Recommended Dental Equipment Maintenance Programme
DAILY MAINTENANCE
1. Wipedown dental unit and surgery surfaces using diluted non alcoholic solution. Do not spray directly onto surface - spray
onto soft cloth. Do not use
excessive solution.
2. Clean vacuum lines using non
foaming vacuum line solution
formulated per product
instructions. Clean or replace
solids collectors in every dental
Clean or replace amalgam trap daily.
unit. Autoclave vacuum
handpieces where possible.
3. Flush dental unit water lines at
beginning and end of each day.
4. Turn off dental unit master
switch at end of day.
Spray handpiece lubricant into
second largest terminal hole
5. Ensure handpieces are lubricated
at back of handpiece.
Ensure lubricant reaches
prior to every autoclave cycle.
the handpiece head.
6. Check distilled water level in
autoclave. Refill where necessary.
7. Turn off compressor and vacuum
at end of day.
8. Turn off O2 and N20 gas tanks
Always utilise handpiece lube
at end of day.
adaptors where available.
WEEKLY MAINTENANCE
1. Disinfect dental unit water lines as
instructed by equipment
manufacturer, then store water
lines dry over weekend.
2. Change solution in ultrasonic
cleaner and clean tank.
3. Clean solids collector at vacuum
pump (wet system suction).
4. Check separator at dental
unit (dry system suction).
Clean or replace as necessary.
5. Check oil level in compressor
(for lubricated models).
Replenish as necessary.
6. Drain compressor tank
(may be required daily).
7. Check oil level in dental chair
(for hydraulic models).
Disinfect water lines, then leave
dry over weekend by fitting an
empty bottle and flushing unit
until water runs out.
MONTHLY MAINTENANCE
1. Check filter for mains water (where fitted).
2. Clean Plaster trap.
Recommended handpiece air pressures
3. Check air
Instrument
Recom. pressure
pressure
Turbines
Dabi Atlante Maxi 400,
2.4 bar (35psi)
MS - 350 & RS - 350
settings to
KaVo 640 B
2.2 bar (32psi)
handpieces.
W & H 700 series
2.2 bar (32psi)
4. Check rubber
NSK
2.2 bar (32psi)
goods on
Midwest Tradition
2.2 bar (32psi)
Nitrous Oxide
KaVo 650 B
2.8 bar (41psi)
System.
W & H 898 LE
2.8 bar (41psi)
Replace split
Air motors
Most brands
3.1 bar (45psi)
tubings.
Ensure handpiece air pressures are set per handpiece
5. Check contents
manufacturers instructions
of emergency resuscitation kit.
SEMI-ANNUAL MAINTENANCE
1. Change compressor air filters.
2. Lubricate joints on dental light
and delivery systems.
3. Clean x ray developer tanks
with a commercial cleanser.
Tighten
loose
screws and
lubricate
joints
MISCELLANEOUS MAINTENANCE
1. Save all equipment invoices for reference.
2. Keep maintenance log.
3. Mark down name of serviceman
that did the work for you.
4. Save work orders in case
of ‘call backs’.
5. Save all booklets, manuals,
trouble shooting guides and
circuit diagrams for equipment.
6. Sterilise / autoclave all
handpieces and instruments
before sending to repair company.
Save all Manuals
For more information please contact:Clean or replace solids collector
at vacuum pump.
TWICE MONTHLY MAINTENANCE
1. Change solution in x ray developer and clean tanks.
2. Empty and Clean handpiece exhaust oil collector on dental unit.
Phone (02) 9153 6700 Fax (02) 9153 6677