Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets

Transcription

Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion and Disengagement Instruments
Technique Guide
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Recommended Archwire Insertion Instruments
Single point torquing key for tight spots
Double-end torquing key end for rectangular wires
SmartClip™ Appliance Wire Insertion
Hand Instrument
REF 804-152 .022 Bracket Instrument
REF 804-053 .018 Bracket Instrument
2
© 2009 3M. All Rights Reserved.
Weingart utility pliers can be
used for feeding the wire
through traditional tubes
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion
Using the SmartClip™ Appliance Wire Insertion Instrument
ƒ Double torquing key end
Torque to
match slot
Tip:
Consider using fingers instead of
instruments on round and smaller
rectangular wires. It may be easier to
feel the force levels and the proper
seating of the wire.
1. Align archwire over clip opening and bracket slot.
2. Position instrument on archwire so the torquing key straddles the bracket.
3. Torque rectangular wires by rolling the instrument until the wire torque matches the slot
torque, which will allow the wire to naturally fall into place with little force.
4. Guide the torqued wire into the bracket slot gently while providing lingual support to the
teeth.
3
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion
Using the SmartClip™ Appliance Wire Insertion Instrument
ƒ Single torquing key end
Torque to
match slot
1. Align archwire over clip opening and bracket slot.
2. Position instrument on archwire on either mesial or distal side of bracket. Avoid applying
pressure at the midpoint of the inter-bracket distance. Instead try to aim the single point as
close to the rotation arm as possible without coming into contact with it.
3. Torque rectangular wires by rolling the instrument until the wire torque matches the slot
torque, which will allow the wire to naturally fall into place with little force.
4. Guide the torqued wire into the bracket slot gently while providing lingual support to the
teeth.
4
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion Procedural Tips
5
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Procedural Tips
ƒ
3M self-ligating brackets are designed to make wire insertion and
disengagement quick and consistent.
ƒ
However, the incorrect technique could result in the following:
1. Debonded brackets
2. Broken rotation arms
3. Deformed archwires
4. Patient discomfort
5. Wire will not insert or disengage
ƒ
The following slides will demonstrate the proper techniques to help
avoid the aforementioned problems.
6
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion: Recommended Sequence
When inserting archwires it is important to follow the recommended insertion sequence.
ƒ Recommended Sequence
1) Engage molar brackets or buccal tubes
2) Align the midline, but do not engage centrals
3) Then work from posterior to anterior
ƒ Incorrect Sequence
ƒ
Anterior to posterior
ƒ Potential Consequences
ƒ
Deformed Archwires
ƒ
Patient Discomfort
7
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion: Recommended Sequence
When inserting archwires it is important to follow the recommended sequence.
Step 3: Insert
from posterior
to anterior
Step 2: Align
Midline but
Don’t Insert
Step 1: Insert
Molars
8
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion: Lingual Support
When inserting archwires it is important to provide lingual support to optimize patient comfort, or
have the patient bite down on a cotton roll.
ƒ Correct Technique
ƒ
Regardless of insertion technique
used, it is important to provide
lingual support
ƒ Incorrect Technique
ƒ
No use of lingual support
ƒ Potential Consequences
ƒ
Patient discomfort
9
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion: Slot Alignment
Before inserting large rectangular archwires it is important to make sure that the slots are
aligned vertically.
ƒ Correct Technique
ƒ
Make sure slots are aligned before
moving to large, rectangular wires
ƒ Incorrect Technique
ƒ
Not leveling the teeth before
inserting large wires
ƒ Potential Consequences
ƒ
Patient discomfort
ƒ
Wire will not engage
10
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion: Instrument Positioning
Regardless of the insertion technique used, it is important to position the instrument
correctly when inserting the archwire.
ƒ Correct Positioning
ƒ
Insertion instrument should
straddle the mesial and distal
aspects of the bracket
ƒ Incorrect Positioning
ƒ
Insertion instrument hits bracket,
not wire
ƒ Potential Consequences
ƒ
Broken Rotation Arms
11
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Insertion: Archwire Torque
When inserting rectangular archwires torque the wire to match the torque of the bracket. Most
preprogrammed brackets have torque built-in. For example, an MBT™ Appliance System
central bracket has 17 degrees of torque.
ƒ Correct Technique
ƒ
Gently torque large rectangular
wires to match the torque and
insertion path of the bracket
ƒ Incorrect Technique
ƒ
Not adding torque to the wire to
match torque and insertion path
of the bracket
ƒ Potential Consequences
ƒ
Patient discomfort
ƒ
Wire will not insert
12
© 2009 3M. All Rights Reserved.
Torque
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Sensitive Patients
ƒ For sensitive patients, have them bite down on a
cotton roll to provide occlusal support
13
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Disengagement
Instructions for Use
14
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Disengagement Instrument
Using the SmartClip™ Appliance Wire Disengagement Instrument REF 804-160
ƒ
Routinely check the disengagement instrument hook
ƒ
If there is significant “play” in the hook, tighten the screw that
holds the hook mechanism in place before you proceed with
the archwire disengagement
Note: Tighten hook screw after each sterilization
15
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Prepare for Archwire Disengagement
ƒ Remove any bendbacks distal
to the molars
ƒ For sensitive patients use a
cotton roll
ƒ
ƒ
Have the patient bite down on
the cotton roll while using the
disengagement instrument
Biting on the cotton roll provides
occlusal support to the teeth
16
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Disengagement
Using the SmartClip™ Appliance Wire Disengagement Instrument
2
1
3
4
1. Approach archwire with disengagement instrument from occlusal or gingival side
(it is recommended to work from the occlusal side whenever possible).
2. Place instrument hooks under archwire keeping the bracket between the
instrument hooks, avoiding the mesial and distal protrusions and tie-wings.
3. Squeeze handles (Note: It is not necessary to fully squeeze the instrument. Only
squeeze until the wire releases from the clips.)
Follow the recommended disengagement sequence in the following slides to help avoid archwire deformation.
17
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Disengagement Procedural Tips
18
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Archwire Disengagement: Recommended Sequence
The recommended disengagement sequence is designed to optimize
patient comfort and provide a dependable procedure.
ƒ Tactics to optimize patient comfort:
ƒ Lengthen inter-bracket distance
ƒ Reduce amount of deflection by utilizing inter-bracket
distance
ƒ Distribute force to teeth best equipped to handle it
19
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Disengagement Sequence
It is important to follow the recommended sequence when disengaging the archwire.
ƒ Recommended Sequence:
1) Upper right lateral
2) Upper left lateral
3) Upper right central
4) Upper left central
5) Upper cuspids
6) Upper bicuspids
ƒ Incorrect Sequence
ƒ It may be possible to slide
out the wire from the posterior
after disengaging the anterior.
ƒ Adjacent brackets might
release when disengaging
large stainless steel wires.
ƒ It is viable to start with
any tooth mesial of the
bicuspids, as long as
lingual support is provided.
ƒ
Starting in the posterior (i.e.. 4’s, 5’s, and 6’s)
ƒ
Moving sequentially from tooth to tooth along the arch
ƒ Potential Consequences
20
Please Note:
ƒ
Patient discomfort
ƒ
Deformed archwire
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Disengagement Sequence
Step 2:
Disengage
Lateral
Step 3:
Disengage
Central
Step 1:
Disengage
Lateral
Step 4:
Disengage
Central
Step 5:
Disengage
Cuspids
Step 6:
Disengage 1st
Bicuspids
21
© 2009 3M. All Rights Reserved.
Step 7: Slide out
Archwire
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Disengagement: Lingual Support
It is important to provide lingual support when disengaging the archwire, or have the patient bite
down on a cotton roll.
ƒ Correct Technique
ƒ
Use lingual support as you
disengage
ƒ Incorrect Technique
ƒ
No lingual support
ƒ Potential Consequences
ƒ
Patient discomfort
22
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Disengagement: Instrument Approach
It is important to use a vertical approach when seating the disengagement instrument on the
bracket and the wire.
ƒ Correct Technique
ƒ
Vertical approach (occlusal or gingival side)
ƒ
Getting the hook to straddle the bracket and rotation arm
ƒ
Avoid the mesial and distal protrusions and tie-wings
ƒ Incorrect Technique
ƒ
Non-vertical approach
ƒ
Poorly seated disengagement
instrument on the wire
ƒ Potential Consequences
ƒ
Broken rotation arms
ƒ
Debonded brackets
ƒ
Patient discomfort
ƒ
Wire will not disengage
23
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Disengagement: Instrument Hook Position
It is important to place the hook of the disengagement instrument precisely on the wire to avoid
contact with the rotation arms and tie-wings.
ƒ Correct Technique
ƒ
Disengagement instrument hooks must straddle
the bracket and rotation arm, avoiding the
mesial and distal protrusions and tie-wings
ƒ Incorrect Technique
ƒ
Disengagement hook under rotation arm
ƒ
Disengagement hook under tie-wing
ƒ Potential Consequences
24
ƒ
Broken rotation arms
ƒ
Broken tie-wings
ƒ
Debonded brackets
ƒ
Patient discomfort
© 2009 3M. All Rights Reserved.
Clarity™ SL and SmartClip™ SL3 Self-Ligating Brackets
Disengagement: Light Round Wires
Light round wires may not disengage with the disengagement instrument.
ƒ Correct Technique
ƒ
ƒ
Use the disengagement instrument as directed
If the round wire does not disengage with the instrument, cut the
wire at the midline and slide it out
ƒ Incorrect Technique
ƒ
Using the disengagement instrument to “pull-out” light round wires
ƒ Potential Consequences
ƒ
Wire will not disengage
ƒ
Patient discomfort
ƒ
Deformed archwire
25
© 2009 3M. All Rights Reserved.