Tek-CARE® NC300 II Nurse Call System

Transcription

Tek-CARE® NC300 II Nurse Call System
Tek-CARE® NC300 II
Nurse Call System
Submittal Package
as of 3/15/2016
www.tektone.com
Phone:
(828) 524-9967
Toll-Free: (800) 327-8466
Sales:
Option 2
Tech Support: Option 3
277 Industrial Park Road
Franklin, NC 28734
[email protected]
Fax: (828) 524-9968
Copyright © 2016 TekTone® Sound & Signal Mfg., Inc., All rights reserved.
No part of these publications may be copied without the express written permission of TekTone®
Sound & Signal Mfg., Inc. The contents of these specification sheets are furnished for informational
use only, are subject to change without notice, and should not be construed as a commitment by
TekTone® Sound & Signal Mfg., Inc. TekTone® Sound & Signal Mfg., Inc. assumes no responsibility
or liability for any errors or inaccuracies that may appear in this documentation.
TekTone, the TekTone logo, Tek-Alert, Tek-Bridge, Tek-Call, Tek-Care, TekCD, Tek-Check-In,
Tek-Com, Tek-Entry, Tek-Guard, Tek-Micro, Tek-MMARS, Tek-Paging, Tek-Safe and Tek-Status
are either registered trademarks or trademarks of TekTone® Sound & Signal Mfg., Inc. in the United
States and/or other countries. All other trademarks are the property of their respective owners.
NC305LCD Master Station
IL693 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 8 - 07/2013
Features
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
UL® 1069 Listed
Supervised
Serial & Keyboard ports
Spill-proof membrane switch with alphabetic
characters
Easy access to commonly used menu functions
Simultaneous display of 20 calls
Clear, distinct voice communications
Simple, menu-driven programmable functions
Master-to-master intercom capability
Push-to-Talk button or handset intercom
Calls displayed by priority and origination
Calls identified by room and bed number
Programmable patient priority and privacy
Calls-waiting, calls-overtime and cord-out indication
View station configuration capability
Initiate stat staff emergency request
Upgrade patient call
Distinct tones for high, medium and low-priority calls
Nurse follower function
Paging by zone or to all staff
Patient monitor single or group
Manual tone-silencing switch (for Routine calls only)
High priority calls override tone-off feature
Simplified multiplex wiring uses plug-in type
connectors
Continuous display of nurse/aide location and nurse/
aide requested
Use masters in parallel or independently
Description
A major component of the Tek-CARE®NC300II System,
the NC305LCD master station supplies the user interface
to respond to calls, issue staff requests and alter room
programming status to accommodate differing patient
needs.
The master console allows for room and staff communication
by either open speaker or handset for additional privacy. It
includes an integrated backlit color LCD VGA display with
NC305LCD Master Station
10.4" viewing area to display up to 20 active patient, staff
or duty calls, plus columns for two levels of staff presence
and STAT priority calls in English language format. Master
function buttons are provided to select the system functions
as follows:
reset
aide
nurse
stat
menu
page
follow
talk
alpha
(black) Terminates most operations.
(yellow) Enters an aide service request.
(green) Enters a nurse service request.
(red)
Enters an urgent staff emergency
request.
(gray)
Initiates paging operations.
(blue) Displays functions menu for
programming additional features.
(orange) Initiates nurse-follower operations.
(white) Press to speak, press again to
listen when using loudspeaker for
communications.
(white) Changes all function keys to the
specifically labeled alpha characters.
monitor (blue)
www.tektone.com
Enables room monitoring.
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
capture (blue)
view
(blue)
priority (blue)
Allows for zone assignment.
Allows review of station information.
Activates menu for changing room
priority.
tone on (white) Turns normal call-tone on/off.
+
–
>
<
(white) Adjusts volume up
(white) Adjusts volume down
(white)Space
(white)Backspace
Additional programmable functions include: set station
priority, set privacy, zone assignments, view station
configurations, set password, reset time of day, time display,
set overtime, transfer for master stations, paging, system
configuration, and system re-initialization.
The master station is equipped with volume control for
adjustment of incoming voice communication; tone on/off
switch for controlling the lowest priority tone signal only;
call light with indication of low, medium and high priority
calls; CE fault LED; busy and alphanumeric keypad used
for entry of station numbers, zone and station selection,
and for selection of menu functions.
The master station also provides a serial port for connection
to ancillary devices. System programming is accessible
to the user via the keys on the master station or via the
keyboard port which accepts any standard USB keyboard.
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
Dimensions: Capacity:
Console: 13.2″ × 9″ × 5.5″
(335mm × 229mm × 140mm)
Address up to 512 single or dual
patient stations in up to 16 zones
Environments:26°C, relative humidity should not
exceed 80%
Terminations:
Plug-on connectors
System Protection:
All outputs are protected by selfresetting current-limiting devices
Current Requirements: 0.75A @ 16VAC
Required Components
Central Equipment:
NC351(A)—4 masters/256 stations
–or–
NC351(A)/2—8 masters/512 stations
IR300C-series Patient Stations
NC351A and NC351A/2 Central Equipment
IL696 Specification Sheet
Tek-CARE® NC300II Nurse Call System • Rev. 10 - 04/2013
Features
•
•
•
•
•
•
UL® 1069 Listed
Supports 4 masters and 256 stations (NC351A) or
8 masters and 512 stations (NC351A/2)
Plug-in connections with multiplexed common wiring
Self-contained electronics and power supply
Serial port for Tek-PAGING® radio pocket paging system
Serial port for Tek-MMARS ® 300 management
monitoring and reporting system
NC351A and NC351A/2 Central Equipment
Description
A major component of the Tek-CARE® NC300II Nurse Call
System, the NC351A and NC351A/2 central equipment
supply the central point for system component integration.
The NC351A supports up to 4 master consoles and 256
patient/staff stations, while the NC351A/2 supports up to
8 master consoles and 512 patient/staff stations.
The NC351A and NC351A/2 house the necessary firmware
and control circuits to operate all the Tek-CARE® NC300II
system functions. From a single durable metal enclosure,
the central equipment supports 2 isolated serial ports, field
station ports, master console ports, and power requirements
for up to 256 patient, staff and duty stations (512 with
external PK305A Supply).
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
Dimensions: Capacity:
15.25″ × 13″ × 6.25″
(387 mm × 330 mm × 159 mm)
up to 4 (NC351A) or 8 (NC351A/2)
masters, up to 256 (NC351A) or 512
(NC351/2A) stations
Power Input: 115 VAC, 60Hz, 140W
Environments: 26°C, relative humidity maximum 80%
Terminations: Plug-on connectors
System
Supervision:
Continuous polling of all stations. Stations
or wiring faults displayed at masters.
System
All outputs are protected by self-resetting
Protection:
current-limiting devices.
Audio:
Up to 4 (NC351A) or 8 (NC351A/2)
audio paths provided. Amplifiers rated
at 20 watts. Overload and short circuit
protection included.
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PK305A Supplemental Power Supply
IL698 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 6 - 07/2013
Features
•
•
•
Plug-in connectors for easy connection to NC351A/2
Central Equipment
Provides additional power to the NC351A/2 Central
Equipment to support 512 patient stations
UL® 1069 Listed
Description
The PK305A Supplemental Power Supply is used to provide
additional power to the NC351A/2 Central Equipment to
support 512 patient stations.
The supplemental power supply is housed in a durable
metal enclosure with provisions for plug-in field wiring
connections.
Listings
UL® 1069 listed; UL® File Number E71599.
PK305A Supplemental Power Supply
Specifications
Dimensions: Console: 10.75″ × 8.75″ × 3.25″
115 VAC, 60Hz, 140W
Power Input: Environments:26°C, maximum 80% relative humidity
Plug-in connectors
Terminations:
System Protection: In-line fuses are provided on primary
input
Required Components
Tek-CARE®NC300II Nurse Call System
with NC351A/2 Central Equipment
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
NC385 Tek-BRIDGE® with Tek-MMARS®300
Management Monitoring and Reporting System
IL504 Specification Sheet
Tek-CARE® NC300/II Nurse Call Systems • Rev. 18 - 03/2016
Features
•
•
•
•
•
•
•
Proprietary server with built-in speaker, flat panel
monitor, keyboard and mouse.
Includes reporting, event monitor, and one remote
event monitor.
Includes paging software module for automatic and
manual pages to staff.
Includes software module to connect one Tek-CARE®
NC300/II nurse call system. Additional licenses are
available separately.
Optional software modules include up to eight
additional remote event monitors, Tek-CARE®400
and Tek-CARE®500ES connections, and email output.
3 network interface cards/ports provide connections
to a LAN and a Tek-CARE®400 central equipment.
6 RS232 serial ports provide connections for a paging
transmitter, plus up to 4 directly connected Tek-CARE®
NC300/II nurse call systems.
Description
NC385 Tek-BRIDGE® with Tek-MMARS®300
•
The NC385 Tek-BRIDGE® with Tek-MMARS®300 software
is the management monitoring and reporting system for
the Tek-CARE® NC300/II nurse call system.
The Tek-BRIDGE® server also supplies ports for a local
area network (LAN) or the facility’s LAN, and the radio
pocket paging system’s paging transmitter. Each Tek-CARE®
NC300/II nurse call system connects either to one of the
server’s unused serial ports, or to the LAN using a CT601
Serial to IP Converter.
•
The Tek-MMARS®300 software includes these modules:
•
•
LS301 Tek-CARE® NC300/II Connection enables
Tek-MMARS®300 to monitor one nurse call system.
Purchase an LS302 license for each additional
Tek-CARE® NC300/II system to be monitored.
LS576 Event Monitor displays all activity of monitored
Tek-CARE® NC300/II systems on the Tek-BRIDGE®.
•
LS586 Remote Event Monitors display Tek-CARE®
NC300/II nurse call system activity. This software
may be installed on up to nine of the facility’s personal
computers (PCs) running Microsoft® Windows®, and
networked to the Tek-BRIDGE® server. One license is
included with the NC385. Purchase additional licenses
separately as one license (LSMON1) or a block of five
licenses (LSMON5).
Tek-CARE® Reporting System records all nurse call
activity on the Tek-BRIDGE®, and provides custom
reports on this activity for followup review and
evaluation by management. Reports are viewed using
a web browser (such as Microsoft® Internet Explorer).
They may also be printed, saved as HTML files for later
viewing in a web browser, or saved as CSV files for
viewing in a spreadsheet program (such as Microsoft®
Excel).
When connected to a facility’s LAN, the Tek-CARE®
Reporting System can be accessed by any networked
computer via a web browser.
LS450 Config Tool is used to set up and customize
the Tek-MMARS®300 software, and to save the system
configuration.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by DQS to the ISO 9001 standard. (Reference #10001510.)
Optional Software
Required Components
Available separately, these optional software modules run
on the Tek-BRIDGE® server, adding ancillary features to the
Tek-CARE® NC300/II nurse call system.
Tek-CARE®NC300II nurse call system
–or–
Tek-CARE®NC300 with Tek-NIOS software version 1.96,
3.06 or 4.06. (Earlier versions and “S” versions of the
Tek-NIOS software are not compatible. To determine a
system’s software version, push the menu button on the
nurse master station.The menu version is displayed at
the top of the Menu box, for example: TEK-NIOS 1.96.)
•
•
•
•
•
LS453 Email Output enables email addresses to be
added as staff devices. The software sends calls and
other events via email to the assigned staff members.
Events can also be sent to anything that has an email
gateway—for example, sending an SMS message
to a cellphone. The LS453 Email Output software
also enables the Tek-CARE® Reporting System to
automatically send reports via email.
LS400 Tek-CARE®400 Connection Software enables
the Tek-BRIDGE® to monitor Tek-CARE®400 nurse call
system activity.
LS501ES Tek-CARE®500 allows the Tek-BRIDGE® to
act as the master station for our wireless emergency
call system. It also enables use of the NC502 PC-Alarm
Panel to auto dial a central monitoring facility.
LS601 Tek-ALERT® Integration Manager imports
events from other systems (such as nurse call and fire
alarm) and displays them on Tek-CARE® NC300/II
event monitors.
LS701K Tek-CARE®700 Software plus Starter Kit
enables the Tek-BRIDGE® to display and store events
from the wander management system.
Optional Components
Software:
• LS302 One Tek-CARE® NC300/II Connection License
• LS453 Email Output Software
• LSMON1 One Remote Event Monitor License
• LSMON5 Five Remote Event Monitor Licenses
• LS400 Tek-CARE®400 Nurse Call System Connection
(requires Tek-CARE®400 hardware)
• LS501ES Tek-CARE®500 EchoStream Connection
Software (requires Tek-CARE®500 hardware)
• LS601 One Tek-ALERT ® Foreign Device License
(requires Tek-ALERT® hardware)
• LS701K Tek-CARE®700 Software plus Starter Kit
(requires Tek-CARE®700 hardware)
Hardware:
• CT601 Serial to IP Converter
Tek-PAGING® UHF Pocket Paging System:
• NC365B Paging Transmitter
• NC397A Alphanumeric Pagers
NOTE: The Tek-BRIDGE® with Tek-MMARS®300 software is ancillary
to the nurse call system and must not be used as the primary means
of annunciation for the Tek-CARE®NC300/II nurse call system. To
respond to calls, staff must have access to features that are available
only on the nurse call system’s master stations.
CT601 Serial to IP Converter
IL918 Specification Sheet
Tek-ALERT™ Alert Integration Manager, Tek-MMARS®300 System • Rev. 2 - 07/2013
Description
The CT601 Serial to IP Converter is used to connect a TekCARE® NC300/II nurse call system or a monitored foreign
system to a facility’s LAN. Events and messages from the
connected system are sent across the LAN to either an
NC385 Tek-MMARS®300 or a Tek-BRIDGE™ running TekALERT™ Software.
The CT601 Serial to IP Converter becomes a virtual COM
port on the Tek-MMARS®300 or Tek-BRIDGE™, and provides
long distance communication between it and the monitored
serial device. The CT601 eliminates the otherwise expensive
long-distance cabling required for these connections.
Configuring the CT601 is accomplished using software
provided with the device. The unit can be wall or DIN-rail
mounted. The CT601 incorporates surge protection.
Specifications
Serial Port:
Male DB9 serial port, 1200 bps to 115
Kbps, hardware flow control
RS232 Signals: Rx, Tx, RTS, CTS, DTR, DSR, GND, DCD
Network Port: 10Base-T Ethernet
Protocols: ARP, IP/ICMP, TCP/IP, Telnet, DHCP client,
DNS
Security:
User ID & Password, IP Address filtering
Power:
7.5 ~ 15VDC, 300mA 7.5V
Environment: 90% non-condensing humidity;
0˚C ~ 50˚C operating temperature.
Dimensions:3.9″×2.8″×1.0″ (100mm×72mm×25mm)
Certification: FCC (A), CE, MIC
CT601 Serial to IP Converter
Required Components
One of the following:
• NC385 Tek-MMARS®300 Management Monitoring and
Reporting System
• NC470 Tek-BRIDGE™ with Tek-CARE®400 Software
• NC501ES Tek-BRIDGE™ with Tek-CARE®500 Software
• NC601 Tek-BRIDGE™ with Tek-ALERT™ Software
Plus one for each monitored system:
• LS302 Tek-CARE® NC300/II Connection License –or–
• LS601 Tek-ALERT™ Software License
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PK250B Power/Standby UPS
IL472 Specification Sheet
Tek-CARE® and Tek-ALERT® Systems • Rev. 17 - 05/2014
Description
The PK250B Uninterruptable Power Supply provides
uninterrupted power in the event of utility power failures,
brownouts and blackouts. It is for use with the Tek-CARE®
NC300II Nurse Call System and the Tek-BRIDGE® Server.
Its economical cost and compact size will enhance the
operation of your system.
When power is interrupted, the PK250B immediately cuts in
to provide square wave power for the Tek-CARE® NC300II
Central Equipment or the Tek-BRIDGE® until an emergency
generator is activated or power is restored. In the event of
a prolonged power outage, the PK250B will provide power
to the limit of its battery. The green indicator on the front
panel indicates power line output when steady, and battery
output when flashing. The unit also provides an audible
tone during battery output operation.
Other features include output receptacles for connection
of equipment, and power cord for standard 120 volt, 15
amp power circuit.
PK250B Power/Standby UPS
Specifications
Listings
Output Rating
Wattage:
Voltage:
Frequency:
Waveform:
UL® 1778 Listed and cUL® Listed to CSA C22.1.
Input
Voltage:
Frequency:
300 W
115 VAC
60 Hz
Stepped square wave
115 VAC
60 Hz
Backup Operation
At full load:
2 minutes
At half load:
9 minutes
Transfer time: 6 ms typical
Battery
Type:
Recharge:
Auto cut-off:
Size
Overall:
Weight
Required Components
Tek-CARE®NC300II Nurse Call System
–or–
Tek-BRIDGE® Server
Maintenance-free 12V 5Ah
8 hours
Yes
10.9″ × 3.5″ × 10.8″
13.5 lbs. (6 kg)
Operating temperature range is 0°C–40°C.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IR300E Multipurpose Station
IL491 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 10 - 07/2013
Features
•
•
•
•
•
•
UL® 1069 Listed
Supervised
Microprocessor design
Solid-state multiplex circuitry
Simplified multiplex wiring using plug-in connectors
Flame-retardant (UL® 94V-0) plastic faceplate
Description
The IR300E Multipurpose Station provides for audible,
visual and digital communications to and from the
Tek-CARE®NC300II central equipment (CE) when receiving
or sending signals to remotely connected devices. The
IR300E includes inputs for contact closures and all types of
emergency call switches for these systems. It also provides
outputs to speakers or paging amplifiers, and dome or zone
lamps. It does not have the controls, LEDs, or speakers
normally associated with patient, staff or duty stations.
Some practical applications for the IR300E are as follows:
provides input for code emergency call stations in an ICU;
a PA amplifier interface for the nurse call master; a control
point for exit/entry doors and narcotics cabinets; provides
input for bath emergency call switches in public bath areas;
provides input/output to a remote door intercom station
for access control.
Listings
UL 1069 listed; UL File Number E71599.
®
®
IR300E Multipurpose Station
Specifications
Dimensions:
Construction:
Connections:
Housing:
Wiring:
Height: 4.5″
(114 mm)
Width:
6.375″ (162 mm)
Depth:
2.75″
(70 mm)
Projects:
0.375″ (9.5 mm)
Flame retardant (UL® 94V-0) plastic
Plug-in type
Steel City #H3BD with Steel City #3GC plaster ring, or exact equal. Horizontal
mount.
Multiplex.
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IR310E Staff Station
IL493 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 11 - 07/2013
Features
•
•
•
•
•
•
•
•
•
UL® 1069 Listed
Supervised
Microprocessor design
Solid-state multiplex circuitry
Field programmable for staff or patient station
operation
Programmable call priority
All stations interchangeable
Simplified multiplex wiring using plug-in type
connectors
Attractive flame retardant (UL® 94V-0) plastic panel
IR310E Staff Station
Description
The IR310E Staff Station provides reliable two-way,
hands-free communication between the nurse master
station and utility room, examination rooms, kitchen, or
any location where staff members may need to originate
calls or where cord sets are not required. Calls are placed
by use of a momentary push button. Solid-state lights on
the panel indicate that the call has been placed and that
the intercom is in use.
As an additional feature, the staff station is field
programmable to function as a patient station, if desired.
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
Dimensions:
Construction:
Finish:
Speaker:
Connections:
Housing:
Height: 4.5″
(114 mm)
Width: 6.375″
(162 mm)
Depth:
2.75″
( 70 mm)
Projects:
0.5″
(12.7 mm)
Flame retardant (UL® 94V‑0) plastic
Light texture, white
2.5″ × 2.5″ (64 mm × 64 mm)
Plug-in type
Steel City #H3BD with Steel City #3GC
plaster ring or exact equal. Horizontal
mount.
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IR315F Duty Station
IL495 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 12 - 07/2013
Features
•
•
•
•
•
•
UL® 1069 Listed
Supervised
All stations interchangeable
Simplified multiplex wiring using plug-in connectors
Attractive, flame retardant (UL® 94V-0) plastic faceplate
Selective annunciation of call priority
Description
The IR315F Duty Station provides reliable two-way, handsfree communication between the nurse master station and
utility rooms, examination rooms, kitchen, or any location
where a nurse may perform a function out of hearing or
sight of the master station. Calls are placed to the master
station by use of a momentary push button. When the call
is answered at the master station, the red in-use indicator
on the duty station lights, and the nurse may then carry on
a conversation from the remote location.
A call originating from a patient station is indicated both
audibly and visually at the duty station to which it is
assigned. Normal calls are indicated by a steady illumination
of the yellow call-placed indicator and a slow repeating
tone. Emergency/bath calls cause alternate flashing of the
yellow call-placed and red emergency indicators and a rapid
repeating tone. Code calls function the same as emergency
calls, but at twice the pulse rate. Dome light indication is
also provided. All calls initiated from stations other than
the duty station can be reset only at the master station or
point of origin.
Listings
UL 1069 listed; UL File Number E71599.
®
®
IR315F Duty Station
Specifications
Dimensions:
Construction:
Speaker:
Connections:
Housing:
Height:
4.5″
(114 mm)
Width:
6.375″
(162 mm)
Depth:
2.75″
(70 mm)
Projects: 0.5″
(12.7 mm)
Flame retardant (UL® 94V-0) plastic
2.5″ × 2.5″ (64 mm × 64 mm), 60 ohm
Plug-in type
Steel City #H3BD with Steel City #3GC
plaster ring, or exact equal. Horizontal
mount.
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IR318B Single Patient Station
IL689 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 9 - 07/2013
Features
•
•
•
•
•
•
•
•
UL® 1069 Listed
Supervised
Microprocessor design
Solid-state multiplex circuitry
All stations are interchangeable
Simplified multiplex wiring using plug-in connectors
Flame-retardant (UL® 94V-0) plastic faceplate
Six-foot washable TPU plastic pull cord
Description
The IR318B single patient station provides reliable twoway hands-free communication between patient and nurse,
and provides for programmable privacy, paging and patient
priority from the master station.
Calling the nurse is accomplished by pulling on the end of
the pull cord. Solid-state LEDs on the panel indicate the
call has been placed and the intercom is in use.
Listings
UL® 1069 listed; UL® File Number E71599.
IR318B Single Patient Station
Specifications
Dimensions:Height: 4.5″
(119 mm)
Width:
6.375″
(162 mm)
Depth:2.75″
(70 mm)
Projects:
0.5″
(12.7 mm)
Construction: Flame-retardant (UL® 94V-0) plastic.
Speaker:2.5″ × 2.5″ (64 mm × 64 mm)
Connections: Plug-in type
Housing:
Steel City #H3BD with Steel City #3GC
plaster ring, or exact equal. Horizontal
mount.
Wiring:Multiplex
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IR319B Single and IR320B Dual Patient Stations
IL713 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 4 - 07/2013
Features
•
•
•
•
•
•
UL® 1069 Listed
Supervised
All stations interchangeable
Simplified multiplex wiring and plug-in connectors
Input available for Hill-Rom® Sidecom® beds or remote
call-in button applications
Uses pillow speakers or push-button call cords
Description
The IR319B Single Patient Station provides reliable twoway, hands-free communication between patient and
nurse, and provides for programmable privacy, paging
and patient priority from the master station. Calling the
nurse is accomplished by pressing the call cord button or
the area indicated nurse on the pillow speaker. Solid state
LEDs on the panel indicate the call has been placed and
the intercom is in use.
If the pillow speaker or call cord is accidentally removed
from its receptacle, the nurse master station registers a
non-cancellable cord out call. The call may be reset only by
returning the call cord or pillow speaker to its receptacle.
The IR320B Dual Patient Station has the same specifications
and functions as the IR319B, plus an additional
independently operating call cord/pillow speaker
receptacle and a second call-placed indicator.
Listings
UL® 1069 listed; UL® File Number E71599.
IR319B Single and IR320B Dual Patient Stations
Specifications
Dimensions:
Height:
4.5″
(114 mm)
Width:
6.375″
(162 mm)
Depth:
2.75″
(70 mm)
Projects:
1.0″
(25 mm)
above mounting surface
Construction: Flame-retardant (UL® 94V-0) plastic
faceplate
2.5″ × 2.5″ (64 mm × 64 mm)
Speaker:
Connections: Plug-in type
Housing: Steel City #H3BD and #3GC plaster ring,
or exact equal. Horizontal mount.
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF401A Oxygen-Safe Call Cord
IL767 Specification Sheet
Tek-CARE®400 P5 Nurse Call System • Rev. 5 - 11/2012
Features
•
•
•
•
•
UL® 1069 Listed
Oxygen safe
Impact resistant and waterproof
Small, lightweight pendant fits comfortably in hand
Bed clips included
Description
The TekTone® SF401A single call cord is for use with
the Tek-CARE ® 400, Tek-CARE ® 400 P5, and TekCARE®NC300II nurse call systems, where 8-pin DIN jacks
are used. The SF401A is an oxygen safe call cord that
is certified safe for use by all patients receiving oxygen
therapy. This waterproof and impact resistant call cord
has a completely sealed push button.
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
Connection:
Construction:
Dimensions:
SF401A Oxygen-Safe Call Cord
Required Components
for the Tek-CARE®400 nurse call system:
IR411, IR411EPB, IR412, IR412EPB patient stations
for the Tek-CARE®400 P5 nurse call system:
IR411P5, IR412P5 patient stations
for the Tek-CARE®NC300II nurse call system:
IR319B, IR320B patient stations
8-pin DIN plug
SPST1-A switch
Integral brass, fine silver contacts
1 AMP, 115 watt maximum, AC
7' (2.13 m) long polyvinyl-covered cord
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF401G Call Cord
IL829 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 3 - 11/2014
Features
•
•
•
•
•
Sensitive touch activation
Oxygen tent safe
Ideal for geriatric use
Momentary, air-activated switch
Adjustable bed clip
Description
TekTone® SF401G Call Cords are used with 8-pin DIN jack
patient stations, including Tek-CARE®400 IR411- and
IR412-series patient stations, Tek-CARE®400 P5 IR411P5
and IR412P5 patient stations, and Tek-CARE® NC300II
IR319- and IR320-series patient stations.
The SF401G Call Cord is specifically designed for use in
geriatric and oxygen environments, but may also be used in
general applications. To place a call, the patient need only
apply slight pressure to the squeeze bulb using any part of
the body. The force of air through the vinyl tubing closes a
switch in the plug, thus placing a call. An adjustable bed clip
is supplied for the SF401G and is attached to the call cord.
Specifications
Switch:
Plug:
Cord:
Electrical:
Dimensions:
Momentary contact, air activated
Stainless steel 8-pin DIN
Vinyl tubing
Rated current 1/3 amp, 24 VAC-VDC
Cord:6′ long
(1.83 m)
Bulb: 1.5" diameter (38 mm)
SF401G Call Cord
Required Components
for Tek-CARE®NC300II Nurse Call Systems:
IR319- or IR320-series Patient Stations
for Tek-CARE®400 Nurse Call Systems:
IR411- or IR412-series Patient Stations
for Tek-CARE®400 P5 Nurse Call Systems:
IR411P5 or IR412P5 Patient Stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF401GP Geriatric Pad Call Cord
IL989 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 0 - 10/2015
Features
•
•
•
•
•
Non-powered pneumatically operated switch
Large 3.5″ diameter pad for easy activation
Oxygen safe
Latex free
Includes security clip
Description
The SF401GP Geriatric Call Cord is a round, flattened soft
vinyl pad with gripping feet to prevent it from slipping off
the patient’s bed. The SF401GP works well for residents
who have limited hand dexterity. It has a mechanical switch
that is activated by slight pressure from the hand, arm or
body, or can even be positioned under the resident’s chin.
The SF401GP has an 8-foot cord and an 8-pin DIN plug.
Also available is the SF401GP/10 with a 10-foot cord.
Listings
SF401GP Geriatric Pad Call Cord
UL® 1069 Listed and cUL® Listed to CSA C22.2
No. 205; UL® File Number E61802.
Specifications
Cord Length: 8′ (2.44 meters)
Connections: 8-pin DIN plug
Switch:
Momentary contact
Electrical Rating:24 volts AC or DC at 0.33 amps
Required Components
for Tek-CARE®NC300II Nurse Call Systems:
IR319B or IR320B Patient Station
SF401EX Strain Relief Cable
for Tek-CARE®400 Nurse Call Systems:
IR411-series or IR412-series Patient Station
SF401EX Strain Relief Cable
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF401EX Strain Relief Cable
IL886 Specification Sheet
Tek-CARE®400 P5 Nurse Call System • Rev. 2 - 11/2012
Features
•
•
•
•
UL® 1069 Listed
Molded connectors
Clean break-away action
Increases life of call cords, pillow speakers and room
stations
Description
The SF401EX Strain Relief Cable is used to increase the life
span of call cords, pillow speakers and room stations. The
8-pin DIN cable is placed between the call cord or pillow
speaker and the room station. If the call cord or pillow
speaker is attached to a bed and then the bed is pulled away,
the strain relief cable disconnects. This prevents damage to
both the room station and the call cord or pillow speaker.
Listings
UL® 1069 listed; UL® File Number E71599.
SF401EX Strain Relief Cable
Specifications
Required Components
Length:
Connections:
Construction:
for Tek-CARE®NC300II Nurse Call Systems:
IR319- or IR320-series Patient Stations
SF4-series Call Cords or Pillow Speakers
6" (153 mm)
8-pin DIN plugs
24-gauge polyvinyl-covered cord with
molded ends
for Tek-CARE®400 Nurse Call Systems:
IR411- or IR412-series Patient Stations
SF4-series Call Cords or Pillow Speakers
for Tek-CARE®400 P5 Nurse Call Systems:
IR411P5 or IR412P5 Patient Stations
SF4-series Call Cords or Pillow Speakers
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF41-series Pillow Speakers
IL824 Specification Sheet
Tek-CARE®400 P5 Nurse Call System • Rev. 7 - 11/2012
Features
•
•
•
•
•
•
•
•
SF41ZKL (standard)
for Zenith Digital TVs, includes
direct channel access, digital volume
controls and 2 light controls
UL® 1069 Listed
Oxygen safe
Fully sealed for easy cleaning
Flame-retardant (UL® 94V-0) high-impact plastic
Long lasting switches surpass UL®1069 requirements
8-foot cord with DIN plug and security clip
Provides pillow speaker intercom and light control
(when used with the appropriate modules)
Special Order: RCA or Philips TV controls, analog
volume control, 10-foot cord, earphone jack
Description
SF41-series Pillow Speakers are used with Tek-CARE®400
IR411- and IR412-series patient stations, Tek-CARE®400
P5 IR411P5 and IR412P5 patient stations, and Tek-CARE®
NC300II IR319- and IR320-series patient stations.
The pillow speakers provide call placement to the master
station, as well as remote control for hospital-grade digital
TVs (Zenith is standard; RCA and Philips are available
by special order). When used in conjunction with the
appropriate TekTone® pillow speaker modules, the pillow
speakers also supply two-way, hands-free communication
with the master station, and automatic silencing of TV audio
while the intercom is in use.
Push buttons are provided to call the nurse, plus TV power,
channel selection up/down, Closed Captioning, mute and
two light controls. Buttons are also provided for volume
up/down (analog volume controls are available by special
order), and for direct channel access (in some models).
Listings
UL® 1069 listed; UL® File Number E165723.
SF41ZDL (standard)
fo r Z e n i t h D i g i t a l T Vs ,
includes channel up/down,
digital volume controls and
2 light controls
SF41-series Pillow Speakers
Required Components
Hospital-grade Zenith Digital TV
for Tek-CARE®NC300II Nurse Call Systems:
IR319- or IR320-series Patient Station
PM311C or PM312C Pillow Speaker Intercom Module
PM321B or PM322B Light Control Module
for Tek-CARE®400 Nurse Call Systems:
IR411- or IR412-series Patient Station
PM421 or PM422 Pillow Speaker & Bed Interface Module
for Tek-CARE®400 P5 Nurse Call Systems:
IR411P5 or IR412P5 Patient Station
PM421P5 or PM422P5 Pillow Speaker Module
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PM311C, PM312C
Pillow Speaker Intercom Modules
IL429 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 12 - 07/2013
Features
•
•
•
UL® 1069 Listed (when used with IR319B, IR320B,
IR311E or IR312E patient stations)
Provides hands-free pillow speaker intercom
Mounts directly to patient stations
Description
The PM311C and PM312C Pillow Speaker Intercom Modules
are for use with the Tek-CARE®NC300II Nurse Call System
using IR319B, IR320B, IR311E or IR312E patient stations
and SF41ZDL, SF41ZKL, SF31ZDL or SF31ZKL pillow
speakers. The modules provide for reliable, two-way handsfree communication from the pillow speaker to the master
station. Calling the nurse is accomplished by pressing the
push button identified with the nurse symbol. Call placed
assurance is indicated at the patient station.
If the pillow speaker is removed from its receptacle during
communication, the call is automatically transferred to
the patient station speaker. If the pillow speaker or call
cord is removed from its receptacle when not in use, a
non-cancellable cord-out call is registered at the master
station. Standard SF401-series or SF311-series call cords
may be used without removing modules. Communication
is automatically transferred to the patient station speaker.
Listings
UL® 1069 listed; UL® File Number E71599.
PM312C Pillow Speaker Intercom Module
Specifications
Dimensions:Width: 5″
(127 mm)
Depth:1.875″
(47.6 mm)
Height:0.5″
(12.7 mm)
Construction: PC board with quality electronic
components
Mounting:
Plug-in connectors mounted to patient
station
Wiring:Multiplex
Required Components
Tek-CARE®NC300II Nurse Call System
IR319B or IR320B Patient Stations
with SF41ZDL or SF41ZKL Pillow Speakers
–or–
IR311E or IR312E Patient Stations
with SF31ZDL or SF31ZKL Pillow Speakers
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PM321B, PM322B Multiplex Lamp Control Modules
IL461 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 11 - 07/2013
Features
•
•
•
UL® 1069 Listed (when used with IR319B, IR320B,
IR311E or IR312E patient stations)
Provides on/off control of two lamp sources
Mounts directly to patient station or to pillow speaker
intercom module (if used).
Description
The PM321B and PM322B Multiplex Lamp Control Modules
are for use with the Tek-CARE®NC300II Nurse Call System
using IR319B, IR320B, IR311E or IR312E patient stations
and SF41ZDL, SF41ZKL, SF31ZDL or SF31ZKL pillow
speakers. The modules provide on/off control through the
pillow speaker of two lamp sources, typically the overhead
room lamp and a separate reading lamp. Circuitry for low
voltage lamp switching is provided by others. Turning the
lamps on or off is accomplished by pressing the appropriate
push buttons on the pillow speaker.
The lamp control module can be mounted directly to the
patient station, or can be mounted to the station’s PM311C
or PM312C pillow speaker intercom module. Standard
SF401-series or SF311-series call cords may be used
without removing modules.
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
PM322B Multiplex Lamp Control Module
Required Components
Tek-CARE®NC300II Nurse Call System
IR319B or IR320B Patient Stations
with SF41ZDL or SF41ZKL Pillow Speakers
–or–
IR311E or IR312E Patient Stations
with SF31ZDL or SF31ZKL Pillow Speakers
Dimensions:Width: 5″
(127 mm)
Depth:1.625″
(41 mm)
Height:0.5″
(13 mm)
Construction: PC board with quality electronic
components
Mounting:
Plug-in connectors mounted to patient
station or to PM311C/PM312C (if used)
Wiring:Multiplex
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
IR311E Single & IR312E Dual Patient Stations
IL494 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 9 - 07/2013
Features
•
•
•
•
•
•
UL® 1069 Listed
Fully supervised
All stations interchangeable
Simplified multiplex wiring & plug-in type connectors
Input available for Hill-Rom® Sidecom® beds or remote
call-in button applications
Uses pillow speakers or push button call cords
Description
The IR311E Single Patient Station provides reliable twoway, hands-free communication between patient and
nurse, and provides for programmable privacy, paging, and
patient priority from the master station. Calling the nurse
is accomplished by pressing the call cord button or the area
indicated nurse on the pillow speaker. Solid-state LEDs on
the panel indicate the call has been placed and the intercom
is in use. If the pillow speaker or call cord is accidentally
removed from its receptacle, the nurse master station
registers a non-cancelable call with cord out indication.
The call may be reset only by returning the call cord or
pillow speaker to its receptacle. The pillow speaker also
provides television control.
The IR312E Dual Patient Station has the same specifications
and functions as the IR311E, and has an additional
independently operating call cord/pillow speaker
receptacle, and another call-placed indicator.
Listings
UL® 1069 listed; UL® File Number E71599.
IR312E Dual Patient Station
Specifications
Dimensions:
Height:
4.5″
(114 mm)
Width: 6.375″
(162 mm)
Depth: 2.75″
(70 mm)
Projects:
0.5″
(12.7 mm)
above mounting surface
Construction: Flame-retardant (UL® 94V-0) plastic
faceplate
Speaker:
2.5″ × 2.5″ (64 mm × 64 mm)
Connections: Plug-in type
Steel City #H3BD and #3GC plaster ring,
Housing:
or exact equal. Horizontal mount.
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF31ZDL, SF31ZKL Pillow Speakers
IL930 Specification Sheet
Tek-CARE® NC300II Nurse Call Systems • Rev. 1 - 04/2013
Features
•
•
•
•
•
•
•
•
UL® 1069 Listed
Oxygen safe
Fully sealed for easy cleaning
Flame-retardant (UL® 94V-0) high-impact plastic
Long lasting switches surpass UL®1069 requirements
8-foot cord with 8-pin 8P8C plug and security clip
Provides pillow speaker intercom and light control
(when used with the appropriate modules)
Special Order: RCA or Philips TV controls, analog
volume control, 10-foot cord, earphone jack
SF31ZKL for Zenith Digital TVs,
includes direct channel access,
digital volume controls and
2 light controls
Description
The SF31ZDL and SF31ZKL Pillow Speakers are used with
Tek-CARE® NC300II IR311-series and IR312-series patient
stations. The pillow speakers provide call placement to the
master station, as well as remote control for hospital-grade
digital TVs (Zenith is standard; RCA and Philips are available
by special order). When used in conjunction with the
appropriate TekTone® pillow speaker modules, the pillow
speakers also supply two-way, hands-free communication
with the master station, and automatic silencing of TV audio
while the intercom is in use.
Push buttons are provided to call the nurse, plus TV power,
channel selection up/down, Closed Captioning, mute and
two light controls. Buttons are also provided for volume
up/down (analog volume controls are available by special
order), and for direct channel access (in some models).
SF31ZDL for Zenith Digital TVs,
includes channel up/down,
digital volume controls and
2 light controls
SF31ZDL, SF31ZKL Pillow Speakers
Required Components
Hospital-grade Zenith Digital TV
IR311- or IR312-series Patient Station
PM311C or PM312C Pillow Speaker Intercom Module
PM321B or PM322B Light Control Module
Listings
UL® 1069 listed; UL® File Number E71599.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LI380 Corridor Dome Light
IL519 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 9 - 07/2013
Features
•
•
•
•
•
UL® 1069 Listed
Easy bulb replacement
Snap-in colored lens sections
Excellent visibility
Interchangeable lenses
Description
TekTone®’s LI380 Corridor Dome Light provides visual
indication of calls originating from patient, staff, duty, bath,
emergency, presence and code call stations. Patient priority
and staff presence are indicated by amber, green, red and
white colored lenses. When used with any room station or
IR310E Staff Station, the LI380 indicates all calls originated
from that room. When used with the IR300E Multipurpose
Station or the IR315F Duty Station, the LI380 may be used
as a zone light to indicate all calls from the zones to which
those stations are assigned.
Listings
UL® 1069 listed; UL® File Number E71599.
LI380 Corridor Dome Light
Required Components
Tek-CARE®NC300II Nurse Call System
Specifications
Dimensions:Height: 4.5625″
(115 mm)
Width:4.5625″
(115 mm)
Depth:3.3″
(84 mm)
Bulbs:
#1829 28v. 70mA
Power:
24 VDC, 205 mA
Finish:
Base is flame-retardant (UL® 94V-0)
plastic. Amber, green and red lenses are
flame-retardant translucent polycarbonate plastic. White lens is flame-retardant
translucent polystyrene plastic.
Housing:
Fits standard dual gang box with either
single gang ring or dual gang ring.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LI380LED Corridor Dome Light
IL913 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 2 - 07/2013
Features
•
•
•
•
•
UL® 1069 Listed
Long lasting LEDs
No bulb replacement required
Plug-in connector supplied
Attractive flame-retardant plastic panel
Description
TekTone®’s LI380LED Corridor Dome Light provides visual
indication of calls originating from patient, staff, duty, bath,
emergency, presence and code call stations. Patient call
priority and staff presence are indicated by amber, green,
red and white colored LEDs. When used with any room
station or IR310E Staff Station, the LI380LED indicates
all calls originated from that room. When used with the
IR300E Multipurpose Station or the IR315F Duty Station,
the LI380LED may be used as a zone light to indicate all
calls from the zones to which those stations are assigned.
Listings
UL® 1069 listed; UL® File Number E71599.
LI380LED Corridor Dome Light
Specifications
Dimensions:Height:
4.8″
(122 mm)
Width:5.0″
(127 mm)
Depth:2.755″
( 70 mm)
LEDs:
2V 20mA per LED
Power:
24 VDC
Finish:
Base is flame-retardant (UL ® 94HB)
plastic. Lens is flame-retardant translucent
polycarbonate plastic.
Housing:
Fits standard one- or two-gang box with
one- or two-gang ring.
Required Components
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LI381 Corridor Light & LI382 Corridor Zone Light
IL518 Specification Sheet
Tek-CARE® Nurse Call & CM800 Emergency Call Systems • Rev. 18 - 07/2013
Features
•
•
•
•
•
UL® 1069 Listed
Snap-in lens
Easy bulb replacement
Attractive flame-retardant HIPS30 plastic panel
Excellent visibility
Description
The LI381 Corridor Light and LI382 Corridor Zone Light
provide visual indication of calls when used with a nurse
call system, emergency call system, or similar application.
Emergency calls are indicated by the rapid flashing of the
LI381 Corridor Light associated with the calling station.
Normal calls produce a steady illumination.
The LI382 Corridor Zone Light indicates an emergency call
by the rapid flashing of the red-covered bulb associated with
the zone or area from which an emergency call has been
placed. Normal calls are indicated by a steady illumination
of the white bulb. If an emergency and normal call are
placed in the same zone at the same time, the red covered
bulb connected to the emergency station will flash rapidly,
while the white bulb associated with the normal call will
maintain a steady illumination.
The standard mounting plate accommodates both models
and features an easily removable white flame-retardant
translucent polystyrene plastic lens for bulb replacement.
Listings
UL® 1069 listed; UL® File Number E71599.
LI381 Corridor Light & LI382 Corridor Zone Light
Specifications
Dimensions:
Bulbs:
Power:
Finish:
Housing:
Height:
4.5625″
(115 mm)
Width:
4.5625″
(115 mm)
Depth:
3.3″
( 84 mm)
#1829 28V 70 mA
LI381:
24 VDC, 51mA
LI382:
24 VDC, 102mA
White flame-retardant translucent
polystyrene plastic lens mounted on a
flame-retardant (UL® 94V-0) plastic panel.
Fits either standard dual gang box and
ring or single gang box and ring.
Required Components
Tek-CARE® NC110, NC150, NC200 or NC300II Nurse Call
System
–or–
CM800 Emergency Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LI382LED Corridor Zone Light
IL898 Specification Sheet
Tek-CARE® Nurse Call & CM800 Emergency Call Systems • Rev. 3 - 11/2014
Features
•
•
•
•
•
UL® 1069 Listed
Long lasting LEDs
No bulb replacement required
Plug-in connector supplied
Attractive flame-retardant plastic panel
Description
The LI382LED Corridor Zone Light provides visual
indication of calls when used with a nurse call system,
emergency call system, or similar application.
The LI382LED Corridor Zone Light indicates an emergency
call by the rapid flashing of the two red LED bulbs associated
with the zone or area from which an emergency call
has been placed. Normal calls are indicated by a steady
illumination of the two white LEDs. If an emergency and
normal call are placed in the same zone at the same time,
the red LEDs connected to the emergency station will flash
rapidly, while the white LEDs associated with the normal
call will maintain a steady illumination.
The standard mounting plate features both one- and twogang mounting options.
LI382LED Corridor Zone Light
Specifications
Dimensions:Height: 4.8″
(122 mm)
Width:5.0″
(127 mm)
Depth:2.755″
( 70 mm)
LEDs:
2V 20mA per LED
Power:
24 VDC
Finish:
Base is flame-retardant (UL ® 94HB)
plastic. Lens is flame-retardant
translucent polycarbonate plastic.
Housing:
Fits standard one- or two-gang box with
one- or two-gang ring.
Listings
UL® 1069 listed; UL® File Number E71599.
Required Components
Tek-CARE® NC110, NC150, NC200 or NC300II Nurse
Call System with NC110-, NC150-, or NC200-series
master station
–or–
CM800 Emergency Call System with CM800-series
speaker panel
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LI386C Addressable Dome/Zone Light
IL547 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 12 - 07/2013
Features
•
•
•
•
•
•
•
•
UL® 1069 Listed
Supervised
Easy bulb replacement
Excellent visibility
Lamp supervision
Interchangeable lens caps
Simplified multiplex wiring
Plug-in connectors
Description
TekTone ®’s LI386C Dome/Zone Light with built-in
multipurpose station provides visual indication of calls
originating from bath, emergency and code call switches
when used as a dome light. The LI386C also indicates calls
originating from patient, staff and duty stations when
used as a zone light. The LI386C also indicates all calls
from the zones to which those stations are assigned and
provides communication to and from the central equipment,
including the nurse master station and monitor. It is used
with remotely-mounted devices such as switches, contact
closures, or other signal-originating equipment. No controls
are provided.
Listings
UL® 1069 listed; UL® File Number E71599.
LI386C Addressable Dome/Zone Light
Specifications
Dimensions:
Bulbs:
Power:
Connections:
Finish:
Housing:
Height:
4.5625″
(115 mm)
Width:
4.5625″
(115 mm)
Depth:
3.3″
(84 mm)
#1829 28v. 70mA
24 VDC, 51 mA
Ramp lock plug-in type
Base is flame retardant (UL® 94V-0)
plastic. Amber, red and green lenses are
flame-retardant translucent polycarbonate plastic. White lens is flame retardant
translucent polystyrene plastic.
Standard dual gang box and ring.
Required Components
Tek-CARE® NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
LI386LED Addressable Dome/Zone Light
IL914 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 3 - 01/2015
Features
•
•
•
•
•
•
•
UL® 1069 Listed
Long lasting LEDs
No bulb replacement required
Plug-in connector supplied
Attractive flame-retardant plastic panel
Addressable station with data signal supervision
Simplified multiplex wiring
Description
TekTone®’s LI386LED Dome/Zone Light with built-in
multipurpose station provides visual indication of calls
originating from bath, emergency and code call switches
when used as a dome light. The LI386LED indicates calls
originating from patient and staff stations when used as a
zone light. The LI386LED also indicates all calls from the
zones to which those stations are assigned and provides
communication to and from the central equipment,
including the nurse master station and monitor. No controls
are provided.
Listings
UL® 1069 listed; UL® File Number E71599.
LI386LED Addressable Dome/Zone Light
Required Components
Specifications
Dimensions:Height: 4.8″
(122 mm)
Width:5.0″
(127 mm)
Depth:2.755″
( 70 mm)
LEDs:
2V 20mA per LED
Power:
24 VDC
Finish:
Base is flame-retardant (UL ® 94HB)
plastic. Lens is flame-retardant
translucent polycarbonate plastic.
Housing:
Fits standard one- or two-gang box with
one-or two-gang ring.
Tek-CARE® NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
RY351B Hill-Rom® SideCom® Adapter
IL477 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 12 - 07/2013
Features
•
•
•
•
UL® 1069 Listed
Provides for the following functions from a Hill-Rom®
SideCom® Unit:
◊ Call-placed indication
◊ Monitor indication
◊ Hands-free two-way communication
◊ Nurse Call
Supervised
Ramp lock plug-in connectors
Description
The RY351B SideCom® adapter is designed for use with
IR311E, IR312E, IR319B and IR320B patient stations, and
a Hill-Rom® SideCom® unit. When used with the above
patient stations and the Hill-Rom® SideCom® unit, the
RY351B adapter will provide all the necessary circuitry
for a Hill-Rom® SideCom® unit to control all functions
normally controlled by a pillow speaker. This includes, but
is not limited to, nurse call and patient communications.
Note: There is one RY351B SideCom® adapter required
per SideCom® unit.
Listings
UL® 1069 listed; UL® File Number E71599.
RY351B Hill-Rom® SideCom® Adapter
Specifications
Dimensions:Length: 4.25″
(108 mm)
Width:2.00″ (50.8 mm)
Height:1.25″ (31.8 mm)
Construction: Circuit board mounted on a rugged
aluminum mounting bracket.
Connections: Reliable ramp lock plug-ins.
Housing:
Single gang ring, or single gang ring and
double gang box, or exact equivalent.
Required Components
Tek-CARE® NC300II Nurse Call System
JK401 Dummy plug for IR319B or IR320B patient
stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF350B Presence Station
IL407 Specification Sheet
Tek-CARE®400 P5 & NC300II Nurse Call Systems • Rev. 11 - 11/2012
Features
•
•
•
•
•
UL® 1069 Listed
Push button and green LED for nurse presence
indication
Push button and yellow LED for aide presence indication
Automatic nurse follower transfer
Flame retardant (UL® 94V-0) plastic faceplate
Description
The SF350B Presence Station provides push button
registration of nurse presence (green LED indication), or
aide presence (yellow LED indication) at remote locations.
The nurse/aide presence indicator unit can also be used
to transfer nurse follower feature from room to room as
nursing staff proceeds with rounds. Depressing the nurse
presence registration button automatically transfers
indication of all subsequent patient calls to the remote
location of the nurse, as well as to the master station. Plugin connectors provide for simple wiring.
SF350B Presence Station
Specifications
Dimensions:
Construction:
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.50"
(114 mm)
2.75"
(70 mm)
1.50"
(38 mm)
0.4375"
(11 mm)
from mounting surface
Flame retardant (UL® 94V-0) plastic
faceplate. Switches and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square box
(IH358), or exact equal.
Listings
UL® 1069 listed; UL® File Number E71599.
Applicable Systems
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF337C Emergency Shower Switch
IL629 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 13 - 10/2015
Features
•
•
•
•
•
•
•
UL® 1069 Listed
Waterproof if installed per manufacturer’s
instructions
Call-placed indicator
Call cancel at point of origin
Attractive, flame-retardant (UL ® 94V-0) plastic
faceplate
1.1″ (28mm) wide pull-down lever
Washable TPU plastic pull cord kit is included
Description
The SF337C Emergency Shower Switch is for any location
where activation of an emergency signal is required. The
design is such that it may be located anywhere in a shower,
bath or tub room, regardless of environmental conditions.
The combination of rubber-grommetted mounting
screws, rubber mounting gasket, and completely sealed
front faceplate make it waterproof (if properly installed).
Electronic components are mounted on a circuit board,
including an industry-proven reliable reed switch for call
initiation. The activating magnet is mounted on the large
pull-down lever, which is completely sealed in plastic. The
indicator LED is also completely sealed.
Due to the flexibility of its design, the SF337C may be
used as a pull cord or as a pull down activated emergency
switch. The pull cord supplied may be used for emergency
call activation, or if desired, the large 1.1″ (28 mm) red
pull-down lever may be used. Dome light and system
annunciation vary according to the manner in which the
SF337C is connected.
Listings
UL® 1069 listed; UL® File Number E71599.
SF337C Emergency Shower Switch
(cord not shown)
Specifications
Dimensions:
Construction:
Connections:
Flush Housing:
Height:
4.5″
(114.3 mm)
Width:
2.8125″ (71.4 mm)
Overall Depth: 1.75″
(44.4 mm)
Projects Front: 0.71″
(18 mm)
Projects Rear: 1.04″
(26.4 mm)
®
Flame-retardant (UL 94V-0) plastic
Plug-in type.
Single-gang ring with square box; for
Tek-CARE®500 use plastic single-gang box
with minimum 2.875″ × 2.125″ opening.
Surface Housing: TekTone® IH527 plastic back box for
Tek-CARE®500.
Applicable Systems
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®NC300II Nurse Call System
Tek-CARE® NC200, NC150, NC110 Nurse Call Systems
Tek-CARE®500 Wireless Emergency Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF339 Emergency Switch
IL744 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 6 - 10/2015
Features
•
•
•
•
•
•
UL® 1069 Listed
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame-retardant (UL® 94V-0) plastic faceplate
0.9″ diameter (23 mm) red mushroom-style button
Description
The SF339 Emergency Switch includes a locking mushroomstyle switch, and is for use in any location where activation of
an emergency signal is required. All electronic components
are mounted on a circuit board, including a reliable industryproven DPDT switch for circuit initiation. The DPDT switch
and the indicator LED are completely sealed in plastic.
The SF339 may be used as an emergency switch or as a
staff-assist emergency call switch. The large mushroomstyle button is easily activated, so that calls may be placed
using an elbow, shoulder, etc., when using the hands is
inconvenient or inappropriate. Calls are cancelled by turning
the mushroom-style button clockwise.
Dome light indication and system annunciation vary
according to the manner in which the SF339 is installed.
SF339 Emergency Switch
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
Dimensions:
Construction:
Height:
4.5″
(114.3 mm)
Width:
2.8125″ (71.4 mm)
Overall Depth: 2″
(50.8 mm)
Projects Front: 0.95″
(24.1 mm)
Projects Rear: 1.05″
(26.7 mm)
Flame-retardant (UL® 94V-0) plastic
faceplate. DPDT switch and electronic
components are circuit board mounted.
Connections: Plug-in connectors.
Flush Housing: Single-gang ring with square box; for
Tek-CARE®500 use plastic single-gang box
with minimum 2.875″ × 2.125″ opening.
Surface Housing: TekTone® IH527 plastic back box for
Tek-CARE®500.
Applicable Systems
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®NC300II Nurse Call System
Tek-CARE® NC200, NC150, NC110 Nurse Call Systems
Tek-CARE®500 Wireless Emergency Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF340B Emergency Switch
IL509 Specification Sheet
Tek-CARE®400 P5 & NC300II Nurse Call Systems • Rev. 17 - 11/2012
Features
•
•
•
•
•
•
•
UL® 1069 Listed
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame-retardant (UL® 94V-0) plastic faceplate
1.1" (28 mm) wide pull-down lever
Washable TPU plastic pull cord kit is included
Description
The SF340B Emergency Switch is for use in any location
where activation of an emergency signal is required. All
electronic components are mounted on a circuit board.
The specially mounted LED is completely sealed in plastic.
Due to the flexibility of its design, the SF340B may be used
as a pull cord emergency switch, a pull down activated
emergency switch, or a staff assist emergency call switch.
The pull cord supplied may be used for emergency call
activation or, if desired, the large 1.1" (28 mm) red pulldown lever may be used.
Dome light and system annunciation vary according to the
manner in which the SF340B is connected.
SF340B Emergency Switch
(pull cord not shown)
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
Dimensions:
Construction:
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.5"
(114 mm)
2.8125"
(72 mm)
1.625"
(41 mm)
0.6875"
(17 mm)
from mounting surface
(including pull-down lever)
Flame-retardant (UL ® 94V-0) plastic
faceplate. Switch and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square box
(IH358), or exact equal.
Applicable Systems
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF341B Code Call Switch
IL510 Specification Sheet
Tek-CARE®400 P5 & NC300II Nurse Call Systems • Rev. 15 - 11/2012
Features
•
•
•
•
•
•
•
•
•
UL® 1069 Listed
Supervised
Registers audible and visual calls at master station
Indicates local call origination on dome lights
Simple and flexible operation
Call-placed indicator
Call cancel at point of origin
Flame retardant (UL® 94V-0) plastic faceplate
1.1" (28 mm) wide pull-down lever
Description
The SF341B Code Call Switch operates all system code
signals, and is designed for use in ICU/CCU, physical therapy
rooms, post-op recovery, or any location where code call
initiation may be required. The large 1.1" (28 mm) blue
pull-down lever is used for code call activation. The red callplaced LED indicator flashes to provide for call assurance.
Calls are registered on the master station (and duty station,
if used) via audible and visual signals pulsing at a rate of
twice that of a standard emergency signal. Local indication
on dome lights is also provided. Call cancellation is possible
only at the originating location.
SF341B Code Call Switch
Listings
UL® 1069 listed; UL® File Number E71599.
Specifications
Dimensions:
Construction:
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.50"
(114 mm)
2.8125"
(72 mm)
1.625"
(41 mm)
0.6875"
(17 mm)
from mounting surface
(including pull down lever)
Flame retardant (UL® 94V-0) plastic
faceplate. Switch and electronic
components are circuit board mounted.
Plug-in connectors.
Single gang ring (IH357) with square box
(IH358), or exact equal.
Applicable Systems
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF381 Emergency & SF382 Code
Standalone Stations
IL753 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 4 - 07/2013
Features
•
•
•
•
•
UL® 1069 Listed
Plug-on connection
Supervised connections
Standalone installation
SF381 includes a washable TPU plastic pull cord kit
Description
The SF381 Emergency Station and the SF382 Code Call
Station are for use in any location where an emergency or
code call switch is needed without the need to install another
device (such as a multipurpose station or other addressable
device). The stations are designed to be connected directly
to the station bus wiring of the Tek-CARE® NC300II system.
The stations include dome lamp outputs, and are addressed
using dip switches mounted on the rear of the stations.
Listings
UL® 1069 listed; UL® File Number E71599.
SF381 Emergency & SF382 Code Standalone Stations
Specifications
Dimensions: Height
4.5″
(114 mm)
Width:2.8125″
(75 mm)
Depth:1.625″
(41 mm)
Projects:0.6875″
(17 mm)
(Including pull-down lever)
Construction: Flame-retardant (UL® 94V-0) plastic
faceplate. Electronic components are
mounted on a printed circuit board.
Connections: Plug-on connectors.
Housing:
Single gang ring or double gang box with
single gang ring.
Required Components
Tek-CARE® NC300II Nurse Call System
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF380A Visual-Only Patient Station
IL866 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 4 - 04/2014
Features
•
•
•
•
•
Call-placed indicator
Signaling for routine, bath emergency, code, and fire
with smoke detectors
Low power requirements
Addressable—requires no IR-series station
Station-bus connector pin-compatible with existing
IR-series stations
Description
The SF380A Visual-Only Patient Station controls two dome
light bulbs, senses button presses and switch closures, and
provides routine, bath emergency, and code calls (when
partnered with the proper peripheral switches) to the
Tek-CARE® NC300II central equipment for display at the
system master stations.
The SF380A provides routine call input via a push-button
call cord, and can signal a cord out call. When used with
an emergency switch, it provides a medium priority bath
emergency call. When used with a code switch, it provides a
high priority code call. The SF380A also has a smoke detector
input. Station addressing uses a 7-position dipswitch.
Six switches set the station address; the seventh switch
determines whether lamps are supervised. The SF380A
is a drop-in replacement for the SF380C, with dipswitchselectable lamp supervision.
Listings
UL® 1069 listed; UL® File Number E71599.
SF380A Visual-Only Patient Station
Specifications
Dimensions:Height: 4.5″
(114 mm)
Width:2.75″
(69 mm)
Depth:2.5″
(64 mm)
Projects:0.4375″
(11 mm)
Construction: Flame-retardant (UL® 94V-0) plastic
faceplate. All electronic components are
circuit-board mounted.
Connections: Plug-in connectors for all wiring.
Housing:
Single-gang ring & double-gang back box.
Required Components
Tek-CARE® NC300II Nurse Call System
LI380, LI382 or LI384A Corridor Dome Light
SF301A Push-button or SF301G Geriatric Call Cord
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF301A, SF302 Call Cords
IL338 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 18 - 06/2013
Features
•
•
•
•
•
UL® 1069 Listed
Oxygen safe
Impact resistant and waterproof
Small, lightweight pendant fits comfortably in hand
Bed clips included
Description
The TekTone® SF301A (single) or SF302 (double) call cords
are for use with the Tek-CARE® NC110N, NC150N, NC200N
and NC300™II Nurse Call Systems, and the Tek-CARE®500
Wireless Emergency Call System, where ¼″ phono jacks
are used. Ten-foot cords are also available as SF301A/10
and SF302/10. The SF301A and SF302 are oxygen safe call
cords certified safe for use by all patients receiving oxygen
therapy. These waterproof and impact resistant call cords
have completely sealed push buttons.
Listings
UL® 1069 Listed; UL® File Number E71599.
Specifications
Connection:¼″ (6.35 mm) plug-in phono jack
Construction: SPST-1A switch
Integral brass, fine silver contacts
1 AMP, 115 watt maximum, AC
Dimensions:
7′ (2.13 m) long polyvinyl covered cord
SF301A, SF302 Call Cords
Required Components
Tek-CARE®500 Wireless Emergency Call System
with SF530 Room Station
or
Tek-CARE® NC300™II Nurse Call System
with SF380A Patient Stations
or
Tek-CARE® NC200N or NC150N Nurse Call Systems
with IR151B or IR152B Patient Stations
or
Tek-CARE® NC110N Nurse Call System
with SF100C or SF102 Patient Stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SF301G, SF302G Call Cords
IL434 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 15 - 06/2013
Features
•
•
•
•
•
•
UL® 1069 Listed
Sensitive touch activation
Oxygen tent safe
Ideal for geriatric use
Momentary, air-activated switch
Adjustable bed clip(s)
Description
TekTone® SF301G and SF302G Call Cords are for use with
the Tek-CARE® NC110N, NC150N, NC200N and NC300II
Nurse Call Systems, where ¼″ phono jacks are used. Patient
stations include SF100C, IR151B, IR152B, SF380C, SF380P
and SF530.
SF301G and SF302G Call Cords are specifically designed
for use in geriatric and oxygen environments, but may also
be used in general applications. To place a call, the patient
applies slight pressure to the squeeze bulb using any part
of the body. The force of air through the vinyl tubing closes
a switch in the plug, thus placing a call. An adjustable bed
clip is supplied for the SF301G and is attached to the call
cord. Two adjustable bed clips are supplied with the SF302G.
Listings
UL® 1069 Listed; UL® File Number E71599.
Specifications
Switch:
Momentary contact, air activated
Plug:
Stainless steel
Cord:
Vinyl tubing
Electrical:
Rated current 1/3 amp, 24 VAC-VDC
Dimensions:Cord:
6′ long
(1.83 m)
Plug:1.25″ long (32 mm)
Bulb:1.5″ diameter(38 mm)
SF301G Call Cord
Required Components
Tek-CARE®500 Wireless Emergency Call System
with SF530 Room Station
or
Tek-CARE® NC300™II Nurse Call System
with SF380A Patient Stations
or
Tek-CARE® NC200N or NC150N Nurse Call Systems
with IR151B or IR152B Patient Stations
or
Tek-CARE® NC110N Nurse Call System
with SF100C or SF102 Patient Stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
JK300A, JK300/2 Monitor Call Interface
IL558 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 10 - 11/2012
Features
•
•
•
Easy installation
Mounted on a flame-retardant (UL® 94V-0) plastic panel
Simple, reliable operation
Description
The JK300A Monitor Call Interface is for use in applications
where a simple interface is needed to connect the monitor
input of an IR300-series or IR400P5-series station to an
external contact closure. The JK300A accepts a standard
¼" phono-type jacks and registers a monitor call if a plug
is removed. A ¼" dummy plug is included for use when the
monitor point is not connected to an external device. The
JK300/2 Monitor Call Interface provides two ¼" jacks and
two dummy plugs.
Specifications
Dimensions:
Construction:
Connections:
Housing:
Height:
Width:
Depth:
Projects:
4.5"
(114 mm)
2.75"
(70 mm)
1.5"
(38 mm)
0.25"
(6.5 mm)
from mounting surface
Reliable components mounted on a flameretardant (UL® 94V-0) plastic panel
Wire lead pigtails
Single-gang electrical box or ring
JK300A, JK300/2 Monitor Call Interface
Applicable Systems
Tek-CARE®400 P5 Nurse Call System
IR400P5-series Multipurpose, Patient, Resident, CheckIn, Staff or Duty Stations
-or-
Tek-CARE®400 Nurse Call System
PM423 Auxiliary Input Module
IR400-series Multipurpose, Patient, Resident, Check-In,
Staff or Duty Stations
-or-
Tek-CARE®NC300II Nurse Call System
IR300-series Multipurpose, Patient, Staff or Duty
Stations
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
PM368K Short Haul Modem Kit
IL833 Specification Sheet
Tek-CARE®NC300II Nurse Call System • Rev. 4 - 07/2013
Features
•
•
•
•
Converts RS232 signaling to RS422
Transmits up to 2000′
Uses CAT5 cabling
Provides CTS/RTS for supervision functionality.
Description
TekTone®’s PM368K Short Haul Modem Kit increases
the distance of serial signaling data by converting RS232
signaling to RS422 signaling. This extends the distance
(using CAT5 cable) to 2000′ between a Tek-BRIDGE™ and
a serial printer, Tek-PAGING® system, Tek-CARE®NC300II
Central Equipment, or a monitored foreign system.
Specifications
Electrical:
Rated 12VDC@100mA DC connectors
Dimensions:
Transformers:
2.2″×1.7″×1.5″ with 5.6′ cable
Short Haul Modems:
2.2″×1.7″×0.65″
CT549 with CA132:
2′
Other Cable Assemblies: 1′
Cable distance: up to 2000′ CAT5
PM368K Short Haul Modem Kit
Required Components
for the Tek-CARE® NC300II Nurse Call System:
Tek-PAGING® Paging System, or
Serial Printer, or
Tek-MMARS®300 Management Monitoring and
Reporting System
for the Tek-CARE®500 Wireless Emergency Call System:
Tek-PAGING® Paging System, or
Serial Printer
for the Tek-ALERT™ Integration Manager:
Foreign System to be monitored,
Tek-PAGING® Paging System, or
Serial Printer
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
NC365B Paging Transmitter
IL670 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 12 - 10/2015
Description
The NC365B Paging Transmitter receives alphanumeric
messages and patient call information via serial interface
and transmits the information directly to pagers.
Information is sent to the transmitter at 9600 baud to
minimize bottlenecks. Call type and location are determined
by the system supplying information to the transmitter. The
NC365B Paging Transmitter is pre-programmed for use
with TekTone® systems for simplified plug-and-play setup.
The system’s range depends upon the site and the
transmitter’s location. To determine the exact coverage,
conduct a site survey. Under ideal conditions, the NC365B
Paging Transmitter’s range is approximately one mile. The
unit includes a ¼-wave BNC-terminated antenna to provide
coverage that meets most requirements.
Specifications
Dimensions:7.48″ × 12.99″ × 2.76″
(190 mm × 330 mm × 70 mm)
NC365B Paging Transmitter
Clearance:7.87″ (200 mm) minimum clearance on
all sides
Ports:
Scope protocol RS232 serial DB9
COMP2 protocol RS232 serial DB9
8 dry contact inputs configured as
Normally Open (#8 is preprogrammed
to send RANGE TEST to CAP code 400
using included dummy plug.)
Housing:
Surface mounted
Frequency:
457.55 MHz
Channel Spacing: 12.5 KHz
Adjacent Channel: better than 200nW @ 4.5 KHz deviation
Baud Rate:
9600 serial input
1200 transmit
Power: 115 VAC 0.8A, 50/60 Hz input
2 Watt output
Consumption less than 50mA standby;
up to 2.4A transmit
13.8 Volts DC operating voltage
Optional Components
BA365K Battery Backup Kit
Required Components
NC397A Pagers
plus one of these systems:
• Tek-CARE®400 P5 Nurse Call System
• Tek-CARE®400 Nurse Call System
• Tek-CARE®NC300II Nurse Call System
• Tek-CARE ® NC110/150/200 Nurse Call System
(requires NC377-series Paging Interface Adapter/
Voltage Interface)
• Tek-CARE®500 Wireless Emergency Call System
• Tek-ALERT® Alert Integration System
Each paging system installed in the United States is required by law
to have an FCC license. The FCC license can be obtained directly by
the end user or through an authorized agent. The FCC application
process takes approximately two weeks and provides the end user
with a ten-year renewable license. To apply for an FCC license, visit:
wireless.fcc.gov/index.htm?job=online_filing.
Paging is an ancillary method of call annunciation, and is not intended
to replace the primary annunciation provided by the nurse call system.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
NC397A Alphanumeric Pager
(Scope Geo 40A9)
IL734 Specification Sheet
Tek-CARE® Nurse Call Systems • Rev. 9 - 03/2015
Features
•
•
•
•
•
•
•
•
•
Alphanumeric
2-line 20-character backlit display
Easy to use large 4-button design
Preprogrammed with up to 8 user-specified CAP codes
4 vibrate patterns and 5 tone alert styles
Real-time clock
Message time stamping
Low battery indicators
Includes holster belt clip and bungee safety cord
Description
The NC397A alphanumeric pager is a radio pager with
a large 2-line display. It includes a holster/belt clip and
bungee safety cord. The pager is designed for use with
TekTone®’s Tek-CARE® nurse call systems, Tek-CARE®500
wireless emergency call system, and Tek-ALERT® alert
integration system.
The pager stores up to 60 messages of 239 characters each.
Messages may be selectively deleted or locked into memory.
Messages not locked into memory are automatically deleted
as new messages arrive (beyond the 60 message limit),
with the oldest messages deleted first.
The NC397A Pager may be configured for annunciation
via vibration or audible tones, and has one alarm for daily
annunciation. It can be programmed to turn on and off at
preset times each day.
The NC397A Pager is pre-programmed to receive call
information from TekTone® systems.
Specifications
Dimensions:
2.087″ × 2.99″× 0.67″
(53 mm × 76 mm × 17 mm)
without holster belt clip
Weight:
1.76 oz. (50 g) without battery
Protocol:
POCSAG
Operating Freq.:457.550 Mhz
Power:
1 AAA battery
NC397A Alphanumeric Pager
Required Components
NC365B Paging Transmitter
plus one of these systems:
• Tek-CARE®400 P5 Nurse Call System
• Tek-CARE®400 Nurse Call System
• Tek-CARE®NC300II Nurse Call System
• Tek-CARE ® NC110/150/200 Nurse Call System
(requires NC377-series Paging Interface Adapter/
Voltage Interface)
• Tek-CARE®500 Wireless Emergency Call System
• Tek-ALERT® Alert Integration System
Paging is an ancillary method of call annunciation, and is not intended
to replace the primary annunciation provided by the nurse call system.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)
SI005 LED Messaging Sign
IL890 Specification Sheet
Tek-ALERT® and Tek-CARE®400, 400 P5, 500 & 550 Systems • Rev. 5 - 04/2015
Features
•
•
•
•
•
•
One-line backlit LED display
Displays 10 to 40 alphanumeric characters per line
Configurable for audible alerts
Programmable for up to 6 CAP codes
Wireless data input
Receives text messages from Tek-CARE® and TekALERT® systems
SI005 LED Messaging Sign
Description
The SI005 LED Message Sign is a radio paging receiver
with a large one-line alphanumeric display. It is designed
for use with TekTone®’s Tek-ALERT® Alert Integration
Manager, Tek-CARE ®400 Nurse Call, Tek-CARE ®500
Wireless Emergency Call, and Tek-CARE®550 Wireless
Nurse Call Systems.
The SI005 stores up to 24 messages. The messages are
deleted after a predetermined length of time. Messages
are automatically deleted as new messages arrive (beyond
the 24 message limit), with the oldest messages being
deleted first.
The SI005 may be configured for annunciation via audible
tones. The SI005 has been pre-programmed to receive
call information from TekTone® systems, and it can be
programmed for up to six CAP codes.
Specifications
Dimensions:
Weight:
Construction:
28″ L × 6.75″ H (711 mm × 171 mm)
8.4 lbs.
LED display board with RF paging circuitry
enclosed in a plastic housing.
Operating Freq.:457.550 Mhz
Power supply: Input 115 VAC
Required Component
NC365B Paging Transmitter
Applicable Systems
Tek-ALERT® Alert Integration Manager
Tek-CARE®400 P5 Nurse Call System
Tek-CARE®400 Nurse Call System
Tek-CARE®500 Wireless Emergency Call System
Tek-CARE®550 Wireless Nurse Call System
The LED message sign is an ancillary method of call annunciation,
and is not intended to replace the primary annunciation provided by
the nurse call system.
www.tektone.com
277 Industrial Park Road • Franklin, NC 28734 • [email protected]
Phone: (828) 524-9967 • Fax: (828) 524-9968 • Sales: Choose option 2 • Tech Support: Choose option 3
TekTone®’s quality system is registered by UL® to the ISO 9001 standard. (Reference #10001510.)