view article - Portland Veterinary Specialists
Transcription
view article - Portland Veterinary Specialists
John M M G M U Q PVM. DipWVIM Clinical Situation in Cardiology iCjlI\l»|n^v) Laor* E, Iblton DVM Several episodes of collapse MASSACHUSETTS \ HTF.RJ N AR*i REFERRAL HOSPITAL CLINICAL CASE: A 12-year-old spayed female Shih Tzu, weighing 10.3 kg, presented to the Cardiology Service because of a three-week history of lethargy, exercise intolerance, and several episodes of collapse. She had a history of mild, asymptomatic, myxomatous mitral valve disease with minimal left atrial enlargement but no other significant medical problems. She was receiving no therapy for her valvular disease. The collapse episodes consisted of slowing down dramatically, becoming unsteady on her feet, and then collapsing. She would recover within several seconds and act as if nothing had happened. During her physical examination (PE), her heart rate was 110 beats per minute (bpm) and her rhythm was irregular with several one-second pauses. A grade 2/6 systolic left apical murmur was ausculted, consistent with previous examination findings. No other abnormalities were noted on thoracic auscultation. Her pulses were strong and synchronous and her mucous membranes were pink with a capillary refill time of less than 2 seconds. Her neurologic examination was normal. Screening bloodwork showed no abnormalities. Her packed cell volume (PCV) was 4 2 % and her total solids were 7.2 mg/dl, both of which are considered normal. Thoracic radiographs were taken (Figure 1), which demonstrate mild cardiomegaly but no other significant thoracic abnormalities. An echocardiogram was performed (Figure 2), which demonstrated a mildly leaky mitral valve with mild left atrial and left ventricular enlargement. An electrocardiogram (ECG) was performed as well (Figure 3). Woburo, MA 01801 Figure i (a) and (b). Thoracic radiographs of the patient. Right lateral (Fig. 1 (a) left image) and DV (Fig. 1 (b) right image) thoracic radiographs. Note mild cardiomegaly and mild left atrial enlargement (arrows). considta • 1 ANSWERS Figure 2 (a) and (b). 20 right parasternal long-axis echocardiography images. The left ventricle (LV), left atrium (LA), interventricular septum (IVS). left ventricular free wall (FW), and anterior leaflet of the mitral valve (MV) are labelled in Fig 2 (a) left Note mild to moderate left ventricular and left atrial enlargement and the thickened mitral valve. Fig. 2 (b) right shows the same structures as Fig 2 (a) left but this image is in systole and demonstrates mild to moderate mitral regurgitation (MR) with mild mitral valve prolapse. 1 Br • 33 ; LA," T » - Figure 3. ECG from current visit (lead II, 25 mm/sec, imV • 10 mm). Questions 1. What are some potential causes for the lethargy and collapse? 2. What rhythm is present in the ECG seen in Figure 3 ? 3. What additional testing modalities could be used to diagnose/confirm that this rhythm is the cause of the collapse and how are they best used? 4. What is a diagnostic test that could reveal whether high vagal tone could be partially responsible for the rhythm seen in Figure 3? If there is a positive response to this test, w h a t are potential therapies for this patient? 1 - POTENTIAL CAUSES FOR COLLAPSE There are multiple potential causes of any collapse episode. Broadly, the collapses could be due to neurologic causes, causes related to the heart that directly compromise the pumping function (pericardial effusion, pulmonary hypertension), causes related to heart rate/rhythm, causes related to blood loss (internal or external), or metabolic causes (which might include an Addisonian crisis among other potential causes). In this case, blood loss and common metabolic problems were given lower priority as potential causes of collapse since Woodwork was normal. Neurologic causes were thought to be less likely, given the description of the collapse episodes (no tonic/clonic motion, no preor post-ictal phase) and the fact that the patient's neurologic examination was normal. Structural cardiac pump problems were not considered likely, as the echocardiogram showed no evidence of either pulmonary hypertension or pericardial effusion. 2 • consutta Collapse can be caused by several different types of syncope. Tussive syncope ("cough drop" syncope) is a common cause of syncope, which typically occurs in small breed dogs with tracheal collapse, chronic pulmonary disease, and/or brachycephalic syndrome. Episodes of syncope typically occur during or immediately after coughing. One mechanism proposed to explain "cough drop" syncope is that coughing causes an increase in intrathoracic pressure, which causes a transient increase in intracranial pressure with subsequent diminished cerebral blood flow. Coughing may also cause a decrease in cardiac venous return and, therefore, limit cardiac output. Alternatively, coughing causes vagal stimulation to the heart and blood vessels, resulting in bradycardia and hypotension (i.e. reflex cardiac slowing +/- peripheral vasodilation similar to neurocardiogenic syncope). In this patient, tussive syncope was thought to be unlikely, as coughing was not a feature of the collapse episodes. Autonomic dysfunction (vasodepressor or neurocardiogenic syncope vs. reflex-mediated syncope) is the result of withdrawal in sympathetic tone with a concomitant increase in parasympathetic activity. The result is hypotension and bradycardia causing a decrease in cerebral blood flow with subsequent syncope. It is considered an abnormal and inappropriate activation of the baroreceptor reflex, which in humans typically occurs in hypovolemic or hypotensive patients. In dogs, these episodes are typically brought on by exertion or excitement and it is not apparent that dehydration is a predisposing factor. Neurocardiogenic syncope can occur in normal healthy animals or in animals with cardiac disease. It commonly occurs in small breed dogs with advanced mitral valve disease. In these patients a hyperdynamic left ventricle may occur with severe mitral regurgitation. This may mimic the hypovolemic human patient in that high sympathetic tone will increase cardiac contractility and, with concurrent severe mitral regurgitation, will cause an "empty ventricle". Sympathetic withdrawal then occurs in these patients, resulting in vasodilation, increased vagal activity, and subsequent bradycardia and collapse. Neither cause of autonomic dysfunction was thought to be a cause of syncope in this patient. ANSWERS In this patient, arrhythmia was detected during the PE and was consistent with several types of arrhythmias that can cause collapse when the severity is sufficient. 2 - THE RHYTHM PRESENT IN THE ECG SEEN IN FIGURE 3 B t i 1 • ' t- T • Figure 3 (labelled). ECG from current visit (lead II, 25 mm/sec, lmV - 1 0 mm). Arrows indicate P waves that are blocked (no QRS complex follows the P waves). The arrowhead demonstrates a ventricular escape beat. The 'A' distance Is a period of sinus arrest of 3.6 seconds. Complex "8" is a premature supraventricular complex. The rhythm present is second-degree Mobitz type II atrioventricular (AV) block with a period of sinus arrest, a premature supraventricular beat, and a ventricular escape beat. Atrioventricular block is characterized by either the delay or complete extinguishing of an electrical impulse through the AV node and is generally due to AV-node injury or fibrosis. Normal sinus beats originate in the sinus node, travel through the atria via Bachmann's bundle and reach the AV node. The depolarization of the atria that this causes is seen on the ECG as a P wave. The impulse travels through the AV node at a somewhat slower speed, allowing the ventricles to be filled by the contracting atria prior consutta • 3 to ventricular depolarization. In first-degree AV block, the atrial impulse is abnormally delayed, resulting in a prolonged P-R interval. In second-degree AV block, one or more of the atrial impulses do not pass through the AV node. There are two types of second-degree AV block - Mobitz type I and Mobitz type II. In Mobitz type I second-degree AV block, there is successive prolongation of the P-R interval with each beat until one P wave is blocked completely. First-degree AV block and Mobitz type I AV block do not cause clinical signs and, generally, are not treated. In Mobitz type II second-degree AV block, prolongation of the P-R interval does not occur prior to the P wave being blocked. Mobitz type II second-degree AV block can be described by the ratio of total P waves to conducted P waves. As an example, when only every 9th P wave is conducted, it is called 9:1 second-degree AV block (Figure 4). Mobitz type II second-degree AV block can cause clinical signs depending on the severity of the AV block. ANSWERS t•i H ^ ^ ^ ^ Figure 4. Example of 9:1 second-degree AV block (lead II, 25 mm/sec, ImV - 1 0 mm). Single-headed arrows indicate P waves and arrowheads indicate QRS complexes. Double-headed arrows demonstrate the P-R interval from the conducted (ninth) P waves. In third-degree AV block, none of the P waves are conducted through the AV node (Figure 5). Ventricular depolarization occurs secondary to development of an "escape rhythm" that emanates from pacemaker cells either in the more distal part of the AV junction (the His bundle) or in the ventricles. When the escape rhythm emanates from the AV node or Bundle of His, the QRS complexes are narrow with an escape rate of 40 to 60 bpm. When the escape pacemaker is in the ventricular myocardium, the QRS complexes appear wide and bizarre and the escape rate is 20 to 40 beats/minute. Third-degree AV block is more common in older animals, but can occur at any age. Patients may present for lethargy, exercise intolerance, or collapse. Some animals will develop signs of congestive heart failure causing dyspnoea, tachypnoea, coughing, or ascites. Physical exam findings reveal bradycardia (HR typically < 50 beats/minute in dog). Intermittent jugular pulsations called "cannon a waves" may be present, which are caused by intermittent right atrial contraction against a closed tricuspid valve. The first heart sound may vary in intensity secondary to variability in the end-diastolic ventricular volume, a phenomenon called "bruit de canon". Some animals will also have signs of low-output failure (i.e. forward heart failure) or congestive heart failure. At times it is difficult to distinguish between high-grade second-degree AV block and third-degree AV block and calling it advanced AV block suffices, as it describes both conditions. ii t • t • M • t y •*—M T I \ • Figure 5. Third-degree AV block (tead II, 25 mm/sec, ImV • 10 mm). Single-headed arrows indicate P waves. Double-headed arrows demonstrate the P-R interval. Note the variable P-R interval, which Is consistent with third-degree (complete) AV block. 4 •consutta i Sinus arrest is defined as a failure of a normal impulse to be formed in the sinus node owing to a depression of automaticity in the sinus node. This is often difficult to distinguish from sinus node block, in which the impulse is formed normally but a conduction abnormality exists, preventing the impulse from reaching the rest of the heart. Both result in pauses where no P wave is seen. In sinus block, the R-R length of the pause is an exact multiple of the R-R intervals preceding the pause, whereas in sinus arrest the pause can be any length greater than 2 R-R intervals. In practice it is often quite difficult, and clinically irrelevant, to distinguish between the two underlying causes of lack of P wave formation and the resultant ECG is simply called sinus arrest. 3 - ADDITIONAL TESTING MODALITIES THAT COULD CONFIRM THAT THE CAUSE OF THE COLLAPSE EPISODES WAS DUE TO THE RHYTHM DEMONSTRATED A N D HOW THE TEST SYSTEMS ARE BEST USED. A 24-hour ambulatory ECG monitor (Holter monitor) or an event recorder could be used to demonstrate that the rhythms observed in the room second-degree AV block or sinus arrest - were the cause of the collapsing episodes. A Holter monitor records 24-72 hours of continuous ECG data. These data are recorded either on a cassette tape or digitally. The patient has 5-7 electrodes placed on their skin and then carries the monitor for the duration of the testing period (Figure 6). The data from the monitor are then analysed and a catalogue of the rhythm is generated via a computer program. A Holter monitor is most useful for determining the severity of the arrhythmia present in the testing period or to determine the efficacy of the anti-arrhythmic therapy. It is less useful for figuring out the cause of collapsing episodes that happen relatively infrequently. Figure 6.24-hour ambulatory ECG monitor. Note the cassette tape where ECG data are stored, battery compartment, and ECG cable. The ends of the ECG leads are attached to selfadhering electrode patches on either side of the thoracic cavity at the level of the 4-6" intercostal spaces. In order to determine accurately whether the cause of the episode is due to the arrhythmia seen on the Holter monitor, the patient diary needs to be synchronized relatively precisely to the time on the Holter monitor. In contrast, an external event loop recorder (Figure 7 (a)) has 2 electrodes and is worn for up to one week at a time. It records continuously but does not save data until a button is pressed on the recorder by the pet owner during an event. The recorder then saves an ECG loop. Since the data are saved in a loop, the recorder contains the ECG recording for several seconds prior to the button being pushed, the ECG when the button was pushed, and the ECG for several seconds after the button was pushed. This enables the event recorder to "reach back in time" to capture the ECG that was most likely to be associated with the clinical signs exhibited by the patient. These event recorders can be worn for up to a week at a time and are most useful when the patient is having 2 or more events weekly. The final option is an implantable loop recorder (Figure 7 (b)). This is a device that is implanted subcutaneously and can be programmed to record all low and high heart rate events. The data are read and programmed via an external unit (7 (c)) and can be used to determine whether rhythm events are the cause of rare collapse events. Figure 7 (a). External event loop recorder. Note the 2 electrodes that are attached to self-adhering electrode patches on either side of the thorax. Electrodes are placed over the heart in the middle of the thorax. Note the "Record" button in the centre of the recorder. In the event of a collapse episode, the patient's owner presses the button and the ECG loop that was recorded is stored. These data are then transferred transteiephonicaliy and an ECG is generated. Figure 7 (b). Implantable loop recorder. Note the US dime (18 mm) placed in the picture for size reference. The event recorder is placed subcutaneously and stays in the patient for many months. The event recorder can be programmed to record high and low heart rate Incidents and a remote control device can be used by the owner to mark events in the memory. Figure 7 (c). Pacemaker programmer. This is used to program the event recorder and to download information from the implantable event recorder. The programming head (arrow) is placed over the event recorder while the event recorder is still in the patient and information is transmitted from the event recorder to the programmer for analysis. 4 - DIAGNOSTIC TEST ANSWERS An atropine response test can be used to determine the influence of parasympathetic (vagal) tone on the heart rate and rhythm. Atropine is a competitive antagonist of the muscarinic acetylcholine receptors, which acts as a parasympatholytic and, as such, abolishes or diminishes the influence of the vagus nerve on the heart rate. This leads to increased firing of the sinus node and can increase conduction through the AV node, if this is being inhibited by high vagal tone. In this test, a baseline ECG is recorded and then 0.04 mg/kg of atropine is administered intravenously. Ten to 15 minutes later, a follow-up ECG is recorded. If the rhythm abnormality is due to high vagal tone, the heart rate in the follow-up ECG should be significantly faster and the abnormalities should normalize. The result is often sinus tachycardia. In this patient, atropine administration did result in sinus tachycardia (Figure 8). Figure 8. ECG obtained 15 minutes after atropine administration as part of an atropine response test (lead II, 25 mm/sec ImV - 1 0 mm). The heart rate is 150 beats per minute and the rhythm is sinus tachycardia. This indicates a positive response to atropine and demonstrates that the arrhythmia was at least partially due to high vagal tone. Potential therapies for this patient if there is a positive response to this test If there is a positive response to atropine, several oral medications may be used to help normalize the rhythm. Propatheline bromide is a quaternary antimuscarinic antichoienergic that can be given orally three times daily at a dose starting at 7.5 mg/patient q 8 hours. This can be increased to 30 mg/patient q 8 hours as needed. Common side-effects are dry mouth and eyes, anxious behaviour and Gl signs. Hycosamine, an anticholinergic alkaloid, can also be used at 0.003 to 0.006 mg/kg q 8 hours. In this patient. Propantheline bromide was initiated at 7.5 mg q 8 hours. At a recheck examination, the owner reported that the patient did not have any more collapse episodes and an ECG showing intermittent Mobitz type II 2:1 AV block (Figure 9) was recorded. No change was instituted in the therapy and the patient was discharged with instructions to return in 1 month for a recheck ECG. j i: - * i ' I -1 _ ;. ; * — i j r t 1'.Z' — j — : i 1 4 1 pi 1 i •ft • - Figure 9. ECG obtained after 1 week of propantheline bromide therapy (lead II, 25 mm/sec, ImV • 10 mm). The heart rate is 75-150 beats per minute and the rhythm is sinus tachycardia with Mobitz type II second4egree AV block present. Arrows represent blocked P waves. FOLLOW-UP The patient re-presented prior to the scheduled recheck examination with a return of collapsing episodes. An ECG was obtained (Figure 10 (a)) that was consistent with third-degree AV block. At the owner's request, the atropine response test was repeated but a satisfactory response was not obtained. The patient's heart rate rose slightly from 15-18 beats per minute to 20-25 beats per minute, but this was not considered to be sufficient improvement to have the patient lead a life with normal activity and not collapse (Figure 10 (b)). consutta • 7 ANSWERS Figure 10 (a). Third-degree AV block (lead II, 25 mm/sec, ImV • 10 mm). Red arrows denote junctional escape beats and blue arrows represent premature ventricular complexes. Note the variable P-R interval and extremely slow ventricular rate. t\ " A !i k Figure 10 (b). Third-degree AV block after the atropine response test (lead II, 25 mm/sec, ImV - 1 0 mm). Red arrows denote P waves "buried" in the QRS complex. Note the variable P-R interval consistent with third-degree AV block. Note also that the atrial rate has increased and that the junctional escape rate has increased as well. This Indicates that there was some vagal influence on the junctional escape rate. 5. What treatment options are available at this juncture? 6. What are some of the potential methods to stabilize the patient prior to surgery? 7. Describe the components of a pacemaker system and some of the potential pacemaker modalities that could be used in this patient. 5 - TREATMENT OPTIONS AVAILABLE AT THIS JUNCTURE In most cases of complete AV block, an underlying cause is not found and it is therefore presumed to be secondary to idiopathic fibrosis; however, an attempt should be made to address or diagnose any possible causes for this condition. Potential aetiologies include congenital defects (e.g. aortic stenosis, ventricular septal defect), infiltrative diseases (e.g. neoplasia or amyloidosis), excessive vagal tone, bacterial endocarditis, myocardial infarction, hypothyroidism, infection (e.g. Trypanasoma cruzi (Chagas disease) and B. Burgdorferi), hyperkalaemia, drugs (e.g. digitalis, beta-blockers or calcium-channel blockers). In these cases, treatment of the underlying condition should be attempted if possible. However, in this patient, none of the above conditions were suspected. The goals of treatment are to restore normal cardiac output and/or resolve congestive heart failure if this is present. Positive chronotropic medications can be administered, such as theophylline (5-15 mg/kg PO BID of the extended release formulation) or terbutaline (0.2 mg/kg PO BID-TID). However, response is variable and almost always unrewarding with third-degree AV block. Chronic treatment requires permanent artificial pacemaker implantation to resolve clinical signs and provide a more favourable long-term prognosis. In this case, pacemaker implantation was chosen. 6 - SOME OF THE POTENTIAL METHODS TO STABILIZE THE PATIENT PRIOR TO SURGERY Temporary artificial pacemaker implantation may be considered in patients that are at high risk for anaesthesia or require stabilization prior to implantation of a permanent pacemaker. This decision should be based on the stability of the patient, experience of the clinician in pacemaker placement, and availability of equipment. There is no agreed-upon standard 8 •consutta in veterinary medicine as to when temporary pacing should be used. Temporary pacing can use either transvenous or transthoracic approaches, based on the clinician's preference. Many cases that are haemodynamically stable can proceed directly to permanent pacemaker implantation. Typically some type of temporary pacing system is placed prior prepping patients for permanent pacemaker implantation in order to control the rhythm should the rate become even more bradycardic during induction. In this case, transthoracic pacing was used. Transthoracic pacing uses a defibrillator with pacing capabilities and surface electrodes that are attached bilaterally over the patient's thoracic wall. General anaesthesia is used to prevent pain associated with the electrical stimulation itself and, typically, transthoracic pacing is only administered if the patient is not haemodynamically stable during permanent pacemaker placement. ANSWERS 7 - DESCRIPTION OF THE COMPONENTS OF A PACEMAKER SYSTEM AND SOME OF THE POTENTIAL PACEMAKER MODALITIES THAT COULD BE USED IN THIS PATIENT. The critical components of a pacing system include a pulse generator that has been programmed appropriately and a pacing lead (Figure 11). The cardiac pacemaker functions as an electrical circuit, whereby the pulse generator (battery) provides electrical stimulation that travels through the pacing lead to the myocardium and then back to the battery to form a complete circuit. All pacemaker systems have two poles. Bipolar pacemaker leads are most commonly used and create a smaller circuit with both poles located near the end of the pacemaker lead. Unipolar leads have one pole located near the end of the lead, with the other pole encompassing the metal of the pacemaker generator. Unipolar leads therefore contain a larger circuit of electricity and can trigger activation of surrounding musculature. Bipolar leads only trigger stimulation of the ventricular myocardium and are less likely to pick up stray signals from the environment Figure 11. Pacemaker generator (GEN) and lead (Lead). Note active fixation, screw-in type tip (AF), and bipolar sites (arrows). Pacemaker modes are described using a 3- to 5-letter classification system (Table 1). This system categorizes pacing based on the site and mode of cardiac pacing and sensing. The first letter indicates the chamber paced, the second letter indicates the chamber sensed, and the third position describes the expected response to sensing. The fourth describes the programmable features of the device. The fifth indicates whether an anti-tachyarrhythmia function is available. consutta* g Pacemaker modes commonly used in small animals ! 1 Chamber(s) paced II Chamber(s) sensed III Response to sensing rv Programmable functions V • ventricle V = ventricle T = triggers pacing P = simple programmable A = atrium A = atrium 1 = inhibits pacing M - multiprogrammable 0 • dual chamber D = dual chamber 0 = triggers and inhibits pacing C = communicating functions 0 = none O-none O = none S*AOfV* S = AorV* R = rate modulating 0 = none *S: used by some manufacturers to indicate single-chamber (A or V) The most common pacing mode is W l in which the ventricle is paced (V); the pacemaker senses native cardiac depolarizations in the ventricle (V) and then inhibits pacemaker discharge d u r i n g these periods (I). W i t h conventional W l pacing, the patient's heart rate is fixed and does not change w i t h variable cardiac output demands of the patient. In contrast, W I R pacing is rate responsive and uses an activity sensor that attempts to correlate patient activity w i t h heart rate. Rate-responsive pacemakers (R) change the pacing rate based on parameters such as blood temperature, oxygen saturation or pH, ventilation rate, gravitation sensors, or accelerometer functions. The most common method of pacemaker placement uses a single-chamber pacemaker programmed to ventricular demand ( W l or WIR) modes. The pacing lead is inserted into the right ventricular apex through the jugular vein. An endocardial lead is advanced into the right ventricular apex using fluoroscopic guidance. The tip of the lead is either actively or passively attached to the RV endocardium. After the lead is securely in place, its proximal end is attached to the pacemaker generator. The pacemaker generator is placed in a subcutaneous pocket dorsal to the incision over the jugular vein. A subcutaneous tunnel is created that will pass from the jugular vein to the generator. Dual-lead chamber pacing (DDD, DDDR) uses a second lead placed within the right auricular appendage, which senses and paces this chamber. DDD pacing is beneficial in that it maintains synchrony between atrial and ventricular contractions. It is the most commonly used pacing mode in humans. DDD pacing has been described in veterinary patients, but is not as widely used due to technical limitations in small animals, expense, and complication associated with placing a separate atrial lead. This modality is particularly useful in advanced heart block, as it allows the patient's intrinsic sinus rhythm to control the rate of ventricular depolarization. Care must be taken to set an appropriately high baseline heart rate, as sinus node dysfunction often accompanies AV node dysfunction. Single-lead synchronous VDD pacing allows maintenance of AV synchrony without the need of two leads as in the DDD pacing system. These leads use a proximal set of electrodes on the outer surface of the pacing lead, which are able to detect atrial electrical activity as it is conducted through the right atrium. The atrial electrodes do not come into direct contact with the atrial tissue, but are sensed through "floating" atrial electrodes (Figure 12). Following proper programming, atrial depolarization will lead to ventricular pacing that mimics the normal AV conduction period. VDD pacing has been shown to increase stroke volume and cardiac output as well as to decrease left atrial size, pulmonary capillary wedge pressures, and circulating biomarkers. Overall, VDD or DDD systems should be considered for their haemodynamic advantage in patients with advanced atrioventricular block and normal sinus node function. Final Outcome In this patient a single-chamber pacemaker in WIR mode was implanted through the right jugular vein. The bottom rate was set at 80 bpm and the top rate was set at 140 bpm. Placement was confirmed with post-op radiographs (Figure 13 (a)). The patient was monitored overnight with continuous ECG monitoring and the rate was never below 80 bpm. The patient went home the following morning. In 10 days, the skin incision sutures were removed and an ECG was checked (Figure 13 (b)). Figure 13 (a) and (b). Thoracic radiographs post-pacemaker implantation. Fig 13 (a) left: right lateral thoracic radiograph showing the pacemaker generator (Gen) within the right lateral neck. The pacemaker lead then courses from the generator through the subcutaneous tissues, right jugular vein (Right Jugular), cranial vena cava (CrVC),rightatrium (RA), and finally into the right ventricular apex (RV) where the lead tip comes into contact with the right ventricular endocardium. ANSWERS References Tilley LP. Analysis of common canine cardiac arrhythmias. In Tilley LP (ed): Essentials of Canine and Feline Electrocardiography. Malvern, Lea & Febiger, 1992, 127-207. Cote E, Ettinger SJ. Electrocardiography and Cardiac Arrhythmias. In SJ Ettinger and Feldman EC (eds): Textbook of Veterinary Internal Medicine. St. Louis, Elsevier Saunders, 2005, 1040-1076. Moise NS. Diagnosis and Management of Canine Arrhythmias. In Fox PR, Sisson D, Moise NS (eds): Textbook of Canine and Feline Cardiology. Philadelphia, W.B. Saunders, 1999, 331-385. Hyoscyamine Sulfate. In Plumb DC: Plumb's Veterinary Drug Handbook. Ames, Blackwell Publishing, 2008. 469-470. Propantheline Bromide. In Plumb DC: Plumb's Veterinary Drug Handbook. Ames, Blackwell Publishing, 2008. 775-776. Kittleson MD: Syncope. In Kittleson (ed): Medicine. St Louis, Mosby, 1998 pp 495-501. Small Animal Cardiovascular Rush JE: Syncope and episodic weakness. In Fox PR, Sisson D, Moise NS (eds): Textbook of Canine and Feline Cardiology. Philidelphia, Saunders, 1999, pp 446-454. Moise NS: Pacemaker Therapy. In Fox PR, Sisson D, Moise NS (eds): Textbook of Canine and Feline Cardiology. Philidelphia, Saunders, 1999, pp 400-425. Petrie JP: Permanent Transvenous Cardiac Pacing: Clinical Techniques in Small Animal Practice 20: 164-172, 2005. Bulmer BJ et al: Physiologic VDD versus non-physiologic W l pacing in canine third degree atrioventricular block, J Vet Intern Med 20: 257, 2006. Oyama, MA, Sisson D: Permanent Cardiac Pacing in Dog. In Bonagura JD, Twedt DC (eds): Current Veterinary Therapy XIV. St. Louis, Saunders, 2009, pp 717-721. Kraus MS, Calvert CA: Syncope. In Bonagura JD, Twedt DC (eds): Current Veterinary Therapy XIV. St. Louis, Saunders, 2009, pp 709-712. Aminophylline, Theophylline. In Plumb DC: Plumb's Veterinary Drug Handbook. Ames, Blackwell Publishing, 2008. 37-40. Terbutaline Sulfate. In Plumb DC: Plumb's Veterinary Drug Handbook. Ames, Blackwell Publishing, 2008. 859-860. 12 •consutta