MR-TOF at ISOLDE Frank Wienholtz - University of Greifswald -

Transcription

MR-TOF at ISOLDE Frank Wienholtz - University of Greifswald -
GUI – 2014-11-03
MR-TOF at ISOLDE
Frank Wienholtz
- University of Greifswald for the ISOLTRAP Collaboration
http://isoltrap.web.cern.ch
[email protected]
ISOLTRAP overview
ISOLTRAP uncertainty:
2010
dm
» 10-8 ; >500 short-lived nuclides investigated
m
1987
1999
1994
2
ISOLTRAP overview: MR-ToF-MS1
 mean kinetic energy Ekin=2.1keV
 ToF separation due to different m/q
RFQ:
 Δt≈100ns
 ΔEkin/Ekin≈3%
several applications possible:
•
•
•
high-resolution mass separation with
Bradbury-Nielsen gate
for subsequent experiments
observing and gating on separated ion-ofinterest to perform further studies
high-precision mass measurements with
reference masses
Bradbury-Nielsen gate
(BNG)2,3
MR-ToF-MS
mass resolving power (FWHM)
m/∆m=100 000 at 12ms
m/∆m=200 000 at 30ms
transmission
≈50% at 30ms
ion capacity
≈1000 per cycle
≈100 000 per second
1: Wollnik & Przewloka, Int. J. Mass Spectrom. Ion Proc. 96, 267 (1990); 2: Bradbury & Nielsen, Phys. Rev. 49, 388 (1936); 3: Plass et al., NIM B 266, 4560 (2008)
Wolf et al., IJMS 313, 8 (2012);
Wolf et al., IJMS 349-350, 123 (2013);
3
MR-ToF-MS at ISOLTRAP: in-trap lift
 capture and ejection with one electrode
 simple technique, stable mirror potentials
 decouple MR-ToF-MS and adjacent beamline
 independent optimization
 adjust ions’ kinetic energy
 ToF focusing, max. mass resolving power
in-trap lift
3
2
1
Injection
Storage
Ejection
 only one parameter to adjust
4
Wolf et al., IJMS 313, 8 (2012)
MR-ToF ion-beam analysis
Ion-beam composition analysis





direct feedback for target/line optimization
sampling of release curve possible
single ion sensitivity to detect lowest yields
no upper limit on half-life as with decay station
not hindered by decay branching ratio
target release curve
99Rb+
target heating increased
by 70K
T1/2=54ms
5
Wolf et al., IJMS 349-350, 123 (2013); Kreim et al., NIM B, accepted for publication (2013)
MR-ToF ion-beam analysis
15
number of ions
−−> roughly 250 ions/s form ISOLDE
10
5
0
0
10
20
30
40
50
60
Increasing current of OVEN 1
70
80
90
6
100
MR-ToF  RILIS yield optimization
MR-ToF analyzer to investigate resonant laser
ionization of nuclides far from stability
 fast, sensitive tool to improve ionization eff.
 high dynamic range: 1-10e5 counts/s
 counts free from background contamination
 not limited by decay branching ratio
 help to provide isomerically pure beams
185Tl+
185Au+
185Au+
185Tl+
scan: RILIS first excitation step
7
Wolf et al., IJMS 349-350, 123 (2013); Kreim et al., NIM B, accepted for publication (2013)
MR-ToF <--> laser scans of hyperfine structure
(Total counts - 149Dy) /
149Dy
≈ 70
MR-ToF <--> laser scans of hyperfine structure
250
149Dy
200
counts
150
100
50
0
15971
15972
15973
wavenumber
15974
MR-ToF half life measurements
A=97
?
97Sr
Cooling time in the buncher
In Steps of 60 ms
610 ms
10 ms
Time of flight after 500 revs
97Rb
MR-ToF half life measurements
1000
97
Rb
T1/2
Counts
800
T1/2
=173(8)ms
NUBASE
=169.1(0.6)ms
600
400
200
0
200
300
400
500
Cooling time buncher (ms)
600
Possible applications @ ISOLDE
 Versatile tool for beam analysis especially for ion yields which are not detectable by FCs or
accessible by decay spectroscopy
 Continues observation of the beam composition possible
 Beam optimisation
 Varying different target parameters
 Beam line optimisation (transport)
 Fast response to laser on/off or protons on/off
 Laser frequency optimisation
 Delivery of highly pure beams for experiments with such needs
 Beam purification for REX-Trap and EBIS
 Fast stacking in the Penning trap possible
12
Thanks to…
S. George, M. Rosenbusch, R.N. Wolf, L. Schweikhard
P. Ascher, D. Atanasov, Ch. Böhm, Ch. Borgmann, R. B. Cakirli,
S. Eliseev, T. Eronen, D. Kissler, S. Naimi, K. Blaum
D. Beck, F. Herfurth, A. Herlert, E. Minaya-Ramirez, D. Neidherr, Y. Litvinov
G. Audi, V. Manea, M. Wang, D. Lunney
M. Kowalska, S. Kreim, J. Kurcewicz
J. Stanja, A. Welker, K. Zuber
N. Althubiti, T. Cocolios
Grants No.:
05P12HGCI1
05P12HGFNE
M. Breitenfeldt
http://isoltrap.web.cern.ch
[email protected]
13
MR-ToF-MS: voltage stability
mirror electrode 5: turn-around point
 limiting long-term stability
 mass resolving power
M4
M5
 Temperature stabilization of supply voltages to <100mK
 ToF temperature coefficient:
14
Wolf et al., NIM A 686, 82 (2012)