DMS-Prefiltered Mass Spectrometry for the Detection of Biomarkers

Transcription

DMS-Prefiltered Mass Spectrometry for the Detection of Biomarkers
DMS-Prefiltered Mass Spectrometry for the Detection of Biomarkers
Stephen L. Coy, Evgeny V. Krylov, Erkinjon G. Nazarov
Sionex Corporation, 8-A Preston Ct., Bedford, MA 01730
ABSTRACT
Technologies based on Differential Mobility Spectrometry (DMS) are ideally matched to rapid, sensitive, and selective
detection of chemicals like biomarkers. Biomarkers linked to exposure to radiation, exposure to CWA’s, exposure to
toxic materials (TICs and TIMs) and to specific diseases are being examined in a number of laboratories. Screening for
these types of exposure can be improved in accuracy and greatly speeded up by using DMS-MS instead of slower
techniques like LC-MS and GC-MS. We have performed an extensive series of tests with nanospray-DMS-mass
spectroscopy and standalone nanospray-DMS obtaining extensive information on chemistry and detectivity. DMS-MS
systems implemented with low-resolution, low-cost, portable mass-spectrometry systems are very promising. Lowresolution mass spectrometry alone would be inadequate for the task, but with DMS pre-filtration to suppress
interferences, can be quite effective, even for quantitative measurement.
Bio-fluids and digests are well suited to ionization by electrospray and detection by mass-spectrometry, but signals from
critical markers are overwhelmed by chemical noise from unrelated species, making essential quantitative analysis
impossible. Sionex and collaborators have presented data using DMS to suppress chemical noise, allowing detection of
cancer biomarkers in 10,000-fold excess of normal products1,2. In addition, a linear dynamic range of approximately
2,000 has been demonstrated with accurate quantitation3. We will review the range of possible applications and present
new data on DMS-MS biomarker detection.
Keywords: DMS, MS, biomarkers, detection,
OVERVIEW
Differential Mobility Spectrometry (DMS) is a rapidly gaining universal acceptance for gas phase ion separation. The
interfacing of DMS with Mass Spectrometry (MS) offers advantages over the use of a mass spectrometer alone. Such
advantages include improvements in mass spectral signal/noise, an ion separation orthogonal/complementary to mass
spectrometry, enhanced ion structural analysis and the potential for analyte quantitation without LC. This manuscript
introduces a new ESI-DMS-MS system and utilizes it to aid in the understanding of DMS separation and the advantages
of integrating DMS with MS for applications such as biomarker detection and identification.
Differential mobility spectrometry (DMS) has been called by many names. These include ion drift non-linearity
spectrometry4,5 field ion spectrometry (FIS)6, radio frequency ion mobility spectrometry7 and high-field asymmetric
waveform ion mobility spectrometry (FAIMS)8.
The DMS method of ion separation is related to mobility based ion separation in gas media9. A coefficient of ion
r
r
mobility K determines velocity ϑ of an ion under the effect of a DC electric field E between two electrodes.
Coefficient of mobility is a characteristic of individual ion species because it depends on reduced masses, cross section
and ion-molecular interaction10. Identification and/or separation of ions at ambient pressure condition through
measurement of ion velocities is used in conventional time of flight type ion mobility spectrometry (IMS)9. The principle
of ion separation in IMS is well known. A packet or swarm of ions is injected into the drift tube and due to effect of a
DC electric field, all ions start moving toward the detector. Due to differences in mobility coefficient, the swarm of ions
is separated into packages of ion with the same coefficient of mobility and ion identification is provided by drift time
measurement.
In contrast to IMS, differential mobility spectrometry does not exploit the absolute value of mobility but uses
K ( E ) = K (0)[1 + α ( E )]
i
i
. There are a number
dependence of mobility from an electric field, the alpha parameter in i
of chemical-physical mechanisms which could affect electric field dependence, such as the collision energy, ion
Chemical, Biological, Radiological, Nuclear, and Explosives (CBRNE) Sensing IX,
edited by Augustus Way Fountain III, Patrick J. Gardner, Proc. of SPIE Vol. 6954,
695411, (2008) · 0277-786X/08/$18 · doi: 10.1117/12.782437
Proc. of SPIE Vol. 6954 695411-1
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
declustering at high electric field11, dependence of polarization cross section on relative kinetic energy, resonant charge
transfer and ion dipole alignment under the effect of a strong electric field12
Ion Mobility as an Analytical Chemical Property
V1
+
υi
υi = K i * E = K i *
V2
K =
Polarization
interaction
Hardsphere
collisions
3e
2π 1 / 2
Z
(
)
16 N µ kTeff
Ω 1.1 (Teff )
Ki ( E) = Ki (0)[1 + α i ( E)]
Ω ~ πr 2
Ωpol ~1/εi ;
V 2 − V1
L
~ 1/ E
K(E) comprehensive coefficient of mobility is an analytical
“finger print” for ion identification.
αi (E) ≈
∆Κi (Ε)
Κ i (0)
Mason, E.A. ; McDaniel,E.W. “Transport Prperties of Ions in Gases” Wiley,N-Y,1988
Figure 1. The ion mobility coefficient depends on collisional cross-section and on the effective collision energy.
Operationally, the differential mobility spectrometer functions as a tunable ion filter.
DMS Ion Species Filtration in Continuous Mode
Negative ion
detector
Flow
to MS
RF
Vc
Positive ion
detector
DC voltage
Figure 2. Schematic of operation of a miniature DMS sensor as an atmosphere pressure tunable, continuously operated ion
filter
Ion species are carried through a narrow analytical gap in a gas transport stream where a transverse motion is
superimposed by a strong non-symmetric waveform high frequency RF voltage (1-2MHz.) and a static DC compensation
voltage. Positive and negative ions move with the transport gas through the analytical gap to the detector and exhaust.
Ion trajectories depend on the alpha parameter5. The rf voltage causes the ions to move up or down in the gap; if the ions
move such that they touch the upper or lower plate in the gap, they are neutralized and thus not detected the faraday
detectors at the end of the gap. By controlling the compensation voltage Vc, a desired ion species can be kept in the
center of the gap so that it is not neutralized and can be detected, either at the positive or negative detector for the
Proc. of SPIE Vol. 6954 695411-2
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
different ion polarities. If the field is zero between the two faraday detectors electrodes, filtered ions can be directed to a
second analytical device, such as a mass spectrometer, where they can be mass identified.
As a result, a miniature DMS sensor can be use as a continuous, tunable ion filter for mass spectrometry. Advantages of
the DMS prefiltering include:
Positive or negative ions are filtered without the need of retuning of the filter.
The filter can operate in “transparent” mode by turning off the RF voltage and setting Vc=0V, thereby allowing all ions
to pass into the mass spectrometer.
•
DMS devices are continuous-mode ion filters unlike the standard IMS which operates as a pulsed filter. As such
DMS enhances the total ion transfer efficiency of a selected ion to the MS.
•
Ion separation occurs according to the alpha parameter which provides enhanced chemical identification.
•
This filter allows rf voltage strengths up to E=30 kV/cm which enables to reach optimal separation for defined
ion species by regulation rf voltage intensity.
EXPERIMENTAL SETUP FOR AN ATMOSPHERIC PRESSURE MS INTERFACE
BASED ON DMS
Figure 3 shows conceptual schematics of two different types of planar DMS couplings to a mass spectrometer inlet.
These are described as inline and right angle configurations, indicating the location of the MS orifice relative to the gas
stream direction in the DMS analytical gap. An in-line configuration provides high coupling efficiency between DMS
prefilter and MS. The disadvantage of this configuration is that all un-filtered ions (neutralized in DMS) and other
impurities are also introduced into the MS spectrometer along with the filtered ions. Due to adiabatic expansion of the
gas stream in the interface area behind of orifice region, the clusterization processes between filtered ions and neutral
analyte molecules may significantly transform recorded mass-spectrum. Right-angle configuration allows exclusion of
neutral molecules from the MS inlet stream but requires some tuning of ion optics for introduction of all ions into the
MS. Coefficient of ion transmission may be different for light and heavy ion species due to their different mobilities.
Either interface can be used with most types of mass spectrometers. In our work we use the JEOL AccuTOF time of
flight mass spectrometer and the single quadrupole Waters Micromass ZQ.
MicroDMx™ MS Pre-filter Designs
Tested in Micromass and JOEL
++ +++ + +
+++
+
++
+++
+
+
+
Curtain gas
+
+
+
+
+
+
++
++
+++
+
+
+ ++ +
Curtain gas
Figure 3. Configurations for right-angle and in-line DMS prefilters.
Additionally, we used DMS pre-filters with different ion sources. For samples with sufficient vapor pressure, we used a
UV lamp to provide atmospheric pressure photoionization13. In major experiments related to study components related to
Proc. of SPIE Vol. 6954 695411-3
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
Homeland security interests (CWA agent simulants, and explosives) we used a radioactive 63Ni ion source. For all
biomarker and peptide detection experiments, we used a DMS interface with electrospray/nanospray ionization sources.
ORTHOGONALITY OF DMS AND MS SEPARATIONS
A goal of mass spectrometry and ion mobility spectrometry is chemical identification. The general principals and
operational schematics for both analytical methods are similar. Even the sequence of procedures for these two methods is
the same: generate ions from sample molecules by employing any suitable ionization method, separate formed ions, and
detect them qualitatively or quantitatively to determine abundance for different ion species. The differences between MS
and IMS technology lie in the conditions where ion separation occurs. In mass spectrometer ion separation occurs in
vacuum conditions with no interaction between formed ions and neutral gas molecules. Therefore MS provides ions
separation by mass-to-charge ration (m/z). In contrast, in IMS technology ions are separated in ambient pressure
conditions where the time between collisions is less then 10-9 s. As a result ions continuously interact with neutral gas
molecules and ion separation is based on ion size and conformation rather than simply on mass-to-charge ratio. So,
comparison of the conditions of ion separation in IMS and DMS shows these to be complimentary methods. In addition,
the enormous amount of interaction between ions and molecules dramatically increases the contribution of diffusion
processes in gases. Again, the direct comparison MS and IMS spectra show differences. Mobility based spectra show
very broad peaks for ion species in comparison with MS. Formal calculation of resolving power according to the
resolving power formula (R= peak position / peak width ) for DMS yields very small values (20-30) in comparison with
MS (at least 500). But on the other hand, DMS based technology can provide effective resolution for isomeric14 and
isobaric compounds15.
Figure 4 shows that effect of structure of ions in DMS and illustrates one specific advantage of using DMS as a pre-filter
to MS isomeric ions. Direct experiments show that DMS can provide separation for not only isomeric ions; it can
provide separation of many isobaric components also. Identification of ions with close m/z but with different structures
(formulas) can be a problem for mass spectrometry with moderate resolved power.
Separation of Isomers in DMS
p-xylene
Io n In te n s ity , a .u .
o-
o-xylene
m-xylene
o-
RF=1100V
RF=1300V
pm-
m-
p-
-15
-10
-5
0
Compensation Voltage (V)
5
-7
-6
-5
-4
-3
-2
-1
0
Compensation Voltage (V)
Figure 4. Superimposed DMS spectra for meta, para, and ortho xylene shows that DMS can resolve structural isomer that
cannot be distinguished by mass-spectrometry measurements.
To solve many problems in a single mass spectrometric measurement, accurate exact mass measurements may be
needed, which can only be done in expensive high resolution mass spectrometers. Since a DMS-MS system combines
two complimentary methods, it can solve such problems. In our first presentation in this meeting we have illustrated the
power of such system for example to separate five isobaric ion species which have the close m/z~316 ions by using mass
spectrometer with resolution ~ 500 and a miniature DMS prefilter15. In this presentation we will present results of our
investigation related to using DMS-MS system for selective detection: explosives, CWA simulants and some small
molecules which are considered as candidates for biomarkers for radiation exposure (biodosimetry). Analysis of all
Proc. of SPIE Vol. 6954 695411-4
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
obtained experimental data proves our expectation and provides information that characterizes the power of a DMS-MS
system. Pre-filtration of ions enhances the performance of mass spectrometric sniffers due to: improved signal/chemical
noise ratio, enhanced limit of detection (LOD) of analyte trace detection, enriched chemical information, reduced
chemical noise (interferences) and consequently reduced false alarm rates.
APPLICATION DMS-MS FOR DETECTION TRACES OF EXPLOSIVES
In Figures 5a and 5b, we present results from the DMS-MS interface prototype. DMS operates without change for
positive and negative ion species and provide measurement simultaneously for both ion polarities. Figure 5a,b shows that
the DMS pre-filter totally suppresses all background negative ion O2-, and (H2O)O2-. Filtration of reactant ions with
m/z=32, and 50 occurs when compensation voltage in DMS is turned on Vc = -10V (see bottom DMS spectrum on left
frame of figure 5a). When compensation voltage of DMS is turned on Vc= -1.5V which are shown in the left frames only
the negative molecular ions (M-) of 2,4-DNT (right frame of figure 5a) is selected. Estimated limit of detection for DNT
in this case is less than 1 ppbv.
2,4-Dinitro Toluene.
Negative mode.
2,4-Dinitro Toluene.
Positive mode
2.4-DNT also appears in the (+) channel as MH+.
O2- (32),
O2-•H2O (50)
M- (182)
RIP is 19, 29, 37.
MH+
RA0602061725.avi
•
Complete suppression of chemical noise is possible. (DNT ~ 1 ppbv in sensor)
a)
b)
Figure 5. DMS spectra (bottom frames) and MS spectra (upper frames) for API formation of a) negative ions and b) positive
ions 2,4 DNT.
DMS is able to filter and detect positive and negative ions simultaneously. Surprisingly for us, with increasing analyte
concentration of 2,4 DNT , positive polarity DMS channel shows an additional peak (Vc~-1V), Figure 5b. By switching
mass-spectrometer polarity on positive mode we have identified this ion species as 2.4-DNT protonated ions (MH+).
Intensity of these ions was about five times lower in comparison with negative molecular ions (M-). Peak position in
DMS spectrum for these ion differ also for negative molecular ions Vc= -1.5V and for positive ions Vc= -0.95V.
APPLICATION DMS-MS FOR DETECTION OF TRACES OF CWA SIMULANTS
Figure 6 shows an example of ion species separation in DMS and MS spectra for a gas mixture which contains a mixture
of nerve and blister chemical warfare agent stimulants: Methyl Salicylate (MS, M.W=152) and
dimethymethylphosphonate (DMMP M.W=124).
By tuning the compensation voltage of DMS pre-filter system, each component can be selectively recorded with
enhanced signal/noise ratio. The first mass spectrum (at the top of the 4 spectra) was obtained without DMS pre-filtering.
Therefore, it contains all expected ion species16: DMMP protonated monomer ions (MH+, 125 Da), Methyl Salicylate
protonated monomer ions (MH+, 153Da), and proton bounded DMMP dimmer ions (M2H+, 250Da). The bottom three
mass spectra show that by appropriately tuning DMS compensation voltage, it is possible to selectively record DMMP
monomer ions with Vc=-5V, methyl salicylate monomer with Vc=-2V, and DMMP cluster ions with Vc=+2.5V.
This efficient ion filtration can be used in combination with other analytical devices in front of the DMS. For example, it
might be different electrospray array systems. All mass-spectra which would be obtained after filtration will not show
any background ion species.
Proc. of SPIE Vol. 6954 695411-5
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
DMS–MS Detection of Mixture of Methyl Salicylate
and DMMP Vapors
MS
DMMP
monomer
10
-8
-6
-4
-2
Compensation voltage
DMMP
cluster
0
2
4
No DMS separation
Vc=-5V
Vc=-2V
Vc=+2.5V
Figure 6. DMS (upper spectrum) and MS spectra (bottom four spectra) for gas mixture which contains mixture of Methyl
Salicylate (1ppb) and DMMP(5ppb).
DEMONSTRATION OF CHEMICAL NOISE REDUCTION FOR DETECTION OF
CAFFEINE IONS.
Decreasing the number of ion species by ion prefiltering before the MS helps to clean up the MS spectrum. A smaller
number of ion species is introduced into the MS resulting in fewer fragment ions and a lower probability of interferences
with target ions. Figure 7 shows an example. DMS prefiltering reduces chemical noise and consequently it records only
selected ions, in this case the monomer ions. Mass spectrum presented on left frame shows that our caffeine sample
contains certain amount of impurity (sodium). Therefore without separation, we can see a number of peaks associated
with cluster and fragment ions. Complex DMS spectrum (bottom spectrum on right side) shows that in air it is possible
to form many complexes. Yet only the peak located in Vc=-12V can be associate with caffeine monomer peak.
i
Selective Detection of Caffeine Ions
Cafline nESI all ions
10
MH
CaItais 14OOV
fff
3-5
MNa
2.5
IIIiIII±
M2Na
I-s
' 50
LJL
100
150
200
i.L
250
Sz
300
Without filtration: RF is off
350
400
LII
450
With filtration: RF1s on, and Vc=-12V
Figure7. Selective detection of caffeine monomer peak in DMS-MS system..
Proc. of SPIE Vol. 6954 695411-6
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
To show the power of filtration we performed the following control experiment (Figure 8). We intentionally introduced
into the caffeine sample chemicals which could be considered as a source of chemical noise. These are a polyethylene
glycol (PEG) species with different molecular weights. This procedure with PEG is often used for mass calibration.
Standalone mass spectrum in this case contains a lot of peaks with high intensity and as a result the caffeine peak
(m/z=195Da) intensity close to chemical noise level. When DMS compensation voltage was tuned to filter caffeine
monomer ions, the mass spectrum became very clean and contains only one peak related to caffeine monomer. And
absolute magnitude of peak intensity is almost the same as on the mass spectrum on left side (~1V).
Filtration Caffeine Ion from Chemical Noise
(Polyethylene Glycol Species (PEG), which is Used for Mass Calibration)
Caffeine With PEG
101
.
7
With DMS
No DMS
DMS
Vc=-11.5V
n
l
— IL
0
100
200
300 400
000
600
700
Figure8. Selective detection caffeine monomer peak in complex mixture which containe sufficient amount of chemical
noise.
The prefiltering benefits are especially useful for seeking target ions in complex mixtures, for example, detection and
measurement of the biomarkers.
BIOMARKER DETECTION AND SEPARATION WITH nESI-DMS-MS
Screening for radiation exposure is a critical technology related to Homeland Security. In the event of widespread
exposure to radioactive material, a large number of people would have to be screened rapidly. The ideal technology
would be non-invasive, and provide selective and specific detection. Metabolomic work is ongoing in collaboration with
Sionex that has identified a few candidate biomarkers for radiation exposure. These biomarkers are present in simple
biofluids like urine. DMS is being investigated as a much more rapid separation technique than LC as a pre-separation
method for mass spectrometric analysis.
As an example of separation and detection of candidate biomarkers, we will show results for azelaic acid, C9H16O4, MW
= 188.10485, CAS 123-99-9, obtained using nESI-DMS-MS. Azelaic acid is a medium chain linear dicarboxylic acid
containing 7 CH2 groups between the carboxylic acid groups. It is used topically for skin conditions, and is involved in
fatty acid metabolism. Azelaic acid is observed in negative ion mode. The ions observed are predominantly the
deprotonated anion (AZA – H)- at m/z 187, and for higher mass spec inlet cone voltages as a fragment at m/z 125. Figure
9 shows that azelaic acid can be effectively separated from the fragment by DMS-MS.
The azelaic acid sample contained as impurities several related dicarboxylic acids of varying chain length. Analyzing the
DMS-MS spectrum for these related species provides a good demonstration of the ability of DMS to separate related
compounds, the extended dynamic range of the DMS-MS combination, and the suppression of chemical noise by DMS.
Proc. of SPIE Vol. 6954 695411-7
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
DM3 Separation of Azelaic Acid Conformers
m/z I 25
DM3-MS
DM3 scan
c!( N
Azc Acid ICCO YR/Vc =
/''->
:.LMSMS
lii !J.
to -8.65)
mlz I 87
DM3
separation
:[ 7T1:i
100
200
OQ
400
Figure 9. DMS is able to select individual different conformers of azelaic acid, and can suppress one ion species or the other
by compensation voltage tuning.
DMS SEPARATION OF RELATED COMPOUNDS AND DYNAMIC RANGE
Because of the high S/N achieved in nESI – DMS – MS, we were able to detect and resolve 5 additional related
dicarboxylic acids that were also present in the azelaic acid sample at quite low concentrations. This presented an
interesting demonstration of DMS separation of a related series of compounds. This ability of DMS to separate these
compounds is illustrated in the following intensity-normalized DMS plot.
DMS-MS
Acids
Figure10. DMS separation of dicarboxylic acids occurring in the azelaic acid sample. The number of connecting CH2 groups
is indicated in the legend. Ion signals for contaminant ions range from 0.2% to 8% of the azelaic acid signal.
DMS REDUCTION OF CHEMICAL NOISE
Biofluids are complex mixtures that may show mass peaks at every unit m/z. Overlapping mass spectra from unwanted
compounds and from isotope peaks of unwanted compounds cause errors in quantitative measurements, and result in
Proc. of SPIE Vol. 6954 695411-8
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
false alarms. Because the azelaic acid contains minor impurities, it is possible to use DMS filtration to select one of the
minor components and suppress chemical noise at nearby m/z values. This illustrates improvement in quantative
accuracy because the restoration of normal isotopic ratios shows that even the smallest peaks are due only to the target
ion. This is demonstrated for n=6 of the previous figure (suberic acid, 4.5% of the azelaic acid concentration) in Figure.
Chemical Noise Suppression for Suberic Acid
Present at 4.5% in Azelaic Acid Sample
cs.FII,dl I7 IOOOV(V •.l2.!l 1)
xj
.
DM3
separation
]rj
170
171
171
ITS
17S
.
110
IU
74
170
17
Figure11. Suberic acid is detected at a 4.5% level in electrospray of our azelaic acid sample.
The presence of several other species contributes chemical noise to the suberic acid signal. This chemical noise is
completely removed by DMS filtration. The remaining peaks at m/z 174 and 175 are the expected isotope peaks of
suberic acid (predicted intensities 100:9:1.2, observed 100:11:3)).
These results on azelaic acid and related species give a dramatic illustration of the separation and noise suppression
capabilities of DMS.
SUMMARY
In summary, the results obtained in this work shows that DMS prefiltering provides number advantages:
•
Enhances chemicals identification information
•
Reduces chemical noise by a factor of 10-30. Consequently enhancing the detection limit and quantitation
accuracy
•
For complex mixtures DMS pre-separation can replace or reduce the requirement and time for separation steps
in GC or HPLC.
REFERENCES
[1] Black T., Coy S., Miller R., Nazarov E. and Vouros P., “Rapic Sceening of DNA Adducts Using Differential
Mobility Spectrometry / Mass Spectrometry”, 55th ASMS WP-108, June 3-7, 2007, Indianapolis, IN.
[2] Coy S.L., Krylov E.V., Miller R.A., Nazarov E.G., “Miniature Planar Differential Mobility Spectrometry (DMS)
Prefilters for Rapid High-Resolution Pre-separation of Peptides and Other Molecular Ions by Mass Spectrometry”, 55th
ASMS, June 3-7,2007, Indianapolis, IN.
[3] Levin D.S., Miller R.A., Nazarov E.G., and Vouros P.P., "Rapid Separation and Quantitative Analysis of Peptides
Using a New Nanoelectrospray-Differential Mobility Spectrometer-Mass Spectrometer System", Anal. Chem.2006, 78,
5443-5452.
Proc. of SPIE Vol. 6954 695411-9
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms
[4] Buryakov I.A., Krylov E.V., Nazarov E.G., and Rasulev U.Kh., “A New Method of Separation of Multi-Atomic Ions
by Mobility at Atmospheric Pressure Using a High-Frequency Amplitude-Asymmetric Strong Electric Field, ”Int. J. of
Mass-spectrometry and Ion Processes”,128, 143-148 (1993).
[5] Buryakov I.A., Krylov E.V., Makas A. L., Nazarov E.G., Pervukhin V.V., and Rasulev U.Kh., “Separation of ions
according to their mobility in a strong alternating current electric field” Pis’ma Zh.Tech.Fiz 17(12), 61-65 (1991).
[6] Carnahan B., Day S., Kouznetsov V., Matyjaszczyk M., and A. Tarassov, Proc. 41st Annual ISA Analysis Division
Symposiym, Framingham, MA, 21-24 April, 1996
[7] Miller R.A., Nazarov E.G., Eiceman G.A., King A.T. “A MEMS radio-frequency ion mobility spectrometer for
chemical vapor detection”. Sensors and Actuators, A 91 (2001) 301-312.
[8] Purves R. W.; Guevremont R.; Day S.; Pipich C. W.; Matyjaszcyk M. S. Rev. Sci. Instrum. 1998, 69, 4094-4105.
[9] Eiceman G. A.; Karpas Z. [Ion Mobility Spectrometry]; CRC Press: Boca Raton, FL, 1993.
[10] Mason E. A.; McDaniel E. W. [Transport Properties of Ions in Gases]; John Wiley: New York, 1988.
[11] Krylova N., Krylov E., Eiceman G. A., Stone J. A. “Effect of Moisture on the Field Dependence of Mobility for
Gas-Phase Ions of Organophosphorus Compounds at Atmospheric Pressure with Field Asymmetric Ion Mobility
Spectrometry”. J. Phys. Chem. A 2003, 107, 3648-3654
[12] Shvartsburg A. A., Bryskiewicz T., Purves R.W., Tang K., Guevremont R., and Smith R.D. J.Phys.Chem. B.
2006,110,21966-21980)
[13] Nazarov E.G., Miller R.A., Eiceman G.A., Stone J.A. “Miniature differential mobility spectrometry using
atmospheric pressure photoionization”. Analytical Chemistry, 2006, 78(13):4553-4563
[14] Borsdorf H., Nazarov E.G., Miller R.A.. „Atmospheric pressure ionization studies and field dependence of ion
mobilities of isomeric hydrocarbons using a miniature differential mobility spectrometer”. Analytica Chimica Acta,
2006, 575, 76-88
[15] Anderson A.G., Markoski K.A., Shi Q., Coy S.L., Krylov E.V., Nazarov E.G. “DMS-IMS2, GC-DMS, DMS-MS:
DMS Hybrid Devices Combining Orthogonal Principles of Separation for Challenging Applications”, SPIE, Paper 695416
[16] Nazarov E.G., Coy S.L., Krylov E.V., Miller R.A., Eiceman G.A. “Pressure Effects in Differential Mobility
Spectrometry”. Anal. Chemistry, 78, 7697-06. 2006.
Proc. of SPIE Vol. 6954 695411-10
Downloaded from SPIE Digital Library on 02 Mar 2010 to 156.145.140.84. Terms of Use: http://spiedl.org/terms