Wavetek

Transcription

Wavetek
Model 9100
Universal Calibration System
W
Model 9100
Model 9100 — the world’s best value
multi-product calibrator
◆ Calibrates Over 15 Different Categories of General-Purpose Test Equipment
◆ Options for Power Meter, Insulation / Continuity Tester and Oscilloscope
Calibration — Internally Installed and Retrofittable
◆ Semi-Automated and Fully Automated Procedure Modes for Maximum
Calibration Throughput
◆ Fully Supported by Portocal-II/9010 Calibration Software and Procedure
Libraries (updates via the World Wide Web from www.wavetek.com)
◆ Rapid Return on Investment
◆ Zero-Downtime Support Using Wavetek’s Model 4950 MTS
◆ Intuitive Front Panel Operation for Ease of Use
2
W
Unmatched
Workload Coverage
f you’re constantly being asked
to calibrate more and more with
less and less, Wavetek’s Model
9100 Calibration System is the
answer you’ve been waiting for. Its
exceptional performance in both
the analog and digital domain
allows it to calibrate an extremely
wide range of test and measurement
equipment – quickly, efficiently
and to ISO9000 requirements.
For the Model 9100, calibrating
everything from handheld digital
multimeters to high performance
digital-storage oscilloscopes is all part
of a day’s work.
Take advantage of the Model
9100’s unique procedure mode,
which guides operators step-by-step
through the entire calibration
process, and you’ll not only calibrate
more with less – you’ll do it faster.
I
The Model 9100 is a multifunction calibrator with a breadth
and depth of outputs never before
available from a single calibration
source. In addition to DC and AC
Voltage to 1050V, variable Resistance
to 400MΩ and DC and AC Current
to 20A (1000A via the optional
current coils), the Model 9100
delivers continuously variable
Capacitance values to 40mF and
Conductance values to 2.5
milliSiemens. It also generates
digitally synthesized and phaselocked Sine, Square, Triangle,
Impulse and Trapezoidal waveforms,
variable amplitude Pulses to 10MHz,
Pulse Widths to 2 seconds, and Duty
Cycles between 0.05% and 99.95%.
Add one of the two oscilloscope
calibration options and it generates
all the waveforms required to
calibrate oscilloscopes up to
250MHz or 600MHz.
Fit the insulation/continuity
tester option and it synthesizes
resistance values as high as 2GΩ at
test voltages up to 1000V. Fit the
power meter option and it
simultaneously generates variable
phase angle voltages and currents
that allow you to calibrate power
meters up to 1MW or 1MVAR.
No other multi-product calibrator
gives you such wide workload
coverage and versatility. Weighing
in at only 41 lbs (18.5kg), it’s also
the ideal solution to on-site
calibration.
Model 9100 – the low-cost solution
to calibrating:Handheld Multimeters
Bench Multimeters (up to 6-1/2 digit scale length)
Analog Meters
Clamp Meters
Panel Meters
Power Meters
Harmonic Analyzers
Oscilloscopes
Combination Scope / Multimeters
Insulation / Continuity Testers
Counters
Electronic Thermometers
Chart Recorders
Oscillograph Recorders
XY Recorders
Data Loggers
3
Model 9100
Faster Throughput
Calibration
o cope with the varied and
demanding workload of a
modern calibration department,
you not only need a calibrator that’s
versatile, you also need one that
maximizes throughput.
That’s why the Model 9100
incorporates a unique procedure
mode that guides untrained
operators through the entire
calibration process, with direct
T
printout of calibration results on a
connected printer.
By utilizing PCMCIA cards to
import procedures into the Model
9100, we’ve eliminated the added
cost and complexity of a separate
computer – so operators don’t have
to be conversant with Windows™
software to enjoy the benefits of
semi-automated calibration.
For commonly used instruments,
the chances are you’ll find the
calibration procedures you need in
our comprehensive procedure
library. Already containing over 800
Direct printer connection for
immediate print-out of
calibration certificates
4
Anti-static workmat organizes
connections to UUT without
compromising operator safety
or degrading accuracy
fully tested procedures, this library
is continually being added to as
part of Wavetek’s ongoing
commitment to calibrate the widest
range of equipment possible.
If you want to write your own
procedures, or you want to fully
automate the calibration of
instruments that can be controlled
and interrogated via the IEEE-488
bus, you can integrate the Model
9100 into a PC-based calibration
workstation, by running Wavetek’s
Portocal-II or 9010 calibration
software. In addition to directly
Fully shrouded terminal pod
eliminates the possibility of
incorrect lead connections
to the Model 9100.
Optional trackerball
simplifies output adjustment
and go/no-go decision
making
LCD display clearly indicates
output conditions and prompts
the operator during procedure
execution
Dynamic soft-keys provide
complete control of secondary
functions
W
controlling a Model 9100 calibrator,
both Portocal-II and the 9010
software let you transfer calibration
procedures to PCMCIA cards so
that you can run them on a standalone Model 9100.
To make the operator’s task as
simple as possible, the Model 9100
solves other aspects of the calibration process as well.
Before running a procedure, all
the additional information required
for ISO9000 compliance – such as
the operator’s name and the
instrument-under-test serial number
Spinwheel allows coarse and fine
slewing of output values. Cursor and
increment/decrement keys allow
digit-by-digit editing of output values
– can be entered using the Model
9100’s alphanumeric keypad.
An optional trackerball provides
finger-tip control of output values,
go/no-go decision making and
procedure stepping. And after the
procedure has been completed, a
single keystroke prints out a full
certificate of calibration results on
any Centronics compatible printer.
The whole process is as simple
as A-B-C, taking less than five
minutes for the average test
instrument.
Even when you use the Model
9100 in manual mode, we’ve given
it a simplicity of operation that
minimizes human error and speeds
the calibration process.
Frequently used functions such
as Voltage, Current and Resistance
have dedicated front-panel keys
that allow you to activate them at
the touch of a button. Less
commonly used functions are
selected using screen menus and
soft-keys to avoid cluttering the
front-panel with too many keys.
Three different methods of
adjusting the output value
– numeric entry, increment/
decrement keys or spinwheel
control – let you choose one that
suits the calibration operation. Soft
keys allow you to implement any
of the commonly used range
sequences, while ∆ and ∆% modes
give you an instantaneous display
of output deviation.
Keypad provides direct
entry of alphanumeric
information
Single keystroke selection
of primary output functions
Output controls isolate
the unit-under-test and
provide safety
interlocking for dangerous
output conditions
Mode key allows selection of the Model 9100’s
Procedure, Manual, Configuration, Calibration and SelfTest modes.
Calibration procedures
on PCMCIA cards
eliminate the need for a
control computer
PCMCIA SRAM card transfers
calibration results to the
optional 9010 Support
Software
PC-based Portocal-II or 9010
support software for new
procedure generation, fully
automated calibration and
inventory management
5
Model 9100
Multimeters and Panel Meters
lmost every year, DMM
manufacturers introduce new
models with more ranges and
functions than ever before. To
protect your investment in
calibration equipment you need a
calibrator that will cope with nextgeneration instruments, as well as
those which make up today’s
calibration workload. To be really
flexible, you also need one that will
deal with analog
multimeters as well.
The Model 9100’s
A
Specifications
comprehensive range of outputs,
which includes DC and AC
Voltage, DC and AC Current,
Resistance, Conductance,
Capacitance, Frequency, Duty
Cycle, Pulse Width, Logic Level
and RTD/ Thermocouple
simulation, covers all the functions
you’re likely to find on modern
DMMs. Every one of these
functions has sufficient span to test
virtually any multimeter to its
performance limits. No other
calibrator in this class can deliver
1050V at 10kHz, 350V at 30kHz or
20A at 10kHz, without the use of
external power boosters or
transconductance amplifiers.
On its Resistance function, it
copes with the high sense currents
(up to 350mA) used by older
analog multimeters.
And with Portocal-II’s ability to
use a precision DMM as a system
reference, you can even enhance
the Model 9100’s accuracy
specification to the level required
to calibrate 6-1/2 digit benchtop and systems DMMs.
Uncertainties are for 1 year, Tcal ± 5°C.
FUNCTION
RANGE
DC Voltage
Zero to ±1050V
AC Voltage
Zero to 1050V
10Hz to 100kHz
DC Current
Zero to ±20A (Up to 1000A
via current coil*)
AC Current
Zero to 20A (Up to 1000A
via current coil*)
10Hz - 30kHz
DC Power
1mW to 20kW (Up to 1MW
via current coil*)
AC Power
1mW/mVAR to 20kW/kVAR
(Up to 1MW/MVAR via
current coil*)
Resistance
Zero to 400MΩ
Conductance
2.5nS to 2.5mS
Capacitance
500pF to 40mF
Frequency
0.5Hz to 10MHz
Duty Cycle
0.05% to 99.95%
Pulse Width
0.30µs to 1999.99ms
Logic Level and Pulse
TTL, CMOS and ECL
Waveforms
Sine, square, triangle,
trapezoid and impulse
Phase
±180°
Temperature (IPTS68 or ITS90)
Thermocouples
Type
B, C, E, J, K, L, N, R, S, T
Temperature
-250°C to +2320°C
Type
Pt385, Pt392
RTD
0°C Resistance
10Ω to 2kΩ
Temperature
-200°C to +850°C
*The 50-turn coil has been designed for optimum accuracy and
inductance. With some clamp meters and power meters, especially those
using Hall effect sensors, the increase in inductance due to the design of
Power Meters
itted with its Power Meter Calibration
option (Option PWR), the Model 9100
simultaneously generates voltages and
currents as high as 1000V and 1000A at any
phase angle between ±180°, allowing you to
calibrate power meters up to 1MW and 1MVAR.
Its ability to output squarewave, impulse,
triangular and trapezoidal waveshapes as well
as sinewaves allows you to evaluate power
F
6
W
Clamp Meters
RESOLUTION
BEST
UNCERTAINTY
1µV
0.006%
1µV
0.04%
1nA
0.014%
1nA
0.07%
1mW
0.03%
1mW or 1mVAR
0.125%
100µΩ
0.015%
0.1pS
0.04%
0.1pF
0.3%
1mHz
25ppm (0.25ppm with Option 100)
0.01%
35ns
0.1µs
25ppm (0.25ppm with Option 100)
0.01°
0.07°
0.1°C
0.17°C
0.01°C
0.08°C
he Model 9100’s unique
Current Coil option (Option
200) overcomes all the magnetic
circuit problems normally associated
with clamp-on ammeter calibration.
The coil module incorporates
x10 and x50 coils, both of which
feature internal magnetic shielding
to eliminate interference from stray
flux. Even the x50 coil accepts the
full 20A output of the Model 9100
with sufficient voltage compliance
to calibrate most popular clamp
meters at currents up to 1000Arms.
Yet the whole module is small
enough and light enough to sit
comfortably on the bench.
Because the coils can be driven
with AC or DC currents, they are
just as suitable for calibrating
meters based on Hall-Effect*
sensors as they are for calibrating
meters that use a current
transformer.
T
Once you’ve selected the x10 or
x50 coil from the Model 9100’s onscreen menu, all outputs are
automatically scaled to the correct
values, as are the calibration results
that are printed out or transferred
onto PCMCIA cards.
the current clamp may limit the Current/Hertz profile obtainable from the
Model 9100. In some cases, 1000A may not be attainable at higher
frequencies.
meter performance with simulated ‘real-world’
currents and voltages — such as those drawn
by switchmode power supplies and lighting
ballasts.
To calibrate harmonic analyzer functions,
the Model 9100 generates harmonics up to
3kHz at amplitudes as high as 3.2Vrms.
Power meter calibration has never been as
easy or as thorough.
7
Model 9100
Insulation / Continuity Testers
he high stimulus voltages delivered by insulation testers make
particular demands on a calibration system. Demands that previously required the use of complicated switched resistor networks
and high-impedance DMMs.
T
The Insulation/Continuity
Tester option for the Model 9100
(Option 135) changes all that by
employing a unique High-Voltage
Active Resistance Technology that
simulates variable resistances as high
as 2GΩ at voltages up to 1350V.
It also provides you with a
simultaneous read-out of the test
voltage and current. These features
are unmatched by any other multiproduct calibrator.
In addition to calibrating
insulation resistance, Option 135
also calibrates the continuity test
functions that are often provided
on insulation testers. Four-Wire
Active Resistance outputs from
zero to 4kΩ allow you to precisely
determine continuity thresholds.
Direct read-out of current allows
you to calibrate unit-under-test
current sources.
Fitted internally, Option 135 is
fully compatible with other Model
9100 options such as the Power
Meter and Scope Calibration
options.
Specifications
Uncertainties are for 1 year, Tcal ± 5°C.
FUNCTION
Insulation Resistance
Resistance
100kΩ to 2GΩ
BEST
UNCERTAINTY
0.1%
Voltage (measured)
Zero to 1350V
0.6%
Current (derived)
1µA to 2.3mA
1.5%
Zero to 4kΩ
0.035%
Continuity
Resistance
8
RANGE
Voltage
Zero to 10V
Current (derived)
100µA to 350mA
1.0%
W
Oscilloscopes
Specifications
FUNCTION
dding one of the Model 9100’s
two Oscilloscope Calibration
options (Option 250 or Option
600) allows you to comprehensively
calibrate oscilloscopes up to 250MHz
or 600MHz.
All the outputs required to
calibrate the gain, linearity and
bandwidth of vertical and horizontal
deflection circuits, and the accuracy
and linearity of timebase circuits,
are delivered through a single pair
of BNC cables (one for the
calibration waveforms and one for
the trigger signal), making complex
lead changes a thing of the past.
Full accuracy is maintained right
up to the scope’s BNC inputs.
These scope calibration options
provide precision DC levels and
1kHz squarewaves up to 120V for
vertical and XY deflection
calibration, plus continuously
variable leveled sine-waves from
10Hz to 250MHz (Option 250) or
10Hz to 600MHz (Option 600) for
bandwidth and AC flatness checks.
A
They both generate ultra-fast
low-level edges at repetition rates
high enough to allow even the
shortest persistence traces to be
examined for overshoot, undershoot
and ringing. And they generate fast
high-level edges so that you can
check the AC performance of input
attenuators.
Clearly visible timing markers at
intervals as short as 2ns or as far
apart as 5 seconds calibrate timebase
accuracy, while an optional highstability crystal reference (Option
100) improves basic timing accuracy
from 25ppm to 0.25ppm in order to
calibrate high-performance DSOs.
And if you ever need to calibrate
oscilloscopes with bandwidths
higher than 600MHz, or require
full multi-channel automation,
remember that Wavetek also
produces a range of dedicated
scope calibration workstations – the
Model 9500 Series.
RANGE
Voltage Amplitude*
Into 1MΩ
5mV to 120V pk-pk
dc and 1kHz
Into 50Ω
5mV to 3V pk-pk at 1kHz
Up to ± 2.5V dc
Range Sequence
1-2-5
Adjustment
±10%
Low Edge*
Amplitude into 50Ω
100mV to 1.1V pk-pk
Rise/Fall Time
<1ns
Period
100ns to 10ms
High Edge*
Amplitude into 1MΩ
1V to 50V pk-pk
Rise Time
<100ns
Period
10µs to 10ms
Leveled Sinewave*
Frequency
Option 250
Option 600
Amplitude into 50Ω
10Hz to 250MHz
10Hz to 600MHz
4.5mV up to 5.5V pk-pk
Amplitude Adjustment ±10%
Markers*
Period
Option 250
4ns to 5s
Option 600
2ns to 5s
Range Sequence
1-2-5
Amplitude
Up to 1V into 50Ω
Timing Accuracy
Normal
25ppm
With Option 100
0.25ppm
*External trigger output provided
9
Model 9100
Electronic Thermometers
apable of simulating ten different
thermocouple types, plus RTDs with
nominal resistance values anywhere between
10Ω and 2kΩ, the Model 9100 covers the
calibration requirements of all popular electronic
thermometers.
An exceptionally wide temperature range is
offered from -250°C to +2320°C with
temperature entry in either degrees Centigrade,
degrees Fahrenheit or Kelvin.
And the Model 9100’s conversion algorithms
can be switched to follow either the IPTS68 or
ITS90 temperature scales, producing precision
C
Counters
B
10
y adding the High Stability Crystal Reference option (Option 100) you can
increase the Model 9100’s frequency and timing accuracy to 0.25 ppm, making it
suitable for counter/timer calibration.
W
Specifications
voltage or resistance outputs that accurately
simulate temperature to a resolution of 0.1°C for
thermocouples or 0.01°C for RTDs.
The use of a specially designed thermocouple
connector, which features integral cold junction
sensing, allows thermometers to be calibrated on
the workmat or connected directly into the
Model 9100’s front-panel terminals via a longer
length of thermocouple cable – making it as easy
to calibrate panel mounted temperature meters
as it is to calibrate handhelds. For RTD meter
calibration, the Model 9100 can be used with
either 2-wire or 4-wire lead connections.
Uncertainties are for 1 year, Tcal ± 5°C, and include cold junction compensation errors.
Stated in °C. Selectable IPTS68 or ITS90 temperature scales.
BEST
FUNCTION
RANGE
RESOLUTION
UNCERTAINTY
Thermocouple
Type
B
+500°C to +1820°C
0.1°C
0.34°C
C
0°C to +2320°C
0.1°C
0.27°C
E
-250°C to +1000°C
0.1°C
0.17°C
J
-210°C to +1200°C
0.1°C
0.19°C
K
-250°C to +1372°C
0.1°C
0.19°C
L
-200°C to +900°C
0.1°C
0.18°C
N
-200°C to +1300°C
0.1°C
0.22°C
R
0°C to +1767°C
0.1°C
0.28°C
0°C to +1767°C
S
0.1°C
0.35°C
T
-250°C to +400°C
0.1°C
0.17°C
Pt 385
10Ω to 2kΩ
-200°C to +850°C
0.01°C
0.08°C
Pt 392
10Ω to 2kΩ
-200°C to +630°C
0.01°C
0.08°C
RTD
Recorders
ecause modern data recorders can be configured to accept a wide range of signal inputs –
including thermocouples, 20mA current loops, unipolar/bipolar voltages, and digital bit streams
– you need a versatile calibrator to calibrate them.
The Model 9100’s comprehensive coverage of functions, ranges and waveforms, in both the
analog and digital domains, makes it an ideal unit for calibrating these instruments.
B
11
Model 9100
9010 Calibrator Support Software
unning under Windows™, the 9010 Support
Software allows you to automate the calibration
process, either by using the Model 9100’s
PCMCIA-based procedure mode or by
integrating the calibrator into a PC-based
IEEE-488 bus system. It also gives you the
ability to generate custom calibration
procedures, analyze and archive calibration
results and print custom reports and
certificates, while its powerful inventory
management capabilities maintain
calibration records and help you schedule
instruments through your calibration
laboratory. To ensure you get the productivity
benefits from the moment you install the
software, it comes complete with a library of over
800 calibration procedures for popular test and
measurement equipment.
R
New Procedure Generation
Easy Procedure Execution
If you run the Model 9100 in an IEEE-488 bus system, you execute
calibration procedures directly from the PC. If you use procedure mode,
you transfer procedures and results to and from the Model 9100 on
PCMCIA cards. The 9010’s Procedure Card Programming and Results
Card Reading utilities support a wide range of PCMCIA card types,
reducing procedure/results selection to simple point-and-click operations.
You can even control the language in which annotation on the Model
9100’s LCD is displayed. A suitable PCMCIA card drive can be supplied
with the 9010 software.
12
Wavetek is
continuously generating new
procedures for the Model 9100,
making them available for instant
download from its World Wide Web
site (www.wavetek.com) as part of
the Portocal-II/9010 Option 10
support program.
However, if you do need to write
your own procedures, the 9010
W
Sophisticated Data Management
In addition to supporting the Model 9100 Calibrator, the 9010 Software is a complete
inventory management package in its own right. It can store complete information,
including extended repair histories or special notes, on any instrument – not only those
calibrated by a Model 9100.
Calibration results are uniquely attached to the instrument’s inventory record,
together with all the information required by stringent quality standards such as ISO9000
to ensure both forward and backward traceability.
Full network access to the entire instrument inventory database means you can
integrate the 9010 Software into sophisticated information management systems.
We’ve also recognised the need for users to add corporate identity to their
documentation. That’s why the 9010 Software has a built-in report generator that allows
you to design your own certificates and reports, complete with company logos, custom
headers and footers etc.
Portocal-II
Software contains a powerful yet
easy to use procedure generator.
Powerful because it enables you to
write procedures entirely logically
using a hierarchical approach. Easy
to use because the complete process
is menu driven so that you won’t
have to remember a single
programming language instruction.
The Portocal-II Software does everything the 9010 Support Software can
do, plus it allows you to integrate the Model 9100 alongside other
calibrators in a fully automated PC-based calibration system. Portocal-II
supports all 9000-Series, 4800-Series and 4000-Series calibrators from
Wavetek, plus a range of calibrators from other manufacturers. It also
supports the use of measuring instruments as system references,
making it possible to enhance the Model 9100’s accuracy specification.
By monitoring the output of the Model 9100 with a precision DMM,
such as a Wavetek Model 1281, the higher accuracy of this DMM can
be transferred to the Model 9100’s outputs. Portocal-II calibration
procedures that use this technique to allow a Model 9100 to calibrate
6-1/2 digit DMMs (for example, the Hewlett Packard HP34401A) are
already in the procedure library.
13
Model 9100
Reliability and Support
he Model 9100’s built-in
reliability not only ensures that
the unit gives you years of
trouble-free service, it also ensures
maximum operator safety.
Comprehensive power-on selftest routines check all the major
internal circuits for correct operation,
while internal watchdogs
continuously monitor the status of
the calibrator for fault conditions.
Automatic shut-down of the output
under fault conditions protects the
operator and prevents equipment
damage.
T
14
Zero Downtime Recalibration
In common with all other
calibrators from Wavetek’s Test &
Measurement Division, the Model
9100 can be calibrated using our
Model 4950 Multifunction Transfer
Standard System.
And because we offer a unique
on-site calibration service based
around the Model 4950, your Model
9100 won’t even have to leave the
bench to be calibrated – so you
won’t incur expensive downtime
and shipment costs. For more
information on our Model 4950 Onsite Calibration Service, contact
your local Wavetek Service Center.
W
Global Customer Care
n addition to providing
innovative, customer-focused
solutions for calibration and test,
our mission is to provide the
highest quality of customer service
and care worldwide.
Wavetek has Master Service
Centers and Calibration facilities in
the United States, United Kingdom,
France, Germany, China and
Singapore, that work closely with a
network of service partners
throughout the world.
I
This global network provides a
range of preventative maintenance,
repair, calibration and other
value-added services –
proof of our
commitment to
deliver services
that meet the
highest
standard of
customer
satisfaction.
General Specification
Environment
Temperature:
Operating:
5°C to 40°C
Storage:
0°C to 50°C
Humidity (non-condensing):
Operating:
<90% over 5°C to 30°C; <75% over 30°C to 40°C.
Storage:
<95% over 0°C to 50°C.
Warm-up Period:
20 minutes
Power
Voltage:
Frequency:
Consumption:
100V/120V/220V/240V ±10%
48Hz to 63Hz
450VA maximum
Dimensions
H x W x D:
Weight:
133 x 427 x 460 mm (5.24 x 16.8 x 18.1 inches)
18.5kg (41lbs)
Safety
Designed to UL3111 and EN61010-1-1:1993/A2:1995
CE Marked
EMC
Emissions:
EN55011:1991
Generic Immunity: EN50082-1:1992
FCC Rules part 15 sub-part J class B
Detailed Specification
For detailed performance specifications see separate booklet
‘9100 Specifications’
Ordering Information
Model 9100
Option 10
Option 30
Option 50
Option 60
Option 65
Option 70
Option 90
Option 100
Option 135
Option 200
Option 250*
Option 600*
Option PWR
Universal Calibration System including Lead Kit, Workmat
and Calibration Certificate
Blank 256-Kbyte FLASH card
Blank 256-Kbyte battery-backed SRAM card
Tracker ball
Carry case
Ruggedized Transit Case
NAMAS Calibration Certificate
Rack Mounting Kit
High Stability Crystal Reference
Insulation/Continuity Tester Calibration Module (fitted internally)
x10/x50 Current Coils
250MHz Oscilloscope Calibration Module (fitted internally)
600MHz Oscilloscope Calibration Module (fitted internally)
Power Meter Calibration Module (fitted internally)
* Option 250 and Option 600 cannot be fitted together
Software
9010
Portocal-ll
Windows™ Automated Calibration Software for Wavetek
9000-Series Calibrators
Windows™ Automated Calibration Software for Wavetek
Calibrators and Calibrators from Other Manufacturers
(See separate Portocal-II brochure)
Procedure Library
Access www.wavetek.com to view the latest list of procedures
15
Model 9100
W
®
Other Calibration Instruments from Wavetek
Worldwide Sales Offices
4800-Series DMM Calibrators
Austria
Wavetek Gesellschaft m.b.H.
Pharos Haus
Nordbahnstrasse 36/TOP 1.4
A-1020 Vienna, Austria
Tel:
Fax:
(43) 1-214-5110
(43) 1-214-5109
Tel:
Fax:
(86) 10-6592-8044
(86) 10-6500-8199
Tel:
Fax:
(33) 1-41-90-6666
(33) 1-41-90-6650
Tel:
Fax:
(49) 89-996-410
(49) 89-996-41160
Tel:
Fax:
(852) 2788-6221
(852) 2788-6220
Tel:
Fax:
(81) 552-43-0311
(81) 552-43-0396
Tel:
Fax:
(65) 356-2522
(65) 356-2553
Tel:
Fax:
(44) 1603-256-600
(44) 1603-483-670
Tel:
Fax:
(1) 619 279 2955
(1) 619 450 0325
China
DC & AC Voltage, DC & AC Current and Ohms. Calibration of DMMs to 8-1/2
digits. Two levels of precision.
Model 9500 Oscilloscope Calibration Workstations
High accuracy calibration of analog and digital-storage oscilloscopes up to
3.2 GHz.
Model 4950 Multi-Function Transfer Standard
Wavetek Corporation
Room 2701, Citic Building
No. 19 Jianguomenwai Dajie
Beijing 100004, P. R. China
France
Wavetek S. A.
Immeuble le Seine St Germain
12, Bd des îles
Bat B 3ème étage
92441 Issy les Moulineaux, Cedex
France
Germany
Wavetek GmbH
Gutenbergstrasse 2-4
85737 Ismaning
Germany
Hong Kong
Ultra-stable transfer measurements for on-site calibrator support.
Wavetek Hong Kong Ltd.
3A HKPC Building
78 Tat Chee Avenue
Kowloon, Hong Kong
Japan
Model 4920 Alternating Voltage Measurement Standard
Yokogawa Electric Corporation
Measurement Division
155 Takamuro-cho, Kofu-shi
Yamanashi-ken, 400-0057 Japan
Singapore
The world’s most advanced AC Voltage measuring instrument.
Model 1281 Precision Digital Multimeter
Wavetek Asia-Pacific Pte Ltd
51 Goldhill Plaza
#14-04/05
Singapore 308900
United Kingdom
Wavetek Ltd
Hurricane Way
Norwich, Norfolk NR6 6JB, U.K.
Ultra-low noise, 8-1/2 digit precision, multi-function measurements.
United States
Wavetek Corporation
9045 Balboa Avenue
San Diego, CA 92123, U.S.A.
Web Site
www.wavetek.com
Specifications may be subject to change without notice
© Wavetek Corporation 1994, 1996, 1997, 1998
16
DS239 - FEB98