Contracer CV

Transcription

Contracer CV
CATALOG No. E4221-218
®
CONTRACER
Contour Measuring Instruments
That's digital!
All models feature a Z-axis
with integrated digital-scale detecting unit.
Mitutoyo's Contracer CV Series is a high-accuracy contour measuring instrument
with a digital scale on the X- and Z-axis. Digitization of the Z-axis gives you the following:
• No need to calibrate the Z-axis according to measurement magnification
• No need to change the Z-axis detection zone according to measurement magnification
• High-accuracy measurement for large-displacement contours
That's Mitutoyo!
A wide selection of models designed to make measuring easier, more comfortable and more accurate!
Outstanding measurement support, thanks to
Windows OS-compatible "FORMPAK -1000"
Contour Analysis
®
®
Windows®95/Windows®98/Windows®NT4.0/Windows®2000 is used as the OS. Program commands
can be accessed via icons and a powerful Help function, making it easy for anyone to use . . . even a novice!
2
CV-2000M4
CV-H628
X-axis/Z-axis
measuring range:
200mm (8")/25mm (1")
Vertical travel:
250mm (10")
Base size (WxD):
600x450mm
(24"x18")
X-axis/Z-axis
measuring range:
100mm (4")/40mm (1.5")
Vertical travel:
320mm (13")
Base size (WxD):
600x450mm
(24"x18")
CV-1000N2
X-axis/Z-axis
measuring range:
50mm (2")/25mm (1")
Vertical travel:
—
Base size (WxD):
—
CV-3000S4
/4000S4
X-axis/Z-axis
measuring range:
100mm (4")/50mm (2")
Vertical travel:
250mm (10")
Base size (WxD):
610x450mm
(24"x18")
CV-3000W8
/4000W8
X-axis/Z-axis
measur ing range:
200mm (8")/50mm (2")
Vertical travel:
450mm (18")
Base size (WxD):
1000x450mm
(40"x18")
CV-3000H4
/4000H4
X-axis/Z-axis
measuring range:
100mm (4")/50mm (2")
Vertical travel:
450mm (18")
Base size (WxD):
610x450mm
(24"x18")
CV-3000H8
/4000H8
CV-3000W4
/4000W4
X-axis/Z-axis
measuring range:
100mm (4")/50mm (2")
Vertical travel:
450mm (18")
Base size (WxD):
1000x450mm
(40"x18")
CV-3000S8
/4000S8
X-axis/Z-axis
measuring range:
200mm (8")/50mm (2")
Vertical travel:
250mm (10")
Base size (WxD):
610x450mm
(24"x18")
X-axis/Z-axis
measuring range:
200mm (8")/50mm (2")
Vertical travel:
450mm (18")
Base size (WxD):
610x450mm
(24"x18")
Mitutoyo offers a range of models for every requirement.
There's sure to be one that fits your needs!
Contracer CV Series
3
Contracer
Equipped the X- and Z-axis with
Laser Holoscale
Small-footprint design . . . smart,
simple operation!
Mitutoyo's holographic scale, which is based on the
interference phenomenon of diffracted light, is
incorporated into the detector as a measuring unit for
the X- and Z-axes. The Laser Holoscale offers the world's
highest level of resolution along the X-axis 0.00625µm
(.25µinch), Z-axis 0.0125µm (.5µinch) over the entire
measurement range of the X-axis at 200 mm (8"), and
the Z-axis 25 mm (1").
The electronic unit is housed within the
measuring unit's vibration isolator, and the
vertical side table is a standard configuration
that houses the data analysis unit.
Principale
of Laser
Holoscale
Detector
Laser beam
Beam splitter
[b]
[a]
Mirror
Mirror
Scale
Mirror
Mirror
Beam splitter
Laser diode
Laser
Holoscale
4
Interference
fringe
Glass
plate
0.5µm (.02µinch)
World-class resolution and accuracy
(Z-axis)
Photo: CV-H628 with personal computer system
FORMPAK®-1000 software is easy to
operate, and capable of powerful
analysis, concurrently.
Contracer CV Series
5
Contracer
Featuring a digital scale unit
installed in the Z-axis detection
unit (CV-3000 Series)
Ceramic guides have been
employed for abrasion resistance.
The CV-3000 Series, which uses a digital scale unit in its
Z-axis detection unit, achieves a resolution of 0.0002
mm (.000008") at a full stroke of ±25 mm (±1"). And
unlike conventional, differential transformer units, there
is no change in resolution as the range of magnification
changes. The result: a dramatic improvement in
operation and performance!
Photo: CV-3000S4 with
personal computer system
6
Contracer
A dedicated control box for ease
of operation.
• The control box, which is independent of the main
unit, gives you remote control over positioning,
measurement start/stop, returning, and other key
operations.
• The drive unit's up/down position and the X-axis
traverse can be fine-tuned under manual control.
Mitutoyo’s Holographic Scale
delivers super-high accuracy
(CV-4000 Series)
• The new, top-of-the-line Holography Scale is based on
the interference phenomenon of diffracted light,
ensuring the absolute in accuracy. It allows
dimensional measurement at a resolution of
0.00005mm (.000002") over the entire measurement
range of 50mm (2").
Photo: CV-4000S4 with
personal computer system
FORMPAK®-1000 software is easy to
operate, and capable of powerful
analysis, concurrently.
Contracer CV Series
7
Contracer
Mitutoyo digitizes the X- and Z-axes over the full range, giving you a
wider range of measurement with higher resolution. No more reliance
on measurement magnifications!
• CV-1000N2 can be attached
to the optical stand.
The CV-1000 is portable and can be carried to the
machine-shop for measurement of large workpieces.
Even at a site where a power source may not be nearby,
the CV-1000 can work with the optional display unit
(built-in rechargeable battery).
Display unit and XY plotter
Icons and graphics on the LCD and touch panel make it
easy for anyone—even a novice—to perform key analysis
operations. The contour profile can be confirmed with
the display unit prior to output to an XY plotter.
Contour measurement data can be stored on optional RAM
cards and read into the personal computer.
8
Contracer
Arc Scale
on the Z-axis
Light emission
device
Connected to a personal computer, the FORMPAK-1000
contour analysis program provides various modes of
measurement and analysis. When connected to an XY
plotter, the contour profile can be checked with the
display unit prior to output.
Arc scale
Receiver IC
*The following plotters are recommened
•MP5100-51 (Graphtec)
•Y-1x50 (DXY-1150, 1250, 1350) (Roland)
Photo: CV-2000M4 with personal computer system
FORMPAK®-1000 software
is easy to operate, and
capable of powerful
analysis, concurrently.
Contracer CV Series
9
FORMPAK-1000 is a powerful program that has an enhanced user interface, including mouse
operation, a variety of graphic displays, and online help which is convenient if the user experiences a
problem. In addition to being able to perform dimensional measurements such as angle, radius, and
distance measurements, the contour tolerancing function for evaluating "form" and the report
creation function for outputting measurement results in a specific format are provided as standard,
ensuring both ease of operation and powerful analysis capabilities.
Realtime
Measurement
window
Analysis
Result
window
Online help function
The appropriate help window displaying operating procedures, the definition of icons,
etc., can be opened as required. This function eliminates the need to find information
in the manual.
Report creation function
Geometric measured data, measurement conditions, contour tolerancing results, and
comments can be laid out as desired and printed out in a report. Once a layout has
been saved, it can be reused as a glossary if the same measurement is made. A report
can be printed in A3 size or from a color printer (option).
Layout
Edit
window
User customization function
The size and position of operation buttons on a window can be modified as desired.
This function allows the counter display, each window, the location of analysis
command buttons, or the display color of measured data and auxiliary lines to be
modified.
External output function
Calculation results and tolerancing results can be downloaded in a readable format to
commercial spreadsheet software. This function increases processing capabilities
including statistical calculation, saving, and management of measurement results. It
also allows measured point group data to be output in the text format.
10
Printout
examples
Contour tolerancing function
This function compares measured data as "form" with nominal data, displays the
deviation, and then stores it, being different from dimensional measurement of angle,
radius, etc. In addition to values obtained from an ideal contour in a CAD drawing,
nominal values can be created by converting the measured data of a master.
Contour
tolerancing flow
Measured data
• Creation by the function
• Conversion from the master
measured data
• Inputting from CAD
Tolerancing
Tolerancing
Best-fit
processing
Developed
Error Display
window
Tolerancing
calculation
Tolerancing
Result
window
Tolerancing
results display
Specifications
Operation commands
1. Point measurement: Point, peak, contact,
intersection, bisector-point (between elements/
pieces of measured data)
2. Line measurement: Line, tangent line,
perpendicular, parallel line, bisector
3. Circle measurement: Circle (multi-points, range,
center and radius)
4. Position, difference: Coordinate difference, true
position judgment
5. Distance, angle: Distance, groove length, step,
pitch, distance reference, angle
6. Coordinate system setup: Origin setup, coordinate
system rotation
7. Calculation, statistics: Arithmetic calculation,
absolute value, square root, maximum value,
minimum value, mean, sum, standard deviation,
unbiased standard deviation, area
8. Editing of measured points: Deletion, translation,
rotation, mirror image, positioning, combination,
separation, projection
Contour tolerancing
1. Nominal value data creation
• Function specification (line/circle/aspheric
surface, etc.), CAD input (IGES, DXF), measured
data conversion, text file conversion
• Tolerancing direction setup (normal direction/Xaxis direction/Z-axis direction)
2. Reference coordinate system translation/rotation
3. Best-fit
• X-axis translation, Z-axis translation, rotating
direction, several kinds in combination
4. Nominal value tolerancing
5. Results display
• Result list display, graphic display
Special functions
a. Sub-part program: Calculation procedure can be
registered as a Sub-part program. The Sub-part
program can be integrated into a part program
facilitating measurement of batch of workpieces.
b. Measured data/conditions saving: Measured data
and analysis procedure can be saved in the hard
disk.
c. Layout Edit function: A measured profile,
measurement conditions, etc., can be arranged in
a form and laid out as desired. This function
allows an original result report to be created and
printed out.
d. External output of files: Analysis results can be
output in the CSV format, and therefore, can be
easily accessed by commercial spreadsheet
software. Also, measured data (point group data)
can be output in the ASCII code.
e. Part program: Measurement procedures complete
with measurement, analysis, contour tolerance
judgment, print, and save as a file, can be
registered as a part program. The part program
can be executed for automatic measurements.
f. Polar coordinate development display: Develops a
circular form on to the polar coordinate system,
where the range for circle calculation can be
specified.
Contracer CV Series
11
Rugged tables
Rotary vises
Rotary table
Holder with clamp
Cross-travel table
Rotary vise
Rotary table
Holder with clamp
• Table top: 280x180mm
(11"x7")
• XY travel: 100x50mm
(4"x2")
• Two-slide jaw type.
• Max. workpiece size:
ø60mm (2.4")
• Minimum reading: 1°
• Table top: ø146mm (5.7"
DIA)
• Minimum reading: 2'
• Holder with clamp (176107) can be mounted.
• Used with a cross-travel
table or rugged table.
• Max. workpiece height:
35mm (1.3")
176-107
218-001 (mm)
218-011 (inch)
218-003
Cross-travel table
Rotary vise
• Table top: 280x152mm
(11"x6")
• XY travel: 50x25mm (2"x1")
• One-slide jaw type.
• Max. workpiece size:
ø60mm (2.4")
• Minimum reading: 5°
Center supports
172-196
Swivel center support
Leveling tables
XY leveling table with
swivel
• Table top: 130x100mm
(5"x4")
• Leveling range: ±1.5°
• XY travel: ±12.5mm (±.5")
218-041 (mm)
218-051 (inch)
172-144
178-043-1
(mm)
178-053-1
(inch)
Rugged table
• Table top: 280x180mm
(11"x7")
• Provided with T-grooves for
mounting workpiece
fixtures.
V-blocks
• Max. workpiece diameter:
80mm (3.1")*
*65mm (2.5") when swiveled 10°
• Max. workpiece length:
140mm (5.5")
172-197
Center support
• Max. workpiece diameter:
120mm (4.7")
• 60mm (2.3") rizer is
optional (172-143).
V-block with clamp
• Used with a cross-travel
table or rugged table.
• Max. workpiece diameter:
50mm (2")
218-002
Digimatic XY leveling
table with swivel
• Table top: 130x100mm
(5"x4")
• Leveling range: ±1.5°
• XY travel: ±12.5mm (±.5")
172-142
Center support riser
• Used with a center support.
• Max. workpiece diameter:
240mm (9.4")
172-234
V-block with clamp
Vibration isolator
Vibration isolator
• Dimensions (WxDxH):
750x550x59mm
(30"x22"x2.3")
• Stand is optional
(178-024).
• Used with a cross-travel
table or rugged table.
• Max. workpiece diameter:
25mm (1")
172-143
178-042-1 (mm)
178-052-1 (inch/mm)
Leveling table
• Table top: 130x100mm
(5"x4")
• Leveling range: ±1.5°
• Height: 40mm (1.57")
Vise
Precision vise
• Max. workpiece size: 36mm
(1.42")
• Can be mounted on a
leveling table.
172-378
178-016
178-023
12
178-019
System diagram
172-196
218-003
172-144
176-107
172-234
172-378
172-142
172-197
172-143
Optional arms
(See page 14)
Optional styli
(See page 15)
218-041 (mm)
218-051 (inch)
218-001 (mm)
218-011 (inch)
218-002
178-047
178-016
CV Series main unit
and standard accessories
CV-H628
CV-3000/4000
CV-1000/2000
218-010
218-008
178-043-1
178-053-1
178-042-1
178-052-1
218-007
178-023
218-024
(CV-1000 only)
211-016
997996
178-019
998291
V-block
166-215
Symmetry calibration gages
V-block
• Workpiece diameter: 1mm
to 160mm (.04" to 6.3")
• Can be mounted on a
leveling table.
Others
CV-H628
Used to calibrate a slight circular error due to attachment errors
of the stylus and arm. These gages are usually required to
perform high-accuracy measurement of a lens, etc.
Windshield cover
Symmetry calibration
gage (angle
measurement type)
Display unit
Symmetry calibration
gage (sphere
measurement type)
12AAC168
CV-1000/2000
218-341■
■ Suffix: A for 120V, D for 220V, E for 240V
or non for 100V
RAM card
12AAA841
997996
998291
Pin gage unit
(CV-1000/2000)
211-016
Measuring stand
(CV-1000)
218-024
998862 (mm)
998861 (inch)
Contracer CV Series
13
— ARMS —
For CV-3000/4000
Eccentric type
Stylus
185
")
(7.28
15
(.59")
185
")
(7.28
Stylus
(1.
996504 for 6mm (.24") high stylus
996505 for 12mm (.47") high stylus
996506* for 20mm (.79") high stylus
996507 for 30mm (1.18") high stylus
996508 for 42mm (1.65") high stylus
Unit: mm (inch)
996509
996510
996511
996512
996513
40
57
ø3
(.12")
ø8
(.31")
ø8 (.31")
Unit: mm (inch)
For small hole stylus
15
(.59")
Stylus
130
")
(5.12
15
(.59")
")
for 6mm (.24") high stylus
for 12mm (.47") high stylus
for 20mm (.79") high stylus
for 30mm (1.18") high stylus
for 42mm (1.65") high stylus
Unit: mm (inch)
ø8
(.31")
Straight type
15
(.59")
996514
*Standard accessory
For CV-1000/2000
155
")
(6.10
15
(.59")
935111 for 6mm (.24") high stylus
935112 for 12mm (.47") high stylus
935113* for 20mm (.79") high stylus
935114 for 30mm (1.18") high stylus
935115 for 42mm (1.65") high stylus
Stylus
(1. 40
57
")
935116
935117
935118
935119
935120
For small hole stylus
Unit: mm (inch)
ø8
(.31")
Unit: mm (inch)
ø8 (.31")
Stylus
Eccentric type
ø8
(.31")
Straight type
15
155
9")
") (.5
(6.10
Unit: mm (inch)
95
") 25
Stylus
(3.74
(.98")
145
")
(5.71
for 6mm (.24") high stylus
for 12mm (.47") high stylus
for 20mm (.79") high stylus
for 30mm (1.18") high stylus
for 42mm (1.65") high stylus
935110
*Standard accessory
— ARM AND STYLUS SET —
• Consists of three special arms and five special styli.
For CV-3000/4000
Order No.
Components
in set
996660
14
996660
996504 Straight arm for 6mm high stylus
996511 Eccentric arm for 20mm high stylus
996514 Arm for small hole stylus
354882 6mm high one-sided cut stylus
354889 20mm high intersecting cut stylus
354894 20mm high cone stylus
996540 Small hole stylus
996542 Small hole stylus
— STYLI —
For CV-3000/4000
For CV-1000/2000 (
CV-H628)
One-sided cut
Intersecting cut
H
H
H
Tip angle: 12°
Tungsten carbide
H
6mm (.24")
12mm (.47")
20mm (.79")
30mm (1.18")
42mm (1.65")
Cone
Order No.
354882
354883
354884**
354885
354886
Tip angle: 30°
Sapphire or
Tungsten carbide
Tip angle: 20°
Tungsten carbide
H
6mm (.24")
12mm (.47")
20mm (.79")
30mm (1.18")
42mm (1.65")
Order No.
354887
354888
354889
354890
354891
H
6mm (.24")
12mm (.47")
20mm (.79")
30mm (1.18")
42mm (1.65")
** CV-1000/2000/3000/4000
Standard accessory
Knife edge
Sapphire tip
Tungsten tip
Diamond
H
H
Tip angle: 20°
Edge width: 3mm
Tungsten carbide
Order No.
354897
354898
354899
354900
354901
Order No.
12AAA566
12AAA567
12AAA568
12AAA569
12AAA570
Ball
H
H
6mm (.24")
12mm (.47")
20mm (.79")
30mm (1.18")
42mm (1.65")
Order No.
354892
354893
354894
354895
354896
Ball dia.: 1mm
Tungsten carbide
H
6mm (.24")
12mm (.47")
20mm (.79")
30mm (1.18")
42mm (1.65")
Order No.
354902
354903*
354904
354905
354906
Cone
Tip angle: 50°
High accuracy type
Diamond
H
6mm (.24")
12mm (.47")
20mm (.79")
30mm (1.18")
42mm (1.65")
Order No.
—
—
355129
—
—
*CV-H628 Standard accessory
For CV-3000/4000
20
(.79")
996540 Tip angle: 20°
Tungsten carbide
ø4
(.16")
ø5
(.2)
ø3
(.12")
ø4
(.16)
55
")
(2.17
2.5 (.10")
13
(.51")
Unit: mm (inch)
1 (.04")
0.4 (.02")
ø1.6
(.06")
SMALL HOLE
55
)
(2.17
55
")
(2.17
996541 Tip angle: 20°
Tungsten carbide
996542 Tip angle: 20°
Tungsten carbide
For CV-1000/2000
932693
Tip angle: 20° (one-sided cut)
Tungsten carbide
9")
(.7
20
ø8
(.32")
(ø4.5mm or larger)
50 ")
97
(1.
932694
Tip angle: 20° (one-sided cut)
Tungsten carbide
2.5 (.10")
0
9") 5 7")
(.4
9
.5
(1.
12
ø4
(.16")
ø6
(.24")
ø3
(.12")
ø6
(.16")
(ø2.5mm or larger)
1 (.04")
0.4 (.02")
ø1.6
(.06")
SMALL HOLE
Unit: mm (inch)
(ø7mm or larger)
")
58
(1.
")
40 1.97
(
50
932695
Tip angle: 20° (one-sided cut)
Tungsten carbide
Contracer CV Series
15
TECHNICAL DATA
Contracer CV-H628
Model Number
Order Number Selection
100V
110V
120V
220/230V
240V
Specifications
Model
Measuring
range
Resolution
X-axis
Z-axis
X-axis
Z-axis
Measuring
X-axis
accuracy (20°C) Z-axis
Straightness of drive unit
Stylus tip orientation
Traversing direction
Measuring force
Maximum traceable angle
Measuring speed
Moving speed
X-axis
Column
Digital data output
Operating temperature
Power supply
Power consumption
Dimensions Main unit
(WxDxH)
Control unit
Vibration
isolation table
Side desk
Mass
Main unit
Control unit
Vibration
isolation table
Side desk
Standard accessories
CV-H628
200mm (8")
25mm (1")
0.00625µm (.25µinch)
0.0125µm (.5µinch)
(0.5+0.003L)µm, (20+3L)µinch
±(0.5+ | 0.02H |)µm, ±(20+ | 20H |)µinch
(0.3+0.002L)µm, (11.8+2L)µinch
Downward
Forward/backward
30mN (3gf)
60 degrees for upward/downward slope***
0.02mm/s (.0008"/s) to 2mm/s (.08"/s),
Can be specified by 7 steps.
0.1mm/s (.004/s") to 5mm/s (0.2"/s),
Can be adjusted manually with a joystick box.
0.1mm/s (.004"/s) to 5mm/s (.2"/s),
Can be adjusted manually with a joystick box.
Via GPIB interface
10°C to 30°C
100V AC - 240V AC
150VA (at maximum)
765x450x890mm (30.1"x17.7"x35")
250x380x300mm (9.8"x15"x11.8")
1130x630x760mm (44.5"x24.8"x30")
620x940x1200mm (24.4"x37"x47.2")
140kg (308 Ibs.)
7kg (15.4 Ibs.)
130kg (286 Ibs.)
997123
Pin gage and holder
12AAA567 Cone stylus
354903
Ball stylus
12AAC169 Reversal mounting arm
211-016
Reference hemisphere
546-113
Dial tension gage
537976
Carrying bolts
178-042-1 Digimatic XY leveling table with swivel*
178-052-1 Digimatic XY leveling table with swivel**
178-016
Leveling table
355130
Machine cover
Gauge blocks
613615-016 (5mm)* 613192-221 (.2")**
613671-016 (10mm)* 613194-221 (.4")**
613635-016 (25mm)* 613201-221 (1")**
User's manual (1set), Fuse, Wrenches, Screw driver, Power
cord, GND wire
Gage blocks are provided with Grade K of /ISO/JIS/DIN or Grade 1of Federal Standard
* for metric
** for inch/mm
Dimensions
Unit: mm (inch)
940 (37.01")
Order number
suffix denotes the
AC power cord
type equipped.
CV-H628
mm
inch/mm
—
218-797
218-787C
—
218-787A
218-797A
218-787D
218-797D
218-787E
218-797E
630 (24.8")
Model
Type
Order No.
40kg (88 Ibs.)
1430 (56.30")
840 (33.07")
760 (29.92")
1600 (63")
* L = Traverse length (mm [inch])
** H = Measuring height (within ±25mm [±1inch])
***Depends on the workpiece surface condition.
1130 (44.49")
16
620 (24.41")
TECHNICAL DATA
Contracer CV-1000/2000
Model Number
Order Number Selection
Model
Type
Order No.
Order number
suffix denotes the
AC power cord
type equipped.
CV-1000N2
mm
inch/mm
218-611
—
218-611C
—
218-611A
218-621A
218-611D
218-621D
218-611E
218-621E
100V
110V
120V
220/230V
240V
CV-2000M4
mm
inch/mm
218-631
—
218-631C
—
218-631A
218-641A
218-631D
218-641D
218-631E
218-641E
Specifications
Model
Measuring
range
Resolution
Measuring
accuracy
X-axis
Z-axis
X-axis
Z-axis
X-axis
Z-axis
Straightness
Stylus tip orientation
Traversing direction
Measuring force
Stylus traceable angle
Stylus speed
Digital data output
Operating temperature
Power supply
Power consumption
Dimensions Main unit
(WxDxH)
Display unit
Mass
Main unit
Display unit
Standard accessories
CV-1000N2
CV-2000M4
50mm (2")
100mm (4")
25mm (1")
40mm (1.5")
0.2µm (.008µinch)
0.4µm (.016µinch)
0.5µm (.02µinch)
(3.5+0.02L)µm [(138+20L)µinch]*
±(3.5+| 4H |/25)µm [ ±(138+ |157H | )µinch]**
3.5µm/50mm
3.5µm/100mm
(.14µinch/2")
(.14µinch/4")
Downward
Forward
10 to 30mN (1 to 3gf)
Ascent:77º
(depends on the workpiece surface condition)
Descent: 87º
(depends on the workpiece surface condition)
0.2mm/s (.008"/s) or 0.5mm/s (.02"/s) for tracing
0.2mm/s (.008"/s) or 1mm/s (.04"/s)
or manual feeding for retracing)
Via RS-232C interface
0°C to 40°C
100V AC - 240V AC
150VA (at maximum)
490x178x138mm
600x450x740mm
(19.30"x7.01"x5.43") (23.62"x17.71"x18.80")
306x165x79mm
(12.05"x6.50"x3.11")
5kg(11 Ibs.)
115.8kg( 255 Ibs.)
5kg(11 Ibs.)
357651
AC adapter
354884
High one-side cut stylus
935113
Straight arm
546-113
Dial tension gage,
537392
Thread gage (p=1mm)*
538333
Thread gage (p=.5")**
Gauge blocks
(CV-1000) 611615-041 (5mm)* 611192-231 (.2")**
611671-041 (10mm)* 611194-231 (.4")**
611625-041 (15mm)* 611196-231 (.6")**
611635-041 (25mm)* 611201-231 (1")**
611673-041 (30mm)* 611673-041 (30mm)**
(CV-2000) 611671-041 (10mm)* 611194-231 (.4")**
611672-041 (20mm)* 611198-231 (.8")**
611674-041 (40mm)* 611201-231 (1")**
User's manual (1set), Wrenches, Screw driver, Power cord,
Connecting cables
Gage blocks are provided with Grade 2 of /ISO/JIS/DIN or Federal Standard
* for metric
** for inch/mm
* L = Measuring length (mm [inch])
** H = Measuring height from the horizontal (within ±12.5mm [±.5inch])
118 (4.65")
[39 (1.54")]
[18.5(.73")]
20 (.79")
220 (8.66")
[18.5
(.73")]
205 (8.07")
[X= 25 (1") position]
CV-1000N2
224 (9.6")
260 (10.24")
127 (5")
21
(.83")
[24.7
(.97")]
450 (17.72")
100 (3.94")
44 (1.73")
113 (4.45")
15 (.6")
[24.7 (.97")]
3 (.12")
12
(.47")
739.8 (29.13")
127 (5")
[12 (.47")]
32
19
32
(.75") (1.26") (1.26")
105
(4.34")
20 (.79")
3 (.12")
34 (1.34") (Z= 0 position)
34 (1.34") 46 (1.81")
20
(.79")
4
(.16")
(Z= 0 position)
39 (1.54") 61 (2.4")
Unit: mm (inch)
34 (1.34")
Dimensions
255 (10.04")
310
(12.2")
[X= 50 (2") position]
92 (3.62")
CV-2000M4
63
(2.48")
600 (23.62")
Contracer CV Series
17
TECHNICAL DATA
Contracer CV-3000/4000
Model Number
Order Number Selection
Model
Type
Order No.
Order number
suffix denotes the
AC power cord
type equipped.
Model
Type
Order No.
Order number
suffix denotes the
AC power cord
type equipped.
Model
Type
Order No.
Order number
suffix denotes the
AC power cord
type equipped.
100V
110V
120V
220/230V
240V
CV-3000S4
mm
inch/mm
218-652
218-662
218-652C
—
218-652A
218-662A
218-652D
218-662D
218-652E
218-662E
CV-3000H4
mm
inch/mm
218-653
218-663
218-653C
—
218-653A
218-663A
218-653D
218-663D
218-653E
218-663E
CV-3000W4
mm
inch/mm
218-654
218-664
218-654C
—
218-654A
218-664A
218-654D
218-664D
218-654E
218-664E
CV-3000S8
mm
inch/mm
218-657
218-667
218-657C
—
218-657A
218-667A
218-657D
218-667D
218-657E
218-667E
100V
110V
120V
220/230V
240V
CV-3000H8
mm
inch/mm
218-658
218-668
218-658C
—
218-658A
218-668A
218-658D
218-668D
218-658E
218-668E
CV-3000W8
mm
inch/mm
218-659
218-669
218-659C
—
218-659A
218-669A
218-659D
218-669D
218-659E
218-669E
CV-4000S4
mm
inch/mm
218-672
218-682
218-672C
—
218-672A
218-682A
218-672D
218-682D
218-672E
218-682E
CV-4000H4
mm
inch/mm
218-673
218-683
218-672C
—
218-673A
218-683A
218-673D
218-683D
218-673E
218-683E
100V
110V
120V
220/230V
240V
CV-4000W4
mm
inch/mm
218-674
218-684
218-674C
—
218-674A
218-684A
218-674D
218-684D
218-674E
218-684E
CV-4000S8
mm
inch/mm
218-677
218-687
218-677C
—
218-677A
218-687A
218-677D
218-687D
218-677E
218-687E
CV-4000H8
mm
inch/mm
218-678
218-688
218-678C
—
218-678A
218-688A
218-678D
218-688D
218-678E
218-688E
CV-4000W4
mm
inch/mm
218-679
218-689
218-679C
—
218-679A
218-689A
218-679D
218-689D
218-679E
218-689E
Spesifications
Model
Measuring
range
Resolution
CV-3000S4/4000S4
X-axis
Z-axis
X-axis
Z-axis
X-axis
CV-3000H4/4000H4 CV-3000W4/4000W4 CV3000S8/4000S8
100mm (4")
50mm (2")
CV-3000H8/4000H8 CV-3000W8/4000W8
200mm (8")
50mm (2")
0.00005mm (.000002")
CV-3000 models: 0.0002mm (.000008")
CV-4000 models: 0.00005mm (.000002")
Measuring
CV-3000S4/H4/W4: ±(1+2L/100)µm [±(40+20L)µinch]* CV-3000S8/H8/W8: ±(1+4L/200)µm [±(40+20L)µinch]*
accuracy (20°C)
CV-4000S4/H4/W4: ±(0.8+2L/100)µm [±(31+20L)µinch]* CV-4000S8/H8/W8: ±(0.8+4L/200)µm [±(31+20L)µinch]*
Z-axis
CV-3000 models: ±(3+|2H|/25)µm [±(120+|40H|)µinch]**
CV-4000 models: ±(0.8+|0.5H|/25)µm [ ±(31+|10H|)µinch]**
Linearity of drive unit
0.8µm/100mm (32µinch4/4")
2µm/200mm (80µinch/8")
Drive unit vertical travel
250mm (10")
450mm (18")
250mm (10")
450mm (18")
Drive unit inclinating range
±45 degrees
Stylus tip orientation
Downward/upward
Traversing direction
Forward/backward
Measuring force
30mN (3gf)
Maximum traceable angle
77 degrees for upward slope***
87 degrees for downward slope***
Stylus speed
Tracing
0.02mm/s (.0008"/s) to 2mm/s (.08"/s), Can be specified by 7 steps.
Retracting
5mm/s (.2"/s)
Positioning
0.1mm/s (.004"/s) to 5mm/s (.2"/s), Can be adjusted manually with a operation box.
Digital data output
Via RS-232C interface
Operating temperature
15°C to 25°C
Power supply
100V AC – 240V AC
Power consumption
150VA (at maximum)
Dimensions
Main unit
741x450x905mm
741x450x1105mm 1118x450x1105mm 767x450x905mm
767x450x1105mm 1144x450x1105mm
(WxDxH)
(29.17"x17.7"x35.63") (29.17"x17.7"x43.5") (44.02"x17.7"x43.5") (30.2"x17.7"x35.63") (30.2"x17.7"x43.5") (45.04"x17.7"x43.5")
Control unit
180x380x300mm (7.1"x15"x11.8")
Operation box
245x116x78mm (9.6"x4.6"x3.1")
Mass
Main unit
140kg (308 lbs.)
150kg (330 lbs.)
150kg (330 lbs.)
145kg (319 lbs.)
155kg (341 lbs.)
155kg (341 lbs.)
Control unit
6.5kg (14.3 lbs.)
Operation box
0.8kg (1.76 lbs.)
* L = Traverse length (mm [inch])
** H = Measuring height (within ±25mm [±1inch])
***Depends on the workpiece surface condition.
18
Standard accessories
354884
High one-side cut stylus
996506
Straight arm
546-113
Dial tension gage
178-043-1 Leveling table (CV-4000 only, for metric)
178-053-1 Leveling table (CV-4000 only, for inch/mm)
211-016
Symmetry calibration gage (CV-4000 only)
537976
Carrying bolts
352637
Lubricant
998862
Pin gage (mm) and holder*
537392
Thread gage (p=1mm)*
998861
Pin gage (inch) and holder**
538333
Thread gage (p=.05")**
Gauge blocks
(CV-3000) 611671-021 (10mm)* 611672-021 (20mm)* 611675-021 (50mm)*
611194-231 (.4")**
611198-231 (.8")**
611202-231 (2")**
(CV-4000) 611671-017 (10mm)* 611672-017 (20mm)* 611675-017 (50mm)*
611194-221 (.4")**
611198-221 (.8")**
611202-221 (2")**
User's manual (1set), Fuse, Wrenches, Screw driver, Connecting cables, Power cord, GND wire, Machine cover, Driving cable assy,
Column cable assy
Gage blocks are provided with Grade 0 of ISO/JIS/DIN or Grade 2 of Federal Standard for CV-3000 and Grade K of ISO/JIS/DIN or Grade 1 of Federal Standard for CV-4000
* for metric
** for inch/mm
Dimensions
(Example:
CV-3000S4/
CV-4000S4)
Main unit
300 (11.81")
905 (35.63")
Unit: mm (inch)
180 (7.09")
380 (15")
78
(3.07")
Control unit
245 (9.65")
610 (24.02")
450 (17.72")
116
(4.57")
Operation box
741 (29.17")
Contracer CV Series
19
Specifications are subject to change without notice.
Kyung-hwa Tools & Machinery Co., Ltd
Pukun BLD, 642 Omgung-dong, Sasang-gu, PUSAN, KOREA
Tel : 051- 313-9300 Fax : 051-316-9092
Homepage : http://www.gagemall.com
E-Mail : [email protected]
Printed in Japan
88.330201 (1) CBU