Performance Profile

Transcription

Performance Profile
Publication No. 1734 Rev 0802
Sample Pump
Performance Profile
•Grab sample pump with constant flow from 5 to 15 L/min
•Ideal for use with spore trap cassettes such as VersaTrap®, impactors, wall
cavity samplers, and filter cassettes including microvacuum cassettes
•VersaTrap cassette fits easily onto pump inlet
Flow range: 5 to 15 L/min
Flow compensation accuracy:
± 5% of set flow
Compensating back pressure range:
(inches of water)
15 L/min at 8
10 L/min at 15
5 L/min at 20
•Run times up to 4 hours on one Li-Ion battery charge
•Easy hand-held operation; only 32 oz (0.9 kg)
•Convenient programmable timer for unattended sampling
–Timed run presets, continuous run, and intermittent sampling
•No-tool calibration using any reliable calibrator
Power: Rechargeable 7.2-volt lithium-ion
(Li-Ion) battery, standard AC charger/adapter
(115 V), or deluxe AC charger/adapter (100
to 240 V)
Typical run time: VersaTrap Cassette:
4 hours at 15 L/min (pump and battery
performance may vary)
•End-of-cycle visual and audible alarms
Run time features: User-selectable
features, user-adjustable presets
•High-contrast digital display
•CE marked model available
The constant flow SKC QuickTake® 15 Sample Pump is ideal for easy short-run, high-flow
sampling from 5 to 15 L/min. Constant flow and a flow fault feature ensure sample integrity.
Convenient features include no-tool calibration, an inlet that easily accommodates spore trap
cassettes, end-of-cycle alarms, large keypad buttons, high-contrast digital display, and programmable timer. Use the versatile accessories for measurements at the breathing zone and in
ductwork and other hard-to-reach areas.
Description
QuickTake 15 Sample Pump*† and Charger
115 V
CE marked model with Cat. No. 223-245 charger
100-240 V
QuickTake 15 Sample Pump*†, Rotameter, and Charger
115 V
QuickTake 15 Sample Pump*† Kit includes pump, rotameter,
charger, wand, tripod stand, and carry case
115 V
CE marked model with Cat. No. 223-245 charger
100-240 V
Accessories
Defender Primary Standard Calibrator, 300 ml/min to 30 L/min, 100-240 V
Sampling Wand, telescopes to 34 in (0.9 m) for ductwork and hard-to-reach
locations, includes 6 ft (1.8 m) of tubing
Tripod Stand, telescopes to 5 ft (1.5 m) for mounting sample media or QuickTake 15
Rotameter for QuickTake 15 and 30
Preset timed runs: 1, 2, 5, 10, or
15 minutes
Manually set continuous run with
manual shutoff: 1 to 999 minutes
(repeats 1 to 999-minute runs
indefinitely until shut off).
Intermittent sampling: User-set sample
periods and sample delays
Cat. No.
228-9515*
228-9515B*
228-9515A*
Dimensions: ABS plastic housing is 10.3 x
4.6 x 2.4 inches (26.2 x 11.7 x 6.1 cm)
228-9515K*
228-9515KB*
Approvals: CE (models with 223-245
charger only)
Weight: 32 oz (0.9 kg)
717-510H
228-9521
225-9536
320-100
* QuickTake sample pumps contain Li-Ion batteries and may be subject to special shipping regulations dependent upon quantity.
† Use in non-explosive environments only. Not UL Listed for intrinsic safely.
SKC Inc. 724-941-9701
SKC West 714-992-2780
SKC Gulf Coast 281-859-8050
SKC South 434-352-7149
www.skcinc.com
Performance Profile
Flow range: 10 to 30 L/min
Flow compensation accuracy:
± 5% of set flow
Sample Pump
•Ideal for use with BioStage® Impactors, spore trap cassettes such as
VersaTrap®, asbestos cassettes, wall cavity samplers, or other samplers
that require flows to 30 L/min
•Long run times on one Li-Ion battery charge
Compensating back pressure range:
(inches of water)
10 L/min at 90
20 L/min at 50
30 L/min at 15
Power: Rechargeable 7.2-volt lithium-ion
(Li-Ion) battery or deluxe AC charger/adapter
(100 to 240 V)
Typical run time:
VersaTrap Cassette:
5 hours at 15 L/min
•Weighs only 4.8 lbs (2.2 kg)
–Handle for easy portability
37-mm, 0.8 µm MCE filter:
14+ hours at 10 L/min
•Convenient programmable timer for
unattended sampling
–Timed run presets, continuous
run, and intermittent sampling
Viable cascade impactors:
4 hours at 28.3 L/min
25-mm, 1.2 µm MCE filter:
For maximum air volumes during asbestos
clearance sampling, use QuickTake 30 with
a 25-mm, 1.2-µm MCE filter at 10 L/min
with AC power to achieve 9+ hours of run
time, a total air volume of 5400 L.
•No-tool calibration using any
reliable calibrator
•End-of-cycle visual and
audible alarms
(pump and battery performance may vary)
Run time features: User-selectable
features, user-adjustable presets
•High-contrast digital display
Preset timed runs: 1, 2, 5, 10, or
15 minutes
Take high flow sampling on the go with the portable SKC QuickTake® 30 sample pump.
QuickTake 30 features constant flows from 10 to 30 L/min to meet most applications. A
programmable timer and large keypad allow the user to select time presets, continuous run
with manual shut-off, or intermittent sampling.
Description
QuickTake 30 Sample Pump*† and Charger
100-240 V
QuickTake 30 Sample Pump*†, Rotameter, and Charger
100-240 V
QuickTake 30 BioStage Pump*† Kit includes pump, charger,
rotameter, Standard BioStage, mounting bracket with inlet adapter,
calibration adapter, and deluxe carry case
100-240 V
Accessories
Defender Primary Standard Calibrator, 300 ml/min to 30 L/min, 100-240 V
Mounting Bracket for BioStage includes inlet adapter
Sampling Wand, telescopes to 34 in (0.9 m) for ductwork and hard-to-reach
locations, includes 6 ft (1.8 m) of tubing
Rotameter, for QuickTake 15 and 30
Tripod Stand, telescopes to 5 ft (1.5 m) for mounting sample media or QuickTake 15
Cat. No.
228-9530*
228-9530A*
Manually set continuous run with
manual shutoff: 1 to 999 minutes
(repeats 1 to 999-minute runs
indefinitely until shut off).
Intermittent sampling: User-set sample
periods and sample delays
Dimensions: ABS plastic housing is 9.3 x
228-9530K*
717-510H
228-9531
228-9521
320-100
225-9536
* QuickTake sample pumps contain Li-Ion batteries and may be subject to special shipping regulations dependent upon quantity.
† Use in non-explosive environments only. Not UL Listed for intrinsic safety.
SKC Inc. 724-941-9701
SKC West 714-992-2780
SKC Gulf Coast 281-859-8050
SKC South 434-352-7149
www.skcinc.com