CK40-RFL
Transcription
CK40-RFL
INSTRUCTIONS CK40-RFL REFLECTED LIGHT ATTACHMENT This instruction manual is for the Olympus CK40-RFL Reflected Light Attachment. To ensure the safety, obtain optimum performance and familiarize yourself fully with the use of this product, we recommend that you study this manual thoroughly before operating the equipment. Retain this instruction manual in an easily accessible place near the work desk for future reference. AX7170 IMPORTANT This attachment employs an LB optical design, and should be used only with equipment using the LB optical design, such as the CK40 microscope frame and its eyepiece, objective and condenser. Less than optimal performance may result if inappropriate accessories are used. SAFETY PRECAUTIONS 1. Make sure that the main switch on the power supply unit is set to “\” (OFF) before connecting the power cord to the AC mains. 2. The power supply unit contains high voltage components. Never attempt to disassemble the unit, or you may run the risk of electric shock. 3. Always be sure to ground (earth) the equipment. Otherwise Olympus will no longer warrant the electrical safety of the equipment. 4. The attachment should be installed on a flat surface so that the ventilation opening on the bottom panel is not blocked. Do not install the attachment on a surface which is soft or which may collapse under the weight of the microscope, for the subsequent blockage of the bottom panel ventilation opening could create a fire hazard. 5. The high pressure mercury burner (mercury arc lamp) used should be a HBO50W/AC (OSRAM) or CS50W4 (PHILIPS) lamp. 6. Before opening the lamp housing for replacement of the burner or any other internal part, set the main switch to “\” (OFF), unplug the lamp housing´s connecting cord plug from the output connector on the power supply unit, and wait 10 minutes or more until the lamp housing cools down. 7. Do not open the lamp housing while it is turned on or for at least 10 minutes after it has been turned off. Lamp housing parts are extremely hot and cause burns if touched. (See page 9.) 8. Do not remove the lamp housing while the burner is lit. Also, do not turn the burner on while the lamp housing is removed. 9. Do not install the lamp housing upside down or at an angle. The lamp housing surface becomes very hot during operation, so ensure that there is ample free space around the lamp housing, especially above and below. 10. If the microscope power cord comes in contact with the lamp housing or its surroundings, the cord may melt and subsequently cause electric shock. Be sure to position the power cord at a safe distance from the lamp housing. 11. A used mercury burner should be disposed of in compliance with the ordinances or regulations of your national or local government. 12. To replace the lithium battery (for the hour counter) built into the power supply unit, consult your local Olympus representative. As the power supply unit incorporates a lithium battery, it should be disposed of in the same way as a used mercury burner, i.e., in compliance with the ordinances or regulations of your national or local government. 13. Before placement of the dust cover for protection after operation, wait until the lamp housing cools down sufficiently and unplug the power cord. i Safety Symbols The following symbols are found on the equipment. Study the meaning of the symbols, and always use the equipment in the safest possible manner. Symbol Explanation Indicates that the surface becomes hot, and should not be touched with bare hands. Before use, carefully read the instruction manual. Improper use could result in personal injury to the user and/or damage to the equipment. Indicates a potential fire hazard; when replacing fuses, be sure replacement fuses are of the specified rating. I \ Indicates that the main switch is ON. Indicates that the main switch is OFF. Warning Indications Warning indications are placed at parts where special precaution is required when handling and using the equipment. Always heed the warnings. Warning indication position 1 Lamp housing [Warning against high temperature] Getting Ready 1. This attachment is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact. 2. Do not use the attachment where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations. (For operating environment conditions, see Section 7, “SPECIFICATIONS” on page 20.) 3. Make certain that the burner is installed correctly and that all cords are correctly connected. 4. Do not apply excessive force to the limiting mechanism provided with all functions. ii 2 Maintenance and Storage 1. Be careful to avoid leaving dirt or fingerprints on the lenses, filters or high pressure mercury burner, for contamination of these parts may cast the shadow of a foreign object in the field of view or compromise burner performance. If contaminated, clean by wiping gently with a piece of gauze. To remove fingerprints or oil smudges, wipe with gauze slightly moistened with a mixture of ether (70%) and alcohol (30%), EE System Cleaner (Olympus EE-6310), alcohol or xylene. Since ether, alcohol and EE System Cleaner are highly flammable, be careful to keep these chemicals away from open fire and potential sources of electrical sparks, such as sparks from main switches switching ON or OFF. 2. The mercury burner has an average service life of 100 hours. When the hour counter on the power supply unit indicates 100 hours, replace the burner with a new one (see page 14). Using a burner after its service life expires could lead to the burner exploding, though this occurs very rarely. 3. If a dichroic mirror cube is not going to be used for a while, place it in its container and store it in a safe place. 4. The dichroic mirror cube surface is very delicate. When it becomes necessary to clean it, please consult us or your local Olympus representative. 5. Never disassemble any part of the attachment, for this may degrade performance. 3 Caution If the equipment is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the equipment may also be damaged. Always operate the equipment as outlined in this instruction manual. The following symbols are used to set off text in this instruction manual. : Indicates that failure to follow the instructions in the warning could result in bodily harm to the user and/or damage to equipment (including objects in the vicinity of the equipment). # : Indicates that failure to follow the instructions could result in damage to equipment. } : Indicates commentary (for ease of operation and maintenance). iii CK40-RFL CONTENTS 1 PRINCIPLE 1 2 NOMENCLATURE 2 3 ASSEMBLY 3 3-1 Assembly Diagram ............................................................................................................................. 3 3-2 Detailed Assembly Procedure ........................................................................................... 4 4 CONTROLS OF EACH MODULE 12 5 PREPARATION BEFORE OBSERVATION 14 6 OBSERVATION 19 7 SPECIFICATIONS 20 8 TROUBLESHOOTING GUIDE 21 ■ PROPER SELECTION OF POWER SUPPLY CORD ..................................... 23 PRINCIPLE The design of the reflected light fluorescent microscope features a dichroic mirror* which directs the excitation light through the objective to illuminate the specimen and provide efficient fluorescent light observation. Specimen Emitted light Objective Excitation light Dichroic mirror Collector Light source Exciter filter Barrier filter Eyepiece * The dichroic mirror enables efficient fluorescent light microscopy by making use of its characteristics of short wavelength reflection and long wavelength transmission. Reflectance Absorbance 50% 1 The spectral characteristics of the dichroic mirror when it is positioned at an inclination of 45° to the optical axis of incident light is shown in the figure on the left. Because crossover exists between transmittance and reflectance, it is necessary Transmittance to employ an appropriate combination of exciter and barrier filters in conjunction with the dichroic mirror to obtain a wellcontrasted image. Fluorescence When the dichroic mirror is inclined at 45° to the optical axis of the incident excitation light, the excitation light is reflected towards the objective and other unwanted wavelengths pass through the mirror, practically in their entirety. When the specimen is irradiated by the excitation wavelength, it emits a visible, longer wavelength in accordance with Stokes´ λM Wavelength Law. The dichroic mirror passes almost all of this light to the eyepiece. The barrier filter mounted between the dichroic mirCrossover ror and the eyepiece blocks out unwanted wavelengths to provide a black background. CK40-RFL NOMENCLATURE Excitation light protective shield Fluorescent vertical illuminator CK40-RFA Lamp housing U-LH50HG Microscope CK40 Fluorescent mirror cubes CX-DMB-2 CX-DMG-2 Power supply unit U-RFLT50 2 ASSEMBLY 3-1 Assembly Diagram The diagram below shows how to assemble the various modules for reflected light fluorescence observation. The numbers in the diagram indicate the assembly step order. # When assembling the microscope system, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching the glass surfaces. Lamp housing U-LH50HG Mercury burner Fluorescent vertical illuminator CK40-RFA Cover Power cord Microscope CK40 Excitation light protective shield Power supply unit U-RFLT50 RCUBE indication Light shield cap* Right light excluding plate Cube Removal lever Right cover Left cover Required Tool Removal lever Cube LCUBE indication Allen wrench Left light excluding plate * The light shield cap is provided only with the left cover. 3 CK40-RFL 3-2 Detailed Assembly Procedure 1 | ² ³ @ Fig. 1 … Mounting the Fluorescent Vertical Illuminator (CK40-RFA) (Figs. 1 - 3) 1. Set the main switch of the microscope (CK40) to “\” (OFF). For safety, unplug the power cord @ until the fluorescent vertical illuminator has been mounted completely. (Fig. 1) 2. Unplug the lamp cord socket ² of the illumination support because it gets in the way of removing the cover ³. (Fig. 1) 3. Move the revolving nosepiece to the upper limit by turning the coarse and fine focus adjustment knobs. 4. Pull out the cover ³ by hooking a finger into the removal hole |. (Fig. 1) 5. Loosen one of the filter slider knobs ƒ and pull out the filter slider …. (Fig. 2) ƒ Fig. 2 6. Insert the fluorescent vertical illuminator † gently into the cover removed above, and tighten the provided clamping screws (x 3) securely using the Allen wrench. (Fig. 3) 7. Mount the filter slider to the original position and attach the filter slider knob. † Fig. 3 4 2 Mounting the Fluorescent Mirror Cubes (Figs. 4 - 8) } 2-wavelength observation using excitations B and G or transmitted-light brightfield observation using one excitation can be selected according to mounted cubes. Mount cubes to meet the desired observation type. # While the CX-DMB-2 and CX-DMG-2 fluorescent mirror cubes can be mounted either on the left or right, the light excluding plates and cube covers for the left and right are provided with different designs. Check the assembly diagram to ensure design and orientation before using these parts. # The cube covers must always be mounted, for they are important for glare protection during cube switching and cube positioning. Right cube 1. Place a cube @ as shown in Fig. 4. 2. Orient the right light excluding plate as shown in Fig. 4, and using the Allen wrench, fit the stopper pin ³ into the mounting threaded hole | on the cube. ³ ² @ | Fig. 4 3. Place the cube with the right light excluding plate in the normal position shown in Fig. 5, and insert it gently into the dovetail ƒ on the illuminator so that it is not inclined. # Be careful that the cube is oriented right side up. The cube cannot be inserted if it is held upside down. ƒ Fig. 5 5 CK40-RFL 4. Mount the right cover … by aligning the clamping screws † with the threaded holes ‡ and tightening using the Allen wrench. (Fig. 6) … ‡ † Fig. 6 Š ‰ Œ 5. Insert the cube installation/removal lever Š, which is marked for use with the cube mounted on the right, into the lever hole ‰, and tighten the lever in the mounting threaded hole on the cube. (Fig. 7) 6. Orient the indication plate ‹ of the cube installation/removal lever in the forward direction by loosening the indication plate clamping knob Œ, then tightening it. ‹ Fig. 7 <<Removing the cube installation/removal lever>> If the cube installation/removal lever cannot be removed by turning it counterclockwise, remove the indication plate clamping knob and turn the internal screw counterclockwise using a flat-blade screwdriver. Left cube 1. In the same way as for the right cube, place the left cube as shown in Fig. 8 and mount the left light excluding plate onto it. 2. Remove the light shield cap on the left cover by pushing it outward. # When only the left cube is mounted, re-attach the light shield cap onto the right cover. 3. In the same way as for the right cube, mount the left cube on the side opposite to the right cube and attach the left cover and cube installation/removal lever. Fig. 8 6 3 ² Mounting the Excitation Light Protective Shield (Fig. 9) Insert the excitation light protective shield @ from the bottom of the condenser ² until it stops, then tighten the clamping screw ³. The wider end of the protective shield should be on the front side of the microscope. @ ³ Fig. 9 4 Mounting the U-LH50HG Lamp Housing ² ³ ƒ @ Fig. 10 7 | (Fig. 10) 1. Insert the mercury burner housing collector @ into the vertical illuminator until it reaches the click-stop. 2. Turn the lamp housing so that its top panel ² becomes horizontal, then tighten the clamping screw ³ (x 1) on the illuminator using the Allen wrench. As the lamp housing surface becomes very hot during operation, ensure that there is ample free space around the lamp housing, especially above and below. 3. Plug the CK40 power cord | and the lamp cord socket ƒ (removed as above) into the original positions. If the CK40 power cord comes in contact with the lamp housing or its surroundings, the cord could melt and cause electric shock. Be sure to position the power cord at a safe distance from the lamp housing. CK40-RFL 5 Mounting the High Pressure Mercury Burner (Figs. 11 - 14) <<Applicable burners>> · HBO50W/AC (OSRAM) · CS50W4 (PHILIPS) @ Fig. 11 ² | ³ Fig. 12 # Burner type information must be set in the power supply unit, so check the burner box or burner base to confirm whether it is of type L1 or L2. 1. Using the Allen wrench, loosen the burner socket clamping screw @ and flap open the burner socket section by about 90° in the direction of the arrow. (Fig. 11) 2. Loosen the burner clamping screws ² and ³ and remove the transport post attached to the burner socket. }To replace the burner, remove the used burner. (Fig. 12) 3. Loosen the clamping screw ² and attach the extremity without the “UP” marking of the mercury burner | so that it comes on the lower side. Then, loosen the clamping screw ³ and attach the extremity marked “UP”. (Fig. 12) # Do not mount the mercury burner upside down. Otherwise the light may become dim, the original performance will not be achievable, and burner service life may be shortened. # Contamination of the burner surface with dirt may cause its shadow to be cast in the field of view. Be sure to remove any dirt from the burner surface. 4. Close the burner socket with burner into the original position and tighten the socket clamping screw @. (Fig. 11) }A click sound is generated when the socket clamping screw @ is tightened. This sound indicates the normal action of an interlock switch safety mechanism. (Fig. 11) 8 ƒ Fig. 13 5. Set the burner type switch ƒ on the rear of the power supply unit according to the type of the mounted burner (L1 or L2). (Fig. 13) # If the clamping screw @ is accidentally loosened while the burner is operating, the interlock switch turns off the burner. (Fig. 11) To restart the burner, first set the main switch on the power supply unit to “\” (OFF) and wait about 10 minutes*. Then, after firmly tightening the clamping screw @, set the main switch to “ l ” (ON) again. * The burner cannot be re-ignited for about 10 minutes; that is, until the mercury vapor inside it has cooled down and condensed to liquid. Caution for burner replacement during observation The burner, lamp housing and its surroundings are very hot during operation. Before replacing the burner, set the main switch to “\” (OFF), unplug the power cord and wait until the burner, lamp housing and its surroundings cool down completely. … Fig. 14 9 6. Press the reset button … on the front of the power supply unit to reset the hour counter display to “0.00”. (Fig. 14) } The hour counter display registers hour units. Replace the lamp when the displayed time reaches “100.00” hours. Note that the service life of a mercury burner may expire before the above-mentioned average depending on how frequently the “ l ” (ON) and “\” (OFF) switch is used. When start-up of the burner degrades or flickering becomes noticeable, replace the burner regardless of the hour counter indication. # Be sure to reset to “0.00”. Otherwise the burner may not be ignitable. Clear more than 10 cm of space around the power supply unit to assure proper heat radiation. The power supply unit should be installed so that the main switch can easily be accessed and set to “\” (OFF) in case of emergency. CK40-RFL 6 Setting Input Power Voltage ² @ (Fig. 15) 1. Using a flat-blade screwdriver, set the voltage selector switch @ on the rear of the power supply unit to your local power voltage. 2. Set the frequency switch ² to your local power frequency. # Improper setting of the above items may degrade burner performance, or in the worst (although very rare) case, cause the burner to explode. Fig. 15 7 Connecting the Cords | ² ³ @ Fig. 16 (Figs. 16 - 18) Do not forcefully bend or twist the cords. They are sensitive to such application of force, which can cause internal wire disconnection and short-circuiting. Be sure to set the main switch on the power supply unit to “\” (OFF) before connecting the cords. 1. Plug the connector @ from the burner socket securely into the connector ² on the power supply unit. (Fig. 16) Always use the U-LH50HG Lamp Housing. Always use the power cord provided by Olympus. If no power cord is provided, please select the proper power cord by referring to the section “PROPER SELECTION OF THE POWER SUPPLY CORD” at the end of this instruction manual. 2. Plug the connector ³ of the power cord securely into the connector | on the power supply unit. (Fig. 16) 10 ƒ Fig. 17 … 3. Insert the power cord plug ƒ into the wall outlet …. (Fig. 17) Connect the power cord correctly and ensure that the ground terminal of the power supply and that of the wall outlet are properly connected. If the equipment is not grounded, Olympus can no longer warrant the electrical safety and performance of the equipment. If the power cord or burner socket cord comes in contact with the lamp housing or its surroundings, the cord may melt and cause electric shock. Be sure to position the power cord and burner socket cord at a safe distance from the lamp housing. Fuse replacement † Fig. 18 # Set the main switch to “\” (OFF) and unplug the power cord before replacing fuses. 1. Remove each of the fuse holders † by turning it counterclockwise using a flat-blade screwdriver and pulling out. (Fig. 18) 2. Replace both fuses with new ones. # Make sure that the replacement fuses are of the properly specified rating. Otherwise a fire hazard may result. U-RFLT50-100: T4A(H)250V (LITTELFUSE 215004), 2 pieces Applicable fuses U-RFLT50-200: T2.5A(H)250V (LITTELFUSE 21502.5), 2 pieces 11 CK40-RFL CONTROLS OF EACH MODULE Fluorescent Vertical Illuminator CK40-RFA Field iris diaphragm lever Open: Clockwise Close: Counterclockwise Field iris centering knob Filter slider Shutter/Open aperture/Filter pocket · Accommodates the ND filter. Cube installation/removal lever · B excitation mirror cube lever · G excitation mirror cube lever Index groove Fluorescent mirror cubes · B excitation: CX-DMB-2 · G excitation: CX-DMG-2 ND Filter (32ND25) 12 Lamp Housing U-LH50HG Collector focusing knob Burner up/down centering knob Rear Mirror left/right centering screw Mirror focusing screw Mirror up/down centering screw Burner left/right centering knob Power Supply U-RFLT50 # The depth dimensions of the 100 V and 200 V models differ from each other. Frequency switch Reset button Hour counter Voltage switch Burner type switch Fuse holders Power cord connector Main switch I : ON \: OFF 13 Starter reset switch Lamp housing connector For exclusive use with the U-LH50HG. CK40-RFL PREPARATION BEFORE OBSERVATION 1 Turning ON the Mercury Burner @ Fig. 19 (Fig. 19) # Confirm that the power supply unit has been set to the proper power supply voltage, power supply frequency and burner type. Improper setting of these items may degrade burner service life considerably. · Set the main switch on the power supply unit to “ I ” (ON). After the burner is ignited, between 5 and 10 minutes are required before the arc stabilizes. # Some mercury burners may not ignite the first time the power is turned ON, and the safety mechanism in the starter of the power supply unit is activated to reset the starter in such a case. If this occurs, set the main switch to “\” (OFF) once, then insert a thin object such as a ballpoint pen tip into the starter reset hole @ on the right panel of the power supply unit to press the internal switch, and set the main switch to “ I ” (ON) again. # To avoid shortening the power supply unit´s operating service life, avoid turning power on while the burner is not mounted, or turning power on and off within short time intervals. # To avoid shortening burner life, do not turn the burner off within 15 minutes after ignition. Use the shutter Instead. # After the burner is turned off, it cannot be re-ignited before the mercury vapor cools and condenses to liquid. Wait about 10 minutes before restarting the burner. If the lamp housing is opened while the burner is ignited, the safety interlock will activate and switch off the power automatically. In this case, set the main switch to “\” (OFF) and wait for more than 10 minutes before restarting the burner. The lamp housing can be opened only after it has cooled down fully. When resetting the hour counter, be sure to press the reset button until “0.00” is displayed. When the hour counter indicates “100.00”, set the main switch to “\” (OFF) for safety, wait for more than 10 minutes, then replace the lamp burner after making sure that the housing has cooled down. A mercury burner seals high-pressure gas inside. If the burner is used beyond its service life, stress may accumulate inside the burner, and in the worst (but very rare) case, the burner could explode. 14 2 Centering the Mercury Burner ² @ ³ Fig. 20 | ƒ Fig. 21 15 (Figs. 20 - 24) }Before attempting to center the burner, wait for the arc image to stabilize. To protect against glare during arc image centering, it should be viewed across the excitation light protective shield (see page 7). 1. Remove the cap (or objective) from the revolving nosepiece and engage the open aperture into the light path. 2. Push the “B” cube installation/removal knob @ to engage it into the light path. (Fig. 20) 3. Turn the field iris diaphragm lever ² clockwise to open the iris diaphragm. (Fig. 20) 4. Place a piece of thin white paper, such as copy paper, on the top of the stage, set the filter slider ³ to the central position, open aperture position, and project the arc image. (Fig. 20) }The arc image can be confirmed more easily by using the U-CST centering target screen. # If the arc image is too bright, insert the ND filter | into the filter pocket ƒ and engage the filter pocket into the light path. (Fig. 21) 32ND25: Reduces transmittance to 25%. # If the arc image focusing position cannot be identified easily, close the iris diaphragm. CK40-RFL Š ‹ ‰ … ‡ † Fig. 22 Terminal Arc image Terminal Fig. 23 ¬ Mirror reflected image ® 5. Project the arc image on the white paper, then center and sharpen it by manipulating the collector focusing knob …, burner left/right centering knob † and burner up/down centering knob ‡ on the lamp housing. (Figs. 22 & 23) # There may be cases in which the arc image projected on the white paper is a mirror reflected image. If the arc image moves when the mirror left/right centering screw Š or mirror up/down centering screw ‰ is turned, it is a mirror reflected image; bring it out of the field and center the arc image as described above. (Fig. 22) 6. Manipulate the burner left/right centering knob † slightly to move the arc image to the right (or left). (Figs. 22 & 24¬) 7. Using the Allen wrench provided with the vertical illuminator, turn the mirror focusing screw ‹, mirror left/right centering screw Š and mirror up/down centering screw ‰ to form the mirror reflected arc image on the white paper. (Figs. 22 & 24®) 8. Adjust the mirror centering screws Š and ‰ to move the direct and mirror reflected images in the symmetrical position between each other, and turn the mirror focusing screw ‹ to adjust the size of the mirror reflected image until it is identical to that of the direct image. (Figs. 22 & 24¯) 9. Turn the burner left/right centering screw † to overlap the direct image with the mirror reflected image. (Figs. 22 & 24¶) 10. Engage the 10X objective into the light path while continuing to use the “B” excitation mirror cube. 11. Place a specimen on the stage and bring it into approximate focus. 12. Look into the eyepiece and manipulate the collector focusing knob … to make the field as bright and regular as possible, then turn the knob clockwise to tighten. (Fig. 22) }Maintain this condition until the next time the burner is replaced. ¯ ¶ Fig. 24 16 3 Centering the Field Iris Diaphragm @ ² Fig. 25 (Figs. 25 & 26) 1. Engage the 10X objective into the light path, push the “B” cube installation/removal knob into the light path, and set the filter slider to the central open aperture position or the position of the filter pocket accommodating the ND filter. 2. Place a specimen on the stage and bring into approximate focus. 3. Turn the field iris diaphragm lever @ counterclockwise, little by little, and confirm the iris diaphragm position. (Figs. 25 & 26) 4. Manipulate the 2 field iris centering knobs ² to move the image of the diaphragm to the center. (Figs. 25 & 26) 5. After moving the image to the center, open the field iris diaphragm. As this makes slight deviation noticeable, adjust the centering precisely. 6. After completion of centering, enlarge the iris diaphragm diameter until it just circumscribes the field of view. Adjusting the field iris diaphragm Fig. 26 17 To obtain good image contrast, adjust the diameter of the illuminating beam, according to the objective in use. The field iris diaphragm can be adjusted only within the required area for observation, to prevent fading of specimen areas outside the observed field. According to the objective in use, adjust the diaphragm image using the field iris diaphragm lever so that the field of view is circumscribed by the field iris diaphragm to exclude stray light. CK40-RFL 4 Selecting Mirror Cubes }Select the mirror cube which matches the purpose of observation. Cube Dichroic Mirror/Filter Combinations Excitation Mirror Cube B CX-DMB-2 G CX-DMG-2 Dichroic Mirror DM500 DM570 Exciter Filter BP460-490 BP475-545 Barrier Filte Applicable Reagents OG515IF FITC Acridine orange Auramine GFP OG590 Rhodamine B200 TRITC Feulgen Propidium iodide Filter Characteristics G excitation (Super wide band) B excitation (Wide band) Spectral Transmittance (%) Spectral Transmittance (%) 100 DM500 BP460-490 OG515IF 50 0 300 400 500 600 Wavelength (mm) 700 800 100 BP475-545 DM570 BP475-545 50 OG590 0 300 400 500 600 700 800 Wavelength (mm) 18 OBSERVATION }The excitation light from parts other than the excitation light protective shield may be glaring to your eyes. However, this light is not hazardous because a UV cut filter is built into the lamp housing. # The objectives provided with the CK40 microscope can be used for fluorescent observation. However, when the phase contrast objective or relief phase contrast objective is used, the image may become slightly dark or the contrast may become degraded, due to the built-in phase plate. ƒ ³ | ² Fig. 27 @ 1. Pull out the filter slider @ fully toward the right side to engage the shutter into the light path. (Fig. 27) 2. Pull the left and right cube installation/removal levers ² fully outward to remove the cubes from the reflected fluorescent light path. (Fig. 27) # Pull out each lever until the index groove ³ at its deeper end is visible. (Fig. 27) 3. Perform transmitted light observation of the CK40 and move the specimen section to be observed to the center of the field of view. }The transmitted light observation can be either phase contrast observation or brightfield observation. If a specimen is hard to observe in brightfield observation, adjust the condenser aperture iris toward the closing direction. 4. Turn off the transmitted light, push in the installation/removal lever of the fluorescent mirror cube to be used in observation (until the index groove | on the lever end is aligned with the side panel ƒ of the cover), slide the filter slider to disengage the shutter from the light path, and perform fluorescent observation. (Fig. 27) }Adjust the collector focusing knob on the lamp housing to make the entire field of view uniform and as bright as possible. }Adjust the field iris diaphragm lever so that the field of view is circumscribed by the field iris diaphragm. }To interrupt observation for a short period, use the shutter and keep the burner ignited. Repeated switching ON and OFF of the burner will shorten its service life. Caution for fluorescence observation This unit provides maximum excitation intensity to enable image observation even of specimens with low fluorescence. Therefore, during observation with a high-power objective, fading may occur, resulting in loss of image brightness and contrast. To prevent this, reduce the intensity of the excitation light to reduce the speed of fading. It is advisable that the intensity of the excitation light is reduced by using an ND filter. Also, use the shutter so that the specimen will not be exposed to excitation light for longer than is necessary. By using a commercially available anti-fading agent (e.g., DABCO), fading of the specimen can be slowed. For high-magnification observation, in particular, use of an anti-fading agent is recommended. # Note that anti-fading agents are not applicable to some specimens. 19 CK40-RFL SPECIFICATIONS Module Specification Fluorescent vertical illuminator CK40-RFA Fluorescent lighting method: Reflected fluorescent lighting based on slide switching of dichroic mirror cubes. Number of fields of view: 20 (when WHK10X is used). Intermediate tube magnification: 1X. Shutter: Switchable using filter slider (shutter/open aperture/filter pocket). Excitation protective shield: Fixed on condenser with a screw. Possible observation modes: B excitation/G excitation. Lamp housing U-LH50HG Burner centering: Up/down, left/right. Mirror centering: Up/down, left/right. Burner focusing: Collector focusing, mirror focus, screw system. Safety mechanism: Interlock switch, UV cut filter built in. <<Applicable high-pressure mercury burners>> · HBO50W/AC (OSRAM) · CS50W4 (PHILIPS) Power supply U-RFLT50-100 U-RFLT50-200 Ignition system: Auto ignition. Hour counter: Displays accumulated hours of operation. Input rating: Input voltage switchable. U-RFLT50-100: 100/110/120 V AC , 2.5/2.3/2.0 A, 50/60 Hz. U-RFLT50-200: 220/230/240 V AC , 1.7 A, 50/60 Hz. Dimensions U-RFLT50-100: 150(W) x 320(D) x 150(H) mm. U-RFLT50-200: 150(W) x 210(D) x 150(H) mm. Weight U-RFLT50-100: 12 kg (26.4 lb.). U-RFLT50-200: 7 kg (15.4 lb.). Operating environment Indoor use. Altitude: Max. 2000 m. Ambient temperature: 5°C to 40°C. (41°F to 104°F) Maximum relative humidity: 80% for temperatures up to 31°C (88°F) decreasing linearly through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F). Supply voltage : Fluctuations not to exceed ±10% of nominal voltage. Pollution degree: 2 (in accordance with IEC664). Installation/overvoltage category: II (in accordance with IEC664). 20 TROUBLESHOOTING GUIDE Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If problems occur, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local Olympus representative for assistance. Problem Cause Remedy Page ND filter is engaged. Set filter slider to open aperture position. 15 Cube is not engaged properly in light path. Engage it properly in light path. Shutter is engaged in light path. Set filter slider to open aperture position. 15 The wrong 40X objective is mounted, or the objective’s correction collar has not been adjusted. Optimum fluorescence performance cannot be achieved with a 40X objective that does not have a correction collar. Use the LWDCDPlan40XPL-6 or LWDCDPlan40X with correction collar and adjust the collar. – Objectives and/or filters are dirty. Clean them. iii Field iris diaphragm is not correctly opened. Adjust field iris diaphragm so that its image circumscribes the field of view. 17 Objective is not engaged properly in light path. Be sure to stop the revolving nosepiese at a click position. – Cube is not engaged properly in light path. Engage it properly in light path. Filter slider is located in the middle of light path. Set filter slider to open aperture position. ND filter is located in the middle of light path. Set it to proper position. Mercury burner is not centered properly. Center mercury burner correctly. Collector focus deviates from correct position. Adjust to correct position. Power cord is not connected properly. Connect correctly. A fuse is blown. Replace it. 1. Optical System a) Mercury burner lights, but field of view remains dark. b) Image is unclear, blurred or has insufficient contrast. c) Field of view is vignetted, or it is not evenly illuminated. 19 19 15 15 15 16 2. Electrical System a) Main switch indicator does not light up. 21 10 11 CK40-RFL Problem Cause b) Main switch indicator lights, but the mercury burner does not ignite. Connectors are not connected properly. Connect correctly. Mercury burner is not mounted. Install a burner. 8 Lamp housing interlock switch is activated. Tighten the burner socket clamping screw. 9 Auto ignition is malfunctioning. Set main switch on the power supply unit to “\” (OFF), then set to “ l ” (ON) again. (Repeat as necessary.) 14 Push the switch inside the starter reset hole on the right side of power supply unit, then set main switch to “ l ” (ON) again. 14 Wait for at least 10 minutes after re-igniting the burner. 14 c) Mercury burner flickers or is dark. Auto ignition safety mechanism is activated. Insufficient time has elapsed since the burner was turned on. d) Mercury burner life is very short. Remedy Page 10 Input voltage and frequency Set them correctly. switches are not set properly. 10 Burner service life has expired. Replace the mercury burner. 8 Input voltage is not set properly. Set correctly. 10 This device complies with the requirements of both directive 89/336/EEC concerning electromagnetic compatibility and directive 73/23/EEC concerning low voltage. The CE marking indicates compliance with the above directives. 22 ■ PROPER SELECTION OF THE POWER SUPPLY CORD If no power supply cord is provided, please select the proper power supply cord for the eqipment by referring to “ Specifications ” and “ Certified cord ” below: CAUTION: In case you use a non-approved power supply cord for Olympus products, Olympus can no longer warrant the electrical safety of the eqipment. Specifications Voltage Rating Current Rating Temperature Rating Length Fitting Configuration 125V AC (for 100-120V AC area) or, 250V AC (for 220-240V AC area) 6A minimum 60°C minimum 3.05 m maximum Grounding type attachment pulg cap. Opposite terminates in molded-on IEC configuration appliance coupling. Table 1 Certified Cord A power supply cord should be certified by one of the agencies listed in Table 1 , or comprised of cordage marked with an agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of agencies listed in Table 1. In case you are unable to buy locally in your country the power supply cord which is equivalent and authorized agencies in your country. Country 23 Agency Certification Mark Country Agency Australia SAA Italy IMQ Austria ÖVE Japan MITI Belgium CEBEC Netherlands KEMA Canada CSA Norway NEMKO Denmark DEMKO Spain AEE Finland FEI Sweden SEMKO France UTE Switzerland SEV Germany VDE United Kingdom ASTA BSI Ireland NSAI U.S.A. UL Certification Mark CK40-RFL Table 2 HAR Flexible Cord APPROVAL ORGANIZATIONS AND CORDAGE HARMONIZATION MARKING METHODS Approval Organization Printed or embossed Har- Alternative Marking Utilizing monization Marking (May be Black-Red-Yellow Thread (Length located on jacket or insula- of color section in mm) tion of internal wiring) Black Red Yellow Comite Electrotechnique Belge (CEBEC) CEVEC <HAR> 10 30 10 Verband Deutscher Elektrotechniker (VDE) e.V. Prüfstelle <VDE> <HAR> 30 10 10 Union Technique de d´Electricite´ (UTE) USE <HAR> 30 30 10 Instituto Italiano del Marchio di Qualita´ (IMQ) IEMMEQU <HAR> 10 30 50 British Approvals Service for Electric Cables (BASEC) BASEC <HAR> 10 10 30 N.V. KEMA KEMA-KEUR <HAR> 10 30 30 SEMKO AB Svenska Elektriska Materielkontorollanstalter SEMKO <HAR> 10 10 50 Österreichischer Verband für Elektrotechnik (ÖVK) <ÖVE> <HAR> 30 10 50 Danmarks Elektriske Materielkontrol (DEMKO) <DEMKO> <HAR> 30 10 30 National Standards Authority of Ireland (NSAI) <NSAI> <HAR> 30 30 50 Norges Elektriske Materiellkontroll (NEMKO) NEMKO <HAR> 10 10 70 Asociacion Electrotecnica Y Electronica Espanola (AEE) <UNDE> <HAR> 30 10 70 Hellenic Organization for Standardization (ELOT) ELOT <HAR> 30 30 70 Instituto Portugues da Qualidade (IPQ) l np l <HAR> 10 10 90 Schweizerischer Elektro Technischer Verein (SEV) SEV <HAR> 10 30 90 Elektriska Inspektoratet SETI <HAR> 10 30 90 Underwriters Laboratories Inc. (UL) Canadian Standards Association (CSA) SV, SVT, SJ or SJT, 3 X 18AWG SV, SVT, SJ or SJT, 3 X 18AWG 24 MEMO 43-2, Hatagaya 2-chome, Shibuya-ku, Tokyo Japan (Premises/Goods delivery) Wendenstasse 14-16, D-20097 Hamburg, Germany (Letters) Postfach 10 49 08, 20034 Hamburg, Germany 2 Corporate Center Drive, Melville, N.Y. 11747-3157, U.S.A. 2-8 Honduras Street, London EC1Y OTX, United Kingdom This publication is printed on recycled paper. Printed in Japan 9712 M 020–@