Adjusting the Cap
Transcription
Adjusting the Cap
Support Bulletin A70-B10A As Needed February 2003 Adjusting the Cap-piercing Needle ADVIA® 70 Autosampler Product ADVIA 70 autosampler Applicability All customers with optional autosampler Estimated Time ½ hour Special Materials Phillips head screwdriver, 0.040- and 0.060-inch flat feeler gauge, 5-ml Hemogard sample tubes, utility software version 2.x.x Purpose This bulletin describes how to recognize when the sample cap-piercing punch is too deep, too shallow, or off center, and it contains the associated adjustment procedures. Problem The carousel is properly aligned when it is lined up with the hole on the cap piercer ejector plate. With the carousel in this position, extending the cap-piercing needle should not cause it to hit the carousel or the outside of the Hemogard tube. When you observe that the needle does come in contact with the carousel, you must perform an adjustment. When the cap-piercing punch is too deep or too shallow, you must perform an adjustment. When the cap-piercing punch is too shallow, the stopper blocks the tube vent. If it is too deep, the needle will not retract far enough and is exposed when running the autosampler. Procedure: Adjusting Needle Depth and Centering the Needle Punch BIOHAZARD All products or objects that come in contact with human body fluids should be handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear facial protection, gloves, and protective clothing. The operator should follow the recommendations to prevent the transmission of infectious agents in health-care settings as recommended for potentially infectious specimens in Protection of Laboratory Workers from Infectious Disease Transmitted by Blood, Body Fluids, and Tissue, 2d edition; Approved Guideline (1997) Document M29-A, National Committee for Clinical Laboratory Standards (NCCLS). This document contains complete information on user protection and it can be used as reference material for instructions on laboratory safety. If you are in a biohazardous situation and you need assistance, please contact your branch HES representative. For internal use only. Do not distribute outside of Bayer Corporation. This information is provided as an aid in Technical Support for trained support personnel. Do not attempt to service the system without the appropriate technical training and the proper tools. 083B0008-02 Rev. A, 2003-02 Page 1 of 8 Adjusting the Cap-piercing Needle 1. Close the user interface. 2. Start the utility software and connect to the autosampler. 3. Remove the carousel and front cover from the autosampler and unhook the strap joining the front cover to the autosampler. The cap-piercing mechanism is exposed. Figure 1 Cap-piercing mechanism 4. Remove the two screws from the rear hinged autosampler door and tilt the door downward. For internal use only. 083B0008-02 Rev. A, 2003-02 Page 2 of 8 Adjusting the Cap-piercing Needle 5. Loosen the Phillips-head screws on the cap piercer PCB, which is mounted on the platform above the needle. Figure 2 Cap piercer PCB For internal use only. 083B0008-02 Rev. A, 2003-02 Page 3 of 8 Adjusting the Cap-piercing Needle 6. Insert a 0.040-inch flat feeler gauge between the cap-piercer motor mounting block and the lead screw platform. Figure 3 Feeler gauge inserted 7. Use the utility software to extend the needle. WARNING Use caution when handling sharp objects such as needles and system probes. The tips are sharp and can cause skin punctures. a. Click Tool on the menu bar. A drop-down menu appears. b. Click Service. The Hardware Test window appears. c. Double-click Autosampler to expand it. d. Highlight Cap Piercer and right-click. A menu appears. e. Click Extend. The needle extends. For internal use only. 083B0008-02 Rev. A, 2003-02 Page 4 of 8 Adjusting the Cap-piercing Needle NOTE After about 12 seconds, the needle will retract. You must work quickly or you will have to perform step 7 again. 8. Slowly slide the cap piercer PCB downward toward the drive belt. 9. When the tension on the gauge is tight, tighten the Phillips head screws on the cap piercer PCB, and remove the gauge. NOTE If the system frequently looses the air gap, you may need to use a 0.060 feeler gauge instead of a 0.040 gauge. 10. Close the utility software but do not turn off either the analyzer or the sampler. 11. Restart the user interface. 12. From the run menu, select the Sample mode and then Autosampler. 13. Click Stats Auto Sampler. 14. Reinstall the carousel, ensuring it is locked into place. 15. When prompted, insert one 5-ml Hemogard sample into position 0 (the stat position) on the carousel. 16. At the end of the stat cycle, click OK to close the Stat message. 17. Inspect the punch location and needle depth. a. Remove the carousel and inspect the Hemogard stopper to see if the punch is in the center of the stopper. (If the punch is not centered, go to step 18a.). tube cap rubber stopper hole created in stopper center Figure 4 Punch properly placed in center of cap For internal use only. 083B0008-02 Rev. A, 2003-02 Page 5 of 8 Adjusting the Cap-piercing Needle b. Unscrew and swing aside the stainless steel stripper plate and inspect the needle position. When the needle is properly adjusted, it is flush with the top surface of the white delrin (plastic) cap piercer assembly (figure 5, B). If the tip is lower than the top surface, it is seated too low. If the tip of the needle can be seen, the needle is seated too high (figure 5, A). If the position is too low, go to step 18b. If it is too high, go to step 18c. Figure 5 Needle position in relationship to cap piercer assembly (stripper plate moved aside) For internal use only. 083B0008-02 Rev. A, 2003-02 Page 6 of 8 Adjusting the Cap-piercing Needle TUBE TUBE CAP RUBBER STOPPER NEEDLE Figure 6 Needle position in relationship to Hemogard sample tube 18. Adjust the needle depth and punch location on the Hemogard cap. Needle depth and Hemogard cap punch adjustment chart If the problem is… Then move the… a. The punch is not centered. Cap piercer assembly—loosen the setscrews on the cap piercer assembly and move it. If the punch is still not centered, move the optics sensor up and down until the proper punch is achieved. NOTE The optical sensor is located on the upper-left side of the stripper plate. b. The needle is too low in relationship to the cap piercer assembly, which creates a shallow punch (figure 6, E). Loosen the Phillips head screws behind the stripper plate and move the sensors. (The needle is lower than the top surface of the piercer assembly.) For internal use only. 083B0008-02 Rev. A, 2003-02 Page 7 of 8 Adjusting the Cap-piercing Needle Needle depth and Hemogard cap punch adjustment chart (continued) If the problem is… Then move the… c. The needle is too high, which creates a deeper punch. Loosen the Phillips head screws behind the stripper plate and move the sensors. (The needle extends beyond the top surface of the cap-piercer assembly (figure 5, A).) 19. Reassemble the autosampler components in the reverse order in which you disassembled it. NOTE You may have to repeat this process several times to achieve proper alignment. Hemogard is a trademark of Becton-Dickinson. For internal use only. 083B0008-02 Rev. A, 2003-02 Page 8 of 8