Page 1 D06025_04_Insert_CK (IFCC)_MU EN CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) THIS PACKAGE INSERT IS APPLICABLE FOR USE OUTSIDE THE US. INTENDED USE For in vitro diagnostic use in the quantitative determination of creatine kinase (ATP: Creatine NPhosphotransferase (CK), EC 2.7.3.2) activity in human serum or plasma on Konelab and Indiko analyzers. Any reference to the Konelab systems also refers to the T Series. SUMMARY (1,2) Creatine kinase activity is greatest in striated muscle, brain, and heart tissue. Other tissues, such as the kidneys, the diaphragm, and the prostate, contain significantly less activity. CK activity is elevated in many diseases, including those involving skeletal muscle, the heart, the central nervous system, and the thyroid. After acute muscle damage the total CK activity in the blood begins to rise, reaching pathological levels after a mean duration of 4 - 5 hours in 50% of patients. The values elevate regularly from the 8th - 24th hour after onset of infarction and return to the reference range with great interindividual fluctuations. PRINCIPLE OF THE PROCEDURE The CK assay is two part liquid assay that utilizes three enzymes to determine acitivity of CK. Creatine kinase catalyses the reaction between creatine phosphate and ADP (adenosine-5’diphosphate) with formation of creatine and ATP (adenosine-5’-triphosphate). The latter compound phosphorylates glucose to G-6-P (glucose-6-phosphate) in the presence of HK (hexokinase). G-6-P is oxidised to 6-PG (gluconate-6-phosphate) and in the same time an equimolar amount of NADP (nicotinamide adenine dinucleotide phosphate) is reduced to NADPH in a reaction catalysed by G6P-DH (glucose-6-phosphate dehydrogenase). The increase in absorbance at 340 nm, resulting from the formation of NADPH, is proportional to the activity of creatine kinase in the sample (3). CK Creatine phosphate + ADP → Creatine + ATP HK ATP + Glucose → ADP + G-6-P Sample 4 hours at 20…25 °C, 1 month at 4…8 °C, or 1 month at -20 °C; protected from light. CK-BB not stable. TEST PROCEDURE Refer to the Application Notes for an automated procedure on your analyzer. Any application which has not been validated by Thermo Fisher Scientific Oy cannot be performance guaranteed and therefore must be evaluated by the user. Indiko instrument settings Enter the application parameters via barcode found in the Indiko application sheet or via electronical file as appropriate. Materials provided Reagents as described above. Materials required but not provided Calibrator and controls as indicated below. Calibration The response (dA/min) is converted to result units by a calculation factor. Or use eCal (981830) as a calibrator according to the instructions given in the package insert of eCal (981830). Quality Control Use quality control (QC) samples at least once a day and after each calibration and every time a new bottle of reagent is used. It is recommended to use at least two levels (low and high) of controls or sample pools. Always follow the local, state and federal regulations in performing QC. Available controls: CK-MB Control, code 981908 CK-MB Control High, code 981255 Nortrol, code 981043 Abtrol, code 981044 The control (QC) intervals and limits must be adapted to the individual laboratory requirements. The results of the quality control samples should fall within the limits pre-set by the laboratory. CALCULATION OF RESULTS The results are calculated automatically by the analyzer as follows: G-6-P-DH G-6-P + NADP → 6-PG + NADPH + H+ + Activity in U/l = Factor x dA/min REAGENT INFORMATION Kit code 981828 contains 2 vials of 52 ml Reagent A 2 vials of 13 ml Reagent B Kit code 981829 contains 2 vials of 20 ml Reagent A 2 vials of 5 ml Reagent B Concentrations Reagent A Buffer (Imidazole), pH 6.5 D-glucose N-acetyl-L-cysteine (NAC) Magnesium acetate EDTA-Na2 Nicotinamide adenine dinucleotide phosphate (NADP) Hexokinase (HK) NaN3 Reagent B Buffer (Imidazole) Adenosine-5’-diphosphate (ADP) Adenosine-5’-monophosphate (AMP) Di(adenosine-5’)penta-phosphate (AP5A) Glucose-6-phosphate dehydrogenase (G-6-P-DH) EDTA -Na2 Creatine phosphate NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1 % TV x 1000 Theoretical factor = 6.22 x SV x P TV SV 6.22 P Total reaction volume in ml Sample volume in ml Extinction coefficient of NADPH at 340 nm Cuvette pathlength = 1 Conversion factor µkat/l = (U/l)/60 If eCal (981830) is used as a calibrator, the results are calculated automatically by the analyzer using a calibration curve. Calibration Curve (example) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Precautions Reagent B is hazardous. See separate sheet inside the kit for Hazardous- and Precautions-phrases: H315, P280, P302+P352, P305+P351+P338. The reagents contain sodium azide as preservative. Exercise the normal precautions required for handling all laboratory reagents. The product has to be disposed of as laboratory chemical in accordance with local regulations. Preparation Reagent A and B are ready for use. Note: Check that there are no bubbles in the bottleneck or on the surface of the reagent when you insert the reagent vials or vessels in the analyzer. Storage and Stability Reagents in unopened vials are stable at 2…8 °C until the expiration date printed on the label. Refer to the Application Notes of your analyzer for the on board stability of reagents. If the reagents show turbidity, do not use them. SPECIMEN COLLECTION Note: When processing samples in sample collection tubes, follow the instructions of the tube manufacturer carefully to avoid erroneous results. Pay special attention to the preanalytical variables such as mixing, standing time before centrifugation and centrifuge settings. Sample Type Unhemolyzed serum or heparin plasma can be used. Note: Hemolysis can lead to deviating results. Avoid exposure of samples to strong light. Precautions Human samples should be handled and disposed of as if they were potentially infectious. The calibration curve is lot dependent. LIMITATIONS OF THE PROCEDURE For diagnostic purposes, the results should always be assessed in conjunction with the patient’s medical history, clinical examination and other findings. Interference A study was performed using the CLSI document EP7-A2 as a guideline. Criterion: Difference within ± 10% of initial values. Konelab Interfering Substance Conjugated bilirubin Unconjugated bilirubin Hemoglobin in hemolysate Lipemia: Intralipid (trademark of Fresenius Kabi AB) Interferent Concentration Target CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Difference from Target (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Storage (4) The sample can be stored as shown below. D06025_04_Insert_CK (IFCC)_MU Indiko Interfering Substance 4) Conjugated bilirubin Unconjugated bilirubin Hemoglobin in hemolysate Lipemia: Intralipid (trademark of Fresenius Kabi AB) Interferent Concentration Target CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Difference from Target (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Avoid using hemolyzed specimens. There is a poor correlation between turbidity and triglycerides concentration. In very rare cases gammopathy, in particular type IgM (Waldenström’s macroglobulinemia), may cause unreliable results (6). Not tested at Thermo Fisher Scientific Oy. For other interfering substances, please refer to the reference (5). EXPECTED VALUES (3) Male < 171 U/l (< 2.85 µkat/l) at 37 °C Female < 145 U/l (< 2.41 µkat/l) at 37 °C The quoted values should serve as a guide only. It is recommended that each laboratory verify this range or derives a reference interval for the population that it serves. Conversion factor µkat/l = (U/l)/60 MEASURING RANGE 10 – 1000 U/l (0.17 – 17 µkat/l). Extended measuring range after secondary dilution: 10 – 9000 U/l (0.17 – 150 µkat/l). PERFORMANCE CHARACTERISTICS The results obtained in individual laboratories may differ from the given performance data. Limit of Detection A study was performed using document EP17-A as a guideline. Konelab LoB: 0.6 U/l (0.01 µkat/l) The Limit of Blank (LoB) represents the highest measurement result that is likely to be observed for an analyte-free sample (n = 60). LoD: 1.5 U/l (0.03 µkat/l) The Limit of Detection (LoD) represents the lowest amount of analyte in sample that can be detected (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) The Limit of Quantitation (LoQ) represents the lowest actual concentration in a sample that can be quantitatively determined (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) The Limit of Blank (LoB) represents the highest measurement result that is likely to be observed for an analyte-free sample (n = 60). LoD: 2.2 U/l (0.04 µkat/l) The Limit of Detection (LoD) represents the lowest amount of analyte in sample that can be detected (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) The Limit of Quantitation (LoQ) represents the lowest actual concentration in a sample that can be quantitatively determined (n = 42). Imprecision The imprecision of the CK assay is ≤ 5.0% Total CV. Konelab CK Repeatability (Within run) Within device (Total) Mean 92 U/l SD CV% 1.7 1.8 Mean 226 U/l SD CV% 2.0 0.9 Mean 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 The precision study was performed using CLSI Document EP5-A2 as a guideline at one site with three Konelab 60 analyzers, one operator, 2 reagent lots, one calibrator lot and 6 calibrations for 21 days, with the number of measurements being n = 84. Results from each analyzer were combined. Indiko CK Repeatability (Within run) Within device (Total) Mean 95 U/l SD CV% 1.2 1.2 Mean 226 U/l SD CV% 2.0 0.9 Mean 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 The precision study was performed using CLSI Document EP5-A2 as a guideline at one site with three Indiko analyzers, one operator, 2 reagent lots, one calibrator lot and 7 calibrations for 21 days, with the number of measurements being n = 84. Results from each analyzer were combined. Method Comparison Konelab A comparison study was performed on Konelab Prime 60 using CLSI Document EP9-A2 as a guideline and the previous 981372 CK(IFCC) method (x) as the reference. Linear regression (Deming) (result unit U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 The sample concentrations were between 17 and 2253 U/l. Indiko 1) A comparison study was performed on Indiko using CLSI Document EP9-A2 as a guideline and the previous 981372 CK(IFCC) method on Konelab Prime 60 (x) as the reference. Linear regression (Deming) (result unit U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 The sample concentrations were between 17 and 2253 U/l. 2) A comparison study was performed on Indiko using CLSI Document EP9-A2 as a guideline and the same CK(IFCC) method on Konelab Prime 60 (x) as the reference. Linear regression (Deming) (result unit U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 The sample concentrations were between 14 and 2450 U/l. BIBLIOGRAPHY 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 5) 6) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. MANUFACTURER Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Date of revision (yyyy-mm-dd) 2015-05-08 Changes from previous version The line beside the text indicates changes. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 2 D06025_04_Insert_CK (IFCC)_MU CS Zvláštní upozornění S lidskými vzorky je nutné manipulovat a likvidovat je, jako by byly potenciálně infekční. CK (IFCC) 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) Uskladnění (4) Vzorky lze uchovávat podle dále uvedených pokynů. 4 hodiny při 20…25 °C, 1 měsíc při 4…8 °C nebo 1 měsíc při-20 °C. Vzorek je nutno chránit před světlem. CK-BB nestabilizovaná. Vzorek TATO PŘÍBALOVÁ INFORMACE JE URČENA PRO POUŽITÍ MIMO ÚZEMÍ USA. POUŽITÍ Pro diagnostické použití in vitro při kvantitativním stanovení aktivity kreatinkinázy (ATP: kreatin-Nfosfotransferáza (CK), EC 2.7.3.2) v lidském séru nebo plazmě na analyzátorech Konelab a Indiko. Každá zmínka o systémech Konelab se rovněž týká série T. SHRNUTÍ (1,2) Aktivita kreatinkinázy je nejvyšší v příčně pruhovaných svalech, mozku a srdeční tkáni. V dalších tkáních, jako jsou ledviny, bránice a prostata, je jejich aktivita podstatně nižší. Aktivita CK je zvýšená při mnoha onemocněních, včetně onemocnění kosterních svalů, srdce, centrální nervové soustavy a štítné žlázy. Po vzniku akutního poškození svalu se celková aktivita CK v krvi začne zvyšovat, a u 50 % pacientů dosahuje patologických hladin průměrně po uplynutí 4–5 hodin. Hodnoty se pravidelně zvyšují počínaje 8.–24. hodinou po začátku infarktu a vracejí se do referenčního rozmezí s velkými individuálními odchylkami. PRINCIP POSTUPU Stanovení CK je dvoudílné tekuté stanovení, které pro stanovení aktivity CK využívá tří enzymů. Kreatinkináza katalyzuje reakci mezi kreatinfosfátem a ADP (adenosin-5’-difosfát), při níž se vytváří kreatin a ATP (adenosin-5’-trifosfát). Druhá z těchto sloučenin fosforyluje glukózu na G-6-P (glukózo-6-fosfát) za přítomnosti HK (hexokinázy). G-6-P je oxidována na 6-PG (glukonát-6-fosfát) a zároveň je ekvimolární množství NADP (nikotinamid adenin dinukleotid fosfát) redukováno na NADPH v reakci katalyzované G-6P-DH (glukózo-6-fosfát dehydrogenáza). Zvýšení absorbance při 340 nm, které je důsledkem vytvoření NADPH, proporcionálně odpovídá aktivitě kreatinkinázy ve vzorku (3). POSTUP STANOVENÍ Údaje o automatizovaném postupu práce na analyzátoru naleznete v aplikačních poznámkách. Nelze zaručit účinnost žádné aplikace, která nebyla validována společností Thermo Fisher Scientific Oy. Taková aplikace musí být proto hodnocena uživatelem. Nastavení analyzátorů Indiko Zadejte aplikační parametry pomocí čárového kódu, který naleznete ve specifikacích analyzátoru Indiko, nebo pomocí elektronického souboru. Dodávané materiály Reagencie uvedeny výše. Potřebné materiály, které nejsou součástí dodávky Kalibrátor a kontrolní vzorky uvedené dále. Kalibrace Reakce (dA/min) je přepočtena pomocí faktoru přepočtu na výsledné jednotky. Nebo použijte jako kalibrátor eCal (981830), a to v souladu s návodem uvedeným na příbalovém letáku eCal (981830). G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Řízení kvality Vzorky pro řízení kvality (QC) používejte alespoň jednou denně, po každé kalibraci a vždy, když použijete novou lahvičku činidla. Doporučuje se používat kontrolní materiály nebo směsné vzorky nejméně pro dvě hladiny (nízká a vysoká). Při řízení kvality vždy dodržujte platné místní předpisy. Dodávané kontrolní vzorky: CK-MB Control, kód 981908 CK-MB Control High, kód 981255 Nortrol, kód 981043 Abtrol, kód 981044 Intervaly a limitní hodnoty kontrolních vzorků musejí být upraveny podle požadavků jednotlivých laboratoří. Výsledky vzorků pro řízení kvality by měly vyhovovat mezním hodnotám přednastaveným laboratoří. INFORMACE O ČINIDLECH VÝPOČET VÝSLEDKŮ Výsledky jsou automaticky vypočteny analyzátorem, a to dále uvedeným způsobem: CK Kreatinfosfát + ADP → Kreatin + ATP HK ATP + Glukóza → ADP + G-6-P Obsah sady označené kódem 981828 2 lahvičky s 52 ml činidla A 2 lahvičky s 13 ml činidla B Obsah sady označené kódem 981829 2 lahvičky s 20 ml činidla A 2 lahvičky s 5 ml činidla B Koncentrace Činidlo A Pufr (imidazol), pH 6.5 D-glukóza N-acetyl-L-cystein (NAC) Octan hořečnatý EDTA-Na2 Nikotinamid adenin dinukleotid fosfát (NADP) Hexokináza (HK) NaN3 Činidlo B Pufr (imidazol) Adenosin-5’-difosfát (ADP) Adenosin-5’-monofosfát (AMP) Di(adenosin-5’)penta-fosfát (AP5A) Glukózo-6-fosfát dehydrogenáza (G-6-P-DH) EDTA-Na2 Kreatinfosfát NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1 % Aktivita v U/l = Faktor x dA/min TV x 1000 Teoretický faktor = 6.22 x SV x P TV SV 6.22 P Celkový reakční objem v ml Objem vzorku v ml Extinkční koeficient NADPH při 340 nm Optická tloušťka kyvety = 1 Faktor přepočtu µkat/l = (U/l)/60 Je-li použit jako kalibrátor eCal (981830), jsou výsledky vypočteny automaticky analyzátorem pomocí kalibrační křivky. Kalibrační křivka (příklad) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Zvláštní upozornění Činidlo B je nebezpečné. Viz zvláštní list uvnitř balení obsahující bezpečnostní opatření: H315, P280, P302+P352, P305+P351+P338. Činidla obsahují jako konzervační látku azid sodný. Používejte běžná bezpečnostní opatření vyžadovaná pro manipulaci se všemi laboratorními činidly. Výrobek musí být zlikvidován jako laboratorní chemikálie v souladu s místními předpisy. Příprava Činidla A a B jsou připravena k použití. Poznámka: Překontrolujte, zda při vkládání lahviček nebo nádob s činidly do analyzátoru nejsou v hrdle lahvičky nebo na povrchu činidla bubliny. Uchovávání a stabilita Činidla v neotevřených lahvičkách jsou stabilní při teplotě 2…8 °C do data ukončení použitelnosti vytištěného na štítku. Údaje o stabilitě činidel v analyzátoru („on-board“) naleznete v aplikačních poznámkách svého analyzátoru Konelab. Jeví-li činidla známky turbidity, nepoužívejte je. ODBĚR VZORKŮ Poznámka: Při zpracování vzorků v odběrových zkumavkách důsledně dodržujte pokyny výrobce zkumavek, aby se zabránilo chybným výsledkům. Zvláštní pozornost věnujte preanalytickým faktorům, jako je promíchání, doba stání před odstředěním a nastavení odstředění. Typ vzorků Lze použít nehemolyzované sérum nebo heparinizovanou plazmu. Poznámka: Hemolýza může vést k odchylce výsledků. Zabraňte expozici vzorků silnému světlu. Kalibrační křivka je závislá na konkrétní šarži. OMEZENÍ POSTUPU Pro diagnostické účely by výsledky měly být vždy hodnoceny v souvislosti s anamnézou pacienta, klinickými vyšetřeními a dalšími poznatky. Rušení Byla provedena studie podle pokynů uvedených v dokumentu CLSI EP7-A2. Kritérium: Rozdíl v rozmezí ± 10 % počátečních hodnot. Konelab Látka rušící stanovení Konjugovaný bilirubin Nekonjugovaný bilirubin Hemoglobin v hemolyzátu Lipémie: Intralipid (ochranná známka společnosti Fresenius Kabi AB) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l Cílová konc. CK (U/l) 188 498 180 495 184 505 Rozdíl vůči cílové konc. (%) - 5.3 - 5.8 −1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Koncentrace rušící látky Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 3 D06025_04_Insert_CK (IFCC)_MU Indiko Látka rušící stanovení 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l Cílová konc. CK (U/l) 179 477 175 469 187 502 Rozdíl vůči cílové konc. (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Koncentrace rušící látky Konjugovaný bilirubin Nekonjugovaný bilirubin Hemoglobin v hemolyzátu Lipémie: Intralipid (ochranná známka společnosti Fresenius Kabi AB) Nepoužívejte hemolyzované vzorky. Mezi turbiditou a koncentrací triglyceridů existuje nízká korelace. Ve velmi vzácných případech může gamopatie, zvláště typu IgM (Waldenströmova makroglobulinémie), způsobit nespolehlivé výsledky (6). Společnost Thermo Fisher Scientific Oy takové testy neprovedla. Údaje o dalších látkách rušících stanovení naleznete v odkazu č. 5. PŘEDPOKLÁDANÉ HODNOTY (3) Muži < 171 U/l (< 2.85 µkat/l) při 37 °C Ženy < 145 U/l (< 2.41 µkat/l) při 37 °C Uvedené hodnoty by měly sloužit pouze jako vodítko. Doporučuje se, aby každá laboratoř toto rozpětí ověřila nebo odvodila referenční interval pro populaci, jíž poskytuje služby. Faktor přepočtu µkat/l = (U/l)/60 Změny oproti předchozí verzi Svislá čára podél textu označuje změny. PRACOVNÍ CHARAKTERISTIKY Výsledky získané v jednotlivých laboratořích se mohou od uvedených údajů lišit. Mez detekce Byla provedena studie podle pokynů uvedených v dokumentu EP17-A. Konelab LoB: 0.6 U/l (0.01 µkat/l) Mez slepého vzorku (LoB) představuje nejvyšší naměřený výsledek, který bude pravděpodobně zjištěn pro vzorek bez analytu (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Mez detekce (LoD) představuje nejnižší množství analytu ve vzorku, které lze detekovat (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Mez stanovitelnosti (LoQ) představuje nejnižší skutečnou koncentraci ve vzorku, kterou lze kvantitativně stanovit (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Mez slepého vzorku (LoB) představuje nejvyšší naměřený výsledek, který bude pravděpodobně zjištěn pro vzorek bez analytu (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Mez detekce (LoD) představuje nejnižší množství analytu ve vzorku, které lze detekovat (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Mez stanovitelnosti (LoQ) představuje nejnižší skutečnou koncentraci ve vzorku, kterou lze kvantitativně stanovit (n = 42). Nepřesnost Nepřesnost stanovení CK je ≤ 5.0% celkového CV. Opakovatelnost (v rámci série) V rámci analyzátoru (celkem) Střední hodnota 92 U/l SD % CV 1.7 1.8 Střední hodnota 226 U/l SD % CV 2.0 0.9 Střední hodnota 616 U/l SD % CV 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Byla provedena studie přesnosti podle pokynů v dokumentu CLSI EP5-A2 na jednom pracovišti a třech analyzátorech Konelab 60, s jedním laborantem, dvěma šaržemi reagencií, jednou šarží kalibrátorů a 6 kalibracemi trvající 21 dní, v níž byl počet měření n = 84. Výsledky z každého analyzátoru byly zkombinovány. Indiko CK Opakovatelnost (v rámci série) V rámci analyzátoru (celkem) Střední hodnota 95 U/l SD % CV 1.2 1.2 Střední hodnota 226 U/l SD % CV 2.0 0.9 Střední hodnota 637 U/l SD % CV 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Byla provedena studie přesnosti podle pokynů v dokumentu CLSI EP5-A2 na jednom pracovišti a třech analyzátorech Indiko, s jedním laborantem, dvěma šaržemi činidel, jednou šarží kalibrátorů a 7 kalibracemi trvající 21 dní, v níž byl počet měření n = 84. Výsledky z každého analyzátoru byly zkombinovány. Srovnání metod Konelab Byla provedena srovnávací studie na analyzátoru Konelab Prime 60 podle pokynů dokumentu CLSI EP9-A2. Jako referenční metoda byla použita předchozí metoda 981372 CK(IFCC). Lineární regrese (dle Deminga) (jednotky výsledků U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Koncentrace vzorků se pohybovaly v rozmezí 17 až 2253 U/l. Indiko 1) Byla provedena srovnávací studie na analyzátoru Indiko podle pokynů dokumentu CLSI EP9-A2. Jako referenční metoda byla použita předchozí metoda 981372 CK(IFCC) na analyzátoru Konelab Prime 60 (x). Lineární regrese (dle Deminga) (jednotky výsledků U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Koncentrace vzorků se pohybovaly v rozmezí 17 až 2253 U/l. 2) Byla provedena srovnávací studie na analyzátoru Indiko podle pokynů dokumentu CLSI EP9-A2. Jako referenční metoda byla použita stejná metoda CK(IFCC) na analyzátoru Konelab Prime 60 (x). Lineární regrese (dle Deminga) (jednotky výsledků U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Koncentrace vzorků se pohybovaly v rozmezí 14 až 2450 U/l. VÝROBCE Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel.: +358 10 329 200 www.thermoscientific.com Datum revize (rrrr-mm-dd) 2015-05-08 ROZSAH MĚŘENÍ 10 – 1000 U/l (0.17 – 17 µkat/l). Rozšířený rozsah měření po sekundárním ředění: 10 – 9000 U/l (0.17 – 150 µkat/l). Konelab CK SEZNAM LITERATURY 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5.vydání, W B Saunders Company, Philadelphia, 2001, str. 357 - 358. 2) Thomas, L. Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1. vydání, TH-Books Verlagsgesellschaft mbH, Frankfurt n. Mohanem, Německo, 1998, p. 73. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 4 Page 5 D06025_04_Insert_CK (IFCC)_MU DE CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) DIE VORLIEGENDE FACHINFORMATION IST FÜR DEN GEBRAUCH AUSSERHALB DER USA VORGESEHEN. ANWENDUNGSBEREICH Zur In-vitro--Diagnose für den quantitativen Nachweis von Kreatinkinase (ATP: Kreatin-NPhosphotransferase (CK), EC 2.7.3.2) in Humanserum oder -plasma mit Konelab- oder IndikoAnalysegeräten. Jeder Verweis auf Konelab-Systeme betrifft auch die T-Serie. ZUSAMMENFASSUNG (1.2) Die Kreatinkinaseaktivität ist in der quergestreiften Muskulatur, im Gehirn und im Herzgewebe höher. In anderen Geweben, wie beispielsweise Nieren, Zwerchfell und Prostata, liegt eine signifikant verminderte Aktivität vor. Die CK-Aktivität ist bei vielen Erkrankungen erhöht, einschließlich solcher, welche die Skelettmuskulatur, das Herz, das zentrale Nervensystem und die Schilddrüse betreffen. Nach einem akuten Muskelschaden steigt die totale CK-Aktivität im Blut an und erreicht nach einem durchschnittlichen Zeitraum von 4–5 Stunden bei 50 % der Patienten eine pathologische Konzentration. Die Werte erhöhen sich in der Regel zwischen der 8. und 24. Stunde nach Infarktbeginn und kehren mit großen individuellen Schwankungen in den Referenzbereich zurück. TESTPRINZIP Dieses CK-MB-Assay ist ein Flüssig-Assay aus zwei Teilen, das drei Enzyme zur Bestimmung der CK-Aktivität verwendet. Die Kreatinkinase katalysiert die Reaktion zwischen Kreatinphosphat und ADP (Adenosin-5’diphosphat), bei der Kreatin und ATP (Adenosin-5’-triphosphat) gebildet wird. Die letztgenannte Verbindung phosphoryliert Glucose in Gegenwart von HK (Hexokinase) zu G-6-P (Glucose-6Phosphat). G-6-P wird zu 6-PG (Gluconat-6-Phosphat) oxidiert und gleichzeitig wird eine äquimolare Menge NADP (Nicotinamid-adenin-dinucleotid-phosphat) zu NADPH in einer Reaktion reduziert, die durch G-6P-DH (Glucose-6-Phosphat-Dehydrogenase) katalysiert wird. Die aus der Bildung von NADPH resultierende Zunahme der Extinktion bei 340 nm ist proportional zur Kreatinkinaseaktivität in der Probe (3). CK Kreatinphosphat + ADP → Kreatin + ATP HK ATP + Glucose → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Probe 4 Stunden bei 20...25 °C, 1 Monat bei 4...8 °C oder 1 Monat bei -20 °C; vor Licht geschützt. CK-BB instabil. TESTVERFAHREN Angaben zur Automatisierung mit dem Analysengerät den Applikationshinweisen entnehmen. Bei Verwendung von Applikationen, die nicht durch Thermo Fisher Scientific Oy validiert wurden, kann keine Garantie für die angegebenen Leistungsdaten übernommen werden. Für die Validierung derartiger Applikationen ist der Anwender daher selbst verantwortlich. Indiko-Geräteeinstellungen Die Applikationsparameter mit Hilfe des Barcodes auf dem Indiko-Applikationsblatt oder aus einer elektronischen Datei einlesen. Packungsinhalt Reagenzien wie oben beschrieben. Erforderliche, jedoch nicht im Lieferumfang enthaltene Materialien Kalibrator und Kontrollen wie nachstehend angegeben. Kalibrierungs Das Messergebnis (dA/min) wird durch einen Umrechnungsfaktor in Ergebniseinheiten umgerechnet. Oder eCal (981830) als Kalibrator gemäß den Anweisungen in der Packungsbeilage von eCal (981830) verwenden. Qualitätskontrolle Mindestens einmal täglich sowie nach jeder Kalibrierung und bei jeder Verwendung eines neuen Reagenzfläschchens eine Qualitätskontrolle (QK) durchführen. Es wird empfohlen, mindestens zwei Konzentrationen (niedrig und hoch) von Kontrollen oder Probenmischungen zu verwenden. Bei der Qualitätskontrolle sind stets alle örtlichen, bundesstaatlichen und landesweiten Vorschriften zu berücksichtigen. Erhältliche Kontrollen: CK-MB Control, Bestellnr. 981908 CK-MB Control High, Bestellnr. 981255 Nortrol, Bestellnr. 981043 Abtrol, Bestellnr. 981044 Die Intervalle und Grenzwerte der Kontrolle (QC) müssen an die Anforderungen der einzelnen Laboratorien angepasst werden. Die Ergebnisse der Qualitätskontrollen müssen innerhalb der vom Labor vorgegebenen Grenzwerte liegen. BERECHNUNG DER ERGEBNISSE Die Ergebnisse werden vom Analysegerät automatisch wie folgt berechnet: REAGENZDATEN Aktivität in U/l = Faktor x dA/min Kit Bestellnr. 981828 enthält 2 Fläschchen mit 52 ml Reagenz A 2 Fläschchen mit 13 ml Reagenz B Kit Bestellnr. 981829 Bestellnr 2 Fläschchen mit 20 ml Reagenz A 2 Fläschchen mit 5 ml Reagenz B Konzentrationen Reagenz A Buffer (Imidazol), pH 6,5 D-Glucose N-Acetyl-L-Cystein (NAC) Magnesiumacetat EDTA-Na2 Nicotinamid-adenin-dinucleotid-phosphat (NADP) Hexokinase (HK) NaN3 Reagenz B Puffer (Imidazol) Adenosin-5’-diphosphat (ADP) Adenosin-5’-monophosphat (AMP) Di(adenosin-5’)-pentaphosphat (AP5A) Glucose-6-Phosphat-Dehydrogenase (G-6-P-DH) EDTA -Na2 Kreatinphosphat NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2,7 mmol/l ≥ 5000 U/l < 0.1 % 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % TV x 1000 Theoretischer Faktor = 6.22 x SV x P TV SV 6,22 P Gesamtes Reaktionsvolumen in ml Probenvolumen in ml = Extinktionskoeffizient von NADPH bei 340 nm Schichtdicke der Küvette = 1 Umrechnungsfaktor µkat/l = (U/l)/60 Wenn eCal (981830) als Kalibrator verwendet wird, werden die Ergebnisse vom Analysegerät automatisch mit einer Bezugskurve berechnet. Bezugskurve (Beispiel) Vorsichtsmaßnahmen Reagenz B ist gefährlich. Informationen zu den Gefahren und Vorsichtsmaßnahmen finden Sie im Kit auf einem gesonderten Blatt: H315, P280, P302+P352, P305+P351+P338. Die Reagenzien enthalten Natriumazid als Konservierungsmittel. Es sind die üblichen Vorsichtsmaßnahmen für den Umgang mit Laborreagenzien zu befolgen. Das Produkt muss gemäß den lokalen Regulierungsvorschriften als Laborchemikalie entsorgt werden. Vorbereitungen Reagenz A und B sind gebrauchsfertig. Hinweis: Kontrollieren Sie, dass sich im Flaschenhals oder an der Reagenzoberfläche keine Luftblasen befinden, wenn Sie die Reagenzfläschchen bzw. -gefäße einsetzen. Lagerung und Stabilität In ungeöffneten Fläschchen sind die Reagenzien bei 2...8 °C bis zum Ablauf des Verfallsdatums auf dem Etikett haltbar. Informationen zur Haltbarkeit der Reagenzien nach Einlegen in das Analysegerät finden Sie in den entsprechenden Applikationshinweisen. Verwenden Sie die Reagenzien nicht, wenn sie eine Trübung aufweisen. PROBENGEWINNUNG Hinweis: Befolgen Sie bei der Verarbeitung von Proben in Probenröhrchen genau die Anweisungen des Röhrchenherstellers, um fehlerhafte Ergebnisse zu vermeiden. Achten Sie besonders auf die präanalytischen Variablen wie das Mischen, die Standzeit vor der Zentrifugation und die Zentrifugeneinstellungen. Probenart Es kann nicht hämolysiertes Serum oder Heparinplasma verwendet werden. Hinweis: Hämolyse kann zu abweichenden Ergebnissen führen. Die Proben dürfen keiner starken Lichteinwirkung ausgesetzt werden. Vorsichtsmaßnahmen Proben humanen Ursprungs sind als potenziell infektiös zu betrachten und dementsprechend zu behandeln und zu entsorgen. Die Bezugskurve ist chargenabhängig. GRENZEN DES VERFAHRENS Zu diagnostischen Zwecken sollten die Ergebnisse immer in Verbindung mit der Anamnese, den klinischen Untersuchungen und anderen Befunden des Patienten bewertet werden. Interferenz Es wurde eine Studie durchgeführt, bei der das CLSI-Dokument EP7-A2 als Richtlinie diente. Kriterium: Differenz innerhalb von ±10 % der Ausgangswerte. Konelab Störsubstanz Konjugiertes Bilirubin Nicht-konjugiertes Bilirubin Hämoglobin im Hämolysat Lipämie: Intralipid (Warenzeichen der Fresenius Kabi AB) Konzentration der Störsubstanz Zielwert für CK (U/l) 592 µmol/l (34,6 mg/dl) 601 µmol/l (35,2 mg/dl) 621 µmol/l (36,3 mg/dl) 677 µmol/l (39,6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Abweichung vom Zielwert (%) -5.3 -5.8 -1.7 -2.9 +8.5 +4.6 10 g/l 10 g/l 181 504 +2.8 +7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Lagerung (4) Die Proben können wie nachfolgend dargestellt gelagert werden. D06025_04_Insert_CK (IFCC)_MU Indiko Störsubstanz Konjugiertes Bilirubin Nicht-konjugiertes Bilirubin Hämoglobin im Hämolysat Lipämie: Intralipid (Warenzeichen der Fresenius Kabi AB) Konzentration der Störsubstanz Zielwert für CK (U/l) 592 µmol/l (34,6 mg/dl) 601 µmol/l (35,2 mg/dl) 621 µmol/l (36,3 mg/dl) 676 µmol/l (39,5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Abweichung vom Zielwert (%) -0,8 -2,4 +0,2 -0,1 +8.6 +3.4 10 g/l 10 g/l 185 505 +1.0 -0.8 Kein hämolysiertes Untersuchungsmaterial verwenden. Es besteht eine schwache Korrelation zwischen der Trübung und der Konzentration der Triglyceride. In sehr seltenen Fällen kann es infolge einer Gammopathie, insbesondere vom IgM-Typ (Waldenströmsche Makroglobulinämie), zu unzuverlässigen Ergebnissen kommen (6). Von Thermo Fisher Scientific Oy nicht bestätigte Daten. Siehe Literaturhinweis (5) für weitere Substanzen, die Störfaktoren darstellen können. ERWARTUNGSWERTE (3) Männlich < 171 U/l (< 2.85 µkat/l) bei 37 °C Weiblich < 145 U/l (< 2.41 µkat/l) bei 37 °C Die angegebenen Werte gelten nur als Richtlinie. Es wird empfohlen, dass jedes Labor diesen Bereich überprüft oder ein Referenzintervall für die betroffene Population ableitet. Umrechnungsfaktor µkat/l = (U/l)/60 Änderungen gegenüber der vorherigen Fassung Die Linie neben dem Text weist auf Änderungen hin. LEISTUNGSDATEN Die Ergebnisse einzelner Laboratorien können von den angegebenen Leistungsdaten abweichen. Nachweisgrenze Es wurde eine Studie durchgeführt, bei der das Dokument EP17-A als Richtlinie diente. Konelab LoB: 0.6 U/l (0.01 µkat/l) Die Blankgrenze (LoB, Limit of Blank) stellt das höchste Messergebnis dar, das für eine Probe ohne Analyt erhalten wird (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Die Nachweisgrenze (LoD, Limit of Detection) stellt die geringste Menge an Analyten in einer Probe dar, die noch nachgewiesen werden können (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Die Bestimmungsgrenze (LoQ, Limit of Quantitation) stellt die niedrigste Konzentration in einer Probe dar, die noch quantitativ bestimmt werden kann (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Die Blankgrenze (LoB, Limit of Blank) stellt das höchste Messergebnis dar, das für eine Probe ohne Analyt erhalten wird (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Die Nachweisgrenze (LoD, Limit of Detection) stellt die geringste Menge an Analyten in einer Probe dar, die noch nachgewiesen werden kann (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Die Bestimmungsgrenze (LoQ, Limit of Quantitation) stellt die niedrigste Konzentration in einer Probe dar, die noch quantitativ bestimmt werden kann (n = 42). Präzision Die Impräzision des ALT-Tests ist ≤ 5.0 % des Gesamt-VK. Wiederholgenauigkeit (in der Serie) Intra-Gerät (Gesamt) Mittelwert 92 U/l SD CV% 1.7 1.8 Mittelwert 226 U/l SD CV% 2.0 0.9 Mittelwert 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Unter Verwendung des CLSI-Dokuments EP5-A2 als Richtlinie wurde an einem Standort, mit drei Konelab 60-Analysengeräten, einem Bediener, 2 Reagenzienchargen, einer Kalibriercharge und 6 Kalibrierungen 21 Tage lang eine Präzisionsstudie durchgeführt, wobei die Anzahl der Messungen (n) 84 betrug. Die Ergebnisse von jedem Analysegerät wurden kombiniert. Indiko CK Wiederholgenauigkeit (in der Serie) Intra-Gerät (Gesamt) HERSTELLER Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finnland Tel.: +358 10 329 200 www.thermoscientific.com Datum der Überarbeitung (JJJJ-MM-TT) 2015-05-08 MESSBEREICH 10 – 1000 U/l (0.17 – 17 µkat/l). Erwarteter Messbereich nach Sekundärverdünnung: 10 – 9000 U/l (0.17 – 150 µkat/l). Konelab CK LITERATURVERWEISE 1) Burtis, CA und Ashwood, ER (Hrsg.), Tietz Fundamentals of Clinical Chemistry, 5. Auflage, WB Saunders Company, Philadelphia, 2001, S. 357–358. 2) Thomas, L. (Hrsg.) Clinical Laboratory Diagnostics: Use and Assessment of Clinical Laboratory results, 1. Auflage, TH-Books Verlagsgesellschaft, Frankfurt/Main, Deutschland, 1998, S. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Teil. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Aspartate Aminotransferase, Clin.Chem.Lab.Med. 2002, 40(6): 635–642. 4) WHO/DIL/LAB/99.1 Ausg. 2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Genf, 15. Januar 2002, S. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fünfte Auflage, AACC Press, Washington, D.C., 3-234–3-239, 2000. 6) Bakker A.J. et al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 –1243. Mittelwert 95 U/l SD CV% 1.2 1.2 Mittelwert 226 U/l SD CV% 2.0 0.9 Mittelwert 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Unter Verwendung des CLSI-Dokuments EP5-A2 als wurde an einem Standort auf drei IndikoAnalysegeräten, mit einem Bediener, 2 Reagenzienchargen, einer Kalibriercharge und 7 Kalibrierungen 21 Tage lang eine Präzisionsstudie durchgeführt, wobei die Anzahl der Messungen (n) 84 betrug. Die Ergebnisse von jedem Analysegerät wurden kombiniert. Vergleich der Methoden Konelab Auf einem Konelab Prime 60-Analysegerät wurde unter Verwendung des CLSI-Dokuments EP9-A2 als Richtlinie mit der vorherigen 981372 CK(IFCC)-Methode (x) als Referenz eine Vergleichsuntersuchung durchgeführt. Lineare Regression (Deming) (Ergebnisse in U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Die in den Proben enthaltenen Konzentrationen lagen zwischen 17 und 2253 U/l. Indiko 1) Es wurde eine Vergleichsuntersuchung auf einem Indiko-Analysegerät unter Verwendung des CLSI-Dokuments EP9-A2 als Richtlinie mit der vorherigen 981372 CK(IFCC)-Methode (x) auf einem Konelab Prime 60-Analysegerät als Referenz durchgeführt. Lineare Regression (Deming) (Ergebnisse in U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Die in den Proben enthaltenen Konzentrationen lagen zwischen 17 und 2253 U/l. 2) Es wurde eine Vergleichsuntersuchung auf einem Indiko-Analysegerät unter Verwendung des CLSI-Dokuments EP9-A2 als Richtlinie und mit der gleichen CK-MBCK(IFCC)-Methode auf einem Konelab Prime 60-Analysegerät als Referenz durchgeführt. Lineare Regression (Deming) (Ergebnisse in U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Die in den Proben enthaltenen Konzentrationen lagen zwischen 14 und 2450 U/l. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 6 D06025_04_Insert_CK (IFCC)_MU EL CK (IFCC) 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) Προφυλάξεις Τα ανθρώπινα δείγματα θα πρέπει να χρησιμοποιούνται και να απορρίπτονται ως δυνητικώς μολυσματικά. Φύλαξη (4) Το δείγμα μπορεί να φυλαχθεί όπως περιγράφεται παρακάτω. Δείγμα 4 ώρες στους 20...25 °C, 1 μήνα στους 4...8 °C ή 1 μήνα στους -20 °C, προστατευμένο από το φως. Το ισοένζυμο CK-BB δεν είναι σταθερό. ΤΟ ΠΑΡΟΝ ΕΝΘΕΤΟ ΣΥΣΚΕΥΑΣΙΑΣ ΙΣΧΥΕΙ ΓΙΑ ΧΡΗΣΗ ΕΚΤΟΣ ΤΩΝ Η.Π.Α. ΠΡΟΒΛΕΠΟΜΕΝΗ ΧΡΗΣΗ Για in vitro διαγνωστική χρήση στον ποσοτικό προσδιορισμό της ενεργότητας της κινάσης της κρεατίνης (ATP: N-φωσφοτρανσφεράση της κρεατίνης (CK), EC 2.7.3.2) σε ανθρώπινο ορό ή πλάσμα με αναλυτές Konelab και Indiko. Οποιαδήποτε αναφορά στα συστήματα Konelab ισχύει και για τη Σειρά T. ΠΕΡΙΛΗΨΗ (1, 2) Η ενεργότητα της κινάσης της κρεατίνης είναι υψηλότερη στο γραμμωτό μυϊκό, τον εγκεφαλικό και τον καρδιακό ιστό. Σε άλλους ιστούς όπως οι νεφροί, το διάφραγμα και ο προστάτης, απαντάται σημαντικά χαμηλότερη ενεργότητα. Η ενεργότητα της CK είναι αυξημένη σε πολλές παθήσεις, συμπεριλαμβανομένων εκείνων που αφορούν το σκελετικό μυ, την καρδιά, το κεντρικό νευρικό σύστημα και το θυρεοειδή. Έπειτα από οξεία μυϊκή βλάβη, η ενεργότητα της ολικής CK στο αίμα αρχίζει να αυξάνεται, αγγίζοντας παθολογικά επίπεδα μετά από μια μέση διάρκεια 4 - 5 ωρών, στο 50% των ασθενών. Οι τιμές αυξάνονται με ομαλό ρυθμό την 8η - 24η ώρα μετά την εμφάνιση εμφράγματος και επανέρχονται στο εύρος αναφοράς με μεγάλες διακυμάνσεις από άτομο σε άτομο. ΑΡΧΗ ΤΗΣ ΔΙΑΔΙΚΑΣΙΑΣ Ο προσδιορισμός CK είναι μια ανάλυση υγρών δειγμάτων σε δύο μέρη, που χρησιμοποιεί τρία ένζυμα για τον προσδιορισμό της ενεργότητας της CK. Η κινάση της κρεατίνης καταλύει την αντίδραση ανάμεσα στη φωσφορική κρεατίνη και την ADP (5'-διφωσφορική αδενοσίνη) για το σχηματισμό κρεατίνης και ATP (5'-τριφωσφορική αδενοσίνη). Η τελευταία ένωση φωσφορυλιώνει τη γλυκόζη σε G-6-P (6-φωσφορική γλυκόζη) παρουσία του ενζύμου HK (εξοκινάση). Η G-6-P οξειδώνεται σε 6-PG (6-φωσφορικό γλυκονικό οξύ) και, ταυτόχρονα, μια ισομοριακή ποσότητα NADP (φωσφορικό νικοτιναμιδο-αδενινο-δινουκλεοτίδιο) ανάγεται σε NADPH σε μια αντίδραση που καταλύεται από το ένζυμο G-6P-DH (αφυδρογονάση της 6-φωσφορικής γλυκόζης). Η αύξηση της απορρόφησης στα 340 nm, ως αποτέλεσμα του σχηματισμού NADPH, είναι ανάλογη προς την ενεργότητα της κινάσης της κρεατίνης στο δείγμα (3). ΔΙΑΔΙΚΑΣΙΑ ΕΞΕΤΑΣΗΣ Για αυτοματοποιημένες διαδικασίες στον αναλυτή σας, ανατρέξτε στις Σημειώσεις εφαρμογής. Η απόδοση οποιασδήποτε εφαρμογής που δεν έχει επικυρωθεί από τη Thermo Fisher Scientific Oy δεν μπορεί να είναι εγγυημένη και επομένως πρέπει να αξιολογείται από το χρήστη. Ρυθμίσεις οργάνου Indiko Εισαγάγετε τις παραμέτρους εφαρμογής μέσω του γραμμωτού κώδικα που θα βρείτε στο φύλλο εφαρμογής Indiko, ή μέσω ηλεκτρονικού αρχείου, όπως κρίνεται κατάλληλο. Παρεχόμενα υλικά Αντιδραστήρια, όπως αυτά περιγράφονται παραπάνω. Υλικά που απαιτούνται αλλά δεν παρέχονται Βαθμονομητής και υλικά ελέγχου, όπως περιγράφεται παρακάτω. Βαθμονόμηση Η απόκριση (dA/min) μετατρέπεται σε μονάδες αποτελέσματος με ένα συντελεστή υπολογισμού. Εναλλακτικά, χρησιμοποιήστε ως βαθμονομητή μια μονάδα eCal (981830), ακολουθώντας τις οδηγίες που αναγράφονται στο ένθετο συσκευασίας της μονάδας eCal (981830). HK ATP + Γλυκόζη → ADP + G-6-P Ποιοτικός έλεγχος Χρησιμοποιείτε δείγματα ποιοτικού ελέγχου τουλάχιστον μία φορά την ημέρα και μετά από κάθε βαθμονόμηση, καθώς και κάθε φορά που χρησιμοποιείτε νέα φιάλη αντιδραστηρίου. Συνιστάται να χρησιμοποιείτε υλικά ελέγχου ή συνδυασμό δειγμάτων τουλάχιστον δύο επιπέδων (χαμηλής και υψηλής συγκέντρωσης). Κατά τη διεξαγωγή ποιοτικού ελέγχου, τηρείτε πάντοτε τους τοπικούς, περιφερειακούς και εθνικούς κανονισμούς. Διαθέσιμα υλικά ελέγχου: CK-MB Control, κωδικός 981908 CK-MB Control High, κωδικός 981255 Nortrol, κωδικός 981043 Abtrol, κωδικός 981044 Τα διαστήματα και τα όρια των ελέγχων (QC) πρέπει να προσαρμόζονται στις απαιτήσεις του εκάστοτε εργαστηρίου. Τα αποτελέσματα των δειγμάτων ποιοτικού ελέγχου θα πρέπει να εμπίπτουν στα όρια που έχει προκαθορίσει το εργαστήριο. G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ ΥΠΟΛΟΓΙΣΜΟΣ ΤΩΝ ΑΠΟΤΕΛΕΣΜΑΤΩΝ Τα αποτελέσματα υπολογίζονται αυτομάτως από τον αναλυτή ως εξής: ΠΛΗΡΟΦΟΡΙΕΣ ΑΝΤΙΔΡΑΣΤΗΡΙΩΝ Eνεργότητα σε U/l = Συντελεστής x dA/min CK Φωσφορική κρεατίνη + ADP → Κρεατίνη + ATP Το κιτ με κωδικό 981828 περιέχει 2 φιαλίδια Αντιδραστηρίου A των 52 ml 2 φιαλίδια Αντιδραστηρίου B των 13 ml Το κιτ με κωδικό 981829 περιέχει 2 φιαλίδια Αντιδραστηρίου A των 20 ml 2 φιαλίδια Αντιδραστηρίου B των 5 ml Συγκεντρώσεις Αντιδραστήριο Α Ρυθμιστικό διάλυμα (Ιμιδαζόλη), pH 6.5 D-γλυκόζη N-ακετυλ-L-κυστεΐνη (NAC) Οξικό μαγνήσιο EDTA-Na2 Φωσφορικό νικοτιναμιδο-αδενινο-δινουκλεοτίδιο (NADP) Εξοκινάση (HK) NaN3 Αντιδραστήριο Β Ρυθμιστικό διάλυμα (Ιμιδαζόλη) 5'-διφωσφορική αδενοσίνη (ADP) 5'-μονοφωσφορική αδενοσίνη (AMP) Πενταφωσφορική δι(αδενοσίνη-5') (AP5A) Αφυδρογονάση της 6-φωσφορικής γλυκόζης (G-6-P-DH) EDTA -Na2 Φωσφορική κρεατίνη NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1 % TV x 1000 Θεωρητικός συντελεστής = 6.22 x SV x P TV SV 6.22 P Ολικός όγκος αντίδρασης σε ml Όγκος δείγματος σε ml Συντελεστής εξασθένησης NADPH στα 340 nm Μήκος διαδρομής κυψελίδας = 1 Συντελεστής μετατροπής µkat/l = (U/l)/60 Αν χρησιμοποιείται ως βαθμονομητής μια μονάδα eCal (981830), τα αποτελέσματα υπολογίζονται αυτόματα από τον αναλυτή με χρήση καμπύλης βαθμονόμησης. Καμπύλη βαθμονόμησης (παράδειγμα) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Προφυλάξεις Το αντιδραστήριο Β είναι επικίνδυνο. Δείτε το ξεχωριστό φυλλάδιο για τις φράσεις κινδύνου και προφυλάξεων που εσωκλείεται στο κιτ: H315, P280, P302+P352, P305+P351+P338. Τα αντιδραστήρια περιέχουν αζίδιο του νατρίου ως συντηρητικό. Λαμβάνετε τις συνήθεις προφυλάξεις που απαιτούνται για το χειρισμό όλων των εργαστηριακών αντιδραστηρίων. Το προϊόν πρέπει να απορρίπτεται ως εργαστηριακή χημική ουσία σύμφωνα με τους τοπικούς κανονισμούς. Προετοιμασία Τα Αντιδραστήρια Α και Β είναι έτοιμα για χρήση. Σημείωση: Ελέγξτε ότι δεν υπάρχουν φυσαλίδες στο λαιμό του φιαλιδίου ή στην επιφάνεια του αντιδραστηρίου όταν εισάγετε τα φιαλίδια ή τα δοχεία αντιδραστηρίων στον αναλυτή. Φύλαξη και σταθερότητα Τα αντιδραστήρια σε φιαλίδια που δεν έχουν ανοιχτεί είναι σταθερά στους 2...8°C μέχρι την ημερομηνία λήξης που αναγράφεται στην ετικέτα. Για πληροφορίες σχετικά με τη σταθερότητα των αντιδραστηρίων όταν αυτά είναι τοποθετημένα στο μηχάνημα, ανατρέξτε στις Σημειώσεις εφαρμογής του αναλυτή. Εάν τα αντιδραστήρια παρουσιάζουν θολερότητα, μην τα χρησιμοποιήσετε. ΣΥΛΛΟΓΗ ΔΕΙΓΜΑΤΩΝ Σημείωση: Κατά την επεξεργασία δειγμάτων σε σωληνάρια δειγματοληψίας, ακολουθήστε προσεκτικά τις οδηγίες του κατασκευαστή των σωληναρίων, για να αποφύγετε τη λήψη εσφαλμένων αποτελεσμάτων. Να είστε ιδιαίτερα προσεκτικοί με τις προαναλυτικές μεταβλητές, όπως είναι η ανάμιξη, ο χρόνος ηρεμίας πριν από τη φυγοκέντρηση και οι ρυθμίσεις της φυγοκέντρου. Τύπος δείγματος Μπορεί να χρησιμοποιηθεί μη αιμολυμένος ορός ή ηπαρινισμένο πλάσμα. Σημείωση: Η αιμόλυση μπορεί να οδηγήσει σε αποκλίνοντα αποτελέσματα. Αποφεύγετε την έκθεση των δειγμάτων σε δυνατό φως. Η καμπύλη βαθμονόμησης εξαρτάται από την παρτίδα. ΠΕΡΙΟΡΙΣΜΟΙ ΤΗΣ ΔΙΑΔΙΚΑΣΙΑΣ Για διαγνωστικούς σκοπούς, τα αποτελέσματα θα πρέπει να εκτιμώνται πάντοτε σε συνδυασμό με το ιατρικό ιστορικό του ασθενούς, τα ευρήματα κλινικής εξέτασης και άλλα ευρήματα. Παρεμβολές Διεξήχθη μελέτη με οδηγό το έγγραφο ΕΡ7-Α2 του CLSI. Κριτήριο: Διαφορά εντός ± 10% των αρχικών τιμών. Konelab Ουσία παρεμβολής Συζευγμένη χολερυθρίνη Μη συζευγμένη χολερυθρίνη Αιμοσφαιρίνη σε αιμόλυμα Λιπαιμία: Intralipid (εμπορικό σήμα της Fresenius Kabi AB) Συγκέντρωση ουσίας παρεμβολής Στόχος CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Διαφορά από στόχο (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 7 D06025_04_Insert_CK (IFCC)_MU Indiko Ουσία παρεμβολής Συζευγμένη χολερυθρίνη Μη συζευγμένη χολερυθρίνη Αιμοσφαιρίνη σε αιμόλυμα Λιπαιμία: Intralipid (εμπορικό σήμα της Fresenius Kabi AB) Συγκέντρωση ουσίας παρεμβολής Στόχος CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Διαφορά από στόχο (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Αποφεύγετε τη χρήση αιμολυμένων δειγμάτων. Υπάρχει μικρή συσχέτιση ανάμεσα στη θολερότητα και τη συγκέντρωση τριγλυκεριδίων. Σε πολύ σπάνιες περιπτώσεις γαμμαπάθειας, και συγκεκριμένα τύπου IgM (μακροσφαιριναιμία Waldenström), μπορεί να προκληθούν αναξιόπιστα αποτελέσματα (6). Δεν έχει δοκιμαστεί στη Thermo Fisher Scientific Oy. Για άλλες ουσίες παρεμβολής, ανατρέξτε στην παραπομπή (5). ΒΙΒΛΙΟΓΡΑΦΙΑ 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. ΑΝΑΜΕΝΟΜΕΝΕΣ ΤΙΜΕΣ (3) Άνδρες < 171 U/l (< 2.85 µkat/l) στους 37 °C Γυναίκες < 145 U/l (< 2.41 µkat/l) στους 37 °C Οι προαναφερθείσες τιμές θα πρέπει να λαμβάνονται υπόψη μόνον ενδεικτικά. Συνιστάται κάθε εργαστήριο να επαληθεύει αυτό το εύρος τιμών ή να ορίζει ένα διάστημα αναφοράς για τον πληθυσμό με τον οποίο ασχολείται. ΚΑΤΑΣΚΕΥΑΣΤΉΣ Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Τηλ. +358 10 329 200 www.thermoscientific.com Συντελεστής μετατροπής µkat/l = (U/l)/60 Ημερομηνία αναθεώρησης (εεεε-μμ-ηη) 2015-05-08 ΕΥΡΟΣ ΜΕΤΡΗΣΗΣ 10 – 1000 U/l (0.17 – 17 µkat/l). Εκτεταμένο εύρος μέτρησης μετά από δευτερογενή αραίωση: 10 – 9000 U/l (0.17 – 150 µkat/l). Αλλαγές από την προηγούμενη έκδοση Η γραμμή δίπλα στο κείμενο υποδεικνύει αλλαγές. ΧΑΡΑΚΤΗΡΙΣΤΙΚΑ ΑΠΟΔΟΣΗΣ Τα αποτελέσματα που λαμβάνονται σε μεμονωμένα εργαστήρια ενδέχεται να διαφέρουν από τα παρεχόμενα δεδομένα απόδοσης. Όριο ανίχνευσης Διεξήχθη μελέτη με οδηγό το έγγραφο ΕΡ17-Α. Konelab LoB: 0.6 U/l (0.01 µkat/l) Το όριο τυφλού (LoB) αντιστοιχεί στο υψηλότερο αποτέλεσμα μέτρησης που είναι πιθανό να παρατηρηθεί για δείγμα χωρίς προσδιοριζόμενη ουσία (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Το όριο ανίχνευσης (LoD) αντιστοιχεί στη χαμηλότερη ποσότητα προσδιοριζόμενης ουσίας στο δείγμα, η οποία μπορεί να ανιχνευτεί (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Το όριο ποσοτικοποίησης (LoQ) αντιστοιχεί στη χαμηλότερη πραγματική συγκέντρωση στο δείγμα, η οποία μπορεί να ποσοτικοποιηθεί (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Το όριο τυφλού (LoB) αντιστοιχεί στο υψηλότερο αποτέλεσμα μέτρησης που είναι πιθανό να παρατηρηθεί για δείγμα χωρίς προσδιοριζόμενη ουσία (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Το όριο ανίχνευσης (LoD) αντιστοιχεί στη χαμηλότερη ποσότητα προσδιοριζόμενης ουσίας στο δείγμα, η οποία μπορεί να ανιχνευτεί (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Το όριο ποσοτικοποίησης (LoQ) αντιστοιχεί στη χαμηλότερη πραγματική συγκέντρωση στο δείγμα, η οποία μπορεί να ποσοτικοποιηθεί (n = 42). Ανακρίβεια Η ανακρίβεια του προσδιορισμού CK είναι ≤ 5.0% του συνολικού CV. Konelab CK Επαναληψιμότητα (Σε μία ανάλυση) Επί της συσκευής (Σύνολο) Μέση τιμή 92 U/l SD 1.7 CV% 1.8 Μέση τιμή 226 U/l SD CV% 2.0 0.9 Μέση τιμή 616 U/l SD CV% 7.1 1.2 2.5 2.8 3.1 22.3 1.4 3.6 Διεξήχθη μελέτη ακριβείας με οδηγό το έγγραφο EP5-A2 του CLSI, σε ένα κέντρο με τρεις αναλυτές Konelab 60, ένα χειριστή, δύο παρτίδες αντιδραστηρίων, μία παρτίδα βαθμονομητή και έξι βαθμονομήσεις σε 21 ημέρες, με αριθμό μετρήσεων n = 84. Τα αποτελέσματα από κάθε αναλυτή συνδυάστηκαν. Indiko CK Επαναληψιμότητα (Σε μία ανάλυση) Επί της συσκευής (Σύνολο) Μέση τιμή 95 U/l SD 1.2 CV% 1.2 Μέση τιμή 226 U/l SD CV% 2.0 0.9 Μέση τιμή 637 U/l SD CV% 4.1 0.6 1.7 1.7 2.8 11.7 1.2 1.8 Διεξήχθη μελέτη ακριβείας με οδηγό το έγγραφο EP5-A2 του CLSI, σε ένα κέντρο με τρεις αναλυτές Indiko, ένα χειριστή, δύο παρτίδες αντιδραστηρίων, μία παρτίδα βαθμονομητή και επτά βαθμονομήσεις σε 21 ημέρες, με αριθμό μετρήσεων n = 84. Τα αποτελέσματα από κάθε αναλυτή συνδυάστηκαν. Σύγκριση μεθόδων Konelab Διεξήχθη συγκριτική μελέτη στον αναλυτή Konelab Prime 60 με οδηγό το έγγραφο EP9-A2 του CLSI και με την προηγούμενη μέθοδο 981372 CK (IFCC) (x) ως μέθοδο αναφοράς. Γραμμική παλινδρόμηση (Deming) (μονάδα αποτελεσμάτων U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Οι συγκεντρώσεις του δείγματος ήταν μεταξύ 17 και 2253 U/l. Indiko 1) Διεξήχθη συγκριτική μελέτη στον αναλυτή Indiko με οδηγό το έγγραφο EP9-A2 του CLSI και με την προηγούμενη μέθοδο 981372 CK (IFCC) στον αναλυτή Konelab Prime 60 (x) ως μέθοδο αναφοράς. Γραμμική παλινδρόμηση (Deming) (μονάδα αποτελεσμάτων U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Οι συγκεντρώσεις του δείγματος ήταν μεταξύ 17 και 2253 U/l. 2) Διεξήχθη συγκριτική μελέτη στον αναλυτή Indiko με οδηγό το έγγραφο EP9-A2 του CLSI και με την ίδια μέθοδο CK (IFCC) στον αναλυτή Konelab Prime 60 (x) ως μέθοδο αναφοράς. Γραμμική παλινδρόμηση (Deming) (μονάδα αποτελεσμάτων U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Οι συγκεντρώσεις του δείγματος ήταν μεταξύ 14 και 2450 U/l. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 8 Page 9 D06025_04_Insert_CK (IFCC)_MU ES CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) ESTE PROSPECTO ES PARA USO FUERA DE EE. UU. INDICACIONES Para uso diagnóstico in vitro en la determinación cuantitativa de la actividad de la creatina-cinasa (ATP: creatina N-fosfotransferasa (CK), EC 2.7.3.2) en suero o plasma humano en analizadores Konelab e Indiko. Las referencias a los sistemas Konelab son extensibles a la serie T. RESUMEN (1,2) La actividad de la creatina-cinasa es mayor en los músculos estriados, el cerebro y el tejido cardíaco. Otros tejidos, como los riñones, el diafragma y la próstata, presentan una actividad notablemente inferior. La actividad de la CK es elevada en muchas enfermedades, incluidas las que afectan al músculo esquelético, el corazón, el sistema nervioso central y la tiroides. La actividad total de la CK en sangre comienza a elevarse tras una lesión muscular aguda y alcanza niveles patológicos en el 50 % de pacientes tras una duración media de 4-5 horas. Los valores aumentan regularmente a partir de 8 a 24 horas tras el comienzo del infarto y recuperan el rango de referencia con grandes fluctuaciones según el individuo. PRINCIPIO DEL PROCEDIMIENTO El ensayo CK es un ensayo líquido en dos partes que utiliza tres enzimas para determinar la actividad de la CK. La creatina-cinasa cataliza la reacción entre la fosfocreatina y la ADP (adenosina-5'-difosfato) y forma creatina y ATP (adenosina-5'-trifosfato). Este último compuesto fosforila la glucosa a G-6-P (glucosa-6-fosfato) en presencia de HK (hexocinasa). La G-6-P se oxida a 6-PG (gluconato-6fosfato) y, al mismo tiempo, una cantidad equimolar de NADP (nicotinamida adenina dinucleótido fosfato) se reduce a NADPH en una reacción catalizada por G-6P-DH (glucosa-6-fosfato deshidrogenasa). El incremento de absorbencia a 340 nm causado por la formación de NADPH es proporcional a la actividad de la creatina-cinasa en la muestra (3). CK Fosfocreatina + ADP → Creatina + ATP HK ATP + glucosa → ADP + G-6-P 4 horas a 20…25 °C, 1 mes a 4…8 °C o 1 mes a -20 °C; protegida de la luz. CK-BB no estable. PROCEDIMIENTO DEL ENSAYO Consulte las notas de aplicación para obtener información sobre el uso de un procedimiento automatizado con el analizador. No puede garantizarse la fiabilidad de ninguna aplicación no aprobada por Thermo Fisher Scientific Oy, por lo que deberá evaluarla el usuario. Ajustes del analizador Indiko Para introducir los parámetros de aplicación, utilice el código de barras que aparece en la hoja de aplicación de Indiko o el archivo electrónico, según corresponda. Material suministrado Los reactivos antes descritos. Material requerido pero no suministrado El calibrador y los controles indicados más abajo. Calibración La respuesta (dA/min) se convierte en unidades de resultado con un factor de cálculo. También se puede usar eCal (981830) como calibrador según las instrucciones del prospecto del paquete. Control de la calidad Utilice muestras de control de la calidad al menos una vez al día, después de cada calibración y cada vez que use un nuevo frasco de reactivo. Se recomienda utilizar al menos dos niveles (alto y bajo) de controles o mezclas de muestras. Siga siempre las normas locales, estatales y federales al realizar los controles de calidad. Controles disponibles: CK-MB Control, código 981908 CK-MB Control High, código 981255 Nortrol, código 981043 Abtrol, código 981044 Los intervalos y límites de los controles deben adaptarse a los requisitos de cada laboratorio. Los resultados de las muestras de control de la calidad deben estar dentro de los límites establecidos por el laboratorio. CALCULO DE RESULTADOS El analizador calcula los resultados automáticamente de la manera siguiente: G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Actividad en l/U = Factor x dA/min INFORMACIÓN DE REACTIVOS El kit código 981828 contiene 2 viales con 52 ml de reactivo A 2 viales con 13 ml de reactivo B Muestra El kit código 981829 contiene 2 viales con 20 ml de reactivo A 2 viales con 5 ml de reactivo B Concentraciones Reactivo A Tampón (imidazol), pH 6.5 D-glucosa N-acetil-L-cisteína (NAC) Acetato de magnesio EDTA-Na2 Nicotinamida adenina dinucleótido fosfato (NADP) Hexocinasa (HK) NaN3 Reactivo B Tampón (imidazol) Adenosina-5'-difosfato (ADP) Adenosina-5'-monofosfato (AMP) Di(adenosina-5')pentafosfato (AP5A) Glucosa-6-fosfato deshidrogenasa (G-6-P-DH) EDTA -Na2 Fosfocreatina NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/I < 0.1 % TV x 1000 Factor teórico = 6.22 x SV x P TV SV 6.22 P Volumen total de reacción en ml Volumen de muestra en ml Coeficiente de extinción de NADPH a 340 nm Longitud de recorrido de cubeta = 1 Factor de conversión µkat/l = (U/l)/60 Si se utiliza eCal (981830) como calibrador, el analizador calcula automáticamente los resultados con una curva de calibración. Curva de calibración (ejemplo) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Precauciones El reactivo B es peligroso. Consulte la hoja separada dentro del kit para ver las indicaciones de peligrosidad y las precauciones: H315, P280, P302+P352, P305+P351+P338. Los reactivos contienen azida sódica como conservante. Deben adoptarse las precauciones normales requeridas en el manejo de reactivos en laboratorio. El producto debe desecharse como producto químico de laboratorio según la normativa local. Preparación Los reactivos A y B están listos para su uso. Nota: compruebe que no haya burbujas en el cuello de la botella ni en la superficie del reactivo cuando introduzca viales o recipientes de reactivos en el analizador. Conservación y estabilidad Los reactivos en viales sin abrir son estables a 2...8 °C hasta la fecha de caducidad impresa en la etiqueta. Consulte las notas de aplicación del analizador para determinar la estabilidad de los reactivos una vez abiertos y puestos en el analizador. Si los reactivos presentan turbidez, no los utilice. RECOGIDA DE MUESTRAS Nota: cuando se procesen muestras en tubos de recogida de muestras, siga atentamente las instrucciones del fabricante del tubo para evitar resultados incorrectos. Preste especial atención a las variables preanalíticas, como la mezcla, el tiempo de verticalidad antes de la centrifugación y los ajustes de la centrífuga. Tipo de muestra Puede utilizarse suero no hemolizado o plasma con heparina. Nota: la hemólisis puede dar lugar a resultados desviados. No exponga las muestras a una iluminación intensa. Precauciones Las muestras humanas se deben manipular y desechar como si se tratase de material potencialmente infeccioso. La curva de calibración depende del lote. LIMITACIONES DEL PROCEDIMIENTO Para fines diagnósticos, los resultados deben evaluarse siempre conjuntamente con la anamnesis del paciente, las exploraciones físicas y otros datos. Interferencia Se ha realizado un estudio según las directrices del documento EP7-A2 de CLSI. Criterios: diferencia de ± 10 % respecto a los valores iniciales. Konelab Sustancia interferente Bilirrubina conjugada Bilirrubina no conjugada Hemoglobina en hemolizado Lipidemia: Intralipid (marca comercial de Fresenius Kabi AB) Concentración interferente CK objetivo (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Diferencia respecto del objetivo (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Conservación (4) Las muestras pueden conservarse como se indica abajo. D06025_04_Insert_CK (IFCC)_MU Indiko Sustancia interferente Bilirrubina conjugada Bilirrubina no conjugada Hemoglobina en hemolizado Lipidemia: Intralipid (marca comercial de Fresenius Kabi AB) Concentración interferente CK objetivo (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Diferencia respecto del objetivo (%) - 0.8 - 2.4 + 0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 No utilice muestras hemolizadas. Existe una escasa correlación entre la turbidez y la concentración de triglicéridos. Rara vez la gammopatía, en particular el tipo IgM (macroglobulinemia de Waldeström), puede generar resultados no fiables (6). No probado en Thermo Fisher Scientific Oy. Si desea obtener información sobre la interferencia de otras sustancias, consulte la referencia (5). VALORES PREVISTOS (3) Hombre < 171 U/l (< 2.85 µkat/l) a 37 °C Mujer < 145 U/l (< 2.41 µkat/l) a 37 °C Los valores mencionados sirven sólo de referencia. Se recomienda que cada laboratorio verifique este rango u obtenga un intervalo de referencia para la población a la que atiende. Factor de conversión µkat/l = (U/l)/60 INTERVALO DE MEDICION 10 – 1000 U/l (0.17 – 17 µkat/l). Intervalo de medición aumentado después de la segunda dilución: 10 – 9000 U/l (0.17 – 150 µkat/l). CARACTERISTICAS DE RENDIMIENTO Los resultados obtenidos en cada laboratorio pueden diferir de los datos de resultados presentados. Límites de detección Se realizó un estudio según las directrices del documento EP17-A. Konelab LdB: 0.6 U/l (0.01 µkat/l) El límite del blanco (LdB) representa el resultado de medición más alto que es probable que se observe con una muestra sin analito (n = 60). LdD: 1.5 U/l (0.03 µkat/l) El límite de detección (LdD) representa la cantidad más baja de analito que puede detectarse en una muestra (n = 60). LdC: 3.2 U/l (0.05 µkat/l) El límite de cuantificación (LdC) representa la concentración real más baja que puede determinarse cuantitativamente en una muestra (n = 41). Indiko LdB: 0.2 U/l (0.003 µkat/l) El límite del blanco (LdB) representa el resultado de medición más alto que es probable que se observe con una muestra sin analito (n = 60). LdD: 2.2 U/l (0.04 µkat/l) El límite de detección (LdD) representa la cantidad más baja de analito que puede detectarse en una muestra (n = 60). LdC: 3.8 U/l (0.06 µkat/l) El límite de cuantificación (LdC) representa la concentración real más baja que puede determinarse cuantitativamente en una muestra (n = 42). Precisión La imprecisión del ensayo de CK es ≤ 5.0 % del coeficiente de variación (CV) total. Konelab CK Repetibilidad (intraserie) Intradispositivo (total) Media 92 U/l DE CV% 1.7 1.8 Media 226 U/l DE CV% 2.0 0.9 Media 616 U/l DE CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 En el estudio de precisión realizado en un centro según las directrices del documento EP5-A2 de CLSI se usaron tres analizadores Konelab 60, un operador, 2 lotes de reactivos, un lote de calibradores y 6 calibraciones durante 21 días, siendo el número de medidas n = 84. Los resultados de cada analizador se combinaron. Indiko CK Repetibilidad (intraserie) Intradispositivo (total) Media 95 U/l DE CV% 1.2 1.2 Media 226 U/l DE CV% 2.0 0.9 Media 637 U/l DE CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 En el estudio de precisión realizado en un centro según las directrices del documento EP5-A2 de CLSI se usaron tres analizadores Indiko, un operador, 2 lotes de reactivos, un lote de calibradores y 7 calibraciones durante 21 días, siendo el número de medidas n = 84. Los resultados de cada analizador se combinaron. Comparación de métodos Konelab Se realizó un estudio comparativo con Konelab Prime 60 según las directrices del documento EP9-A2 de CLSI y con el método 981372 CK(IFCC) (x) anterior como referencia. Regresión lineal (Deming) (unidad de resultado U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Las concentraciones de las muestras oscilaron entre 17 y 2253 U/l. Indiko 1) Se realizó un estudio comparativo con Indiko según las directrices del documento EP9-A2 de CLSI y con el método 981372 CK(IFCC) utilizado anteriormente con Konelab Prime 60 (x) como referencia. Regresión lineal (Deming) (unidad de resultado U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Las concentraciones de las muestras oscilaron entre 17 y 2253 U/l. 2) Se realizó un estudio comparativo con Indiko según las directrices del documento EP9-A2 de CLSI y con el mismo método CK(IFCC) utilizado en Konelab Prime 60 (x) como referencia. Regresión lineal (Deming) (unidad de resultado U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Las concentraciones de las muestras oscilaron entre 14 y 2450 U/l. BIBLIOGRAFÍA 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. FABRICANTE Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa (Finlandia) Tel. +358 10 329 200 www.thermoscientific.com Fecha de revisión (aaaa-mm-dd) 2015-05-08 Cambios desde la versión anterior Una línea junto al texto indica modificaciones. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 10 Page 11 D06025_04_Insert_CK (IFCC)_MU ET CK (IFCC) 2 × 65 ml (Konelab™) 2 × 25 ml (Konelab™, Indiko™) PAKENDI INFOLEHT ON KOOSTATUD KASUTAMISEKS VÄLJASPOOL USA-D. Proov 4 tundi temperatuuril 20…25 °C, 1 kuu temperatuuril 4…8 °C või 1 kuu temperatuuril -20 °C kaitstuna valguse eest. CK-BB ei ole stabiilne. MÕÕTEPROTSEDUUR Teave analüsaatori automaatse mõõteprotseduuri kasutamise kohta on esitatud tehnilistes märkustes. Thermo Fisher Scientific Oy poolt kinnitamata rakendusviiside sooritusnäitajaid tagada ei saa, seetõttu peab neid hindama lõppkasutaja. SIHTOTSTARVE Kreatiini kinaasi (ATP: kreatiini N-fosfotransferaas (CK), EC 2.7.3.2) aktiivsuse in vitro kvantitatiivseks määramiseks inimese seerumis või plasmas Konelabi ja Indiko analüsaatoritel. Konelabi süsteemide viited kehtivad ühtlasi T-seeria kohta. Indiko seadme sätted Sisestage rakenduse parameetrid Indiko rakenduse lehel oleva vöötkoodi või elektroonilise faili kaudu. KOKKUVÕTE (1, 2) Kreatiini kinaasi aktiivsus on kõige kõrgem vöötlihastes, ajus ja südamekoes. Muudes kudedes, nagu neerudes, diafragmas ja eesnäärmes, on aktiivsus palju madalam. CK aktiivsus on kõrgem paljude haiguste, sh skeletilihaste, südame, kesknärvisüsteemi ja kilpnäärmega seotud haiguste puhul. Pärast ägedat lihasekahjustust hakkab CK koguaktiivsus veres tõusma, jõudes 50% patsientidest haigusliku tasemeni keskmiselt 4–5 tunni jooksul. Väärtused tõusevad korrapäraselt alates 8.–24. tunni möödumisest infarkteerumise algusest ja naasevad normaalvahemikku suurte individuaalsete erinevustega. Kaasasolevad materjalid Eespool kirjeldatud reaktiivid. PROTSEDUURI PÕHIMÕTE Ck analüüs on kaheosaline vededlikanalüüs, mis kasutab CK aktiivsuse määramiseks kolme ensüümi. Kreatiini kinaas katalüüsib reaktsiooni kreatiinfosfaadi ja ADP (adenosiin-5’-difosfaat) vahel, reaktsioonis tekivad kreatiin ja ATP (adenosiin-5’-trifosfaat). Viimane fosforüülib glükoosi HK (heksokinaas) juuresolekul glükoos-6-fosfaadiks (G-6-P). G-6-P oksüdeerub 6-PG-ks (glükonaat-6fosfaat), samal ajal redutseerub ekvimolaarne NADP (nikotiinamiidadeniindinukleotiidfosfaat) kogus G-6P-DH (glükoos-6-fosfaadi dehüdrogenaas) katalüüsitavas reaktsioonis NADPH-ks. NADPH tekkest tingitud neelduvuse tõus lainepikkusel 340 nm on võrdeline kreatiini kinaasi aktiivsusega proovis (3). Vajaminevad materjalid, mida kaasas pole Allpool kirjeldatud kalibraator ja kontrollmaterjalid. Kalibreerimine Vastus (dA/min) teisendatakse ühikuliseks tulemuseks vastava arvutusteguri abil. Teine võimalus on kasutada eCal (981830) kalibraatorit, järgides eCal (981830) pakendi infolehes toodud juhiseid. HK ATP + Glükoos → ADP + G-6-P Kvaliteedikontroll Kvaliteedikontrolli materjale kasutatakse vähemalt kord päevas, pärast igat kalibreerimist ja iga kord, kui kasutatakse uut reaktiivipudelit. Kasutage vähemalt kahe tasemega kontrollmaterjale (madal ja kõrge) või proovisegusid. Kvaliteedikontrolli (QC) läbiviimisel järgige alati kohalikke, riiklikke ja föderaalseid regulatsioone. Saadaval on järgmised kontrollmaterjalid: CK-MB Control, kood 981908 CK-MB Control High, kood 981255 Nortrol, kood 981043 Abtrol, kood 981044 Kontrollmaterjalide (KK) vahemikud ja piirid tuleb kohaldada konkreetse labori vajadustele. Kontrollmaterjalide vahemikud ja piirid tuleb kohaldada individuaalsete laborite vajadustele. G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ TULEMUSTE ARVUTAMINE Tulemused arvutab analüsaator automaatselt järgmisel viisil REAKTIIVI INFO Aktiivsus U/l = tegur × dA/min CK Kreatiinfosfaat + ADP → Kreatiin + ATP Komplekt koodiga 981828 sisaldab 2 viaali 52 ml reaktiivi A 2 viaali 13 ml reaktiivi B Komplekt koodiga 981829 sisaldab 2 viaali 20 ml reaktiivi A 2 viaali 5 ml reaktiivi B TV SV 6.22 P Kontsentratsioonid Reaktiiv A Puhver (imidasool), pH 6.5 D-glükoos N-atsetüül-L-tsüsteiin (NAC) Magneesiumatsetaat EDTA-Na2 Nikotiinamiidadeniindinukleotiidfosfaat (NADP) Heksokinaas (HK) NaN3 Reaktiiv B Puhver (Imidasool) Adenosiin-5’-difosfaat (ADP) Adenosiin-5’-monofosfaat (AMP) Di(adenosiin-5’)pentafosfaat (AP5A) Glükoos-6-fosfaadi dehüdrogenaas (G-6-P-DH) EDTA -Na2 Kreatiinfosfaat NaN3 TV × 1000 Teoreetiline tegur = 6.22 × SV × P 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1% Reaktsiooni kogumaht (ml) Proovi maht (ml) NADPH ekstinktsioonikoefitsient lainepikkusel 340 nm Küveti optiline teepikkus = 1 Teisendustegur µkat/l = (U/l)/60 eCal (981830) kasutamisel kalibraatorina arvutab analüsaator tulemused kalibreerimiskõvera järgi automaatselt Kalibreerimiskõver (näide) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14 000 U/l 2 mmol/l 160 mmol/l < 0.1% Ettevaatusabinõud Reaktiiv B on ohtlik. Vaadake ohu- ja ettevaatuslausete komplektis olevat eraldi lehte: H315, P280, P302+P352, P305+P351+P338. Reaktiivid sisaldavad säilitusainena naatriumasiidi. Kõigi laborireaktiivide käsitsemisel tuleb rakendada tavapäraseid ettevaatusabinõusid. Toode tuleb kõrvaldada laborikemikaalina kooskõlas kohalike eeskirjadega. Ettevalmistus Reaktiivid A ja B on kasutusvalmis. Märkus: enne reaktiiviviaalide või -nõude viimist analüsaatorisse tuleb kontrollida, et pudelikaelas ega reaktiivi pinnal ei ole mulle. Säilitamine ja stabiilsus Avamata viaalides püsivad reaktiivid temperatuuril 2…8 °C stabiilsena sildile trükitud aegumistähtajani. Teave reaktiivide seadmesisese stabiilsuse kohta on esitatud analüsaatori tehnilistes märkustes. Kui reaktiivides esineb hägusust, ärge kasutage neid. PROOVIDE VÕTMINE Märkus: väärade tulemuste vältimiseks tuleb proovide töötlemisel proovivõtukatsutites tähelepanelikult järgida katsuti valmistaja juhiseid. Pöörake kindlasti tähelepanu analüüsile eelnevatele teguritele nagu näiteks segamine, säilitamisaeg enne tsentrifuugimist ja tsentrifuugi sätted. Proovi tüüp Kasutada võib hemolüüsumata seerumi- või hepariiniga plasmaproove. Märkus: hemolüüs võib viia tulemuste kõrvalekalleteni. Proove tuleb hoida tugeva valguse eest. Ettevaatusabinõud Inimpäritolu proove tuleb käsitsemisel ja kõrvaldamisel lugeda võimalikeks nakkusallikateks. Kalibreerimiskõver on eri partiide puhul erinev. PROTSEDUURI PIIRANGUD Diagnostilistel eesmärkidel tuleb analüüsi tulemusi kasutada alati kombinatsioonis haigusloo, kliinilise läbivaatuse ja teiste leidudega. Segavad mõjud Uuring viidi läbi CLSI dokumendi EP7-A2 suuniseid järgides. Kriteerium: erinevus algväärtusest ±10%. Konelab Segav aine Konjugeeritud bilirubiin Konjugeerimata bilirubiin Hemoglobiin hemolüsaadis Lipeemia: Intralipid (Fresenius Kabi AB kaubamärk) Segav kontsentratsioon 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l CK sihtväärtus (U/l) 188 498 180 495 184 505 Erinevus sihtväärtusest (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Säilitamine (4) Proove võib säilitada alljärgnevalt kirjeldatud viisil. D06025_04_Insert_CK (IFCC)_MU Indiko Segav aine Segav kontsentratsioon Konjugeeritud bilirubiin Konjugeerimata bilirubiin Hemoglobiin hemolüsaadis Lipeemia: Intralipid (Fresenius Kabi AB kaubamärk) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l CK sihtväärtus (U/l) 179 477 175 469 187 502 Erinevus sihtväärtusest (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Hoiduge kasutamast hemolüüsitud proove. Korrelatsioon hägususe ja triglütseriidide sisalduse vahel on väike. Väga harvadel juhtudel võib gammopaatia, ennekõike IgM-tüüpi (Waldenströmi makroglobulineemia), põhjustada ebatõeseid tulemusi (6). Ei ole testitud Thermo Fisher Scientific Oy-s. Muid segavaid aineid on käsitletud viites 5. EELDATAVAD VÄÄRTUSED (3) Mees < 171 U/l (< 2.85 µkat/l) temp. 37 °C Naine < 145 U/l (< 2.41 µkat/l) temp. 37 °C Need väärtused on mõeldud suunitluslikena. Kõigil laboritel on soovitatav seda vahemikku kontrollida või tuletada ise etalonvahemik teenindatava populatsiooni seas. Teisendamistegur µkat/l = (U/l)/60 SOORITUSNÄITAJAD Individuaalsetes laborites saadavad tulemused võivad erineda esitatud sooritusnäitajatest. Avastamise piirmäär Uuring viidi läbi CLSI dokumendi EP17-A suuniseid järgides. Konelab LoB: 0.6 U/l (0.01 µkat/l) Puhas piirmäär (LoB) on suurim tõenäoline mõõtmistulemus analüüdivabast proovist (n=60). LoD: 1.5 U/l (0.03 µkat/l) Avastamise piirmäär (LoD) on väikseim avastatav analüüdikogus proovis (n=60). LoQ: 3.2 U/l (0.05 µkat/l) Koguse piirmäär (LoQ) on madalaim tegelik kontsentratsioon proovis, mida on võimalik kvantitatiivselt määrata (n=41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Puhas piirmäär (LoB) on suurim tõenäoline mõõtmistulemus analüüdivabast proovist (n=60). LoD: 2.2 U/l (0.04 µkat/l) Avastamise piirmäär (LoD) on väikseim avastatav analüüdikogus proovis (n=60). LoQ: 3.8 U/l (0.06 µkat/l) Koguse piirmäär (LoQ) on madalaim tegelik kontsentratsioon proovis, mida on võimalik kvantitatiivselt määrata (n=42). Ebatäpsus CK analüüsi ebatäpsus on ≤ 5.0% üld-CV-st. Korratavus (mõõtmisseeria sisene) Seadmesisene (summaarne) Keskmine 92 U/l SD 1.7 CV% 1.8 Keskmine 226 U/l SD CV% 2.0 0.9 Keskmine 616 U/l SD CV% 7.1 1.2 2.5 2.8 3.1 22.3 1.4 3.6 Vastavalt CLSI dokumendi EP5-A2 suunistele viidi ühes laboris kolme Konelab 60 analüsaatori peal ühe operaatori poolt 2 reaktiivipartiid, ühte kalibraatoripartiid ja 6 kalibreerimist kasutades 21 päeva jooksul läbi täpsusuuring, mõõtmiste arv uuringus oli n = 84. Kõigi analüsaatorite tulemused ühendati. Indiko CK Korratavus (mõõtmisseeria sisene) Seadmesisene (summaarne) Keskmine 95 U/l SD 1.2 CV% 1.2 Keskmine 226 U/l SD CV% 2.0 0.9 Keskmine 637 U/l SD CV% 4.1 0.6 1.7 1.7 2.8 11.7 1.2 TOOTJA Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Teksti läbivaatamise kuupäev (aaaa-kk-pp) 2015-05-08 MÕÕTEPIIRKOND 10 – 1000 U/l (0.17 – 17 µkat/l). Laiendatud mõõtepiirkond pärast teisest lahjendust: 10 – 9000 U/l (0.17 – 150 µkat/l). Konelab CK KIRJANDUS 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. 1.8 Vastavalt CLSI dokumendi EP5-A2 suunistele viidi ühes laboris kolme Indiko analüsaatori peal ühe operaatori poolt 2 reaktiivipartiid, ühte kalibraatoripartiid ja 7 kalibratsiooni kasutades 21 päeva jooksul läbi täpsusuuring, mõõtmiste arv uuringus oli n = 84. Kõigi analüsaatorite tulemused ühendati. Meetodite võrdlus Konelab Vastavalt CLSI dokumendi EP9-A2 suunistele viidi Konelab Prime 60 analüsaatoril läbi võrdlusuuring, kasutades võrdlusalusena varasemat 981372 CK (IFCC) meetodit (x). Lineaarne regressioon (Deming) (tulemus ühikutes U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Proovide kontsentratsioonid jäid vahemikku 17 kuni 2253 U/l. Indiko 1) Vastavalt CLSI dokumendi EP9-A2 suunistele viidi Indiko analüsaatoril läbi võrdlusuuring, kasutades võrdlusalusena varasemat 981372 CK (IFCC) meetodit Konelab Prime 60 analüsaatoril (x). Lineaarne regressioon (Deming) (tulemus ühikutes U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Proovide kontsentratsioonid jäid vahemikku 17 kuni 2253 U/l. 2) Vastavalt CLSI dokumendi EP9-A2 suunistele viidi Indiko analüsaatoril läbi võrdlusuuring, kasutades võrdlusalusena sama CK (IFCC) meetodit Konelab Prime 60 (x) analüsaatoril (x). Lineaarne regressioon (Deming) (tulemus ühikutes U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Proovide kontsentratsioonid jäid vahemikku 14 kuni 2450 U/l. Muudatused võrreldes eelmise versiooniga Teksti kõrval olev joon tähistab muudatusi. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 12 Page 13 D06025_04_Insert_CK (IFCC)_MU FR CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) CETTE NOTICE EST VALABLE POUR UNE UTILISATION EN DEHORS DES ÉTATS-UNIS. UTILISATION PREVUE Pour usage diagnostique in vitro basé sur la détermination quantitative de l’activité de la créatine kinase (ATP : créatine N-phosphotransférase (CK), EC 2.7.3.2) dans le sérum ou le plasma humain sur les analyseurs Konelab et Indiko. Toute référence aux systèmes Konelab fait également référence à la série T. RESUME (1, 2) L'activité de la créatine kinase est maximale au niveau des muscles striés, du cerveau et du tissu cardiaque. Les autres tissus comme les reins, le diaphragme et la prostate contiennent une activité significativement inférieure. L'activité de la CK est élevée dans de nombreuses maladies, notamment celles impliquant les muscles squelettiques, le cœur, le système nerveux central et la thyroïde. Après un dommage musculaire aigu, l'activité totale de la CK dans le sang commence à augmenter pour atteindre des niveaux pathologiques après un délai moyen de 4 à 5 heures chez 50 % des patients. Les valeurs augmentent régulièrement entre la 8ème et la 24ème heure après le début d'un infarctus, pour revenir ensuite à l'intervalle de référence, avec des variations importantes entre les individus. PRINCIPE DE LA PROCEDURE Le dosage CK est un dosage liquide en deux parties utilisant trois enzymes pour déterminer l’activité de la CK. La créatine kinase catalyse la réaction entre la créatine phosphate et l'ADP (adénosine-5'diphosphate) avec une formation de créatine et d'ATP (adénosine-5'-triphosphate). Ce dernier composé phosphoryle le glucose en G-6-P (glucose-6-phosphate) en présence d'hexokinase (HK). Le G-6-P est oxydé en 6-PG (gluconate-6-phosphate) et, en même temps, une quantité équimolaire de NADP (nicotinamide adénine dinucléotide phosphate) est réduite en NADPH au cours d'une réaction catalysée par la G-6P-DH (glucose-6-phosphate déshydrogénase). L'augmentation de l'absorbance à 340 nm résultant de la formation de NADPH est proportionnelle à l'activité de la créatine kinase dans l'échantillon (3). Echantillon 4 heures à 20…25 °C, 1 mois à 4…8 °C ou 1 mois à -20 °C, protégé de la lumière. CK-BB non stable. PROCÉDURE DE TEST Pour une description de la procédure automatisée sur votre modèle d'analyseur, se référer à la fiche d'application de l'appareil. Toute application n’ayant pas été validée par Thermo Fisher Scientific Oy ne peut pas être garantie en ce qui concerne ses performances et doit par conséquent être évaluée par l’utilisateur. Paramètres de l'instrument Indiko Entrer les paramètres d'application via le code-barres qui se situe dans la fiche d'application Indiko ou via le fichier électronique, selon le cas. Matériel fourni Réactifs décrits ci-dessus. Matériel requis mais non fourni Étalon et contrôles indiqués ci-dessous. Étalonnage La réponse (dA/min) est convertie en unités de résultat à l’aide d'un facteur de calcul. Ou utiliser l'étalon eCal (981830) conformément aux instructions figurant dans la notice de eCal (981830). CK Créatine phosphate + ADP → Créatine + ATP Contrôle qualité Utiliser les échantillons de contrôle qualité (QC) au moins fois par jour, après chaque étalonnage et à chaque fois qu'un nouveau flacon de réactif est entamé. Il est conseillé d’utiliser au moins deux niveaux (bas et élevé) de contrôles ou de pools d’échantillons. Toujours se conformer aux réglementations locales, fédérales et nationales applicables en matière de contrôle qualité. Contrôles disponibles : CK-MB Control, N° de réf. 981908 CK-MB Control High, N° de réf. 981255 Nortrol, N° de réf. 981043 Abtrol, N° de réf. 981044 Les intervalles et les limites de contrôle (QC) doivent être adaptés aux besoins de chaque laboratoire. Les résultats des échantillons de contrôle qualité doivent se situer dans la plage de tolérance prédéfinie par le laboratoire. HK ATP + Glucose → ADP + G-6-P CALCUL DES RESULTATS Les résultats sont calculés automatiquement par l'analyseur de la manière suivante : G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Activité en U/l = Facteur x dA/min TV x 1000 Facteur théorique = 6.22 x SV x P INFORMATIONS SUR LES RÉACTIFS Contenu de la trousse 981828 2 flacons de 52 ml de réactif A 2 flacons de 13 ml de réactif B Contenu de la trousse 981829 2 flacons de 20 ml de réactif A 2 flacons de 5 ml de réactif B Concentrations Réactif A Tampon (Imidazole), pH 6.5 D-glucose N-acétyl-L-cystéine (NAC) Acétate de magnésium EDTA-Na2 Nicotinamide adénine dinucléotide phosphate (NADP) Hexokinase (HK) NaN3 Réactif B Tampon (imidazole) Adénosine-5’-diphosphate (ADP) Adénosine-5’-monophosphate (AMP) Di(adénosine-5’)penta-phosphate (AP5A) Glucose-6-phosphate déshydrogénase (G-6-P-DH) EDTA -Na2 Créatine phosphate NaN3 TV SV 6.22 P Volume total de la réaction en ml Volume d'échantillon en ml Coefficient d'extinction du NADPH à 340 nm Longueur du trajet de la cuvette = 1 Facteur de conversion µkat/l = (U/l)/60 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5 000 U/l < 0.1 % Si l'on utilise l'étalon eCal (981830), les résultats sont calculés automatiquement par l'analyseur à l'aide d'une courbe d'étalonnage. Courbe d’étalonnage (exemple) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14 000 U/l 2 mmol/l 160 mmol/l < 0.1 % Précautions Le réactif B est dangereux. Voir aussi la liste séparée des mentions de danger et de précautions à l’intérieur du kit : H315, P280, P302 + P352, P305 + P351 + P338. Les réactifs contiennent de l’azoture de sodium comme conservateur. La courbe d'étalonnage dépend du lot. Prendre les précautions habituelles requises lors de la manipulation de tout réactif de laboratoire. Ce produit doit être éliminé comme un produit chimique de laboratoire conformément aux réglementations locales. LIMITES DE LA PROCEDURE Pour l’établissement d'un diagnostic, les résultats doivent toujours être évalués en association avec les antécédents médicaux du patient, les examens cliniques et les autres observations. Préparation Les réactifs A et B sont prêts à l'emploi. Remarque : s’assurer de l’absence de bulles au niveau du goulot du flacon ou à la surface du réactif lors de la mise en place des flacons ou des récipients de réactifs dans l’analyseur. Interférence Une étude a eu lieu conformément aux directives du document CLSI EP7-A2. Critère : différence se situant dans les limites de ±10 % des valeurs initiales. Conservation et stabilité Les réactifs contenus dans les flacons scellés sont stables à 2…8 °C jusqu’à la date de péremption figurant sur l’étiquette. Se référer à la fiche d'application de l'analyseur en ce qui concerne la stabilité des réactifs dans l'appareil. Si les réactifs montrent une turbidité, ne pas les utiliser. PRELEVEMENT DES ECHANTILLONS Remarque : si les échantillons sont traités dans des tubes de prélèvement d’échantillons, suivre attentivement les instructions du fabricant des tubes afin d’éviter des résultats erronés. Prêter une attention particulière aux variables préanalytiques comme le mélange, le temps de repos avant centrifugation et les réglages de la centrifugeuse. Type d'échantillon On peut utiliser des échantillons de sérum ou de plasma hépariné non hémolysés. Remarque : une hémolyse peut entraîner une déviation des résultats. Éviter l'exposition des échantillons à la lumière vive. Précautions Les échantillons d'origine humaine doivent être manipulés et éliminés comme du matériel potentiellement infectieux. Konelab Susbtance interférente Bilirubine conjuguée Bilirubine non conjuguée Hémoglobine dans l'hémolysat Lipémie : Intralipid (marque de Fresenius Kabi AB) Concentration de substance interférente CK cible (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Différence depuis la cible (%) -5.3 -5.8 -1.7 -2.9 +8.5 +4.6 10 g/l 10 g/l 181 504 +2.8 +7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Stockage (4) Les échantillons peuvent être conservés comme indiqué ci-dessous. D06025_04_Insert_CK (IFCC)_MU Indiko Susbtance interférente Bilirubine conjuguée Bilirubine non conjuguée Hémoglobine dans l'hémolysat Lipémie : Intralipid (marque de Fresenius Kabi AB) Concentration de substance interférente CK cible (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Différence depuis la cible (%) -0.8 -2.4 +0.2 -0.1 +8.6 +3.4 10 g/l 10 g/l 185 505 +1.0 -0.8 Éviter d'utiliser des échantillons hémolysés. Il existe une faible corrélation entre turbidité et concentration en triglycérides. De très rares cas de gammapathie, notamment de type IgM (macroglobulinémie de Waldenström), peuvent être à l’origine de résultats non fiables (6). Non testé par Thermo Fisher Scientific Oy. Pour les autres substances interférentes, se reporter à la référence 5. VALEURS ATTENDUES (3) Homme < 171 U/l (< 2.85 µkat/l) à 37 °C Femme < 145 U/l (< 2.41 µkat/l) à 37 °C Les valeurs citées sont données à titre indicatif uniquement. Il est recommandé que chaque laboratoire vérifie cet intervalle ou établisse son propre intervalle pour la population qu'il dessert. Facteur de conversion µkat/l = (U/l)/60 INTERVALLE DE MESURES 10 – 1000 U/l (0.17 – 17 µkat/l). Intervalle de mesures étendu après dilution secondaire: 10 – 9000 U/l (0.17 – 150 µkat/l). CARACTERISTIQUES DE PERFORMANCE Les résultats obtenus dans chaque laboratoire peuvent différer des données de performance indiquées. Limite de détection Une étude a eu lieu conformément aux directives du document EP17-A. Konelab LoB: 0.6 U/l (0.01 µkat/l) La limite de blanc (LoB) représente le résultat de mesure le plus élevé qui est susceptible d’être observé pour un échantillon sans analyte (n = 60). LoD: 1.5 U/l (0.03 µkat/l) La limite de détection (LoD) représente la plus faible quantité d’analyte qu’il est possible de détecter dans un échantillon (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) La limite de quantification (LoQ) représente la plus faible concentration réelle qu’il est possible de quantifier dans un échantillon (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) La limite de blanc (LoB) représente le résultat de mesure le plus élevé qui est susceptible d’être observé pour un échantillon sans analyte (n = 60). LoD: 2.2 U/l (0.04 µkat/l) La limite de détection (LoD) représente la plus faible quantité d’analyte qu’il est possible de détecter dans un échantillon (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) La limite de quantification (LoQ) représente la plus faible concentration réelle qu’il est possible de quantifier dans un échantillon (n = 42). Imprécision L’imprécision du dosage CK est ≤ 5.0 % du CV total. Konelab CK Répétabilité (intra-série) Intra-analyseur (totale) Moyenne 92 U/l SD % CV 1.7 1.8 Moyenne 226 U/l SD % CV 2.0 0.9 Moyenne 616 U/l SD % CV 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 L’étude de précision a été réalisée conformément aux directives du document CLSI EP5-A2, sur un seul site et sur trois analyseurs Konelab 60, avec un opérateur, 2 lots de réactifs, un lot d’étalons et 6 étalonnages pendant 21 jours, le nombre de mesures étant de n = 84. Les résultats de chaque analyseur ont été combinés. Indiko CK Répétabilité (intra-série) Intra-analyseur (totale) Moyenne 95 U/l SD % CV 1.2 1.2 Moyenne 226 U/l SD % CV 2.0 0.9 Moyenne 637 U/l SD % CV 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 L’étude de précision a été réalisée conformément aux directives du document CLSI EP5-A2, sur un seul site et sur trois analyseurs Indiko, avec un opérateur, 2 lots de réactifs, un lot d’étalons et 7 étalonnages pendant 21 jours, le nombre de mesures étant de n = 84. Les résultats de chaque analyseur ont été combinés. Comparaison des méthodes Konelab Une étude de comparaison a été réalisée sur un analyseur Konelab Prime 60 conformément aux directives du document CLSI EP9-A2 en utilisant comme référence la méthode 981372 CK(IFCC) précédente. Régression linéaire (Deming) (unité du résultat : U/l) : y = 1.07 x – 6.3 r = 0.999 n = 95 Les concentrations des échantillons se situaient entre 17 et 2253 U/l. Indiko 1) Une étude de comparaison a été réalisée sur un analyseur Indiko conformément aux directives du document CLSI EP9-A2 et en utilisant comme référence la méthode 981372 CK(IFCC) précédente sur le Konelab Prime 60 (x). Régression linéaire (Deming) (unité du résultat : U/l) : y = 1.10 x – 4.7 r = 0.999 n = 95 Les concentrations des échantillons se situaient entre 17 et 2253 U/l. 2) Une étude de comparaison a été réalisée sur un analyseur Indiko conformément aux directives du document CLSI EP9-A2 en utilisant comme référence la même méthode CK(IFCC) que sur le Konelab Prime 60 (x). Régression linéaire (Deming) (unité du résultat : U/l) : y = 1.03 x + 1.4 r = 1.00 n = 98 Les concentrations des échantillons se situaient entre 14 et 2450 U/l. BIBLIOGRAPHIE 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. FABRICANT Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tél. +358 10 329 200 www.thermoscientific.com Date de révision (aaaa-mm-jj) 2015-05-08 Modifications par rapport à la version précédente Le trait en regard du texte indique les modifications. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 14 Page 15 D06025_04_Insert_CK (IFCC)_MU HU CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) EZ A TÁJÉKOZTATÓ AZ EGYESÜLT ÁLLAMOKON KÍVÜLI HASZNÁLATRA VONATKOZIK. RENDELTETÉS In vitro diagnosztikai felhasználásra az emberi szérum, illetve plazma kreatin kináz (ATP: Kreatin N-foszfotranszferáz (CK), EC 2.7.3.2) aktivitásának kvantitatív meghatározásában Konelab és Indiko analizátorokban. A Konelab rendszerekre tett minden hivatkozás a T sorozatra is vonatkozik. ÖSSZEGZÉS (1, 2) A kreatin kináz aktivitás a harántcsíkolt izom-, az agy-, valamint a szívizomszövetben a legmagasabb. Más szervek szövetei, mint például a vesék, a rekeszizom, vagy a prosztata szignifikánsan kevesebb aktivitással rendelkeznek. A CK aktivitás sok betegségben emelkedett, beleértve a vázizmok, a szív, a központi idegrendszer és a pajzsmirigy betegségeit. Akut izomkárosodás után a vér CK aktivitása emelkedni kezd, a betegek 50%-ában átlagosan 4-5 óra alatt éri el a kóros szintet. Infarktus kialakulása után az értékek rendszeresen 8-24 órával emelkednek, visszatérésük a referenciatartományba egyénenként nagy eltéréseket mutat. AZ ELJÁRÁS ALAPELVE A CK assay egy kétrészes folyadék-assay, amely három enzimet használ a CK-aktivitás meghatározására. A kreatin kináz a kreatin-foszfát és az ADP (adenozin-5’-difoszfát) közti reakciót katalizálja, mely során kreatin és ATP (adenozin-5’-trifoszfát) keletkezik. Az utóbbi vegyület foszforilálja a glukózt G-6-P-tá (glukóz-6-foszfát) HK (hexokináz) jelenlétében. A G-6-P 6-PG-tá (6-foszfoglukonát) oxidálódik, miközben ekvimoláris mennyiségű NADP (nikotinamid-adenin-dinukleotid-foszfát) NADPH-vá redukálódik a G-6P-DH (glukóz-6-foszfát dehidrogenáz) által katalizált reakcióban. A 340 nm-en mért abszorbancia NADPH-képződés következtében létrejött növekedése arányos a minta kreatin kináz aktivitásával (3). Minta 4 óráig 20…25 °C-on, 1 hónapig 4…8 °C-on, illetve 1 hónapig -20 °C-on, fénytől védett helyen CK-BB nem stabil. A VIZSGÁLATI ELJÁRÁS Az automatizált eljárás kivitelezéséhez olvassa el az analizátorához tartozó Alkalmazási tudnivalókat. Nem garantálható semmilyen olyan alkalmazás eredménye, amelyet a Thermo Fisher Scientific Oy nem hagyott jóvá, ezért ezeket a felhasználónak kell értékelnie. Indiko készülék beállításai Adja meg az alkalmazási paramétereket az Indiko alkalmazási lapon lévő vonalkód, illetve elektronikus fájl alapján. Szolgáltatott anyagok A fent leírt reagensek. Szükséges, de nem szolgáltatott anyagok Az alább leírt kalibrátorok és kontrollok. Kalibrálás A válasz (dA/min) egy számítási tényező segítségével eredményegységekre átszámítva jelenik meg. Ettől eltérően használhatja az eCal (981830) vegyületet kalibrálásra az eCal (981830) csomagjában található tájékoztató szerint. HK ATP + glukóz → ADP + G-6-P Minőség-ellenőrzés Használjon minőségellenőrző (QC) mintákat naponta legalább egyszer, minden kalibrálás után, illetve minden új üveg reagens felnyitásakor. Legalább két különböző szintű (alacsony és magas) kontroll vagy mintapool alkalmazása javasolt. Mindig tartsa be a minőség-ellenőrzésre vonatkozó helyi, állami és szövetségi szabályozásokat. Rendelkezésre álló kontrollminták: CK-MB Control, kód 981908 CK-MB Control High, kód 981255 Nortrol, kód 981043 Abtrol, kód: 981044 Az ellenőrzési intervallumokat (QC) és határértékeket az aktuális laboratóriumi követelményekhez kell igazítani. A minőségellenőrző minták eredményeinek a laboratórium által előre beállított határértékek közé kell esniük. G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ AZ EREDMÉNYEK KISZÁMÍTÁSA Az eredményeket az analizátor automatikusan kiszámítja a következők szerint: A REAGENSEK ADATAI Aktivitás U/l = faktor x dA/min CK Kreatin-foszfát + ADP → Kreatin + ATP A 981828 kódszámú készlet a következőket tartalmazza 2 darab 52 ml A reagenst tartalmazó üvegcse 2 darab 13 ml B reagenst tartalmazó üvegcse A 981829 kódszámú készlet a következőket tartalmazza 2 darab 20 ml A reagenst tartalmazó üvegcse 2 darab 5 ml B reagenst tartalmazó üvegcse Koncentrációk A reagens Puffer (imidazol), pH 6.5 D-glukóz N-acetil-L-cisztein (NAC) Magnézium-acetát EDTA-Na2 Nikotinamid-adenin-dinukleotid-foszfát (NADP) Hexokináz (HK) NaN3 B reagens Puffer (imidazol) Adenozin-5’-difoszfát (ADP) Adenozin-5’-monofoszfát (AMP) Di-(adenozin-5’)-pentafoszfát (AP5A) Glukóz-6-foszfát dehidrogenáz (G-6-P-DH) EDTA -Na2 Kreatin-foszfát NaN3 TV x 1000 Elméleti faktor = 6.22 x SV x P TV SV 6.22 P Teljes reagáló térfogat ml-ben Minta térfogata ml-ben A NADPH extinkciós együtthatója 340 nm-en Küvetta úthossza = 1 Átváltási faktor µkat/l = (U/l)/60 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1% eCal (981830) alkalmazásával történő kalibráláskor az eredményeket a analizátor egy kalibrációs görbe segítségével automatikusan kiszámítja. Kalibrációs görbe (példa) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1% Óvintézkedések A B reagens veszélyes. Tekintse meg a készlet belsejében található különálló lapot a veszélyes anyagokra- és óvintézkedésekre vonatkozó mondatokért: H315, P280, P302+P352, P305+P351+P338. A reagens tartósítószerként nátrium-azidot tartalmaz. A kalibrációs görbe cikkszámtól függ. Tartsa be a laboratóriumi reagensek kezelésére vonatkozó szokásos előírásokat. A terméket mint laboratóriumi vegyszert a helyi szabályozásoknak megfelelően kell kidobni. Elkészítés Az A és B reagens használatra kész. Megjegyzés: Ellenőrizze, hogy a reagensfiolák vagy -edények analizátorba helyezésénél nincs-e buborék az üveg nyakánál vagy a reagens felszínén. Tárolás és stabilitás A felbontatlan üvegcsében tárolt reagensek 2...8 °C között a címkén feltüntetett lejárati időpontig stabilak. A reagensek elemzőkészüléken belüli stabilitásához olvassa el az analizátorhoz tartozó Alkalmazási tudnivalókat. Amennyiben a reagens zavaros, ne használja. MINTAVÉTEL Megjegyzés: A minták mintavételi csövekben történő feldolgozásakor kövesse pontosan a kémcső gyártójának utasításait a hibás eredmények elkerülése érdekében. Szenteljen külön figyelmet az elemzés előtti változóknak, mint amilyen a keverés, a centrifugálás előtti állási idő, valamint a centrifugabeállítások. A minta típusa Használjon hemolizálatlan szérumot vagy heparinos plazmát. Megjegyzés: A hemolízis hibás eredményekhez vezethet. Tartsa távol a mintákat erős fénytől. Óvintézkedések Az emberi mintákat úgy kell kezelni és kidobni, mintha fertőzőek lennének. AZ ELJÁRÁS KORLÁTAI A diagnózisalkotáskor az eredményeket mindig a beteg kórelőzményével, az klinikai vizsgálatokkal és a többi lelettel együtt kell értelmezni. Interferencia Vizsgálatot végeztek a CLSI EP7-A2 dokumentumának előírásai alapján. Feltétel: Az eltérés a kezdeti értékek ± 10%-án belül van. Konelab Interferáló anyag Konjugált bilirubin Nem konjugált bilirubin Hemoglobin a hemolizátumban Lipémia: Intralipid (a Fresenius Kabi AB védjegye) Interferáló anyag koncentrációja Cél CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Eltérés a célértéktől (%) -5.3 -5.8 -1.7 -2.9 +8.5 +4.6 10 g/l 10 g/l 181 504 +2.8 +7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Tárolás (4) A minta az alább látható ideig tárolható. D06025_04_Insert_CK (IFCC)_MU Indiko Interferáló anyag Konjugált bilirubin Nem konjugált bilirubin Hemoglobin a hemolizátumban Lipémia: Intralipid (a Fresenius Kabi AB védjegye) Interferáló anyag koncentrációja Cél CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Eltérés a célértéktől (%) -0.8 -2.4 +0.2 -0.1 +8.6 +3.4 10 g/l 10 g/l 185 505 +1.0 -0.8 Kerülje hemolizált minták használatát. Gyenge összefüggés van a zavarosság és trigliceridkoncentráció között. Igen ritka esetekben a gammopátiák, különösen az IgM típusúak (Waldenström macroglobulinemia), megbízhatatlan eredményekhez vezethetnek (6). Nem történt vizsgálat a Thermo Fisher Scientific Oy vállalatnál. Egyéb interferenciát okozó anyagok tekintetében lásd az (5) referenciát. BIBLIOGRÁFIA 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. VÁRHATÓ ÉRTÉKEK (3) Férfi < 171 U/l (< 2.85 µkat/l) 37 °C hőmérsékleten Nő < 145 U/l (< 2.41 µkat/l) 37 °C hőmérsékleten A feltüntetett értékek kizárólag tájékoztató jellegűek. Minden laboratórium esetében ajánlott a normál tartomány megerősítése, illetve a kiszolgált populációra jellemző referenciaintervallum kialakítása. GYÁRTÓ Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Átváltási tényező µkat/l = (U/l)/60 Ellenőrzés időpontja (éééé-hh-nn) 2015-05-08 MÉRÉSI TARTOMÁNY 10 – 1000 U/l (0.17 – 17 µkat/l). Másodlagos hígítás utáni kiterjesztett mérési tartomány: 10 – 9000 U/l (0.17 – 150 µkat/l). Változtatások az előző változathoz képest A szöveg melletti vonal a változásokat jelöli. TELJESÍTMÉNYJELLEMZŐK Az egyes laboratóriumokban nyert értékek különbözhetnek a megadott teljesítményadatoktól. Detektálási határ Vizsgálatot végeztek az EP17-A dokumentum előírásai alapján. Konelab LoB: 0.6 U/l (0.01 µkat/l) A vak határérték (Limit of Blank, LoB) azt a legnagyobb mérési eredményt képviseli, melynek mérése a legvalószínűbb egy analitmentes mintában (n = 60). LoD: 1.5 U/l (0.03 µkat/l) A detektálási határérték (Limit of Detection, LoD) azt a mintában lévő legkisebb analitmennyiséget jelenti, amely még detektálható (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) A kvantálási határérték (Limit Of Quantitation, LoQ) azt a legalacsonyabb aktuális mintakoncentrációt jelenti, amely kvantitatívan meghatározható (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) A vak határérték (Limit of Blank, LoB) azt a legnagyobb mérési eredményt képviseli, melynek mérése a legvalószínűbb egy analitmentes mintában (n = 60). LoD: 2.2 U/l (0.04 µkat/l) A detektálási határérték (Limit of Detection, LoD) azt a mintában lévő legkisebb analitmennyiséget jelenti, amely még detektálható (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) A kvantálási határérték (Limit Of Quantitation, LoQ) azt a legalacsonyabb aktuális mintakoncentrációt jelenti, amely kvantitatívan meghatározható (n = 42). Pontatlanság A CK assay pontatlansága ≤ 5.0% teljes CV. Konelab CK Ismételhetőség (Futtatás alatt) Eszközön belül (összes) Középérték 92 U/l SD CV% 1.7 1.8 Középérték 226 U/l SD CV% 2.0 0.9 Középérték 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 A pontosságot a CLSI EP5-A2 dokumentumának előírásai alapján vizsgáltuk 21 napon keresztül egy helyszínen három Konelab 60 analizátorral, egy kezelővel, két reagenstétellel, egy kalibrátortétellel és hat kalibrálással, n = 84-es mérésszámmal. Az egyes analizátorokból származó eredményeket egyesítettük. Indiko CK Ismételhetőség (Futtatás alatt) Eszközön belül (összes) Középérték 95 U/l SD CV% 1.2 1.2 Középérték 226 U/l SD CV% 2.0 0.9 Középérték 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 A CLSI EP5-A2 dokumentumának előírásai alapján pontossági mérést végeztek 21 napon keresztül egy helyszínen három Indiko analizátorral, egy felhasználóval, 2 reagenstétellel, egy kalibrátortétellel és 7 kalibrációval, n = 84-es mérésszámmal. Az egyes analizátorokból származó eredményeket egyesítettük. Módszer-összehasonlítás Konelab A CLSI EP9-A2 dokumentumának figyelembe vételével összehasonlító vizsgálat készült Konelab Prime 60 analizátoron, a korábbi 981372 CK-MB módszer (x) referenciaként való felhasználásával. Lineáris regresszió (Deming) (eredmény U/l egységben): y = 1.07 x – 6.3 r = 0.999 n = 95 A mintakoncentrációk 17 és 2253 U/l között voltak. Indiko 1) Összehasonlító vizsgálat készült Indiko analizátoron a CLSI EP9-A2 dokumentumának figyelembe vételével és a korábbi, Konelab Prime 60 (x) analizátoron végzett 981372 CK(IFCC) módszer referenciaként való felhasználásával. Lineáris regresszió (Deming) (eredmény U/l egységben): y = 1.10 x – 4.7 r = 0.999 n = 95 A mintakoncentrációk 17 és 2253 U/l között voltak. 2) Összehasonlító vizsgálat készült Indiko analizátoron a CLSI EP9-A2 dokumentumának figyelembe vételével és ugyanannak a Konelab Prime 60 (x) analizátoron végzett CK(IFCC) módszernek a referenciaként való felhasználásával. Lineáris regresszió (Deming) (eredmény U/l egységben): y = 1.03 x + 1.4 r = 1.00 n = 98 A mintakoncentrációk 14 és 2450 U/l között voltak. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 16 Page 17 D06025_04_Insert_CK (IFCC)_MU ID CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) Penyimpanan (4) Sampel bisa disimpan seperti ditunjukkan di bawah ini. Sampel 4 hari pada suhu 20…25 °C 1 bulan pada suhu 4…8 °C atau 1 bulan pada suhu -20 °C; terlindung dari cahaya. CK-BB tidak stabil. SISIPAN PAKET INI DIIZINKAN UNTUK DIGUNAKAN DI LUAR AMERIKA SERIKAT. TUJUAN PENGGUNAAN Untuk penggunaan diagnostik in vitro dalam penetapan kuantitatif kreatin kinase (ATP: Aktivitas Kreatin N-Fosfotransferase (CK), EC 2.7.3.2) dalam serum atau plasma manusia pada penganalisis Konelab dan Indiko. Semua acuan ke sistem Konelab juga mengacu ke Seri T. RINGKASAN (1,2) Aktivitas kreatin kinase paling besar di dalam otot lurik, otak, dan jaringan jantung. Jaringan lainnya, seperti ginjal, diafragma, dan prostat, memiliki aktivitas yang jauh lebih sedikit. Aktivitas CK meningkat dalam banyak penyakit, termasuk penyakit yang menyangkut otot rangka, jantung, sistem saraf pusat, dan tiroid. Setelah penyakit otot akut, total aktivitas CK dalam darah mulai naik, mencapai level patologis setelah rata-rata durasi 4 - 5 jam pada 50% pasien. Nilainya meningkat secara teratur mulai jam ke-8 - ke-24 setelah mulainya infark dan kembali ke kisaran referensi dengan fluktuasi yang besar antar-individu. PRINSIP PROSEDUR Asai CK merupakan asai cairan dua-bagian yang memanfaatkan tiga enzim untuk menentukan aktivitas CK. Kreatin kinase mengkatalisis reaksi antara kreatin fosfat dan ADP (adenosin-5'-difosfat) dengan pembentukan kreatin dan ATP (adenosin-5'-trifosfat). Senyawa yang belakangan ini memfosforilasi glukosa menjadi G-6-P (glukosa-6-fosfat) dengan adanya HK (heksokinase). G-6P dioksidasi menjadi 6-PG (glukonat-6-fosfat) dan secara bersamaan sejumlah ekuimolar NADP (nikotinamida adenin dinukleotida fosfat) direduksi menjadi NADPH dalam reaksi yang dikatalisis oleh G-6P-DH (glukosa-6-fosfat dehidrogenase). Kenaikan absorbansi pada 340 nm, akibat pembentukan NADPH, adalah proporsional dengan aktivitas kreatin kinase dalam sampel (3). CK Kreatin fosfat + ADP → Kreatin + ATP HK ATP + Glukosa → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ INFORMASI REAGEN Kode kit 981828 mengandung 2 vial yang masing-masing berisi 52 ml Reagen A 2 vial yang masing-masing berisi 13 ml Reagen B Kode kit 981829 mengandung 2 vial yang masing-masing berisi 20 ml Reagen A 2 vial yang masing-masing berisi 5 ml Reagen B Konsentrasi Reagen A Buffer (Imidazole), pH 6,5 D-glukosa N-asetil-L-sistein (NAC) Magnesium asetat EDTA-Na2 Nikotinamida adenin dinukleotida fosfat (NADP) Heksokinase (HK) NaN3 Reagen B Buffer (Imidazole) Adenosin-5’-difosfat (ADP) Adenosin-5’-monofosfat (AMP) Di(adenosin-5’)penta-fosfat (AP5A) Glukosa-6-fosfat dehidrogenase (G-6-P-DH) EDTA -Na2 Kreatin fosfat NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2,7 mmol/l ≥ 5000 U/l < 0.1 % PROSEDUR PENGUJIAN Lihat Catatan Aplikasi untuk mengetahui prosedur otomatis pada penganalisis Anda. Pengaplikasian yang belum divalidasi Thermo Fisher Scientific Oy tidak menjamin kinerjanya dan harus dievaluasi oleh pengguna. Pengaturan instrumen Indiko Masukkan parameter pengaplikasian via barcode yang terdapat pada lembar aplikasi Indiko atau melalui file elektronik yang sesuai. Bahan yang disediakan Reagen seperti dipaparkan di atas. Bahan yang diperlukan tetapi tidak disediakan Kalibrator dan kontrol seperti disebutkan di bawah. Kalibrasi Respons (dA/mnt) dikonversi ke satuan hasil oleh faktor penghitungan. Atau gunakan eCal (981830) sebagai kalibrator menurut petunjuk yang tercantum pada sisipan kemasan eCal (981830). Kontrol Kualitas Gunakan sampel kontrol kualitas (QC) setidaknya sehari sekali dan setelah setiap kalibrasi serta setiap kali botol reagen baru digunakan. Sebaiknya gunakan setidaknya dua level kontrol (rendah dan tinggi) atau pool sampel. Selalu patuhi peraturan lokal, provinsi, dan pemerintah pusat dalam melakukan QC. Kontrol yang tersedia: Kontrol CK-MB, kode 981908 Kontrol Tinggi CK-MB, kode 981255 Nortrol, kode 981043 Abtrol, kode 981044 Interval dan batas kontrol (QC) harus disesuaikan dengan persyaratan masing-masing laboratorium. Hasil sampel kendali mutu harus masuk dalam batas yang telah ditetapkan oleh laboratorium. PENGHITUNGAN HASIL Hasilnya dihitung secara otomatis oleh penganalisis sebagai berikut: Aktivitas dalam U/l = Faktor x dA/mnt TV x 1000 Faktor teoretis = 6,22 x SV x P TV SV 6,22 P Volume reaksi total dalam ml Volume sampel dalam ml Koefisien ekstingsi NADPH pada 340 nm Panjang lintas cahaya dalam kuvet = 1 Faktor konversi µkat/l = (U/l)/60 Jika eCal (981830) digunakan sebagai kalibrator, hasilnya dihitung secara otomatis oleh penganalisis menggunakan kurva kalibrasi. Kurva Kalibrasi (contoh) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Tindakan Pencegahan Reagen B berbahaya. Lihat lembar terpisah di dalam kit untuk penjelasan tentang Berbahaya dan Pencegahan: H315, P280, P302+P352, P305+P351+P338. Reagen ini mengandung natrium azida sebagai pengawet. Lakukan tindakan pencegahan normal yang diperlukan untuk menangani semua reagen laboratorium. Produk harus dibuang sebagai bahan kimia laboratorium sesuai dengan peraturan setempat. Persiapan Reagen A dan B siap untuk digunakan. Catatan: Pastikan tidak ada gelembung di leher botol atau pada permukaan reagen saat Anda memasukkan vial atau bejana reagen di penganalisis. Penyimpanan dan Stabilitas Reagen dalam ampul yang tidak terbuka stabil pada suhu 2…8 °C sampai tanggal kedaluwarsa yang tercetak pada label. Lihat Catatan Aplikasi penganalisis Anda untuk mengetahui stabilitas reagen dalam kondisi baik. Jika reagen menunjukkan turbiditas (kekeruhan), jangan gunakan. PENGUMPULAN SPESIMEN Catatan: Saat memproses sampel di tabung pengumpulan sampel, patuhi petunjuk produsen tabung secara cermat untuk menghindari hasil yang keliru. Berikan perhatian khusus terhadap variabel praanalisis seperti waktu pencampuran dan pendiaman sebelum pengaturan sentrifugasi dan sentrifus. Jenis Sampel Serum tidak dihemolisis atau plasma heparin bisa digunakan. Catatan: Hemolisis dapat menyebabkan hasil yang menyimpang. Hindari keterpaparan sampel terhadap cahaya kuat. Kurva kalibrasi bergantung pada lot. BATASAN PROSEDUR Untuk keperluan diagnostik, hasilnya harus selalu dinilai beserta riwayat medis pasien, pemeriksaan klinis, dan temuan lain. Interferensi Penelitian dilakukan menggunakan dokumen CLSI EP7-A2 sebagai panduan. Kriteria: Perbedaan dalam ± 10% nilai awal. Konelab Zat Pengganggu Bilirubin terkonjugasi Bilirubin tak terkonjugasi Hemoglobin dalam hemolisat Lipemia: Intralipid (merek dagang dari Fresenius Kabi AB) Konsentrasi Gangguan 592 µmol/l (34,6 mg/dl) 601 µmol/l (35,2 mg/dl) 621 µmol/l (36,3 mg/dl) 677 µmol/l (39,6 mg/dl) 2,1 g/l 2,4 g/l Target CK (U/l) 188 498 180 495 184 505 Selisih dari Target (%) - 5,3 - 5,8 - 1,7 - 2,9 + 8,5 + 4,6 10 g/l 10 g/l 181 504 + 2,8 + 7,1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Tindakan Pencegahan Sampel manusia harus ditangani dan dibuang seakan-akan berpotensi menular. D06025_04_Insert_CK (IFCC)_MU Indiko Zat Pengganggu Konsentrasi Gangguan Bilirubin terkonjugasi Bilirubin tak terkonjugasi Hemoglobin dalam hemolisat Lipemia: Intralipid (merek dagang dari Fresenius Kabi AB) 592 µmol/l (34,6 mg/dl) 601 µmol/l (35,2 mg/dl) 621 µmol/l (36,3 mg/dl) 676 µmol/l (39,5 mg/dl) 2,5 g/l 2,4 g/l Target CK (U/l) 179 477 175 469 187 502 Selisih dari Target (%) - 0,8 - 2,4 +0,2 - 0,1 + 8,6 + 3,4 10 g/l 10 g/l 185 505 + 1,0 - 0,8 Hindari menggunakan spesimen terhemolisis. Korelasi antara turbiditas dan konsentrasi trigliserida buruk. Dalam kasus gamopati yang sangat jarang, terutama jenis IgM (makroglobulinemia Waldenström) tertentu, bisa menyebabkan hasil yang tidak andal (6). Tidak diuji pada Thermo Fisher Scientific Oy. Untuk zat penginterferensi lainnya, harap baca referensi (5). PERKIRAAN NILAI (3) Laki-laki < 171 U/l (< 2,85 µkat/l) pada suhu 37 °C Perempuan < 145 U/l (< 2,41 µkat/l) pada suhu 37 °C Nilai yang dikutip harus difungsikan sebagai panduan saja. Masing-masing laboratorium sebaiknya memverifikasi kisaran ini atau memperoleh interval referensi untuk populasi yang dilayani. Faktor konversi µkat/l = (U/l)/60 KISARAN PENGUKURAN 10 – 1000 U/l (0,17 – 17 µkat/l). Kisaran pengukuran yang diperpanjang setelah pengenceran sekunder: 10 – 9000 U/l (0,17 – 150 µkat/l). KARAKTERISTIK KINERJA Hasil yang diperoleh dari masing-masing laboratorium mungkin berbeda dengan data kinerja yang diberikan. Batas Deteksi Penelitian dilakukan menggunakan dokumen EP17-A sebagai panduan. Konelab LoB: 0,6 U/l (0,01 µkat/l) Limit of Blank (LoB) merupakan hasil pengukuran paling tinggi yang mungkin teramati untuk sampel bebas-analit (n = 60). LoD: 1,5 U/l (0,03 µkat/l) Limit of Detection (LoD) merupakan jumlah analit paling rendah dalam sampel yang dapat dideteksi (n = 60). LoQ: 3,2 U/l (0,05 µkat/l) Limit of Quantitation (LoQ) merupakan konsentrasi aktual paling rendah di sampel yang dapat ditentukan secara kuantitatif (n = 41). Indiko LoB: 0,2 U/l (0,003 µkat/l) Limit of Blank (LoB) merupakan hasil pengukuran paling tinggi yang mungkin teramati untuk sampel bebas-analit (n = 60). LoD: 2,2 U/l (0,04 µkat/l) Limit of Detection (LoD) merupakan jumlah analit paling rendah dalam sampel yang dapat dideteksi (n = 60). LoQ: 3,8 U/l (0,06 µkat/l) Limit of Quantitation (LoQ) merupakan konsentrasi aktual paling rendah di sampel yang dapat ditentukan secara kuantitatif (n = 42). Impresisi Ketidaktepatan pemeriksaan kadar CK adalah ≤ 5,0% Total CV. Konelab CK Keterulangan (Dalam siklus) Dalam perangkat (Total) Rata-rata 92 U/l SD CV% 1,7 1,8 Rata-rata 226 U/l SD CV% 2,0 0,9 Rata-rata 616 U/l SD CV% 7,1 1,2 2,5 3,1 22,3 2,8 1,4 3,6 Penelitian presisi dilakukan menggunakan Dokumen CLSI EP5-A2 sebagai panduan di satu lokasi dengan tiga penganalisis Konelab 60, satu operator, 2 lot reagen, satu lot kalibrator, dan 6 kalibrasi selama 21 hari, dengan jumlah pengukuran n = 84. Hasil dari masing-masing penganalisis digabungkan. Indiko CK Keterulangan (Dalam siklus) Dalam perangkat (Total) Rata-rata 95 U/l SD CV% 1,2 1,2 Rata-rata 226 U/l SD CV% 2,0 0,9 Rata-rata 637 U/l SD CV% 4,1 0,6 1,7 2,8 11,7 1,7 1,2 1,8 Penelitian presisi dilakukan menggunakan Dokumen CLSI EP5-A2 sebagai panduan di satu lokasi dengan tiga penganalisis Indiko, satu operator, 2 lot reagen, satu lot kalibrator, dan 7 kalibrasi selama 21 hari, dengan jumlah pengukuran n = 84. Hasil dari masing-masing penganalisis digabungkan. Perbandingan Metode Konelab Penelitian perbandingan dilakukan atas Konelab Prime 60 menggunakan Dokumen CLSI EP9A2 sebagai panduan dan metode 981372 CK(IFCC) sebelumnya (x) sebagai referensi. Regresi linear (Deming) (satuan hasil U/I): y = 1,07 x + 6,3 r = 0,999 n = 95 Konsentrasi sampelnya antara 17 dan 2253 U/l. Indiko 1.) Penelitian perbandingan dilakukan atas Indiko menggunakan Dokumen CLSI EP9-A2 sebagai panduan dan metode 981372 CK(IFCC) atas Konelab Prime 60 (x) sebagai referensi. Regresi linear (Deming) (satuan hasil U/I): y = 1,10 x + 4,7 r = 0,999 n = 95 Konsentrasi sampelnya antara 17 dan 2253 U/l. 2.) Penelitian perbandingan dilakukan atas Indiko menggunakan Dokumen CLSI EP9-A2 sebagai panduan dan metode CK(IFCC) yang sama atas Konelab Prime 60 (x) sebagai referensi. Regresi linear (Deming) (satuan hasil U/I): y = 1,03 x + 1,4 r = 1,00 n = 98 Konsentrasi sampelnya antara 14 dan 2450 U/l. BIBLIOGRAFI 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. PRODUSEN Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Telp. +358 10 329 200 www.thermoscientific.com Tanggal revisi (yyyy-mm-dd) 2015-05-08 Perubahan dari versi sebelumnya Garis di sebelah teks menunjukkan perubahan. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 18 Page 19 D06025_04_Insert_CK (IFCC)_MU IT CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) Conservazione (4) I campioni possono essere conservati come specificato di seguito. Campione IL PRESENTE FOGLIETTO ILLUSTRATIVO È APPLICABILE AL DI FUORI DEGLI STATI UNITI. USO PREVISTO Per uso diagnostico in vitro nella determinazione quantitativa della creatinchinasi (ATP: creatina N-fosfotransferasi (CK), EC 2.7.3.2) nel siero o nel plasma umano su analizzatori Konelab e Indiko. Eventuali riferimenti ai sistemi Konelab fanno riferimento anche alla serie T. SOMMARIO (1,2) L'attività della creatinchinasi è più elevata nella muscolatura striata, in quella cardiaca e nel cervello. In altri tessuti come i reni, il diaframma e la prostata, l'attività di questa sostanza è significativamente inferiore. Livelli elevati di CK si riscontrano in caso di patologie dei muscoli scheletrici, del cuore, del sistema nervoso centrale e della tiroide. Conseguentemente a danno muscolare acuto, l'attività della CK totale nel sangue inizia a salire, fino a raggiungere livelli patologici in media dopo 4-5 ore, nel 50% dei pazienti. I livelli si innalzano regolarmente entro 8-24 ore dall'insorgenza dell'infarto miocardio e ritornano all'intervallo di riferimento con elevate fluttuazioni interindividuali. PRINCIPIO DELLA PROCEDURA Per la CK si effettua un dosaggio liquido in due parti che utilizza tre enzimi per determinarne l'attività. La creatinchinasi catalizza la reazione tra creatinfosfato e ADP (adenosin-5'-difosfato) con formazione di creatina e ATP (adenosin-5'-trifosfato). In presenza di esochinasi (HK), il glucosio viene fosforilato dall'ATP prodotto nella reazione primaria formando G-6-P (glucosio-6-fosfato). Il G-6-P viene infine ossidato in 6-PG (gluconato-6-fosfato) e contemporaneamente una quantità equimolare di NADP (nicotinamide adenin dinucleotide fosfato) viene ridotta a NADPH in una reazione catalizzata da G-6P-DH (glucosio-6-fosfato deidrogenasi). L'aumento dell'assorbanza a 340 nm, dovuta alla formazione di NADPH, è proporzionale all'attività della creatinchinasi nel campione esaminato (3). CK Creatinfosfato + ADP → Creatina + ATP HK ATP + Glucosio → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ INFORMAZIONI SUI REAGENTI Il kit con codice 981828 contiene 2 fiale da 52 ml di Reagente A 2 fiale da 13 ml di Reagente B Il kit con codice 981829 contiene 2 fiale da 20 ml di Reagente A 2 fiale da 5 ml di Reagente B Concentrazioni Reagente A Tampone (imidazolo), pH 6.5 D-glucosio N-acetil-L-cisteina (NAC) Magnesio acetato EDTA-Na2 Nicotinamide adenin dinucleotide fosfato (NADP) Esochinasi (HK) NaN3 Reagente B Tampone (imidazolo) Adenosina-5'-difosfato (ADP) Adenosina-5'-monofosfato (AMP) Di(adenosin-5')penta-fosfato (AP5A) Glucosio-6-fosfato deidrogenasi (G-6-P-DH) EDTA -Na2 Creatinfosfato NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l <0.1% 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l <0.1% 4 ore a 20…25°C, 1 mese a 4…8°C o 1 mese a -20°C al riparo dalla luce CK-BB non stabile PROCEDURA ANALITICA Per le procedure automatiche sull'analizzatore in uso, consultare le note applicative. Tutte le applicazioni non esplicitamente approvate da Thermo Fisher Scientific Oy non possono essere garantite in termini di prestazioni e dovranno pertanto essere valutate dall'utente. Impostazioni dello strumento Indiko Inserire i parametri dell'applicazione utilizzando il codice a barre presente nella scheda dell'applicazione Indiko o tramite il file elettronico, in base alle esigenze. Materiali inclusi nel kit I reagenti descritti in precedenza. Materiali necessari ma non inclusi nel kit Calibratore e controlli indicati di seguito. Calibrazione La risposta (dA/min) è convertita nelle unità di misura del risultato per mezzo di un fattore di calcolo. In alternativa utilizzare eCal (981830) come calibratore in base alle istruzioni riportate nel foglietto illustrativo accluso alla confezione di eCal (981830). Controllo qualità Utilizzare campioni di controllo qualità (CQ) almeno una volta al giorno, dopo ogni calibrazione e ogni volta che si utilizza un nuovo flacone di reagente. È consigliabile utilizzare almeno due livelli (basso e alto) di controlli o pool di campioni. Osservare sempre le normative locali, nazionali e federali applicabili quando si esegue il controllo qualità (CQ). Controlli disponibili: CK-MB Control, codice 981908 CK-MB Control High, codice 981255 Nortrol, codice 981043 Abtrol, codice 981044 Gli intervalli e i limiti del controllo (QC) devono essere adattati ai requisiti dei singoli laboratori. I risultati dei campioni di controllo qualità devono rientrare nei limiti di variabilità stabiliti dal laboratorio. CALCOLO DEI RISULTATI I risultati vengono calcolati automaticamente dall'analizzatore come segue: Attività in U/l = Fattore x dA/min TV x 1000 Fattore teorico = 6.22 x SV x P TV SV 6.22 P Volume di reazione totale in ml Volume del campione in ml Coefficiente di estinzione del NADPH a 340 nm Tracciato della cuvetta = 1 Fattore di conversione µkat/l = (U/l)/60 Se come calibratore viene utilizzato eCal (981830), i risultati vengono calcolati automaticamente dall'analizzatore in base a una curva di calibrazione. Curva di calibrazione (esempio) Precauzioni Il reagente B è pericoloso. Vedere il foglio separato fornito nel kit per le frasi H (indicazioni di pericolo) e le frasi P (precauzioni): H315, P280, P302+P352, P305+P351+P338. I reagenti contengono azoturo di sodio come conservante. Rispettare le normali precauzioni previste per l'utilizzo di tutti i reagenti di laboratorio. Il prodotto deve essere smaltito come sostanza chimica da laboratorio conformemente alle disposizioni locali. Preparazione I reagenti A e B sono pronti all'uso. Nota: verificare che non siano presenti bolle nel collo del flacone o sulla superficie del reagente quando si inseriscono fiale o recipienti di reagente nell'analizzatore. Conservazione e stabilità I reagenti in fiale intatte sono stabili fino alla data di scadenza indicata sull'etichetta, se conservati a una temperatura di 2…8°C. Per la stabilità dei reagenti nello strumento, fare riferimento alle Note applicative dell'analizzatore in uso. Se i reagenti presentano torbidità non utilizzarli. La curva di calibrazione dipende dal lotto. LIMITI DELLA PROCEDURA Per scopi diagnostici, i risultati devono sempre essere valutati congiuntamente a un'anamnesi medica del paziente, agli esami clinici e ad altri reperti. Interferenze È stato eseguito uno studio utilizzando il documento EP7-A2 del CLSI come linea guida. Criterio: differenza entro ± 10% dei valori iniziali. Konelab Sostanza interferente RACCOLTA DEL CAMPIONE Nota: per l'analisi dei campioni in provette, seguire attentamente le istruzioni fornite dal produttore delle provette al fine di evitare risultati errati. Prestare particolare attenzione alle variabili preanalitiche quali la miscelazione, il tempo di attesa prima della centrifugazione e le impostazioni di centrifugazione. Bilirubina coniugata Tipo di campione È possibile utilizzare siero non emolizzato o plasma eparinizzato. Nota: l'emolisi può produrre risultati fuorvianti. Evitare di esporre i campioni alla luce diretta. Emoglobina nell'emolisato Bilirubina non coniugata Lipemia: Intralipid (marchio commerciale di Fresenius Kabi AB) Concentrazione interferente CK target (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Deviazione dal valore di riferimento (%) - 5.3 - 5.8 -1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Precauzioni Tutti i campioni umani devono essere maneggiati e smaltiti come campioni potenzialmente infetti. D06025_04_Insert_CK (IFCC)_MU Indiko Sostanza interferente Bilirubina coniugata Bilirubina non coniugata Emoglobina nell'emolisato Lipemia: Intralipid (marchio commerciale di Fresenius Kabi AB) Concentrazione interferente CK target (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Deviazione dal valore di riferimento (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Evitare l'uso di campioni emolizzati. Vi è una scarsa correlazione fra torbidità e concentrazione dei trigliceridi. In rarissimi casi la presenza di gammopatia, in particolare il tipo IgM (macroglobulinemia di Waldenström), può determinare risultati inaffidabili (6). Non testato da Thermo Fisher Scientific Oy. Per le altre sostanze interferenti, vedere la bibliografia (5). VALORI PREVISTI (3) Maschi < 171 U/l (< 2.85 µkat/l) a 37°C Femmine < 145 U/l (< 2.41 µkat/l) a 37°C I valori citati dovranno servire esclusivamente come riferimento. Si raccomanda a ogni laboratorio di verificare l'applicabilità di questo intervallo alla propria popolazione di pazienti e, se necessario, di determinare intervalli di riferimento propri. INTERVALLO DI MISURAZIONE 10 – 1000 U/l (0.17 – 17 µkat/l). Intervallo di misurazione ampliato dopo la seconda diluizione: 10 – 9000 U/l (0.17 – 150 µkat/l). CARATTERISTICHE PRESTAZIONALI I risultati ottenuti nei singoli laboratori possono differire dai dati sulle prestazioni riportati. Limiti di rilevamento È stato eseguito uno studio utilizzando il documento EP17-A come linea guida. Konelab LoB: 0.6 U/l (0.01 µkat/l) Il LoB (Limit Of Blank) rappresenta il risultato della misurazione massima che è probabile osservare per un campione privo di analita (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Il LoD (Limit of Detection) rappresenta la concentrazione minima rilevabile di analita in un campione (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Il LoQ (Limit of Quantitation) rappresenta la concentrazione minima effettiva in un campione che è possibile determinare quantitativamente (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Il LoB (Limit Of Blank) rappresenta il risultato della misurazione massima che è probabile osservare per un campione privo di analita (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Il LoD (Limit of Detection) rappresenta la concentrazione minima rilevabile di analita in un campione (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Il LoQ (Limit of Quantitation) rappresenta la concentrazione minima effettiva in un campione che è possibile determinare quantitativamente (n = 42). Imprecisione L'imprecisione del dosaggio della CK è ≤ 5.0% del CV totale. Ripetibilità (Entro la serie) Entro il dispositivo (totale) Media 92 U/l SD CV% 1.7 1.8 Media 226 U/l SD CV% 2.0 0.9 Media 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Lo studio di precisione è stato eseguito utilizzando il documento EP5-A2 del CLSI come linea guida, in una sede con tre analizzatori Konelab 60, un operatore e due lotti di reagente, un lotto di calibratore e sei calibrazioni per un periodo di 21 giorni, per un numero di misurazioni pari a n = 84. I risultati dei singoli analizzatori sono stati combinati. Indiko CK Ripetibilità (Entro la serie) Entro il dispositivo (totale) PRODUTTORE Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finlandia Tel. +358 10 329 200 www.thermoscientific.com Data della revisione (aaaa-mm-gg) 2015-05-08 Fattore di conversione µkat/l = (U/l)/60 Konelab CK BIBLIOGRAFIA 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. Media 95 U/l SD CV% 1.2 1.2 Media 226 U/l SD CV% 2.0 0.9 Media 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Lo studio di precisione è stato eseguito utilizzando il documento EP5-A2 del CLSI come linea guida, in una sede con tre analizzatori Indiko, un operatore e due lotti di reagente, un lotto di calibratore e sette calibrazioni per un periodo di 21 giorni, per un numero di misurazioni pari a n = 84. I risultati di ciascun analizzatore sono stati combinati. Metodo di confronto Konelab È stato eseguito uno studio comparato sull'analizzatore Konelab Prime 60, utilizzando il documento EP9-A2 come linea guida e il precedente metodo 981372 CK (IFCC) (x) come riferimento. Regressione lineare (di Deming) (risultato espresso in U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Le concentrazioni dei campioni erano comprese tra 17 e 2253 U/l. Indiko 1) È stato eseguito uno studio comparato sull'analizzatore Indiko, utilizzando il documento EP9-A2 del CLSI come linea guida e il precedente metodo 981372 CK (IFCC) sull'analizzatore Konelab Prime 60 (x) come riferimento. Regressione lineare (di Deming) (risultato espresso in U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Le concentrazioni dei campioni erano comprese tra 17 e 2253 U/l. 2) È stato eseguito uno studio comparato sull'analizzatore Indiko, utilizzando il documento EP9-A2 del CLSI come linea guida e lo stesso metodo CK (IFCC) sull'analizzatore Konelab Prime 60 (x) come riferimento. Regressione lineare (di Deming) (risultato espresso in U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Le concentrazioni dei campioni erano comprese tra 14 e 2450 U/l. Modifiche rispetto alla versione precedente La linea accanto al testo indica le modifiche apportate. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 20 Page 21 D06025_04_Insert_CK (IFCC)_MU LT 981828 981829 2 x 65 ml (Konelab™ ) 2 x 25 ml (Konelab™, Indiko™) ŠIS PAKUOTĖS LAPELIS SKIRTAS NAUDOTOJAMS NE JAV TERITORIJOJE. NUMATYTOJI PASKIRTIS Skirtas in vitro diagnostikai, kiekybiškai nustatant kreatino kinazės (ATP: kreatino N-fosfotransferazės (CK), EC 2.7.3.2) aktyvumą žmogaus serume arba plazmoje, naudojant „Konelab“ ir „Indiko“ analizatorius. Nurodant bet kokias „Konelab“ sistemas taip pat nurodomi T serijos gaminiai. SANTRAUKA (1, 2) Kreatino kinazės aktyvumas yra didžiausias skersaruožiuose raumenyse, smegenyse ir širdies audinyje. Kituose audiniuose, pavyzdžiui, inkstuose, diafragmoje ir prostatoje, šis aktyvumas yra žymiai mažesnis. CK aktyvumas padidėja sergant daugeliu ligų, įskaitant tas, kurios susijusios su skeleto raumenimis, širdimi, centrine nervų sistema ir skydliauke. Po stipraus raumenų pažeidimo bendras CK aktyvumas kraujyje pradeda didėti, vidutiniškai po 4–5 valandų 50 % ligonių jis pasiekia patologinį lygį. Vertės kyla reguliariai nuo 8–24 - os valandos po infarkto priepuolio ir grįžta į atskaitos diapazoną su dideliais individualiais svyravimais. PROCEDŪROS PRINCIPAS CK tyrimas yra dviejų dalių skysčių tyrimas, per kurį CK aktyvumui nustatyti naudojami trys fermentai. Kreatino kinazė katalizuoja kreatino fosfato ir ADP (adenozin-5’-difosfato) reakciją, sudarydama kreatiną ir ATP (adenozin-5’-trifosfatą). Pastarasis junginys fosforilina gliukozę į G-6-P (gliukozės-6fosfatą) esant HK (heksokinazės). G-6-P oksiduojama į 6-PG (gliukonato-6-fosfatą), ir tuo pat metu toks pat NADP (nikotinamidadenindinukleotido fosfato) kiekis redukuojamas į NADPH reakcijoje, kurią katalizuoja G-6P-DH (gliukozės-6-fosfato dehidrogenazė). Dėl susidariusio NADPH padidėjusi absorbcijos geba 340 nm bangų ruože yra proporcinga kreatino kinazės aktyvumui mėginyje (3). CK Kreatino fosfatas + ADP → Kreatinas + ATP HK ATP + Gliukozė → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Mėginys Rinkinį, kurio kodas 981829, sudaro 2 buteliukai po 20 ml A reagento 2 buteliukai po 5 ml B reagento TYRIMO PROCEDŪRA Skaitykite analizatoriaus naudojimo nuorodas apie įrenginiu vykdomą automatinę procedūrą. Naudojant gaminį bendrovės „Thermo Fisher Scientific Oy“ nepatvirtintai paskirčiai, nurodytų techninių savybių užtikrinti neįmanoma ir tokį naudojimą turi įvertinti naudotojas. „Indiko“ prietaiso parametrai Naudojimo parametrus įveskite nuskaitydami brūkšninį kodą, esantį „Indiko“ naudojimo duomenų lape, arba iš elektroninio failo (kaip tinkama). Teikiamos medžiagos Anksčiau aprašyti reagentai. Būtinos, bet neteikiamos medžiagos Toliau nurodytas kalibratorius ir kontrolės medžiagos. Kalibratorius Atsako rodiklis (dA/min) konvertuojamas į rezultato pateikimo vienetus pritaikius perskaičiavimo koeficientą. Arba naudokite „eCal“ (981830) kaip kalibratorių pagal „eCal“ (981830) pakuotės lapelyje pateiktas instrukcijas. Kokybės kontrolė Kokybės kontrolės (angl. QC) mėginius tirkite ne mažiau nei vieną kartą per dieną ir po kiekvieno kalibravimo, taip pat kaskart naudodami naują reagento butelį. Rekomenduojama naudoti ne mažiau nei dviejų lygių (didelės ir mažos koncentracijos) kontrolės medžiagas arba jungtinius mėginius. Atlikdami kokybės kontrolę, visuomet vadovaukitės vietos, valstijos ir federaliniais reglamentais. Kontrolės medžiagos, kurias galima įsigyti: „CK-MB Control“, kodas 981908 „CK-MB Control High“, kodas 981255 „Nortrol“, kodas 981043 „Abtrol“, kodas 981044 Kontrolės (QC) intervalus ir apribojimus kiekviena laboratorija turi pritaikyti pagal savo reikalavimus. Kokybės kontrolės mėginių rezultatai turi patekti į laboratorijos iš anksto nustatytų verčių intervalą. Aktyvumas, išreikštas U/I = koeficientas x dA/min TV x 1000 Teorinis koeficientas = 6.22 x SV x P Koncentracija A reagentas Buferis (imidazolas), pH 6.5 D-gliukozė N-acetil-L-cisteinas (NAC) Magnio acetatas EDTA-Na2 Nikotinamidadenindinukleotido fosfatas (NADP) Heksokinazė (HK) NaN3 Reagentas B Buferis (imidazolas) Adenozin-5’-difosfatas (ADP) Adenozin-5’-monofosfatas (AMP) Di(adenozin-5’)penta-fosfatas (AP5A) Gliukozės-6-fosfato dehidrogenazė (G-6-P-DH) EDTA -Na2 Kreatino fosfatas NaN3 4 valandas 20…25 °C temp., 1 mėnesį 4…8 °C temp. arba 1 mėnesį -20 °C temp.; apsaugoti nuo šviesos. CK-BB nėra stabilus. REZULTATŲ SKAIČIAVIMAS Analizatorius automatiškai apskaičiuoja rezultatus taip: INFORMACIJA APIE REAGENTUS Rinkinį, kurio kodas 981828, sudaro 2 buteliukai po 52 ml A reagento 2 buteliukai po 13 ml B reagento Laikymas (4) Mėginį galima laikyti toliau nurodytomis sąlygomis. 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1 % TV SV 6.22 P Bendrasis reakcijos tūris ml Mėginio tūris ml NADPH ekstinkcijos koeficientas 340 nm bangų ruože Kiuvetės trajektorijos ilgis = 1 Konversijos koeficientas µkat/l = (U/l)/60 Jeigu naudojamas kalibratorius „eCal“ (981830), analizatorius apskaičiuoja rezultatus automatiškai pagal kalibravimo kreivę. Kalibravimo kreivė (pavyzdys) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Atsargumo priemonės B reagentas yra pavojingas. Žr. atskirą lapą rinkinyje apie pavojaus ir atsargumo priemonių frazes: H315, P280, P302+P352, P305+P351+P338. Reagentų sudėtyje yra konservanto natrio azido. Laikykitės standartinių atsargumo priemonių, būtinų dirbant su visais laboratorijos reagentais. Gaminį reikia išmesti kaip laboratorinį cheminį preparatą pagal vietinius reglamentus. Paruošimas Reagentas A ir B jau paruošti naudojimui Pastaba. Prieš dėdami reagentų buteliukus ar indus į analizatorių, patikrinkite, ar buteliuko kaklelyje arba reagento paviršiuje nėra burbuliukų. Su kiekviena partija teikiama jai tinkanti kalibravimo kreivė. Laikymas ir stabilumas Neatidarytuose buteliukuose 2…8 °C temperatūroje laikomi reagentai išlieka stabilūs iki tinkamumo datos, nurodytos etiketėje. Apie prietaise esančių reagentų stabilumą žr. analizatoriaus naudojimo nuorodose. Jei reagentai atrodo susidrumstę, jų nenaudokite. Neigiama sąveika Tyrimas atliktas vadovaujantis CLSI dokumento EP7-A2 gairėmis. Kriterijus: skirtumas nuo pradinių verčių ± 10 % diapazone. ĖMINIO PAĖMIMAS Pastaba. Apdorodami mėginius mėginių ėmimo mėgintuvėliuose, atidžiai vadovaukitės mėgintuvėlio gamintojo instrukcijomis, kad išvengtumėte klaidingų rezultatų. Būkite itin dėmesingi prieš analizę veikiantiems kintamiesiems, kaip antai maišymas, nusistovėjimo trukmė prieš centrifugavimą ir centrifugos parametrų sąranka. Mėginio tipas Galima naudoti nehemolizuotą serumą arba heparino plazmą Pastaba. Dėl hemolizės galite gauti rezultatus su dideliu nuokrypiu. Saugokite mėginius nuo stiprios šviesos. Atsargumo priemonės Su žmonių mėginiais reikia dirbti ir juos šalinti taip, lyg jie būtų potencialiai užkrečiami. PROCEDŪROS APRIBOJIMAI Diagnostikos tikslais rezultatus visada reikia vertinti atsižvelgiant į paciento anamnezę, klinikinius tyrimus ir kitas išvadas. „Konelab“ Sąveikaujanti medžiaga Konjuguotas bilirubinas Nekonjuguotas bilirubinas Hemoglobinas hemolizate Lipemija: „Intralipid“ („Fresenius Kabi AB“ prekės ženklas) Sąveiką sukelianti koncentracija Tikslinis CK kiekis (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Skirtumas nuo tikslinio kiekio (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. CK (IFCC) D06025_04_Insert_CK (IFCC)_MU „Indiko“ Sąveikaujanti medžiaga Konjuguotas bilirubinas Nekonjuguotas bilirubinas Hemoglobinas hemolizate Lipemija: „Intralipid“ („Fresenius Kabi AB“ prekės ženklas) Sąveiką sukelianti koncentracija Tikslinis CK kiekis (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Skirtumas nuo tikslinio kiekio (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Nenaudokite hemolizuotų ėminių. Drumstumo ir trigliceridų koncentracijos koreliacija labai nedidelė. Labai retais atvejais dėl gamopatijos, ypač IgM tipo (Waldenström makroglobulinemijos), galima gauti nepatikimus rezultatus (6). „Thermo Fisher Scientific Oy“ netirta. Daugiau apie galinčias sąveikauti medžiagas žr. nuorodoje (5). NUMATOMOS VERTĖS (3) Vyrai < 171 U/l (< 2.85 µkat/l) 37 °C temperatūroje Moterys < 145 U/l (< 2.41 µkat/l) 37 °C temperatūroje Nurodytos vertės gali būti traktuojamos tik kaip pavyzdys. Rekomenduojama, kad kiekviena laboratorija patikrintų šį diapazoną arba išvestų atskaitos intervalą populiacijai, kurią ji aptarnauja. Konvertavimo koeficientas µkat/l = (U/l)/60 MATAVIMO DIAPAZONAS 10 – 1000 U/l (0.17 – 17 µkat/l). Išplėstinis matavimo reikšmių diapazonas po antrojo praskiedimo: 10 – 9000 U/l (0.17 – 150 µkat/l). VEIKIMO CHARAKTERISTIKOS Atskirose laboratorijose gauti rezultatai gali skirtis nuo pateiktų veikimo duomenų. Aptikimo riba Tyrimas atliktas vadovaujantis CLSI dokumento EP17-A gairėmis. „Konelab“LoB: 0.6 U/l (0.01 µkat/l) Tuščioji riba (angl. „Limit of Blank“, LoB) reiškia didžiausią matavimo rezultatą, kurį galima nustatyti mėginyje, kuriame nėra analitės (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Aptikimo riba (angl. „Limit of Detection“, LoD) reiškia mažiausią analitės kiekį, kurį galima aptikti mėginyje (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Kiekio nustatymo riba (angl. „Limit of Quantitation“, LoQ) yra mažiausia faktinė medžiagos koncentracija mėginyje, kurios kiekį galima nustatyti (n = 41). „Indiko“ LoB: 0.2 U/l (0.003 µkat/l) Tuščioji riba (angl. „Limit of Blank“, LoB) reiškia didžiausią matavimo rezultatą, kurį galima nustatyti mėginyje, kuriame nėra analitės (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Aptikimo riba (angl. „Limit of Detection“, LoD) reiškia mažiausią analitės kiekį, kurį galima aptikti mėginyje (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Kiekio nustatymo riba (angl. „Limit of Quantitation“, LoQ) yra mažiausia faktinė medžiagos koncentracija mėginyje, kurios kiekį galima nustatyti (n = 42). Netikslumas CK tyrimo rezultatų nuokrypis yra ≤ 5.0 % bendrojo CV. „Konelab“ CK Pakartojamumas (vieno ciklo metu) Atliekant tyrimą vienu analizatoriumi (bendrasis) Vidurkis – 92 U/l SD 1.7 CV% 1.8 Vidurkis – 226 U/l SD CV% 2.0 0.9 Vidurkis – 616 U/l SD CV% 7.1 1.2 2.5 2.8 3.1 22.3 1.4 3.6 Tikslumo tyrimas buvo atliktas vadovaujantis CLSI EP5-A2 dokumento gairėmis vienoje laboratorijoje naudojant tris „Konelab 60“ analizatorius, dirbant vienam operatoriui ir naudojant 2 partijų reagentus, vienos partijos kalibratorių bei atlikus 6 kalibravimus per 21 dieną; matavimų skaičius n = 84. Kiekvienu analizatoriumi gauti rezultatai buvo sujungti. „Indiko“ CK Pakartojamumas (vieno ciklo metu) Atliekant tyrimą vienu analizatoriumi (bendrasis) Vidurkis – 95 U/l SD 1.2 CV% 1.2 Vidurkis – 226 U/l SD CV% 2.0 0.9 Vidurkis – 637 U/l SD CV% 4.1 0.6 1.7 1.7 2.8 11.7 1.2 1.8 Tikslumo tyrimas buvo atliktas vadovaujantis CLSI EP5-A2 dokumento gairėmis vienoje laboratorijoje, naudojant tris „Indiko“ analizatorius, dirbant vienam operatoriui ir naudojant 2 partijų reagentus, vienos partijos kalibratorių bei atlikus 7 kalibravimus per 21 dieną; matavimų skaičius buvo n = 84. Kiekvienu analizatoriumi gauti rezultatai buvo sujungti. Metodų palyginimas „Konelab“ Palyginamasis tyrimas buvo atliktas pagal CLSI EP9-A2 dokumento gaires su „Konelab Prime 60“, kaip etaloną naudojant ankstesnįjį 981372 CK (IFCC) metodą (x). Tiesinė regresija (pagal Demingą) (rezultatų matavimo vienetas – U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Analitės koncentracija mėginiuose buvo nuo 17 iki 2253 U/l. „Indiko“ 1) Palyginamasis tyrimas buvo atliktas pagal CLSI EP9-A2 dokumento gaires su „Indiko“, kaip etaloną naudojant ankstesnįjį 981372 CK (IFCC) metodą su „Konelab Prime 60“ (x). Tiesinė regresija (pagal Demingą) (rezultatų matavimo vienetas – U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Analitės koncentracija mėginiuose buvo nuo 17 iki 2253 U/l. 2) Palyginamasis tyrimas buvo atliktas pagal CLSI EP9-A2 dokumento gaires su „Indiko“, kaip etaloną naudojant tą patį CK (IFCC) metodą su „Konelab Prime 60“ (x). Tiesinė regresija (pagal Demingą) (rezultatų matavimo vienetas – U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Analitės koncentracija mėginiuose buvo nuo 14 iki 2450 U/l. LITERATŪRA 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. GAMINTOJAS Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Redakcijos data (MMMM-mm-dd) 2015-05-08 Pakeitimai nuo ankstesnės versijos Linija šalia teksto reiškia, kad tekstas pakeistas. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 22 Page 23 D06025_04_Insert_CK (IFCC)_MU LV 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) ŠIS IEPAKOJUMA IELIKTNIS IR LIETOJAMS ĀRPUS ASV. LIETOŠANA In vitro kvantitatīvai kreatinīna kināzes (ATP: kreatīna N-fosfortransferāzes (CK), EC 2.7.3.2) aktivitātes noteikšanai cilvēka serumā vai asinīs, izmantojot Konelab un Indiko analizatorus. Jebkura atsauce uz Konelab sistēmām ir attiecināma arī uz T sēriju. KOPSAVILKUMS (1,2) Vislielākā kreatīna kināzes ektivitāte ir šķērssvītrotajos muskuļos, smadzenēs un sirds audos. Citos audos, piemēram nierēs, diafragmā un prostatā, ir ievērojami mazāka aktivitāte. CK aktivitāte ir palielināta daudzu slimību gadījumos, tostarp to, kas saistītas ar skeleta muskuļiem, sirdi, centrālo nervu sistēmu un vairogdziedzeri. Pēc akūta muskuļu bojājuma kopējā CK aktivitāte asinīs sāk palielināties, 50 % pacientu sasniedzot patoloģiskus līmeņus aptuveni pēc 4–5 stundām. Šīs vērtības regulāri palielinās no 8. līdz 24. stundai pēc infarkta simptomu parādīšanās un atgriežas atsauces diapazonā, kas ievērojami atšķiras atkarībā no individuālām svārstībām. PROCEDŪRAS PRINCIPS CK pārbaude ir divdaļīga šķidrā pārbaude, kas izmanto trīs enzīmus CK aktivitātes noteikšanai. Kreatīna kināze katalizē reakciju starp kreatīna fosfātu un ADP (adenozīna-5-difosfātu), kā rezultātā veidojas kreatīns un ATP (adenozīna-5-trifosfāts). ATP fosforilizē glikozi par G-6-P (glikozes-6fosfātu) HK (heksokināzes) klātbūtnē. G-6-P oksidējas par 6-PG (glikonāta-6-fosfātu), un tajā pašā laikā tāds pats molu skaits NADP (nikotīnamīda adenīna dinukleotīda fosfāts) tiek reducēts par NADPH tādā reakcijā, kuru katalizē G-6-P-DH (glikozes-6-fosfāta dehidrogenāze). Absorbcijas palielinājums pie 340 nm, kas rodas no NADPH izveidošanās, ir proporcionāls kreatīna fosfokināzes aktivitātei paraugā (3). CK Kreatīna fosfāts + ADP → Kreatīns + ATP HK ATP + Glikoze → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Komplekta ar kodu 981829 saturs 2 pudelītes 20 ml reaģenta A 2 pudelītes 5 ml reaģenta B Koncentrācijas Reaģents A Buferšķīdums (Imidazole), pH 6.5 D-glikoze N-acetil-L-cisteīns (NAC) Magnija acetāts EDTA-Na2 Nikotīnamīda adenīna dinukleotīda fosfāts (NADP) Heksokināze (HK) NaN3 Reaģents B Buferšķīdums (Imidazols) Adenozīn-5’-difosfāts (ADP) Adenozīn-5’-monofosfāts (AMP) Di(adenozīn-5)pentafosfāts (AP5A) Glikozes-6-fosfāta dehidrogenāze (G-6-P-DH) EDTA -Na2 Kreatīna fosfāts NaN3 4 stundas 20…25 °C temperatūrā, 1 mēnesis 4…8 °C temperatūrā vai 1 mēnesis -20 °C temperatūrā; paraugs jāuzglabā tumsā. CK-BB nav stabils. Paraugs PĀRBAUDES PROCEDŪRA Lai iegūtu informāciju par analizatora automātisko procedūru, skatiet lietošanas norādījumus. Jebkāda veida lietošana, ko nav apstiprinājis uzņēmums “Thermo Fisher Scientific Oy”, nevar garantēt atbilstošus rezultātus, tādēļ lietotājam ir pašam jānovērtē tādas lietošanas nepieciešamība. Indiko instrumentu iestatījumi Lietošanas parametrus ievadiet, izmantojot Indiko lietošanas norādījumos pieejamo svītrkodu vai elektronisko failu. Piegādātie materiāli Iepriekš aprakstītie reaģenti. Nepieciešamie, bet nepiegādātie reaģenti Tālāk norādītais kalibrators un kontroles. Kalibrācija Reakcija (dA/min.) tiek pārveidota rezultāta vienībās, izmantojot aprēķinu koeficientu. Var arī izmantot eCal (981830) kā kalibrētāju saskaņā ar instrukcijām eCal (981830) iepakojuma ieliktnī. Kvalitātes kontrole Izmantojiet kvalitātes kontroles paraugus vismaz reizi dienā, pēc katras kalibrācijas un ikreiz, kad tiek izmantota jauna reaģenta pudelīte. Ieteicams izmantot vismaz divus kontroles vai paraugu kopu līmeņus (zemāko un augstāko). Veicot kvalitātes kontroli, vienmēr ievērojiet vietējos, valsts un pašvaldības noteikumus. Pieejamās kontroles CK-MB Control, kods 981908 CK-MB Control High, kods 981255 Nortrol, kods 981043 Abtrol, kods 981044 Kontroles intervāli un ierobežojumi ir jāpielāgo individuālajām laboratorijas prasībām. Kvalitātes kontroles paraugu rezultātiem ir jāatbilst laboratorijas iepriekš noteiktajiem ierobežojumiem. REZULTĀTU APRĒĶINĀŠANA Analizators automātiski aprēķina rezultātus šādi: INFORMĀCIJA PAR REAĢENTU Komplekta ar kodu 981828 saturs 2 pudelītes 52 ml reaģenta A 2 pudelītes 13 ml reaģenta B Uzglabāšana (4) Paraugus var glabāt, kā norādīts tālāk. 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1 % Aktivitāte U/I = faktors x dA/min TV x 1000 Teorētiskais faktors = 6.22 x SV x P TV SV 6.22 P Kopējais reakcijas apjoms ml Parauga apjoms ml NADPH zuduma koeficients pie 340 nm Kivetes ceļa garums = 1 Konversijas koeficients µkat/l = (U/l)/60 Ja kā kalibrators tiek izmantots eCal (981830), analizators rezultātus aprēķina automātiski, izmantojot kalibrēšanas līkni. Kalibrēšanas līkne (piemērs) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Piesardzības pasākumi Reaģents B ir bīstams. Informāciju par apdraudējumiem un piesardzības pasākumiem skatiet atsevišķajā komplektā iekļautajā ieliktnī: H315, P280, P302+P352, P305+P351+P338. Reaģents satur nātrija azīdu kā konservantu. Nodrošiniet parastos piesardzības pasākumus, kas jāievēro attiecībā uz rīkošanos ar visiem laboratorijas reaģentiem. Izstrādājums ir jāizmet kā laboratorijām paredzēta ķīmiskā viela saskaņā ar vietējiem noteikumiem. Sagatavošana Reaģenti A un B ir gatavi lietošanai. Piezīme. Ievietojot reaģenta pudelītes vai traukus analizatorā, pārbaudiet, vai pudelītes kaklā un uz reaģenta virsmas nav burbuļu. Uzglabāšana un stabilitāte Reaģenti neatvērtās pudelītēs ir stabili 2…8 °C temperatūrā līdz uzglabāšanas laika, kas uzdrukāts uz etiķetes, beigām. Ievietoto reaģentu stabilitāti skatīt analizatora lietošanas norādījumos. Neizmantojiet, ja reaģents ir duļķains. Kalibrēšanas līkne ir atkarīga no partijas. PROCEDŪRAS IEROBEŽOJUMI Diagnostiskos nolūkos rezultāti vienmēr jāizvērtē saistībā ar pacienta slimības vēsturi, klīniskajiem izmeklējumiem un citiem iegūtajiem datiem. Mijiedarbība Izpēte veikta, par pamatnorādi izmantojot CLSI dokumentu EP7-A2. Kritērijs: atšķirība ± 10 % robežās no sākotnējām vērtībām. Konelab Interferējošā viela PARAUGU IEGŪŠANA Piezīme. Apstrādājot paraugus savākšanas stobriņos, precīzi ievērojiet stobriņu ražotāja norādījumus, lai rezultāti nebūtu kļūdaini. Pievērsiet īpašu uzmanību pirmsanalīzes vērtībām, piemēram, sajaukšanai, noturēšanas laikam pirms centrifugēšanas un centrifugēšanas iestatījumiem. Saistīts bilirubīns Paraugu veids Var izmantot nehemolizētu serumuvai heparīna plazmu. Piezīme. Hemolīze var sniegt kļūdainus rezultātus. Nepakļaujiet paraugus pārmērīgai gaismas ietekmei. Lipēmija: Intralipid (uzņēmuma Fresenius Kabi AB preču zīme) Piesardzības pasākumi Ar paraugiem, kas iegūti no cilvēkiem ir jārīkojas un tie jālikvidē tā, it kā tie būtu iespējami infekciozi. Nesaistīts bilirubīns Hemoglobīns hemolizātā Interferences koncentrācija 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l Mērķa CK (U/l) 188 498 180 495 184 505 Atšķirība no mērķa (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. CK (IFCC) D06025_04_Insert_CK (IFCC)_MU Indiko Interferējošā viela Saistīts bilirubīns Nesaistīts bilirubīns Hemoglobīns hemolizātā Lipēmija: Intralipid (uzņēmuma Fresenius Kabi AB preču zīme) Interferences koncentrācija 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l Mērķa CK (U/l) 179 477 175 469 187 502 Atšķirība no mērķa (%) - 0.8 - 2.4 + 0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Neizmantojiet hemolizētus paraugus. Pastāv vāja korelācija starp duļķainumu un triglicerīdu koncentrāciju. Ļoti retos gadījumos gammopātija, īpaši IgM tips (Valdenstrēma makroglobulinēmija), var sniegt neuzticamus rezultātus (6). Nav pārbaudīts Thermo Fisher Scientific Oy. Lai iegūtu informāciju par citām interferējošām vielām, skatiet uzziņas materiālu (5). PAREDZAMĀS VĒRTĪBAS (3) Vīriešiem < 171 U/l (< 2.85 µkat/l) 37 °C temperatūrā Sievietēm < 145 U/l (< 2.41 µkat/l) 37 °C temperatūrā Norādītās vērtības ir jāuztver tikai kā uzziņas vērtības. Ieteicams katrai laboratorijai verificēt šo diapazonu vai izstrādāt references intervālu saviem pacientiem. BIBLIOGRĀFIJA 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. Konversijas koeficients µkat/l = (U/l)/60 RAŽOTĀJS Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tālr. +358 10 329 200 www.thermoscientific.com MĒRĪŠANAS DIAPAZONS 10 – 1000 U/l (0.17 – 17 µkat/l). Extended measuring range after secondary dilution: 10 – 9000 U/l (0.17 – 150 µkat/l). Pārskatīšanas datums (gggg-mm-dd) 2015-05-08 VEIKTSPĒJAS RAKSTUROJUMS Rezultāti, kas iegūti atšķirīgās laboratorijās, var atšķirties no dotajiem veikuma datiem. Izmaiņas no iepriekšējās versijas Līnija blakus tekstam norāda uz izmaiņām. Noteikšanas robežvērtības Izpēte veikta, par pamatnorādi izmantojot dokumentu EP17-A. Konelab LoB: 0.6 U/l (0.01 µkat/l) Tukšā parauga robežvērtība (LoB) attēlo augstāko mērījumu rezultātu, ko var novērot paraugam bez analizējamām vielām (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Atklāšanas robežvērtība (LoD) attēlo mazāko analizējamās vielas daudzumu paraugā, ko var konstatēt (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Kvantitātes robežvērtība (LoQ) attēlo mazāko faktisko koncentrāciju paraugā, ko var noteikt kvantitatīvi (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Tukšā parauga robežvērtība (LoB) attēlo augstāko mērījumu rezultātu, ko var novērot paraugam bez analizējamām vielām (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Atklāšanas robežvērtība (LoD) attēlo mazāko analizējamās vielas daudzumu paraugā, ko var konstatēt (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Kvantitātes robežvērtība (LoQ) attēlo mazāko faktisko koncentrāciju paraugā, ko var noteikt kvantitatīvi (n = 42). Neprecizitāte CK pārbaudes neprecizitāte ≤ 5.0% kopējā CV. Konelab CK Atkārtojamība (darbības laikā) Ierīcē (kopējā) Vidēji 92 U/l SD CV% 1.7 1.8 Vidēji 226 U/l SD CV% 2.0 0.9 Vidēji 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Precizitātes izpēte tika veikta vienā izpētes veikšanas vietā 21 dienu, izmantojot trīs Konelab 60 analizatorus, vienu operatoru, divas reaģentu partijas un vienu kalibratora partiju, veicot sešas kalibrēšanas, saskaņā ar CLSI dokumentā EP5-A2 sniegtajām vadlīnijām. Mērījumu skaits: n = 84. Tika apvienoti rezultāti no visiem analizatoriem. Indiko CK Atkārtojamība (darbības laikā) Ierīcē (kopējā) Vidēji 95 U/l SD CV% 1.2 1.2 Vidēji 226 U/l SD CV% 2.0 0.9 Vidēji 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Precizitātes izpēte tika veikta vienā izpētes veikšanas vietā 21 dienu, izmantojot trīs Indiko analizatorus, vienu operatoru, divas reaģentu partijas un vienu kalibratora partiju, veicot septiņas kalibrēšanas, saskaņā ar CLSI dokumentā EP5-A2 sniegtajām vadlīnijām. Mērījumu skaits: n = 84. Tika apvienoti rezultāti no visiem analizatoriem. Metožu salīdzināšana Konelab Salīdzināšanas izpēte tika veikta analizatoram Konelab Prime 60, saskaņā ar CLSI dokumentā EP9-A2 sniegtajām vadlīnijām, kā atsauci izmantojot iepriekšējo 981372 CK(IFCC) metodi (x). Lineārā regresija (pēc Deminga) (rezultātu mērvienība U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Paraugu koncentrācija bija no 17 līdz 2253 U/l. Indiko 1) Salīdzināšanas izpēte tika veikta analizatoram Indiko saskaņā ar CLSI dokumentā EP9-A2 sniegtajām vadlīnijām, kā atsauci izmantojot iepriekšējo analizatoram Konelab Prime 60 (x) izmantoto 981372 CK(IFCC) metodi (x). Lineārā regresija (pēc Deminga) (rezultātu mērvienība U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Paraugu koncentrācija bija no 17 līdz 2253 U/l. 2) Salīdzināšanas izpēte tika veikta analizatoram Indiko saskaņā ar CLSI dokumentā EP9-A2 sniegtajām vadlīnijām un kā atsauci izmantojot analizatoram Konelab Prime 60 (x) izmantoto CK(IFCC) metodi. Lineārā regresija (pēc Deminga) (rezultātu mērvienība U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Paraugu koncentrācija bija no 14 līdz 2450 U/l. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 24 D06025_04_Insert_CK (IFCC)_MU PL CK (IFCC) 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) NINIEJSZA ULOTKA INFORMACYJNA DOŁĄCZANA DO OPAKOWANIA PRZEZNACZONA JEST DO STOSOWANIA POZA GRANICAMI USA. PRZEZNACZENIE Do ilościowego oznaczania w diagnostyce in vitro aktywności kinazy kreatynowej (N-fosfotransferaza ATP: kreatyna (CK), EC 2.7.3.2) w surowicy lub osoczu ludzkim w analizatorach Konelab i Indiko. Wszelkie odniesienia do systemów Konelab dotyczą również serii T. PODSUMOWANIE (1,2) Aktywność kinazy kreatynowej jest największa w mięśniach poprzecznie prążkowanych, mózgu i sercu. W innych tkankach, takich jak nerki, przepona i prostata, aktywność kinazy kreatynowej jest znacznie niższa. Aktywność CK jest podwyższona w wielu stanach chorobowych, w tym w chorobach dotyczących mięśni szkieletowych, serca, ośrodkowego układu nerwowego i tarczycy. Po wystąpieniu ostrego uszkodzenia mięśni całkowita aktywność CK we krwi zaczyna rosnąć, osiągając nieprawidłowe poziomy po średnio 4–5 godzinach u 50% pacjentów. Wartości rosną systematycznie od 8. do 24. godziny po zawale i powracają do zakresu wartości prawidłowych z dużą zmiennością osobniczą. ZASADA DZIAŁANIA TESTU Oznaczenie CK jest dwuczęściowym testem w postaci płynnej, w którym do określenia aktywności kinazy kreatynowej wykorzystywane są trzy enzymy. Kinaza kreatynowa katalizuje reakcję pomiędzy fosforanem kreatyny i ADP (adenozyno-5’difosforanem), w wyniku której powstaje kreatyna i ATP (adenozyno-5’-trójfosforan). Ten ostatni związek jest wykorzystywany podczas fosforylacji glukozy do glukozo-6-fosforanu (G-6-P) w obecności heksokinazy (HK). G-6-P zostaje utleniony do 6-fosfoglukonianu (6-PG) i w tym samym czasie równomolowa ilość NADP (fosforanu dinukleotydu nikotynamidoadeninowego) zostaje redukowana do NADPH w reakcji katalizowanej przez dehydrogenazę glukozo-6fosforanową (G-6P-DH). Wzrost absorbancji przy długości fali 340 nm, wynikający z powstania NADPH, jest proporcjonalny do aktywności kinazy kreatynowej w próbce (3). CK Kinaza kreatynowa + ADP → kreatyna + ATP HK ATP + glukoza → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ INFORMACJE O ODCZYNNIKACH Zawartość zestawu nr kat. 981828 2 fiolki po 52 ml odczynnika A 2 fiolki po 13 ml odczynnika B Zawartość zestawu nr kat. 981829 2 fiolki po 20 ml odczynnika A 2 fiolki po 5 ml odczynnika B Stężenia Odczynnik A Bufor (imidazol), pH 6.5 D-glukoza N-acetylo-L-cysteina (NAC) Octan magnezu EDTA-Na2 Fosforan dinukleotydu nikotynamidoadeninowego (NADP) Heksokinaza (HK) NaN3 Odczynnik B Bufor (imidazol) Adenozyno-5’-difosforan (ADP) Adenozyno-5’-monofosforan (AMP) Di(adenozyno-5’)pentafosforan (AP5A) Dehydrogenaza glukozo-6-fosforanowa (G-6-P-DH) EDTA -Na2 Fosforan kreatyny NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥5000 U/l < 0.1 % 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Środki ostrożności Odczynnik B jest niebezpieczny. Należy zapoznać się z informacjami dotyczącymi niebezpieczeństwa i środków ostrożności znajdującymi się w odrębnym arkuszu: H315, P280, P302+P352, P305+P351+P338. Odczynniki zawierają środek konserwujący – azydek sodu. POBIERANIE PRÓBEK Uwaga: Podczas przygotowywania próbek w probówkach do pobierania próbek należy uważnie przestrzegać instrukcji producenta probówki, aby uniknąć błędnych wyników. Szczególną uwagę należy zwrócić na zmienne występujące przed analizą, takie jak mieszanie, czas przechowywania przed wirowaniem oraz ustawienia wirowania. Typ próbki Można używać niezhemolizowanej surowicy lub osocza z heparyną. Uwaga: Hemoliza może doprowadzić do nieprawidłowych wyników. Należy unikać ekspozycji próbek na silne światło. Środki ostrożności Próbki pochodzenia ludzkiego powinny być traktowane i utylizowane jako potencjalnie zakaźne. Przechowywanie (4) Próbkę można przechowywać zgodnie z poniższymi instrukcjami. Próbka 4 godziny w temperaturze 20…25 °C, 1 miesiąc w temperaturze 4…8 °C lub 1 miesiąc w temperaturze -20 °C; próbki należy chronić przed światłem. Izoenzym CK-BB niestabilny. PROCEDURA ANALITYCZNA Procedurę zautomatyzowaną w danym analizatorze przedstawiono w uwagach dotyczących obsługi. Wszelkie zastosowania, które nie posiadają aprobaty firmy Thermo Fisher Scientific Oy, nie są objęte gwarancją poprawności funkcjonowania. Ustawienia przyrządu Indiko Wprowadzić parametry aplikacyjne przy użyciu kodu kreskowego znajdującego się na arkuszu aplikacyjnym Indiko lub za pomocą pliku elektronicznego. Materiały dostarczone Odczynniki zgodnie z opisem powyżej. Materiały niezbędne, lecz niedostarczone Kalibrator i odczynniki kontrolne podane niżej. Kalibracja Odpowiedź (dA/min) jest konwertowana na jednostki wynikowe przy użyciu współczynnika przeliczeniowego. Można również użyć preparatu eCal (981830) jako kalibratora, zgodnie z instrukcjami podanymi na ulotce opakowania preparatu eCal (981830). KONTROLA JAKOŚCI Próbki kontroli jakości (QC) należy badać co najmniej raz dziennie oraz po każdej kalibracji i po rozpoczęciu nowej butelki odczynnika. Zaleca się użycie próbek zbiorczych lub odczynników kontrolnych o co najmniej dwóch poziomach (niskim i wysokim). Kontrolę jakości należy zawsze wykonywać zgodnie z przepisami lokalnymi, krajowymi i federalnymi. Dostępne preparaty kontrolne: CK-MB Control, nr kat. 981908 CK-MB Control High, nr kat. 981255 Nortrol, nr kat. 981043 Abtrol, nr kat. 981044 Przedziały czasowe i limity kontroli jakości (QC) należy dostosować do wymogów poszczególnych pracowni analitycznych. Wyniki próbek kontroli jakości powinny się zawierać w granicach ustalonych w danej pracowni analitycznej. OBLICZANIE WYNIKÓW Wyniki są obliczane automatycznie w analizatorze przy użyciu następującej metody: Aktywność w U/l = współczynnik x dA/min TV x 1000 Współczynnik teoretyczny = 6.22 x SV x P TV SV 6.22 P Całkowita objętość reagentów w ml Objętość próbki w ml Współczynnik ekstynkcji NADPH przy długości fali 340 nm Droga przez kuwetę = 1 Współczynnik przeliczeniowy µkat/l = (U/l)/60 Jeżeli jako kalibrator używany jest preparat eCal (981830), to wyniki obliczane są automatycznie przez analizator Konelab, w oparciu o krzywą kalibracyjną. Krzywa kalibracyjna (przykład) Należy przestrzegać standardowych środków ostrożności obowiązujących podczas pracy ze wszystkimi odczynnikami laboratoryjnymi. Produkt należy traktować jako laboratoryjny odczynnik chemiczny i usuwać zgodnie z lokalnie obowiązującymi przepisami. CK (IFCC) A/min. U/l Przygotowanie Odczynnik A i B są gotowe do użycia. Uwaga: Umieszczając fiolki lub naczynia z odczynnikami w analizatorze, należy upewnić się, że w szyjce naczynia ani na powierzchni odczynnika nie ma pęcherzyków powietrza. Przechowywanie i stabilność Odczynniki w nieotwartych fiolkach, przechowywane w temperaturze 2…8°C zachowują stabilność do terminu ważności wydrukowanego na etykiecie. Dane dotyczące stabilności odczynników znajdujących się w analizatorze przedstawiono w uwagach dotyczących obsługi analizatora. Jeśli odczynniki są mętne, nie należy ich używać. Przebieg krzywej kalibracyjnej zależy od partii odczynników. OGRANICZENIA PROCEDURY Do celów diagnostycznych wyniki należy zawsze oceniać łącznie z historią zdrowotną pacjenta, badaniami klinicznymi i innymi wynikami. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 25 D06025_04_Insert_CK (IFCC)_MU Zakłócenia Przeprowadzono badanie, używając dokumentu CLSI EP7-A2 w charakterze wytycznych. Kryterium: Różnica w zakresie ±10% wartości początkowych. Konelab Substancja zakłócająca Bilirubina związana Bilirubina niezwiązana Hemoglobina w hemolizacie Lipemia: Intralipid (znak handlowy firmy Fresenius Kabi AB) Indiko Substancja zakłócająca Bilirubina związana Bilirubina niezwiązana Hemoglobina w hemolizacie Lipemia: Intralipid (znak handlowy firmy Fresenius Kabi AB) Stężenie substancji zakłócającej Wartość docelowa CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Różnica wzgl. wartości docelowej (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Stężenie substancji zakłócającej Wartość docelowa CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Różnica wzgl. wartości docelowej (%) - 0.8 - 2.4 + 0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Unikać używania próbek zhemolizowanych. Istnieje słaba korelacja pomiędzy stopniem zmętnienia a stężeniem triglicerydów. W bardzo rzadkich przypadkach gammapatii, w szczególności typu IgM, (makroglobulinemia Waldenströma), wyniki mogą być nierzetelne (6). Nie testowano w Thermo Fisher Scientific Oy. Informacje dotyczące innych substancji zakłócających przedstawiono w pozycji (5) piśmiennictwa. WARTOŚCI OCZEKIWANE (3) Mężczyźni <171 U/l (<2.85 µkat/l) przy 37 °C Kobiety <145 U/l (<2.41 µkat/l) przy 37 °C Podane wartości należy traktować jedynie jako orientacyjne. Zaleca się, by w każdej pracowni analitycznej zweryfikować podane zakresy lub ustalić własny przedział referencyjny dla badanej populacji. Współczynnik przeliczeniowy µkat/l = (U/l)/60 ZAKRES POMIAROWY 10 – 1000 U/l (0.17 – 17 µkat/l). Rozszerzony zakres pomiarowy po wtórnym rozcieńczeniu: 10 – 9000 U/l (0.17 – 150 µkat/l). CHARAKTERYSTYKA DZIAŁANIA Wyniki uzyskane w poszczególnych pracowniach analitycznych mogą się różnić od podanych parametrów wydajnościowych. Granica wykrywalności Przeprowadzono badanie, używając dokumentu EP17-A w charakterze wytycznych. Konelab LoB: 0.6 U/l (0.01 µkat/l) Limit próby pustej (LoB) oznacza najwyższy wynik pomiaru, jaki może być zmierzony dla próbki niezawierającej analitu (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Granica wykrywalności (LoD) oznacza najniższą ilość analitu w próbce, którą można wykryć (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Granica oznaczalności (LoQ) to najniższe rzeczywiste stężenie w próbce, które można określić ilościowo (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Limit próby pustej (LoB) oznacza najwyższy wynik pomiaru, jaki może być zmierzony dla próbki niezawierającej analitu (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Granica wykrywalności (LoD) oznacza najniższą ilość analitu w próbce, którą można wykryć (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Granica oznaczalności (LoQ) to najniższe rzeczywiste stężenie w próbce, które można określić ilościowo (n = 42). Niedokładność Niedokładność testu do oznaczania CK wynosi ≤5.0% całkowitego CV. Konelab CK Powtarzalność (W jednej serii) W jednym urządzeniu (sumarycznie) Średnia 92 U/l SD CV% 1.7 1.8 Średnia 226 U/l SD CV% 2.0 0.9 Średnia 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Badanie precyzji przeprowadzone zostało w jednej placówce przez jednego operatora, z wykorzystaniem dokumentu CLSI EP5-A2 w charakterze wytycznych, z użyciem trzech analizatorów Konelab 60, 2 partii odczynników, jednej partii kalibratora i 6 kalibracji, w okresie 21 dni, w ciągu których wykonano 84 pomiary (n = 84). Wyniki z każdego analizatora zostały skompilowane. Indiko CK Powtarzalność (W jednej serii) W jednym urządzeniu (sumarycznie) Średnia 95 U/l SD CV% 1.2 1.2 Średnia 226 U/l SD CV% 2.0 0.9 Średnia 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Badanie precyzji przeprowadzone zostało w jednej placówce przez jednego operatora, z wykorzystaniem dokumentu CLSI EP5-A2 w charakterze wytycznych, użyciem trzech analizatorów Indiko i 2 partii odczynników, jednej partii kalibratora i 7 kalibracji, w okresie 21 dni, w ciągu których wykonano 84 pomiary (n = 84). Wyniki z każdego analizatora zostały skompilowane. Porównanie metod Konelab Przeprowadzono badanie porównawcze przy użyciu analizatora Konelab Prime 60, z wykorzystaniem dokumentu CLSI EP9-A2 w charakterze wytycznych i poprzedniej metody oznaczania CK 981372 (IFCC) jako odniesienia. Regresja liniowa (Deming) (jednostka wyników: U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Wartości stężeń w próbkach wynosiły od 17 U/l do 2253 U/l. Indiko 1) Przeprowadzono badanie porównawcze przy użyciu analizatora Indiko, z wykorzystaniem dokumentu CLSI EP9-A2 w charakterze wytycznych i poprzedniej metody oznaczania CK 981372 (IFCC) na analizatorze Konelab Prime 60 (x) jako odniesienia. Regresja liniowa (Deming) (jednostka wyników: U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Wartości stężeń w próbkach wynosiły od 17 U/l do 2253 U/l. 2) Przeprowadzono badanie porównawcze przy użyciu analizatora Indiko, z wykorzystaniem dokumentu CLSI EP9-A2 w charakterze wytycznych i tej samej metody oznaczania CK (IFCC) na analizatorze Konelab Prime 60 (x) jako odniesienia. Regresja liniowa (Deming) (jednostka wyników: U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Wartości stężeń w próbkach wynosiły od 14 U/l do 2450 U/l. PIŚMIENNICTWO 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. PRODUCENT Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Data zatwierdzenia (rrrr-mm-dd) 2015-05-08 Zmiany w stosunku do poprzedniej wersji Linia obok tekstu oznacza zmiany. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 26 Page 27 D06025_04_Insert_CK (IFCC)_MU PT CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) ESTE FOLHETO INFORMATIVO É APLICÁVEL PARA UTILIZAÇÃO FORA DOS EUA. USO PRETENDIDO Para uso em diagnóstico in vitro na determinação quantitativa da atividade da creatina quinase [ATP: Creatina N-Fosfotransferase (CK), EC 2.7.3.2] no soro ou plasma humanos em analisadores Konelab e Indiko. Qualquer referência aos sistemas Konelab também se refere à série T. RESUMO (1, 2) A atividade da creatina quinase é mais elevada nos músculos estriados, no cérebro e no tecido cardíaco. Outros tecidos, tais como os rins, o diafragma e a próstata, contêm um nível de atividade significativamente mais reduzido. A atividade da CK é elevada em diversas doenças, incluindo as que envolvem o músculo esquelético, o coração, o sistema nervoso central e a tiroide. Após a ocorrência de danos musculares graves, a atividade total de CK no sangue começa a aumentar, alcançando níveis patológicos após uma duração média de 4 a 5 horas em 50% dos pacientes. Os valores aumentam regularmente entre a 8.ª e a 24.ª hora após o início do enfarte e regressam ao intervalo de referência com grandes flutuações interindividuais. PRINCIPIO DO PROCEDIMENTO Este ensaio da CK é um ensaio líquido de duas partes que utiliza três enzimas para determinar a atividade da CK. A creatina quinase catalisa a reação entre a creatina fosfato e a ADP (adenosina-5'-difosfato) com a formação de creatina e de ATP (adenosina-5'-trifosfato). Este último composto fosforila glucose para G-6-P (glucose-6-fosfato) na presença da HK (hexocinase). A G-6-P é oxidada em 6-PG (gluconato-6-fosfato) enquanto simultaneamente uma quantidade equimolar de NADP (nicotinamida adenina dinucleotideo fosfato) é reduzida a NADPH numa reação catalisada pela G-6P-DH (glucose-6-fosfato desidrogenase). O aumento na absorvência a 340 nm, resultante da formação de NADPH, é proporcional à atividade da creatina quinase na amostra (3). CK Creatina fosfato + ADP → Creatina + ATP HK ATP + Glucose → ADP + G-6-P Amostra 4 horas a 20…25°C, 1 mês a 4…8°C ou 1 mês a -20°C; protegida da luz. CK-BB instável. PROCEDIMENTO DE TESTE Consulte as Notas de aplicação para obter mais informações sobre o procedimento automático do analisador. Qualquer aplicação não validada pela Thermo Fisher Scientific Oy não pode ter o desempenho garantido e, por conseguinte, tem de ser avaliada pelo utilizador. Programação do aparelho Indiko Insira os parâmetros da aplicação através do código de barras existente na folha da aplicação do Indiko ou através de ficheiro eletrónico, conforme apropriado. Materiais fornecidos Reagentes, conforme descrito acima. MATERIAIS NECESSÁRIOS, MAS NÃO FORNECIDOS Calibrador e controlos conforme indicado abaixo. Calibragem A resposta (dA/min) é convertida para unidades de resultado através de um fator de cálculo. Ou utilize o eCal (981830) como calibrador de acordo com as instruções fornecidas no folheto da embalagem do eCal (981830). Controlo de qualidade Utilize amostras de controlo de qualidade (CQ) pelo menos uma vez por dia, depois de cada calibragem e sempre que utilizar um frasco de reagente novo. Recomenda-se a utilização de, pelo menos, dois níveis (baixo e alto) de controlos ou aglomerados (pools) de amostras. Proceda sempre de acordo com as normas locais, distritais e nacionais quando efetuar o controlo de qualidade. Controlos disponíveis: CK-MB Control, código 981908 CK-MB Control High, código 981255 Nortrol, código 981043 Abtrol, código 981044 Os intervalos e limites do controlo (CQ) devem ser adaptados aos requisitos individuais de cada laboratório. Os resultados da amostra de controlo de qualidade devem situar-se dentro dos limites predefinidos pelo laboratório. CÁLCULO DE RESULTADOS Os resultados são automaticamente calculados pelo analisador da seguinte forma: G-6-P-DH G-6-P + NADP → 6-PG + NADPH + H+ + Atividade em U/l = Fator x dA/min INFORMAÇÃO SOBRE OS REAGENTES O kit com o código 981828 contém 2 frascos de 52 ml de Reagente A 2 frascos de 13 ml de Reagente B O kit com o código 981829 contém 2 frascos de 20 ml de Reagente A 2 frascos de 5 ml de Reagente B Concentrações Reagente A Tampão (Imidazola), pH 6.5 D-glucose N-acetil-L-cisteína (NAC) Acetato de magnésio EDTA-Na2 Nicotinamida adenina dinucleótido fosfato (NADP) Hexocinase (HK) NaN3 Reagente B Tampão (Imidazola) Adenosina-5'-difosfato (ADP) Adenosina-5'-monofosfato (AMP) Di(adenosina-5')penta-fosfato (AP5A) Glucose-6-fosfato desidrogenase (G-6-P-DH) EDTA-Na2 Creatina fosfato NaN3 TV x 1000 Fator teórico = 6.22 x SV x P TV SV 6.22 P 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1% Volume total da reação em ml Volume da amostra em ml Coeficiente de extinção do NADPH a 340 nm Comprimento do percurso da cuvete = 1 Fator de conversão µkat/l = (U/l)/60 Se utilizar o eCal (981830) como calibrador, os resultados são calculados automaticamente pelo analisador através de uma curva de calibragem. Curva de calibragem (exemplo) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1% Precauções O reagente B é perigoso. Consulte os avisos relacionados com o perigo e as precauções na folha fornecida em separado no kit: H315, P280, P302+P352, P305+P351+P338. Os reagentes contêm azida de sódio como conservante. Adote as precauções habitualmente requeridas para o manuseamento de reagentes de laboratório. O produto tem ser eliminado como se fosse um químico de laboratório, de acordo com as normas locais. A curva de calibragem depende do lote. Preparação Os reagentes A e B estão prontos a usar. Nota: certifique-se de que não há nenhuma bolha no gargalo do frasco nem na superfície do reagente quando inserir os frascos ou as ampolas do reagente no analisador. LIMITAÇÕES DO PROCEDIMENTO Para fins de diagnóstico, os resultados devem ser sempre avaliados em conjunto com o historial clínico, os exames clínicos e outros dados do paciente. CONSERVAÇÃO E ESTABILIDADE Os reagentes que se encontram em frascos fechados permanecem estáveis até à data indicada no rótulo, desde que mantidos a 2...8°C. Consulte as Notas de aplicação do analisador para obter mais informações sobre a estabilidade dos reagentes quando inseridos no analisador. Se os reagentes apresentam turvação, não os utilize. Interferência Foi realizado um estudo no qual se utilizou o documento EP7-A2 do CLSI como referência. Critério: diferença dentro de ± 10% dos valores iniciais. Konelab Substância interferente Bilirrubina conjugada COLHEITA DA AMOSTRA Nota: ao processar as amostras nos tubos de colheita de amostra, siga cuidadosamente as instruções do fabricante dos tubos, de forma a evitar resultados erróneos. Preste especial atenção às variáveis pré-analíticas, tais como a mistura, o tempo de repouso antes da centrifugação e as velocidades de centrifugação. Tipo de amostra Pode utilizar soro não hemolisado ou plasma com heparina. Nota: a hemólise pode causar resultados erróneos. Evite a exposição das amostras a luzes intensas. Precauções As amostras humanas devem ser manuseadas e eliminadas como se fossem potencialmente infecciosas. Bilirrubina não conjugada Hemoglobina no hemolisado Lipemia: Intralipid (marca comercial da Fresenius Kabi AB) Concentração interferente 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l CK alvo (U/l) 188 498 180 495 184 505 Diferença do alvo (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Armazenamento (4) As amostras podem ser conservadas como indicado a seguir. D06025_04_Insert_CK (IFCC)_MU Indiko Substância interferente Bilirrubina conjugada Bilirrubina não conjugada Hemoglobina no hemolisado Lipemia: Intralipid (marca comercial da Fresenius Kabi AB) Concentração interferente 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l CK alvo (U/l) 179 477 175 469 187 502 Diferença do alvo (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Evite a utilização de amostras hemolisadas. Há uma correlação fraca entre a turvação e a concentração de triglicéridos. Em casos muito raros, a gamopatia, em particular o tipo IgM (macroglobulinemia de Waldenström), pode causar resultados inexatos (6). Não testado na Thermo Fisher Scientific Oy. Para outras substâncias interferentes, consulte a referência (5). VALORES ESPERADOS (3) Homem < 171 U/l (< 2.85 µkat/l) a 37°C Mulher < 145 U/l (< 2.41 µkat/l) a 37°C Os valores indicados devem servir apenas como uma referência. Recomendamos que cada laboratório verifique este intervalo, ou obtenha um intervalo de referência para a população que serve. Fator de conversão µkat/l = (U/l)/60 INTERVALO DE MEDIÇÃO 10 – 1000 U/l (0.17 – 17 µkat/l). Intervalo de medição alargado depois da diluição secundária: 10 – 9000 U/l (0.17 – 150 µkat/l). CARACTERÍSTICAS DE DESEMPENHO Os resultados obtidos em laboratórios individuais podem diferir dos dados de desempenho fornecidos. Limite de deteção Foi realizado um estudo de acordo com o documento EP17-A. Konelab LoB: 0.6 U/l (0.01 µkat/l) O Limite do Branco (LoB) representa o resultado de medição mais elevado provável de ser observado numa amostra sem analito (n = 60). LoD: 1.5 U/l (0.03 µkat/l) O Limite de Deteção (LoD) representa a quantidade mais baixa de analito numa amostra que pode ser detetada (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) O Limite de Quantificação (LoQ) representa a concentração mais baixa concreta numa amostra que pode ser determinada quantitativamente (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) O Limite do Branco (LoB) representa o resultado de medição mais elevado provável de ser observado numa amostra sem analito (n = 60). LoD: 2.2 U/l (0.04 µkat/l) O Limite de Deteção (LoD) representa a quantidade mais baixa de analito numa amostra que pode ser detetada (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) O Limite de Quantificação (LoQ) representa a concentração mais baixa concreta numa amostra que pode ser determinada quantitativamente (n = 42). Imprecisão A imprecisão do ensaio de CK é ≤ 5.0% do total do CV. Konelab CK Repetibilidade (intraensaio) Intra-aparelho (total) Média 92 U/l SD % do CV 1.7 1.8 Média 226 U/l SD % do CV 2.0 0.9 Média 616 U/l SD % do CV 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Foi realizado um estudo de precisão no qual se utilizou o documento EP5-A2 do CLSI como referência, num local com três analisadores Konelab 60, um operador, dois lotes de reagentes, um lote de calibrador e seis calibragens durante 21 dias, sendo o número de medições de n= 84. Foram combinados os resultados de cada analisador. Indiko CK Repetibilidade (intraensaio) Intra-aparelho (total) Média 95 U/l SD % do CV 1.2 1.2 Média 226 U/l SD % do CV 2.0 0.9 Média 637 U/l SD % do CV 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Foi realizado um estudo de precisão no qual se utilizou o documento EP5-A2 do CLSI como referência, num local com três analisadores Indiko, um operador, dois lotes de reagentes, um lote de calibrador e sete calibragens, durante 21 dias, sendo o número de medições de n = 84. Foram combinados os resultados de cada analisador. Comparação de métodos Konelab Foi realizado um estudo comparativo no analisador Konelab Prime 60, tendo-se utilizado o documento EP9-A2 do CLSI e o anterior método 981372 CK(IFCC) (x) como referências. Regressão linear (Deming) (unidade dos resultados U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 As concentrações das amostras situam-se entre 17 e 2253 U/l. Indiko 1) Foi realizado um estudo comparativo no analisador Indiko, tendo-se utilizado o documento EP9-A2 do CLSI e o anterior método 981372 CK(IFCC) no analisador Konelab Prime 60 (x) como referências. Regressão linear (Deming) (unidade dos resultados U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 As concentrações das amostras situam-se entre 17 e 2253 U/l. 2) Foi realizado um estudo comparativo no analisador Indiko, tendo-se utilizado o documento EP9-A2 do CLSI e o mesmo método CK(IFCC) no analisador Konelab Prime 60 (x) como referências. Regressão linear (Deming) (unidade dos resultados U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 As concentrações das amostras situam-se entre 14 e 2450 U/l. REFERÊNCIAS 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. FABRICANTE Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Data de revisão (aaaa-mm-dd) 2015-05-08 Alterações em relação à versão anterior A linha junto ao texto indica alterações. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 28 Page 29 D06025_04_Insert_CK (IFCC)_MU RO 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) ACEST PROSPECT ESTE APLICABIL PENTRU FOLOSIREA ÎN AFARA S.U.A. UTILIZAREA RECOMANDATA Pentru diagnostic in vitro utilizată în determinarea cantitativă a activităţii creatin-kinazei (ATP: Creatin-N-fosfotransferaza (CK), EC 2.7.3.2) în ser uman sau plasmă pe analizoarele Konelab şi Indiko. Orice referinţă la sistemele Konelab se referă şi la seria T. REZUMAT (1,2) Activitatea creatin-kinazei e mai intensă în muşchiul striat, în creier şi în ţesutul cardiac. Alte ţesuturi, cum ar fi rinichii, diafragmul şi prostata, au o activitate mult mai scăzută. Activitatea CK e crescută în multe boli, inclusiv cele ale muşchilor scheletici, ale inimii, ale sistemului nervos central şi ale tiroidei. După traumatisme musculare masive, activitatea totală a CK în sânge începe să crească, atingând niveluri patologice după o durată medie de 4-5 ore la 50% dintre pacienţi. Valorile cresc regulat după 8-24 ore de la instalarea infarctului şi revin la valorile de referinţă în intervale care variază mult de la un individ la altul. PRINCIPIUL ANALIZEI Analiza CK este o analiză lichidă în două părţi care utilizează trei enzime pentru a determina activitatea CK. Creatin-kinaza catalizează reacţia dintre creatin-fosfat şi ADP (adenozin-5’-difosfat) cu formare de creatină şi ATP (adenozin-5’-trifosfat). Ultimul compus fosforilează glucoza la G-6-P (glucozo-6fosfat) în prezenţa HK (hexokinazei). G-6-P este oxidat la 6-PG (gluconat-6-fosfat) şi, totodată, o cantitate echimolară de NADP (nicotinamid-adenin-dinucleotid fosfat) e redusă la NADPH printr-o reacţie catalizată de G-6P-DH (glucozo-6-fosfat dehidrogenază). Creşterea absorbanţei la 340 nm, care rezultă din formarea de NADPH, este proporţională cu activitatea creatin-kinazei din probă. (3). CK Creatin-fosfat + ADP → Creatină + ATP HK ATP + Glucoză → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Probă 4 ore la 20...25 °C, 1 lună la 4...8 °C, sau 1 lună la -20 °C; ferită de lumină. CK-BB nu este stabil. PROCEDURA DE ANALIZARE Consultaţi Notele de Aplicare pentru procedura automată pe analizorul dv. Toate aplicaţiile care nu au fost validate de Thermo Fisher Scientific Oy nu garantează performanţa şi, prin urmare, trebuie evaluate de către utilizator. Setările instrumentului pentru Indiko Introduceţi parametrii aplicaţiei prin intermediul codului de bare care se află în fişa aplicaţiei Indiko sau printr-un fişier electronic, conform necesităţilor. Materiale furnizate Reactivii mai sus-menţionaţi. Materiale necesare, dar care nu sunt furnizate Calibratorii şi martorii indicaţi mai jos. Calibrarea Răspunsul (dA/min) este convertit în unităţile rezultatului, cu ajutorul unui factor de calcul. Sau folosiţi eCal (981830) drept calibrator, conform instrucţiunilor din prospectul pentru eCal (981830). Controlul calităţii Trebuie folosite probe pentru controlul calităţii cel puţin o dată pe zi, după fiecare calibrare şi de fiecare dată când se foloseşte un flacon nou de reactiv. Se recomandă să utilizaţi cel puţin două niveluri (scăzut şi ridicat) de martori sau baterii de probe. Respectaţi întotdeauna reglementările locale, regionale şi naţionale la efectuarea controlului calităţii. Martori disponibili: CK-MB Control, cod 981908 CK-MB Control High, cod 981255 Nortrol, cod 981043 Abtrol, cod 981044 Intervalele (QC) şi limitele martorilor trebuie adaptate la necesităţile individuale ale laboratorului. Rezultatele probelor pentru controlul calităţii trebuie să se încadreze în limitele stabilite de laborator. CALCULAREA REZULTATELOR Rezultatele sunt calculate automat de către analizor, după cum urmează: INFORMAŢII DESPRE REACTIVI Trusa cod 981828 conţine 2 fiole de 52 ml reactiv A 2 fiole de 13 ml reactiv B Păstrarea (4) Probele pot fi păstrate aşa cum se arată mai jos. Activitatea în U/l = Factor x dA/min Trusa cod 981829 conţine 2 fiole de 20 ml reactiv A 2 fiole de 5 ml reactiv B Concentraţii Reactiv A Soluție-tampon (Imidazol), pH 6.5 D-glucoză N-acetil-L-cisteină (NAC) Acetat de magneziu EDTA-Na2 Nicotinamid-adenin-dinucleotid fosfat (NADP) Hexokinază (HK) NaN3 Reactiv B Soluție-tampon (Imidazol) Adenozin-5’-difosfat (ADP) Adenozin-5’-monofosfat (AMP) Di(adenozin-5’)penta-fosfat (AP5A) Glucozo-6-fosfat dehidrogenaza (G-6-P-DH) EDTA -Na2 Creatin-fosfat NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1% TV x 1000 Factor teoretic = 6.22 x SV x P TV SV 6.22 P Volumul total de reacţie în ml Volumul probei în ml Coeficientul de extincţie al NADPH la 340 nm Lungimea drumului optic al cuvei = 1 Factor de conversie µkat/l = (U/l)/60 Dacă se foloseşte eCal (981830) drept calibrator, rezultatele sunt calculate automat de către analizor folosind o curbă de calibrare. Curbă de calibrare (exemplu) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1% Precauţii Reactivul B este periculos. Consultaţi fişa separată din interiorul kitului pentru frazele privind Pericolul şi Precauţiile: H315, P280, P302+P352, P305+P351+P338. Reactivii conţin azidă de sodiu drept conservant. Adoptaţi măsurile normale de precauţie necesare pentru mânuirea tuturor reactivilor de laborator. Produsul trebuie aruncat ca produs chimic de laborator, în conformitate cu reglementările locale. Pregătirea Reactivii A şi B sunt gata de folosire. Notă: Verificaţi să nu existe bule în gâtul flaconului sau la suprafaţa reactivului când introduceţi fiolele sau recipientele cu reactiv în dispozitivul de analiză. Păstrarea şi stabilitatea Reactivii din fiolele nedeschise sunt stabili la 2…8 °C până la data de expirare tipărită pe etichetă. Consultaţi Notele de Aplicare ale analizorului dv. pentru stabilitatea reactivilor păstraţi pe instrument. Nu folosiţi reactivii dacă aceştia prezintă turbiditate. RECOLTAREA PROBELOR Notă: Când prelucraţi probele în eprubete de recoltare, urmaţi cu grijă instrucţiunile producătorului eprubetelor, pentru a evita rezultatele eronate. Acordaţi o atenţie deosebită variabilelor preanalitice precum omogenizarea, intervalul de repaus înainte de centrifugare şi setările centrifugei. Tipul de probe Pot fi folosite ser nehemolizat sau plasmă heparinizată. Notă: Hemoliza poate conduce la rezultate deviate. Evitaţi expunerea probelor la lumină puternică. Precauţii Probele umane trebuie mânuite şi aruncate ca şi cum ar fi potenţial infecţioase. Curba de calibrare depinde de lot. LIMITARILE PROCEDURII În scopul stabilirii unui diagnostic, rezultatele trebuie evaluate întotdeauna luând în considerare istoricul medical al pacientului, examenele clinice şi alte investigaţii. Interferenţe S-a efectuat un studiu folosind Documentul CLSI EP7-A2 ca ghid. Criteriu: Diferenţă în limite de ± 10% din valorile iniţiale. Konelab Substanţă interferentă Bilirubină conjugată Bilirubină neconjugată Hemoglobină în hemolizat Lipemie: Intralipid (marcă înregistrată a firmei Fresenius Kabi AB) Concentraţie interferentă Nivel-ţintă CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Diferenţă faţă de nivelul-ţintă (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. CK (IFCC) D06025_04_Insert_CK (IFCC)_MU Indiko Substanţă interferentă Bilirubină conjugată Bilirubină neconjugată Hemoglobină în hemolizat Lipemie: Intralipid (marcă înregistrată a firmei Fresenius Kabi AB) Concentraţie interferentă Nivel-ţintă CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Diferenţă faţă de nivelul-ţintă (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Evitaţi folosirea probelor hemolizate. Există o corelaţie scăzută între turbiditate şi concentraţia trigliceridelor. În cazuri foarte rare de gamopatii, mai ales cele de tip IgM (macroglobulinemia Waldenström), rezultatele pot fi inexacte (6). Nu a fost testat la Thermo Fisher Scientific Oy. Pentru alte substanţe care pot interfera, vă rugăm să consultaţi referinţa (5). VALORI ESTIMATE (3) Bărbaţi < 171 U/l (< 2.85 µkat/l) la 37 °C Femei < 145 U/l (< 2.41 µkat/l) la 37 °C Valorile menţionate au doar caracter orientativ. Se recomandă ca fiecare laborator să verifice acest interval sau să elaboreze un interval de referinţă pentru populaţia pe care o deserveşte. Factor de conversie µkat/l = (U/l)/60 INTERVAL DE MASURARE 10 – 1000 U/l (0.17 – 17 µkat/l). Interval extins de măsurare după diluţia secundară: 10 – 9000 U/l (0.17 – 150 µkat/l). CARACTERISTICILE PERFORMANTEI Rezultatele obţinute în laboratoarele individuale pot fi diferite faţă de datele indicate referitoare la performanţă. Limitele de detecţie S-a efectuat un studiu folosind documentul EP17-A ca ghid. Konelab LoB: 0.6 U/l (0.01 µkat/l) Limita blankului (LoB) reprezintă rezultatul cel mai mare al măsurătorii care este probabil să fie observat pentru o probă fără analit (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Limita de detecţie (LoD) reprezintă cantitatea cea mai mică de analit dintr-o probă care poate fi detectată (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Limita de cuantificare (LoQ) reprezintă concentraţia cea mai mică existentă într-o probă care poate fi determinată cantitativ (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Limita blankului (LoB) reprezintă rezultatul cel mai mare al măsurătorii care este probabil să fie observat pentru o probă fără analit (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Limita de detecţie (LoD) reprezintă cantitatea cea mai mică de analit dintr-o probă care poate fi detectată (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Limita de cuantificare (LoQ) reprezintă concentraţia cea mai mică existentă într-o probă care poate fi determinată cantitativ (n = 42). Imprecizie Imprecizia analizei pentru CK este ≤ 5.0 % din CV total. Konelab CK Repetabilitate (În cadrul aceleiaşi analize) Pe acelaşi dispozitiv (Total) Medie 92 U/l SD CV% 1.7 1.8 Medie 226 U/l SD CV% 2.0 0.9 Medie 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 S-a efectuat un studiu asupra preciziei folosind Documentul CLSI EP5-A2 ca ghid, la o unitate medicală, cu trei analizoare Konelab 60, un operator, două loturi de reactivi, un lot de calibratori şi 6 calibrări, timp de 21 de zile, cu numărul măsurătorilor n = 84. Au fost combinate rezultatele de la fiecare analizor. Indiko CK Repetabilitate (În cadrul aceleiaşi analize) Pe acelaşi dispozitiv (Total) Medie 95 U/l SD CV% 1.2 1.2 Medie 226 U/l SD CV% 2.0 0.9 Medie 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 S-a efectuat un studiu asupra preciziei folosind Documentul CLSI EP5-A2 ca ghid, la o unitate medicală, cu trei analizoare Indiko, un operator, două loturi de reactivi, un lot de calibratori şi 7 calibrări, timp de 21 de zile, cu numărul măsurătorilor n = 84. Au fost combinate rezultatele de la fiecare analizor. Comparaţia metodelor Konelab S-a efectuat un studiu comparativ asupra Konelab Prime 60 folosind Documentul CLSI EP9-A2 ca ghid şi metoda 981372 CK(IFCC) anterioară (x) ca referinţă. Regresia liniară (Deming) (unitatea rezultatului U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Concentraţiile probelor au fost cuprinse între 17 şi 2253 U/l. Indiko 1) S-a efectuat un studiu comparativ asupra Indiko folosind Documentul CLSI EP9-A2 ca ghid şi metoda 981372 CK(IFCC) anterioară asupra Konelab Prime 60 (x) ca referinţă. Regresia liniară (Deming) (unitatea rezultatului U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Concentraţiile probelor au fost cuprinse între 17 şi 2253 U/l. 2) S-a efectuat un studiu comparativ asupra Indiko folosind Documentul CLSI EP9-A2 ca ghid şi aceeaşi metodă CK(IFCC) asupra Konelab Prime 60 (x) ca referinţă. Regresia liniară (Deming) (unitatea rezultatului U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Concentraţiile probelor au fost cuprinse între 14 şi 2450 U/l. BIBLIOGRAFIE 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry (Principii fundamentale ale chimiei clinice), ediţia a 5-a, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results (Diagnoza în laboratoarele clinice; Utilizarea şi evaluarea rezultatelor clinice de laborator), prima ediţie, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germania, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. (Proceduri primare de referinţă pentru măsurarea concentraţiilor activităţii catalitice a enzimelor la 37 °C). 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. ((Proceduri de referinţă pentru măsurarea concentraţiilor catalitice ale creatin-kinazei). 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples (Utilizarea anticoagulanţilor în investigaţiile şi stabilitatea diagnosticului de laborator al probelor de sânge, plasmă şi ser). Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition (Efectele medicamentelor asupra testelor clinice de laborator, Ediţia a cincea), AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention (Interferenţa gamopatiei în analizele chimice clinice: mecanisme, detecţie şi prevenţie). Clin Chem Lab Med 2007; 45(9), 1240 -1243. PRODUCĂTOR Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Data reviziei (aaaa-ll-zz) 2015-05-08 Modificări faţă de versiunea precedentă Linia de lângă text indică modificări. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 30 D06025_04_Insert_CK (IFCC)_MU RU CK (IFCC) КК (IFCC) 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) ДАННЫЙ ЛИСТОК-ВКЛАДЫШ ПРЕДНАЗНАЧЕН ДЛЯ ИСПОЛЬЗОВАНИЯ ЗА ПРЕДЕЛАМИ США. НАЗНАЧЕНИЕ Используется в диагностике in vitro для количественного определения активности креатинкиназы (ATP: креатин N-фосфотрансфераза (КК), EC 2.7.3.2) в человеческой сыворотке или плазме крови с помощью анализаторов Konelab и Indiko. Любая информация, касающаяся систем Konelab, также относится к анализаторам серии T. КРАТКАЯ СПРАВКА (1,2) Наибольшую активность креатинкиназа имеет в поперечных мышцах, мозге и сердечной ткани. Другие ткани, такие как почки, диафрагма и простата, содержат значительно меньше креатинкиназы. Активность КК увеличивается при многих заболеваниях, включая повреждения скелетных мышц, сердца, центральной нервной системы и щитовидной железы. После острого повреждения мышц общая активность КК в крови начинает увеличиваться и достигает патологических уровней в среднем через 4–5 часов у 50 % пациентов. Активность регулярно увеличивается с 8-го по 24-й час после начала инфаркта и возвращается в нормальный диапазон в зависимости от индивидуальных особенностей пациента. ПРИНЦИП АНАЛИЗА Данный тест представляет собой двухэтапный анализ жидкости с использованием трех ферментов для определения активности КК. Креатинкиназа катализирует реакцию между креатинфосфатом и АДФ (аденозин-5'дифосфат), в результате чего образуется креатин и АТФ (аденозин-5'-трифосфат). Последнее соединение фосфорилирует глюкозу в G-6-P (глюкозо-6-фосфат) в присутствии HK (гексокиназы). G-6-P окисляется до 6-PG (глюконат-6-фосфат). Одновременно с этим эквимолярное количество НАДФ (никотинамид аденин динуклеотид фосфат) восстанавливается до НАДФH в реакции, катализируемой G-6P-DH (глюкозо-6-фосфатдегидрогеназа). Увеличение поглощения при 340 nm в результате образования НАДФH пропорционально активности креатинкиназы в образце (3). КК Креатинфосфат + АДФ → Креатин + АТФ HK АТФ + Глюкоза → АДФ + G-6-P G-6-P-DH + + G-6-P + НАДФ → 6-PG + НАДФH + H ИНФОРМАЦИЯ О РЕАГЕНТЕ Комплект с кодом 981828 содержит 2 флакона с 52 ml реагента А 2 флакона с 13 ml реагента В Комплект с кодом 981829 содержит 2 флакона с 20 ml реагента А 2 флакона с 5 ml реагента В Концентрация Реагент А Буфер (Имидазол), pH 6.5 D-глюкоза N-ацетил-L-цистеин (NAC) Магния ацетат EDTA-Na2 Никотинамид-аденин-динуклеотидфосфат (НАДФ) Гексокиназа (HK) NaN3 Реагент В Буфер (Имидазол) Аденозин-5’-дифосфат (АДФ) Аденозин-5’-монофосфат (АМФ) Ди(аденозин-5’)пента-фосфат (АФ5A) Глюкозо-6-фосфатдегидрогеназа (G-6-P-DH) EDTA -Na2 Креатинфосфат NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2,7 mmol/l ≥ 5000 U/l < 0.1 % 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Меры предосторожности Реагент В является опасным веществом. Сведения о возможных рисках и мерах предосторожности содержатся на отдельном вкладыше: H315, P280, P302+P352, P305+P351+P338. Реагенты содержат азид натрия в качестве консерванта. Необходимо соблюдать стандартные меры предосторожности при работе со всеми лабораторными реагентами. Продукт необходимо утилизировать как лабораторный химический реактив в соответствии с местными нормативными требованиями. СБОР ОБРАЗЦОВ Примечание: При обработке образцов в пробирках соблюдайте все инструкции производителя пробирок чтобы избежать получения ошибочных результатов. Особое внимание на преаналитическом этапе следует уделять таким факторам, как смешивание, время от момента взятия образца до центрифугирования и настройки центрифуги. Тип образца Можно использовать негемолизированную сыворотку или плазму крови с гепарином. Примечание: Это может привести к отклонениям в результатах. Не подвергайте образцы воздействию интенсивного света. Меры предосторожности При исследовании и утилизации все образцы должны рассматриваться как потенциально инфицированные. Хранение (4) Образцы можно хранить, как указано ниже. Образец 4 суток при температуре 20…25 °C, 1 месяц при температуре 4…8 °C, или 1 месяц при температуре -20 °C; защищенные от света. КК-BB не стабильна. ПРОЦЕДУРА ТЕСТИРОВАНИЯ Подробная информация содержится в указаниях по выполнению автоматической процедуры тестирования для используемой модели анализатора. Результаты любого тестирования, не подтвержденного компанией Thermo Fisher Scientific Oy пользователь оценивает самостоятельно. Настройки аппарата Indiko Введите соответствующие области применения параметры с помощью штрихкода в краткой инструкции по применению или из электронного файла в зависимости от ситуации. Входящие в комплект материалы Перечисленные выше реагенты. Необходимые реагенты, не входящие в набор Калибровочные и контрольные образцы, указанные ниже. Калибровка Показатель реакции (dA/min) преобразуется в итоговые единицы измерения с помощью коэффициента пересчета. Либо используйте в качестве калибратора eCal (981830) в соответствии с инструкциями, приведенными на листке-вкладыше калибратора eCal (981830). Контроль качества Используйте образцы для контроля качества (КК) как минимум один раз в день, а также после каждой калибровки и замены флакона с реагентом. Рекомендуется использовать образцы или группы образцов для контроля как минимум двух уровней (низкого и высокого). При проведении процедуры контроля качества соблюдайте все местные, региональные и федеральные нормативные требования. Имеющиеся в наличии контрольные материалы: CK-MB Control (код 981908) CK-MB Control High (код 981255) Nortrol (код 981043) Abtrol (код 981044) Периодичность анализа контрольных образцов и предельные значения должны соответствовать требованиям, установленным в конкретной лаборатории. Результаты анализа образцов для контроля качества должны находиться в пределах, заранее установленных лабораторией. РАСЧЕТ РЕЗУЛЬТАТОВ Анализатор автоматически рассчитывает результаты следующим образом: Активность в U/l = коэффициент x dA/min TV x 1000 Теоретический коэффициент = 6.22 x SV x P TV SV 6.22 P Общий объем реакции в ml Объем образца в ml Коэффициент поглощения НАДФH при 340 nm Длина оптического пути кюветы = 1 Коэффициент преобразования µkat/l = (U/l)/60 Если в качестве калибратора используется eCal (981830), результаты рассчитываются анализатором автоматически с использованием кривой калибровки. Кривая калибровки (пример) КК (IFCC) Подготовка Реагенты А и В готовы к использованию. Примечание: Помещая флаконы или другие емкости с реагентами в анализатор, убедитесь в том, что в горлышке флакона или на поверхности реагента нет пузырьков воздуха. Хранение и стабильность реагентов Реагенты в невскрытых флаконах остаются стабильными при температуре 2…8 °C до истечения указанного на этикетке срока годности. Сведения о стабильности реагентов в анализаторе см. в указаниях по применению конкретного анализатора. Не используйте мутные реагенты. Кривая калибровки зависит от лота реагента. ОГРАНИЧЕНИЯ При постановке диагноза результаты лабораторных исследований следует всегда оценивать совместно с данными анамнеза, клинического обследования и другими показателями. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 31 D06025_04_Insert_CK (IFCC)_MU Искажение результатов анализа Исследование проводилось в соответствии с рекомендациями EP7-A2 Института клинических и лабораторных стандартов (CLSI). Критерий: Разница в пределах ± 10 % от первоначальных значений. Konelab Интерферент Концентрация интерферента Связанный билирубин Несвязанный билирубин Гемоглобин в гемолизате Липемия: Intralipid (товарный знак компании Fresenius Kabi AB) Indiko Интерферент 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l Целевая концентрация КК (U/l) 188 498 180 495 184 505 Отклонение от значения (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Концентрация интерферента Связанный билирубин Несвязанный билирубин Гемоглобин в гемолизате Липемия: Intralipid (товарный знак компании Fresenius Kabi AB) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l Целевая концентрация КК (U/l) 179 477 175 469 187 502 Отклонение от значения (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Не используйте гемолизированные образцы. Мутность образцов слабо коррелирует с уровнем концентрации триглицеридов. В очень редких случаях наличие гаммапатии, в частности обусловленной нарушением синтеза иммуноглобулинов класса IgM (макроглобулинемия Вальденстрема), может привести к получению недостоверных результатов анализа (6). Это не проверялось компанией Thermo Fisher Scientific Oy. Информация о других интерферентах содержится в публикации (5). ОЖИДАЕМЫЕ ЗНАЧЕНИЯ (3) Мужчины < 171 U/l (< 2.85 µkat/l) при 37 °C Женщины < 145 U/l (< 2.41 µkat/l) при 37 °C Приведенные значения носят исключительно справочный характер. Каждой лаборатории рекомендуется подтвердить этот диапазон значений или установить собственный референтный диапазон. Коэффициент преобразования µkat/l = (U/l)/60 ДИАПАЗОН ИЗМЕРЕНИЙ 10 – 1000 U/l (0.17 – 17 µkat/l). Расширенный диапазон измерений после вторичного разведения: 10 – 9000 U/l (0.150 – 17 µkat/l). РАБОЧИЕ ХАРАКТЕРИСТИКИ Результаты, полученные в конкретных лабораториях, могут отличаться от приведенных характеристик. Предел обнаружения Исследование проводилось в соответствии с рекомендациями EP17-A Института клинических и лабораторных стандартов (CLSI). Konelab LoB: 0.6 U/l (0.01 µkat/l) Предел холостой пробы (LoB) представляет собой максимальный результат измерения, который можно получить в образце без аналита (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Предел обнаружения (LoD) представляет собой наименьшее количество аналита в образце, которое можно обнаружить (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Предел количественного определения (LoQ) представляет наименьшую фактическую концентрацию в образце, которую можно определить количественно (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Предел холостой пробы (LoB) представляет собой максимальный результат измерения, который можно получить в образце без аналита (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Предел обнаружения (LoD) представляет собой наименьшее количество аналита в образце, которое можно обнаружить (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Предел количественного определения (LoQ) представляет наименьшую фактическую концентрацию в образце, которую можно определить количественно (n = 42). Погрешность Погрешность результатов анализа CK ≤ 5.0 % от общего коэффициента вариации (CV). Konelab КК Повторяемость (в пределах одного цикла измерений) В пределах устройства (всего) Среднее значение 92 U/l SD 1.7 CV% 1.8 Среднее значение 226 U/l SD CV% 2.0 0.9 2.5 2.8 3.1 1.4 Среднее значение 616 U/l SD CV% 7.1 1.2 22.3 3.6 Исследование точности измерений проводилось в соответствии с Руководством по оценке прецизионности количественных методов измерения (EP5-A2) Института клинических и лабораторных стандартов (CLSI) в одном исследовательском центре с использованием трех анализаторов Konelab 60, одного оператора, двух партий реагента, одной партии калибратора и шести процедур калибровки на протяжении 21 дня, за которые было выполнено восемьдесят четыре измерения (n = 84). Результаты, полученные на всех анализаторах, объединялись. Indiko КК Повторяемость (в пределах одного цикла измерений) В пределах устройства (всего) Среднее значение 95 U/l SD CV% 1.2 1.2 Среднее значение 226 U/l SD CV% 2.0 0.9 Среднее значение 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Исследование точности измерений проводилось в соответствии с Руководством по оценке прецизионности количественных методов измерения (EP5-A2) Института клинических и лабораторных стандартов (CLSI) в одном исследовательском центре с использованием трех анализаторов Indiko, одного оператора, двух партий реагента, одной партии калибратора и семи процедур калибровки на протяжении 21 дня, за которые было выполнено восемьдесят четыре измерения (n = 84). Результаты, полученные на всех анализаторах, объединялись. Сравнение методов Konelab Сравнительное исследование было проведено в соответствии с Руководством по сравнению методов исследования биологических образцов и расчету стандартных ошибок (EP9-A2) Института клинических и лабораторных стандартов (CLSI) с использованием анализатора Konelab Prime 60 и предыдущего теста 981372 CK (IFCC) (х) в качестве эталонного метода. Линейная регрессия (Deming) (единица измерения — U/l): y = 1.07x – 6.3 r = 0.999 n = 95 Концентрация аналита в образцах находилась в пределах от 17 до 2253 U/l. Indiko 1) Сравнительное исследование было проведено в соответствии с Руководством по сравнению методов исследования биологических образцов и расчету стандартных ошибок (EP9-A2) Института клинических и лабораторных стандартов (CLSI) с использованием анализатора Indiko и предыдущего теста 981372 CK (IFCC) и анализатора Konelab Prime 60 (x) в качестве эталонного метода. Линейная регрессия (Deming) (единица измерения — U/l): y = 1.10x – 4.7 r = 0.999 n = 95 Концентрация аналита в образцах находилась в пределах от 17 до 2253 U/l. 2) Сравнительное исследование было проведено в соответствии с Руководством по сравнению методов исследования биологических образцов и расчету стандартных ошибок (EP9-A2) Института клинических и лабораторных стандартов (CLSI) с использованием анализатора Indiko и этого теста CK (IFCC) и анализатора Konelab Prime 60 (x) в качестве эталонного метода. Линейная регрессия (Deming) (единица измерения — U/l): y = 1.03x + 1.4 r = 1.00 n = 98 Концентрация аналита в образцах находилась в пределах от 14 до 2450 U/l. СПИСОК ЛИТЕРАТУРЫ 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. e. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. ПРОИЗВОДИТЕЛЬ Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Тел.: +358 10 329 200 www.thermoscientific.com Дата редактирования (гггг-мм-дд) 2015-05-08 Изменения по сравнению с последней версией Изменения указаны линией возле текста. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 32 Page 33 D06025_04_Insert_CK (IFCC)_MU SK 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) Bezpečnostné opatrenia S ľudskými vzorkami zaobchádzajte a likvidujte ich ako potenciálne infekčný materiál. Uskladnenie (4) Vzorky je možné uskladňovať podľa nižšie uvedeného návodu. Vzorka 4 hodiny pri teplote 20…25°C, 1 mesiac pri teplote 4…8 °C alebo 1 mesiac pri teplote -20 °C; chránená pred svetlom. CK-BB nestabilná. TENTO PRÍBALOVÝ LETÁK JE URČENÝ PRE KRAJINY MIMO USA. POUŽITIE Na diagnostické použitie in vitro pri kvantitatívnom stanovení aktivity kreatín kinázy (ATP: kreatín N-fosfotransferáza (CK), EC 2.7.3.2) v ľudskom sére alebo plazme v analyzátoroch Konelab a Indiko. Akákoľvek zmienka o systémoch Konelab sa vzťahuje aj na sériu T. SÚHRN (1, 2) Aktivita kreatín kinázy je najväčšia v priečne pruhovanom svale, mozgu a srdcovom tkanive. V iných tkanivách, ako sú obličky, bránica a prostata, je aktivita značne nižšia. Aktivita CK sa zvyšuje pri mnohých ochoreniach, ku ktorým patria ochorenia kostrového svalstva, srdca, centrálneho nervového systému a štítnej žľazy. U 50 % pacientov sa po akútnom poškodení svalov začne zvyšovať aktivita CK v krvi a patologické hladiny dosahuje v priemere po 4-5 hodinách. Hodnoty sa pravidelne zvyšujú od ôsmej do dvadsiatej štvrtej hodiny po začiatku poškodenia a znižujú sa na pôvodné hodnoty v referenčnom rozpätí s veľkými rozdielmi medzi jednotlivcami. PRINCÍP METÓDY Test CK predstavuje tekutý test zložený z dvoch častí, ktorý využíva tri enzýmy na určenie aktivity CK. Kreatín kináza katalyzuje reakciu medzi kreatínfosfátom a ADP (adenozín-5'-difosfát), pri ktorej sa tvorí kreatín a ATP (adenozín-5'-trifosfát). Nová zlúčenina za prítomnosti HK (hexokinázy) fosforyluje glukózu na G-6-P (glukóza-6-fosfát). G-6-P sa oxiduje na 6-PG (glukonát-6-fosfát) a súčasne sa ekvimolárne množstvo NADP (nikotínamid adenín dinukleotidfosfát) redukuje na NADPH v reakcii katalyzovanej G-6P-DH (glukóza-6-fosfát dehydrogenáza). Zvýšenie absorbancie pri 340 nm, ktoré je výsledkom vzniku NADPH, je priamo úmerné aktivite kreatín kinázy vo vzorke (3). CK Kreatín fosfát + ADP → Kreatín + ATP HK ATP + Glukóza → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ Súprava s kódom 981829 obsahuje 2 fľaštičky s 20 ml činidla A 2 fľaštičky s 5 ml činidla B Koncentrácie Činidlo A Pufer (Imidazol), pH 6.5 D-glukóza N-acetyl-L-cysteín (NAC) Octan horečnatý EDTA-Na2 Nikotínamid adenín dinukleotidfosfát (NADP) Hexokináza (HK) NaN3 Činidlo B Pufer (Imidazol) Adenozín-5’-difosfát (ADP) Adenozín-5’-monofosfát (AMP) Di(adenozín-5’)penta-fosfát (AP5A) Glukóza-6-fosfátdehydrogenáza (G-6-P-DH) EDTA -Na2 Kreatín fosfát NaN3 Nastavenie prístroja Indiko Pomocou čiarového kódu na aplikačnom formulári Indiko alebo pomocou elektronického súboru zadajte aplikačné parametre. Poskytnuté materiály Činidlá popísané vyššie. Materiály potrebné, ale nedodané Kalibrátor a kontroly uvedené nižšie. Kalibrácia Odozva (dA/min) sa konvertuje na výsledné jednotky pomocou koeficientu prepočtu. Ako kalibrátor môžete použiť aj eCal (981830) podľa inštrukcií uvedených v príbalovom letáku eCal (981830). Kontrola kvality Používajte vzorky na kontrolu kvality (QC) aspoň raz denne, po každej kalibrácii a vždy, keď použijete novú fľaštičku s činidlom. Odporúčame použiť kontroly alebo skupiny vzoriek aspoň dvoch úrovní (nízkej a vysokej). Pri kontrole kvality vždy dodržiavajte miestne, štátne a federálne predpisy. Dostupné kontroly: CK-MB Control, kód 981908 CK-MB Control High, kód 981255 Nortrol, kód 981043 Abtrol, kód 981044 Intervaly a hraničné hodnoty kontroly (QC) sa musia prispôsobiť individuálnym požiadavkám laboratória. Výsledné hodnoty kontrolných vzoriek majú byť v rozmedzí hodnôt vopred stanovených laboratóriom. VÝPOČET VÝSLEDKOV Výsledky automaticky vypočíta analyzátor takýmto spôsobom: INFORMÁCIE O ČINIDLÁCH Súprava s kódom 981828 obsahuje 2 fľaštičky s 52 ml činidla A 2 fľaštičky s 13 ml činidla B TESTOVACÍ POSTUP Postup automatizácie analyzátora nájdete v Používateľskej príručke. Účinnosť postupu, ktorý nebol schválený spoločnosťou Thermo Fisher Scientific Oy, nemôže byť garantovaná, a preto musí byť vyhodnotená používateľom. 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < 0.1 % Aktivita v U/l = faktor x dA/min TV x 1000 Teoretický faktor = 6.22 x SV x P TV SV 6.22 P Celkový reagujúci objem v ml Objem vzorky v ml Absorpčný koeficient NADPH pri 340 nm Veľkosť kyvety = 1 Konverzný pomer µkat/l = (U/l)/60 Ak sa eCal (981830) používa ako kalibrátor, výsledky sa automaticky vypočítajú analyzátorom pomocou kalibračnej krivky. Kalibračná krivka (príklad) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < 0.1 % Bezpečnostné opatrenia Reagencia B je nebezpečná. V samostatnom hárku vnútri súpravy nájdete výstražné a bezpečnostné upozornenia: H315, P280, P302+P352, P305+P351+P338. Reagencie obsahujú azid sodný ako konzervačnú látku. Dodržujte štandardné bezpečnostné opatrenia, ktoré sú nevyhnutné pri manipulácii so všetkými laboratórnymi reagenciami. Produkt likvidujte ako laboratórnu chemickú látku v súlade s platnými predpismi. Príprava Činidlá A a B sú pripravené na použitie. Poznámka: Pred vložením fľaštičiek alebo nádobiek s činidlom do analyzátora skontrolujte, či sa v hrdle nádobky alebo na povrchu činidla nenachádzajú žiadne bubliny. Uskladnenie a stabilita Činidlá v neotvorených fľaštičkách sú stabilné pri teplote 2…8 °C až do exspiračnej doby vytlačenej na štítku. Informácie o stabilite činidiel v prístroji nájdete v Používateľskej príručke analyzátora. Ak sú činidlá zakalené, nepoužívajte ich. ODBER VZORIEK Poznámka: Pri spracovaní vzoriek v skúmavkách na odber vzoriek dôkladne dodržiavajte pokyny výrobcu skúmavky, aby ste sa vyhli chybným výsledkom. Zvláštnu pozornosť venujte predanalytickým premenným, ako napr. čas miešania a státia pred odstreďovaním, ako aj nastaveniam odstreďovania. Typ vzorky Môžete používať nehemolyzované sérum alebo heparinizovanú plazmu. Poznámka: Hemolýza môže viesť k nesprávnym výsledkom. Vzorky nevystavujte silnému svetlu. Kalibračná krivka závisí od šarže. OBMEDZENIA METÓDY Pre diagnostické účely majú byť výsledky vždy vyhodnocované spolu s anamnézou pacienta, klinickým vyšetrením a ďalšími zisteniami. Interferencie Vykonala sa štúdia s použitím CLSI dokumentu EP7-A2 ako smernice. Kritérium: Rozdiel do ± 10 % počiatočných hodnôt. Konelab Interferujúca substancia Konjugovaný bilirubín Nekonjugovaný bilirubín Hemoglobín v hemolyzáte Lipémia: Intralipid (obchodná známka spoločnosti Fresenius Kabi AB) Koncentrácia interferujúcej substancie Cieľová hodnota CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l 188 498 180 495 184 505 Rozdiel oproti cieľovej hodnote (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. CK (IFCC) D06025_04_Insert_CK (IFCC)_MU Indiko Interferujúca substancia Konjugovaný bilirubín Nekonjugovaný bilirubín Hemoglobín v hemolyzáte Lipémia: Intralipid (obchodná známka spoločnosti Fresenius Kabi AB) Koncentrácia interferujúcej substancie Cieľová hodnota CK (U/l) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l 179 477 175 469 187 502 Rozdiel oproti cieľovej hodnote (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Nepoužívajte hemolyzovanú vzorku. Korelácia medzi zakalenosťou vzorky a koncentráciou triglyceridov je nízka. Vo veľmi vzácnych prípadoch môže gamopatia, najmä typ IgM (Waldenströmova makroglobulinémia), spôsobiť nespoľahlivosť výsledkov (6). Nebolo testované spoločnosťou Thermo Fisher Scientific Oy. Viac o ďalších interferujúcich substanciách, pozri odkaz (5). OČAKÁVANÉ HODNOTY (3) Muži < 171 U/l (< 2.85 µkat/l) pri teplote 37 °C Ženy < 145 U/l (< 2.41 µkat/l) pri teplote 37 °C Tieto hodnoty slúžia len ako pomôcka. Odporúčame, aby si každé laboratórium overilo tento rozsah alebo prispôsobilo referenčný interval pre konkrétnu populáciu. Zmeny od predchádzajúcej verzie Čiara vedľa textu naznačuje zmeny. MERACÍ ROZSAH 10 – 1000 U/l (0.17 – 17 µkat/l). Rozšírený merací rozsah po sekundárnom zriedení: 10 – 9000 U/l (0.17 – 150 µkat/l). PREVÁDZKOVÉ CHARAKTERISTIKY Výsledky získané v jednotlivých laboratóriách sa môžu líšiť od daných prevádzkových dát. Limit detekcie Vykonala sa štúdia s použitím dokumentu EP17-A ako smernice. Konelab LoB: 0.6 U/l (0.01 µkat/l) Limit slepej vzorky (LoB) predstavuje najvyšší výsledok merania, ktorý je možné pozorovať pri vzorke bez analytu (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Limit detekcie (LoD) predstavuje najnižšie množstvo analytu vo vzorke, ktoré je možné detegovať (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Limit stanoviteľnosti (LoQ) predstavuje najnižšiu skutočnú koncentráciu vo vzorke, ktorú je možné stanoviť kvantitatívne (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Limit slepej vzorky (LoB) predstavuje najvyšší výsledok merania, ktorý je možné pozorovať pri vzorke bez analytu (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Limit detekcie (LoD) predstavuje najnižšie množstvo analytu vo vzorke, ktoré je možné detegovať (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Limit stanoviteľnosti (LoQ) predstavuje najnižšiu skutočnú koncentráciu vo vzorke, ktorú je možné stanoviť kvantitatívne (n = 42). Nepresnosť Nepresnosť analýzy CK je ≤ 5.0 % celkovej hodnoty CV. Opakovateľnosť (v rámci cyklu analýz) Pre dané zariadenie (celkovo) Priemer 92 U/l SD CV % 1.7 1.8 Priemer 226 U/l SD CV % 2.0 0.9 Priemer 616 U/l SD CV % 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 V súlade s dokumentom CLSI EP5-A2 použitým ako smernica sa konala štúdia presnosti na jednom pracovisku s tromi analyzátormi Konelab 60, jedným operátorom, 2 šaržami činidiel, jednou šaržou kalibrátorov a 6 kalibráciami po dobu 21 dní a s počtom meraní n = 84. Výsledky z každého analyzátora boli skombinované. Indiko CK Opakovateľnosť (v rámci cyklu analýz) Pre dané zariadenie (celkovo) VÝROBCA Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Dátum revízie (rrrr-mm-dd) 2015-05-08 Konverzný faktor µkat/l = (U/l)/60 Konelab CK BIBLIOGRAFIA 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. Priemer 95 U/l SD CV % 1.2 1.2 Priemer 226 U/l SD CV % 2.0 0.9 Priemer 637 U/l SD CV % 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 V súlade s dokumentom CLSI EP5-A2 použitým ako smernica sa konala štúdia presnosti na jednom pracovisku s tromi analyzátormi Indiko, jedným operátorom, 2 šaržami činidiel, jednou šaržou kalibrátorov a 7 kalibráciami po dobu 21 dní a s počtom meraní n = 84. Výsledky z každého analyzátora boli skombinované. Porovnanie metódy Konelab Vykonala sa porovnávacia štúdia na analyzátore Konelab Prime 60 s použitím CLSI dokumentu EP9-A2 ako smernice a predchádzajúcej metódy 981372 CK (IFCC) ako referencie. Lineárna regresia (Deming) (výsledok v jednotkách U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Koncentrácie vzoriek sa pohybovali v rozmedzí od 17 do 2253 U/l. Indiko 1) Vykonala sa porovnávacia štúdia na analyzátore Indiko s použitím CLSI dokumentu EP9-A2 ako smernice a predchádzajúcej metódy 981372 CK (IFCC) na analyzátore Konelab Prime 60 (x) ako referencie. Lineárna regresia (Deming) (výsledok v jednotkách U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Koncentrácie vzoriek sa pohybovali v rozmedzí od 17 do 2253 U/l. 2) Vykonala sa porovnávacia štúdia na analyzátore Indiko s použitím dokumentu CLSI EP9-A2 ako smernice a tej istej metódy CK (IFCC) na analyzátore Konelab Prime 60 (x) ako referencie. Lineárna regresia (Deming) (výsledok v jednotkách U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Koncentrácie vzoriek sa pohybovali v rozmedzí od 14 do 2450 U/l. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 34 Page 35 D06025_04_Insert_CK (IFCC)_MU SV CK (IFCC) 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) DENNA BIPACKSEDEL GÄLLER KUNDER UTANFÖR USA. AVSEDD ANVÄNDNING För in vitro-diagnostik vid kvantitativ bestämning kreatinkinasaktivitet (ATP: kreatin-Nfosfotransferas (CK), EC 2.7.3.2) i humant serum eller plasma på Konelab- och Indikoanalysatorer. Eventuella hänvisningar till Konelab-system avser även T-serien. SAMMANFATTNING (1, 2) Kreatinkinasaktivitet är störst i tvärstrimmig muskulatur och hjärn- och hjärtvävnad. Annan typ av vävnad, till exempel njurar, diafragma och prostata, har avsevärt mindre aktivitet. CK-aktiviteten är förhöjd vid många sjukdomar, till exempel där skelettmuskulaturen, hjärtat, centrala nervsystemet eller sköldkörteln påverkas. Efter en akut muskelskada börjar den totala CK-aktiviteten i blodet att öka och når patologiska nivåer efter i genomsnitt 4–5 timmar för 50 % av patienterna. Värdena ökar regelbundet från den 8:e till den 24:e timmen efter att hjärtinfarkten inträffar och återgår till referensintervallet med stora fluktuationer mellan enskilda värden. TESTPRINCIP CK-analysen är en vätskeanalys i två delar där tre enzymer utnyttjas för bestämning av CK-aktivitet. Kreatinekinas katalyserar reaktionen mellan kreatinfosfat och ADP (adenosin-5’-difosfat) genom att kreatin och ATP (adenosin-5’-trifosfat) bildas. Den senare substansen fosforylerar glukos till G-6-P (glukos-6-fosfat) vid förekomst av HK (hexokinas). G-6-P oxideras till 6-PG (glukonat-6-fosfat) och inom samma tidsperiod reduceras samma molmängd NADP (nikotinamid-adenindinukleotid-fosfat) till NADPH i en reaktion som katalyseras av G-6P-DH (glukos-6-fosfatdehydrogenas). Ökningen i absorbans vid 340 nm, som uppstår genom att NADPH bildas, är proportionell mot aktiviteten för kreatinkinas i provet (3). Prov 4 timmar i 20…25 °C, 1 månad i 4…8 °C eller 1 månad i -20 °C; skyddat från ljus. CK-BB inte stabilt. TESTUTFÖRANDE Se Application Notes för automatiskt utförande på aktuell analysator. Varje applikation som inte har validerats av Thermo Fisher Scientific Oy kan inte garanteras vad gäller prestanda och måste därför utvärderas av användaren. Indiko - instrumentinställningar Mata in parametrar för tillämpningen via streckkoden på Indiko tillämpningsblad eller via elektronisk fil, efter behov. Bifogat material Reagens enligt ovan. Erforderligt material som ej medföljer Kalibrator och kontroller enligt nedan. Kalibrering Svaret (dA/min) omvandlas till resultatenheter med en omräkningsfaktor. Använd alternativt eCal (981830) som en kalibrator, enligt anvisningarna som ges på bipacksedeln för eCal (981830). HK ATP + Glukos → ADP + G-6-P Kvalitetskontroll Använd kvalitetskontrollprover (QC) minst en gång om dagen, efter varje kalibrering och varje gång en ny reagensflaska används. Vi rekommenderar att kontroller eller prov-pooler på minst två nivåer används (låg och hög). Följ alltid lokala, regionala och statliga bestämmelser vid kvalitetskontroll. Tillgängliga kontroller: CK-MB Control, kod 981908 CK-MB Control High, kod 981255 Nortrol, artikelnr. 981043 Abtrol, artikelnr: 981044 Kontrollens intervall och gränser måste anpassas till laboratoriets egna krav. Resultaten från kvalitetskontrollproverna bör ligga inom de av laboratoriet fastställda gränserna. G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ RESULTATBERÄKNING Resultaten beräknas automatiskt av analysatorn enligt följande: REAGENSINNEHÅLL Aktivitet i U/l = Faktor x dA/min CK Kreatinfosfat + ADP → Kreatin + ATP Satsartikelnr 981828 innehåller 2 flaskor med 52 ml Reagens A 2 flaskor med 13 ml Reagens B Satsartikelnr 981829 innehåller 2 flaskor med 20 ml Reagens A 2 flaskor med 5 ml Reagens B TV SV 6.22 P Koncentrationer Reagens A Buffert (Imidazole), pH 6.5 D-glukos N-acetyl-L-cystein (NAC) Magnesiumacetat EDTA-Na2 Nikotinamid-adenindinukleotid-fosfat (NADP) Hexokinas (HK) NaN3 Reagens B Buffert (Imidazole) Adenosin-5’-difosfat (ADP) Adenosin-5’-monofosfat (AMP) Di(adenosin-5’)pentafosfat (AP5A) Glukos-6-fosfatdehydrogenas (G-6-P-DH) EDTA -Na2 Kreatinfosfat NaN3 TV x 1000 Teoretisk faktor = 6.22 x SV x P 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5 000 U/l < 0.1 % Total reaktionsvolym i ml Provvolym i ml Extinktionskoefficient för NADPH vid 340 nm Kyvettväglängd = 1 Konverteringsfaktor µkat/l = (U/l)/60 Om eCal (981830) används som en kalibrator kommer resultaten att beräknas automatiskt av analysatorn med hjälp av en kalibreringskurva. Kalibreringskurva (exempel) 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14 000 U/l 2 mmol/l 160 mmol/l < 0.1 % Försiktighetsåtgärder Reagens B är farlig. Se separat blad inuti satsen för varnings- och försiktighetsfraser: H315, P280, P302+P352, P305+P351+P338. I reagenser finns natriumazid som konserveringsmedel. Vidta normala försiktighetsåtgärder som vid all hantering av laboratoriereagenser. Produkten måste kasseras som laboratoriekemikalier enligt lokala bestämmelser. Preparering Reagens A och B är färdiga för användning. Anmärkning: Kontrollera att det inte finns bubblor i flaskhalsen eller på ytan av reagenset då behållarna placeras i analysatorn. Förvaring och hållbarhet Reagens i oöppnad behållare är hållbar vid 2…8 °C till på etiketten angivet utgångsdatum. Se Application Notes för aktuell analysator för information om reagensernas hållbarhet i instrumentet. Om reagenserna är grumliga ska de inte användas. PROVTAGNING Anmärkning: Vid bearbetning av prover i provrör ska du följa anvisningarna från rörtillverkare noggrant för att undvika felaktiga resultat. Var särskilt uppmärksam på preanalytiska variabler, t.ex. blandning, väntetid före centrifugering och centrifugens inställningar. Provmaterial Ohemolyserat serum eller heparinplasma kan användas. Anmärkning: Hemolys kan leda till avvikande resultat. Exponera inte prover för starkt ljus. Försiktighetsåtgärder Humana prover ska behandlas som potentiellt smittförande, både vid hantering och kassering. Kalibreringskurvan är batchberoende. BEGRÄNSNINGAR I UTFÖRANDET För diagnostik ska resultaten alltid bedömas tillsammans med patientens anamnes, kliniska undersökningar och andra resultat. Interferens En studie har utförts enligt anvisningarna i CLSI dokument EP7-A2. Kriterium: Skillnad inom ± 10 % av initialvärdena. Konelab Interfererande ämne Konjugerat bilirubin Okonjugerat bilirubin Hemoglobin i hemolysat Lipemi: Intralipid (varumärket tillhör Fresenius Kabi AB) Interferenskoncentration 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l CKmålvärde (U/l) 188 498 180 495 184 505 Skillnad från mål (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. 981828 981829 Förvaring (4) Provet kan förvaras enligt nedan. D06025_04_Insert_CK (IFCC)_MU Indiko Interfererande ämne Interferenskoncentration Konjugerat bilirubin Okonjugerat bilirubin Hemoglobin i hemolysat Lipemi: Intralipid (varumärket tillhör Fresenius Kabi AB) 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l CKmålvärde (U/l) 179 477 175 469 187 502 Skillnad från mål (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Använd ej hemolyserat prov. Det är dålig korrelation mellan grumlighet och triglyceridkoncentration. I mycket sällsynta fall kan gammopati, särskilt typ IgM (Waldenströms makroglobulinemi), ge otillförlitliga resultat (6). Ej testat hos Thermo Fisher Scientific Oy. För andra interfererande ämnen, se referens (5). REFERENSOMRÅDE (3) Män < 171 U/l (< 2.85 µkat/l) vid 37 °C Kvinnor < 145 U/l (< 2.41 µkat/l) vid 37 °C Angivna värden är endast avsedda som vägledning. Vi rekommenderar att varje laboratorium verifierar detta område eller fastställer ett referensintervall för populationen som betjänas. REFERENSER 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 - 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 - 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 - 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 -1243. Konverteringsfaktor µkat/l = (U/l)/60 TILLVERKARE Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa, Finland Tel. +358 10 329 200 www.thermoscientific.com MÄTOMRÅDE 10 – 1000 U/l (0.17 – 17 µkat/l). Utökat mätområde efter sekundär spädning: 10 – 9000 U/l (0.17 – 150 µkat/l). Revisionsdatum (åååå-mm-dd) 2015-05-08 UTFÖRANDETS KARAKTERISTIKA Resultaten som erhålls vid varje enskilt laboratorium kan skilja sig från angivna data för prestanda. Detekteringsgräns En studie har utförts enligt anvisningarna i EP17-A-dokumentet. Konelab LoB: 0.6 U/l (0.01 µkat/l) Gräns för blankprov (LoB) representerar det högsta mätresultat som sannolikt kan förekomma för ett analytfritt prov (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Detekteringsgränsen (LoD) representerar den lägsta mängden analyt i ett prov som kan detekteras (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Kvantiteringsgränsen (LoQ) representerar den lägsta faktiska koncentrationen i ett prov som kan bestämmas kvantitativt (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Gräns för blankprov (LoB) representerar det högsta mätresultat som sannolikt kan förekomma för ett analytfritt prov (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Detekteringsgränsen (LoD) representerar den lägsta mängden analyt i ett prov som kan detekteras (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Kvantiteringsgränsen (LoQ) representerar den lägsta faktiska koncentrationen i ett prov som kan bestämmas kvantitativt (n = 42). Imprecision Imprecisionen för CK-analysen är ≤ 5.0 % Total CV. Konelab CK Repeterbarhet (inom serie) Inom instrument (totalt) Medel 92 U/l SD CV% 1.7 1.8 Medel 226 U/l SD CV% 2.0 0.9 Medel 616 U/l SD CV% 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Precisionsstudien utfördes enligt anvisningarna i CLSI:s dokument EP5-A2 på en plats, med tre Konelab 60-analysatorer, en användare, två reagenspartier, ett kalibratorparti och sex kalibreringar under 21 dagar. Antalet mätningar var n = 84. Resultaten från varje analysator kombinerades. Indiko CK Repeterbarhet (inom serie) Inom instrument (totalt) Medel 95 U/l SD CV% 1.2 1.2 Medel 226 U/l SD CV% 2.0 0.9 Medel 637 U/l SD CV% 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Precisionsstudien utfördes enligt anvisningarna i CLSI Dokument EP5-A2 på en plats, med tre Indikoanalysatorer, en användare, två reagenspartier, ett kalibratorparti och sju kalibreringar under 21 dagar. Antalet mätningar var n=84. Resultaten från varje analysator kombinerades. Metodjämförelse Konelab En jämförelsestudie har utförts på en Konelab Prime 60-analysator enligt anvisningarna i CLSI:s dokument EP9-A2 och med föregående 981372 CK(IFCC)-metod (x) som referens. Linjär regression (Deming) (resultatenhet U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Provens koncentration låg mellan 17 och 2253 U/l. Indiko 1) En jämförelsestudie har utförts på en Indiko-analysator enligt anvisningarna i CLSI:s dokument EP9-A2 och med föregående 981372 CK(IFCC)-metod på Konelab Prime 60 (x) som referens. Linjär regression (Deming) (resultatenhet U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Provens koncentration låg mellan 17 och 2253 U/l. 2) En jämförelsestudie har utförts på en Indiko-analysator enligt anvisningarna i CLSI:s dokument EP9-A2 och med samma CK(IFCC)-metod på Konelab Prime 60 (x) som referens. Linjär regression (Deming) (resultatenhet U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Provens koncentration låg mellan 14 och 2450 U/l. Ändringar från tidigare utgåva Raden vid sidan av texten anger ändringar. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 36 Page 37 D06025_04_Insert_CK (IFCC)_MU TR 981828 981829 2 x 65 ml (Konelab™) 2 x 25 ml (Konelab™, Indiko™) BU PROSPEKTÜS ABD DIŞINDA KULLANIM İÇİN GEÇERLİDİR. KULLANIM AMACI Konelab ve Indiko analizörlerinde insan serum veya plazmasında kreatin kinaz aktivitesinin miktar tayininde in vitro tanı amaçlı kullanım içindir (ATP: Kreatin N-Fosfotransferaz (CK), EC 2.7.3.2). Konelab sistemlerine yönelik referanslar aynı zamanda T Serisi için de geçerlidir. ÖZET (1,2) En yüksek kreatin kinaz aktivitesi çizgili kas, beyin ve kalp dokusunda görülür. Böbrekler, diyafram ve prostat gibi diğer dokular anlamlı ölçüde daha az aktivite içerir. İskelet kası, kalp, merkezi sinir sistemi ve tiroid ile ilişkili olanlar da dahil olmak üzere birçok hastalıkta CK aktivitesi yükselir. Akut kas hasarından sonra kandaki total CK aktivitesi yükselmeye başlayarak hastaların %50'sinde ortalama 4 – 5 saat içinde patolojik düzeylere ulaşır. Değerler enfarktüsün başlangıcından itibaren 8. – 24. saatten itibaren düzenli olarak yükselir ve referans aralığa geri döner, bu süreçte hastalar arasında büyük dalgalanmalar görülür. PROSEDÜR PRENSIBI CK testi, CK aktivitesini belirlemek üzere üç enzimin kullanıldığı iki bölümden oluşan bir sıvı testidir. Kreatin kinaz; kreatin ve ATP (adenozin-5’-trifosfat) oluşumuyla kreatin fosfat ile ADP (adenozin-5’-difosfat) arasındaki reaksiyonu katalize eder. ATP bileşiği, HK (heksokinaz) varlığında glukozu G-6-P'ye (glukoz-6-fosfat) fosforile eder. G-6-P, 6-PG'ye (glukonat-6fosfat) oksitlenir ve aynı zamanda, eşmolar miktarda NADP (nikotinamid adenin dinükleotid fosfat), G-6P-DH (glukoz-6-fosfat dehidrojenaz) tarafından katalize edilen bir reaksiyonda NADPH'ye indirgenir. NADPH oluşumu nedeniyle 340 nm'deki absorbansta meydana gelen artış, numunedeki kreatin kinaz aktivitesiyle orantılıdır (3). CK Kreatin fosfat + ADP → Kreatin + ATP HK ATP + Glukoz → ADP + G-6-P G-6-P-DH G-6-P + NADP+ → 6-PG + NADPH + H+ REAKTIF BILGILERI Kit kodu 981828 şunları içerir: 2 flakon 52 ml Reaktif A 2 flakon 13 ml Reaktif B Kit kodu 981829 şunları içerir: 2 flakon 20 ml Reaktif A 2 flakon 5 ml Reaktif B Konsantrasyonlar Reaktif A Tampon (İmidazol), pH 6.5 D-glukoz N-asetil-L-sistein (NAC) Magnezyum asetat EDTA-Na2 Nikotinamid adenin dinükleotid fosfat (NADP) Heksokinaz (HK) NaN3 Reaktif B Tampon (İmidazol) Adenozin-5’-difosfat (ADP) Adenozin-5’-monofosfat (AMP) Di(adenozin-5’)penta-fosfat (AP5A) Glukoz-6-fosfat dehidrojenaz (G-6-P-DH) EDTA -Na2 Kreatin fosfat NaN3 60 mmol/l 27 mmol/l 27 mmol/l 14 mmol/l 2 mmol/l 2.7 mmol/l ≥ 5000 U/l < %0.1 160 mmol/l 11 mmol/l 28 mmol/l 55 µmol/l ≥ 14000 U/l 2 mmol/l 160 mmol/l < %0.1 Önlemler İnsan numuneleri kullanırken veya atılırken potansiyel olarak enfeksiyöz nitelikte olduğu kabul edilmelidir. Saklama (4) Numuneler aşağıda gösterildiği gibi saklanabilir. Numune 20…25°C'de 4 saat, 4…8°C'de 1 ay veya -20°C'de 1 ay; ışıktan korunmak kaydıyla. CK-BB stabil değildir. TEST PROSEDÜRÜ Analizörünüzdeki otomatik bir prosedür için Uygulama Notlarına bakın. Thermo Fisher Scientific Oy tarafından valide edilmemiş uygulamaların performansına ilişkin teminat verilememektedir; bu nedenle söz konusu uygulamalar kullanıcı tarafından değerlendirilmelidir. Indiko cihaz ayarları Indiko uygulama talimatları sayfasındaki barkod veya elektronik dosyada (duruma göre) bulunan bilgilere göre uygulama parametrelerini girin. Sağlanan malzemeler Yukarıda açıklanan reaktifler. Gerekli olan ancak sağlanmayan malzemeler Kalibratör ve aşağıda belirtilen kontroller. Kalibrasyon Yanıt (dA/dak.), bir hesaplama faktörüyle sonuç birimlerine dönüştürülür. Ya da eCal'in (981830) prospektüsünde verilen talimatlara göre kalibratör olarak eCal'i (981830) kullanın. Kalite Kontrol Günde en az bir kez ve her kalibrasyondan sonra kalite kontrol (QC) numunelerini kullanın; her seferinde yeni bir reaktif şişesi kullanılır. En az iki kontrol seviyesi (düşük ve yüksek) veya numune havuzu kullanılması önerilir. Kalite kontrol gerçekleştirirken daima yerel ve federal düzenlemelerle devlet düzenlemelerine uyun. Kullanılabilir kontroller: CK-MB Control, kod 981908 CK-MB Control High, kod 981255 Nortrol, kod 981043 Abtrol, kod 981044 Kontrol (QC) aralıkları ve limitleri, ayrı laboratuvar gereksinimlerine uyarlanmalıdır. Kalite kontrol numunelerinin sonuçları, laboratuvar tarafından önceden ayarlanmış limitler dahilinde olmalıdır. SONUÇLARIN HESAPLANMASI Sonuçlar, analizör tarafından otomatik olarak aşağıdaki gibi hesaplanır: U/l cinsinden aktivite = Faktör x dA/dak. TV x 1000 Teorik faktör = 6.22 x SV x P TV SV 6.22 P ml cinsinden toplam reaksiyon hacmi ml cinsinden numune hacmi 340 nm'de NADPH'nin sönüm katsayısı Küvet yol uzunluğu = 1 Dönüştürme faktörü µkat/l = (U/l)/60 eCal (981830), kalibratör olarak kullanılırsa sonuçlar, bir kalibrasyon eğrisi kullanılarak analizör tarafından otomatik olarak hesaplanır. Kalibrasyon Eğrisi (örnek) Önlemler Reaktif B tehlikelidir. Tehlike ve Önlem ifadeleri için kitin içindeki ayrı belgeye bakın: H315, P280, P302+P352, P305+P351+P338. Reaktifler koruyucu olarak sodyum azit içerir. Tüm laboratuvar reaktiflerinin kullanımı için gerekli normal önlemleri uygulayın. Ürün, yerel düzenlemelere göre laboratuvar kimyasalı olarak atılmalıdır. Preparasyon Reaktif A ve B kullanıma hazırdır. Not: Reaktif flakonlarını veya kaplarını analizöre yerleştirirken şişe boynu veya reaktif yüzeyinde kabarcık olmadığını kontrol edin. Saklama ve Stabilite Açılmamış flakonlardaki reaktifler 2…8°C sıcaklıkta etiket üzerinde belirtilen son kullanma tarihine kadar kararlıdır. Reaktiflerin cihazdaki kararlılığı hakkında detaylı bilgi almak için analizörünüzün Uygulama Notları kısmına bakın. Türbidite gösteren reaktifleri kullanmayın. NUMUNE TOPLAMA Not: Numune toplama tüplerinde numune işlerken hatalı sonuçları önlemek için tüp imalatçısının talimatlarını dikkatli bir şekilde izleyin. Karıştırma, santrifügasyon öncesi bekleme süresi ve santrifüj ayarları gibi preanalitik değişkenlere ayrıca dikkat edin. Numune Tipi Hemolize olmayan serum veya heparin plazma kullanılabilir. Not: Hemoliz farklı sonuçlara neden olabilir. Numunelerin güçlü ışığa maruz kalmasını önleyin. Kalibrasyon eğrisi lota bağlıdır. PROSEDÜR SINIRLAMALARI Tanı amaçlı olarak, sonuçlar daima hastanın tıbbi geçmişi, klinik muayenesi ve diğer bulgularıyla bağlantılı olarak değerlendirilmelidir. Etkileşim Kılavuz olarak CLSI dokümanı EP7-A2'nin kullanıldığı bir çalışma yürütülmüştür. Kriter: Başlangıç değerlerinin %10'u içinde ± fark. Konelab Etkileşen Madde Konjuge bilirubin Konjuge olmayan bilirubin Hemolizat içinde hemoglobin Lipemi: Intralipid (Fresenius Kabi AB'nin ticari markası) Etkileşime Neden Olan Konsantrasyon 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 677 µmol/l (39.6 mg/dl) 2.1 g/l 2.4 g/l Hedef CK (U/l) 188 498 180 495 184 505 Hedefe göre Fark (%) - 5.3 - 5.8 - 1.7 - 2.9 + 8.5 + 4.6 10 g/l 10 g/l 181 504 + 2.8 + 7.1 Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. CK (IFCC) D06025_04_Insert_CK (IFCC)_MU Indiko Etkileşen Madde Konjuge bilirubin Konjuge olmayan bilirubin Hemolizat içinde hemoglobin Lipemi: Intralipid (Fresenius Kabi AB'nin ticari markası) Etkileşime Neden Olan Konsantrasyon 592 µmol/l (34.6 mg/dl) 601 µmol/l (35.2 mg/dl) 621 µmol/l (36.3 mg/dl) 676 µmol/l (39.5 mg/dl) 2.5 g/l 2.4 g/l Hedef CK (U/l) 179 477 175 469 187 502 Hedefe göre Fark (%) - 0.8 - 2.4 +0.2 - 0.1 + 8.6 + 3.4 10 g/l 10 g/l 185 505 + 1.0 - 0.8 Hemolize örnek kullanmaktan kaçının. Türbidite ve trigliserit konsantrasyonu arasında zayıf bir korelasyon vardır. Çok nadir durumlarda gamopati, özellikle tip IgM (Waldenström makroglobülinemisi) güvenilir olmayan sonuçlara neden olabilir (6). Thermo Fisher Scientific Oy'da test edilmemiştir. Etkileşen diğer maddeler için lütfen referansa bakın (5). BEKLENEN DEĞERLER (3) Erkek 37 °C'de < 171 U/l (< 2.85 µkat/l) Kadın 37 °C'de < 145 U/l (< 2.41 µkat/l) Alıntılanan değerlerin yalnızca kılavuz görevi görmesi amaçlanmıştır. Her laboratuvarın bu aralığı doğrulaması veya hizmet ettiği popülasyon için bir referans aralığı elde etmesi önerilir. Dönüştürme faktörü µkat/l = (U/l)/60 ÖLÇÜM ARALIĞI 10 – 1000 U/l (0.17 – 17 µkat/l). İkincil dilüsyondan sonra genişletilmiş ölçüm aralığı: 10 – 9000 U/l (0.17 – 150 µkat/l). PERFORMANS ÖZELLIKLERI Ayrı laboratuvarlarda elde edilen sonuçlar, sağlanan performans verilerinden farklılık gösterebilir. Saptama Limiti Kılavuz olarak EP17-A'nın kullanıldığı bir çalışma yürütülmüştür. Konelab LoB: 0.6 U/l (0.01 µkat/l) Boşluk Limiti (LoB), analit içermeyen bir numune için gözlenebilecek en yüksek ölçüm sonucunu temsil eder (n = 60). LoD: 1.5 U/l (0.03 µkat/l) Saptama Limiti (LoD), numune içinde saptanabilen en düşük analit miktarını temsil eder (n = 60). LoQ: 3.2 U/l (0.05 µkat/l) Miktar Tayini Limiti (LoQ), numune içinde kantitatif olarak belirlenebilen en düşük gerçek konsantrasyonu temsil eder (n = 41). Indiko LoB: 0.2 U/l (0.003 µkat/l) Boşluk Limiti (LoB), analit içermeyen bir numune için gözlenebilecek en yüksek ölçüm sonucunu temsil eder (n = 60). LoD: 2.2 U/l (0.04 µkat/l) Saptama Limiti (LoD), numune içinde saptanabilen en düşük analit miktarını temsil eder (n = 60). LoQ: 3.8 U/l (0.06 µkat/l) Miktar Tayini Limiti (LoQ), numune içinde kantitatif olarak belirlenebilen en düşük gerçek konsantrasyonu temsil eder (n = 42). Hatalı ölçüm CK testinin hatalı ölçümü, Toplam CV'nin %5.0'inden az veya buna eşittir. Konelab CK Tekrarlanabilirlik (Çalışma dahilinde) Cihaz içinde (Toplam) Ortalama 92 U/l SD CV %'si 1.7 1.8 Ortalama 226 U/l SD CV %'si 2.0 0.9 Ortalama 616 U/l SD CV %'si 7.1 1.2 2.5 3.1 22.3 2.8 1.4 3.6 Üç Konelab 60 analizörü, bir operatör, 2 reaktif lotu, bir kalibratör lotu ve 6 kalibrasyon yardımıyla bir merkezde 21 günlük süre içinde n = 84 ölçümün yapıldığı ve CLSI Dokümanı EP5-A2'nin kılavuz olarak kullanıldığı bir hassasiyet çalışması yürütülmüştür. Her analizörün verdiği sonuçlar birleştirilmiştir. Indiko CK Tekrarlanabilirlik (Çalışma dahilinde) Cihaz içinde (Toplam) Ortalama 95 U/l SD CV %'si 1.2 1.2 Ortalama 226 U/l SD CV %'si 2.0 0.9 Ortalama 637 U/l SD CV %'si 4.1 0.6 1.7 2.8 11.7 1.7 1.2 1.8 Üç Indiko analizörü, bir operatör, 2 reaktif lotu, bir kalibratör lotu ve 7 kalibrasyon yardımıyla bir merkezde 21 günlük süre içinde n = 84 ölçümün yapıldığı ve CLSI Dokümanı EP5-A2'nin kılavuz olarak kullanıldığı bir hassasiyet çalışması yürütülmüştür. Her analizörün verdiği sonuçlar birleştirilmiştir. Yöntem Karşılaştırması Konelab Konelab Prime 60 üzerinde, CLSI Dokümanı EP9-A2'nin kılavuz olarak ve önceki 981372 CK (IFCC) yönteminin (x) referans olarak kullanıldığı bir karşılaştırma çalışması yürütülmüştür. Doğrusal regresyon (Deming) (sonuç birimi U/l): y = 1.07 x – 6.3 r = 0.999 n = 95 Numune konsantrasyonları 17 ve 2253 U/l arasında olmuştur. Indiko 1) Indiko üzerinde, CLSI Dokümanı EP9-A2'nin kılavuz olarak ve Konelab Prime 60'ta uygulanan önceki 981372 CK (IFCC) yönteminin (x) referans olarak kullanıldığı bir karşılaştırma çalışması yürütülmüştür. Doğrusal regresyon (Deming) (sonuç birimi U/l): y = 1.10 x – 4.7 r = 0.999 n = 95 Numune konsantrasyonları 17 ve 2253 U/l arasında olmuştur. 2) Indiko üzerinde, CLSI Dokümanı EP9-A2'nin kılavuz olarak ve Konelab Prime 60'ta uygulananla aynı CK (IFCC) yönteminin (x) referans olarak kullanıldığı bir karşılaştırma çalışması yürütülmüştür. Doğrusal regresyon (Deming) (sonuç birimi U/l): y = 1.03 x + 1.4 r = 1.00 n = 98 Numune konsantrasyonları 14 ve 2450 U/l arasında olmuştur. BIBLIYOGRAFYA 1) Burtis, CA and Ashwood, E R (ed.), Tietz Fundamentals of Clinical Chemistry, 5th edition, W B Saunders Company, Philadelphia, 2001, pp. 357 – 358. 2) Thomas, L. (ed.), Clinical Laboratory Diagnostics; Use and Assessment of Clinical Laboratory Results, 1st edition, TH-Books Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998, p. 73. 3) IFCC 2002/3: IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C, Part. 2. Reference Procedure for the Measurement of Catalytic Concentrations of Creatine Kinase, Clin.Chem.Lab.Med. 2002, 40(6): 635 – 642. 4) WHO/DIL/LAB/99.1 Rev.2, 2002: Use of Anticoagulants in Diagnostic Laboratory Investigations & Stability of blood, plasma and serum samples. Geneva, 15 January 2002, p. 28. 5) Young, D.S., Effects of Drugs on Clinical Laboratory Tests, Fifth Edition, AACC Press, Washington, D.C., 3-234 – 3-239, 2000. 6) Bakker A.J. et. al. Gammopathy interference in clinical chemistry assays: mechanisms, detection and prevention. Clin Chem Lab Med 2007; 45(9), 1240 – 1243. İMALATÇI Thermo Fisher Scientific Oy Analyzers & Automation Clinical Diagnostics Ratastie 2, P.O. Box 100, FI-01621 Vantaa Tel. +358 10 329 200 www.thermoscientific.com Revizyon tarihi (yyyy-aa-gg) 2015-05-08 Önceki versiyona göre değişiklikler Metnin yanındaki satır değişiklikleri ifade eder. Template: D01162_5 © 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. Page 38