Fluke`s Artifact Calibration and Its Use in Calibration of 8.5 Digit DMMs
Transcription
Fluke`s Artifact Calibration and Its Use in Calibration of 8.5 Digit DMMs
Fluke's Artifact Calibration and Its Use in Calibration of 8.5 Digit DMMs Jack Somppi Product Marketing Manager Fluke Calibration NCSLI 2014, Orlando, Florida 1 Today’s Objectives • Review the concept of artifact calibration as seen in various commercial instruments • Evaluate its history as an effective method of maximizing performance of a calibrator • Present the performance improvements it offers over traditional periodic calibration with higher level primary standards • Offer an approach to use to make calibration of 8.5 digit dmms more efficient with improved confidence 2 Today’s challenge with 7½ & 8½ digit dmms • Supporting anything better than a 6½ digital multimeter requires more from a calibrator than the performance based on it’s normal “1-year” specifications. 3 The practical challenge – supporting commonly used digital multimeters Manufacturer’s Specifications (parts in 106) Example Cal. Points Popular 6½ & 8½ Digit DMMs Calibrator Agilent 34401A (1 year) Fluke 8846A (1 year) Agilent 3458A (1 year) Fluke 8508A (1 year) Fluke 5730A (95% conf. 1 year) 1.0 V dc 47 32 10.3 5.0 5.7 10.0 V dc 40 29 10.05 5.0 3.75 10.0V 1 kHz ac 900 900 92.0 102.0 47.0 100 Ω 110 140 20.0 10.6 10.0 10 kΩ 110 110 13.5 10.6 6.5 10 mA dc 550 700 30.0 22.0 39.0 0.14% 0.21% 505 610 138 10 mA 1 kHz ac 4 Tools to address this challenge… • Common tools: – Manual verification of the calibrator using primary lab class standards – Improved a calibrator’s specs by more frequent calibration – Improved specs + guardbanding the test limits – Custom specs through control charting & characterization of the calibrator • A less commonly used tool: – Using Artifact Calibration on a frequent basis to operate at improved 90 day specs or even 24 hour spec levels. Benefits from Artifact Calibration • Artifact Cal will give a 25% to 30% spec improvement Support a shortened calibration interval to utilize a better standard specification – calibrate every 90 days with Artifact Cal Selected as the most practical specs to use Up to 2 X better, often preferred, but not usually used 6 What is this “Artifact Calibration”? Simplified calibration using self adjustment through primary level standards Fluke’s Technique: • A traceable full calibration & optimized “self adjustment” of all functions and ranges • Traceable to the SI through 3 primary level standards • With Artifact calibration the “As Left” condition of the instrument will meet its best specified performance (24 hour specs) – and is documented • Every 57XX calibrator uses Artifact Calibration, it is the only way to adjust and compensate for natural performance errors. First introduced in 1988 with the 5700A and continued with the new 5730A in 2013 Agilent’s similar use of self adjustment • At about the same time (~1988) Agilent (Hewlett Packard at that time) introduced a dmm with self adjusting capabilities – the HP 3458A • Both instruments intended to simplify the calibration/adjustment process of their high end dmms and calibrators Using “self adjustment” • Agilent 3458A DMM – a “must do” to keep performance at the level of its published specs – “ACAL” – • Used daily to maintain standard performance to its published specs within any 24 hour period or corrects for temperature shifts larger than +/-1 degree – “Calibration” – • Routinely used for a full calibration to correct any and all output errors and return instrument to its 24 hour spec level – using a 10V and a 10kΩ artifacts • Fluke 5700 Calibrator Series – a discretionary technique intended to optimize performance at a better specified level – “Artifact Cal” – • Adjust: To correct for any and all output errors and return instrument uncertainty to its 24 hour spec level – “Cal Check” – • Measure only: To test and document all changes in the output levels since its last Artifact Cal Traditional Adjustment through measuring output parameters with primary lab standards Requires a primary lab type of measurement process with basic, traceable, primary standards, ~ 235 tests, taking many manhours of time Primary Voltage Standards Primary Resistance Standards Primary AC:DC Transfer Standards Linear Ratio Devices Comparison Measurement Volts Amps &Ω Working Standards 10 Artifact Calibration Innovation #1 – A primary level measurement system in the working standard Benefit: Now full performance verification becomes a routine part of the calibrator’s quality system. 11 Artifact Calibration Innovation #1 – A primary level measurement system in the working standard Volts Amps &Ω Working Standards Calibrator Circuitry 12 Artifact Calibration Innovation #1 – A primary level measurement system in the working standard 6.5 V DC Reference. 13 V DC Reference 1 Ohm Reference 1.9 Ohm Reference 10 k Ohm Reference Ultra Linear Ratios with a 26 Bit DAC Volt Meter for Measurement & Comparison (2) Multi-Range TVC Standards 10Hz -1MHz Volts Amps &Ω Working Standards Calibrator Circuitry All the necessary measurement standards are built in, to fully measure & properly verify the calibrator’s full specified performance 13 Artifact Calibration Innovation #2 – Traceable adjustments use external primary level standards to calibrate through the internal measurement system 6.5 V DC Reference. 10 V DC Standard 1 Ohm Standard 13 V DC Reference 1 Ohm Reference 1.9 Ohm Reference 10 kOhm Standard 10 k Ohm Reference Ultra Linear Ratios with a 26 Bit DAC Volt Meter for Measurement & Comparison (2) Multi-Range TVC Standards 10Hz -1MHz Volts Amps &Ω Working Standards Calibrator Circuitry Benefit: Self support of the best performance becomes very practical 14 A Documented Traceability Chain The electrical traceability chain 15 The documentation of traceability A Laboratory Verification Report DC VOLTAGE DC DC DC DC DC DC DC DC DC DC DC DC Range 220 mV 220 mV 2.2V 2.2V 11V 11V 22V 22V 220V 220V 1100V 1100V Zero Shift Magnitude 1.16E-07 2.20E-01 1.16E-07 -2.20E-01 4.07E-09 2.20E+00 5.81E-12 -2.20E+00 2.03E-08 1.10E+01 0.00E+00 -1.10E+01 -2.85E-07 2.20E+01 5.29E-07 -2.20E+01 -2.66E-05 2.20E+02 -2.70E-05 -2.20E+02 -4.81E-11 1.10E+03 -2.03E-06 -1.10E+03 Abs Shift -3.60E-07 5.93E-07 -3.01E-06 3.01E-06 -7.85E-06 7.87E-06 -9.79E-06 1.00E-05 7.24E-06 -6.09E-05 7.78E-04 -7.80E-04 V V V V V V V V V V V V Rel Shift Spec -1.637 9.2727 2.6932 9.2727 -1.3684 4.3636 1.3702 4.3636 -0.714 3.2727 0.7158 3.2727 -0.4452 3.2273 0.4563 3.2273 0.0329 4.2273 -0.2766 4.2273 0.7071 6.4545 -0.7089 6.4545 ppm ppm ppm ppm ppm ppm ppm ppm ppm ppm ppm ppm Pct of Spec -17.654 29.0446 -31.3585 31.4009 -21.8166 21.873 -13.7948 14.1385 0.7781 -6.5433 10.9546 -10.9833 AC VOLTAGE AC AC AC AC AC AC AC AC AC AC AC AC AC AC AC Range 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV 2.2 mV Magnitude Frequency 2.20E-03 All Freqs. 2.20E-03 2.00E+04 2.20E-03 5.00E+04 2.20E-03 1.00E+05 2.20E-03 1.20E+05 2.20E-03 1.20E+05 2.20E-03 2.00E+05 2.20E-03 3.00E+05 2.20E-03 4.00E+05 2.20E-03 5.00E+05 2.20E-03 6.00E+05 2.20E-03 7.00E+05 2.20E-03 8.00E+05 2.20E-03 9.00E+05 2.20E-03 1.00E+06 Abs Shift -5.39E-09 5.59E-09 -1.20E-08 -1.10E-08 -6.27E-09 -1.07E-08 1.33E-08 -2.41E-08 1.48E-08 3.59E-08 -1.73E-08 -1.65E-08 -2.70E-08 -1.30E-08 -1.91E-08 V V V V V V V V V V V V V V V Rel Shift Spec -0.0002 0.2363 0.0003 0.2363 -0.0005 0.2503 -0.0005 0.3267 -0.0003 0.6655 -0.0005 0.6655 0.0006 0.6655 -0.0011 0.6655 0.0007 1.2864 0.0016 1.2864 -0.0008 1.4464 -0.0007 1.4464 -0.0012 1.4464 -0.0006 1.4464 -0.0009 1.4464 % % % % % % % % % % % % % % % Pct of Spec -0.1036 0.1075 -0.2174 -0.1531 -0.0428 -0.0728 0.0906 -0.165 0.0521 0.1268 -0.0543 -0.0518 -0.0847 -0.0408 -0.0599 DC CURRENT DC DC DC DC DC DC DC DC DC DC Range 220 uA 220 uA 2.2 mA 2.2 mA 22 mA 22 mA 220 mA 220 mA 2.2A 2.2A Zero Shift Magnitude 5.83E-11 2.20E-04 6.64E-11 -2.20E-04 9.81E-10 2.20E-03 1.06E-09 -2.20E-03 2.27E-09 2.20E-02 3.08E-09 -2.20E-02 4.52E-08 2.20E-01 5.33E-08 -2.20E-01 -1.12E-07 2.20E+00 -3.04E-08 -2.20E+00 Abs Shift -1.63E-10 2.88E-10 -1.23E-09 3.28E-09 1.11E-08 -5.71E-09 1.11E-07 -1.21E-08 1.16E-06 -1.30E-06 A A A A A A A A A A Rel Shift Spec -0.7415 73.8182 1.3081 73.8182 -0.5608 38.6364 1.4896 38.6364 0.5026 37.2727 -0.2595 37.2727 0.5025 58.3164 -0.0548 58.3164 0.5264 125.2182 -0.5911 125.2182 ppm ppm ppm ppm ppm ppm ppm ppm ppm ppm Pct of Spec -1.0044 1.772 -1.4515 3.8553 1.3485 -0.6963 0.8617 -0.094 0.4204 -0.472 AC CURRENT AC AC AC AC AC AC AC AC AC AC AC AC Range 220 uA 220 uA 220 uA 220 uA 220 uA 220 uA 220 uA 2.2 mA 2.2 mA 2.2 mA 2.2 mA 2.2 mA Magnitude Frequency 2.20E-04 All Freqs. 2.20E-04 5.00E+02 2.20E-04 1.00E+03 2.20E-04 2.00E+03 2.20E-04 5.00E+03 2.20E-04 7.00E+03 2.20E-04 1.00E+04 2.20E-03 All Freqs. 2.20E-03 5.00E+02 2.20E-03 1.00E+03 2.20E-03 2.00E+03 2.20E-03 5.00E+03 Abs Shift -2.68E-10 -3.61E-08 -7.77E-10 -6.01E-10 3.35E-10 1.14E-09 2.88E-09 -2.68E-09 -1.36E-09 2.34E-09 -6.47E-10 -5.03E-09 A A A A A A A A A A A A Rel Shift -1.2192 -164.091 -3.5299 -2.7309 1.5229 0.0005 0.0013 -1.2196 -0.6189 1.0645 -0.2941 -2.2845 ppm ppm ppm ppm ppm % % ppm ppm ppm ppm ppm Pct of Spec -0.7505 -101.0073 -2.1729 -0.7035 0.3923 0.3555 0.8947 -0.9022 -0.4578 0.7875 -0.1054 -0.8185 Spec 162.4545 162.4545 162.4545 388.1818 388.1818 0.1464 0.1464 135.1818 135.1818 135.1818 279.0909 279.0909 16 A 5730A “Cal Check” Verification Report That is the Artifact Calibration • Artifact Calibration is the process of traceably adjusting the assigned value(s) of a large array of multidimensional parameters with full documentation • Standards required – 10 Volt DC reference (±1.5 µV/V) – 1 Ω resistor (±10 µΩ/Ω) – 10 kΩ resistor (±2 µΩ/Ω) • Measures > 390 internal points & calibrates 182 operating function/range points • Performance is adjusted to well within the 24 hour specifications • ~1 Hour of time to artifact calibrate an entire calibrator • Also has a “measure only” capability to evaluate and report all drift since the previous calibration (called Cal Check) – no adjustments are made if not desired 17 Artifact Calibration Philosophy • External references are used to establish values for internal references through built-in comparator and scaling circuitry • The same circuitry is used to check all of the other functions and ranges of the 5700A • This duplicates the metrology tasks and techniques done in a primary lab • Full measurement and shift analysis information is recorded for every function, range, and specification point • The information is available for view and stored for review. • If desired the 5700A stores internal calibration factors for each function and range during artifact calibration making the adjustment permanent. 18 Benefits of Artifact Calibration • Performance Benefit: • Once complete, allows use of 24hour, 90-day, 180-day or 1 year specs for up to 2 years between external measurements • Economical Traceability Benefit: • 3 external standards check drift of internal standards, calibration constants are adjusted, and record is kept internally to maintain traceability Artifact Cal - does it work? • Yes, the concept works by design • Yes, every individual calibrator is proven by testing – Fluke tests each and every calibrator for adjustment accuracy and time stability – Testing continues, a two year cycle of external verification is recommended (measure only with external standards) • Yes, the process works and has been proven by ongoing Fluke and other independent evaluations – Fluke has performed many studies of large samples (several hundred) units by comparing the artifact calibration results to the full external performance verification of 5700A’s in for calibration – The national measurement institutes including Germany, Netherlands and Sweden performed numerous studies that thoroughly studied and validated Artifact Calibration • Yes, it’s proven performance is a historical fact – And 20+ years and thousands of 57XX’s prove the technology works 20 International NMI studies • In 1999 it was evaluated and confirmed by multiple European NMI’s – The Artifact Calibration process – The detailed calibration results • Confirmed it works equal to or better than required • Report is available for review from the www.flukecal.com website Recent Published Data from PTB 5730A Measured by a Quantum AC JV System Spec Limit Spec Limit is ± 53 to 78 µV/V measured error was less than ±15 µV/V Different amplitudes are consistent and in spec Operation is consistent over the full bandwidth Fluke’s Process Metrology – Artifact Cal Example of As Left Measurements A recent histogram of tests on 59 Calibrators 1 Year guaranteed Specs are ± 43 µV or (4.3 ppm) 24 hour guaranteed specs are ± 28 µV or (2.8 ppm) Production Pass/Fail limits are at 80% of 24 hour spec or ± 22.4 µV 23 Fluke’s Process Metrology – Artifact Cal As Left Measurements of ACV, I & Ohms Pass/Fail limits are at 80% of 24 hour spec Histogram of 59 Calibrators 24 Benefits from internal measurements Use of Cal Check for in depth drift analysis & control charting • Cal Check is designed to measure that small changes are well within the spec limits • On newly made 57X0A calibrators Fluke uses Cal Check to confirm proper noise performance and stability over a 1 to 2 week period 25 Back to the 8½ digit dmm challenge • What Artifact Cal offers as an improved approach to make calibration of 8.5 digit dmms more efficient with improved confidence 26 A solution for the Agilent 3458A • A calibration process jointly developed by Agilent & Fluke • Multi Function Calibrator with specs based on a 90 day calibration interval • Plus guardbanding the test to insure confident decisions – Refer to this Fluke Calibration application note for details The best approach to efficiently use 90 day specs for the calibrator is… Rather than send it out for calibration every 3 months…. The solution for simple support of 90 day specs is using the Artifact Calibration capabilities of Fluke’s 5700 Series of calibrators. Benefits from Artifact Calibration • Artifact Cal can give a 25% to 30% spec improvement Support a shortened calibration interval to utilize a better standard specification – calibrate every 90 days with Artifact Cal Selected as the most practical specs to use Up to 2 X better, often preferred, but not usually used 29 Test Accuracy Ratio Improvements of 60% or more • Examples of DC volts and Resistance Verification Points Setting 100 mV 1.0 V 10 V 100 V 1000 V 10 ohms 100 ohms 1k ohms 10k ohms 100k ohms 1M ohms 100M ohms 1 year 1.84 1.75 2.38 2.06 3.47 1.22 2.31 2.18 2.18 1.67 1.61 1.76 TAR 5730A vs. 3458A 90 day improvement 2.12 15% 2.38 36% 3.24 36% 2.86 38% 4.89 41% 1.33 10% 2.57 11% 2.49 14% 2.58 18% 1.89 13% 1.90 18% 2.27 29% improvement 10% to 41% 24 hour improvement 2.65 44% 2.70 54% 3.96 67% 3.28 59% 5.45 57% 1.40 15% 2.89 25% 2.58 18% 2.84 30% 2.58 55% 2.09 30% 2.60 48% improvement 30 15% to 67% To summarize Artifact Calibration’s benefits • Frequent Use of Cal Check – Measurement Assurance – be sure your calibrator performs at expected levels every day, week, and/or month during the time between calibrations through regular Cal Checks • Use Artifact Cal to self support your calibrator – Self support your 57XX series calibrator with the 3 artifacts – an economical way to reduce calibration down time • Use Artifact Cal Regularly – Support a shortened calibration interval to utilize a better standard specification – calibrate every 90 days or even daily as needed with Artifact Cal • Improve your specs by control charting your calibrators true output over time using Cal Check – Let Cal Check track the changes over time and confirm your calibrator’s true performance 31 Thank You Any questions or comments? • You may reach me at: Jack Somppi Fluke Calibration [email protected] 32