Swema 3000 - Elma Instruments
Transcription
Swema 3000 - Elma Instruments
Manual for Swema 3000 Version 0604 4.91 Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 1 Contents Points to note before starting 3 The different programs used in Swema 3000 4 The different probes for Swema 3000 5 Instrument Swema 3000 with air velocity sensor SWA 31 6 How to put on/off holders for the probes 7 Measuring with APF (Hot wire anemometer and pito static pipe) 8 Recomended measuring points for circular and rectangular ducts 9 Measuring with DPF (Differential pressure) 11 Measuring with AF (Flow capture) 13 Measuring with BP (Flow capture) 14 Measuring with AP (All probes) 16 Measuring with AS (All probes) 17 Measuring with ASF (Hot wire anemometer) 18 Measuring with CO (Draught probe) 20 Measuring with LOG (All probes) 21 Measuring with LOGP (All probes) 22 Measuring with SwemaTwin - Balancing of ventilation 23 Settings 24 Note Book, Log Book and File 25 Parameters 26 Swema Terminal PC-program 27 Swema MultiPoint PC-progarm 28 Technical Data 30 Guarantee conditions 35 Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 2 Points to note before starting The Swema 3000 is a multi-purpose instrument for measuring air velocities, air flows, relative humidity, temperatures and differential pressure. Swema 3000 has a high calculation capacity, and you can store single or double measurement results in the memory. Measurement reports can easily be transfered to Windows or SwemaTerminal. The Swema 3000 has a number of different modes, each adapted to the different measurement you are about perform. Swema 3000 has completely replaceable probes which are each individually calibrated. If you have replaced a probe or connected it for the first time, the instrument always selects the program which is standard for the connected probe. In addition, the Swema 3000 reads in the calibration and only makes available the programs which are relevant to that particular type of probe. If you switch off the instrument and turn it on again, you will return to the program you were using. At delivery the instrument is set to use k2-factors that is reducing the duct surface when calculating flow in APF-mode. USB - Swema does not include any driver for USB on the delivered CD. The USB-port is used for upgrading of the software inside Swema 3000 and for external power supply. 220v adapter can be ordered (art.nr. 764.610). For updating the program contact Swema. RS232-cable (for Com-port) can be ordered. (Part.no. 759.030) The PC-Program SwemaTerminal is free and can be downloaded from www.swema.se. The PC-program SwemaMultipoint can be ordered. (Part.no. 763.710) Swema 3000 ver. 4.96 Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 3 The different programs used in Swema 3000 Choose program by pressing the MENU-KEY and move the marker to MODE with the up/down arrow key. Press ENTER, SET or right ARROW KEY and choose program with the Up/down arrow keys. Confirm your choice with ENTER or OK. Set the parameter for the current program in the same way. Press EXIT to start measuring. APF (Average Point air Flow) Measured points are taken with ENTER Average, max, min and measured points are showed on the display. The Swema 3000 calculates the air flow in l/s or m3/h. Enter the area as diameter, height x width or area direct. If you choose to indicate the height and width or diameter of the duct, the Swema 3000 calculates the effective area = the actual area times the duct factor k2 in accordance with the recommended measurement methods. The duct factor can be turned off in the settings. Use the APF mode for measuring the flow in ducts and the flow through certain grilles. DPF (differential Pressure Flow). Only available with a pressure probe connected. The flow in the measurement of pressure drop across devices is directly obtained in l/s or m3/h. You enter the kfactor, and the Swema 3000 calculates the air flow according to the formula: q = k sqrt ∆P AF (Average Flow) Only available with SwemaFlow 2000, 125 and 65. Average Flow is calculated during selectable measuring time and presented on the display. BP (Back pressure) Only available with SwemaFlow 125. This mode is for measurement over devices with low pressure drop and high flows. BP mode calculates the real flow without the capture over the device. AP (Average Point). Measures the actual values of the connected probe. Average, max, min and number of measure points are showed on the display. AP mode is used for measuring air velocity, flow, temperature, moisture and pressure. AS (Auto Sampling). The Swema 3000 automatically collects measured values at the interval set by the time constant. Use AS mode for example if you want to measure average velocity, max, min, temperature and standard deviation (turbulence) at a point. ASF (Auto Sampling Flow). Same as AS, but the Swema 3000 calculates the flow in l/s or m3/h. You enter the area as area direct. In ASF the Swema 3000 calculates the area without reduction. Use ASF mode for example to measure the flow and air velocity in fume cupboards, intake grilles and flat devices. ASF is not intended for measuring in ducts, and diameters and duct coefficient are therefore omitted. CO (Comfort). Only available with the omni directional draught probe SWA 03 connected. CO mode measures average velocity and temperature and standard deviation over three minutes, (period of time - selectable.) After storage and on printing out, the Swema 3000 calculates the DR (Draught Rating). The DR indicates as a percentage how many people find the air velocity unpleasant. The DR is based on experiments conducted at the Technical University in Copenhagen, directed by Professor Fanger. DR is included as a measure of undesirable air movements in the ISO 7730 standard adopted for indoor climates. LOG/LOGP Collects readings at user-selectable intervals and time constant. Use LOG/LOGP mode to measure throughout longer time intervals (Ex: variations of air velocity and temperature throughout a 24-hour.). LOG collects readings made at direct test points and LOGP series of readings presented in a report as max, min, avg and the like. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 4 The different probes for Swema 3000 Hot Wire Anemometers SWA 31 or SWA 31E. Ø 8...10mm, 66 cm long, SWA 31E extendable up to 116 cm. Measure range: air speed 0...10 m/s (option 10...30 m/s) and temperature -20...80°C. Possible measuring programs: APF, AP, AS, ASF, LOG and LOGP. Differential pressure SWA 07 or SWA 10. Measuring range: SWA 07 ±7000Pa, SWA 10 -100...1500Pa. SWA 10 has a built in valve that automatically zero checks the pressure. Possible measuring programs: APF, DPF, AP, AS, LOG and LOGP. Flow capture hoods SwemaFlow 65, 125 or 2000. Measuring range: SWF 65 2...65 l/s, SWF 125 2...125 l/s and SWF 2000 4...900 l/s. Possible measuring programs: AF, AP, AS, LOG, LOGP and BP (only with SWF 125). Draught probe SWA 03. Measuring range 0,05...3,0 m/s (omni directional) and 10...34°C. SWA 03 fulfils ISO 7726. Possible measuring programs: CO, AP, AS, LOG and LOGP. Relative humidity and temperature HygroClip S, SWA 12, 13 and 16. Measuring range: All probes 0....100%RH and temperature HygroClip S -40...85°C, SWA 12 and 13 -20...60°C and SWA 16 -20....150°C. Possible measuring programs: AP, AS, LOG and LOGP. Temperature Black globe, SWA 53, SWT 315, SWT 215 and with temperature handle SWA 25 also SWT 14, 18, 22, 28, 39, 50, 51 and 53. Measuring ranges from -50°C to 280°C depending on probe. Possible measuring programs: AP, AS, LOG and LOGP. Radio modem SwemaTwin transfers measured values on the reference valve from one Swema 3000 to another through radio modem or 40 meter cable. The flow relation between them is displayed, which makes adjustment according to the proportional method an easy task for just one person. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 5 Instrument Swema 3000 with air velocity sensor SWA 31 Probe connector RS232/SwemaTwin connector USB connector Battery hatch 2 x AA Batteries SWA 31 Display Menu-keys Arrow Keys Enter Key Display light On / Off Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 6 How to put on/off holders for the probes Start with removing the rubber list by griping it with your fingers. Pull it up about 1 cm to the opening and pull it out. Remove the rubber list by just pull it out. When you have removed the list put in the appropriate holders and slide them to the wanted positions. Now you can attach the probe. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 7 Measuring with APF (Hot wire anemometer and pito static pipe) Possible probes with APF are: SWA 31, SWA 31E or SWA 07 or SWA 10 with pito static pipe. MODE APF (Average Point Flow). Swema 3000 calculates the flow in l/s, m3/h or CFM. You enter the area as diameter, height x width or area direct. If you choose to indicate the height and width or diameter of the duct, the Swema 3000 calculates the effective area = the actual area less the duct coefficient k2 (see page 34) in accordance with the recommended measurement methods. To turn off the k2, see page 23. Use the APF mode for measuring the flow in ducts and the flow through certain grilles. When a pressure probe is connected the Swema 3000 calculate the air speed with the formula: V = sqrt (2 ∆P /ρ) ∆P= dynamic pressure =stagnation - static pressure (Pa) ρ= density of air = 1,293 x (B x 273)/(1013 x (273 + t)) (kg/m3) B = barometric pressure (calculated standard value 1013hPa) t = Air temperature ºC (calculated standard value 20ºC) Choose mode by pressing the MENU- KEY and move the marker to ”Mode” with Connect the air velocity or pressure probe and press ON. After a the up/down arrow key. Press ENTER or SET and choose mode with the up/down few seconds Swema 3000 is in measuring mode. Choose program by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose program with the up/down arrow key. Confirm your choice with ENTER or OK. arrow key. Confirm your choice with ENTER or OK. Set the parameters and press EXIT to start measuring. Set the PARAMETERS for measuring mode APF: Time constant, i.e. the damping or inertia of the instrument. Move the marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the up/down arrow key. Confirm your choice with ENTER or OK. Ø set the diameter of the duct you want to measure. Move the marker to ”Ø” with the up/down arrow key. Press ENTER or SET and choose the diameter. Choose among standard diameters with the up/down arrow key. To set the diameter to your choice press EDIT and change with the arrow keys. Press ENTER or OK to confirm. H x W set the height and width of the duct you want to measure. Move the marker to ”H x W” with the up/down arrow key. Press ENTER or SET and choose height with the arrow keys. Press When measuring, collect values with the ENTER button. When values are colENTER or OK and set the width with the arrow keys. Press EN- lected the display will show average-, TER or OK to confirm. maximum-, minimum-values and number of measurements. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 8 Recomended measuring points for circular and rectangular ducts From ”Methods for measuring air flow in ventilation systems”, by the Nordic Ventilation Group and the Building Research Council, 1998, ISBN 91-540-5827-9. For D≤400, k2 values are from NBI report Nordtest-proj. No. 1463-99. Circular cross-section: *) D≤ 160 mm → k2 = 0,92 (SWA 31), 0,89 (Pito Static Pipe) 160 < D ≤ 400 mm → k2 = 0,96 (SWA31), 0,95 (Pito Static Pipe) 400 < D ≤ 1250 mm → k2 = 0,98 Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 9 Necessary measuring straight before and after measuring position: According to ”Methods for measuring air flow in ventilation systems” Dh=Hydraulic diameter Round ducts Rectangular ducts Before measuring position a ≥5 x D a ≥6 x Dh After measuring position a ≥2 x D a ≥2 x Dh Reduction of Flow When selecting a circular or rectangular duct it is recommended to reduce the flow by a k2-factor. (According to ”Methods for measuring air flow in ventilation systems”) A reduction is recommended because the air velocity is reduced close to the walls of the duct because of friction. Set the k2-factor to on in the Settings for automatic compensation. See Settings. Area set the area directly. Move the marker to ”Area” with the up/down arrow key. Press ENTER or SET and choose area with the arrow keys. Press ENTER or OK to confirm. Unit choose the unit you want to measure with l/s, m3/h or CFM. See Settings. Atmospheric pressure see Parameters Temperature see Parameters see Note Book File see Note Book When all parameters are set press the EXIT-button to get back to measuring position. Swema 3000 will show selected area. When all values are collected you can choose if you want to clear or save your measurement. Collect values with the ENTER-button. When values are collected the display will show average-, maximum-, minimumvalues and number of measurements. When all values are collected you can choose if you want to clear or save your measurement. To clear press the CLEAR-button and to save press the SAVEbutton. If you press the SAVE-button the instrument saves your measurement at the first free memory place in the note book memory. The instrument also give a short signal and displays the memory place for a few seconds. See Note Book. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 10 Measuring with DPF (Differential pressure) Possible probes with DPF are: SWA 07 or SWA 10. MODE DPF (Differential Pressure Flow). The flow in the measurement of pressure drop across devices is directly obtained in l/s , m3/h or CFM. (Always use a k-factor that is used to calculate in l/s. When choosing m3/h the Swema 3000 will calculate into m3/h). You enter the k-factor, and the Swema 3000 calculates the air flow according to the formula: q = k sqrt ∆P Connect the pressure probe and press ON. After a few seconds Swema 3000 is in measuring mode. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode DPF: Time constant, i.e. the damping or inertia of the instrument. Move the marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the Up/down arrow key. Confirm your choice with ENTER or OK. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the parameters and press EXIT to start measuring. K-factor, set the K-factor you want Swema 3000 to calculate with. In Swema 3000 you have a memory for 20 K-factors. Move the marker to ”K Factor” with the up/down arrow key. Press ENTER or SET and choose with the up/down arrow keys which K-factor memory place (1-20) you want to set. Press EDIT and edit your K-factor with the arrow keys. Confirm your choice by pressing ENTER or OK two times. Unit choose the unit you want to measure with l/s, m3/h or CFM. See settings. Atmospheric pressure see Parameters Temperature see Parameters see Settings see Note Book When all parameters are set press the EXIT-button to get back to measuring position. Swema 3000 will show selected K-factor the pressure and the air flow. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 When measuring, collect values with the enter button. When values are collected the display will show average-, maximum-, minimum-values and number of measurements. SE-123 56 FARSTA [email protected] www.swema.com 11 Collect values with the ENTER-button. When values are collected the display will show average-, maximum-, minimumvalues and number of measurements. When all values are collected you can choose if you want to clear or save your measurement. To clear press the CLEAR-button and to save press the SAVEbutton. If you press the SAVE-button the instrument saves your measurement at the first free memory place in the note book memory. The instrument also give a short signal and displays the memory place for a few seconds. When all values are collected you can choose if you want to clear or save your measurement. Use the k-factor to calculate the air flow. Certain suppliers calculate the correction factor using the formula: q = sqrt (Ik ∆P) In such case you must enter: k = sqrt Ik in Swema 3000. If the supplier has not specified a k-factor, You calculate it easily if you have a pressure-drop diagram. Proceed as follows: q = k sqrt ∆P → k = q / sqrt ∆P. Now go into the diagram for example: ∆P = 100 Pa → sqrt ∆P = sqrt 100 = 10 for Ø 100: 100Pa correspond to 50 l/s Use the calculation for k: k= 50/10 → 5,0 So the k factor in this case was 5,0. Set this in Swema 3000 to obtain the flow in l/s, m3/h or CFM. If the diagram does not extend to 100 Pa, select for example 25 Pa (sqrt 25 = 5) Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 12 Measuring with AF (Flow capture) Possible probes with AF are: SwemaFlow 65, SwemaFlow 125 or SwemaFlow 2000. MODE AF (Average Flow) Average Flow is calculated during selectable measuring time and presented on the display (time constant + selected measuring time). Connect the flow capture hood and press ON. After a few seconds Swema 3000 is in measuring position. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode AF: Time constant, in AF the time constant is the waiting time before measuring, to stabilising the flow in the flow capture. Move the marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the Up/down arrow key. Confirm your choice with ENTER or OK. Measure time set the sampling time for how long you want to take an average value. Move the marker to ”Measure time” with the up/down arrow key. Press ENTER or SET and choo- Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the se time constant with the Up/down arrow key. Confirm your up/down arrow key. Press ENTER or SET choice with ENTER or OK. Flow Factor For SwemaFlow 2000 and as option for SwemaFlow 125 it is possible to multiply the flow with a flow factor. Flow factor=1.00 will not alter the flow. Warning: Changing to anything else will alter the reading from calibarted value. This can be used to compensate for example for leakage. and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the parameters and press EXIT to start measuring. Unit choose the unit you want to measure with l/s, m3/h or CFM. See Settings. Atmospheric pressure see Parameters Temperature see Parameters see Settings See Note Book File see Note book. When all parameters are set, press EXIT to start measuring. Start measuring with the ENTER-button (or button on SwemaFlow 125/2000). Swema 3000 will collect values and calculate the minimum, maximum and average air flow and standard deviation during the selected measuring time. When all values are collected you can choose if you want to clear or save your measurement. To clear press the CLEAR-button and to save When measuring, Start the measuring with press the SAVE-button. If you press the SAVE-button (or long the enter button. Then you can choose if press onn button on SwemaFlow 125/2000) the instrument you want to clear or save your measuresaves your measurement at the first free memory place in the ment. Note Book memory. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 13 Measuring with BP (Flow capture) Possible probe with BP is: SwemaFlow 125 MODE BP (Back pressure) When measuring in some ventilation systems (especially systems with low pressure drops) corrections of measured values have to be made for throttling of the systems that the air capture does. That correction the SwemaFlow 125 can do for you. By measuring two times with SwemaFlow 125, with and without an extra throttleing ring. Swema 3000 calculates the flow that reflects the true flow, with no capture interference. In the same way as AF mode the BP mode makes an average measurement over time. Connect the SwemaFlow 125 and press ON. After a few seconds Swema 3000 is in measuring position. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode BP: Time constant, in BP the time constant is the waiting time before measuring, to stabilising the flow in the flow capture Move the marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the Up/down arrow key. Confirm your choice with ENTER or OK Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the parameters and press Exit to start measuring. Measure time set the sampling time for how long you want to take an average value. Move the marker to ”Measure time” with the up/down arrow key. Press ENTER or SET and choose time constant with the Up/down arrow key. Confirm your choice with ENTER or OK. Unit choose the unit you want to measure with l/s, m3/h or CFM. See settings for more information. Atmospheric pressure see Parameters Temperature see Parameters See Settings See Note Book Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 For example. Start with the throttle ring on and press Enter to collect the first value, the display will show PO 1. When ready with the first value the display will show PO 2. Remove the throttle ring and coolect the second value by pressing Enter. The button on SwemaFlow 125 can also be used as a Enter-button. A long press will save a measurment. SE-123 56 FARSTA [email protected] www.swema.com 14 When all parameters are set press the EXIT-button to get back to measuring mode. When measuring with the throttle ring, the ring shall always be placed after the hot wires in the sense of flow direction. The display will show PO 1. Place the throttle ring on the measuring unit and place the capture over the valve and press enter. When ”Point 1” is measured you hear a beep and the display shows PO 2 . Within one minute you must measure ”Point 2” without the throttle ring otherwise the instruments returns to ”Point 1”. Remove the throttle ring and place the capture over the valve once more and press enter. It is also possible to take the first measure without the throttle ring and the second with the throttle ring. The display will now show the true value and measuring point 1 & 2. You can now choose if you want to clear or save your measurement. To clear press the CLEAR-button and to save press the SAVEbutton. If you press the SAVE-button the instrument save your measurement at the first free memory place in the note book memory. The instrument also give a short signal and display the memory place for a few seconds. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 After PO 2 is measured the diplay will calculate the true value and show it on the display. The display will also show point 1 & 2. Max is without the ring and min is with the ring. Then you can choose if you want to clear or save your measurement. SE-123 56 FARSTA [email protected] www.swema.com 15 Measuring with AP (All probes) Possible with all probes MODE AP (Average Point). Measures the actual values of the connected probe. Calculates avgerage and display Max and Min. Connect the selected probe and press ON. After a few seconds Swema 3000 is in measuring position. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode AP: Time constant, i.e. the damping or inertia of the instrument. Move the marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the Up/down arrow key. Confirm your choice with ENTER or OK. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the parameters and press Exit to start measuring. Atmospheric pressure see Parameters Temperature see Parameters See Settings See Note Book When all parameters are set press the EXIT-button to get back to measuring mode. Collect values with the ENTER-button. When values are collected the display will show average-, maximum-, minimumvalues and number of measurements. When all values are collected you can choose if you want to clear or save your measurement. To clear press the CLEAR-button and to save press the SAVEbutton. If you press the SAVE-button the instrument save your measurement at the first free memory place in the note book memory. The instrument also give a short signal and display the memory place for a few seconds. When measuring, collect values with the enter button. When values are collected the display will show average-, maximum-, minimum-values and number of measurements. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 16 Measuring with AS (All probes) Possible with all probes MODE AS (Auto Sampling). The Swema 3000 coutinuesly collects measured values at 0,5s. If the time constant is longer than 0,5s, the measured values overlap each other. Set the time constant to 0,5s or lower to sample at the rate of the time constant. Use AS mode for example if you want to measure average, max, min and standard deviation at a point. Connect any probe and press ON. After a few seconds Swema 3000 is in measuring mode. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode AS: Time constant, i.e. the damping or inertia of the instrument. Choose mode by pressing the MENU-KEY In AS mode it also set the sampling speed. Move the marker and move the marker to ”Mode” with the to ”Time Constant” with the up/down arrow key. Press ENTER up/down arrow key. Press ENTER or SET or SET and choose time constant with the Up/down arrow key. and choose mode with the up/down arrow key. Confirm your choice with ENTER or Confirm your choice with ENTER or OK. OK. Set the parameters and press Exit to start measuring. Atmospheric pressure see Parameters Temperature see Parameters See Settings See Note Book When all parameters are set press the EXIT-button to get back to measuring mode. Start and stop measuring with the ENTER-button. Swema 3000 will collect values and calculate the average. It will also calculate standard deviation, number of measuring points and maximum and minimum. When using a differential pressure probe air velocity will be measured. Connect a pito static pipe. When all values are collected you can choose if you want to clear or save your measurement. To clear press the CLEAR-button and to save press the SAVEbutton. If you press the SAVE-button the instrument save your measurement at the first free memory place in the note book memory. The instrument also gives a short signal and displays the memory place for a few seconds. Start and stop the sampling by pressing Enter. When the sampling are done you can choose if you want to clear or save your measurement. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 17 Measuring with ASF (Hot wire anemometer) Possible probes with ASF are: SWA 31 or SWA 31E. MODE ASF (Auto Sampling Flow). Same as AS, but the Swema 3000 calculates the flow in l/s, m3/h or CFM. In ASF mode the area is not reduced. Use ASF mode for example to measure the flow and air velocity in fume hoods and over grilles. ASF is not intended for measuring in ducts and diameters and duct coefficient are therefore omitted. Connect the air velocity probe and press ON. After a few seconds Swema 3000 is in measuring mode. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode ASF: Time constant, i.e. the damping or inertia of the instrument. Choose mode by pressing the MENU-KEY In ASF mode it also set the sampling speed. Move the marker and move the marker to ”Mode” with the to ”Time Constant” with the up/down arrow key. Press ENTER up/down arrow key. Press ENTER or SET or SET and choose time constant with the Up/down arrow key. and choose mode with the up/down arrow key. Confirm your choice with ENTER or Confirm your choice with ENTER or OK. Area set the area directly. Move the marker to ”Area” with the up/down arrow key. Press ENTER or SET and choose area with the arrow keys. Press ENTER or OK to confirm. OK. Set the parameters and press Exit to start measuring. Unit choose the unit you want to measure with l/s, m3/h or CFM. See Settings for more information. Start and stop the auto sampling flow by pressing Enter. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 18 Atmospheric pressure see Parameters Temperature see Parameters See Settings See Note Book Start and stop measuring with the ENTER-button. Swema 3000 will collect values and calculate the average air flow and temperature. It will also calculate standard deviation, number of measuring points and maximum and minimum air flow. When all values are collected you can choose if you want to clear or save your measurement. To clear press the CLEAR-button and to save press the SAVE-button. If you press the SAVE-button the instrument saves your measurement at the first free memory place in the note book memory. The instrument also gives a short signal and displays the memory place for a few seconds. When the sampling are done you can choose if you want to clear or save your measurement. Measure the total flow by gently sweeping the probe across the surface. Set the time constant as low as possible. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 19 Measuring with CO (Draught probe) Possible probe with CO is: SWA 03. MODE CO (Comfort). CO mode measures average velocity and temperature and standard deviation. In the note book you can also view DR (Draught Rating). DR indicates as a percentage how many people find the draught unpleasant. DR is included as a measure of undesirable air movements in the ISO 7730 standard adopted for indoor climates. DR is based on experiments conducted at the Technical University in Copenhagen, directed by Professor Fanger. Connect the draught probe and press ON. After a few seconds Swema 3000 is in measuring mode. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET Time constant, In CO mode each measurment value is 0,1s and choose mode with the up/down arrow key. Confirm your choice with ENTER or long so in CO mode the time constant is only meaningfull to set if you want to update the display faster than 0,5s. Move the OK. Set the parameters and press Exit to start measuring. Set the PARAMETERS for measuring mode CO: marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the up/down arrow key. Confirm your choice with ENTER or OK. Measure time here you select the length of the reading series. Swema 3000 calculate standard deviation, number of measuring points and maximum and minimum air speed and draught rating. ISO 7730 suggests 3 minutes. Atmospheric pressure see Parameters Temperature see Parameters See Settings See Note Book The measurment is automatically saved in the note book. To see the draft rate, press the MENU-key and move the marker to Note Book and press enter. When all parameters are set press the EXIT-button to get back to measuring mode. Start the draught measurment with the ENTER-button. Swema 3000 will take measurments during the selected measuring time and will automatically save to the first free memory place in the note book. The instrument also gives a short signal and displays the memory place for a few seconds. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 20 Measuring with LOG (All probes) Possible with all probes MODE LOG Collects readings at user-selectable intervals and time constant. Use LOG mode to measure throughout longer time intervals (Ex: variations of air flow and pressure throughout 24-hours.). Connect an 220V adapter if you want to log longer time than 12 hours. Connect any probe and press ON. After a few seconds Swema 3000 is in measuring position. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode LOG: Time constant, i.e. the damping or inertia of the instrument. Move the marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the Choose program by pressing the MENUUp/down arrow key. Confirm your choice with ENTER or OK. KEY and move the marker to ”Mode” with Interval set the log interval by moving the marker to ”Interval” with the up/down arrow key. Press ENTER or SET and choose interval with the up/down arrow key. Confirm your choice with ENTER or OK. the up/down arrow key. Press ENTER or SET and choose program with the Up/down arrow key. Confirm your choice with ENTER or OK. Set the parameters and press Exit to start measuring when your done. Atmospheric pressure see Parameters Temperature see Parameters See Settings See Note Book When all parameters are set press the EXIT-button to get back to measuring position. Start and stop the log with the ENTER-button. After the second push on ENTER the Swema 3000 will save the collected measurments to the first free memory place in the log book. The instrument also gives a short signal and displays the memory place for a few seconds. Note that together with a differential pressure probe air velocity will be measured. Connect a pito static pipe. Start and stop the log with the ENTER-button. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 21 Measuring with LOGP (All probes) Possible with all probes MODE LOGP Log series of readings presented in notes with maximum, minimum, average and the standard deviation. Each note represents one logging. Connect any probe and press ON. After a few seconds Swema 3000 is in measuring mode. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the PARAMETERS for measuring mode LOGP: Time constant, the time of each measurment value. Move the marker to ”Time Constant” with the up/down arrow key. Press ENTER or SET and choose time constant with the Up/down arrow key. Confirm your choice with ENTER or OK. Measure time here you select the length of the readings series. Swema 3000 calculate average, maximum, minimum and standard deviation. In this example: Each serie is set to 3 minutes. After one serie is finished another one start immediately. Interval here you select how often you want a reading series to start. Set the log interval by moving the marker to ”Interval” with the up/down arrow key. Press ENTER or SET and choose time constant with the Up/down arrow key. Confirm your choice with ENTER or OK. Atmospheric pressure see Parameters Temperature see Parameters See Settings See Note Book When all parameters are set press the EXIT-button to get back to measuring mode. Start and stop the log with the ENTER-button. Swema 3000 will log reading series and save to the first free memory place in the log book. The instrument also give a short signal and display the memory place for a few seconds. If you stop the logging before the end of the first serie, no data is saved. Start and stop the log with the ENTER-button. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 22 Measuring with SwemaTwin - Balancing of ventilation Connect 1 modem each to 2 Swema 3000 or a long cable in between the 2 Swema 3000:s. Always start by turning on the master first and select a mode that shows flow DPF, AF or APF. To get the relations in % you need to have the same unit on both instruments. If you have the 4000mAh modems you also have to start them manualy. Choose mode by pressing the MENU-KEY and move the marker to ”Mode” with the up/down arrow key. Press ENTER or SET and choose mode with the up/down arrow key. Confirm your choice with ENTER or OK. Set the Master - unit (The one that you will carry with you) Check that ”Communication ” is set to RS232 in the settings menu. Choose Settings by pressing the MENU-KEY and move the marker to ”Settings” with the up/down arrow key. Move the marker to ”Communication” with the up/down arrow key. Press ENTER or SET and choose ”RS232” with the up/down arrow key, confirm with Enter. Select the baud rate in the Settings menu. (Modems are normally set to 4800 (2AsxE or 1AsxE)). Press ENTER or SET and choose ”Baud Rate” with the up/down arrow key. Press enter and change the baud rate to 4800 with the up/down arrow keys, confirm with Enter. In the settings menu you should also set the Swema Twin function at ”ON”. Move the marker to ”Swema Twin” with the up/down arrow key. Press ENTER or SET and choose ”ON” with the Up/down arrow key, confirm with Enter. Set the Slave - unit Check that ”Communication ” is set to RS232. Choose Settings by pressing the MENU-KEY and move the marker to ”Settings” with the up/down arrow key. Move the marker to ”Communication” with the up/down arrow key. Press ENTER or SET and choose ”RS232” with the Up/down arrow key, confirm with Enter. Select the baud rate in the Settings menu. Press ENTER or SET and choose ”Baud Rate” with the up/down arrow key. Press enter and change the baud rate to 4800 with the up/down arrow keys, confirm with Enter. Leave the SwemaTwin function at ”OFF”. If you have the 4000mAh modems you also have to start them manualy. When measuring, the master shows the flow of both the Master and the Slave. It also shows the relation: Master in % of the reference instrument: (Master / Slave)*100%. The relation is the same calculation regardless of what the Master instrument displays with big digits. The relation is very handy when doing ventilation system adjustments according to the proportional method. Balancing a ventilation system according to the Proportional method: First decide where to place the reference instrument. The flow according to the drawing at this place is the nominal reference flow. The other measurement points are adjusted with the Master instrument. We call these nominal flow values from the drawing nominal Master flows. Use projected flows from drawing. Calculate for each Master measurement point: (nominal Master flow/nominal reference flow x %100). These are the relations that you shall obtain when you have balanced two diffusers/inlets/outlets. Measure with the reference instrument on the reference in/outlet and the Master instrument on the in/outlet you want to adjust. The relation is important and not the actual flow. The final flows will occur when the final adjustment of the total flow in the ventilation system is done. When no connection is establish between the Master and Reference instruments show ”zzz”. Charging of modem: 1.5hours, charger is constantly lit - charging. Charger blinks - battery is charged. If the charger blinks after 0,5 hours you need to start charging once more. The battery packs delivered by Swema supplies the Swema 3000 with power. One charge is good for 5-10 hours use with 1800mAh battery and 10-20 hours use with 4000mAh battery. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 23 Settings In the ”Settings” you can choose among following alternative: Move in the menu with the up/down arrow keys. Press ENTER, SET or right arrow key to select. Change with the up/down arrow keys. Swema Twin ON or OFF Only when you us SwemaTwin, set ON at your master unit Communication USB or RS232 Communicate via USB (cable 764.430) or RS232 (cable 759.030) to your PC Auto Zero ON or OFF When you have a SWA 10 pressure probe connected you can choose automatic zeroing when you collect values. Unit System SI or US Set which units you want to use Metric (SI) or American (US). Add Flow ON or OFF If set to ON. Swema 3000 will add the flow from each measurement in one file and present a total flow. Protocol SHORT or LONG Here you specify if you want a long or short protocol at PC-printout. 1 Unit Decimal 0, 1, 2, 3 or 4, Select number of decimals for the big digits 2 Unit Decimal 0, 1 or 2, Select number of decimals for the small digits Baud Rate 2400, 4800, 9600, 19200 or 57600 Normally 4800 for SwemaTwin, 9600 for PC-transfer (RS232 cable). Language English, German, Swedish, Finnish, French, Japanese, Danish, Norwegian, Dutch or Polish. Print Out Select if you want comma or point separation at PC-print out. Excel uses either comma or point. K2 Factor ON or OFF Set ON if you want Swema 3000 to calculate with a K2-factor. See page 9 ON: Flow = air velocity * k2 * area OFF: Flow = air velocity * area Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 Move in the menu with the up/down arrow keys. Press Enter, Set or right arrow key to select. Change with the up/down arrow keys. SE-123 56 FARSTA [email protected] www.swema.com 24 Note Book, Log Book and File To access the Note Book or Log Book press ”Menu” key. If you have selected one of the following modes: AP, APF, AS, ASF, DPF or CO. You can access the ”Note Book”. If you have selected one of the following modes: LOG or LOGP. You can access the ”Log Book”. To acess the measurments in the active file open Notes or Log file by move the marker to ”Note Book” or ”Log Book” with the arrow keys. To see your measurements or Log files press Enter, Set or right arrow. Browse between your measurements with the UP/ DOWN arrow keys. To transfer a note or log file to a PC just press Print. A long press on print will transfer the complete log book or note book. Open your Note or Log files by moving the For software’s see SwemaTerminal and Swema Multipoint. marker to ”Note Book” or ”Log Book” with Erase with the Erase-button and Exit with the Exit-button. the arrow keys and press Enter. If you have more than one file or if you want to start a new file. Move the marker to ”File” with the UP/DOWN arrow keys. To see your Log Files press Enter or Set. Browse between your log files with the UP/DOWN arrow keys. When you have selected the file you want to see, press Exit. Now you can see you measurements our log files by going to the ”Note Book” or ”Log Book” as described above. To start a new file, just press Enter while your in the Filemenu. To transfer to a PC, just press Print while you are in the Filemenu. A long press will transfer the complete note/log book. To erase a file, just press Erase while you are in the Filemenu. Press Exit to leave the file menu. Alter between your notes with UP/DOWN arrow keys. Example of a print out to PC. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 25 Parameters Atmospheric pressure is used when Swema 3000 does calculations for air velocity and air flow. Swema 3000 is available in two different models, with a built in barometer Swema 3000d or without Swema 3000. Swema 3000d automatically compensates for the air density (atmospheric pressure and temperature). In Swema 3000 the atmospheric pressure can be set manually by moving the marker to ”Atmos.P” with the up/down arrow key. Press ENTER or SET and choose atmospheric pressure with the up/down arrow key and alter position with left/right arrow key. Confirm your choice with ENTER or OK Temperature is used when Swema 3000 does calculations for air velocity and air flow. Swema 3000 is available in two different models, with an extra temperature thermocouple input probe Swema 3000d or without Swema 3000. Both Swema 3000 and Swema 3000d compensates for correct temperature when a probe with a built in temperature sensor is connected. Suchs as hot wire anemometers, flow capture, draught probes and relative humidity and temperature probes. Highlight the parameter you want to change and press Enter. When a differential pressure probe (SWA 10 or 07) is connected. Swema 3000d automatically compensates for the correct temperature when a thermocouple probe is connected. In Swema 3000 the temperature can be set manually when a differential pressure probe is connected. Move the marker to ”Temp” with the up/down arrow key. Press ENTER or SET and choose temperature with the arrow keys and alter position with left/right arrow key. Confirm your choice with ENTER or OK Change the parameter with the arrow keys and press Enter to confirm. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 26 Swema Terminal Swema Terminal is a freeware that can bee downloaded at www.swema.com SwemaTerminal only works with RS232 communication, cable 759.030 between Swema 3000 - PC. USB - Swema does not include any driver for USB on the delivered CD. The USB-port is used for upgrading of the software inside Swema 3000 and for external power supply. 220v adapter can be ordered (art.nr. 764.610). For updating the program contact Swema. This software reads data from Swema instruments through the serial port on the chosen COM port, sort data from clean logs from the Swema 3000 from some sensors and makes it easy to copy and paste data into Word, Excel or other programs. The serial port settings must be according to the Swema 3000 manual, baud rate 9600, 1 stop bit, no parity, no flow control and 8 data bits. Start with setting the right COM number corresponding to the one connected to the Swema 3000. Click on ’Connected’ to open the port. Press the ’Print’ button on the Swema 3000 according to its manual. The data will be read and added to the bottom of the text window to the left. This text can be saved, copied to the clipboard or cleared. It is also possible to open earlier saved data. The data can be edited in this window. If the read data from the Swema 3000 is of one single sort corresponding to one single log mode and one single sensor, it is possible to sort the data for some sensors according to the choices below the left text box. The decimal sign must be the same as the language settings in Windows that is changed in the Control Panel. If the input data have different form, it is possible to sort the first part. Then the program gives an error but data until the discontinuity will be sorted. Handle the sorted data, then remove the first part from the input data textbox and sort again. After sorting the data, the user can save the sorted data with tabs (ASCII 0d09) separating the variables. The clipboard can be used. The time interval is read from the Swema 3000 and the time in days is calculated with start at the given start time for the first reading. Time as float value is defined by Excel where the unit means 24 hours and 1 corresponds to 1900-01-01 00:00:00. By changing the cell format in Excel to time, it is possible to get the seconds but obviously not together with the date. Graphs can be made based on the sorted data, copied to clipboard or saved to a bitmap file. Graphs are zoomable by holding down the left mouse button and making a window. Zoom out is made by making a window with the right down corner first. Pan is made by holding down the right mouse button. Note that all units in tables, graphs and in the program window is given in the standard form where the unit is divided from the quantity in the header. The button ’Log’ in the lower left corner instructs the Swema 3000 to send the display values at the specified interval by sending a space (ASCII 0d32). An instrument program memory update may be needed. The maximum reading frequency is 10 Hz, that means an interval of 0.1 s. Swema AB Pepparvägen 27 SE-123 56 FARSTA 27 Tel: +46 8 94 00 90 Fax: +46 8 93 44 93 [email protected] www.swema.com Swema MultiPoint (Part.no. 763.710) Swema MultiPoint can monitor and log up to 8 Swema 3000 and present it in a live graph. See Screenshot on next page. SwemaMultipoint can only be used with Com-ports. To access more Com-ports an USB adpater (driver is included) with 8 Com-ports is available (Part.no. 763.700). The adapter is connected to a USBport on the PC. The adapter has 8 9 pole d-sub connectors. The PC-cable (Part no. 759.030) for Swema 3000 connects to the adapter. The Swema Multipoint Pro can be used to control the Swema 3000 to log data on the two variables showed on the LCD of the instrument. The bigger value is called Variable 1 and the smaller value is called Variable 2. The instrument, SwemaAir 300, may need an updated program memory. Use USB-8 com adapter (763.700) and up to 8 comport cable s to Swema 3000 (759.030). Firstly, set the channels that will be used and their corresponding COM port numbers. It is not possible to set the same number for more than one channel. 8 channels are available. It can be possible to run the program twice to get 16 channels if COM ports are available. The serial port settings must be according to the Swema 3000 manual, baud rate 9600, 1 stop bit, no parity, no flow control and 8 data bits. Click ’Open ports’ to open the ports. Set the interval described below. Click on ’Activate log’. Data is logged with the time interval given in the ’Time between readings’-box. After the ’Measuring period’, the logged data is analysed with respect to mean, standard deviation, minimum and maximum period values. Thereafter, another log is started after ’Log interval’ counted from the former start. Draught index is calculated based on speed and temperature. If other sensors are used, the value is calculated anyway but is only relevant if Variable 1 contains speed in m/s and Variable 2 contains temperature in °C. Data is saved in a number of files. Set the directory for these files. There are two files, 1_all.dat and 2_all.dat that contains a matrix of time and the different channels for the two variables. The file 1_mean.dat contains a matrix of time and channels for the analysed data and so on. The files s1_all.dat and so on contains the channel 1 and so on data with a matrix of time and the two variables for all logged points. The file s1_int.dat contains channel 1 analysed data in a matrix with time and mean, standard deviation, min and max for the two variables respectively. All files are text files that are tab separated. If a file exists, new rows are added in the end of the file. If the computer does not keep up with the necessary speed, try to switch off the antivirus program or switch off the saving of all points that can be made in two steps. Next step could be to close other processes or increase the priority of this program. In Windows XP it is important to give priority. Press (START), right click on (My Computor), then choose properties, choose (Advanced) then in the first part choose (Program). The other choice (Background Services) should be choosen, press (OK). Time as a float value is defined by Excel where the unit means 24 hours and 1 corresponds to 1900-01-01 00:00:00. By changing the cell format in Excel to time, it is possible to get the seconds but obviously not together with the date. The input values must have decimal separator according to the language settings of Windows, changed in the Control Panel. Graphs are created and can be copied to Clipboard or saved as bitmap files. The number of points is set and when the number is reached, the earliest points are erased. If too many points are present, the computer can be slowed down. Zooming is possible if the mouse left button is held down. Make a window from upper left to zoom in and from down right to zoom out. It is possible to remove channels from the graphs and by that save one graph for each channel. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 28 Swema MultiPoint Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 29 Technical Data: Extend the probe to the desired length. Both holes must be clear for air to pass through. AIR VELOCITY SENSOR (THERMISTOR) DO NOT TOUCH THE THERMISTOR! The sensor must be replaced if it breaks. The Swema 3000 has individually calibrated probes. For repair, it is sufficient to send in the probe. TEMPERATURE SENSOR Ni 100 Check that the sensors are free of dust. Blow carefully! Note: Turn the arrow so that the market on the bottom part of the telescope stem points in the same direction as the arrow. Do not pull on the cable when retracting the telescope stem! Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 30 General The uncertainty of measurement for each probe is stated below as the uncertainty for the probe together with a calibrated Swema 3000. That is, included in the uncertainty for each probe is the accuracy for Swema 3000, interchangeablity, calibration, temperature dependence, hysteresis and repeatability. The user should correct the measured values with the correction on the probe calibration protocol to obtain stated accuracy. Swema 3000 instrument: Instrument temperature: 0...+50°C Memory capacity, notebook - logbook: approx. 1 600 notes or 14 000 pairs of values or 20 000 singel values Interchangeability accuracy: Negligible in relation to accuracy of probes. Digital Output: RS232, send : Swema 3000 sends the displayed value (big digits) (Transmitting rate max 4 times/second) USB Battery: Battery life with SWA 31 at 1 m/s: 2 x 1,5V AA 13 hours with NiMH, 17 hours with alkaline Use the 220V-adapter (764.610) for continuous operation. Battery with backlight on: 5,5 hours with NiMH, 8,5 hours with alkaline Instrument Calibration: Recommended every 5 year Only on Swema 3000d part no. 764.201 Barometer 600...1200hPa, ±2,5hPa Thermo couple Type K -270...1370°C ±0,3°C at -10...70°C Draught probe, SWA 03: Measuring range of velocity Measuring media Accuracy at 23°C: 0,05...3,0 m/s at +10...+34°C Dry and moist air, non-aggressive gases ±0,03 m/s at 0,05... 1,00 m/s ±3% read value at 1,00...3,00 m/s ±0,04 m/s at 0,05... 1,00 m/s ±4%read value at 1,00...3,00 m/s at 10…34°C: Response time of velocity sensor: Measuring range of temperature: 0,2 sec. +10...+40°C Accuracy: ±0,3°C at 20°C ±0,5°C at 10...+40°C Probe Calibration: Recommended every 6 months. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 31 Air velocity probe, SWA 31: Ø8mm at top, Ø10mm first step on telescope (handle) Measuring media Dry and moist air, non-aggressive gases Measuring range of velocity 0,1...30 m/s at -10...+45°C Accuracy at 20°C: 0,10...1,33 m/s ±0,04 m/s 1,33...30 m/s ±3% of read value 0,10...1,10 m/s ±0,05 m/s 1,10...30 m/s ±4,5% of read value at other: Measuring range of temperature: -20...+80°C Accuracy: ±0,3°C at 20°C ±1,0°C at -20...+80°C Probe Calibration: Recommended every 12 months. Pressure probe SWA 10: SWA 07: Pressure range: -300…1500Pa ±7 000 Pa Max. overload: ±20 000 Pa ±35 000 Pa Accuracy: ±1% read value ±1 Pa ±2% read value Minimum 0,3Pa After zeroing: ±0,3 Pa ±2% read value Temperature dependence: 0,2Pa/°C Long-term stability, Zero point: Full measuring range: 0,1 %FSO/year 0,1 %FSO/year Measuring media: Dry and moist air, nonaggressive gases Probe Calibration: Recommended every 12 months. 0,4 Pa/°C Temperature SWA 25 / 53 / SWT 315 / 215 / 221: Measuring range (probe depending): Interchangeability accuracy: -50...+280°C max ±0.1°C Probe Calibration: Sensors: Recommended every 24 months. According DIN 43760 (approx. ±0.5 °C at room temperature.) Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 32 SwemaFlow 125: Flow measurement range: 2-125 l/s (7 - 450 m3/h) The instrument always displays 0 when the flow is below 2 l/s (7 m3/h). Measuring media: Dry and moist air, non-aggressive gases Accuracy: ca ± 3% read value, min ± 1 l/s Temperature range -10...+50 °C Accuracy: ± 0.3 °C at 20 °C ± 0.5 °C at other Probe Calibration: Recommended each 6:e month. Battery: In the Swemaflow 125 there is a rechargeble battery pack of nickel-metal-hydride type (NiMH). A full charged battery last a normal working day. At high Air flow it may last a shorter time. Indication when battery has to be recharged. When the battery is nearly empty the battery symbol on the Swema 3000 blinks. When the battery is completely empty the Swema 3000 turns itself off. (When the Swema 3000 battery is nearly empty the battery symbol is continously lit. When the battery is empty Swema 3000 turns itself off.) Charge the battery when you want. Use the charger that is delivered together with the SwemaFlow 125. Do not use another charger which may be dangerous. Connect first the charger to a wall connector and secondly to the SwemaFlow 125. A red diod is lit on the SwemaFlow 125. If the red diod is blinking disconnect and try to connect once again. When the battery is fully charged the diod blinks. It takes one hour to charge the battery. If the Swema 3000 is connected when charging the SwemaFlow 125 it should be turned off. The battery is protected by over charging and when the battery is charged, trickel charging will go on until the battery charger is disconnected. Charging the batteries should take place at an instrument temperature between 10 - 45°C Pressure drop SwemaFlow 125 Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 33 SwemaFlow 65: Flow measurement range: Measuring media Accuracy at 18…30°C: at 10…18, 30…40°C: 2-65 l/s (7 - 230 m3/h) below 2 l/s the instrument displays 0. Above 65 l/s the display blinks - then the value is uncertain. Dry and moist air, non-aggressive gases ± 4% read value, minimum ± 1 l/s ± 6% read value, minimum ± 1 l/s Probe Calibration: Recommended each 6:e month. Battery: The Swemaflow 65 has a rechargeble nickel-metallhydrid (NiMH) battery pack. A full charged battery last a normal working day. At high Air flow it may last a shorter time. When the battery is nearly empty the battery symbol on the Swema 3000 blinks and the red diod on SwemaFlow 65 is shut off. When the battery is completely empty the Swema 3000 turns itself off. Use the charger that is delivered together with the SwemaFlow 65. Do not use another charger which may be dangerous. It takes approx. 3 hours to charge the battery. The battery is protected by over charging and it may also be charged before it is empty. SwemaFlow 2000: Flow measurement range: 4-900 l/s (15 - 3240 m3/h) below 4 l/s the instrument displays 0. -10...+50 °C Dry and moist air, non-aggressive gases ± 4% read value, minimum ± 1 l/s ± 0,3°C at 20°C, ± 0.5°C at -10...+50°C Recommended each 6:e month. about 2 hours Temperature range: Measuring media: Accuracy: Probe Calibration: Battery charge time: Weight including flow capture 650x650mm: 3,9 kg (4,3 kg including the Swema 3000) Battery: Same charger as for SwemaFlow 125. See page 32 Humidity and Temperature SWA 12-13 and 16 + cable and Hygroclip + cable: Relative Humidity range: Measuring media: Accuracy SWA 12…13 and 16 at 23°C: Hygroclip at 23°C: 0...100%RH Dry and moist air, non-aggressive gases ± 1,6% at 10...90%RH others ± 2,6%RH ± 1,6% at 0...100%RH Temperature range SWA 12…13: SWA 16: Hygroclip S (other models) : -20...+60 °C -20...+150 °C -40…+85°C (-50…+200°C) Accuracy temperature: ± 0,3 °C Probe calibration: Recommended each 12 months. Calculations: Dew Point and Water content (mg water / kg dry air) The instrument shows water content on the display by g/kg Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 34 Guarantee conditions 1. SWEMA provides the original purchaser of this instrument with a guarantee covering original defects in materials or manufacture for one (1) full year from the original delivery date. Utilization of the guarantee for any part of the instrument does not extend the guarantee period beyond the limit of one year from the original delivery date. Batteries are not covered by the guarantee nor are any defects which arise out of failure to follow the instrctions nor defects arising as a result of normal usage, wear and tear, omitted maintenance, unsuitable storage or neglect, unauthorized tampering and repairs carried out by a technician not recommended by SWEMA. 2. Claims under the guarantee must be made in writing before the expiration of the guarantee period. 3. Should SWEMA find the product defective as indicated and claimed for under the guarantee conditions, SWEMA has the right to choose between reparing the product free of charge or replacing it with another product in perfect condition. If a claim under the guarantee is made the product must be returned to SWEMA, carriage paid. SWEMA will pay the return carriage. Goods forwarded COD will not be accepted. 4. SWEMA’s liability for defects is limited to what is stated above. Accordingly, the purchaser is not entitled to claim compensation for direct or indirect damage or loss or submit any other claims in consequence of defects besides those specified in the foregoing. Swema AB Tel: +46 8 94 00 90 Pepparvägen 27 Fax: +46 8 93 44 93 SE-123 56 FARSTA [email protected] www.swema.com 35