Hair removal - TFD Implantologia

Transcription

Hair removal - TFD Implantologia
Eterna Giovinezza
Silk Light
Twain IPL
Eterna
Giovinezza
The Stand Alone Systems
DEPILAZIONE
DEFINITIVA
Eterna Giovinezza
Silk Light
Twain IPL
The TWAIN IPL opzioni per laser
della serie Matisse
Eterna Giovinezza
Silk Light
Twain IPL
TWAIN IPL ovvero manipolo aggiuntivo per
espandere i trattamenticome IPL (depilazione,
Acne , trattamentiVasculari
Universal TWAIN connector
IPL Applications
• DEPILAZIONE permanante
• FOTORINGIVANIMENTO
• lesioni Vasculari/couperose
• lesioni Pigmentate
• trattamento Acne acuta
Eterna Giovinezza
Silk Light
Twain IPL
Scala dei fototipi (di Fitzpatric)
Eterna Giovinezza
Silk Light
Twain IPL
N°
colore
particolarita
1
Pelle bianca
Non si abbronza mai , si scotta
sempre – NON TRATTABILE IPL
2
bianca
Si abbronza con difficoltà
3
Bianco\Asia
Capelli scuri, si abbronza
moderatamente.
4
Bianco\italico si abbronza lentamente,e la perde
non completamente in inverno.
5
Ambrato
Diviene molto scuro, ma non si
scotta mai. Insensibile al sole.
6
Nero \Africa
NON TRATTABILE CON IPL
I vantaggi della tecnologia
Quanta System IPL
Eterna Giovinezza
Silk Light
Twain IPL
• lampada flasch raffreddata ad acqua forzata
• due formati della lente, piccolo e grosso per adattarsi al corpo
• sistema per calibrazione fine del manipolt
• Flasch a ripetizione per trattamenti rapidi un lampo ogni 2 sec
• cristallo per guida d’onda di drande dimensione
• protocolli di trattamento già inseriti nella macchina
• Sei differenti filtri con diverse lunghezze d’onda
• 10.000 macchine vendute nel mondo operatori soddisfatti
• marchio CE , approvato FDA
How the IPL works ?
High Power Filtered Flashlamp
Electromagnetic Spectrum of emission
0-100 nm
100-400nm
400-780nm
780nm1mm
Gamma
Rays
X-rays
Ultraviolet
Visible
Infrared
>1mm
Microwaves
Radiowaves
Eterna Giovinezza
Silk Light
Twain IPL
Pulsed-Light Systems
• Polychromatic
• Flexible system
• Less expensive than lasers
• Smaller and easier to use
than lasers systems
Skin
Waveguide
Cut Off Filter
Filtered Light
Useful spectrum
Filtro di taglio delle frazioni UV
Intense Light dannose per la pelle,
Wide spectrum
Optics of the
IPL
Water cooled High
Power Xenon
FlashLamp
The Handpieces
Manipolo fisso raffreddato ad acqua –ottica
fissa per depilazione definitiva
Eterna Giovinezza
Silk Light
Twain IPL
The Handpieces
Manipolo con ottica intercambiabile per:
Vascular, Pigmentations, acne, rejuvenation, hair
removal in small areas
Eterna Giovinezza
Silk Light
Twain IPL
Eterna Giovinezza
Silk Light
Twain IPL
The Handpieces
4.8 cm
Big Spot size
Ideale per la rimozione dei pelie
rigenerazione della pelle in grandi aree
1.3 cm
6.2 cm2 spot
2.4 cm
Small Spot size
Ideale per Vascular, Pigmentationi
e Acne treatmenti
1.3 cm
3.1 cm2 spot
Skin’s structure
Target / Chromophores
Eterna Giovinezza
Silk Light
Twain IPL
Targets:
•Oxihemoglobin
(vascular)
•Melanin (hair,
pigment)
•Water (skin)
IPL’s filters
GRAFICO curve assorbimento melanina, emoglobina
e acqua
Hair Removal
650-1200 nm
Oxihemoglobin
Extincion coefficient (cm -1)
Eterna Giovinezza
Silk Light
Twain IPL
Hair Removal
625-1200 nm
1000
Hair Removal/Skin Rejuv dark skin
590-1200 nm
Skin Rejuv ligth skin
570-1200 nm
Vascular/Pigments
550-950 nm
100
Active Acne
400-1200 nm
Water
Melanin
10
400
500
600
700
800
900
1000
1100
1200
Wavelength (nm)
(From: Lask G, Elman M, Slatkine M. Laser-assisted hair removal by selective photothermolysis. Dermatol Surg 1997; 23: 737-9)
Curve di affinità per calibrazione
di assorbimento e penetrazione
Eterna Giovinezza
Silk Light
Twain IPL
Eterna Giovinezza
Silk Light
Twain IPL
Temperatura di lavoro
Quanta System’s doppio flasch
ravvicinato non visibile a occ. n.
Eterna Giovinezza
Silk Light
Twain IPL
• n.pulsazioni (1–5)
• durata impulso (5-40ms)
• tempo di riposo tra 2 lampi
STRUTTURA DEL LAMPO CON UNA O DUE SCARICHE.
Il tempo di pausa tra due lampi serve alla pelle scura per
raffreddarsi
The Applications
• depilazione
• fotoringiovanimento
• Vasculari /couperose
• lesioni pigmentate
(lentiggini)
• trattamento Acne
Eterna Giovinezza
Silk Light
Twain IPL
Hair removal procedure through
Intense Pulsed Light (IPL)
Eterna Giovinezza
Silk Light
Twain IPL
La melanina contenuta nel follicolo assorbe
l’energia fornita dall’ipl
Il repintino innalzamento della temperatura
danneggia la struttura cellulare portando alla
distruzione del bulbo.
Affinchè il bulbo si riscaldi alla giusta
temperatura si deve calibrare
esattamente il lampo del flash
Le variabili di cui tenere conto sono:
•fototipo
•Livello di abbronzatura
•Colore del capello
• dimensione del pelo
Hair removal procedure through
Intense Pulsed Light (IPL)
Eterna Giovinezza
Silk Light
Twain IPL
La depilazione anche se
ben fatta interviene solo
sui bulbi attivi in quel
momento.
I peli hanno varie fasi di
vita dette anagen 1 e 2
catagen e telogen
Se un bulbo con contiene
almeno una minima
quantità di pigmento scuro
non subirà alcun danno e
potrà in seguito generare
un pelo. Occorrono diversi
trattamenti per arrivare a
una depilazione completa .
Si ripetono di soloto ogni
30 \35 giorni.
Wavelengths for Hair removal
Black hair
Fair hair
Eterna Giovinezza
Silk Light
Twain IPL
In funzione del fototipo la lunghezza
d’onda può variare per ottimizzare la
procedura.
Per pelli scure , meridionali , è
indicato usare 650 nm, mentre nelle
regioni padane va bene la 625 nm.
Le bionde naturali possono essere
trattate anche con 590 nm
Treatment procedure
Eterna Giovinezza
Light
Tutti i trattamentimenti contra-indicazioniSilk
Twain IPL
Patient selection
Contra-indications:
•Disturbi o cure ormonali o
disordini endocrini.
•Epilessia
•History of Melanoma
•Tatuaggi nell’area da trattare
•Immunosuppresion
- nessun trattamento della pelle
per 30 giorni successivi, non
esporsi al sole o simili.
• predisposto alla formazione di
cheloidi
• condizioni generali della pelle
non buone.
• eccesiva abbronzatura
1Controindicazioni
Eterna Giovinezza
Gravidanza – dagli studi non sono emersi riscontri di danni arrecati al feto da unSilksistema
a luce pulsata; i
Light
risultati del trattamento potrebbero non essere affidabili a causa delle fluttuazioni dei
Twainlivelli
IPL ormonali.
Disfunzioni mestruali – i pazienti con disfunzioni mestruali o iperandrogenismo accertato devono rivolgersi
a un endocrinologo per sottoporsi a una valutazione ed individuare un possibile trattamento di tipo medico.
Questi pazienti tendono a manifestare un irsutismo secondario al processo della malattia che potrebbe
rispondere al trattamento medico.
Epilessia – sono possibili reazioni epilettiche durante il trattamento.
Storia di melanoma: non effettuare il trattamento
Tatuaggi – non effettuare il trattamento su tatuaggi: rischio di bruciature e danneggiamento di pelle e
pigmenti.
Allergie – accertarsi della presenza di eventuali allergie a medicinali, lattice, alimenti o altre sostanze.
Non è un criterio di esclusione assoluto ma va valutato per eventuali contatti con allergeni.
Medicinali – con o senza prescrizione. Accertarsi di includere qualsiasi tipo di rimedio, in quanto alcuni
potrebbero causare fotosensibilità. I pazienti non devono assumere Accutane, anticoagulanti e farmaci
fotosensibilizzanti.
Storia di formazione di cheloidi – anche se esiti cicatriziali sono estremamente rari, essi possono essere
provocati dall’atto di togliere eventuali croste. Per questo motivo, si consiglia di escludere dal trattamento
persone soggette alla formazione di cheloidi.
Infezioni attive – infezioni attive e immunosoppressione compromettono la capacità di guarigione
del corpo. Riprogrammare il trattamento una volta eliminata l’infezione.
Condizioni generali della pelle precarie – il trattamento deve essere effettuato esclusivamente su pelle
sana e intatta.
Herpes nell’area da sottoporre al trattamento – il paziente deve consultare anzitutto il proprio medico per
individuare una possibile profilassi prima di iniziare il trattamento.
Uso di epilatori meccanici – interrompere l’uso almeno 4 settimane prima del trattamento.
procedura di lavoro
Eterna Giovinezza
Silk Light
Twain IPL
•
•
•
Anamnesi - Informare il paziente , chiedere che si presenti
depilato (anche per non sporcare in studio)
Compilare una scheda promemoria e fotografare la zona;
•
Preparare occhiali e allontanare chi ne è sprovvisto;
•
Proteggere nei o lesioni con adesivi o creme bianche;
•
Stendere il gel trasparente raffreddato su una area adeguata;
•
Selezionare i parametri;
•
Iniziare il trattamento.
Importante per la depilazione
Premere il manipolo sulla superficie
• premere il manipo avvicina i bulbi e
ovviamente allontana i fluidi facilitando il
passaggio dell’onda energetica
• spesso la radice del pelo
si trova a 4 \ 5 mm sotto
cute. Negli altri trattamenti
non occorre premere.
Eterna Giovinezza
Silk Light
Twain IPL
Treatment procedure
Hair Removal
Eterna Giovinezza
Silk Light
Twain IPL
Dal secondo trattamento l’esposizione ideale deve generare un lieve rossore. Se
non compare questo rossore la fluenza potrebbe essere troppo bassa.
Tuttavia entro 1\2 giorni tutto ritorna come prima (si può sentire anche un lieve
prurito ) .Le uniche creme da applicare sono quelle idratanti, a meno che si
abbia una reazione avversa eccessiva , in tal caso si utilizzano quelle con ossido
di zinco tipo Pasta Fissan Ovviamente se così avviene si deve ridurre la fluenza.
Ulteriuori informazioni sul manuale applicativo.
Post Trattamento
Hair Removal











Eterna Giovinezza
Silk Light
Twain IPL
L’utilizzo di impacchi freddi e/o creme all’ossido di zinco o aloe vera dopo il trattamento riduce la
possibilità di comparsa di eritema ed edemi e arreca sollievo al paziente.
È normale provare una sensazione simile a quella di una lieve bruciatura solare. Generalmente dura
2-24 ore, ma può persistere anche fino a 72 ore, in questo caso applicare impacchi freddi o un
panno umido sulle aree sottoposte al trattamento per 15 minuti ogni 2-4 ore, fino a quando i
sintomi non scompaiono.
Fare il bagno o la doccia, come al solito. Le aree trattate potrebbero essere sensibili alla
temperatura: bagni o docce tiepide o fredde daranno sollievo.
Non utilizzare scrubber aggressivi o esfolianti, spugne naturali o scrub brushes, fino a quando l’area
interessata non sarà tornata alla condizione antecedente il trattamento.
Non esporre le aree trattate ai raggi solari. Utilizzare uno schermo totale se l’esposizione è
inevitabile. Evitare l’esposizione al sole ridurrà la probabilità di cambiamenti del colore della pelle.
7-21 giorni dopo il trattamento si noterà una sorta di ricrescita: non si tratta tuttavia di una ricrescita
dei peli, ma sono i peli trattati che vengono espulsi, rimossi dalla pelle.
I peli che erano nella fase di riposo (telogen) al momento del trattamento entreranno nella fase
attiva di crescita (anagen) dopo 1-6 mesi, a seconda della zona in questione.
I trattamenti di follow-up potranno essere effettuati entro 1-2 settimane dalla nuova ricrescita, in
media 35 giorni per il viso, ascelle e inguine; 45 giorni per il resto del corpo.
Evitare rasatura, ceretta, nuoto, hot tubs, e un esercizio fisico eccessivo nei 7 giorni successivi al
trattamento, o comunque fino a quando non scompare il rossore.
Se si formano vesciche o lievi abrasioni, applicare una pomata antibiotica sull’area interessata.
Non cercare di togliere eventuali croste, per ridurre il rischio di infezioni o cicatrizzazioni.
Eterna Giovinezza
Silk Light
Twain IPL
Hair Removal
Female labial, skin type II
Handpiece:
625-1200 nm
Average Fluence:
20 J/Cm2
Pulse width:
Burst 2 pulses
of 20 ms each
Pulse pause:
10 ms
N. Of treatments:
5
Eterna Giovinezza
Silk Light
Twain IPL
Hair Removal
Female labial, skin type III
Handpiece:
625-1200 nm
Average Fluence:
17 J/Cm2
Pulse width:
35 ms
N. Of treatments:
5
Eterna Giovinezza
Silk Light
Twain IPL
Hair Removal
Female labial, skin type II
Handpiece:
625-1200 nm
Average Fluence:
20 J/Cm2
Pulse mode:
Burst 2
Pulse width:
15 ms
Pulse pause:
15 ms
N. Of treatments:
5
Eterna Giovinezza
Silk Light
Twain IPL
Hair Removal
Before
After 4 sessions
Male lumbar area, skin type III
After 2 sessions
Handpiece:
625-1200 nm
Average Fluence:
18 J/Cm2
Pulse mode:
Burst 2
Pulse width:
20 ms
Pulse pause:
15 ms
N. Of treatments:
4
Hair removal
Eterna Giovinezza
Silk Light
Twain IPL
Female axilla, skin type II
Handpiece:
625-1200 nm
Average Fluence:
20 J/Cm2
Pulse mode:
Burst 2
Pulse width:
20 ms
Pulse pause:
10 ms
N. Of treatments:
4
Photorejuvenation Treatment
•
•
•
•
Small Wrinkles
Dyschromies
Dilated pores
Elastosis
Eterna Giovinezza
Silk Light
Twain IPL
Photorejuvenation treatment
through Eterna
The treatment consists of
an irradiation, at correct
wavelengths and pulse
structure, of the superficial
dyscromies and the deeper
dermis, in order to get a
relevant improvement of the
skin texture and
pigmentation with the
leveling of the small
wrinkles
Eterna Giovinezza
Silk Light
Twain IPL
Treatment procedure
All treatments contra-indications
Eterna Giovinezza
Silk Light
Twain IPL
Patient selection
Contra-indications:
•Pregnancy
•Menstrual dysfunction
•Epilepsy
•History of Melanoma
•Tattoos on the area to be
treat
•Mechanical and chemical
epilation 30 days before
treatments
•History of keloid formation
•General poor skin conditions
•High tanning (bronzing)
Treatmenti Pigmentazioni e
fotoringiovanimento pelle.
Eterna Giovinezza
Silk Light
Twain IPL
1.1 Raccomandazioni post-trattamento






È normale provare una sensazione simile a quella di una lieve bruciatura solare. Generalmente dura
2-24 ore, ma può persistere anche fino a 72 ore. A questo si può accompagnare un leggero gonfiore
e/o rossore, entrambi risolvibili nel giro di 2-3 giorni. In alcuni casi, il rossore può risultare
prolungato e possono comparire delle piccole croste in prossimità di lentigo molto scure.
Le lentigo possono inizialmente sembrare rialzate e/o più scure con un perimetro arrossato.
Applicare del ghiaccio o un impacco freddo sulle aree trattate per 10-15 minuti ai fini di diminuire la
sensazione di calore.
La lesione apparirà sempre più scura nelle 24-48 ore successive. Potrà diventare marrone scura o
addirittura nera.
La lesione assumerà infine la forma di croste che inizieranno a staccarsi dopo 2-3 giorni. Non
cercare di togliere le eventuali croste per ridurre il rischio di effetti collaterali indesiderati.
Le lesioni pigmentate guariranno solitamente nel giro di 7-10 giorni. Continuerà a sbiadire nelle
2-4 settimane successive.
L’intero ciclo prevede 5-8 trattamenti, con una pausa di 3-4 settimane tra le varie sedute.


Evitare l’esposizione diretta della zona trattata ai raggi solari. Si raccomanda l’utilizzo quotidiano di uno
schermo totale per prevenire nuove lesioni o una nuova iper-pigmentazione.
Treatment Procedure
Photorejuvenation
Endpoint:
The ideal endpoint is a
moderate edema and erithema
of the treated area and a light
darkening of the pigmented
lesions, without whitening of
the skin, and low pain. This
endpoint is more visible in low
skin type (I to III)
Eterna Giovinezza
Silk Light
Twain IPL
Eterna Giovinezza
Silk Light
Twain IPL
FOTORINGIOVANIMENTO
Male facial sundamage, skin type III
Pulse mode:
burst 2 pulses
Pulse width:
20 ms
Handpiece:
570-1200 nm
Pulse pause:
20 ms
Average Fluence:
15 J/Cm2
N. Of treatments:
5
Eterna Giovinezza
Silk Light
Twain IPL
Photorejuvenation
Female facial sundamage, skin type II
Pulse mode:
burst 2 pulses
Pulse width:
20 ms
Handpiece:
570-1200 nm
Pulse pause:
30 ms
Average Fluence:
16 J/Cm2
N. Of treatments:
4
Eterna Giovinezza
Silk Light
Twain IPL
Female facial sundamage, skin type III
Handpiece:
570-1200 nm
Average Fluence:
16 J/Cm2
Pulse mode:
burst 2 pulses
Pulse width:
20 ms
Pulse pause:
20 ms
N. Of treatments:
4
Eterna Giovinezza
Silk Light
Twain IPL
Female facial sundamage, skin type III
Pulse mode:
burst 2 pulses
Handpiece:
590-1200 nm
Pulse width:
20 ms
Pulse pause:
20 ms
Average Fluence:
16 J/Cm2
N. Of treatments:
3
FOTORINGIOVANIMENTO
DEL VISO
Female facial sundamage, skin type III
Eterna Giovinezza
Silk Light
Twain IPL
Pulse mode:
burst 2 pulses
Pulse width:
20 ms
Handpiece:
570-1200 nm
Pulse pause:
20 ms
Average Fluence:
15 J/Cm2
N. Of treatments:
3
Vascular Treatments
–
–
–
–
Rosacea
Telangiectasia
Dyschromias
Red Scars
Eterna Giovinezza
Silk Light
Twain IPL
Vascular Treatments
Vascular lesions like
telangectasia and
rosacea can be
coagulated by the light
energy of the IPL,
choosing the right range
of wavelengths and pulse
duration.
The correct handpiece, is
the 550 nm able to treat
the skin up to skintype IV.
Eterna Giovinezza
Silk Light
Twain IPL
Treatment Procedure
Vascular Treatments
The small area (3.1Cm2) is highly suggested
Do not treat near the eye area. Always use eye protection
during treatments.
 Wash and dry the area to be treated. The area should be free
of makeup, ointment or any hair prior to treatment.
 Apply Quanta Transparent Gel to treatment area.
 Start spot tests with the lowest setting listed in the
corresponding Test Spot Table and increase to next step,
without overlapping of pulses, after checking for skin reaction. If
there is evidence of darkening, strong pain or deep reddening of
the skin reduce the setting to a lower one, and re-test.
 Do not use pressure on the skin with the handpiece during the
treatment.
 Treatment may be administered at settings, as indicated by
the test spots and below the threshold for adverse skin
reactions.
Eterna Giovinezza
Silk Light
Twain IPL
Treatment Procedure
Vascular Treatments
Endpoint:
The ideal endpoint is a
moderate erythema with
disappearing or color changes
of the vessel without whitening
of the skin, and low pain.
Eterna Giovinezza
Silk Light
Twain IPL
Treatment Procedure
Vascular Treatments
Eterna Giovinezza
Silk Light
Twain IPL
Post Treatment:
 Mild sunburn-like sensation is expected. This usually lasts 2-24 hours but can
persist up to 72 hours. For better patient’s comfort, cool the skin after
treatment.
 Avoid aggressive scrubbing and use of exfoliants, scrub brushes and loofah
sponges until the treatment area has returned to its pretreatment condition
 Avoid sun exposure to treated areas. Use a sunblock if sun exposure is
unavoidable. Sun avoidance will decrease likelihood of skin color changes
 The lesion will gradually turn darker over the next 24-48 hours
 The lesion will progress to scabs/crusting and will start flaking off in 2-3
days.
 Do not pick, scratch or remove scabs. This will prevent unwanted side
effects.
 Apply cicatrizing cream at the end of the treatment. Do not use
corticosteroids.
 The lesion is usually healed in 7-10 days.
 In average 1-3 sessions every 20 days are necessary.
Treatment Procedure
Vascular Treatments
Eterna Giovinezza
Silk Light
Twain IPL
Female facial rosacea, skin type I
Treatment parameters with Silk Light
Handpiece:
550-950 nm
Average Fluence:
14 J/Cm2
Spot:
3.1 cm2
Pulse mode:
single
Pulse width:
15 ms
N. Of treatments:
4
Treatment Procedure
Vascular Treatments
Eterna Giovinezza
Silk Light
Twain IPL
Female facial telangectasia, skin type II
Pulse mode:
single
Treatment parameters with Silk Light
Pulse width:
25 ms
Handpiece:
550-950 nm
Average Fluence:
16 J/Cm2
N. Of treatments:
4
Spot:
3.1 Cm2
Pigmented lesions treatments
– Cafè au lait spots
– Benign lentigo
– Freckles
Eterna Giovinezza
Silk Light
Twain IPL
Pigmented lesions procedure through
IPL
Eterna Giovinezza
Silk Light
Twain IPL
Wavelengths from 550 nm must be
used to heat benign pigmented
lesions, such as the lentigo or age
spots.
The pulse width must be very short
to prevent the abnormal thermal
diffusion from the pigment patch to
the surrounding tissue. Pulse width
≤ 15 ms
Treatment procedure
Pigmented lesions
endpoint:
The ideal endpoint is a
moderate darkening of the
pigmented lesions, without
whitening of the skin or
darkening also of the
surrounding areas.
Eterna Giovinezza
Silk Light
Twain IPL
Original color
Darkening after 15 min
Pigmented lesions procedure through
IPL
Male forehead lentigo, skin type II
Pulse mode:
single
Pulse width:
10 ms
Handpiece:
550-950 nm
Average Fluence:
16 J/Cm2
N. Of treatments:
2
Spot:
3.1 Cm2
Eterna Giovinezza
Silk Light
Twain IPL
Pigmented lesions procedure through
IPL
Male facial pigment patch, skin type III
Eterna Giovinezza
Silk Light
Twain IPL
Pulse mode:
single
Pulse width:
10 ms
Handpiece:
550-950 nm
Average Fluence:
14 J/Cm2
N. Of treatments:
2
Spot:
3.1 Cm2
Pigmented lesions procedure through
IPL
Female facial melanine patch, skin type IV
Handpiece:
550-950 nm
Average Fluence:
13 J/Cm2
Spot:
3.1 Cm2
Pulse mode:
single
Pulse width:
15 ms
N. Of treatments:
2
Eterna Giovinezza
Silk Light
Twain IPL
Acne Vulgaris treatments
– Active acne
– Comedonal acne
Eterna Giovinezza
Silk Light
Twain IPL
Acne procedure through IPL
Eterna Giovinezza
Silk Light
Twain IPL
The treatment acts in two ways:
1. Photo-chemical
2. Photo-thermal
The first trough the Porfirin
absorption of the light that
produce free-radicals able to
oxidize the bacteria
The second with a reduction of the
sebaceous gland activity
Active acne
SYSTEM:
SKIN TYPE
AREA:
SPOT SIZE:
EV. FLUENCE:
SESSIONS:
FOLLOW UP:
Eterna Giovinezza
Silk Light
Twain IPL
IPL 400-1200 nm filter
2
face
4.8x1.3 cm
7 J/Cm2
5
2 months
Active acne
SYSTEM:
SKIN TYPE
AREA:
SPOT SIZE:
EV. FLUENCE:
SESSIONS:
FOLLOW UP:
IPL 400-1200 nm filter
2
face
4.8x1.3 cm
7 J/Cm2
5
2 months
Eterna Giovinezza
Silk Light
Twain IPL
Eterna Giovinezza
Silk Light
Twain IPL
Technical specifications
Eterna Giovinezza
Hair Removal
Photorejuvination
Vascular
Pigmentations
Acne
Wavelength
590 or 625 or
650-1200
570-1200 or
590-1200
550-950
400-1200
Spot size
48 x 13
25 x 13
48 x 13
25 x 13
48 x 13
25 x 13
48 x 13
25 x 13
Pulses
Pulse length
(with burst)
Single or burst
Up to 240 ms
Single or burst
Up to 240 ms
Single or burst
Up to 240 ms
Single or burst
Up to 240 ms
Fluence
Up to 40 J/Cm2
Up to 40 J/Cm2
Up to 40 J/Cm2
Up to 20 J/Cm2
Pulse delay
Rep. rate
Up to 50 ms
2 sec
Up to 50 ms
2 sec
Up to 50 ms
2 sec
Up to 50 ms
2 sec
Eterna Giovinezza
Silk Light
Twain IPL
Minimum session cost
Treatment cost
Hair removal
Average treatment time (min.)
Minimum
Face
15
120,00
Moustaches
10
60,00
Chin
10
110,00
Arms
15
150,00
Armpits
12
120,00
Chest
25
180,00
Abdomen
25
180,00
Shoulders
25
160,00
Back
30
200,00
Bikini line
15
150,00
Leg (complete) each
Hands and feet
35
250,00
10
100,00
Eterna Giovinezza
Silk Light
Twain IPL
Minimum sessions cost
Skin Rejuvenation
Acne
Average treatment time
(min.)
Cost
MIN €
15
80,00
Average treatment time
(min.)
Cost
MIN €
15
60,00
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
General project – www.generalproject.com
Beauty Flash from 20.000$
Hair Removal
Wavelength
Spot size
Photorejuvinatio
n
590-1200
35 x 25 mm
Pulses
590-1200
35 x 25 mm
(20x10 mm)
Single
Pulse length
20 ms
20 ms
Fluence
15-30 J/Cm2
15-30 J/Cm2
1-6 sec
1-6 sec
Single
Pulse delay
Rep. rate
Sales argumentations:
•Large Spot Size
•No skin cooling required
•Plug-in hand piece
•Automatic recognition of filtration in use
•Flash Card shots recharge
•Light weight 7.5 Kg
•Conformity FDA e CE standards
Acne
Vascular
No
No
Comments:
•Lamp with not enough cooling (air cooling) with
consequent production of heating on the treated
skin, short flashlamp lifetime and slow repetition
rate (up to 6 sec)
•Cartridge with only 10.000 shots
•Only single and fixed shot available
•Long delays of the pulses causing long
treatments time
•Small spot size (20 x 10 mm) built with a plastic
gate on the emission window that produces hand
piece heating.
•No Clinical documentation and literature
available with the machine
•Very noisy device
•Software without detailed information
•The standard filter at 590-1200 nm is indicated
for hair removal only in low skin type I - II and III
•Equipment not indicated for vascular,
pigmentations and acne treatments
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
General project – www.generalproject.com
Flash 1 From 25.000$
Hair Removal
Photorejuvination
Acne
Vascular
Wavelength
590-1200
590-1200
390-520
590-1200
Spot size
50 x 25
50 x e 20 x 10
Pulses
Single o burst
Single o burst
Pulse length
20 ms
20 ms
Fluence
18-38 J/Cm2
18-38J/Cm2
Pulse delay
5 ms
5 ms
Rep. rate
1-6 sec
1-6 sec
Sales argumentations:
•Large Spot Size
•No skin cooling required
•Plug-in hand piece
•Automatic recognition of filtration in use
•Flash Card shots recharge
•CE Certified
COMMENTS:
•Lamp with not enough cooling (air cooling) with
consequent production of heating on the treated
skin, short flashlamp lifetime and slow repetition
rate (up to 6 sec)
•Fixed length and pause pulses
•Cartridge with only 10.000 shots
•Long delays of the pulses causing long treatments
time
•Small spot size (20 x 10 mm) built with a plastic
gate on the emission window that produces hand
piece heating.No Clinical documentation and
literature available with the machine
•Very noisy device
•Small black and white display
•The standard filter at 590-1200 nm is indicated
for hair removal only in low skin type I - II and III
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
General project – www.generalproject.com
Flash 1 PLUS from 28.000$
Wavelength
Spot size
Hair Removal
590-1200
50 x 25
Photorejuvination
590-1200
50 x or 20 x 10
Pulses
Single or burst
Single or burst
Pulse length
Fluence
20 ms
10-45 J/Cm2
Pulse delay
20 ms
2-40 J/Cm2 only
with filter 5201200 at Flash
lamp level
5 ms
Rep. rate
1-5 sec
1-5 sec
Sales argumentations:
5 ms
•Large Spot Size
•10.4 inches color display
•Skin cooling not required
•Plug-in hand piece
•Automatic recognition of filtration in use
•Flash Card shots recharge
•FDA, KFDA e CE certified
•Telediagnosis
Acne
390-510
50 x or 20 x
10
Single or
burst
Vascular
520-1200
50 x or 20 x 10
Single or burst
Comments:
•Lamp with not enough cooling (air cooling)
with consequent production of heating on
the treated skin, short flashlamp lifetime and
slow repetition rate (up to 6 sec)
•Cartridge with only 10.000 shots
•Long delays of the pulses causing long
treatments time
•Small spot size (20 x 10 mm) made with a
plastic gate on the emission window that
produces hand piece heating.
•No Clinical documentation and literature
available with the machine
•The standard filter at 590-1200 nm is
indicated for hair removal only in low skin
type I - II and III
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Energist – www.energist-international.com
Ultra From 35.000$
Wavelength
Spot size
Hair Removal
610-1000
50 x 10
Photorejuvination
530-1000
50 x 10
Pulses
burst
burst
Pulse length
3-7 ms
3-7 ms
Fluence
6-51 J/Cm2
6-51 J/Cm2
Pulse delay
1-20 ms
1-20 ms
Rep. rate
Up to 6 sec
Up to 6 sec
Sales argumentations:
Acne
No
•Large Spot Size
•Painless and safe hair removal and photorejuvination
•Indicated for treatments on skin type I to V
•Marketed as VPL (Variable Pulse Light) and not as IPL,
presented as a new technology different from the
traditional one (IPL)
•Large color display
Vascular
No
COMMENTS
•Only Burst mode, it is not possible to
deliver a single pulse, necessary for effective
treatments of vascular lesions
•VPL is a very long burst mode with very
short pulses, each shots are too close to the
TRT of the epidermis (< 10ms). Each pulse
is consequently absorbed very well by the
epidermis that has to be contrarily
protected. It is much safer to use burst with
each pulse longer than the epidermis TRT
(>15 ms). Energist burst safety is not
proven
•Long rep. rate causes long treatment time
•Only Spot size available of 50 x 10 mm
(5cm2 Eterna Giovinezza as 6.2 Cm2)
•No Clinical literature available with the
machine
•The equipment is not indicated for vascular
treatments, hyper pigmentations and acne
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Radiancy – www.radiancy.com
Skin Station From 20.000$
Hair Removal
Photorejuvin Acne
ation
Vascular
Wavelength
400-1200
400-1200
430-1100
400-1200
Spot size
55 x 22
55 x 22
55 x 22
35 x 22
Pulses
Pulse length
1
35 ms
1
10 ms
1
35 ms
1
10 ms
Fluence
3-10 J/Cm2
3-7 J/Cm2
3-9 J/Cm2 3-7 J/Cm2
Pulse delay
Rep. rate
-
-
-
Sales argumentations:
•Large Spot Size
•Painless and safe
•Thermal and photo thermal mixed effect
marketed as LHE
•Indicated for treatments on skin type I to V
•Very low fluence for safety and long life flashlamp
-
Comments:
•Primitive software.
•The only advantage of this system is the safety
thanks to absolutely insufficient fluence
•Only single pulse emission
•Very low fluence. Theoretically permanent hair
reduction (hair removal) is not possible.
•LHE is heat and photo thermal combination.
Hair are targeted combining both this effects.
This is proven that don’t work because the
spectrum of emission is to high from 400 to
1200 nm that the most of the energy delivered
by the flashlamp (that is very low max 9 J/Cm2)
is visible and only the last part of the visible light
start go deep (es. 600 nm reach 1mm depth,
500 only 0.4 mm). considering scattering, only
2-3 J/Cm2 reach the depth of the hair follicles in
hair removal treatment. In second way the heat
is by conduction from the handpiece to the skin
so the epidermis must be at higher temperature
than dermis, exactly the opposite of the
necessity for safe and effective treatments.
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Radiancy – www.radiancy.com
Spa Touch II From 15.000$
Wavelength
Hair
Removal
400-1200
Spot size
55 x 22
Pulses
1
Pulse length
35 ms
Fluence
3-9 J/Cm2
Pulse delay
Rep. rate
-
Sales argumentations:
Photorejuvination
-
•Large Spot Size
•Painless and safe
•Thermal and photo thermal mixed effect
marketed as LHE
•Indicated for treatments on skin type I to V
•Very low fluence for safety and long life
flashlamp
Acne
-
Vascular
-
Comments:
•Primitive software.
•The only advantage of this system is the safety
thanks to absolutely insufficient fluence
•Only single pulse emission
•Very low fluence. Theoretically permanent hair
reduction (hair removal) is not possible.
•LHE is heat and photo thermal combination. Hair are
targeted combining both this effects. This is proven
that don’t work because the spectrum of emission is
to high from 400 to 1200 nm that the most of the
energy delivered by the flashlamp (that is very low
max 9 J/Cm2) is visible and only the last part of the
visible light start go deep (es. 600 nm reach 1mm
depth, 500 only 0.4 mm). considering scattering, only
2-3 J/Cm2 reach the depth of the hair follicles in hair
removal treatment. In second way the heat is by
conduction from the handpiece to the skin so the
epidermis must be at higher temperature than dermis,
exactly the opposite of the necessity for safe and
effective treatments.
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Palomar – www.palmed.com
StarLux from 75.000$
Hair Removal
Photorejuvination
Wavelength nm 525-1200
525-1200
650-1200
Spot size
1.2 x 2.8 or 1.6 x 1.6 x 4.6 cm
4.6 cm
Pulses
1
1
Pulse length
Up to 500 ms
Fluence
Up to 40 J/Cm2
Up to 40 J/Cm2
Up to 45 J/Cm2
Pulse delay
Rep. rate
-
Sales argumentations:
•Large Spot Size
•Integrated skin cooler
•Big color touch screen display
•Long life flashlamp
•Compact design
•Platform for optional laser handpiece
Acne
400-700/8701200
1.6 x 4.6 cm
Vascular
500-670/870-1400
1
1
Up to
25J/Cm2
-
Up to 60 J/Cm2
1.0 x 1.5 cm
-
Comments:
•Quanta System Light B TWAIN IPL has
yag 1064 nm laser more powerful at
cheaper cost
•The standard handpiece for hair removal
is too aggressive for dark skins (525nm)
•Only single pulse emission. The burst
mode is safer when darker skins are
treated
•The contact integrated skin cooler doesn't
guarantee the suitable protection with
standard handpiece for hair removal
because of the too large spectrum of
emission
•Very expensive system
•The software don’t suggest operative
protocols
•Expensive flashlamp replacement
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Palomar – www.palmed.com
MediLux from 40.000$
Hair Removal
Wavelength
nm
525-1200
650-1200
Spot size
Pulses
Pulse length
Fluence
1.6 x 4.6
1
Up to 100 ms
Up to 20
J/Cm2
Up to 15
J/Cm2
-
Pulse delay
Rep. rate
Sales argumentations:
•Large Spot Size
•Integrated skin cooler
•Easy to use
•Long life flashlamp
•Compact design
Photorejuvinati
on
525-1200
1.6 x 4.6
1
Acne
Vascular
400700/8701200
1.6 x 4.6
1
500-670/8701400
1.2 x 1.2
1
Up to 20 J/Cm2 Up to 20
J/Cm2
Up to 33 J/Cm2
-
-
-
Comments:
•The standard handpiece for hair removal is too
aggressive for dark skins
•The filters that allow wavelength up to 1200 nm are
dangerous especially for dark skin because of the
absorption by the skin’s water is higher than the water
absorption with filters cutted at 950 nm. With this filters,
many energy is absorbed by all the skin, not only by
chromophore melanin that is the target for hair removal
treatment, increasing the risk of side effects
•Only single pulse emission. The burst mode is safer
when darker skins are treated
•The contact integrated skin cooler doesn't guarantee the
suitable protection with standard handpiece for hair
removal because of the too large spectrum of emission
•The fluences declared are reached with the maximum
pulse duration. This time is too long for effective
treatment of the hair especially if the hair are thin (see
selective photothermolisis). This system, in many
conditions, cannot reach the theoretical performances for
effective treatments.
•Very expensive system
•The software don’t suggest operative protocols
•Expensive handpiece replacement
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Palomar – www.palmed.com
EsteLux from 30.000$
Hair Removal
Photorejuvinati
on
525-1200
Acne
Vascular
500-670/8701400
1
Wavelength
nm
525-1200
650-1200
Spot size
1.6 x 4.6
1.6 x 4.6
400700/8701200
1.6 x 4.6
Pulses
1
1
1
Pulse length
Up to 100 ms
Fluence
Up to 15
J/Cm2
Up to 12
J/Cm2
Up to 15 J/Cm2 Up to 16
J/Cm2
Up to 33 J/Cm2
Pulse delay
-
-
-
-
Rep. rate
-
-
-
-
Sales argumentations:
•Large Spot Size
•Easy to use
•Long life flashlamp
•Compact design
1.2 x 1.2
Comments:
•The standard handpiece for hair removal is too
aggressive for dark skins
•The filters that allow wavelength up to 1200 nm are
dangerous especially for dark skin because of the
absorption by the skin’s water is higher than the water
absorption with filters cutted at 950 nm. With this filters,
many energy is absorbed by all the skin, not only by
chromophore melanin that is the target for hair removal
treatment, increasing the risk of side effects
•Only single pulse emission. The burst mode is safer
when darker skins are treated
•The roller for skin cooling is uncomfortable and
ineffective
•The fluences declared are reached with the maximum
pulse duration (100 ms). This time is too long for
effective treatment of the hair especially if the hair are
thin (see selective photothermolisis) where the pulse
must be shorter than 60 ms. Anyway this machine
cannot reach the theoretical performances for effective
treatments.
•Expensive system compared with systems with the
same performances
•The software don’t suggest operative protocols
•Expensive handpiece replacement
Comments:
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Lumenis – www.lumenis.com
Quantum SR from 55.000$
Wavelength
Spot size
Pulses
Pulse length
Fluence
Pulse delay
Rep. rate
Hair Removal
560-1200
8 x 34
Only burst 2 -3
6-26 ms
15-45 J/Cm2
5-60 ms
2 sec
Sales argumentations:
Photorejuvination
590-1200
8 x 34
Only burst 2-3
6-26 ms
15-45 J/Cm2
5-60 ms
2 sec
•Large Spot Size
•Integrated contact skin cooling
•Touch screen display with programs
•Compact design
•Fast repetition rate
Acne
no
Vascular
560-1200
8 x 34
Only burst 2-3
6-26 ms
15 – 45 J/Cm2
5-60 ms
2 sec
Comments:
•Very small spot size. This system is especially
for skin rejuvenation. The area of the spot size
is only 2.72 Cm2 (less than half spot area of
eternal or silk light) too small for hair removal
in big areas
•The pulse duration is limited to 26 ms. This
way the system is too aggressive and the
using of burst mode is a must.
•Expensive flash lamp replacement4
•The standard hair removal hadpiecce 5601200 is too aggressive for dark skin
treatments.
•The vascular handpiece is cutted at 1200 nm.
The Eterna Giovinezza vascular handpiece is
cutted at 950 nm to limit the water absorption,
dangerous for this application.
•Expensive system
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
DDD – www.3d.dk
Ellipse Light from 45.000$
Wavelength
Spot size
Pulses
Pulse length
Fluence
Hair Removal
600-950
10 x 48
1 or burst
-
Photorejuvination Acne
530-750
no
10 x 48
1 or burst
-
Pulse delay
-
-
Rep. rate
2 sec
2 sec
Sales argumentations:
•Large Spot Size
•Advanced software
•Touch screen display
•Two handpieces connected
•Fast repetition rate
•Dual filtering technology (I2PL)
Vascular
no
Comments:
•Small spot size. This system as only 4.8 Cm2 spot
area, Eterna Giovinezza has 6.2 Cm2 spot area
(20% more). This mean more shots, more
treatment time and less flashamp lifetime for
Ellipse.
•Expensive flash lamp replacement
•The dual filtering is not an exclusive technology of
DDD. Eterna Giovinezza use double filtering only for
vascular and pigmentation where the water
absorption is really dangerous. In the hair removal
treatment the water absorption improve the
response of fair hair and in Skin rejuvenation
treatment, the water absorption is very important
for the dermal heating with consequent new
collagen formation
•The handpiece for photorejuvenation is cutted to
low for effective dermis heating and neocollagen
formation
•Expensive system
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
DDD – www.3d.dk
Ellipse Flex from 65.000$
Photorejuvination
Acne
Vascular
Wavelength
Spot size
Pulses
Pulse length
Fluence
Pulse delay
Hair
Removal
600-950
10 x 48
1 or burst
26 J/Cm2
-
530-750
10 x 48
1 or burst
26 J/Cm2
-
no
555-950
10 x 48
1 or burst
Rep. rate
2 sec
2 sec
Sales argumentations:
•Large Spot Size
•Advanced software
•Touch screen display
•Two handpieces connected
•Fast repetition rate
•Dual filtering technology (I2PL)
26 J/Cm2
2 sec
Comments:
•Small spot size. This system as only 4.8 Cm2 spot
area, Eterna Giovinezza have 6.2 Cm2 spot area
(20% more). This mean more shots, more
treatment time and less flashamp life for Ellipse.
•Expensive flash lamp replacement
•The maximum fluence is only 26 J/Cm2 it mean
that the flaslamp is used always at high levels
reducing the lifetime
•The handpiece for photorejuvenation is cutted to
low for effective dermis heating and neo-collagen
formation
•The dual filtering is not an exclusive technology
of DDD. Eterna Giovinezza use double filtering
only for vascular and pigmentation where the
water absorption is really dangerous. In the hair
removal treatment the water absorption improves
the response of fair hair and in Skin rejuvenation
treatment the water absorption is very important
for the dermal heating with consequent new
collagen formation
•Very expensive system
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Luxsano – www.luxsano.com
Plasmalite FPL
from 25.000$
Photorejuvination
Acne
Wavelength
Spot size
Pulses
Pulse length
Fluence
Pulse delay
Hair
Removal
615-1200
10 x 20
1
2-500 ms
180 J/Cm2
-
550-1200
10 x 20
1
2-500 ms
180 J/Cm2
-
515-1200
10 x 20
535-1200
10 x 20
1
2-500 ms 2-500 ms
180 J/Cm2 180 J/Cm2
Rep. rate
1 sec
1 sec
1 sec
Sales argumentations:
•Fluorescent technology
•Wide range of applications
•Easy flashlamp replace
•Pulse duration up to 500ms
•Fast repetition rate 1Hz
Vascular
1 sec
Comments:
•Small spot size. This system as only 2 Cm2
spot area, Eterna Giovinezza have bigger
spot area (300% more). This mean more
shots, more treatment time, less results
because of the difficult overlap of the pulses
and less flashamp life for Plasmalite.
•Very long treatments time also with 1 pulse
every second because of the small spot size
makes the pulse spacing very difficult
•Expensive flash lamp replacement
•Short life flashlamp
•The maximum fluence is 180 J/Cm2. with
the same spot size and pulse duration,
Eterna Giovinezza could delivery >300J/Cm2.
•The handpiece for photorejuvenation is
cutted to low (550 nm) it could be too
aggressive for higher skintype.
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Alma Lasers – www.almalasers.com
Harmony from 40.000$
Hair
Removal
Rejuvenation Acne
Pigment
Vascular and
restoration pigmentation
Wavelength (nm) 650-950
570-950
420950
Spot size (mm)
40 x 16
40 x 16
40 x 16
Pulses
Pulse length
1
12, 30, 50
1
10, 12, 30
40 x
40 x 16
16
1
1
30, 40, 30, 40, 50
50
5-20
No
3 sec
5-25
No
3 sec
Fluence (J/Cm2) 5-25
Pulse delay
No
Rep. rate
3 sec
Sales argumentations:
5-25
No
3 sec
•Multi-application
•Upgradeable
•Large Spot Size
•Transportable
•Can be upgraded with laser guns
•AFT technology
300-380
0.2-1
No
6 sec
540-950
1
10, 12, 15
Comments:
•Only one spot size. Vascular and
pigmentations treatment needs smaller
spot size to improve the comfort and
reduce the possibility to damage the
surrounding health skin.
•The pulse duration is limited to few
selections only in single pulse. This way
the system could be too aggressive
especially in darker skins.
•Expensive flash lamp replacement
•The AFT technology is only marketing. It
is important that the wavelength must be
correct. Our IPL with hair removal
handpiece has 30J/Cm2, 5 more than this
machine.
•The laser guns are not powerful but
heavy and not easy to be used. Our
combined laser as Light B TWAIN is much
more powerful and professional
Competitors
Eterna Giovinezza
Silk Light
Twain IPL
Linton – www.lynton.co.uk
Luminette Advance from 35.000$
Wavelength
(nm)
Spot size
(mm)
Pulses
Pulse length
Fluence
(J/Cm2)
Pulse delay
Rep. rate
Hair
Rejuvenation Acne
Removal
625-1100 550-1100
or 7301100
35 x 10
35 x 10
Vascular and
pigmentation
550-1100
Burst
50-190
Burst
50-190
Burst
50-190
40
40
40
No
Up to 1
sec
No
Up to 1 sec
No
Up to 1 sec
Sales argumentations:
•Compact
•Upgradeable
•Multi Spot Size
•Transportable
•ILG technology (interchangeable
light guide)
10 x 10
Comments:
•The biggest spot size is half dimension of the
standard Quanta System spot size (6.2 cm2). This
way, the flashlamp life of this system must be very
long (more than 80.000 shots) to compete with the
Quanta System flashlamp life.
•The pulse duration is limited to the burst, the
shortest time is 50 ms in burst
•Expensive flash lamp replacement
•The handpiece for pigment restoration has no
literature or important testimonial that can
demonstrate that it works
•The AFT technology is only marketing. It is important
that the wavelength must be correct. Our IPL with
hair removal handpiece has 30J/Cm2, 5 more than
this machine.
•The laser guns are not powerful, heavy, and not
easy to use. Our combined laser as Light B Plus is
much more powerful and professional
Competitors
Deka
Eterna Giovinezza
Silk Light
Twain IPL
– www.dekamela.com
Photosilk from 22.000$
Hair Removal
Wavelength
(nm)
Spot size (mm)
Pulses
550-950
650-950
46 x18
(standard)
46x10
Burst
Pulse length
10-340
Fluence (J/Cm2) 30 (46x10)
17 (46x18)
Rep. rate
Up to 1 sec
Sales argumentations:
•Patented U shape flashlamp
•Upgradeable
•Multi Spot Size
•Large spot size
•Up to one pulse per second
Rejuvenation Acne
Vascular and
pigmentation
Comments:
•The biggest spot size has very low fluence
theoretically at the lower limit to be effective
•Cut-off filter of 550 nm of the standard
handpiece is too low for dark skin types
•Only Hair removal treatments.
•Too big for the performances
Eterna Giovinezza
Silk Light
Twain IPL
Eterna Giovinezza