Incubator CO2 series - KW apparecchi scientifici

Transcription

Incubator CO2 series - KW apparecchi scientifici
GreenLine
Incubator CO2 series
model W180CCI.IR
K W A P PA R E C C H I S C I E N T I F I C I
Incubator CO2 series model W180CCI.IR
W180CCI.IR
control panel
W180CCI.IR - opened
Incubator CO2 series model W180CCI.IR
GENERAL CHARACTERISTICS:
Structure in electrolytically galvanized sheet steel with high resistance
epoxy powder coating. Internal chamber in smooth AISI 304 stainless
steel, with rounded edges to avoid contamination, without any
surface discontinuity. Any internal parts are removable without the
need for tools for maximum cleanliness. Internal door in glass with
silicone rubber sealing.
Exclusive KW system for controlled heating and ventilation of the
internal door in glass sheet: eliminates condensation, ensuring
quick temperature recovery after opening the door.
Elimination of the condensate on the internal walls and internal door,
with an adequate distribution of the thermal flow and with a distinct
temperature regulation of the door and the internal chamber.
Heating takes place by convective exchange with the air.
Separate, easily removable fan. Humidification is obtained through
a stainless steel container placed in the bottom of the internal tub
which allows it to saturate the chamber atmosphere.
CO2 injection is with a PWM solenoid valve system, through an easily
replaceable sterilizing filter. Upon opening the door, the μP system
controls the thermal energy supply to the minimum level of stability
of Tset and interrupts both the flow of CO2 and the microventilation.
Upon closing the door, the PWM system ensures a quick restoration
without overshooting the set value under the centralized and
integrated μP management, which also controls the delay of any
alarm activation.
The W180CCI.IR is available as a bench or floor model, with or without
tubular support, to allow an optimization of space, adapting to every
laboratory configuration. The incubator W180CCI.IR is simple and
safe to use, through:
• Liquid crystal display, alphanumeric with 2 rows of 16 characters
(usable area: 20x95 mm);
• Messages and indications in Italian or English
• Simultaneous indication on display of temperature and CO2
concentration;
• Indication on display of alarm status with remote setup;
• CO2 and °C display and setting with two dedicated keys;
• Password for access to setup of set points and other system
parameters;
• Differentiated temperature regulation for chamber and door;
• Open door sensor with visual acoustic alarm;
• CO2 and temperature concentration recovery algorithms
optimized to minimize thermal shocks (in particular T overshoot)
and concentration;
• Mute-capable visual and acoustic alarms of movement from set
points: T (min/max) and CO2 % (min/max);
• Calibratable overtemperature safety thermostat; CO2 supply
pressure from 0.7 to 3 bar; a regular membrane reducer is
sufficient for the CO2 tank.
CO2 REGULATION:
• NDIR infrared beam sensor;
• Measurement cell thermostatted at 45°C ±0,2°C, so that ambient
temperature and humidity become uninfluential
• Regulation range: 0.0÷10.0% standard; 0.0÷20.0% optional;
• Precision of measurement and regulation: better than 2% of
maximum value;
• Resolution and repetitivity: better than 1% of maximum value;
• Stability < ± 0.1%;
• Uniformity < ± 0.2%;
• Programmable min/max alarm thresholds;
• Acoustic-visual alarms capable of remote setup;
• Zero drift of sensor: less than 2% of maximum value, cleared every
24 hours by automatic calibration system which can be called up by
keyboard;
• Procedure for span value calibration with reference tank =>
programme installed in the μP software complete with instruction
messages, also for self-maintenance;
• Proportional PWM regulation with solenoid valve;
• Time to reach set point after opening 30 sec door: < 10 min in
standard range;
• RH (Relative Humidity) % values > 90% to avoid culture desiccation;
• RS485 or RS232 serial port for transfer of signal to PC (optional);
• Connection and socket for manual control of CO2.
TEMPERATURE REGULATION:
• RTD Pt 100 sensor;
• Optimized PID control;
• Regulation range: from + 5.0 °C above ambient to + 42.0 °C;
• Precision of measurement and regulation: better than 0. 1 °C;
• Stability < ± 0.1 °C;
• Resolution and repetitivity: better than 0. 1 °C;
• Uniformity < ± 0. 5 °C of the working volume;
• Time to reach set-point: after opening door for 30 sec: < 6 min;
• Programmable min/max alarm thresholds;
• Acoustic-visual alarms capable of remote setup;
• RS485 or RS232 serial output for transfer of signal (optional);
• Electronic strip-chart recorder (optional).
ACCESSORIES UPON REQUEST:
• Tubular floor support;
• Tubular floor support with wheels;
• CO2 pressure regulator;
• Fyrite CO2 analyzer 0-20%;
• Supplementary RTD Pt 100 Ohm probe, for connection to external
acquisition system and T recorder, such as T-GUARD® and
SensiNet®, or similar systems
• Additional RTD Pt 100 Ohm probe, complete with 4-20 mA converter,
mounted on DIN rails, for connection to external recording system
• Electronic strip-chart temperature recorder;
• RS485 or RS232 serial port
An I.Q. (Installation Qualification) and O.Q. (Operational
Qualification) can be carried out on this series of equipment;
for an evaluation of the costs of these activities, consult the KW
commercial office.
KW is also available for ISO calibration certification services for SIT
sample comparisons.
Model
External
dimensions (WxDxH)
Internal
dimensions (WxDxH)
Capacity
shelves
N.
Power supply
Maximum
power
Weight
W180CCI.IR
cm. 92x64x74 h
cm. 60x55x56 h
litres 180
5
V 220 Hz 50
800 VA
Kg. 110
K W A P PA R E C C H I S C I E N T I F I C I
Contact us:
Managing director: [email protected]
Production and technician office: [email protected]
Sales management: [email protected]
Tenders and export office : [email protected]
Export support: [email protected]
Service and support: [email protected]
[email protected]
Logistics and shipments: [email protected]
Administration office: [email protected]
= Min. / max. temperature alarm
= Internal light
= Power failure alarm
= Open door alarm
= Lock
= Temperature programmer for thermal cycling
= Graphic temperature recorder
adviseritalia.com
= Wheels
• W-LAB WR-LAB series legend
• WR 90 - 102 series
thermostatic chambers / precision incubators
refrigerated with water jacket
• WI series
natural guided convection incubators
• W.80 - W.82 series
precision water bath
• WPL series INCUBATORS
table and floor incubators
• W.82/O - W.84/O series
precision water bath
with linear and/or orbital shaking
• WPLR series INCUBATORS
refrigerated table and floor incubators
• W180CCI.IR model CO2 Incubator series
• W-LAB series INCUBATORS
forced air circulation incubators
• WRS 96 - 85 series
thermal refrigerating chambers with photoperiod control for environmental simulation
• GreenLine Introduction
• WR-LAB series INCUBATORS
forced air circulation
refrigerated incubators
• WRC series
forced air circulation
thermal refrigerator
for modern scientific
and industrial laboratories
• W 90 - 102 series
thermostatic chambers / precision incubators
with water jacket
• WR UR700C model Climate Chamber series
for pharmaceutical stability tests
and environmental simulation
• KW 20/B-100/B KW 6/B-12/B-18/B series
modular glass or plastic bottle rollers ø80-120mm
• Walk-in incubator
• Walk-in stability chamber
KW APPARECCHI SCIENTIFICI s.r.l. • 53035 Monteriggioni (SI) • Via della Resistenza, 119 • Tel. +39 0577 309143 / 309145