ACE SHEAR (Harmonic Scalpel) – Need for Speed
Transcription
ACE SHEAR (Harmonic Scalpel) – Need for Speed
ACE SHEAR (Harmonic Scalpel) – Need for Speed Mr. Mahesh Patwardhan Consultant Gynaecologist Queen Elizabeth Hospital Woolwich, London Rationale for HARMONIC technology in surgery Cutting and coagulation can be achieved with minimal tissue damage and reduced risk TECHNOLOGY Electrical energy is passed to the handpiece Piezo-electric ceramic disks in handpiece become excited Disks transfer electrical energy into mechanical energy Mechanical (longitudinal) motion is amplified at nodes in shaft Reaches max. amplitude of 55,500 cycles/sec at blade tip TISSUE EFFECT Blade comes into contact w/tissue, pressure causes coaptation of blood vessel H+ bonds are broken, protein in cells is denatured Denatured protein forms a sticky coagulum This allows for simultaneous cutting and coagulation to take place at a lower temperature than electrocautery Cutting and hemostasis can be achieved with minimal tissue damage and minimal risk BALANCE OF TISSUE EFFECTS Cavitation Cutting Balance Coaptation Coagulation CUT n n The active blade vibrates back and forth (50 - 100 µm) with 55.500 Hertz The vibrating blade edge cuts tissue by stretching it beyond its elastic limit CAVITATION Motion of the blade creates an area of low pressure, causing water to boil at a lower temperature. vapor at 37°C tissue plane energy cavitation Water vapor expands, causing the layers to separate which enhances visualization of avascular plane of dissection. COAPTATION HARMONIC Blade Blood Vessel Ultrasound Pressure Coapted Blood Vessel COAGULATION n n n Secondary heat from friction causes protein denaturation (> 63° C) Depends on exposure time (coupling of tissue with ultrasonic energy) In practice: fluent transition between coaptation and coagulation COAPTIVE COAGULATION Vessel walls are collapsed/approximated due to pressure. The blade is then activated... …this causes the proteins within the cells to denature, forming a sticky coagulum that coagulates the vessels. ACTIVE SURGICAL TEMPERATURES Electrosurgery/Laser Surgery HARMONIC Protein Coagulation 50º C Proteins disorganize to form a coagulum Tissue Dessication 100º C 150º C Vaporization of water dessicates tissues Eschar (oxidation) 400º C+ Eschar forms when tissues burn LATERAL SPREAD OF COAGUALTION 70 60 Lateral spread (cc²) Electrosurgery 50 40 30 20 10 0 UltraCision 0 1 2 3 4 5 Time (sec) Amaral JF, Chrostek CA. SAGES, 1995. 6 7 8 9 10 COMPARISON OF INCISION Hepatic Tissue in Porcine Model HARMONIC Max Power = 35 Watts Monopolar Pencil 50 watts Blend 5 FACTORS AFFECTING CUTTING AND COAGULATION ! Power Level ! Tissue Tension ! Grip Force ! Blade Sharpness ! Time POWER LEVEL TISSUE TENSION GRIP FORCE Strong grip Strong grip Light grip Light grip BLADE SHARPNESS Sharper blade • Faster cutting • Less hemostasis Blunt surface • Slower cutting • More hemostasis TIME Less time More time • Faster cutting • Less hemostasis • Slower cutting • More hemostasis Raising the Standards u u I want to decrease Operating time I want reliable vessel sealing « “I « I want to sleep at night” want no leakers at the seal « I want minimal surrounding thermal damage u u I want to easily rotate the instrument 360 degrees and be a one handed instrument I would like hand activation HARMONIC ACE Convex Surface Blunt Nose Concave Surface Knife-Down FEATURES SEALS VESSELS UP TO 5mm IN DIAMETER HALVES THE TRANSECTION TIME GRASPING FORCE INCREASED BY 2 TIMES SEAL STRENGTH IN EXCESS OF 2 TIMES SYSTOLIC PRESSURE OPTIMAL OCCLUSION PRESSURE AUDIBLE FEEDBACK “CLICK” ALL PRODUCTS INCLUDE HAND ACTIVATION HARMONIC ACE INVENTIONS n n Finger actuation vs. Thumb actuation eliminates “numb thumb”. 7 Hand activation buttons aligned to prevent inadvertent activation. 8 9 1 Pressure Profile 2 Pad Material 3 4 Force limiting Spring Mech. 5 Improved end effector 6 Click 7 Thumb activation 8 9 1 0 1 1 1 2 Ergo handle, Trigger, & Mechanism Pad Life Clamp Force SEALS VESSELS 66% LARGER *IN ANIMAL MODEL * SURGICAL STEPS GENERATOR GEN04 Ready Indicator Increase Standby Power Level Graphic Display Test HandActivation Capability Alarm Indicator Power switch Volume Decrease UserPower Settable Level Minimum Power Setting Maximum Power Level Hand Piece Receptacle Has ACE really changed my practice? … any questions? Thank you