Installation, Setup and Operating Instructions v2.14

Transcription

Installation, Setup and Operating Instructions v2.14
Vmed
Installation, Setup and Operating Instructions v2.14
PC-VETGARD WIRELESS PATIENT MONITOR
Installation, Setup and Operating Instructions
v2.14
PC-VetGard Wireless Patient Monitor
© Vmed Technology
16149 Redmond Way • Suite 108 Redmond, WA 98052 USA
1
Table of Contents
PRECAUTIONS
Chapter 1 Installation and Setup ............................................................................... 4
Installing the VetGard Application software ................................................................................................... 4
Installing Bluetooth..................................................................................................................................... 7
ƒ
Do not plug VetGard sensors into any other equipment.
Bluetooth Installation and Discovery, Windows XP Service Pack 2.................................................................. 9
ƒ
VetGard is not registered with the FDA for human use.
Bluetooth Installation and Discovery; Computers Not Using Windows XP Service Pack 2. ............................... 14
ƒ
Do not use in the presence of flammable anesthetics.
Setup VetGard Defaults............................................................................................................................ 35
ƒ
Do not immerse the VetGard, sensors or cables in water or other fluids. Avoid
spilling any fluids on the VetGard or accessories. Review the Service and
Maintenance Chapter for proper cleaning procedures.
Setup Patient Information.......................................................................................................................... 38
Using the VetGard with Windows XP Service Pack 2................................................................................... 13
Multiple Installations on Computers not using Windows XP Service Pack 2.................................................... 26
Setup Vet/Clinic Information...................................................................................................................... 37
Computer Video Resolution ...................................................................................................................... 38
Chapter 2 Description .............................................................................................. 39
Hardware .............................................................................................................................................. 39
ƒ
Do not gas sterilize or autoclave the VetGard or sensors.
Software................................................................................................................................................. 41
ƒ
Do not disassemble. The VetGard contains no operator serviceable components.
Sweep and Scale .................................................................................................................................... 42
ƒ
Electrical Interference: Strong electromagnetic or radio frequency interference
(RFI) may affect the performance of all ECG equipment including the VetGard.
RFI may result in distorted ECG signals. Do not operate near electro-cautery,
diathermy equipment, or cellular phones.
Windows Display..................................................................................................................................... 41
Display Toolbar ....................................................................................................................................... 42
ƒ
ƒ
ƒ
ƒ
ƒ
CAUTION: Conductive parts such as electrodes and associated connectors and
leads, including the neutral electrode, should not contact other conductive parts
including ground when using the VetGard.
Ensure that electrodes are connected only to the patient and do not touch any
other conductive part or equipment to prevent possible shock to patient from
leakage current.
Avoid cable and connector damage by taking care when connecting and
disconnecting sensors and leads from the VetGard. Grasp plug, not the cable
and pull straight out and push straight in.
Do not pull ECG clip leads or Sp02 sensors from the patient without opening
clip.
Do not use the rectal temperature probe while using the esophageal probe.
Remote Function Selection....................................................................................................................... 43
Chapter 3 Operating Instructions............................................................................. 45
Preparation ............................................................................................................................................. 45
Taking the ECG....................................................................................................................................... 46
Using Chest Probe ............................................................................................................................... 47
Using Leg Clips .................................................................................................................................... 48
Using Esophageal Probe.......................................................................................................................... 48
Using Disposable Chest Electrodes ........................................................................................................... 49
Using the Rectal Temperature Probe......................................................................................................... 50
Using the Oxygen Saturation (Sp02) Sensor............................................................................................... 50
Chapter 4 Reviewer Program .................................................................................. 55
Description.............................................................................................................................................. 55
Modes of Operation .............................................................................................................................. 55
Using the Reviewer Program................................................................................................................ 56
Chapter 5 Service and Maintenance ....................................................................... 65
Service and Support ................................................................................................................................ 65
Cleaning and Maintenance ....................................................................................................................... 65
Troubleshooting....................................................................................................................................... 66
Appendix A Computer Requirements and Specifications ........................................ 68
Computer Requirements ...................................................................................................................... 68
Physical Specifications ......................................................................................................................... 68
System Specifications .......................................................................................................................... 69
Appendix B Warranty............................................................................................... 70
2
3
1
Chapter
Chapter 1 Installation and Setup
DO NOT INSTALL THE BELKIN BLUETOOTH SOFTWARE.
Refer to Bluetooth installation instructions in this manual.
Installing the VetGard Application software
1. Put the VetGard installation CD into the computer’s CD ROM drive. The Vmed
VetGard installation program should automatically start and the window below should
appear:
NOTE: In the software does not start automatically, click on the START icon in
the lower left hand corner of the screen and select ‘Run’ from the menu. When
the ‘Run’ dialog box appears, click the Browse button. A Windows Explorer
window will appear that will allow you to navigate to your computer’s CD ROM
device. The name of the CD should contain “VetGard”. Select this CD and
double click on the file named “Setup” then click on “OK” in the ‘Run’ dialog
box.
3. Select “Yes”, which should cause the following window to appear::
NOTE: Vmed recommends that you use the default installation location.
4. Click “Next” and the following window should appear:
2. Click on “Next” to open the following window:
4
5
Note: A Shortcut to the VetGard program has been placed on the windows
desktop and on the Start menu.
7. Remove the VetGard Installation CD and turn off the VetGard by pressing and holding
the ON/OFF key down until the device turns off.
Installing Bluetooth
IF YOUR COMPUTER HAS A PRE-INSTALLED BLUETOOTH DEVICE,
DO NOT INSTALL THE BLUETOOTH ADAPTOR SOFTWARE. USE
YOUR COMPUTER’S BLUETOOTH DEVICE.
5. Select “Typical” and then click “Next”. The following window should appear:
IF YOUR COMPUTER USES THE WINDOWS XP OPERATING SYSTEM
WITH SERVICE PACK 2 (SP2) INSTALLLED, DO NOT INSTALL THE
BLUETOOTH ADAPTOR SOFTWARE. IT IS MUCH EASIER TO SET UP
AND USE BLUETOOTH AND TO DISPLAY MULTIPLE MONITORS IF
WINDOWS XP WITH SP2 IS USED.
FOR THIS REASON, VMED
RECOMMENDS USING A WINDOWS XP COMPUTER AND
DOWNLOADING SP2 AND THE LATEST CRITICAL UPDATES.
IF YOU HAVE PREVIOUSLY INSTALLED THE BELKIN BLUETOOTH
SOFTWARE, UNINSTALL IT AND RE-BOOT YOUR COMPUTER.
NOTE: If you are not using a Windows XP computer or you do not wish to
update your XP system with SP2, refer to “Bluetooth Installation and Discovery,
Computers Not Using Windows XP Service Pack 2” on page 14 of this manual.
STEP 1: Determine the version of your operating system
1. From the “START” button in the lower left hand corner of the screen, Click on
“Control Panel” and then double click on “System”. A window similar to the one
below should appear:
6. Highlight “Accessories” and press “Next” to start the installation of the VetGard
software from the CD-ROM device. Click Finish when the installation is complete.
6
7
STEP 3
Bluetooth Installation and Discovery, Windows XP Service Pack 2
1. Plug the Bluetooth device (Dongle) into a USB port on your computer.
2. Windows should automatically detect the device. The computer should respond with
an audio prompt and a balloon confirming detection of the device should appear in
the lower right hand corner of the display. Close this balloon if it appears.
3. Double click the “Bluetooth Devices” icon on your desktop.
NOTE: You may also select “Bluetooth Devices” from the
“Control Panel” or from the Windows “Start” menu.
4. The following window will appear:
2. This window will tell you if your PC has Service Pack 2 installed. NOTE: If your
operating system is not Windows XP with SP2, refer to page 11 for Bluetooth
installation instructions.
STEP 2: Download Service Pack 2 for your Windows XP operating system.
3. Vmed recommends that Windows XP computers be upgraded to include Service Pack
2 since Bluetooth setup is much quicker. This requires internet access for the PC.
4. If you already have Service Pack 2 installed, proceed to STEP 3
5. If you are upgrading to Service Pack 2 and have previously installed Bluetooth
software, select “Add or Remove Programs” in the Control Panel and remove this
software.
5. Click on the “Options” tab and make sure that each box is checked as below:
6. Connect your PC to the internet
7. From the Start Menu select Windows Update and follow Microsoft’s instructions for
Service Pack 2 installation or “critical downloads.
8. After installation, restart your computer if required.
8
9
7. Click “Next” and your computer will search for the VetGard. When the VetGard is
discovered, a window like the one below will appear. The serial number following
“Vmed” and beginning with EG will be the same as the serial number that is printed
on the back label of the VetGard.
6. Click the “Devices” tab again and turn on your VetGard. Wait for the status indicator
on the VetGard to turn yellow and then click the “Add” button. The following
window should appear.
8. Highlight the identified VetGard and click “Next”, which will open the following
window:
10
11
Using the VetGard with Windows XP Service Pack 2
1. Power on the previously discovered VetGard. The STATUS light on the VetGard
will turn red and then yellow. NOTE: Every device has to be discovered once.
If you have never been through the discovery procedure with this particular
device on this PC, do so before proceeding.
2. Start the VetGard application by double clicking the VetGard icon on the desktop.
3. The window below will open over the VetGard display screen:
9. Select the “Don’t use a passkey” button and click “Next”. The window below will
open:
10. Click “Finish” to conclude installation and discovery
12
NOTE: If you have discovered multiple VetGard monitors, the device last
discovered or connected to your computer will appear in the text box in the
above window. Press the “down” arrow to view the other devices, highlight
the one you want to use and press “Connect” As many as eight VetGard
devices can be connected to a computer at one time. The practical limitation,
however, is the size of the display and the computing power of your computer.
To use multiple devices, go to step #1 and repeat the discovery process for each.
The Vmed software automatically selects the COM Port if Windows XP with
SP2 is installed. The appropriate selection will be made automatically as
shown in the ”Configure VetGard” window under the “Tools” button in the
application software.
13
Bluetooth Installation and Discovery; Computers Not Using
Windows XP Service Pack 2.
1. Install the Bluetooth software from the CD included with the adaptor and per
instructions with the Bluetooth adaptor. NOTE: Do not plug the Bluetooth adaptor
into the USB port until prompted in the installation sequence.
2. After installation is complete, close all windows and re-boot your computer.
3. Double click the “My Bluetooth Places” icon on your desktop and the window below
will open:
6. Deselect all services EXCEPT Bluetooth Serial Port by clicking on the box to the
left of the text describing each service and then select finish. Note: Be sure to
scroll down using the down arrow to ensure that all other boxes are NOT
checked.
7. Click “Finish” and when the next window opens click SKIP and close all windows.
Set up of the Bluetooth Adaptor
1. The Bluetooth installation procedure will place a “My Bluetooth Places” icon on
the Windows desktop. Double click on this icon and the following window will
appear:
4. Make sure that all the boxes in the window above are checked and click NEXT. A
window will appear for you to name your computer. Please do so and then click
NEXT which will open the “Initial Bluetooth Configuration Wizard” screen.
5. Click NEXT and the window below will open:
14
15
2. Click on the “View or modify configuration” hyperlink at the left side of the window,
alternatively select “Bluetooth” from the top menu and then select “Advanced
Configuration”. The following window should appear:
3. Choose the “General” tab of the setup box, then choose a name for your computer
and enter it into the computer name box. Select the computer type: Personal
Computer or Laptop Computer.
4. Select the “Accessibility “ tab and the following window should appear:
16
17
6. Verify that the “Look for other Bluetooth devices” box is checked and that the pull
down menu is set to “Report all Bluetooth devices”. Configure the Bluetooth to discover
new devices every [1] minutes. Next select the “Local Services” tab which should
generate the following window:
5. Make sure the “Let other Bluetooth devices discover this computer” box is NOT
checked. Allow all devices to connect to this computer by selecting “All devices” in the
pull down menu. Next, select the “Discovery” tab which should generate the following
window:
18
19
THIS STEP IS CRITICAL TO PROPER OPERATION OF THE PC-VET.
Note: The next part of the configuration sets up a COM port local to the computer
which will host the actual Bluetooth COM port. In the example shown above COM7
was chosen by the Bluetooth software as the Bluetooth drivers were installed. The
actual COM port number displayed on your particular VetGard may be different.
This COM port number is NOT the one that is used by the VetGard application but
is only an internal host for the COM port that will be used. Even though it is not
directly used by Vmed’s software, it is important that this host COM port be
configured properly
9.
In the Bluetooth configuration window select the “Client Applications” tab to open the
following window::
7. Select the Bluetooth Serial Port which will highlight the text as shown above. If the
Secure connection column of this item says anything other than “Not Required”, please
click on the “Properties” button which will bring up the following window:
Note: This part of the configuration deals with setting up the actual COM port
used by the VetGard software. It is very similar to the setup for “Local Service”
except now the COM port number will represent the actual COM port used to
communicate to the VetGard. In the example shown above, COM5 was chosen by
the Bluetooth software when the drivers were installed. The actual COM port
number displayed on your particular PC may be different. If this is the very first
time this window has been opened no COM port number will appear until
“Properties” is clicked on.
8. Verify that the “Secure Connection” box is NOT checked. If changes need to be made,
un-check the “Secure Connection” box and then click “Apply” followed by “OK”
which should cause the Bluetooth Serial Port item to be configured to read “Not
Required” in the “Secure Connection” column of the Bluetooth Configuration window.
20
10. Click on the Bluetooth Serial port which will highlight the text as shown above. If
the Secure connection column of this item says anything other than “Not Required” or
no COM port number has been assigned to the Bluetooth Serial Port, then click on the
“Properties” button which will again bring up the following window:
21
Bluetooth Discovery
1. Bluetooth Discovery will be less confusing to setup when starting from a point of “No
devices in range.” As a starting point turn OFF all Bluetooth devices within 150 meters.
This includes other PC’s in the realm that have Bluetooth devices.
2. Double click the “My Bluetooth Places” icon on the desktop and the window below will
open:
11. Vmed recommends that the port be renamed “Bluetooth COMX” where X is the
COM port number shown in the COM port box. Verify that the “Secure Connection”
box is NOT checked. Click on “Apply” followed by “OK” which will configure the
Bluetooth Serial Port to read “Not Required” in the “Secure Connection” column of the
Bluetooth Configuration window and the new comport number will also appear.
NOTE: This step is critical to the proper operation of the VetGard. Write down
this comport number (In this case, COM 5). It will be used later to configure the
VetGard for connection to your computer.
12. Click “Apply” and “OK” in the above window then close the “My Bluetooth
Places” Window.
22
3. Click on the “View devices in range” on the left side menu of the box or alternatively
select the Bluetooth pull down menu and select “Search for devices”. This should cause
the Bluetooth device to search for any devices in range. None should be found and the
following window should be displayed:
23
Note: Should this view not be seen, click on the right most icon (just under the word
Help) on the toolbar and select ‘Details’ view. The text that appears under device
name should be Vmed-EGXXXX, where XXXX is the unique I.D. number preprogrammed into each VetGard at manufacture. This number is stamped on the back
label of each VetGard.
5. Double click on the Vmed EGXXXX icon to discover services available.
4. Turn on the VetGard monitor by depressing and releasing the red ON/OFF power
switch. The VetGard STATUS indicator will turn from red to yellow. Press the “OK”
button in window above. Click on the “View devices in range” text to the left side of the
box or alternatively select the Bluetooth pull down menu and select “Search for devices”.
This should again cause the Bluetooth device to search for any devices in range. This
time the window shown below should be displayed:
6. Right click on the Vmed EGXXXX text; this will bring up a menu of possible
selections. Select “Connect COM0” which should cause the following window to
appear:
Note: If the battery level in the VetGard is low, the Replace Battery indicator will
blink red. If the battery is critically low the indicator will be steady red. Do
not proceed until the batteries have been replaced. In order to save power,
the VetGard will turn itself off after 3 minutes of inactivity.
Note: The icon will now be green in color and the word “Connected” should
appear. The VetGard I.D. number will also appear near the logo. The “Status”
light on the VetGard will also blink green.
Discovery is now complete and the computer should remember that the connection of a
particular VetGard to a particular COM port. Exit the Window and turn off the
VetGard Monitor.
24
25
Multiple Installations on Computers not using Windows XP Service
Pack 2
Several wireless monitors may be installed on one computer however each new monitor being
displayed must be rediscovered each time. See the “Discovery” instructions in Chapter
1. Double click on the “My Bluetooth Places” shortcut located on the computers
desktop. Within a few seconds this should cause the following window to appear:
3. Select the “Client Applications” tab as shown and then click on the “Add COM
port” button in the lower right hand side of the box. The following screen should
appear.
2. Click on the “View or modify configuration” text to the left side of the box. The
following screen should appear.
26
27
4. Verify that the “Secure Connection” box is NOT checked. An additional COM Port
can be chosen from a pull down menu. In this case, the next available COM Port is
COM10. Click on “Apply” followed by “OK”
Note: Up to four serial ports can be added in this manner; for the purpose of
this document only one is added. The COM port numbers available may be
different on your particular PC. They are a function of your PC’s hardware
and software configuration. In this case COM5 and COM10 are available. For
your PC these numbers may be different.
7. Turn one device off and then view devices in range. The results should look similar
to that below. In this case the VetGard with an ID of 0047 has been turned off.
5. The window shown above should now appear listing the 2 available serial ports.
6. Turn on both VetGard devices and wait for the Status indicators on both devices to
turn from red to amber. Then view devices in range. The results should look similar
to that below. In this case a VetGard with an ID of 0018 and one with an ID of 0047
have been discovered. The ID number will match the one stamped on the back label
of your particular VetGard. Each VetGard has a unique ID number which will
undoubtedly be different than the ones shown in this example
8. Double click on the icon and then double click again to discover the device the
screen shown below should appear:
28
29
9. The VetGard with an ID of 0018 has now been assigned to COM5. Turn the
VetGard off and then on again. Wait for the status indicator to turn amber and
double click on the VetGard desktop icon.
10. Select “File” on the toolbar above and click on the “Connect VetGard” text to
remove the check mark. This will allow the “Tools” menu to become active.
11. Select “Tools” at the top of the screen, click “Configure VetGard” and the
following screen will appear:
VetGard v2.12.lnk
30
31
16. Double click on the icon and then double click again to discover the device the
screen shown below should appear:
12. Verify that the COM port number is what you have previously determined it should
be, change it if necessary and then select “OK”
13. Select “File” and again click on the “Connect VetGard” text to replace the check
mark. This will again allow the VetGard to connect to your computer.
14. Within a minute or so the VetGard should connect on the designated COM port
and the status indicator should be green.
15. Leave this VetGard running, turn on the second VetGard and wait for the Status
indicator to turn amber and then return to the Bluetooth discovery window and
search for devices. The screen should look similar to the one shown below:
32
33
19. The VetGard unit with an ID of 0047 has now been assigned to COM10. Turn the
VetGard off and then on again. Wait for the status indicator to turn amber and again
double click on the VetGard desktop icon.
VetGard v2.12.lnk
20. A second instance of the VetGard program should run. The following error box
should appear because COM5 is in use by 0018. Click OK
22. Verify that the COM port number is as previously determined, change if necessary
and then select “OK”
23. Select the File menu and click “Connect VetGArd”. Verify that a check mark is next
to the “Connect VetGard” text.
24. Within a minute or so the VetGard should connect on the designated com port and
the VetGard should operate normally. The Status indicator should be green. Leave the
VetGard running.
25. Your PC is now configured to communicate with two VetGards at once. To use
either device, turn on the VetGard you want to use and connect to the COM port that
has been assigned to it.
Setup VetGard Defaults
This section describes the procedure to change the default settings that will apply each time
the VetGard program is opened. Feature default settings such as TONE, ALARM,
ON/OFF, GRID, etc. may be overridden by clicking the applicable icon on the upper
toolbar only if the VetGard is connected to your computer.
21. Next select the “Tools” menu at the top of the screen and click “Configure PCVetGard ” and the following screen will open:
1. Double click the VetGard icon on your desktop. The following window will open:
2. Click File and remove the check mark from “Connect VetGard” option.
3. Click Tools and the following dialog box will open:
34
35
Setup Vet/Clinic Information
1. Click the “Tools” menu on the upper tool bar of the Windows display and then “Edit Vet
Information” and the following window will appear:
Figure 1 VetGard default configuration menu.
4. Fill in the appropriate COM port number for the VetGard as determined in item 11 of
Step #2 (Set up of the Belkin Bluetooth Adaptor) of this procedure.
5. The default ECG notch filter, used to filter AC line interference, can be set here also. For
Domestic use (North America) choose 60 Hz.; for international use choose 50 Hz.
6. Select the desired sweep rate (Chart speed) and other monitoring parameters as desired.
Most veterinarians use a sweep speed of 50 mm/sec. Filter AC line noise by selecting a
“Notch” filter of 60 Hz. for North America and 50 Hz. for international use.
7. Select the desired Reviewer recording mode by placing a check mark in the appropriate
selection. NOTE: The default recording mode for the VetGard is “Continuous
Event Only” as checked above.
2. Type in your vet/clinic information as above.
Note: Up to seven lines of information specific to your particular veterinary clinic
can be entered in this window. It will be remembered each time the program is
started and contains text that will be printed on the ECG print report and
Reviewer generated reports.
3. Press OK.
NOTE: This heading will appear on all reports generated.
8. Click “OK” to save these settings. These settings will apply until changed with a new
default selection or with input from the “Select” button on the VetGard or the window
toolbar.
36
37
2
Chapter
Setup Patient Information
NOTE: You may elect to enter patient information after completing the monitoring
procedure if desired. This information may also be edited in the Reviewer Program.
1. Select the “Tools” on the upper tool bar and then “Edit Patient Information”. The
following window will appear:
Chapter 2 Description
Hardware
The PC-VetGard is a battery operated, Bluetooth enabled, wireless, multi-parameter patient
monitor providing ECG, Sp02, Respiration and Temperature information and displaying realtime waveform and digital information on a computer monitor. Each VetGard includes the
following sensors:
™ A chest probe with three build-in metallic electrodes for ECG screening. When placed
over the heart, accurate rate and rhythm information is derived and if oriented to
approximate the Lead II axis, an estimation of true Lead II amplitude can be achieved.
™ A set of three clip electrodes that, when attached to the legs of the patient as described,
provide Lead II information.
Note: Since it is not always possible to obtain an adequate signal using the
chest probe, clip leads may be required when monitoring some patients. Clip
leads can also be used to continuously monitor the ECG signal during surgery or
for short term critical care monitoring.
3. Select “File” from the top toolbar and replace the check mark at “Connect to VetGard”
to enable automatic connection to the VetGard. NOTE: The VetGard will not
connect to your computer unless the check mark is replaced.
™ Two internal esophageal probes to monitor ECG and temperature of the anesthetized
patient. Three electrodes are affixed to a flexible plastic tube that is inserted into the
esophagus and positioned over the heart with the distal electrode positioned behind
the left ventricle and the second electrode behind the left atrium. The third electrode
is ground. Each esophageal probe includes a temperature sensor for the accurate
detection and display of core body temperature
Computer Video Resolution
™ A cable yoke and set of 3 leads with spring clips for attachment to pre-gelled,
disposable electrodes for long term monitoring.
2. Type in desired information and press OK.
NOTE: Ensure that the resolution on your computer display is set to 1024 x 768 so
that the entire PC-Vet display is viewable. Select Start>Settings>Control
Panel>Display>Settings
38
™ Large and small spring clips to hold a single oxygen saturation sensor at the
appropriate site to monitor oxygen saturation and pulse rate.
™ A rectal temperature probe with 5 foot cable
39
The VetGard is equipped with an integral wireless Bluetooth radio transmitter. Bluetooth
enables the wireless transmission of information, including real-time patient data,
between the VetGard and a Bluetooth enabled Windows based computer.
All functions not requiring keyboard input can be controlled from the VetGard front
panel or. The Windows display toolbar, however commands are only active if a wireless
connection between the VetGard and the computer is established.
Bluetooth operates at approximately 2.4 GHz. and steps up and down frequency
constantly making the VetGard free of interference including the radio frequency
interference common in hospital and clinic settings. Bluetooth transmissions are omnidirectional, operate at low power and travel through walls and around corners to a radius
of 50-100 ft. It is not necessary to aim the VetGard at the computer or printer to initiate
commands. The computer and/or printer can be in another room if desired. The print
report can be printed from any printer connected to the computer being used with the
VetGard.
Power is supplied by three AA alkaline batteries and these batteries will last for over 50
hours of continuous use. Battery life increases markedly by substituting long-life type AA
batteries. Rechargeable AA batteries may also be used however a charger is not available
through Vmed. Refer to the maintenance instructions at back of this manual for battery
changing instructions.
Software
VetGard software
The Windows application program installed on your computer for the processing and
display of real time patient information. VetGard software can be installed on multiple
computers without limit and simultaneous instances of the program can be run subject to
computer capability.
Reviewer
The Vmed Reviewer software provides automatic storage and retrieval of ECG sweep
and patient information, provides and assist in diagnosing ECG signals and facilitates
email transmission of patient files. Refer to Chapter 4 for a complete description.
Windows Display
To ensure that the entire screen is displayed, set computer screen resolution to
1024 x 768 by selecting START>SETTINGS>CONTROL
PANEL>DISPLAY>SETTINGS tab.
VETGARD KEYPAD CONTROLS
REPLACE
BATTERIES
CONNECTION
STATUS
PC-VetGard
Wireless Surgical Monitor
VETERINARY USE ONLY
Vmed Technology
SELECT
ON/OFF
Press and release to turn ON; press and hold
down to turn OFF. VetGard will shut off automatically if a
computer wireless connection is not established within three
minutes.
ON/OFF SWITCH:
SELECT SWITCH: Connects to icons on the display tool bar
for remote selection of options such as SPECIES, SWEEP
SPEED, ALARMS, TONE, SCALE, GRID, etc.
UP/DOWN ARROWS: Scroll through the various menus available with
the SELECT switch. Press the SELECT after scrolling to the
desired selection.
Turns off the audio portion of the
SILENCE ALARM SWITCH:
alarm display during and alarm event.
SILENCE
ALARM
STOP/
START
PRINT
STOP/START SWITCH:
Stops and re-starts the ECG sweep
PRINT:
Sends information on the screen to the printer that is
connected to the computer and to the Reviewer.
40
Figure 2 Computer display of VetGard information.
41
Sweep and Scale
Sweep: Sweep moves across the display left to right at the selected sweep speed. Default
sweep speed is 50 mm/sec. Sweep is magenta if sensors are off or not properly attached to
the patient. ECG sweep is green with good electrode-patient contact; Spo2 is blue with good
sensor placement; and respiration is yellow if the ECG electrodes is detecting respiration.
Vertical Scale:
The vertical scale on the display and print report measures signal
amplitude. The default gain setting is 1 for all animal species. With the gain set at 1, the
signal amplitude of 1 mV has a vertical scale deflection of 10 mm by convention. Vertical
scale deflection increases by a multiple of the gain. The amplitude of the square gain reference
marker will change with gain setting so that the signal amplitude can always be referenced to a
gain of 1.
Remote Function Selection
1. Turn on the VetGard and establish connection to the computer.
2. Press the SELECT switch on the VetGard keypad and the following menu appears:
Horizontal Scale: Time is measured on the horizontal scale and marked on the
display and print report every second.
Display Toolbar
SWEEP R WAVE
RATE
MARKER
GRID
DISPLAY
SELECT
Sp02 WAVE ALARM
ECG
GAIN
PATIENT
INPUT
SILENCE
ALARM
SILENCE DISPLAY
BEEP
RESPIRATION
WAVE
CONTINUOUS
REVIEWER RECORDING
3. Use the UP/DOWN arrows on the VetGard keypad to select the function and then
press the SELECT switch again to choose the option. Repeat the process for each
function
PRINT
ABOUT VETGARD
SOFTWAR
4. These functions may also be selected from the icons on the toolbar.
NOTE: These functions may also be selected from the icons on the toolbar
Figure 3 display toolbar
42
43
SELECT
Silence Alarms
Species
Sweep Rate
Gain
R-Wave
Marker
Grid
Beat Tone
Record ECG
Alarms
OPTIONS
Press to silence audio
alarm
Canine
Feline
Equine
Other
25 mm/sec
50 mm/sec
100 mm/sec
0.5: 55 mm/mV
1.0: 10 mm/mV
2.0: 20 mm/mV
3.0: 30 mm/mV
4.0: 40 mm/mV
ON
DEFAULT
Keypad & Toolbar
OFF
ON
OFF
ON
OFF
Press for continuous
Reviewer recording
ON
OFF
OFF
3
Chapter
Gain 1
Gain 1
Gain 1
Gain 1
50 mm/sec
Chapter 3 Operating Instructions
Preparation
1.0: 10 mm/mV
PROCEED ONLY AFTER YOU HAVE SUCCESSFULLY INSTALLED THE SOFTWARE AND
HAVE COMPLETED THE BLUETOOTH DISCOVERY PROCESS.
Note: Ensure that the Bluetooth icon at the lower right corner of your desktop is white in
the center and not red. If the icon is red, the Bluetooth device is disabled. To correct
this, right click on the icon and select “Start the Bluetooth Device”.
Table 1 “Select” key options and factory default values
1. Plug the sensor of choice into the appropriate VetGard receptacle.
OFF
2. Double click the VetGard icon on your desktop to start the program. This will initiate the
search process.
ON
Keypad & Toolbar
3. Turn on the VetGard by briefly depressing and releasing the red ON/OFF power switch
on the keypad (Do not hold the power switch down). The STATUS light will turn red
and then yellow.
ON
4. Wait for the STATUS light to turn from yellow to green. The process may take as long as
30 seconds to complete the first time after initial discovery.
5. When the VetGard status light blinks green, the notation in the STATUS box on the
screen will change to “Connected” and the notation at the lower right corner of the screen
will change to: “VetGard Remote Connected” on a green background.
NOTE: To preserve battery life, the VetGard will shut off automatically if
connection to your computer is not established within 3 minutes.
44
45
When connection is complete, the sweep appears on the display as a flat magenta
colored line moving left to right at the selected sweep speed. Upon acquisition of a
satisfactory signal, the waveform will appear and the trace will change to a color
corresponding to each sensor:
Green: ECG
Blue: Sp02
Yellow: Respiration
White: (digits only): Temperature
Using Chest Probe
1. Place the patient in the right lateral recumbent position and pull the front legs cranial.
2. Liberally apply alcohol or electrolyte solution to the left lateral chest wall where the
electrodes will make contact. Position the chest probe over the heart, cable toward the
right shoulder, and press firmly enough that all three electrodes make good contact
with the body wall.
NOTE: A magenta ECG sweep represents a leads off condition and may occur
even if a waveform is present, however, the computer will not recognize a valid
ECG and the digital heart rate display will not appear. You should, in this case,
reposition the sensor or clip leads until a green sweep appears.
Taking the ECG
Important comments on ECG:
The VetGard obtains two types of clinical ECG information (1) rate and rhythm and (2)
QRS and P characteristics. For surgical and long term monitoring, rate and rhythm only
is needed to monitor patient status since changes in heart rate or significant variation of
the periodic interval of the beating heart may be indications of patient compromise.
Accurate Rate and rhythm can be obtained by any of the available ECG sensors provided
with the VetGard, leg clips and the internal positioned esophageal probe being the most
often used in surgery. For long term patient monitoring, leg clips or disposable chest
electrodes should be used.
Accurate diagnostic lead II ECG, on the other hand, requires information on both the
rate and rhythm of the heart and evaluation of the ECG complexes themselves.
Conductive abnormalities, for example are manifested by higher than normal R and S
amplitude. For routine ECG and pre-surgical screening, the chest sensor is most
convenient and is recommended for this application, however the method may produce
inaccuracies in amplitude if the chest probe is not positioned at the correct angle to the
heart. For this reason, whenever a beat characteristic or signal amplitude seems abnormal
when using the chest sensor, a follow up trace should be taken using the leg clips to
confirm the reading.
Note: The VetGard is packaged with a small bottle of saline electrolyte. Alcohol is
the best contact medium for animal ECG, so you might want to refill the bottle with
alcohol if the electrolyte does not give you good results.
46
Figure 4 Correct position of the chest probe for ECG exams.
3. Rotate the probe until the “R” amplitude is at its highest and a good tracing appears on
the screen. If a good tracing is not obtained, re-saturate the chest wall with more
electrolyte and reposition the chest probe. In some cases excess hair should be clipped.
NOTE: Orientation of the chest probe is important to achieve good lead II
information. Confirm all suspect “R” wave signals with the leg clips.
4. Record the ECG until three good lines of tracing are obtained. NOTE: The three
panels stacked vertically on the display represent a continuous trace with the
earliest information being replaced by the current sweep. At the default chart
speed of 50 mm/sec, the three panels represent 12 seconds of data.
5. Depress the Start/Stop button on the chest probe or on the VetGard keypad to stop
the trace. Depress the Print button on the keypad to print this display and save it to
the Reviewer. You may also print directly from the moving trace by pressing the
PRINT switch on the keypad.
47
Using Leg Clips
The three lead clips are color coded according to American Heart Association (AHA) standards.
The following table provides the list of AHA color codes for Lead II connections modified to
identify animal legs. See the table on the next page.
LEAD
BIPOLAR
LEAD & REFERENCE
COLOR
II
RF-Right Front
Negative Electrode
White
RH-Right Hind
Reference
Green
LH-Left Hind
Positive
Red
NOTE: For an accurate Lead II reading, each pair of legs must be parallel to one
another but not touching. Use standing or right lateral recumbence position.
1.
2.
3.
4.
Plug the clip lead cable into the PC-Vet receptacle
Place animal in the right lateral recumbent or standing position
Saturate the area where clips attach with alcohol or electrolyte
Connect leg clips to the appropriate limb. Connect two clips to the hind legs, Red to
left hind and Green to right hind, on the posterior aspect of the stifle just proximal to
or distal to the point of the elbow. Connect the remaining clip, White to the right front
leg on the anterior aspect of the leg just proximal to or distal to the elbow.
Using Esophageal Probe
Vmed esophageal probes have sensors for monitoring ECG and temperature. Choose the
right probe for the size animal. If the baseline tends to wonder or deviate from a straight
line, the electrode spacing is too large and a probe with a shorter spacing should be used.
If the ECG signal is intermittent, usually with respiration, the electrode diameter is too
small for the animal and a probe with larger diameter electrodes should be used.
NOTE: Do not try to evaluate ECG traces using the esophageal probe. Sweep is
accurate for rate and rhythm only.
CAUTION: Do not use the rectal temperature probe while monitoring with
esophageal probe as inaccurate temperatures will be displayed.
1. Lay the probe on the animal simulating it being in the esophagus.
2. Place end of probe one finger width from the last rib.
3. Note the distance marked on the probe in cm at the nose.
4. Slide the probe into the esophagus until the distance marked at the animal’s nose is
at the mouth.
5. Move probe cranial and caudal until the best ECG signal is obtained.
Using Disposable Chest Electrodes
NOTE: The clip leads provided with the VetGard connect to the metal tabs on
disposable pre-gelled electrodes. Use 3M 2258-3 infant, long-term monitoring
electrodes for small animals; 3M 2270 for large animals. Disposable chest electrodes
are best for long term monitoring when the animal is under or recovering from
anesthesia. It may be necessary to shave excess hair and to place a protective
bandage around the animal covering the electrodes so they are not dislodged or
chewed off by the patient. Clip leads connect to the metal tabs.
1. Plug the three clip leads into the corresponding colored receptacle on the yoke cable.
2. Connect a clip to each of three electrodes.
3. Prepare the electrode site by clipping excess hair.
4. Wet area with alcohol and wipe dry.
5. Remove the liner from three electrodes and on each leg indicated by the color code on
the cable, preferably above the hock and elbow. Electrodes may also be placed on the
chest, over the heart and oriented to the Lead II axis.: White on the right chest toward
the right front leg, Red in the middle and Green on the left chest toward the left hind
leg. NOTE: A protective bandage should be applied over the electrodes to
prevent the animal from dislodging or chewing off the electrode or lead. Lead
II information is most accurate when electrodes are attached to the patien’s legs.
6. Smooth down the electrodes with a circular motion. Avoid pressing on center of
electrode.
48
49
™ Slide one of the sensors alignment buttons along the clip slot until the sensor is fully
engaged in the clip. (B)
Using the Rectal Temperature Probe
Plug the rectal probe into the receptacle at the front of the VetGard. Insert the tip into the
rectum to a safe depth. Wrap the Velcro strap around the base of the tail to stabilize and
secure the rectal probe. CAUTION: Do not use with esophageal probe inserted or
inaccurate temperatures will be displayed.
Using the Oxygen Saturation (Sp02) Sensor
™ Slide the second sensor button along the slot in the other side of the clip until fully
engaged as in step 3. (C) NOTE: Be sure the optical components face each other.
The sensor is now ready to be applied to the animal. The preferred application site for
canine, feline and equine patients is on the tongue with the sensor positioned at the center
of the tongue as shown (D). Alternatively, the sensor may be applied to the animals lip,
toe, ear, prepuce or vulva.
™ Failure to apply sensor properly may cause incorrect measurements.
™ Exposure of bright light at the point of sensor attachment may cause inaccurate
measurements. Cover sensor with an opaque material to block ambient light.
™ Move sensor to a new site at least every four hours. It may be necessary to move the
sensor more often with some patients.
™ Do not use supplemental tape or straps to secure sensor to the site since this may
restrict blood flow and cause inaccurate readings.
™ Use of Intravascular dyes may cause inaccurate measurements.
™ Motion, particularly repetitive motion, may compromise performance. If readings
cannot be obtained, change the site.
™ Carefully route all monitoring cables to reduce the possibility of entanglement or
strangulation.
™ Do not use during MRI scanning.
™ Do not alter or modify the sensor in any way since accuracy will be affected.
™ Do not use any other brands of Sp02 sensor with the VetGard.
NOTE: Each Sp02 reusable sensor comes with two clips that are used to apply the
sensor to different sized animals and to different sites. The sensor and clips are
indicated for use when continuous non-invasive monitoring of arterial oxygen
saturation is desired for canine, feline and equine animals. Reusable sensors may
be used on the same site for a maximum of 4 hours provided that the site is
inspected routinely to ensure skin integrity and correct positioning. Some patients
may require moving the sensor more often.
™ Clean the sensor and clip separately before and after each use (Refer to cleaning
instructions).
NOTE: If the sensor does not track the pulse reliably, it may be incorrectly
positioned-or the sensor site may be too thick, thin, or deeply pigmented to
permit appropriate light transmission. If any of these situations occur, reposition
the sensor or try another sensor site. If the sensor site is covered with furor or
hair, try shaving the site and reapplying the sensor.
1. Attach the sensor as described above. Ensure that the sensor cable is positioned along
the side of the animal’s face and body to avoid entanglement with the animal. (E)
2. Plug the sensor connector into the receptacle on the VetGard and verify proper
operation.
NOTE: To verify correct placement of the sensor on the site, manually take a
pulse rate reading at another site on the animal. If the sensor is placed
correctly, the manual reading should correspond to the pulse rate reading
displayed on the VetGard computer screen.
CAUTION : For the comfort of the animal and to avoid damaging the sensor, do
not pull on the cable when removing the sensor and clip from the sensor site.
3. To remove the sensor and clip from the animal, press the clip open and remove.
4. To remove the sensor from the clip, grasp the end of each sensor pad and pull it through
to the inside of the clip. The sensor should pop out of the clip easily. Do not pull on the
cable.
™ Open the clip by pressing with the thumb and forefinger as shown (A)
50
51
Printing and Storing Records
Once you have recorded a satisfactory ECG trace you may print the information directly
from the VetGard keypad or later from the Reviewer program provided the ECG sweep was
the selected mode of recording. NOTE: See Reviewer instructions in Chapter 4.
TO PRINT REPORT WHILE MONITORING
1. From the VetGard keypad: Press Print, or….
2. From the display upper tool bar: Click the print icon, or…
3. From “File” choose print. (Check “For Print” on the printer setup dialog box to
ensure best print resolution.)
NOTE: Printing from the screen requires wireless connection between the
VetGard and your computer. In other words, you can only print from the
screen when there is an active connection between the VetGard and your
computer. Printing from saved and recalled files from the Reviewer does not
require a wireless connection.
4. You may save reports to a PDF file by selecting “PDF Acrobat” or other PDF
program in the “Print To” selection box of the print dialog box. (See screen below)
5. Press OK to store the file in print format.
Figure 5 Computer’s print dialog box
A sample of the print report appears on the next page. Note the Vet/Clinic information
(shown as Vmed), automatic date and time, heart rate, sweep rate, ECG gain and filter.
Temperature is displayed only if esophageal or rectal probes are used. A color printer is
required to reproduce a colored grid.
52
A sample of the print report appears on the next page. Note the Vet/Clinic information
(shown as Vmed), automatic date and time, heart rate, sweep rate, ECG gain and filter.
Temperature is displayed only if esophageal or rectal probes are used. A color printer is
required to reproduce a colored grid.
53
4
Chapter
Chapter 4 Reviewer Program
Description
The Vmed Reviewer program is included with each Vmed wireless monitoring product
and is also available free to any interested person in order to facilitate ECG interpretation
by third party consultants. The Reviewer program performs the following functions:
™ Provides a patient database where information recorded from Vmed wireless
monitoring products may be stored automatically.
™ Allows archived ECG sweep, patient and episode data to be recalled for review,
editing, measurement and printing.
™ Provides for the electronic measurement of ECG complexes
™ Allows real-time playback of previously recorded ECG sweep data.
™ Makes available, through the Help menu, a table of normal ECG measurements
for cats and dogs and guidelines for measuring ECG complexes.
™ Provides an easy method of summarizing and storing ECG measurements, vital
signs and diagnostic notes.
™ Provides and easy way of sending email transmission of stored reports.
™ Automatically documents and stores vital signs every 5 minutes.
Modes of Operation
ECG and related data are stored automatically in one of three available modes. Record
modes are selected in the PC-Vet Configuration window:
1. Continuous Event Recording (Typically used in surgery
monitoring)
Figure 6 Sample VetGard ECG print report [Note signal amplitude (1 mV) reference scale
and time reference points on horizontal axis]. Temperature readout requires use of
esophageal probe.
54
™ Records vital signs only on event (Print command, alarm activation, or change
of record mode).
™ Does not record ECG sweep
™ Records Vital signs every five minutes.
NOTE: This is default mode for the VetGard patient monitor
55
2. Event Initiated ECG Sweep Recording (Typically used in ECG
exams)
3. Highlight the patient of interest and then the episode as above and press OK and
the next screen will appear:
™ Vital signs and 30 second ECG sweep are recorded on event (Print command,
alarm activation, change of record mode). NOTE: This is the factory default
mode for the PC-Vet.
3. Continuous ECG Sweep Recording (Typically used in long-term
monitoring)
™ Continuous recording of ECG sweep and event initiated vital signs for the life
of the batteries (about 50 hours).
™ Activated with Record icon on the PC-Vet toolbar
NOTE: This mode selection overrides and supersedes any previously
selected mode.
Using the Reviewer Program
1. Click the Reviewer icon on the desktop and the Reviewer window will open:
2. Click “File” then “Patient episode on the upper toolbar and the window below will
open:
NOTE: This screen summarizes the patient file for the episode selected along with
cursor measurement of the QRS complex. The measurements selected have been
dropped into the diagnostics notes and comments box. Note the red rubber band
vector and the measurement crosshairs above.
Using the electronic cursor measurement feature:
1. First click on the complex you wish to measure in the “Measurements” box at the top
of the screen. Note: For a description of each complex, click on the “Help” menu
in the upper toolbar.
NOTE: Patients are filed alpha-numerically. If a patient is not pre-identifie,
patient files are stored numerically by year, month and day. Patient data can be
entered prior to, during, or after the monitoring procedure and edited by using the
Reviewer toolbar. Episodes are identified by date and time of the event.
56
57
2. Next, left click the mouse at the beginning point of your measurement, then drag
the mouse to the right and up if amplitude is to be included. Note the red rubber
band in the above illustration. Place the blue crosshair at the end of the interval
and the top of the amplitude and release the mouse button. The one step
measurement will result in an interval and amplitude calculation if appropriate for
that complex. NOTE: For best accuracy, increase the gain of the signal
before beginning the measurement procedure.
™ Manage Patient Episodes: Selecting this choice opens the screen below:
3. Repeat the procedure in step 2 for each measurement.
4. Click the “Copy to Diagnostic Notes” box to drop your measurements into the
diagnostics notes section of the report.
Upper Tool Bar Icons
This screen allows you to remove unwanted patients and episodes. To delete a
patient, highlight the patient and press the “Delete Patient” button. To delete an
episode, highlight that episode and press the “Delete Episode” button. NOTE:
to delete episodes, you must first highlight a patient.
™ Configure Vmed Reviewer Selecting this option opens the screen below:
NOTE: The first row contains the command buttons like any Windows
toolbar. The “Tools” button deserves some comment, however. The “Tools”
command includes the following options:
58
NOTE: This window allows you to select the directory where patient
files are stored. The default directory is “My Documents” You can also
change the color of the printed report grid as well as the grid color in the
Reviewer.
59
The second row of the upper toolbar displays the icons listed below. NOTE:
Some of these icons are duplicates of commands that appear under the “File”,” Edit”
and “View” menus.
NOTE: You can print one or both of the following reports from the Reviewer
toolbar: The following window will open when the print command is initiated
allowing you to make your selection:
™ Open Episode: Opens patient and episode window
™ Save Episode: Saves the information on the window displayed
™ Email Episode: Opens the Email window with the file as an attachment
NOTE: If your computer is equipped with MAPI capable email like
Outlook or Outlook Express, you can email the report using this button.
Otherwise, the appropriate file may be opened from the “My
Documents”>”Vmed Episode” file. Select the patient and episode of
interest and select “Send To”> “Email Recipient” or drag and drop the
file directly into your email “Attachments box”.
™ Edit Clinic/Vet information: Opens Clinic/vet window for any changes
NOTE: The controls described below are only visible if you have
selected an episode that includes an ECG sweep, otherwise these icons
will be grayed out.
™ Change Sweep Speed: Select chart sweep speed in drop down window
™ Change Gain Setting: Select desired ECG gain from drop down window.
Again, it is best to increase the gain setting in order to obtain more accurate
cursor measurements.
™ Add “R” Wave Markers: Places triangular marker over each “R” wave to
help identify this complex.
™ Up/Down Arrows: Changes the position of QRS complex displayed in
the sweep area and brackets the sweep to be printed. Use this control to
show the entire sweep complex in the window before printing.
™ Episode time display: The time that the episode was recorded is
displayed here.
™ Video Controls: for the real-time playback of recorded sweep and to stop,
start and scroll the ECG record.
™ Print Command: Prints to any printer connected to the computer being
used.
™ Question Mark: displays version of the Reviewer program
60
Figure 7 ECG Waveform report printed from the Reviewer
61
Figure 8 Event Summary Report includes vital signs, measurements and clinical
comments.
Selecting “Help” from the Reviewer upper toolbar allows you to open a window for
assistance in defining and measuring ECG complexes and to display a table of normal
measurements for feline and canine patients. These selections are shown below:
Figure 10 Table of Normal ECG measurements for canine patients from the “Help”
menu.
Figure 9 Instructions for measuring ECG complexes from the “Help” menu
62
63
5
Chapter
Chapter 5 Service and Maintenance
Service and Support
If you purchased your Vmed product directly from Vmed
Phone:
FAX:
Email:
WEB:
425 497-9149
424 497-0260
[email protected]
www.vmedtech.com
If you purchased your Vmed product from DVM Solutions or your Dealer:
Phone: 866 373-9627
FAX: 830 438-7041
WEB: www.dvmsolutions.com
Cleaning and Maintenance
Cleaning
Except for the Sp02 sensor (Refer to using the Sp02 sensor on page 46 of this manual), clean the
VetGard, sensors and cables by wiping surfaces with any of the following solutions:
• Soap and water
• Alcohol
• Peroxide solution
Figure 11 Table of Normal ECG measurements for feline patients from the “Help” menu
Information in Tables 7, 8 and 9 from Electrocardiography for the Small Animal Practitioner, Larry P. Tilley,
DVM, Diplomate, ACVIM (Internal Medicine), Naomi L. Burtnick, MT (ASCP). “Made Easy Series”,
Teton New Media.
64
Precautions
™
™
™
™
Do not immerse or soak the device or cables in any fluid
Do not use bleach, bleach solution or keytones
Do not steam autoclave or gas sterilize
Do not use petroleum based solvents other than alcohol
65
Maintenance
There are no field serviceable components inside the VetGard. Opening the housing voids the
warranty. Store the VetGard in the storage pouch when not in use.
Changing Batteries
To change the three AA batteries, open the battery compartment lid on the back of the VetGard housing.
Replace ALL three batteries using Alkaline AA or long-life cells only. Note polarity when replacing
batteries. Insure that each new battery makes proper spring contact and is pushed all the way into the
bottom of the battery compartment.
SITUATION
POSSIBLE CAUSE
Cannot make Bluetooth
connection with computer.
Weak battery on VetGard
Improper Bluetooth discovery.
Bluetooth not enabled
Bluetooth connection has
range <50ft.
Cannot acquire satisfactory
ECG signal
Excessive noise in the
ECG signal
Cable Connectors
Sensor plugs should be inserted into the receptacle opening by pushing the plug straight into the
receptacle. Note the Vmed logo is on the top and the plug key is on the bottom.
Troubleshooting
Refer to the table on the next page to trace common problem areas.
Lost wireless connection
with computer
Low Batteries
Signal must pass through high
attenuation barrier such as
concrete floor/walls or heavy
metal objects
Not enough electrolyte fluid
used.
Not making adequate skin
contact
Corroded electrodes.
Broken or corroded lead wire,
cable or plug.
Bad cable connection
Inappropriate filter selected.
Poor electrode-skin contact.
Excessive motion.
GAIN set too high
RFI or electromagnetic
interference.
VetGard out of range of
computer or behind signal
attenuator.
VetGard lost power
Bluetooth device loose in USB
port.
Computer turned off or
VetGard program closed
Table 2 Trouble Shooting Guide
66
67
CORRECTIVE ACTION
Replace with three fresh AA
batteries
Re-initiate discovery procedure
Enable Bluetooth on lower
toolbar icon
Replace both batteries
Clear signal path
Re-apply electrolyte.
Re-position electrodes and clip
excessive hair.
Replace electrodes.
Replace lead wire or cable
Remove and re-insert all plugs
Change filter selection in the
“Tools” drop down menu.
Re-apply electrolyte and reposition electrodes. Reduce
patient motion.
Set GAIN to 1.
Shut off electrical devices near
the VetGard
Move toward computer or clear
attenuating obstacle.
Check VetGard power on
switch
Check /Replace Batteries
Check Bluetooth device/USB
connection.
Re-start VetGard program.
System Specifications
Appendix A
Appendix A Computer Requirements and Specifications
Computer Requirements
•
•
•
•
•
•
•
•
>200 MHz. PC desktop, laptop or tablet computer
Windows XP, 2000
32 MB RAM
SVGA display, 256 colors, 1024 x 768 resolution
Spare USB 1.1 port
CD Rom Drive
60 MB available hard disk space for VetGard program, plus approximately 100k Bytes per patient stored
Windows supported printer
• MAPI compliant E-Mail system such as Microsoft Outlook or Outlook Express.
Physical Specifications
TYPE: Multi-parameter patient monitor with integrated Bluetooth wireless module
CONSTRUCTION: High impact ABS-PC plastic.
DIMENSIONS OVERALL
LENGTH:
6.0”
WIDTH:
3.75”
THICKNESS:
1.1”
WEIGHT (WITH BATTERIES)
14 oz.
CONTROLS: FLAT PANEL KEYPAD SWITCHES, TACTILE PUSH TYPE:
ON/OFF
SELECT
UP/DOWN ARROWS
STOP/START
SILENCE ALARMS
PRINT
INDICATORS:
CONNECTION STATUS LED
REPLACE BATTERIES LED
CABLE CONNECTION: ECG: Circular mini-DIN, 6 pin. Sp02: RS 232, Rectal temp.: 2.5 mm phone plug
BATTERIES:
TYPE: AA Alkaline, Long life or rechargeable
QUANTITY: 3
BATTERY LIFE: >50 hours continuous operation
PATIENT CABLES AND LEADS:
36” Interface cable with three lead receptacle yoke and connector
Three electrode chest probe with 32” cable and connector
Three wire, 36” Zip cable with low force skin clips.
Rectal temperature sensor with 5’ cable
Esophageal probes with 36” cable
Sp02 sensor with 36” cable.
68
ECG GAIN/SENSITIVITY: 5, 10, 20, 30 and 40 mm/mV
FREQUENCY RESPONSE:
COMMON MODE REJECION: -70dBV CMRL
SAFETY STANDARDS: IEC 601, AAMI EC-11
INPUT IMPEDANCE: 1 M Ohm
INPUT DYNAMIC RANGE: 15 mV
DC OFFSET CORRECTION: -0.5 Hz.
A/D SAMPLING RATE: 400 Hz.
SAMPLE RESOLUTION: 12 bit
ECG SWEEP SPEED, SELECTABLE: 25, 50, 100 mm/sec. (Default: 50 mm/sec.)
QRS DETECTION RANGE: >0.2 mV
FILTERS:
PERMANENT: Pass Band: High pass: 70 Hz
Low pass: 40 Hz
USER SELECTABLE:
Notch (AC line noise rejection): 50/60 Hz
ALARMS:
HEART RATE, SP02, RESPIRATION: High/Low, selectable limits
TEMPERATURE: Low limit selectable with auto step-down
FEATURES:
ALARMS:
ON/OFF
TONE:
ON/OFF
GRID:
ON/OFF
ALARMS:
ON/OFF
R-WAVE DETECTOR: ON/OFF
SILENCE ALARMS: USER ACTIVATED
WIRELESS: Bluetooth Class I radio
RANGE: 50-100 ft. in practice
FREQUENCY: 2.4-2.4835 GHz
OUTPUT POWER: 4 to 20 dBm
PHYSICAL INTERFERANCE: USB UHCI/OHCI 1.1 compliant
MANUFACTURER: Belkin
MODEL: F8T001
69
Appendix B
Appendix B Warranty
Vmed Technology warrants the VetGard monitor to be free of defects in material and
workmanship for a period of TWO YEARS from the date of purchase by the original end
user customer. Sensors including cables, lead wires and connectors sold with the VetGard
or purchased as an optional accessory, are warranted for a period of SIX MONTHS from
the date of original purchase. If in the judgment of Vmed Technology, the VetGard or
accessories are proven to be defective during the warranty period, it will be repaired or
replaced with no charge for parts or labor.
This warranty does not cover consumable supplies or products damaged by accident,
abuse, or misuse or products altered or repaired by anyone other than Vmed Technology
or its designated repair agents. OPENING THE VETGARD ENCLOSURE VOIDS
THE WARRANTY.
This warranty is exclusive and is in lieu of all other warranties whether expressed or
implied including warranties of merchantability and fitness for a particular purpose. The
obligations under this warranty are limited to repair or replacement.
A return authorization number is required before returning any product for warranty or
non-warranty repair. Products or parts shipped to Vmed for warranty service must be
shipped prepaid and insured via a carrier that provides package tracking such as UPS or
FedEx. Vmed will pay the return freight on products repaired under warranty.
70