Performance Profile
Transcription
Performance Profile
Publication No. 1734 Rev 0802 Sample Pump Performance Profile •Grab sample pump with constant flow from 5 to 15 L/min •Ideal for use with spore trap cassettes such as VersaTrap®, impactors, wall cavity samplers, and filter cassettes including microvacuum cassettes •VersaTrap cassette fits easily onto pump inlet Flow range: 5 to 15 L/min Flow compensation accuracy: ± 5% of set flow Compensating back pressure range: (inches of water) 15 L/min at 8 10 L/min at 15 5 L/min at 20 •Run times up to 4 hours on one Li-Ion battery charge •Easy hand-held operation; only 32 oz (0.9 kg) •Convenient programmable timer for unattended sampling –Timed run presets, continuous run, and intermittent sampling •No-tool calibration using any reliable calibrator Power: Rechargeable 7.2-volt lithium-ion (Li-Ion) battery, standard AC charger/adapter (115 V), or deluxe AC charger/adapter (100 to 240 V) Typical run time: VersaTrap Cassette: 4 hours at 15 L/min (pump and battery performance may vary) •End-of-cycle visual and audible alarms Run time features: User-selectable features, user-adjustable presets •High-contrast digital display •CE marked model available The constant flow SKC QuickTake® 15 Sample Pump is ideal for easy short-run, high-flow sampling from 5 to 15 L/min. Constant flow and a flow fault feature ensure sample integrity. Convenient features include no-tool calibration, an inlet that easily accommodates spore trap cassettes, end-of-cycle alarms, large keypad buttons, high-contrast digital display, and programmable timer. Use the versatile accessories for measurements at the breathing zone and in ductwork and other hard-to-reach areas. Description QuickTake 15 Sample Pump*† and Charger 115 V CE marked model with Cat. No. 223-245 charger 100-240 V QuickTake 15 Sample Pump*†, Rotameter, and Charger 115 V QuickTake 15 Sample Pump*† Kit includes pump, rotameter, charger, wand, tripod stand, and carry case 115 V CE marked model with Cat. No. 223-245 charger 100-240 V Accessories Defender Primary Standard Calibrator, 300 ml/min to 30 L/min, 100-240 V Sampling Wand, telescopes to 34 in (0.9 m) for ductwork and hard-to-reach locations, includes 6 ft (1.8 m) of tubing Tripod Stand, telescopes to 5 ft (1.5 m) for mounting sample media or QuickTake 15 Rotameter for QuickTake 15 and 30 Preset timed runs: 1, 2, 5, 10, or 15 minutes Manually set continuous run with manual shutoff: 1 to 999 minutes (repeats 1 to 999-minute runs indefinitely until shut off). Intermittent sampling: User-set sample periods and sample delays Cat. No. 228-9515* 228-9515B* 228-9515A* Dimensions: ABS plastic housing is 10.3 x 4.6 x 2.4 inches (26.2 x 11.7 x 6.1 cm) 228-9515K* 228-9515KB* Approvals: CE (models with 223-245 charger only) Weight: 32 oz (0.9 kg) 717-510H 228-9521 225-9536 320-100 * QuickTake sample pumps contain Li-Ion batteries and may be subject to special shipping regulations dependent upon quantity. † Use in non-explosive environments only. Not UL Listed for intrinsic safely. SKC Inc. 724-941-9701 SKC West 714-992-2780 SKC Gulf Coast 281-859-8050 SKC South 434-352-7149 www.skcinc.com Performance Profile Flow range: 10 to 30 L/min Flow compensation accuracy: ± 5% of set flow Sample Pump •Ideal for use with BioStage® Impactors, spore trap cassettes such as VersaTrap®, asbestos cassettes, wall cavity samplers, or other samplers that require flows to 30 L/min •Long run times on one Li-Ion battery charge Compensating back pressure range: (inches of water) 10 L/min at 90 20 L/min at 50 30 L/min at 15 Power: Rechargeable 7.2-volt lithium-ion (Li-Ion) battery or deluxe AC charger/adapter (100 to 240 V) Typical run time: VersaTrap Cassette: 5 hours at 15 L/min •Weighs only 4.8 lbs (2.2 kg) –Handle for easy portability 37-mm, 0.8 µm MCE filter: 14+ hours at 10 L/min •Convenient programmable timer for unattended sampling –Timed run presets, continuous run, and intermittent sampling Viable cascade impactors: 4 hours at 28.3 L/min 25-mm, 1.2 µm MCE filter: For maximum air volumes during asbestos clearance sampling, use QuickTake 30 with a 25-mm, 1.2-µm MCE filter at 10 L/min with AC power to achieve 9+ hours of run time, a total air volume of 5400 L. •No-tool calibration using any reliable calibrator •End-of-cycle visual and audible alarms (pump and battery performance may vary) Run time features: User-selectable features, user-adjustable presets •High-contrast digital display Preset timed runs: 1, 2, 5, 10, or 15 minutes Take high flow sampling on the go with the portable SKC QuickTake® 30 sample pump. QuickTake 30 features constant flows from 10 to 30 L/min to meet most applications. A programmable timer and large keypad allow the user to select time presets, continuous run with manual shut-off, or intermittent sampling. Description QuickTake 30 Sample Pump*† and Charger 100-240 V QuickTake 30 Sample Pump*†, Rotameter, and Charger 100-240 V QuickTake 30 BioStage Pump*† Kit includes pump, charger, rotameter, Standard BioStage, mounting bracket with inlet adapter, calibration adapter, and deluxe carry case 100-240 V Accessories Defender Primary Standard Calibrator, 300 ml/min to 30 L/min, 100-240 V Mounting Bracket for BioStage includes inlet adapter Sampling Wand, telescopes to 34 in (0.9 m) for ductwork and hard-to-reach locations, includes 6 ft (1.8 m) of tubing Rotameter, for QuickTake 15 and 30 Tripod Stand, telescopes to 5 ft (1.5 m) for mounting sample media or QuickTake 15 Cat. No. 228-9530* 228-9530A* Manually set continuous run with manual shutoff: 1 to 999 minutes (repeats 1 to 999-minute runs indefinitely until shut off). Intermittent sampling: User-set sample periods and sample delays Dimensions: ABS plastic housing is 9.3 x 228-9530K* 717-510H 228-9531 228-9521 320-100 225-9536 * QuickTake sample pumps contain Li-Ion batteries and may be subject to special shipping regulations dependent upon quantity. † Use in non-explosive environments only. Not UL Listed for intrinsic safety. SKC Inc. 724-941-9701 SKC West 714-992-2780 SKC Gulf Coast 281-859-8050 SKC South 434-352-7149 www.skcinc.com