syngo MammoReport - Region Syddanmark

Transcription

syngo MammoReport - Region Syddanmark
syngo MammoReport
XP
Startup
Installation
RIS Configuration
© Sieme ns
2011
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: SPB7-420.815.51.06.02
Replaces: SPB7-420.815.51.05.02
Doc. Gen. Date: 12.11
0-2
Revision
Chapter
Page
All
All
Revision
06
Document revision level
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents correspond to the
version at the time of system delivery and/or printout. Versions to hardcopy documentation are not
automatically distributed. Please contact your local Siemens office to order current version or refer to
our website http://www.healthcare.siemens.com.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Copyright notice
The DICOM functionalities in the syngo MammoReport system are based on the libraries from Merge
Technologies Inc.
Celsius is a trademark of Fujitsu Siemens Computers.
Windows XP Professional is a trademark of Microsoft Corporation.
syngo is a trademark of Siemens AG.
Softcopy reading software by MeVis BreastCare,
Copyright © 2001 - 2011 by MeVis BreastCare.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
Table of Contents
0-3
1 _______General Information ____________________________________________ 1 - 1
Purpose of Document . . . . . . . . . . . . . . . . . . .
Target Group . . . . . . . . . . . . . . . . . . . . . . .
Requirements . . . . . . . . . . . . . . . . . . . . . . .
Required Tools, Measurement, and Auxiliary Devices
Required Documents . . . . . . . . . . . . . . . . .
Writing Conventions . . . . . . . . . . . . . . . . . . . .
Text Emphasis
. . . . . . . . . . . . . . . . . . .
Safety Information and Protective Measures . . . . . . .
Acronyms and Abbreviations . . . . . . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.1- 1
.1- 1
.1- 1
.1- 1
.1-1
.1- 2
.1- 2
.1- 2
.1- 3
2 _______Overview______________________________________________________ 2 - 1
Preconditions . . . . . . . . . . . . . . . . . .
Restrictions . . . . . . . . . . . . . . . . . . .
Delivery . . . . . . . . . . . . . . . . . . . . .
Scope of this Document . . . . . . . . . . . . .
Supported RIS Systems. . . . . . . . . . . . .
RIS Bridge Specification . . . . . . . . . . .
Location of the RIS Bridge . . . . . . . . . . .
Procedure Options . . . . . . . . . . . . . . .
RIS Bridge directly on the syngo Computer .
RIS Bridge on a Virtual Machine. . . . . . .
RIS Bridge on a Stand-alone RIS Computer
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.2- 1
.2- 1
.2- 1
.2- 2
.2- 2
.2- 3
.2- 4
.2-5
.2- 5
.2- 6
.2- 7
3 _______RIS Bridge on syngo Workstation _________________________________ 3 - 1
Preconditions . . . . . . . . . . . . . . . . . . .
Installing the RIS Bridge . . . . . . . . . . . . .
Starting the Installation . . . . . . . . . . . .
Basic Bridge Setup . . . . . . . . . . . . . .
Different Configuration Steps . . . . . . . . .
Uninstalling a RIS Bridge . . . . . . . . . . . . .
RIS bridge installed on RIS client . . . . . . .
RIS bridge installed on syngo MammoReport .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.3- 1
.3-1
.3- 1
.3- 3
.3- 5
.3-5
.3- 5
.3- 5
4 _______RIS on Virtual Machine __________________________________________ 4 - 1
Procedure Steps . . . . . . . . . . . . . . . .
Installing the Virtual Machine . . . . . . . . . .
Disabling the Auto Login for syngo meduser
Basic Installation. . . . . . . . . . . . . . .
Starting the Virtual Machine . . . . . . . .
Windows XP System Preparation . . . . . .
VMware Mouse Driver . . . . . . . . . . . .
Configuration . . . . . . . . . . . . . . . .
Network Connection . . . . . . . . . . . .
Siemens
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
SPB7-420.815.51.06.02
12.11
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
Page 3 of 6
CS PS SP
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.4- 1
.4- 2
.4- 2
.4- 2
.4- 5
.4- 8
4 - 10
4 - 12
4 - 16
syngo MammoReport
0-4
Table of Contents
Checking the VMware Player . . . . . . . . . . . . .
Using the Virtual Machine with syngo MammoReport
Enabling the Auto Login for syngo meduser . . . . .
Installing the RIS Application
. . . . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.4 - 18
.4 - 18
.4 - 19
.4 - 20
5 ______ RIS Bridge on Virtual Machine ____________________________________5 - 1
Installing the RIS Bridge . . . . .
Starting the Installation . . . .
Basic Bridge Setup . . . . . .
Different Configuration Steps .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
5-1
5-1
5-2
5-5
6 ______ RIS Bridge on RIS Computer _____________________________________ 6 - 1
Preconditions . . . . . . . . . . .
Installing the RIS Bridge . . . . .
Installer Setup . . . . . . . .
Starting the Installation . . . .
Basic Bridge Setup . . . . . .
Different Configuration Steps .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
6-1
6-1
6-1
6-2
6-3
6-5
7 ______ File Polling ____________________________________________________7 - 1
General . . . . . . . . . . . . . . . . . . . .
Configuring the syngo MammoReport system
Configuring the RIS Computer . . . . . . . .
Checking the File Polling RIS Connector . . .
Troubleshoot Filesharing . . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
7-1
7-2
7-8
7-9
7-9
8 ______ HTTP Server RIS Bridge _________________________________________8 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Configuring the syngo MammoReport system . . . . . . . . . . . . . . . . . . . . . 8 - 1
Preparing the RIS computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 5
9 ______ MedBroker RIS Bridge___________________________________________9 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Configuring the syngo MammoReport system . . . . . . . . . . . . . . . . . . . . . 9 - 1
Preparing the RIS computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 4
10 _____ Medos RIS Bridge _____________________________________________10 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 - 1
Configuring the syngo MammoReport system . . . . . . . . . . . . . . . . . . . . .10 - 1
Checking the MEDOS RIS Connector . . . . . . . . . . . . . . . . . . . . . . . . .10 - 5
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
0-5
Table of Contents
11 _____ Mitra Context Server ___________________________________________11 - 1
General . . . . . . . . . . . . . . . . . . . . . . .
Prerequisites . . . . . . . . . . . . . . . . . . . .
Network . . . . . . . . . . . . . . . . . . . . .
RIS Computer. . . . . . . . . . . . . . . . . .
Configuring the syngo MammoReport system . . .
Configuring the RIS Bridge . . . . . . . . . . . . .
Checking the Mitra Context Server RIS Connector.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.11 - 1
.11 - 1
.11 - 1
.11 - 1
.11 - 2
.11 - 5
.11 - 6
12 _____ PACSBridge RIS Connector _____________________________________12 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 - 1
Configuring the syngo MammoReport system . . . . . . . . . . . . . . . . . . . . .12 - 1
Preparing the RIS computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 - 7
13 _____ PenRad Serial RIS Bridge _______________________________________13 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 - 1
Configuring the syngo MammoReport system . . . . . . . . . . . . . . . . . . . . .13 - 1
Preparing the RIS computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 - 4
14 _____ PenRad XML RIS Bridge ________________________________________14 - 1
General . . . . . . . . . . . . . . . . . . . .
Configuring the syngo MammoReport system
Configuring the RIS Computer . . . . . . . .
Checking the PenRad XML RIS Connector .
Troubleshoot File Sharing . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.14 - 1
.14 - 1
.14 - 5
.14 - 5
.14 - 5
15 _____ Process Starter _______________________________________________15 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 - 1
Configuring the syngo MammoReport system . . . . . . . . . . . . . . . . . . . . .15 - 1
16 _____ Appendix A___________________________________________________16 - 1
Find a Workable RIS Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 1
Basic Data Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 2
17 _____ Appendix B___________________________________________________17 - 1
Importing the License File . . . . . . . . . . . . . . . . . . . . .
Disabling the Automatic Start of Virtual Machine. . . . . . . . . .
Backup of RIS Configuration Settings . . . . . . . . . . . . . . .
Backup with RIS bridge on syngo MammoReport Workstation.
Backup with RIS bridge on Virtual Machine . . . . . . . . . .
Backup with RIS bridge on RIS “Standalone” . . . . . . . . .
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.17 - 1
.17 - 1
.17 - 2
.17 - 2
.17 - 2
.17 - 3
Siemens
0-6
Table of Contents
Restoring the RIS Configuration Settings . . . . . . . . . . . . . .
Restoring with RIS bridge on syngo MammoReport Workstation
Restoring with RIS bridge on Virtual Machine . . . . . . . . . .
Restoring with RIS bridge RIS “Standalone” . . . . . . . . . . .
Backup of Virtual Machine Settings . . . . . . . . . . . . . . . . .
Restoring the Virtual Machine Settings . . . . . . . . . . . . . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.17 - 4
.17 - 4
.17 - 4
.17 - 5
.17 - 6
.17 - 6
18 _____ Trouble Shooting ______________________________________________18 - 1
General File Sharing . . . . . . . . . . . . . .
Check the File Sharing Settings . . . . . .
Installation of RIS bridge failed . . . . . . . .
Patient synchronization MammoReport to RIS .
Troubleshooting General synchronization .
Troubleshooting XAppInterface . . . . . . .
XAppSettings.xml . . . . . . . . . . . . . .
Troubleshooting Bridge . . . . . . . . . . .
Patient synchronization RIS to MammoReport .
Troubleshooting. . . . . . . . . . . . . . .
Troubleshooting General synchronization .
Troubleshooting Bridge . . . . . . . . . . .
Troubleshooting XAppInterface . . . . . . .
Connectivity problems with bridge in VM . . .
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.18 - 1
.18 - 1
.18 - 1
.18 - 1
.18 - 1
.18 - 2
.18 - 2
.18 - 3
.18 - 4
.18 - 4
.18 - 4
.18 - 5
.18 - 6
.18 - 6
19 _____ Changes to previous version ____________________________________19 - 1
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
General Information
1
1-1
Purpose of Document
1
The purpose of syngo MammoReport RIS Configuration is to provide instructions on how
to install and configure RIS connections with syngo MammoReport system.
Target Group
1
This manual is intended for customer support engineers.
Requirements
1
Required Tools, Measurement, and Auxiliary Devices
1
NOTE
•
All tools, measurement and auxiliary devices marked “ * ”, are
listed along with their specifications in the STC (Service Tools
Catalogue).
Standard service equipment*
Required Documents
1
• Safety Information
• RIS Connectivity
Siemens
TD00-000.860 ...
SPB7-420.850 ...
SPB7-420.815.51.06.02
12.11
Page 1 of 4
CS PS SP
syngo MammoReport
1-2
General Information
Writing Conventions
1
Text Emphasis
1
WARNING
Short description of hazard!
Consequence - optional:
WARNING indicates a risk of danger that may lead to death or
serious injury.
♦ Advice to avoiding the hazard.
CAUTION
Short description of hazard!
Consequence - optional:
CAUTION with the safety alert symbol indicates a risk of danger
that leads to slight or moderate physical injury and /or damage to
property.
♦ Advice to avoiding the hazard.
NOTICE
Short description of hazard!
Consequence - optional:
NOTICE used without the safety alert symbol indicates a risk of
danger that if disregarded leads or may lead to a potential situation, which may result in an undesirable result or state other than
death, physical injury or property damage.
♦ Advice to avoiding the hazard.
NOTE
NOTE contains information provided with special emphasis to
facilitate proper use of the equipment or proper execution of a
procedure, i.e. hints, tips.
Safety Information and Protective Measures
1
The product specific safety information contained in this document, as well as the general
safety information must be observed, see “Safety Information” document., TD00-000.860
...
NOTE
syngo MammoReport
Be aware that this product is intended to be used in a non-patient
environment.
SPB7-420.815.51.06.02
12.11
Page 2 of 4
CS PS SP
Siemens
General Information
1-3
Acronyms and Abbreviations
ABVS
AE Title (AET)
AWS
Conformance
CSE
DICOM
DHCP
Dmax
FFDM
HR
IM
IP Address
MIS
MWLS
OS
Q/R
Port
RIS
Subnet mask
Router
Service Class
SCP
SCU
SMPTE
UI
Siemens
1
Automated Breast Volume Scanner
DICOM Application Entity Title. Must be unique in a DICOM network
Acquisition Workstation
Each manufacturer of a DICOM device has to provide such a statement
which gives an overview of the products DICOM capability.
Customer Support Engineer
Digital Imaging and Communication in Medicine
Dynamic Host Configuration Protocol. Reduces the complexity of configuring computers for TCP/IP networks. The following TCP/IP configuration
can be dynamically assigned by a DHCP server: IP Address, Subnet
masks, gateway and additional parameters like domain name.
The darkest area of an image that a device can reproduce and still have
detail. A Dmax value of 4.0 would be considered perfect.
Full-Field Digital Mammography
High Resolution Display
Port address of image management to query/retrieve images on a
MagicStore PACS
32-bit address assigned to hosts using TCP/IP. An IP address belongs to
one of four classes (A, B, C, D) and is written as 4 pairs separated with
periods (decimal point format).
The IP is part of the socket address. Here IP means Internet Protocol.
Mammography Information System
Modality Worklist Server
Operating System
Query / Retrieve is a service class which provides the ability to retrieve/
transfer a well-identified set of images.
Part of the socket address. Different DICOM services on the same host
PC may use different port numbers.
Radiology Information System
The subnet mask is used to extract network and sub network information
from the IP address.
A device that connects different networks. In a TCP/IP environment a
router is used to connect computers that are located in different IP
address ranges.
A structured description of a service which is supported by cooperating
DICOM Application Entities, e.g. Storage Service Class.
Service Class Provider, e.g. a Storage Service Class Provider provides
the capability to receive DICOM images from a DICOM AET Storage
SCU.
Service Class User, e.g. a DICOM Storage Service Class User can send
an image to a DICOM Service Class Provider.
Society of Motion Picture and Television Engineers
User Interface
SPB7-420.815.51.06.02
12.11
Page 3 of 4
CS PS SP
syngo MammoReport
1-4
General Information
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 4
CS PS SP
Siemens
Overview
2
2-1
Preconditions
2
At first check the following items:
•
RIS computer is in the same network as syngo MammoReport
•
RIS computer requires a different static IP address in the same subnet as syngo
MammoReport
•
MBC-RIS license is available on syngo MammoReport
(to import the license file, refer to the section Importing the License File on page 17 - 1)
•
Access to syngo MammoReport “Local Service” is granted
•
Access to syngo MammoReport “SCR Settings” is granted
•
Access to RIS computer as administrator is granted
•
RIS bridges and installer are available on CD
•
If the RIS application shall be installed on a Virtual Machine, additionally the
administrator password for Virtual Machine is required.
NOTE
All necessary passwords will be communicated via training
courses and Siemens Intranet only.
Restrictions
NOTE
2
VMware supports 256MB RAM only.
Consider this before any installation of third party software.
Delivery
2
The RIS Connection package must be ordered additionally to an existing syngo MammoReport system VB24 or higher.
The RIS Connection package contains:
- syngo MammoReport RIS Bridge CD with setup and installation for the RIS bridges
- MBC-RIS license
If the RIS application shall be installed on a Virtual Machine which runs on the syngo
MammoReport system, additionally the Package of VMware for RIS Client must be
available.
The VMware Package contains:
- syngo MammoReport VMware DVD with pre-configured Virtual Machine
(VMware 6.0)
- VMware serial number as license
- Windows XP license label for the Virtual Machine
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 8
CS PS SP
syngo MammoReport
2-2
Overview
Scope of this Document
2
First, read the functional description RIS Connectivity, SPB7-420.850 ... on page 1 - 1 to
clarify the customer’s requirements, infrastructure and equipment before starting any RISconfiguration.
Supported RIS Systems
2
The following table lists all RIS systems to be connected to syngo MammoReport
System
Vendor
Bridge
Configuration
syngo Workflow MLR
(Magic SAS)
syngo Workflow SLR
(Novius)
syngo WorkflowPortal
RadCentre
MammaSoft
GE Centricity
CRIS
NBSS
MaSc
MIS
ImageFlow
syngo Workflow
Portal (MLR)
MedOutlook
MEDOS
Kodak RIS 2010
TRIS2000
CRIS
Conrad (1)
Siemens
File Polling
see page 7 - 1
Siemens
iSoft
KV Bayern
General Electrics
Healthcare Software Systems
Temenos
KV No / KV WL
HTTP Server
Commit
ImageFlow
Siemens
Keymind
MEDOS
Kodak
Risco
Healthcare Software Systems
Sectra, Siemens, Sweden
PowerScribe
Dictaphone
(Unknown Dictation
Lanier/Medquist
System)
(Unknown Dictation
Converter by
System) (1)
Siemens, Canada
(Unknown)
PenRad
PenRad RIS
PenRad
MagView Information MagView
Systems
MIS used by Grand
(Unknown)
River Hospital, Canada
Radiant
EPIC
Tab. 1
syngo MammoReport
MedBroker
MEDOS
Mitra Context
Server
MR-DDEBridge
(3. party bridge)
PACSBridge
see page 8 - 1
see page 9 - 1
see page 10 - 1
see page 11 - 1
(1)
see page 12 - 1
(1)
PACSBridge +
3. party converter
PenRad Serial
see page 14 - 1
PenRad XML
see page 13 - 1
Supported RIS systems
SPB7-420.815.51.06.02
12.11
Page 2 of 8
CS PS SP
Siemens
Overview
2-3
System
Vendor
n.a.
Tab. 1
Bridge
n.a.
Process Starter
Configuration
(2)
see page 15 - 1
Supported RIS systems (Forts.)
The RIS bridge configuration for systems marked with (1) are not described in detail,
because 3. party information is affected. Refer to section above: Scope of this Document
on page 2 - 2.
(2)
Process Starter allows starting applications/processes that are usable for a RIS
connection.
NOTE
If you can’t find your RIS in this table please refer to Appendix A:
Find a Workable RIS Configuration on page 16 - 1.
RIS Bridge Specification
2
The following table gives an overview of the recommended port settings and the synchronization type (sMR = syngo MammoReport).:
Name
Recommended Port
Synchronization Type
Epic Radiant
File Polling
HTTP Server
MedBroker
Mitra Context Server
MEDOS
MR-DDEBridge
PACSBridge
PACSBridge + Converter
PenRad Serial
PenRad XML
Process Starter
4701
4701
4701
4701
4701
4701
4701
4701
4701
4701
4701
4701
unidirectional RIS -> sMR
unidirectional RIS -> sMR
unidirectional RIS -> sMR
bidirectional sMR <-> RIS
bidirectional sMR <-> RIS
bidirectional sMR <-> RIS
bidirectional sMR <-> RIS
unidirectional sMR -> RIS
unidirectional sMR -> RIS
unidirectional sMR -> RIS
bidirectional sMR <-> RIS
unidirectional sMR -> RIS
Tab. 2
Siemens
Names and Port Numbers for Bridges
SPB7-420.815.51.06.02
12.11
Page 3 of 8
CS PS SP
syngo MammoReport
2-4
Overview
Location of the RIS Bridge
2
The procedure of installing and configuring the RIS connection depends on the decision,
where the RIS bridge shall be installed.
NOTE
As first option, RIS bridges shall be installed on syngo MammoReport workstation.
Only for specified exeptions, use Virtual Machine RIS computer to
install the RIS bridge on.
Use the RIS “Standalone” only as last resort to install the RIS
bridge on.
There are 3 options where to install the RIS bridge:
•
As standard option, RIS bridges shall be installed directly on syngo MammoReport
workstation.
Continue with overview on page 2 - 5 or refer to details on page 3 - 1.
•
It is strongly recommended NOT to install the Mitra Context Server bridge on the syngo
MammoReport workstation. Install this RIS bridge on a Virtual Machine (VMware),
where the RIS application is running separately on the syngo MammoReport computer.
Continue with overview on page 2 - 6 or refer to details on page 4 - 1.
•
Only as last resort a RIS bridge can also be installed on a stand-alone
RIS computer, where the RIS application is already running.
Continue with overview on page 2 - 7 or refer to details on page 6 - 1.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 8
CS PS SP
Siemens
Overview
2-5
Procedure Options
2
RIS Bridge directly on the syngo Computer
2
The required steps for this option are:
1. RIS application already installed on the RIS computer (by RIS administrator/service)
2. Installing the RIS bridge.
3. Configuring the SCR settings for the RIS bridge
- done automatically from syngo MammoReport RIS Bridge CD
- different settings, depending on RIS bridge
4. Configuring the settings of the RIS bridge
- to be done manually on syngo MammoReport workstation
- different settings, depending on RIS bridge
5. Configuring the RIS computer (by RIS administrator/service)
syngo MammoReport system
RIS computer:
including RIS license
RIS bridge:
installed from CD RIS
for syngo MammoReport
RIS application:
installed by RIS service
(2)
(1)
bridge specific settings
entered manually
RIS connection
SCR Settings RIS:
RIS application:
(5)
configured by RIS service
SCR configuration:
(3)
Name, Host, Port
RIS bridge configuration
of RIS bridge
entered automatically
RIS configuration:
open the configuration
dialog on the RIS bridge
Fig. 1
(4)
Overview: RIS bridge directly on the syngo workstation
Refer to details on page 3 - 1.
Siemens
SPB7-420.815.51.06.02
12.11
Page 5 of 8
CS PS SP
syngo MammoReport
2-6
Overview
RIS Bridge on a Virtual Machine
2
The required steps for this option are:
1. Installing the Virtual Machine (VMware) on the syngo MammoReport computer using
the syngo MammoReport VMware DVD
2. Installing the RIS application on the VM workstation (by RIS administrator/service)
3. Installing the RIS bridge on the VM workstation using the syngo MammoReport RIS
Bridge CD
4. Configuring the SCR settings for the RIS bridge
- to be done manually on the syngo MammoReport workstation
- different settings, depending on RIS bridge
5. Configuring the settings of the RIS bridge
- to be done manually in VMware (running on syngo MammoReport workstation)
- different settings, depending on RIS bridge
6. Configuring the RIS computer (by RIS administrator/service)
syngo MammoReport system
including RIS license
Virtual Machine:
(1)
VM installed from DVD
VMware syngo MammoReport
SCR Settings RIS:
SCR configuration:
Name, Host, Port
of RIS bridge
entered manually
(4)
RIS configuration:
open the configuration
dialog on the RIS bridge
RIS application:
(2)
to be installed by RIS service
RIS connection
RIS bridge configuration
RIS bridge:
installed from CD RIS
for syngo MammoReport
bridge specific settings
entered manually
(3)
(5)
RIS application:
(6)
configured by RIS service
Fig. 2
Overview: RIS bridge on a Virtual Machine
Refer to details on page 4 - 1.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 8
CS PS SP
Siemens
Overview
2-7
RIS Bridge on a Stand-alone RIS Computer
2
The required steps for this option are:
1. RIS application already installed on the RIS computer (by RIS administrator/service)
2. Performing the initial setup on the RIS computer using syngo MammoReport RIS
Bridge CD
3. Installing the RIS bridge on the RIS computer using syngo MammoReport RIS Bridge
CD
4. Configuring the SCR settings for the RIS bridge
- to be done manually on the syngo MammoReport workstation
- different settings, depending on RIS bridge
5. Configuring the settings of the RIS bridge
- to be done manually on the RIS computer
- different settings, depending on RIS bridge
6. Configuring the RIS computer (by RIS administrator/service)
RIS computer:
syngo MammoReport system
including RIS license
SCR Settings RIS:
SCR configuration:
Name, Host, Port
of RIS bridge
entered manually
(4)
RIS configuration:
open the configuration
dialog on the RIS bridge
Fig. 3
RIS connection
RIS application:
installed by RIS service
(1)
initial setup from CD RIS
for syngo MammoReport
(2)
RIS bridge:
installed from CD RIS
for syngo MammoReport
RIS bridge configuration
(3)
bridge specific settings
entered manually
(5)
RIS application:
configured by RIS service
(6)
Overview: RIS bridge on a stand-alone RIS computer
Refer to details on page 6 - 1.
Siemens
SPB7-420.815.51.06.02
12.11
Page 7 of 8
CS PS SP
syngo MammoReport
2-8
Overview
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 8 of 8
CS PS SP
Siemens
RIS Bridge on syngo Workstation
3
3-1
Preconditions
3
•
The hospital network administrator has to be contacted to check that the network
communication is opened for the required data exchange.
•
The setup for the RIS computer should be carried out by the RIS administrator/service.
•
The RIS computer must be reachable via TCP/IP from the syngo MammoReport
system.
•
The RIS Connector is a license protected feature of the syngo MammoReport system.
NOTE
The license number that is requested directly by the production
line is supplied via file.
NOTE
The RIS computer may be a Virtual Machine also.
For setup see RIS on Virtual Machine on page 4 - 1.
Installing the RIS Bridge
NOTE
3
It is strongly recommended NOT to install the Mitra Context
Server bridge on the syngo MammoReport workstation:
This bridge shall be installed on a Virtual Machine RIS computer.
Use the RIS “Standalone” only as last resort.
Starting the Installation
3
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows.
5. Insert the syngo MammoReport RIS Bridge CD.
6. In the Explorer change to drive D:\.
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
3-2
RIS Bridge on syngo Workstation
7. Start the installation by double-clicking on Installer.exe.
The bridge installer opens with required preset values for SCR Settings.
Fig. 1
RIS Bridge Installer
As the SCR settings are already done by installation routine, the default values are:
•
Name: SCR
•
Host: localhost
•
Port: 4700
NOTE
Delete any bridge that is not needed.
If deletion is required, perform the following steps:
• First, select existing item in RIS Bridges Configured section.
• Click Delete to free the list.
A confirmation dialog is opened:
• Click Yes in the Confirm Unconfigure dialog.
• To free the RIS Bridge Installed section, repeat these steps.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
RIS Bridge on syngo Workstation
3-3
Basic Bridge Setup
3
The following description exemplifies the setup of a RIS bridge.
1. Select the bridge you want to install from the Bridge list.
(The example shows File Polling.)
Fig. 2
RIS Bridge Installer
NOTE
Siemens
In the following we will name the bridge “File Polling” as some
dialogs may appear and are relating to this bridge. Choose an
applicable name for the bridge to be installed.
We will also describe the port with “4701”. Even this setting is
free, we recommend to use 4701, as this port number is not used
by other applications.
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
3-4
RIS Bridge on syngo Workstation
2. Enter an applicable name for the bridge to be installed and 4701 for Port (1).
(See Fig. 3 below. The example shows File Polling.)
3. Click Install (2) to start the installation procedure. The installation is finished with the
according entries for RIS Bridges Installed / RIS Bridges Configured:
(2)
(1)
Fig. 3
RIS Bridge Installer
4. Click Exit to close the installation window and remove the CD from drive D:
5. Restart the syngo MammoReport system.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
RIS Bridge on syngo Workstation
3-5
Different Configuration Steps
3
The next configuration steps are described for each bridge separately:
•
File Polling, refer to chapter File Polling on page 7 - 1.
•
HTTP Server, refer to chapter HTTP Server RIS Bridge on page 8 - 1.
•
MedBroker, refer to chapter MedBroker RIS Bridge on page 9 - 1.
•
MEDOS, refer to chapter Medos RIS Bridge on page 10 - 1.
•
Mitra Context Server, refer to chapter Mitra Context Server on page 11 - 1.
•
PACSBridge, refer to chapter PACSBridge RIS Connector on page 12 - 1.
•
PenRad Serial, refer to chapter PenRad Serial RIS Bridge on page 14 - 1.
•
PenRad XML, refer to chapter PenRad XML RIS Bridge on page 13 - 1.
•
Process Starter, refer to chapter Process Starter on page 15 - 1.
Uninstalling a RIS Bridge
NOTE
3
The de-installation of a syngo MammoReport RIS bridge depends
on the location, where you have installed the RIS bridge.
RIS bridge installed on RIS client
3
•
On the RIS client, delete the folder
C:\Programs\MeVisBreastCare\XAPP\Bridges\HttpServer0 and all it's content.
•
In the RIS client AutoStart folder [all users], delete the file HttpServer0.bat.
RIS bridge installed on syngo MammoReport
3
•
On the syngo MammoReport, delete the folder C:\MBCXapp\Bridges\HttpServer0
and all it's content.
•
Delete the file C:\MBCXapp\pm\HttpServer0.pm.
In every case:
1. On the syngo MammoReport, run syngo, login as MammoAdmin and start the SCR
settings dialog.
2. Delete the RIS bridge in the SCR Settings.
3. Restart syngo MammoReport
NOTE
Siemens
Install new RIS bridges only with a released version of the RIS
bridge Installation CD.
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
3-6
RIS Bridge on syngo Workstation
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
RIS on Virtual Machine
4
4-1
Procedure Steps
NOTICE
4
Limited memory available.
Do only install the RIS application (by RIS service) and if required
the RIS bridge (by SIEMENS service).
Do not install other software on the Virtual Machine. This can
raise problems because of limited memory.
1. Installing the Virtual Machine (VMware) on the syngo MammoReport computer using
syngo MammoReport VMware DVD
2. Installing the RIS application on the Virtual Machine (by RIS administrator/service)
syngo MammoReport system
including RIS license
Virtual Machine:
SCR RIS interface:
VM installed from DVD VMware
VMware syngo MammoReport
SCR configuration:
Name, Host, Port
of RIS bridge
entered manually
RIS application:
to be installed by RIS service
RIS configuration:
open the configuration
dialog on the RIS bridge
Fig. 1
RIS connection
RIS bridge (1):
installed from CD RIS
for syngo MammoReport
RIS bridge configuration
bridge specific settings
entered manually
Overview: RIS bridge on a Virtual Machine
(1) = Optional
If the RIS bridge is installed on the Virtual Machine, refer to details on page 5 - 1.
Siemens
•
Installing the RIS bridge on the Virtual Machine using syngo MammoReport RIS
Bridge CD
•
Configuring the SCR settings for the RIS Bridge
- to be done manually on the syngo MammoReport workstation
- different settings, depending on RIS bridge
•
Configuring the settings of the RIS bridge
- to be done manually from syngo MammoReport workstation
- different settings, depending on RIS bridge
SPB7-420.815.51.06.02
12.11
Page 1 of 20
CS PS SP
syngo MammoReport
4-2
RIS on Virtual Machine
Installing the Virtual Machine
4
The installation comprises the following steps:
•
Disabling the Auto Login for syngo meduser
•
Basic Installation
•
Configuration
•
Using the Virtual Machine with syngo MammoReport
•
Enabling the Auto Login for syngo meduser
The installation takes about 45 minutes.
Disabling the Auto Login for syngo meduser
NOTE
4
Using the shift key while booting to enable the log in window
does NOT work on systems with USB keyboard installed.
1. Log in to syngo as Administrator.
2. In the Local Service windows click Configuration.
3. Go to Local Host ⇒ Users
4. Uncheck the Enable Autologin function for meduser.
5. Click Save, then Home and close the Local Service.
6. Restart the system via the End Session dialog.
Basic Installation
4
1. Log in as OS administrator to the syngo MammoReport computer.
2. Disconnect the network cable from the computer.
3. Insert the syngo MammoReport VMware DVD to DVD drive D:\.
4. In the Explorer change to D:\ and wait until the disk is loaded. This may take about a
minute.
5. Start the installation by double-clicking install_VM.bat.
VMware files are copied from the DVD.
This takes up to 10 minutes.
No progress indicator is shown while copying a ~ 3.5 GB file.
Fig. 2
syngo MammoReport
VMware Installation
SPB7-420.815.51.06.02
12.11
Page 2 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4-3
The Virtual Machine installation is started automatically:
Fig. 3
VMware Installation
The installation takes about 3 minutes:
Fig. 4
VMware Installation
6. Wait until the installation opens the Registration Information window.
The serial number / license is delivered together with the syngo MammoReport
VMware DVD.
NOTE
Siemens
Never use the same serial number for another Virtual Machine.
SPB7-420.815.51.06.02
12.11
Page 3 of 20
CS PS SP
syngo MammoReport
4-4
RIS on Virtual Machine
7. Enter the Serial Number (License Key) and click Enter.
Fig. 5
VMware Installation
8. On the Installation Wizard Completed window click Finish.
Fig. 6
VMware Installation
9. Remove syngo MammoReport VMware DVD from drive D:\.
10. On the following message, click Yes.
Fig. 7
VMware Installation
11. After restart of the computer, login again as OS administrator.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4-5
Starting the Virtual Machine
NOTE
4
With the first start of the VMware player, the Virtual Machine will
try to connect to the network.
The network cable must be disconnected before continuing.
As long as the network connection is disabled, the following message about missing network connection will appear during start process of Virtual Machine. There is no need to
react.
Fig. 8
VMware Player Message
1. Open the Explorer and browse to C:\VMRIS.
2. Select Windows XP Professional.vmx via right mouse click and choose Open with...
Fig. 9
Siemens
VMware Player
SPB7-420.815.51.06.02
12.11
Page 5 of 20
CS PS SP
syngo MammoReport
4-6
RIS on Virtual Machine
3. In the Open With window click Browse... and go to
C:\Program Files\VMware\VMware Workstation.
Fig. 10 VMware Player
4. Select vmplayer.exe and click Open.
5. Check Always use the selected program to open this kind of file.
Fig. 11 VMware Player
6. Click OK. The VMware Player starts.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4-7
7. On the License Agreement window select Yes, I accept the terms of the license
agreement.
Fig. 12 VMware Player
8. Click OK.
While the starting the Virtual Machine several settings are checked:
- If the following VMware Player message appears, make sure that I copied it is
selected. Then click OK.
Fig. 13 VMware Player Message
- If the following VMware Player message about missing virtual floppy device appears,
click OK.
Fig. 14 VMware Player Message
- The following message about missing network connection appears in the VMware
Player. There is no need to react.
Fig. 15 VM Player Message
- The following message about missing sound device may appear in the VMware
Player. There is no need to react.
Fig. 16 VM Player Message
Siemens
SPB7-420.815.51.06.02
12.11
Page 7 of 20
CS PS SP
syngo MammoReport
4-8
RIS on Virtual Machine
Windows XP System Preparation
4
With the first start the Windows XP Professional setup makes the Virtual Machine unique.
1. After some minutes the setup opens the window for setting the computer name and the
administrator password.
NOTE
To change the cursor location in-between syngo MammoReport
display and VMware Player display press Ctrl+Alt for back and
Ctrl+G for forth direction.
2. Click twice into the Windows XP Setup window and enter the Computer name,
e.g. VM_<host name>.
Fig. 17 VMware Player
NOTE
Do NOT change the administrator password here.
NOTE
The default password for the administrator account of the Virtual
Machine / RIS computer is the syngo administrator password.
The syngo administrator password is used only for installation
purpose. It shall be changed later.
3. Click Next and wait for System Preparation finishs and restarts the Virtual Machine.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 8 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4-9
4. After the reboot a file named DskCnfg.ini is opened automatically:
close the file without any further actions via mouse click on the cross in the upper right
window corner.
Fig. 18 VMware Player
System Setup:
- A DOS box is opened while the installation of syngo security files is running.
- After several minutes the Virtual Machine reboots automatically.
- The Windows login dialog is displayed.
Fig. 19 VMware Player
5. Log in to the Virtual Machine as administrator with the syngo administrator password.
Siemens
SPB7-420.815.51.06.02
12.11
Page 9 of 20
CS PS SP
syngo MammoReport
4 - 10
RIS on Virtual Machine
6. Maximize the VMware Player window to fit the VMware Player Windows desktop to the
syngo display.
Fig. 20 VMware Player
The Windows Start button is now visible.
VMware Mouse Driver
4
1. In the VMware Player select Start ⇒ Settings ⇒ Control Panel and select Mouse.
Fig. 21 VM Mouse Driver
2. Double-click to open the Mouse Properties window.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 10 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4 - 11
3. Change to the Hardware tab and click Properties.
Fig. 22 VM Mouse Driver
4. Change to the Driver tab and click Roll Back Driver.
Fig. 23 VM Mouse Driver
5. On the following message confirm the roll back with Yes.
Fig. 24 VM Mouse Driver
Siemens
SPB7-420.815.51.06.02
12.11
Page 11 of 20
CS PS SP
syngo MammoReport
4 - 12
RIS on Virtual Machine
6. Check that VMware, Inc is displayed as Driver Provider.
Fig. 25 VM Mouse Driver
7. Click Close.
8. On the following message click Yes to confirm the restart of the Virtual Machine.
Fig. 26 VM Mouse Driver
9. After restart, log in to the Virtual Machine as administrator.
Configuration
4
1. On the upper border inside the VMware Player open the Main Menu ⇒ VMware Player
⇒ Preferences and make sure that Confirm before existing the application is NOT
checked.
Fig. 27 VM Configuration
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 12 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4 - 13
2. Remove the tick at VMware Web update.
3. Click OK to save. The dialog is closed.
4. In the VMware Player select Start ⇒ Settings ⇒ Control Panel and select the VMware
Tools.
Fig. 28 VM Configuration
5. Double-click the VMware Tools to open the Properties window.
6. Check that the Time synchronization between the virtual machine and the host
operating system is active.
Fig. 29 VM Configuration
7. Remove the tick at Notify if upgrade is available.
8. Click Apply and OK. The dialog is closed.
Siemens
SPB7-420.815.51.06.02
12.11
Page 13 of 20
CS PS SP
syngo MammoReport
4 - 14
RIS on Virtual Machine
9. In the Control Panel window open the User Accounts by double-clicking.
Fig. 30 VM Configuration
10. In the User Accounts window open Change the way users log on or off.
Fig. 31 VM Configuration
11. Remove the tick for Use the Welcome screen and click Apply Options.
Fig. 32 VM Configuration
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 14 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4 - 15
12. In the User Accounts window open the RISUser (this is a demo user for testing
purpose).
Fig. 33 VM Configuration
13. Open Change the password.
Fig. 34 VM Configuration
14. Enter a password for the RIS test user.
Fig. 35 VM Configuration
15. Click Change Password to save the password.
Siemens
SPB7-420.815.51.06.02
12.11
Page 15 of 20
CS PS SP
syngo MammoReport
4 - 16
RIS on Virtual Machine
16. Back in the User Accounts window open the Administrator account.
Fig. 36 VM Configuration
17. Open Change the password.
18. Enter the password for the Virtual Machine administrator.
19. Confirm the new password and click Change Password to save it.
20. Close the User Accounts window, but leave the Control Panel window open.
Network Connection
NOTE
4
The Virtual Machine used for RIS requires a different static IP
address in the same subnet as syngo MammoReport. This shall
be configured here.
Contact to the IT department of the hospital may be required.
1. In the Control Panel window of Virtual Machine open the Network Connections by
double-clicking
Fig. 37 VM Configuration
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 16 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4 - 17
2. Via right mouse click on Network Connections open the context menu and click
Properties.
Fig. 38 VM Configuration
3. In General tab select Internet Protocol (TCP/IP) and click Properties.
Fig. 39 VM Configuration
4. Enter a static IP address and the same subnet as used for the syngo MammoReport
workstation.
Fig. 40 VM Configuration
5. Click OK to save the Internet Protocol Properties.
6. Click Close to save the Network Connections Properties.
7. Restart the Virtual Machine via Start ⇒ Shutdown ⇒ Restart.
Siemens
SPB7-420.815.51.06.02
12.11
Page 17 of 20
CS PS SP
syngo MammoReport
4 - 18
RIS on Virtual Machine
Checking the VMware Player
4
1. Wait until the Log On to Windows prompt of the Virtual Machine appears.
Fig. 41 VM Configuration
2. Log in as RISUser. The RISUser desktop is opened with a black background.
3. Exit the VMware Player via Main Menu ⇒ VMware Player ⇒ Exit.
4. Observe the task bar: wait until VMware Player window is closed automatically.
5. From C:\VMRIS open again Windows XP Professional.vmx by double-clicking.
The VMware Player restores the status from before last exit: the RISUser is logged in,
the desktop shows black background.
6. Again, exit the VMware Player via Main Menu ⇒ VMware Player ⇒ Exit.
Using the Virtual Machine with syngo MammoReport
4
The Virtual Machine needs to be configured also for the meduser account.
1. On the desktop of the syngo MammoReport computer open the Explorer and browse to
C:\MBCSCR\recovery.
2. Run autologin_meduser.exe by double-clicking. Wait until the auto login for meduser is
set.
3. Re-connect the network cable.
4. Restart the syngo MammoReport system via Start ⇒ Shut Down ⇒ Restart.
5. Login to syngo as Administrator.
6. On the VMware Player License Agreement window select Yes, I accept the terms of
the license agreement.
Fig. 42 VMware Player: Licence Agreement
The Virtual Machine starts and finally the last state is restored with the RISUser logged in.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 18 of 20
CS PS SP
Siemens
RIS on Virtual Machine
4 - 19
7. Maximize the VMware Player window to fit the VMware Player Windows desktop to the
syngo monitor.
8. Open inside the VMware Player the Main Menu ⇒ VMware Player ⇒ Preferences and
make sure that Confirm before existing the application is NOT checked.
Fig. 43 Meduser account: VMware Player Preferences
9. Remove the tick at VMware Web update.
10. Click OK to confirm settings and close this dialog.
NOTE
Removing the tick at VMware Web update is recommended for all
RIS accounts, that will be created later on the Virtual Machine and
shall use the syngo MammoReport RIS connection.
11. Minimize the VMware Player.
Enabling the Auto Login for syngo meduser
4
1. Open Main menu ⇒ Options ⇒ Service ⇒ Local Service
2. Go to Configuration ⇒ Local Host ⇒ Users
3. Check the function Enable Auto login for meduser.
4. Click Save, then Home and close the Local Service.
5. Archive the VMware license string together with syngo MammoReport VMware
DVD.
NOTE
Siemens
If the customer does NOT want to start the Virtual Machine automatically together with the syngo MammoReport, refer to Appendix B: Disabling the Automatic Start of Virtual Machine on page 17
- 1.
SPB7-420.815.51.06.02
12.11
Page 19 of 20
CS PS SP
syngo MammoReport
4 - 20
RIS on Virtual Machine
Installing the RIS Application
4
NOTE
The Virtual Machine used for RIS requires a different static IP
address in the same subnet as syngo MammoReport. If not
already done, this may be configured here.
NOTE
The RIS application needs to be installed by the RIS administrator
or service.
NOTE
The contact to the IT department of the hospital may be required.
NOTE
MBC-RIS license on syngo MammoReport is required.
1. Open the syngo Patient Browser and click the RIS button:
Fig. 44 RIS Button
2. All Windows XP settings and installation steps for the RIS application can now be done
by RIS administrator or service.
3. Log in to the Virtual Machine as a RIS user for testing purpose.
NOTE
The RIS user for testing shall have the same access rights as a
normal RIS user.
A RIS user needs full access to the folder
C:\MeVisBreastCare\XApp.
syngo MammoReport
NOTE
You need administrator rights on the Virtual Machine to create a
new RIS user.
NOTE
For backup of the Virtual Machine, refer to Appendix B: Backup of
Virtual Machine Settings on page 17 - 6.
SPB7-420.815.51.06.02
12.11
Page 20 of 20
CS PS SP
Siemens
RIS Bridge on Virtual Machine
5
5-1
Installing the RIS Bridge
NOTE
5
Skip this section, if the RIS bridge is already installed on the
syngo MammoReport workstation according to RIS Bridge on
syngo Workstation on page 3 - 1.
Starting the Installation
5
1. If not already done: Log in to the Virtual Machine as administrator.
2. Insert the syngo MammoReport RIS Bridge CD.
3. Open the Explorer and change to DVD drive R:.
NOTE
For updating the Explorer display, click into the free area of the
right frame and press the F5 key.
4. Start the installation by double-clicking on Installer.exe.
The bridge installer for the SCR Settings opens:
Fig. 1
RIS Bridge Installer
5. Enter settings of the SCR system:
- SCR for Name (this is the default and must be used)
- the IP address of the syngo MammoReport computer for Host
- 4700 for Port (this is the default and must be used)
This dialog looks like:
Fig. 2
RIS Bridge Installer
6. Click OK.
The Bridge Installer opens the setup window and allows to select the Bridge to be
installed.
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
5-2
RIS Bridge on Virtual Machine
Basic Bridge Setup
5
The following description exemplifies the setup of a RIS bridge.
1. Select the bridge you want to install from the Bridge list.
(The example shows PenRad XML.)
Fig. 3
RIS Bridge Installer
NOTE
syngo MammoReport
In the following we will name the bridge “PenRad XML” as some
dialogs may appear and are relating to this bridge. Choose an
applicable name for the bridge to be installed.
We will also describe the port with “4701”. Even this setting is
free, we recommend to use 4701, as this port number is not used
by other applications.
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
RIS Bridge on Virtual Machine
5-3
2. Enter an applicable name for the bridge to be installed and 4701 for Port.
(The example shows PenRadXML.)
Fig. 4
RIS Bridge Installer
3. Click Install to start the installation. The installation is finished with the according entry
for RIS Bridges Installed.
(The example shows PenRadXML.)
Fig. 5
Siemens
RIS Bridge Installer
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
5-4
RIS Bridge on Virtual Machine
4. Click Exit and remove CD from drive R:
Fig. 6
RIS Bridge Installer
NOTE
To start the bridge, the user must log off and log in again.
5. Log off administrator from Virtual Machine.
6. Log in again as administrator and check on the task bar that the bridge is running.
The bridge runs as a task in systray. An icon stands for this application.
Fig. 7
Systray icon
7. Log off administrator from Virtual Machine.
8. Log in as a RIS user for testing purpose.
NOTE
syngo MammoReport
If connectivity is not successful refer also to Connectivity problems with bridge in VM on page 18 - 6.
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
RIS Bridge on Virtual Machine
5-5
Different Configuration Steps
5
The next configuration steps are described for each bridge separately:
Siemens
•
File Polling, refer to chapter File Polling on page 7 - 1.
•
HTTP Server, refer to chapter HTTP Server RIS Bridge on page 8 - 1.
•
MedBroker, refer to chapter MedBroker RIS Bridge on page 9 - 1.
•
MEDOS, refer to chapter Medos RIS Bridge on page 10 - 1.
•
Mitra Context Server, refer to chapter Mitra Context Server on page 11 - 1.
•
PACSBridge, refer to chapter PACSBridge RIS Connector on page 12 - 1.
•
PenRad Serial, refer to chapter PenRad Serial RIS Bridge on page 14 - 1.
•
PenRad XML, refer to chapter PenRad XML RIS Bridge on page 13 - 1.
•
Process Starter, refer to chapter Process Starter on page 15 - 1.
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
5-6
RIS Bridge on Virtual Machine
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
RIS Bridge on RIS Computer
6
6-1
Preconditions
NOTE
6
Use the RIS “Standalone” only as last resort for installing RIS
bridges.
For details refer to Location of the RIS Bridge on page 2 - 4.
Perform this section only if the RIS bridge can not be installed on the syngo MammoReport workstation or the Virtual Machine.
The operating system of the RIS computer must be Windows XP.
The RIS application is already installed (done by RIS administrator / service).
Installing the RIS Bridge
6
Installer Setup
6
1. Log in to the RIS computer as administrator.
2. Insert the syngo MammoReport RIS Bridge CD.
3. Open the Explorer and change to the CD drive.
4. Start the installation by double-clicking on Installer.exe.
The bridge installer for the Path Settings opens.
Fig. 1
RIS Bridge Installer
5. Check the path of the installation directory. You may change it, if necessary.
6. Click OK. The path settings are executed and the installer closes.
7. Log off administrator and log in again as administrator.
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
6-2
RIS Bridge on RIS Computer
Starting the Installation
6
1. If not already done: log in to the RIS computer as administrator.
2. Insert the syngo MammoReport RIS Bridge CD.
3. Open the Explorer and change to the CD drive.
4. Start the installation time by double-clicking on Installer.exe.
The bridge installer for the SCR Settings opens.
Fig. 2
RIS Bridge Installer
5. Enter as settings of the SCR system:
- SCR for Name (this is the default and must be used)
- the IP address of the syngo MammoReport computer for Host
- 4700 for Port (this is the default and must be used)
This dialog looks like:
Fig. 3
RIS Bridge Installer
6. Click OK.
The Bridge Installer opens the configuration window and allows to select the Bridge to be
installed.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
RIS Bridge on RIS Computer
6-3
Basic Bridge Setup
6
The following description exemplifies the setup of a RIS bridge.
1. Select the bridge you want to install from the Bridge list.
(The example shows Mitra Context Server.)
Fig. 4
RIS Bridge Installer
NOTE
In the following we will name the bridge “Mitra Context Server” as
some dialogs may appear and are relating to this bridge. Choose
an applicable name for the bridge to be installed.
We will also describe the port with “4701”. Even this setting is
free, we recommend to use 4701, as this port number is not used
by other applications.
2. Enter an applicable name and 4701 for Port:
Fig. 5
Siemens
RIS Bridge Installer
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
6-4
RIS Bridge on RIS Computer
3. Click Install to start the installation. The installation is finished with the according entry
for RIS Bridges Installed:
Fig. 6
RIS Bridge Installer
4. Click Exit and remove CD from the CD drive.
NOTE
To start the bridge, the user must log off and log in again.
5. Log off administrator from RIS computer.
6. Log in again as administrator and check on the task bar that the bridge is running.
The bridge runs as a task in systray. An icon stands for this application.
Fig. 7
syngo MammoReport
Systray icon
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
RIS Bridge on RIS Computer
6-5
Different Configuration Steps
6
The next configuration steps are described for each bridge separately:
Siemens
•
File Polling, refer to chapter File Polling on page 7 - 1.
•
HTTP Server, refer to chapter HTTP Server RIS Bridge on page 8 - 1.
•
MedBroker, refer to chapter MedBroker RIS Bridge on page 9 - 1.
•
MEDOS, refer to chapter Medos RIS Bridge on page 10 - 1.
•
Mitra Context Server, refer to chapter Mitra Context Server on page 11 - 1.
•
PACSBridge, refer to chapter PACSBridge RIS Connector on page 12 - 1.
•
PenRad Serial, refer to chapter PenRad Serial RIS Bridge on page 14 - 1.
•
PenRad XML, refer to chapter PenRad XML RIS Bridge on page 13 - 1.
•
Process Starter, refer to chapter Process Starter on page 15 - 1.
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
6-6
RIS Bridge on RIS Computer
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
File Polling
7
7-1
This chapter describes the configuration of the File Polling RIS Connector, named File
Polling bridge for syngo MammoReport VB24 or higher.
General
7
The syngo MammoReport is unidirectional controlled by the RIS computer using one of
the following File Polling systems from different vendors:
•
syngo Workflow MLR (MagicSAS)
•
syngo Workflow SLR (Novius)
•
MammaSoft
•
RadCentre
•
GE Centricity
•
HSS CRIS
•
NBSS UK
Every time patient information is copied to the shared folder on the RIS, certain information (i.e. patient ID, accession number) is polled by the syngo MammoReport, so that the
same patient is opened for reviewing on the high resolution monitors.
With GE Centricity additionally a Close Patient command can be used.
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 10
CS PS SP
syngo MammoReport
7-2
File Polling
Configuring the syngo MammoReport system
7
The File Polling RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
SCR Settings
4. On the SCR Settings tab in the RIS section, click Configure...
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 2 of 10
CS PS SP
Siemens
File Polling
7-3
5. In the RIS applications window, select FilePolling (respectively the name entered
during the RIS installation) and click Edit.
Fig. 2
File Polling RIS: Select
NOTE
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
6. In the Edit RIS Bridge window check Active and click OK to accept settings.
Key assigned
not applicable with this bridge
Fig. 3
Siemens
File Polling RIS: Edit
SPB7-420.815.51.06.02
12.11
Page 3 of 10
CS PS SP
syngo MammoReport
7-4
File Polling
7. Back in the RIS applications window select FilePolling and click Echo.
Fig. 4
File Polling RIS: Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
File Polling RIS: Ping
NOTE
syngo MammoReport
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
SPB7-420.815.51.06.02
12.11
Page 4 of 10
CS PS SP
Siemens
File Polling
7-5
9. Back in the RIS applications window select FilePolling again and click Configure.
Fig. 6
File Polling RIS: Configure
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The FilePolling Bridge Configuration window opens.
Fig. 7
Siemens
File Polling: File Polling Bridge Configuration
SPB7-420.815.51.06.02
12.11
Page 5 of 10
CS PS SP
syngo MammoReport
7-6
File Polling
11. Select the requested File format from the list box, e.g. RadCentre.
Fig. 8
File Polling: File format
The following dialog looks different depending on selected file format.
12. Check that the Directory and the File name are as requested. Leave settings in File
monitoring and Data items as shown in screen shots.
RadCentre settings:
Fig. 9
RadCentre: File name and File monitoring
syngo Workflow MLR (MagicSAS) settings:
Fig. 10 syngo Workflow MLR (MagicSAS): File specification and File monitoring
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 10
CS PS SP
Siemens
File Polling
7-7
MammaSoft, syngo Workflow SLR (Novius) or GE Centricity settings:
Fig. 11 MammaSoft / syngo Workflow SLR (Novius): File specification and File monitoring
HSS CRIS or NBSS settings:
Fig. 12 HSS CRIS/NBSS: File specification and File monitoring
13. Leave the default settings or change to other suitable settings.
14. Click OK to save the entries, OK to close RIS applications dialog (Fig. 6), and Exit SCR
Settings (Fig. 1).
NOTE
If the file format HSS CRIS is configured on syngo MammoReport,
the System Environment Variable PATH must be completed by:
%MBCXAPP%\Bridges\FilePolling0\ImagingClient
If more than one RIS bridge is installed, please check the correct
sequential number of the folder C:\MBCXAPP\Bridges\FilePolling<number> (refer to UID tab, Fig. 3 page Fig. 3 - 4).
Siemens
SPB7-420.815.51.06.02
12.11
Page 7 of 10
CS PS SP
syngo MammoReport
7-8
File Polling
NOTE
If the file format HSS CRIS is configured on syngo MammoReport,
the user check is enable per default:
- syngo user name and HSSCRIS user name must be identical or
- Every user must enter his RIS login values in SCR User
Preferences or
Alternatively the user check can be disabled in C:\MBCXAPP\
Bridges\FilePolling0\ImagingClient\ImagingClient.ini
Configuring the RIS Computer
7
If the RIS bridge is installed on syngo MammoReport, the RIS application needs to be
enabled to write the polling files into the specified folder on the syngo MammoReport
workstation:
- C:\MBCXAPP\Bridges\FilePolling0\FileTransferIn
•
The folder C:\MBCXAPP\Bridges\FilePolling0 on the syngo MammoReport
workstation is already prepared for the file sharing.
•
This folder must be connected from the RIS computer by using the Windows Explorer
function mapping network drive. Therfore use the following path:
- \\<IP address>\FilePolling0
(<IP address> = IP address of the syngo MammoReport workstation)
•
Check the file sharing, e.g. by creating a text file in the shared folder.
NOTE
If the file format NBSS is configured on syngo MammoReport, the
NBBS RIS must be configured to run the following batch on the
syngo MammoReport: computer:
- C:\MBCXAPP\Bridges\FilePolling0\NBSS\
NBSS2MammoReport.bat
If more than one RIS bridge is installed, please check the correct
sequential number of the folder C:\MBCXAPP\Bridges\FilePolling<number> (refer to UID tab, Fig. 3 page Fig. 3 - 4).
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 8 of 10
CS PS SP
Siemens
File Polling
7-9
Checking the File Polling RIS Connector
7
1. Login to syngo MammoReport as user with Radiologist rights.
2. Open a case with the patient ID and accession number on the RIS computer and check
that this patient is opened for reviewing on the high resolution displays of the syngo
MammoReport system.
3. An RIS error message may appear after log in (or user change) and it will reappear as
long as the connection does not work correctly.
4. Click OK on the error message being displayed.
NOTE
Connection error message will reappear upon log in or user
change in case the network connection does not work correctly.
NOTE
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
Troubleshoot Filesharing
7
In case of sharing problems refer to General File Sharing on page 18 - 1.
Siemens
SPB7-420.815.51.06.02
12.11
Page 9 of 10
CS PS SP
syngo MammoReport
7 - 10
File Polling
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 10 of 10
CS PS SP
Siemens
HTTP Server RIS Bridge
8
8-1
This chapter describes the configuration of the HTTP Server RIS Connector for syngo
MammoReport VB24 or higher.
General
8
The RIS and the syngo MammoReport are connected uni-directional, i.e. information
(accession number) is sent from the RIS to syngo MammoReport, but no information
according to patients is sent from syngo MammoReport to the RIS.
If a case is opened for reviewing on the high resolution displays (e.g. by a ViewNow on a
patient, or by a case change in the current work list), certain information is sent over to the
RIS, so that the same case is opened on the RIS.
Configuring the syngo MammoReport system
8
The HTTP Server RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
SCR Settings
4. On the SCR Settings tab in the RIS section, click Configure...
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
8-2
HTTP Server RIS Bridge
5. In the RIS applications window select HTTPServer (respectively the name entered
during the RIS installation) and click Edit.
Fig. 2
RIS HTTP Server: Edit
NOTE
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
6. In the Edit RIS Bridge window tick Active and click OK to accept settings.
Key assigned
not applicable with this bridge
Fig. 3
syngo MammoReport
HTTP Server: Edit RIS Bridge
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
HTTP Server RIS Bridge
8-3
7. Back in the RIS applications window select HTTPServer again and click Echo:
Fig. 4
HTTP Server: Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
HTTP Server: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
9. Back in the RIS applications window select HTTPServer again and click Configure:
Fig. 6
Siemens
HTTP Server: Configure
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
8-4
HTTP Server RIS Bridge
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The HTTP Bridge Configuration window opens.
Fig. 7
HTTP Server: Bridge Configuration
NOTE
Do not confuse the port number for HTTP Server (8888) with port
number for syngo MammoReport (4701).
For more technical information about connection requirements,
refer to RIS Connectivity, SPB7-420.850 ...
11. In the HTTP Bridge Configuration window
- leave the default settings for Local HTTP server port (8888)
- choose the Configuration: MaSc or Commit MIS or ImageFlow
- leave the default settings for the SCR response time out (30 seconds)
NOTE
For syngo Workflow Portal select Commit MIS.
12. Click OK. The window is closed.
13. Back in the RIS applications window (see Fig. 6) click OK to close and Exit SCR
Settings (see Fig. 1).
NOTE
syngo MammoReport
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
HTTP Server RIS Bridge
8-5
Preparing the RIS computer
8
Only for MIS:
For synchronizing with syngo MammoReport the RIS administrator must configure the
MIS to use the following HTTP GET request:
- http://<IPAddresss>:8888/index.html?loadCurrent=<StudyInstanceUID>
Explanation:
- <IP address> = IP address of the system where the HTTP Bridge is running on
(normally syngo MammoReport workstation)
- <StudyInstanceUID> = DICOM StudyInstanceUID of the study to open in syngo
MammoReport
If a problem with the RIS connection occurred, the RIS administrator needs to check the
following settings:
Siemens
•
RIS application (MaSc, Commit MIS or ImageFlow) must be configured for sending to
port 8888 on syngo MammoReport system.
•
The RIS computer must be reachable via TCPI/IP from the syngo MammoReport
system.
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
8-6
HTTP Server RIS Bridge
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
MedBroker RIS Bridge
9
9-1
This chapter describes the configuration of the MedBroker RIS Connector for syngo MammoReport VB24 or higher.
General
9
The RIS and the syngo MammoReport are connected bidirectional, i.e. information
(accession number) is sent from syngo MammoReport to the RIS, and information
according to patients is sent from the RIS to syngo MammoReport.
If a case is opened for reviewing on the high resolution displays (e.g. by a ViewNow on a
patient, or by a case change in the current work list), certain information is sent over to the
RIS, so that the same case is opened on the RIS.
Configuring the syngo MammoReport system
9
The MedBroker RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
SCR Settings
4. On the SCR Settings tab in the RIS section, click Configure...
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 4
CS PS SP
syngo MammoReport
9-2
MedBroker RIS Bridge
5. In the RIS applications window select MedBroker (respectively the name that has
been chosen during the RIS installation) and click Edit.
Fig. 2
RIS MedBroker: Edit
NOTE
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
6. In the Edit RIS Bridge window tick Active and click OK to accept settings.
Key assigned
not applicable with this bridge
Fig. 3
syngo MammoReport
MedBroker: Edit RIS Bridge
SPB7-420.815.51.06.02
12.11
Page 2 of 4
CS PS SP
Siemens
MedBroker RIS Bridge
9-3
7. Back in the RIS applications window select MedBroker again and click Echo:
Fig. 4
MedBroker: Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
MedBroker: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
9. Back in the RIS applications window select MedBroker again and click Configure:
Fig. 6
Siemens
MedBroker: Configure
SPB7-420.815.51.06.02
12.11
Page 3 of 4
CS PS SP
syngo MammoReport
9-4
MedBroker RIS Bridge
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The MedBroker Configuration window opens.
Customize
It might be useful to exclude a prefix
from the received invitation number.
Fig. 7
MedBroker: Bridge Configuration
NOTE
Do not confuse the port number for MedBroker Service (1801)
with port number for syngo MammoReport (4701).
For more technical information about connection requirements,
refer to RIS Connectivity, SPB7-420.850 ...
11. In MedBroker Configuration window leave the settings for Endpoints unchanged (see
screen shot). Enter the IP address for Host and the Port number in the MedBroker
section. Click OK to save the settings and close this window.
12. Back in the RIS applications window (see Fig. 6) click OK to close and Exit SCR
Settings (see Fig. 1).
NOTE
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
Preparing the RIS computer
9
If a problem with the RIS connection occurred, the RIS administrator needs to check the
following settings:
•
RIS computer with MedOutlook and MedBroker Service running is required.
•
Endpoint name must be set in MedOutlook: RIS (as default).
•
MedBroker Service must be configured with open port 1801 to receive the information
from the syngo MammoReport system.
•
The RIS computer must be reachable via TCPI/IP from the syngo MammoReport
system.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 4
CS PS SP
Siemens
Medos RIS Bridge
10
10 - 1
This chapter describes the configuration of the MEDOS RIS Connector for syngo MammoReport VB24 or higher.
General
10
The RIS and the syngo MammoReport are connected bi-directional by the XAppInterface:
•
If a case is opened for reviewing on the high resolution monitors (e.g. by a ViewNow on
a patient, or by a case change in the current work list), certain information is sent over
to the RIS, so that the same case is opened on the RIS.
•
If a case is opened on the RIS computer, certain information (depends on the
functionalities of the RIS) is sent over to the syngo MammoReport, so that the same
case is opened on the syngo MammoReport.
Configuring the syngo MammoReport system
10
The MEDOS RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
Siemens
SCR Settings
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
10 - 2
Medos RIS Bridge
4. On the SCR Settings tab in the RIS section, click Configure...
5. In the RIS applications window, select MEDOS (respectively the name entered during
the RIS installation) and click Edit.
Fig. 2
MEDOS RIS: Select
NOTE
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
6. In the Edit RIS Bridge window check Active and click OK to accept settings.
Key assigned
(useful with bi-directional bridges)
Tick the check boxes (r, o, p) to enable
syngo MammoReport to send the current patient ID to the RIS, when the
appropriate key is pressed.
It is possible to tick more than one key
for this feature.
Fig. 3
syngo MammoReport
MEDOS RIS: Edit
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
Medos RIS Bridge
10 - 3
7. Back in the RIS applications window select MEDOS and click Echo.
Fig. 4
MEDOS RIS: Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
MEDOS RIS: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
9. Back in the RIS applications window select MEDOS again and click Configure.
Fig. 6
Siemens
MEDOS RIS: Configure
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
10 - 4
Medos RIS Bridge
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The MEDOS Bridge Configuration window opens.
Fig. 7
MEDOS: File Polling Bridge Configuration
11. Enter suitable settings in the MEDOS Bridge Configuration window. Therefore also
contact your MEDOS RIS administrator in order to provide a correct configuration:
- MEDOS Server: Required = Name or IP address of the MEDOS Server
- MEDOS Clientname: Required = Name of the MEDOS client that is supposed to be
synchronized
- Type: Optional = Identifier for the calling system (here: syngo MammoReport)
- Command Auto Sync: Optional = Command that is supposed to be executed on
MEDOS system after patient change on syngo MammoReport
- Command Manual Sync: Optional = Command (manually triggered) that is supposed
to be executed on MEDOS system after pressing synchronization key on syngo
MammoReport
- Test connection:
Use command from: Optional = Make selection as demanded
Study Instance UID: Optional = Enter value of test image
- Port: Required = Port used by the program tell MammoReport (shall be different from
the port used by the MEDOS Bridge=4701), use e.g. 4702
12. Click OK to save the entries and to close the configuration window.
13. Click OK to close RIS applications dialog (Fig. 6), and Exit SCR Settings (Fig. 1).
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
Medos RIS Bridge
10 - 5
Checking the MEDOS RIS Connector
10
1. Login to syngo MammoReport as user with Radiologist rights.
2. Open a test case on the RIS computer and check that this patient is opened for
reviewing on the high resolution displays of the syngo MammoReport system.
3. Open a test case on syngo MammoReport system and check that this patient is opened
also on the RIS computer.
4. An RIS error message may appear after log in (or user change) and it will reappear as
long as the connection does not work correctly.
5. Click OK on the error message being displayed.
Siemens
NOTE
Connection error message will reappear upon log in or user
change in case the network connection does not work correctly.
NOTE
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
10 - 6
Medos RIS Bridge
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
Mitra Context Server
11
11 - 1
This chapter describes the configuration of the Mitra Context Server RIS Connector
(Kodak2010, TRIS2000) for syngo MammoReport VB24 or higher.
NOTE
It is strongly recommended NOT to install the Mitra Context
Server bridge on the syngo MammoReport workstation:
This bridge shall be installed on a Virtual Machine.
Use the RIS “Standalone” only as last resort.
General
11
The RIS and the syngo MammoReport are connected bi-directional by the Mitra Context
Server:
•
If a case is opened for reviewing on the high resolution monitors (e.g. by a ViewNow on
a patient, or by a case change in the current work list), certain information is sent over
to the RIS, so that the same case is opened on the RIS.
•
If a case is opened on the RIS computer, certain information (depends on
functionalities of the RIS) is sent over to the syngo MammoReport, so that the same
case is opened on the syngo MammoReport.
Prerequisites
11
Network
11
The hospital network administrator has to be contacted to check that the network communication is opened for the required data exchange.
RIS Computer
11
The setup for the RIS computer should be carried out by the RIS administrator:
Siemens
•
The RIS computer must be reachable via TCP/IP from the syngo MammoReport
system.
•
The Mitra Context Server must be installed and configured on the RIS computer. The
Mitra Context Server service must be started.
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
11 - 2
Mitra Context Server
Configuring the syngo MammoReport system
11
The Mitra Context Server RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
SCR Settings
4. On the SCR Settings tab in the RIS section, click Configure...
NOTE
syngo MammoReport
As the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first.
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
Mitra Context Server
11 - 3
5. In the RIS applications window, click Add.
Fig. 2
Mitra Context Server RIS: Select
6. In the Add RIS Bridge window enter the required values: Name (as used when the
bridge was installed), Host (IP address of RIS computer) and Port (4701).
7. Check Active and click OK to accept settings.
Key assigned
(useful with bi-directional bridges)
Tick the check boxes (r, o, p) to enable
syngo MammoReport to send the current patient ID to the RIS, when the
appropriate key is pressed.
It is possible to tick more than one key
for this feature.
Fig. 3
Siemens
Mitra Context Server: Add
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
11 - 4
Mitra Context Server
8. Back in the RIS applications window select Mitra Context Server and click Echo.
Fig. 4
Mitra Context Server RIS: Echo
9. The following message shall appear. Click OK to confirm.
Fig. 5
Mitra Context Server RIS: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
10. Back in the RIS applications window select Mitra Context Server again and click
Configure.
Fig. 6
syngo MammoReport
Mitra Context Server RIS: Configure
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
Mitra Context Server
NOTE
11 - 5
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
Configuring the RIS Bridge
11
1. Change to the RIS computer (Virtual Machine or RIS “Standalone”) and select the open
MitraContextServer Bridge Configuration:
Fig. 7
Mitra Context Server: Bridge configuration
2. Select the Predefined Configuration, i.e. TRIS2000, Kodak2010 or HSS CRIS.
NOTE
Siemens
As the Mitra Context Server bridge is installed on the RIS computer where Mitra Context Server is running, no entries for Host,
User and Password are necessary.
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
11 - 6
Mitra Context Server
Checking the Mitra Context Server RIS Connector
11
This RIS Connector can be used in both directions, i.e. open a patient and add notifications, which are directly sent to the RIS, or initiate a review from RIS, that a message window on syngo MammoReport is opened and informs the user to review this patient.
1. Login to syngo MammoReport as user with Radiologist rights.
2. Open a test case with appropriate patient ID and accession number on the RIS
computer and check that this patient is opened for reviewing on the high resolution
displays of the syngo MammoReport system.
3. Open a test case with appropriate patient ID and accession number on syngo
MammoReport system and check that this patient is opened also on the RIS computer.
4. An RIS error message may appear after log in (or user change) and it will reappear as
long as the connection does not work correctly
5. Click OK on the error message being displayed.
syngo MammoReport
NOTE
Connection error message will reappear upon log in or user
change in case the network connection does not work correctly.
NOTE
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
PACSBridge RIS Connector
12
12 - 1
This chapter describes the configuration of the PACSBridge RIS Connector for syngo
MammoReport VB24 or higher.
General
12
The RIS and the syngo MammoReport are connected uni-directional, i.e. information
(accession number) is sent from syngo MammoReport to the RIS, but no information
according to patients is sent from the RIS to syngo MammoReport.
If a case is opened for reviewing on the high resolution displays (e.g. by a ViewNow on a
patient, or by a case change in the current work list), certain information is sent over to the
RIS, so that the same case is opened on the RIS.
Configuring the syngo MammoReport system
12
The PACSBridge RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
SCR Settings
4. On the SCR Settings tab in the RIS section, click Configure...
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 8
CS PS SP
syngo MammoReport
12 - 2
PACSBridge RIS Connector
5. In the RIS applications window select PACSBridge (respectively the name entered
during the RIS installation) and click Edit.
Fig. 2
RIS PACSBridge: Edit
NOTE
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
6. In the Edit RIS Bridge window tick Active and click OK to accept settings.
Key assigned
not applicable with this bridge
Fig. 3
syngo MammoReport
PACSBridge: Edit RIS Bridge
SPB7-420.815.51.06.02
12.11
Page 2 of 8
CS PS SP
Siemens
PACSBridge RIS Connector
12 - 3
7. Back in the RIS applications window select PACSBridge again and click Echo:
Fig. 4
PACSBridge: Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
PACSBridge: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
9. Back in the RIS applications window select PACSBridge again and click Configure:
Fig. 6
Siemens
PACSBridge: Configure
SPB7-420.815.51.06.02
12.11
Page 3 of 8
CS PS SP
syngo MammoReport
12 - 4
PACSBridge RIS Connector
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The PACSBridge Configuration window opens.
Fig. 7
PACSBridge Configuration
11. Make the following settings in the PACSBridge Configuration window:
Host:
IP address of the computer where the PACSBridge RIS application is running on.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 8
CS PS SP
Siemens
PACSBridge RIS Connector
12 - 5
Port:
Enter the same port as used by PACSBridge RIS application.
Fig. 8
Siemens
PACSBridge Configuration: Host and Port
NOTE
Do not confuse the port number for PACSBridge RIS application
(5555) with port number for syngo MammoReport (4701).
For more technical information about connection requirements,
refer to RIS Connectivity, SPB7-420.850 ...
NOTE
Delete old entries from the Format text box before inserting a new
one.
SPB7-420.815.51.06.02
12.11
Page 5 of 8
CS PS SP
syngo MammoReport
12 - 6
PACSBridge RIS Connector
12. Click Insert... to configure the Format by selecting items from the list as requested:
Fig. 9
PACSBridge Configuration
13. Make sure that check box Ping before send remains ticked, and leave the default
value for Timeout (ms) as it is.
14. In the Test Connection tile enter arbitrary test data for the items, which are configured
as Format in the Settings section.
15. Make sure, that the RIS application is already running and press Send.
The box right from Send button gives the result of the send command:
Fig. 10 PACSBridge Configuration: Send test data
16. After successful test result, click OK to close the Configuration window.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 8
CS PS SP
Siemens
PACSBridge RIS Connector
12 - 7
17. Back in the RIS applications window (see Fig. 6) click OK to close and Exit SCR
Settings (see Fig. 1).
NOTE
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
Preparing the RIS computer
12
If a problem with the RIS connection occurred, the RIS administrator needs to check that
the following services are running on the RIS computer:
•
AsiRecog Server
•
HSG-FileOutputClient: PowerScribe
•
HSG-HL7 Client: PowerScribe
•
HSG-HL7 Server: PowerScribe
The RIS computer must be reachable via TCP/IP from the syngo MammoReport system.
Siemens
SPB7-420.815.51.06.02
12.11
Page 7 of 8
CS PS SP
syngo MammoReport
12 - 8
PACSBridge RIS Connector
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 8 of 8
CS PS SP
Siemens
PenRad XML RIS Bridge
13
13 - 1
This chapter describes the configuration of the PenRad XML RIS Connector for syngo
MammoReport VB24 or higher.
General
13
The RIS and the syngo MammoReport are connected bi-directional by the XAppInterface:
•
If a case is opened for reviewing on the high resolution monitors (e.g. by a ViewNow on
a patient, or by a case change in the current work list), certain information is sent over
to the RIS, so that the same case is opened on the RIS.
•
If a case is opened on the RIS computer, certain information (depends on
functionalities of the RIS) is sent over to the syngo MammoReport, so that the same
case is opened on the syngo MammoReport.
Configuring the syngo MammoReport system
13
The PenRad XML RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
Siemens
SCR Settings
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
13 - 2
PenRad XML RIS Bridge
4. On the SCR Settings tab in the RIS section, click Configure...
5. In the RIS applications window select PenRadXML (respectively the name entered
during the RIS bridge installation).
Fig. 2
PenRad XML: Edit
NOTE
6.
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
Click Edit in the Edit RIS Bridge window tick Active and click OK to accept settings.
Key assigned
(useful with bi-directional bridges)
Tick the check boxes (r, o, p) to enable
syngo MammoReport to send the current patient ID to the RIS, when the
appropriate key is pressed.
It is possible to tick more than one key
for this feature.
Fig. 3
syngo MammoReport
PenRad XML: Edit RIS Bridge
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
PenRad XML RIS Bridge
13 - 3
7. Back in the RIS applications window select PenRadXML again and click Echo:
Fig. 4
PenRad XML: Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
PenRad XML RIS: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
9. Back in the RIS applications window select PenRadXML again and click Configure.
Fig. 6
Siemens
PenRad XML RIS: Configure
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
13 - 4
PenRad XML RIS Bridge
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The XAppBridge Configuration window opens.
Fig. 7
PenRad XML: File Polling Bridge Configuration (left) or Epic Radiant (right)
11. Make the following settings in the XAppBridge Configuration for PenRad XML resp.
Epic.
12. Select the RIS from the combo box
13. In Writing files ⇒ Destination ⇒ Filename enter PenRad.xml resp. response.xml.
14. In Reading files ⇒ Source ⇒ Filename enter PenRad.xml.
For Epic the filename is automatically filled out.
15. Click OK to save the entries and to close the Configuration window.
16. Click OK to close RIS applications dialog (Fig. 6), and Exit SCR Settings (Fig. 1).
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
PenRad XML RIS Bridge
13 - 5
Configuring the RIS Computer
13
The RIS application needs to be enabled to write the polling files into the specified folder
on the syngo MammoReport workstation:
- C:\MBCXAPP\Bridges\PenRadXML0\FileTransferIn
- C:\MBCXAPP\Bridges\PenRadXML0\FileTransferOut
•
These folder on the syngo MammoReport workstation are already prepared for the file
sharing.
•
These folder must be connected from the RIS computer by using the Windows Explorer
function mapping network drive. Therefore use the following paths:
- \\<IP address>\c$\MBCXAPP\Bridges\PenRadXML0\FileTransferIn
- \\<IP address>\c$\MBCXAPP\Bridges\PenRadXML0\FileTransferOut
(<IP address> = IP address of the syngo MammoReport workstation)
Checking the PenRad XML RIS Connector
13
1. Login to syngo MammoReport as user with Radiologist rights.
2. Open a test case with appropriate patient ID and accession number on the RIS
computer and check that this patient is opened for reviewing on the high resolution
displays of the syngo MammoReport system.
3. Open a test case with appropriate patient ID and accession number on syngo
MammoReport system and check that this patient is opened also on the RIS computer.
4. An RIS error message may appear after log in (or user change) and it will reappear as
long as the connection does not work correctly.
5. Click OK on the error message being displayed.
NOTE
Connection error message will reappear upon log in or user
change in case the network connection does not work correctly.
NOTE
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
Troubleshoot File Sharing
13
In case of sharing problems refer to General File Sharing on page 18 - 1.
Siemens
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
13 - 6
PenRad XML RIS Bridge
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
PenRad Serial RIS Bridge
14
14 - 1
This chapter describes the configuration of the PenRad Serial RIS Connector for syngo
MammoReport VB24 or higher.
General
14
The RIS and the syngo MammoReport are connected uni-directional, i.e. information
(accession number) is sent from syngo MammoReport to the RIS, but no information
according to patients is sent from the RIS to syngo MammoReport.
If a case is opened for reviewing on the high resolution displays (e.g. by a ViewNow on a
patient, or by a case change in the current work list), certain information is sent over to the
RIS, so that the same case is opened on the RIS.
Configuring the syngo MammoReport system
14
The MedBroker RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
SCR Settings
4. On the SCR Settings tab in the RIS section, click Configure...
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 4
CS PS SP
syngo MammoReport
14 - 2
PenRad Serial RIS Bridge
5. In the RIS applications window select Penrad (respectively the name entered during
the RIS installation) and click Edit.
Fig. 2
RIS PenRad Serial: Edit
NOTE
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
6. In the Edit RIS Bridge window tick Active and click OK to accept settings.
Key assigned
not applicable with this bridge
Fig. 3
syngo MammoReport
PenRad Serial: Edit RIS Bridge
SPB7-420.815.51.06.02
12.11
Page 2 of 4
CS PS SP
Siemens
PenRad Serial RIS Bridge
14 - 3
7. Back in the RIS applications window select Penrad again and click Echo:
Fig. 4
PenRad Serial: Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
PenRad Serial: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
9. Back in the RIS applications window select Penrad again and click Configure:
Fig. 6
Siemens
PenRad Serial: Configure
SPB7-420.815.51.06.02
12.11
Page 3 of 4
CS PS SP
syngo MammoReport
14 - 4
PenRad Serial RIS Bridge
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The Serial Configuration window opens.
Fig. 7
PenRad Serial: Bridge Configuration
11. Make the following settings in the Serial Configuration window: Select COM1 as Port
(if this is used for the null modem cable). Select 4800 Baud. Enter 8 data bits and 1 stop
bit. Make the other settings as required and close this window by clicking OK.
12. Back in the RIS applications window (see Fig. 6) click OK to close and Exit SCR
Settings (see Fig. 1).
NOTE
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
Preparing the RIS computer
14
If a problem with the RIS connection occurred, the RIS administrator needs to check the
following settings:
•
The RIS computer must be configured to receive the information from the syngo
MammoReport system. The default settings for data connection should be:
- Baud: 4800
- Non-parity
- Data bits: 8
- Stop bits: 1
- Flow Control: None
•
The RIS computer must be connected via COM port and null modem cable to the
syngo MammoReport system.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 4
CS PS SP
Siemens
Process Starter
15
15 - 1
This chapter describes the configuration of the Process Starter RIS Connector for syngo
MammoReport VB24 or higher.
General
15
The RIS and the syngo MammoReport are connected unidirectional, i.e. information
(accession number) is sent from syngo MammoReport to the RIS, but no information
according to patients is sent from the RIS to syngo MammoReport.
If a case is opened for reviewing on the high resolution displays (e.g. by a ViewNow on a
patient, or by a case change in the current work list), certain information is sent over to the
RIS, so that the same case is opened on the RIS.
Configuring the syngo MammoReport system
15
The Process Starter RIS Connector must be configured in the SCR Settings:
1. Login to syngo MammoReport as mammoadmin (the user needs
MammoAdministration rights).
2. Select Main Menu ⇒ Options ⇒ Configuration...
3. In syngo Configuration Panel double-click SCR Settings:
Fig. 1
SCR Settings
4. On the SCR Settings tab in the RIS section, click Configure...
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
15 - 2
Process Starter
5. In the RIS applications window select Process Starter (respectively the name entered
during the RIS installation) and click Edit.
Fig. 2
RIS Process Starter: Edit
NOTE
If the RIS bridge is NOT installed on the syngo MammoReport
workstation, it must be added first. Click Add and enter the
required values: Name (as used when the bridge was installed),
Host (IP address of RIS computer) and Port (4701).
6. In the Edit RIS Bridge window tick Active and click OK to accept settings.
Key assigned
not applicable with this bridge
Fig. 3
syngo MammoReport
Process Starter: Edit RIS Bridge
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
Process Starter
15 - 3
7. Back in the RIS applications window select Process Starter again and click Echo:
Fig. 4
Process Starter Echo
8. The following message shall appear. Click OK to confirm.
Fig. 5
Process Starter: Ping
NOTE
If Ping command results in error, check all RIS bridge settings
carefully and then retry.
9. Back in the RIS applications window select Process Starter again and click
Configure:
Fig. 6
Siemens
Process Starter: Configure
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
15 - 4
Process Starter
NOTE
The configuration dialog is part of the bridge application.
Therefore this dialog will be opened on the computer where the
bridge is installed and running, but it's essential that the network
connection from bridge to syngo MammoReport is established
successfully.
10. The Serial Configuration window opens.
Fig. 7
Process Starter: Bridge Configuration
Make the following settings:
11. Process: enter the path or browse to the path of the application/process, that shall be
started by the Process Starter bridge. (For example you may use the Internet Explorer:
C:\Program Files\Internet Explorer\iexplore.exe.)
12. Command Line Parameter Configuration String: click Insert to add one or more of
the values given in the context menu:
- PatientID
- PatientName
- StudyInstanceUID
- StudyID
- AccessionNumber
- Percent Sign% (Converter for special character in an URL string. To be used only
together with URL Parameter marked active. Therefore refer to next step.)
13. URL Parameter: to be checked, if the application/process is a browser.
14. Close process method: choose a value to control the application/process when a
case is closed on the HR displays. Values are:
- Do not close
- Close windows
- Close windows and terminate
- Terminate
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
Process Starter
15 - 5
You may check the Process Starter connection by entering Test data in the Test connection section and clicking Send.
NOTE
The access to the application/process that shall be triggered must
be granted.
15. Click OK for saving. The bridge configuration window is closed.
16. Back in the RIS applications window (see Fig. 6) click OK to close and Exit SCR
Settings (see Fig. 1).
NOTE
Siemens
For backup of the RIS configuration settings, refer to Appendix B:
Backup of RIS Configuration Settings on page 17 - 2.
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
15 - 6
Process Starter
This page intentionally left blank.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
Appendix A
16
16 - 1
Find a Workable RIS Configuration
16
Discuss a possible solution with the RIS administrator in the hospital (see more details in
the “Functional Description RIS Connectivity” document SPB7-420.850 ...) based on the
reusable bridges (see table below).
The syngo MammoReport RIS Bridge CD contains various bridges, which can be reused.
The following matrix presents the supported features of the most important reusable RIS
bridges:
Bridge /
Protocol
ASCII File Polling
(File)
HTTP Server
(HTTP request)
PenRad XML
(File)
Mitra Context Server (COM,
License)
Process Starter
(Process invocation)
Tab. 1
Login
Logout
Open Patient Open Patient
sMR -> RIS
sMR <- RIS
X
X
X
X
X
X
X
X
X
Reusable bridges
If you are able to establish a proper RIS communication, re-using one of possibilities
above, please inform Service via the established service processes (RSC -> HSC).
HSC can provide you an appropriate Integration test form to validate the interface (allows
you to official document this solution).
•
Please provide basic data -> see Basic Data Form.
•
Perform the test at site and send the test document back.
If no appropriate bridge can be used in your situation, please also inform Service via the
established service processes (RSC -> HSC) to get further support.
•
Please provide also basic data -> see Basic Data Form.
New developments may then be required, either at the RIS Vendor and/or Siemens side.
Extended resources are needed as well as close contact between both vendors (a fast
solution can not be expected).
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 2
CS PS SP
syngo MammoReport
16 - 2
Appendix A
Basic Data Form
syngo MammoReport
Serial number
Software version
RIS license available?
Name of Siemens service contact
RIS Information
RIS vendor
RIS product name and version
Supported Interfaces (HTTP Request, File Polling,
command line ...)
Name of RIS vendor's service contact
Workflow
Please describe the desired workflow briefly
What products are involved (provide a scheme that
shows the relationships and data transfer)
Additional comments
Specific Integration Information
Is single login on syngo MammoReport required?
Is Reporting additionally to RIS input on MammoReport required, e.g. enter BIRADS?
Where does the customer select the Patient: RIS or
MammoReport?
Where is the RIS Application running (separate PC,
VMware on syngo MammoReport)?
Which information can be used to identify the study
to be opened, e.g. Study Instance UID, patient ID,
accession number?
Does the site support unique Patient IDs and/or
Accession numbers?
Can a RIS bridge be reused? If so which one?
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 2 of 2
CS PS SP
Siemens
Appendix B
17
17 - 1
Importing the License File
17
1. Login to syngo as administrator.
2. Insert the License CD into drive D:\.
3. Open via Main Menu ⇒ Options ⇒ Service ⇒ Local Service the Configuration
service function.
4. From the System Options select Service ⇒ Licensing.
5. Click the Select New License file link, then click Browse.
6. Browse to the License CD and select the license file (*.dat).
7. Click Open to load the file into the path display.
8. Click Save on the action bar and wait until saving is completed.
9. Eject the License CD from drive D:\
10. Click Home on the Main Menu and wait until the report creation is finished.
11. On the appearing message click OK to reboot the computer.
12. After restart, login to syngo as administrator.
Disabling the Automatic Start of Virtual Machine
17
The Virtual Machine is always started together with the syngo MammoReport. This
behavior is defined in SCR_System.pm to be found under C:\MBCSCR\syngo\config.
If the customer does not want to start the Virtual Machine together with the syngo MammoReport, the according entry in SCR_System.pm must be commented out:
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows.
5. Open the file C:\MBCSCR\syngo\config\SCR_System.pm by double-clicking.
6. Scroll down to the end of the file.
7. Comment out the last line by adding # in front:
# include %LIGHTBOX%\syngo\config\dependent\VMWatchDog.pm
8. Save the changes and close the file.
9. Restart the computer and check that the Virtual Machine is not started together with the
syngo MammoReport.
NOTE
Siemens
If the customer wants to re-enable the VMware to be started
together with the syngo MammoReport, uncomment the according entry by remove the #.
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
17 - 2
Appendix B
Backup of RIS Configuration Settings
17
Backup with RIS bridge on syngo MammoReport Workstation
17
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows.
5. In the Explorer change to drive C:\.
6. Create a backup folder, e.g. C:\Backup_RIS.
7. Copy the file C:\MBCSCR\XAppSettings.xml to the backup folder.
8. Copy the folder C:\MBCXAPP into the backup folder.
9. Save the backup folder, e.g. as Backup_RIS_<SW version>_<Computer Name> to an
USB stick or another computer.
10. For archiving burn the folder on a CD.
11. Make sure that the backup files are readable.
For restoring, refer to Restoring with RIS bridge on syngo MammoReport Workstation on
page 17 - 4.
Backup with RIS bridge on Virtual Machine
17
This backup consists of two steps:
Saving XAppSettings.xml from syngo MammoReport Workstation
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows.
5. In the Explorer change to drive C:\.
6. Create a backup folder, e.g. C:\Backup_RISfromMammoReport.
7. Copy the file C:\MBCSCR\XAppSettings.xml to the backup folder.
8. Save the backup folder to an USB stick or another computer.
Saving the RIS Settings from Virtual Machine
1. Start the Virtual Machine by double-clicking
C:\VMRIS\Windows XP Professional.vmx.
2. Log in to VMware as administrator.
3. In the Explorer change to drive C:\.
4. Create a backup folder, e.g. C:\Backup_RISfromVM.
5. Copy the folder C:\MevisBreastCare into the backup folder.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
Appendix B
17 - 3
6. Change to C:\Documents and Settings\All Users\Startmenu\Programs\Startup.
7. Copy (all) batch file(s) *.bat (*= Name of RIS bridge) into the backup folder.
8. Save the backup folder to an USB stick or another computer.
9. For archiving burn both backup folder on a CD.
10. Make sure that the backup files are readable.
For restoring, refer to Restoring with RIS bridge on Virtual Machine on page 17 - 4.
Backup with RIS bridge on RIS “Standalone”
17
This backup consists of two steps.
Saving XAppSettings.xml from syngo MammoReport Workstation
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows.
5. In the Explorer change to drive C:\.
6. Create a backup folder, e.g. C:\Backup_RISfromMammoReport.
7. Copy the file C:\MBCSCR\XAppSettings.xml to the backup folder.
8. Save the backup folder to an USB stick or another computer.
Saving the RIS Settings from RIS “Standalone”
1. Log in to RIS computer as administrator.
2. In the Explorer change to drive C:\.
3. Create a backup folder, e.g. C:\Backup_RISfromStandalone.
4. Copy the folder C:\Program Files\MevisBreastCare into the backup folder.
5. Change to C:\Documents and Settings\All Users\Startmenu\Programs\Startup.
6. Copy (all) batch file(s) *.bat (*= Name of RIS bridge) into the backup folder.
7. Save the backup folder to an USB stick or another computer.
8. For archiving burn both backup folder on a CD.
9. Make sure that the backup files are readable.
For restoring, refer to Restoring with RIS bridge RIS “Standalone” on page 17 - 5.
Siemens
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
17 - 4
Appendix B
Restoring the RIS Configuration Settings
17
Restoring with RIS bridge on syngo MammoReport Workstation
17
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows and open the Explorer.
5. Load the backup media (insert CD or connect USB stick).
6. Copy the backup folder from CD or USB stick to C:\.
7. From the backup folder on C:\, copy the file XAppSettings.xml to C:\MBCSCR.
8. Confirm the overwriting, if necessary.
9. From the backup folder on C:\, copy the folder MBCXAPP to C:\.
10. Confirm the overwriting, if necessary.
11. Insert the syngo MammoReport RIS Bridge CD into drive D:\
12. Start the D:\Installer.exe and wait until the application is loaded.
13. Select the bridge in the RIS Bridge Installed section and click Update (refer to Fig. 3 on
page 3 - 4.) Repeat this for every installed bridge.
14. Restart the computer.
15. Check the RIS connection according the description given in the configuration chapter.
Restoring with RIS bridge on Virtual Machine
17
Restoring XAppSettings.xml to syngo MammoReport Workstation
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows and open the Explorer.
5. Load the backup media (insert CD or connect USB stick).
6. Copy the MammoReport backup folder from CD or USB stick to C:\.
7. From the backup folder on C:\, copy the file XAppSettings.xml to C:\MBCSCR.
Restoring the RIS Settings from Virtual Machine
1. Start the Virtual Machine by double-clicking
C:\VMRIS\Windows XP Professional.vmx.
2. Log in to VMware as administrator.
3. Load the backup media (insert CD or connect USB stick).
4. In the Explorer change to drive C:\.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
Appendix B
17 - 5
5. Copy the VM backup folder from CD or USB stick to C:\.
6. From the backup folder on C:\, copy the folder MevisBreastCare to C:.
7. Confirm the overwriting, if necessary.
8. From the backup folder on C:\, copy (all) batch file(s) *.bat (*= Name of RIS bridge) to
C:\Documents and Settings\All Users\Startmenu\Programs\Startup.
9. Confirm the overwriting, if necessary.
10. Insert the syngo MammoReport RIS Bridge CD into drive D:\
11. Start the D:\Installer.exe and wait until the application is loaded.
12. Select the bridge in the RIS Bridge Installed section and click Update (refer to Fig. 3 on
page 3 - 4.) Repeat this for every installed bridge.
13. Log off Administrator from VMware and login as RISUser to VMware.
14. Restart the syngo MammoReport computer.
15. Check the RIS connection according the description given in the configuration chapter.
Restoring with RIS bridge RIS “Standalone”
17
Restoring XAppSettings.xml to syngo MammoReport Workstation
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows and open the Explorer.
5. Load the backup media (insert CD or connect USB stick).
6. Copy the MammoReport backup folder from CD or USB stick to C:\.
7. Rename C:\MBCSCR\XAppSettings.xml to XAppSettings.backup
8. From the backup folder on C:\, copy the file XAppSettings.xml to C:\MBCSCR.
9. Restart the computer.
Restoring the RIS Settings to RIS “Standalone”
1. If not already done: log in to the RIS computer as administrator.
2. Load the backup media (insert CD or connect USB stick).
3. Copy the Standalone backup folder from CD or USB stick to C:\.
4. From the backup folder on C:\, copy the folder MevisBreastCare to
C:\Program Files\.
5. Confirm the overwriting, if necessary.
6. From the backup folder on C:\, copy (all) the batch file(s) *.bat (*= Name of RIS bridge)
to C:\Documents and Settings\All Users\Startmenu\Programs\Startup.
7. Confirm the overwriting, if necessary.
8. Insert the syngo MammoReport RIS Bridge CD.
Siemens
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
17 - 6
Appendix B
9. Start the installation by double-clicking on Installer.exe.
The bridge installer for the Path Settings opens.
10. Click OK. The path settings are executed and the configuration window is opened.
11. Select the bridge in the RIS Bridge Installed section and click Update (refer to Fig. 3 on
page 3 - 4.) Repeat this for every installed bridge.
12. To start the RIS bridge: Log off administrator and log in again as a RIS user.
13. Check the RIS connection according the description given in the configuration chapter.
Backup of Virtual Machine Settings
17
The Virtual Machine is located in C:\VMRIS with a size of about 3.5 GB. For a backup,
copy the complete folder to a USB drive and burn the folder on a DVD for archiving:
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows.
5. Connect a USB drive to the computer.
6. Copy the folder C:\VMRIS to the USB drive or another computer.
This may take about 10 minutes.
7. For archiving burn the backup folder on a DVD.
8. Make sure that the backup files are readable.
Restoring the Virtual Machine Settings
NOTE
17
Before restoring a Virtual Machine by copying the VMRIS folder,
make sure that the VM workstation is installed
- Check that C:\Program Files\VMware\VMware Workstation\
vmplayer.exe is available
- If not, you need to do a complete installation according RIS on
Virtual Machine on page 4 - 1.
1. Restart Application via Main Menu Options ⇒ End Session.
2. Click Yes to confirm the application restart on the appearing message.
3. When the syngo screen saver appears, press and hold the left shift key until the
Windows login appears.
4. Log in as user Administrator in Windows.
5. Insert VM backup DVD to DVD drive.Connect a USB drive to the computer.
6. From the DVD, copy the folder VMRIS to C:\. This may take about 10 minutes.
Make sure that the auto start of the Virtual Machine is not disabled. Therefore refer to
Disabling the Automatic Start of Virtual Machine on page 17 - 1.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
Trouble Shooting
18
18 - 1
General File Sharing
This chapter is only valid for the File Polling bridge and the PenRad XML RIS connector
for syngo MammoReport VB24 or higher.
It describes what has to be checked in case of sharing problems.
Check the File Sharing Settings
18
•
Check read and write permission of the share.
•
Sometimes you have to add a user if the file share is running on the RIS PC:
- Add local account with user meduser and password.
•
Otherwise please contact IT department (e.g. in case of different domain name,
different subnet etc.).
Installation of RIS bridge failed
18
Possible cause
Solution
Installation directory was not correct.
Refer to chapter Uninstalling a RIS Bridge on
page 3 - 5 to get information how to uninstall a
RIS bridge.
Patient synchronization MammoReport to RIS
Find out where the problem is:
•
XAppInterface
•
XAppInterface – Bridge communication
•
Bridge
•
Bridge – RIS communication
•
RIS
Troubleshooting General synchronization
Problem
Solution
Find out if bridge receives request from
MammoReport.
• Check log file of bridge
- Check if date/time of log file changes when
you open another patient
• Bridges on MammoReport:
- see C:\syngo\Log\MBCLog\Bridge***.log
for log file
• Bridges on other system:
- see %MBCXAPP%\Log\Bridge***.log for
logfile
- %MBCXAPP% is „C:/Program Files/
MeVisBreastCare/XApp“ by default
Siemens
SPB7-420.815.51.06.02
12.11
Page 1 of 6
CS PS SP
syngo MammoReport
18 - 2
Trouble Shooting
Troubleshooting XAppInterface
Problem
Solution
If bridge does not receive request
• Check log file of XAppInterface
- C:\syngo\Log\MBCLog\XAppInterface.log
If log file indicates this as the cause
Check MBC-RIS license
Check User Preferences and User Rights
Check XAppInterface settings (in SCR
Settings)
• XAppInterface settings include
- MammoReport Logical Name and Port (not
accessible from GUI)
- Logical Name, Host and Port of each bridge
- Active flag for each bridge
- Additional settings like time-outs, etc. (not
accessible from GUI)
• Check communication MammoReport to
Bridge
- Open XAppInterface settings (in SCR
Settings)
- Send Echo to bridge
• See also C:\MBCSCR\XAppSettings.xml
XAppSettings.xml
<XAppSettings>
<MBCCommSetting>
<Peer>
<Name Type="String">SCR</Name>
<Port Type="Int">4700</Port>
</Peer>
<LogLevel Type="String">Trace</LogLevel>
<TimeOutConnect Type="Int">10000</TimeOutConnect>
</MBCCommSetting>
<TimeOutResponse Type="UInt">10</TimeOutResponse>
<XAppSetting>
<XAppPeer>
<Name Type="String">ScreeningRIS</Name>
<Host Type="String">192.168.0.2</Host>
<Port Type="Int">4701</Port>
</XAppPeer>
</XAppSetting>
</XAppSettings>
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 2 of 6
CS PS SP
Siemens
Trouble Shooting
18 - 3
Troubleshooting Bridge
Problem
Solution
If bridge receives request from MammoReport
Check log file of bridge
Check configuration of bridge
Bridges on MammoReport:
• Open SCR Settings/RIS Configuration
• Select Bridge and press „Configure“
• See also
C:\MBCXApp\Bridges\Bridge***\*.ini for
configuration
Bridges on other system:
• Select from System Tray menu of bridge
„Configure“
• See also
%MBCXAPP%\Bridges\Bridge***\*.ini
for configuration
• %MBCXAPP% is „C:/Program Files/
MeVisBreastCare/XApp“ by default
Some bridges allow to test the connectivity
to the RIS from their configuration dialog
Siemens
SPB7-420.815.51.06.02
12.11
Page 3 of 6
CS PS SP
syngo MammoReport
18 - 4
Trouble Shooting
Patient synchronization RIS to MammoReport
Troubleshooting
Find out where the problem is
•
RIS
•
RIS - Bridge communication
•
Bridge
•
Bridge - XAppInterface communication
•
XAppInterface
Troubleshooting General synchronization
Problem
Solution
Find out if Bridge receives request from
RIS
• Check log file of bridge
- Check if date/time of log file changes when
you open another patient
• Bridges on MammoReport:
- see C:\syngo\Log\MBCLog\Bridge***.log
for log file
• Bridges on other system:
- see %MBCXAPP%\Log\Bridge***.log for
log file
- %MBCXAPP% is „C:/Program Files/
MeVisBreastCare/XApp“ by default
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 4 of 6
CS PS SP
Siemens
Trouble Shooting
18 - 5
Troubleshooting Bridge
Problem
Solution
If bridge does not receive request from
MammoReport
Check log file of bridge
Check configuration of bridge
Bridges on MammoReport:
• Open SCR Settings/RIS Configuration
• Select Bridge and press „Configure“
• See also
C:\MBCXApp\Bridges\Bridge***\*.ini for
configuration
Bridges on other system:
• Select from System Tray menu of bridge
„Configure“
• See also
%MBCXAPP%\Bridges\Bridge***\*.ini
for configuration
• %MBCXAPP% is „C:/Program Files/
MeVisBreastCare/XApp“ by default
Simulate RIS to send request to bridge,
e.g. by manually placing a synchronization
file into the bridges input folder
Contact RIS service for RIS settings, local
admin for network connectivity, etc.
Siemens
SPB7-420.815.51.06.02
12.11
Page 5 of 6
CS PS SP
syngo MammoReport
18 - 6
Trouble Shooting
Troubleshooting XAppInterface
Problem
Solution
If bridge received request
• Check if XAppInterface receives request
from bridge
- C:\syngo\Log\MBCLog\XAppInterface.log
If log file indicates this as the cause
Check MBC-RIS license
Check User Preferences and User Rights
Check XAppInterface settings (in SCR
Settings)
• XAppInterface settings include
- MammoReport Logical Name and Port (not
accessible from GUI)
- Logical Name, Host and Port of each bridge
- Active flag for each bridge
- Additional settings like time-outs, etc. (not
accessible from GUI)
• Check communication MammoReport to
Bridge
- Open XAppInterface settings (in SCR
Settings)
- Send Echo to bridge
• See also C:\MBCSCR\XAppSettings.xml
Connectivity problems with bridge in VM
Problem
Solution
If Echo is not successful.
Re-configure the setting as follows.
The problem is caused by the Realtek
Network Adapter driver.
• Login as Windows administrator:
The on board checksum calculation
(default setting) fails under some circumstances.
- Open <Control Panel><Administrative
Tools><Computer Management><Device
Manager><Network Adapters>
• Select “RealTek PCIe GBE ...“, Submenu
entry “Properties”
- A new dialog window displays.
• Select “Rx enabled” from the list and
confirm the new settings with “OK”.
- The Echo shall work from now on.
• Reboot computer to ensure that all
components have registered the new
network settings.
syngo MammoReport
SPB7-420.815.51.06.02
12.11
Page 6 of 6
CS PS SP
Siemens
Changes to previous version
Siemens
Page
Changes
all
text corrections
16-1 ff
added Appendix A
18-1 ff
added Trouble Shooting
SPB7-420.815.51
Rev. 06
12.11
19
Page 1 of 2
CS PS SP
19 - 1
syngo MammoReport
19 - 2
Changes to previous version
This page intentionally left blank.
syngo MammoReport
RIS Configuration
SPB7-420.815.51
Rev. 06
12.11
Page 2 of 2
CS PS SP
Siemens