Instruction Manual - Gilson Engineering Sales

Transcription

Instruction Manual - Gilson Engineering Sales
Hydra SX40
INDUSTRY APPROVALS
The Polysonics, Hydra SX430 is approved by CSA and CE
INSTRUCTION MANUAL
HYDRA SX30
PORTABLE DUAL FREQUENCY DOPPLER FLOWMETER
JANUARY 2000
REV. D
NOTICE
Read this manual before working with the product. For personal and system
safety, and for optimum product performance, make sure you thoroughly
understand the contents before installing, using, or maintaining this instrument.
For equipment service needs outside the United States, contact your nearest
Polysonics representative.
Within the United States, the Polysonics Service Center is your single-point
contact for all Polysonics equipment service needs. If at any time you are not
sure what to do, you have a question about using the product, or you have a
service or support request, call the Center at (281) 879-3700. This contact is
your fastest link to quick and complete answers about any Polysonics product
or service.
Polysonics, Inc.
10335 Landsbury Dr., Suite 300
Houston, TX 77099-3407 USA
Telephone: (281) 879-3700
Fax:
(281) 498-7721
Web:
www.polysonics.com
C
Polysonics, Inc.
Hydra SX30
Table of Contents
1. Guidelines and Checklists .................................................. 1
1.1 Guidelines .......................................................................................... 1
1.1.1 General Application Guidelines ................................................................... 1
1.2 Checklist ............................................................................................ 1
1.2.1 Standard Equipment Checklist .................................................................... 1
2. Basic Installation & Operation Procedures ....................... 3
2.1 Install the meter. ................................................................................ 3
2.1.1
2.1.2
2.1.3
2.1.4
Select an installation location. ..................................................................... 3
Choose a transducer orientation. ................................................................ 5
Attach the transducers to the pipe strap. .................................................... 6
Mount the transducers on the pipe. ............................................................ 7
2.2 Measure flow...................................................................................... 9
2.2.1 Obtain a flow reading. ................................................................................. 9
3. Hydra SX30 Features
Installation and Operation Procedures. .......................... 12
3.1 Power source procedures. ............................................................. 12
3.1.1 Recharge the battery. ................................................................................ 12
3.1.2 Connect an auxiliary DC power source. .................................................... 12
3.2 Screen Contrast............................................................................... 13
3.2.1 Change the screen contrast. ..................................................................... 13
3.3 Trend ................................................................................................ 13
3.3.1
3.4.1
3.3.2
3.4.3
View the flow trend. ................................................................................... 13
Set the totalizer units. ................................................................................ 14
Change the totalizer multiplier. .................................................................. 14
Reset the totalizer. ..................................................................................... 14
3.4 Totalizer............................................................................................ 14
3.5 The 4-20 mA Loop ........................................................................... 15
3.5.2 Range the 4-20 mA loop. .......................................................................... 15
3.5.3 Calibrate the 4-20 mA loop........................................................................ 17
3.6 Datalogger ....................................................................................... 18
3.6.1
3.6.2
3.6.3
3.6.4
3.6.5
3.6.6
3.6.7
3.6.8
Set the data logger time and date. ............................................................ 18
Set up and start the data logger. ............................................................... 19
Download data from the logger. ............................................................... 21
Zoom in on data......................................................................................... 22
Zoom in on a selection of data. ................................................................. 22
Save the log files. ...................................................................................... 23
Retrieve a saved file. ................................................................................. 24
Erase all data in the log files. .................................................................... 25
4. Maintenance ....................................................................... 26
4.1 Flowmeter Maintenance .................................................................. 26
4.3 Upgrade Meter Software ................................................................. 27
5. Troubleshooting ................................................................. 28
5.1 Different Modes of Operation ......................................................... 28
5.2 Learn Mode Warnings ..................................................................... 28
5.3 Operating Mode Warnings .............................................................. 29
5.4 Service.............................................................................................. 31
5.4.1 Contacting the Polysonics Service Center ................................................ 31
5.4.2 Factory Service .......................................................................................... 32
6. Troubleshooting Procedures ............................................ 33
6.1 Check the FFT ................................................................................. 33
6.2 Set the low flow cutoff. ................................................................... 36
6.3 Perform an on-site calibration. ...................................................... 36
6.4 Reset the factory settings. ............................................................. 37
6.5 Set the damping parameter. ........................................................... 37
7. Appendix ............................................................................ 38
7.1 Hydra SX30 Manual Symbol Descriptions .................................... 38
7.2 Pipe ID .............................................................................................. 38
7.2.1 Obtain the Pipe ID. .................................................................................... 38
7.2.2 Pipe ID Charts ........................................................................................... 39
7.3 Application Signal Quality and Signal Strength ........................... 42
7.4 Hydra SX30 Meter Specifications .................................................. 43
7.5 Hydra SX 30 menu tree ................................................................... 44
1. Guidelines and Chec
klists
Checklists
1.1 Guidelines
1.1.1 General Application Guidelines
The Hydra SX30 Dual Frequency Doppler (DFD) Portable Flowmeter is suited for
measuring the flow of fluids with undissolved, entrained solids or dispersed gases
such as:
Wastewater
• primary sludge
• activated sludge
Food
• dairy
• fluidized foods
Industrial
• crude oil
• dredging
1.2 Chec
klist
Checklist
1.2.1 Standard Equipment Checklist
The following equipment is included with your new Hydra SX30 DFD Flowmeter.
•
•
•
•
•
•
•
•
•
•
•
•
(1) Hydra SX30 Portable DFD Flowmeter
(1) battery, fully charged
(1) carrying case
(2) non-submersible DFD Transducers
(2) 32 inch pipe straps
(1) 67 inch pipe strap
(1) 120 volt AC battery recharger
(2) 16 foot cables with BNC connectors
(2) 3.5" disk containing HydraLink datalog software
(1) tube of sonic coupling compound
(1) RS232 Cable
(1) 3/8 inch Hex socket driver
1
A
A
B
C
D
YES
E
FIGURE 2.1A – RECOMMENDED INSTALLATION LOCATIONS
Do not install the transducers at the following locations:
A – Pipes may not be full
B – Down flow
C – Too close to the elbow
D – Air collects at the top of horizontal pipe
E – Sediment collects at the bottom of horizontal pipe
Conditions existing at these locations can interfere with the
transmission of the ultrasonic wave and yield
inaccurate or unreliable flow readings.
2
2. Basic Installation & Operation Pr
ocedures
Procedures
2.1 Install the meter
meter..
2.1.1 Select an installation location.
Selecting the proper location to install the transducers is essential to flow
measurement reliability. Each application is unique and may require
variations in the installation location.
NOTE
1.
2.
Review Figure 2.1a (Recommended Installation Location), on the facing page.
Select an installation location using Figure 2.1a and the following guidelines:
•
•
•
full pipe
flow in the pipe
5 pipe diameters upstream and 3 pipe diameters downstream from:
- a direction change on the pipe
- pipe joints
- a narrowing or widening of the pipe
3
HORIZONTAL PIPES
Best: - 3 o’clock, 9 o’clock
12
Good: - 2 o’clock, 4 o’clock
- 10 o’clock, 8 o’clock
9
OK:
3
- 3 o’clock, 4 o’clock
- 9 o’clock, 8 o’clock
6
FIGURE 2.1B – TRANSDUCER ORIENTATIONS ON HORIZONTAL PIPE
VERTICAL PIPES
Best: - 3 o’clock, 9 o’clock
- another directly opposite position
12
3
9
6
Good: - 2 o’clock, 4 o’clock
- 10 o’clock, 8 o’clock
OK:
- 2 o’clock, 3 o’clock
- 10 o’clock, 9 o’clock
FIGURE 2.1C – TRANSDUCER ORIENTATIONS ON VERTICAL PIPE
4
2.1.2 Choose a transducer orientation.
Choosing the proper transducer orientation is essential to flow
measurement reliability. Each application is unique. You may need to
vary the orientation with each application, to achieve optimum meter
performance.
NOTE
1.
2.
Review Figures 2.1b,c (Transducer Orientations on Horizontal and Vertical
Pipe), on the facing page.
Meet the following guidelines to ensure optimum meter performance:
•
•
•
•
•
3.
3 o’clock and 9’oclock orientation
transducers parallel to one another along the pipe circumference
generous amounts of coupling compound
secure cable connections
tight pipe strap
IF your application does not enable you to mount the transducers at the
recommended “Best” orientation, THEN choose a “Good” or “OK” transducer
orientation OR choose a different location to install the transducers.
5
2.1.3 Attach the transducers to the pipe strap.
1.
2.
3.
4.
5.
6.
See Figure 2.1d - Attach Transducers to the Pipe Strap, below.
Loosen the mounting screw on each transducer using a screwdriver.
Slide the lock back on each transducer.
Place the pipestrap across the mounting.
Slide the lock back in place.
Tighten the screw.
1
2
3
4
5
FIGURE 2.1D – ATTACH TRANSDUCERS TO THE PIPE STRAP
6
2.1.4 Mount the transducers on the pipe.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
NOTE
Wrap the pipe strap around the pipe.
Slide the pipe strap end through the tension nut assembly.
Pull the strap tight AND clamp the tension nut down to secure the pipe strap to
the pipe.
Loosen the mounting screw of one transducer.
Move the transducer at the 3 o’clock position on the pipe. (See Figures 2.1b,c
Transducer Orientations on Horizontal and Vertical Pipe, page 4)
Tighten the mounting screw to secure the transducer in place.
Repeat steps 4-6 for the second transducer, now at the 9 o’clock position.
Remove the pipe strap from the pipe AND generously apply sonic coupling
compound to each transducer surface.
Put the pipe strap on the pipe, placing the transducers in the proper position.
Tighten the tension nut assembly with a 3/8” hex nut driver securing the pipe
strap.
Increasing the pipe strap tension improves signal transmission.
7
FIGURE 2.2A - POLYSONICS HYDRA SX SCREEN
FIGURE 2.2B - PIPE ID SCREEN
FIGURE 2.2C - FLOW UNITS SCREEN
8
2.2 Measure flo
w
flow
2.2.1 Obtain a flow reading.
NOTE
1.
2.
You must know the pipe internal diameter to ensure reliable flow
measurements AND there must be flow in the pipe when you start the
meter.
IF you do not know the pipe ID, THEN go to section 7.2 Pipe ID on page 38.
Press ON to start the meter.
You are now in Start Up Mode
3.
4.
5.
6.
NOTE
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
Wait for the Polysonics Hydra SX scren to appear. (See Figure 2.2.a-Hydra SX
Screen, facing page)
Check the battery life gauge when the Polysonics Hydra SX screen appears
(appears for 5 seconds).
IF the battery life is less than the anticipated hours of use, THEN see
procedures 3.1.1-Recharge the battery, OR 3.1.2-Connect an auxiliary DC
power source, on page 12.
Wait for the Pipe ID screen to appear. (See Figure 2.2b-Pipe ID Screen, facing
page.)
Use the BACK-SPACE key, to the right of the number keys, for
correcting errors when entering numbers.
Press SCROLL until the cursor moves to the correct pipe ID units.
Press ENTER.
Enter the Pipe ID using the number keys.
Press ENTER
Press the Menu ---> key to the left of Next.
Wait for the Flow Units screen to appear. (See Figure 2.2c - Flow Units
Screen, facing page)
Press SCROLL until the cursor moves to the correct volume units.
Press ENTER.
Press SCROLL until the cursor moves to the correct time base units.
Press ENTER.
Press the Menu ---> key to the left of Next.
9
FIGURE 2.2D - LEARN MODE SCREEN
FIGURE 2.2E – FLOW SCREEN
10
(procedure 2.2.1 Obtain a flow reading continued.........)
You are now in Learn Mode.
18.
NOTE
19.
20.
Wait for the Learn Mode screen to appear. (See Figure 2.2d - Learn Mode
Screen, facing page)
IF you are operating the meter without installing it, THEN press the
ENTER key four times to bring up the flow screen.
Review the diagnostic information that appears in the Learn Mode screen.
(appears for 60 seconds)
IF any of the following warnings appear:
•
Invalid Signal, Can’t Learn
•
No Flow, Can’t Learn
•
Warning Low S Strength
•
Warning Poor S Strength
THEN see section 5.2 (Learn Mode Warning), on page 28.
You are now in Operating Mode
21.
22.
23.
24.
Wait for the Flow screen to appear. (See Figure 2.2e - Flow Screen , facing
page)
Read the flow.
IF the flow reading is not accurate, AND you can verify the flow from a
calibrated flowmeter or other certified source, THEN see procedure
6.3 - Perform an on-site calibration, on page 36.
IF any of the following warnings appear:
•
Warning
•
Alarm
•
Failed/No Flow
THEN see section 5.3 - Operating Mode Warnings, on page 29.
11
3. Hydra SX30 Features
Installation and Operation Pr
ocedures.
Procedures.
3.1 P
ower sour
ce pr
ocedures.
Po
source
procedures.
3.1.1 Recharge the battery.
NOTE
You can recharge the battery while you are operating the meter, even
under low battery conditions.
Overcharging will degrade the battery. Do not charge the batter for more
than eight hours. The battery is fully charged at the factory.
1.
2.
3.
IF you need to run the meter at one location for more than 12 hours, THEN
use an auxiliary DC power source. (See procedure 3.1.2 Connect an auxiliary
DC power source, below)
Plug the power supply cable into the CHARGE socket on the meter face, to
the right of the display window.
Charge a depleted battery for 8 hours to fully recharge.
3.1.2 Connect an auxiliary DC power source.
NOTE
1.
2.
The battery will not charge while the meter is operating from an auxiliary
DC power source.
Use a minimum 10.5-volt power supply that supplies 0.25 amps to power the
flowmeter.
Plug the power supply cable into the appropriate socket on the side of the
meter.
DO NOT use power supplies greater than 18-volts, to avoid damaging
the flometer.
12
3.2 Screen Contrast
3.2.1 Change the screen contrast.
1.
2.
3.
4.
5.
6.
Press the +/- key on the number key pad.
Wait for the adjust contrast bar to appear. (See Figure 3.2 - Adjust Contrast
Bar, below)
Press the +/- key on the number key pad to start the contrast bar scrolling.
WHEN the screen is the desired contrast, THEN press the +/- key once more,
to stop the scrolling.
Click the SCROLL and BACK-SPACE keys to fine adjust the contrast.
Press the Menu ---> key to the left of Flow to return to the Flow screen.
FIGURE 3.2 - ADJUST CONTRAST BAR
3.3 Trend
3.3.1 View the flow trend.
NOTE
1.
2.
3.
Trend runs at real time rate of action. To change the update rate turn on
the datalogger and set the log rate. (See procedure 3.6.2 - Set up and
start the datalogger, page 19)
Press the Menu ---> key to the left of Trend in the Flow Screen.
Wait for the Trend screen to appear.
Press the Menu ---> key to the left of Flow to return to the Flow screen.
13
3.4 Totaliz
er
otalizer
3.4.1 Set the totalizer units.
1.
2.
3.
4.
5.
6.
7.
Press the Menu ---> key to the left of Diags in the Flow screen.
Wait for the Limits screen to appear.
Press the Menu ---> key to the left of Total.
Wait for the Total screen to appear. (See Figure 3.4 - Totalizer Screen, below)
Press SCROLL until the cursor moves to the correct volume units.
Press ENTER.
Press the Menu ---> key to the left of Flow to return to the Flow Screen.
3.3.2 Change the totalizer multiplier.
1.
2.
3.
4.
5.
6.
7.
8.
Press the Menu ---> key to the left of Diags in the Flow screen.
Wait for the Limits screen to appear.
Press the Menu ---> key to the left of Total.
Wait for the Total screen to appear. (See Figure 3.4 - Totalizer Screen, below)
Press ENTER until the cursor moves to the multiplier box.
Press SCROLL until the desired multiplier appears in the box.
Press ENTER, to set the multiplier.
Press the Menu ---> key to the left of Flow to return to the Flow screen.
3.4.3 Reset the totalizer.
1.
2.
3.
4.
5.
6.
7.
8.
9.
Press the Menu ---> key to the left of Diags in the Flow screen.
Wait for the Limits screen to appear.
Press the Menu ---> key to the left of Total.
Wait for the Total screen to appear. (See Figure 3.4 - Totalizer Screen, below)
Press ENTER twice, to move the cursor to “Reset?”.
Press the SCROLL key until “Reset?” reads “Yes”.
Press ENTER.
Wait while the totalizer resets.
Press the Menu ---> key to the left of Flow to return to the Flow Screen.
FIGURE 3.4 - TOTALIZER SCREEN
14
3.5 The 4-20 mA Loop
3.5.1 Range the 4-20 mA loop.
1.
2.
NOTE
3.
4.
5.
6.
7.
Press the Menu ---> key to the left of 4-20 mA in the Flow screen.
Wait for the 4-20 mA Range screen to appear. (See Figure 3.5a - 4-20 mA
Range Screen, below)
Use the BACK-SPACE key, to the right of the number keys, for correcting
errors when entering numbers.
Enter the minimum flow using the number keys.
Press ENTER.
Enter the maximum flow using the number keys.
Press ENTER.
Press the Menu ---> key to the left of Flow to return to the Flow screen.
FIGURE 3.5A – 4-20 MA RANGE SCREEN
15
Current Loop Terminals
Black
-
Red
+
Remote Device
or
Current Meter
-
+
Battery or
Power Supply
FIGURE 3.5B– CURRENT LOOP
FIGURE 3.5C - 4-20 MA CALIBRATION SCREEN
FIGURE 3.5D - 4-20 MA CALIBRATION TEST SCREEN
16
3.5.2 Calibrate the 4-20 mA loop.
NOTE
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
You need a calibrated current meter to perform this procedure. In the
loop powered configuration, you will also need a DC power source.
Set up a calibration current loop. (See Figure 3.5b - Current Loop, facing
page)
IF the meter is not on a pipe with flow, THEN press the Menu ---> key to the
left of Next two times, AND press the ENTER key four times to access the Flow
screen.
Press the Menu ---> key to the left of 4-20 mA in the Flow Screen.
Wait for the Range screen to appear. (See Figure 3.5a - 4-20 mA Range
Screen, page 15)
Press the Menu ---> key to the left of Calib.
Wait for the Calibration screen to appear. (See Figure 3.5c - 4-20 mA
Calibration Screen, facing page)
Enter the value, in mA, from the current meter that corresponds to the zero
scale, using the number keys.
Press ENTER
Enter the value, in mA, from the current meter that corresponds to the full
scale, using the number keys.
Press ENTER.
Wait for the Test screen to appear. (See Figure 3.5d - 4-20 mA Calibration Test
Screen, facing page)
Press the SCROLL key or the BACK-SPACE key to move the current bar in the
Test screen.
Verify that the flowmeter follows the changes.
IF the flowmeter does not follow the changes, THEN press the Menu ---> key
to the left of Calib AND repeat steps 6-13.
IF the flowmeter follows the changes, THEN press the Menu ---> key to the left
of Flow to return to the Flow screen.
17
3.6 Datalog
ger
Datalogg
3.6.1 Set the data logger time and date.
1.
2.
3.
4.
5.
6.
7.
8.
NOTE
9.
10.
11.
12.
13.
14.
15.
Press the Menu ---> key to the left of Log in the Flow screen.
Wait for the Log screen to appear. (See Figure 3.6b - Log Screen, facing page)
Press the Menu ---> key to the left of Time.
Wait for the Time screen to appear. (See Figure 3.6a - Time Screen, below)
Type the correct hour using the number keys.
Press ENTER.
Type the correct minutes using the number keys.
Press ENTER.
Use the BACK-SPACE key, to the right of the number keys, to correct
errors when entering numbers.
Type in the day using the number keys.
Press ENTER.
Press SCROLL until the correct month appears.
Press ENTER.
Type in the correct year using the number keys.
Press ENTER.
Press the Menu ---> key to the left of Flow to return to the Flow screen OR Log
to go to the Log screen
FIGURE 3.6A - TIME SCREEN
18
3.6.2 Set up and start the data logger.
NOTE
1.
2.
3.
4.
5.
6.
7.
8.
You cannot select the file to store a specific log set. The flowmeter
saves data in sequenced files 0 to 9. The flowmeter sequences to the
next file each time the logger is turned on or a file logs more than 9000
points.
See procedure 3.6.1 (Set the data logger time and date) on page 18, to set the
time and date.
Press the Menu ---> key to the left of Log in the Flow screen.
Wait for the Log screen to appear. (See Figure 3.6b - Log Screen, below)
Press SCROLL until the cursor moves to the desired time between points.
(0.5 - 60 min/point)
Press ENTER.
Press SCROLL until the cursor moves to ON, next to Log, at the bottom of the
screen.
Press ENTER
Press the Menu ---> key to the left of Flow to return to the Flow Screen.
FIGURE 3.6B - LOG SCREEN
19
FIGURE 3.6C- COMM PORT PROPERTIES
FIGURE 3.6D - SELECT LOG FILE
20
3.6.3 Download data from the logger.
NOTE
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
NOTE
The following procedure assumes:
- HydraLink software is installed
- You have basic operating knowledge of computers and Windows
Connect the communication cable from the flowmeter to the RS232 port on
the computer.
Start the HydraLink program.
Wait for the “Communications Established” window to appear.
Click on the “OK” button.
IF the “Communication Established window appears, THEN go the step 11.
IF the “Communications Failed” window appears, THEN check the security of
the cable, AND make sure the flowmeter is on.
Click on OK.
WHEN the CommPort Properties window appears, THEN make sure the
correct comm port number appears in the Port option, AND enter the correct
CommPort properties, show in Figure 3.6c - CommPort Properties, on the
facing page.
Wait for the “Communication Established” window to appear.
IF the Communication Established Window still does not appear, THEN
contact your Polysonics representative.
WHEN the HydraLink window appears, THEN click on the Log File 0 button.
(See Figure 3.6d - Select Log File, facing page)
Select the log file you want to download.
Click on the Get Log button in the upper right corner of the HydraLink window.
Wait for HydraLink to complete downloading.
The logger saves data in sequenced files 0 through 9. Once all the files
are full the logger restarts the sequence and overwrites the files
beginning with log file 0. Perform procedure 3.6.6 (Save the log files), on
page 23, to ensure data is not lost.
21
3.6.4 Zoom in on data.
1.
2.
3.
Click on the Zoom button. (See Figure 3.6e - Zoom Button, below)
Press the arrow buttons at the bottom of the data graph to view data.
Return to the original view by clicking on the Zoom button again.
3.6.5 Zoom in on a selection of data.
1.
2.
3.
4.
5.
6.
Move the arrow to a point on the screen.
Depress the mouse button.
Drag the mouse to enclose a range of data you want to zoom in on. (See
Figure 3.6f - Data Zoom, below)
Release the mouse button.
Repeat steps 1-4 until you achieve your desired zoom view.
Return to the previous view by double clicking on the Zoom button. (See
Figure 3.6e - Zoom Button, below)
FIGURE 3.6E - ZOOM BUTTON
FIGURE 3.6F - DATA ZOOM
22
3.6.6 Save the log files.
NOTE
1.
2.
3.
4.
NOTE
Saved datalog files are delimited by commas, as it applies to opening
the files into a spread sheet programs.
Perform procedure 3.6.3 (Download data from the logger), on page 21.
Click on File in the HydraLink window. (See Figure 3.6g - Save Log, below)
Select “Save Log”.
Use standard Windows procedures to save the log file on your computer or
floppy disk.
The logger saves data in sequenced files 0 through 9. When all the files
are full the logger restarts the sequence and overwrites the files
beginning with log file 0.
FIGURE 3.6G SAVE LOG
23
3.6.7 Retrieve a saved file.
NOTE
1.
2.
3.
NOTE
If the meter is not connected to your PC, then click on OK in the
“Communications Failed” AND Cancel in the CommPort Properties”
windows.
Click on File in the HydraLink window. (See Figure 3.6h - Load Log, below)
Select “Load Log”.
Use standard Windows procedures to retrieve the log file.
The logger saves data in sequenced files 0 through 9. When all the files
are full the logger restarts the sequence and overwrites the files
beginning with log file 0.
FIGURE 3.6H - LOAD LOG
24
3.6.8 Erase all data in the log files.
Performing the following procedure erases ALL the log information
stored in the flowmeter. You cannot erase a specific log file.
NOTE
1.
2.
3.
4.
It is not necessary to create space in log files for new log information by
erasing data. The logger saves data in sequenced files 0 through 9.
When all the files are full the logger restarts the sequence and
overwrites the files beginning with log file 0. Perform procedure 3.6.6
(Save the log files), on page 23to ensure data is not lost.
Click on Log in the HydraLink window. (See Figure 3.6i - Erase All, below)
Select “Erase All” from the pull down menu.
IF you are sure you want to erase all 10 log files, THEN click on the Yes button
in the Erase All window, AND wait while the logger erases all the files.
IF you do not want to erase all 10 log files THEN click on the No button in the
Erase All window, AND click on OK in the Erase Aborted Window.
FIGURE 3.6I – ERASE ALL
25
4. Maintenance
4.1 Flo
wmeter Maintenance
Flowmeter
The Hydra SX30 Portable Flowmeter is easy to maintain. The DFD transducers
and flowmeter are factory service only components and maintenance free. The
following Flowmeter Maintenance table describes:
•
•
•
system components
appropriate maintenance actions
a recommended maintenance schedule
TABLE 4.1 –COMPONENT MAINTENANCE AND SCHEDULE
Component
Transducers
Recommended Maintenance
! none, this is a service item
How Often
---whenever:
! compound diminishes
! repositioning transducers
! relocating meter
Coupling compound
! add more compound
Flowmeter
! none, this is a service item
Pipe strap
! re-tighten
! once yearly
Cable connections
! make sure connections are secure
! remove any residue build-up within the
connection
! as part of your facility's
maintenance schedule
26
----
4.3 Upgrade Meter Software
The most current software is installed in the meter at shipment. Upgrade the
software as newer versions become available, using the RS232 port and a remote
terminal. The most current versions of the following software:
•
•
•
Win Loader - used to upgrade SX30 meter software
SX30 meter software
HydraLink datalogger software,
are available on the Polysonics web-site, at the address below, or contact your
Polysonics Service Representative to inquire about software upgrades.
Polysonics, Inc.
Attn: Service Center
10335 Landsbury Dr., Suite 300
Houston, TX 77099-3407 USA
Telephone:
Fax:
Web:
(281) 879-3700
(281) 498-7721
www.polysonicsinc.com
27
5. Troub
leshooting
oubleshooting
5.1 Diff
erent Modes of Operation
Different
1.
2.
3.
Start Up Mode - initialize meter
Learn Mode - meter “learns” the system parameters -- THERE MUST BE
FLOW in the pipe during learn mode
Operating Mode - read flow and monitor conditions
5.2 Learn Mode Warnings
Table 5.2, below, describes the Learn Mode warning and appropriate corrective
actions. Perform the corrective actions in the sequence that is presented.
TABLE 5.2 – LEARN MODE WARNINGS AND RECOMMENDED CORRECTIVE ACTIONS
Warning
Corrective Action
Invalid Signal,
Can't Learn
1. A
2. B
3. C
4. D
5. E
No Flow,
Can't Learn
1. Check for flowing fluid
2. Make sure pipe is full
3. Check BNC connections
4. Restart meter go into learn mode, check to see if warning is gone
5. E (see Table 5.3c - Action Codes, page 30)
Warning Low S Strength
1. Shut meter down
2. A
3. B
4. C
(see Table 5.3c - Action Codes, page 30)
5. D
6. E
Warning Poor S Quality
1. Let meter enter operating mode
2. See procedure 6.1 (Check the FFT), page 33
28
(see Table 5.3c - Action Codes, page 30)
5.3 Operating Mode Warnings
In operating mode, the warnings indicate a level of severity and a reason for fault.
Table 5.3a (Operating Mode Warnings and Indicated Level of Severity), below,
describes:
• operating mode warning messages
• the level of severity indicated.
Table 5.3b (Operating Mode Faults & Corrective Actions), on page 30, describes:
• faults that caused the warning
• recommended corrective actions.
TABLE 5.3A – OPERATING MODE WARNINGS AND INDICATED LEVEL OF SEVERITY
Warning
Level of Severity
Warni ng
! flow readi ng i s margi nal
Alarm
! flow readi ng i s unreli able
! the meter wi ll not read flow
! the meter wi ll conti nue to read zero flow unti l the fault clears
Fai led/No Flow
NOTE! A "Fai led" warni ng i s normal under no-flow condi ti ons.
The meter wi ll recover when the flow resumes.
29
TABLE 5.3B – OPERATING MODE FAULTS & CORRECTIVE ACTIONS
Faults
Corrective Actions
Note! If more than one of the faults appear then perform the actions for each in
order
PQ
Poor Signal Quality
1. See procedure 6.1 (Check FFT), page 33
2. Perform sequence B,C,D,E (see Table 5.3c - Action Codes, below)
FS
Failed Signal Status
1. Perform sequence B,C,D,E (see Table 5.3c - Action Codes, below)
2. See procedure 6.1 (Check FFT), page 33
3. See section 5.4 (Service), page 31
IS
Invalid Signal
1. See procedure 6.1 (Check FFT), page 33
2. See section 5.4 (Service), page 31
TABLE 5.3C – ACTION CODES
Action
30
Action Description
A
1. Turn off meter
2. Apply more coupling compound
3. Tighten pipe strap
4. Start the meter see if warning clears
B
1. Reposition transducers
2. See if signal strength increases
3. See if fault clears
4. Restart the meter
C
1. Relocate transducers
2. See if signal strength increases
3. See if fault clears
4. Restart the meter
D
1. Turn off meter
2. If pipe surface is rough or coated, then grind the pipe surface
3. Remount the transducers
4.Start the meter see if warning clears
E
1. See section 5.4 (Service), page 31
2. Contact service representative
5.4 Ser
vice
Service
5.4.1 Contacting the Polysonics Service Center
To expedite service by helping the service technician to better diagnose the
problem, please have the following information ready before contacting the factory.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
Signal Strength (%SS - from the bottom of Flow screen)
Process Temperature
Transducer Mounting Configuration
Power Consumption
Pipe Orientation
Fluid Type
Pipe OD and ID
Pipe Material
Liner Material
Liner Thickness
Model Number
Serial Number
Polysonics, Inc.
Attn: Service Center
10335 Landsbury Dr., Suite 300
Houston, TX 77099-3407 USA
Telephone:
Fax:
Web:
(281) 879-3700
(281) 498-7721
www.polysonicsinc.com
31
(section 5.4 Service continued...........)
5.4.2 Factory Service
If Polysonics determines that the problem cannot be resolved over the phone,
return the entire unit to Polysonics.
Contact the Polysonics Service Center before returning an instrument for repair.
Mark the RMA number issued by the service representative on the outside of
the shipping box. The receiving dock will not accept shipments without an RMA
number.
You must include a letter fully explaining the symptoms of the failure as well as
details describing the application where the unit was being operated (type of fluid,
pipe size, pipe material, fluid velocity, temperature, etc.). Service cannot be
adequately performed until the Service Center receives this written information.
To ship an instrument to Polysonics:
1.
2.
3.
4.
Ensure that the instrument is well packed (in its original shipping box, if
available).
Include the letter of explanation.
Write the RMA number of the outside of the shipping box.
Send the unit freight-paid to the Service Center:
Polysonics, Inc.
Attn: Service Center
10335 Landsbury Dr., Suite 300
Houston, TX 77099-3407 USA
Telephone:
Fax:
Web:
32
(281) 879-3700
(281) 498-7721
www.polysonicsinc.com
6. Troub
leshooting Pr
ocedures
oubleshooting
Procedures
6.1 Chec
k the FFT
Check
1.
2.
3.
4.
5.
6.
Press the Menu ---> key to the left of Diags in the Flow screen.
Wait for the Limits screen to appear.
Press the Menu ---> key to the left of FFT in the Limits screen.
See Figure 6.1a-f (FFT Diagrams) on pages 34 and 35 to determine the FFT
condition.
See Table 6.1 (FFT Analysis and Corrective Actions), below, to determine the
corrective action.
Press the Menu ---> key to the left of Flow to return to the Flow screen.
Table 6.1 – FFT Analysis and Corrective Actions
F F T S h ap e
Ideal Doppler
Possible Causes
----
Corrective Actions
! restart the the meter
Broad Band
Noise
! A,B,C,D (see Table 5.3c
! poor pipe coupling
Action Codes, page 30)
! multiple moise generators ! relocate transducers
! shut off noise source
Steady Noise
S p i ke
! electromagnetic noise
! shut off noise source
! relocate transducers
Fluctuating
Doppler
! constantly changing flow
! noise
! increase damping, see
procedure 6.5 (Set the
damping parameter), page 37
! relocate transducers
! ultrasound signal is not
No Doppler
transmitting
Reflection I or II
! A,B,C,D (see Table 5.3c Action
C o d e s, p a g e 3 0 )
33
Ideal Doppler shape
Steady noise spike with Doppler
Broad band noise
FIGURES 6.1 A,B,C – FFT DIAGRAMS
34
Fluctuating Doppler profile
No Doppler reflection I
No Doppler reflection II
FIGURES 6.1 D,E,F – FFT DIAGRAMS
35
6.2 Set the lo
w flo
w cutoff
low
flow
cutoff..
1.
2.
3.
4.
5.
6.
Press the Menu ---> key to the left of Diags in the Flow screen.
Wait for the Limits screen to appear. (See Figure 6.5 - Limits Screen, facing
page)
Press ENTER until the blinking cursor appears in the Low Flow Cutoff box.
Enter the rate for low flow using the number keys.
Press ENTER
Press the Menu ---> key to the left of Flow to return to the Flow screen.
6.3 P
erf
orm an on-site calibration.
Perf
erform
NOTE
1.
2.
3.
4.
5.
6.
7.
36
Determine the flow using another source before performing the following
procedure.
Calculate the site calibration adjustment. (actual flow÷Hydra flow)
Press the Menu ---> key to the left of Diags in the Flow screen.
Wait for the Limits screen to appear. (See Figure 6.5 - Limits Screen, facing
page)
Press ENTER until a cursor appears in the Site Cal box.
Enter the site calibration adjustment, calculated in step 1.
Press ENTER
Press the Menu ---> key to the left of Flow to return to the Flow screen.
6.4 Reset the factor
y settings.
factory
1.
2.
NOTE
Shut off the meter.
Turn the meter back on.
All of the settings will reset except the 4-20 mA calibration, the pipe ID
and the flow units/time base
6.5 Set the damping parameter
parameter..
1.
2.
3.
4.
NOTE
Press the Menu ---> key to the left of Diags in the Flow screen.
Wait for the Limits screen to appear. (See Figure 6.5 - Limits Screen, below)
Enter the amount of damping (0-99 units, 1 unit = 15 seconds) using the
number keys.
Press the Menu ---> key to the left of Flow to return to the Flow screen.
The Smart Filter used by this meter, allows the damping function to
smooth small fluctuations without diminishing the meter’s response to
large flow changes, even at high damping settings.
FIGURE 6.5 – LIMITS SCREEN
37
7. Appendix
7.1 Hydra SX30 Man
ual Symbol Descriptions
Manual
Symbol
NOTE
Description
Note: Indicates information
that is important for the proper
operation of the instrument.
Caution: Indicates information
that is important to prevent
damage to the instrument.
See additional information
WARNING
Warning: Indicates
information that is important to
prevent harm to the operator
and/or damage to the
instrument.
7.2 Pipe ID
7.2.1 Obtain the Pipe ID.
If you do not know the pipe ID, then try to get the Outside Diameter (OD) by one of
the following methods:
1.
2.
3.
4.
5.
Read the pipe Outside Diameter (OD) and schedule on the pipe
Read the OD and schedule from system drawings
Measure the pipe circumference using a tape measure, and divide the
measured circumference by 3.14 (Circumference ÷ 3.14)
For an approximate pipe ID use the OD with the charts in section 7.2.2 (Pipe
ID Charts), on pages 39-41.
For a more accurate pipe ID use an ultrasonic pipe thickness gauge (part
#0704/0187, available from Polysonics), AND perform the following
calculation:
pipe ID = pipe OD – (2 x pipe thickness)
38
1.1 85
1.5 30
1.7 70
2.2 45
2.7 09
1.3 15
1.6 60
1.9 00
2.3 75
2.8 75
3.5 00
4.0 00
4.5 00
5.5 63
6.6 25
8.6 25
10 .75 0
12 .75 0
14 .00 0
16 .00 0
18 .00 0
20 .00 0
24 .00 0
30 .00 0
36 .00 0
42 .00 0
48 .00 0
1
1.2 5
1.5
2
2.5
3
3.5
4
5
6
8
10
12
14
16
18
20
24
30
36
42
48
17 .67 0
19 .63 4
23 .56 4
29 .50 0
15 .67 0
8.4 07
10 .48 2
12 .43 8
3.3 34
3.8 34
4.3 34
5.3 45
6.4 07
S c he d. 5
N om in al
O u ts id e
P ipe S ize ( ) D iam e te r
(O D )
17 .50 0
19 .50 0
23 .50 0
29 .37 6
35 .37 6
8.3 29
10 .42 0
12 .39 0
13 .50 0
15 .50 0
3.2 60
3.7 60
4.2 60
5.2 95
6.3 57
1.0 97
1.4 42
1.6 82
2.1 57
2.6 35
S c he d. 10
(L igh t
W all)
8.0 71
10 .13 6
12 .09 0
13 .25 0
15 .25 0
17 .12 4
19 .00 0
22 .87 6
28 .75 0
34 .75 0
40 .75 0
17 .37 6
19 .25 0
23 .25 0
29 .00 0
35 .00 0
41 .00 0
S c he d.
30
8.1 25
10 .25 0
12 .25 0
13 .37 6
15 .37 6
S c he d.
20
40 .50 0
47 .25 0
16 .87 6
18 .81 2
22 .62 4
28 .50 0
34 .50 0
7.9 81
10 .02 0
11 .93 8
13 .12 4
15 .00 0
3.0 68
3.5 48
4.0 26
5.0 47
6.0 65
1.0 49
1.3 80
1.6 10
2.0 67
2.4 69
S c he d.
40
16 .50 0
18 .37 6
22 .06 2
7.8 13
9.7 50
11 .62 6
12 .81 2
14 .68 8
S c he d.
60
16 .12 4
17 .93 8
21 .56 2
7.6 25
9.5 62
11 .37 4
12 .50 0
14 .31 2
2.9 00
3.3 64
3.8 26
4.8 13
5.7 61
0.9 57
1.2 78
1.5 00
1.9 39
2.3 23
S c he d.
80
15 .68 8
17 .43 8
20 .93 8
7.4 37
9.3 12
11 .06 2
12 .12 4
13 .93 8
S c he d.
10 0
15 .25 5
17 .00 0
20 .37 6
7.1 87
9.0 62
10 .75 0
11 .87 6
13 .56 2
3.6 24
4.5 63
5.5 01
S c he d.
12 0
14 .43 8
16 .06 2
19 .31 2
31 .31 2
31 .87 6
6.8 13
8.5 00
10 .12 6
11 .18 8
12 .81 2
3.4 38
4.3 13
5.1 87
2.6 24
0.8 15
1.1 60
1.3 38
1.6 87
2.1 25
S c he d.
16 0
14 .87 6
16 .50 0
19 .87 6
7.0 01
8.7 50
10 .50 0
11 .50 0
13 .12 4
S c he d.
14 0
T a b le B -1
S te el, S tainle ss S te el an d P V C P ip e S ta nd a rd S che d ule s
Ins id e D ia m eter (ID ) an d O u ts id e D ia m ete r (O D) in In ch es
41 .25 0
47 .25 0
17 .25 0
19 .25 0
23 .25 0
29 .25 0
35 .25 0
7.9 81
10 .02 0
12 .00 0
13 .25 0
15 .25 0
3.0 68
3.5 48
4.0 26
5.0 47
6.0 65
1.0 49
1.3 80
1.6 10
2.0 67
2.4 69
S td. W all
41 .00 0
47 .00 0
17 .00 0
19 .00 0
23 .00 0
29 .00 0
35 .00 0
7.6 25
9.7 50
11 .75 0
13 .00 0
15 .00 0
2.9 00
3.3 64
3.8 26
4.8 13
5.7 61
0.9 57
1.2 78
1.5 00
1.9 39
2.3 23
X ST G
7.2.2 Pipe ID Charts
39
40
3.80
4.80
6.90
9.05
11.10
13.20
15.30
17.40
19.50
21.60
25.80
31.74
37.96
44.20
50.50
56.66
62.80
75.34
87.54
3
4
6
8
10
12
14
16
18
20
24
30
36
42
48
54
60
72
84
53.96
60.02
72.10
84.10
24.28
28.98
35.98
42.00
47.98
12.12
14.16
16.20
18.22
20.26
3.02
3.96
6.02
8.13
10.10
ID
57.10
63.40
76.00
88.54
25.80
32.00
38.30
44.50
50.80
13.20
15.30
17.40
19.50
21.60
3.96
5.00
7.10
9.05
11.10
OD
54.00
60.06
72.10
84.10
24.02
29.94
36.00
41.94
47.96
11.96
13.98
16.00
18.00
20.00
3.12
4.10
6.14
8.03
9.96
ID
Class B
57.80
64.20
76.88
26.32
32.40
38.70
45.10
51.40
13.50
15.65
17.80
19.92
22.06
3.96
5.00
7.10
9.30
11.40
OD
54.00
60.20
72.10
24.22
30.00
35.98
42.02
47.98
12.14
14.17
16.20
18.18
20.22
3.06
4.04
6.08
8.18
10.16
ID
Class C
58.40
64.82
26.32
32.74
39.16
45.58
51.98
13.50
15.65
17.80
19.92
22.06
3.96
5.00
7.10
9.30
11.40
OD
53.94
60.06
24.00
30.00
36.00
42.02
48.00
12.00
14.01
16.02
18.00
20.00
3.00
3.96
6.00
8.10
10.04
ID
Class D
26.90
33.10
39.60
13.78
15.98
18.16
20.34
22.54
7.22
9.42
11.60
OD
26.90
33.46
40.04
13.78
15.98
18.16
20.34
22.54
7.22
9.42
11.60
OD
24.00
30.00
36.00
12.00
14.00
16.00
18.00
20.00
6.00
8.10
10.00
ID
Class F
27.76
14.08
16.32
18.54
20.78
23.02
7.38
9.60
11.84
OD
24.26
12.14
14.18
16.18
18.22
20.24
6.08
8.10
10.12
ID
Class G
27.76
14.08
16.32
18.54
20.78
23.02
7.38
9.60
11.84
OD
24.00
12.00
14.00
16.00
18.00
20.00
6.00
8.00
10.00
ID
Class H
Standard and double cementthicknesses
lining
are listed in Table B-3.
24.28
30.00
36.00
12.14
14.18
16.20
18.20
20.24
6.06
8.10
10.12
ID
Class E
N OT E:For pipes with cement linings, reduce the pipe inside diamete
r by two times the lining thickness.
OD
Class A
Nom inal
Pipe Size
( )
Table B-2
Cast Iron Pipe Standard Classes
Inside Diam eter (ID) and Outside Diam eter (OD) in Inches
41
3.96
4.80
6.90
9.05
11.10
13.20
15.30
17.40
19.50
21.60
25.80
32.00
38.30
44.50
50.80
57.10
3
4
6
8
10
12
14
16
18
20
24
30
36
42
48
54
31.22
37.44
43.56
49.78
55.96
14.64
16.72
18.80
20.88
25.04
6.40
8.51
10.52
12.58
Class 50
31.14
37.34
43.44
49.64
55.80
14.58
16.66
18.74
20.82
24.98
3.46
4.28
6.34
8.45
10.46
12.52
Class 51
31.06
37.06
43.32
49.50
55.64
14.52
16.60
18.68
20.76
24.92
3.40
4.22
6.28
8.39
10.40
12.46
Class 52
30.98
37.14
43.20
49.36
55.48
14.46
16.54
18.62
20.70
24.86
3.34
4.16
6.22
8.33
10.34
12.40
Class 53
30.90
37.04
43.08
49.22
55.32
14.40
16.48
18.56
20.64
24.80
3.28
4.10
6.16
8.27
10.28
12.34
Class 54
Note: For pipes with cem ent linings, reduce the pipe inside diam eter by two tim es the lining thickness listed above.
Outside
Diam eter
(OD)
Nominal
Pipe Size
Inside Diam eter
30.82
36.94
42.96
49.08
55.16
14.34
16.42
18.50
20.58
24.74
3.22
4.04
6.10
8.21
10.22
12.28
Class 55
30.74
36.84
42.84
48.94
55.00
14.28
16.36
18.44
20.52
24.68
3.16
3.98
6.04
8.15
10.16
12.22
Class 56
Table B-3
Ductile Iron Pipe Standard Classes
Inside Diameter (ID) and Outside Diameter (OD) in Inches
0.250
0.1875
0.125
Standard
Thickness
0.500
0.375
0.250
Double
Thickness
Cem ent Lining (See Note)
7.3 Application Signal Quality and Signal Strength
Signal quality and signal strength are indicators of how well suited an application is
to a Doppler flowmeter. To determine the quality of your application:
1.
2.
3.
Look at the bottom right corner of the Flow screen.
Note the SQ number and the %SS number.
Refer to Figure 7.3b - Signal Quality/Signal Strength Chart, below, to
determine the quality of your application.
Signal Quality (SQ)
Excellent
5
Good
3.5
Adequate
2.5
Failure
0.25
2.5
10
Signal Strength (%SS)
FIGURE 7.3B – SIGNAL QUALITY/SIGNAL STRENGTH CHART
Failure
Adequate
Good
Excellent
– the meter is not reading flow
– minimum quality, flow readings may be inconsistent or inaccurate
– flow readings are fairly accurate and consistent
– flow readings are accurate and consistent
To improve signal quality and signal strength see Table 5.3c (Action Codes), on
page 30.
42
Functional
Specifications
Performance
Specifications
Physical
Specifications
7.4 Hydra SX30 Meter Specifications
Outputs
Power Supply
"
"
Enclosure
High impact structural copolymer polypropylene.
Watertight and corrosion proof. MIL-C-4150J
Transducers
Enclosure and cable are suitable for non-submersible
service.
Weight
Approximately 11 lbs (4.9 kg) with 12-hour battery
Flow Range
0.2 to 32 FPS (0.06 to 9.7 MPS)
Accuracy
+/- 2% of flow under controlled conditions
Pipe Size
"
"
200 in (5 m) maximum
0.5 in (13 mm) minimum
4-20 mA (i nto 100 ohms), 12 bi ts, 5kV opti cally i solated, loop or self powered
RS232 seri al i nterface usi ng D B25 connector
C harger i nput voltage supply:
" 15 volt @ 1 amp typi cal
" 2 amp maxi mum
" mai ns supply voltage fluctuati ons to charger not to exceed +/- 10% of the
nomi nal supply voltage
Auxi li ary supply voltage:
" 15 volt @ 2 amp typi cal
" must be an S E LV source
"
auxi li ary i nput termi nals are suppli ed from an SELV source as per
IEC 1010-1 Annex H
Keypad
19-key wi th tacti le acti on
D i splay
240x60 graphi c LC D di splay wi th backli ght
D atalogger
90,000 data poi nts i n 10 fi les of 9,000 poi nts each.
Programmable i n 0.5-60 mi nute i ntervals. D ata i s ti me/date stamped
5 oC to 40 oC , C SA approved
-20 oC to +60 oC , non C SA approved
Transmi tter
Temperature Range
"
Transducer
Temperature Range
"
Alti tude
Maxi mum 2000 m
Humi di ty Range
0 to 99% relati ve humi di ty, non condensi ng
Envi ronmental
Operati ng
C ondi ti ons
"
"
"
"
-40 to +250 oF (-40 to +122 oC )
-40 to +350 oF usi ng opti onal hi gh temperature transducer blocks
Polluti on degree: 2
Installati on category: I for meter II for battery charger
43
44
FLOW
TEST
CALIB
RANGE
5 SEC
POLYSONICS
ENTER
PIPE ID
4-20MA
LO G
DIAGN
SET UP
FLOW
FLOW
TIME
TREND
LO G
LEARN MODE
FLO W UNITS/
TIME BASE
ENTER
FLOW
TO TALZ
FFT
LIMITS
7.5 Hydra SX 30 men
u tree
menu