE SIGNALS DEVICE TYPE: KC/5 OPTIC: CPU XXC XXF YYC YYF FRICTION x.x Nm POWER: MOTOR: x.x V/ xxC x.x V/x.xA SERIAL # TAG TEXTPROM FIRMWARE VERSION VERSION HART ID BINARY INPUT STATUS: OOOO RAW: xxxxxx SPD: xxx.x RPM N:xxxx M: xxxx DATALOG SINCE yyyy-mm-dd hh:mm Cs min: Cs max: T: yyyy Cs: xx.x% xx.x % xx.x % mNm min: xxxxx mNm max: xxxxx OPTIC xx C MIN: yy F OPTIC xx C MAX: yy F >1 Nm SHOCK COUNT XX Date Time Value Duration <-100 mNm SHOCK COUNT XX Date Time Value Duration RESET LOG DATA ENTER CONFIRMS TORQUE SETUP LOW TORQUE xxxxxxx / xx mNm EDIT LOW BRAKE xxx mNm VALUE: HIGH TORQUE xxxxxxx / xx mNm SIGNAL SIMULATION CHECK ALARMS (XX) RESET TO FACTORY DEFAULT VALUES September 2006 EDIT HIGH BRAKE xxx mNm VALUE: Cs SIMULATION SET Cs: xx.x % OUT xx mA mNm SIMULATION xxx SET mNm: Cs xx.xx% xx.xx mA Alarm 1 APPLY xx mNm TORQUE BRAKE RESULT: xxxxxxx ENTER CONFIRMS EDIT RAW VALUE xxxxxx EDIT APPLY xx mNm TORQUE BRAKE RESULT: xxxxxxx ENTER CONFIRMS EDIT RAW VALUE xxxxxx EDIT xxxxxx xxxxxx 2 3 PRESS ENTER TO CONFIRM RESET W41040300 V3.01 Page 28 of 67 SW Cs 0.00 % Out 4.0 mA CALIBRATION SET UP MAINTENANCE FACTORY SETTINGS PASSWORD (633) EDIT SN XXXX SERIAL NUMBER XXXX DEFAULT ELEMENT PADDLE CURRENT OUTPUT CALIBRATION PLANE FRICTION ALARM LIMIT: 1.5 Nm CHECK TEMP COMP FACTORS TEMP FACTOR 0 0C xx MOTOR CONTROLS September 2006 4-PADDLE EDIT VALUES S: xxxx Z: xxx FRICTION ALARM LIMIT: 1.5 Nm TEMP SENSOR DIAGNOSTIC SELECT DEFAULT: PADDLE 1 9 EDIT FACTORS 0-9 00C 0000 RPM SET POINT xxx RPM SET POINT xxx MOTOR DIRECTION CW +/- XX mNm MOTOR DIRECTION: CW CCW AUTO REVERSE 10s / 3 min / 0.8 SET AUTO REVERSE XX s / XX min / X.X CONTROL LOOP SPEED: 40 EDIT CONTROL LOOP SPEED: XX 157/187/ORD ID: 1059736B0008012 W41040300 V3.01 Page 29 of 67 5.3 Set-up Perform following checklist before configuration of the sensor: 1. Check that the sensor unit is properly inserted into the process. 2. Check that the seal water is flowing. 3. Check that the interconnect cable and power cable to the sensor unit are connected as per the wiring diagram. 4. Turn on the switch in display unit. The motor starts to run unless “process stop” input is activated. For normal installation, follow the steps in the KC/5 set-up checklist as follows: 1. Select grade: The user can change grade manually or by binary inputs from a remote device. The current grade selection mode (local or remote) is shown below the menu name. Each element has a predefined S (Slope) and Z (Zero) parameters for 8 different pulp grades (SW, HW, CTMP, TMP, GW, RECYCLE, EUCALYPT and USER DEF). Sensing Element: Grade: SW HW TMP CTMP GW Recycle Eucalypt User def Paddle (Standard) 2 Paddle S Z 1.6 -1.9 1.7 -1.5 1.6 -1.9 1.5 -1.9 1.2 -0.6 1.9 -2.1 1.9 -2.1 2.0 -5.0 4-Paddle S 1,9 2,0 1,9 1,8 1,5 2,1 2,1 2.0 Plane Z -3.4 -2.9 -3.4 -3.2 -2.2 -3.5 -3.5 -5.0 S 1.5 1.5 1.5 1.5 1.2 1.8 1.8 2.0 Z -1.5 -1.0 -1.5 -1.8 -0.8 -1.5 -1.5 -5.0 Table 6.1. Default values for S and Z. Grade names may be edited (see Menu Diagram, Section 5.2). Edited grade names remain in use until factory reset is activated, at which time, the original factory settings are restored in the language selected. Factory default values are stored permanently in KC/5 memory. When the KC/5 is started for the first time, the factory default values are in use. The user may then change these parameters as necessary. The KC/5 will use the most recent parameters selected for each grade. 2. Select low limit of measurement span The low limit may be set between 0.00 and 19.99 %/Cs. This value corresponds to output current 4 mA. Low limit must be smaller than high limit. 3. Select high limit of measurement span The high limit may be set between 0.00 and 19.99 %/Cs. This value corresponds to output current 20 mA. High limit must be larger than low limit. 4. Output filter The user may elect to filter out process noise, or abnormal spikes, from the output signal by using this function. Filtering time is expressed in seconds (default = 10 seconds). 5. Set clock Date and time are given in “YYYY-MM-DD and HH:MM” format. 6. Grade selection mode The user may choose either local selection of grade type or remote selection using binary inputs. September 2006 W41040300 V3.01 Page 30 of 67 7. Select menu language Select between English (default) and Finnish. 8. Select sensing element The user selectable sensing element types are Paddle (default), 4-Paddle and Plane. The user has to select the element that is installed on the sensor. 9. Error output The User may select one of four responses when the KC/5 detects an error. These are: • NO EFF: measured value is outputted even through the value of output can be erroneous (default). • 3.5 mA: analog output goes to 3.5 mA during error. • 22 mA: analog output goes to 22 mA during error. • FREEZE: output freezes on the last accepted measurement value. See more information from chapter 7.2 - Check alarms. 10. Process stop output When the “Process stop” input is activated the KC/5 sensor motor stops running. The user can select how the output behaves in this case. Alternatives are: • 4.0 mA: analog output goes to 4.0 mA during stoppage. • 20 mA: analog output goes to 20 mA during stoppage. • FREEZE: output freezes on the last accepted measurement value. • 11. Password If other than 000 is selected this password is required to move from the main display to the menus. 633 overrides any set password and can be used always. 12. Auto reverse When auto reverse is activated the motor runs at set interval in reversed direction. There are 3 options to select from: • “OFF”: Motor runs all the time in the same direction • “AUTO ZERO”: Motor runs at set interval for a certain time in reversed direction. The measured reversed torque is used for automatic zeroing of the torque measurement system. The reversed run frequency and time are set in the “factory settings” menu. • “SELF CLEANING”: Motor runs at set interval for a certain time in reversed direction. No automatic zeroing is done. The purpose to remove wrapped material from the sensing element. 6. Calibration A separate set of calibration values is needed for each type of sensing element and pulp grade measured. The KC/5 may be calibrated against laboratory values by taking samples or by choosing values from the sensor memory. Initial calibration of KC/5 consists of following steps: • A. “Torque Offset”: While the process is running in the normal conditions the motor is run in revered direction with the sensing element inserted in the measurement position. KC/5 calculates the torque offset for zeroing the system. Press ENTER to confirm. (See 6.2). • B. “Single-Point Calibration”: KC/5 tunes calibration parameters Slope (S) and Zero (Z) based on one laboratory sample (see Section 6.3). • C. “Calibration”: Fine-tuning of the calibration using multiple points is described in Sections 6.4-6.6. For standard installations, follow Steps A and B. September 2006 W41040300 V3.01 Page 31 of 67 6.1 Calibration menu Consistency calibration is made by linearized curve. Cs = S x M + Z Cs=Consistency, S = Slope, M = Linearized measurement and Z = Zero. 1. Change zero and slope: The user can review and adjust the Z and S values from this display setting. Zero Z can be adjusted from 99.99 to +99.99. Slope S can be adjusted from 0.00 to +19.99. 2. Lab values: The user may review the last 10 samples stored in the memory; including sampling times, measured values, and lab values. Through this display the new lab values are entered or old values edited. Use the delete (“DEL”) function to remove old sample.. 3. Calibration calculation: The KC/5 is able to calculate new Z and S values based upon a minimum of three (3) sets of stored calibration data. Recommendation: Use a spreadsheet program (e.g. Excel™) to record and track calibration data in place of using KC/5-aided calculation. The spreadsheet provides the user better visibility to data and simplifies removing of unreliable information. 4. Start up: 1. Torque Offset: To zero any offset in torque measurement system. Used normally in the initial set up of the unit 2. Edit torque offset: The zero adjustment offset received in the above procedure can be edited. Used only in special cases. 3. Target Cs: Used to quickly start-up a new unit. See 6.3 for details. 4. Edit target Cs: To edit the “presumed” consistency. See 6.3. for details 6.2 Torque Offset (Zeroing of the torque measurement system) A: Low consistency applications below 5% Cs. Insert the sensor into the process. Make sure that the process is running in the normal operating conditions. 1. With the sensor in-line and the process running, seal water flowing, turn power “On.” 2. Go to “Start up – Torque offset – Meas. Torq. Offset,” and press ENTER. 3. The display asks to “Wait a moment” while it is checking the zero point of the torque measurement system. 4. Result of the zero check is shown in the display. Press ENTER to complete the zeroing procedure. Write down the result. 5. Repeat the Meas. Torq. Offset procedure. Compare the results, if difference is bigger than 2mNm, proceed according the following procedure: B: Medium consistency applications. B: Medium consistency applications above 5% Cs. 1. Remove the sensor from the process according chapter 7.5, Removal of the sensor, or make sure the process line is empty or filled with water. 2. Hold the sensor on your lap and make sure the sensing element can rotate freely. 3. Go to “Start up – Torque offset – Meas. Torq. Offset,” and press ENTER. 4. The display asks to “Wait a moment” while it is checking the zero point of the torque measurement system. September 2006 W41040300 V3.01 Page 32 of 67 5. Result of the zero check is shown in the display. Press ENTER to complete the zeroing procedure. Write down the result. 6. Note: sensing element has to rotate freely. Anything touching the sensing element will interfere with measurement 7. Reinstall the sensor according Chapter 3.5, Inserting the sensor. 6.3 Single-point calibration Use the “Target Cs” procedure in the “Start-up” Menu to quickly start-up a new unit. • 1. Go to “Target Cs” menu while KC/5 is measuring the consistency (Cs%) value of the process. 2. “Enter” the ‘presumed’ consistency of the process. KC/5 averages 30 seconds of measured torque and adjusts calibration parameters S and Z to match the consistency value of the process. While averaging, the display reads “Wait a moment”. 3. Take a sample while KC/5 is averaging and perform laboratory analysis to verify the ‘presumed’ consistency settings. 4. If lab consistency differs more than ±10 % from the ‘presumed’ consistency, go to “Edit Target Cs” and change the original, ‘presumed’ consistency to the correct value and repeat tuning by selecting “ENTER”. 5. Continue calibration fine-tuning by following steps described in Sections 6.5-6.6 6.4 .Zero adjustment After initial calibration described in chapter 6.2-6.3 only zero adjustment (offset change) is usually needed. This is done by changing Z value. 1. Take the sample. 2. Read the transmitter’s consistency reading when the sample is taken. 3. Make laboratory analysis. 4. Adjust Z to make lab and transmitter readings match. Example: KC/5 reading Lab result New Z = old Z + 0.3 3.2% 3.5% If old Z = -7.0 New Z = -6.7 6.5 Sampling procedure Laboratory samples are taken and stored in the KC/5 memory as follows: 1. Press “SAMPLE” button. The display indicates “Sampling Time Left in Seconds”. During a 30second countdown, the KC/5 averages 30 seconds worth of measurement value and stores that data and the time of sampling, into the calibration data memory. 2. You must take the laboratory sample during this 30-second count down to match the stored data. 3. The KC/5 display indicates: measured Cs %, the linearized measurement M, and the MIN-MAX value. MIN - MAX value identifies the stability of the process during the sampling time. If the consistency value is unstable during the sampling time, the sample extracted from the process will not reflect the real consistency value – this sample is not a reliable basis to determine KC/5 calibration. Discard this sample. REPEAT the sampling process until a uniform sample is obtained. The measured value is stored in memory by pressing “ENTER” or automatically after 5 minutes have elapsed. By pressing “ESC” the sample value is discarded, and the display returns to the main menu. September 2006 W41040300 V3.01 Page 33 of 67 4. Analyze the sample in laboratory. 5. Enter the laboratory results into the KC/5 as follows: • a. Select “Lab Values” from “Calibration” menu. The display indicates the time of the last sample and the Cs average and the variation in the reading (“DIF”) during the time of sampling. Use the “ENTER” button to scroll between “DIF”, “Lab Cs%”, or “M” values. When the value of “DIF” is > 1 % Cs, the extracted sample should be rejected, and not used, by pressing “ESC”. • b. Scroll with “↑” or “↓” buttons to reach appropriate sample data. “Time” provides identification of the sample. • c. Press “→” button to select desired data. Menu prompts “ENTER LAB VALUE”. Enter by the “→” button, enter the laboratory value, then press “ENTER” to confirm the value or “ESC” to discard it. • d. When the sample data is not acceptable (for example, there is a large swing in consistency during the sampling), go to “DELETE SAMPLE” mode and press “ENTER” to discard the unacceptable sample values from KC/5 memory. 6.6 Changing calibration parameters There are two possible ways of adjusting the calibration values: 1. Change Z and S manually. 2. Ask KC/5 to calculate a new Z and S based on stored calibration data. 6.6.1. Manual adjustment Manual adjustment allows the operator to change the “Z” and “S” of the curve in use. This is performed in the “Calibration” menu. To achieve the highest degree of accuracy in calculating calibration parameters, KPM recommends use of a spreadsheet-type program. Contact KPM for a copy of spreadsheet file compatible with Excel. 6.6.2. Transmitter calculated parameter change When minimum of 3 laboratory samples have been collected and results entered into KC/5 memory, the transmitter is ready to calculate “new” calibration parameters. Make certain that Cs varies ± 2% of set point to ensure that the calculation is accurate. For a more accurate calibration, KPM recommends the use of minimum five (5) sets of sample data. 1. Select “Calibration” menu. Use the “→” button to go to “Calibration calculation”. 2. The display shows “# SAMPLES OK ENTER CONTINUES”. The value # must be 3 (three) or higher for the calculation to be performed. When fewer than three samples are entered, the display indicates “Invalid Samples”. KPM recommends extracting minimum five (5) sets of sample data spread at ± 2% Cs to determine correct calibration information 3. Press “ENTER” to calculate new S and Z using the laboratory samples. The display shows the new values, and the estimated error of calibration. Please check that S is at realistic level, typically 1,5 -5.0. Limits to enter new value are 0.0 – 19.99. If the result is unrealistic, discard and do not use the new parameters. Troubleshoot the calibration data. 4. Press “ENTER” to accept the new calibration parameters or press “ESC” to retain the old values. 5. The program automatically returns to the Main menu. Note! KC/5 requires a minimum of 3 laboratory samples to calculate calibration parameters. The calibration data is stored in a stack of 10 data sets that follow the FIFO (First In First Out) principle. When calculating calibration parameters, the KC/5 uses all data stored in memory at that time. The “Estimated Calibration Error” indicates the reliability of the new data and helps determine the validity of the calculated parameters. Therefore all unacceptable sample data must be deleted from memory. September 2006 W41040300 V3.01 Page 34 of 67 7. Maintenance 7.1 Regular maintenance of the KC/5 Safety Warning: Jack and Lock Ring must be secured and attached in place after re-inserting the sensor, as per chapter 3.5. The KC/5 requires no regularly scheduled maintenance other than checking that the seal water flow rate stays between 0,5 – 2l/min (1/8 – ½ gal/min. When using the seal water reservoir, keep the water level between 30%-70% full. Seal water reservoir (optional, when mill seal water is not suitable) The mechanical seal generates heat and requires cooling and lubrication. The seal water reservoir provides water to protect he seal when mill seal water cannot be used. Note: KPM recommends turning off the KC/5 motor during long process outages using “Process Stop Output” (see Figure 4.1, Electrical connections, “DCS Connections”. Check the seal water reservoir regularly – at least every other week, to determine the appropriate replacement rate. The reservoir should be kept above 30% full. If fibres appear in the reservoir, drain immediately and refill – fibres plug fluid channels and prevent seal fluid circulation. A large increase in fibre quantity indicates that the front seal may need replacement. Note! It is normal for mechanical seal to leak, but this should be kept at a minimum for longer seal life. Seal water may become cloudy when fibres, fines or fillers mix with water. 7.2 Maintenance menu The Maintenance Menu allows the user to perform two important functions: • To evaluate performance of the KC/5. • To evaluate the behaviour of the process. 1. Device identification: The user may view the following product information: • device type • serial number • tag number • firmware version. • textprom version • Hart ID • The tag number can be edited by the user. All other information is stored in permanent memory. 2. Check on-line signals: The user can view following values in real time: • sensor temperatures (optics and CPU) in Celsius and Fahrenheit • friction [Nm] • power module output voltage and temperature (V) (C) • motor voltage and current (V) (A) • binary input status (O = open, I = Closed) • raw-signal from sensor (RAW) • motor speed (RPM) (SPD) September 2006 W41040300 V3.01 Page 35 of 67 • • • • • torque (N) temperature compensated torque (T) linearized measurement signal (M) consistency measurement (Cs % = Slope x M + Zero). Displayed Torque (N) is the sum of measured torque + torque offset. 3. Data log since: The user may view the following data: • Cs min/max • mNm min/max • optics temp min/max • the number of abnormal shocks (hits) on the sensing element (high and low force ranges < –100 mNm and > 1 Nm). • The time since the last reset is shown with the data. The data logs may be cleared in the “Reset Log Data” menu. 4. Torque setup: Used to bench calibrate the sensor after service. Calibration is performed using a torque brake. The zero level is set when sensing element is spinning freely in air. 1. Set Low Torque, when in air, set 0,0 mNm 2. Perform Low Torque set up 3. Set High Torque value specific for the torque brake used. 4. Perform High torque set up Low torque: Displays the existing low torque calibration numbers – raw value and used torque. Torque value may be edited. Use to re-zero the sensor signal. The “New Low Torque” calibration value is accepted by pressing “ENTER” button. Calibration is cancelled by “ESC. High torque: Displays the existing high torque calibration numbers – raw value and used torque. Torque value can be edited. Recalibration requires user acceptance (“ENTER”). 5. Signal simulation: The correct output signal performance (4 to 20 mA) may be checked with this function. The simulated output current corresponds to the consistency value and output scaling. Simulated signal may be either the Cs% reading (“CS simulation”) or the mNm (“Torque simulation”) corresponding to 0-999 mNm. 6. Check alarms: Current active alarms (count of alarms shown in brackets). User can view following internal alarm flags: Alarm name Possible cause Action Temp sensor error Temp too high Temp too low Pr. Eeprom empty Gap too small Gap too big Torque setup err Cs>high limit Cs<low limit No signal Temperature sensor failure. Temp compensation not used Sensor overheating, >80 C Temp < 0 C (32 F) EEPROM, optics board failure; loose wire The optical discs out of alignment, check spring The optical discs out of alignment, check spring Torque set up failed; sensitivity out of limits Consistency over set range Consistency below set range Optics board failure, discs overlapping, loose wire Temp compensation not used Temp compensation not used Temp compensation not used Output set to alarm mode Output set to alarm mode Output set to alarm mode Output to 20 mA Output to 4 mA Output set to alarm mode September 2006 W41040300 V3.01 Page 36 of 67 Motor jam Motor Problem Power over temp. High friction Power problem Too high friction; seal tight, bearing worn out Motor RPM < set value; bearing, power supply Power supply to motor overheating Friction >1,5Nm; bearings or seal damaged Power supply, Hall-sensors of motor,connection cable Output set to alarm mode Output set to alarm mode Output set to alarm mode Output set to alarm mode Output set to alarm mode 7. Error messages: In addition to “Alarm” messages, KC/5 indicates the following “Error messages: Message Possible reason Invalid samples Laboratory result not in memory. Consistency varies more than 1 Cs % during sampling. Fewer than three (3) samples in memory. Slope too high Calculated Slope in calibration is greater than 19.99. Result will be discarded Invalid Cs Editing of consistency (in torque tuning) denied – S or Z has been changed; see 5.4 AUTO RVS Motor runs counter-clockwise in “Auto-reverse” mode 8. Reset to factory default values: The user may reload the default values of S (Slope) and Z (Offset) for all pulp grades. Default values are shown in Table 6.1. Other parameters that are returned to factory values are: sensing element type, output filtering, and local grade selection mode. Grade names are changed to default in the selected language. 7.3 Motor controls These functions are under the “Factory settings”-menu. Password 633 is required to enter. 1. Change of rotation speed: The user may change RPM of the motor within 300-650 RPM. Display shows set speed, actual speed and torque value. By pressing “Enter” the set value becomes permanent speed. Measurement sensitivity and signal-to-noise ratio improves when RPM is increased. As a rule of thumb: Cs range Rotation speed <3% >3% 600 RPM 420 RPM 2. Change sensor rotation direction: The user may change the direction of the motor rotation. Normal direction is ClockWise (CW) viewed from the motor end of the sensor. Counter ClockWise direction is indicated as CCW. Direction change is useful to remove wrapped material from sensing element. Reverse torque reading is indicated only. This reading helps to determine if the sensing element has dislodged foreign materials. Exiting the “Motor Controls” menu automatically returns the direction to CW). 3. Auto-reverse The rotating direction of the sensor may be changed manually, or programmed to automatically reverse after a pre-set interval. The parameters are set in the form of A sec / B min / C: September 2006 W41040300 V3.01 Page 37 of 67 A: selection of the time of reversed operation (default 10 sec). Time of reverse is adjustable from 1-120 seconds at 5-second intervals. B: how often reversed operation occurs (default 3 min). The time is adjustable up to 180 minutes. C: when auto-zero has been selected (set up menu) this number sets how much of the measured zero point shift of the torque is corrected at a time. 0 means 100%, 1 means 0%. Default is 0.8 (= zero point deviation is corrected in 20% steps). NOTE: “Auto-Reverse” is interrupted, when: • The “Sample” button is activated. • Display unit is not showing the Main display window - KC/5 is operating. The interrupted “AutoReverse” function restarts one minute after returning to the required position. The Output signal holds the last value displayed when “Auto-Reverse” is activated. Use the “Auto-Reverse” mode in recycled pulp applications to remove foreign matter that may attach to the sensing element. Reversing the rotation dislodges this material from the sensor. This helps reduce the need to remove the sensor from the line to clean the sensor head. 7.4 Troubleshooting The KC/5 features useful diagnostic functions to help find defective components. These easy-to-use diagnostics are located in the “Maintenance / Check on-line signals” section. First, check the following: 1. Motor is running (RPM is as set). 2. Friction is less than 1 Nm. (FRICTION) 3. Motor voltage is 24 – 48 VDC (POWER) If these values are not correct, use the wiring drawing (Figure 7-14) to help locate faulty component(s). Check following voltages: AC/DC Power supply outlet voltage is between 48±2 VDC. Measure at the AC Power Connection Board. Motor Supply voltage is 24-48 VDC. Measure at Connection Board AND at the Sensor Board. Sensor Supply Voltage is 12 VDC ± 1 VDC. Measure at Connection Board AND at the Sensor Board. Problem Potential source of problem & solution A. Motor is not running 1. Check that the process stop binary input is not activated (if active main display reads “Process stop”) 2. Check the power switch in display unit is ON 3. Check fuse in the display unit and reset if needed. (fig 7.14, 7.17 & 4.1) 4. Check that drive shaft is not jammed •If jam is reason alarm message reads motor jam •Check from on-line signals if motor tries to start (motor current) •Rotate drive shaft from sensing element and try if it helps •If not, replace mechanical seal or bearing assembly 5. Check power to the motor. See figure 7.14. (pins 9 & 10) •If there is no power to the motor replace AC/DC power unit. •If power ok change sensor board (See section. 7.13). B. Erratically spinning sensing element. 1. Sensing element needs to be re-aligned (See section 7.10). Or damaged one changed 2. If gate valve has been closed while sensing element is inside it, the moment shaft might also be bent. Check straightness of the head of the moment shaft. Replace moment shaft if not straight. C. Seal water reservoir Refill reservoir to line level using good quality water. September 2006 W41040300 V3.01 Page 38 of 67 has a low fluid level. D. Seal water is dripping from drain hole If the rear seal of dual seal system has failed, seal fluid will leak out of the drain channels. Replace rear seal immediately (See section 7.12) E. No RS485 signal. 1. Check the connector from sensor electronics to display unit. 2. Check the cable connections at pin 6 and 7 of Display Unit connector board 3. Replace the Sensor Board F. Raw output reading is erratic 1. Take out KC/5 unit and check that the sensing element is clean. If not clean - proceed as described in section 7.9 2. Check that area between sensing element and drive shaft does not have a build-up of hard debris. Clean if necessary (See section 7.9 7.10). 3. Sensing element is spinning erratically (See section problem B). September 2006 W41040300 V3.01 Page 39 of 67 7.5 Removing the KC/5 sensor Safety Warning: Jack and Lock Ring must be secured and attached in place after re-inserting the sensor as per section 3.5. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. Turn power switch OFF in the display unit. Unplug the interconnect cable (See figure 7.1) Close seal water line Loosen the seal assembly locking ring Jack the sensor to FULLY extracted position until jack stops Close the gate valve Open the drain valve in the mounting flange (verifying gate valve has good seal) Disconnect the seal water hose connectors Remove the locking bins and mounting bolts holding the sensor in the jack Remove the sensor Note! Position the sensor so that the sensing element does not rest on the locking ring while pulling out the sensor unit. This may bend the sensing element and the moment shaft and cause severe damage. Gate Valve Insertion Jack 4 5 Sensor 9 6 Pipe 1 2 Drain 7 Valve Locking Ring 3 Mounting Bolts 8 Figure 7-1. Removing the KC/5 sensor. September 2006 W41040300 V3.01 Page 40 of 67 7.6 On-line Zero check KC/5 measurement zero reading may be checked while the KC/5 is operating the following way. 1. Go to “Factory settings”, Motor Controls” menu and press”→” Read the mNm value and record it when sensor is operating normally. 2. Change the direction of rotation to CCW by pressing “↓” 3. Read the CCW mNm value and record it 4. Change the direction of rotation back to CW by pressing “↓” 5. Repeat the test by reading and recording the CW mNm value again. 6. Change the direction of rotation to CCW by pressing “↓” 7. Read the CCW mNm value again and record it. 8. The absolute value of the average CW torque readings and absolute value of the average CCW torque readings should be closely on the same level indicating the zero point of torque measurement has not drifted. 9. If absolute values differ more than 2 mNm, adjust Torque offset according to section 6.2. 10. Exiting the “Motor Controls” menu automatically returns the direction to CW Note! When process had quick consistency changes, the zero check may give faulty results due to rapidly changing torque measurement. In Medium consistency applications, changes in torque are typically so big and fast, the on-line zero check does not give good results. In MC applications, check the zero according to section 7.8. Checking the functionality of the KC/5. 7.7 On-line sensitivity check KC/5 sensitivity may be checked with a Torque Brake while the KC/5 is operating. The Torque Brake is available from KPM, part number A41040222. 1. 2. 3. 4. 5. 6. 7. 8. Open the “Check torque” plug in the end of the sensor to expose the torque socket Check the torque setup value from the torque brake (KPM brake 85 mNm). Read the torque value “T” mNm -reading from ”maintenance” menu, “on-line signals Connect the Torque Brake to the socket - this may be done without stopping the motor. Hold the Torque Break firmly (not with excessive power) against the socket while avoiding contact between shaft and case Read the “T” mNm -reading from ”maintenance” menu, “on-line signals” The T value mNm -reading should increase about the same than the brake set value from the nobrake level. If not, proceed as per section 7.7. Reinstall the ”Check Torque” cap tightly Figure 7-2. Sensitivity check with torque brake September 2006 W41040300 V3.01 Page 41 of 67 7.8 Checking the functionality of the KC/5 by removing the sensor from the process Remove the sensor from the process (see section 7.5) 1. Check the elastic protective shield and clean the front end of the probe 2. Place the KC/5 sensor unit into the horizontal position so the sensing element is free to rotate. 3. Connect seal water to the sensor 4. Connect the interconnect cable to the Sensor. 5. In the display unit, switch the power to the “On” position. The sensing element will rotate ClockWise – CW – when looking from the Motor end of the KC/5. 6. The sensor is properly calibrated when “No-Torque T”-reading (the sensing element rotating in air) is +/-2 mNm. T-value can be read from Maintenance/On-line Signals menu or from Calibration/Start up/Torque Offset menu. 7. Clean the front part of the sensor and check to verify that, behind the sensing element, the elastic shield is intact (see Section 7.10). 8. If the performance has not improved, then perform torque sensitivity calibration (see Section 7.9 below) 7.9 Torque sensitivity calibration Torque calibration is performed by applying a measured torque force to the sensing element. The KC/5 is factory calibrated using 0-85mNm. Low torque value, 0 mNm, is measured while the sensing element is spinning in air. 1. Place sensor unit on the workbench with the display unit next to it. 2. Position the KC/5 so that the drive shaft is horizontal. 3. Connect seal water to the sensor Note! When the KC/5 transmitter is operated in air, the mechanical seal will heat up if seal water is not connected. This can damage the seal. To prevent this, make sure there is seal water in the reservoir, or connect mill water to the seal water connections on the sensor. 4. Switch power ON in the display unit. 5. Select “torque setup” in maintenance menu and type in the high torque value to be used. Press “ENTER” and place torque brake to the torque socket (See figure 7-2). Press “ENTER” to initiate the calibration. After a while calibration raw value appears in display. “ENTER” accepts the new number, “ESC” cancels. 6. Repeat above with low torque value (sensing element rotating freely in air). 7.10 Cleaning the sensing element 7.10.1. Cleaning debris wrapped around the sensing element using “Auto Reverse Occasionally, debris catches on the sensing element – plastic, wire, string, etc. – and may even wrap around the sensing element. This results in distortion of the measurement. Reversing the direction of the rotation aids in removing the debris and foreign material. 1. Go to “Motor Controls” under “Factory Settings”. Select “Motor Direction” block, display reads “CW” = clockwise. Select “CCW” = counter clockwise rotation direction. 2. Press ENTER and direction is reversed. Display returns to show RPM and T. 3. Allow the sensing element to run in reverse direction. Watch the “T” value decrease as debris is released. September 2006 W41040300 V3.01 Page 42 of 67 4. Return to normal rotation “CW”. Check the “T” value – if it still shows the original value, repeat reversed rotation a few times until “T” reading indicates the removal of the foreign particles. 7.10.2. Cleaning the sensing element manually Safety Warning: Jack and Lock Ring must be secured and attached in place after re-inserting the sensor as per chapter 3.5. The sensing element must be free to rotate with reference to the drive shaft. This twisting action is the essence of the torque measurement. If hardened debris or chemical additives build up on elastic shield or in the region between the sensing element and the shield, it can cause the sensing element to lock-up on to the end of the shaft. This can cause a loss of sensitivity, and in the worst case, cause the output to be frozen at a constant reading. To clean, proceed as follows: 1. Remove sensor from the process line (See section 7.5). 2. Remove the sensing element by unscrewing the fixing screw (See figure 7-3). 3. Remove shield locking ring, if necessary for cleaning 4. Clean and remove any build-up on the sensing element, elastic shield, and the area in between them. 5. Check the condition of elastic shield; if cut or broken replace with a new one (See section 7.10) 6. Reattach the sensing element to the moment shaft and tighten the fixing screw. Figure 7-3. Sensor head and the shafts. 7.11 Replacing elastic shield, rod seal and needle roller bearing Safety Warning: Jack and Lock Ring must be secured and attached in place after re-inserting the sensor as per chapter 3.5. If the protective elastic EPDM shield is damaged in any way, replace immediately. KC/5 elastic shield replacement set contains the parts needed. Content of Elastic Shield Replacement Set A41040302V1.1 Order Code Type H41040158 Elastic Shield 2650016 Rod Seal 2750001 Needle Roller Bearing 2700015 O-ring 19.1x1.6 FPM September 2006 W41040300 V3.01 Page 43 of 67 7.11.1. Removing elastic shield 1. Remove the sensing element (See figure 7-3). 2. Open the screws in the shield locking ring. Remove the ring. 3. Open the set screws in the shield cover and pull off the shield cover. Elastic shield is removed with the cover (see figure 7-4b). Figure 7-4a. Sensor head Figure 7-4b. Removing the elastic shield. 7.11.2. Removing rod seal and needle roller bearing 1. Pull out moment shaft bushing using two screw drivers (figure 7-5a) 2. Remove the seal washer to get access to the rod seal (figure 7-5b). Figure 7-5a. Removing moment shaft bushing Figure 7-5b. Removing seal washer 3. Remove moment shaft bushing (figure 7-5c) . Remove the needle roller bearing by tipping the sensor head downwards and moving moment shaft radially (figure 7-5e). Check the condition of the needle roller bearing (figure 7-5d). Replace the bearing if needed. September 2006 W41040300 V3.01 Page 44 of 67 Figure 7-5c. The rod seal location Figure 7-5d. Location of the needle roller bearing Figure 7-5e. Removing the needle roller bearing 4. Clean the moment shaft of any dried material. Pay extra attention to area under the rod seal. 7.11.3. Installing elastic shield, rod seal and needle roller bearing 1. 2. 3. 4. 5. Install the needle roller bearing. Use silicone oil as lubricant. Place the new rod seal (Shield Replacement Kit) inside the moment shaft bushing. Replace the moment shaft bushing O-ring with a new one from the replacement set Reinstall the seal washer. Note that the bevelled side is facing the rod seal. (see figures.7-5b,c) Slide the assembled bushing over the moment shaft, and slide it all the way down the shaft until it stops on the drive shaft. September 2006 W41040300 V3.01 Page 45 of 67 Figure 7-6. Installation of the moment shaft bushing and the seal washer. 6. Place a new elastic shield into the shield cover. Pull the shield over the drive and moment shafts. To reduce friction, wet the shaft surfaces with soapy water). Push -do not twist - the shield cover all the way until the shield cover stops against the drive shaft. 7. Tighten the set screws on the shield cover (see figure 7-4a) 8. Place the shield locking ring on top of the elastic shield –do not fasten the screws yet. 9. Reattach sensing element to the moment shaft and tighten the fixing screw 10. Lift the shield locking ring until it rests against the sensing element and tighten the screws 7.12 Replacement of the mechanical seals Safety Warning: Jack and Lock Ring must be secured and attached in place after re-inserting the sensor as per chapter 3.5. The KC/5 features tandem mechanical seals. The seal water is introduced between the front (process side) and rear (motor side) seals. The front seal prevents pulp from getting inside the KC/5 probe housing. The rear seal keeps seal water from getting to drive shaft bearings. The seals are manufactured according to DIN 24960 standard and are available as stock items. Front seal (process side) “ Rear seal (motor side) “ 2650017 2650018 Order Code Type Bellows Seal T502, SiC/EPDM 2650005 Stationary Seat C606/25, SiC/EPDM Bellows Seal T502, SiC/FPM 2650007 Stationary Seat T6/25, SiC/FPM Note: If Rear seal (motor side) has been in service more than three years, KPM recommends replacement when the Front Seal needs replacement. Mechanical seal has an estimated lifetime of five years. Rear mechanical seal should be replaced if 1. When seal water/fluid is leaking out from the draining duct 2. When at the time of replacing the front mechanical seal, and the rear seal has been in service over 3 years September 2006 W41040300 V3.01 Page 46 of 67 3. When the rear seal has been in service for 5 years. Figure 7-7. Location of the draining duct. 7.12.1. Removing front mechanical seal 1. Remove first the sensing element and the elastic shield (section 7.10.1 and figure 7-4) 2. Remove the front bellows seal by sliding it out with symmetrically placed screw drivers as shown in figure 7-8a. 3. Unfasten the stationary seat by removing the locking ring (4 screws) (figure 7-8b) 4. Remove the front stationary seat with two screw drivers (figure 7-8c) Figure 7-8a. Removing the bellows seal. September 2006 Figure 7-8b. Removing the locking ring. W41040300 V3.01 Page 47 of 67 Figure 7-8c. Removing the stationary seat with 2 screw drivers. 7.12.2. Removing rear mechanical seal 1. Anchor the sealing cover in a bench vice with lock ring. Break free the sleeve from the installation assembly or use Sealing Cover Removal Tool H41040314 V1.0). Rotate the sensor head until the head is completely open. Figure 7-9. Opening the sealing cover 2. 3. 4. 5. Open two set screws and remove Seal Locking Ring (see figure 7-10). Remove the bellows seal by lifting it with symmetrically placed screw drivers. Open the 4 screws of the seat adapter holding the rear stationary seat. Remove the rear stationary seat from seat adapter by pushing forwards. Figure 7-10. Removing the rear mechanical seal 6. Wipe the drive shaft clean and check that the draining duct is open. 7.12.3. Installation of front and rear mechanical seals Wipe the drive shaft clean with soapy water before installing the rear stationary seal. September 2006 W41040300 V3.01 Page 48 of 67 Note: Use clean, soapy water to lubricate the seals and to reduce friction when installing seals. NEVER use oil, grease or alcohol for this purpose. 1. Install seat adapter with 4 screws. Install the adapter so that seal water ducts are not blocked. 2. Push the stationary seat into seat adapter with polished side facing up. Use your hands to press the stationary seal into place 3. Push the rear bellows seal on top of the stationary seat with polished face down Figure 7-11. Rear seal compression adjustment. 4. Place the seal locking ring on top of the bellows seal and adjust the seal compression by pressing the ring. Set installation length to 27 mm ±0.5 mm from the top of the seat adapter to the bottom of the ring (see figure 7-11). 5. Lock the locking ring with two (2) set screws 6. Apply small amount of graphite grease (e.g. molykote) to the sealing cover threads. 7. Install the sealing cover. Anchor the sealing cover to a bench vice using the locking ring and brake sleeve, or use the Sealing Cover Removal Tool, H41040314 V1.0, and fasten the sensor head (see Figure 7-9). 8. Place the front stationary seat on the sealing sleeve. The polished side faces up. Use your hands to press the mating ring into place (see figure 7-8c) 9. Install the locking ring and fasten the four (4) screws (see figure 7-8b) Figure 7-12. Installing front seal. 10. Slide the front bellows seal over the drive shaft. Compress front seal to its smallest size. September 2006 W41040300 V3.01 Page 49 of 67 Figure 7-13. Installing shield cover. 11. Install the shield cover and the elastic shield (see fig 7-13). Push – do not twist – the shield cover until the shield cover stops against the drive shaft. 12. Fasten 3 set screws. 13. Uncompress front sealing spring until it reaches the shield cover. Now installation length is correct. 14. Place the shield locking ring on top of the elastic shield. Leave screws loose 15. Reattach the sensing element to the moment shaft and tighten the fixing screw. 16. Lift the shield locking ring until it rests against the sensing element and fasten the locking ring screws 7.13 Electronics KC/5 electronics contains the following replaceable boards/units: 1. 2. 3. 4. 5. Part name Optics Board Sensor Board Connection Board AC/DC Power Supply LC Display Board Order Code A41040062V1.0 A41040070V1.0 A410402202V1.0 3100003 A41080023V2.0 The KC/5 features helpful diagnostics to quickly locate faulty components. These easy-to-use diagnostics are located in the “Maintenance / Check on-line signals” Section. Check following first: 1. Motor is running (RPM is as set). 2. Friction is less than 1.5 Nm. (FRICTION) 3. Motor voltage is 24 – 48 VDC (POWER) If not, use wiring layout drawing (figure 7-14) to locate faulty component. To help pinpoint the problem, check the following voltages: AC/DC Power supply outlet voltage is 48±2VDC, measure from Connection Board. Motor Supply voltage is 24 – 48 VDC, measure from Connection Board and from Sensor Board. Sensor Supply Voltage is 12 VDC ± 1 VDC, measure from Connection Board and from Sensor Board. September 2006 W41040300 V3.01 Page 50 of 67 Jumper position: JP1: Current output mode Motor drive Active (default) Passive Sensor Board 1 Optical disk J1 9 Motor supply 24-48VDC J7 _ + Photo detector 8 1 2 9 7 1 6 3 Optics Board 4 5 6 7 8 1 2 9 6 3 5 Measurement signal 4 Quick Connector Commutation pulses Interconnect Cable AC/DC Power supply 48 VDC Mains Connector 85-264 AC in Connection Board Resettable fuse Signal Connn. Board J1 Interconnect Cable wire Shield 1 shield Current Loop + 2 white Quick Connector pin Sensor Cable wire Sensor Board J7 J1 1 white 8 Current loop - 3 brown 2 brown 9 Sensor Supply (+12V) 4 green 3 green 3 Sensor Supply GND 5 yellow 4 yellow 1 RS 485 A 6 grey 5 grey 6 RS 485 B 7 pink 6 pink 7 PWM Control (pulse) 8 blue 7 blue 5 Motor Supply + 9 black + gray/pink 8 black + gray/pink 2 Motor Supply GND 10 violet + red/blue 9 violet + red/blue 1 Figure 7-14. Wiring layout. September 2006 W41040300 V3.01 Page 51 of 67 7.14 Replacing Sensor Board Refer to figure 7-15. To change the sensor board, proceed as follows: 1. Remove the KC/5 sensor from process line (See section 7.5). 2. Open the back of the enclosure by opening the 4 screws 3. Remove Plastic insulation cover 4. Disconnect the cable connectors (3 pcs) on the Sensor Board. 5. Unscrew the six screws which attach the measurement board to the end housing. 6. Uncouple the Sensor Cable wires on the board 7. Replace with a new board. Note the orientation of connectors (fig 7-15). 8. Reconnect the cables on the board. Figure 7-15. Sensor electronics. September 2006 W41040300 V3.01 Page 52 of 67 7.15 Replacing Optics Board Refer to figure 7-16. 1. Open the small cover on side of the sensor housing, see fig 7-15. 2. Disconnect the optics cable connector on the optics board 3. Open 4 mounting screws and remove the board 4. Install a replacement board 5. Align the board so that the optics disks are in the middle of the alignment slot (see figure 7-16) 6. Connect the cable Figure 7-16. Optics Board Alignment 7.16 Replacing Power Supply Unit Disconnect AC power supply to Display Unit before starting the replacement work Power supply unit provides 48 VDC operating voltage to the motor and 12 VDC for display and measurement electronics. 1. Disconnect AC supply to Display Unit 2. Disconnect the cables between the connection board and the power supply unit and remove the mounting screws (4 pcs) and the mounting bars (2 pcs), figure 7-17. 3. Mount a new unit and reconnect the cables September 2006 W41040300 V3.01 Page 53 of 67 Figure 7-17. Display Unit. 7.17 Replacing Connection Board Disconnect AC power supply to Display Unit before starting the replacement work 1. Disconnect AC supply from Display Unit 2. Disconnect 110/220 VAC mains power wires from the connection board 3. Remove the power supply unit (see section 7.15) 4. Disconnect all the cables, open the 4 mounting nuts of the board and the screw fastening the regulator on the heat sink (fig 7-17) 5. Remove the board 6. Mount a replacement board 7. Install the screw fixing the regulator on the heat sink. Note the insulation pad between the heat sink and the regulator, and the insulation bushing between the screw and the regulator (fig 7-18) 8. Check insulation between the regulator body and the display unit body with an ohmmeter (must be over 1 MΩ) 9. Install 4 mounting nuts. 10.Re-install the removed power supply unit and connect the cables September 2006 W41040300 V3.01 Page 54 of 67 Figure 7-18. Regulator isolation (bottom view of the connection board) 7.18 Replacing LC Display Board 1. Disconnect all the wires, open the 7 mounting nuts of the board and remove the board, see fig 717 2. Check the height of the LC Display board, the new version with backlight is thicker that the version without backlight. 3. If you replace older type Board with new type, please install new spacers (7 pcs) delivered with the new LC display board. 4. Mount the new board and connect the cables. September 2006 W41040300 V3.01 Page 55 of 67 Appendix 1.1: Installation drawing - Meas. chamber PN10/16 3 4 This document must not be copied without our written permission, and the contents thereof must not be imparted to a third party or be used for any unauthorized purpose. (c) Kajaani Process Measurements Ltd. 2002 2 1 Description Vers. Date Prepd D C PN10/16: a4 PN25: a7 A B 82± 2.5 Tätä asiakirjaa ei saa ilman meidän lupaamme jäljentää. Sitä ei myöskään saa esittää toiselle tai muutoin asiattomasti käyttää. (c) Kajaanin Prosessimittaukset Oy. 2002 A ø 215 A-A Cut chamber to same radius with existing prosess pipe. A September 2006 2 H41040324 1 H41040323 Item Part No. V1.1 Pipe V1.0 Flange Description Vers. Treatment Material KAJAANI PROCESS MEASUREMENTS Ltd Cross Sectional View of Installation Chamber Assembly. Notes Tolerance Title Scale KC/5 Meas.Chamber V1.1 PN 10/16, SS Prepd With cutting W41040300 V3.01 1:4 12.09.06MLa Drawing No. E41040326 V1.1 Page 56 of 67 Appendix 2: BTG weld-in stud adapter and flange dimensions September 2006 W41040300 V3.01 Page 57 of 67 Appendix 3: Assembly drawings and parts lists Appendix 3.1.: Standard Meas. Chamber, PN10 Gate Valve 21 6 22 20 17 12 10 3 8 11 7 13 14 1 9 18 15 4 2 16 19 5 Item Part No. Description Value Device Code Qty Unit 1 2000022 Vers. Hex Screw M8x100 DIN 931 A4 2 pcs 2 2000183 Screw, Socket Head Cap M8x80 DIN 912 A4 6 pcs 3 2000097 Hex nut M16 DIN 934 A4 4 pcs 4 2450004 Douple-nipple R 1/4 Aisi316 DIN 2990 1 pcs 5 2450005 Ball valve, Reduced bore R 1/4 Aisi 316 (onninen 2017k) 1 pcs 6 2600008 Knife Gate Valve PN10 Fabri C67S316EBU ANSI 125/150-3" Aisi 316 1 pcs 7 8 2700011 2700012 O-ring O-ring 69.85x3.53 FPM 80x5 FPM 3 1 pcs pcs pcs 9 E41040326 V1.1 Meas. Chamber PN10/16 1 10 H31040129 V1.0 Mounting Flange, SS 1 pcs 11 H41040408 V1.0 Insertion Housing 67, SS 1 pcs 12 H41040128 V1.0 Flange Bolt 4 pcs 13 H41040131 V1.0 Lock Ring 1 pcs 14 H41040132 V1.0 Brake Sleeve 1 pcs 15 H41040134 V1.0 Flange Guide Ring 1 pcs 16 H41040135 V1.0 Housing Guide Ring 2 pcs 17 H41040140 V1.0 Guide Bol 4 pcs 18 H41040160 V1.0 2 pcs 19 2000158 20 H41040189 V1.0 Jack Bolt 21 E31040175 V1.2 Jack Assembly 22 2000115 September 2006 Flange Gasket Screw, Socket Head Cap Locking Pin 3 M8x40 DIN 912 A4 LSA 4223 SS2331-06 W41040300 V3.01 1 pcs 4 pcs 1 pcs 4 pcs Page 58 of 67 Appendix 3.2.: Standard Meas. Chamber, PN16 Gate Valve 21 6 22 20 17 12 10 3 8 11 7 13 14 1 9 18 15 4 2 19 16 5 Item Part No. Description Value Device Code Qty 1 2000022 Vers. Hex Screw M8x100 DIN 931 A4 2 Unit pcs 2 2000183 Screw, Socket Head Cap M8x80 DIN 912 A4 6 pcs 3 2000097 Hex nut M16 DIN 934 A4 4 pcs 4 2450004 Douple-nipple R 1/4 Aisi316 DIN 2990 1 pcs 5 2450005 Ball valve, Reduced bore R 1/4 Aisi 316 (onninen 2017k) 1 pcs 6 26000011 Knife Gate Valve PN16, hand operated metal seat, wafer PN16 DN 80 1 pcs 7 8 2700011 2700012 O-ring O-ring 69.85x3.53 FPM 80x5 FPM 3 1 pcs pcs pcs 9 E41040326 V1.1 Meas. Chamber PN10/16 1 10 H31040129 V1.0 Mounting Flange, SS 1 pcs 11 H41040408 V1.0 Insertion Housing 67, SS 1 pcs 12 H41040128 V1.0 Flange Bolt 4 pcs 13 H41040131 V1.0 Lock Ring 1 pcs 14 H41040132 V1.0 Brake Sleeve 1 pcs 15 H41040134 V1.0 Flange Guide Ring 1 pcs 16 H41040135 V1.0 Housing Guide Ring 2 pcs 17 H41040140 V1.0 Guide Bol 4 pcs 18 H41040160 V1.0 Flange Gasket 2 pcs 19 2000158 1 pcs pcs Screw, Socket Head Cap M8x40 DIN 912 A4 20 H41040189 V1.0 Jack Bolt 4 21 E31040175 V1.2 Jack Assembly 1 pcs 22 2000115 4 pcs September 2006 Locking Pin 3 LSA 4223 SS2331-06 W41040300 V3.01 Page 59 of 67 Appendix 3.3.: PN25 Installation Parts 2 1 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 Part No. 20000158 2000175 2000176 2000177 2000178 2000179 2450004 2450005 2600017 2700011 2700012 E41040371 H31040372 H31040373 H41040132 H41040135 H41040374 H41040391 September 2006 Vers. V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 Description Screw, Socket Head Cap Hex nut Hex Screw Hex Screw Screw, Socket Head Cap Hex Screw Douple-nipple Ball valve, Reduced bore Knife Gate Valve O-ring O-ring Installation Cone PN25, DN100 Mounting Flange, PN25, DN100 Insertion Housing PN25 Brake Sleeve Housing Guide Ring Lock Ring PN25 Flange Gasket DN100 PN25 Value Device Code M8x40 DIN 412 A4 M20 DIN 985 A4 M20x45 DIN 933 A4 M20x145 DIN 931 A4 M10x70 DIN 912 A4 M10x110 DIN 931 A4 R 1/4 Aisi316 DIN 2990 R 1/4 Aisi 316 (onninen 2017k) PN25, DN80, SMO 254, DN 100 Flange 69.85x3.53 FPM 80x5 FPM W41040300 V3.01 Qty 1 6 4 6 6 2 1 1 1 3 1 1 1 1 1 2 1 2 Unit pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs Page 60 of 67 Appendix 3.4.: BTG Adapter, PN10 Gate Valve 22 6 23 18 10 13 21 3 11 15 14 7 1 X 12 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 X 9 Part No. 2000022 2000183 2000097 2450004 2450005 2600008 2700011 2700012 H31040233 H31040129 H41040408 H31040223 H41040128 H41040131 H41040132 H41040134 H41040135 H41040140 H41040160 2000158 H41040189 E31040175 2000115 2000123 September 2006 16 Vers. V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.2 19 4 Description Hex Screw Screw, Socket Head Cap Hex nut Double-nipple Ball valve, Reduced bore Knife Gate Valve PN10 O-ring O-ring Installation Flange, BTG, SS Mounting Flange, SS Insertion Housing 67, SS Flange Gasket, BTG Flange Bolt Lock Ring Brake Sleeve Flange Guide Ring Housing Guide Ring Guide Bolt Flange Gasket Screw, Socket Head Cap Jack Bolt Jack Assembly Locking Pin 3 Screw, Hex cap 8 5 17 2 17 Value Device Code Qty Unit M8x100 DIN 931 A4 2 pcs M8x80 DIN 912 A4 6 pcs M16 DIN 934 A4 4 pcs R 1/4 Aisi316 DIN 2990 1 pcs R 1/4 Aisi 316 (onninen 2017k) 1 pcs Fabri C67S316EBU ANSI 125/150-3" Aisi 316 1 pcs 69.85x3.53 FPM 3 pcs 80x5 FPM 1 pcs 1 pcs 1 pcs 1 pcs 1 pcs 4 pcs 1 pcs 1 pcs 1 pcs 2 pcs 4 pcs 2 pcs M8x40 DIN 912 A4 1 pcs 4 pcs 1 pcs LSA 4223 SS2331-06 4 pcs M8x25 DIN 933 A4 16 pcs W41040300 V3.01 20 Notes Not included in ass’y Page 61 of 67 Appendix 3.5.: BTG Adapter, PN16 Gate Valve 22 6 23 18 10 13 21 3 11 15 14 7 1 X 12 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 X 9 Part No. 2000022 2000183 2000097 2450004 2450005 26000011 2700011 2700012 H31040233 H31040129 H41040408 H31040223 H41040128 H41040131 H41040132 H41040134 H41040135 H41040140 H41040160 2000158 H41040189 E31040175 2000115 2000123 16 Vers. V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.0 V1.2 September 2006 19 4 Description Hex Screw Screw, Socket Head Cap Hex nut Double-nipple Ball valve, Reduced bore Knife Gate Valve PN16, hand operated O-ring O-ring Installation Flange, BTG, SS Mounting Flange, SS Insertion Housing 67, SS Flange Gasket, BTG Flange Bolt Lock Ring Brake Sleeve Flange Guide Ring Housing Guide Ring Guide Bolt Flange Gasket Screw, Socket Head Cap Jack Bolt Jack Assembly Locking Pin 3 Screw, Hex cap 8 17 5 2 20 17 Value Device Code M8x100 DIN 931 A4 M8x80 DIN 912 A4 M16 DIN 934 A4 R 1/4 Aisi316 DIN 2990 R 1/4 Aisi 316 (onninen 2017k) metal seat, wafer PN16 DN 80 69.85x3.53 FPM 80x5 FPM M8x40 DIN 912 A4 LSA 4223 SS2331-06 M8x25 DIN 933 A4 W41040300 V3.01 Qty 2 6 4 1 1 1 3 1 1 1 1 1 4 1 1 1 2 4 2 1 4 1 4 16 Unit pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs pcs Notes Not included in ass’y Page 62 of 67 Appendix 3.6.: Sensor Front Assembly, PN16 SS Sensor Front Assembly, PN16 SS 17 14 16 5 3 6 13 2 15 7 1 12 10 9 11 4 8 17 16 14 5 3 6 13 4 2 15 7 1 12 10 9 11 8 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 September 2006 Part No. 2000036 2000106 2000114 2000129 2650004 2650005 2650006 2650007 2700010 2700013 E31040153 H41040118 H41040119 H41040158 H41040182 H41040188 H41040244 Description Screw,Slotted cheese head Set Screw Lock Ring Screw, Slotted cheese head Bellows Seal Stationary Ring Bellows Seal Stationary Ring O-ring O-ring Main Assembly Seat Adapter Sealing Cover Protection Rubber Snap Ring Shield Cover Shield Locking Ring W41040300 V3.01 Page 63 of 67 Appendix 3.7.: Moment shaft and motor assembly Moment shaft and motor assembly 19 8 9 13 2 5 11 1 10 18 4 16 17 12 14 3 15 6 C 7 Item 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 Part No. 2000093 2000109 2000111 2000112 2100004 2100005 2600007 2650008 2750001 2750002 2750005 2800001 E31040148 E41040145 E41040146 E41040152 H41040038 H41040108 H41040154 September 2006 Description Lock Ring Scerw, Socket Head Cap Screw,Slotted cheese head Scerw, Socket Head Cap Wave Spring Extension Spring 2.0xDu12x46 Cone Clamp Radial Shaft Seal Needle Roller Bearing Thrust Ball Bearing Groove Ball Bearing Rotor Bearing Assembly Drive Disk Assy Moment Disk Assy Body Assembly Lock Cone Moment Shaft Seal Washer W41040300 V3.01 Page 64 of 67 Appendix 4: KC/5 Spare parts Kit (Recommended Spare Parts) KC/5 Spare parts Description Order Code H41040158V1.0 2650016 2750001 2700015 2650017 2650005 2650018 2650007 A41040062V1.0 A41040070V1.0 A41040202V2.20 A41080023V2.0 3100003 2700011 2700012 A41040222V1.0 2000036 2000106 2000114 2000129 2700010 H41040118V1.0 H41040119V1.0 H41040182V1.0 H41040188V1.0 H41040244V1.0 H41040154 A41040175V1.0 E41040326V1.0 2600013 H41040327V1.0 2600008 2400001 E31040322V1.1 2600014 2600015 x x x x A41040302V1.1 A41040312V1.0 A41040313V1.0 x September 2006 xx xx xx xx xx xxx xxx xxx xxx xxx xxx xxx xxx xxx xxx xxx xxx xxx xxx xxx xxx Elastic shield EPDM Rod Seal Needle Roller Bearing O-ring 19.1*1.6 FPM Front seal, Bellows Seal T502/GJ-1, SiC/EPDM Front seal, Stationary Seat C606/25, SiC/EPDM Rear seal, Bellows Seal T502/AJ-1, SiC/FPM Rear seal, Stationary Seat T6/25, SiC/FPM Optics Board Sensor Board Connection Board LC Display Board AC/DC Power Supply O-ring, Seal sleeve, 69.85x3.53 FPM O-ring, Mounting flange, 80x5 FPM Torque brake Screw, M4x10 DIN 84 A4 Set Screw M5x6 DIN 913 A4 Lock ring Screw M3x8 DIN 84 A4 O-Ring 63x2, FPM Seat Adapter Sealing Cover Snap Ring Shield Cover Shield Locking Ring Seal Washer Jack Assembly KC/5 Meas. Chamber PN16 (precut for process pipe) Blind Flange DN80 (PN25) Blind Flange DN80 (PN10 / 16) Knife gate valve AISI316 PN10 Knife gate valve AISI316 PN16 Meas. Chamber PN 10 / 16, SS, Not cutted 3"C67 Gate Valve EPDM gasket (P/N F153811) 3"C67 Gate Valve Acrylic Packing (P/N F137098) KC/5 Elastic Shield Replacement Kit KC/5 Mechanical Seal Replacement Kit xxx KC/5 Complete Spare part kit W41040300 V3.01 Page 65 of 67 Appendix 5: Model selection table Type KC5-S Description KC/-S Rotating Consistency Sensor KC/5 Display Unit with Power Supply 85 - 264 VAC, 47 - 63 Hz, 150W Sensor Wetted Parts Materials S T SS316L Titanium Sensor Pressure class 16 25 PN16 (232 psi), Standard PN16 Sensor PN25 (362 psi) PN25 Sensor, (mechanical seal rating PN25) Process connection KPM Measument Chamber V1.1 SS316L, PN10 / 16 (150 / 232 psi) KPM Measument Chamber V1.1 SS316L, PN25 (362 psi) or Flange Mounted KPM Measument Chamber V1.1 Titanium Gr2, PN10 / 16 (150 / 232 psi) KPM Measument Chamber V1.1 Titanium Gr2, PN25 (362 psi) or Flange Mounted KPM Measument Chamber V1.1 SAF2205, PN10 / 16 (150 /232 psi) KPM Measument Chamber V1.1 SAF2205, PN25 (362 psi) or Flange Mounted KPM Measument Chamber V1.1 SMO254, PN10 / 16 (150 /232 psi) KPM Measument Chamber V1.1 SMO254, PN25 (362 psi) or Flange Mounted Adapter for BTG MEK SS316L, PN10 / 16 (150 / 232 psi) Adapter for BTG MEK SS316L, PN25 (362 psi), includes insertion assy Adapter for BTG MEK Titanium, PN10 / 16 (150 / 232 psi) K1 K2 K3 K4 K5 K6 K7 K8 B1 B2 B3 Gate Valve or Flange Mounted G1 G2 G3 G4 G5 G6 G7 T1 T2 Gate Valve SS316L PN10 (150 psi) Gate Valve SS316L PN16 (232 psi) Gate Valve 254SMO PN10 (150 psi) Gate Valve SS316L PN25 (362 psi) Gate Valve 254SMO PN25 (362 psi) Gate Valve 254SMO PN16 (232 psi) Gate Valve SAF2205 PN16 (232 psi) Gate Valve Titanium PN10 (150 psi) NOTE 1 Gate Valve Titanium PN16 (2320 psi) NOTE 1 Insertion Housing Assy (Seal sleeve + Locking ring) S1 S2 S3 T1 S4 S5 S6 Insertion Housing 67 PN16, SS316L Insertion Housing 67 PN16, 254SMO Insertion Housing 67 PN16, SAF2205 Insertion Housing 67 PN16, Titanium Gr2 Insertion Housing PN25, SS316L Insertion Housing PN25, 254SMO Insertion Housing PN25, SAF2205 Installation Jack Assembly J1 J2 J0 Jack Assembly Jack Assembly PN25 No Installation Jack (Flange mounted) Seal Water System W I Low pressure seal water circulation (provided by customer) Integrated seal water tank Process pH-range E V EPDM Elastic shield and front mechanical seal (pH-range 5-14) Viton Elastic shield and front mechanical seal (pH-range 1-7) Interconnect Cable 1 2 3 Interconnect cable 10m Interconnect cable 20m Interconnect cable 30m Installation Position H V Horizontal Vertical Process Pipe Diameter for pre-cutting the chamber 0 No pre-cutting or no measurement chamber xxx Type in pipe size in inches or mm September 2006 W41040300 V3.01 Page 66 of 67 Appendix 6: Technical specifications SENSOR TYPE Rotating Consistency Transmitter OUTPUT SIGNAL Analog output 4 - 20 mA + HART®, FDT/DTM, Foundation Fieldbus and Profibus PA optional BINARY INPUTS 24 VDC (supplied from DCS), three for built-in calibration curves, one for process stop, one for sample button BINARY OUTPUT Alarm output; 24 VDC, 2A; Opening or closing dry contact POWER REQUIREMENTS Single phase, 84-264 VAC, 47-63 Hz, 2.5 Amp. 10 AT fuse required. POWER CONSUMPTION Max. 150 W MOTOR 150 W integrated Direct Drive Servo Motor MEASURING RANGE 1.5 % to 16 % consistency, full range with same sensing element SENSITIVITY Better than 0.003 % Cs PRESSURE RATING Sensor PN25 (232 psi) Gate Valve PN10 (150 psi), PN16 (232 psi) or PN25 (362 psi). PROCESS TEMPERATURE 0 - 120 °C (32 - 248 °F) AMBIENT TEMPERATURE Sensor 0 - 60 °C (32 - 140 °F), Display unit 0 - 50 °C (32 - 122 °F) FLOW VELOCITY 0 - 5 m/s (0 - 16.4 feet/s) SEAL COOLING Seal water, R1/8” internal thread (compatible with NPT R1/8”). Optional Selfcontained seal fluid system. HOUSING Sensor IP66 (better than NEMA 4X), Display IP65 (NEMA 4X) enclosure. CABLING 10 m (30’) interconnect cable from Sensor to Display Unit, 20m and 30m optional GATE VALVE DN80 (3") opening 316L standard, Duplex, SMO and Titanium available. PROCESS CONNECTION Measurement Chamber fits pipe diameters of 150 mm (6") and larger, AISI 316L standard, Duplex, SMO and Titanium optional. Adapter available for other manufacturer’s measuring vessels. All required hardware included with the transmitter. SENSOR MATERIAL Wetted parts AISI 316L or Titanium. DIMENSIONS (L*H*W) AND WEIGHTS Sensor unit: 520 x 140 x 180 mm (20.5“ x 5.5“ x 7“), 14.8 kg (32 lbs) Installation assembly: 430 x 560 x 200 mm (17 x 22 x 8”), 19 kg (42 lbs) Display Unit: 200 x 300 x 150 mm (7.9 x 11.8 x 5.9“), 6 kg (13 lbs) September 2006 W41040300 V3.01 Page 67 of 67