T121 Universal Transmitter

Transcription

T121 Universal Transmitter
IOS-T121
Universal Temperature Input,
p
2-Wire Transmitter
• Thermocouple Input (J , K, R, S, T, B, E, N)
• RTD Input: 2, 3, or 4 Wire RTD Sensors
- Lead Length Compensation (3 & 4 wire RTD Sensors)
• Potentiometer Input (Resistance)
• Millivolt Input
•
•
•
•
•
2-Wire Transmitter Output (0-20 mA, 4-20 mA)
0.1% Precision Class
16 Bit Resolution
Wide Operating Temperature Range
Easy PC Configuration
The IOS-T121 is a programmable, head mount 2-wire transmitter that features statee of the art microprocessor based design.
The microprocessor based design enables full configuration with Windows based software. The T121 accepts 2-wire, 3-wire,
or 4-wire RTD, thermocouple, potentiometer or millivolt input types. The T121 has a 4-20 mA, 2-wire transmitter output. The
T121 drops 7 V (from a typical 24 V dc powered current loop) to power its electronics and provide the linear current output.
All input range parameters are easily configurable using the Easy LP software, which is included in model: IOS-T12x SETUP
along with the S117P modem and USB cable, and a cable to connect from the modem directly to the T121.. There is no need
for a power supply, calibrator, screwdriver, jumpers or handheld programmers to configure the input range or Thermocouple
type. The S117P is powered from the PC’s USB port and provides power to the T121. All configuration parameters can be
stored in a configuration file on the PC for future use or to facilitate repeated configurations. Units already in service may have
their configuration “up-loaded” to the PC if the file is lost or misplaced.
Signal Conditioning made simple from ioSelect and powered by Seneca, the fastest growing signal conditioning company in
Europe.
General Specifications
Input
RTD
Standard
Range
RTD
Range
RTD
Range
RTD
Range
Wire Resistance
Thermocouple
Potentiometer
Millivolt
Minimum span
Ordering Information:
Accessories:
877 3 GET IOS (877.343.8467)
Fax: (800) 303-5381
Per formance Specifications
Pt100
EN 60751/A2 (ITS-90)
-200 to 650 °C
(-328 to 1202 °F)
Ni100
-60 to 650 °C
(-76 to 1202 °F)
Pt500
-200 to 650 °C
(-328 to 1202 °F)
Pt1000
-200 to 200 °C
(-328 to 392 °F)
Max 25 Ω per wire
Type J , K, R, S, T, B, E, N
0 to 1700 Ω, 100 kΩ with
parallel resistance
-150 to 150 mV
20°C (RTD), 50 °C(t/c), 2.5 mV
Response Time
Output
Power
Isolation
Load Resistance
Resloution
Fault Output
Error
EMI
Temp Coefficient
Humidity
Operating
Temperature
Storage
Temperature
Wire Size
Dimensions
Weight
<1 s
4-20 mA or 20-4 mA
7-30 V dc (5 V loop drop)
1500 VAC
850 Ω @ 24 V dc, 20 mA
~ 2 µA
105% of Full Scale
0.1 °C
< 0.5%
<100 ppm (30 ppm typical)
30-90% @ 40 °C
-40 to +85 °C
(-40 to 185 °F)
-40 to +105 °C
(-40 to 221 °F)
15 to 26 AWG
43.7 mm diameter
(1.72 in)
30 g ( 1.06 oz)
IOS-T121
Universal Input, Head Mount 2-Wire Transmitter
IOS-T12X SETUP
PC Configuration Kit (Modem, Cables & Software)
www.ioselect.com
Rev: 2/23/2011
IOS-T121
Universal Input, Head Mount 2-Wire Transmitter
Configuration
Factory Configuration
Input
Range
Filter
Output
K type TC
0-1000 °C
50/60 Hz, enabled
Normal, Out of Scale 2..5% enabled, Output high @ error, Filter excluded
The T121 is easily configured with the IOS-T12x SETUP which includes the Easy LP software (on a mini CD), the
communications modem (S117P) and cables (USB and PM002411). The Easy LP software is available on the ioSelect
website: http://www.ioselect.com/Downloads. After installing the Easy LP program on your PC, go to the Start menu in
Windows, select All Programs and look for the Seneca folder and “Easy LP”. Hold your mouse pointer on the folder and
Easy LP will appear to the right, click to open Easy LP configuration software.
USB Cable
IOS-T121
S117P
PM002411
Connect the USB cable and PM002411 to the S117P and the T121 and connect to the PC. Windows will assign a COM
port to the modem.
1. Select Next
2. Select Automatic Search (Offline is useful for a “demo” mode of the software.)
3. Select New Configuration, Load Existing Configuration or Configuration Test.
New Configuration
4. Edit Tag - A version, serial number and Tag may be entered here. Click Next to proceed.
5. Select: Sensor Type, Unit of Measure, Offset, Begin Scale (zero -0%) End Scale (span -100%) Internal Compensation
(CJC) and Sensor Break Detection (Yes or No). Click Next to proceed.
6. Select Output Mode, Scale Over range, Fault (for error indication), and Filter. Click Next to proceed)
7. Save Configuration, Send Configuration to Device or Show a report.
Done.
If there are multiple devices with the same configuration, including information from step 3 above, another unit may
be substituted and the configuration sent. The data on any screen may be changed with out affecting the other
entries so configuration of multiple units may be accomplished quickly.
877 3 GET IOS (877.343.8467)
Fax: (800) 303-5381
www.ioselect.com
Rev: 2/23/2011
IOS-T121
Universal Input, Head Mount 2-Wire Transmitter
Configuration (continued)
Load Existing Configuration
1. Configurations may be downloaded from a device and saved to a file or loaded from a file
2. Selecting the File method allows modification or to send the configuration as is.
Configuration Test
A Test function is included with the software. The test section also provides information about the sensor status and allows
reading the module’s status, the values of the input and output. It also allows the user to set the module into manual mode
(clicking on Automatic Manual button) and to force the value of the analog output to compensate for any offset.
Once the parameters are set they need to be downloaded into the T121. The T121 is then ready to provide years of service
with little or no maintenance.
Connections
Push Wire Connectors
2-wire RTD Sensors - Best used for short distances between
module and probe (<10 m or 33 ft).
push wire to insert
push button and
pull wire to remove
3-wire RTD Sensors - Best for general purpose. Lead length
compensation assumes all 3 wire lengths to sensor are equal.
4-wire RTD Sensors - The most accurate RTD sensor. Lead length
compensation assumes all 4 wire lengths to sensor are equal.
Potentiometer input - A pot with resistance between 500 and 1700
Ω can be connected directly to the input. For pots greater than 1.7
kΩ a parallel resistance of 1.5 kΩ is required.
-
DC Power Supply
(5-30 VDC)
+
Dimensions
+
I
I out
I out
+
5
44 mm
33 mm
(1.3 “)
Series device,
display or PLC, etc.
-
-
6
(1.73 “)
4
3
2
1
4-wire
RTD
3-wire
RTD
4
3
4
3
2
1
1
TC & mV
16.6 mm
(0.65”)
Programming
Connector
4
3
2
4
3
2-wire
RTD
-
Short:
1 to 2
& 3 to 4
@
transmitter
1
Pot
≤1.7 kΩ
Pot > 1.7 kΩ
(remove Ra for < 1.7 kΩ)
4
3
4
3
2
2
2
1
1
1
+
Ra = 1.5 kΩ
20.0 mm
(0.79”)
EN61000-6-4/2002-10
EN61000-6-2/2002-10
ioSelect, Incorporated
877 3 GET IOS (877.343.8467)
Fax: (800) 303-5381
P: 858-537-2060, F: 800-303-5381, [email protected]
www.ioselect.com
rev: 7/09