Guidelines for Using Lyophilized TaqMan Assays with Master Mix

Transcription

Guidelines for Using Lyophilized TaqMan Assays with Master Mix
Guidelines for Using Lyophilized TaqMan® Assays with Master Mix
For 7500/7300 System users
To run an experiment on your real-time PCR instrument using strip-tubes, one of the
following precision holders is required.
Table 1. 7300/7500 System Precision Holders
Instrument
Precision Holders
Part No.
7300 and
7500
7500 Fast
Precision Plate Holder for
0.2ml tubes and strips, for
7300 and 7500 Systems
Precision Plate Holder for
0.1ml tubes and strips, for
7500 Fast System
Reference Image
4367033
4403809
Steps:
1. Open the real-time PCR instrument tray and remove the “Precision Plate Holder
for plates” from the instrument tray if necessary.
2. Place the “Precision Plate Holder for tubes and strips” (refer to Table 1) in the
instrument tray as shown in Figure 1.*
Step 1: Open tray and remove holder for plates.
Step 2: Insert the holder for strip-tubes.
Figure 1. Preparing the real-time PCR system for Lyophilized TaqMan® Gene Expression
Assays with Master Mix products.
For instructions on how to prepare experiments in strip-tubes, please refer to the
“Lyophilized TaqMan® Gene Expression Assays with Master Mix” experiment protocol
provided in either the Fast or Standard format product inserts.
_______________________________________________________________
* If the adaptor does not fit your PCR instrument, please contact technical support, as some older model
instruments are not compatible. http://www.appliedbiosystems.com/support/contact/
For StepOne™ and StepOnePlus™ System users
To run an experiment on your real-time PCR instrument using strip-tubes, one of the
following adaptor trays is needed.
Table 2. StepOne™/StepOnePlus™ System Accessories
Instrument
Adaptor Trays
Part No.
StepOne™
StepOnePlus™
StepOne™ MicroAmp®
Fast 48-Well Tray
Reference Image
4375282
StepOnePlus™
MicroAmp® 96-Well Tray
for VeriFlex™ Blocks
4379983
Use only Fast consumables (reaction plates, strip-tubes, and tubes) with the StepOne™
and StepOnePlus™ Systems.
Steps:
1. Open the real-time PCR instrument tray.
2. If you have been running plates, place the adaptor tray (refer to Table 2) directly
on top of the tray of the real-time PCR instrument in the correct orientation (A1
well position on the top left) as noted in Figure 2.
Step 1: Open the tray.
Step 2: Place strip-tube adaptor directly on top.
Figure 2. Preparing the real-time PCR system for Lyophilized TaqMan® Gene Expression
Assays with Master Mix products.
For instructions on how to prepare experiments in strip-tubes, please refer to the
“Lyophilized TaqMan® Gene Expression Assays with Master Mix” experiment protocol
provided in either the Fast or Standard format product inserts.
For ViiA™ 7 System users
To run an experiment on the ViiA™ 7 Real-Time PCR instrument using strip-tubes, the
following trays and adaptors are needed. You can find product details for the Retainer
Trays on our website, while ViiA™ 7 adaptors are only bundled with the ViiA™ 7
instruments.
Table 3| ViiA™ 7 Trays and Adaptors
Instrument
ViiA™ 7
Standard
96-well Block
ViiA™ 7
Fast
96-well Block
Trays and Adaptor
Part No.
Retainer Tray for Standard 96-well (0.2mL)
tubes and strips
4381850
ViiA™ 7 96-Well Tube Adaptor
4459845
Retainer Tray for Fast 96-well (0.1mL)
tubes and strips
4379983
ViiA™ 7 Fast 96-Well Tube Adaptor
4462078
Steps:
1. Touch
to eject the tray arm.
2. When the tray arm opens, pull the latch, then lift and remove the plate adaptor
(Figure 3, Steps 2 and 3).
3. Replace with the appropriate Tube Adaptor (see Table 3).
4. Cap and place the strip-tubes in the corresponding Retainer Tray (Table 3) and
place the Retainer Tray with strip-tubes on the Tube Adaptor (See “Loading
0.2mL Strip-Tubes on the Retainer Tray” section on the following page).
5. Make sure “A1” of the Tube Adaptor aligns with “A1” of the Retainer Tray.
6. Touch
again to close the tray arm.
Steps 2 & 3: Replace the Plate Adaptor
Steps 4 & 5: Place the Retainer Tray on Adaptor
(Retainer tray in picture is for 0.2mL tube strips)
Figure 3. Preparing the ViiA™ 7 Real-Time PCR System for Lyophilized TaqMan® Gene
Expression Assays with Master Mix strip-tubes.
Loading 0.2mL Strip-Tubes on the Retainer Tray
The Retainer Tray for Standard 96-well (0.2mL) tubes and strips (Part No. 4381850) is
comprised of two parts; a top unit and a bottom unit.
Steps:
1.
2.
3.
4.
Snap open the top unit of the Retainer Tray (Figure 4, Step 1).
Put the un-capped 0.2mL strip-tubes into the bottom unit of the retainer tray.
Snap the top unit of the retainer tray back on top (over the un-capped strip-tubes).
Cap the strip-tubes, the optical strip-tube caps should be on top of the top unit of
the Retainer Tray.
Step 1: Snap open the top unit of the Retainer
Tray.
Steps 2 to 4: Assemble the tube strip with
the Retainer Tray.
Figure 4. Preparing and loading Lyophilized TaqMan® Gene Expression Assays with
Master Mix strip-tubes into the Retainer Tray.
For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use.